Millar Instruments HU72 Short range telemetry transceiver User Manual Telemetry Unit

Millar Instruments Ltd Short range telemetry transceiver Telemetry Unit

USERS MANUAL

Download: Millar Instruments HU72 Short range telemetry transceiver User Manual Telemetry Unit
Mirror Download [FCC.gov]Millar Instruments HU72 Short range telemetry transceiver User Manual Telemetry Unit
Document ID2228150
Application IDX++J3ejz4pgLg908/rw82w==
Document DescriptionUSERS MANUAL
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize327.54kB (4094218 bits)
Date Submitted2014-03-28 00:00:00
Date Available2014-03-28 00:00:00
Creation Date2014-02-27 16:08:13
Producing SoftwareMicrosoft® Word 2013
Document Lastmod2014-02-27 16:08:13
Document TitleTelemetry Unit
Document CreatorMicrosoft® Word 2013
Document Author: mdon056

Wireless
Solutions
for
Physiological
Monitoring
User Manual for Small
Animal Systems
Important Information
Disclaimer
Millar believes that information in this user manual is correct at the time of publishing. In the event errors may exist,
Millar reserves the right to make changes to future editions without notice to holders of this edition. The reader
should consult Millar if errors are suspected. In no event shall Millar be held liable for damages arising out of the
use of this document or information contained within. Except as specified below, Millar makes no warranties,
express or implied, and specifically disclaims any warranty of the products fitness for a particular purpose.
The customer’s right to recover damages by fault or negligence of Millar are limited to the amount paid by the
customer. Any action against Millar must be brought within one year after the cause of action accrues. Millar will
not be held liable for any loss of data, research funding or profits and neither will it be liable for incidental or
consequential damages even if advised of the possibility thereof.
Millar expresses that its products are for research purposes only. They are not designed with components
and testing intended for use in medical treatment or diagnosis of humans or animals and doing so may
result in serious harm.
Warranty
Telemeter Warranty
Millar’s comprehensive telemeter warranty guarantees that telemeters will function out-of-box for the first six months
after factory shipment. Additionally, Millar provides a sixty (60) day in-animal telemeter warranty following the first
implantation within the six month warranty period. Subsequent implantations are outside the warranty. Any
telemeter not registered within six months after factory shipment will be out of warranty.
In order to provide the highest level of customer service, Millar requires all new telemeters to be registered
upon the first implantation. To register the telemeter’s in-animal warranty, simply send an email to
support@millar.com and include the following:

