Nihon Kohden ZM-931PA Medical Telemetry Transmitter User Manual
Nihon Kohden Corporation Medical Telemetry Transmitter Users Manual
Users Manual
Transmitter ZM-920PA/ZM-921PA ZM-930PA/ZM-931PA If you have any comments or suggestions on this manual, please contact us at: www.nihonkohden.com 0614-007205B Copyright Notice The entire contents of this manual are copyrighted by Nihon Kohden. All rights are reserved. No part of this document may be reproduced, stored, or transmitted in any form or by any means (electronic, mechanical, photocopied, recorded, or otherwise) without the prior written permission of Nihon Kohden. GENERAL HANDLING PRECAUTIONS This device is intended for use only by qualified medical personnel. Use only Nihon Kohden approved products with this device. Use of non-approved products or in a non-approved manner may affect the performance specifications of the device. This includes, but is not limited to, batteries, recording paper, pens, extension cables, electrode leads, input boxes and AC power. Please read these precautions thoroughly before attempting to operate the instrument. 1. To safely and effectively use the instrument, its operation must be fully understood. 2. When installing or storing the instrument, take the following precautions: (1) Avoid moisture or contact with water, extreme atmospheric pressure, excessive humidity and temperatures, poorly ventilated areas, and dust, saline or sulphuric air. (2) Place the instrument on an even, level floor. Avoid vibration and mechanical shock, even during transport. (3) Avoid placing in an area where chemicals are stored or where there is danger of gas leakage. (4) Pay extra attention when the instrument is in combination with other instruments to avoid misdiagnosis or other problems. (5) Choose a room where a proper grounding facility is available. 3. Before Operation (1) Check that the instrument is in perfect operating order. (2) Check that the instrument is grounded properly. (3) Check that all cords are connected properly. (4) Pay extra attention when the instrument is in combination with other instruments to avoid misdiagnosis or other problems. (5) All circuitry used for direct patient connection must be doubly checked. (6) Check that battery level is acceptable and battery condition is good when using battery-operated models. Operator’s Manual ZM-920PA/921PA/930PA/931PA 4. During Operation (1) Both the instrument and the patient must receive continual, careful attention. (2) Turn power off or remove electrodes and/or transducers when necessary to assure the patient’s safety. (3) Avoid direct contact between the instrument housing and the patient. 5. To Shutdown After Use (1) Turn power off with all controls returned to their original positions. (2) Remove the cords gently; do not use force to remove them. (3) Clean the instrument together with all accessories for their next use. 6. The instrument must receive expert, professional attention for maintenance and repairs. When the instrument is not functioning properly, it should be clearly marked to avoid operation while it is out of order. 7. The instrument must not be altered or modified in any way. 8. Maintenance and Inspection: (1) The instrument and parts must undergo regular maintenance inspection at least every 6 months. (2) If stored for extended periods without being used, make sure prior to operation that the instrument is in perfect operating condition. (3) Technical information such as parts list, descriptions, calibration instructions or other information is available for qualified user technical personnel upon request from your Nihon Kohden representative. 9. When the instrument is used with an electrosurgical instrument, pay careful attention to the application and/or location of electrodes and/or transducers to avoid possible burn to the patient. 10. When the instrument is used with a defibrillator, make sure that the instrument is protected against defibrillator discharge. If not, remove patient cables and/or transducers from the instrument to avoid possible damage. ii Operator’s Manual ZM-910PA/921PA/930PA/931PA WARRANTY POLICY Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and workmanship for one year from the date of delivery. However, consumable materials such as recording paper, ink, stylus and battery are excluded from the warranty. NKC or its authorized agents will repair or replace any products which prove to be defective during the warranty period, provided these products are used as prescribed by the operating instructions given in the operator’s and service manuals. No other party is authorized to make any warranty or assume liability for NKC’s products. NKC will not recognize any other warranty, either implied or in writing. In addition, service, technical modification or any other product change performed by someone other than NKC or its authorized agents without prior consent of NKC may be cause for voiding this warranty. Defective products or parts must be returned to NKC or its authorized agents, along with an explanation of the failure. Shipping costs must be prepaid. This warranty does not apply to products that have been modified, disassembled, reinstalled or repaired without Nihon Kohden approval or which have been subjected to neglect or accident, damage due to accident,fire, lightning, vandalism, water or other casualty, improper installation or application, or on which the original identification marks have been removed. In the USA and Canada other warranty policies may apply. CAUTION United States law restricts this device to sale by or on the order of a physician. Equipment Authorization Requirement Operation of this equipment requires the prior coordination with a frequency coordinator designated by the FCC for the Wireless Medical Telemetry Service. Operator’s Manual ZM-920PA/921PA/930PA/931PA iii EMC RELATED CAUTION This equipment and/or system complies with IEC 60601-2 International Standard for electromagnetic compatibility for medical electrical equipment and/or system. However, an electromagnetic environment that exceeds the limits or levels stipulated in IEC 60601-1-2, can cause harmful interference to the equipment and/or system or cause the equipment and/or system to fail to perform its intended function or degrade its intended performance. Therefore, during the operation of the equipment and/or system, if there is any undesired deviation from its intended operational performance, you must avoid, identify and resolve the adverse electromagnetic effect before continuing to use the equipment and/or system. The following describes some common interference sources and remedial actions: 1. Strong electromagnetic interference from a nearby emitter source such as an authorized radio station or cellular phone: Install the equipment and/or system at another location. Keep the emitter source such as cellular phone away from the equipment and/or system, or turn off the cellular phone. 2. Radio-frequency interference from other equipment through the AC power supply of the equipment and/or system: Identify the cause of this interference and if possible remove this interference source. If this is not possible, use a different power supply. 3. Effect of direct or indirect electrostatic discharge: Make sure all users and patients in contact with the equipment and/or system are free from direct or indirect electrostatic energy before using it. A humid room can help lessen this problem. 4. Electromagnetic interference with any radio wave receiver such as radio or television: If the equipment and/or system interferes with any radio wave receiver, locate the equipment and/or system as far as possible from the radio wave receiver. iv Operator’s Manual ZM-910PA/921PA/930PA/931PA 5. Interference of lightning: When lightning occurs near the location where the equipment and/or system is installed, it may induce an excessive voltage in the equipment and/or system. In such a case, disconnect the AC power cord from the equipment and/or system and operate the equipment and/or system by battery power, or use an uninterruptible power supply. 6. Use with other equipment: When the equipment and/or system is adjacent to or stacked with other equipment, the equipment and/or system may affect the other equipment. Before use, check that the equipment and/or system operates normally with the other equipment. 7. Use of unspecified accessory, transducer and/or cable: When an unspecified accessory, transducer and/or cable is connected to this equipment and/or system, it may cause increased electromagnetic emission or decreased electromagnetic immunity. The specified configuration of this equipment and/or system complies with the electromagnetic requirements with the specified configuration. Only use this equipment and/or system with the specified configuration. 8. Use of unspecified configuration: When the equipment and/or system is used with the unspecified system configuration different than the configuration of EMC testing, it may cause increased electromagnetic emission or decreased electromagnetic immunity. Only use this equipment and/or system with the specified configuration. 9. Measurement with excessive sensitivity: The equipment and/or system is designed to measure bioelectrical signals with a specified sensitivity. If the equipment and/or system is used with excessive sensitivity, artifact may appear by electromagnetic interference and this may cause mis-diagnosis. When unexpected artifact appears, inspect the surrounding electromagnetic conditions and remove this artifact source. If the above suggested remedial actions do not solve the problem, consult your Nihon Kohden representative for additional suggestions. Operator’s Manual ZM-920PA/921PA/930PA/931PA Conventions Used in this Manual and Instrument Warnings, Cautions and Notes Warnings, cautions and notes are used in this manual to alert or signal the reader to specific information. WARNING A warning alerts the user to the possible injury or death associated with the use or misuse of the instrument. CAUTION A caution alerts the user to possible injury or problems with the instrument associated with its use or misuse such as instrument malfunction, instrument failure, damage to the instrument, or damage to other property. NOTE A note provides specific information, in the form of recommendations, prerequirements, alternative methods or supplemental information. vi Operator’s Manual ZM-910PA/921PA/930PA/931PA Explanations of the Symbols in this Manual and Instrument The following symbols found in this manual/instrument bear the respective descriptions as given. Symbol Description Symbol Description Power On Attention, consult operator’s manual Power Off Nurse call Defibrillation proof type BF applied part Replace battery Defibrillation proof type CF applied part Check electrode Direct current Operator’s Manual ZM-920PA/921PA/930PA/931PA vii Introduction The ZM-920PA/ZM-921PA/ZM-930PA/ZM-931PA transmitter transmits ECG and other data from a patient to a Nihon Kohden monitor for continuous monitoring. Available parameters and functions vary between the models. Read the operator’s manual for the monitor before operation. Model Parameters Functions The following information is indicated ZM-920PA • ECG by LED. ZM-921PA • Impedance respiration • Check ECG electrodes • Replace batteries The following information is indicated on LCD. • SpO2 value • ECG ZM-930PA • Pulse wave amplitude • Impedance respiration ZM-931PA • Replace batteries • SpO2 The following information is indicated by LED. • Check ECG electrodes The transmission frequency range is as follows. ZM-920PA/930PA: 608.0250 MHz (channel number 9002) to 613.9750 MHz (channel number 9478) ZM-921PA/931PA: 1395.0250 MHz (channel number E002) to 1399.9750 MHz (channel number E398) 1427.0250 MHz (channel number E502) to 1431.9750 MHz (channel number E898) The transmitter channel can be changed by the QI-901PK Channel Writer. To change the channel number, refer to the channel writer manual. Operator’s Manual ZM-920PA/921PA/930PA/931PA WARNING The following actions must be taken to properly receive the transmitter signal of the correct patient on the receiving instrument. Otherwise, there may be signal loss or signals may mix causing a serious accident, such as monitoring a different patient. • Assign a channel administrator in the hospital and only he or she should manage channel assignment. • The channel administrator must manage the channels in the facility so that there is no signal interference. • When the transmitter channel is changed, the channel administrator must check that the channel on the receiving monitor is also changed and the signal is properly received. • The channel administrator must replace the channel number label on the transmitter with the new one after changing the channel. WARNING Do not use the same transmitter on more than one patient at the same time. Do not connect different sensors on different patients to the same transmitter. CAUTION • Do not use the same channel for different patients. If the same channel is used for two patients, the two patients’ data will be lost due to mutual modulation interference, or another patient’s data may appear on the receiving monitor screen. • Do not use transmitters of adjacent channels in a hospital. If a transmitter of an adjacent channel is used, radio waves from one