GN Hearing A S BE70 Wireless Hearing Instrument User Manual BE70

GN Hearing A/S Wireless Hearing Instrument BE70

BE70_User manual

USER GUIDEBEHIND-THE-EAR HEARING INSTRUMENTS400346000-GB-14.11-Rev.A.indd   1 10-12-2014   10:03:01
2Hearing instrument type designations for models included in this user guide are: BE70, FCC ID: X26BE70, IC 6941C-BE70; LO85, FCC ID: X26LO85, IC: 6941C-LO85. Please see page 10 for a list of models refer-ring to these types.Statement:This device complies with part 15 of the FCC rules and ICES-003 of the IC rules.Operation is subject to the following two conditions: (1) this device may not cause harmful inter ference, and (2) this device must accept any interference received, including interference that may cause undesired operation.Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pur-suant to part 15 of the FCC rules and CAN ICES-003(B)/NMB-003(B) of the IC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of thefollowing measures:•  Reorient or relocate the receiving antenna.•  Increase the separation between the equipment and receiver.•  Connect the equipment into an outlet on a circuit different from the one in which the receiver is con-nected.•  Consult the dealer or an experienced radio/TV technician for help.Changes or modifications can void the user´s authority to operate the equipment 400346000-GB-14.11-Rev.A.indd   2 10-12-2014   10:03:02
3Intended use Generic air-conduction hearing instruments are wearable sound-amplifying devices intended to compen-sate for impaired hearing. The fundamental operating principle of hearing instruments is to receive, amplify, and transfer sound to the ear drum of a hearing impaired person. The products are in compliance with the following regulatory requirements: •  In EU: the device conforms to the Essential Requirements according to Annex I of Council Directive 93/42/EEC for medical devices (MDD) and essential requirements and other relevant provisions of Direc-tive 1999/5/EC (R&TTE). •  The declaration of conformity may be consulted at www.resound.com•  In US: FCC CFR 47 Part 15, subpart C.•  Other identified applicable international regulatory requirements in countries outside the EU and US. Please refer to local country requirements for these areas.•  Canada: these hearing instruments are certified under the rules of IC.•  Japanese Radio Law and Japanese Telecommunications Business Law Compliance. This device is grant-ed pursuant to the Japanese Radio Law (電波法) and the Japanese telecommunications Business Law  (電気通信事業法) This device should not be modified (otherwise the granted designation number will become invalid)•  Patents  US 7,593,537 US 8,00,849ReSound is a registred trade mark of GN ReSound A/S400346000-GB-14.11-Rev.A.indd   3 10-12-2014   10:03:02
4Prescription use of this TSG hearing instrumentThe TSG module should be used as prescribed by your doctor, audiologist or hearing healthcare profes-sional. In order to avoid permanent hearing damages, the maximum daily usage depends on the level of the generated sound.Should you develop any side effects from using the sound generator, such as dizziness, nausea, head-aches, perceived decrease in auditory function or increase in tinnitus perception, you should discontinue use of sound generator and seek medical evaluation.The target population is primarily the adult population over 18 years of age. This product may also be used with children 5 years of age or older. However, children and physically or mentally challenged users will re-quire training by a doctor, audiologist, hearing healthcare professional or the guardian for the insertion and removal of the hearing instrument containing the TSG module.Important notice for prospective sound generator usersA tinnitus masker is an electronic device intended to generate noise of sufficient intensity and bandwidth to mask internal noises. It is also used as an aid in hearing external noises and speech.Good health practice requires that a person with a tinnitus condition have a medical evaluation by a licensed physician (preferably a physician who specializes in diseases of the ear) before using a sound generator.Licensed physicians who specialize in diseases of the ear are often referred to as otolaryngologists, otologists or otorhinolaryngologists. The purpose of medical evaluation is to assure that all medically treatable conditions that may affect tinnitus are identified and treated before the sound generator instrument is used.400346000-GB-14.11-Rev.A.indd   4 10-12-2014   10:03:02
5The sound generator instrument is a tool to generate sounds to be used with appropriate counselling and/or in a tinnitus management program to relieve patients suffering from tinnitus.400346000-GB-14.11-Rev.A.indd   5 10-12-2014   10:03:02
6IntroductionCongratulations on the purchase of your new hearing instruments. ReSound’s innovative sound technology and design, combined with the customized programming selected by your hearing care professional, will make hearing a more enjoyable experience. Hearing instruments will enable you to hear sounds that you may not have heard in years because of your hearing loss. Practice and a positive attitude are important in learning to use hearing instruments. Your ReSound instruments have been adjusted according to your individual hearing loss and needs. Some people adjust quickly to wearing hearing instruments in their ears and hearing new sounds; other people may need more time.Please read this manual carefully in order to wholly benefit from the use of your hearing instruments. With proper care, maintenance, and usage, your hearing instruments will aid you in better communication for many years. Ask your hearing care professional if you have any questions.Hearing instrument model:Model 77: Battery size 13Model 88: Battery size 13Tube type:Dome size:Left serial number:Right serial number:400346000-GB-14.11-Rev.A.indd   6 10-12-2014   10:03:02
7400346000-GB-14.11-Rev.A.indd   7 10-12-2014   10:03:02
8Important notice for prospective  sound generator users ...................4Introduction............................6Getting started ........................12SmartStart ...........................12Inserting/Replacing the battery............13Low battery indicator ...................14Low battery indicator when paired  with wireless accessories only ............14Inserting/Removing hearing instruments.....15Operation of the hearing instrument . . . . . . . . 18Flight mode* ..........................21Telephone use.........................22Listen to radio or TV ....................22Using ReSound Hearing Instruments with  iPhone®, iPad®, and iPod touch®  (ReSound LiNX2).......................23Cellular phones........................23PhoneNow  ...........................23Telecoil (optional).......................26Tele-loop systems......................26FM and Direct audio input (optional) ........26Connecting/Disconnecting direct audio boot  or integrated FM .......................27Important points for FM  .................28Battery door lock (optional)...............29Care and maintenance ..................30Daily maintenance......................31Cleaning earmoulds ....................31Cleaning thin tubes and domes ...........31How to apply domes....................32Using ReSound hearing instruments with  smart phone apps......................33General precautions ....................34General warnings ......................35Tinnitus Sound Generator (TSG) module ....37Intended use for the TSG module ..........37User instructions for the TSG module  ......37Contents400346000-GB-14.11-Rev.A.indd   8 10-12-2014   10:03:02