Telemeter Model and Serial Number

Facility Name

Primary Investigator / Laboratory Head name and phone/email

Primary lab contact name and phone/email

Implantation Date
Multiple telemeters can be activated with a single email.
For complete telemeter handling and storage instructions, refer to this manual.
Telemetry Hardware Warranty
Millar warrants that at the time of sale to the original Purchaser, the telemetry hardware (excluding telemeters – see
above) shall be free from defects in materials and workmanship for a period of one (1) years from its date of
shipment to the original purchaser. If there is such a defect, Millar will, at no charge and at its option, either repair
or replace the equipment as appropriate.
Should the hardware become damaged after expiration of the one year warranty, Millar will accept the product for
a no charge evaluation with a signed RMA (return material authorization). If repairs are required, Millar will provide
a detailed quote and lead time for the repair process. Repair work will not commence until approval is received
from the customer.
As part of Millar’s continued commitment to customer support, free online, phone and email support in the use of
all telemetry equipment is available beyond all warranty periods.
The warranties above do not cover damage caused by failure to follow instructions for use, owner’s abuse, misuse
or negligence, user tampering or modifications, power failures or surges and events or accidents considered force
majeure such as flooding, fire, etc. outside the reasonable control of Millar. In no case shall liability exceed the
purchase price of the original product.
Millar provides free product evaluations during the warranty period and free return to customer shipment. All
returned products should be packed safely, preferably in original packaging and shipped to Millar with the RMA
number visible on the packaging. In order to receive the highest level of support, the customer is responsible for
notifying Millar immediately of a problem within the warranty period. Following product evaluation, Millar will replace
or repair (telemeters are non-repairable) any product found to be defective and covered within the warranty period,
while operated in accordance with instructions for use and specifications.
For complete product instructions, general insights and surgical recommendations, visit millar.com and access
Millar’s free online Knowledge Center. http://millar.com/knowledge-center Millar is committed to delivering the
collaborative support you expect in facilitating scientific insight and advancing medical innovation.
Copyright © Millar March 2013
This publication may not be reproduced or transmitted in any form, electronic or mechanical, including
photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the
prior written consent of Millar.
Contents
FCC Compliance……………………………………………………………………………..1
Introduction.................................................................................................................. 3
Telemetry System Components .................................................................................. 4
Configurator System Setup ......................................................................................... 7
Configurator Software Installation ....................................................................................................... 7
Connecting the Configurator ............................................................................................................. 10
SmartPad Setup ........................................................................................................ 13
The Telemeter ........................................................................................................... 16
Configurator Software - Overview ............................................................................. 17
Pairing a SmartPad and Telemeter ........................................................................... 19
Troubleshooting and frequently asked questions ............................................................................. 24
Calibration values ...................................................................................................... 26
Telemeter general use .............................................................................................. 28
Telemeter Modes ...................................................................................................... 29
Placing a Telemeter into Safe Mode ......................................................................... 30
Positioning the Telemeter during surgery .................................................................. 31
Construction of Mesh Pouches.................................................................................. 33
Surgical Mesh for BP Catheter Implantation.............................................................. 34
Telemeter Care and Handling ................................................................................... 35
Telemeter Removal and Preparation for Reuse ........................................................ 36
Sterilization ................................................................................................................ 37
Telemeter Diagnostics ............................................................................................... 39
SmartPad Diagnostics ............................................................................................... 40
SmartPad Technical Specifications ........................................................................... 42
Telemeter Technical Specifications ........................................................................... 43
Oxygen Telemeter Technical Specifications.............................................................. 44
Frequently Asked Questions........ ............................................................................. 45
Contact Information ................................................................................................... 46
To all users;
We pride ourselves on helping you achieve results. If you have any questions related to the
operation of your system, please contact us support@millar.com
We are here to help.
Contact Information
US Head Office
New Zealand Office
Houston, Texas
Auckland, New Zealand
Toll free (in USA) 1 800 669 2343
TEL: +64 9 367 7126
TEL: +1 832 667 7000
FAX: +64 9 367 7157
FAX: +1 713 714 8497
EMAIL support@millar.com or sales@millar.com
WEB www.millar.com
FCC Compliance
Telemeter (TRM54 series)
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference that may cause
undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
 Reorient or relocate the receiving antenna.
 Increase the separation between the equipment and receiver.
 Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
 Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by Millar Inc. could void the user's authority
to operate the equipment.
TR190 Configurator
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference that may cause
undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
 Reorient or relocate the receiving antenna.
 Increase the separation between the equipment and receiver.
 Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
 Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by Millar Inc. could void the user's authority
to operate the equipment.
Millar Telemetry User Manual for Small Animal Systems
v2.1
TR181 SmartPad
Millar Instruments Ltd TR181 SmartPad
The Wireless Power Transfer System is FCC Part 18 Compliant
Note: This equipment has been tested and found to comply with part 18 of the FCC limits for
non-consumer equipment. These limits are designed to provide reasonable protection against
harmful interference in an industrial installation. This equipment generates uses and can
radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:—
 Reorient or relocate the receiving antenna.
 Increase the separation between the equipment and receiver.
 Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
 Consult the dealer or an experienced radio/TV technician for help.
Note: There is no required maintenance of this device from a FCC compliance perspective.
The 2.4GHz Data Transmitter is FCC Part 15 Compliant
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions:
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including interference that may cause
undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or
more of the following measures:
 Reorient or relocate the receiving antenna.
 Increase the separation between the equipment and receiver.
 Connect the equipment into an outlet on a circuit different from that to which the
receiver is connected.
 Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by Millar Inc. could void the user's authority
to operate the equipment.
Millar Telemetry User Manual for Small Animal Systems
v2.1
Introduction
This User Manual contains information relating to the proper installation and basic use of your
telemetry system. It is strongly recommended that users visit our online Knowledge Center for
more detailed information and to view instructional videos showing implantation, explantation
and cleaning techniques. http://millar.com/knowledge-center
All small animal telemeters use a unique patented inductive power transfer system for
recharging and making long-term recordings in small animals 175 grams and over.
Please note that if your system has been customized in any way parts of this manual may not
apply to your system. If you have any questions please contact us at support@millar.com
Key Features
Accuracy:
For long term, high frequency physiological recordings. Digital wireless recording of data with
the highest quality signal integrity.
24/7:
Data can be collected continuously - 24 hours a day while the animal is on the SmartPad, with
no need to replace batteries.
Versatility:
Telemeters send data wirelessly from within the animal to the SmartPad where it is converted
to an analog signal compatible with any data acquisition system.
Low cost of ownership:
Battery replacement is not required. With user care, telemeters can be re-charged and reused.
Efficient housing:
Animal cages can be next to each other in one location for simultaneous data collection from
multiple animals. No specialized cages or shielding is required as each telemeter operates
digitally using one of 30 independent frequencies with no interference to recordings.
Service & support:
Millar prides itself on being responsive to our customers’ current and future needs. We have
experienced engineers and scientists on staff to provide professional, expert and timely
support and advice.
Millar Telemetry User Manual for Small Animal Systems
v2.1
Telemetry System Components
The Telemeter
Care must be taken not to damage the telemeter body, catheter or electrodes. Telemeters are
not repairable! Depending on the model, your telemeter may look slightly different to images
throughout this manual; however all telemeter models have the same basic construction and
operation. All telemeters are shipped individually packaged and sterile ready for use. All
telemeters will need to be activated and configured prior to implantation. PLEASE NOTE the
telemeter does NOT need to be removed from its packaging for Activation and Configuration.
The SmartPad
Do not operate the SmartPad on or within 10 cm of a metal surface or within 40 cm above or
below another SmartPad or Wireless Power Charger.
The Configurator System
The Configurator System includes the Configurator hardware and ConfigSoft software.
Together they are used to wirelessly control telemeters and SmartPads to:
 Change/configure a telemeter and SmartPad Channel frequency.
 Run and provide diagnostic information about telemeters (such as battery charge) and
SmartPads.
 Place telemeters into Safe Mode (for storage).
 Control the charge-field of the SmartPad.
ONLY a single Configurator System is required to control all telemeters and SmartPads in a
laboratory. The Configurator System is not required for data acquisition.
** IMPORTANT: The ConfigSoft software supplied on the USB drive must be installed on a PC
before connecting the Configurator hardware to the computer.
Millar Telemetry User Manual for Small Animal Systems
v2.1
Millar Telemetry User Manual for Small Animal Systems
v2.1
6
Millar Telemetry User Manual for Small Animal Systems
v2.1
Configurator System Setup
Introduction
The Configurator System (Configurator hardware and ConfigSoft), wirelessly communicates
with all telemeters and SmartPads. The system allows the user to:
1)
2)
3)
4)
5)
Change/configure Channel frequencies.
Turn telemeters into Safe Mode for storage.
Check telemeter battery level.
Enable or disable the charging field of the SmartPad.
Run telemeter and SmartPad Diagnostics.
Only a single Configurator System is required for each laboratory and is NOT required for data
acquisition.
Configurator Software Installation
NOTE: ConfigSoft must be installed on the PC before connecting the Configurator hardware.
System requirements; Windows XP, Vista (32 bit) or 7 (32 & 64 bit) and
USB 2.0 port (Windows 8 and USB 3.0 are not supported).
Internet connection is necessary to access and install any drivers not
already installed.
1)
2)
3)
Insert the supplied USB drive containing ConfigSoft into a PC USB
port.
Open the folder “ConfigSoft”.
Double click on “setup” to start the installation.
Depending on the files and drivers already installed on your computer some of the following
steps may not be necessary.
4)
Accept the license agreement for.NET Framework.
Millar Telemetry User Manual for Small Animal Systems
v2.1
5)
The Microsoft .NET Framework will be downloaded from the internet if not already
installed (this may take some time). Follow all instructions to complete the installation.
6)
Follow the instructions for installing ConfigSoft. Click Next.
7)
Click Next to use the default folder or choose another folder to install ConfigSoft.
Note: We suggest that you choose the option to install ConfigSoft for anyone who uses this
computer.
Choose this option
Millar Telemetry User Manual for Small Animal Systems
v2.1
8)
Confirm installation. Click Next
9)
Install the Driver for the TR190 Configurator. Click Install.
10)
Ignore the software Windows Logo warning. ConfigSoft and drivers are safe to install.
Do not hit the Enter key but make sure to Click “Continue Anyway”.
11)
Success: Click OK
Millar Telemetry User Manual for Small Animal Systems
v2.1
12)
ConfigSoft is now installed. Click Close
Connecting the Configurator
Front
Back
Antenna
connection
Status light
USB port
1)
Attach the antenna to the back of the Configurator.
2)
After installation of the software ConfigSoft, connect the Configurator to your PC using the
supplied USB cable. The status light on the front of the Configurator should glow red
indicating that it is receiving power from the PC.
The status light (front) on the Configurator indicates the following information:
Off
Not connected, USB cable unplugged Configurator is off
Red
Configurator is receiving power but not communicating with PC
Green
Configurator is communicating with the PC
10
Millar Telemetry User Manual for Small Animal Systems
v2.1
3)
The PC will open the “Found New Hardware” Wizard window. Choose the “No, not this
time” option and click Next.
4)
Click Next to install the driver automatically.
5)
Ignore the driver signing warning. Configurator drivers are safe to install. Click “Continue
Anyway”.
Millar Telemetry User Manual for Small Animal Systems
v2.1
11
6)
The driver may take a few moments to install.
7)
If successful the Status light on the front of the Configurator should now be green
indicating that it is communicating with the PC. Click Finish to close the Window.
8)
The Configurator System is now ready for use with all your telemeters and SmartPads.
12
Millar Telemetry User Manual for Small Animal Systems
v2.1
SmartPad Setup
The SmartPad is a receiver and charger in one device. Do not operate a SmartPad near a
metal surface (unless the charging field has been disabled using the Configurator System).
All SmartPads are shipped set to Channel “Setup”. Before using a SmartPad with a telemeter
you need to configure and pair each SmartPad with the same Channel of the telemeter to be
used. It is important that only one SmartPad is used on any particular channel or interference
will occur.
Front
Status light
Back
Power socket
Analog Outputs
Status Light information:
Orange
The SmartPad is turned on but is not communicating with a telemeter
Green
The SmartPad is communicating with a telemeter on the same Channel
Flashing orange The SmartPad is in diagnostic mode. No telemeter data will be output.
Red
The SmartPad has automatically disabled the charging field due to
detection of a high current. In this mode data output will continue but
battery charging is not occurring. Cause: may be due to the SmartPad
being placed too close to a metal surface or to another SmartPad. Move
the SmartPad away from any metal and then reset by unplugging the power
and plugging it back in (see page 13).
1)
Connect the supplied Power Supply to the Power Socket on the back of the SmartPad.
2)
The Status light will be Orange.
3)
Use the Configurator System to select the Channel to match the telemeter being used.
4)
Status light will change to Green when SmartPad and telemeter Channel are matched.
5)
By default the charge field of the SmartPad is enabled, if you need to disable the charge
field use the Configurator System.
Millar Telemetry User Manual for Small Animal Systems
v2.1
13
Analog Outputs
After pairing the SmartPad/telemeter Channels the SmartPad automatically detects the type
of telemeter it is communicating with and the analog outputs on the back will correspond to the
following signals (all analog outputs are low pass filtered at 1 kHz).
Telemeter model
TRM54P
TRM54PB
TRM54PP
Output 1
Pressure
Pressure
Pressure
Biopotential 1
(Green/Yellow)
Unused
Oxygen
Pressure
TR50BB
TR50B
TR57Y
TRM56SP
Output 2
Unused
Biopotential
Pressure
Biopotential 2
(Red/Black)
Biopotential
Unused
SNA
Output 3
Temperature
Temperature
Temperature
Temperature
Temperature
Temperature
Temperature
Dual Pressure Telemeter (TRM54PP) Catheter Outputs
When looking at the TRM54PP dual pressure telemeter with the label facing down (as if the
telemeter is placed in the animal), the SmartPad output 1 will correspond to the pressure
catheter furthest from the charging coil.
Charging coil
Output 2
Output 1
Suture tabs
14
Millar Telemetry User Manual for Small Animal Systems
v2.1
Cage Positioning on SmartPad
The SmartPad is designed to charge small
animal telemeters while they are implanted in
the animal. To do this place the animal cage
on the SmartPad so that it is against the
raised edges in the back left hand corner.
To protect the telemeter battery, the
SmartPad automatically controls the strength
of the charging field using information from
the telemeter. It is important that telemeters
are only placed on SmartPads set to the
same channel otherwise over or under charging may damage the battery. The Charging field
of the SmartPad is also set to cycle automatically and turns off for 30 minutes every two hours,
this however does not significantly affect recharging of the telemeter battery.
Shelving Setup
The SmartPad charges the telemeter using an electromagnetic field generated above and
below the SmartPad. For this reason it is important that a SmartPad is NOT placed on a metal
surface or above or below other SmartPads (unless 40 cm apart). If the SmartPad is to be
used near metal, for data collection only, then the charging field must be disabled using the
Configurator System. (Please note that with the field disabled, the telemeter will not recharge
and recording time will be limited to 4-6 hours).
Plastic shelving is best for your SmartPads.
Arrange your SmartPads so that there is at
least 10cm horizontal and 40 cm vertical
separation between them. If using shelving
that has metal supports or framing then the
SmartPads must be >10cm away from
these as well (don’t forgot to check the
underside of the shelves for metal
supports). There should also be at least 10
cm separation between the top of a metal
cage lid and the SmartPad on the shelf
above.
>10cm
>10cm
>10cm
>40cm
If metal shelves are used, while the charging
field is enabled, the SmartPad needs to be
raised at least 10 cm from the shelf surface
but you must maintain a vertical separation
of at least 40 cm between SmartPads.
If your SmartPad shows a red light, then you need to separate your SmartPads further
(vertically and horizontally) and/or move them further away from any metal brackets, supports,
cables or power packs. Reset the SmartPad by unplugging the power and plugging it back in.
If the problem continues, please contact us support@millar.com
Millar Telemetry User Manual for Small Animal Systems
v2.1
15
The Telemeter
Telemeters are shipped sterile ready for implantation. ALL
telemeters are shipped on Channel “Setup” and each
telemeter must be configured to one of 30 different
channels before use. There is no need to unpack the
telemeter from the sterile packaging (see right) to change
the channel. Please note that the communication
channels of the current telemeters and SmartPads are the
same as the old model telemeters, receivers or wireless power chargers (pre 2011). Any other
devices on the same channel will interfere with SmartPad communication.
The telemeters are fragile electronic devices. Care must be taken not to damage the telemeter
body, catheter or electrode as they cannot be repaired. Please read the section about
Handling and Care of your telemeters. Depending on the model, your telemeters may look
slightly different to those pictured throughout the manual; however their care, handling,
preparation and use are the same.
Model number
Serial number
Charging coil
Antenna
Suture tabs
Pressure Sensor
Telemeter models that include pressure
measurement are fitted with a solid state pressure
sensor from Millar. The sensor is at the end of the
catheter (see below) allowing highly accurate
measurement of pressure signals directly where
the tip is located. The sensing tip should be
handled carefully. Never lift the telemeter by the tip, pierce the tip with a sharp object or hit
the tip against a hard surface.
Temperature measurement
All telemeters record temperature. When collecting temperature data from the telemeter at the
same time as charging, temperature output form the SmartPad is updated once every two
hours. Continuous temperature measurement is available if the SmartPad charging field is
disabled (see page 21).
Telemeter Storage and Battery Charge
If a telemeter is to be stored for any length of time it is recommended that the telemeter be
turned on and the battery level checked. If the telemeter has not been used for up to 6 months
it should be turned on and the battery checked using the Configurator System. After turning
on the telemeter, the battery should be checked using the Configurator system and while it is
away from the charge field of the SmartPad. Before placing the telemeter back into Safe Mode
it should be fully charged (90-100%).
16
Millar Telemetry User Manual for Small Animal Systems
v2.1
Configurator Software - Overview
ConfigSoft can be started from either the

