Instruction ManualPortable ECG Monitor
Thank you for purchasing XYZlife portable ECG monitor. Please read this entire instruction manual.  If you are using this device for the very first time, please follow the instructions and use it correctly.   Please keep the instruction manual for future reference.XYZlife does not bear legal responsibilities for any casualty or product damage caused by users’misuse.
Caution ...........................................................................................................2Safety ConditionsProduct Description and Indication of Use .........................................4Components and AccessoriesProcedure of Set-up ....................................................................................7Initial Set UpXYZlife BC1 Patch Procedure (with Optional Electrode Patch) Battery .............................................................................................................9Important Notes on Taking a Reading  .................................................9Care and Maintenance .............................................................................10Sound and Lighting MessagesApp Error Messages DisplayTrouble ShootingStorageFCC Statement ............................................................................................13Specification ................................................................................................15Table of Contents
Caution2Safety ConditionsThis product is designed to monitor ECG (Electrocardiogram) only. Do not use for any other purpose: such as monitoring life signal or emergency.Use for self-diagnosis or self- treatment is dangerous, please always consult your doctor if you notice any change in your condition, regardless of your measurement result.The measurement result is used for monitor purpose only. Do not read it yourself and conduct any treatment without consult from medical professionals.This product does not support Pacing Spikes function.Patients with cardiac pacemaker are not recommended to use this product.  Notification• Do not sterilize this device in an autoclave or gas sterilizer   (EO, formaldehyde, high density ozone, etc.)• Use at temperature within +10 to +45°C, and humidity between 30 and 85%RH. • Do not wash this device with water.• Do not store the device in the following ambient conditions:   Locations subject to high temperatures and high humidity.   Near fires or open flames.   Locations exposed to strong vibration.   Locations exposed to strong electromagnetic fields.   Locations exposed to direct sunlight.   Wet or damp locations where water may get on the device.   Dusty locations.  Attention  • Do not expose the unit to static environment.   Always avoid any static electricity from your body before handling the device.• The chest electrode patch will be influenced by body hair. Regularly remove    hair at the electrode patch contact point.• Do not take measurement with used electrode patch.• Do not take measurements in a moving vehicle.• Do not disassemble, repair, or modify the device.
Caution3WARNING!• Do not use the device in ICU or operation room.• Do not use the device in hospital or other medical institutions without the   permission of physicians. This product could impact other medical devices.• Do not use the device together with defibrillator (AED).• Avoid flammable materials, anesthetics, drugs or pressurized oxygen    (such as in a hyperbaric chamber, ultraviolet sterilizer or oxygen tent).• Do not attempt self-diagnosis or self-treatment based on the measurement results.    Always consult your doctor. Self-diagnosis or self-treatment may lead to    deterioration in your condition.• Do not use this device on people with sensitive skin or allergic condition.    With some situations, it may cause rashes, skin allergies or eczema.• Please pay special attention to the cable of the charger to avoid the risk    for tangling around the kid’s neck.            • Do not strongly shock or vibrate the device. Do not drop or strike on the device.• Avoid the areas that conduct electricity on the device to touch those of others.• Please make sure your skin is clean and dry, avoid to use after bathing and sweating.• Keep out of reach of children, self-care deficits or people who cannot express their consent.• Do not use for any purpose other than obtaining an electrocardiogram.           • Do not work the device together with other test and measurement    equipments to avoid risks.• Do not take measurements where the device will be exposed to strong    electromagnetic environment.• Do not use the device on the children under age 12.
Single lead portable ECG monitor, is addressed as“device”below.The device is used to measure user’s ECG continuously.Recordings can be transmitted, and viewed by the collaborative App.Product Description4The device consists of three main components:(1) BC1 main unit. It attaches to user’s body via electrode patch (optional)  , using blue tooth to keep communication with mobile devices. (2) App: This App is used to collect, view, store, and wirelessly transfer the readings from device, it must be applied with electrode patch  when taking readings.(3) A user-supplied compatible mobile device.Intended use of the product:BC1, is a portable Electrocardiograph recording device which is designed to be used in combination with an individual’s smartphone to create a portable health monitoring system. The BC1 record and store user’s physical data, wirelessly transmit such data to and display on user’s smartphone. BC1 is intended to be use on users who require monitoring for non-life threatening arrhythmias at home. This device has features of ECG recordings that are available to prescription users only. This device enables a user to:• Detect and store single-channel ECG recordings using the mobile device.• Store user information data associated with the recording.• Request professional clinical interpretation and analysis of your ECG recordings.• Wirelessly transmit ECG recordings in real time to your mobile device.*Determing Heart Rate: use 5 second ECG to caluate R-peak inveral time, take reciprocal and then multiply 60.Important Notification:Please notice that ECG recordings cannot detect all heart symptoms, and are used for physicians’ references only.  Individuals who discover any change on measurement result, or diagnosed with heart disease, regardless of stable readings on electrocardiogram, should always consult doctor.*Photos might have slightly difference on the appearance compared to the real product.