9The scientific concepts that form the basis for the device .........................38Technical Specifications .................39TSG warnings   ........................40TSG precautions.......................40Battery warning information ..............42Hearing instrument expectations ..........42Warning to hearing aid dispensers (US Only) . 43Troubleshooting Guide ..................49Warranty and repairs  ...................52Temperature test, transport and storage  information ...........................52400346000-GB-14.11-Rev.A.indd   9 10-12-2014   10:03:02
10Behind-the-ear  (BTE) hearing instru-ments of type BE70  with FCC ID X26BE70,  IC number 6941C-BE70 and size 13 battery are available in the following variants: LS977-DW, LS777-DW, LS577-DWBehind-The-Ear (BTE) hearing instruments type  LO85 with FCC ID X26LO85, IC number 6941C-LO85 and size 13 battery are available in following variants:LS988-DW, LS788-DW, LS588-DWThe identification number for the BE70 and LO85 device models can be found at location “10” as indi-cated in the illustrations on the next page.400346000-GB-14.11-Rev.A.indd   10 10-12-2014   10:03:02
6154328/91078/911121516.a14 13 16.c16.b111.  Programme button2.  Battery compartment & On/Off switch3.  Sound outlet4.  Front microphone inlet5.  Back microphone inlet6.  Volume control7.  Direct audio input8.  Left/Right indicator (Left=Blue/Right=Red)9.  Battery lock (optional)10.  Model (in battery chamber)11.  Manufacturer12.  Serial number (in battery chamber)13.  Earhook14.  Earmold and tubing15.  Thin Tube16.  Domes/earmolda. Open domeb. Tulip-Domec. Custom earmoldTraditional fitting(earmold and tube)Open fitting(Thin Tube and dome)400346000-GB-14.11-Rev.A.indd   11 10-12-2014   10:03:03
12OnOffGetting startedOn/Off function1.  When the battery door is closed, the hearing instrument turns on, and the default program will be activ-ated.2.  To turn off the hearing instrument, open the battery door. Many individuals can use their fingernail to pull it open.SmartStartHearing instruments can be turned on once you have placed them on your ears. If you prefer to turn them on just prior to placing them on your ear, your hearing care professional can activate a function called Smart-Start. This function will delay the time in which the hearing instruments turn on after the battery compartment is closed. With SmartStart, a beep will be heard for each second of the delay period.400346000-GB-14.11-Rev.A.indd   12 10-12-2014   10:03:03
21313Inserting/Replacing the battery1.  Open the battery door completely by using your fingernail. Remove the used battery if present.2.  Prepare the new battery (please refer to page 5 for information on appropriate battery type/size for your hearing instrument). Remove the protective foil and wait 2 minutes before inserting the battery into the hearing instrument to allow activation of the battery.3.  Insert the new battery with the positive side in the correct position. 4.  Gently close the battery door.i Tip: 1.  Always use new Zinc-Air batteries that have a minimum remaining shelf life of 1 year.2.  Whenever the hearing instruments are not in use, remember to turn them off to avoid unnecessary battery consumption.3.  Remove the sticker from the battery and let it air for 2 minute before you insert it.4.  If hearing instrument is experiencing frequent loss of connection to ReSound Unite accessories, contact your hearing care professional for a list of low impedance batteries.400346000-GB-14.11-Rev.A.indd   13 10-12-2014   10:03:04
14Low battery indicatorYour hearing care professional can activate a low battery indicator in your hearing instruments. The hear-ing instrument will reduce amplification and play a melody if battery power gets too low. This signal will recur every 15 minutes until the hearing instrument automatically switches off. The timing of the low battery indicator can vary slightly, depending on the type of battery used. It is recommended that you keep spare batteries on hand.Low battery indicator when paired with wireless accessories onlyActive usage of the ReSound Unite accessories (Remote Control 2, Phone Clip+, TV Streamer 2 and Mini Microphone) or an FM receiver requires more battery power from the hearing instruments than when they are working without accessories which means that battery life is highly dependent on the amount of wire-less accessory usage. When the battery in the hearing instrument has depleted to a level at which use of ReSound Unite TV Streamer 2, Phone Clip+ and Mini Microphone cannot be supported, the hearing instru-ment will play two sets of descending tones.After this, your hearing instrument and ReSound Unite Remote Control 2 will continue to work as usual, but you will not be able to use your ReSound Unite TV Streamer 2, Phone Clip+ and Mini Microphone. At some point the battery level will also be too low to support the remote control as well and you will once again hear the descending tones. The hearing instruments will continue to work as usual. Once a new battery is inserted, full operation of the accessories will resume.400346000-GB-14.11-Rev.A.indd   14 10-12-2014   10:03:04
15Inserting/Removing hearing instrumentsInsertion (custom earmoulds)1.  Hold the earmould between your thumb and index finger and position its sound outlet in your ear canal.2.  Slide the earmould all the way into your ear with a gentle, twisting movement.3.  Turn the top part of the earmould gently backwards and forwards so that it tucks behind the fold of skin above your ear canal.4.  Move the earmould up and down and gently press to ensure it is positioned correctly in the ear. Opening and closing your mouth can ease insertion.5.  Make sure the hearing instrument is seated behind the ear.400346000-GB-14.11-Rev.A.indd   15 10-12-2014   10:03:04
16By experimenting, an easier method may be discovered. With proper insertion, hearing instruments should fit snugly but comfortably. If the hearing instruments cause irritation of the ears, contact your hearing care professional. i Note: Never attempt to modify the shape of the hearing instrument, earmoulds, or tubing yourself.i Tip: It may be helpful to pull your ear up and outward with your opposite hand during insertion.Removal (custom earmolds)1.  Hold a portion of your earmould towards the back of the ear.2.  Pull the earmould outward and simultaneously rotate the earmould forward.3.  Consult your hearing care professional if you have difficulties removing the hearing instrument.400346000-GB-14.11-Rev.A.indd   16 10-12-2014   10:03:04
