Fukuda Denshi Co LX8100 ECG & Respiration Transmitter User Manual 4L0114970 LX 8100 FA V01

Fukuda Denshi Co Ltd ECG & Respiration Transmitter 4L0114970 LX 8100 FA V01

05 (Short-Term Confidential) User Manual

i Thank you for purchasing our product. Before using this product, read this operation manual thoroughly for correct handing and operation. Safety Precautions Read the “Safety Precautions” thoroughly before use to ensure correct and safe use of the product. Make sure to follow the precautions indicated below, as these are important messages related to safety. DANGER Failure to follow this message may cause immediate threat of death or serious injury, or complete failure of the equipment. WARNING Failure to follow this message may result in death or serious injury, or complete failure of the equipment. CAUTION Failure to follow this message may cause injury or failure of the equipment. NOTE A note is not related to product safety, but provides information about the correct use and operating procedures to prevent incorrect operation and malfunction of the equipment.   Precaution from Fukuda Denshi Fukuda Denshi is responsible for the safety, reliability, and performance of its equipment only if; Maintenance, modifications, and repairs are carried out by authorized personnel. Components are used in accordance with Fukuda Denshi operating instructions. If the equipment is used incorrectly and become unusable, Fukuda Denshi is not liable for the malfunction. Use the equipment only for the purpose specified in this manual.
ii  Intended Use of this Equipment This equipment is designed for the following <Intended Use>.  <Intended Use> This equipment is intended for measuring parameters of ECG, respiration in surgery room, ICU, ward, emergency room in the medical facility and transmitting the measured data by wireless network to the central monitor continuously. This equipment is intended for monitoring one patient. It is not intended for monitoring multiple patients. For specification of this equipment, refer to "Chapter 14 Specification" of this manual.   This equipment is intended to be used by healthcare professionals. Users should have a thorough knowledge of the function and operation before using this equipment. The maintenance of this equipment should be performed by skilled personnel who received a training of possible hazards and measures to avoid those hazards. Also, your local regulation must be followed. If this equipment is used for the purpose other than intended, or if the user does not follow the safety instructions, the following hazard may result. ȷHazard to the Life and Health of the Patient or the User ȷA Problem Related to Medical Practice ȷDamage to the Equipment
iii  Graphic Symbol The following symbols are used for this equipment. LX-8100 Symbols indicated on the main unit Symbol Description  Warning (indicated in yellow) Follow operating instructions (Warning); (indicated in blue) Indicates that the failure to follow operating instructions could place the patient or operator at risk. Type CF Applied Part with Defibrillation-Proof Indicates that the degree of protection against electric shock is Type CF Applied Part with defibrillation-proof. Indicates the power ON/OFF status.  Indicates the point to close the battery compartment lid. Indicates the battery type and direction  Name and Address of Manufacturer Date of Manufacture Indicates the name and address of manufacturer, and date of manufacture.  Non-ionizing electromagnetic radiation Indicates the radio transmitting device. 
iv LX-8100 Symbols displayed on the display screen Symbol Description  Synchronized Mark This mark flashes synchronizing to the heartbeat. Indicates the remaining battery level.
v Precautions for Safe Operation of Medical Electrical Equipment Cautions described here are regarding the general instructions for safety use to the patient and the users. As for cautions about the LX-8100, please refer to the following pages.  CAUTION 1.  Users should have a thorough knowledge of the operation before using this equipment. 2.  Pay attention to the following when installing or storing the equipment.y  Do not install or store in a place where the equipment will be subjected to splashing water. y  Do not install or store in an area where the environmental conditions, such as atmospheric pressure, temperature, humidity, ventilation, sunlight, dust, sodium, sulfur, will adversely affect the system. y  Place the equipment on a stable surface where there is no inclination, vibration, or shock (including during transportation). y  Do not install or store in an area where chemicals are stored or gas may generate. 3.  Before operating the equipment, verify the following items. y  Check the cable connection and polarity to ensure proper operation of the equipment. y  Ensure that all cables are firmly and safely connected. Especially, recheck the attachment and connection condition of electrodes and transducers. y  Pay special attention when the equipment is used in conjunction with other equipment as it may cause erroneous diagnosis and danger. y  Check the remaining battery level. y  When replacing the battery, make sure that the battery polarity is correct. Do not charge the battery. 4.  During operation of the equipment, verify the following items. y  Do not operate the equipment beyond the time period required for diagnosis and medical care. y  Do not pick up and/or swing the equipment pulling/grabbing the probe (sensor) or cable part. It may damage the equipment and lead to measurement error. y  Always observe the equipment and patient to ensure safe operation of the equipment. y  If any abnormality is found on the equipment or patient, take appropriate measures such as ceasing operation of the equipment and/or detaching the probe (sensor) and/or electrode, in the safest way for the patient. y  Do not allow the patient to come in contact with other equipment.
vi CAUTION 5.  After using the equipment, verify the following items. y  Return all operating switches, knobs etc to the position before using the equipment, and then switch off the power. y  When unplugging the cables, do not apply excessive force on the cable and pull from its connector. y  Clean the accessories and cables, and keep them together in one place. y  Keep the equipment clean to ensure proper operation for the next use. y  Make sure to remove the battery if the equipment is not used for a long time. The leakage from the battery may damage the equipment or an explosion from the battery may occur. 6.  If the equipment is damaged and in need of repair, ensure patient safety by immediately turning the equipment off and remove the electrodes and/or probe from the patient. Users should not attempt service. Label the unit “OUT OF ORDER” and contact Fukuda Denshi representative. 7.  Do not disassemble or remodel the eqipment. 8. Maintenance y  Make sure to periodically check the equipment and accessories. y  Before reusing the equipment that has been left unused for a while, make sure that the equipment works normally and safely. 9.  When using electrosurgical knives or defibrillator with this equipment, take care of the following. y  To prevent burn injury to the patient, verify proper attachment of the patient ground plate, the ECG electrode type for the electrosurgical knives, and the quantity of gel, output energy for the defibrillator. Also, verify that a proper ground is selected. y  Some types of equipment other than above may cause accidental hazards to the patient and operator due to the conditions of the equipment. Read the operation manual attached to each equipment and understand the precautionary instructions prior to use. Non-Explosion Proof DANGER z  Never operate the equipment in the presence of flammable anesthetics, high concentration of oxygen. It may cause an explosion or fire. z  Never operate the equipment inside a hyperbaric chamber. It may cause an explosion or fire. z  Never operate the equipment where flammable gas or fluid such as anesthetic, oxygen, and hydrogen are used. It may cause an explosion or fire.
vii Precautions about Magnetic Resonance Imaging (MRI) WARNING z  Do not operate this equipment in magnetic resonance imaging (MRI) environments. z  When conducting MRI test, remove the electrodes and sensors connected to the patient (test subject). The local heating caused by the induced electromotive force may cause burn injury to the patient (subject). For details, refer to the operation manual for the MRI testing device.  Electrosurgery Safety WARNING The monitoring system contains protection against interference generated by electrosurgical instruments. However, depending on the operating conditions, surgery site with respect to the location of ECG electrodes, ground plate attachment condition, or the type of instrument used, it may cause burn injury at the electrode site or noise on the ECG.    The noise is generated at the tip of the electrosurgical knife and is difficult to completely eliminate because of the frequency components of the ECG. To reduce electrosurgical interference, take the following precautions:  Location: Locate the electrosurgical unit as far as possible from this equipment and the patient cable. This will help reduce interference on the ECG through the monitor or cables. Electrode Placement: The amount of noise interference is considerably different depending on the electrode position and surgery site. Place the ECG electrodes as far away as possible from the surgery site and the ground plate. Do not place electrodes in the path between the surgery site and the ground plate. If the electrodes are placed in this path, the amount of interference will be quite large. Position (+) and (–) electrodes as close as possible to each other. Ground Plate: When using electrosurgical instruments, make sure the contact between the patient and the ground plate is secure. If the connection is incomplete, the patient may suffer from burn at the electrode site. 
viii Defibrillation Safety WARNING z  Use only the lead cable specified by Fukuda Denshi when defibrillating. If used by unspecified lead cable, the equipment may be damaged, resulting in a safety hazard. z  When using the defibrillator, keep away from the electrodes or medicament applied to the patient chest. If this is not possible, remove the electrodes or medicament before using it. If the defibrillator paddles are directly in contact with the electrodes or medicament, an electrical shock may result from the discharged energy. z  When using the defibrillator, do not touch the patient and the metal part of the equipment or cables. Electric shock may result from the discharged energy.  Precautions about the Pacemaker WARNING z  Minute ventilation rate-adaptive implantable pacemakers can occasionally interact with certain cardiac monitoring and diagnostic equipment, causing the pacemakers to pace at their maximum programmed rate. The cardiac monitoring and diagnostic equipment may possibly send wrong information. If such event occurs, disconnect the cardiac monitoring and diagnostic equipment, or follow the procedures described in the operation manual of the pacemaker. (For more details, contact FUKUDA DENSHI personnel, your institution’s professionals, or your pacemaker distributors.) z  Rate meters may continue to count the pacemaker rate during occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter alarms. Keep pacemaker patients under close surveillance.  Reference “Minute Ventilation Rate-Adaptive Pacemakers” FDA alerts health professionals that minute ventilation rate-adaptive implantable pacemakers can occasionally interact with certain cardiac monitoring and diagnostic equipment, causing pacemakers to pace at their maximum programmed rate. [October 14, 1998 – FDA]
ix Precautions about the LX-8100 WARNING z  Do not connect cables not authorized by Fukuda Denshi to any I/O connector. If done so by mistake, the LX-8100 cannot deliver its maximum performance and may be damaged, resulting in a safety hazard. z  Do not use this equipment with multiple patients simultaneously. z  This equipment itself has no alarm function. Do not use it if an alarm function is necessary. The alarm function with the receiving monitor, refer to the operation manual of the receiving monitor. CAUTION Do not pick up and/or swing the LX-8100 pulling/grabbing the probe (sensor) or cord part. The cable could break or get disconnected from the LX-8100. And it may hit people or damage other equipment around. Precautions about Waterproof CAUTION z  Replace the “Battery Compartment Lid” of the LX-8100 regularly to keep the performance of waterproof. If not regularly replaced, the quality of the lid will deteriorate and cannot keep the waterproof performance. For details about the regular replacement, contact your local Fukuda Denshi service representative. z  The lid may be damaged from high impact. If the LX-8100 is dropped or is subjected to a high impact, make sure that the lid is not damaged. z  Do not use the LX-8100 wet. Always wipe the LX-8100 with a soft cloth and dry it thoroughly before use. 
x Precautions about ECG CAUTION z  When removing electrodes from the patient, remove them carefully and slowly. Do not apply excessive force to remove them. Otherwise, it may damage the skin. z  If any electrodes get detached from the patient after being connected to the lead cable and the patient monitor, pay attention that the metal part of the electrode does not get in touch with any metal parts of the bed or any conductive parts. Also, the operator should not touch any conductive parts with bare hands. Otherwise, it may cause electric shock to the patient and/or operator due to excessive leakage current. z  The indication for continuous use of the electrode is about one day. z  Replace the electrode if the skin contact gets loosen due to perspiration, etc. z  When an electrode is attached to the same location for a long period, some patients may develop skin irritation. Check the patient's skin condition periodically and change the electrode site as required. z  For stable ECG monitoring, verify proper electrode placement, lead, and waveform size. If not properly selected, it may cause erroneous detection. z  There are some cases when the pacemaker pulse cannot be detected depending on the pacemaker type, pulse voltage, pulse width, electrode lead type (unipolar, bipolar), or electrode placement which causes the pacemaker pulse amplitude to decrease, and disables the pacemaker pulse detection. z  If signals similar to a pacemaker pulse are present, such as electric blanket noise or excessive AC frequency noise, these may be erroneously detected and displayed as a pacemaker pulse. In this case, check the condition of the electrodes and ECG lead cable to resolve the cause or turn off the pacemaker detection setting on the receiving monitor. z  If a pacemaker pulse is continuously detected due to AC frequency interference, QRS detection will be suspended and the heart rate will be reduced.  Precautions about Output Signal WARNING Do not use the output signal of the monitor that receives radio wave signal from the LX-8100 as the trigger signal for IABP, MRI echocardiographic, or defibrillator for the following reasons. It may lead to a delay of operating timing due to the delay time of waveform transmission. A trigger signal unrelated to the heart rate may be generated due to the interfusion of spike noise at weak electric field.
