Mitutoyo Microscope And Magnifier Tm 500 Users Manual Toolmakers Microscopes

2015-02-09

: Mitutoyo Mitutoyo-Mitutoyo-Microscope-And-Magnifier-Tm-500-Users-Manual-557234 mitutoyo-mitutoyo-microscope-and-magnifier-tm-500-users-manual-557234 mitutoyo pdf

Open the PDF directly: View PDF PDF.
Page Count: 42

MANUAL No.4769AԘ
SERIES No.176
TM-500 Series
TOOLMAKER’S
MICROSCOPE
User
s Manual
Read this User's Manual thoroughly
before operating the instrument. After reading,
retain it close at hand for future reference.
PREFACE
Thank you for purchasing the Mitutoyo TM-500 series Measuring Microscope.
This User's Manual explains the hardware operations of the Mitutoyo TM-500 series and the
precautions to be observed during operation. To obtain the best possible performance and
longest service life from your TM-500 series Measuring Micro-scope, please read this user's
manual thoroughly. Be sure to read PRECAUTIONS before using this Measuring
Microscope.
After reading this user's manual retain it for future reference.
CONVENTIONS USED IN THIS MANUAL
The following visual cues are used throughout this manual to identify different types of information:
DANGER
Indicates that the operator should exercise care to avoid danger of fire, explosion, o
r
personal injury.
CAUTION
Indicates that operating the instrument in this manner may damage it or may cause it
to malfunction.
Draws attention to important information, recommended operation techniques, o
r
helpful tips.
Gives reference location for further information on topic.
Highlights installation, measurement, or other procedures.
㩷㩷㩷
PREFACE /CONVENTIONS USED IN THIS MANUAL i
PRECAUTIONS
ii
PRECAUTIONS
(1) Relocation
The Measuring Microscope is a precision instrument. Handle it with care when
transporting/relocating. Do not touch the movable members, which must be secured, during
transportation.
Place a cushioning pad between the Micrometer Head and the XY stage, as shown below, to
prevent the Micrometer Head spindle from hitting the table.
Cushioning pad
(2) Installation
Install the TM-500 in a site with:
as little dirt, dust, and humidity as possible. Put the supplied dust cover on it if it is not
going to be used for an extended period of time.
no vibrations.
no sudden temperature changes, which may result if the site is subjected to direct sunlight.
(3) Explosive Hazard
Do not operate the Measuring Microscope in the presence of explosive gas.
(4) Power Supply
Depending on the installation site, voltage fluctuation may cause a change in light intensity
or flickering of the illumination bulb. Although this does not affect measurement, use a
voltage stabilizer if a stable light intensity is required.
Use an AC power supply which is isolated from high-voltage, large-current machinery.
High-voltage, large-current machinery often emits a substantial amount of electric noise due
to surge currents, which can adversely affect this measuring system if it draws power from
an AC outlet often provided on such equipment. Therefore, the Measuring Microscope
should draw power from a separate outlet. The power cable of the Measuring Microscope
should run as far apart as possible from such noise sources.
Only use the power cord that is supplied with the instrument. For a replacement power cord,
contact Mitutoyo.
Only use the supplied replacement fuse or a fuse with identical specifications (type, nominal
voltage and current).
(5) Ground Positively
Ground the Measuring Microscope positively, especially if it must be installed near
equipment emitting substantial amounts of electric noise. To do this connect the grounding
wire from the Measuring Microscope to the body of a piece of equipment, which also must
be grounded. The grounding wire should be as short as possible and should not be wound
into a loop or coil.
(6) Do Not Disassemble
㩷䃂There are high-voltages inside the Measuring Microscope. Do not attempt to disassemble it
unless otherwise specified.
WARRANTY
In the event that the Mitutoyo Measuring Microscope TM-500 series should prove
defective in workmanship or material, within one year from date of purchase for use, it
will be repaired or replaced, at our option, free of charge upon its prepaid return to us.
PRECAUTIONS/WARRANTY iii
CONTENTS
PREFACE...........................................................................................................................
γ
CONVENTIONS USED IN THIS MANUAL...................................................................
γ
PRECAUTIONS.................................................................................................................
δ
WARRANTY .....................................................................................................................
ε
1. Outline ..............................................................................................................................7
1.1 Description.............................................................................................................. 7
1.