17Insertion (domes/custom earmoulds with thin tubes)1.  Hold the thin tube where it bends, and gently place the dome into the ear canal. Push the dome far enough into the ear canal so that the thin tube lies flush with the head.2.  It is important that the tube and the dome fit correctly into your ear.3.  When the dome is placed correctly, you should not be able to see the thin tube sticking out when facing a mirror.i Note: You should never attempt to bend or modify the shape of the thin tube.Removal (domes/custom earmoulds with thin tubes)1.  Hold the thin tube with your thumb and forefinger and remove the tube.2.  For thin tube custom earmoulds, grasp the removal string and pull the earmould outward.3.  Consult your hearing care professional if you have difficulties removing the hearing instrument. 400346000-GB-14.11-Rev.A.indd   17 10-12-2014   10:03:05
18Operation of the hearing instrumentVolume control (optional)The volume control will allow the volume of hearing instruments to be increased or decreased.1.  To increase the volume, push the volume control up.2.  To decrease the volume, push the volume control down.When volume is increased or decreased, a beep signal will be heard for each incremental change. When the upper or lower limits of the volume range are reached, a beep signal with a longer duration will be heard.If you have two hearing instruments with the synchronization function enabled, volume control adjustments to one instrument will automatically repeat in the second instrument. When a volume control adjustment is made in one instrument, you will hear a confirmation beep. A beep in the second instrument will follow.Your hearing care professional can disable the volume control function if it is not desired. The volume con-trol can also be replaced with a non-functional cover.400346000-GB-14.11-Rev.A.indd   18 10-12-2014   10:03:05
19Programme button Depending on your experience level with hearing instruments, in-dividual hearing needs, and the type of listening environments you experience, your hearing care professional may activate additional programmes. If additional programmes have been activated, the fol-lowing list explains how they work.Programme Description of when to use12341.  You can switch between programmes by pushing the push button once.2.  You will then hear one or more beeps. The number of beeps indicates which programme you have selected (one beep = programme one, two beeps = programme two, etc.). 3.  When the hearing instruments are turned off and then back on, the hearing instrument always returns to the default setting (programme one).400346000-GB-14.11-Rev.A.indd   19 10-12-2014   10:03:06
20If you have two hearing instruments with the synchronization function enabled, program changes to one instrument will automatically repeat in the second instrument. When a program change is made in one instrument, you will hear the same amount of confirmation beeps in the second instrument.400346000-GB-14.11-Rev.A.indd   20 10-12-2014   10:03:06
21*Wireless models onlyi Flight mode*When boarding a flight or entering an area where RF transmitters are prohibited, wireless functional-ity must be deactivated, as it is not allowed to radiate radio signals during flights or in otherwise restricted areas.For LiNX2 wireless hearing instruments follow the following instructions to enter and leave flight mode:It is possible to disable wireless operation by opening and closing the battery compartment three times within a ten second period (open-close, open-close, open-close). Your instruments will now be in flight mode.  If the hearing instrument is in flight mode, the hearing instrument must have been operating in flight mode  for at least 10 seconds before attempting to enable wireless again. it is possible to re-enable wireless opera-tion by opening and closing the battery door once. 10 seconds after this operation is completed, wireless operation will begin again.i Note: It is important to wait an additional 15 seconds after wireless function resumes before open-ing and closing the battery compartment again for any reason. If the battery compartment is opened and closed during this 15 second window, flight mode will resume.400346000-GB-14.11-Rev.A.indd   21 10-12-2014   10:03:06
22Telephone useFinding the optimal position for holding a telephone may require practice for some individuals, and one or more of the following suggestions may be helpful.1.  Hold the telephone as you would normally.2.  Hold the telephone towards the top of the ear (closer to where the microphones are).3.  If whistling occurs, it may take a few seconds of holding the telephone in the same position before the hearing instrument eliminates the feedback.4.  Any whistling may also be decreased by holding the telephone slightly away from the ear.5.  Depending on your individual needs, your hearing care professional may activate a programme specifically for telephone use.Listen to radio or TVWhen listening to the TV or the radio, start out by listening to news commentators since they usually speak clearly, then try other programmes. If you find it difficult to listen to TV or radio, your hearing care profes-sional will be able to give you advice on available wireless accessories to enhance your listening capabilities for TV and radio.400346000-GB-14.11-Rev.A.indd   22 10-12-2014   10:03:06
23Using ReSound Hearing Instruments with iPhone®, iPad®, and iPod touch® (ReSound LiNX2)ReSound LiNX2 is a Made for iPhone instrument and allows for direct communication and control with an iPhone, iPad, or iPod touch. For assistance in pairing and using these products with your ReSound LiNX2 device, please contact your hearing care professional or visit our support site at resound.com/linx.Cellular phonesYour hearing instrument is designed to comply with the most stringent Standards of International Electro-magnetic Compatibility. However, not all cell phones are hearing instrument compatible. The varying degree of disturbance can be due to the nature of your particular cellular phone or of your wireless telephone service provider.If you find it difficult to obtain a good result while using your cellular phone, your hearing care professional will be able to give you advice on available wireless accessories to enhance listening capabilities.PhoneNow The PhoneNow function, allows your hearing instrument to automatically switch to your tele phone pro-gramme when a telephone receiver is raised to the ear. When the telephone receiver is removed from the ear, the hearing instrument automatically returns to the previous listening programme.400346000-GB-14.11-Rev.A.indd   23 10-12-2014   10:03:06
