Fluke Temperature
Calibrators
Technical Data
Fluke 724 Temperature Calibrator
Test temperature sensors and transmitters and
gauges with one tool
Now you can carry one tool to test all temperature
sensors and transmitters in your plant. The Fluke
724 can measure and source 12 thermocouple types
and seven RTD types, plus volts and ohms. The 724
even handles high-speed pulsed RTD circuits and
provides loop power.
The dual display lets you source temperature and
view loop current at the same time. With its simple,
“no menus” controls, it is easy to operate, too.
Easy to read dual display lets you view input and
output simultaneously
Measure RTDs, thermocouples, ohms, and volts to
test sensors and transmitters
Source/simulate thermocouples, RTDs, volts, and
ohms to calibrate transmitters
Perform fast linearity tests with 25 % and
100 % steps
Execute remote tests with auto step and auto ramp
Power transmitters during test using loop power
supply with simultaneous mA measurement
Store frequently-used test setups for later use
Backlight lets you work in poor light
Large battery capacity of four AA cells
Battery door for easy changes
Mechanical and General Specifications
Size: 96 mm x 200 mm x 47 mm
Weight: 650 g
Batteries: Four AA alkaline batteries
Warranty: Three-years
Battery life: 25 hours typical
Shock & Vibration: Random, 2G, 5 Hz to 500 Hz
2 Fluke Temperature Calibrators
Measurement Accuracy
Voltage dc 30.000 V 0.02 % + 2 counts
(upper display)
20.000 V 0.02 % + 2 counts
(lower display)
100.00 mV 0.02 % + 2 counts
-10.00 mV to
75.00 mV
0.025 % + 1 count
(via TC connector)
Current dc 24.000 mA 0.02 % + 2 counts
Resistance 0.0 Ω to
400.0 Ω
0.1 Ω (4-wire)
0.15 Ω (2- and 3-wire)
401 Ω
to 1500 Ω
0.5 Ω (4-wire)
1 Ω (2- and 3-wire)
1500 Ω
to 3200 Ω
1 Ω (4-wire)
1.5 Ω (2- and 3-wire)
Source Accuracy
Voltage DC 100.00 mV 0.02 % +2 counts
10.000 V 0.02 % +2 counts
–10.00 mV to
75.00 mV
0.025 % + 1 count
(via TC connector)
Resistance 15.0 Ω
to 400.0 Ω
0.15 Ω (exc. current
0.15 mA to 0.5 mA), 0.1
Ω (exc. current 0.5 mA
to 2 mA)
401 Ω
to 1500 Ω
0.5 Ω (excitation current
0.05 mA to 0.8 mA)
1500 Ω
to 3200 Ω
1 Ω (excitation current
0.05 mA to 0.4 mA)
Specifications
Ramp functions Source functions: Voltage, current,
resistance, frequency, temperature
Ramps: Slow ramp, Fast ramp,
25 % step-ramp
Loop power function Voltage: 24 V
Accuracy: 10 %
Maximum current: 22 mA, short circuit
protected
Step functions
Source functions: voltage, resistance,
temperature
Steps: 25 % of range, 100 % of range
Environmental Specifications
Operating
temperature
–10 ºC to 55 ºC
Storage temperature –20 ºC to 71 ºC
Humidity
(Without
Condensation)
90 % 10 ºC to 30 ºC
75 % 30 ºC to 40 ºC
45 % 40 ºC to 50 ºC
35 % 50 ºC to 55 ºC
Safety Specifications
Safety rating CSA C22.2 No. 1010.1:1992
EMC EN50082-1:1992 and
EN55022:1994 Class B
RTDs and Thermocouples
Measure accuracy NI-120 0.2 ºC
PT-100 (385) 0.33 °C
PT-100 (393) 0.3 °C
PT-100 (JIS) 0.3 °C
PT-200 (385) 0.2 °C
PT-500 (385) 0.3 °C
PT-1000 (385) 0.2 °C
Resolution 0.1 °C
J 0.7 °C
K 0.8 °C
T 0.8 °C
E 0.7 °C
R 1.8 °C
S 1.5 °C
B 1.4 °C
L 0.7 °C
U 0.75 °C
N 0.9 °C
Resolution J, K, T, E, L, N, U: 0.1 °C,
0.1 °F B, R, S: 1 °C, 1 °F
XK 0.6°C
BP 1.2 °C
Source accuracy NI-120 0.2 °C
PT-100 (385) 0.33 °C
PT-100 (393) 0.3 °C
PT-100 (JIS) 0.3 °C
PT-200 (385) 0.2 °C
PT-500 (385) 0.3 °C
PT-1000 (385) 0.2 °C
Resolution 0.1 °C
Note Accuracy stated for
4-wire measurement.
