Intromedic INTROMEDIC Capsule Endoscope Receiver User Manual

Intromedic Co., Ltd. Capsule Endoscope Receiver

Contents

User Manual

MiroCam User Manual                                                                1  한글판 버전 1.00   Contents 1 SAFETY.............................................................................................................................2 1.1 Warnings ................................................................2 1.2 Symbols for Safety ...................................................... 3 1.3 Function Symbols ........................................................ 4 1.4 Remarks for Safe Use .................................................... 6 2 Overview ..........................................................................................................................12 2.1 Description of MiroCam Capsule Endoscope System ........................ 12 2.2 System Overview ........................................................ 13 2.3 Feature of MiroCam Capsule Endoscope System ............................ 14 2.4 Function Explanation ................................................... 15 2.5 Intened Pupose and Side Effect ......................................... 18 3 Product.............................................................................................................................19 3.1 Pictures ............................................................... 19 3.2 Supply Items ........................................................... 21 4 Installation.......................................................................................................................22 4.1 Installation ........................................................... 22 5 Technical Data.................................................................................................................24 5.1 Overview ............................................................... 24 5.2 Hardware ............................................................... 24 5.3 Software ............................................................... 25 5.4 COMPLIANCE/APPROVALS ................................................... 25 6 Environment condition for storage and transport. ........................................................27 6.1 Safety environment condition for storage ............................... 27 6.2 Environment condition for transport .................................... 27 7 Troubleshooting...............................................................................................................28 7.1 Introduction ........................................................... 28 7.2 Who should perform repairs ............................................. 28 7.3 Obtain replacement parts ............................................... 28 7.4 Troubleshooting guide .................................................. 28
MiroCam User Manual                                                                2 1 SAFETY  1.1 Warnings  The MiroCam® has been manufactured to conform to the International Standard for Medical Electrical Equipment: General Requirements for Safety IEC 60601-1, together with the Collateral Standard for Electromagnetic Compatibility Requirement and Tests IEC 60601-1-2. The MiroCam® has been manufactured to conform Medical Equipment with respect to electric shock fire and mechanical hazard only CAN/CSA C22.2 NO.601.1.  To guarantee safety and reliability of the equipment, special attention should be paid to all the precautions directed throughout this manual, such as " WARNING", " CAUTION", and " NOTE". Also, to prevent incorrect operations and a failure in maintenance, it is essential that you should be familiar with the functions, operation and maintenance instructions by thoroughly reading this manual before operating the equipment.  WARNING "WARNING" is used to alert the user of a potential serious hazard that may lead to death or serious injury of patients and personnel.   CAUTION "CAUTION' is used to indicate information the user should know to avoid a hazard that may lead to device malfunction, device failure, damage to the device, or damage to other property. NOTE "NOTE' is used to indicate important information on installation, operation, or maintenance which you should to know for the safe and effective use of the equipment.  한글판 버전 1.00
MiroCam User Manual                                                                3 1.2 Symbols for Safety  This section describes a set of symbols that IEC (The International Electrotechnical Commission) has established for medical electronic equipment to classify a connection or warning of any potential hazards.  IEC 348: Notice for the user to pay special attention to the following details   IEC 878-02-03: Indicates that this is classified into Type BF equipment  EN 980: Denotes Date of Manufacture EN 980: Denotes Date of Manufacture Address IEC60601, ANNEX D: Denotes “ON” status of main power switch IEC60601, ANNEX D: Denotes “OFF” status of main power switch SN  EN 980: Denotes serial number IEC 417-5031: Denotes DC (Direct Current) IEC 417-5032: Denotes AC (Alternating Current) A  Denotes Ampere, the unit of current V  Denotes Volt, the unit of Voltage Hz  Denote Herz, the unit of Frequency IEC 417-5021: Denotes potential equalization terminal Single Use Only The Available Period   한글판 버전 1.00