A. Power ButtonABCD EFMain UnitComponent 1 / Main UnitFunctionB. Battery LEDC. Status LEDD. Calling KeyE. Charging holeF. Electroded    BuckleLED light to display current battery life Green light: Full power /  Flashing red light: Charging /Red light: Battery hour is less than 6 hours LED light to show system statusLED light to display current battery life Green light: Power above 60% Red light: Power less than 20%(low power)  Flashing red light: in chargingPress the 1st time to call, 2nd time to cancel callingUnder battery power sufficient circumstance, calling Key is working even the power is off. For charging purpose onlyBuckle on electrode patch  use5Press 1st time to power ON,  2nd time to power OFFProduct Description
Component 2 / Carry BagComponent 3 / ChargerElectrode PatchUSB Connector* Please use charger designated for    XYZlife Patch onlyBuckle connecting pointEasy carry and protection of device6Model:GS2U-006-050-AInput: 100-240V~0.2A 60/50Hz Output: 5V 1200mA 6W Max Charger is complied with 60601-1 3.0 regulationDevice Name: Medi-Trace 200 and Medi-Trace 200-30 ECG Electrodes 510(k) Number: K960968Product Description
Initial Set UpInitial Set Up1.ChargingTake out BC1 main unit, and use USB cable start charging.2.Download BC1 App to your mobile device (The device supports Android4.4 or more updated version, as well as iOS7.0 or more updated version.) Please find and download Apps for XYZlife Patch on Google Play or App store).3.Follow APP instruction to complete initial set up and Bluetooth connection.Set up the device according to selected options7XYZlife Patch  mode set up procedure(with optional electrode patch)1. Sync main unit and Bluetooth connection-Press the power button to turn on power. (see the photos on the right) make sure that battery LED is green.-Sync the device with your mobile APP (1st time initial setting please follow the APP instruction) and make sure the status LED is green (connect successfully). If fail to sync APP, please connect manual following the APP instruction.    2.Electrode patch and adhesive method-At patch contact point, remove body hair or grease.-Buckle up main unit with patch.Remove electrode patch protect film.Stick patch on the center of your chest (see the right photo, The electrode patch should stick vertically with calling key at downward side. WARNING: Calling key must be placed downward; otherwise will lead incorrect readings. Device sticks on center of chest, calling key in parallel with nipples
Initial Set Up83. Calling Key-Press the calling key on the front of the main unit to make warning beep sounds.-After 30 seconds, the main unit will send calling texts to preset contact persons.WARNING:Press calling key again to stop mis-trigger(Warning beep sounds will stop)4. How main unit works with APP-Please keep main unit and mobile device within 8 meters distance-Farer distance or additional blocks might cause data transmitting or reading issuesNote: At lease 5 minutes to accumulate storage data. Data cannot be saved if removed device within first 5 minutes.       5. How to turn off the devicePress the power button on the side ofthe device; the main unit will disconnect automatically to APP on mobile device.
