Johnson and Johnson Surgical Vision SIGREMM3 WHITESTAR Signature Pro System User Manual signature users manual
Abbott Medical Optics WHITESTAR Signature Pro System signature users manual
User Manual
WHITESTAR SIGNATURE PRO System 8 • Error Messages Troubleshooting and Diagnostics 513 IV Pole communications error. Probable cause: Communication error. Corrective action: 1. Prime the system. 2. If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service. 514 IV Pole error. Probable cause: Communication error. Corrective action: 1. Prime the system. 2. If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service. 515 IV Pole calibration error. Probable cause: Communication error. Corrective action: 1. Run Self Test. 2. If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service. 516 IV Pole communication error. Probable cause: Communication error. Corrective action: 1. Prime the system. 2. If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service. RX Only –NGP Z370500 Rev. B 1116 8-38 WHITESTAR SIGNATURE PRO System 8 • Error Messages Troubleshooting and Diagnostics 517 IV Pole is jammed. Probable cause: • IV Pole motor failure. • There is an obstruction that does not allow the IV pole to move. Corrective action: 1. Check the IV pole for obstructions. 2. Prime the system. 3. If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service. 601 Tune excessive vacuum error. Probable cause: There is too much vacuum. Corrective action: 1. Check tubing connections. 2. Check the handpiece. Replace the tip if needed. 3. Retune the handpiece. 4. If the error still does not clear, replace the handpiece with a new handpiece, and tune the handpiece. 5. If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service. 605 Tune no handpiece error. Probable cause: Hardware failure. Corrective action: 1. Attach handpiece to the system. 2. Retune the handpiece. 3. If the error still does not clear, replace the handpiece with a new handpiece, and tune the handpiece. 4. If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service. RX Only –NGP Z370500 Rev. B 1116 8-39 9 WARRANTY AND MAINTENANCE Warranty Statement RX Only –NGP Z370500 Rev. B 1116 9-1 WHITESTAR SIGNATURE PRO System Warranty Statement 9 • Warranty and Maintenance Abbott Medical Optics Inc. warrants for a period of one year (12 months) from the date of installation of the WHITESTAR SIGNATURE PRO System console, foot pedal, wireless remote control, the programmable power pole, and the phaco handpiece to be free from defects in materials and workmanship when properly installed, maintained, and used for the intended purpose. In no event shall the date of installation be deemed to occur later than six months from the date of delivery of the WHITESTAR SIGNATURE PRO System by Abbott Medical Optics Inc. to the common carrier. You are required to provide AMO with prompt written notice of any defect or defects and permit AMO to have access to the WHITESTAR SIGNATURE PRO System within a reasonable time after notification of any defect. In the event that you do not provide AMO with prompt notification of a defect or permit access to the WHITESTAR SIGNATURE PRO System within a reasonable time after notification, AMO is hereby released from all liability with reference to such defect and subsequent damage, if any, resulting from the inability of AMO to correct such defect. MISUSE AND MISHANDLING ARE NOT COVERED UNDER WARRANTY. AMO’s sole obligation is to repair or replace, at AMO’s option, the defective part(s), which shall be returned freight prepaid to AMO by you and then returned to you freight collect once repaired or replaced. The irrigation/aspiration handpiece set, diathermy forceps and diathermy cord are warranted for ninety (90) days. This warranty applies only to the original purchaser/user of the device and may not be transferred. This warrant will be void if and only so long as the equipment is used in any country other than the country to which it was originally shipped by AMO. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WITHOUT LIMITATION THE WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL AMO BE LIABLE TO THE PURCHASER/USER FOR CONSEQUENTIAL, SPECIAL, INCIDENTAL, OR PUNITIVE DAMAGES, LOSS OF PROFITS OR LOSS OF USE. The liability of AMO for breach of this warranty shall, in any case, be limited to the total amount paid by you for the WHITESTAR SIGNATURE PRO System furnished by AMO and any work performed by AMO in connection therewith. Extended Warranty Extended warranty contracts (excluding the phaco handpiece) are available. Contact AMO for information on the availability of an extended warranty contract. Note: If outside the United States, contact your local AMO office or representative for warranty information. RX Only –NGP Z370500 Rev. B 1116 9-2 WHITESTAR SIGNATURE PRO System 9 • Warranty and Maintenance Maintenance User maintenance of the WHITESTAR SIGNATURE PRO System is only for those adjustments and corrective actions in the Error Messages and Troubleshooting, and Diagnostics sections of this manual. There are no user serviceable components within the console and you must not attempt to access the internal components. Any attempt to do so will void the warranty. AMO recommends, at least annually, routine or periodic maintenance of the WHITESTAR SIGNATURE PRO System by an AMO representative. AMO recommends the measurement of PE resistance and leakage current according to IEC 601-1 every two years. If a problem continues following setup, check-out and troubleshoot as per the procedures in this manual, then contact AMO for corrective action (1-877-AMO-4-LIFE in USA 1-877-266-4543 or 1-800-511-0911 for outside the USA). RX Only –NGP Z370500 Rev. B 1116 9-3 10 SPECIFICATIONS Physical Specifications Environmental Specifications Electrical Specifications Phacoemulsification Specifications Irrigation and Aspiration Specifications Vitrectomy Specifications Diathermy Specifications Diathermy Power Graphs Diathermy Power versus Load Impedance Phaco Power Graphs Phaco Tip Velocity Bottle Height Pressure RX Only – NGP Z370500 Rev. B 1116 10-1 WHITESTAR SIGNATURE PRO System Physical Specifications 10 • Specifications Table 10.1 – System Physical Specifications English / U.S. Metric: Width 24 inches 61 cm Depth 24 inches 61 cm Height 59 inches 150 cm Power cord length 20 feet 6 meters Electrical enclosure current leakage IEC 60601-1 compliance UL 60601-1 Weight (including IV pole) 185 pounds 84 kg Width 12 inches 31 cm Length 10.5 inches 27 cm Height 5.5 inches 14 cm 10.0 pounds 4.5 kg Width 10.5 inches 27 cm Length 14 inches 36 cm Height 5.5 inches 14 cm Weight 15 pounds 7 kg Cord Length 11.8 feet 36 cm Width 9.82 inches 25 cm Length 15.57 inches 39.6 cm Height 4.88 inches 12.4 cm Weight 13 pounds 5 kg Cord Length 12 feet 366 cm Maximum travel 41 inches 106 cm Velocity 2.4 inches/sec 6 cm/sec. Maximum lift weight 2.43 pounds 1.1 kg WHITESTAR SIGNATURE PRO System Console Console dimensions (with display) Foot Pedal Single Linear Foot Pedal Dimensions Weight Advanced Control Pedal Dimensions Advance Linear Pedal Dimensions Programmable IV Pole RX Only – NGP Z370500 Rev. B 1116 10-2 WHITESTAR SIGNATURE PRO System 10 • Specifications English / U.S. Metric: Width 5 inches 13 cm Length 5 inches 13 cm Height 1.5 inches 4 cm Weight 2.0 pounds .9 kg Wireless Remote Control Environmental Specifications Table 10.2 – Environmental Specifications Environmental Specifications Storage/Transportation Conditions Electrical Specifications System Operating Temperature 10 to 40°C Humidity Up to 95% RH, non-condensing Ambient Pressure See Vitrectomy Specifications for additional information. Storage Temperature -20 to 60°C Humidity Up to 95% RH, non-condensing Table 10.3 – Electrical Specifications Voltage Frequency Rated Power Fuse Rating Enclosure Current Leakage 100-240 Vac 50-60 Hz 750 VA 6.3A, 250V, Bussman GDA <500 µA Phacoemulsification Specifications Table 10.4 – Phacoemulsification Specifications Peristaltic Pump Venturi Pump Phaco Power 1%-5% in 1% increments, 5-100%, in 5% increments Vacuum Level 0–650 mmHg in 5 mmHg increments Pump Flow Rate 0-60 cc/minute Vacuum Level 0-600 mmHg in 5 mmHg increments RX Only – NGP Z370500 Rev. B 1116 10-3 WHITESTAR SIGNATURE PRO System Irrigation and Aspiration Specifications 10 • Specifications Table 10.5 – Irrigation and Aspiration Specifications Peristaltic Pump Venturi Pump Vitrectomy Specifications Vacuum Level 0–650 mmHg in 5 mmHg increments Pump Flow 0-60 cc/minute Vacuum Level 0-600 mmHg in 5 mmHg increments Table 10.6 – Vitrectomy Specifications Peristaltic Pump Vacuum Level 0–650 mmHg in 5 mmHg increments Pump Flow Rate 10-60 cc/minute, in 1 cc increments Venturi Pump Vacuum Level 0-600 mmHg in 5 mmHg increments Atmospheric Pressure/Altitude Density Altitude (DA) conditions are 2,500 cpm from sea level to 3,000 feet (914.4 m) elevation and shall be de-rated by 100cpm per 1,000ft (304.8 m) additional elevation up to a maximum of 10,000 ft above sea level. Indicated Cut Rate Actual Cut Rate Upper Limit (cpm) Actual Cut Rate Lower Limit (cpm) 2500 3000 2000 RX Only – NGP Z370500 Rev. B 1116 10-4 WHITESTAR SIGNATURE PRO System Diathermy Specifications 10 • Specifications Table 10.7 – Diathermy Specifications Power Adjustment 5% to 100%, in 5% increments Diathermy Power (maximum power into rated load) 8.5 Watts into 350 Ohms 386 kHz Diathermy Type Bipolar Rated Peak Voltage 160 Volts Connector of diathermy forcep/ probe Ø.072 RECEPTACLE (2) All measurements are in inches. RX Only – NGP Z370500 Rev. B 1116 10-5 WHITESTAR SIGNATURE PRO System 10 • Specifications Diathermy Power Graphs Figure 10.1 – Diathermy Output Power (Typical) 10 Ohm Y Axis = Watts X Axis = Settings Figure 10.2 – Diathermy Output Power (Typical) 50 Ohm Y Axis = Watts X Axis = Settings RX Only – NGP Z370500 Rev. B 1116 10-6 WHITESTAR SIGNATURE PRO System 10 • Specifications Figure 10.3 – Diathermy Output Power (Typical) 200 Ohm Y Axis = Watts X Axis = Settings Figure 10.4 – Diathermy Output Power (Typical) 350 Ohm Y Axis = Watts X Axis = Settings RX Only – NGP Z370500 Rev. B 1116 10-7 WHITESTAR SIGNATURE PRO System 10 • Specifications Figure 10.5 – Diathermy Output Power (Typical) 500 Ohm Y Axis = Watts X Axis = Settings Figure 10.6 – Diathermy Output Power (Typical) 1000 Ohm Y Axis = Watts X Axis = Settings RX Only – NGP Z370500 Rev. B 1116 10-8 WHITESTAR SIGNATURE PRO System Diathermy Power versus Load Impedance 10 • Specifications Figure 10.7 – Diathermy Output Power (50% Setting) Y Axis = Watts X Axis = Impedance (Ohms) Figure 10.8 – Diathermy Output Power (100% Setting) Y Axis = Watts X Axis = Impedance (Ohms) RX Only – NGP Z370500 Rev. B 1116 10-9 WHITESTAR SIGNATURE PRO System Phaco Power Graphs 10 • Specifications Figure 10.9 – Longitudinal Phaco Power Y Axis = Output Power (Watts) X Axis = GUI Settings % Figure 10.10 – ELLIPS FX Phaco Power Linear Mode Y Axis = Power (Watts) X Axis = GUI Settings % RX Only – NGP Z370500 Rev. B 1116 10-10 WHITESTAR SIGNATURE PRO System Phaco Tip Velocity Bottle Height Pressure 10 • Specifications Nominal