MiroCam User Manual                                                                4 1.3 Function Symbols 1.3.1  Apllication Function Symbols Symbol Description Symbol Description  Functions & Windows for Receiver Unit Functions & Windows for Diagnosis Report  Function & Windows for Patient Database Listing Function & Windows for Export Patient Image Data  Function & Windows for Image Display and Diagnosis Function & Windows for Backup Patient Image Data  Capturing current patient image to the thumbnail Landmark current time of patient image to the thumbnail  Single previous image     Single next image  Previous Selected image   Next Selected image  Display image continuously    Display image continuously reverse direction  Pause image display   Enlarge image display  Single image display mode   Dual image display mode  Quad image display mode   Display full image  Display selected image   Display red image  Grouping image display in   Thumbnail   Single image display in   Thumbnail  Minimize application software   Close application software  Mark circle in current image   Mark arrow in current image  Erase circle and arrow in current image    Select marker color   한글판 버전 1.00
MiroCam User Manual                                                                5 Symbol Description Symbol Description  Undo mark in current image   Redo mark in current image  Doctor login  Cancel login    Patient’s diagnosis history  Delete current patient information  Progress reporting current patient   Complete reporting current patient  Complete Export current patient     Complete backup current patient  Doctor logout  Display full patient list  1.3.2  Receiver Function Symbols Symbol Description SIG    Status of capsule signal receiving   Green : Good Yellow : Bad INI   Status of initialization of Receiver Unit Green : Good Yellow : Bad BAT   Status of Battery Green : Full charged   Yellow : Not charged     한글판 버전 1.00
MiroCam User Manual                                                                6  한글판 버전 1.00 1.4 Remarks for Safe Use    Follow these safety instructions and the clinically adopted precautions.   The manufacturer is not liable for harm or damage caused by improper, unauthorized, unprofessional or inexpert use of the device   IntroMedic.Co.Ltd., is NOT responsible for physical harm or equipment problems caused by the user’s careless operation or mismanagement.   The user MUST have read and understood the user's manual. ONLY trained and qualified persons may operate the system.   The user's manual must ALWAYS be with the equipment. It is the USER’S RESPONSIBILITY to ensure this!   CAUTION is advised that the foreign substances, ironware or fluids, such as water or disinfecting fluid, not enter the equipment.   ONLY authorized personnel may perform any type of repair. Never attempt to open the covers, panels or casings.   DO NOT crease, bend, fold or twist the probe cables and take care to guard them against mechanical stress (e.g. wheels or heels)!     The probes must not be exposed to mechanical shock (e.g. by dropping). Any damage caused that way invalidates the warranty.   CAUTION: Damage/injuries to the probe or cable may lead to safety hazard.  It is HIGHLY advised to have them repaired IMMEDIATELY.   DO NOT handle fluids in the vicinity of the system.   When using a cart purchased elsewhere, implement the latch guard method to prevent the wheels from rolling.   DO NOT USE in moist or damp places.     DO NOT operate the equipment with wet hands.   Avoid using the equipment in extreme temperatures or humid environments.   DO NOT keep the equipment or carry out the procedure in places such as areas exposed to direct sunlight, vicinity of heaters, vicinity of chemical materials or gases, areas moist/damp or dusty, or poorly ventilated areas.   DO NOT disassemble or open the equipment without permission. The free repair warranty will be automatically invalidated.     DO NOT carry out the procedure in areas with high vibrations or in environments where high electro-magnetic waves are generated.   DO NOT pull out the power cord by grabbing the cable. When disconnecting the power cord, grab the plug and then pull out.  This prevents short-circuits, disconnection, or cord damage.
MiroCam User Manual                                                                7  한글판 버전 1.00   CAUTION: Verify that the power voltage supplied from the power receptacle, matches with the voltage the system requires. Check Voltage and Frequency on the AC/DC adaptor.   CAUTION: Verify that all connection terminals are securely connected to the system.   CAUTION: Turn off the power switch before connecting the probe.   DO NOT discard probes, cables and connectors with general waste. Separately discard as industrial waste or medical waste.     CAUTION: Discard the battery according to the regulations of industrial waste.  DO NOT discard with general waste.     DO NOT carry out the procedure simultaneously with other procedures using medical products or equipment.   DO NOT carry out the procedure on patients with pacemakers and/or defibrillators.   DO NOT use for purposes other than medical treatment.   DO NOT connect the USB cable to the receiver while the receiver’s patch probes are still connected   DO NOT charge the receiver while the receiver’s patch probes are still connected.   Connect USB cable to receiver only after mounting it on charger.   DO NOT install any other programs onto the workstation utilized for review and diagnosis of patient image data.   DO NOT carry out the procedure on patients with known or suspected gastrointestinal tract obstruction, perforation, stricture or fistular.   DO NOT carry out the procedure on patients who have difficulty swallowing food or pills(dysphagia).   DO NOT carry out the procedure on patient who have difficulty communicating.   DO NOT carry out the procedure on patient with indigestion or delay in gastric emptying.   DO NOT carry out the procedure on patient who may be affected by electromagnetic radiation, such as pregnant women, infants, and patients with heart disease or epileptics.   DO NOT carry out the procedure on patients with diverticulosis in the Small Bowel.   DO NOT carry out the procedure on patients who are recommended against having the procedure by a Physician.   The capsule is disposable and should not be reused.