2Name of Each Part .................................................................................................8
2. Installation and Setup ..................................................................................................... 11
2.1 Package Contents..................................................................................................11
2.2 Installation Site .....................................................................................................11
2.3 Setting-up.............................................................................................................. 12
2.4 Checking and Adjustment.....................................................................................13
2.4.1 Checking the reticle position against the XY stage movement direction .. 13
2.4.2 Checking the centering of the reticle .........................................................13
2.4.3 Adjusting the reticle...................................................................................14
3. Measurement ..................................................................................................................17
3.1 Preparations for Measurement.............................................................................. 17
3.1.1 Precautions for measurement ..................................................................... 17
3.1.2 Replacing the lenses................................................................................... 18
3.1.3 Fixing the workpiece.................................................................................. 18
3.1.4 Illumination modes..................................................................................... 19
3.1.5 Replacing reticles.......................................................................................20
3.1.6 Adjusting the diopter..................................................................................20
3.1.7 Bringing the measuring surface into focus.................................................21
3.1.8 Positioning the workpiece .......................................................................... 21
CONTENTS
iv
CONTENTS v
3.2 Measurement.........................................................................................................22
3.2.1 Dimensional measurement ......................................................................... 22
3.2.2 Angle measurement.................................................................................... 23
3.2.3 Template matching inspection ................................................................... 24
3.2.4 Step measurement ......................................................................................24
4. Maintenance ...................................................................................................................25
4.1 Cleaning and Lubrication......................................................................................25
4.2 Inspection..............................................................................................................25
4.3 Replacing Consumable Parts ................................................................................27
4.4 Consumable Parts..................................................................................................29
5. Troubleshooting..............................................................................................................31
6. Specifications .................................................................................................................35
7. Optional Accessories...................................................................................................... 37
8.
Reference Diagrams...................................................................................................... 39
SERVICE NETWORK.......................................................................................................41
CONTENTS
vi
Outline
This chapter outlines the Mitutoyo Toolmaker’s Microscope TM-500 series and
gives the name of each part.
1.1 Description
The Mitutoyo Toolmaker’s Microscope TM-500 series (TM- 505 and TM-510) are easy-to-use,
compact-size Toolmaker’s Microscopes that feature a vertical supporting column. Designed with
measurement of workpiece contours and inspection of surface features in mind, the TM-500 series
supports a wide range of applications from shop-floor inspection, measurement of tools and
machined parts, to precision measurement of test tools in a measuring room.
zA long vertical working distance (67mm) and erecting image have improved operability.
zThe XY stage is mounted on a compact body to support wide travel ranges (50mm50mm
for TM-505, 100mm50mm for TM-510) for effective measurement.
zFine/coarse focusing can be performed with a single focusing knob.
zThe angle dial built into the eyepiece portion of the optical tube allows easy angle
measurement.
1.
Outline 7
1.Outline
8
1.2 Name of Each Part
8
9
10
11
12
13
14
1
2
3
4
5
6
7
1. Eyepiece
2. Diopter adjustment ring
3. Eyepiece mount
4. Optical tube
5. Surface illuminator
6. Objective
7. XY stage
8. Angle dial
9. Angle dial clamp screw
10.Vernier
11.Vernier clamp screw
12.Focusing knob
13.Control panel
14.Power panel
1. Outline 9
13. Control panel
Contour illumination switch
Surface illumination switch
Light control knob
Power switch
1.Outline
10
14. Power panel (for standard model)
Surface illuminator
cable connector
GND terminal
The voltage setting
method display
AC inlet
Fuse holder
Voltage selector
Power panel (for CE correspondences)
Fuse holder
Surface illuminator
cable connector
AC inlet
Voltage selector GND terminal
Installation and Setup
This chapter lists the contents of the TM-500 series package and describes the
setup and adjustment procedures.
2.1 Package Contents
Standard accessories
Name TM-505 TM-505R TM-510 TM-510R
1. Objective2㧕 176-138 1111
2. Eyepiece15㧕 176-116 1111
3. Lens cap 511187 1 1 1 1
4. Cross-hair line reticle 176-126 1 1 1 1
5. Surface illuminator 1 1 1 1
6. Micrometer Head 22
7. Mounting screw 380597 㧙㧙2* 2*
8. Reticle setting screw 200624 1 1 1 1
9. Bulb setter 511188 1 1 1 1
10. Allen key nominal 3 538616 1 1 1 1
11. Vinyl cover 512555 1 1 1 1
12. Spare bulb 24V 383038 1111
Spare fuse1Anot CE 350279 1111
13. Spare fuse1Afor CE 384204 1111
14. Power cord 1 1 1 1
15. Grounding wire 1 1 1 1
16. User’s Manual 1 1 1 1
17. Warranty 1 1 1 1
*Keep the mounting screws in the TM-510/510R package. They are provided to install the optional
mounting jigs on the XY stage.