24Placement of PhoneNow magnetsPlace PhoneNow magnet on your telephone receiver to allow operation of the PhoneNow function. In order to place PhoneNow magnet properly:1.  Clean the telephone receiver thoroughly.2.  Hold the telephone vertically, in a position similar to when making a telephone call.3.  Place the magnets just below the telephone receiver. Make sure not to cover the microphone openings. If necessary, move the magnet to another position to improve ease of use and comfort while speaking.4.  If you are not satisfied with the strength of PhoneNow, you can reposition the PhoneNow magnet or add additional PhoneNow magnets.PhoneNow usageTelephones can be used in a normal manner. A short melody will indicate that the PhoneNow feature has automatically switched the hearing instrument to your telephone programme. Initially, you may need to move the telephone receiver slightly to find the best position for reliable PhoneNow activation and good hearing on the telephone.If you have two hearing instruments with the synchronization function enabled, the volume of hearing instru-ment on the non-phone ear will be turned down.i Only use recommended cleaning agent to clean the telephone prior to placing the magnet on the phone in order to obtain best possible adherence.400346000-GB-14.11-Rev.A.indd   24 10-12-2014   10:03:06
25i PhoneNow warnings1.  Keep magnets out of reach of pets, children and people with mental disabilities. If a magnet is swallowed, please seek advice from a medical practitioner.2.  The magnet may affect some medical devices or electronic systems. The manufacturer of any magnetically sensitive devices (e.g. pacemakers) should advise you regarding appropriate safety precautions when using your hearing instrument and magnet in close proximity to the medical device or electronic system in question.If the manufacturer cannot issue a statement, we recommend keeping the magnet or a telephone equipped with the magnet 30 cm (12”) away from magnetically sensitive devices (e.g. pacemakers).i PhoneNow precautions1.  High distortion during dialing or phoning may mean that the magnet is not in the optimal position relative to the telephone receiver. To avoid the issue, please move the magnet to another place on the telephone receiver.2.  Only use magnets supplied by ReSound.400346000-GB-14.11-Rev.A.indd   25 10-12-2014   10:03:06
26Telecoil (optional)A telecoil can be activated by your hearing care professional and accessed through one of the additional programmes. A telecoil picks up a telephone’s magnetic signal and converts it to sound. An optional tel-ephone programme may help to improve speech understanding on the telephone. When using a telecoil programme, the receiver of the telephone may need to be held closer to the hearing instrument. The hand-set of the telephone may need to be moved to slightly different positions in order to find the best reception.Tele-loop systemsMany places, such as theatres, houses of worship, and schools are equipped with tele-loop systems. When using a telecoil programme with tele-loop systems, sound is picked up directly and may improve speech understanding. If there is no sound from the hearing instruments in a tele-loop system and with a telecoil programme activated, the tele-loop system may not be turned on or is not operating correctly. If a facility is not equipped with a tele-loop system, sitting as close as possible to the front may be helpful.FM and Direct audio input (optional)An FM system can greatly improve speech understanding in many situations such as when a speaker is a long distance away or a room is noisy or reverberant. The direct audio input (DAI) can be used with a univer-sal FM receiver or together with a cable to connect to a radio, computer or television. The sound source is connected to the hearing instruments by a cable or a wireless FM system to the audio boot. This accessory connects to the bottom of the hearing instruments, and once properly clicked into place, the hearing instru-ments switches to DAI automatically.400346000-GB-14.11-Rev.A.indd   26 10-12-2014   10:03:06
1325427Connecting/Disconnecting direct audio boot or integrated FMConnecting audio boot and integrated FM1.  Align the tip of the audio boot or integrated FM boot with the groove just above the battery compartment and below the model number.2.  Once in place, move the boot in the direction of the battery compartment.3.  Gently click the boot onto the hearing instrument.Disconnecting audio boot and integrated FM4.  Press and hold the button on the front side of the audio boot. For an FM boot, slide the latch downward with your finger nail.5.  Gently remove the boot from the hearing instrument.Direct audio Integrated FM400346000-GB-14.11-Rev.A.indd   27 10-12-2014   10:03:09
28i Important points for FM •  Do not use two transmitters on the same FM channel.•  Do not use water or fluids for cleaning FM boot.•  Do not use an FM transmitter in locations where it is forbidden to use electronic devices, for instance in airplanes.•  Be aware that FM signals might also be picked up and overheard by other receivers.•  Before using the system in another country, contact your hearing care professional to make sure your radio channel is permitted in that country.•  Your FM boot and transmitter may only be repaired by an authorized service center. 400346000-GB-14.11-Rev.A.indd   28 10-12-2014   10:03:09
29Battery door lock (optional)Your hearing instrument can be equipped with a lock to prevent children or mentally challenged persons from accidently ingesting the battery. Ask your hearing care professional to demonstrate the locking system on your hearing instruments. Standard battery door lockYour hearing care professional can install a battery door lock. Once the battery door has been locked, the hearing instrument can be turned on and off, but the battery cannot be changed unless the battery door is unlocked again.To lock the battery door:1.  Open the battery door to it’s off position.2.  With the battery door in it’s off position, slide the battery lock slider to the left by pressing from the right side of the battery door (see illustration)To unlock the battery door:3.  With the battery door in it’s off position, slide the battery lock slider to the right by pressing from the left side of the battery door (see illustration)400346000-GB-14.11-Rev.A.indd   29 10-12-2014   10:03:10