transmitter affect the receiver of the adjacent channel’s transmitter and there may be interference. Operator’s Manual ZM-920PA/921PA/930PA/931PA NOTE • To prevent interference between channels, assign a channel administrator in the hospital and only he or she should manage channel assignment. • Use Nihon Kohden parts and accessories to assure maximum performance from your instrument. • For stable signal reception, it is recommended to use a diversity antenna system on the receiving monitor. Otherwise, spike noise from transient fading of electric field strength (for example, people moving) may interfere with the transmitter signal and may be mistaken as an arrhythmia on the receiving monitor. • Do not diagnose a patient based on only part of the monitoring data on the transmitter or only on the data acquired by the transmitter. Overall judgement must be performed by a physician who understands the features, limitations and characteristics of the transmitter by reading this operator’s manual thoroughly and by reading the biomedical signals acquired by other instruments. • For details on the receiving monitor and upgrade information, contact your Nihon Kohden representative. Operator’s Manual ZM-920PA/921PA/930PA/931PA Panel Description Top Panel For attaching a strap ZM-930PA/931PA only SpO2 socket Refer to the warnings below Connects to the SpO2 probe. Refer to the symbol page Refer to the symbol page Refer to the warnings below ECG/RESP socket Connects to the electrode lead for measuring ECG and/or respiration by the impedance method. WARNING When performing defibrillation, discharge as far as possible from electrodes, patches and any gel, cream or medicine on the chest of the patient. If there is a possibility that the defibrillator paddle could touch these materials, remove them from the patient. If the defibrillator paddle directly contacts these materials, the discharged energy may cause skin burn to the patient. WARNING Before defibrillation, all persons must keep clear of the bed and must not touch the patient or any equipment or cord connected to the patient. Failure to follow this warning may cause electrical shock or injury. Operator’s Manual ZM-920PA/921PA/930PA/931PA Front Panel ZM-920PA/921PA Power switch Turns transmitter power on or off. Replace batteries LED Lights when the batteries need replacement. Check electrodes LED Lights when the electrode is detached from the patient. RA LA LL Va Vb USAch E002 1395.025 MHz Refer to the warning below. CALL key When this key is pressed, a “peep” sounds at the transmitter, and “CALL” message appears at the monitor. Depending on the settings on the monitor, an ECG waveform is recorded when this key is pressed. Battery case Contains two 1.5 V dry cell batteries (AA). WARNING Close the battery case cover during operation. If the transmitter is used with the battery case cover open, anyone who touches the opened battery case may receive an electrical shock when defibrillation is performed. Touching the opened battery case may cause electrostatic discharge and intermittently interfere with the waveform or data. CAUTION Only use your finger to press the CALL key. Do not press the key with a sharp object. Otherwise the key may be damaged. Operator’s Manual ZM-920PA/921PA/930PA/931PA ZM-930PA/931PA For the descriptions except for the LCD, refer to the “ZM-920PA/921PA” section on the previous page. RA LA LL Va Vb USAch E002 1395.025 MHz LCD %SpO2: Battery replacement mark: Appears when the batteries are weak. Immediately replace the batteries when this appears. Displayed when the power is turned on. This indication is not displayed when SpO2 display is turned off. Pulse level bar graph: Displays pulse level in 7 steps. SpO2 data: Displays SpO2 data. When SpO2 is 41 to 100%, the value is displayed. (ex. ) When SpO2 is under 40%, “ ” is displayed. When the detected pulse is too small to measure, “ ” is displayed. Operator’s Manual ZM-920PA/921PA/930PA/931PA Important Safety Information General WARNING Never use the transmitter in the presence of any flammable anesthetic gas or high concentration oxygen atmosphere. Failure to follow this warning may cause explosion or fire. WARNING Never use the transmitter in a hyperbaric oxygen chamber. Failure to follow this warning may cause explosion or fire. WARNING Do not take this transmitter into the MRI test room. This transmitter is not designed to be used during MRI tests. WARNING When performing MRI test, remove all electrodes from the patient which are connected to this transmitter. Failure to follow this warning may cause skin burn on the patient. For details, refer to the MRI manual. WARNING When performing defibrillation, discharge as far as possible from electrodes, patches and any gel, cream or medicine on the chest of the patient. If there is a possibility that the defibrillator paddle could touch these materials, remove them from the patient. If the defibrillator paddle directly contacts these materials, the discharged energy may cause skin burn to the patient. Operator’s Manual ZM-920PA/921PA/930PA/931PA WARNING Before defibrillation, all persons must keep clear of the bed and must not touch the patient or any equipment or cord connected to the patient. Failure to follow this warning may cause electrical shock or injury. WARNING When the transmitter is used with an electrosurgical unit (ESU), firmly attach the entire area of the ESU return plate. Otherwise, the current from the ESU flows into the electrodes of the transmitter, causing electrical burn where the electrodes are attached. For details, refer to the ESU manual. WARNING Close the battery case cover during operation. If the transmitter is used with the battery case cover open, anyone who touches the opened battery case may receive an electrical shock when defibrillation is performed. Touching the opened battery case may cause electrostatic discharge and intermittently interfere with the waveform or data. WARNING If detergent or liquid spills into the transmitter, clean it and dry it completely before use. If a wet transmitter is used, the patient or operator may receive an electrical shock or injury. WARNING Do not use the same transmitter on more than one patient at the same time. Do not connect different sensors on different patients to the same transmitter. Operator’s Manual ZM-920PA/921PA/930PA/931PA CAUTION Only use Nihon Kohden specified electrodes, electrode leads and SpO2 probes. Otherwise, the maximum performance from the transmitter cannot be guaranteed. CAUTION Do not reuse disposable parts and accessories. CAUTION Do not shake or swing the transmitter while holding the leads or cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments. CAUTION Attach a strap to the transmitter to prevent the transmtiter from falling. CAUTION Turn off the power of mobile phones, small wireless devices and other devices which produce strong electromagnetic interference around a patient (except for devices allowed by the hospital administrator). Radio waves from devices such as mobile phones or small wireless devices may be mistaken as pulse waves and the displayed data may be incorrect. Operator’s Manual ZM-920PA/921PA/930PA/931PA CAUTION • Do not use the same channel for different patients. If the same channel is used for two patients, the two patients’ data will be lost due to mutual modulation interference, or another patient’s data may appear on the receiving monitor screen. • Do not use transmitters of adjacent channels in a hospital. If a transmitter of an adjacent channel is used, radio waves from one transmitter affect the receiver of the adjacent channel’s transmitter and there may be interference. Battery WARNING • Keep the batteries away from fire. They may explode. • Keep the batteries away from patients. • Never short-circuit the + and – terminals on the battery. It may cause overheating and fire. • Do not damage, disassemble, drop or give impact to the battery. WARNING If the battery is damaged and the substance inside the battery contacts the eyes or skin, wash immediately and thoroughly with water and see a physician. Never rub your eyes, because you may lose your eyesight. CAUTION Battery replacement must be performed by the operator. When replacing batteries of the transmitter currently used for a patient, disconnect electrode leads from the transmitter before replacing batteries or do not touch the patient during replacement. 10 Operator’s Manual ZM-920PA/921PA/930PA/931PA CAUTION The battery charger must be used outside the patient environment. CAUTION Refer to the battery and battery charger manuals for details on handling the batteries. Transmitter Channel Management WARNING The following actions must be taken to properly receive the transmitter signal of the correct patient on the receiving instrument. Otherwise, there may be signal loss or signals may mix causing a serious accident, such as monitoring a different patient. • Assign a channel administrator in the hospital and only he or she should manage channel assignment. • The channel administrator must manage the channels in the facility so that there is no signal interference. • When the transmitter channel is changed, the channel administrator must check that the channel on the receiving monitor is also changed and the signal is properly received. • The channel administrator must replace the channel number label on the transmitter with the new one after changing the channel. Operator’s Manual ZM-920PA/921PA/930PA/931PA 11 For Patients Using Implantable Pacemaker WARNING Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac Monitoring and Diagnostic Equipment The bioelectric impedance measurement sensor of a minute ventilation rate-adaptive implantable pacemaker may be affected by transmitter which is connected to the same patient. If this occurs, the pacemaker may pace at its maximum rate and the transmitter may give incorrect data to the monitor. If this occurs, disconnect the electrode leads from the patient or change the setting on the pacemaker by referring to the pacemaker’s manual. For more details, contact your pacemaker representative or Nihon Kohden representative. Output Signal WARNING Do not use the output signal from the receiving monitor as the synchronization signal for other equipment such as IABP, MRI, echocardiography or defibrillator. There may be time delay between the monitor and the other equipment caused by waveform transmission delay and spike noise may interfere on the output signal and be mistaken as a trigger. 12 Operator’s Manual ZM-920PA/921PA/930PA/931PA ECG Monitoring CAUTION Only use Nihon Kohden specified electrodes and electrode leads. When other type of electrodes or electrode leads are used, the “CHECK ELECTRODES” message may be displayed and monitoring may stop. CAUTION When the “ELECTRODE OFF” or “CHECK ELECTRODE” message is displayed on the receiving monitor, ECG is not monitored properly and the ECG alarm does not function. Check the electrode, electrode leads, and if necessary, replace with new ones. SpO2 Monitoring WARNING SpO2 measurement may be incorrect in the following cases. • When the patient’s carboxyhemoglobin or methemoglobin increases abnormally. • When dye is injected in the blood. • When using an electrosurgical unit. • During CPR. • When measuring at a site with venous pulse. • When there is body movement. • When the pulse wave is small (insufficient peripheral circulation). Operator’s Manual ZM-920PA/921PA/930PA/931PA 13 WARNING Check the circulation condition by observing the skin color at the measurement site and pulse waveform. Change the measurement site every 8 hours for disposable probes and every 4 hours for reusable probes (every 8 hours for TL-630T/TL-631T series probe). The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree. • Patient with a fever • Patient with peripheral circulation insufficiency • Neonate or low birth weight infant with delicate skin • Patient who is receiving photodynamic therapy* * Photodynamic therapy is a treatment to remove the affected tissue by using a photosensitizing agent and exposing the tissue to light. This treatment has a side effect of photosensitivity and the light from the finger probe sensor may cause a burn. This probe uses two light wavelengths in the range from 650 to 950 nm. The maximum light intensity is less than 5.5 mW/sr. WARNING When not monitoring SpO2, disconnect the SpO2 cable from the transmitter. Otherwise, noise from the probe sensor may interfere and incorrect data is displayed on the screen. 