Fluke 707Ex Users Manual Text
2015-09-09
: Fluke Fluke-707Ex-Users-Manual-808355 fluke-707ex-users-manual-808355 fluke pdf
Open the PDF directly: View PDF .
Page Count: 10
Download | |
Open PDF In Browser | View PDF |
® 707Ex mA Calibrator Calibration Information XW Warning Read Safety Information before using the mA Calibrator. Introduction Contacting Fluke The Calibration Information for the 707Ex mA Calibrator (hereafter referred to as the mA Calibrator) provides the information necessary to verify the performance and adjust the calibration of the mA Calibrator. • Contacting Fluke • Specifications To contact Fluke, call one of the following telephone numbers: • Technical Support USA: 1-800-44-FLUKE (1800-443-5853) • Calibration/Repair USA: 1-888-99-FLUKE (1888-993-5853) • Canada: 1-800-36-FLUKE (1-800-363-5853) • Europe: +31 402-675-200 • Japan: +81-3-3434-0181 • Singapore: +65-738-5655 • Anywhere in the world: +1-425-446-5500 • Safety Information Or, visit Fluke's website at www.fluke.com. • Maintenance • User-Replaceable Parts To register your product, visit http://register.fluke.com. • Recommended Test Equipment • Performance Verification • Calibration Adjustment This document includes the following: To view, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals. Address correspondence to: For complete operating instructions, refer to the 707Ex Users Manual located on the CD that came with the mA Calibrator. Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A. Warranty This Fluke product comes with a 3-year warranty for material and workmanship. For warranty information, refer to the 707Ex Users Manual that came with the mA Calibrator. PN 2086601 May 2003 Rev. 2, 5/09 © 2003, 2009 Fluke Corporation, All rights reserved. Printed in U.S.A. Specifications are subject to change without notice. All product names are trademarks of their respective companies. 1 Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands Fluke 707Ex Calibration Information Accuracy Specifications Accuracy is specified for 1-year after calibration at operating temperatures of 18 °C to 28 °C and is given as: ± ( [ % of reading ] + [ counts ] ) MEASURE V dc: Range: + 28 V max Resolution: 1 mV Input Impedance: 1 MΩ Accuracy: ± (0.015 % of reading + 2 counts) MEASURE mA dc: Range: 20 mA (24 mA max) Resolution: 1 μA Accuracy: ± (0.015 % of reading + 2 counts) SOURCE / SIMULATE mA dc: Range: 0 mA to 20 mA (24 mA max) Resolution: 1 μA Accuracy: ± (0.015 % of reading + 2 counts) Temperature Coefficient: ± 0.005 % of range per °C for temperatures of −10 °C to 18 °C and 28 °C to 50 °C Relative Humidity: 95 % up to 30 °C 75 % up to 40 °C 45 % up to 50 °C Vibration: Random 2 g, 5 to 500 Hz Shock: 1 meter drop test Safety Compliance: Complies with IEC 61010-1-95 CAT I, 28 V; CSA C22.2 No. 1010-92 NRTL; ANSI/ISA S82.02.011994; ( Directive 94/9/EG and p NEC 500: Uo = 27.6 V, Io = 96.13 mA, Co = 76 nF, Lo = 2.5 mH, Ui = 30 V, Ii = 24 mA, Ci = 10 nF, Li = 0 mH CE: Source Mode: Complies with EN61010-1 and EN61326 Compliance: To 700 Ω at 20 mA Power Requirements: Simulate Mode: Single 9 V battery (See Table 2 in the User-Replaceable Parts section) External loop voltage requirement: 24 V nominal, 28 V maximum, 12 V minimum Battery Life (typical): Loop Power: Maximum Load: 700 Ω Percent Display: − 25 % to 125 % (4 to 20 mA mode) Input/Output Protection: Fused; not replaceable SOURCE mode: 18 hours; 12 mA into 500 Ω; MEASURE / SIMULATE mode: 50 hours Size: 69.85 mm (W) x 142.87 mm (L) x 50.80 mm (H) [ 2.75 in (W) x 5.625 in (L) x 2.00 in (H) ] With red holster and Flex-Stand: 76.20 mm (W) x 158.75 mm (L) x 54.61mm (H) [ 3.00 in (W) x 6.25 in (L) x 2.15 in (H) ] Weight: General Specifications Maximum voltage between any terminal and earth ground or between any two terminals: 28 V Storage Temperature: − 30 °C to 60 °C Operating Temperature: − 10 °C to 50 °C Operating Altitude: 3000 meters maximum Pollution Degree 2 2 0.28 kg (0.62 lb) With red holster and Stand: 0.42 kg (0.93 lb) Safety Information In this manual a Warning identifies conditions and actions that pose hazards to the user. A Caution identifies conditions and actions that may damage the mA Calibrator or equipment under test. International symbols used in this manual are identified in the Symbols section. Read the entire 707Ex Users Manual and the 707Ex mA Calibrator CCD (Concept Control Drawing) before using the mA