MiroCam User Manual                                                                8 1.4.1  Environmental Conditions for Operation  Temperature    : -10 ℃ - +70 ℃ (Operating : +10 ℃ - +40℃)   Relative humidity    : 10% - 80% (Operating : 45% - 75%)   Atmospheric pressure   : 700hPa to 1060hPa  WARNING DO NOT operate the equipment in the vicinity of generators, power stations, X-ray devices, and broadcasting stations where high levels of electro-magnetic waves are generated. The electro-magnetic waves can cause equipment malfunctions.   CAUTION If the equipment has been brought in from a cold environment (stock room, airfreight) into a warm room, initial activation should take place after a few hours, to allow for temperature adjustment and balance and evaporation of condensed humidity. WARNING DO NOT operate the equipment in the vicinity of heat sources, strong electric or magnetic fields (close to a transformer), or near instruments generating high-frequency signals.   WARNING -  Do not use this device when the output contains d.c. components. -  Do not use the unit with h.f. surgical equipment. It may result in burns at the site of the electrodes and possible damage to the capsule and receiver. -  Do not use the unit in close radius (within 1 m) of short wave or microwave therapy equipment. It may produce instability in the captured image. WARNING This device is a Class B device according to EN60601-1-2 standards. This equipment can cause radio interference in residential areas. In this case, the owner (or operator) can be held responsible to take appropriate measures or take proper measures for compensation.   한글판 버전 1.00
MiroCam User Manual                                                                9 1.4.2 Safety Precautions  CAUTION -  Make sure the environment is without interference from electromagnetic fields. -  Make sure the environment is without noise and vibration. -  DO NOT carry out the procedure while using other equipments, devices or products. -  The instruction for use of the probes MUST be observed. -  DO NOT use on patients with pacemakers or defibrillators. NOTE This device has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of U.S. FCC Rules. CAUTION EQUIPOTENTIAL BONDING: In hospital, doctors and patients are subjected to dangerous, uncontrollable compensating currents. These currents are due to the potential differences between connected equipment and touchable conducting parts as found in medical rooms. The safest solution to the problem is consistent equipotential bonding.    Medical equipment is connected with connecting leads made up with angle sockets to the equipotential bonding network in medical rooms. CAUTION DO NOT use the capsule if the package is unsealed. If capsule package is unsealed, it can be infected.    DO NOT reuse a used capsule.   To prevent unexpected accidents like fire or explosion, do not use it near or in the presence of inflammable or ignitable substances.   DO NOT disassemble the equipment case nor open the cover. In case of problems in equipment or need for other service, please contact our head office or a customer support staff immediately.   Only the accessories authorized and designed by our company should be used with this equipment. Faults resulting from the usage of unapproved or inappropriate accessories are not guaranteed against.   This equipment may have an effect on other products or be effected by them.   Follow your doctor’s instructions and abide by the guidelines in the user manual.   DO NOT try to upload the data while the patch probes are still connected to the receiver.   DO NOT charge the rechargeable battery in the receiver while the patch probes are still connected to the receiver.   Stay away from high frequency radiation sites (such as high voltage, radar, installation power plants, MRI, CT or electric blankets etc.) during your capsule endoscope procedure. (It may result in serious side effects requiring an emergency operation.)  한글판 버전 1.00
MiroCam User Manual                                                                10  한글판 버전 1.00   In case of any symptoms of abdominal pain, vomiting, fever, heart trouble, dizziness or seizure during or after the capsule endoscope procedure, please notify your doctor.   Always check the connection between the receiver and the patch probes.   Always check that the battery in the receiver is fully charged before use.   DO NOT use the capsule if the package is unsealed.     After ingesting the capsule, always check whether the capsule has been excreted.     Prior to undergoing the capsule endoscopy procedure, patients with diabetes must control their insulin doses.   For more accurate data and better analysis, patients can have solid food for lunch, but must have liquid food as dinner on the day before the procedure and need to fast for 12 hours before the procedure.   12 hours before the procedure, take a laxative such as Sodium phosphate or PEG solution.   DO NOT smoke for 12 hours prior to the capsule endoscope procedure.   DO NOT apply body lotion to your body before the procedure.   DO NOT bite the capsule before swallowing.   Avoid excessive physical activity during the capsule endoscope procedure.   2 hours after swallowing the capsule, patients may drink water, and after 4 hours, patients may have liquid food.   When undergoing the capsule endoscope procedure, DO NOT makes physical contact with another person undergoing this same procedure.   During operation of the receiver, DO NOT touches it or gets it wet.   Use the provided batteries only and never take it out of the receiver body during the procedure.   During upload of the data recorded in the receiver to the PC, avoid disconnecting the USB(It may damage the patient’s data).   Always confirm that the USB is connected by checking on the screen of the MiroCam application program.