2.2 Installation Site
Install the Measuring Microscope in a place which is free from vibration and dust. Exposing the
instrument to vibrations over long periods of time can deteriorate its measuring accuracy. Dust
adversely affects the optical parts, the XY stage, and moving parts.
2. Installation and setup 11
2.3 Setting-up
1.Install the Micrometer Head on the XY stage.
Loosen the hex-socket head screw on the bracket. Fully insert the stem of the Micrometer Head . Then,
tighten the hex-socket head screw with the stem in place.
The clamp screw is provided to secure the spindle of the Micrometer Head. If using a Micrometer Head
with a fitting hole in the stem, insert the stem so the hole is aligned with the clamp screw. If this puts the
scale on the Micrometer head in a poor position for viewing the zero graduation, adjust the scale position
by turning the Micrometer Head sleeve. If using a Micrometer Head with no hole to clamp the spindle,
lightly tighten the clamp screw.
2. Connect the surface illuminator cable to the connector on the power panel, which is at the back of the
Measuring Microscope. Insert the cable firmly into the connector and secure it with the nut.
3. Set the voltage selector plug to the supply voltage as follows.
XY stage
Clamp screw
Bracket
Hex-socket head screw
When the voltage selector on the back panel is as in the
upper left figure, the combination of two slide switches on
the upper and lower sides performs a voltage setup. Set up
the voltage according to the voltage setting method display.
When the voltage selector on the back panel is as in the
lower left figure, remove the fuse holder by turning it
counterclockwise using a screw driver.
Turn the voltage selector with a coin, etc., and align the
rated voltage display with the position of the arrow. Replace
the fuse holder as before.
Voltage setting
method display.
Voltage selector
Fuse holderVoltage selector
2. Installation and setup
12
2.4 Checking and Adjustment
2.4.1 Checking the reticle position against the XY stage
movement direction
1Place a small workpiece on the stage glass and bring it into focus.
2Turn the Micrometer Heads to align an edge of the workpiece with the center of the cross-hair.
3While turning the Micrometer Head to move the workpiece left and right, turn the angle dial so
that the horizontal cross-hair is oriented to coincide with the direction of the stage movement.
㪋㪅 Loosen the vernier clamp screw. Align the “0” graduation on the angle dial with that of the
vernier scale. Ensure the margin is sufficient for adjusting the vernier scale position. If space for
adjusting the vernier scale position is limited, re-adjust the vernier scale position by referring to
"2.4.3 (1) Adjusting the reticle to the XY stage movement direction".
2.4.2 Checking the centering of the reticle
Small workpiece
To perform dimensional measurement by turning the angle dial or after replacing the reticle, align the
cross-hair with the center of rotation of the angle dial, as follows.
1Place a small workpiece on the stage glass and bring it into focus.
2Turn the Micrometer Heads to align an edge of the workpiece with the center of the cross-hair.
3Turn the angle dial 180q. Make sure the edge of the workpiece remains within 3μm of the center of the
cross-hair. If it is not within 3μm, adjust the center of the reticle by referring to "2.4.3 (2) Centering
the reticle".
2. Installation and setup 13
2.4.3 Adjusting the reticle
(1) Adjusting the reticle with the XY stage moving direction
Angle dial cover
㩷㩷㩷㩷㩷
Screws (4 pcs.) Eyepiece mount Angle dial
Fixing screws (4 pcs.) Adjustment screws (4 pcs.)
Angle dial clamp
knob
Vernier scale
Vernier scale
clamp knob
㪈䋮 Remove the clamp knobs from the angle dial and vernier scale.
㪉䋮 Remove the four screws from the angle dial cover and remove the cover.
3Screw-in the clamp knobs on the angle dial and vernier scale.
4Approximately center the vernier scale in the adjustable range. Then secure it with the clamp
knob.
5Align the “0” graduation of the angle dial with that on the vernier scale. Then, secure the angle
dial with the clamp knob.
6Loosen the adjustment screws and fixing screws (4 pcs. each) so that the eyepiece mount can be
moved manually.
㪎䋮 While looking into the eyepiece, adjust the position of the eyepiece mount so that the horizontal
cross-hair is oriented to coincide with the direction of the stage movement.
8Temporarily secure the eyepiece mount by lightly tightening the fixing screws.
㪐䋮 Centering the reticle as described in "(2) Centering the reticle" below. Firmly secure the
eyepiece mount by fully tightening the fixing screws.
㪈㪇䋮Remove the clamp knobs from the angle dial and vernier scale.
㪈㪈䋮Replace the angle dial cover and secure it to the optical tube with the four screws.
