Fall Sensor – FS319.5GE
This fall sensor is designed for a user to push the button for help at any time. It can also
automatically activate an emergency alarm if it detects a fall in the event that the user
is unable to trigger an alarm by pressing the button.
A. Identifying the Parts
Length Adjustment
1. Pendant Button
Pressing the Pendant Button for 0.5 sec will activate the Control Panel, causing it to
dial an emergency call or alarm (CID event code: 101).
2. LED
LED Light
LED Off
Standby Mode
3 Red Flashes
Powering On
1 Green Flash
Transmitting with Good Battery
1 Red Flash
Transmitting with Low Battery
Emergency
Button
Breakaway Clasp
3. Lanyard
When wearing the fall sensor, it is recommended to adjust the lanyard so the fall
sensor is resting at the center of the breastplate.
4. Battery Compartment Cover
B. Low Battery Detection and Supervision
Fall sensor features Auto Low Battery detection and Supervision.
When a battery is detected to be below the low battery threshold, fall sensor
will automatically transmit a Low Battery detection & Supervisory code to detect
Low Battery every 24 hours. Once the Control Panel receives the Low Battery
Note: fall sensors do not detect 100% of falls. If you are able to do so, you should
always press the button if you need help.
CR2477
Battery
Battery
Compartment Cover
detection & Supervisory code, it will then notify the user accordingly.
Auto Low Battery Detection & Supervision functions can only be activated as a pair.
Once Auto Low Battery Detection & Supervision functions are activated, they cannot be deactivated later in any circumstances.
C. How to Learn In Fall sensor
Fall sensor has a unique numeric code called “ID code”. The ID code enables the Control Panel to identify the signal is transmitted from the fall
sensor. 1.Put the Control Panel into Device Learning Mode
2.Press the Button on fall sensor, a radio signal will be transmitted to the control panel.
3.Please refer to the user manual for your control panel for further information on learning-in the fall sensor to complete the process.
D. Battery Life
The fall sensor uses one 3V CR2477 lithium battery as its power source. The fall sensor will have a typical battery life of 2 years with an average of one activation per day.
When the battery voltage drops below a certain threshold, a Low Battery signal will be sent to the Control Panel to notify the user.
When the fall sensor is triggered in a low battery state, the red LED will flash once to indicate the battery needs to be replaced.
It is not recommended to learn-in the fall sensor to the control panel while in a low battery state.
E. Usage Recommendation
1.Wear the fall sensor on the outside of all articles of clothing; do not cover with any clothes during use.
2.Carefully place the fall sensor on a table or nightstand when you are not using it in order to avoid activating an alarm.
3.This device is rated IP45: protected from low pressure water jets in any direction.
F. Testing
During testing, do not activate the fall sensor twice within a 10-second interval.
FCC Statement
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1)This device may not cause harmful interference, and
(2)This device must accept any interference received, including interference that may cause undesired operation.
FCC Caution:
To assure continued compliance, any changes or modifications not expressly approved by the party responsible for compliance may void the user's authority to operate this equipment.

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 2
Language                        : en-US
Tagged PDF                      : Yes
Author                          : Windows User
Creator                         : Microsoft® Word 2010
Create Date                     : 2018:04:03 09:07:02-07:00
Modify Date                     : 2018:04:03 09:07:02-07:00
Producer                        : Microsoft® Word 2010
EXIF Metadata provided by EXIF.tools
FCC ID Filing: XO8-IC319FS-1

Navigation menu