30i Care and maintenancePlease follow the following instructions to prolong the durability of your hearing instruments:1.  Keep your hearing instrument clean and dry. Wipe the case with a soft cloth or tissue after use to remove grease or moisture. Do not use water or solvents, as these can damage the hearing instrument(s).2.  Never immerse hearing instruments in water or other liquids, as liquids may cause permanent damage to the hearing instruments.3.  Avoid rough handling of hearing instruments or dropping them on hard surfaces or floors.4.  Do not leave hearing instruments in or near direct heat or sunlight, such as in a hot, parked car, as excessive heat can cause damage or deform the casing. 5.  Do not wear your instrument while showering, swimming, in heavy rain or in a moist atmosphere such as a steam bath or sauna.6.  If your instrument does get wet, or if it has been exposed to high humidity or perspiration, it should be left to dry out overnight with the battery out and the battery compartment open. It is also a good idea to put the instrument and battery in a sealed container together with a drying agent (desiccator) overnight. Do not use the instrument until it is completely dry. Consult your hearing care professional as to which drying agent to use.7.  Remove your hearing instrument when applying such things as cosmetics, perfume, aftershave, hair spray, and suntan lotion. These might get into the instrument and cause damage.400346000-GB-14.11-Rev.A.indd   30 10-12-2014   10:03:10
31i Daily maintenanceIt is important to keep your hearing instrument clean and dry. On a daily basis, clean the hearing instruments using a soft cloth or tissue. In order to avoid damage due to humidity or excessive perspiration, the use of a drying kit is recommended. i Cleaning earmoulds1.  Remove the earmould and attached tubing from the hearing instruments prior to cleaning.2.  Clean the earmould using a mild soap, and rinse with lukewarm water.3.  After cleaning, dry earmoulds thoroughly and remove any residual water and debris from the tubing utilising an air bulb and wire loop.i Note: Earmould tubing may become stiff, brittle, or discoloured over time. Con-tact your hearing care professional regarding tube changes.i Cleaning thin tubes and domes1.  Remove thin tubes from hearing instruments before cleaning by  unscrewing them counter clockwise.2.  Wipe down thin tubes and domes with a damp cloth.3.  In order to clear the thin tube of moisture and debris, push the black cleaning rod through the thin tube, beginning at the end opposite the dome.400346000-GB-14.11-Rev.A.indd   31 10-12-2014   10:03:11
1232Note: Thin tube and dome systems should be changed every three months or sooner, should the compo-nents become stiff or brittle.How to apply domesIt is recommended that your hearing care professional change domes, as incorrect dome replacement could result in the dome falling out in the ear.ReSound Standard domes1.  Push the new dome over the flange on the thin tube.2.  Make sure that the new dome is properly and securely mounted.ReSound Tulip domesTulip domes are mounted in a similar manner to standard domes, but a few extra steps are required. Tulip domes consist of two “petals”. It is important to note that the largest petal is the outermost petal. To ensure this:400346000-GB-14.11-Rev.A.indd   32 10-12-2014   10:03:11
331.  Push the largest petal away from the thin tube using a finger. This bends the petal forward.2.  Then push the largest petal back, and it will be placed on top of the smaller petal.Using ReSound hearing instruments with smart phone appsi Intended use of smart phone apps:GN ReSound smart phone apps are intended to be used with GN ReSound wireless hearing aids. GN Re-Sound smart phone apps send and receive signals from the GN ReSound wireless hearing aids via smart phones for which the apps have been developed.Use with smart phone apps:•  Notifications of app updates should not be disabled, and it is recommended that the user installs all updates to ensure that the app will function correctly and will be kept up to date.•  The app must only be used with GN ReSound devices for which it is intended, and GN ReSound take no responsibility if the app is used with other devices.•  If you would like a printed version of the user guide for a smart phone app please consult customer sup-port or our website to obtain a printed user guide.400346000-GB-14.11-Rev.A.indd   33 10-12-2014   10:03:12
34i General precautions1.  When wireless function is activated, the device uses low-powered digitally coded transmissions in order to communicate with other wireless devices. Although unlikely, nearby electronic devices may be affected. In that case, move the hearing instrument away from the affected electronic device.2.  When using wireless functionality and the devices are affected by electromagnetic interference, move away from the source.3.  Use only original GN ReSound consumables e.g. tubes and domes. 4.  Never attempt to modify the shape of the hearing instrument, ear-moulds, or tubing yourself.5.  Only connect ReSound hearing instruments to ReSound accessories intended and qualified to be used with ReSound hearing instruments.400346000-GB-14.11-Rev.A.indd   34 10-12-2014   10:03:12
35i General warnings1.  Consult a hearing care professional if you think there may be a foreign object in your ear canal, if you experience skin irritation, or if excessive ear wax accumulates with the use of the hearing instrument. 2.  Different types of radiation, from e.g. NMR, MRI, or CT scanners, may damage hearing instruments. It is recommended not to wear hearing instruments during these or other similar procedures. Other types of radiation, such as burglar alarms, room surveillance systems, radio equipment, mobile telephones, contain less energy and will not damage hearing instruments. However, they have the potential to momentarily affect the sound quality or temporarily create strange sounds from hearing instruments.3.  Do not wear hearing instruments in mines, oil fields, or other explosive areas unless those areas are certified for hearing instrument use.4.  Do not allow others to use your hearing instruments. This may cause damage to the hearing instruments or to the hearing of the other individual.5.  Instrument usage by children or mentally challenged persons should be supervised at all times to ensure their safety. The hearing instrument contains small parts that could be swallowed by children. Please be mindful not to leave children unsupervised with this hearing instrument.6.  Hearing instruments should be used only as prescribed by your hearing care professional. Incorrect use may result in hearing loss. 7.  Warning to hearing care professionals: Special care should be exercised in selecting and fitting hearing instruments with maximum sound pressure level that exceeds 132dB SPL with an IEC 60711:1981 occluded ear simulator. There may be a risk of impairment of the remaining hearing. 8.  Be careful when boarding flights to deactivate the wireless functionality. Turn off your wireless functionality by using the flight mode in areas where radio frequency emission is prohibited.9.  If device is broken, do not use.10.  External devices connected to the electrical input must be safe according to the requirements of IEC 60601-1-1, IEC 60065, or IEC 60950-1, as appropriate (wired connection, f.ex. HI-PRO), SpeedLink).400346000-GB-14.11-Rev.A.indd   35 10-12-2014   10:03:12