Calibrator. mA Calibrator Safety Information XW Warnings and Cautions Faults and Damage To avoid electric shock, injury, or damage to the mA Calibrator: • Use the mA Calibrator only as described in this Calibration Information, the Users Manual, and the Fluke 707Ex mA Calibrator CCD (Concept Control Drawing) or the protection provided by the mA Calibrator may be impaired. • Inspect the mA Calibrator before use. Do not use it if it appears damaged. • Check the test leads for continuity, damaged insulation, or exposed metal. Replace damaged test leads. • Never apply more than 28 V between the input terminals, or between any terminal and earth ground. • Applying more than 28 V to the input terminals invalidates the mA Calibrator’s Ex Approval and may result in permanent damage to the unit so it can no longer be used. XW Warning Applying a voltage greater than 28 V to the input of the mA Calibrator invalidates its Ex Approval and may impair its safe operation in an Ex-hazardous area. If there is any reason to suspect that the safe operation of the mA Calibrator has been affected, it must be immediately withdrawn from use, and precautionary measures must be taken to prevent any further use of the mA Calibrator in an Ex-hazardous area. The safety features and integrity of the unit may be compromised by any of the following: • External damage to the housing • Internal damage to the mA Calibrator • • Exposure to excessive loads Use the proper terminals, mode, and range for your measuring or sourcing application. • Incorrect storage of the unit • To prevent damage to the unit under test, put the mA Calibrator in the correct mode before connecting the test leads. • Damage sustained in transit • Correct certification is illegible When making connections, connect the COM test probe before the live test probe. When disconnecting, disconnect the live probe before the COM probe. • Using the product with the red holster removed • Functioning errors occur • Permitted limitations are exceeded • Functioning errors or obvious measurement inaccuracies occur which prevent further measurement by the mA Calibrator • • Never use the mA Calibrator with the red holster removed. • Never use the mA Calibrator with the case open. Opening the case violates Ex Approval. • Make sure the battery door is closed before using the mA Calibrator. • Replace the battery as soon as the M (low battery) symbol appears to avoid false readings that can lead to electric shock. • Remove test leads from the mA Calibrator before opening the battery door. • This equipment is specified for use in measurement category I (CAT I) pollution degree 2 environments and should not be used in CAT II, CAT III, or CAT IV environments. Voltage transients should not exceed 300 V for the CAT I applications where this product is used. Measurement transients are defined in IEC1010-1 as 2 µs rise time with a 50 µs duration at 50 % of the maximum amplitude height. • • Measurement Category I (CAT I) is defined for measurements performed on circuits not directly connected to the mains. Safety Regulations The use of this 707Ex mA Calibrator meets the requirements of the regulations providing that the user observes and applies the requirements as laid down in the regulations and that improper and incorrect use of the unit is avoided. • Use must be restricted to the specified application parameters. • Do not open the mA Calibrator. • Do not remove or install the battery within the Ex-hazardous area. • Do not carry additional batteries within the Ex-hazardous area. • Use only type-tested batteries. The use of any other battery will invalidate the Ex-certification and present a safety risk. Refer to Table 2. • Do not use the mA Calibrator in an Ex-hazardous area unless it is completely and securely fitted in its accompanying red holster. Do not use in a damp or wet environment. 