xi Precautions about Accessories and Optional Accessories WARNING Use only the accessories such as disposable electrodes and electrode codes specified by Fukuda Denshi. Otherwise, this equipment cannot deliver its maximum performance and may be damaged, resulting in a safety hazard.  CAUTION z  Do not reuse disposable products.   z  Store the disposable products properly as mentioned in their user manuals.  Precautions about Battery WARNING z  Use new "AA" size (“LR6” size) alkaline cell. z  Install the battery with the correct polarity. z  Do not charge the battery. Any attempt to charge the battery may cause it to leak or break. z  Do not short the (+) and (-) terminals. It may result in exothermic heat and fire. z  Do not throw the battery into fire. It may explode.  Precautions about Disposing of Equipment, Accessories, or Components CAUTION z  When disposing of equipment, accessories, or components, use an industrial waste distributor. Do not dispose of as ordinary waste. z  Used disposal items (ECG electrodes, etc.) shall be discarded as medical waste. Precautions about Disposing of Battery CAUTION Follow the local municipal rule to dispose the used dry cell batteries.
xii  Precautions for Use of Medical Telemeter WARNING z  The LX-8100 transmitter must not be co-located or operated in conjunction with any other antenna or transmitter. z  This equipment complies with FCC/IC radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure Guidelines and RSS-102 of the IC radio frequency (RF) Exposure rules. This equipment has very low levels of RF energy that are deemed to comply without testing of specific absorption rate (SAR).z  Operation of LX-8100 requires the prior coordination with a frequency coordinator designated by the FCC for the Wireless Medical Telemetry Service.  z  This radio frequency device is susceptible to interference from outside sources. Interference may prevent the monitoring of patients connected to this equipment. If a problem exists, contact your local service representative. z  The LX-8100 transmits vital signs to the receiving monitor using radio wave signal. Under unstable radio wave signals, the receiving monitor will not generate any alarms. This situation may miss sudden change in the patient's condition and may cause a serious accident. Under unstable radio wave signals, check the patient status consistently under this situation. To get stable radio wave signals, make sure to have a proper telemetry installation.  CAUTION For installation, make sure of the following. y  The medical institution (hereinafter referred to as the “Institution”) must decide the telemetry installation plan for the medical department in order to prevent interference and interference between transmitters (telemetry based on destination country’s radio law). When telemetry has already been installed and been used, radio format, frequency, and antenna power are required to be examined to prevent interference. y  When laying receiver antenna for each transmitter, the Institution has to examine the installation so that electronic interference does not occur. y  Based on the above examination result, the Institution should install each receiver antenna as required. 
xiii CAUTIONFor management, make sure to follow the precautions below. y  The Institution should appoint a person (hereinafter referred to as the “Coordinator”) to manage the wireless channels for the whole Institution. y  The Coordinator must be selected from people who understand the characteristics and functionality of telemetry systems, and are skilled in operating telemetry. y  When installing telemetry, the Coordinator has to understand the precautions for use of telemetry in advance. y  The Coordinator is responsible for maintenance of wireless channels and storage and maintenance of telemeter in the overall medical facilities to give proper instructions to the telemetry users. y  The Coordinator should create a management log (hereinafter referred to as the “log”), which contains a list of the management status of the wireless channels for the whole Institution. When changing a wireless channel, register it in the log and give proper instructions to the user. y  The telemetry user verifies operation of the transmitter/receiver before use.y  When interference or breakdown occurs in telemetry communication, the user is required to inform the Coordinator of the problems. The Coordinator is to deal with the problem properly and/or contact their nearest Fukuda Denshi representative for service. 
xiv Electromagnetic Compatibility This equipment complies with IEC 60601-1-2 (2014), safety standard regarding the electromagnetic distrubances of medical electrical equipment. To ensure maximum performance against the electromagnetic distrubances, make sure to follow the precautions for installation and usage described in this manual.  ŪThis equipment is intended for use in the medical facility (except inside the shield room of MRI device), and satisfies the immunity level for professional healthcare facility environment stipulated in IEC 60601-1-2.  ŪAn excessive magnetic disturbance may degrade the HR measurement accuracy (refer to 14. Specification, ECG, HR measurement detection), which is the essential performance of this equipment, and may cause delay in treatment or inaccurate diagnosis.  ŪWhen using this equipment, interference with other medical electrical equipments or non-medical electrical equipments may oocur. Make sure that no interference is present before usage.  ŪTo ensure basic safety and essential performance related to electromagnetic distrubances during the expected service life of this equipment, “Daily Check” and “Periodic Check” must be performed. (refer to 12.Maintenance and Inspection)  Precautions for Safe Operation under Electromagnetic Influence CAUTION If any sorts of electromagnetic wave, magnetic field, or static electricity exist around the equipment, noise interference or malfunction of the equipment may occur. If any unintended malfunction or noise occurs during monitoring, check the magnetic influence and take appropriate countermeasures.  The followings are examples of the common cause and countermeasures.ەMobile Phone The radio wave may cause malfunction to the equipment.   Mobile phones and radio sets should be turned off in the room (building) where medical device is located. ەStatic Electricity In a dry environment (room), static electricity is likely to occur. Take the following countermeasures. y  Both operator and patient should remove any static electricity before entering the room. y Humidify the room. 
xv CAUTION z  If this equipment is installed close to, or stacked with other equipment, malfunction may occur. Make sure to verify that the equipments operate properly in a used location. z  Use of accessories, probes, or cables other than specified may cause increase in electromagnetic emission or decrease in electromagnetic immunity resulting in malfunction of the equipment. z  The portable RF communications equipment (including antenna cable and peripheral equipment such as external antenna) with the specified cable should be used in a location at least 30 cm apart from any part of this equipment. Otherwise, it may result in performance degradation of this equipment. 
xvi EMC Guidance This equipment complies with IEC 60601-1-2 (2014). However, if portable transmitter or wireless LAN equipment is used extremely nearby, the electromagnetic influence may largely exceed the compliance level and may cause unexpected phenomenon such as noise interference on the waveform, etc. Therefore, this equipment should be used in a location specified by each medical institution. If any unexpected noise interference on the waveform or failure to the peripheral device occurs, stop using the equipment and follow the instruction of the technician.  The following is the information relating to EMC (Electromagnetic Compatibility). (When using this equipment, verify that it is used within the environment specified below.)  䖃Compliance to the Electromagnetic Emissions This equipment complies with the following emission standard. Emission test  Compliance RF Emission CISPR 11  Group 1 Class A CAUTION The emission performance of this equipment is suitable for use in industrial environment and hospital environment (CISPR 11 Class A). To use in home environment (generally, CISPR 11 Class B is required), this equipment may not be properly protected from wireless frequency communication service. It may be necessary to take measures such as changing the installation location or equipment orientation. 
xvii 䖃Compliance to the Electromagnetic Immunity The LX-8100 is intended for use in the electromagnetic environment specified below. The customer or the user of the LX-8100 should assure that it is used in such an environment. Basic EMC standard or test method  Immunity test levels Electrostatic discharge IEC 61000-4-2  r8kV  contact r2kV, r4kV, r8kV, r15kV air Radiated RF EM fields IEC 61000-4-3  10V/m 80MHz to 2.7GHz 1kHz80%AM Proximity fields from RF wireless communications equipment IEC 61000-4-3 Refer to the following table. Conducted disturbances induced by RF fields IEC 61000-4-6 3V 0.15MHz to 80MHz 1kHz80%AM 6V 0.15MHz to 80MHz (in ISM bands between 0.15 MHz and 80MHz) 1kHz 80%AM Rated power frequency magnetic fields IEC 61000-4-8 30A/m 50Hz Immunity test specifications for RF wireless communications equipment. Test frequency(MHz)  Modulation  Maximum power (W) Distance (m)  Immunity test level (V/m) 710, 745, 780  PM, 217Hz  0.2  0.3  9 810, 870, 930  PM, 18Hz  2  0.3  28 1720, 1845, 1970 PM, 217Hz  2  0.3  28 2450 PM, 217Hz 2 0.3 28 5240, 5500, 5785 PM, 217Hz  0.2  0.3  9  
xviii Contact If you need more information, please contact the following.  (1)  Fukuda Denshi Co., Ltd., Head Office 3-39-4 Hongo, Bunkyo-ku, Tokyo, Japan Tel: +81-3-5684-1455  Fax: +81-3-3814-1222 E-mail: info@fukuda.co.jp Home Page: http://www.fukuda.com (2) Sales Representative Write the name, address, phone, fax number of your local sales representative. (Name of Sales Representative, Address, Phone/Fax) 
 Contents Safety Precautions ............................................................................................. i Precaution from Fukuda Denshi ................................................................ i Intended Use of this Equipment ............................................................... ii Graphic Symbol ...................................................................................... iii Precautions for Safe Operation of Medical Electrical Equipment ............. v Non-Explosion Proof ............................................................................... vi Precautions about Magnetic Resonance Imaging (MRI) ........................ vii Electrosurgery Safety .............................................................................  vii Defibrillation Safety ............................................................................... viii Precautions about the Pacemaker ........................................................  viii Precautions about the LX-8100 ............................................................... ix Precautions about Waterproof ................................................................ ix Precautions about ECG ........................................................................... x Precautions about Output Signal ............................................................. x Precautions about Accessories and Optional Accessories ..................... xi Precautions about Battery ....................................................................... xi Precautions about Disposing of Equipment, Accessories, or Components ........................................................................................................... xi Precautions about Disposing of Battery .................................................. xi Precautions for Use of Medical Telemeter ............................................. xii Electromagnetic Compatibility ................................................................... xiv Precautions for Safe Operation under Electromagnetic Influence ......... xiv EMC Guidance ...................................................................................... xvi Contact ......................................................................................................... xviii 1. General Description ...................................................................................... 1 2. Names of Parts and Their Functions ............................................................. 3 3. Preparation ................................................................................................... 5 (1) Installing the Battery ........................................................................... 5 (2) Operating Power Switch ..................................................................... 9 4. ECG Monitoring .......................................................................................... 11 ŠConnecting the ECG Lead Cable and Electrodes .............................. 11 ŠAttaching the Electrodes .................................................................... 14 ŠConnecting the ECG Lead Cable to the LX-8100 ............................... 15 5. Respiration Monitoring ................................................................................ 17 6. Measurement .............................................................................................. 19 ŠStarting Screen ................................................................................... 19 ŠWaveform Display Screen .................................................................. 19 ŠBattery Level Check ........................................................................... 21 ŠWaveform Display .............................................................................. 22 7. Operation .................................................................................................... 31 ŠChanging Setup .................................................................................. 31 ŠRestarting the display ......................................................................... 37
 ŠPressing the EVENT button ................................................................ 37 8. Other Setting Items ..................................................................................... 39 ŠChanging the Time Constant .............................................................. 40 ŠChanging the Detection Sensitivity of the Pacemaker Pulse .............. 40 ŠChanging QRS Detection ................................................................... 41 ŠChanging the Respiration Detection Signal ON/OFF ......................... 42 ŠChanging the display Brightness ........................................................ 42 ŠChanging Display Timeout Duration ................................................... 42 ŠChanging Sound ON/OFF .................................................................. 43 9. Changing the Transmitter Channel and Group ID ....................................... 45 ŠChanging the Transmitter Channel ..................................................... 45 ŠChanging the Group ID ...................................................................... 45 10. Troubleshooting ........................................................................................ 47 ŠList of Displayed messages ................................................................ 47 ŠIn Case of Dropping the LX-8100 into Water ...................................... 51 11. Cleaning and Disinfection ......................................................................... 53 ŠCleaning ............................................................................................. 53 ŠDisinfection ......................................................................................... 53 ŠCleaning the ECG lead cable ............................................................. 54 12. Maintenance and Inspection ..................................................................... 55 ŠRepairing the Equipment .................................................................... 63 ŠReplacing the Battery Compartment Lid Unit ..................................... 63 13 . Standard and Optional Accessories ......................................................... 67 ŠStandard Accessories ........................................................................ 67 ŠOptional Accessories .......................................................................... 68 14. Specification .............................................................................................. 69 ŠSpecification ....................................................................................... 69 ŠDisplays .............................................................................................. 73 ŠDetails of the “Electrode” Message .................................................... 75 ŠList of Setup Items .............................................................................. 76 
1 1. General Description The LX-8100 is a telemetry transmitter designed to measure the ECG and respiration waveform with a “AA” size (“LR6” size) alkaline battery. Information such as ECG measurements, respiration waveform, battery level, and the conditions of the ECG electrodes are displayed on the front panel. ECG lead selection is available using the two buttons (ENTER and  ť) on the front panel (In case of using a 3-electrode lead cable or a 5-electrode chest lead cable). Before using the LX-8100, read also the operation manual of the patient monitor at the receiving side thoroughly.   External Appearance 
1. General Description 2  
 3 2. Names of Parts and Their Functions         ECG/RESP Input Connector   Connects the ECG lead cable to measure ECG and respiration waveform.  Refer to “Safety   Precautions in this manual’s preface” EVENT ButtonTransmits EVENT InformationDisplay Displays measurement waveform and transmitter information. Power Switch Turns ON/OFF the power Electrode Position LabelIndicates standard ECG electrode position. Channel Number Indicates transmitter channel number. Power Supply LEDIndicates the power supply statusx  Light ON: In nomal operation x  Light OFF: Power Off ENTER ButtonFor SetupۃButtonFor Setup
2. Names of Parts and Their Functions 4 Strap Hole Attaches the strap. Battery Lid Lock Button Unlocks the battery compartment lid.Refer to “Safety Precautions” in this manual’s preface. Battery Compartment Lid Open/Close Lever To open/close the battery compartment lid, slide this lever up/down.