Velocity for the ELLIPS FX handpiece and the WHITESTAR handpiece at 100% power, continuous, with the OPOR3020L phaco tip is shown in the following table. Handpiece Nominal Longitudinal Velocity (m/s) ELLIPS FX (690880 (A)) 8.85 WHITESTAR (690697 (A)) 11.13 The WHITESTAR SIGNATURE PRO System has a built in IV pole to generate positive pressure using balanced salt solution to provide positive irrigation fluid. Measurement points: • All measurements are taken with a pressure meter at the same height as the pressure sensor in the WHITESTAR SIGNATURE PRO System. • This position is defined as 41 inches or 101.6 cm from the floor. 1. Center of Drip Chamber 2. Center of pressure sensor is 41 inches or 104.2 cm from the floor RX Only – NGP Z370500 Rev. B 1116 10-11 WHITESTAR SIGNATURE PRO System 10 • Specifications An IV bottle height reported as 0 cm includes an offset of 7.9 inches or 20 cm to make sure that a positive pressure is always provided. Table 10.8 – Expected Pressure as a function of Bottle Height Bottle Height cm Theoretical Pressure mmHg Corrected Pressure At Patient Eye Level mmHg 0.00 14.80 20 14.80 29.60 40 29.60 44.40 60 44.40 59.20 80 59.20 74.00 100 74.00 88.80 105 77.70 92.50 RX Only – NGP Z370500 Rev. B 1116 10-12 11 ACCESSORIES AND PARTS REORDERING Accessories and Parts List RX Only – NGP Z370500 Rev. B 1116 11-1 WHITESTAR SIGNATURE PRO System 11 • Accessories and Parts Reordering Table 11.1 – WHITESTAR SIGNATURE PRO System Part Number Name WHITESTAR SIGNATURE PRO Console List of Accessories with Part Numbers Description An ophthalmic microsurgical system. Part number NGP680301 The following table is a list of accessories that you can use to reorder parts for your WHITESTAR SIGNATURE PRO System. Your AMO representative can advise you of recommended inventory levels based your facility's volume of phacoemulsification procedures. You can order any of the items listed below through your AMO representative or directly through AMO customer service. Your AMO representative can confirm the availability of these items in your area. Table 11.2 – Accessories and Parts List System Accessories Phaco Handpiece I/A Handpiece Part Number Description 690697 WHITESTAR Phaco Handpiece, Coaxial 690880 ELLIPS FX Phaco Handpiece OPOR0019L 19 Gauge LAMINAR Flow Phaco Tip, 0° Round Tip OPOR1519L 19 Gauge LAMINAR Flow Phaco Tip, 15° Round Tip OPOR3019L 19 Gauge LAMINAR Flow Phaco Tip, 30° Round Tip OPOR4519L 19 Gauge LAMINAR Flow Phaco Tip, 45° Round Tip OPOR0020L 20 Gauge LAMINAR Flow Phaco Tip, 0° Round Tip OPOR1520L 20 Gauge LAMINAR Flow Phaco Tip, 15° Round Tip OPOR3020L 20 Gauge LAMINAR Flow Phaco Tip, 30° Round Tip OPOR4520L 20 Gauge LAMINAR Flow Phaco Tip, 45° Round Tip OPOCR3019L Disposable LAMINAR Flow Phaco Tip, 19 Gauge, 30° Curved OPOCR3020L Disposable LAMINAR Flow Phaco Tip, 20 Gauge, 30° Curved OPOS19L 19 Gauge LAMINAR Flow Irrigation Sleeve and Test Chamber OPOS20L 20 Gauge LAMINAR Flow Infusion Sleeve and Test Chamber OPOMTWL LAMINAR Flow Tip Wrench OPOIA20CRV SOLO I/A Handpiece, 20 Ga., Curved Silicone Sleeve Tip OPOIA2045D SOLO I/A Handpiece, 20 Ga., 45° Silicone Sleeve Tip OPOIA20STR SOLO I/A Handpiece, 20 Ga., Straight Silicone Sleeve Tip OPOIA20KIT SOLO I/A Handpiece Kit OM055002 PHACOFIT Titanium Handle OM05510110 PHACOFIT I/A Tip Curved 0.5 mm Port RX Only – NGP Z370500 Rev. B 1116 11-2 WHITESTAR SIGNATURE PRO System System Accessories Diathermy Handpiece Vitrectomy Cutter 11 • Accessories and Parts Reordering Part Number Description OM05510112 PHACOFIT O-Ring's I/A Tip (5PK) OM05510113 PHACOFIT Titanium, Multi-Tip Set OM05510114 PHACOFIT Cleaning/Flushing Kit OM05510115 PHACOFIT Tip, Luer Adapter (Aspiration) OM05510116 