Jawon Medical ECHOH Wrist Type NIBP User Manual ECHO H Wrist

Jawon Medical Co., Ltd. Wrist Type NIBP ECHO H Wrist

Users Manual

 1  WARNING  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙∙···········∙∙∙·∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙…………········· 3 ABOUT BLOOD PRESSURE  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙∙··············∙∙∙·∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙…………······∙∙∙··· 4 1. What is blood pressure?  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙∙………·········∙∙∙∙∙∙∙∙∙∙∙………………∙∙∙∙∙  4 2. What makes my blood pressure change?  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙··········∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙·····∙∙∙∙∙∙…………………∙·· 4 3. Classification of the blood pressure?  ∙∙∙∙······∙∙···∙∙∙∙∙∙∙∙∙·······∙∙………………∙∙…∙……………∙∙…·4 4. Why do I have to check blood pressure at home?  ∙∙∙∙∙∙·∙··················……………………… 5   5. What is benefit of ECHO-H the data transmissible blood pressure monitor use?……… 6 6. What is the benefit of ECHO-H date and time recorded monitor use? ……………………6 7. What is the benefit of ECHO-H data transmissible wrist blood pressure monitor use?…6 INTRODUCTION  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙·················∙∙∙·················…………··……  7 WORD DEFINITION  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·····························∙∙∙∙∙∙∙······∙∙∙∙∙∙∙∙∙∙∙∙··············…………·····…… 8 NOTE TO USER  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·············∙∙∙∙∙∙∙∙∙∙∙∙∙∙······················∙∙∙∙····················…………··……  9 CLASSIFICATION & COMPLIANCE WITH STANDARD    ∙∙∙∙∙∙∙∙∙∙·················∙∙······…………·····  9 SAFETY PRECAUTIONS  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙………·····… 10 SAFETY SYMBOLS  ∙∙∙∙······················∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙·…………····…·…… 11 PART NAMES & FUNCTIONS  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙················∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙…………∙∙∙∙·……12 1. Main unit and accessories∙∙∙∙∙∙∙∙∙∙∙∙∙∙··················∙∙∙∙∙∙∙······················……………………… 12 2. Symbols in LCD indicator  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙······∙∙∙∙∙∙∙∙∙∙∙∙∙∙····························…………······……… 13 PRECAUTIONS FOR MEASUREMENT  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙········∙∙∙∙∙∙∙∙∙∙·····················…………·······……14 BATTERY INSERTION/REPLACEMENT········∙································…………······………… 16 1. How to insert the batteries  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·············∙∙∙∙∙∙·······························………··…·…… 16 2. Replacement of the batteries  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙··················∙∙∙∙∙∙∙∙∙∙∙···················………··……·… 16 SETTING THE DATE AND TIME  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙················∙∙∙∙······························…………·····…17 1. Entering the mode to set date and time ·············∙∙∙∙∙∙∙∙∙∙∙∙∙∙·················∙∙∙∙∙∙∙∙……∙…∙…… 17 2. Setting up  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙∙···∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙·∙················∙∙∙∙∙∙∙∙……∙…………17 SOUND ON/OFF  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙···∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙·∙·····················∙∙∙∙∙∙………………………………………19 SET INFLATION LEVEL……………………………………………………………………………… 19 DATA TRANSMISSION·············∙∙∙∙∙∙∙∙···∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙·∙·····················∙∙∙∙∙∙∙…………∙……………20 MEASUREMENT  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙···········∙∙∙∙∙∙∙∙∙∙∙∙∙∙······················∙∙∙∙∙∙····················……………·……22 1. Wrapping the cuff  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙········…∙∙∙∙∙∙∙∙∙∙···········………·…………22 2. Correct measuring posture  ∙∙∙∙∙∙∙∙∙∙∙∙∙····················∙∙∙∙∙∙∙·······················……………·…… 22 3. Measuring method    ∙∙∙∙∙∙∙∙∙∙∙∙∙∙············∙∙∙∙∙∙∙∙∙∙∙∙∙∙·················∙∙∙∙∙∙∙∙∙∙∙∙∙∙·········………………… 24 4. Memorizing  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙∙··········…·∙∙∙∙∙∙∙∙∙∙∙∙∙···········……………………26 MAINTENANCE ··············∙∙∙∙∙·························∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙∙···········∙∙∙∙∙∙∙∙………………27 TROUBLE SHOOTING  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙∙·················∙∙∙∙······…………··……· 29 1. Against errors  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·······················∙∙∙∙∙∙∙∙∙∙∙∙∙∙···∙∙∙∙∙∙∙∙∙∙∙∙∙·∙·········∙∙∙∙∙∙∙……………………∙  29 2. Against trouble  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·····················∙∙∙∙∙∙∙∙∙∙∙∙∙∙···∙∙∙∙∙∙∙∙∙∙∙∙∙·∙···············∙∙∙∙∙…………∙………30 WARRANTY  ∙∙∙∙∙∙∙∙∙∙∙∙∙∙·················∙∙∙∙∙··············∙∙······································………………… 31 SPECIFICATION  ∙∙∙∙∙∙∙∙∙∙∙∙∙············∙∙∙∙∙∙∙∙∙∙∙∙∙∙················∙∙∙∙∙∙∙∙·························…………··…… 33  TABLE OF CONTENTS
 2    We highly appreciate you choose this JAWON product.  Those who will use this blood pressure monitor are strongly advised to read this manual thoroughly and understand all instructions and warnings before using it.    Information in this manual is made for informational purpose and this manual and product  are  not  meant  to  be  a  substitute  for  the  advice  provided  by  your  own physician or other medical professional. You should not use this information contained in this manual or this equipment for diagnosis or treatment of health problem or prescription of medication.  If you have or suspect that you have a medical problem, consult  your  physician promptly.                        WARNING