 5 3. Preparation (1) Installing the Battery The LX-8100 functions with an “AA” size (“LR6” size) alkaline battery. Use the battery specified by Fukuda Denshi.  Use a newßAA” size (“LR6” size) alkaline battery.  Specified types of batteries Type Product Manufacturer Operation time “AA” size (“LR6” size) alkaline battery  MX1500 DURACELL Approximately  6 days*1 *1: When using a new battery. (ECG measurement, RESP measurement, Default setting, Operating environments in 23qC)  *The operation times are estimated for operations with specified types of batteries by Fukuda Denshi.  WARNING z  Unplug the ECG lead cable when the battery compartment lid is opened. Otherwise, patient leakage current beyond the allowable value may occur. z  Use a new ”AA” size (“LR6” size) alkaline battery. z  Do not short out the (+) and (-) terminals. It may result in exothermic heat and fire, the leakage from the batteries may damage the equipment, or an explosion from the batteries may occur. z  Install the batteries with the correct polarity.   z  Do not charge alkaline batteries. Any attempt to charge the batteries may cause them to leak or break. z  Do not use disassembled or damaged batteries due to drop or shock. The leakage from the batteries may damage the equipment, or an explosion from the batteries may occur.
3. Preparation 6  WARNING z  Remove the exhausted batteries immediately. The leakage from the batteries may damage the equipment, or an explosion from the batteries may occur. z  If the equipment is not in use for a long period of time, remove the batteries and store the equipment in an appropriate place. If the batteries are left in the equipment for a long period of time, the leakage from the batteries may damage the equipment or an explosion from the batteries may occur.  CAUTION z  Use alkaline battery (AA) specified by Fukuda Denshi. z  Do not replace the batteries with wet hands. z  In case of storing the used or unused batteries, make sure that the terminals are not touching other batteries or metal parts. z  Use the same type of battery as its setting otherwise the accurate battery level cannot be displayed.
3. Preparation 7  Unlock and open the battery compartment lid by sliding the open/close lever towards OPEN while pressing the lock button. Install a new battery according to the polarity indication inside the battery compartment. After installing the battery, lock the battery compartment lid by sliding the open/close lever towards CLOSE while pressing over “PUSH” on the lid. Ĭ ĭ Ĭ ĭ
3. Preparation 8  Make sure that the battery compartment lid is locked. (If you can still see red, then it is not locked properly.)  CAUTION z  Make sure that any foreign particles, such as hairs, are not held on the battery compartment lid and dust is not adhered to the edge of the lid to prevent water entering into the battery compartment area. z  Do not keep the compartment lid unlocked as the battery may unexpectedly get out from the compartment.
3. Preparation 9 (2) Operating Power Switch Turning the power switch to “ON” Press the power switch.  Display screen turns ON and measurement starts. Regarding the display screen, refer to page 19.  The display screen automatically turns itself OFF by following the setting. After the power is turned ON, make sure to check the remaining battery level on the display screen. Refer to the following symbol about the remaining battery level. Battery SymbolRemaining Battery LevelFullGetting low but still availableNearly empty; Replace the battery  Turning the power switch to “OFF” Press the power switch for two seconds, then display screen displays as the left illustration to confirm. Choose “Yes” and press the  ENTER  button.
3. Preparation 10
4. ECG Monitoring 11 4. ECG Monitoring 䕔Connecting the ECG Lead Cable and Electrodes The optional ECG lead cables for LX-8100 are as follows. ECG Lead Cables Item No.  Applicable Lead  Remark CMT-01CTH-0.8DA  Limb Lead (1CH)  3-electrode Clip Type CMT-02CTH-0.8DA  Limb Lead (2CH)  4-electrode Clip Type CMT-03CTH-0.8DA  Limb Lead (1CH) +Chest (1CH)  5-electrode Clip Type CMT-01HTH-0.8DA  Limb Lead (1CH)  3-electrode Hook Type CMT-02HTH-0.8DA  Limb Lead (2CH)  4-electrode Hook Type CMT-03HTH-0.8DA  Limb Lead (1CH) +Chest (1CH)  5-electrode Hook Type  WARNING Use only the specified lead cable by Fukuda Denshi. Otherwise, proper monitoring may not be performed, and also it may fail defibrillation or cause a malfunction of the equipment when the equipment is used with a defibrillator. 
4. ECG Monitoring 12 The relations between the attached electrode positions and lead method are as follows. Attach the electrodes to monitor proper waveform.  For 3-electrode lead cable  Limb leads Standard Limb leads can be selected from lead I, lead II, or lead III under the setting of the equipment.  Refer to “7. Operation  ŠChanging Setup  ŪECG Display Screen (1) <<Switching Lead>>” .   For 4-electrode lead cable  Limb leads Two leads such as lead I and II can be measured. Lead III, aVR, aVL, and aVF can be also displayed from the setting on the receiving monitor. For details, refer to the operation manual of the receiving monitor  Black (LA) Red (LL) White (RA) Green (RL) White (RA)  Black (LA) Red (LL)
4. ECG Monitoring 13 For 5-electrode (Chest) lead cable   Limb Lead +Chest  One limb lead and one chest lead (Brown) measurements are available. Standard Limb leads can be selected from lead I, lead II, or lead III under the setting of the equipment. Refer to “8. Operation  ŠChanging Setup  ŪECG Display Screen (1) <<Switching Lead>>” . The chest lead waveform is measured from the chest lead (Brown) positioned on the chest. Brown(V)Red (LL) Green (RL) White (RA) Black (LA)
4. ECG Monitoring 14 䕔Attaching the Electrodes  CAUTION z  Always use the same type of electrodes. If different types of electrodes are used at the same time, the difference between the polarization potential from each electrode may interfere with monitoring. z  Do not reuse the disposable electrodes. It is intended for single patient use only.   Clean the electrode sites with alcohol wipes or other skin preparation. If necessary,   shave the electrode sites to remove excessive hair. Peel off the disposable electrode   Pay attention not to touch the electrode gel.   Attach the lead cable end to the electrode (convex part). Turn right and left to verify that it is securely attached.
4. ECG Monitoring 15 䕔Connecting the ECG Lead Cable to the LX-8100 Insert the ECG lead cable firmly into the ECG/RESP input connector matching the transmitter’s connector guide and the direction of the notched part on the connector. ECG Lead cable
4. ECG Monitoring 16 CAUTION z  There are some cases when pacemaker pulse cannot be detected depending on the pacemaker type, pulse voltage, pulse width, electrode lead type (unipolar, bipolar), electrode placement, or lead method which causes the pacemaker pulse amplitude to decrease and disables pacemaker pulse detection. z  If signals similar to a pacemaker pulse are present, such as electric blanket noise or excessive AC frequency noise, these may be erroneously detected and displayed as a pacemaker pulse. In this case, check the condition of the electrodes and ECG lead cable to resolve the cause or turn off the pacemaker detection setting on the receiving monitor. z  Time constant of this equipment is shorter than Fukuda Denshi monitors (direct ECG connection). Therefore, there is a difference in the ST measurement value between them. Pay attention to the difference when monitoring a patient from a transmitter or a monitor. z  When an electrode is attached on the same location for a long time, some patients may develop skin irritation. Check the patient’s skin condition periodically and change the electrode position as required. z  The indication for continuous use of an electrode is about one day. z  Replace the electrode if the skin contact gets loosen due to perspiring, etc. z  Make sure to use new disposable electrodes. Otherwise, the waveform quality may become poor and it may fail to perform correct monitoring z  When “Check Electrode” message is displayed on the screen of the receiving monitor or the display of this equipment, check the condition of the electrodes and ECG lead cable to resolve the cause. z  When removing electrodes from the patient, remove them carefully and slowly. Do not apply excessive force to remove them. Otherwise, it may damage the skin. z  It may not perform a correct measurement due to the attached position of the electrodes. Attach the electrodes on the patient referring to page 12 and 13 and make sure that the correct waveform is measured on the display.
 17 5. Respiration Monitoring Follow the preparation of “4. ECG Monitoring” to allow the respiration monitoring.  This respiration monitoring is performed with impedance method. The ECG electrodes are also used for detecting the respiration. Each lead cable specifies the electrodes to detect the respiration. For 3-electrode and 5-electrode (chest) lead cable, the electrodes to detect the respiration are fixed as follows. Even if lead method is switched, they are no changes.  Lead Cable  Color of Electrode 3-electrode  White (RA) and Red (LL) 4-electrode  White (RA) and Red (LL) 5-electrode (Chest) White (RA) and Red (LL) WARNING Minute ventilation rate-adaptive implantable pacemakers can occasionally interact with certain cardiac monitoring and diagnostic equipment, causing the pacemakers to pace at their maximum programmed rate. The cardiac monitoring and diagnostic equipment may possibly send wrong information. If such event occurs, please disconnect the cardiac monitoring and diagnostic equipment, or follow the procedures described in the operation manual of the pacemaker. (For more details, contact FUKUDA DENSHI personnel, your institution’s professionals, or your pacemaker distributors.) Reference “Minute Ventilation Rate-Adaptive Pacemakers” FDA alerts health professionals that minute ventilation rate-adaptive implantable pacemakers can occasionally interact with certain cardiac monitoring and diagnostic equipment, causing pacemakers to pace at their maximum programmed rate. [October 14, 1998 – FDA] CAUTION z  Even if the electrodes are attached on the proper positions for ECG monitoring, it may not be always the proper ones for respiration monitoring as well. z  When a defibrillator is used during respiration monitoring, a large offset voltage will be placed on the ECG electrodes, which may cause interruption of monitoring for a few seconds.
5. Respiration Monitoring 18 
 19 6. Measurement Turn ON the power and the measurement starts. 䕔Starting Screen When the power is turned ON, the channel number configured on the LX-8100 is displayed at the top of the display.        Make sure whether the channel number on the display matches the channel number indicated on the label of the LX-8100 and the channel number configured on the receiving monitor.   䕔Waveform Display Screen ECG waveform (1CH when using 3-electrode lead cable, 2CH when using other lead cable), heart rate, pacemaker marker, respirogram, respiration rate, remaining battery level, and various messages are displayed. CAUTION z  The LX-8100 does not have a diagnostic function. Check the diagnostic function on the receiving monitor. z  The LX-8100 does not have an alarm function. Check the alarm function on the receiving monitor. z  The ECG waveform size and sweep speed settings displayed on the display of the LX-8100 do not interface with the ones displayed on the screen of the receiving monitor. z  The heart rate and respiration rate displayed on the display of the LX-8100 may be different from the ones displayed on the receiving monitor because the algorithm of the ECG and respiration rate is different. 