PHACOFIT Tip, 45 Degree Silicone (No Sleeve) OM0551011 PHACOFIT I/A Tip Straight 0.3 MM Port OM0551012 PHACOFIT I/A Tip Curved 0.3 mm Port OM0551013 PHACOFIT I/A Tip 45 Degree 0.3 mm Port OM0551014 PHACOFIT I/A Tip 90 Degree 0.3 mm Port OM0551015 PHACOFIT I/A Tip Binkhorst 0.3 mm Port OM0551016 PHACOFIT I/A Tip Straight SI 0.3 mm Port OM0551017 PHACOFIT I/A Luer Adapter, F/F OM0551018 PHACOFIT I/A Tip Curved, SI, 0.3 mm Port OM0551019 PHACOFIT I/A Tip Straight 0.5 MM Port OM0551020J PHACOFIT Athlete I/A Tip Straight Type OM0551021J PHACOFIT Athlete I/A Tip 45 Degree Type DUA02020 DUET STARTER KIT, 20GA DU02100 DUET ASPIRATION HANDLE DU02200 DUET IRRIGATION HANDLE DU02301 DUET 20G CURVED IRRIGATOR CLOSED END DU02302 DUET 20G CURVED ASPIRATION 0.3MM PORT CAP POL DU02303 DUET 19G FINE IRRIGATING CHOPPER DU02305 DUET 20G OLSON II IRRIGATING CHOPPER DU02325 DUET MST Aguirre Duet Tip DU02335 DUET MST Fine/Nagahara Tip SCP680600 Diathermy Power Cable K121085 Bipolar Forceps, Curved Iris 0.5 mm Tip K147000 18 Gauge Straight Diathermy Pencil, reusable NGP0020 20 gauge Vitreous Cutter NGP0023 23 gauge Vitreous Cutter NGP0025 25 gauge Vitreous Cutter RX Only – NGP Z370500 Rev. B 1116 11-3 WHITESTAR SIGNATURE PRO System System Accessories FUSION Pack 11 • Accessories and Parts Reordering Part Number Description OPO70 FUSION Pack (disposable, sterile) OPO71 FUSION Dual Pump Pack (disposable, sterile) Wireless Remote Control NGP680137 WHITESTAR SIGNATURE PRO System Remote Control 0100-1834 AC Charger, SIGNATURE Wireless Remote Foot Pedal NGP680701 WHITESTAR SIGNATURE System Single Linear Foot Pedal NGP680702 WHITESTAR SIGNATURE System Advanced Control Pedal NGP680703 WHITESTAR SIGNATURE PRO System Advanced Linear Foot Pedal 0100-1835 AC Charger, SIGNATURE Wireless Foot Pedal 0100-0055 CABLE ASSY, WHITESTAR SERIES FOOT PEDAL (Single Linear Foot Pedal and Advanced Control Pedal) Miscellaneous Operator’s Manual 0100-6303 Cable Assy, Advanced Linear Foot Pedal to Console, Signature 0100-1836 LI-BATTERY, SIGNATURE PRO Advanced Linear Pedal 2410-0049-L Power cord, hospital grade, 20 ft., 6.1 meters OM05510114 I/A Cleaning Kit 0100-0750 WHITESTAR SIGNATURE System Dust Cover MSR309 Sterilization Tray NGP Z370494 WHITESTAR SIGNATURE PRO System Operator’s Manual (Bulgarian) NGP Z370495 WHITESTAR SIGNATURE PRO System Operator’s Manual (Chinese) NGP Z370496 WHITESTAR SIGNATURE PRO System Operator’s Manual (Czech) NGP Z370497 WHITESTAR SIGNATURE PRO System Operator’s Manual (Croatian) NGP Z370498 WHITESTAR SIGNATURE PRO System Operator’s Manual (Danish) NGP Z370499 WHITESTAR SIGNATURE PRO System Operator’s Manual (Dutch) NGP Z370500 WHITESTAR SIGNATURE PRO System Operator’s Manual (English) NGP Z370501 WHITESTAR SIGNATURE PRO System Operator’s Manual (Estonian) NGP Z370502 WHITESTAR SIGNATURE PRO System Operator’s Manual (Finnish) NGP Z370503 WHITESTAR SIGNATURE PRO System Operator’s Manual (French) RX Only – NGP Z370500 Rev. B 1116 11-4 WHITESTAR SIGNATURE PRO System System Accessories 11 • Accessories and Parts Reordering Part Number Description NGP Z370504 WHITESTAR SIGNATURE PRO System Operator’s Manual (German) NGP Z370505 WHITESTAR SIGNATURE PRO System Operator’s Manual (Greek) NGP Z370506 WHITESTAR SIGNATURE PRO System Operator’s Manual (Hungarian) NGP Z370507 WHITESTAR SIGNATURE PRO System Operator’s Manual (Italian) NGP Z370508 WHITESTAR SIGNATURE PRO System Operator’s Manual (Japanese) NGP Z370509 WHITESTAR SIGNATURE PRO System Operator’s Manual (Korean) NGP Z370510 WHITESTAR SIGNATURE PRO System Operator’s Manual (Latvian) NGP Z370511 WHITESTAR SIGNATURE PRO System Operator’s Manual (Lithuanian) NGP Z370512 WHITESTAR SIGNATURE PRO System Operator’s Manual (Norwegian) NGP Z370513 WHITESTAR SIGNATURE PRO System Operator’s Manual (Polish) NGP Z370514 WHITESTAR SIGNATURE PRO System Operator’s Manual (Portuguese) NGP Z370515 WHITESTAR SIGNATURE PRO System Operator’s Manual (Portuguese-Brazilian) NGP Z370516 WHITESTAR SIGNATURE PRO System Operator’s Manual (Romanian) NGP Z370517 WHITESTAR SIGNATURE PRO System Operator’s Manual (Russian) NGP Z370518 WHITESTAR SIGNATURE PRO System Operator’s Manual (Serbian) NGP Z370519 WHITESTAR SIGNATURE PRO System Operator’s Manual (Slovakian) NGP Z370520 WHITESTAR SIGNATURE PRO System Operator’s Manual (Slovenian) NGP Z370521 WHITESTAR SIGNATURE PRO System Operator’s Manual (Spanish) NGP Z370522 WHITESTAR SIGNATURE PRO System Operator’s Manual (Swedish) NGP Z370523 WHITESTAR SIGNATURE PRO System Operator’s Manual (Turkish) RX Only – NGP Z370500 Rev. B 1116 11-5 WHITESTAR SIGNATURE PRO System System Accessories 11 • Accessories and Parts Reordering Part Number Description NGP Z370526 WHITESTAR SIGNATURE PRO System Operator’s Manual (Traditional Chinese) NGP Z370524 WHITESTAR SIGNATURE PRO System Operator’s Manual CD, Multilingual RX Only – NGP Z370500 Rev. B 1116 11-6 12 INDEX Accessories 11-1 Add a Surgeon 4-3 Advanced Control Pedal 1-4, 1-18, 2-5, 10-2 Calibration 4-23 Connections 4-17 Default Settings 4-16 Setup 4-16 Switch Assignment 4-17 Testing 4-23 Threshold Configuration 4-22 Advanced Linear Pedal 1-4, 2-5 Battery Replacement 8-4 Connections 4-25 Handle 2-7 Setup 4-25 Switch Assignment 4-26 Test 4-32 Threshold Configuration 4-30 Anterior chamber too deep 8-9 Anterior Segment Surgery 3-2 Aspiration Specifications 10-4 Aspiration Rate Occlusion 5-24 Assign Order Program 5-29 Surgeon 4-35 Back-light button 8-7 Backup All 6-21 Balance Irrigation/Aspiration 3-14 Batteries Remote control 2-9 Battery Charging 4-16, 4-25, 8-6 Replacement 8-4 Begin Tune 3-13 Blocked Aspiration port 5-26 BLUETOOTH 8-7 Bottle Height IA 5-11 Phaco 5-8 Vitrectomy 5-12 Bypass 3-11, 3-15 Calibrate Foot Pedal Advanced Control Pedal 4-23 Single Linear 4-14 Touch Screen 6-11 CASE Mode 5-26 CASE ONE Touch 1-7, 5-27 CASE Vacuum 5-27 Chamber partially collapses 8-11 Changing Mode Parameters 3-16 Charging Battery 4-16, 4-25, 8-6 Checkout System 3-17 Cleaning Touch Screen 7-2 Cleaning/Care 7-2 Common Problems 8-2 Connecting Handpieces 1-9 RX Only – NGP Z370500 Rev. B 1116 12-1 WHITESTAR SIGNATURE PRO System 12 • Index Connections Advanced Control Pedal 4-17 Front Panel 3-2 Rear Panel 3-3, 4-8 Console Dimensions 10-2 Continuous Burst 5-13 Phaco Power 5-15 Continuous Irrigation 1-5, 3-11 Cup Fill 3-10 Continuous Phaco Power 5-14 Copy a Program 5-29 Create Surgeon 5-2 Cup Fill 3-10 Custom Duty Cycles 5-16 Cut Mode Vitrectomy 5-12 Cut Rate Vitrectomy 5-12 Cutter Vitrectomy 3-7 Diathermy 1-7, 1-10 Default Settings 5-12 Forceps 3-6 Handpiece 3-6, 7-2 Output Power 10-9 Diathermy Handpiece Part Numbers 11-3 Dimensions 10-2 Dual Pump 1-5 Duty Cycles 5-15 Database Delete 6-25 Import/Export 6-19 Restore 6-23, 6-25 Date Set 6-16 Default Settings Diathermy 5-12 Foot Pedal Advanced Control Pedal 4-16 Single Linear 4-7 Irrigation/Aspiration 5-11 Phaco 5-8 Vitrectomy 5-12 Delete Database 6-25 Program 5-2, 5-28 Delete a Backup Database 6-25 Diagnostics 8-1 Edit Program 5-3 Surgeon 4-4 Effective Phaco Time 3-21 Electrical Specifications 10-3 ELLIPS Handpiece 1-4, 3-4, 3-5 End Case 3-21 Environmental Specifications 10-3 EPT 3-21 Export 6-19 Failed to pair 8-8 Feedback Foot Pedal 4-7 Fluidics 1-3 Fluidics Mode Phaco 5-8 RX Only – NGP Z370500 Rev. B 1116 12-2 WHITESTAR SIGNATURE PRO System 12 • Index Foot Pedal 1-4 Advanced Control Pedal 1-18 Calibrate 4-23 Connections 4-17 Default Settings 4-16 Setup 4-16 Switch