14 Operator’s Manual ZM-920PA/921PA/930PA/931PA WARNING • When using the TL-201T finger probe, do not fasten the probe and cable to the finger by wrapping with tape. This may cause burn, congestion or pressure necrosis from poor blood circulation. • When using probes other than the TL-201T finger probe, to avoid poor circulation, do not wrap the tape too tight. Check the blood circulation condition by observing the skin color and congestion at the skin peripheral to the probe attachment site. Even for shortterm monitoring, there may be burn or pressure necrosis from poor blood circulation, especially on neonates or low birth weight infants whose skin is delicate. Accurate measurement cannot be performed on a site with poor peripheral circulation. CAUTION While a patient is on medication which causes vasodilation, the pulse waveform may change and in rare cases the SpO2 value might not be displayed. CAUTION Normal external light does not affect monitoring but strong light such as a surgical light or sunlight may affect monitoring. If affected, cover the measuring site with a blanket. CAUTION If the attachment site is dirty with blood or bodily fluids, clean the attachment site before attaching the probe. If there is nail polish on the attachment site, remove the polish. Otherwise, the amount of transmitted light decreases, and measured value may be incorrect or measurement cannot be performed. Operator’s Manual ZM-920PA/921PA/930PA/931PA 15 CAUTION Do not use a probe which is deteriorated by aging. Accurate measurement cannot be performed. CAUTION Do not use a damaged or disassembled probe. It causes incorrect measurement and may injure the patient. CAUTION The disposable probe is not sterilized. Use the disposable probe only for a single patient. Never reuse the disposable probe for another patient because it causes cross infection. CAUTION If the skin gets irritated or redness appears on the skin from the probe, change the attachment site or stop using the probe. Take extreme care for the patients with delicate skin. CAUTION Do not pull or bend the probe cable, and do not put caster feet on the probe cable. Do not immerse the probe cable in chemical solutions or water. Failure to follow these instructions may cause cable discontinuity, short circuit, skin burn on the patient and incorrect measurement data. Replace any broken probe with a new one. CAUTION When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one. 16 Operator’s Manual ZM-920PA/921PA/930PA/931PA CAUTION When a message indicates a faulty probe, stop monitoring and replace the probe with a new one. CAUTION When any of the following conditions occurs, the probe may be broken. Replace it with a new one. • The transmitter generates “pip” sounds. • SpO2 data is 85% and blinking. CAUTION Neonatal skin is delicate. Remove the probe and tape carefully and slowly. CAUTION When removing a probe that is taped to the skin, do not pull the probe cable because this can damage the cable. CAUTION When removing the probe from the attachment tape, do not pull the sensor cable because this can damage the cable. CAUTION Do not immerse the disposable probe in detergents or water. If the probe adhesive surface gets wet, adhesiveness becomes weak and the probe cannot be attached to the skin. CAUTION Refer to the probe instruction manual for details. Operator’s Manual ZM-920PA/921PA/930PA/931PA 17 Maintenance CAUTION If detergent or liquid spills into the transmitter, clean it and dry it completely before use. If a wet transmitter is used, the transmitter may malfunction or get damaged. CAUTION Never disassemble or repair the transmitter. Disassembly and repair must be performed by qualified service personnel. CAUTION Before cleaning or disinfection, remove the batteries from the transmitter. Failure to follow this instruction may result in electrical shock or transmitter malfunction. CAUTION The transmitter cannot be sterilized. Sterilizing the transmitter may damage it. 18 Operator’s Manual ZM-920PA/921PA/930PA/931PA Preparation Installing (Replacing) Batteries Use two AA(R6) type alkaline dry cell batteries, manganese dry cell batteries, NiCd rechargeable batteries or NiMH batteries. Battery Lifetime With new alkaline batteries, the transmitter can continuously measure for the following number of days. Operation time depends on the thickness of the SpO2 probe attachment site. Transmitter ZM-920PA/930PA ZM-921PA/931PA Operating time (Measuring parameters) ECG, Resp, SpO2 ECG, Resp 3 days 4 days 2.5 days 3.5 days NOTE The capacity of manganese, NiCd and NiMH batteries is less than that of alkaline batteries, therefore the lifetime of the battery is shorter. Type Manganese NiCd NiMH Lifetime About 1/2 of alkaline batteries About 1/3 of alkaline batteries (when fully charged) About 1/2 of alkaline batteries (when fully charged) Recommended Batteries Alkaline primary: Nihon Kohden Medipower (equivalent to Panasonic LR6 (G)) CAUTION Battery replacement must be performed by the operator. When replacing batteries of the transmitter currently used for a patient, disconnect electrode leads from the transmitter before replacing batteries or do not touch the patient during replacement. Operator’s Manual ZM-920PA/921PA/930PA/931PA 19 If electrode leads are attached to the patient and the person replacing the batteries touches the patient, the patient leakage current over the amount allowed may occur. CAUTION • Replace all batteries at the same time. • Do not use different types of batteries together. NOTE Insert the batteries with the correct polarity (+ and −). Procedure 1. Open the battery case cover. 2. Insert two dry cell batteries (LR6) into the battery case observing the correct polarity. 3. Close the cover and press it gently until it clicks. WARNING and CAUTION for Battery Handling WARNING • Keep the batteries away from fire. They may explode. • Keep the batteries away from patients. • Never short-circuit the + and – terminals on the battery. It may cause overheating and fire. • Do not damage, disassemble, drop or give impact to the battery. 20 Operator’s Manual ZM-920PA/921PA/930PA/931PA WARNING If the battery is damaged and the substance inside the battery contacts the eyes or skin, wash immediately and thoroughly with water and see a physician. Never rub your eyes, because you may lose your eyesight. CAUTION Do not handle the batteries with wet hands. CAUTION When the transmitter is not in use, remove batteries. When batteries are installed, battery power is consumed even if measurement is not performed. Especially, when NiMH batteries remain in the transmitter when the transmitter is not in use, the battery may become unusable from overdischarge and leak liquid which will damage the transmitter. CAUTION Refer to the battery and battery charger manuals for details on handling the batteries. CAUTION The battery charger must be used outside the patient environment. NOTE • When using rechargeable NiCd batteries or NiMH batteries, shallow charging/discharging shortens battery capacity. For details, refer to the battery operator’s manual. • Remove the batteries from the transmitter before disposing of the transmitter. Operator’s Manual ZM-920PA/921PA/930PA/931PA 21 Situations Requiring Battery Replacement Replace the batteries when any of the following occurs. • The “ ” LED lights (ZM-920PA/921PA) or the “ ” mark is displayed on the LCD (ZM-930PA/931PA) on the transmitter. • The transmitter generates a constant alarm (continuous “peep” sound). • The monitor displays the battery replacement message on the screen. • When the power of the LCD transmitter is turned on, no message or icon is displayed. (Only the ZM-930PA/931PA). CAUTION When using batteries other than alkaline, the interval between the battery replacement indication and battery discharge may be shorter. Replace the batteries immediately upon battery replacement indication. Attaching a Strap to the Transmitter To open the clip, firmly pull out the tab in direction of the arrow. CAUTION Attach a strap to the transmitter to prevent the transmtiter from falling. NOTE Do not attach the clip to hard objects such as thick cloth or zippers, or the clip may break. Attach a strap to the transmitter and fasten the clip to the patients’ clothes or bed sheets. 22 Operator’s Manual ZM-920PA/921PA/930PA/931PA If the transmitter falls off, the battery cover may be opened. If the patient touches the terminals of the batteries, patient leakage current over the allowable amount can occur. Turning On/Off the Transmitter Check Items Before Turning On the Power To use the instrument in a safe and optimum condition, before turning on the transmitter power switch, check the following. Appearance • There is no damage or dirt on the outside of the transmitter. (Power switch, LED, LCD, CALL key, junction, battery case cover, battery case, etc.) • The transmitter is completely dry. • The electrode lead is not broken. • There is no damage or dirt on the SpO2 probe or on the disposable electrodes. Battery • The battery polarity is correct. • The battery case spring is firmly fixed and the battery is not loose. • The battery case cover is firmly closed. Channel Setting • The transmitter channel corresponds to those of the receiving monitor. • The same channel is not being used on a different transmitter in the surrounding area. Turning On/Off the Power ON OFF To turn on the power, turn the power switch to the right. After a “peep” sound for about one second, the power is turned on. (There is no “peep” sound when the “ ” LED light or the “ ” are blinking on the LCD). To turn off the power, turn the power switch to the left. Operator’s Manual ZM-920PA/921PA/930PA/931PA 23 Check Items After Turning On the Power After turning on the power, check the following items. Power On • The power switch is not damaged. • The transmitter generates a “peep” sound for about one second. • All LEDs light and values are displayed on the LCD for about one second. • The transmitter does not generate a continuous “pip” sound. • The transmitter does not liberate excessive heat. • The “ ” LED does not light or the “ ” mark is not displayed on the LCD. • The transmitter does not interfere with the operation of medical instruments used near it. Basic Operation • The “signal loss” message is not displayed on the monitor when the transmitter is inside the receiving range of the monitor. • A “peep” sounds at the transmitter and “CALL” message appears at the receiving monitor when the CALL key is pressed and the transmitter is inside the receiving range of the monitor. • The battery replacement message is not displayed on the monitor. Check Items After the Power Off • ECG electrode leads and SpO2 probe are cleaned and disinfected. • When the transmitter gets wet, liquid is wiped off and the transmitter is thoroughly dried. • There are enough consumables, such as disposable electrodes. • The power is turned off. • The batteries are removed from the transmitter when it will not be used for a long time. • Dead batteries are disposed of properly. 24 Operator’s Manual ZM-920PA/921PA/930PA/931PA ECG Monitoring When 6 leads are used on this transmitter, up to 8 lead (I, II, III, aVR, aVL, aVF, Va and Vb) of ECG waveforms can be displayed on the receiving monitor. The heart rate is also measured. When 3 leads are used, one channel ECG waveform of lead II can be displayed on the receiving monitor. Refer to the operator’s manual of the monitor for details. WARNING Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac Monitoring and Diagnostic Equipment* The bioelectric impedance measurement sensor of a minute ventilation rate-adaptive implantable pacemaker may be affected by transmitter which is connected to the same patient. If this occurs, the pacemaker may pace at its maximum rate and the transmitter may give incorrect data to the monitor. If this occurs, disconnect the electrode leads from the patient or change the setting on the pacemaker by referring to the pacemaker’s manual. For more details, contact your pacemaker representative or Nihon Kohden representative. * Minute ventilation is sensed in rate-adaptive pacemakers by a technology known as bioelectric impedance measurement (BIM). Many medical devices in addition to pacemakers use this technology. When one of these devices is used on a patient with an active, minute ventilation rate-adaptive pacemaker, the pacemaker may erroneously interpret the mixture of BIM signals created in the patient, resulting in an elevated pacing rate. For more information, see the FDA web site. http://www.fda.gov/cdrh/safety.html Operator’s Manual ZM-920PA/921PA/930PA/931PA 25 WARNING When the transmitter is used with an electrosurgical unit (ESU), firmly attach the entire area of the ESU return plate. Otherwise, the current from the ESU flows into the electrodes of the transmitter, causing electrical burn where the electrodes are attached. For details, refer to the ESU manual. NOTE • This transmitter is not protected against noise generated from an electrosurgical unit. • If an electric blanket is used and incorrect heart rate is displayed on the monitor, turn off the pacing pulse detection on the monitor. ECG Measurement Procedure 1. Select the type of electrode lead and disposable electrode according to the purpose. 2. Connect the electrode lead to the ECG/RESP socket. 3. Connect disposable electrodes to the electrode lead and attach electrodes to the patient. After steps 1 to 3 are finished, ECG monitoring automatically starts. 