Manual Supplement
© 2010-2013 Fluke Corporation. All rights reserved..
Manual Title: BP Pump 2 Operator Supplement Issue: 3
Part Number: 2196592 Issue Date: 4/13
Print Date: June 2007 Page Count: 1
Revision/Date:
This supplement contains information
necessary to ensure the accuracy of
the above manual.
BP Pump 2 Operators Manual Supplement
4/13 1
Change #1, 55935
On page 1-11, following Pressure Generation add:
Dynamic NIBP Simulation Repeatability…..2 mmHg (at maximal pulse size
independent of device under
test)
Change #2, 56150, 56170
On page 1-11, under Pressure Measurement, change the
High-Accuracy Version to the following:
High-Accuracy Version…………………….0.7 mmHg (0.09 kPa)
Change #3, 63302, 65378
On page 1-4, delete the entire General Safety Considerations
section.
On page 1-4, in the Symbols Table delete the ETL Symbol and
add:
Conforms to relevant North American Safety Standards.
Conforms to relevant South Korean EMC Standards.
On page 1-10, following Specifications, delete Mains Voltage,
Environmental Conditions and add the new section:
General Specifications
Power Line .................................................. 100 - 240 V ac 50/60 Hz, 60 VA
Operating Temperature .............................. 15 °C to 40 °C
Storage Temperature .................................. -20 °C to +65 °C
Relative Humidity ........................................ 90 % max
Operating Altitude ....................................... 2000m
Safety .......................................................... IEC 61010-1: CAT II 300 V, Pollution
Degree 2
Electromagnetic Environment ..................... IEC 61326-1: Basic
Electromagnetic Compatibility .................... Applies to use in Korea only. Class A
Equipment (Industrial Broadcasting &
Communication Equipment) [1]
[1] This product meets requirements for industrial (Class A) electromagnetic
wave equipment and the seller or user should take notice of it. This equipment is
intended for use in business environments and is not to be used in homes.