36i Note:* ReSound wireless devices operate in the frequency range of 2.4 GHz - 2.48 GHz.* ReSound wireless devices include a RF transmitter that operates in the range of 2.4 GHz - 2.48 GHz.*  For use of wireless functionality only use ReSound Unite accessories. For further guidance regarding e.g. pairing, please refer to the user guide of the relevant ReSound Unite accessory.400346000-GB-14.11-Rev.A.indd   36 10-12-2014   10:03:12
37Tinnitus Sound Generator (TSG) moduleIntended use for the TSG moduleYour ReSound hearing instruments includes the Tinnitus Sound Generator function, a tool for generating sounds to be used in tinnitus management programmes to relieve suffering from tinnitus.The Tinnitus Sound Generator can generate sounds adjusted to the specific therapeutic needs and your personal preference as determined by your doctor, audiologist, or hearing care professional. Depending on the selected hearing instrument program and the environment you are in, you will sometimes hear the therapeutic sound resembling a continuous or fluctuating whistling.User instructions for the TSG module Description of the deviceThe Tinnitus Sound Generator (TSG) Module is a software tool that generates sounds to be used in tinnitus management programmes to relieve suffering from tinnitus.Explanation of how the device functionsThe TSG module is a frequency and amplitude shaped white-noise generator. Noise signal level and fre-quency characteristics can be adjusted to the specific therapeutic needs as determined by your doctor, audiologist or hearing healthcare professional.Your doctor, audiologist or hearing healthcare professional can modulate the generated noise with the purpose of making it more pleasant. The noise can then resemble, for example, crushing waves on a shore.Modulation level and speed can also be configured to your likes and needs.If your tinnitus troubles you only 400346000-GB-14.11-Rev.A.indd   37 10-12-2014   10:03:12
38in quiet environments, your doctor, audiologist or hearing healthcare professional can set the TSG Module so that it becomes audible exclusively in such surroundings. The overall sound level can be adjusted via an optional volume control. Your doctor, audiologist or hearing healthcare professional will review with you the need for having such a control.TSG volume controlThe sound generator is set to a specific loudness level by the hearing healthcare professional. When switch-ing the sound generator on, the volume will have this optimal setting. Therefore, it might not be necessary to control the volume (loudness) manually. However, the volume control provides the ability to adjust the volume, or amount of stimulus, to the liking of the user.The scientific concepts that form the basis for the deviceThe TSG module provides sound enrichment with the aim of surrounding the tinnitus sound with a neutral sound which is easily ignored. Sound enrichment is an important component of most approaches to tin-nitus management, such as Tinnitus Retraining Therapy (TRT). To assist habituation to tinnitus, this needs to be audible. The ideal level of the TSG module, therefore, should be set so that it starts to blend with the tinnitus, and so that you can hear both your tinnitus as well as the sound used.In a majority of instances, the TSG module can also be set to mask the tinnitus sound, so to provide tem-porary relief by introducing a more pleasant and controllable sound source.400346000-GB-14.11-Rev.A.indd   38 10-12-2014   10:03:12
39Technical SpecificationsAudio signal technologyDigitalAvailable soundsWhite noise signal which can be shaped with the following configurations:The white noise signal can be modulated in amplitude with an attenuation depth of up to 14dB.High-pass filter Low-pass filter500 Hz 2000 Hz750 Hz 3000 Hz1000 Hz 4000 Hz1500 Hz 5000 Hz2000 Hz 6000 Hz400346000-GB-14.11-Rev.A.indd   39 10-12-2014   10:03:12
40i TSG warnings  •  Sound generators can be dangerous if improperly used.•  Sound generators should be used only as advised by your doctor, audiologist, or hearing healthcare professional.•  Sound generators are not toys and should be kept out of reach of anyone who might cause themselves injury (especially children and pets).i TSG precautions•  Should the user develop any side effects from using the sound generator, such as dizziness, nausea, headaches, perceived decrease in auditory function or increase in tinnitus perception, the user should discontinue use of the sound generator and seek medical evaluation.•  Children and physically or mentally challenged users will require guardian supervision while wearing the TSG hearing instrument.•  The volume control is an optional feature in the TSG module used for adjusting the sound generator output level. To prevent unintended usage by pediatric or physically or mentally challenged users, the volume control must, if enabled, be configured to only provide a decrease of the sound generator output level.400346000-GB-14.11-Rev.A.indd   40 10-12-2014   10:03:12
41i TSG warning to hearing healthcare professionalsA hearing healthcare professional should advise a prospective sound generator user to consult promptly with a licensed physician (preferably an ear specialist) before getting a sound generator if the hearing healthcare professional determines through inquiry, actual observation, or review of any other available information concerning the prospective user that the prospective user has any of the following conditions:(i)  Visible congenital or traumatic deformity of the ear.(ii)  History of active drainage from the ear within the previous 90 days.(iii)  History of sudden or rapidly progressive hearing loss within the previous 90 days.(iv)  Acute or chronic dizziness.(v)  Unilateral hearing loss of sudden or recent onset within the previous 90 days.(vi)  Audiometric air-bone gap equal to or greater than 15dB at 500 hertz (Hz), 1000 Hz, and 2000 Hz.(vii)  Visible evidence of significant cerumen accumulation or a foreign body in the ear canal.(viii)  Pain or discomfort in the ear.i CAUTION: The maximum output of the sound generator falls into the range that can cause hearing loss according to OSHA regulations. The user should not use the sound generator for more than eight (8) hours a day when this is set below 90dB SPL. Above that level, the device should not be used for more than two (2) hours per day. In no case should the sound generator be worn at uncomfortable levels. 400346000-GB-14.11-Rev.A.indd   41 10-12-2014   10:03:12