6. Measurement 20 䖃Display Switch The screen can be switched to the other selected screen such as to ECG/RESP. The screen automatically turns itself OFF by following the setting if no operation is done. To restart the display, refer to “7. Operation   ŠRestarting the display”.  When the display is active, press the  ť  button to move onto the next   screen. The screen will be switched in the following order. Press and hold    ENTER    for 2 seconds. Press and hold    ENTER    for 2 seconds.Press and hold    ENTER    for 2 seconds. Move onto ECG1 Lead/Waveform size setting display. Move onto ECG2 Lead/Waveform size setting display. Move onto RESP waveform size setting display. ťťť
6. Measurement 21 䕔Battery Level Check Check the battery level on the waveform display screen.   Battery Symbol  Remaining Battery Level  Full  Getting low, but still available  Nearly empty Replace the battery.   A message that prompts the battery check appears on the screen of the receiving monitor.
6. Measurement 22 䕔Waveform Display 䖃ECG Display Screen (1) ECG1 waveform, heart rate, pacemaker marker, remaining battery level, and electrode check message are displayed. The screen automatically turns itself OFF by following the setting if no operation is done. To restart the display, refer to “7. Operation   ŠRestarting the display”.  When the display is active, press the  ť  button to move onto the next waveform display screen.  ࠙Descriptions of the Screenࠚ The descriptions of contents displayed on the display are as follows        Indicates ECG 1.          Indicates the vertical scale of the displayed ECG waveform. One scale corresponds to 1mV. In the left illustration, it can display ECG waveform between -1mV and +1mV.         Indicates the horizotal scale of the displayed ECG waveform. One scale corresponds to one second.         Displays the ECG waveform.
6. Measurement 23        Displays the heart rate.   Ɔ is displayed in synchronization with the heart rate.         Displays the detection marker when a pacemaker pulse is detected.         Indicates the measuring lead.   CAUTION z  The ECG waveform size setting displayed on the display does not interact with the one displayed on the screen of the receiving monitor. If the ECG waveform size displayed on the screen of the receiving monitor is changed, follow the instruction in the operation manual of the receiving monitor. z  When Wide is selected in the QRS Detection and in case it is outside of the heart rate range (12 bpm to 300 bpm), 0 bpm will be displayed if 11 bpm and below is measured and 300bpm will be displayed if   300 bpm and above is measured. z  When Narrow is selected in the QRS Detection and in case it is outside of the heart rate range (30 bpm to 300bpm), 0bpm will be displayed if 29 bpm and below is measured and 300 bpm will be displayed if 300 bpm and above is measured.         Indicates the ECG waveform size displayed on the display.
6. Measurement 24   Indicates the remaining battery level. For details of the battery level, refer to page 21.        Displays the electrode check. “Electrode?” message appears when an ECG electrode is detached.          Displays the speaker mark when Sync. Tone setting is active. For details, refer to “7. Operation   ŠChanging Setup”.
6. Measurement 25 䖃ECG Display Screen (2) ECG2 waveform, heart rate, pacemaker marker, remaining battery level, and electrode check message are displayed.  NOTE If a 3-electrode lead cable is used, this screen will not appear.  The screen automatically turns itself OFF by following the setting if no operation is done. To restart the display, refer to “7. Operation  ŠRestarting the display”.  When the display is active, press the  ť  button to move onto the next waveform display screen.  ࠙Descriptions of the Screenࠚ The descriptions of contents displayed on the display are as follows.         Indicates ECG 2.         Indicates the vertical scale of the displayed ECG waveform. One scale corresponds to 1mV. In the left illustration, it can display ECG waveform between -1mV and +1 mV.         Indicates the horizotal scale of the displayed ECG waveform. One scale corresponds to one second.          Displays the ECG waveform.
6. Measurement 26  Displays the heart rate.   Ɔ is displayed in synchronization with the heart rate. Displays the detection marker when a pacemaker pulse is detected. Indicates the measuring lead. CAUTION z  The ECG waveform size setting displayed on the display does not interact with the one displayed on the screen of the receiving monitor. z  If the ECG waveform size displayed on the screen of the receiving monitor is changed, follow the instruction in the operation manual of the receiving monitor. z  When Wide is selected in the QRS Detection and in case it is outside of the heart rate range (12 bpm to 300 bpm), 0 bpm will be displayed if 11 bpm and below is measured and 300 bpm will be displayed if   300 bpm and above is measured. z  When Narrow is selected in the QRS Detection and in case it is outside of the heart rate range (30 bpm to 300 bpm), 0bpm will be displayed if 29 bpm and below is measured and 300 bpm will be displayed if 300 bpm and above is measured. Indicates the ECG waveform size displayed on the display.     Indicates the remaining battery level. For details of the battery level, refer to page 21.
6. Measurement 27 Displays the electrode check. “Electrode?” message appears when an ECG electrode is detached. Displays the speaker mark when Sync. Tone setting is active. For details, refer to “7. Operation   ŠChanging Setup”. 
6. Measurement 28 䖃Respiration Display Screen Respiration waveform, respiration rate, remaining battery level, and electrode check message are displayed.  The screen automatically turns itself OFF by following the setting if no operation is done. To restart the display, refer to “7. Operation  ŠRestarting the display”.  When the display is active, press the  ť  button to move onto the next waveform display screen.   ࠙Descriptions of the Screenࠚ The descriptions of contents displayed on the display are as follows.  Indicates the respiration waveform display screen.Indicates the vertical scale of the displayed respiration waveform. One scale corresponds to 1ȍ. In the left illustration, it can display the change of respiration waveform between -1ȍ and +1ȍ.  Indicates the horizontal scale of the displayed respiration waveform. One scale corresponds to one second.   Displays the respiration waveform. 
6. Measurement 29 Displays the respiration rate.  CAUTION z  The ECG waveform size setting displayed on the display does not interact with the one displayed on the screen of the receiving monitor. z  If the ECG waveform size displayed on the screen of the receiving monitor is changed, follow the instruction in the operation manual of the receiving monitor. z  In case it is outside of the respiration rate range (4 Bpm to 150 Bpm), 0 Bpm will be displayed if 3 Bpm and below is measured and 150 Bpm will be displayed if 150 Bpm and above is measured.  Indicates the respiration waveform size displayed on the display.  Indicates the remaining battery level. For details of the battery level, refer to “6. Measurement  ŠBattery Level Check”.         Displays the electrode check “Electrode?” message appears when an ECG electrode is detached.
6. Measurement 30
 31 7. Operation 䕔Changing Setup 䖃ECG Display Screen (1) In the ECG display screen (1), the ECG waveform size and lead displayed on the display of the LX-8100 can be changed and the Synchronized Tone can be set up.   ࠙Setting Methodࠚ How to enter the setup mode: Press and hold the    ENTER    button for 2 seconds in the ECG display screen (1).  䍾Switching Lead䍿 Lead of ECG 1 can be switched when 3-electrode lead cable or 5-electrode (Chest) lead cable is used. Select an appropriate lead by checking the ECG waveform on the display.         The lead indication of ECG 1 is highlighted.  Pressing the  ENTER  button will sequentially change the lead of ECG 1. Lead I Æ II Æ III Æ I  䍾Changing ECG1 Waveform Size on display䍿         Press the  ť  button to highlight the size indication of ECG 1.    Pressing the  ENTER  button will sequentially change the size of ECG 1. Size ×1 Æ ×1/2 Æ ×1         When changing the size of the ECG waveform on the display, the ECG scale will also change. In the left illustration, it can display the ECG waveform between -2mV and +2mV.
7. Operation 32 CAUTION The ECG waveform size setting displayed on the display of the LX-8100 does not interact with the one displayed on the screen of the receiving monitor. To change the waveform size of the receiving monitor, follow the instruction in the operation manual of the receiving monitor.  䍾Returning to ECG Display Screen (1)䍿        Press the  ť  button to highlight the “Back”. Press the  ENTER  button to return to the ECG display screen (1). CAUTION Do not operate the LX-8100 with the setup screen open because it may lead to an unintended operation and cause setting changes. Make sure to press the “Back” to terminate the setup screen. The display will automatically turn itself OFF by following the setting if the “Back” is not pressed.   ࠙Setting the Synchronized Toneࠚ Press and hold the    ENTER    button and  ť  button for 2 seconds in the ECG display screen (1). The speaker symbol can be switched to display ON/OFF.  䍾Generate a Synchronized Tone䍿 When the speaker symbol is displayed, a Synchronized Tone will generate along with the Synchronized mark. The Synchronized Tone will generate only in the ECG screen.          Press and hold the    ENTER    button and ť  button for 2 seconds then the speaker symbol will be displayed. Press and hold the    ENTER    button and the  ť  button again for 2 seconds, then the speaker mark will disappear and the synchronized tone will stop. The setting synchronizes the ECG screen (2) so this setting is also reflected in the ECG screen (2).
7. Operation 33  䖃ECG Display Screen (2)   In the ECG display screen (2), the ECG waveform size displayed on the display of the LX-8100 can be changed and the synchronized tone can be set up.  ࠙Setting Methodࠚ How to enter the setup mode. Press and hold the    ENTER    button for 2 seconds in the ECG display screen (2).  䍾Changing ECG2 Waveform Size on display䍿        The size indication of ECG 2 is highlighted.  Pressing the  ENTER  button will sequentially change the size of ECG 2. Size ×1 Æ ×1/2 Æ ×1          When changing the size of the ECG waveform on the display, the ECG scale will also change. In the left illustration, it can display ECG waveform between   -2mV and +2mV. CAUTION The ECG waveform size setting displayed on the display of the LX-8100 does not interact with the one displayed on the screen of the receiving monitor. To change the waveform size of the receiving monitor, follow the instruction in the operation manual of the receiving monitor. 
7. Operation 34 䍾Returning to ECG display screen (2)䍿        Press the  ť  button to highlight the “Back”. Press the  ENTER  button to return to the ECG display screen (2). CAUTION Do not operate the LX-8100 with the setup screen open because it may lead to an unintended operation and cause setting changes. Make sure to press the “Back” to terminate the setup screen. The display will automatically turn itself OFF by following the setting if the “Back” is not pressed.   ࠙Setting Methodࠚ Press and hold    ENTER    button and  ť  button for 2 seconds in the ECG display screen (2).     䍾Generate a Synchronized Tone䍿 When the speaker symbol is displayed, a Synchronized Tone will generate along with the Synchronized mark. The Synchronized Tone will generate only in the ECG screen.          Press and hold the    ENTER   button and  ť  button for 2 seconds then the speaker symbol will be displayed. Press and hold the    ENTER   button and the  ť  button again for 2 seconds, then the speaker mark will disappear and the synchronized tone will stop. The setting synchronizes the ECG screen (1) so this setting is also reflected in the ECG screen (1).
7. Operation 35 䖃Respiration Display Screen In the respiration display screen, the respiration waveform size displayed on the display of the LX-8100 can be changed.   ࠙Setting Methodࠚ How to enter the setup mode: Press and hold the    ENTER    button for 2 seconds in the respiration display screen.  䍾Changing Respiration Waveform Size on display䍿          The size indication of the respiration is highlighted.  Pressing the  ENTER  button will sequentially change the size of respiration. Size ×1 Æ ×1/2 Æ ×1     When changing the size of the respiration waveform on the display, the respiration scale will also change. In the left illustration, it can display the respiration waveform until 4ȍ of change.   CAUTION The ECG waveform size setting displayed on the display of the LX-8100 does not interact with the one displayed on the screen of the receiving monitor. To change the waveform size of the receiving monitor, follow the instruction in the operation manual of the receiving monitor.
7. Operation 36 䍾Returning to the Respiration Display Screen䍿         Press the  ť  button to highlight the “Back”. Press the “Back” to return to the Respiration Display Screen.  CAUTION Do not operate the LX-8100 with the setup screen open because it may lead to an unintended operation and cause setting changes. Make sure to press the “Back” to terminate the setup screen. The display will automatically turn itself OFF by following the setting even though the “Back” is not pressed.