MiroCam User Manual                                                                11 1.4.3  Cleaning and maintenance  1.4.3.1  System and accessories    Other units and accessories should be cleanly maintained. For cleaning, rub them lightly with a soft cloth wet with warm water at least once a week. Do not use organic solvents such as lacquer, thinner, ethylene and oxide because they can damage the equipment. Be careful that foreign substances do not enter the main system when cleaning. Please maintain the probe according to the disinfection instructions.   The vaginal manometry probe must not be cleaned with a wet towel before usage.    Use the vaginal probe ONLY after putting a disposable latex condom on it, and then discard only the condom after use.   The patch probe for abdominal attachment must not be rubbed or wiped.   ALWAYS operate the equipment under sanitary environmental conditions. DO NOT use heat or gas for the disinfection of the probes and cables.   CLEAN vaginal probes, anal probes, patch probes and vaginal manometry probe should ALWAYS be used.  1.4.3.2 Service documents  If required, or upon request, the supplier may provide block diagrams, list of spare parts, descriptions, adjustment instructions or other related information which may help qualified technical personnel in repairing specified parts of the equipment which have been defined repairable by the manufacturer.  1.4.3.3  Moving the Equipment    CAUTION when moving equipment.   WARNING: Excessive impact/shock causes internal damage.   If wiring is connected/disconnected when moving, check the exact wiring status after moving.   If damage to the equipment is found out after moving, immediately contact the company or dealer.   WARNING Before moving the system, always make sure to disconnect the monitor from the main system, and then safely move the main system and monitor separately. Connect the main system and monitor only after moving is finished and the systems structure is settled and stabilized.   한글판 버전 1.00
MiroCam User Manual                                                                12 2 Overview 2.1 Description of MiroCam Capsule Endoscope System MiroCam is Capsule Endoscope Systems and accessories to be used to diagnosis of patients.    It consists of the capsule and receiver. 2.1.1 Photos 2.1.1.1 Capsule  2.1.1.2 Receiver  .1.1.3 Accessory  2   한글판 버전 1.00
MiroCam User Manual                                                                13 2.2 System Overview   MiroCam Capsule Endoscope Systems consist of Capsule Receiver and Image Processing Workstation (Software).    Image SensorData Transmittion Analog Amplifier Analog FilterFPGADSP Data CommunicationImage ProcessingImage DisplayCapsule Receiver SoftwareMiroCam System Block Diagram2.2.1 Capsule MiroCam Capsule Endoscope System’ Capsule Unit consists of optical dome, white LED, image sensor, lens and gold courted cage. 2.2.2 Receiver MiroCam Capsule Endoscope System’s Receiver unit consist of 8 channel connector, differential amplifier, filter, data demodulator, data storage memory and USB communication channel. 2.2.3 IPW MiroCam Capsule Endoscope System’s Software is running on the Microsoft Windows XP using the IBM compatible PC platform and provide easy-to-use icon menu. It consist of Image reconstruction part and Image display part.     한글판 버전 1.00
MiroCam User Manual                                                                14  한글판 버전 1.00 2.3 Feature of MiroCam Capsule Endoscope System   2.3.1 Main Feature  Capsule - High Resolution Image (320 X 320) - Widest Field of view (150 degrees) - Smallest Capsule size (11 X 24mm) - Longest operation time  Receiver - Built-in circuit protection functions - Small, light and easy to carry around  IPW - User Friendly software feature - Red Color Detection - Color of images are virtually natural - Dual and quad image views  System - Over Intel P4 3.0GHz - Over 1G Byte Memory - Over 100GByte Hard Disk
