Ge Appliances Ion Exchange Column Mono Q 5 50 Gl Users Manual [PDF] マニュアル 5/50 GL, S

Mono Q 550 GL to the manual 39da0379-eced-4523-86f8-6463f6ec5131

2015-01-23

: Ge-Appliances Ge-Appliances-Ion-Exchange-Column-Mono-Q-5-50-Gl-Users-Manual-253360 ge-appliances-ion-exchange-column-mono-q-5-50-gl-users-manual-253360 ge-appliances pdf

Open the PDF directly: View PDF PDF.
Page Count: 2

DownloadGe-Appliances Ge-Appliances-Ion-Exchange-Column-Mono-Q-5-50-Gl-Users-Manual- [PDF] マニュアル Mono Q 5/50 GL, S GL  Ge-appliances-ion-exchange-column-mono-q-5-50-gl-users-manual
Open PDF In BrowserView PDF
GE Healthcare

First-time use

Sample recommendations

Equilibrate the column for first-time use or after long-term storage as follows:

Instructions 71-5017-88 AC

a)

Ion Exchange Columns

Mono Q 5/50 GL and
Mono S 5/50 GL

5 column volumes (CV) distilled water at 1 ml/min at room temperature.

b)

5 CV start buffer at 2 ml/min at room temperature.

c)

5 CV elution buffer at 2 ml/min at room temperature.

d)

5 CV start buffer at 2 ml/min at room temperature.

Try these conditions first
Elution buffer (Mono Q)*: 20 mM Tris-HCl + 1.0 M NaCl, pH 8.0

TM

Users of ÄKTA design system may select one of the buffer recipes recommended for anion
exchange chromatography at pH 8 or cation exchange chromatography at pH 6.

Flow: 2 ml/min at room temperature
1.

Equilibrate column with 5–10 column volumes (CV) of start buffer or until
baseline, eluent pH and conductivity are stable.

2.

Adjust the sample to the chosen starting pH and ionic strength and apply to
the column (see sample recommendations).

The columns are supplied with two union M6 female/1/16” male for connection
TM
to FPLC System, two fingertight connector 1/16” for connecting 1/16” tubing
to column and ÄKTA, two stop plugs 1/16” male to seal the column (attached to
column when delivered) and instruction.

3.

Wash with 5–10 CV of start buffer or until the baseline, the eluent pH and the
conductivity are stable i.e. when all unbound material has washed through
the column.

4.

Begin elution using a gradient volume of 10–20 CV and an increasing ionic
strengt up to 0.5 M NaCl (50% elution buffer).

Column data

5.

Wash with 2–5 CV of 1 M NaCl (100% elution buffer) to elute any remaining
ionically-bound material.

TM

TM

Mono Q™ 5/50 GL and Mono S 5/50 GL are Tricorn high performance columns.
The columns are pre-packed glass columns for high performance ion exchange
chromatography of proteins, peptides, polynucleotides and other biomolecules.

Matrix

Polystyrene/divinyl benzene

Bead form

Rigid, spherical, porous monodisperse

Particle size

10 μm

Column dimensions

5 × 50 mm

Bed volume

1 ml

Average loading capacity

50 mg

6.

2–12
1–14

▼

To avoid local disturbances in pH caused by buffering ions participating in the ion
exchange process, select an eluent with buffering ions of the same charge as the
substituent groups on the ion exchanger.
Choose the start buffer pH so that substances to be bound to the ion exchanger
are charged, e.g. at least 1 pH unit above the isoelectric point for anion exchangers
and at least 1 pH unit below the isoelectric point for cation exchangers. Figure 2
and Figure 3 list a selection of standard aqueous buffers.
pH 4

6

5

7

8

N-methyl piperazine
Piperazine
bis-Tris

Requilibrate with at least 5–10 CV of start buff er or until eluent pH and
conductivity reach the required values.

