GE Healthcare 6961-MON Novii Intrapartum Maternal/Fetal monitor User Manual Novii FHR Gaps Troubleshooting Guide 107 PT 008 USRev1

GE Healthcare Novii Intrapartum Maternal/Fetal monitor Novii FHR Gaps Troubleshooting Guide 107 PT 008 USRev1

Novii - FHR Gaps Troubleshooting Guide 107-PT-008-USRev1

Download: GE Healthcare 6961-MON Novii Intrapartum Maternal/Fetal monitor User Manual Novii   FHR Gaps Troubleshooting Guide 107 PT 008 USRev1
Mirror Download [FCC.gov]GE Healthcare 6961-MON Novii Intrapartum Maternal/Fetal monitor User Manual Novii   FHR Gaps Troubleshooting Guide 107 PT 008 USRev1
Document ID3684200
Application IDJuJERaoZ2KE3dUGTJrrsug==
Document DescriptionNovii - FHR Gaps Troubleshooting Guide 107-PT-008-USRev1
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize59.84kB (748030 bits)
Date Submitted2017-12-19 00:00:00
Date Available2017-12-14 00:00:00
Creation Date2016-01-25 11:04:57
Producing SoftwareAdobe PDF Library 10.0.1
Document Lastmod2016-02-08 16:08:55
Document TitleNovii - FHR Gaps Troubleshooting Guide 107-PT-008-USRev1
Document CreatorAdobe InDesign CS6 (Macintosh)

