Fluke 754 Ing Users Manual MS2
2015-09-09
: Fluke Fluke-754-ing-Users-Manual-806352 fluke-754-ing-users-manual-806352 fluke pdf
Open the PDF directly: View PDF
.
Page Count: 2

Manual Supplement
© 2012 Fluke Corporation. All rights reserved.
Manual Title: 753/754 Calibration Supplement Issue: 2
Part Number: Web Only Issue Date: 12/12
Print Date: November 2011 Page Count: 1
Revision/Date:
This supplement contains information necessary to ensure the accuracy of
the above manual. Enter the corrections in the manual if either one of the
following conditions exist:
1. The revision letter stamped on the indicated PCA is equal to or higher
than that given with each change.
2. No revision letter is indicated at the beginning of the change.

753/754 Calibration Manual Supplement
12/12 1
Change #1, 61489
On page 8, in the Temperature, Thermocouples Table:
Change: Normal Mode Rejection: 40 dB at 50 Hz or 60 Hz nominal
To: Normal Mode Rejection: 65 dB at 50 Hz or 60 Hz nominal
Change #2, 64190, 64295
On page 3, add the following to the Symbols Table:
Conforms to CAN/CSA-C22.2 No. 61010-1, second edition, including
Amendment 1.
Conforms to relevant South Korean EMC Standards.
On page 4, replace the Standard and Agency Approval Specification with:
Standards and Agency Approval Specifications
Intrusion Protection ............................................. IEC 60529-1, IP 52
Safety ..................................................................... IEC 61010-1, CAT II 300 V, Pollution degree 2
Electromagnetic Environment ............................. IEC 61326-1, Basic
Electromagnetic Compatibility ............................ Accuracy for all function is not specified in RF fields >3 V/m
Applies to use in Korea only. Class A Equipment (Industrial Broadcasting &
Communication Equipment) [1]
[1] This product meets requirements for industrial (Class A) electromagnetic
wave equipment and the seller or user should take notice of it. This
equipment is intended for use in business environments and is not to be