3 Fluke 707Ex Calibration Information • After using the mA Calibrator in a nonintrinsically safe protected circuit, a rest time of 3 minutes minimum duration must occur before the mA Calibrator is taken into and/or used in an Ex-hazardous area. Ex-Certification Data • ATEX Certificate of Conformity: ZELM 02 ATEX 0120 X • Certification: ( II 2 G EEx ia IIC T4 • Permitted for: Zone1, Equipment Group II, gas group C hazardous gases, vapor or mist, Temperature class T4. • Factory Mutual, N.I. Class 1 Div. 2 Groups A-D • Permitted for: Division 2 hazardous gases, vapor or mist, Gas Groups A-D Symbols International symbols used on the mA Calibrator and in this manual are explained below. Symbol ON / OFF button. J Earth ground W Caution: Important information. Refer to instruction sheet ( Conforms to ATEX requirements T 4 XW Warning To avoid electric shock, personal injury, or damage to the mA Calibrator: • Do not service this product. To maintain the integrity of the mA Calibrator in explosive atmospheres, return the mA Calibrator to Fluke for all repairs. • Remove any input signals from the test probes prior to removing test leads from the mA Calibrator. • When servicing the mA Calibrator, use only specified replacement parts described in Table 1. • Use only the battery specified in Table 2. • Do not allow water inside the case. Contact a Fluke Service Center before performing any maintenance procedure not described in the 707Ex Users Manual. Refer to the Contacting Fluke section. Meaning + p Maintenance Conforms to Factory Mutual requirements Double insulated M Battery ) Conforms to relevant Canadian Standards Association directives. Certification # LR110460-2. P Conforms to European Union requirements F Direct current ~ Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. ; Conforms to relevant Australian standards Cleaning XW Warning To avoid electrical shock, remove test leads and input signals before cleaning. To clean the case, wipe it with a cloth lightly dampened with water and mild detergent. WCaution Do not use abrasives, solvents, or alcohol. These materials will damage the case. static awareness A Message From Fluke Corporation Some semiconductors and custom IC's can be damaged by electrostatic discharge during handling. This notice explains how you can minimize the chances of destroying such devices by: 1. Knowing that there is a problem. 2. Learning the guidelines for handling them. 3. Using the procedures, packaging, and bench techniques that are recommended. The following practices should be followed to minimize damage to S.S. (static sensitive) devices. 1. MINIMIZE HANDLING 2. KEEP PARTS IN ORIGINAL CONTAINERS UNTIL READY FOR USE. 3. DISCHARGE PERSONAL STATIC BEFORE HANDLING DEVICES. USE A HIGH RESISTANCE GROUNDING WRIST STRAP. 4. HANDLE S.S. DEVICES BY THE BODY. 5. USE STATIC SHIELDING CONTAINERS FOR HANDLING AND TRANSPORT. 8. WHEN REMOVING PLUG-IN ASSEMBLIES HANDLE ONLY BY NON-CONDUCTIVE EDGES AND NEVER TOUCH OPEN EDGE CONNECTOR EXCEPT AT STATIC-FREE WORK STATION. PLACING SHORTING STRIPS ON EDGE CONNECTOR HELPS PROTECT INSTALLED S.S. DEVICES. 6. DO NOT SLIDE S.S. DEVICES OVER ANY SURFACE. 9. HANDLE S.S. DEVICES ONLY AT A STATIC-FREE WORK STATION. 10. ONLY ANTI-STATIC TYPE SOLDERSUCKERS SHOULD BE USED. 11. ONLY GROUNDED-TIP SOLDERING IRONS SHOULD BE USED. 7. AVOID PLASTIC,VINYL AND STYROFOAM IN WORK AREA. PORTIONS REPRINTED WITH PERMISSION FROM TEKTRONIX INC. AND GERNER DYNAMICS, POMONA DIV. Dow Chemical mA Calibrator User-Replaceable Parts Replacing the Battery User-Replaceable Parts XW Warning To avoid false readings, which could lead to electric shock or injury, replace the battery as soon as M (low battery indicator) appears on the display. User-replaceable parts and accessories are shown in Figure 2 and Table 1. Do not remove or install the battery within the Ex-hazardous area. 1 2 Use only a single 9 V alkaline battery, properly installed, to power the mA Calibrator. See Table 2. 3 7 To replace the battery (refer to Figure 1 and Table 2): 6 1. Remove the test probes from the input signal. 4 2. Press D to turn the mA Calibrator OFF. 3. Remove the test leads from the input terminals. 5 4. Remove the red holster. 5. Lift off the battery door on the back of the mA Calibrator as shown. aqv10f.eps Figure 2. User-Replaceable Parts 6. Remove the battery. Table 1. User-Replaceable Parts 7. Insert the replacement battery and re-install the battery door. Make sure it is securely in place. 8. Return the mA Calibrator to its red holster. Item Part Description Fluke Part No. Qty. A Non-skid foot 885884 1 B Battery, 9 volt, Alkaline 822270 or see Table 2 1 C Battery door 665106 1 D Red holster, with bail 2040228 1 E Alligator clips AC72 1 F Test lead set TL75 1 G Users Manual on CD 2053979 1 Table 2. Approved Batteries Battery Description Manufacturer Type aqv07f.eps Figure 1. Replacing the Battery Alkaline, 9 volt Duracell 6LR61 Alkaline Ultra, 9 volt Duracell 6LR61 Professional Alkaline Battery Procell, 9 volt Duracell 6LR61 Alkaline Energizer, 9 volt Eveready 6LR61 Alkaline Power Line Industrial Battery, 9 volt Panasonic 6LR61 Alkaline, 9 volt Daimon 6LR61 5 Fluke 707Ex Calibration Information Recommended Test Equipment A list of recommended equipment for the performance verification tests and calibration adjustment procedure is shown in Table 3. 