1
902
HVAC Clamp Meter
Calibration Information
Introduction
XWWarning
To avoid electric shock or injury, do not perform the performance tests or
calibration procedures unless you are qualified to do so.
The information provided in this manual is for the use of qualified personnel
only.
The 902 Calibration Information provides the information necessary to verify the performance and adjust
the calibration of the Fluke 902 HVAC Clamp Meter, hereafter known as the Meter.
The following information is included in this document:
Safety Information and International Electrical Symbols
Specifications
Maintenance
Performance Tests
User-Replaceable Parts and Accessories
Warranty Statement
See the 902 Users Manual for complete operating instructions.
Contact Information
To contact Fluke, call:
1-888-99-FLUKE (1-888-993-5853) in USA
1-800-36-FLUKE (1-800-363-5853) in Canada
+31 402-675-200 in Europe
+81-3-3434-0181 Japan
+65-738-5655 Singapore
+1-425-446-5500 in other countries
For additional information about Fluke, its products, and services, visit Fluke’s web site at:
www.fluke.com
To register this product, go to register.fluke.com
902
Calibration Information
2
Safety Information
A “XW Warning” statement defines hazardous conditions and actions that could cause bodily harm or
death.
A “W Caution” statement identifies conditions and actions that could damage the Meter or the
equipment under test.
XWRead First: Safety Information
To avoid possible electric shock or personal injury, and to avoid possible
damage to the Meter or the equipment under test, adhere to the following
practices:
Read the Users Manual before use and follow all safety instructions.
Use the Meter only as specified in the Users Manual; otherwise, the Meter's
safety features may be impaired.
Avoid working alone so assistance can be rendered.
Never use the Meter on a circuit with voltages higher than 600 V or a
frequency higher than 60 Hz fundamental. The Meter may be damaged.
Never measure ac current while the test leads are inserted into the input
jacks.
Do not use the Meter or test leads if they look damaged. Inspect the case.
Do not use the Meter if it is damaged. Look for cracks or missing plastic.
Pay particular attention to the insulation around the connectors.
Use extreme caution when working around bare conductors or bus bars.
Contact with the conductor could result in electric shock.
Use caution when working with voltages above 60 V dc or 30 V ac rms or
42 V ac peak. Such voltages pose a shock hazard.
Clean the case with a damp cloth and mild detergent only. Do not use
abrasives or solvents.
To avoid false readings that can lead to electrical shock and injury, replace
the batteries as soon as the low battery indicator (B) appears.
Verify the Meter’s operation by measuring a known voltage. Do not use the
Meter if it operates abnormally. Protection may be impaired. When in doubt,
have the Meter serviced.
Do not hold the Meter anywhere beyond the tactile barrier, see Figure 1.
Adhere to local and national safety codes. Individual protective equipment
must be used to prevent shock and arc blast injury where hazardous live
conductors are exposed.
Use the proper terminals, function, and range for your measurements.
Do not operate the Meter with the case (or part of the case) removed.
When servicing the Meter, use only specified replacement parts.
HVAC Clamp Meter
Symbols
3
Symbols
The following symbols are found on the Meter or in this manual.
, May be used on hazardous live conductors
W Risk of danger. Important information. See Users Manual.
X Hazardous voltage. Risk of electric shock.
T Double insulation
B Battery
) Complies with Canadian and US Standards
P Conforms to relevant European Union directives
J Earth ground
F DC (Direct Current)
B AC (Alternating Current)
~ Do not dispose of this product as unsorted municipal waste. Contact Fluke or a qualified recycler for disposal.