MiroCam User Manual                                                                15 2.4 Function Explanation 2.4.1 Display flow  Report View Image Review 2.4.2  Function of Control and Display of Signal   All operations start after completing or selecting the patient registration on the default screen.   If select Patient in Patient List for data review and press   button, Review window is displayed.    Press    button for playing image data.  한글판 버전 1.00
MiroCam User Manual                                                                16  Press    button for stop playing image data.  Press    button for save current image to thumbnail.  Press    button for recording diagnosis data and printing paper.  Press   button for operation for receiver unit initialization and image data upload.  2.4.3  Sequence of Operation    Initialize Receiver Unit ① Connect USB Cable between Receiver unit with Workstation Æ ② Click   button Æ ③ Click   button ④ Click Input Patient Info button Æ ⑤ Input Patient information Æ ⑥ Click Save button   Upload Image Data  ① Connect USB Cable between Receiver unit with Workstation Æ ② Click   button Æ ③ Click   button    Display Image Data and Diagnosis ① Select Patient in Patient List for data review Æ ② Click   button Æ ③ Click   button    Print Diagnosis Data  ①  Select Patient in Patient List for data review Æ ② Click   button Æ ③ Input diagnosis result Æ ④ Insert diagnosis point in patient image data     한글판 버전 1.00
MiroCam User Manual                                                                17  한글판 버전 1.00 2.4.4  Connection and Disconnection Part Where  When  Method Patch Probe Connector of upper side of receiver unit    Before process capsule endoscope Connection : Hold the connector on the end of the receiver’s patch probe and put it into the socket of the top of the patch.   Disconnection : Hold the connector on the end of the receiver’s patch probe and pull it out by hands.. USB Cable USB Connector of left side of receiver unit  After process capsule endoscope Connection : Hold the USB cable connector and push it into the USB connector on the side of the receiver. Disconnection : Hold the connector on the tip of the USB cable and pull it out by hands.     Cradle  Lower side of receiver unit  After process capsule endoscope Connection : Place down the receiver into the Cradle as its slot. . Disconnection : Pull out the receiver from the Cradle vertically. Adaptor  Rear side of cradle System Installiation Connection : Put the DC cable of the charger into the DC jack of the Cradle.. Disconnection : Pull out the connector of the DC cable of the charger by hands.
MiroCam User Manual                                                                18  한글판 버전 1.00 2.5 Intened Pupose and Side Effect 2.5.1 Intended Purpose  MiroCam  is intended for the detection of pathologies as an adjunctive tool in the diagnosis of small bowel gastrointestinal disorders and diseases.
MiroCam User Manual                                                                19 3 Product 3.1 Pictures 3.1.1 Main Device  Capsule    Receiver    한글판 버전 1.00
MiroCam User Manual                                                                20 3.1.2 Accessories  Patch Cable - Model : 21-1-19-1, 10P - Manufacturer : Mednis  USB Cable  - Model : AV22201-06 - Manufacturer : BELKIN  Cradle - Model : MR1000-C - Manufacturer : JEC Korea  Adaptor - Model : JMW128KAXXXXNXX(Compliance with requirements of IEC601-1) - Manufacturer : JEC Korea    한글판 버전 1.00