26 Operator’s Manual ZM-920PA/921PA/930PA/931PA Selecting Electrode Lead and Disposable Electrode CAUTION Only use Nihon Kohden specified electrodes and electrode leads. When other type of electrodes or electrode leads are used, the “CHECK ELECTRODES” message may be displayed and monitoring may stop. Option Electrode lead BR-916PA BR-906PA 6 electrodes, snap type 6 electrodes, clip type BR-913PA BR-903PA 3 electrodes, snap type 3 electrodes, clip type Operator’s Manual ZM-920PA/921PA/930PA/931PA 27 Connecting the Electrode Lead to the Transmitter Connect the electrode lead to the ECG/RESP socket on the transmitter. CAUTION Do not shake or swing the transmitter while holding the leads or cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments. CAUTION Hold the connector of the electrode lead when connecting/ disconnecting the electrode lead. If you disconnect the electrode lead by pulling the lead, it damages the electrode lead. Selecting the Electrode Position Follow the physician’s instructions for electrode placement when available. For ECG monitoring, electrodes are attached only on the chest to allow patient movement and obtain continuous stable ECG. Following leads are examples. When also monitoring respiration, refer to the “Electrode Position for Respiration Monitoring” section. NOTE The optimum electrode positions for ECG measurement of a patient are not always optimum for respiration measurement of the patient. Select positions suitable for both ECG and respiration measurements, or positions which have priority for one measurement. 28 Operator’s Manual ZM-920PA/921PA/930PA/931PA Six Electrodes Electrode Position The 6-electrode method with lead II and lead V5 is effective for monitoring myocardial ischemia. You can improve monitoring accuracy considerably by adding lead V4 to this combination. Va and Vb can be at any position of the standard 12 leads V1 to V6, but V4 and V5 are most appropriate for myocardial ischemic monitoring. LA RA Va N (RL) Symbol Vb LL Lead Color (Clip Color) Electrode Position RA White (White) Right infraclavicular fossa LA Black (Black) Left infraclavicular fossa LL Red (Red) N (RL) Green (Green) Va Brown (Brown) (BR-906PA) Brown-blue (BR-916PA) Vb Brown (Brown) (BR-906PA) Brown-orange (BR-916PA) Lowest rib on the left anterior axillary line Right anterior axillary line at the same level as LL Fifth intercostal space on the left midclavicular line. (V4 position of standard 12 leads) Left anterior axillary line at the same level as Va. (V5 position of standard 12 leads) Operator’s Manual ZM-920PA/921PA/930PA/931PA 29 Lead Position Standard limb leads Lead I RA Lead II LA Lead III RA LL N (RL) Monopolar limb leads aVR lead N (RL) aVL lead aVF lead RA LA RA LA N (RL) N (RL) LA LL LL LL LA LL LL N (RL) RA RA LA N (RL) Monopolar chest leads V1 to V6 leads to LA RA LL N (RL) When Using 4 to 6 DIN Type Leads to Monitor 6 Lead ECG ON RA LA LL Va Vb USAch E002 1395.0250 MHz CALL key When the BR-906PA/916PA electrode leads are not used, the transmitter is fixed to 3 lead ECG monitoring. To monitor 6 lead ECG using 4 to 6 DIN type leads, the transmitter must be fixed to 6 lead monitoring. To fix transmitter to the 6 lead ECG monitoring, turn off the transmitter power, press and hold the CALL key and turn on the transmitter power. When the transmitter power is turned off and on again, the monitoring mode returns to the original mode. 30 Operator’s Manual ZM-920PA/921PA/930PA/931PA Three Electrodes By using the optional BR-913PA/903PA electrode lead, 3 lead ECG monitoring is available. Electrode Position • Lead MII, which is similar to standard lead II, used when ECG measurement has priority Electrode Position Symbol Lead Color Left infraclavicular fossa LA (N) Black Right infraclavicular fossa RA (−) White Below lowest rib on the LL (+) Red left anterior axillary line • Lead MI, which is similar to standard lead I Change LL and LA of the lead MII. • Lead MIII, which is similar to standard lead III Change RA and LA of the lead MII. If the electrode position shown above is not available due to chest surgery, attach the electrodes to the root of the limbs or below the clavicles for stable ECG monitoring. Operator’s Manual ZM-920PA/921PA/930PA/931PA 31 Connecting the Electrode Lead and Disposable Electrodes Preparing the Patient Skin Shave off excessive body hair. To reduce skin impedance, clean the electrode site with cream or with a cotton pad moistened with the electrode site with cream or with a cotton pad moistened with alcohol. Thoroughly dry the skin with a clean cotton pad. NOTE • For a patient with frequent body movement, rub the sites with Skinpure skin preparation gel. However, do not use Skinpure skin preparation gel for sensitive skin. • Do not place electrodes on a wound or on an inflamed, wrinkled or uneven skin surface. Attaching Electrodes to the Patient CAUTION Do not reuse disposable parts and accessories. NOTE • To maintain good contact between the electrode and skin, check that the paste of the disposable electrode is not dry. • When contact between the disposable electrode and skin becomes poor, replace electrodes with new ones immediately. Otherwise, contact impedance between the skin and the electrode increases and the correct ECG cannot be obtained. Refer to the electrode operator’s manual for details. 1. Connect the electrode lead to the electrode. 32 Operator’s Manual ZM-920PA/921PA/930PA/931PA 2. Carefully remove the backing paper from the electrode. Avoid touching the adhesive surface. 3. Place the electrode on the previously cleaned skin. Pay attention to the electrode lead color and symbol. 4. Fasten the electrode lead wire with surgical tape with an extra length of wire between the tape and the electrode. This lessens the movement of electrode leads by body movement and helps stable monitoring. Detection and Display of Measurement Condition Electrode Detachment The “ ” LED lights on the transmitter or the “CHECK ELECTRODE” message is displayed on the screen of the monitor in the following cases. • Electrode is detached from skin. • Electrode lead is disconnected from the electrode. • Polarization voltage between the electrode and skin is excessively high. In these cases, check the cause and if necessary, replace electrodes with new ones. CAUTION When the “ELECTRODE OFF” or “CHECK ELECTRODE” message is displayed on the receiving monitor, ECG is not monitored properly and the ECG alarm does not function. Check the electrode, electrode leads, and if necessary, replace with new ones. Operator’s Manual ZM-920PA/921PA/930PA/931PA 33 Respiration Monitoring Respiration is monitored by measuring changes in impedance between the RA and LL ECG electrodes. This transmitter sends the changes in impedance to the monitor as a respiration waveform. The monitor displays the respiration waveform and calculates respiration rate. Refer to the operator’s manual of the monitor for details. WARNING Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac Monitoring and Diagnostic Equipment* The bioelectric impedance measurement sensor of a minute ventilation rate-adaptive implantable pacemaker may be affected by transmitter which is connected to the same patient. If this occurs, the pacemaker may pace at its maximum rate and the transmitter may give incorrect data to the monitor. If this occurs, disconnect the electrode leads from the patient or change the setting on the pacemaker by referring to the pacemaker’s manual. For more details, contact your pacemaker representative or Nihon Kohden representative. * Minute ventilation is sensed in rate-adaptive pacemakers by a technology known as bioelectric impedance measurement (BIM). Many medical devices in addition to pacemakers use this technology. When one of these devices is used on a patient with an active, minute ventilation rate-adaptive pacemaker, thepacemaker may erroneously interpret the mixture of BIM signals created in the patient, resulting in an elevated pacing rate. For more information, see the FDA web site. http://www.fda.gov/cdrh/safety.html 34 Operator’s Manual ZM-920PA/921PA/930PA/931PA Respiration Measurement Procedure 1. Select the electrode lead and disposable electrodes. 2. Connect the electrode lead to the ECG/RESP socket. 3. Connect disposable electrodes to the electrode lead and attach electrodes to the patient. After steps 1 to 3 are finished, respiration monitoring automatically starts. Electrode Position for Respiration Monitoring Place the RA and LL electrodes so that the lungs are between the electrodes. NOTE The optimum electrode positions for ECG measurement of a patient are not always optimum for respiration measurement of the patient. Select positions suitable for both ECG and respiration measurements, or positions which have priority for one measurement. Electrode Position Examples NOTE The following examples are when monitoring with 3 electrodes. ECG cannot be monitored correctly when electrodes are attached as the following examples when monitoring with 6 electrodes. Position 1 In this position, respiration measurement is available; however, there is a difference in amplitude between different patients. RA Right infraclavicular fossa Operator’s Manual ZM-920PA/921PA/930PA/931PA LL Fifth intercostal space on the left midclavicular line, V4 35 Position 2 In this position, the waveform amplitude is usually large and the ECG lead is similar to Lead MII. This position can be generally recommended. RA Right infraclavicular fossa LL Fifth intercostal space on the left midaxillary line, V6 Position 3 In this position, the respiration waveform is optimum, but the ECG lead is unusual. RA LL Right midaxillary at the Fifth intercostal space on the horizontal level of V4 left midaxillary line, V6 Position 4 In this position, the respiration measurement is influenced by the impedance variation of the abdomen, so the cardiac pulse wave included in the respiration wave is reduced. Note that the waveform is inverted in phase compared with the chest movement (the waveform goes down during inspiration). It is difficult to measure the ECG at the same time. RA Lowest rib on the right anterior axillary line 36 LL Lowest rib on the left anterior axillary line Operator’s Manual ZM-920PA/921PA/930PA/931PA SpO2 Monitoring The SpO2 monitoring is only available on the ZM-930PA/931PA transmitter. This transmitter sends SpO2 and pulse waveform to the monitor and displays SpO2 data and pulse level bar graph on the LCD. Refer to the operator’s manual of the monitor for details. WARNING SpO2 measurement may be incorrect in the following cases. • When the patient’s carboxyhemoglobin or methemoglobin increases abnormally. • When dye is injected in the blood. • When using an electrosurgical unit. • During CPR. • When measuring at a site with venous pulse. • When there is body movement. • When the pulse wave is small (insufficient peripheral circulation). WARNING When not monitoring SpO2, disconnect the SpO2 cable from the transmitter. Otherwise, noise from the probe sensor may interfere and incorrect data is displayed on the screen. Operator’s Manual ZM-920PA/921PA/930PA/931PA 37 WARNING Check the circulation condition by observing the skin color at the measurement site and pulse waveform. Change the measurement site every 8 hours for disposable probes and every 4 hours for reusable probes (every 8 hours for TL-630T/TL-631T series probe). The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree. • Patient with a fever • Patient with peripheral circulation insufficiency • Neonate or low birth weight infant with delicate skin • Patient who is receiving photodynamic therapy* * Photodynamic therapy is a treatment to remove the affected tissue by using a photosensitizing agent and exposing the tissue to light. This treatment has a side effect of photosensitivity and the light from the finger probe sensor may cause a burn. This probe uses two light wavelengths in the range from 650 to 950 nm. The maximum light intensity is less than 5.5 mW/sr. CAUTION Turn off the power of mobile phones, small wireless devices and other devices which produce strong electromagnetic interference around a patient (except for devices allowed by the hospital administrator). Radio waves from devices such as mobile phones or small wireless devices may be mistaken as pulse waves and the displayed data may be incorrect. CAUTION Normal external light does not affect monitoring but strong light such as a surgical light or sunlight may affect monitoring. If affected, cover the measuring site with a blanket. 38 Operator’s Manual ZM-920PA/921PA/930PA/931PA CAUTION Do not pull or bend the probe cable, and do not put caster feet on the probe cable. Do not immerse the probe cable in chemical solutions or water. Failure to follow these instructions may cause cable discontinuity, short circuit, skin burn on the patient and incorrect measurement data. Replace any broken probe with a new one. CAUTION When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one. CAUTION While a patient is on medication which causes vasodilation, the pulse waveform may change and in rare cases the SpO2 value might not be displayed. NOTE ZM-930PA only When monitoring SpO2, monitor ECG at the same time. The ECG electrode lead works as an antenna for transmitting data from the transmitter to the receiving monitor. If ECG is not measured, the telemetry signal may not be received. Operator’s Manual ZM-920PA/921PA/930PA/931PA 39 Measurement Procedure 1. Select the SpO2 probe. 2. Connect the SpO2 probe to the SpO2 socket. 3. Attach the SpO2 probe to the patient. After steps 1 to 3 are finished, SpO2 monitoring automatically starts. Selecting SpO2 Probe Select an appropriate probe for the patient. CAUTION Only use Nihon Kohden specified electrodes, electrode leads and SpO2 probes. Otherwise, the maximum performance from the transmitter cannot be guaranteed. CAUTION Do not use a damaged or disassembled probe. It causes incorrect measurement and may injure the patient. 