 3    1. What is blood pressure? Blood  pressure  refers  to  the  force  or  pressure  exerted  on  the  artery  vessel when blood flows through them. In order to allow the blood to supply nutrition and oxygen to every organ and corner of the body, a certain level of pressure is necessary and it comes from the each time the heart beats or contracts. The heart activity is broken down into two phases called Systole and Diastole. Systolic  is  the  pressure  to  send  blood  out  into  the  body  and  Diastolic  is  the pressure creating as the heart pulls blood back through it. Blood pressure is recorded as systolic over diastolic, as shown below.                                                              systole occurs                      diastole occurs  during the contraction                              during the relaxation   and empting of the heart            and filling of the heart   2. What makes my blood pressure fluctuate? Blood pressure changes in response to many factors. These  factors  include  changing  in  the  temperature  and  humidity, physical exercise,  restlessness  due  to  stress,  fear,  anxiety,  anger,  food  and  an individual’s physical condition.   3. Classification of the blood pressure   :  National  High  Blood  Pressure  Education  Program,  National  Heart,  Lung  and    ABOUT BLOOD PRESSURE 115/75
 4Blood Institute, NIH (JNC7)  Diastolic    (㎜Hg)                           100                90                             80                                                 120       140        160                  Systolic(㎜Hg)                          4. Why do I have to check blood pressure at home? Home  monitoring  may  help  alleviate  the  medical  phenomena  ‘White Coat Hypertension’ which causes a sensitive individual blood pressure to be higher when  taken  in  a  medical  office  setting.  The  elevation  of  it  may be due to several  factors,  including  stress,  nervousness,  or  increased  sympathetic activity  associated  with  the  clinical  environment.  It  may  cause  your  blood pressure to rise about 20 to 30mmHg higher than usual. When monitoring  at home,  you  may  get  a  truer  indication  of  your  blood  pressure  in  a  usual, relaxed  state.  Regular  home  monitoring  provides  you  a  valuable  over  time readings  to  give  to  your  physician  to  diagnose  and  optimize  the treatment program  corresponding  your  circadian  rhythm  of  blood  pressure.  It  also provides you the information when your blood pressure falls or rises to enable to  manage  your  health.  Therefore,  it  is  important  to  take  regular  daily  home monitoring and keep clear records. Note    Blood pressure constantly fluctuates during of a day, so that monitoring   should be done at least three times and determined. Occasional high or low readings  do  not mean  your monitor  is  malfunctioning  or  you  are ill. The important thing is that you have information to talk with your doctor about normal and abnormal results corresponding your circadian rhythm. Stage 2 Hypertension  Stage 1 Hypertension  Prehypertension  Normal
 55. What is benefit of ECHO-H User-set inflation level monitor use?    ECHO-H allows that user can set up the inflation level corresponding the blood pressure  of  user-self.  This  function  is  ideal  for  the  hypertensive  while  other kinds of pressurization method are more suitable to the person who is able to estimate own blood pressure or under treatment. For a correct monitoring, the pressurization  need  30㎜Hg  higher  than  your  estimated  blood  pressure. Therefore  the  hypertensive  may  have  frequently  experienced  re-pressurization during  monitoring  with  the  monitor  employed  those  methods.  In  case  of excessive hypertension or a feeble pulse, it is possible to experience it over 2 times a monitoring. That will affect your emotion, blood vessel and eventually monitoring results.   ECHO-H,  the  monitor  of  user-set  inflation  excludes  the  necessity  of  re-pressurization  during  monitoring  of  the  hypertensive  and  makes  it  convenient and correct with the single pressurization trial.   6. What is the benefit of date and time recorded wrist monitor use?       There is no need to remember monitoring date and time if the device   automatically  records  them.  As  ECHO-H  is  built-in  calendar  and  clock  that inconvenience is not any more.   These are more helpful to the device having big memory capacity.    7. What is the benefit of ECHO-H data transmissible wrist blood  pressure monitor use? Without  going  to  hospital,  you  can  have  physician’s  advice  with data transmissible blood pressure monitor.   ECHO-H is a Home to Hospital bi-directional networking blood pressure monitor that you can transmit saved data from the device to your  personal computer through interface then send your physician them. Your physician will read those to find health problems and give proper advice you through computer network. With the support of the analyzing program, ECHO-H tracks your blood pressure changes sophisticatedly and enables you to realize  the characteristics  of  yours  and  control  it. ECHO-H is a time and cost saving healthcare networking watch-type instrument.