6. Refer to Table 4 to verify the readings on the HP 3458A for the following tests. (No adjustment is necessary for test 1.) Test 2: 1. Press B twice to select 12.000 mA. Performance Verification 2. Verify the readings. XW Warning To avoid electric shock, do not perform the verification tests unless the mA Calibrator is fully assembled. Perform the following tests to check the accuracy of each of the mA Calibrator’s functions against its specifications. If the mA Calibrator fails any of these tests, calibration adjustment or repair is required. Test 3: 1. Press B twice to select 20.000 mA. 2. Press and rotate the knob until 24.000 mA is displayed. 3. Verify the readings. DC Current Measurement Mode Performance Verification Procedures Tests 4-6: DC Current Source Mode 1. Press the A key twice to put the mA Calibrator into the mA measurement mode. Test 1: 2. The mA Calibrator’s display should show MEASURE in the top right corner. 1. Turn the mA Calibrator on. 2. After the P.S. power timeout displays, the mA Calibrator should display 4.000 mA. If it is displaying 0.000 mA, turn the mA Calibrator off. Hold down A while turning the mA Calibrator on. Continue to hold A for 2 seconds. When the timeout display ends, the mA Calibrator should display 4.000 mA. 3. Connect the mA Calibrator’s + terminal to the HP 3458A I input, and the COM terminal to the HP 3458A LO input. 3. Disconnect the HP 3458A from the mA Calibrator. 4. Connect the test leads from the red AUX terminal of the Fluke 5520A to the HP 3458A I input. 5. Connect a test lead from the black AUX terminal of the Fluke 5520A to the COM terminal on the mA Calibrator. 6. Connect a test lead from the HP 3458A LO input to the mA Calibrator's + terminal. 4. Verify that the mA Calibrator's display shows SOURCE in the upper-left corner. 7. Adjust the Fluke 5520A to output the value of test 4 in Table 4 as measured by the HP 3458A. 5. Set the HP 3458A to measure DC Amps (DCI). 8. Verify the display readings on the mA Calibrator. 9. Repeat adjustment and reading for steps 5 and 6. Table 3. Recommended Equipment Equipment DC Calibrator Recommended Model or Equivalent DC Voltage: 0 to 30 V Accuracy: ± 0.004 % +0.5 mV Fluke 5520A Multi-Product Calibrator DC Current: 0 to 24 mA Accuracy: ± 0.004% +0.5 μA Digital Multimeter Flexible Test Leads 6 DC Current: 0 to 26 mA Accuracy: ± 0.004% +0.5 μA HP 3458A Fluke TL24 mA Calibrator Calibration Adjustment Table 4. DC Current Source / Current Measure / Voltage Measure Mode Tests Test No. 707Ex mA Calibrator Output Mode HP 3458A Minimum HP 3458A Maximum 1 4.000 mA mA Source 3.9974 mA 4.0026 mA 2 12.000 mA mA Source 11.9962 mA 12.0038 mA 3 24.000 mA mA Source 23.9944 mA 24.0056 mA Test No. 5520A Calibrator Output Mode Fluke 707Ex Minimum Fluke 707Ex Maximum 4 4.000 mA mA Measure 3.997 mA 4.003 mA 5 12.000 mA mA Measure 11.996 mA 12.004 mA 6 24.000 mA mA Measure 23.994 mA 24.006 mA Test No. 5520A Calibrator Output Mode Fluke 707Ex Minimum Fluke 707Ex Maximum 7 0.000 V Volt Measure -0.002 V 0.002 V 8 14.000 V Volt Measure 13.995 V 14.005 V 9 28.000 V Volt Measure 27.993 V 28.007 V DC Voltage Measurement Mode Calibration Adjustment Tests 7-9: Perform the following calibration adjustment procedure if the mA Calibrator fails the performance verification test. 1. Press A twice to get the mA Calibrator to V MEASURE. The mA Calibrator’s display should read "V". 2. Connect the test leads from the NORMAL output terminals of the Fluke 5520A to the input terminals on the mA Calibrator (black to COM and red to +). 3. Set the Fluke 5520A to test 7 in Table 4 and verify the display reading on the mA Calibrator. 4. Repeat for tests 8 and 9. 5. The reading on the display should be within the minimum and maximum values shown in Table 4. 6. Set the 5520A output to 0 V (zero), and set output to STANDBY. 7. Turn the mA Calibrator OFF and disconnect the mA Calibrator from the 5520A. 8. The performance verification tests are now complete. 