40 Operator’s Manual ZM-920PA/921PA/930PA/931PA Reusable Probes When using a TL-201T finger probe, choose the appropriate cable length for attachment. Model Subject (Weight) Finger Probe TL-201T Cable length: 60 cm or 160 cm Multi-site probe TL-220T Attachment tape Finger probe TL-630T1/630T3/631T1/631T3 Attachment tape Attachment Site Adult or child (20 kg or more) Finger Adult or Infant (3 kg or more) Finger or toe Neonate (3 kg or less) Instep and sole TL-630T1/630T3: Adult or child (50 kg or more) Finger or toe TL-631T1/631T3: Adult or child (20 kg or more) Cable length: TL-630T1/631T1: 60 cm TL-630T3/631T3: 160 cm Operator’s Manual ZM-920PA/921PA/930PA/931PA 41 Disposable Probes CAUTION The disposable probe is not sterilized. Use the disposable probe only for a single patient. Never reuse the disposable probe for another patient because it causes cross infection. Model Subject (Weight) Attachment Site Adult Finger or toe (30 kg or more) TL-251T TL-252T Child (3 to 40 kg) Finger or toe TL-253T Neonate (3 kg or less) Instep and sole TL-051S/052S Adult (50 kg or more) Finger Neonate (3 kg or less) Instep and sole Cable length TL-051S: 80 cm TL-052S: 160 cm 42 Operator’s Manual ZM-920PA/921PA/930PA/931PA Model TL-061S/062S Subject (Weight) Attachment Site Adult, child Finger (15 to 50 kg) Child, infant (3 to 15 kg) Cable length TL-061S: 80 cm TL-062S: 160 cm TL-271T/271T3 Cable length TL-271T: 80 cm TL-271T3: 160 cm TL-272T/272T3 Cable length TL-272T: 80 cm TL-272T3: 160 cm TL-273T/273T3 Cable length TL-273T: 80 cm TL-273T3: 160 cm Toe Adult (30 kg or more) Child (10 to 50 kg) Finger or toe Neonate (3 kg or less) Instep and sole Adult (40 kg or more) Finger or toe Operator’s Manual ZM-920PA/921PA/930PA/931PA 43 Model TL-274T/274T3 Subject (Weight) Attachment Site Infant Finger or toe (3 to 20 kg) Cable length TL-274T: 80 m TL-274T3: 160 cm Connecting SpO2 Probe to the Transmitter Connect the probe to the SpO2 socket on the transmitter. CAUTION Do not shake or swing the transmitter while holding the leads or cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments. CAUTION Hold the connector when connecting/disconnecting the SpO2 probe. If you disconnect the SpO2 probe by pulling the cable, it damages the cable. 44 Operator’s Manual ZM-920PA/921PA/930PA/931PA Attaching the Probe to the Patient For details, refer to the operator’s manual of each probe. WARNING • When using the TL-201T finger probe, do not fasten the probe and cable to the finger by wrapping with tape. This may cause burn, congestion or pressure necrosis from poor blood circulation. • When using probes other than the TL-201T finger probe, to avoid poor circulation, do not wrap the tape too tight. Check the blood circulation condition by observing the skin color and congestion at the skin peripheral to the probe attachment site. Even for shortterm monitoring, there may be burn or pressure necrosis from poor blood circulation, especially on neonates or low birth weight infants whose skin is delicate. Accurate measurement cannot be performed on a site with poor peripheral circulation. WARNING Check the circulation condition by observing the skin color at the measurement site and pulse waveform. Change the measurement site every 8 hours for disposable probes and every 4 hours for reusable probes (every 8 hours for TL-630T/TL-631T series probe). The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree. • Patient with a fever • Patient with peripheral circulation insufficiency • Neonate or low birth weight infant with delicate skin • Patient who is receiving photodynamic therapy* * Photodynamic therapy is a treatment to remove the affected tissue by using a photosensitizing agent and exposing the tissue to light. This treatment has a side effect of photosensitivity and the light from the finger probe sensor may cause a burn. This probe uses two light wavelengths in the range from 650 to 950 nm. The maximum light intensity is less than 5.5 mW/sr. Operator’s Manual ZM-920PA/921PA/930PA/931PA 45 CAUTION If the attachment site is dirty with blood or bodily fluids, clean the attachment site before attaching the probe. If there is nail polish on the attachment site, remove the polish. Otherwise, the amount of transmitted light decreases, and measured value may be incorrect or measurement cannot be performed. CAUTION If the skin gets irritated or redness appears on the skin from the probe, change the attachment site or stop using the probe. Take extreme care for the patients with delicate skin. CAUTION When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one. CAUTION Do not use a probe which is deteriorated by aging. Accurate measurement cannot be performed. CAUTION Neonatal skin is delicate. Remove the probe and tape carefully and slowly. CAUTION When removing a probe that is taped to the skin, do not pull the probe cable because this can damage the cable. 46 Operator’s Manual ZM-920PA/921PA/930PA/931PA CAUTION When removing the probe from the attachment tape, do not pull the sensor cable because this can damage the cable. CAUTION Do not immerse the disposable probe in detergents or water. If the probe adhesive surface gets wet, adhesiveness becomes weak and the probe cannot be attached to the skin. CAUTION Refer to the probe instruction manual for details. Starting Measurement When monitoring starts, SpO2 and pulse waveform are sent to the monitors and SpO2 data and pulse level bar graph are displayed on the transmitter LCD. You can turn off the display of SpO2 data and pulse level bar graph on the LCD. Refer to the “Turning SpO2 Data and Pulse Level Bar Graph Display On/Off” section. CAUTION When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one. Operator’s Manual ZM-920PA/921PA/930PA/931PA 47 Turning SpO2 Data and Pulse Level Bar Graph Display On/Off You can turn off the display of SpO2 data and pulse level bar graph on the LCD. SpO2 display off Press the CALL key for more than 3 seconds within 10 seconds after turning transmitter power on (after a “peep” sound). RA LA LL Va Vb USAch E002 1395.025 MHz CALL key “%SpO2” is not displayed. When SpO2 monitoring starts, SpO2 data and pulse level bar graph are not displayed on the LCD. To turn SpO2 display on, turn the transmitter power off and turn the power on again. 48 Operator’s Manual ZM-920PA/921PA/930PA/931PA Detecting and Displaying Measurement Condition External Light Noise Alarm CAUTION Normal external light does not affect monitoring but strong light such as a surgical light or sunlight may affect monitoring. If affected, cover the measuring site with a blanket. Strong external light (surgical light or inverter type fluorescent lamp, etc.), may affect SpO2 monitoring. When external light is too strong to correctly measure SpO2, the transmitter generates an alarm tone (“pip” sound every 0.5 seconds). Cover the probe attachment site with blanket or cloth. Insufficient Light Alarm When sufficient light cannot be obtained from the photo emitter of the probe, the transmitter generates the following sound and indication. (4) (1) “peep” sound (every 1 second) (2) pulse level bar graph is maximum (all bars are lit.) (3) SpO2 data is displayed as “ ” (4) %SpO2 is blinking. (2) (3) In this case, change the attachment site to the appropriate site. Refer to the operator’s manual of the SpO2 probe. Probe Malfunction Alarm CAUTION When any of the following conditions occurs, the probe may be broken. Replace it with a new one. • The transmitter generates “pip” sounds. • SpO2 data is 85% and blinking. Operator’s Manual ZM-920PA/921PA/930PA/931PA 49 When Measurement Condition is Unstable SpO2 data blinks every 1 second when SpO2 signal stability decreases and the transmitter cannot detect correct pulse waveform because of patient body movement, poor attachment condition or poor circulation condition at the probe attachment site. CAUTION SpO2 data blinking every second indicates an unstable pulse waveform and displayed SpO2 value may be inaccurate. The displayed data might not reflect sudden SpO2 changes. 50 Operator’s Manual ZM-920PA/921PA/930PA/931PA Alarm List (1) (5) %SpO2 indication Displayed as LED on ZM-920PA/921PA (4) pulse level bar graph (2) RA LA LL Va Vb Sound Single “peep” sound for 4 s --- Continuous “peep” sound --- Intermittent “pip” sound every 0.5 s Display --- (1) All lights are off (2) --- (3) SpO2 data Cause Countermeasure The CALL key is pressed. The sound lasts while the key is pressed. The battery voltage decreases and battery charge is almost zero. Battery is completely discharged. Electrode lead is disconnected from the electrode. Electrode lead is disconnected from the transmitter. Electrode lead discontinuity Electrode is not firmly attached to the skin. Polarization voltage is abnormally high. SpO2 measurement site is under fluorescent light, surgical light, sunlight, etc. Operator’s Manual ZM-920PA/921PA/930PA/931PA Replace the batteries with new ones. Replace the batteries with new ones. To stop the sound, turn off the power. Firmly connect the electrode lead to the electrode. Firmly connect the electrode lead to the transmitter. Replace the electrode lead with a new one. Replace the electrode with a new one. Cover the measurement site with a blanket or cloth. 51 Sound Intermittent “peep” sound every 1 s Intermittent “pip” sound every 0.25 s --- Display (3) (4) all lit (5) blinking Cause Cannot receive sufficient light from the probe photo emitter. (3) 85%, blinking Broken probe (3) blinking Patient body movement Probe is not attached securely. 52 Countermeasure Attach probe to a site with 6 to 14 mm thickness where sufficient light can be received. Replace the probe with a new one. Remove the cause by checking the patient condition and changing the attachment site. Securely attach the probe. Operator’s Manual ZM-920PA/921PA/930PA/931PA Troubleshooting If the problem still remains after checking the following, contact your Nihon Kohden representative. Problem The power cannot be turned on. Cause Countermeasure Batteries are not Install the batteries correctly. installed correctly. The battery polarity is wrong. Batteries are completely discharged. Replace the batteries with new ones. Nothing is displayed on the LCD after turning the power on. (ZM-930PA/ 931PA only) SpO2 display is turned Turn off the power, and turn on the off. power again. Nothing is displayed on the monitor after turning the transmitter power on. The channel of the transmitter and monitor does not match. Signal receiving condition Electrode lead is is poor. not connected to the transmitter. Set the correct channel on the monitor. Connect the electrode lead to the transmitter. ECG electrode lead works as an antenna for transmitting data to the receiving monitor. If ECG is not measured, the signal may not be received. Another transmitter Turn the transmitter power off. If the of the same channel is monitor still receives a signal, there used near by. is a high probability that another transmitter of the same channel is used nearby. Follow the instruction of your channel administrator and use another transmitter of a different channel. Signals are mixing. Follow the instruction of your channel administrator and use another transmitter of a different channel. The transmitter is damaged. Contact your Nihon Kohden representative. Operator’s Manual ZM-920PA/921PA/930PA/931PA 53 Problem Cause 3 electrode leads are The transmitter properly attached to the is fixed to 6 lead patient but LEDs light. monitoring. Countermeasure Turn off and on the transmitter power. Heart������������� rate of the patient who is using an electric blanket cannot be monitored on the receiving monitor. Pacing pulse detection Turn off the pacing pulse detection. is set to ON on the monitor. ECG baseline is thick. (Hum is overlapping) The gel on the electrode is dried out. Replace the electrode with a new one. The gel on the electrode is coming off. Respiration waveform measurement is unstable. Electric blanket is used. Cover the blanket with a shield cover. Hum filter is set to OFF on the monitor Set the filter to ON. The gel on the electrode is dried out. Replace the electrode with a new one. The gel on the electrode is coming off. SpO2 data is unstable and not reliable. 