Assignment 4-17 Testing 4-23 Threshold Configuration 4-22 Advanced Linear Pedal 2-5 Configuration 4-30 Connections 4-25 Setup 4-25 Test 4-32 Feedback 4-7 Modes Vitrectomy 5-6 Part Number 11-4 Pitch 2-7 Pitch Regions 4-7 Positions 2-7, 5-16 Single Linear 1-4, 2-5 Calibration 4-14 Default Settings 4-7 Setup 4-6 Switch Assignment 4-9 Testing 4-14 Threshold Configuration 4-12 Specifications 10-2 Forceps 9-2 Diathermy 3-6 Front Panel 3-2 Fuse Replacement 8-2 FUSION 5-26 FUSION Fluidics 5-20, 5-22, 5-26 FUSION Tubing Pack 2-3, 11-4 Graphic User Interface 1-5 Graphs Phaco Power 10-10 GUI 1-5 Handle Advanced Linear Pedal 2-7 Handpiece Cleaning 7-2 Connecting 1-9 IA 3-6 Phaco 3-5 Setup Diathermy 3-6 I/A 3-6 Phaco 3-4 Vitrectomy 3-7 Ultrasonic 1-9 Warranty 9-2 Help 8-14 High Power Limit 5-19 High Power Pulse Phaco Power 5-14 IA Handpiece 3-6 IAC 5-6 ICA 5-6 ICE Pulse Shaping WHITESTAR Technology 5-19 Import 6-20 Import/Export Database 6-19 Inspection Instructions 2-2 Install Tubing Cassette 3-9 Irrigation Changing 1-8 Continuous 3-11 Flow does not stop 8-9 Inadequate 1-9 Irrigation/Aspiration 1-7 Balance 3-14 Default Settings 5-11 RX Only – NGP Z370500 Rev. B 1116 12-3 WHITESTAR SIGNATURE PRO System 12 • Index IV Pole Does not work 8-8 Programmable 1-10, 2-8 Setup 3-3 Testing 6-3 Monitor Drape 2-3 Multiple Burst Phaco Power 5-15 Power 5-13 Multiple Mode Programming 1-6 Kick 5-19 No aspiration 8-10 No diathermy or poor diathermy 8-13 No phacoemulsification 8-12 No sound when using diathermy 8-13 No vitrectomy cutting 8-13 LAMINAR Flow 3-6 Language Selection 6-14 Locked Program 5-28 Long Pulse Phaco Power 5-14 Power 5-13 Low batteries 8-8 Low Power Limit 5-19 Low Power Pulse Phaco Power 5-14 Lower Threshold 5-27 Maintenance 9-3 Max Aspiration IA 5-11 Vitrectomy 5-12 Max Power Diathermy 5-12 Max Vacuum IA 5-11 Vitrectomy 5-12 Maximum Height IV Pole 6-2 Maximum Vacuum 5-20 Mayo Tray Drape 2-3 Miscellaneous Part Numbers 11-4 Mode Programming 1-6 Modes Changing Parameters 3-16 Occluded Aspiration Rate 5-21 Occlusion Aspiration Rate 5-24 Vacuum Threshold 5-24 Venting 5-26 Occlusion Mode Phaco Settings 5-21 One Touch CASE 5-27 Operating Modes 1-6 Output Power Diathermy 10-9 Pairing Remote Control 6-10 Wired 6-7 Wireless 6-5 Wireless Remote Control 8-7 Panel Front 3-2 Rear 3-3 Parameter Settings WHITESTAR Technology 5-16 Parameters Programming 1-5 Parts 11-1 RX Only – NGP Z370500 Rev. B 1116 12-4 WHITESTAR SIGNATURE PRO System 12 • Index Passive Reflux 5-28 Pencil Probe Diathermy 3-6 Percent Kick High End of Range 5-19 Low End of Range 5-19 Peristaltic Pump 10-4 Phaco Default Settings 5-8 Handpiece 3-4 Phaco Handpiece 1-4, 1-10, 3-4, 3-5, 7-2 Part Numbers 11-2 Phaco Power Continuous 5-14 Graph 10-10 Phacoemulsification 1-2, 1-6, 1-9, 3-4, 8-12 Physical Specifications 10-2 Pitch 2-7 Pitch Regions Foot Pedal 4-7 Poor aspiration 8-10 Poor phacoemulsification 8-12 Positions Foot Pedal 2-7 Power Continuous Burst 5-13 Cord Length 10-2 Failure 1-9 Long Pulse 5-13 Multiple Burst 5-13 Short Pulse 5-13 Single Burst 5-13 Switch 8-2 Power Delivery Diathermy 5-12 pps 5-14 Precautions Safety 1-8 Prime Bypass 3-11 Prime/Tune 1-5, 3-10, 3-11 Priming Vitrectomy 3-15 Program Assign Order 5-29 Delete 5-2 Lock 5-28 Locked 5-28 Select 3-8 Program Name Edit 5-3 Programmable IV Pole 1-6, 1-10, 2-8 Setup 3-3 Specifications 10-2 Testing 6-3 Pulse Off Time 5-16 On time 5-16 Pulses per second 5-14 Rate 5-16 Pulse Shaping 5-18 Pump Ramp Threshold IA 5-11 Rear Panel 3-3 Reduced/insufficient irrigation flow 8-9 Reflux 5-28 Passive 5-28 Remote Control Pairing 6-10 Wireless 2-8 Remote Control, Wireless Setup 6-5 Reordering 11-1 Replacing Batteries 1-14 Fuse 8-2 Reset Timers 3-14, 3-18 