 6   Thank you for purchasing the ECHO-H. Intended use This home blood pressure monitoring device is indicated for use in the noninvasive measurement of systolic and diastolic blood pressure and heart  rate in adult age 18 and over, in a home care environment. This model is designed to enable a person who suspects some trouble with blood pressure to check it at home. User-set inflation and Oscillometric method   Your unit employs a user-set inflation and an oscillometric method. The user can decide  its  inflation  level  corresponding  the  characteristics  of  the  user’s  blood pressure. Oscillometric for measurement of blood pressure and heart rate, which refers to check blood pressure by measuring oscillation in the cuff while the artery of the arm oscillate reflecting contracts and relaxes every heart beat. Data transmissible and analyzing software This device is data transmissible. You can transmit readings saved memory of the device  to  your  personal  computer  then  mailing  to  your  physician and get advice about it from her/him at home. ECHO-H is a time saving device partially replacing a  necessity  your  visit  to  your  physician.  Additionally  the  data  transmitted  to  your PC can be analyzed and tracked sophisticatedly with analyzing program optionally offered by manufacturer.   Voice massage  Makes you monitor blood pressure easily and know the result clearly without third person’s help especially for the elderly.   Noise detecting As  ECHO-H is  contrived  to  detect  moving  or noising during measurement and display  it  both  on  the  real  time and recalled reading on LCD, you and your physician can determine if the readings are reliable.   Memory feature It has 64 reading memory capacity recorded monitoring date and time supporting built-in calendar and clock which are input them automatically to each reading.   Accuracy Blood pressure measurements determined with this device are equivalent to those obtained by a trained observer using the cuff/stethoscope auscultation method,   within  the  limits  prescribed  by  the  American  National  Standard, Electronic or INTRODUCTION
 7Automated Sphygmomanometers.   To ensure safe operation and durable performance stability, it is essential that you fully understand the functions, operation and maintenance instructions by reading this manual carefully before using the unit. Particular attention must be paid to all warnings, cautions and notes incorporated herein. The following conventions are used throughout the manual  to  denote information of special emphasis.            ※ Property damage refers to a wide range of damage involving your  house, household goods, domestic animals, and pets etc..            WORD DEFINITION  CAUTION Important information to indicate the presence of a hazard in which may arise severe personal injury, death of substantial property damage if the warning is ignored. NOTE Important  information  to  notify  the  user  of  operation  or  maintenance information  in  which  is  important  but  not  hazard-related.  Hazard  warning should never be included under the note signal word.