All commonly used aqueous buffers, pH 2–12
Urea, up to 8 M
Guanidine hydrochloride, up to 6 M
Acetonitrile, up to 30% in aqueous buffers
Non-ionic detergents
Cationic detergents (Mono Q)
Anionic detergents (Mono S)

Flow rate (water at room temperature)
recommended
0.5–3.0 ml/min
maximum
3 ml/min
Pressure over column
maximum

4 MPa, 40 bar, 580 psi

Type of exchanger

Mono Q
Strong anion
+

Charged group

-CH2-N (CH3)3

Ionic capacity

0.27–0.37 mmol
Cl /ml medium

Recommended to have an on-line filter upstream of the injection valve. Buffers
and solvents with increased viscosity will affect the back-pressure and flow rate.
De-gas and filter all solutions through a 0.22 μm filter.

Daily use

4 to 40 ºC

-

Choice of eluent

Buffers and solvent resistance

Temperature
operating

Dissolve the sample in start buffer,
filter through a 0.22 μm filter or
centrifuge at 10 000 × g for 10 min

Read the section ”Optimization” for information about how to optimize a
separation.

(will vary depending on sample and loading conditions)

pH stability
regular use
cleaning

Preparation

The column is delivered in degassed 20% ethanol sealed with two stop plugs to
prevent the column from drying out. For column storage, wash with 5 column
volumes of distilled water followed by 5 column volumes of 20% ethanol. Degas
the ethanol/water mixture thoroughly and apply at a low flow rate to avoid overpressuring the column. Store at room temperature or, for long periods, store at +4°
C to +8 °C. Ensure that the column is sealed well to avoid drying out. Do not freeze.

Separation by gradient elution

Quick information

≤ 45 mg

Delivery/storage

20 mM 2-[N-morpholino] ethanesulphonic acid (MES), pH 6.0

Elution buffer (Mono S)*: 20 mM MES + 1.0 M NaCl, pH 6.0
*

negative (Mono Q), positive (Mono S)

Recommended initial sample load

In-depth information

Start buffer (Mono Q)*: 20 mM Tris-HCl, pH 8.0
Start buffer (Mono S)*:

Net charge of target molecule

Mono S
Strong cation
-CH2-SO3

-

Cleaning

0.12–0.15 mmol
+

H /ml medium

Acetonitrile, up to 100%
Sodium hydroxide, up to 2 M
Ethanol, up to 100%
Methanol, up to 100%
Acetic acid, up to 75%
Isopropanol, up to 100%
Hydrochloric acid, up to 1 M
1% Trifluoroacetic acid

Note: Before connecting the column to a chromatography system, start the pump and remove all air
and debris in the system, particularly in the tubing and valves.

Avoid:
Oxidizing agents
Anionic detergents (Mono Q)
Cationic detergents (Mono S)

Fig 1. Illustration of how to lock the upper adapter. The locking ring (black) must be
in down position to prevent uncontrolled adjustment of the column’s bed height.

Tricorn

™

9

10

pKa
(25 ˚C)

11

4.75
5.33
6.48
6.65; 9.10
bis-Trispropane
7.76
Triethhanolamine
8.07
Tris
8.52
N-methyldiethanolamine
8.88
Propane-1,3-diamino
Ethanolamine
9.50
Piperazine
9.73
Propane-1,3-diamino 10.55
Piperidine
11.12

Fig 2. Recommended buffers for anion exchange chromatography.

▼

▼
pH 2.5

3

4

5

6

7

8

9

Citric acid

pKa
(25 °C)
3.13

Lactic acid
Butanedioic acid
Acetic acid

3.86
4.21
4.75
5.76

Methyl Malonic acid
MES
Phosphate

6.27
7.20
7.56

HEPES
BICINE

8.33

Fig 3. Recommended buffers for cation exchange chromatography.

Table 1 lists suggested volatile buffers that can be used in cases where the purified
substance has to be freeze-dried.
Table 1. Volatile buffer systems.

pH

Substance

3.3–4.3; 4.8–5.8

Pyridine/formic acid

3.3–4.3; 9.3–10.3

Trimethylamine/formic acid

4.3–5.8

Pyridine/acetic acid

3.3–4.3; 8.8–9.8

Ammonia/formic acid

4.3–5.3; 8.8–9.8

Ammonia/acetic acid

5.9–6.9; 9.3–10.3

Trimethylamine/carbonate

5.9–6.9; 8.8–9.8

Ammonium carbonate/ammonia

4.3–5.3; 7.2–8.2

N-ethylmorpholine/acetate

Optimization

Troubleshooting

Perform a first run as described in the section “Try these conditions first”. If the
results obtained are unsatisfactory, consider the following:

Symptom

Remedy

Increased back-pressure
over the column

Reverse the flow direction and pump 5 ml elution
buffer at a flow rate of 0.5 ml/min through the
column. Return to normal flow direction and run for
5 minutes at 1 ml/min through the column. If high
backpressure persists, clean the column.