BatteryCaution during measurementBattery Life and Charge1. The product is using Li ion rechargeable battery. 2. Use designated charger to charge the battery. Please charge fully in first time use.3. Battery LED displays GREEN light when charging is completed.4. The battery with full power works continuously for 8 hours.5. Battery can be recharged at 500 times.6. It takes around 2.5 hours to fully charge the battery.To obtain correct ECG readings, make sure you wear it correctly.Please read below notes carefully when using the device for the very first timeMake sure to pay attention to the following notes at each use :• Make sure the electrical buckle up firmly and correctly on patch.  • Excessive body movement could influence the measurement result: including moving, exercising, sneezing, or coughing etc.• Electronic devices can cause interference to measurement readings.• If you have questions about the product, consult local agent or physicians.9
Care and MaintenanceSound and FlashingApp Error Messages   Beep/ 1 short beepContinuous beeps CallingPlease refer to troubleshootingBattery power is less than 60% Sound Message NoteNote1. Confirm POWER ON/OFF2.Detect ECG signalSystem Error Battery LED / Red Lighting  / Red Flashing  / OFF / Red Light / Red Light FlashingStatus LED / Green LightingFlashing Light MessageBattery is less than 20%ChargingConnecting with APPAPP disconnectionSystem Error Please refer to APP Error MessagesECG signal disconnectionError 001Error 002Error 003Error 004Battery warning Low batteryAbnormal sensor signal Storage memoryfailed or damagedPlease Charge1. Reboot the device2. Please contact local agent if Step 1 can not solve the problemError#                Message                    Cause                            SolutionTemp. sensor failedG sensor failedBarometer sensor failedError 005 Internal storage memory error10Beep/ 3 long beepsBeep/ 5 long beep
Care and MaintenanceTroubleshooting1. Mobile device    bluetooth is OFF2. Mobile device bluetooth     version is older than 4.03. Android system is older     than 4.4 or iOS older     than 7.0Problem                               Cause                                             SolutionUnable power ONDevice unable to sync with mobileAbnormal Bluetooth pairingECG data is disappeared while usingThe ECG readings appear upside down, Or unusual readings To many noise or interference in readings. Battery is exhausted and voltage is too lowPlease charge the device, waiting for GREEN light.Make sure the distance between device and mobile device is within 8meters.Sync APP with device manuallyDistance between the device and mobile is too far.1. Battery is exhausted. 2. Bluetooth connection is    not working properly1. Charging battery 2. Make sure that mobile Bluetooth is on and sync with your device. 1. Make sure the electrode patch is    positioned correctly2. Unusual readings appear:- make sure whether you use it    correctly- consult your physicianNoise may come from interference from outside Circumstances.Possible interferences might be:-the device attaches   body improperly-excessive body movement-EMC interference1. Please make sure Bluetooth    is working2. Please make sure mobile    device and Bluetooth version    is updated1. The device is    positioned incorrectly2. ECG signal is weak1. Make sure electrode patch is new 2. Do not exercise excessively while     using the product3. Make sure electrode patch is     positioned correctly4. Keep yourself away from high EMC     Environment, such as stereo, video,     or similar equipments5. Remove body hair and clean the skin     in electrode patch contact point11
Care and Maintenance Storage1. Do not place any object on the top of the main unit, or strike it.   This could damage it.2. Do not wipe the device or electrode connecting point with benzene,     gasoline, paint thinner, concentrated alcohol, or other solvent detergents.3. Do not disassemble the device.  Repairs should only be carried out by    authorized personnel.  Disassemble and repair are for authorized personnel only.4. Do not sterilize this device in an autoclave or gas sterilizer     (EO, formaldehyde, high density ozone, etc.)5. The product does not need regular maintenance or calibration.     If user requests the repair, please contact your local agent or our website. 12
FCC statementEmission test                       Compliance           Electromagnetic environment - guidanceGuidance and manufacturer’s declaration - electromagnetic EmissionsThe BC1 ECG monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the BC1 ECG monitor should assure that it is used in such an environment. Recommended separation distances between portable and mobile RF communications equipment and BC1 monitor BC1 monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the BC1 ECG monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters), the BC1 ECG monitor as recommended below, according to the maximum output power of the communications equipment.               RF emissions CISPR 11 RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations / flicker emissions IEC 61000-3-3 Group1Class BN/AN/AThe device’s RF emissions are very low and are not likely to cause any interferencein nearby electronic equipment. BC1 is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Rated maximum output power of transmitter (W)0.010.11101000.120.381.23.8120.120.381.23.8120.230.732.37.323Separation distance according to frequency of transmitter (M)150 kHz to 80 MHz 80 kHz to 800 MHz 80 kHz to 2.5 GHz13
14FCC StatementImmunity test                        Compliance        Electromagnetic environment - guidanceConducted RF IEC 61000-4-6Radiated RF IEC 61000-4-3 Guidance and manufacturer’s declaration - electromagnetic immunityIEC 60601Test level3V/m3VNOTE 1 - At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which BC1 ECG Monitor is used exceeds the applicable RF compliance level above, BC1 ECG Monitor should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the BC1 ECG Monitor. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.3 V/m 80 MHz to 2.5 GHz 3 Vrms 150 kHz to  80 MHzThe BC1 ECG monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the BC1 ECG monitor should assure that it is used in such an environment.Portable and mobile RF communications equipment should be used no closer to any part of BC1 ECG monitor, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distanceD =1.2√PD =1.2√P 80 MHz to 800 MHz D =2.3√P 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and D is the recommended separation distance in meters (m).Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range. Interference may occur in thevicinity of equipment marked with the right side symbol:
This Device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) This device must accept any interference received, including interference that may cause undesired operation. Notice:The changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.   IMPORTANT NOTE: To comply with the FCC RF exposure compliance requirements, no change to the antenna or the device is permitted. Any change to the antenna or the device could result in the device exceeding the RF exposure requirements and void user’s authority to operate the device.