 8   • For all damages or injuries caused by user’s incorrect operation or failure in maintenance the unit relieves, the manufacturer and its agent of the system do not have any responsibility and non-compliance with specifications.  •  Please  keep  this  manual  together  with  the  appliance  since  it  is  compiled  to provide the maximal convenience to those who use this device.  • Defective unit or accessories that is shipped off from customer  to  Jawon medical Co. or its authorized dealer for service must be packed in the safe way and  the  replacement  carton.  Shipping  and  insurance  cost  for  returning  of defective unit must be prepaid by customer.                       NOTE TO USER
 9   Classification and Compliance with Standards Model ECHO-H manufactured  by Jawon medical is classified as ;  - ECHO-H is specified as Class 2 type BF unit under the standard of IEC60601-1 (Safety of Medical Electrical Equipment).   - Internally powered equipment (Battery : 2  ×  LR6-AAA) - Ordinary equipment without protection against ingress of water - Equipment not suitable for use in presence of a flammable anesthetic mixture by standard of IEC60601-1 (Safety of Medical Electrical Equipment) - Equipment that complies with Class B for Noise-Emission, Level B  for Noise-immunity  by  standard  of  IEC60601-1-2  (Electromagnetic Compatibility Requirements) This blood pressure meter complies with the EN1060-1 : 1995 Non-invasive Sphygmomanometers  general  requirements  as  well  as  EN1060-3  :  1997 supplementary  requirements  for  electro-mechanical  blood  pressure measuring system.    CLASSIFICATION & COMPLIANCE WITH STANDARD
 10
 11  ECHO-H  are  designed  and  manufactured  in  deep  consideration  for  users’ and operators’ safety and completing reliability of the unit based International Standards for medical equipment.   The following precautions must be observed for additional safety ;  1. User must not touch or handle the inner side the system at anytime.  It  may cause an electrical shock and flame out. 2. The unit has been factory-adjusted for optimum performance. Do not attempt to  modify  or  adjust  any  preset  controls  or  switches  except  those  specified  in this manual for operation. If any modification is desired, ask manufacturer or its authorized dealer for service. 3.  If  you  have  any  trouble  with  the  unit,  switch  it  off  immediately,  and  contact manufacturer or its authorized dealer for assistance. 4. Avoid the following environments for storage and operation   - where the ambient temperature deviates -20℃  ~ 60℃  for storage - where the ambient temperature deviates 10℃  ~ 40℃  for operation - where the atmospheric pressure deviates 70kPa(700mbar)~106kPa(1060mbar) - where the humidity is above 95% for storage and below 30% or above 75% for operation - where the unit is exposed to spray or splashing water - where the unit is exposed to dust - where the unit is exposed to water vapor - where the unit is exposed to high density oil vapor - where the unit is exposed to salty atmosphere - where the unit is exposed to explosive gas - where the unit is exposed to excessive shocks or vibrations - where the angle of inclination of mounting surface is in excess of 10 degrees - where the unit is exposed to direct sunlight 5. Do not use the unit on infants or persons who cannot express one’s intension 6. Do not use the unit for any purpose other than measuring blood pressure 7. Do not use a cellular phone near the unit improper operation may result 8. Do not force to bend the arm cuff or the air tube excessively 9. When removing the air tube, pull at the connector  SAFETY PRECAUTIONS
 1210. Do not apply strong shock or drop the main unit 11. Do not inflate the arm cuff without it being wrapped on the arm 12.  The  battery  liquid  may  leak  and  damage  the  main  unit.  Please  observe  the following points. - When you are not going to use the unit for an extended period of  time (approximately three months or more), remove the batteries - Replace the exhausted batteries with new ones immediately - Do not use exhausted and new batteries together - Do not insert the batteries with their polarities in wrong direction       The  International  Electro-technical  Commission  (IEC)  has  established  a  set  of symbols for medical electrical equipment that classify a connection or warn of any potential hazards. The symbols are shown below.  Save these instructions for your safety.  •  Isolated patient connection (IEC 60601-1 Type BF)     •  I and O on power switch represent ON and OFF               respectively   •  This symbol identifies a safety note. Ensure you understand function of this control before using it. Control function is described in operation manual.   SAFETY SYMBOLS   I/O