Action

Effect

Change pH/buffer salt (see
Figure 1 and Figure 2 for buffers)

Changes selectivity, gives
weaker/stronger binding.

Change salt, counter ions
and/or co-ions

Changes selectivity.

Function test of Mono S 5/50 GL

Clean the column according to the procedure
described in the section “More rigorous cleaning”.

mAU (280 nm)

Reverse the flow direction and pump 10 ml well degassed start buffer through the column at a flow
rate of 0.5 ml/min.

200

Decrease the sample load

Improves resolution.

Decrease the flow rate

Improves resolution.

Air in the column

Change gradient slope

Shallower gradients improve selectivity but
broaden peaks (decrease efficiency).
A steeper gradient will sharpen peaks, but
move them closer together.

Sample:

Regular cleaning
Sample volume:
Gradient:
Start buffer:
Elution buffer:
Flow rate:

Flow: 0.5 ml/min at room temperature

3.

Wash with at least 2 CV of 2 M NaCl

4.

Rinse with at least 2CV of distilled water until the UV-baseline and the eluent
pH are stable.

4.

Wash with at least 4 CV of start buffer or storage buffer until pH and
conductivity values have reached the required values.

100

50

Function test of Mono Q 5/50 GL

It is recommended to reverse the direction of flow during column cleaning so that
contaminants do not need to pass through the entire length of the column.

Wash with 2 column volumes (CV) of 2 M NaCl.

1. Conalbumin (3 mg/ml)
2. α-lactalbumin, bovine milk (4 mg/ml)
3. Soybean trypsin inhibitor (6 mg/ml)
200 μl
0–100% elution buffer in 20 CV
20 mM Tris-HCl, pH 7.0
20 mM Tris-HCl + 0.25 M NaCl, pH 7.0
1.0 ml/min (room temperature)

mAU (280 nm)

0

Depending on the nature of contaminant cleaning solution in the section ”Buffers
and solvent resistance” may be appropriate. After cleaning the column wash
with at least 2 CV of distilled water and 4 CV of start buffer or storage buffer. For
more information on how to clean your column, please refer to the handbook ”Ion
exchange chromatography & Chromatofocusing, Principles & Methods”.
As an alternative to more rigorous cleaning or if column performance still not
restored change the filter at the top of the column. (Since contaminants are
introduced with the liquid flow, many of them are caught by the filter.) Instructions
for changing the filter are supplied with the Filter Kit. Clean the column after filter
change according to regular cleaning.

5.0

10.0

15.0

20.0

ml

Fig 5. Typical chromatograms from a function test of Mono S 5/50 GL.

Ordering information

% Elution Buffer
100

Product

200

No. per pack

Code No.

Mono Q 5/50 GL

1

17-5166-01

Mono S 5/50 GL

1

17-5168-01

No. per pack

Code No.

Related products

150

▼

0
0.0

More rigorous cleaning
Remove strongly hydrophobically bound proteins, lipoproteins and lipids by
washing with 4 column volumes (CV) of 30% isopropanol or 70% ethanol at
0.25 ml/min. Remove precipitated proteins with 1 CV of 1 mg/ml pepsin in
0.5 M NaCl, 0.1 M acetic acid (leave overnight) or wash with 2 CV of 6 M Guanidine
hydrochloride at 0.25 ml/min.

% Elution Buffer
100

150

Check the performance of the column when new by running the separation
described in Figures 4 and 5. Figures 4 and 5 shows a typical chromatogram run
on an optimized system. Since the system can profoundly affect the resolution,
it is meaningful to compare runs done on the same system. Check the column at
regular intervals and whenever you suspect a problem.