7. Operation 37 䕔Restarting the display The display automatically turns itself OFF after the preprogrammed duration if no operation is done. Press the    ENTER    button or press and hold the  ť  button to restart the display. The starting screen with the telemetry channel number appears, and then the waveform display screen appears. The setting for the timeout duration of the display screen can be changed. For details, refer to page 50.   CAUTION z  The heart rates and respiration rates are not measured during display OFF. Therefore, the display screen may not accurately display heart rates and respiration rates right after restarting. z  If battery level 1    is displayed, the display will not automatically turn OFF. In such case, replace the battery.     䕔Pressing the EVENT button Press and hold the    EVENT    button for 2 seconds to activate the function assigned on the receiving monitor. The following message appears on the display while the “EVENT” is transmitted. After the transmission is completed, the waveform display screen appears.  “EVENT” operation is available as a remote recording.  For details of the receiving monitor operation and settings related to the “EVENT” function, refer to the operation manual of the receiving monitor.   
7. Operation 38
 39 8. Other Setting Items The following settings are available for the LX-8100 depending on the use and condition of the patient. For details of the settings, contact our service representative.  Items Selection Default Backup Time Constant  0.4 sec, 0.1 sec  0.4 sec  Ũ Detection Sensitivity of Pacemaker Pulse   (Pace Sens.) Low, Mid, High  Mid  Ũ QRS Detection (QRS Width)  Wide, Narrow  Wide  Ũ Respiration Detection Signal (Display)  ON, OFF  ON  Ũ Display Brightness (Brightness)  8 levels  5  Ũ Turn Off Display Time (Display OFF)  1 min, 3 min, 10 min, OFF  3 min  Ũ Sound (Sound)  ON, OFF  ON  Ũ Transmitter Channel (CH)  One from following channels.  0801 to 0879 0900 to 0979 1000 to 1079 1100 to 1179 1200 to 1279 1300 to 1379  Ყ Ũ Group ID  One from 00 to 63,  00  Ũ 
8. Other Setting Items 40 䕔Changing the Time Constant The default setting of the time constant is “0.4 seconds”. If a stable monitoring is difficult with excessive change in the baseline due to excessive body motion of the patient or an interference noise, such as AC frequency, by changing the time constant to “0.1 second” the monitoring may become relatively stable.  For details of the setting, contact your local Fukuda Denshi service representative.   CAUTION z  When changing the time constant to “0.1 seconds”, the lower frequency characteristic becomes 1.6 Hz ± 25%. This setup does not meet IEC 60601-2-27 standard. It may lead to a change in the ECG waveform and the ST measurement value may be especially affected. Fukuda Denshi recommends “0.4 seconds” setting in normal use. z  The display screen in normal use does not indicate the selection of time constant. Make sure to take measures, such as marking on the   LX-8100, to distinguish whether the selection of time constant has changed.  䕔Changing the Detection Sensitivity of the Pacemaker Pulse The default setting of pacemaker pulse detection sensitivity is “Mid”. The “Mid” setting can detect and reject the following pacemaker pulse specified in IEC 60601-2-27 standard.  Detection/Rejection of Pacemaker Pulse: a) Pacemaker Pulse without Over/Undershoot: Capable to detect pulses of pulse width 0.1 ms to 2 ms, amplitude ±2 mV to ±700 mV. b) Pacemaker Pulse with Over/Undershoot:   Rejection is not possible.  Fukuda Denshi recommends the “Mid” setting in normal use.  There may be some cases when the pacemaker pulse cannot be detected depending on the pacemaker type, pulse voltage, pulse width, electrode lead type (unipolar, bipolar). In this case, change the lead or the position of the electrodes to be able to detect the pacemaker pulse. Nonetheless, if the detection is still undetectable, change the setting to “High” in order to increase the detection sensitivity. So that smaller pacemaker pulse can be detected. However, the “High” setting may lead to erroneous detection due to interference noise, such as AC frequency.
8. Other Setting Items  41 If erroneous detections occur due to interference noise, such as AC frequency, turn OFF the setting of the pacemaker pulse detection in the receiving monitor. If erroneous detections occur due to interference noise, such as AC frequency, while monitoring a patient with a pacemaker and the setting of the pacemaker pulse detection cannot be turned OFF, replace the electrodes or change the lead to remove the interference noise, such as AC frequency.  Nonetheless, if erroneous detections still occur, change the setting to “Low” in order to decrease the detection sensitivity. It makes the LX-8100 less likely to be interfered by noise, such as AC frequency. The “Low” setting decreases the detection sensitivity. Therefore, it cannot detect the pacemaker pulse specified in IEC 60601-2-27 standard.  For details of the setting, contact your local Fukuda Denshi service representative.  CAUTION The display screen in normal use does not indicate the setting status of the pacemaker pulse detection. Make sure to take measures, such as marking on the LX-8100, to distinguish whether the setting of the pacemaker pulse detection has changed.  䕔Changing QRS Detection The QRS detection mode of the LX-8100 is initially set as “Wide”. The setting can be changed to “Narrow” if it cannot detect the heart rates due to narrow QRS amplitude.  For details of the setting, contact your local Fukuda Denshi service representative.  CAUTION z  This setting is effective only for the display screen of the LX-8100 and it is not reflected in the QRS detection setting of the receiving monitor. To change the QRS detection in the receiving monitor, refer to the operation manual of the receiving monitor. z  The display screen in normal use does not indicate the setting status of the QRS detection mode such as ”Wide/Narrow”. Make sure to take measures, such as marking on the LX-8100, to distinguish whether the setting of the QRS detection mode has changed.
8. Other Setting Items 42 䕔Changing the Respiration Detection Signal ON/OFF The default setting of the respiration detection signal is “ON”. The respiration waveform can be detected when the setting of the respiration detection signal is turned “ON”.  WARNING If the LX-8100 is used with minute ventilation rate-adaptive implantable pacemaker, the respiration detection signal may cause the pacemaker to pace at its maximum programmed rate. If such event occurs, change the setting to “OFF” to prevent an occurrence of erroneous pacing rate. For details of the setting, contact your local Fukuda Denshi service representative. CAUTION z  The respiration waveform cannot be measured if the setting of the respiration detection signal is turned “OFF”. z  Make sure to turn OFF the respiration measurement function on the receiving monitor to prevent an erroneous detection of the respiration alarm (on the receiving monitor side) z  The display screen in normal use does not indicate the setting status of the respiration detection signal ON/OFF. Make sure to take measures, such as marking on the LX-8100, to distinguish whether the setting of the respiration detection signal ON/OFF has changed.  䕔Changing the Display Brightness The display brightness of the LX-8100 can be changed in 8 levels.  For details of the setting, contact your local Fukuda Denshi service representative.  䕔Changing the Display Timeout Duration The time to automatically turn OFF the display while not in operation can be selected from 1 min, 3 min, 10 min or OFF (The display will not turn off). The default setting is “3 min”.  For details of the setting, contact your local Fukuda Denshi service representative.  CAUTION If battery level 1    is displayed, the display will not automatically turn OFF. In such case, replace the battery.
8. Other Setting Items  43䕔Changing Sound ON/OFF When the sound setting is “ON”, Alarm will generate in the following situation. ȷECG Lead Off  The default setting is “ON”. Alarm will not generate with Display OFF status. For details of the setting, contact your local Fukuda Denshi service representative.
8. Other Setting Items 44
 45 9. Changing the Transmitter Channel and Group ID 䕔Changing the Transmitter Channel The LX-8100 is a transmitter of PLL synthesizer type, and its transmitter channel can be programmed. It can be set up with an arbitrary channel among the channels assigned by the Telemetry Laws (according to each country).  For details of the setting, contact your local Fukuda Denshi service representative. WARNING z  If the transmitter channel is changed, follow the instruction by the person in charge of the radio telemetry channel in your facility. Mismanagement may result in a serious accident, such as interference and mixing up patients. z  Replace promptly with new channel label if the transmitter channel has been changed.䕔Changing the Group ID The LX-8100 transmits its group ID, which it belongs to, to prevent interference with neighboring hospital's transmitter. The receiving monitor checks whether the incoming group ID is the same as the programmed one that the receiving monitor has. There are 64 group codes available. The default setting is “00”. The transmitter group ID can be changed if there is interference with a neighboring hospital's transmitter.  For details of the setting, contact your local Fukuda Denshi service representative.   CAUTION Possible causes of interference other than radio telemetry from neighboring hospital's transmitter, are the proximity of mobile phone, amateur radio station, radio taxi, and illegal citizens band, which may be a cause of interference. In such a case, the situation should be carefully observed to find the cause of interference. 
9. Changing the Transmitter Channel and Group ID 46  
 47 10. Troubleshooting 䕔List of Displayed messages Transmitter (Main Unit) Message CauseSolutionFailed to transmit waveform and value.Contact your local Fukuda Denshi service representative. Faulty EEPROM. Failed to initialize CPU. Faulty Speaker. 
10. Troubleshootimg 48 ECG Message Cause  Solution Electrode is off.Check the electrode condition.Character string displayed, such as LA, depends on the detached electrode position
10. Troubleshootimg 49䕔Troubleshooting Make sure of the following. However, if there is no improvement in the phenomenon, contact your local Fukuda Denshi service representative.  Transmitter (main unit) Phenomenon Cause  Solution Nothing is displayed on the display when the power switch is turned ON. No battery or wrong polarity.  Install the battery correctly. Battery level is empty.  Replace the battery with a new one. Restart  Battery level is empty.  Replace the battery with a new one. Nothing is displayed on the display even though LED is on when the power switch is turned ON. Battery level is empty.  Replace the battery with a new one. Nothing is displayed on the receiving screen. The channel number between the transmitter and the receiving monitor do not match up. Set the same channel number for the transmitter and the receiving monitor. Transmission problem. Same channel number is already used. Make sure to not duplicate channel numbers.   Follow the instruction by the person in charge of radio telemetry channel in your facility and use the LX-8100 with the correct channel setting. Channel interference.  Follow the instruction by the person in charge of radio telemetry channel in your facility and use the LX-8100 with the correct channel setting. Transmitter failure.  Contact your local Fukuda Denshi service representative. 
10. Troubleshootimg 50 ECG Phenomenon Cause  Solution  “ELECTRODE?” message is displayed. Lead cable is off.  Check the connection between the lead cable and the LX-8100. Check the connection between the lead cable and the electrode. Faulty Lead cable.  Replace the ECG cable with a new one. Electrode is peeling off.  Replace the electrode with a new one.  Polarization potential of the electrode is too high. ECG waveform contains noise.  Electrode gel is dry  Replace the electrode with a new one. Electrode is peeling off. Electric blanket is used.  Cover the electric blanket with a shield cover. AC filter setting of the receiving monitor is OFF. Set the AC filter up as ON. Respiration waveform cannot be measured.  Electrode gel is dry  Replace the electrode with a new one. Electrode is peeling off. The positions of the electrodes are improper. Attach the electrodes where the respiration waveform can be measured appropriately.
10. Troubleshootimg 51 䕔In Case of Dropping the LX-8100 into Water In case of dropping the LX-8100 into water containing disinfectant, pick up the LX-8100 quickly from it. Rinse it well with running water, and dry it thoroughly with a soft cloth  CAUTION z  Do not use a dryer. The LX-8100 shape may change or be broken. z  When the LX-8100 is rinsed with running water, make sure to close the battery compartment lid. 
10. Troubleshootimg 52  
 53 11. Cleaning and Disinfection The cleaning and disinfection of the LX-8100 and ECG lead cable shall be performed as follows.  CAUTION Do not sterilize the LX-8100 and ECG lead cable in any manners, such as radioactive rays, steam, or ethylene oxide.  䕔Cleaning Clean the LX-8100 using a squeezed gauze or an absorbent cotton cloth dampened with alcohol or a neutral cleanser.  CAUTION z  Clean the equipment frequently so stains can be removed easily. z  To prevent injury, it is recommended to wear gloves when cleaning the equipment. z  Do not allow any chemical solution to enter the inside of LX-8100 or connectors z  The LX-8100 cannot be sterilized. z  Do not use organic solvents, thinner, toluene and benzene to avoid damaging the resin case. z  Do not polish the housing with abrasive or chemical cleaner. z  Use only neutral detergent to clean the housing. Do not use chemical cloth, scrub brush, abrasive, polishing powder, hot water, volatile solvent and chemicals (cleanser, thinner, benzine, benzol, and synthetic detergent for house and furniture), or sharp-edged tools. The surface resin coating may be damaged, resulting in discoloration, scratches, and other problems. 䕔Disinfection If there is a possibility of being infected, clean the LX-8100 using a squeezed gauze or an absorbent cotton cloth dampened with alcohol or invert soap.  CAUTION z  Do not immerse the connector parts of the LX-8100 in any chemical solution to prevent connection failure. z  When disinfecting the entire room using a spray solution, pay close attention not to have liquids get into the LX-8100 or connectors.