Restore Database 6-25 Restore All 6-24 Restore Database 6-23 Restriction of Hazardous Substances 1-22 RoHS 1-22 RX Only – NGP Z370500 Rev. B 1116 12-5 WHITESTAR SIGNATURE PRO System 12 • Index Screen, Touch Calibrate 6-11 Selecting a Language 6-14 Self Test 3-8 Service Interval 6-18 set vacuum units 1-7 Settings Occlusion Mode 5-21, 5-25 Post Laser Cataract 5-10 Setup Advanced Control Pedal 4-16 Surgical Media Center (SMC) 6-15 System 3-2 Wireless Remote Control 6-5 Short Pulse Phaco Power 5-14 Power 5-13 Shutdown Sequence 3-20 System 3-20 Side vitrectomy 5-6 Single Burst Phaco Power 5-15 Power 5-13 Single Linear Foot Pedal Calibration 4-14 Default Settings 4-7 Setup 4-6 Switch Assignment 4-9 Testing 4-14 Threshold Configuration 4-12 Sleep Mode 8-7 SMC (Surgical Media Center) 2-9 Software Versions View 6-12 SOLO Handpiece 3-6 Sounds Set and test 4-33 Specifications Electrical 10-3 Environmental 10-3 Foot Pedal 10-2 Fuse 8-2 IA 10-4 IV Pole 10-2 Phacoemulsification 10-3 Physical 10-2 Vitrectomy 10-4 Startup 3-8 Status 8-14 Sterilization 3-20 Storage Conditions 10-3 Surgeon Add 4-3, 5-2 Assign Order 4-35 Copy 5-29 Surgical Media Center (SMC) 2-9 Setup 6-15 Switch Assignment Advanced Control Pedal 4-17 Advanced Linear Pedal 4-26 Single Linear Foot Pedal 4-9 Symbol Definitions 1-15 System Checkout 3-17 Components 2-2 Console 1-5 Description 1-2 Disposal 1-22 Does not turn on 8-7 Self Test 3-8 Shutdown 3-20 System Date and Time 6-16 System Self Test 6-13 Technical Service 8-2 Test Advanced Control Pedal 4-23 Advanced Linear Pedal 4-32 Foot Pedal Single Linear 4-14 System 6-13 Wireless Remote Control 6-10 Test Chamber 2-3, 11-2 RX Only – NGP Z370500 Rev. B 1116 12-6 WHITESTAR SIGNATURE PRO System 12 • Index Threshold Vacuum 5-24 Threshold Configuration Advanced Control Pedal 4-22 Advanced Linear Pedal 4-30 Single Linear Foot Pedal 4-12 Time Set 6-16 Timers Reset 3-14 Touch Screen Calibrate 6-11 Cleaning 7-2 Does not respond 8-8 Using 3-16 Tubing Cassette 2-3 Install 3-9 Tune 3-11 Begin 3-13 Venturi pump 5-27 View Software Versions 6-12 Vitrectomy 1-7, 1-10, 3-7 Advanced Control Pedal Modes 4-17 Cutter 1-10 Default Settings 5-12 Foot Pedal Modes 4-7 Handpiece 3-7 Prime 3-15 Ultrasonic (U/S) Time 3-21 Ultrasonic Handpiece 1-9 Units of Measure 6-17 Units of measure Set units of measure for vacuum 6-17 Unpair 8-7 Up Time Threshold 5-27 Upper Threshold 5-27 Using too much fluid 8-10 Touch Screen 3-16 Wake-up 2-5 Warnings 1-8, 1-11 WHITESTAR ICE Technology 1-3 WHITESTAR Technology 1-2, 5-15 Wired Pairing 6-7 Wireless Pairing 6-5 Wireless Remote Control 1-4, 1-11, 2-8 Does not pair 8-7 Part Number 11-4 Wireless Remote Test 6-10 Wireless Setup Foot Pedal 6-5 Remote Control 6-5 Yaw 2-7 Vacuum monitoring 5-20 pressure 5-26 Set units of measure 6-17 Vacuum units 1-7 Venting Occlusion 5-26 RX Only – NGP Z370500 Rev. B 1116 12-7 WHITESTAR SIGNATURE PRO System 12 • Index RX Only – NGP Z370500 Rev. B 1116 12-8
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.7 Linearized : No Author : trippek Create Date : 2016:11:22 09:17:27-08:00 Modify Date : 2016:11:22 13:43:32-08:00 Tagged PDF : Yes XMP Toolkit : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26 Metadata Date : 2016:11:22 13:43:32-08:00 Creator Tool : Acrobat PDFMaker 10.1 for Word Document ID : uuid:f06fa182-a2e4-4022-93f2-f1f9c199b7d0 Instance ID : uuid:8757910b-494f-403d-bb1f-88c348cc19d9 Subject : 16 Format : application/pdf Title : signature users manual.book Creator : trippek Producer : Adobe PDF Library 10.0 Source Modified : D:20161122171607 Company : Abbott Laboratories Created : D:20150614 Last Saved : D:20161115 Page Layout : OneColumn Page Count : 33EXIF Metadata provided by EXIF.tools