42i Battery warning informationBatteries, although very small, contain dangerous substances, and should be disposed of carefully. This is for the safety of you and the environment. Please note: 1.  Do not attempt to recharge batteries (Zinc Air) which are not specifically designated as rechargeable because they may leak or explode.2.  DO NOT attempt to dispose of batteries by burning them. Used batteries are harmful to the environment. Please dispose of them according to local regulations or return them to your hearing care practitioner.3.  DO NOT place batteries in your mouth. Consult a physician immediately if a battery has been swallowed, as they can be harmful to your health.4.  Keep batteries away from pets, children and individuals who are mentallly challenged.5.  Remove the batteries to prevent leakage when the hearing instruments are not in use for an extended period of time.i Hearing instrument expectationsA hearing instrument will not restore normal hearing and will not prevent or improve a hearing impairment re-sulting from organic conditions. Consistent use of the hearing instrument is recommended. In most cases, infrequent use does not permit you to attain full benefit from it.The use of a hearing instrument is only part of hearing rehabilitation and may need to be supplemented by auditory training and instructions in lip-reading.400346000-GB-14.11-Rev.A.indd   42 10-12-2014   10:03:12
43i Warning to hearing aid dispensers (US Only)A hearing aid dispenser should advise a prospective hearing aid user to consult promptly with a licensed physician (preferably an ear specialist) before dispensing a hearing aid if the hearing aid dispenser de-termines through inquiry, actual observation, or review of any other available information concerning the prospective user, that the prospective user has any of the following conditions:(i)  Visible congenital or traumatic deformity of the ear.(ii)  History of active drainage from the ear within the previous 90 days.(iii)  History of sudden or rapidly progressive hearing loss within the previous 90 days.(iv)  Acute or chronic dizziness.(v)  Unilateral hearing loss of sudden or recent onset within the previous 90 days.(vi)  Audiometric air-bone gap equal to or greater than 15 decibels at 500 hertz (Hz), 1,000 Hz, and 2,000 Hz.(vii)  Visible evidence of significant cerumen accumulation or a foreign body in the ear canal.(viii)  Pain or discomfort in the ear.400346000-GB-14.11-Rev.A.indd   43 10-12-2014   10:03:12
44Important notice for prospective hearing aid users (US Only)Good health practice requires that a person with a hearing loss have a medical evaluation by a licensed physician (preferably a physician who specializes in diseases of the ear) before purchasing a hearing aid. Licensed physicians who specialize in diseases of the ear are often referred to as otolaryngologists, otolo-gists or otorhinolaryngologists. The purpose of medical evaluation is to assure that all medically treatable conditions that may affect hearing are identified and treated before the hearing aid is purchased.Following the medical evaluation, the physician will give you a written statement that states that your hearing loss has been medically evaluated and that you may be considered a candidate for a hearing aid. The physi-cian will refer you to an audiologist or a hearing aid dispenser, as appropriate, for a hearing aid evaluation.The audiologist or hearing aid dispenser will conduct a hearing aid evaluation to assess your ability to hear with and without a hearing aid. The hearing aid evaluation will enable the audiologist or dispenser to select and fit a hearing aid to your individual needs. If you have reservations about your ability to adapt to amplifi-cation, you should inquire about the availability of a trial-rental or purchase-option program. Many hearing aid dispensers now offer programs that permit you to wear a hearing aid for a period of time for a nominal fee after which you may decide if you want to purchase the hearing aid.Federal law restricts the sale of hearing aids to those individuals who have obtained a medical evaluation from a licensed physician. Federal law permits a fully informed adult to sign a waiver statement declining the medical evaluation for religious or personal beliefs that preclude consultation with a physician. The exercise of such a waiver is not in your best health interest and its use is strongly discouraged.400346000-GB-14.11-Rev.A.indd   44 10-12-2014   10:03:12
45Children with hearing loss (US Only)In addition to seeing a physician for a medical evaluation, a child with a hearing loss should be directed to an audiologist for evaluation and rehabilitation since hearing loss may cause problems in language develop-ment and the educational and social growth of a child. An audiologist is qualified by training and experience to assist in the evaluation and rehabilitation of a child with a hearing loss.400346000-GB-14.11-Rev.A.indd   45 10-12-2014   10:03:12
46Open ClosedReference test gain (60 dB SPL input) HFA 38 43 dBFull-on gain (50 dB SPL Input) MaxHFA51485753dBdBMaximum output (90 dB SPL input) MaxHFA127116124121dB SPLdB SPLTotal harmonic distortion 500 Hz800 Hz1600 Hz0.20.20.6 0.40.80.7%%%Telecoil sensitivity (SPLIV @ 31.6 mA/m) 100 105 dB SPLEquivalent input noise (w/o noise reduction) 22 20 dB SPLFrequency range (DIN 45605) 100 –6810100 –6140 HzCurrent drain (in test mode) 1.2 1.2 mABTEModels: LS977-DW, LS777-DW, LS577-DWOpen configurationClosed configuration FULL-ON AND REFERENCE TEST GAINFrequency (Hz)Gain (dB)2cc CouplerReference test gain 60 dB SPL inputFull-on gain 50 dB SPL inputMAXIMUM OUTPUT (OSPL 90)Frequency (Hz)Output (dB SPL)2cc CouplerData in accordance with ANSI S3.22-2009; and IEC 60118-7 ed.2.0 Supply Voltage 1.3 V, 2cc coupler400346000-GB-14.11-Rev.A.indd   46 10-12-2014   10:03:13
80706050403020100 1000 100001401301201101009080100 1000 1000047POWER BTEModels: LS988-DW, LS788-DW, LS588-DWData in accordance with ANSI S3.22-2009; and IEC 60118-7 ed.2.0 Supply Voltage 1.3 V, 2cc couplerGain (dB)Reference test gain 60 dB SPL inputFull-on gain 50 dB SPL input2cc CouplerFrequency (Hz)MAXIMUM OUTPUT (OSPL 90)Frequency (Hz)Output (dB SPL)2cc CouplerReference test gain (60 dB SPL input) HFA 51 dBFull-on gain (50 dB SPL Input) MaxHFA6763dBdBMaximum output (90 dB SPL input) MaxHFA132128dB SPLdB SPLTotal harmonic distortion 500 Hz800 Hz1600 Hz0.50.50.3 %%%Telecoil sensitivity (SPLIV @ 31.6 mA/m) 111 dB SPLEquivalent input noise (w/o noise reduction) 22 dB SPLFrequency range (DIN 45605) 100 –6020 HzCurrent drain (in test mode) 1.4 mAFULL-ON AND REFERENCE TEST GAIN400346000-GB-14.11-Rev.A.indd   47 10-12-2014   10:03:13
48Troubleshooting Guide SYMPTOM CAUSEPOSSIBLE REMEDYNo sound Not turned onTurn on by closing the battery doorDead batteryReplace batteryBattery door will not closeInsert battery properlyBlocked earmould or tubeClean earmould or tubeBlocked wax filterReplace wax filter or consult your hearing care professionalNot loud enough Incorrect earmould placementReinsert earmouldBlocked earmould or domeClean earmould, replace dome, replace filterBlocked sound outlet filterChange filter or consult your hearing care professionalChange in hearing sensitivityConsult your hearing care professionalExcessive ear wax Consult your physician Volume set too lowIncrease the volume control if available or consult your hearing care professional400346000-GB-14.11-Rev.A.indd   48 10-12-2014   10:03:13