Desktop icon
OR

Start>Programs>Telemetry Research>ConfigSoft
From the main menu the user has access to the:
1. Telemeter Configuration and Diagnostics
2. SmartPad Configuration and Diagnostics
Millar Telemetry User Manual for Small Animal Systems
v2.1
17
Telemeter Configuration and Diagnostics Dialog
SmartPad Configuration and Diagnostics Dialog
18
Millar Telemetry User Manual for Small Animal Systems
v2.1
Pairing a SmartPad and Telemeter
Before implanting a telemeter into an animal, a SmartPad and telemeter need to have their
Channels paired. Please ensure that only one SmartPad and one telemeter are configured
and paired at any one time. All SmartPads and telemeters are shipped on Channel “Setup”
as default. When pairing a telemeter/SmartPad make sure any other SmartPads (on the same
Channel) are turned off until needed and any other telemeters (on the same Channel) are in
Safe Mode and remain at least 2m away from any active SmartPad.
1)
Turn on the SmartPad (Status light should be orange).
Insert power plug here
2)
Place a telemeter (still in its sterile packaging) on the SmartPad. This will automatically
turn the telemeter into Active Mode (data transmission active).
3)
If the Status light on the front of the SmartPad glows green then the telemeter and
SmartPad are on the same Channel, and paired for data collection.
If the SmartPad Status remains orange the SmartPad and/or telemeter need to have their
Channels changed/paired using the Configurator System.
Changing Telemeter Channel
1) After activating remove the telemeter from the SmartPad and place it 1m away from the
SmartPad.
2) Open ConfigSoft
3) Click on “Telemeter Configuration and Diagnostics”
Millar Telemetry User Manual for Small Animal Systems
v2.1
19
Telemeter Configuration
4)
Enter the serial number of the telemeter. The number is on the outside of the telemeter
box or on the round label on the telemeter under the model number.
Telemeter
serial number
TRM54P
2519
7)
Select or enter the current Channel number of the telemeter (first use: Channel “Setup”).
To change the Channel number, select or enter the new Channel number in “Configure
your Telemeter”.
Click Go to activate the new Channel.
8)
Click OK to confirm the choice.
5)
6)
20
Millar Telemetry User Manual for Small Animal Systems
v2.1
9)
A message confirming the new telemeter Channel should appear. To make it easier for
the future, record the telemeter Serial number and Channel number for all telemeter
configurations.
If using a SmartPad with a Channel other than 3, as in this example the Status light will become
orange. To change the SmartPad Channel see page 19: SmartPad Configuration.
Troubleshooting: Configurator - Telemeter
If the Configurator cannot communicate with the telemeter an error message will appear.
Solution: Do the following and try again
 Check that you have entered the correct telemeter Serial number.
 Check that you have entered the correct telemeter Channel number.
 Move the telemeter closer to the Configurator.
 Check the telemeter is not on an active SmartPad.
 Move other telemeters away.
Millar Telemetry User Manual for Small Animal Systems
v2.1
21
SmartPad Configuration
1)
To change the SmartPad to work on the same Channel as a known telemeter, click on
“SmartPad Configuration and Diagnostics” Tab.
2)
Enter the serial number of the SmartPad (found on the front and back of the SmartPad).
3)
Enter current Channel number of the SmartPad (first use: Channel “Setup”).
4)
Enter or select a new Channel number for the SmartPad (should be same as telemeter
being used).
5)
Ensure “Change transmission channel to” is checked.
6)
Click Go to activate the new Channel.
7)
Click OK to confirm your choice.
22
Millar Telemetry User Manual for Small Animal Systems
v2.1
8)
A message confirming that the SmartPad channel has been changed will appear. Record
the SmartPad serial number and new Channel number for future reference.
9)
If using a telemeter on Channel 3, in this example, the SmartPad Status light should
change to green. The telemeter and SmartPad are now on the same Channel, paired,
and able to communicate for data collection and transmission.
If the Configurator cannot communicate with the SmartPad you will get the following error
message.
Click OK and perform the following then retry:



Check that the light on the front of the SmartPad is glowing either orange or green.
Check that you entered the correct Serial and/or Channel numbers.
Move SmartPad closer to Configurator.
The telemeter is now in Active Mode and sending data. The battery life in this mode is
continuous if on a SmartPad or 4-6 hours away from a SmartPad (dependent on telemeter
type and fully charged status). If data collected is not required immediately, we recommend
you remove the telemeter from the SmartPad and place it into Safe Mode using the
Configurator System.
Repeat the above process for each of the SmartPads and telemeters to configure each pair to
a different Channel.
Your telemetry system is now configured and ready for use.
Charging field
The Charging field option in the SmartPad Configuration and Diagnostics Dialog allows the
user to enable (turn on) or disable (turn off) the charge field of the SmartPad. By default when
the SmartPad is powered the charge field will be turned on. Turning the charge field off may
be useful if recording from an animal while it is in an experimental chamber, running wheel,
water maze etc where the SmartPad will be near metal. The telemeter battery will not charge
when placed on a SmartPad with the field disabled so data collection will only be possible for
approximately 4 continuous hours. After this time, the telemeter will need to be placed on a
SmartPad with the field enabled for a minimum of 5 hours.
Millar Telemetry User Manual for Small Animal Systems
v2.1
23
Troubleshooting and frequently asked questions
What if I do not know the Channel setting of the telemeter?
With the telemeter in Active mode press the “Search for Telemeter” button on the Telemeter
Configuration & Diagnostics dialog of ConfigSoft. The Configurator will search through each
of the available channels and report any telemeters which are in Active Mode and within range
and report the serial number and Channel.
What if I do not know the serial number of a telemeter already implanted in an animal?
It is always best to record the telemeter serial number and Channel at the time of surgery,
however if you do not know the Channel or the serial number of the telemeter use the “Search
for Telemeter” in ConfigSoft and use a process of elimination. The Configurator will search
through each of the channels and report any telemeters active and within range. If you have
more than one unknown telemeter move any other telemeters out of the detection area (more
than 5m away).
24
Millar Telemetry User Manual for Small Animal Systems
v2.1
The Search for Telemeter function does not find the telemeter. What do I do now?
There are 4 possible explanations:
1)
2)
3)
4)
The telemeter is in Safe Mode
The battery of the telemeter is discharged (and the telemeter has switched to Safe Mode)
The telemeter is too far away from the Configurator
The telemeter has been configured to the same channel as another telemeter
Solutions
1)
Move the telemeter closer to the Configurator and run the search again.
2)
Place the telemeter on a SmartPad in the region as shown on page 26 with an active
charging field. Run the search again.
Try turning other telemeters to Safe Mode or move telemeter to another room (>5 m away)
and run the search again.
3)
Now that the telemeter has been found check the battery level (see Telemeter Diagnostics
section). If the battery level is below 30% the telemeter needs to be charged before the
Channel can be changed. Place the telemeter on a SmartPad with the same Channel and
leave it to fully charge (90-100%).
What if I do not know the Channel setting of a SmartPad?
Go to the SmartPad Configuration & Diagnostics dialog. Enter the serial number of the
SmartPad, make sure the SmartPad is turned on and press the “Search for SmartPad” button.
The serial number of the SmartPad can be found on the front and back of the SmartPad.
Millar Telemetry User Manual for Small Animal Systems
v2.1
25
Calibration values
Pressure calibration
Telemeter Pressure Recordings in the Animal
Telemeters are shipped sterile and dry, when implanted pressure values recorded in the first
48 hours after surgery may have an offset of up to 20 mmHg.
Before making any recordings from the implanted telemeter it is important that the telemeter
has been turned on for a minimum of FOUR (4) hours.
Calibration & Offset Testing
During factory calibration, we attempt to ensure the environment is as similar as possible to
what the telemeter will experience when implanted in an animal. Users who wish to follow or
check the offset either pre or post implantation need to precisely set the environment for the
offset test as detailed below. Please Note: Telemeters should remain immersed throughout
the testing procedure.
Key factors that need to be controlled are;
1. Hydration; Telemeters should be soaked for a minimum of 48 hours prior to offset
check. It is important to ensure that, when checking the offset, the sensor tip is no
more than 5 mm under the water so as to avoid a head of pressure effect.
2. Switch on time; On the bench, all telemeters have a signal settling period when first
switched on. In order to correctly account for this “switch-on effect”, telemeters should
be in active mode on a SmartPad for a minimum of FOUR (4) hours prior to checking
the offset pressure.
3. Telemeter Temperature; Telemeters should be held at 38°C for a minimum of 30 min
prior to recording values.
4. Light; Telemeters need to be in the dark.
If testing at the time of explant this process can be simplified:
1. The telemeter can be removed from the animal whilst still in active mode, put
immediately in a Tergazyme solution and placed on the SmartPad.
2. Once cleaned, it can be placed in warm water at 38°C for 30 minutes in the dark. The
output voltage obtained from the SmartPad at this zero pressure level should be close
to 1.0 V.
Telemeters are all factory calibrated using the tightly controlled conditions as listed above.
Under these conditions the voltage analog output from the SmartPad corresponds to the
following pressures:
1V = 0 mmHg
2V = 100 mmHg
Biopotential calibration
Prior to shipping, all telemeters are calibrated so that the voltage analog output from the
SmartPad corresponds to the following voltages:
0V = -2 mV
2V = 0 mV
4V = 2 mV
26
Millar Telemetry User Manual for Small Animal Systems
v2.1
Oxygen calibration
Prior to shipping, all telemeters are calibrated so that the voltage analog output from the
SmartPad corresponds to the following voltages:
1V = 0 nA
2V = -200 nA
Calibrations coefficients will also be supplied with the electrodes which will need to be used in
calculating the correct oxygen concentration.
SNA calibration
Prior to shipping, all telemeters are calibrated so that the voltage analog output from the
SmartPad corresponds to the following voltages:
0V = -60 µV
2V = 0 µV
4V = 60 µV
**SmartPad outputs are low pass filtered at 1 kHz. Additional filtering, rectification and signal
integration can be performed using a data acquisition system.
Temperature calibration
Prior to shipping, all telemeters are calibrated so that the voltage analog output from the
receiver corresponds to the following temperatures:
0V = 0°C
1V = 20°C
Millar Telemetry User Manual for Small Animal Systems
v2.1
27
Telemeter general use
Before implanting a telemeter make sure to write down the serial number and Channel setting.
Channel Number
All telemeters are factory set to Channel “Setup” and need to be activated and configured to
different channels before use. The Channel is set using the Configurator System (page 18).
Telemeter Label (example)
TRM54P
4168
The telemeter model is TRM54P, the Serial number is 4168. The serial number is also on the
outside label of the telemeter packaging.
Telemeter recharging
Telemeters are designed to be recharged either before implantation or inside the animal. It is
recommended to implant the telemeter fully charged (a value 90-100% when the telemeter is
off the SmartPad).
To efficiently charge the telemeter before implantation place the telemeter on the SmartPad in
one of the two areas shaded on the picture below. Make sure the SmartPad and telemeter
are paired (i.e. set to the same channel). Full charging may take up to 5 hours. For the best
estimate of the battery charge remove the telemeter form the SmartPad and use the
diagnostics section of ConfigSoft to determine the battery level. For maximum battery
capacity, we recommend charging the telemeter for a further hour after the battery level
reaches 90-100%.
For effective recharging within the animal, the telemeter body must lie parallel with the
SmartPad when the animal is mobile. The telemeter body needs to be implanted within
the abdominal cavity and attached to the abdominal muscle wall using the suture tabs.
Charging will not be effective if the telemeter is placed on the side or on the back of the
animal.
28
Millar Telemetry User Manual for Small Animal Systems
v2.1
Telemeter Modes
Each telemeter has two modes:
Safe Mode - deactivated, not sending data
Active Mode - sending data
Safe Mode
All telemeters are shipped in Safe Mode. This mode is used when the telemeter is being stored
and has an extremely low current drain
To place a telemeter into Safe Mode:
o The telemeter needs to be at least 1m away from any active SmartPad.
o Use the Configurator System to place the telemeter into Safe Mode.
o Take care not to allow the telemeter to come within 1m of any active SmartPad or the
telemeter will be activated by the field and will enter Active Mode. This will happen
regardless of which Channel the SmartPad is configured to.
Long Term Storage: Always fully charge your telemeter before storing. We recommend
that the battery level of Small animal telemeters be checked regularly while in storage.
At minimum the telemeter should be checked if it has not been used for 6 months. To
do this, activate the telemeter by placing it on a SmartPad and run telemeter diagnostics
in ConfigSoft. If the battery is not fully charged (90-100%) the telemeter should be recharged. Don’t forget to put your telemeter back into Safe Mode before storing.
Active Mode
The telemeter must be in Active Mode to transmit data.
To change the telemeter to Active Mode from Safe Mode:
o Place the telemeter, in its sterile packaging, on the SmartPad and plug the power cable
into the SmartPad. This will automatically switch the telemeter into Active Mode. Be
aware that the telemeter is now sending data and using battery life.
o If the SmartPad is configured to the same transmission channel as the telemeter, the
Status light on the front of the SmartPad will be green.
Safe
Mode
Place on
active
SmartPad
Remove from
SmartPad (1m),
Use Configurator
System
Active
Mode
Mode
If monitoring is not taking place for a period longer than 7 days it is recommended