㪈㪉䋮Screw-in the clamp knobs on the angle dial and vernier scale.
2. Installation and setup
14
(2) Centering the reticle
1Place a small workpiece on the stage glass. Turn the Micrometer Heads to align an edge of the
workpiece with the center of the cross-hair.
2Rotate the angle scale disc 180qand read the displacement between the edge of the workpiece
and the center of the cross-hair.
3Remove the four screws from the angle dial cover and dismount it. Slightly loosen the four fixing
screws.
4Adjust the eyepiece mount position with the four adjustment screws to minimize the
displacement between the edge of the workpiece and the center of the cross-hair. Centering is
easily performed by moving the eyepiece mount by half the displacement in both the X and Y
directions.
Two pairs of adjustment screws are located at the opposite sides (for adjusting the X and Y
displacements). Adjust the screws in pairs. Loosen one on one side, and then tighten its counterpart
on the other side to adjust the displacement.
5Turn the Micrometer Heads to align an edge of the workpiece with the center of the cross-hair.
Rotate the angle dial 180qand check the displacement.
6Repeat steps 1, 2, 4, 5 until the displacement is within 3μm.
7Confirm that the four adjustment screws are fully tightened.
8Tighten the four fixing screws and replace the angle dial cover.
1 2 4
Small workpiece
2. Installation and setup 15
2. Installation and setup
16
Measurement
This chapter describes the preparations for measurement and the
measuring procedures.
3.1 Preparations for Measurement
3.1.1Precautions for measurement
zInstallation site
When selecting an installation site, take vibration, dust and humidity into due consideration.
Vibration can affect measuring accuracy. Dust and humidity can impair optical parts, such as the
objective and prism, and the XY stage and moving parts.
zPrecaution for operation
If focusing, making measurements, or mounting work pieces, take surrounding conditions into
account. Be careful not to bump the objective, stage glass, etc.
zObjective and Eyepiece
The supplied objective and eyepiece were finely adjusted before shipment.
To maintain performance and accuracy, handle them with care and do not disassemble. Do not
allow the surface of the lens to be scratched or to be exposed to machine oil. If the lens is soiled,
clean it as described in “4.1 Dusting and Applying Oil”.
zStage glass
Since workpieces are mounted on it, the stage glass is likely to be scratched or even severely
damaged. Dust the workpiece before placing it on the stage glass. Exercise care so as not to
bump the stage glass with the workpiece. Do not slide the workpiece on the stage.
3. Measurement 17
3. Measurement
18
3.1.2 Replacing the lenses
An objective (2x) and an eyepiece (15x) are provided with a total magnification of 30x.
1The following objectives and eyepieces are available as optional accessories.
zEyepiece (10x, 20x)
z5x Objective (working distance: 33mm), 10x Objective (working distance: 14mm)
2. To mount the eyepiece, insert it into the eyepiece mount. The objective can be screwed into the
optical tube.
Before dismounting the objective, remove the fixture clamping down the surface illuminator from
the objective. The fixture is attached to the objective. The O-ring is between the fixture and the
objective. Pull down the fixture by gently swinging it back and forth.
3.1.3 Fixing the workpiece
Once the workpiece is mounted, secure it to the stage glass. Use the fixing jigs secured to the
T-groove on the XY stage if required.
The workpiece must be positioned so that the measuring surface faces the objective.
The following optional fixing jigs are available.
1. Swivel center support (No.176-105 for TM-505/505R, 172-197 for TM-510/510R)
Used to secure cylindrical workpieces (with center hole) and screws.
2. V-block (No.172-378)
Used to mount cylindrical workpieces (without center hole).
3. Holder with clamp (No.176-107)
Used to mount thin flat workpieces.
4. Chart clip (for XY stage) No.990561 (for TM-505/505R)
Used to mount thin flat workpieces (securing to the T-grooves on the XY stage).
Swivel center support
No.176-105
No.172-197
V-block
No.172-378
Holder with clamp
No.176-107
Chart clipfor XY stage
No.990561
3. Measurement 19
3.1.4 Illumination modes
The TM-500 series supports the following illumination modes. Select appropriate illumination mode
according to your application.
(1) Contour illumination
The contour illumination generates the contour image of a workpiece, and is suited for measurement
and inspection of workpiece contours. The illuminator is equipped with a green filter.
(2) Surface illumination
Surface illumination shows the surface of a workpiece, and is used in observation and inspection of
workpiece surfaces. Adjust the angle and orientation of this illuminator so the workpiece surface can
be observed under optimum conditions.
(3) Simultaneous use of contour and surface illuminations
Both the contour and surface of a workpiece can be observed simultaneously.