11. Cleaning and Disinfection 54 䕔Cleaning the ECG lead cable After using the cable, clean it with neutral detergent or 70% isopropyl alcohol.  CAUTION z  Do not use thinner, toluene, or other organic solvents to clean the cables. z  Do not pull the cable and do not hold the connector part when cleaning.  (It may degrade the cable coating and result in damage. Particularly organic solvents and antiseptic solution such as cresol soap solution will degrade the cable coating.) z  After cleaning, dry it completely before usage. z  Do not use high temperature sterilization such as steam or EOG method.
 55 12. Maintenance and Inspection This section explains the daily checks and periodic checks of the LX-8100. To ensure safety, reliability, and high performance, a “Daily Check” and “Periodic Check” must be performed. We are not liable for any accident arising from lack of maintenance. A full technical description of the LX-8100 is available from your local Fukuda Denshi representative.   CAUTION z  Do not open the housing z  Do not allow excessive moisture or cleaning agents into the connectors or inside the equipment. ڦDaily CheckPerform daily checks using the “Daily Check List” on the next page.  ڦPeriodic CheckPeriodic check of medical electronic equipment is mandatory to prevent failures and accidents, and to ensure safety and reliability. Periodic maintenance may be performed by the medical institution or by a third party by concluding a “Maintenance Contract”.   For more details, contact your local Fukuda Denshi service representative.  Perform Periodic check using the “Periodic Check List”. The periodic check should be performed once a year. If there is an item with “Fail” judgement, the overall judgement will be “Fail”. Make sure to take countermeasures for the “Fail” item. Use the equipment only if the judgements of all the items are “Pass”.  ڦPeriodic Replacement PartsThe “Battery Compartment Lid (Waterproof)” is the only periodic replacement part. To ensure the reliability of waterproof (IPX8) performance of the LX-8100, replace it once a year. Contact your local Fukuda Denshi service representative for replacement. The reliability of water resistance (IPX8) performance will not be ensured without yearly replacement.    CAUTION The periodic replacement parts must be replaced at specified period.
 56 Daily Check List No.                            Inspected Date   Inspected by   Location  Device Type  LX-8100   S/No.   Date of Purchase   Items Details  Criteria Judgment Appearance  Visually check for any damage, cracks, chip, peeled label, and loosen screw on the housing. No abnormality should be found.  ڧOK/  ڧNG Battery Compartment Visually check for the ring condition of the battery compartment lid. No damage, kink, floating, and adhesion of dust should be found. ڧOK/  ڧNG Visually check for the contact springs, inside the LX-8100, to the battery and the lock lever of the battery compartment lid.No deformation, cracks, and rust should be found.ڧOK/  ڧNGPower Supply  Turn the power ON/OFF to verify proper switch operation. With batteries installed, the display should turn ON. ڧOK/  ڧNGECG Connectors  Visually check the connectors of the cable and the LX-8100. No damage, chip, and adhesion of dust should be found. ڧOK/  ڧNGECG Lead cable  Visually check each lead for damages.  No crack and damage should be found. ڧOK/  ڧNGWireless ChannelVerify whether the transmitting channel and group ID are the same with the receiving monitor.Must match the wireless channel check list. ڧOK/  ڧNGTransmission Function Turn the power ON and make sure the information is displayed on the receiving monitor. Waveforms and values should be received without any problem. ڧOK/  ڧNGDisplay Function  Turn the power ON and verify each display condition. All data should be properly displayed. ڧOK/  ڧNGPeriodic Check  Check the date of the previous periodic check.  Should be within one year. ڧOK/  ڧNGComment
12. Maintenance and Inspection 57ڦPeriodic CheckThe periodic maintenance check is intended to check the medical equipment used daily in a medical institution to prevent failures and accidents and to ensure safety and reliability. The check procedures are described for daily and periodic checks. Each check item must be performed according to the described check procedure. The consignee can select the check items according to the product quality, frequency of usage, and maintenance check period. However, electrical safety items must also be performed. For details of the electrical safety check procedure, refer to IEC 60601-1.  䖃Periodic Check Items The periodic check items are as follows.  No. Check Item 1 External Appearance 2 Power Supply Switch 3  Display / Operation / Record 4 ECG 5 Respiration 6 Speaker 7 Electrical Safety   No.  Check Item  Check Procedure  Criteria 1. External Appearance, Accessories 01  Appearance  Visually check the exterior for scratches, cracks, deformation, andrust. No abnormality should be found. 02  Cables  Visually check all cables for any damage or being disconnected. No damage should be found. 03 Operation Manual   Check if the operation manual and other accompanying documents are stored in the specified places. Should be stored in the specified place.  No  Check Item  Check Procedure  Criteria 2Power Supply Switch 01 Power Supply Switch  Turn ON/OFF the power switch.  Should turn ON/OFF the power switch properly.
12. Maintenance and Inspection 58 No  Check Item  Check Procedure  Criteria 3Display, Operation, Record 01  Labels  Visually check the labels, caution labels, etc.  Should be clean, clear and firmly attached. 02 Operation, Switches and keys Check by operating the switches and keys.  Should operate properly. 03  Display  Check that the characters and waveforms appear on the display. The characters and waveforms should be clearly displayed. The brightness should be sufficient.  No  Check Item  Check Procedure  Criteria 4ECG 01 Input Impedance*  According to test procedure of IEC 60601-2-27:2011 201.12.1.101.3 Should be 2.5M or above. 02 Suppression Characteristic ofCommon-Mode Signal* According to test procedure of IEC 60601-2-27:2011 201.12.1.101.10 Should be 10mmp-p or below for standard sensitivity (sensitivity 1). 03 Transient Characteristic*  With comprehensive tester, apply standard voltage of 1mV, and check the time the amplitude natural logarithmically drops and becomes 37% of the waveform of 0.04 sec. after the application of standard voltage. Should be 0.4 ± 0.1 seconds. (When the time constant is set to 0.4seconds.) 04 Frequency Characteristic*  With comprehensive tester, apply sinusoidal voltage. Measure the frequency characteristic at test voltage of 40Hz. According to test procedure of IEC 60601-2-27:2011 201.12.1.101.8 Should be 40Hz (-3dB) or above. 05 Heart Rhythm Detection*  With comprehensive tester, input both positive and negative polarity of 0.3mV and 3mV with sensitivity 1. According to test procedure of IEC 60601-2-27:2011 201.12.101.15 The heart beat rhythm should be detected with sensitivity 1 according to the peak-to-peak signal of 0.3mV and 3mV.
12. Maintenance and Inspection 5906  ECG Sensitivity  With comprehensive tester, apply 1mV voltage and measure the displayed amplitude. According to test procedure of IEC 60601-2-27:2011 201.12.1.101.1 Wave form size on the receiving monitor should be within 0mmd10% at sensitivity1. 07 Heart Rate Accuracy*  With ECG simulator, test heartbeat 60, 180 beats/min., and check the displayed HR value. According to test procedure of IEC 60601-2-27:2011 201.12.101.15 For reference heartbeat signal of 60, 180 beats/min., error of the displayed HR value should be within ±3 beats/min. 08 ECG Lead Switch  With ECG simulator, check that each lead is displayed properly. (Check for 3-electrode, and 4-electrode, 5-electrode.) For each lead cable, lead should be correctly switched, and waveform should be correctly displayed. 09 Lead-Off Indication  Remove each electrode, and check that lead-off message is displayed. Lead-Off message for the corresponded lead should be displayed. 10  ECG Lead CableRecognition  Switch the ECG lead cable or switch the setup of lead cable. Should correctly recognize the connected lead cable. * As these functions are dependent on the design or software, these items are not mandatory for periodic checks. Perform the test as necessary.
12. Maintenance and Inspection 60  No  Check Item  Check Procedure  Criteria 5Respiration 01 Respiration Waveform Sensitivity With comprehensive tester or reference respiration signal generator, input sinusoidal waveform of 0.5Hz with base resistance of 1.5kȍ/1ȍ change. The amplitude displayed on the receiving monitor screen should be within 10 mm ± 2 mm. 02 Respiration RateAccuracy*  Input reference respiration signal to comprehensive tester or respiration simulator, and check the respiration rate display. Test with the respiration reference load signal of 60, 120/min. Error should be within ±5 Bpm. 03 Frequency Characteristic*  With comprehensive tester or reference respiration signal generator, input 0.5Hz and 1.5Hz or 2.5Hz, and measure the frequency characteristic. The crest value at   1.5 Hz should be more than 70% of the crest value at 0.5 Hz input. * As these functions are dependent on the design or software, these items are not mandatory for periodic checks. Perform the test as necessary.  No  Check Item  Check Procedure  Criteria 6Speaker  01  Generation  Generate synchronized tone and check the sound.  Generating synchronized tone.  No  Check Item  Check Procedure  Criteria 7Electrical Safety 01 Contact current  Measure the leakage current that runs through the ground from the enclosure of the device under normal condition using a leak measurement safety tester. According to test procedure of IEC 60601-1 8.7.4 From the enclosure to the ground (NC) ġ0.1mA. 02 Patient leakage current that runs through the ground from the patient connecting part (NC) Measure the patient leakage current that runs through the ground from the patient connecting part using a leak measurement safety tester. According to test procedure of IEC 60601-1 8.7.4 ƂAC/DCƃ From the patient connecting part to the ground (NC) ġ0.01mA.
12. Maintenance and Inspection 61No  Check Item  Check Procedure  Criteria 03 Leakage current when external voltage is applied to the patient connection in the Type F attaching part (SFC) Measure the leakage current when external voltage is applied to the patient connection in the Type F attaching part using a leak measurement safety tester. According to test procedure of IEC 60601-1 8.7.4 Leakage current when external voltage is applied to the patient connection in the Type F attaching part. (SFC) ġ0.05mA. 04 Total patient leakage current that runs through the ground from the total patient connecting part.(NC) Measure the total patient leakage current that runs through the ground from the patient. According to test procedure of IEC 60601-1 8.7.4 ƂAC/DCƃ From the patient connecting part to the ground (NC) ġ0.05mA. 05 Leakage current when external voltage is applied to the patient connection in the Type F attaching part (SFC) Measure the leakage current when external voltage is applied to the patient connection in the Type F attaching part using a leak measurement safety tester According to test procedure of IEC 60601-1 8.7.4 Leakage current when external voltage is applied to the patient connection in the Type F attaching part. (SFC) ġ0.01mA. 06 Patient auxiliary current (NC)  Measure the patient auxiliary current (NC) using a leak measurement safety tester According to test procedure of IEC 60601-1 8.7.4 Patient auxiliary current (NC) ġ0.01mA.  No  Check Item  Check Procedure  Criteria 7  Electrical Safety (*) Perform the following check item as appropriate. Check these items when you have disassembled the equipment to check/ replace the boards or units. 07 Withstand Voltage Test (the enclosure – isolated connecting part)Apply AC1500V for 1 minute between the enclosure and a connecting part. Note: The voltage differs depending on the internal protective circuit composition of the equipment. According to test procedure of IEC 60601-1 8.8.3 Should withstand applied voltage.