54 The probe size is not appropriate for the patient. Use the appropriate probe for the patient. Probe is attached to the same limb that is used for NIBP measurement. Attach the probe to the opposite limb. Avoid a site where blood circulation condition changes greatly. Probe attachment condition is poor. Probe is about to detach from the skin. External light gets in. Firmly attach the probe according to the procedure in the probe operator��� ’��s manual. Measurement site is dirty. Patient is wearing nail polish. Remove dirt and nail polish. Operator’s Manual ZM-920PA/921PA/930PA/931PA Changing the Transmitter Channel The transmitter channel can be changed by the QI-901PK Channel Writer. To change the channel number, refer to the channel writer manual. WARNING The following actions must be taken to properly receive the transmitter signal of the correct patient on the receiving instrument. Otherwise, there may be signal loss or signals may mix causing a serious accident, such as monitoring a different patient. • Assign a channel administrator in the hospital and only he or she should manage channel assignment. • The channel administrator must manage the channels in the facility so that there is no signal interference. • When the transmitter channel is changed, the channel administrator must check that the channel on the receiving monitor is also changed and the signal is properly received. • The channel administrator must replace the channel number label on the transmitter with the new one after changing the channel. Operator’s Manual ZM-920PA/921PA/930PA/931PA 55 Lifetime and Disposal Disposing of Used Batteries Replacement When the “ ” LED is lit or the “ ” is displayed on the LCD during operation, the batteries are running out. Replace the batteries with new ones. When using rechargeable batteries, recharge them. Disposal Before disposing of the batteries, check with your local solid waste officials for details in your area for proper disposal. It may be illegal to dispose of these batteries in the municipal waste stream. Disposing of Electrodes and SpO2 Probes Refer to the manual of each item. 56 Operator’s Manual ZM-920PA/921PA/930PA/931PA Cleaning, Disinfection and Sterilization Transmitter and Electrode Lead CAUTION If detergent or liquid spills into the transmitter, clean it and dry it completely before use. If a wet transmitter is used, the transmitter may malfunction or get damaged. CAUTION Before cleaning or disinfection, remove the batteries from the transmitter. Failure to follow this instruction may result in electrical shock or transmitter malfunction. CAUTION The transmitter cannot be sterilized. Sterilizing the transmitter may damage it. CAUTION Dispose of the transmitter, options and accessories as specified by Nihon Kohden. Otherwise, it causes infection or environmental contamination. Before cleaning or disinfecting, remove the batteries from the transmitter. Be careful not to let any liquid get inside the transmitter. Cleaning Wipe the transmitter and electrode leads with a soft cloth moistened with disinfecting alcohol or neutral detergent diluted with water. After cleaning, dry them completely. Operator’s Manual ZM-920PA/921PA/930PA/931PA 57 Disinfection CAUTION • Do not immerse the electrode lead connector in liquid. • Do not disinfect with hypochlorous acid. • Use the recommended concentration. Wipe the outside surface of the transmitter and electrode lead with a non-abrasive cloth moistened with any of the disinfectants listed on the next page. Use the recommended concentration. Disinfectant Glutaraldehyde solution Alkyldiaminoethylglycine hydrochloride Benzalkonium chloride Benzethonium chloride solution Chlorhexidine gluconate solution Concentration (%) 2.0 0.5 0.2 0.2 0.5 SpO2 Probe Refer to the probe manual. 58 Operator’s Manual ZM-920PA/921PA/930PA/931PA Replacing the Battery Case Cover When the battery case cover is damaged, replace it with a new one. Battery case cover, code no. 6113-046365C. Other than the battery case cover, there are no serviceable parts for the transmitter. Repair Parts Availability Policy Nihon Kohden Corporation (NKC) shall stock repair parts (parts necessary to maintain the performance of the instrument) for a period of 6 years after NKC announces discontinuation of the instrument. In that period, NKC or its distributors will repair the instrument. This period may be shorter than 6 years if the necessary board or part is not available. For discontinuation announcements, contact your Nihon Kohden distributor or representative. Operator’s Manual ZM-920PA/921PA/930PA/931PA 59 Specifications ZM-920PA/930PA ECG measurement Channels: Input range: DC offset: Input impedance: Pacing pulse detection: ±5 mV or more ±500 mV or more 5 MΩ or more (5 Hz) ANSI/AAMI EC13 Based upon Pacemaker pulse rejection Capability Respiration measurement Measuring method: Impedance range: Impedance method 0 to 2 kΩ SpO2 measurement Measuring range: 0 to 100%, in 1% steps Measuring accuracy When the measuring accuracy of the SpO2 probe is not considered: ±1 digit (80% ≤ SpO2 ≤ 100%) ±2 digit (50% ≤ SpO2 ≤ 80%) Less than 50% is not specified When the measuring accuracy of the SpO2 probe is considered: ±2 digit (80% ≤ SpO2 ≤ 100%) ±3 digit (70% ≤ SpO2 ≤ 80%) Less than 70% is not specified Transmitter FCC regulation: Field strength limits: Undesired emission: above 960 MHz: FCC part 95 Subpart-H Wireless Medical Telemetry Service (WMTS) <200 mV/m (at 3 m) below 960 MHz: 200 μV/m (at 3 m) 500 μV/m (at 3 m) 60 Operator’s Manual ZM-920PA/921PA/930PA/931PA Antenna: Transmission channel: Transmission frequency range: Channel spacing: Modulation scheme: Occupied bandwidth: Effective radiated power: ECG electrode lead Indicated on the transmitter 608.0250 to 613.9750 MHz 50 kHz or 37.5 kHz (12.5 kHz when interleaved) FSK (Frequency Shift Keying) <20 kHz 1.0 mW (conducted) Safety standards Safety standard: CSA C22.2 No. 601-1 M90: 1994 IEC 60601-1: 1988 IEC 60601-1 Amendment 1: 1991 IEC 60601-1 Amendment 2: 1995 IEC 60601-1-2: 1993 IEC 60601-2-27: 1994 According to the type of protection against electrical shock: INTERNALLY POWERED EQUIPMENT According to the degree of protection against electrical shock: ECG/impedance method respiration: DEFIBRILLATION-PROOF TYPE CF APPLIED PART SpO2: DEFIBRILLATION-PROOF TYPE BF APPLIED PART According to the degree of protection against harmful ingress of water: IPX0 (Ordinary equipment) According to the degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE: Equipment not suitable for use in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE According to the mode of operation: CONTINUOUS OPERATION Water resistance Water does not get inside the transmitter except for the battery case when immersed in water up to 30 cm deep for 3 minutes. Operator’s Manual ZM-920PA/921PA/930PA/931PA 61 Power requirements Battery type: Battery lifetime: ZM-930PA: ZM-920PA: two AA type alkaline dry cell batteries, manganese dry cell batteries, NiCd rechargeable batteries, NiMH batteries approximately 3 days (with alkaline batteries, measuring ECG, respiration and SpO2 of approximately 60 kg weight adult male patient at the pointing finger) approximately 4 days (with alkaline batteries, measuring ECG and respiration only) approximately 4 days Environment Operating environment Operating temperature: Operating humidity: Operating atmospheric pressure: Operating voltage: 5 to 40°C, 41 to 104°F 30 to 85% (non-condensing) 70 to 106 kPa 1.6 to 3.2 V Storage environment Storage temperature: Storage humidity: Storage atmospheric pressure: −20 to 65°C, −4 to 149°F 15 to 95% 70 to 106 kPa Dimension and Weight Dimension: Weight (without batteries): 62 78 W × 122 H × 26 D (mm) ZM-930PA: about 180 g ZM-920PA: about 165 g Operator’s Manual ZM-920PA/921PA/930PA/931PA ZM-921PA/931PA ECG Measurement Channels: Input range: DC offset: Input impedance: Pacing pulse detection: ±5 mV or more ±500 mV or more 5 MΩ or more (5 Hz) ANSI/AAMI EC13 Based upon Pacemaker pulse rejection Capability Respiration Measurement Measuring method: Impedance range: Impedance method 0 to 2 kΩ SpO2 Measurement Measuring range: 0 to 100%, in 1% steps Measuring accuracy When the measuring accuracy of the SpO2 probe is not considered: ±1 digit (80% ≤ SpO2 ≤ 100%) ±2 digit (50% ≤ SpO2 ≤ 80%) Less than 50% is not specified When the measuring accuracy of the SpO2 probe is considered: ±2 digit (80% ≤ SpO2 ≤ 100%) ±3 digit (70% ≤ SpO2 ≤ 80%) Less than 70% is not specified Transmitter FCC regulation: FCC part 95 Subpart-H Wireless Medical Telemetry Service (WMTS) Field strength limits: <740 mV/m (at 3 m) Undesired emission: below 960 MHz: 200 μV/m (at 3 m) above 960 MHz: 500 μV/m (at 3 m) Antenna: Internal Transmission channel: indicated on the transmitter Transmission frequency range: 1395.0250 to 1399.9750 MHz 1427.0250 to 1431.9750 MHz Operator’s Manual ZM-920PA/921PA/930PA/931PA 63 Channel spacing: Modulation scheme: Occupied bandwidth: Effective radiated power: 50 kHz or 37.5 kHz (12.5 kHz when interleaved) FSK (Frequency Shift Keying) <20 kHz 5.0 mW (factory default setting) Can be changed to 1.0 mW if required Safety Standards Safety standard: CAN/CSA-C22.2 No. 601-1 M90: 1990 CAN/CSA-C22.2 No. 601-1. 1S1-94: 1994 CAN/CSA-C22.2 No. 601-1. 1B-90: R2002 CAN/CSA-C22.2 No. 60601-2-49-04: 2004 CAN/CSA-C22.2 No. 601.2.27-98: 1998 IEC 60601-1: 1988 IEC 60601-1 Amendment 1: 1991 IEC 60601-1 Amendment 2: 1995 IEC 60601-1-2: 2001 IEC 60601-1-2 Amendment 1: 2004 IEC 60601-2-27: 2005 IEC 60601-2-49: 2001 ISO 9919: 2005 According to the type of protection against electrical shock: INTERNALLY POWERED EQUIPMENT According to the degree of protection against electrical shock: ECG/impedance method respiration: DEFIBRILLATION-PROOF TYPE CF APPLIED PART SpO2: DEFIBRILLATION-PROOF TYPE BF APPLIED PART According to the degree of protection against harmful ingress of water: IPX0 (Ordinary equipment) According to the degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE: Equipment not suitable for use in the presence of FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE According to the mode of operation: CONTINUOUS OPERATION 64 Operator’s Manual ZM-920PA/921PA/930PA/931PA Water Resistance Water does not get inside the transmitter except for the battery case when immersed in water up to 30 cm deep for 3 minutes. Power Requirements Battery type: Battery lifetime: ZM-931PA: ZM-921PA: two AA type alkaline dry cell batteries, manganese dry cell batteries, NiCd rechargeable batteries, NiMH batteries approximately 2.5 days (with alkaline batteries, measuring ECG, respiration and SpO2 of approximately 60 kg weight adult male patient at the pointing finger) approximately 3.5 days (with alkaline batteries, measuring ECG and respiration only) approximately 3.5 days Environment Operating environment Operating temperature: Operating humidity: Operating atmospheric pressure: Operating voltage: 5 to 40°C, 41 to 104°F 30 to 85% (non-condensing) 70 to 106 kPa 1.6 to 3.2 V Storage environment Storage temperature: Storage humidity: Storage atmospheric pressure: −20 to 65°C, −4 to 149°F 15 to 95% 70 to 106 kPa Dimension and Weight Dimension: Weight (without batteries): 78 W × 122 H × 26 D (mm) ZM-931PA: about 190 g ZM-921PA: about 170 g Operator’s Manual ZM-920PA/921PA/930PA/931PA 65 Electromagnetic Compatibility IEC 60601-1-2: 2001 IEC 60601-1-2 Amendment 1: 2004 Electromagnetic Emissions This Model ZM-921PA/931PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-921PA/931PA should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 1�1 G������ r����� oup 1 The ZM-921PA/931PA��������������������� ����������������������������������� uses RF energy only for its�������������������������������������� internal ������������������������������������� function.������������������� Therefore, ������������������ its RF emissions are very��������������������������� low����������������������� �������������������������� and are not likely to ���������������������� cause any i��������������������������������� n�������������������������������� ter����������������������������� f���������������������������� erence i�������������������� n������������������� nearby elect������ r����� onic equipment�. RF emissions CISPR �� 1� C������ lass B Th������������������������������������� e������������������������������������ ZM-921PA/931PA��������������������� ����������������������������������� is suitab����������� l���������� e for use in a�������������������������������������� l������������������������������������� l������������������������������������ establishmen����������������������� ����������������������������������� t���������������������� s,�������������������� including domestic ������������������� establishments. Harmonic emissions IEC 61������� 00����� 0-3-�2 Not applicable Voltage fluctuations/ flicker emissions IEC 61000-3-3 Not applicable Electromagnetic Immunity This Model ZM-921PA/931PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-921PA/931PA should assure that it is used in such an environment. Immunity test IEC 60601 test level Electrostatic ±6 kV contact discharge (ESD) ±8 kV air IEC 61000-4-2 66 Electromagnetic environment guidance ±6 kV contact Floors should be wood, concrete or ceramic tile. ±8 kV air If floors are covered with synthetic material, the relative humidity should be at least 30%. Compliance level Operator’s Manual ZM-920PA/921PA/930PA/931PA Immunity test IEC 60601 test level Compliance level Electrical fast transient/ burst IEC 61000-4-4 ±2 kV for power supply lines ±1 kV for input/ output lines Surge IEC 61000-4-5 ±1 kV differential Not applicable mode ±2 kV common mode <5% UT (>95% dip Not applicable in UT) for 0.5 cycle Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 Electromagnetic environment guidance Not applicable — — 40% UT (60% dip in UT) for 5 cycles — 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5 s 3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. NOTE: UT is the AC mains voltage prior to application of the test level Avoiding Electromagnetic Interference (Impedance Respiration) Impedance respiration measurement is very sensitive and affected by electromagnetic interference. Technological limitations do not allow immunity levels higher than 1 V/m for radiated RF electromagnetic fields. Electromagnetic fields with field strengths above 1 V/m may cause measurement error. Do not use electrically radiating equipment near the impedance respiration measurements. Operator’s Manual ZM-920PA/921PA/930PA/931PA 67 Immunity test IEC 60601 Compliance Electromagnetic environment test level level - guidance Portable and mobile RF communications equipment should be