that the Telemeters be fully charged and placed in Safe Mode to protect battery life
(damage to the telemeter can occur if left in a fully discharged state).
Millar Telemetry User Manual for Small Animal Systems
v2.1
29
Placing a Telemeter into Safe Mode
1)
2)
3)
4)
5)
Connect Configurator to the computer with ConfigSoft loaded.
Start ConfigSoft.
Make sure the telemeter is not within 1m of an active SmartPad.
Enter telemeter Serial and Channel numbers.
Under “Put Telemeter into Safe Mode” Press Go.
6) Click OK to confirm.
7) The telemeter is now in Safe Mode. Make sure the telemeter does not come within 1m
of any active SmartPad or the telemeter may become Active again.
30
Millar Telemetry User Manual for Small Animal Systems
v2.1
Positioning the Telemeter during surgery
Charging and Signal Maximization
For best charging performance, the telemeter must be implanted within the abdominal cavity
and parallel with the SmartPad. Securing the telemeter to the abdominal wall using the suture
tabs places it in the best position for recharging and signal communication. Never secure the
telemeter by suturing around the pressure catheter or electrode leads. The pressure catheter
can be inserted into the abdominal aorta for arterial pressure measurement and/or electrodes
can be tunneled subcutaneously to the recording site. More information and surgical videos
are available online at http://millar.com/knowledge-center
Charging
coil
Suture tabs
Antenna
Use of the Suture Tabs
Telemeters are shipped with two suture tabs attached. The suture tabs must be used to secure
the telemeter in the correct position within the abdominal cavity. Care should be taken not to
pierce or cut the suture tabs during either implantation or explantation of the telemeter. Should
you damage either of the suture tabs then a pouch made of surgical mesh can be used to
secure the telemeter in place (see page 30 for instructions).
Biopotential lead separation
Depending on the model, telemeters may be supplied with bipolar electrodes that are suited
to ECG data collection. The bipolar electrodes are supplied connected down their entire
length. While separation of the ends is required to allow appropriate placement, we
recommend that the electrode leads be kept together and run alongside one another for as far
as practical. This will reduce electrical noise in recorded biopotential signals. A free length of
no more than 3-4 cm is recommended for each electrode. It is not recommended that sutures
be placed around the leads other than at the point of contact with the tissue as this may provide
a stress point on the leads and cause them to break.
Millar Telemetry User Manual for Small Animal Systems
v2.1
31
Example
placement of
electrode leads
for recording of
high quality ECG
Sutures
Leads kept
together as
far as
practical
32
Attach to
xiphoid
process
Millar Telemetry User Manual for Small Animal Systems
v2.1
Construction of Mesh Pouches
The suture tabs should be used to secure the telemeter within the abdomen and to the
abdominal wall. However, should either suture tab be damaged, a mesh pouch can be fitted
to the telemeter. Sutures through the mesh pouch can then be used during surgery to attach
the telemeter to the abdominal muscle layer. It is important that the telemeter is not secured
in place by suturing around either the catheter or electrode wires as this may lead to wire
breakages.
The pouch should be made from surgical mesh or similar biocompatible material. Silk or similar
suture material should be used to hold the mesh in place rather than sewing cotton. Suitable
mesh is supplied as part of the BP Consumable Pack from Millar. The piece of mesh supplied
can be used to make 4 small animal pouches. The excess mesh is used for securing the blood
pressure catheter to the vessel.
17.5 cm
8 cm
Use this piece for
BP catheter
implantation
Construction
1) Cut a rectangle of mesh:
4 x 8 cm
2) Fold the mesh over the telemeter body and secure using sutures at a few points around
the telemeter as pictured below.
Care must be taken not to pierce the silicone telemeter body with the needle or to cut or
damage the catheter or electrodes.
Sterilization*/Surgery
Mesh not fitted to telemeter (construct pouch at the time of surgery)
 Autoclave mesh with surgical pack
 Ethylene Oxide (gas sterilization)
 Alcohol/Chemical – mesh must be rinsed thoroughly in sterile saline before construction
and implantation
Mesh fitted to telemeter before sterilization
 DO NOT AUTOCLAVE
 Ethylene Oxide (gas sterilization)
 Chemical Sterilization (NOT ethanol) – Telemeter and mesh must be rinsed thoroughly
with sterile saline
* See Sterilization section for more details of recommended sterilization procedures.
Removal
At the end of the experiment, great care must be taken when removing the telemeter from the
animal. Tissue will grow into the mesh so it may be difficult to see the telemeter body. Never
use a scalpel to cut away the mesh. Scissors should be carefully used taking great care not
to cut the telemeter body, catheter or electrodes. The mesh pouch should be discarded and a
new one fitted for the next use.
Millar Telemetry User Manual for Small Animal Systems
v2.1
33
Surgical Mesh for BP Catheter Implantation
Tissue adhesive can detach from tissue after a period of time allowing the BP catheter to pull
out of the vessel. To avoid this, we recommend the use of a piece of surgical mesh at the time
of implantation. Scar tissue will grow into the mesh fabric holding the catheter in place and
reducing reliance on the glue itself.
1. Insert catheter into the vessel and use the smallest amount of glue to hold the catheter in
place and to stop any blood leakage (see surgical videos online at
http://millar.com/knowledge-center
2. Dry the surrounding area using swab sticks
3. Cut a small square of surgical mesh (5 x 5 mm).
4. Place the mesh over the aorta and catheter so that the slit sits around the catheter.
5. Glue the mesh to the catheter and use small amounts of glue to attach the mesh to the
surrounding tissue. Take care not to put too much glue on the vessel or blood flow may
be restricted when it dries.
34
Millar Telemetry User Manual for Small Animal Systems
v2.1
Telemeter Care and Handling
Warnings
The telemeter pressure catheter (if fitted) is shipped with foam over the catheter tip to protect
the sensor area. The catheter tip should be kept in the foam and should always be stored in
the plastic tray and box provided between uses.
Precautions: Telemeters with Pressure Sensors
Proper handling of the telemeter and pressure catheter is very important to avoid damage and
extend the telemeter life. Take care not to damage the telemeter body, pressure catheter
or electrode as they are non-repairable. It is highly recommended that users view the online
videos on handling the telemeters during implantation, explantation and cleaning.
http://millar.com/knowledge-center
**If handling the pressure catheter using forceps, the forcep tips should always be padded
using soft tubing (e.g. silicon). This will protect the wires inside the catheter from being crushed.
Use and Handling
• Inspect the catheter (using a microscope) for damage (cracking, kinks, etc.) before each
use.
• Clean the telemeter and sensor and electrodes immediately after each use (see Cleaning
Instructions).
• Do not touch the pressure sensor area with sharp objects. Do not make sharp bends in
the catheter.
• Do not apply direct pressure to the pressure sensor area with instruments such as forceps
or tweezers.
• When handling the catheter with either fingertips or surgical instruments, always handle the
catheter 5-10 mm proximal to the sensor area. The sensor area contains very fine wires
which may be damaged or broken if the catheter is gripped too close to the sensor or too
tightly, be GENTLE.
• Always know the location of the catheter tip.
• Do not place heavy objects or metal instruments on top of the catheter or telemeter.
• Take care not to cut the catheter during surgery or when removing the telemeter from the
animal.
Pressure
Sensor
DO:
Clean immediately after use
Protect sensor tip when not in
use
Catheter
Clean immediately after use
Body
After removal Keep catheter and sensor wet
from animal
until cleaned
Disinfection
or
Sterilization
Clean thoroughly with
approved enzymatic cleanser
immediately
Dry catheter before sterilizing
with ethylene oxide
Use a recommended cleaning
agent
DO NOT:
Do NOT clean with stiff-bristled brush
Do NOT clean with high pressure water jet
Do NOT tap the sensor against a hard surface
Do NOT apply excessive force to the sensor tip
Do NOT expose to excessive pressure
Do NOT cut, crease, knot, fold, kink, or crush
with forceps or clamps of any kind
Do NOT expose to alcohol, cresols, phenols,
mercury compounds, hypochlorites, acetone,
peroxide, silicone chlorine, xylenes,
trichloroethylene, or freon
Do NOT use ultrasonic cleaner
Do NOT autoclave, irradiate (gamma/ebeam),
plasma, peroxide or formaldehyde vapor
solutions
Do NOT use Sporox or Cidex PA solutions
Millar Telemetry User Manual for Small Animal Systems
v2.1
35
Telemeter Removal and Preparation for Reuse
If the user takes care, the telemeters are designed to be reused. After implantation for a long
period of time it is common to find the catheter, electrodes and telemeter body covered in
connective tissue. When removing the telemeter take care to avoid damaging catheter,
electrodes or telemeter body (including suture tabs). Never use a scalpel to cut tissue from
around or above the telemeter as it could damage the telemeter. When detaching the telemeter
body from the muscle, take great care to only cut the sutures holding it in place and not the
suture tabs themselves. It may be necessary to first clear attached tissue from the area so that
the suture thread can be visualized. Extreme care should also be taken not to damage the
pressure catheter and sensor tip during telemeter removal.
The telemeter should be soaked in cleaning detergent (see below). After soaking and using a
microscope the sensing tip should be gently wiped with a soft cotton gauze pad or swab sticks
to remove any remaining film or deposits. Care must be taken not to damage the sensing tip,
do not pick up the catheter by the sensor tip or have any sharp objects near this tip.
Cleaning
1. After removing the telemeter from the animal, immediately soak it in a beaker or dish filled
with fresh saline or distilled water. Keep soaking until you are ready to continue cleaning;
DO NOT allow it to dry.
2. Soak in a recommended enzymatic cleaning solution (listed below). This is essential to
prevent protein build-up on the pressure sensor. Without the use of an enzymatic cleaner,
a protein film will form that can result in pressure signal drift.
Type
Trade Name
Manufacturer
Active Ingredient
Soak Time/Temp
Enzymatic Enzol (in UK: Advanced
Propylene Glycol
15 mins / room