Power switch
Contour illumination switch
Surface illumination switch
Light control knob
3. Measurement
20
3.1.5 Replacing reticles
To replace reticles, use the supplied reticle setting screw, as described below.
1. Pull up and remove the eyepiece.
2. Screw the reticle setting screw into the reticle. Pull out the reticle setting screw along with the
reticle.
3. Screw the reticle setting screw into the reticle to be mounted. Insert the reticle into the eyepiece
mount.
The reticle is provided with a positioning pin. Fit the pin in the groove on the eyepiece mount and
insert the eyepiece as far as possible.
4. Remove the reticle setting screw and insert the eyepiece.
Reticle setting screw
Reticle setting screw
Reticle
Reticle
Positioning pin
Positioning pin
3.1.6 Adjusting the diopter
While looking into the eyepiece, turn the diopter adjustment ring until the reticle can be seen
sharply.
Diopter adjustment ring
Focusing knob
3. Measurement 21
3.1.7 Bringing the measuring surface into focus
Bring the measuring surface into focus by moving the optical tube up and down with the focusing
knob. Look into the eyepiece to make sure the cross-hairs are kept in ocular focus during this
focusing operation.
CAUTION
If moving the optical tube, be careful, especially if the workpiece is stepped or is secured with
fixing jigs, not to bump the workpiece.
3.1.8 Positioning the workpiece
Align the measuring direction of the workpiece with the traversing direction of the stage.
䋨䋱䋩Follow the procedure in "2.4.3 (1) Adjusting the reticle with the XY stage moving
direction".
䋨䋱䋩Move the workpiece or the fixing jigs to align the measuring direction of the workpiece with the
cross-hair reticle.
After making the above adjustment, confirm that the workpiece is parallel with the traveling direction
by moving the XY stage.
3. Measurement
22
3.2 Measurement
3.2.1 Dimensional measurement
Align a measuring point on the workpiece with one of the cross-hairs and take the reading from the
Micrometer Head. Then, move the XY stage by turning the Micrometer Head and align another
measuring point with the same cross-hair and take the reading at this point. The difference between
the two readings represents the dimension between the two measuring points.
A Digimatic Head and Digital Counter can be used, in place of the Micrometer Head, for digital
display of the displacement. They also eliminate reading errors. Since the zero-set button zeroes the
counter at any position, the displacement can be read directly.
3. Measurement 23
3.2.2 Angle measurement
Angles are measured with the angle dial, using either of the following two procedures.
(1) (2)
(1) Align an edge of the workpiece with the cross-hair reticle and align the end edge with the center
of the cross-hair. Turn the angle dial to align the cross-hair with the other edge of the workpiece.
Take readings from the angle dial.
(2) Align two edges of the workpiece with the same cross-hair, one after another, by turning the
angle dial and moving the XY stage. Take readings from the angle dial.
1In both procedures, measuring points on the workpiece are aligned with a cross-hair, one after
another. The angle is determined from the difference in readings.
2The resolution of the angle dial is 1qwith the main scale and 6' with the vernier scale.
㩷㩷㩷㩷㩷<Reading the angle dial>
Main scale34¡
34¡42
Vernier scale42
3The zero position of the angle dial can be adjusted by turning the vernier scale. This allows the
angle measurement origin to be set to 0.
CAUTION
After turning the vernier scale, check the reticle position. Refer to "2.4.3 (1) Adjusting the reticle
against the XY stage moving direction" if necessary.
3. Measurement
24
3.2.3 Template matching inspection
This Measuring Microscope allows inspection of screw threads and involute gear teeth using the
optional template reticles.
To perform template matching inspection, set the appropriate optional template reticle according to
the procedure in "3.1.5 Replacing the reticle".
For information about the various template reticles, see "7. Optional Accessories".
3.2.4 Step measurement
Stepped dimensions can be measured using the optional height measurement attachment and a dial
indicator.
CAUTION
This measurement is performed by bringing the measuring faces in focus. A high-precision
measurement cannot be made if using a lens with a large depth of focus. To obtain high-accuracy
measurements, use a high-magnification lens with a shallow depth of focus, as shown below.
30x lens with a depth of focus of approximate 0.12mm
150x lens with a depth of focus of approximate 0.02mm
Attach the height measurement attachment to the top face of the column as shown in the figure below.
Use a gauge block that is appropriate for the optical tube position.
Dial indicator
Mounting screw
Height measurement attachment
Gauge block
Column
Maintenance
This chapter describes the daily and periodical inspection and maintenance
required to maintain the performance of this microscope.