used no closer to any part of the ZM921PA/931PA, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF 3 Vrms IEC 61000-4-6 3 Vrms d = 1.2√P Radiated RF 3 V/m IEC 61000-4-3 80 MHz to 2.5 GHz 3 V/m 80 MHz to 2.5 GHz d = 1.2√P 80 MHz to 800 MHz d = 2.3√P 800 MHz to 2.5 GHz (1 V/m 80 MHz (d = 3.5√P 80 MHz to 800 MHz for to 2.5 GHz for respiration respiration) d = 7.0√P 800 MHz to 2.5 GHz for respiration) where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as deter mined by an electromagnetic site survey*1, should be less than the compliance level in each frequency range*2. Interference may occur in the vicinity of equipment marked with the following symbol: 68 Operator’s Manual ZM-920PA/921PA/930PA/931PA Immunity test IEC 60601 Compliance Electromagnetic environment test level level - guidance NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. *1 Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ZM-921PA/931PA is used exceeds the applicable RF compliance level above, the ZM-921PA/931PA should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ZM-921PA/931PA. *2 Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m for respiration and 3 V/m for all other functions. Operator’s Manual ZM-920PA/921PA/930PA/931PA 69 Recommended Separation Distances between Portable and Mobile RF Communications Equipment The ZM-921PA/931PA is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ZM-921PA/931PA can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ZM-921PA/931PA as recommended below, according to the maximum output power of the communications. Rated maximum output power of transmitter (W) Separation distance according to frequency of transmitter (m) 150 kHz to 80 MHz d = 1.2√P 80 MHz to 800 MHz 800 MHz to 2.5 GHz d = 1.2√P d = 2.3√P (For respiration: d = 3.5√P) (For respiration: d = 7.0√P) 0.01 0.12 0.12 (0.35*) 0.23 (0.7*) 0.1 0.38 0.38 (1.1*) 0.73 (2.2*) 1.2 1.2 (3.5*) 2.3 (7.0*) 10 3.8 3.8 (11*) 7.3 (22*) 100 12 12 (35*) 23 (70*) For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. (* For respiration) NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Recovery Time after Defibrillation The transmitter returns to the normal operating mode within 10 seconds after defibrillation. The stored settings are not affected. 70 Operator’s Manual ZM-920PA/921PA/930PA/931PA System Composition for EMC Test The ZM-921PA/931PA bedside monitor is tested to comply with IEC 60601-1-2: 2001 and IEC 60601-1-2 Amendment 1: 2004 with the following composition. Units Cable length ZM-921PA/931PA transmitter — BR-906P ECG electrode lead 0.8 m TL-201T finger probe 1.6 m Operator’s Manual ZM-920PA/921PA/930PA/931PA 71 Standard Accessories Name Strap 72 Q’ty Supply code Y233 Operator’s Manual ZM-920PA/921PA/930PA/931PA Options CAUTION Use only Nihon Kohden electrodes, electrode leads and SpO2 probes to assure maximum performance from your instrument. Transmitter Channel writer, QI-901PK ECG/RESP Name Electrode lead Application 3 electrodes, clip type, lead length 80 cm 3 electrodes, snap type, lead length 80 cm 6 electrodes, clip type, lead length 80 cm 6 electrodes, snap type, lead length 80 cm Model Q’ty Supply code BR-903PA K911A BR-913PA K910B BR-906PA K912A BR-916PA K915A Operator’s Manual ZM-920PA/921PA/930PA/931PA 73 SpO2 Cable length 0.6 m 1.6 m 1.6 m 0.6 m 1.6 m 0.6 m 1.6 m Name Finger probe (reusable) Multi-site probe (reusable) Finger probe (reusable) SpO2 probe (for adult, disposable) SpO2 probe (for child, disposable) SpO2 probe (for neonate, disposable) SpO2 probe (for adult, disposable) SpO2 probe (for child, disposable) SpO2 probe (for neonate/adult, disposable) SpO2 probe (for child/infant, disposable) SpO2 probe (for adult/neonate, disposable) SpO2 probe (for child/infant, disposable) COTTONY tape Foam tape for TL-051S/052S/ 061S/062S Attachment tape for TL-220T/ 251T/252T/253T/630T/631T Probe fastener 74 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m 0.8 m 1.6 m — Model/ Code No. Q’ty TL-201T TL-220T TL-630T1 TL-630T3 TL-631T1 TL-631T3 TL-251T TL-252T Supply Code No. P225H P225F P225G P310A P310C P311A P311C P201A P201B TL-253T P201C TL-271T TL-271T3 TL-272T TL-272T3 TL-273T TL-273T3 TL-274T TL-274T3 TL-051S TL-052S TL-061S TL-062S 340703 P203A P203E P203B P203F P203C P203G P203D P203H P228A P228B P229A P229B P259 — 24 20 4 × 25 packages 3 × 30 packages YS-093P2 30 P260 P263 P267 Operator’s Manual ZM-920PA/921PA/930PA/931PA Transmission Frequencies Channel: 9002 to 9478 Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. 608.0250 608.0375 608.0500 608.0625 608.0750 608.0875 9002 9003 9004 9005 9006 9007 9008 608.3500 608.3625 608.3750 608.3875 608.4000 608.4125 608.4250 9028 9029 9030 9031 9032 9033 9034 608.6750 608.6875 608.7000 608.7125 608.7250 608.7375 9054 9055 9056 9057 9058 9059 608.1000 608.1125 608.1250 608.1375 608.1500 608.1625 608.1750 9009 9010 9011 9012 9013 9014 608.7500 9060 608.4375 608.4500 608.4625 608.4750 608.4875 608.5000 9035 9036 9037 9038 9039 9040 9015 608.5125 9041 608.7625 608.7750 608.7875 608.8000 608.8125 608.8250 608.8375 9061 9062 9063 9064 9065 9066 9067 608.1875 608.2000 608.2125 608.2250 608.2375 608.2500 608.2625 608.2750 608.2875 608.3000 608.3125 608.3250 608.3375 9016 9017 9018 9019 9020 9021 9022 9023 9024 9025 9026 9027 608.5250 608.5375 608.5500 608.5625 608.5750 608.5875 608.6000 608.6125 608.6250 608.6375 608.6500 608.6625 9042 9043 9044 9045 9046 9047 9048 9049 9050 9051 9052 9053 608.8500 608.8625 608.8750 608.8875 608.9000 608.9125 608.9250 608.9375 608.9500 608.9625 608.9750 608.9875 9068 9069 9070 9071 9072 9073 9074 9075 9076 9077 9078 9079 Operator’s Manual ZM-920PA/921PA/930PA/931PA 75 Transmission frequency (MHz) Channel No. 609.0000 609.0125 609.0250 609.0375 609.0500 609.0625 609.0750 609.0875 609.1000 609.1125 609.1250 609.1375 609.1500 609.1625 609.1750 609.1875 609.2000 609.2125 609.2250 609.2375 609.2500 76 Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. 9080 9081 9082 9083 9084 9085 9086 9087 9088 9089 9090 9091 9092 9093 9094 9095 9096 9097 9098 9099 9100 609.3750 609.3875 609.4000 609.4125 609.4250 609.4375 609.4500 609.4625 609.4750 609.4875 609.5000 609.5125 609.5250 609.5375 609.5500 609.5625 609.5750 609.5875 609.6000 609.6125 9110 9111 9112 9113 9114 9115 9116 9117 9118 9119 9120 9121 9122 9123 9124 9125 9126 9127 9128 9129 609.7500 609.7625 609.7750 609.7875 609.8000 609.8125 609.8250 609.8375 609.8500 609.8625 609.8750 609.8875 609.9000 609.9125 609.9250 609.9375 609.9500 609.9625 609.9750 609.9875 9140 9141 9142 9143 9144 9145 9146 9147 9148 9149 9150 9151 9152 9153 9154 9155 9156 9157 9158 9159 609.6250 9130 610.0000 9160 609.2625 609.2750 609.2875 609.3000 609.3125 9101 9102 9103 9104 9105 9161 9162 9163 9164 9165 9106 9131 9132 9133 9134 9135 9136 610.0125 610.0250 610.0375 610.0500 610.0625 609.3250 609.6375 609.6500 609.6625 609.6750 609.6875 609.7000 610.0750 9166 609.3375 609.3500 609.3625 9107 9108 9109 609.7125 609.7250 609.7375 9137 9138 9139 610.0875 610.1000 610.1125 9167 9168 9169 Operator’s Manual ZM-920PA/921PA/930PA/931PA Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. 610.1250 610.1375 610.1500 9170 9171 9172 610.5000 610.5125 9200 9201 610.8750 610.8875 9230 9231 610.5250 9202 610.9000 9232 610.1625 610.1750 610.1875 610.2000 610.2125 9173 9174 9175 9176 9177 9233 9234 9235 9236 9237 9178 9203 9204 9205 9206 9207 9208 610.9125 610.9250 610.9375 610.9500 610.9625 610.2250 610.5375 610.5500 610.5625 610.5750 610.5875 610.6000 610.9750 9238 610.2375 610.2500 610.2625 610.2750 610.2875 9179 9180 9181 9182 9183 610.6125 610.6250 610.6375 610.6500 610.6625 9209 9210 9211 9212 9213 610.3000 9184 610.6750 9214 610.9875 611.0000 611.0125 611.0250 611.0375 611.0500 9239 9240 9241 9242 9243 9244 610.3125 610.3250 610.3375 610.3500 610.3625 610.3750 610.3875 610.4000 610.4125 610.4250 610.4375 610.4500 610.4625 610.4750 610.4875 9185 9186 9187 9188 9189 9190 9191 9192 9193 9194 9195 9196 9197 9198 9199 610.6875 610.7000 610.7125 610.7250 610.7375 610.7500 610.7625 610.7750 610.7875 610.8000 610.8125 610.8250 610.8375 610.8500 610.8625 9215 9216 9217 9218 9219 9220 9221 9222 9223 9224 9225 9226 9227 9228 9229 611.0625 611.0750 611.0875 611.1000 611.1125 611.1250 611.1375 611.1500 611.1625 611.1750 611.1875 611.2000 611.2125 611.2250 611.2375 9245 9246 9247 9248 9249 9250 9251 9252 9253 9254 9255 9256 9257 9258 9259 Operator’s Manual ZM-920PA/921PA/930PA/931PA 77 Transmission frequency (MHz) Channel No. 611.2500 611.2625 611.2750 611.2875 611.3000 611.3125 611.3250 611.3375 611.3500 611.3625 611.3750 611.3875 611.4000 611.4125 611.4250 611.4375 611.4500 611.4625 611.4750 611.4875 611.5000 78 Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. 9260 9261 9262 9263 9264 9265 9266 9267 9268 9269 9270 9271 9272 9273 9274 9275 9276 9277 9278 9279 9280 611.6250 611.6375 611.6500 611.6625 611.6750 611.6875 611.7000 611.7125 611.7250 611.7375 611.7500 611.7625 611.7750 611.7875 611.8000 611.8125 611.8250 611.8375 611.8500 611.8625 9290 9291 9292 9293 9294 9295 9296 9297 9298 9299 9300 9301 9302 9303 9304 9305 9306 9307 9308 9309 612.0000 612.0125 612.0250 612.0375 612.0500 612.0625 612.0750 612.0875 612.1000 612.1125 612.1250 612.1375 612.1500 612.1625 612.1750 612.1875 612.2000 612.2125 612.2250 612.2375 9320 9321 9322 9323 9324 9325 9326 9327 9328 9329 9330 9331 9332 9333 9334 9335 9336 9337 9338 9339 611.8750 9310 612.2500 9340 611.5125 611.5250 611.5375 611.5500 611.5625 9281 9282 9283 9284 9285 9341 9342 9343 9344 9345 9286 9311 9312 9313 9314 9315 9316 612.2625 612.2750 612.2875 612.3000 612.3125 611.5750 611.8875 611.9000 611.9125 611.9250 611.9375 611.9500 612.3250 9346 611.5875 611.6000 611.6125 9287 9288 9289 611.9625 611.9750 611.9875 9317 9318 9319 612.3375 612.3500 612.3625 9347 9348 9349 Operator’s Manual ZM-920PA/921PA/930PA/931PA Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. 612.3750 612.3875 612.4000 9350 9351 9352 612.7500 612.7625 9380 9381 613.1250 613.1375 9410 9411 612.7750 9382 613.1500 9412 612.4125 612.4250 612.4375 612.4500 612.4625 9353 9354 9355 9356 9357 9413 9414 9415 9416 9417 9358 9383 9384 9385 9386 9387 9388 613.1625 613.1750 613.1875 613.2000 613.2125 612.4750 612.7875 612.8000 612.8125 612.8250 612.8375 612.8500 613.2250 9418 612.4875 612.5000 612.5125 612.5250 612.5375 9359 9360 9361 9362 9363 612.8625 612.8750 612.8875 612.9000 612.9125 9389 9390 9391 9392 9393 612.5500 9364 612.9250 9394 613.2375 613.2500 613.2625 613.2750 613.2875 613.3000 9419 9420 9421 9422 9423 9424 612.5625 612.5750 612.5875 612.6000 612.6125 612.6250 612.6375 612.6500 612.6625 612.6750 612.6875 612.7000 612.7125 612.7250 612.7375 9365 9366 9367 9368 9369 9370 9371 9372 9373 9374 9375 9376 9377 9378 9379 612.9375 612.9500 612.9625 612.9750 612.9875 613.0000 613.0125 613.0250 613.0375 613.0500 613.0625 613.0750 613.0875 613.1000 613.1125 9395 9396 9397 9398 9399 9400 9401 9402 9403 9404 9405 9406 9407 9408 9409 613.3125 613.3250 613.3375 613.3500 613.3625 613.3750 613.3875 613.4000 613.4125 613.4250 613.4375 613.4500 613.4625 613.4750 613.4875 9425 9426 9427 9428 9429 9430 9431 9432 9433 9434 9435 9436 9437 9438 9439 Operator’s Manual ZM-920PA/921PA/930PA/931PA 79 Transmission frequency (MHz) Channel No. 613.5000 613.5125 613.5250 80 Transmission frequency (MHz) Channel No. Transmission frequency (MHz) Channel No. 9440 9441 9442 613.6625 613.6750 9453 9454 613.8250 613.8375 9466 9467 613.6875 9455 613.8500 9468 613.5375 613.5500 613.5625 613.5750 9443 9444 9445 9446 9469 9470 9471 9472 9447 9456 9457 9458 9459 9460 613.8625 613.8750 613.8875 613.9000 613.5875 613.7000 613.7125 613.7250 613.7375 613.7500 613.9125 9473 613.6000 613.6125 613.6250 613.6375 9448 9449 9450 9451 613.7625 613.7750 613.7875 613.8000 9461 9462 9463 9464 613.6500 9452 613.8125 9465 613.9250 613.9375 613.9500 613.9625 613.9750 9474 9475 9476 9477 9478 Operator’s Manual ZM-920PA/921PA/930PA/931PA Channel: E002 to E398 Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 1395.0250 1395.0375 1395.0500 1395.0625 1395.0750 1395.0875 1395.1000 1395.1125 E002 E003 E004 E005 E006 E007 E008 E009 1395.3875 1395.4000 1395.4125 1395.4250 1395.4375 1395.4500 1395.4625 1395.4750 E031 E032 E033 E034 E035 E036 E037 E038 1395.1250 1395.1375 1395.1500 1395.1625 1395.1750 1395.1875 1395.2000 1395.2125 1395.2250 1395.2375 1395.2500 1395.2625 1395.2750 1395.2875 1395.3000 1395.3125 1395.3250 1395.3375 1395.3500 1395.3625 1395.3750 E010 E011 E012 E013 E014 E015 E016 E017 E018 E019 E020 E021 E022 E023 E024 E025 E026 E027 E028 E029 E030 1395.4875 1395.5000 1395.5125 1395.5250 1395.5375 1395.5500 E039 E040 E041 E042 E043 E044 1395.5625 1395.5750 1395.5875 1395.6000 1395.6125 1395.6250 1395.6375 1395.6500 1395.6625 1395.6750 1395.6875 1395.7000 1395.7125 1395.7250 1395.7375 E045 E046 E047 E048 E049 E050 E051 E052 E053 E054 E055 E056 E057 E058 E059 Operator’s Manual ZM-920PA/921PA/930PA/931PA 1395.7500 1395.7625 1395.7750 1395.7875 1395.8000 1395.8125 1395.8250 1395.8375 1395.8500 1395.8625 1395.8750 1395.8875 1395.9000 1395.9125 1395.9250 1395.9375 1395.9500 1395.9625 1395.9750 1395.9875 1396.0000 1396.0125 1396.0250 1396.0375 1396.0500 1396.0625 1396.0750 1396.0875 1396.1000 E060 E061 E062 E063 E064 E065 E066 E067 E068 E069 E070 E071 E072 E073 E074 E075 E076 E077 E078 E079 E080 E081 E082 E083 E084 E085 E086 E087 E088 81 Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 1396.1125 1396.1250 1396.1375 1396.1500 1396.1625 