;
N10140 Conforms to relevant Australian standards
s Inspected and licensed by TÜV Product Services
Electrical Specifications
Function Range Resolution Accuracy
Voltage DC 0 – 600 V 0.1 V 1 % ± 5 counts
Voltage AC (True Rms) 0 – 600 V 0.1 V 1 % ± 5 counts (50/60 Hz)
Current AC (True Rms) 0 – 600 A 0.1 A 2.0 % ± 5 counts (50/60 Hz)
Current DC 0 - 200 µA 0.1 µA 1.0 % ± 5 counts
Resistance 0 – 999
0 – 9999
0.1
1.0 1.5 % ± 5 counts
Continuity < 30
Temperature -40 to 400 °C 0.1 °C
-40 to -10 °C
± 5.0 % + 1.5 °C typical
-10 to 400 °C
± 1.0 % + 0.8 °C typical
Capacitance 1-100 µF
100-1000 µF
0.1 µF
1 µF 1.9 % ± 2 counts
902
Calibration Information
4
General Specifications
Operating Temperature -10 °C to +50 °C
Storage Temperature -40 °C to +60 °C
Operating Humidity
Non condensing (< 10 °C)
90 % RH (10 °C to 30 °C)
75 % RH (30 °C to 40 °C)
45 % RH (40 °C to 50 °C)
(Without Condensation)
Operating Altitude 2500 meters above mean sea level
Storage Altitude 12,000 meters above mean sea level
IP Rating IP 30 per IEC 60529
Vibration Requirements MIL-PRF-28800F Class 2 random vibration
EMI, RFI, EMC
EMI: instrument unspecified for use in EMC field 0.5 V / Meter
EMC: Meets all applicable
requirements in EN61326-1
Temperature
Coefficients
0.1 x (specified accuracy)/ °C
(<18 °C or >28 °C)
Size (H X W X L) 9.1 x 3.8 x 1.7 inches
(240 x 80 x 40 mm)
Weight 1.1 lb
(310 g)
Design Standards and
Compliance EN/IEC 61010-1, EN/IEC 61010-2-032 and relevant U.S. and Canadian Standards
Agency Approvals P ) ; s
N10140
Measurement Category
600 V, CAT III
CAT III equipment is designed to protect against transients in equipment in fixed-equipment
installations, such as distribution panels, feeders and short branch circuits, and lighting
systems in large buildings.
Power Requirements Two AA Batteries, NEDA 15 A, IEC LR6
HVAC Clamp Meter
Maintenance
5
Maintenance
XWWarning
To avoid possible electric shock or personal injury, repairs or servicing not
covered in this manual should be performed only by qualified personnel.
Note
When the Meter displays CAL, the unit will lock and no measurements can be made. The Meter must be
sent to a Fluke Service Center for repair.
Cleaning the Meter
XWWarning
To avoid electrical shock, remove any input signals before cleaning.
W Caution
To avoid damaging the Meter, do not use aromatic hydrocarbons or chlorinated
solvents for cleaning. These solutions will react with the plastics used in the
Meter.
Clean the instrument case with a damp cloth and mild detergent.
Replacing the Batteries
XWWarning
To avoid false readings that could lead to possible electric shock or personal
injury, replace the batteries as soon as the low battery indicator (B) appears.
Disconnect the test leads before replacing the batteries.
To replace the batteries, see Figure 1:
1. Turn the rotary switch to “OFF” and remove the test leads from the terminals.
2. Use a Phillips screwdriver to loosen the battery compartment door screw, and remove the door from
the case bottom.
3. Remove the batteries.
4. Replace the batteries with two new AA batteries observing correct polarity.
5. Reattach the battery compartment door to the case bottom and tighten the screw.
902
Calibration Information
6
adc02f.eps
Figure 1. Replacing the Batteries
Performance Tests
XWWarning
To avoid electric shock, do not perform the performance test procedures unless
the Meter is fully assembled.
The following performance tests verify the complete operation of the Meter and checks the accuracy of
each meter function against the Meter’s specifications. If the Meter fails any part of the test, calibration
adjustment and/or repair is indicated. The Meter must be returned to a Fluke Service Center for repair or
adjustment. See Contacting Fluke.
In the performance tests, the Meter is referred to as the unit under test (UUT).
Preparing for the Performance Test
XWWarning
To avoid possible electric shock or personal injury:
Do not perform the following procedures unless qualified to do so. Some
procedures involve the use of high voltages.
Before handling the test connections and in between tests, make sure the
calibrator is in standby mode (STBY).
To prepare for the performance test:
1. Make sure that you have the required equipment, see Table 1.
2. Warm up the calibrator as required by its specifications.
3. Allow the temperature of the UUT to stabilize at room temperature (23 °C ± 5 °C [73 °F ± 9 °F]).
HVAC Clamp Meter
Performance Tests
7
Table 1. Required Equipment
Equipment Recommended Model
AC Calibrator Fluke 5520A
Digital Multimeter (DMM) Any Fluke model
50-Turn Current Coil Fluke 5500A/Coil
TC Adapter Accessory Fluke 80AK
K-Type TC Extension Wire --
(2) K-type mini-connectors Fluke 80CK
Testing the Display
Test the display by turning the Meter on while holding down the AC/DC button. Check all segments for
clarity and contrast. Refer to Figure 2.
epb01.eps
Figure 2. Testing the Display
Backlight Test
The Meter is equipped with a display backlight. To test the backlight, press b. The backlight will come
on and the unit will beep. To turn off the backlight, press b a second time. Refer to Figure 3.
epb03.eps
Figure 3. Testing the Backlight
902
Calibration Information
8
Keypad Test
To test the keypad, turn the Meter on and push each button separately. Each button push will cause the
Meter to beep.