4.1 Cleaning and Lubrication
(1) Main unit
Periodically apply a thin layer of grease over the slide guide surfaces and rack of the optical tube using
a brush.
(2) XY stage
Apply a thin layer of spindle oil to guide rails. After dusting the stage glass wipe it gently with a soft
cloth.
(3) Eyepiece and objective
Since the optical glass used for lenses is soft and subject to scratches, always use an air-blower or a
feather to dust the lens surface. To remove contaminants such as oil and fingerprints, gently wipe them
in a circular motion with gauze dampened in high-grade alcohol.
4.2 Inspection
To maintain this microscope in prime condition, periodically inspect the parts specified below. If the
microscope is not operating perfectly, inspect it in detail according to the procedure in “5.
Troubleshooting”.
(1)Connecting parts
DANGER
Turn off the power switch and pull out the power cord from the AC outlet to prevent electric
shocks.
zCheck the power cord, input connector (AC inlet), voltage selector, GND terminal, surface
illuminator connector, and other joints for looseness and poor connections.
(2)Power and illumination switch and light control knob
zCheck the power switch
zCheck that the contour and surface illuminators light by turning on the illumination switch.
zCheck that the light intensity of each illuminator changes by turning on the light control knob.
4. Maintenance 25
4. Maintenance
26
(3)Focusing knob
zCheck this knob for any abnormal tightness, play, unevenness, and sound.
(4)XY stage
zCheck the stage glass for scratches and contaminants.
zMove the XY stage over the measuring range by hand to check for any abnormal tightness, play,
unevenness, and sound. (Perform the same cheek on the stage by turning the micrometer heads.)
(5)Angle dial
zLoosen the angle dial clamp knob and turn the dial to check for any abnormal tightness, play,
unevenness, and sound.
(6)Field of view
zLook into the eyepiece under contour illumination and check the entire filed of view for
vignetting and uneven illumination.
(7)XY stage feeding accuracy
zThis inspection will be affected by the measuring environment, alignment error, and other
adverse conditions. Keeping these factors in mind, check the feeding accuracy by measuring a
workpiece or a standard scale with an appropriate dimension of 5mm.
1. Position a workpiece, for which accurate dimensions are known, on the stage glass and
bring it into focus.
2. Align a measuring point on the workpiece with one of the cross-hairs, according to the
moving direction of the workpiece.
3. Obtain the dimension from the readings on the Micrometer Head. Refer to “3.2.1
Dimensional measurement”.
4. Measure the X- and Y-axis dimensions of the workpiece.
If the difference between the measured and nominal dimensions is less than 5Ǵm (for any 5mm
travel range), the XY stage feeding accuracy is adequate.
(8)Resolution
zPosition a workpiece on the XY stage and bring it into focus. Check if any region of the image in
the field of view has poor resolution.
4. Maintenance 27
4.3 Replacing Consumable Parts
(1) Replacing the fuse
(2)Replacing the stage glass
1. While pushing the stage glass in the direction of the leaf spring, raise the glass at the front edge,
and then remove it.
2. Attach the leaf spring in place and mount new stage glass.
Stage glass
Leaf spring
When a fuse holder is as in the upper left figure
1. Turn off the contour and surface illumination switches and the
power switch, and the pull out the power cord.
2. Insert a screwdriver in the rectangular hole of the fuse holder
and raise the screwdriver grip, then a latch will separate and the
fuse holder will come out from the front.
3. Replace the fuse, and replace the fuse holder.
When a fuse holder is as in the lower left figure
1. Turn off the contour and surface illuminator switches and the
power switch and pull out the power cord.
2. Insert a screwdriver in the rectangular hole on the fuse holder
and turn it counterclockwise, then a latch will separate and a
fuse holder will come out from the front.
3. Replace the fuse, and replace the fuse holder.
CAUTION
Since the stage glass is pushed forward with the leaf spring, when you remove the stage glass, be
careful not to lose the leaf spring.
4. Maintenance
28
(3)Replacing the contour illumination bulb
CAUTION
The bulb remains hot after it has been turned off. Do not replace the bulb until it had cooled down.
1. Turn off the contour illumination switch and power switch.
2. Remove the stage glass. (See (2) above.)
3. Turn the green filter counterclockwise to remove it.
4. Remove the bulb by turning it counterclockwise with the supplied bulb setter.
5. Mount a new bulb using the bulb setter.
6. Turn on the power switch and contour illumination switch. Check that the bulb lights.
7. Replace the green filter and the stage glass, in this order.
4. Maintenance 29
(4)Replacing the surface illumination bulb
CAUTION
The bulb remains hot after it has been turned off. Do not replace the bulb until it has cooled down.