 13  1. Main unit and accessories                  Cuff Button  : Power ON/OFF & automatic start,    Enter to sound and volume mode S : An inflation setting button M: Button to memory view, sound off and volume control    : Entering the mode to set date and time  : Number up button to set date and time   : Button to transmit data USB cable : To use data transmission refer to owners’ manual of software ‘Hour 24’ how to use or how to install.         PART NAMES & FUNCTIONS
 142. SYMBOLS IN LCD INDICATOR                ♥    : Displays and flashes when pulse is detected  ▲  : Pressurization    Displays during inflation ▼  : Exhausting  Displays during deflation   :    Displays when the batteries need to be changed     :  Memory indicator  :  Displays when sound is ON  ▲▼  :    Indicates noise or movement sensed   :  Memory indicator  01 : Indicates the number of stored results   DATE(10.01) TIME(12:00) : Month.day    Hour : minute
 15    Exercise or any kind of activity may affect blood pressure. Rest about 5 minutes before monitoring.      Contain one’s urine may elevate blood pressure. Start monitor a few minutes after urination.      Insufficient sleep or constipation may elevate blood pressure.    Cold may elevate blood pressure. Warmyourself at room temperature(about 20℃) before monitoring.       PRECAUTIONS FOR MEASUREMENT
 16      Avoid smoking and drinking caffeine contained beverages such as coffee,tea, and coke, ect.. They may elevate blood pressure.      Moving or talking during monitoring causes error.    To compare the results, choose the readings got under the same condition. (monitoring site, position, time etc.)      Continuous monitoring causes arm congested and inaccurate readings. Allow 3~5 minutes interval between measurements.    BATTERY INSERTION / REPLACEMENT
 171. How to insert the batteries  ①  Pull the battery cover in the arrow direction and open it. ② Insert one AAA type battery to match polarities. ③  Insert other AAA type battery to match polarities as shown below.   ④  Cover the lid of battery case. ⑤  The figure that was finished after inserting batteries             2. Replacement of the Batteries  ①  If the            indicator flashes on LCD, change all batteries. Before batteries are changed or used up, you should transmit memorized readings from your device to your personal computer. If you do not so, you might lose them.       ②  Replace  +  (positive)  and  -  (negative)  polarities  should  be  matched  the polarities of the battery compartment as shown above 1-①     If the polarities of the battery are positioned incorrectly, the unit does not work. ③  When you change the batteries, use the same type. ④  Remove the used batteries immediately to protect the product from damage. ⑤ When you are not going to use the unit for an extended period  of  time (approximately three months or more), remove the batteries.     Note    When batteries are changed or used up, the memorized readings        must be lost. Note   If battery is not disposed properly, it can be harmful. Protect the   environment by taking exhausted batteries to authorized disposal   place.
 18 • Sequence is Year→Month→Day→Hour→Minute. 1. Entering the mode to set date and time. ①  After change batteries, the mode appears on LCD automatically.   ②  At power off, press [ ] button. [00 YEAR] will appear on the screen, and [00] flashes.        2. Setting up   ①  Press [ ] button located in the middle to   set the year. Whenever [ ] button is pressed   or you keep it pressing, the number will increase gradually or rapidly by ‘1’. Its range is 00 to 99. ②  After setting the year, press [ ] button again.        The month indicator, the first two numbers in upper   LCD line, will flash. Set it by pressing [ ]  button. Its range is 01 to 12. ③  After setting the month, press [ ] button once       again. The day indicator, the second two numbers   in upper LCD line, will flash. Set it as same way. Its range is 01 to 31.   ④  After setting the day, press [ ] button again.     The hour indicator, the third two numbers in upper LCD line, will flash. Set it by pressing [ ] button.   Its range is 00 to 23. ⑤  After setting the hour, press [ ] button once   more. The minute indicator, the last two numbers in upper LCD line, will flash. Set it as same way.  SETTING DATE AND TIME
 19Its range is 00 to 59.   ⑥  After setting the minute, please power off then on. Always the second starts from 00.                              Note  When you press I/O button after setting the date and time, they are stored   coinciding with the power off. If you press I/O button on your way to set, what you have done will be stored. The date and time are maintaining during poweroff but they will return to initial status when batteries are changed. Note   When the clock becomes incorrect after using some time, set it again please.