Performance Test Procedure
To test each of the Meter's functions and operating ranges, do the following:
1. Connect the source to the Meter's µA V and COM input jacks.
2. Referring to Table 2, put the Meter into the desired function and range for each test.
3. Apply the indicated output from the source.
4. When using the amp function on the 5520A, make sure LCOMP on the 5520A is ON. See Figure 4 for
current performance test setup.
5. The reading on the Meter display should be within the Meter response limits shown in the table.
6. Repeat steps 1-4 for each function and range in Table 2.
If the Meter fails to perform within the Meter response limits indicated for each test in Table 2, the Meter
must be returned to a Fluke Service Center for repair or calibration adjustment.
epb07.eps
Figure 4. Amps Verification Setup
HVAC Clamp Meter
Performance Tests
9
Table 2. Performance Tests
Meter Response
Test
(Switch Position) 5520A Output Lower Limit Upper Limit
600 V/60 Hz 593.5 606.5
V
AC Voltage 10 V/60 Hz 9.4 10.6
600 V dc 593.5 606.5
0 V dc -0.5 0.5
V
DC Voltage
-600 V dc -606.5 -593.5
P
Continuity
25 Beeper on
9500 9353 9647
1500 1473 1527
950 935.3 964.7
P
Resistance
0 -0.5 0.5
200 µA dc 197.5 202.5
10 µA dc 9.4 10.6
0 µA dc -0.5 0.5
G
DC Current
-200 µA dc -202.5 -197.5
-40 °C -43.5 -36.5
0 °C -0.8 0.8
**F
Temperature
400 °C 395.2 404.8
1 µF 0.8 1.2
90 µF 88.1 91.9
200 µF 195 205
L
Capacitance
1 mF 979 1021
0.4 A, 60 Hz 19.1 20.9
12 A, 60 Hz 587.5 612.5
*K
AC Current
12 A, 50Hz 587.5 612.5
*Test using the 5500A Coil
**Use K-Type TC Extension Wire with 80CK, at both ends, for connection between 5520A and 80AK/Meter
902
Calibration Information
10
Replaceable Parts and Accessories
13
epb08.eps
Item # Description Part Number or
Model Number Qty
A Knob 2280990 1
B Case Top 2556956 1
C Button, Backlight 2281286 1
D LCD 2563370 1
E Button, Hold 2281273 1
F Screw, case (small) 2388412 1
G Battery,1.5V,0-150MA, AA Alkaline 376756 2
H Battery Door 2280907 1
I Screw, Battery Door 2388435 1
J Screw, case (large) 2388382 1
K **Input Receptacle Housing 2280983 1
L Keypad 2556963 1
M *TL75 Test Lead Set TL75 1
Not Shown Softcase 1997276 1
Not Shown Thermocouple Assembly, K-Type, Beaded, Molded Dual Banana Plug (80BK) 1997234 1
Not Shown 902 Users Manual (English-Printed) 2547887 1
Not Shown 902 Manuals CD (Containing English, French, German, Japanese, Simplified
Chinese, and Spanish Users Manuals) 2643264 1
Not Shown 902 Calibration Information 2649733 1
* Fluke accessories are available from your authorized Fluke distributor.
** Use exact replacement only.
Figure 5. Replaceable Parts and Accessories
HVAC Clamp Meter
Replaceable Parts and Accessories
11
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is
three years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only
to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any
product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of
operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days
and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without
interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to
extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke
authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs
of repair/replacement parts when product purchased in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective
product which is returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization information, then send the
product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk
for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke
determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling,
including overvoltage failures caused by use outside the product’s specified rating, or normal wear and tear of mechanical components,
Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be
returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping
Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR
IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR
LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or
consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is
held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or
enforceability of any other provision.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
11/99 To register this product, go to register.fluke.com
902
Calibration Information
12
Page 1 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 2 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 3 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 4 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 5 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 6 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 7 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 8 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 9 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 10 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 11 of 12 - Fluke Fluke-902-Users-Manual 902_caltext
Page 12 of 12 - Fluke Fluke-902-Users-Manual 902_caltext

Navigation menu