15SpecificationCE     Theproduct follows EC 93/42/EEC regulationThe important message regarding EMCWith the increase use of electronic devices, such as personal computer or smart phones, any medical device might have risk of being interfered by electromagnetic from them.  Any interference might lead to the medical abnormal function, causing dangerous situations.This medical device should not interfere other devices. EN60601-1-2 regulation has been applied. The purpose for EMC regulation is to screen out insecure products. This regulation defines the bearable interference level from any medical device, and the level of output electromagnetic.XYZlife manufactured medical device is complied with EN60601-1-2:2012 regulation. However, it is strongly recommended to follow WARNING points below:1.Do not use the device near high volt or EMC filed.  This may lead to device abnormal and cause some safety concern.2.It is recommended to keep the distance of the device between   them more than 8 meters3.More EN60601-1-2:2012 documents, please go to www.xyz-life.com   official website.Information on Disposal for Users of Waste Electrical & Electronic Equipment (WEEE)This symbol on the product(s) and / or accompanying documents means that used electrical and electronic products should not be mixed with general household waste.For proper treatment, recovery and recycling, please take this product(s) to designatedcollection points where it will be accepted free of charge.User and product retail shall contact your local authority for further details of recycle this product safely. The product is not allowed to be mix disposal with other commercial waste. The product does not contain hazardous substances.  Battery disposal method shall comply with locally regulation to proceed.xxxx
Model: BC1ECG Channel: single channel ECG wave measuring range : 0-2mV(    0.1mV error) Heart rate range: 25 to 250 BPM (    5BPM)Frequency response : 0.67Hz to 40Hz  / Sample rate: 200HzIngress Protection Rating: IP22Protection type and range against electrical shock: TYPE BF ECG electrode is applied PartBattery: Rechargeable Lithium ion batteryOperation Temp. : +10℃ to +45℃ Operation Humidity : 30% to 95%Operation atmospheric pressure: 700-1060 mbarStorage Temp. : -20℃ to +60℃ Storage Humidity: 10% to 95%Weight: Approximate 37 g (including batteries) Dimensions : 73mm(W) X 38mm(H) X 16mm(D)Battery Spec. : Li-ion, 620mAh Charger Spec. : Input: 100-240V~0.2A 60/50Hz / Output: 5V 1200mA 6W MaxContents : BC1 main unit, Charger (comply with medical device regulatory), Carry bag, Instruction Manual, Quick guide/ ManualOther symbols: Please carefully read the manualNOTE: Specifications may be changed without prior notice.Manufacture: Kinpo Electronics, Inc. Address: 10F, No. 99, Sec5, NanJing E. Rd, Taipei City 105, TaiwanWebsite: www.xyz-life.comManufactured in Thailand Manufacturer16Specification
17Symbols Used for Product or PackageManufacture: Kinpo Electronics, Inc. Address: 10F, No. 99, Sec5, NanJing E. Rd, Taipei City 105, TaiwanWebsite: www.xyz-life.comManufactured in Thailand Manufacturing batch / lot No.Protection type and range against electrical shock(Body Floating isolated applied part)Date of manufacture,Century year xxxx month xx date xxEuropean Conformity MarkPlease follow WEEE regulation for properly disposalTo signify that the instruction manual/booklet must be readSymbols MeaningsRefer to instruction manual/bookletTEL: +31 658 768 735
Symbols Used for Product or Package18Serial Number of productWARNING and CautionsPlease follow the instructions carefullyTransportation and storage temp. requirementTransportation and storage humanity requirementWarning symbol denoting a device's susceptibility to electrostatic discharge.Radio frequency fields beyond this point may exceed the FCC, etc.The device complies with Federal Communications CommissionRecycling SymbolInternational protecting markSymbols Meanings