 20   ①  If you press [I/O] button around   2 seconds, [Snd On/OFF]   signal appears on LCD.    ②  Press [S] button to change it.   ③  Press [I/O] button to complete it.        ①  If you press a [S] button, the display   as right will appear.                                                   ②  Press [S] button again to set inflation level.       Whenever pressing the [S] button the pressyrization       is varied by 5mmHg from 180~300mmHg rotationally.         It is recommended to set inflation 30~40mmHg higher   than own systolic blood pressure.       Stop pressing at the level what you desire.  ③  Press [I/O] button to start measurement.   The inflation level set is saved the unit automatically   with the button pressing.   화면 SET INFLATION LEVEL SOUND ON/OFF
 21  1. How to transmit monitoring data to your PC                ECHO-H blood pressure monitor is a Home to Hospital bi-directional networking instrument.  You  can  reserve  memorized  data  on  your  personal  computer  then send them  to your physician. If you monitor your blood  pressure periodically and memorize reference for your physician, they will be a good information to enable your  physician  to  optimize  the  treatment  program  and  determine  how  much medication  is  required  to  you  corresponding  your  circadian  rhythm  of  blood pressure.       To  install  an  USB  driver  in  a  computer,  follow  the guidance contained in, manual of software ‘Hour 24’ Turn on the power of two appliances after connecting ECHO-H to the computer running ‘Hour 24’(USB cable should be used the one supplied by JAWON MEDICAL exclusively)   DATA TRANSMISSION   NOTE    Must install an USB driver to use an USB port when connect a computerto an ECHO-H. (Refer to'Hour24')
 222. Data transmission     ①  Execute software 'Hour 24' in a personal computer. ②  Connect USB cable to a computer and a port of   ECHO-H in order to transmit saved data. ③  Press only one time of [ ] button of ECHO-FT. ④  Press a [ ] button and transmit to a computer. (All measurement data saved in memory is transmitted)                         3. How to send the data to your doctor       ①  Retrieve saved data in order to   try to have selected the date and time   in software 'Hour 24'.  ②  An 'E-mail' button will be on right   upper side appeared at various    measurement graphs and right upper side when a data to send.                    ③  A window to get input if click a   'E-mail' button opens. Input a mail   address and title and click a 'Send   E-mail' button.   (A save situation of a file is marked   in E-mail title, and a transmitted   figure file and blood pressure   measurement file are saved in a computer of a user.           ('Hour24' manual reference )      Note      Please transmit the data file which use a [SAVE] function of software 'Hour  24'  use  personal  mail,  and  to  transmit,  and  was  saved.Detailed contents refer to 'Hour 24' manual. Note     Microsoft Outlook Express must be installed in a computer to use a transmission function of 'Hour24'.
 23  Remove thick or tight-fit clothes from your wrist. Sit on a chair with your feet flat on the floor and place your wrist on a table so that the cuff can position as same level as your heart.  1. Wrapping the cuff   ①  Wrap the cuff around your wrist. It should be applied   radius artery(inside wrist) and the monitor should face you.    ②  Position the lower edge of the cuff is 1/4 ~ 1/2   inch above the hand as shown in the picture.  ③  Strain the cuff with about a finger space between   the cuff and the wrist.        2. Correct measuring posture ①  During monitoring, place your wrist on your   chest or a table at the same level as your   heart please. Your palm and monitor should   face you.     Fasten your monitoring arm to your heart   with the another hand.    ②  For monitoring, sit down comfortably on   a chair with your feet flat on the floor and take deep breaths 5 to 6 times and relax.    MEASUREMENT Note  Take  off  thick  coat,  sweater  or  tight-fit  clothes  from  the  monitoring site. The cuff should be touched directly to the skin. When you roll up your sleeve, ensure it does not constrict the blood stream in your wrist.
 24  <Correct measuring posture of wrist type>       Note    Sit comfortably on a chair with your feet flat on the floor and relax   before monitoring starts. Monitoring should be taken in a quiet place.
 253. Measuring method ①  Press the [I/O] button                     After wrapping the cuff, press [I/O] button to start.       There will be a momentary full-segment display.       ②  Automatic pressurization   The LCD panel changes to the momentary screen   displayed all symbol in LCD edge and [0] in main LCD.   Then [▲] symbol remains on LCD panel coinciding   with a start of pressurization. At this time the number   to show the inflated cuff pressure will increase rapidly   as a normal process and the cuff will tighten as air   enters in it. The value of pressurizing follows pre-set.   Do not move your arm and remain still until   the entire measurement completed.         ③  Monitoring is in progress       When the pressurizing is over, [▼] symbol appears   and air is released from the cuff slowly. You can   feel the cuff become less tight gradually.    ④  Monitoring will start with deflation and if the pulse is   sensed during deflation [♥] symbol appears and   flashes  ⑤  When noise is detected during measurement, [▲▼] Note    If a higher inflation is needed in any circumstances, the device re- inflates the cuff up to 30㎜Hg higher   Note   If you want to transmit the real time data to your PC, you should connect the device to the computer  before  the  monitoring  starts  in ECHO-H.
 26appears on its place on the screen. In case that you   move or speak, [▲▼] is accompanied with a voice   warning as “Don’t move or speak please.” but   monitoring will be continued after a pause.  ⑥  When Error occurs due to insufficient inflation,   re-pressurization will be held automatically with a   voice massage as “Can’t measure. We will try again.”  ⑦  Display the results       When monitoring is over, the value of systolic and   diastolic blood pressure and pulse rate will be displayed   on LCD simultaneously, then the voice massage comes out as “Your blood pressure is systolic 000, diastolic   000, pulse 000.” If the noise or movement detected   during monitoring, voice massage says also   “There was movement during measuring.” and shows   the [▲▼] in result view.    ⑧  Switch off the power       To finish the measurement, press the [I/O] button and remove the cuff gently. The monitored reading will be stored automatically into the memory.               Note  If  you  forget  to  press  the  I/O  button,  the  power  will  be  automatically turned off after 1 minute. Caution  1. During measurement, do not move your fingers, hand, or arm, and do not put them on stress. 2. If you want to stop measuring, press the I/O button.       Pressurization will stop immediately and air will be exhausted rapidly, then the power will be turned off.