Detergent Cidezyme )
Sterilization
temperature
Products (J&J)
Ruhoff
Propylene Glycol
15 mins / room
Endozine
Corporation
temperature

Alconox
Sodium
15 mins / room
Terg-A-Zyme
Dodecylbenzene
temperature
3. After soaking (recommended times), gently wipe the sensing tip under a microscope with a
soft cotton gauze pad or swab sticks to remove any remaining film or deposits. Care must
be taken not to damage the sensing tip.
4. After soaking and wiping, thoroughly rinse the telemeter with fresh distilled water or saline.
5. After rinsing is complete, gently dry the catheter as follows:
 Fold a soft tissue or kim wipe
 Use gentle stroking to dry the catheter tip.
 Do not pull the catheter tip through a folded tissue.
 Do not allow the catheter to air dry on a tray, table, or countertop.
6. After the cleaning and drying procedure is finished:
 Protect the sensor area on the catheter tip by carefully placing it in the foam that came
with the telemeter.
 Return the telemeter to its original plastic the tray.
 Slide the tray inside the box and now store the telemeter in a cool, dry place until the
next use.
CAUTION: Delays in rinsing a telemeter after removal will reduce cleaning effectiveness!
Examine the pressure sensor active surface (diaphragm) for blood or materials not removed
by cleaning. A dirty sensor may cause baseline drift when used the next time.
Cleaning Agents
36
Millar Telemetry User Manual for Small Animal Systems
v2.1
An enzymatic detergent can help in removing biological tissue from the catheter and telemeter
body (e.g. Terg-A-Zyme®). These are generally available from most hospital supply
companies and are generally labeled for use on fabrics or surgical equipment/instruments.
The purpose of the detergent is to remove blood, serum proteins, and tissue debris from the
surface of the telemeter. It is important that any product that has a surfactant CANNOT be
used with any pressure telemeters.
We recommend Terg-A-Zyme is an enzyme-active powdered detergent made by Alconox, Inc.
To make a 1% solution, mix 10 grams of powder with 1 Liter of cold or warm water. Allow the
telemeter to soak for a minimum of 15 minutes and a maximum of 24 hours in the solution.
Rinse thoroughly, preferably with running water. Fisher Scientific is a company that supplies
Terg-A-Zyme (catalog #50-821-785, www.fishersci.com) but please refer to the Alconox
website for other companies (www.alconox.com). Terg-A-Zyme is not a sterilant. Telemeters
must be sterilized before re-implantation. Please see following section for more information.
Note: If checking the offset/calibration of the telemeter, the conditions listed under
Calibrations Values must be replicated, i.e. clean, hydrated, turned on for 4 hours, at
38°C and in the dark.
Sterilization
Preventing infection is very important in collecting quality recordings of normal physiology and
animal survival. It is important to make sure the implanted telemeter is sterile before
implantation.
Under no circumstances should the telemeter be autoclaved or subjected to
temperatures over 60°C as this will damage the telemeter.
Do not use alcohol as a sterilizing agent. This will damage the telemeter.
Do not sterilize by radiation (gamma or e-beam), plasma, peroxide or formaldehyde
vapor solutions.
Tissue Oxygen telemeters: The electrodes in particular the carbon paste electrode
should not be sterilized. Please see http://millar.com/knowledge-center for access to
online recommendations.
Disinfection/Sterilization
1. The telemeter must be cleaned, rinsed and dried before disinfection or sterilization. Soil,
debris, proteins, and water can interfere with the effectiveness of the following procedure.
Note that some disinfectants have a limited usable life after activation or opening the
container, failure to take notice of the warnings can prevent the effectiveness of the
disinfection process.
2. Prepare the disinfectant according to the manufacturer’s instructions.
3. Soak the telemeter in the disinfectant at the temperature and time intervals listed.
4. Rinse the telemeter well in sterile pyrogen-free water (or sterile saline) before implantation.
A minimum of three separate rinses is recommended. Do not reuse any of the water used
for rinsing since it will be contaminated with the disinfectant.
CAUTION: Use only the listed recommended cleaners and disinfectants for the
times/temperatures indicated.
Millar Telemetry User Manual for Small Animal Systems
v2.1
37
Chemical Sterilants
Available from most hospital supply companies, chemical sterilants are considered cold
sterilants and should be used for the sterilization of heat sensitive medical equipment such as
Millar telemeters. When used properly, chemical sterilants will destroy all viable forms of
microbial life.
Glutaraldehyde may be used to sterilize the telemeters. Glutaraldehyde (Fisher Scientific)
must be diluted to 2% before use. Check your local chemical supply company for availability.
Do not use gluteraldehyde solutions containing surfactants (i.e. Cidex 7, Cidex Plus 28 Day &
Metrocide 28).
Rinse with Sterile Saline
The telemeter should always be washed in sterile saline before implantation to remove all
traces of the sterilant. Use it to temporarily store (< 48 hours) the telemeter aseptically until
surgical implantation.
Ethylene Oxide Gas Sterilization
These sterilizers operate at a low temperature of 55°C and use chemicals as the sterilant.
Items are packed and dry at the end of the sterilizing cycle. Telemeters are suitable for
Ethylene Oxide sterilization as long as they are not subjected to temperatures above 60°C.
Customers are liable for product replacement if recommended products and instructions are
not used or followed. If there are any questions about a chemical or procedure please contact
Millar (support@millar.com).
Warning: Some examples of chemicals that will cause damage to telemeters include,
but are not limited to: alcohols, phenols, iodophors, and hypochlorite. Please confirm
with Millar before using any product other than the approved products list below.
Type
Trade Name
Manufacturer
Active Ingredient
High-level
disinfectant
Cidex Activated
Dialdehyde
Solution
Cidex OPA
Advanced Sterilization
Products (J&J)
Gluteraldehyde
Advanced Sterilization
Products (J&J)
Metrex
Orthopthalaldehyde
Soak Time /
Temperature
1-2 hours /
25°C (77°F)
16-30 mins /
20°C (68°F)
Gluteraldehyde
1-2 hours /
MetriCide
25°C (77°F)
** Products with similar names are available but may not be suitable. Please use only
products specified or contact Millar for clarification (support@millar.com).
38
Millar Telemetry User Manual for Small Animal Systems
v2.1
Telemeter Diagnostics
Using the Configurator System the user can investigate the status of the telemeter. To run diagnostics
enter the serial and channel numbers of the telemeter and click “Start diagnostics” on the Telemeter
Configuration & Diagnostics dialog. In diagnostic mode the telemeter will continue to transmit data.
Pressure telemeter
Oxygen telemeter
After clicking on the Start Diagnostics you will be asked if you wish to save a log file. This can
be useful to monitor the telemeter over a period of time or if you are asked to send information
to support staff at Millar. The log file and display are both updated every 10 s. A count of the
updates is shown in the bottom left hand corner of the window. If you choose to save a log file
you will not be able to stop the diagnostics for at least 5 minutes. ConfigSoft will ask you to
confirm the name and location for the log file. The default filename includes the telemeter serial
number
and
the
date
and
time
the
diagnostics
was
started
e.g.
Tx5574_log_2013_5_9_16_11.txt
PLEASE NOTE: Depending on the telemeter you have the  may appear if the telemeter does
not have that function i.e. oxygen telemeters will have a  for the pressure sensor.
Battery Level
 = the battery is more than 20% charged.
 = the battery is low and the telemeter needs to be recharged. If the telemeter is not placed
on an active SmartPad it will automatically enter Safe Mode.
Note: The battery level reported during diagnostics may be artificially high while the
telemeter is charging. For an accurate battery level, remove the telemeter from the
SmartPad for 1 minute before noting the battery level.
Charging
 = the telemeter detects a charging field from a SmartPad and charging is occurring.
 = the telemeter cannot detect a charging field from a SmartPad and is not charging. A
telemeter will continue to run off its battery for approximately 4 hours (if fully charged).
Troubleshooting Charging: there are a number of possible situations for a  next to Charging:
1) The Charging field of the SmartPad has been disabled using Configurator System.
2) The telemeter is not within 5 cm of the charging surface of a SmartPad.
3) The SmartPad has temporarily turned the scheduled charging field off for a temperature
update.
4) The telemeter is in an area of the SmartPad with low field strength or it may not be
parallel to the SmartPad surface therefore in the wrong orientation to receive charging
from the field efficiently.
Millar Telemetry User Manual for Small Animal Systems
v2.1
39
Pressure sensor function
This only applies to telemeters with a pressure sensors e.g. TRM54P, TRM54PB, etc
 = the pressure sensor is functioning normally.
 = a problem has been detected with the pressure sensor (contact your distributor or
support@millar.com)
Oxygen Telemeter Potentiostat
 = the oxygen sensor is functioning normally.
 = a problem has been detected with the oxygen sensor.
Please contact Millar for further assistance (support@millar.com).
SmartPad Diagnostics
Using the Configurator System the user can investigate the status of their SmartPad. To run