MiroCam User Manual                                                                21  한글판 버전 1.00 3.2 Supply Items 3.2.1  Basic Supply Item No.  Model  MiroCam Product CodeQuantity Description  Replace Period  1  Capsule  MC1000-C  1 ea    - 2  Receiver  MR1000-R  1 ea    - 3  Cradle  MR1000-C  1 ea    - 4  Adaptor  MR1000-A  1 ea    - 5  Power Code  MR1000-P  1 ea    - 6  Patch Probe  MR1000-S  1 pk    Single use 7  Ground Pad  MR1000-E  1 ea    Single use 8  Patch Cable  MR1000-D  1 ea    6 month 9  Software  MW1000-S  1 ea    - 10  Sack  MR1000-K  1 ea    - 11  PC   MW1000-P  1 ea    - 12  Monitor  MW1000-L  1 ea    - 13  Printer  MW1000-I  1 ea    - 14  Keyboard  MW1000-K  1 ea    - 15  Mouse  MW1000-M  1 ea    - 16  Back  MR1000-G  1 ea    - 17  Cart  MW1000-C  1 ea    - 18  User Manual  MW1000-U  1 ea    -  3.2.2  Optional Supply Item No.  Model  MiroCam Product Code Quantity Description  Replace Period 1  Battery  MR1000-B  1 ea    6 month2           3           4
MiroCam User Manual                                                                22 4 Installation  4.1 Installation 4.1.1  Specification of packing  MiroCam Capsule Endoscope System Box  MiroCam Capsule Endoscope Receiver Set Box  5210.00100.00100.00REMARK---1Q'TY.1/1BoxReceiver BoxMAT'LPulp1No.ITEMN/S MODELBARCODE LABELMiroCamMADE IN KOREA55480.00340.00190.00REMARK---1Q'TY.1/1BoxMAT'LPulp1No.ITEMN/S MODELBARCODE LABELMiroCamMADE IN KOREA5 한글판 버전 1.00
MiroCam User Manual                                                                23  MiroCam Capsule Endoscope Battery Charger Box  MiroCam Capsule Endoscope Accessory Box  5MiroCamMADE IN KOREA350.00150.00150.00REMARK---1Q'TY.1/1BoxCharger BoxMAT'LPulp1No.ITEMN/S MODELBARCODE LABEL55MiroCamMADE IN KOREAREMARK---1Q'TY.1/1BoxAccessory BoxMAT'LPulp1No.ITEMN/S MODELBARCODE LABEL640.00460.00140.005 한글판 버전 1.00
MiroCam User Manual                                                                24  한글판 버전 1.00 5 Technical Data 5.1 Overview Product Name : MiroCam Capsule Endoscope System Model Name : MiroCam Serial Number : Refer to the label which is attaching in the reverse side of the product. 5.2 Hardware 5.2.1 Capsule  Size : 10.8 X 24mm  Weight : 3.25g  Material : Human Compliance Plastic  Light : 6 white LED  View Angle : 150°(in Image)  View Depth : 3 cm  Enlargement Ratio : 1:8  Detectable Range : under 0.1mm  Sampling Ratio : 2.9 fps  Working time : Over 11 Hours  Mechanical Safety : Compatible ISO60601-1-1  Chemical Safety : Safe in pH=2 ~ pH=8  Battery Type : Silver Oxide Cell(3Vdc, 70mA)  Operation Temperature : 20 ~ 40℃  Storage Temperature : 0 ~ 50℃ 5.2.2 Receiver  Operation System : Firmware  Recording Time : 11 Hours  Weight : 350g, include battery  Operation Voltage : 3.7Vdc, 0.45A  Battery Type: Lithium Ion Battery (3.7Vdc, 8.8A)  Battery Weight : 215g  Operation Temperature : 0 ~ 40℃  Storage Temperature : 0 ~ 55℃  Category : Type BF 5.2.3 Image Workstation  Operating System : Windows XP Professional
MiroCam User Manual                                                                25  한글판 버전 1.00  CPU : Core 2 Duo E6300(1.86GHz/2M)  Memory : DDR II 1GByte(667MHz)  Display Adaptor : Geforce 7600GT 256MB  Hard Disk : SATA II 160GB, SATA II 320GB  ODD : DVD-RW  Monitor Resolution : 1280 X 1024  Monitor Contrast : 700 : 1  Printer Resolution : 4800 X 1200 dpi  Printer Paper : A4 5.2.4 Cradle  Input Voltage : 110~220VAC  Input Current : 3A  Output Voltage ; 4.2VDC  Output Current : 4A  Operation Display : LED Display 5.3 Software 5.3.1 Version  MiroCam Application Software Version 1.00 5.3.2 Specification  Language : English  Data Export : JPEG Image, AVI Video Clip, PDF Data Report  Data Display : Single or Multi Image, Time Bar, Color Bar, Diagnosis Data  Event Marker : Small Image with Explanation   Display Ratio : 5 ~ 30 fps  Display Mode : Single View, Dual View, Quad View  Running Mode : Normal Mode, Fast Mode 5.3.3 Feature   Color status Display  User Friendly software feature  Automated detection of GI tract bleeding  Color of images are virtually natural 5.4 COMPLIANCE/APPROVALS The MiroCam capsule endoscope system and accessories complies with the Medical Device Directive 93/42/EEC (CE0843).