12. Maintenance and Inspection 62 Periodic Check List 㻯㼔㼑㼏㼗㻌㻰㼍㼠㼑㻺㼛㻚 㻶㼡㼐㼓㼑 㻯㼔㼑㼏㼗 㻺㼛㻚 㻶㼡㼐㼓㼑 㻯㼔㼑㼏㼗 㻺㼛㻚 㻶㼡㼐㼓㼑 㻯㼔㼑㼏㼗㻝 㻡㻜㻝 㻻㻷㻌㻌㻺㻳㻣 㻱㼘㼑㼏㼠㼞㼕㼏㼍㼘㻌㻿㼍㼒㼑㼠㼥㻜㻞 㻻㻷㻌㻌㻺㻳㻜㻟 㻻㻷㻌㻌㻺㻳 㻜㻞 㻻㻷㻌㻌㻺㻳㻜㻟 㻲㼞㼑㼝㼡㼑㼚㼏㼥㻌㻯㼔㼍㼞㼍㼏㼠㼑㼞㼕㼟㼠㼕㼏 㻻㻷㻌㻌㻺㻳㻞 㻢 㻿㼜㼑㼍㼗㼑㼞㻜㻝 㻻㻷㻌㻌㻺㻳 㻜㻝 㻻㻷㻌㻌㻺㻳㻟㻜㻝 㻻㻷㻌㻌㻺㻳㻜㻞 㻻㻷㻌㻌㻺㻳㻜㻟 㻻㻷㻌㻌㻺㻳㻠㻜㻝 㻻㻷㻌㻌㻺㻳㻜㻟 㻻㻷㻌㻌㻺㻳㻜㻠 㻻㻷㻌㻌㻺㻳㻜㻡 㻻㻷㻌㻌㻺㻳 㻜㻣 㼃㼕㼠㼔㼟㼠㼍㼚㼐㻌㼂㼛㼘㼠㼍㼓㼑㻌㼀㼑㼟㼠㻌 㻻㻷㻌㻌㻺㻳㻜㻢 㻻㻷㻌㻌㻺㻳㻜㻣 㻻㻷㻌㻌㻺㻳㻜㻤 㻻㻷㻌㻌㻺㻳㻜㻥 㻻㻷㻌㻌㻺㻳䝺䠝㽢䠟䠋䠮㼀㼔㼑㻌㼏㼔㼑㼏㼗㻌㼞㼑㼟㼡㼘㼠㻌㼕㼟㻌㼍㼟㻌㼒㼛㼘㼘㼛㼣㼟㻦 㻯㼛㼙㼜㼍㼚㼥䕕㻺㼛㼞㼙㼍㼘㻌㻻㼜㼑㼞㼍㼠㼕㼛㼚䚷䕕㻹㼍㼘㼒㼡㼚㼏㼠㼕㼛㼚㼕㼚㼓㻚䚷䕕㻺㼑㼑㼐㼟㻌㼠㼛㻌㼎㼑㻌㼞㼑㼜㼍㼕㼞㼑㼐㻚㻔㻰㼑㼠㼍㼕㼘㼟㻌㼛㼒㻌㼙㼍㼘㼒㼡㼚㼏㼠㼕㼛㼚㻌㼍㼚㼐㻌㼞㼑㼜㼍㼕㼞䛋㻵㼚㼟㼜㼑㼏㼠㼛㼞㻼㼑㼞㼟㼛㼚㻌㼕㼚㻌㼏㼔㼍㼞㼓㼑㻱㻯㻳㻜㻝㻾㼑㼜㼘㼍㼏㼑㼙㼑㼚㼠㻌㼜㼍㼞㼠㻿㼥㼚㼏㼔㼞㼛㼚㼕㼦㼑㼐㻌㼀㼛㼚㼑㻾㼑㼝㼡㼑㼟㼠㼑㼐㻌㻵㼠㼑㼙㻯㼔㼑㼏㼗㻌㻵㼠㼑㼙㻾㻾㻌㻭㼏㼏㼡㼞㼍㼏㼥㻻㼜㼑㼞㼍㼠㼕㼛㼚㻌㻹㼍㼚㼡㼍㼘㼟㻱㼤㼠㼑㼞㼕㼛㼞㻯㼍㼎㼑㼘㼟㻱㼤㼠㼑㼞㼕㼛㼞㻘㻌㻭㼏㼏㼑㼟㼟㼛㼞㼕㼑㼟㻰㼑㼟㼏㼞㼕㼜㼠㼕㼛㼚㻯㼔㼑㼏㼗㻌㻵㼠㼑㼙㻭㼐㼖㼡㼟㼠㼙㼑㼚㼠㻭㼏㼏㼑㼜㼠㼍㼚㼏㼑㻌㻰㼍㼠㼑㻯㼔㼑㼏㼗㻌㻵㼠㼑㼙㻾㼑㼟㼜㼕㼞㼍㼠㼕㼛㼚㻺㼑㼤㼠㻌㻯㼔㼑㼏㼗㻌㻰㼍㼠㼑㻼㼑㼞㼕㼛㼐㼕㼏㻌㻯㼔㼑㼏㼗㻌㻯㼛㼚㼠㼞㼍㼏㼠䕕㼅㼑㼟䚷䚷䕕㻺㼛㻹㼛㼚㼠㼔㻌㻯㼔㼑㼏㼗㻰㼑㼘㼕㼢㼑㼞㼥㻌㻰㼍㼠㼑㻯㼡㼟㼠㼛㼙㼑㼞㻌㻯㼛㼐㼑㼀㼑㼘㼑㼙㼑㼠㼞㼥㻌㼀㼞㼍㼚㼟㼙㼕㼠㼠㼑㼞㻌㻼㼑㼞㼕㼛㼐㼕㼏㻌㻯㼔㼑㼏㼗㻌㻾㼑㼜㼛㼞㼠㻸㼛㼏㼍㼠㼕㼛㼚㻹㼛㼐㼑㼘㻌㻺㼍㼙㼑 㻿㼑㼞㼕㼍㼘㻌㻺㼛㻚㻦 㻼㼞㼛㼐㼡㼏㼠㻌㻯㼛㼐㼑㻰㼕㼟㼜㼘㼍㼥㻘㻌㻻㼜㼑㼞㼍㼠㼕㼛㼚㻘㻌㻾㼑㼏㼛㼞㼐䕕㻴㼛㼘㼕㼐㼍㼥䚷䚷㻌㻌䕕㻺㼕㼓㼔㼠㻾㼑㼜㼘㼍㼏㼑㼙㼑㼚㼠㻌㼜㼍㼞㼠㼟 㻯㼘㼍㼟㼟㼕㼒㼕㼏㼍㼠㼕㼛㼚䕕㻻㼚㻙㼟㼕㼠㼑䚷䚷㻌㻌䕕㼀㼍㼗㼕㼚㼓㻙㼛㼢㼑㼞㻯㼔㼑㼏㼗㻻㼜㼑㼞㼍㼠㼕㼛㼚㻌㻿㼣㼕㼠㼏㼔㻛㻷㼑㼥㻼㼛㼣㼑㼞㻌㻿㼡㼜㼜㼘㼥㻌㻿㼣㼕㼠㼏㼔㻺㼛㼠㻌㼏㼛㼢㼑㼞㼑㼐㻴㼑㼍㼞㼠㻌㻾㼍㼠㼑㻌㻭㼏㼏㼡㼞㼍㼏㼥㼀㼞㼍㼚㼟㼕㼑㼚㼠㻌㻯㼔㼍㼞㼍㼏㼠㼑㼞㼕㼟㼠㼕㼏㻱㻯㻳㻌㻸㼑㼍㼐㻌㻿㼣㼕㼠㼏㼔㻵㼚㼜㼡㼠㻌㻵㼙㼜㼑㼐㼍㼚㼏㼑㻸㼑㼍㼐㻙㻻㼒㼒㻌㼕㼚㼐㼕㼏㼍㼠㼕㼛㼚㻜㻝㻸㼍㼎㼑㼘㼟㻸㻯㻰㻼㼛㼣㼑㼞㻌㻿㼡㼜㼜㼘㼥㻌㻿㼣㼕㼠㼏㼔㻜㻞㻿㼡㼜㼜㼞㼑㼟㼟㼕㼛㼚㻯㼔㼍㼞㼍㼏㼠㼑㼞㼕㼟㼠㼕㼏㻌㼛㼒㻯㼛㼙㼙㼛㼚㻙㻹㼛㼐㼑㻌㻿㼕㼓㼚㼍㼘㻻㻷㻌㻌㻺㻳㻱㻯㻳㻌㻿㼑㼚㼟㼕㼠㼕㼢㼕㼠㼥㻝㻜 㻱㻯㻳㻌㻸㼑㼍㼐㻌㻯㼍㼎㼘㼑㻾㼑㼏㼛㼓㼚㼕㼠㼕㼛㼚 㻻㻷㻌㻌㻺㻳㻲㼞㼑㼝㼡㼑㼚㼏㼥㻌㻯㼔㼍㼞㼍㼏㼠㼑㼞㼕㼟㼠㼕㼏㻴㼑㼍㼞㼠㻌㻾㼔㼥㼠㼔㼙㻌㻰㼑㼠㼑㼏㼠㼕㼛㼚㻜㻢 㻼㼍㼠㼕㼑㼚㼠㻌㼍㼡㼤㼕㼘㼕㼍㼞㼥㻌㼏㼡㼞㼞㼑㼚㼠㻺㻯㻌㻔㻌㻌㻌㻌㻌㻌㻕㼙㻭㻾㼑㼟㼜㼕㼞㼍㼠㼕㼛㼚㻌㼃㼍㼢㼑㼒㼛㼞㼙㻿㼑㼚㼟㼕㼠㼕㼢㼕㼠㼥 㻻㻷㻌㻌㻺㻳 㻲㼞㼛㼙㻌㼠㼔㼑㻌㼜㼍㼠㼕㼑㼚㼠㻌㼏㼛㼚㼚㼑㼏㼠㼕㼚㼓㼜㼍㼞㼠㻌㼠㼛㻌㼠㼔㼑㻌㼓㼞㼛㼡㼚㼐㻺㻯䠄䚷䚷䚷䚷䠅㼙㻭㻜㻞㻼㼍㼠㼕㼑㼚㼠㻌㼘㼑㼍㼗㼍㼓㼑㻌㼏㼡㼞㼞㼑㼚㼠㻌㼠㼔㼍㼠㼞㼡㼚㼟㻌㼠㼔㼞㼛㼡㼓㼔㻌㼠㼔㼑㻌㼓㼞㼛㼡㼚㼐㻌㼒㼞㼛㼙㼠㼔㼑㻌㼜㼍㼠㼕㼑㼚㼠㻌㼏㼛㼚㼚㼑㼏㼠㼕㼚㼓㻌㼜㼍㼞㼠㻺㻯㻌㻔㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻕㼙㻭㻜㻟㻻㻷㻌㻌㻺㻳㻜㻠㼀㼛㼠㼍㼘㻌㼜㼍㼠㼕㼑㼚㼠㻌㼘㼑㼍㼗㼍㼓㼑㻌㼏㼡㼞㼞㼑㼚㼠㼠㼔㼍㼠㻌㼞㼡㼚㼟㻌㼠㼔㼞㼛㼡㼓㼔㻌㼠㼔㼑㻌㼓㼞㼛㼡㼚㼐㼒㼞㼛㼙㻌㼠㼔㼑㻌㼠㼛㼠㼍㼘㻌㼜㼍㼠㼕㼑㼚㼠㼏㼛㼚㼚㼑㼏㼠㼕㼚㼓㻌㼜㼍㼞㼠㻚㻺㻯㻌㻔㻌㻌㻌㻌㻌㻌㻌㻌㻕㼙㻭㻻㻷㻌㻌㻺㻳㻜㻡㻸㼑㼍㼗㼍㼓㼑㻌㼏㼡㼞㼞㼑㼚㼠㻌㼣㼔㼑㼚㻌㼑㼤㼠㼑㼞㼚㼍㼘㼢㼛㼘㼠㼍㼓㼑㻌㼕㼟㻌㼍㼜㼜㼘㼕㼑㼐㻌㼠㼛㻌㼠㼔㼑㼜㼍㼠㼕㼑㼚㼠㻌㼏㼛㼚㼚㼑㼏㼠㼕㼛㼚㻌㼕㼚㻌㼠㼔㼑㻌㼀㼥㼜㼑㻲㻌㼍㼠㼠㼍㼏㼔㼕㼚㼓㻌㼜㼍㼞㼠㻚㻿㻲㻯㻌㻌㻔㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻕㼙㻭㻻㻷㻌㻌㻺㻳㻸㼑㼍㼗㼍㼓㼑㻌㼏㼡㼞㼞㼑㼚㼠㻌㼣㼔㼑㼚㻌㼑㼤㼠㼑㼞㼚㼍㼘㼢㼛㼘㼠㼍㼓㼑㻌㼕㼟㻌㼍㼜㼜㼘㼕㼑㼐㻌㼠㼛㻌㼠㼔㼑㼜㼍㼠㼕㼑㼚㼠㻌㼏㼛㼚㼚㼑㼏㼠㼕㼛㼚㻌㼕㼚㻌㼠㼔㼑㻌㼀㼥㼜㼑㻲㻌㼍㼠㼠㼍㼏㼔㼕㼚㼓㻌㼜㼍㼞㼠㻚㻿㻲㻯㻌㻔㻌㻌㻌㻌㻌㻌㻌㻌㻌㻌㻕㼙㻭㻻㻷㻌㻌㻺㻳㻯㼘㼑㼍㼚㼕㼚㼓㻾㼑㼜㼍㼕㼞㻻㻷㻌㻌㻺㻳㻻㻷㻌㻌㻺㻳
12. Maintenance and Inspection 63䕔Repairing the Equipment This equipment is basically repaired at Fukuda Denshi factory. If detailed information about the repair is needed, contact Fukuda Denshi.  On-site repair is possible for the following parts.   ȷReplacing the battery compartment Lid Unit  CAUTION Make sure to replace the parts correctly. Otherwise, it may cause damage and heat generation of the equipment.  䕔Replacing the Battery Compartment Lid Unit 䖃Life of the Battery Compartment Lid Unit Life of the waterproof battery compartment lid unit is one year. If this unit is used for more than a year, the waterproof (IPX8) performance cannot be guaranteed. Replace the battery compartment lid unit to maintain its waterproof performance.                CAUTION z  The battery unit must be replaced at specified period. z  Even if the LX-8100 is used less than one (1) year, the unit may be damaged from high impact. If the LX-8100 is dropped or is subjected to a high impact, make sure that the unit is not damaged.       Battery Compartment Lid Unit
12. Maintenance and Inspection 64 䖃Tools ȷPhillips screwdriver (#0)  䖃Assembly and Disassembly Follow the procedure below to remove the battery compartment lid unit.  1.㻌Remove the label.             2.㻌Remove the 2 installation screws then remove the battery compartment lid unit.                            Label Installation screw
12. Maintenance and Inspection 65Follow the procedure below to attach the battery compartment lid unit.  1.㻌Set the battery compartment lid unit adjusting to the front case.                 2.㻌Secure the battery compartment lid unit with new 2 installation screws. Make sure that the screws are securely tightened.                            Installation screw
12. Maintenance and Inspection 66 3.㻌Make sure that the battery compartment lid unit opens/closes smoothly.              4.㻌Attach the label firmly aligning with the groove of the battery compartment lid unit.            (1) (2) Groove Label
 67 13 . Standard and Optional Accessories  WARNING Use only the accessories specified by Fukuda Denshi for the LX-8100. Otherwise, the LX-8100 cannot deliver its maximum performance and may be damaged, resulting in a safety hazard.  CAUTION For quality improvement, specifications are subject to change without prior notice.  䕔Standard Accessories No. Item  Model Type Q’ty Remarks 1 Neck Strap  OAT-03A  1   2  4-electrode ECG lead cable  CMT-02HTH-0.8DA  1  AHA color code,   Hook Type,     Limb Lead (2CH) 3  Disposable portable case  ABT-810D 1   4 Operation Manual    1   (1)  (2)   (3) (4)    㻌This Document
13. Standard and Optional Accessories 68 䕔Optional Accessories The following optional accessories are available for the LX-8100.   Purchase them as required.   䖃LX-8100  Lead cables  Item Model Type RemarksECG Clip Type Lead Cable  CMT-01CTH-0.8DA 3-electrode (White, Black, Red) ECG Clip Type Lead Cable  CMT-02CTH-0.8DA 4-electrode (White, Black, Green, Red) ECG Clip Type Lead Cable  CMT-03CTH-0.8DA 5-electorode (Chest) (White, Black, Green, Red, Brown) ECG Hook Type Lead Cable  CMT-01HTH-0.8DA 3-electrode (White, Black, Red) ECG Hook Type Lead Cable  CMT-02HTH-0.8DA 4-electrode (White, Black, Green, Red) ECG Hook Type Lead Cable  CMT-03HTH-0.8DA 5-electorode (Chest) (White, Black, Green, Red, Brown)  Pochette  Item Model Type RemarksDisposable Portable Case  ABT-810D  5 pieces /pack
14. Specification   6914. Specification 䕔Specification  CAUTION For quality improvement, specifications are subject to change without prior notice. Standard Specification Size:  72.0(W) mm x 67.0(H) mm x 23.4(D) mm   (not including the protrusion) Weight: Approximately 100 grams (with battery) Transmitting Waveform:  ECG 1CH or 2CH (selectable from the ECG lead cables), Respiration waveform. ECG Lead cable Type:  3-electrode, 4-electrode, or 5-electrode (Limb+Chest) lead cable. Automatically detect the type by inserting the lead cable. Transmitting Status Data:   Electrode Off, Low Battery, Event Switch,   Pacemaker OLED: Built-in Waterproof:  IPX8 (If periodic replacements are performed) / IPX5 IPX5: Protection from water. IPX8: Protection from submerge Power Supply:  DC: One “AA” size (“LR6” size) alkaline battery. Continuous Operating Time:  “AA” size (“LR6” size) alkaline battery Approximately6 days with MX1500 (DURACELL) Conditions: ECG measurement, RESP measurement, Set as Initial setting, operating temperature 23qC Operation mode:  Continuous operation *Continuous operating time is assumed when using a new ”AA” size (“LR6”size) Alkaline battery specified by Fukuda Denshi.  