 274. Memorizing ECHO-H has 64 reading memory capacity. When memorized readings are retrieved, they are displayed with monitoring date & time respectively. The reading detected noise or movement during measurement accompanies [▲▼].  1) Storing the readings The readings are stored automatically when the monitoring is completed. When   it is stored by 64 readings, the oldest one will automatically deleted as new one   is added.    2) How to retrieve the stored data ①  Press the [M] button.    ②  The latest reading will be displayed on main   LCD and the number ‘01’ appears in the   memory number displaying part.    ③  When the [M] button is pressed again   the second latest reading will be displayed on LCD with the memory number ‘02’.    ④  Further pressing of the [M] button,   the data will be displayed respectively in reverse order from the latest one to oldest one with memory number increasing.    ⑤  If there is no data reserved, the LCD displays as shown.          Note   1. During power off and after transmission, the stored data in unit is continuously maintained. 2. The memory capacity is subject to change for promotion.
 28  1. Maintaining Avoid exposure to dust, direct sunlight, humidity and extreme changes in temperature.  If this unit is not being used for a long period of time, please remove the batteries.     Do not drop the unit.    Do not disassemble the main body or the cuff.  Do not fold or bend the cuff forcibly.     Do not operate the unit without wrap the cuff around your arm.    MAINTENANCE
 29      2. Cleaning  1) Main body` ①  When cleaning, use a soft cloth, slight   moistened with water or a neutral   detergent, then wipe it with a dry cloth. If the unit is contact with water in outer   case, dry it immediately with a soft   lint-free cloth. ②  Do not use abrasive or corrosive cleaning agent, thinner, alcohol, benzene or   similar solvents which might damage the device. It is recommended to clean the unit monthly.    2) Cuff Do not wash or dip the cuff in the water.              Caution  Do not immerse the unit in water. The device exposed to foreign     material and exceptional environment need to be check whether the device is functioning properly and safely before using. Note    When you need change a cuff worn out or become dirty, ask manufacturer or its authorized dealer for service.
 30  1. Against Error    ERROR  CAUSE  SOLUTION  Err 0  * When inflation level is not sufficient. * When air hose is not connected well. * When air leaks from the cuff.* When battery power is low.   * Please set higher inflation Level. * Check if there is any   ‘CAUSE’.   * After settle it, monitor   again.  Err 1    * When a cuff is not       connected to a monitor. *When  a  cuff  is  positioned incorrectly   * Please check if the cuff is   connected well to the unit   and monitor again.   Err 2  *  When  the  pressurization  is over than 300㎜Hg.   * Please turn the power off   then on and monitor again.   Err 3  * When monitoring is carried out the wrong way.   *When  a  readed  data  is  not output  as  a  result  of  noise  or movement detected.   * Please read ‘wrapping the   cuff’  and  ‘correct  measuring posture’  in  ‘Measurement’ part. (page 18~19) * Monitor again following the way instructed.    Err 4  * When moving or speaking is sensed during monitoring.   * Please do not move or   speak during monitoring.  ※  If any error continues, call its service center.    TROUBLE SHOOTING
 31 2. Against trouble  SYMPTOM  CAUSE  COUNTERMEASURE     The battery power is low.   (When  batteries  are  exhausted completely,  the  mark  does  not appear on the LCD any more.)   Replace all two “AAA” type batteries at the same time. Before change transmit the data to your computer first.  The blood pressure value seems too high or low.    High: When your arm positioned lower than your heart   during monitoring. Low : When your arm positioned higher than your heart   during monitoring.   Make sure that your arm position at the same level as your heart, then monitor it again.  Re-pressurization  In case of insufficient inflation. (In case of excessive hypertension or a feeble pulse, it is possible to occur   re-pressurization over 2 times.)   Relax for a few minutes then monitor it again with correct posture. Do not move during monitoring.  The unit does not start monitor again after error occurs.   The cuff may not be positioned incorrectly.  Monitor it again after wearing the cuff correctly and positioning your arm as your heart level.       TROUBLE SHOOTING