49Troubleshooting Guide SYMPTOM CAUSEPOSSIBLE REMEDYNo sound Not turned onTurn on by closing the battery doorDead batteryReplace batteryBattery door will not closeInsert battery properlyBlocked earmould or tubeClean earmould or tubeBlocked wax filterReplace wax filter or consult your hearing care professionalNot loud enough Incorrect earmould placementReinsert earmouldBlocked earmould or domeClean earmould, replace dome, replace filterBlocked sound outlet filterChange filter or consult your hearing care professionalChange in hearing sensitivityConsult your hearing care professionalExcessive ear wax Consult your physician Volume set too lowIncrease the volume control if available or consult your hearing care professionalIf there are any other problems not mentioned in this guide, please contact your hearing care professional.400346000-GB-14.11-Rev.A.indd   49 10-12-2014   10:03:13
50Troubleshooting Guide SYMPTOM CAUSEPOSSIBLE REMEDYExcessive whistling / feed-backIncorrect earmould placementRe-insert earmould carefullyIncorrect dome placementRe-insert domeExcessive ear waxConsult your hearing care professionalFeedback control may need adjustmentConsult your hearing care professionalEarmould tubing worn or damagedConsult your hearing care professionalThin tube connection looseChange thin tube or consult your hearing care professionalHearing instrument settings not optimalConsult your hearing care professionalSound distorted / not clear Weak batteryReplace batteryImproper earmould or dome fitConsult your hearing care professionalHearing instrument damagedConsult your hearing care professionalHearing instrument settings not optimalConsult your hearing care professionalWireless does not work Possible Root Cause - Device is in flight modeOpen and close the battery compartment once. Wireless will reactivate 10 seconds later. (If Root Cause is device in flight mode)400346000-GB-14.11-Rev.A.indd   50 10-12-2014   10:03:13
51Troubleshooting Guide SYMPTOM CAUSEPOSSIBLE REMEDYExcessive whistling / feed-backIncorrect earmould placementRe-insert earmould carefullyIncorrect dome placementRe-insert domeExcessive ear waxConsult your hearing care professionalFeedback control may need adjustmentConsult your hearing care professionalEarmould tubing worn or damagedConsult your hearing care professionalThin tube connection looseChange thin tube or consult your hearing care professionalHearing instrument settings not optimalConsult your hearing care professionalSound distorted / not clear Weak batteryReplace batteryImproper earmould or dome fitConsult your hearing care professionalHearing instrument damagedConsult your hearing care professionalHearing instrument settings not optimalConsult your hearing care professionalWireless does not work Possible Root Cause - Device is in flight modeOpen and close the battery compartment once. Wireless will reactivate 10 seconds later. (If Root Cause is device in flight mode)400346000-GB-14.11-Rev.A.indd   51 10-12-2014   10:03:13
52Warranty and repairs ReSound provides a warranty on hearing instruments in the event of defects in workmanship or materials, as described in applicable warranty documentation. In its service policy, ReSound pledges to secure func-tionality at least equivalent to the original hearing instrument. As a signatory to the United Nations Global Compact initiative, ReSound is committed to doing this in line with environment-friendly best practices. Hearing instruments therefore, at ReSound’s discretion, may be replaced by new products or products manufactured from new or serviceable used parts, or repaired using new or refurbished replacement parts. The warranty period of hearing instruments is designated on your warranty card, which is provided by your hearing care professional.For hearing instruments that require service, please contact your hearing care professional forassistance. ReSound hearing instruments that malfunction must be repaired by a qualified technician. Do not attempt to open the case of hearing instruments, as this will invalidate the warranty.Temperature test, transport and storage informationGN ReSound Hearing Instruments are subjected to various tests in temperature and damp heating cycling between -25 C and +70C according to internal and industry standards.During transport or storage, the temperature should not exceed the limit values of -20C to 60C and rela-tive humidity of 90% RH, non condensing (for limited time). The air pressure between 500 and 1100 hPa is appropriate.400346000-GB-14.11-Rev.A.indd   52 10-12-2014   10:03:13
i53Be aware of information marked with the warning symbolWARNING points out a situation that could lead to serious injuries, CAUTION indicates a situation that could lead to minor and moderate injuriesi  Advice and tips on how to handle your hearing instrument better.Equipment includes RF transmitterPlease ask your local hearing care professional concerning disposal of your hearing instrumentReSound LiNX2 is compatible with iPhone 6, iPhone 6 Plus, iPhone 5s, iPhone 5c, iPhone 5, iPad Air 2, iPad Air, iPad (4th generation), iPad mini 3,  iPad mini 2, iPad mini with Retina display, iPad mini and iPod touch (5th generation) using iOS 7.X or later. Apple, the Apple logo, iPhone, iPad and iPod touch are trademarks of Apple Inc., registered in the U.S. and other countries.i “Made for iPhone” means that an electronic accessory has been designed to connect specifically to iPhone and has been certified by the developer to meet Apple performance standards. Apple is not responsible for the operation of this device or its compliance with safety and regulatory standards. Please note that the use of this accessory with iPhone may affect wireless performance.400346000-GB-14.11-Rev.A.indd   53 10-12-2014   10:03:14
54400346000-GB-14.11-Rev.A.indd   54 10-12-2014   10:03:14
55400346000-GB-14.11-Rev.A.indd   55 10-12-2014   10:03:14
56400346000-GB-14.11-Rev.AWorldwide HeadquartersReSound A/SLautrupbjerg 7DK-2750 Ballerup, DenmarkTel.: +45 45 75 11 11Fax: +45 45 75 11 19www.resound.comCVR no. 55082715United KingdomGN ReSound Ltd.Kirtlington Business CentrePortwayKirtlingtonOxon OX5 3JATel.: +44 1869 352 800Fax: +44 1869 343 466www.gnresound.co.ukAustraliaGN ReSound Pty. Ltd.Unit R1 Regents Park Estate391 Park RoadRegents Park NSW 2143Tel.: (free) 1800 658 955Fax: +61 2 9743 7472www.gnresound.com.auNew ZealandGN ReSound (NZ) Ltd.Ground Floor, North Entrance4 Fred Thomas DriveTakapunaAuckland, 0622Tel.: (free) 0800 900 126Fax: (free) 0800 007 695www.gnresound.co.nzAny issues relating to the EU Medical Device Directive 93/42/EEC, or Council Directive 1999/5/EC on Radio Equipment and Telecommunications terminal equipment should be directed to ReSound A/S400346000-GB-14.11-Rev.A.indd   56 10-12-2014   10:03:14