14. Specification 70 ECG Particular Standard:  IEC 60601-2-27:2011 (Medical electrical equipment - Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment) *Suitable when using DS-8900 System (Central Monitor) and LW-7080(Central Telemetry Receiver). Numbers of Lead Electrode:  3-electrode, 4-electrode, or 5-electrode (Limb+Chest) lead cable Numbers of Input Channel:1CH (3-electrode) or 2CH Accuracy of Sensitivity:  Complies with IEC 60601-2-27:2011 and 201.12.1.101.1 (±20% or 100uV, whichever is greater.) ECG Input Impedance:  Complies with IEC 60601-2-27:2011 and   201.12.1.101.3  (2.5Mȍ and above) Input dynamic range and Offset voltage:  Complies with IEC 60601-2-27:2011 and 201.12.1.101.2 (Input dynamic range: ±5 mV Offset voltage: ±300 mV Change of amplitude caused by offset voltage: Within ±10%) Common Mode   Rejection Ratio:  Complies with IEC 60601-2-27:2011 and 201.12.1.101.10 (Less than 1mVp-p (RTI)) HR measurment   detection:  Complies with IEC 60601-2-27:2011 and 201.12.1.101.15 (HR Measurement Accuracy: Less than ±10% or ± 5bpm, whichever is greater HR measurement range and accuracy are as follows.  HR measurement range: 0, 12 bpm to 300 bpm) HR Display Range:  QRS Detection Wide0, 12 bpm to 300 bpm (1bpm increment) QRS Detection Narrow0, 30 bpm to 300 bpm   (1 bpm increment) Frequency  Characteristic:  0.5 Hz to 40 Hz (within -3dB) Time Constant:  0.4 sec ± 25% Switching and available to set 0.1 sec ± 25% Rejection of Pacemaker Pulse:  a) Pacemaker Pulse without Over/Undershoot       Capable to reject pulses of pulse width 0.1 ms to 2 ms, amplitude ±2 mV to ±700 mV b) Pacemaker Pulse with Over/Undershoot    Rejection is not possible. Protection to Defibrillation: Meets the requirement of IEC 60601-2-27
14. Specification   71Lead-off Detection Current:  100 nA and below Tall T-wave Rejection Capability:  1.2 mV T-wave can be removed when tested according to IEC 60601-2-27 Average of Heart Rate:  HR measured from 6 seconds of heartbeat for setting QRS width: wide, and 4 seconds of heartbeat for setting QRS width: narrow.   Response time of heart rate meter to change in heart rate: HR change from 80 bpm to 120 bpm:     Range 6 sec. to 8 sec., Average 7 sec. HR change from 80 bpm to 40 bpm:     Range 6 sec. to 8 sec., Average 7 sec. Sweep speed  9.0 mm/s   Heart rate meter accuracy and response to irregular   rhythm: 80 bpm Ventricular Bigeminy: 80 bpm  60 bpm Ventricular Bigeminy: 60 bpm  120 bpm Ventricular Bigeminy: 120 bpm  90 bpm Bidirectional Systoles: 90 bpm   
14. Specification 72 Respiration (Impedance Method) Accuracy of Sensitivity: 10 mm/1ȍ ± 2 mm   (When standard Impedance is 480ȍ.) Resp. Display Range:  0, 4 Bpm to 150 Bpm Display Error of Respiration Rate:  ±3 Bpm Measured Current of Respiration:  Below 100PA (42kHz)    Transmission Method Modulation Mode:  Digital, Frequency shift keying Frequency:  608 MHz to 614 MHz Oscillation Method:  PLL Synthesizer method by crystal control Channel Spacing:  12.5 kHz Occupied Frequency Bandwidth:  Within 8.5 kHz RF Output Power:  1 mW ± 2 dB Transmitting Antenna:  Dielectric Antenna  Safety General Standard: AAMI / ANSI ES 60601-1:2005(R)2012 and A1:2012, C1:2009(r)2012 and A2:2010(r)2012 (Medical electrical equipment - Part 1: General requirements for basic safety and essential performance) EMC Standard:      IEC 60601-1-2: 2014 (Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic disturbances – Requirements and tests) The class of protection against electric shock:  Internally Powered Equipment The type of protection   against electric shock:  Type CF Applied Part Operating Environment Temperature:  10qC to 40qC Humidity:  30% RH to 85% RH (non-condensing) Atmospheric Pressure:  70 kPa to 106 kPa Transport / Storage Environment Temperature:  -10qC to 60qC Humidity:  10% RH to 95% RH (non-condensing) Atmospheric Pressure:  70 kPa to 106 kPa
14. Specification   73䕔Displays The following displays are shown on the display of the LX-8100. Display   Description Starting Screen      Displayed after turning on the power.   Automatically moves onto the channel display screen. Channel Display Screen      Displays the transmitter channel after turning on the power and also when refreshing the screen.   Automatically move s onto the waveform display screen. Event Displayed when the  EVENT  button is pressed.  Electrode?      Displayed when the ECG electrode is disconnected or the ECG/RESP waveform cannot be measured normally. For details about electrode check message, refer to “14. Specification   ŠDetails of the “Electrode” Message”.
14. Specification 74   Display  Description Error Message: If messages are not cleared, contact your local Fukuda Denshi service representative.  Telemeter Error      Displayed when the transmitter is faultyEEPROM Error      Displayed when the EEPROM is faulty CPU Error      Displayed when the CPU is faulty  Speaker Error Displayed when the speaker is faulty. 
14. Specification   75䕔Details of the “Electrode” Message The following “Electrode?” messages are displayed on the display depending on the selected lead cable and lead. Check Position 3-electrode lead cable Lead I display 3-electrode lead cable Lead II display 3-electrode lead cable Lead III display LL     RA     LA     Several Position Simultaneously   Check Position 4-electrode lead cable 5-electrode (Chest) lead cable LL    RA    LA    RL    V    ECG1+    ECG1-    ECG2+    ECG2-    Several Position Simultaneously
14. Specification 76 䕔List of Setup Items Followings are the items that can be set up in the LX-8100.  Items Selection Default Backup ECG Lead  I, II, III  II  Ũ Display Size of ECG (1) ×1, ×1/2  ×1  Ũ Display Size of ECG (2) ×1, ×1/2  ×1  Ũ Display Size of Respiration Waveform  ×1, ×1/2  ×1  Ũ Synchronized Tone  ON, OFF  OFF  g  Items Selection Default Backup Time Constant  0.4 sec, 0.1 sec  0.4 sec Ũ Detection Sensitivity of Pacemaker Pulse   (Pace Sens.) Low, Mid, High  Mid Ũ QRS Detection (QRS WidthᲣ Wide, Narrow  Wide  Ũ Respiration Detection Signal (DISPLAY)  ON, OFF  ON  Ũ Display Brightness (Brightness)  8 levels  5  Ũ Turn Off Display Time (Display OFF)  1 min, 3 min, 10 min, OFF  3 min  Ũ Sound (Sound)  ON, OFF  ON  Ũ Transmitter Channel (CH) One from the following channels.  0801 to 0879 0900 to 0979 1000 to 1079 1100 to 1179 1200 to 1279 1300 to 1379  1100  Ũ Group ID  One from 00 to 63 00  Ũ

Navigation menu