 32   1. Manufacturer’s limited warranty The warranty and remedy provided herein are exclusive and in lieu  of  all  other express  warranties  and  unless  stated  herein,  any  statements  or  representations made by any other person or entity are void. Neither the manufacturer, distributor nor their affiliates shall be liable for any loss, inconvenience, injury, or damage including direct, special, incidental,  or consequential  damages  resulting  from  the  use  or  inability  to  use  this  product, whether arising out of any breach of this warranty or any other legal theory. All express  and  implied  warranties,  including  the  warranties  of  merchantability,  are limited to the applicable warranty period set forth above.  Manufacturer and distributor assume no responsibility for damages  or  penalties incurred resulting from the use of this product in a manner or  location other than that for which it is intended.  Some states do not allow the exclusion or limitation of incidental or consequential damages,  or  limitations  on  how  long  an  implied  warranty  lasts,  so the above exclusions  or  limitations  may  not  apply  to  you.  This  warranty  gives  you  specific legal rights and you may also have other right that vary from state to state.  Should your product prove defective during this warranty period, please return the product in the original carton to your place of purchase for warranty service.  2. Warranty steps Should you experience any technical problems with your product, please follow the steps as below : ①  Re-read  “Battery  installation”  and  “Measurement”  parts  of  the  instruction carefully,  then  take  a  monitoring  again.  If  the  problem  persists,  your  product may be defective.  If  you believe  the product is  defective, follow this warranty steps  ②  or  ③. ②  Manufacturer, at its option, will repair or replace your product (with a same or a comparable  model)  with  new  or  rebuilt  parts,  free  of  charge  within  one  year from the date of original purchase in the event of a defect in materials  or workmanship.  This  warranty  is  extended  only  to  the  original  purchaser.  A  WARRANTY
 33purchase  receipt  or  other  proof  of  date  of  original  purchase  will  be  required before warranty performance is rendered. ③  This  warranty  only  covers  failure  due  to  defects  in  material  or  workmanship which  occur  during  normal  use  and  does  not  cover  abnormal wear of the product. This warranty is void if the product housing has been removed, if the manufacturer’s label, logo, or serial number have been removed, or should the product fail to function properly as a result from accident, misuse,  abuse, neglect,  mishandling,  misapplication,  defective  batteries,  faulty  installation, set-up and adjustments, improper maintenance, alteration, maladjustment of controls,  modification,  power  surges,  commercial  use  of  the  product,  or service  by  anyone  other  than  an  authorized  service  center,  or  acts  beyond manufacturer’s control.  3. Where applying for solving problem Should you be dissatisfied with the service you received from your retailer, please write  JAWON MEDICAL CO.,LTD. •  Seoul office Tel : *82-2-587-4056                 Fax : *82-2-588-1937  •  Factory 13 Lot, 21 Block Jinryang Industrial Complex, Kyungsan-city Kyungsangbuk-do 712-830, R.O.Korea Tel : *82-53-856-0993              Fax : *82-53-856-0995   Include in your correspondence the following information, clearly printed (i) Your name and street address (ii) Daytime telephone number (iii) A detailed description of the problem (iv) A copy of your dated proof of purchase or bill of sale
 34   Model  ECHO-H   Method of Measuring  Oscillometric Method of indicating  Digital Pressurizing Method  PWM(Pulse Width Modulation) Range of Measuring  Blood Pressure : 30 ~ 280mmHg Pulse Rate : 40 ~ 200beats/min Accuracy  Pressure : within  ±2mmHg or  ±2% Pulse Rate : within  ±5% Memory  64 Reading Memory Resolving Power  1mmHg Automation Shut-off System  Performs approximately 1 minutes after measurement. Power DC 3V (AAA type 1.5V x 2),  Operating Environment  Temperature : 10℃ to 40℃ Humidity :    30% to 75% RH, non condensing Storage Environment  Temperature : -20℃ to 60℃ Humidity : less than 95% RH, non condensingDimension  77.0(W) X 57.0(D) X 35.5(H) mm  Weight  Approx. 200g (Including cuff &batteries) Pressure Mode  User-set inflation ◈  Specifications are subject to change without notice.  ※  Life of battery can be reduced according to environment.   ※  This  manual  cannot  do  whole  or  a  part  reproduction  without  notice  without written consent of JAWON MEDICAL according to copyright act. ※  JAWON  MEDICAL  does  not  have  responsibility  about  a  misprint  and  do  not take the responsibility on the results about which do indication in  manual, or did not do presentation.       SPECIFICATION
 35