MiroCam User Manual                                                                26  한글판 버전 1.00 In addition, the product complies with   IEC 60601-1:1988 + A1:1991 + A2:1995 (EN 60601-1:1990 + A1:1992 + A2:1995)  EN 60601-1-2  EN 10993-1  EN 10993-10  The possibility of hazards arising from software errors was minimized in compliance with EN1441 and EN60601-1-4.  5.4.1 FCC STATEMENT    This device complies with Part 15 of the FCC Rules.   Operation is subject to the following two conditions:    (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.   CAUTION : Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.  NOTE : This equipment has been tested and found to comply with the limit for a Cass B digital device, pursuant to Part 15 of the FCC Rules. These limits  are  designed  to  provide  reasonable  protection  against  harmful interference  in  a  residential  installation.  This  equipment  generates, uses and can radiate  radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur ina particular installation, which can be determined by turning the equipment off and on, the user is encourage to try to correct the interference by one or more of the following measures:  - Reorient or relocate the receiving antenna. - Increase the separation between the equipment and receiver. - Connect the equipment into an outlet on a circuit different. from that to which the receiver is connected. - Consult the dealer or an experienced radio/TV technician for help
MiroCam User Manual                                                                27  한글판 버전 1.00 6  Environment condition for storage and transport. 6.1 Safety environment condition for storage 6.1.1 Receiver unit  Environment condition for storage - Temperature     : -10℃ - +70℃  - Relative humidity   : 10% - 80% - Atmospheric pressure   : 700hPa to 1060hPa  Keep in the place which is not contacted with water.  Keep in the place where the direct light dose not shines.  Keep  in  the  place  which  is  not  contacted  with  something  that  is a hazardous to the human body.  Keep in the place which the kid’s hand dose not closes.  Don’t keep it in the (storing) place for chemicals or near gas.  When you clean up the outside of the product, only use gauze with water. Other liquid like alcohol should not be allowed.  When you charge the battery, please use the cradle.  If you don’t use this product for so long, please separate the battery from the body of it. 6.1.2 Image Workstation  Treat this based upon the way how to handle a general PC.  6.2 Environment condition for transport  Temperature     : -10℃ - +70℃   Relative humidity     : 10% - 80%  Atmospheric pressure   : 700hPa to 1060hPa
MiroCam User Manual                                                                28  한글판 버전 1.00 7 Troubleshooting 7.1 Introduction This section explains how to troubleshoot the MiroCam capsule endoscope system if problems arise. Tables are supplied that list of possible difficulties, probable cause, and recommended actions to correct the difficulties. 7.2 Who should perform repairs Only qualified service personnel should open the HMT2100 housing, remove and replace components, or make adjustments. 7.3 Obtain replacement parts IntroMedic Technical Services provides technical assistance information and replacement parts. To obtain replacement parts, contact IntroMedic Co.,Ltd. or your local representative. Refer to part names and part numbers listed on Section 3.2, Supply Items. 7.4 Troubleshooting guide Condition Recommended Action 1. The MiroCam capsule endoscope system workstation fails to power-up when the I/O power switch is on. 1. Ensure that the MiroCam capsule endoscope workstation main system is plugged into an operational AC outlet in accordance with the input specification rated on the side panel of the workstation main system. 2.  If the condition persists, contact IntroMedic Co.,Ltd.. 2. The MiroCam capsule endoscope system receiver unit fails to power-up when the I/O power switch is on. 1.  Recharging battery of receiver unit. 2.  Replacement battery of receiver unit. 3.  If the condition persists, contact IntroMedic Co.,Ltd.. 3. It's not blinking when it's out of the case. 1.  DO NOT use this capsule and retry another capsule. 2.  Contact IntroMedic Co.,Ltd. for replacement capsule.   4. INI LED, on receiver unit, is lighted as yellow. 1.  Initialization receiver unit in MiroCam application software. 2.  If the condition persists, contact IntroMedic Co.,Ltd.. 5. BAT LED, on receiver unit, is lighted as yellow. 1.  Recharging battery of receiver unit. 2.  Replacement battery of receiver unit. 3.  If the condition persists, contact IntroMedic Co.,Ltd..

Navigation menu