Cleaning

Wash with 4 CV of 1 M NaOH

1. Ribonuclease A (1.5 mg/ml)
2. Cytochrome C (0.4 mg/ml)
3. Lysozyme (0.4 mg/ml)
100 μl
0–100% elution buffer in 20 CV
20 mM sodium phosphate, pH 6.8
20 mM sodium phosphate + 0.4 M NaCl, pH 6.8
1.0 ml/min (room temperature)

Column performance control

For more information, please refer to the handbook “Ion exchange
chromatography, Principles & Methods”, which can be ordered from GE Healthcare
or downloaded from our web site.

2.

Sample volume:
Gradient:
Start buffer:
Elution buffer:
Flow rate:

Loss of resolution and/or
decreased sample recovery

1.

Sample:

Product

▼

100

Mono Q 10/100 GL

1

17-5167-01

Mono Q 4.6/100 PE

1

17-5179-01

Mono S 10/100 GL

1

17-5169-01

Mono S 4.6/100 PE

50

TM

HiTrap Desalting

1

17-5180-01

5 × 5 ml

17-1408-01

▼

Accessories

0

0
0.0

5.0

10.0

15.0

20.0

ml

Fig 4. Typical chromatograms from a function test of Mono Q 5/50 GL.

Product

No. per pack

Code No.

Tubing connectors:
Fingertight connector 1/16” male
Tricorn 5 filter kit*
Filter Tool
Union M6 female/1/16” male
On-line filter (1/16”)

10
1
1
8
1

18-1112-55
18-1153-02
18-1153-20
18-1112-58
18-1118-01

Handbook:
Ion Exchange Chromatography & Chromatofocusing,
Principles and Methods

1

11-0004-21

www.gehealthcare.com/protein-purification
www.gehealthcare.com

GE Healthcare Europe GmbH
Munzinger Strasse 5, D-79111 Freiburg, Germany

Tricorn, Mono Q, Mono S, HiTrap, ÄKTA and Drop Design are trademarks of GE Healthcare
companies. GE and GE Monogram are trademarks of General Electric Company.

GE Healthcare UK Ltd
Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK

The Tricorn column and components are protected by US design patents USD500856, USD506261,
USD500555, USD495060 and their equivalents in other countries.

GE Healthcare Bio-Sciences AB
Björkgatan 30
751 84 Uppsala
Sweden

GE Healthcare Bio-Sciences Corp
800 Centennial Avenue, P.O. Box 1327, Piscataway, NJ 08855-1327, USA

All goods and services are sold subject to the terms and conditions of sale of the company within
GE Healthcare which supplies them. General Electric Company reserves the right, subject to any
regulatory and contractual approval, if required, to make changes in specifications and features
shown herein, or discontinue the product described at any time without notice or obligation.

GE Healthcare Bio-Sciences KK
Sanken Bldg, 3-25-1, Hyakunincho, Shinjuku-ku, Tokyo 169-0073, Japan

© 2006 General Electric Company – All rights reserved.
GE Healthcare Bio-Sciences AB, a General Electric Company.

71-5017-88 AC 03/2006

Elanders Östervåla 2006

* includes top and bottom filters and O-rings, 5 of each.



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Encryption                      : Standard V2.3 (128-bit)
User Access                     : Print, Copy, Extract, Print high-res
Page Count                      : 2
Page Mode                       : UseOutlines
Page Layout                     : SinglePage
XMP Toolkit                     : XMP toolkit 2.9.1-13, framework 1.6
About                           : uuid:46cccdde-9fd4-407c-b93c-a9c662a33778
Producer                        : Acrobat Distiller 7.0.5 (Windows)
Creator Tool                    : Adobe InDesign CS (3.0.1)
Modify Date                     : 2007:06:12 11:13:38+09:00
Create Date                     : 2006:03:30 13:38:20+02:00
Metadata Date                   : 2007:06:12 11:13:38+09:00
Document ID                     : uuid:caaf67c2-c686-4527-b40e-bced1498321e
Instance ID                     : uuid:5cd3c95f-8137-42f6-88d5-29a9848f6ef1
Format                          : application/pdf
Title                           : [PDF] マニュアル Mono Q 5/50 GL, Mono S 5/50 GL
Creator                         : 
Author                          : 
EXIF Metadata provided by EXIF.tools

Navigation menu