diagnostics enter the serial and channel numbers of your SmartPad and press the “Start
diagnostics” button on the SmartPad Configuration & Diagnostics dialog. Please note that you
cannot record data from your telemeter using a SmartPad that is in Diagnostics mode instead
a saw tooth pattern is output from the SmartPad analog outputs.
Press OK to continue. You will be asked if you wish to save a log file. This can be useful to
monitor the SmartPad over a period of time or if you are asked to send information to support
staff at Millar. The log file and display are both updated every 10 s. A count of the updates is
shown in the bottom left hand corner of the window. If you choose to save a log file you will not
be able to stop the diagnostics for at least 5 minutes. ConfigSoft will ask you to confirm the
name and location for the log file. The default filename includes the SmartPad serial number
and the date and time the diagnostics was started e.g. Sp6096_log_2013_5_9_16_48.txt
40
Millar Telemetry User Manual for Small Animal Systems
v2.1
The SmartPad will now be in Diagnostics mode and the Status light on the front will flash
orange.
Atmospheric Pressure Sensor function
 = the atmospheric pressure sensor is functioning normally.
 = a problem has been detected with the atmospheric pressure sensor. This can be caused
by a brief loss of communication with the SmartPad. If it occurs frequently or continues
for more than 30s please contact Millar (support@millar.com).
Temperature normal
 = the temperature of the SmartPad is at acceptable levels.
 = the temperature is above acceptable limits. Please contact Millar (support@millar.com).
Power consumption normal
 = SmartPad power consumption is normal.
 = Excess power consumption has been detected. The Status light on the front of SmartPad
will be red. This cause may be that the SmartPad is too close to a metal surface or another
SmartPad. The SmartPad will continue receiving and outputting data but the charging
field will have been turned off. To correct the situation, unplug the SmartPad and try
moving the SmartPad away from any metal surfaces and other SmartPads re-plug in the
power cable to reset the SmartPad. If the problem continues please contact Millar
(support@millar.com).
Millar Telemetry User Manual for Small Animal Systems
v2.1
41
SmartPad Technical Specifications
Weight
Contains ambient pressure monitor,
outputs standard analog voltage compatible with
any data acquisition system (BNC), recharges
the telemeter battery.
2.5 kg
Dimensions
400(w) x 450(d) x 60(h) mm
Cage platform dimensions
295(w) x 425(d) mm
Power input
100-240 V, 50-60 Hz
Max Power draw
40W
Temperature Operating Range
10 to 40 degrees Celsius
Output Connectors
BNC x 3
Output Voltage Range
0 to 4V
Low Pass filtering on signal outputs
Cut off frequency = 1000 Hz
SmartPad
42
Millar Telemetry User Manual for Small Animal Systems
v2.1
Telemeter Technical Specifications
The technical specifications in the table below are for the TRM5* series telemeters. Some specifications
may not be relevant depending on the model of the telemeter.
Pressure sensor accuracy
+2 mmHg
Pressure sensor drift
<4 mmHg per month
Pressure sensor Frequency response
Standard pressure catheter
dimensions
Pressure sensor range
DC to 500 Hz (-3 dB)
Distal tip; 660 µm OD (2Fr), Catheter; 500 µm (1.5Fr), length
9 cm
-20-300 mmHg
Ambient pressure range
650 to 800 mmHg
Maximum operating altitude
1300 m
Biopotential Input range
+2 mV
Biopotential Resolution
12 bit A/D
Biopotential High pass characteristics
AC coupled, single pole, -3dB point at 0.7 Hz
Biopotential Electrode leads
Coiled stainless steel
Biopotential Electrode diameter
1 mm
SNA Input range
+60 µV
SNA High pass filter
-3dB point at 1.5 Hz
SNA Input impedance
500 kOhms at 1000 Hz
SNA Electrode leads
Multi-stranded stainless steel (length 25cm)
SNA Electrode diameter
0.15 mm (wire only)
Temperature operating range
8 to 41 degrees Celsius
Sampling frequency
2000 Hz
Low Pass filtering by SmartPad
Transmitted signal
Cut off frequency = 1000 Hz
Up to 5 m (range may vary depending on laboratory
configuration)
Fully digital at 2.4 GHz
Channels
30 transmission frequencies are available, user set
Minimum animal weight
200 gm
Outer material
Silicone
Weight
Approximately 12.5 gms
Dimensions
31.5 x 24 x 11 mm
Transmission range
Battery life
Recharge time
On-off mechanism
Calibration
Analog output calibration values
Continuous on SmartPad. If the animal is away from the
SmartPad charging field (or the SmartPad charging field is
turned off) battery life is ~4-6 hrs.
Minimum 3 hours
Activation via the SmartPad and deactivation via Configurator
System.
No user intervention required (calibration values stored within
each telemeter).
Pressure: 1V output = 0mmHg, 2V output = 100 mmHg
Biopotential: 2V output = 0mV input, 4V output = 2 mV input
SNA: 2V output = 0µV input, 4V output = 60 µV input
Temperature: 0V output = 0°C, 1V output = 20°C
Millar Telemetry User Manual for Small Animal Systems
v2.1
43
Oxygen Telemeter Technical Specifications
The technical specifications in the table below are for the TR57Y Tissue Oxygen telemeter.
Potentiostat type
Voltage controlled current source
Set potential
-0.65V (-650mV)
Potentiostat current measurement range
0 to -600nA
Potentiostat resolution
12 bit A/D
Potentiostat electrode leads
Coiled stainless steel (~28 cm)
Potentiostat electrode lead diameter
1 mm
Temperature operating range
20 to 41 degrees Celsius
Sampling frequency
2000 Hz
Low Pass filtering in telemeter
Cut off frequency = 1000 Hz
Transmission range
Up to 5 m (range may vary depending on laboratory
configuration)
Transmitted signal
Fully digital at 2.4 GHz
Channels
30 transmission frequencies are available, user set
Minimum animal weight
175 gm
Outer material
Silicone
Weight
Approximately 12.5 gms
Dimensions
31.5 x 24 x 11 mm
Battery life
Continuous on SmartPad. If the animal is away from the
SmartPad charging field (or the SmartPad charging field
is turned off) battery life is ~4-6 hrs.
Recharge time
Minimum 3 hours
On-off mechanism
Calibration
Analog output calibration values
44
Activation via the SmartPad and deactivation via
Configurator System.
User required to compensate for the individual Carbon
Paste Electrode sensitivity coefficient supplied with each
Blue Box electrode. Sensitivity range is 1-2nA/µMol.
Oxygen: 1V output = 0nA, 2V output = -200nA
Temperature: 0V output = 0°C, 1V output = 20°C
Millar Telemetry User Manual for Small Animal Systems
v2.1
Frequently Asked Questions........
Battery and recharging
Would the animal be harmed by staying in the SmartPad charging field?
There are no known biological effects of the inductive field. Studies have shown that a field
strength much higher than the Millar SmartPad have no effect. The level of magnetic field is
quite low and remains useable only ~5 cm above the SmartPad.
Telemeter
How many animals can be monitored at one time?
Each telemeter sends its signal on one of 30 unique transmission channels. Therefore up to
30 telemeters can be used within a 5 m range of one another. If you need to have more
animals in close proximity, contact Millar to discuss a customized solution.
Is there any interference between telemeters?
No. As long as each telemeter is set to use a different channel they use a separate
transmission frequency. No shielding is required between cages. Close proximity is possible.
The transmission frequency is in the 2.4 GHz band.
What is the life expectancy of the telemeters?
The telemeters are designed for the implantation in the abdomen. As this is a relatively hostile
environment users should not expect the same life time as for bench top based laboratory
items. However with appropriate care during implantation and explantation users should
expect to reuse the telemeters.
SmartPad questions
Why does my SmartPad keep showing a red light and my telemeter goes flat?
A red light on your SmartPad suggests that the SmartPad has automatically disabled the
charging field due to detection of a high current. In this mode data output will continue but
battery charging is not occurring. Cause: may be due to the SmartPad being placed too close
to a metal surface or to another SmartPad. Move the SmartPad away from any metal and then
reset by unplugging the power and plugging it back in (see page 13).
I need to record from my animal while it is near metal. How can I do this?
Use the Configurator System to disable the charging field of the SmartPad. Remember that
the telemeter will run of battery power and therefore data can be collected for approximately 4
hours if fully charged.
Further questions?
Millar Instruments is proud to offer unlimited technical support and advice to all its customers.
Our team of engineers and physiologists can advise on specific applications and equipment
configurations. Please contact us
Millar Telemetry User Manual for Small Animal Systems
v2.1
45
Contact Information
US Head Office
Houston, Texas
Toll free (in USA) 1 800 669 2343
TEL: +1 832 667 7000
FAX: +1 713 714 8497
EMAIL
WEB
46
New Zealand Office
Auckland, New Zealand
TEL: +64 9 367 7126
FAX: +64 9 367 7157
support@millar.com or sales@millar.com
www.millar.com
Millar Telemetry User Manual for Small Animal Systems
v2.1

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 50
Language                        : en-US
Tagged PDF                      : Yes
Title                           : Telemetry Unit
Author                          : mdon056
Creator                         : Microsoft® Word 2013
Create Date                     : 2014:02:27 16:08:13+13:00
Modify Date                     : 2014:02:27 16:08:13+13:00
Producer                        : Microsoft® Word 2013
EXIF Metadata provided by EXIF.tools
FCC ID Filing: V58HU72

Navigation menu