Fluke 724 Users Manual

2015-09-09

: Fluke Fluke-724-Users-Manual-806443 fluke-724-users-manual-806443 fluke pdf

Open the PDF directly: View PDF PDF.
Page Count: 4

Manual Supplement
© 2003-2015 Fluke Corporation. All rights reserved.
Manual Title: 724 Users Supplement Issue: 7
Print Date: February 2000 Issue Date: 7/15
Revision/Date: 1, 8/03 Page Count: 3
This supplement contains information necessary to ensure the accuracy of the
above manual. This manual is distributed as an electronic manual on the
following CD-ROM:
CD Title: 724
CD Rev. & Date: 2, 8/2003
CD PN: 1547849
724 Users Manual Supplement
5/07 1
Change #1
On page 41, add the following to the Temperature, Thermocouples table:
Type Range Measure and Source
Accuracies (ITS-90)
XK -200 to 800 °C -200 to -100 ± 0.5 °C
-100 to 800 ± 0.6 °C
BP 0 to 2500 °C 0 to 800 ± 1.2 °C
800 to 2500 ± 2.5 °C
Change #2
On page 16, under the Note, replace the first sentence with the following:
Available with V1.4 Firmware or greater.
Change #3, 41513
On page 43, under General Specifications, add the following:
Protection Class Pollution Degree II
Manual Supplement 724 Users
2 11/14
Change #4
On page 37, Table 7:
Change:
18 Test lead, red
Test lead, black 688051
688066 1
1
To:
18 Fluke-7XX Test Lead Set 3397308 1
Change #5, 67401, 169, 512
On page 7, under Table 2, add:
Symbol Meaning
Consult user documentation.
Conforms to relevant Australian EMC standards.
Conforms to relevant South Korean EMC Standards.
724 Users Manual Supplement
7/15 3
On page 43, under General Specifications, replace the Safety section and add:
Safety IEC 61010-1: Pollution Degree 2
Electromagnetic Compatibility
(EMC)
International IEC 61326-1: Portable Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio
frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and
those directly connected to a low-voltage power supply network that supplies buildings
used for domestic purposes. There may be potential difficulties in ensuring
electromagnetic compatibility in other environments due to conducted and radiated
disturbances.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment
is connected to a test object.
Korea (KCC) Class A Equipment (Industrial Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment
and the seller or user should take notice of it. This equipment is intended for use in
business environments and not to be used in homes.
USA (FCC) 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.

Navigation menu