9. If the mA Calibrator failed any of these tests, calibration adjustment or repair is required. Calibration Adjustment Counter The mA Calibrator contains a calibration adjustment counter. The value in the calibration adjustment counter can be recorded and used to show that no adjustments have been made during a calibration cycle. When Fluke performs the calibration, the value of the calibration adjustment counter will be recorded on the calibration label. To find the value in the calibration adjustment counter, hold down B while turning the unit on. The software revision will display for 2 seconds followed by the calibration counter for another 2 seconds. Note Make sure that the mA Calibrator has a new battery before starting the calibration procedure. Calibration will not function properly if the low battery indicator (M) is on. Because the mA Calibrator incorporates several key hold start up features, entering the Adjust mode through the keypad requires an exact key hold sequence: 1. With the mA Calibrator off, hold down B and C and turn the mA Calibrator on. "CAL" appears on the main display and the calibration adjustment counter is displayed on the 7 Fluke 707Ex Calibration Information secondary display. If the knob is not turned within 2 seconds, the unit returns to the normal operation mode. 2. The mA Calibrator adjustment procedure is password protected. At this point, a password is needed to go forward. "PAS" is displayed on the main display and "000" on the secondary display. Enter the password "297" as follows: 1. Rotate the knob to add a 2 in the LCD of the secondary display push and release the knob. 2. Rotate the knob to get 9 into the LCD, push and release. 3. Rotate the knob to get a 7, push and release. If the password is correct, you'll be led through all the adjust modes (source, mA read, V read), otherwise the unit returns to normal operation. In the adjust mode, the main display will show "1.000 mA". 3. Connect the mA Calibrator + terminal to the HP 3458A I input, and the COM terminal to the HP 3458A LO input. The mA Calibrator will be in mA source cal and "1.000 mA" is displayed. On the secondary LCD, the last 3 digits from the knob value are shown. Adjust the mA output by tuning the knob until "1.000 mA" shows on the HP 3458A. 4. Press C, "22.000 mA" is displayed. Adjust the mA output by turning the knob until "22.000 mA" is shown on the HP 3458A. 5. Press A to go to mA measure adjust. 6. Disconnect the HP 3458A from the mA Calibrator. 8 7. Connect the test leads from the AUX terminals of the Fluke 5520A to the terminals on the mA Calibrator (black to COM and red to +). 8. The mA Calibrator will display "MEASURE" and 0.000 mA is displayed. On the secondary display, the last 3 digits of the A/D reading are shown. Set the 5520A to output 0.000 mA into the mA Calibrator. Press C. 9. After a moment, the mA Calibrator will display 24.000 mA. Set the 5520A to output 24.000 mA into the mA Calibrator. 10. Press A to go to V measure adjust. 11. Connect the test leads from the output NORMAL terminals of the Fluke 5520A to the input terminals on the mA Calibrator. 12. The mA Calibrator will display "MEASURE" and 0.000 V. The Secondary display will show the last 3 digits of the A/D reading. Set the 5520A to output 0.000 V into the mA Calibrator. 13. Press C. After a moment, 25.000 V will display on the mA Calibrator. Set the 5520A to output 25.000 V into the mA Calibrator. 14. Press A to complete the adjustment. The calibration counter is incremented and the calibration data is stored. The calibration count will be on the main display and "CAL" on the secondary display for 2 seconds. The calibration adjustment procedure is complete. To see the value of the calibration counter without calibrating the unit, hold down B at power up. The software revision number is displayed for 2 seconds followed by the calibration counter for 2 seconds.
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : No XMP Toolkit : Adobe XMP Core 4.0-c316 44.253921, Sun Oct 01 2006 17:14:39 Format : application/pdf Title : Microsoft Word - text.DOC Creator : elwood Create Date : 2009:04:14 10:59:32-07:00 Creator Tool : PScript5.dll Version 5.2.2 Modify Date : 2009:04:16 07:34:28-07:00 Metadata Date : 2009:04:16 07:34:28-07:00 Producer : Acrobat Distiller 8.1.0 (Windows) Document ID : uuid:e3a2d3a0-207d-4554-9972-2bac5c1419c1 Instance ID : uuid:04dee5d4-9e14-44ce-99af-c7a766e4a55c Page Count : 10 Author : elwoodEXIF Metadata provided by EXIF.tools