FHR Gaps Troubleshooting Guide
POSSIBLE CAUSE
ACTION & SOLUTION1
l Return patient to bed
l Consider using a maternity belt to support pannus during
ambulation or upright position
2.4"/6cm
HELP/TIP
1"/2-3cm
1"/2-3cm
a) Allow 10-15 minutes of monitoring before starting ambulation.
The patient is ambulating
b) The patient should not be encouraged to ambulate unless the FHR
trace is good and the signal indicator on the Novii Interface shows
2.4"/6cm
2.4"/6cm
3 green squares.
l Use a pillow behind back or head
to make patient more comfortable
Signal
1"/2-3cm
2.4"/6cm
l Return patient to a position where
Novii worked well
1"/2-3cm
1"/2-3cm
l Encourage patient to lie on left or right
Patient position/posture
l If patient on side, support abdomen with
a pillow/rolled blanket to re-position
abdomen so that the Patch is centred
over the uterus
l Check electrodes and move or stick back down
1"/2-3cm
SKIN PREP
SKIN PREPARATION GOOD
SKIN PREPARATION GOODREPEAT SK
REPEAT SKIN PREP
REPEAT SK
REPEAT SKIN PREP
REPEAT SKIN PREP
DO NOT BY
REPEAT
SKIN
PREP
DO NOT BYPASSGOOD
SKIN
DOPREPARATION
NOT BYPASS
l Re-position Patch or electrode to avoid the skin problem
REPEAT SKIN PREP
HELP/TIP
REPEAT SKIN PREP
a) The Interface will alert user with a visual message, but only when electrode isDOfully
NOT detached.
BYPASS
b) Check electrodes after a shower, clinical procedure, ambulation or position change.
c) Electrode(s) should not be placed over a skin lesion, skin fold, umbilicus, stretch mark,
pronounced linea nigra.
Electrode ‘detached’ or makes
poor contact with skin
FHR
FHR
MHR
FHR
MHR
UA
FHR
UA
MHR
MHR
UA
UA
1"/2-3cm
Lost MHR and UA
2.4"/6cm
as well as FHR
d) If necessary use a strip of micropore tape to prevent electrode lifting or detachment.
l Check Interface for help message
l Interface is off
HELP/TIP
FHR
MHR
UA
a) If Interface has lost power Bluetooth pairing with POD is lost.
Remove POD from Patch, place in charging bay and start new monitoring
episode when power is returned.
b) Interface does not have a battery back-up.
l Start-Screen is displayed
HELP/TIP
POD has switched off – remove POD from Patch and place in charging well.
Start new monitoring session with other POD.
1"/2-3cm
l Faulty ‘Y’ Connector - replace
l Exfoliate skin under ‘bad’ electrode
SKIN PREPARATION GOOD
REPEAT SKIN PREP
REPEAT SKIN PREP
DO NOT BYPASS
HELP/TIP
a) Peel the X electrode back, remove excess gel from skin and finish with alcohol wipe.
Wait until skin is dry then exfoliate skin and reapply electrode.
b) May need to use micropore tape to hold in place.
l Restart the monitoring session if location of ‘bad’ electrode not known
HELP/TIP
Bypassed red X on
electrode/skin checkscreen
Remove POD from Patch, place in charging bay and start new monitoring episode.
POSSIBLE CAUSE
ACTION & SOLUTION1
l Re-position electrode
1"/2-3cm
HELP/TIP
2.4"/6cm
2.4"/6cm
b) May need to use micropore tape to hold in place.
Lower electrode is not
2.4"/6cm above the
symphysis pubis
1"/2-3cm
a) Peel the electrode back, remove excess gel from skin and finish with alcohol wipe.
Wait until skin is dry then exfoliate skin again and reapply electrode in correct position.
1"/2-3cm
1"/2-3cm
Pannus covering Symphysis Pubis
1"/2-3cm
l Remove electrode and place it lower or higher on the abdomen
SKIN PREPARATION GOOD
2.4"/6cm
REPEAT SKIN PREP
l Alternatively
place electrode just below the point where
SKIN PREPARATION GOOD
the surface curves back on itself ensuring that the electrode
isREPEAT
notSKIN
folded
PREP
REPEAT SKIN PREP
DO NOT BYPASS
REPEAT SKIN PREP
DO NOT BYPASS
1"/2-3cm
Lower electrode may not
be optimally placed2
SKIN PREPARATION GOOD
l Plug in US transducer for short
FHR ‘filling’ and reassurance
8REPEAT SKIN PREP
REPEAT SKIN PREP
DO NOT BYPASS
l Swap back to conventional
FHR
monitoring modalityNovii
(10 sec delay)
US FHR
HELP/TIP
WARNING: Novii UA/MHR delayed
10 seconds (5mm) from US FHR - see panels below
Novii MHR
(10 sec delay)
10 sec
Novii UA
sec delay)
Using TOCO UA with Novii FHR(10could
make an early deceleration appear late
Using US FHR with Novii UA could
make a late deceleration appear early
Novii FHR
(6 sec delay)
Novii FHR
(10 sec delay)
US FHR
None of the above
Novii MHR
(6 sec delay)
Novii MHR
(10 sec delay)
10 sec
10 sec
TOCO/IUPC UA –
real-time
Novii UA
(6 sec delay)
Novii UA
(10 sec delay)
Novii FHR/MHR/UA are synchronised
FHR trace.
but all trace 5mm (10 sec) behind Novii
US FHR
Novii FHR/MHR/UA are synchronised
but all trace 3 mm (6 sec) behind TOCO UA trace.
10 Second Trace Example
6 Second Trace Example
(6 sec delay)
Novii MHR
(6 sec delay)
High/Low UA Sensitivity Explained
UA High
10 sec
TOCO/IUPC UA –
real-time
Novii UA
(6 sec delay)
UA Low
10
11
12
13
14
15
16
17
18
Data (minutes)
Selecting UA Low sensitivity from the Novii display will decrease the UA trace amplitude, suppressing unwanted low amplitude
UA, but it will also reduce the contraction duration. There will be no change to the location of the peak.
Low amplitude UA is considered to be due to artefact from fetal/maternal movement and unsynchronised myometrial activity.
Notes:
This troubleshooting guide assumes that the patient is supine or semi-supine during Patch placement and Novii set-up.
1. ! CAUTION: any intervention will take 10 seconds before its impact will be seen on the trace.
2. The user is familiar with the placement of Patch and lower mid-line electrode in high BMI patients with a pannus.
107-PT-008-USRev1

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.6-c015 81.157285, 2014/12/12-00:43:15
Create Date                     : 2016:01:25 11:04:57Z
Metadata Date                   : 2016:02:08 16:08:55Z
Modify Date                     : 2016:02:08 16:08:55Z
Creator Tool                    : Adobe InDesign CS6 (Macintosh)
Instance ID                     : uuid:0f771506-2d7a-41de-b8e3-72665d59d254
Original Document ID            : xmp.did:F77F11740720681195FEE98BFF029256
Document ID                     : xmp.id:440123E4092068118C148DA44DB246D7
Rendition Class                 : proof:pdf
History Action                  : converted
History Parameters              : from application/x-indesign to application/pdf
History Software Agent          : Adobe InDesign CS6 (Macintosh)
History Changed                 : /
History When                    : 2016:01:25 11:04:57Z
Derived From Instance ID        : xmp.iid:078CBD49142068118083F8138072F210
Derived From Document ID        : xmp.did:F77F11740720681195FEE98BFF029256
Derived From Original Document ID: xmp.did:F77F11740720681195FEE98BFF029256
Derived From Rendition Class    : default
Format                          : application/pdf
Producer                        : Adobe PDF Library 10.0.1
Trapped                         : False
Page Count                      : 2
Creator                         : Adobe InDesign CS6 (Macintosh)
EXIF Metadata provided by EXIF.tools
FCC ID Filing: YOM-6961-MON

Navigation menu