1. Turn off the surface illumination switch and power switch.
2. Turn the white filter counterclockwise to remove it.
3. Remove the bulb by turning it counterclockwise with the supplied bulb setter. Mount a new bulb
using the bulb setter.
4. Switch the illuminator select switch knob to the surface illumination.
5. Check that the bulb lights.
6. Turn on the power switch and surface illumination switch. Check that the bulb lights.
7. Replace the white filter.
4.3 Consumable Parts
Order No. Name
380405 Stage glass 96x96 (for TM-505/505R)
380495 Stage glass 154x96 (for TM-510/510R)
512555 Vinyl cover
383038 Illumination bulb (24V)
350279 Fuse (1A) for other than CE
384204 Fuse (1A) for CE
380240 Leaf spring
4. Maintenance
30
Troubleshooting
If a problem occurs, diagnose and remedy the problem with the aid of the
following procedures.
(1)The contour illumination bulb will not light.
Check1Is the power switch turned on?
YES No Turn on the power switch.
Check2Is the contour illumination switch turned on?
YES No Turn on the contour illumination switch.
Check3Is the light control knob properly set?
YES No
Turn the knob clockwise to increase the light intensity (if
turned fully counterclockwise).
Check4Is the power cord connected correctly?
YES No Connect the power cord correctly.
Check5Is the voltage selector set to the correct voltage?
YES No Set the voltage selector to the correct voltage.
Check6Is the bulb normal (conductive)?
YES No Replace the bulb.
Check7Is the fuse normal (conductive)?
YES No Replace the fuse.
Check8Is the voltage from the AC outlet normal (supplied)?
YES No Supply power to the outlet.
Check9Is the voltage at the AC inlet on the main unit normal (input)?
YES No Replace the power cord.
Contact Mitutoyo.
5. Troubleshooting 31
5. Troubleshooting
32
㩿㪉㪀The surface illumination bulb will not light.
Check1Is the power switch turned on?
YES No Turn on the power switch.
Check2Is the surface illumination switch turned on?
YES No Turn on the surface illumination switch.
Check3Is the light control knob properly set?
YES No
Turn the knob clockwise to increase the light intensity (if
turned fully counterclockwise).
Check4Is the power cord connected correctly?
YES No Connect the power cord correctly.
Check5Is the voltage selector set to the correct voltage?
YES No Set the voltage selector to the correct voltage.
Check6Is the bulb normal (conductive)?
YES No Replace the bulb.
Check7Is the fuse normal (conductive)?
YES No Replace the fuse.
Check8Is the voltage from the AC outlet normal (supplied)?
YES No Supply power to the outlet.
Check9Is the voltage at the AC inlet on the main unit normal (input)?
YES No Replace the power cord.
Contact Mitutoyo.
(3)The light intensity of contour illumination cannot be adjusted
(if the contour illumination switch is turn on ).
CheckIs the contour illumination bulb lit?
YES No
Refer to "(1) The contour illumination bulb will not light".
Contact Mitutoyo.
5. Troubleshooting 33
(4)The light intensity of surface illumination cannot be adjusted
(if the surface illumination switch is turn on ).
CheckIs the surface illumination bulb lit?
YES No
Refer to "(2) The surface illumination bulb will not light".
Contact Mitutoyo.
(5)The optical tube will not move up and down properly.
If there is any abnormal tightness, play, unevenness, or sound when the optical tube is moved, contact
Mitutoyo. Should this be the case, do not continue to move the optical tube or disassemble it.
(6)The angle dial will not rotate properly.
If there is any abnormal tightness, play, unevenness, or sound when the optical tube is moved, contact
Mitutoyo. Should this be the case, do not continue to move the optical tube or disassemble it.
(7)The XY stage will not properly be fed.
If there is any abnormal tightness, play, unevenness, or sound when the XY stage is fed, contact Mitutoyo.
Should this be the case, do not continue to feed the XY stage or disassemble it. If the feeding accuracy has
deteriorated, determine whether the stage or the Micrometer Head is the problem and contact Mitutoyo.
(8)There is a shadow in the field of view.
Check for:
zObstruction between the contour illumination bulb and the objective .
zContaminated reticle.
zContaminated objective or eyepiece.
If the problem cannot be rectified, contact Mitutoyo.
(9)The resolution in a region of the field of view is poor.
Check for:
zPoor focusing (because of a stepped workpiece).
zOil soiling the workpiece.
zImproperly mounted objective or eyepiece.
zContaminated or damaged (such as scratch) objective or eyepiece.
zContaminated or damaged (such as scratch) reticle.