J 0.7 °C
K 0.8 °C
T 0.8 °C
E 0.7 °C
R 1.4 °C
S 1.5 °C
B 1.4 °C
L 0.7 °C
U 0.75 °C
N 0.9 °C
Resolution J, K, T, E, L, N, U: 0.1 °C,
B, R, S: 1 °C
XK 0.6 °C
BP 1.2 °C
Functional specifications
3 Fluke Temperature Calibrators
Fluke 712 and 714 Temperature Calibrators
The Fluke 712 and 714 temperature calibrators deliver outstanding
performance, durability and reliability. These calibrators are compact,
lightweight and easy to carry and with a push-button interface and
are easy to use. Each calibrator is EMI tolerant, dust- and splash-
resistant and features a removable battery door for quick battery
changes.
Auto-step and auto-ramp features support remote testing.
Fluke 714 Thermocouple Calibrator
Measure temperature from TC probes
Simulate TC output
Operable with nine types of thermocouples
Calibrate linear TC transmitter with mV source function
Selectable °F or °C
Thermocouple mini-jack termination
Available as accessories: Fluke 700TC1 and TC2 Thermocouple
Mini-plug Kits
Fluke 712 RTD Calibrator
Compatible with pulsed current transmitters
Measure temperature from an RTD probe
Simulate RTD output
Operates with seven types of RTD
Measure additional RTDs using Ohms measurement function
Simulate additional RTDs using Ohms source function
°F or °C selectable
Four shrouded banana jacks
Functional Specifications
Range Resolution Accuracy Types
Fluke 712 Measure/simulate RTD -200 °C to 800 °C (Pt
100-385)
0.1 °C, 0.1 °F 0.2 °C, 0.4 °F
(Pt 100-385)
Pt; 100 200 500 1000
(385); Pt 100 (392); Pt 100
(392) JIS; Ni 120 (672)
Measure/simulate resistance 0 Ω to 3200 Ω 0.1 Ω .025 % + 0.1 % to
0.55 %
Fluke 714 Measure/simulate
thermocouple
–200 °C to 1800 °C,
depending on type (K,
–200 °C to 1370 °C)
0.1 °C or °F
(1 °C or °F; BRS)
0.5 °C, 0.8 °F 9 TC types; J K T E R S B
per NIST 175 and ITS-90
L U per DIN 43710 and
IPTS-68
Measure/simulate mV –10 mV to 75 mV 0.01 mV 0.025 % + 1 count
General Specifications
Maximum voltage: 30 V
Non-operating temperature: –40 °C to 60 °C
Operating temperature: –10 °C to 55 °C
Relative humidity: 95 % (10 °C to 30 °C); 75 % (30
°C to 40 °C);
45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C)
Operating altitude: 3,000 m max
Shock: 1 m drop test
Vibration: Random, 2 g, 5 Hz to 500 Hz
Safety: CSA C22.2 No. 1010.1:1992 EMC: EN50082-
1:1992 and EN55022:1994 Class B
Size/weight (HxWxD): 187 mm x 87 mm x 32 mm
(7.35 in x 3.41 in x 1.25 in)
330 g (12 oz)
Size/weight (HxWxD) (with holster and Flex-
Stand™): 201 mm x 98 mm x 52 mm (7.93 in x 3.86
in x 2.06 in) 600 g (21 oz) 992 g (35 oz)
Power: 9 V battery ANSI/NEDA 1604A or IEC
6LR619V alkaline; two batteries in 718
Battery life: 4 to 20 hours, typical, depending
on functions used. Battery timeout (configurable)
extends battery life.
Warranty: Three-years
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2008 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 8/2008 2574114 D -EN-N Rev B
Modification of this document is not permitted
without written permission from Fluke Corporation.
Fluke. Keeping your world
up and running.®

Navigation menu