If the problem cannot be rectified, contact Mitutoyo.
5. Troubleshooting
34
Specifications
(1)Optical tube
zOptical axis:30qfrom vertical
zCross-hair reticule supplied
zAdjustable diopter
zAngle dial: Graduation 1q
Angleofrotation360q
Anglereading6’(vernier)
Vernier zero position adjustable
(2)Eyepiece
zMagnification :15
zField number :13
(3)Objective
zMagnification :2
zWorking distance :67mm (2.63”)
(4)XY stage
Order No. TM-505/TM-505R TM-510/TM-510R
Dimensions 152152mm 240152mm
Stage glass size 9696mm 15496mm
Feeding range
if used with gauge block5050mm 10050mm
Maximum height 115mm 107mm
workpiece 5Kg 5Kg
(5)Transmitted illuminator
zLight source :24V, 2W (special bulb)
zContinuously adjustable light intensity
zGreen filter supplied
6. Specifications 35
6. Specifications
36
(6)Surface illuminator
zLight source : 24V2W (special bulb)
zContinuously adjustable light intensity
(7)Mass
TM-505 TM-505R TM-510 TM-510R
13.5Kg 14.0Kg 14.5Kg 15.0Kg
(8)Main unit dimensions
Optional Accessories
OrderNo. Name
176-105 Swivel-center supportfor TM-505/TM-505R
172-197 Swivel-center support for TM-510/TM-510R
176-106 Rotary tablefor TM-505/TM-505R
172-196 Rotary table for TM-510/TM-510R
176-107 Holder with clamp
172-378 V-block with clamp
176-203 Twin-bulb surface illumination unit
176-204 Height measurement attachment
176-115 Eyepiece 10
176-117 Eyepiece 20
176-137 Objective 10
176-139 Objective 5
176-109 Reticle (Template) Metric screw p=0.251.0
176-110 Reticle (Template) Metric screw p=1.252.0
176-111 Reticle (Template) Concentric circles Ǿ0.054.00mm
176-112 Reticle (Template) Involute gear tooth m0.11.0, pressure angle 20q
reference rac
k
176-113 Reticle (Template) 55qangle line
176-114 Reticle (Template) 60qangle line
176-120 Reticle (Template) Whiworth screw thread (inch) 6026TPI
176-121 Reticle (Template) Whiworth screw thread (inch) 2418TPI
176-122 Reticle (Template) Whiworth screw thread (inch) 1611TPI
176-123 Reticle (Template) Unified screw thread 8028TPI
176-124 Reticle (Template) Unified screw thread 2414TPI
176-125 Reticle (Template) Unified screw thread 1310TPI
7. Optional Accessories 37
7. Optional Accessories
38
176-127 Reticle (Template) Unified screw thread NF8028TPI
176-128 Reticle (Template) Unified screw thread NF2414TPI
176-129 Reticle (Template) Unified screw thread NF1310TPI
176-130 Reticle (Template) Involute gear tooth m0.11.0 pressure 14” 30’ ,
reference rac
k
176-135 Reticle (Template) Concentric circles Ǿ.01” pitch, maximum diameter Ǿ.2”
176-140 Reticle (Template) ISO metric screw thread p=0.0750.7
176-141 Reticle (Template) ISO metric screw thread p=0.752.0
176-142 Reticle (Template) ISO unified screw thread 8028TPI
176-143 Reticle (Template) ISO unified screw thread 2414TPI
176-144 Reticle (Template) ISO unified screw thread 1310TPI
990561 Chart clip (for XY stage)
176-366̪Fiber-optic circular illumination unit
176-344̪Twin fiber optics illumination unit, twin fiber-optic illuminator
264-140ٌ̪
2D data processing unit QM-Data200
164-161 Digimatic micrometer head MHD-50M
164-162
ًDigimatic micrometer head MHD-2̢M
937387 ًConnecting cable (1m): connect between MHD-50M and MUX-10F
965013 ًConnecting cable (2m): connect between MHD-50M and MUX-10F
264-002 ًMUX-10F
12AAD194 ًConnecting cable C: connect between MUX-10F and QM-Data200
̪: The order number differs depending on the shipping destination.
ٌ: When using QM-Data200 it also requires MDH-50M (2 units), connecting cables (2 pcs of either
length), MUX-10F and connecting cable C.
Reference Diagrams
Top view of XY stage
TM-505/505R
Detail drawing of T-groove
8. Reference Diagrams 39
8. Reference Diagrams
40
TM-510/510R
Detail drawing of T-groove

Navigation menu