1396.1750 1396.1875 1396.2000 1396.2125 1396.2250 1396.2375 1396.2500 1396.2625 1396.2750 1396.2875 1396.3000 1396.3125 1396.3250 1396.3375 1396.3500 1396.3625 1396.3750 1396.3875 1396.4000 1396.4125 1396.4250 1396.4375 1396.4500 1396.4625 1396.4750 82 E089 E090 E091 E092 E093 E094 E095 E096 E097 E098 E099 E100 E101 E102 E103 E104 E105 E106 E107 E108 E109 E110 E111 E112 E113 E114 E115 E116 E117 E118 1396.4875 1396.5000 1396.5125 1396.5250 1396.5375 1396.5500 1396.5625 1396.5750 1396.5875 1396.6000 1396.6125 1396.6250 1396.6375 1396.6500 1396.6625 1396.6750 1396.6875 1396.7000 1396.7125 1396.7250 1396.7375 1396.7500 1396.7625 1396.7750 1396.7875 1396.8000 1396.8125 1396.8250 1396.8375 1396.8500 E119 E120 E121 E122 E123 E124 E125 E126 E127 E128 E129 E130 E131 E132 E133 E134 E135 E136 E137 E138 E139 E140 E141 E142 E143 E144 E145 E146 E147 E148 1396.8625 1396.8750 1396.8875 1396.9000 1396.9125 1396.9250 1396.9375 1396.9500 1396.9625 1396.9750 1396.9875 1397.0000 1397.0125 1397.0250 1397.0375 1397.0500 1397.0625 1397.0750 1397.0875 1397.1000 1397.1125 1397.1250 1397.1375 1397.1500 1397.1625 1397.1750 1397.1875 1397.2000 1397.2125 1397.2250 E149 E150 E151 E152 E153 E154 E155 E156 E157 E158 E159 E160 E161 E162 E163 E164 E165 E166 E167 E168 E169 E170 E171 E172 E173 E174 E175 E176 E177 E178 Operator’s Manual ZM-920PA/921PA/930PA/931PA Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 1397.2375 1397.2500 1397.2625 1397.2750 1397.2875 1397.3000 1397.3125 1397.3250 1397.3375 1397.3500 1397.3625 1397.3750 1397.3875 1397.4000 1397.4125 1397.4250 1397.4375 1397.4500 1397.4625 1397.4750 1397.4875 1397.5000 1397.5125 1397.5250 1397.5375 1397.5500 1397.5625 1397.5750 1397.5875 1397.6000 E179 E180 E181 E182 E183 E184 E185 E186 E187 E188 E189 E190 E191 E192 E193 E194 E195 E196 E197 E198 E199 E200 E201 E202 E203 E204 E205 E206 E207 E208 1397.6125 1397.6250 1397.6375 1397.6500 1397.6625 1397.6750 1397.6875 1397.7000 1397.7125 1397.7250 1397.7375 1397.7500 1397.7625 1397.7750 1397.7875 1397.8000 1397.8125 1397.8250 1397.8375 1397.8500 1397.8625 1397.8750 1397.8875 1397.9000 E209 E210 E211 E212 E213 E214 E215 E216 E217 E218 E219 E220 E221 E222 E223 E224 E225 E226 E227 E228 E229 E230 E231 E232 1397.9875 1398.0000 1398.0125 1398.0250 1398.0375 1398.0500 1398.0625 1398.0750 1398.0875 1398.1000 1398.1125 1398.1250 1398.1375 1398.1500 1398.1625 1398.1750 1398.1875 1398.2000 1398.2125 1398.2250 1398.2375 1398.2500 1398.2625 1398.2750 E239 E240 E241 E242 E243 E244 E245 E246 E247 E248 E249 E250 E251 E252 E253 E254 E255 E256 E257 E258 E259 E260 E261 E262 1397.9125 1397.9250 1397.9375 1397.9500 1397.9625 1397.9750 E233 E234 E235 E236 E237 E238 1398.2875 1398.3000 1398.3125 1398.3250 1398.3375 1398.3500 E263 E264 E265 E266 E267 E268 Operator’s Manual ZM-920PA/921PA/930PA/931PA 83 Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 84 1398.3625 1398.3750 1398.3875 1398.4000 1398.4125 1398.4250 1398.4375 1398.4500 1398.4625 1398.4750 1398.4875 1398.5000 1398.5125 1398.5250 1398.5375 1398.5500 1398.5625 1398.5750 1398.5875 1398.6000 1398.6125 1398.6250 1398.6375 1398.6500 E269 E270 E271 E272 E273 E274 E275 E276 E277 E278 E279 E280 E281 E282 E283 E284 E285 E286 E287 E288 E289 E290 E291 E292 1398.6625 1398.6750 1398.6875 1398.7000 1398.7125 1398.7250 E293 E294 E295 E296 E297 E298 1398.7375 1398.7500 1398.7625 1398.7750 1398.7875 1398.8000 1398.8125 1398.8250 1398.8375 1398.8500 1398.8625 1398.8750 1398.8875 1398.9000 1398.9125 1398.9250 1398.9375 1398.9500 1398.9625 1398.9750 1398.9875 1399.0000 1399.0125 1399.0250 1399.0375 1399.0500 1399.0625 1399.0750 1399.0875 1399.1000 E299 E300 E301 E302 E303 E304 E305 E306 E307 E308 E309 E310 E311 E312 E313 E314 E315 E316 E317 E318 E319 E320 E321 E322 E323 E324 E325 E326 E327 E328 1399.1125 1399.1250 1399.1375 1399.1500 1399.1625 1399.1750 1399.1875 1399.2000 1399.2125 1399.2250 1399.2375 1399.2500 1399.2625 1399.2750 1399.2875 1399.3000 1399.3125 1399.3250 1399.3375 1399.3500 1399.3625 1399.3750 1399.3875 1399.4000 1399.4125 1399.4250 1399.4375 1399.4500 1399.4625 1399.4750 E329 E330 E331 E332 E333 E334 E335 E336 E337 E338 E339 E340 E341 E342 E343 E344 E345 E346 E347 E348 E349 E350 E351 E352 E353 E354 E355 E356 E357 E358 Operator’s Manual ZM-920PA/921PA/930PA/931PA Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 1399.4875 1399.5000 1399.5125 1399.5250 1399.5375 1399.5500 1399.5625 1399.5750 1399.5875 1399.6000 1399.6125 1399.6250 1399.6375 1399.6500 E359 E360 E361 E362 E363 E364 E365 E366 E367 E368 E369 E370 E371 E372 1399.6625 1399.6750 1399.6875 1399.7000 1399.7125 1399.7250 1399.7375 1399.7500 1399.7625 1399.7750 1399.7875 1399.8000 1399.8125 1399.8250 E373 E374 E375 E376 E377 E378 E379 E380 E381 E382 E383 E384 E385 E386 Operator’s Manual ZM-920PA/921PA/930PA/931PA 1399.8375 1399.8500 1399.8625 1399.8750 1399.8875 1399.9000 1399.9125 1399.9250 1399.9375 1399.9500 1399.9625 1399.9750 E387 E388 E389 E390 E391 E392 E393 E394 E395 E396 E397 E398 85 Channel: E502 to E898 Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 1427.0250 1427.0375 1427.0500 1427.0625 1427.0750 1427.0875 1427.1000 1427.1125 1427.1250 1427.1375 1427.1500 1427.1625 1427.1750 1427.1875 1427.2000 1427.2125 1427.2250 1427.2375 1427.2500 1427.2625 1427.2750 1427.2875 1427.3000 1427.3125 1427.3250 1427.3375 1427.3500 1427.3625 1427.3750 86 E502 E503 E504 E505 E506 E507 E508 E509 E510 E511 E512 E513 E514 E515 E516 E517 E518 E519 E520 E521 E522 E523 E524 E525 E526 E527 E528 E529 E530 1427.3875 1427.4000 1427.4125 1427.4250 1427.4375 1427.4500 1427.4625 1427.4750 1427.4875 1427.5000 1427.5125 1427.5250 1427.5375 1427.5500 1427.5625 1427.5750 1427.5875 1427.6000 1427.6125 1427.6250 1427.6375 1427.6500 1427.6625 1427.6750 1427.6875 1427.7000 1427.7125 1427.7250 1427.7375 E531 E532 E533 E534 E535 E536 E537 E538 E539 E540 E541 E542 E543 E544 E545 E546 E547 E548 E549 E550 E551 E552 E553 E554 E555 E556 E557 E558 E559 1427.7500 1427.7625 1427.7750 1427.7875 1427.8000 1427.8125 1427.8250 1427.8375 1427.8500 1427.8625 1427.8750 1427.8875 1427.9000 1427.9125 1427.9250 1427.9375 1427.9500 1427.9625 1427.9750 1427.9875 1428.0000 1428.0125 1428.0250 1428.0375 1428.0500 1428.0625 1428.0750 1428.0875 1428.1000 E560 E561 E562 E563 E564 E565 E566 E567 E568 E569 E570 E571 E572 E573 E574 E575 E576 E577 E578 E579 E580 E581 E582 E583 E584 E585 E586 E587 E588 Operator’s Manual ZM-920PA/921PA/930PA/931PA Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 1428.1125 1428.1250 1428.1375 1428.1500 E589 E590 E591 E592 1428.4875 1428.5000 1428.5125 1428.5250 E619 E620 E621 E622 1428.8625 1428.8750 1428.8875 1428.9000 E649 E650 E651 E652 1428.1625 E593 1428.5375 E623 1428.9125 E653 1428.1750 1428.1875 1428.2000 1428.2125 1428.2250 1428.2375 1428.2500 1428.2625 1428.2750 1428.2875 E594 E595 E596 E597 E598 E599 E600 E601 E602 E603 1428.5500 1428.5625 1428.5750 1428.5875 1428.6000 1428.6125 1428.6250 1428.6375 1428.6500 1428.6625 E624 E625 E626 E627 E628 E629 E630 E631 E632 E633 1428.9250 1428.9375 1428.9500 1428.9625 1428.9750 1428.9875 1429.0000 1429.0125 1429.0250 1429.0375 E654 E655 E656 E657 E658 E659 E660 E661 E662 E663 1428.3000 1428.3125 1428.3250 1428.3375 1428.3500 1428.3625 1428.3750 1428.3875 1428.4000 1428.4125 1428.4250 1428.4375 1428.4500 1428.4625 1428.4750 E604 E605 E606 E607 E608 E609 E610 E611 E612 E613 E614 E615 E616 E617 E618 1428.6750 1428.6875 1428.7000 1428.7125 1428.7250 1428.7375 1428.7500 1428.7625 1428.7750 1428.7875 1428.8000 1428.8125 1428.8250 1428.8375 1428.8500 E634 E635 E636 E637 E638 E639 E640 E641 E642 E643 E644 E645 E646 E647 E648 1429.0500 1429.0625 1429.0750 1429.0875 1429.1000 1429.1125 1429.1250 1429.1375 1429.1500 1429.1625 1429.1750 1429.1875 1429.2000 1429.2125 1429.2250 E664 E665 E666 E667 E668 E669 E670 E671 E672 E673 E674 E675 E676 E677 E678 Operator’s Manual ZM-920PA/921PA/930PA/931PA 87 Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 1429.2375 1429.2500 1429.2625 1429.2750 1429.2875 1429.3000 1429.3125 1429.3250 1429.3375 1429.3500 1429.3625 1429.3750 1429.3875 1429.4000 1429.4125 1429.4250 1429.4375 1429.4500 1429.4625 1429.4750 1429.4875 1429.5000 1429.5125 1429.5250 1429.5375 1429.5500 1429.5625 1429.5750 1429.5875 1429.6000 88 E679 E680 E681 E682 E683 E684 E685 E686 E687 E688 E689 E690 E691 E692 E693 E694 E695 E696 E697 E698 E699 E700 E701 E702 E703 E704 E705 E706 E707 E708 1429.6125 1429.6250 1429.6375 1429.6500 1429.6625 1429.6750 1429.6875 1429.7000 1429.7125 1429.7250 1429.7375 1429.7500 1429.7625 1429.7750 1429.7875 1429.8000 1429.8125 1429.8250 1429.8375 1429.8500 1429.8625 1429.8750 1429.8875 1429.9000 1429.9125 1429.9250 1429.9375 1429.9500 1429.9625 1429.9750 E709 E710 E711 E712 E713 E714 E715 E716 E717 E718 E719 E720 E721 E722 E723 E724 E725 E726 E727 E728 E729 E730 E731 E732 E733 E734 E735 E736 E737 E738 1429.9875 1430.0000 1430.0125 1430.0250 1430.0375 1430.0500 1430.0625 1430.0750 1430.0875 1430.1000 1430.1125 1430.1250 1430.1375 1430.1500 1430.1625 1430.1750 1430.1875 1430.2000 1430.2125 1430.2250 1430.2375 1430.2500 1430.2625 1430.2750 1430.2875 1430.3000 1430.3125 1430.3250 1430.3375 1430.3500 E739 E740 E741 E742 E743 E744 E745 E746 E747 E748 E749 E750 E751 E752 E753 E754 E755 E756 E757 E758 E759 E760 E761 E762 E763 E764 E765 E766 E767 E768 Operator’s Manual ZM-920PA/921PA/930PA/931PA Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 1430.3625 1430.3750 1430.3875 1430.4000 1430.4125 1430.4250 1430.4375 1430.4500 1430.4625 1430.4750 1430.4875 1430.5000 1430.5125 1430.5250 1430.5375 1430.5500 1430.5625 1430.5750 1430.5875 1430.6000 1430.6125 1430.6250 1430.6375 1430.6500 1430.6625 1430.6750 1430.6875 1430.7000 1430.7125 1430.7250 E769 E770 E771 E772 E773 E774 E775 E776 E777 E778 E779 E780 E781 E782 E783 E784 E785 E786 E787 E788 E789 E790 E791 E792 E793 E794 E795 E796 E797 E798 1430.7375 1430.7500 1430.7625 1430.7750 1430.7875 1430.8000 1430.8125 1430.8250 1430.8375 1430.8500 1430.8625 1430.8750 1430.8875 1430.9000 1430.9125 1430.9250 1430.9375 1430.9500 1430.9625 1430.9750 1430.9875 1431.0000 1431.0125 1431.0250 1431.0375 1431.0500 1431.0625 1431.0750 1431.0875 1431.1000 E799 E800 E801 E802 E803 E804 E805 E806 E807 E808 E809 E810 E811 E812 E813 E814 E815 E816 E817 E818 E819 E820 E821 E822 E823 E824 E825 E826 E827 E828 Operator’s Manual ZM-920PA/921PA/930PA/931PA 1431.1125 1431.1250 1431.1375 1431.1500 1431.1625 1431.1750 1431.1875 1431.2000 1431.2125 1431.2250 1431.2375 1431.2500 1431.2625 1431.2750 1431.2875 1431.3000 1431.3125 1431.3250 1431.3375 1431.3500 1431.3625 1431.3750 1431.3875 1431.4000 1431.4125 1431.4250 1431.4375 1431.4500 1431.4625 1431.4750 E829 E830 E831 E832 E833 E834 E835 E836 E837 E838 E839 E840 E841 E842 E843 E844 E845 E846 E847 E848 E849 E850 E851 E852 E853 E854 E855 E856 E857 E858 89 Transmission Transmission Transmission Channel Channel Channel frequency frequency frequency No. No. No. (MHz) (MHz) (MHz) 1431.4875 1431.5000 1431.5125 1431.5250 1431.5375 1431.5500 1431.5625 1431.5750 1431.5875 1431.6000 1431.6125 1431.6250 1431.6375 1431.6500 90 E859 E860 E861 E862 E863 E864 E865 E866 E867 E868 E869 E870 E871 E872 1431.6625 1431.6750 1431.6875 1431.7000 1431.7125 1431.7250 1431.7375 1431.7500 1431.7625 1431.7750 1431.7875 1431.8000 1431.8125 1431.8250 E873 E874 E875 E876 E877 E878 E879 E880 E881 E882 E883 E884 E885 E886 1431.8375 1431.8500 1431.8625 1431.8750 1431.8875 1431.9000 1431.9125 1431.9250 1431.9375 1431.9500 1431.9625 1431.9750 E887 E888 E889 E890 E891 E892 E893 E894 E895 E896 E897 E898 Operator’s Manual ZM-920PA/921PA/930PA/931PA
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.4 Linearized : No Create Date : 2008:10:05 11:00:08+09:00 Creator : Adobe InDesign CS2_J (4.0.5) Modify Date : 2008:10:31 10:57:23+09:00 Tagged PDF : Yes Producer : Adobe PDF Library 7.0 Trapped : False Creation Date : 2008:10:05 11:00:08+09:00 Mod Date : 2008:10:31 10:57:23+09:00 Metadata Date : 2008:10:31 10:57:23+09:00 Creator Tool : Adobe InDesign CS2_J (4.0.5) Thumbnail Format : JPEG Thumbnail Width : 256 Thumbnail Height : 256 Thumbnail Image : (Binary data 4430 bytes, use -b option to extract) Instance ID : uuid:f75713da-59f4-40f7-8eac-de913e457989 Document ID : adobe:docid:indd:b55aa884-520c-11dd-8b80-c4ce250e5b3b Rendition Class : proof:pdf Derived From Instance ID : 0d474829-48cc-11dd-922d-988cdb1385a6 Derived From Document ID : adobe:docid:indd:65025cdf-48c8-11dd-9fc3-9f2e64bb5881 Manifest Link Form : ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream Manifest Placed X Resolution : 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00 Manifest Placed Y Resolution : 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00, 72.00 Manifest Placed Resolution Unit : Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches, Inches Manifest Reference Instance ID : uuid:2e417386-ab42-41f5-8495-389d35511147, uuid:ef3970d2-0dcf-40d2-b11a-ddb297cbf420, uuid:ea825821-ce7d-4f7b-956d-3867c12e1c4d, uuid:faaa4aa4-9c9f-44ed-a612-2d6d5e3efe4c, uuid:ff582ca2-ed3a-4234-884f-14dcfe98071b, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:1626a4ab-df3d-40e7-8041-6e2c498bf892, uuid:7250e782-3db2-46b3-bf0e-98374995ea75, uuid:d5e93ca6-c8df-4836-90eb-ffd9d5bcee01, uuid:76d420dd-ebc6-4bbf-b35f-b079c7325b4d, uuid:1f79f104-5eff-4b23-bcfe-f3104e1f1cc2, uuid:58b67b48-837d-4b6c-bd4b-817fdb0893b5, uuid:30fbb454-0f04-4fe9-94b1-3be4b5f7a5a0, uuid:920eb132-0d51-48a8-bffe-f97175bdf603, uuid:abcd517a-cb0a-4154-bd00-cedc28e5b5a3, uuid:3cf5b001-6173-468f-9d4f-9c6790d455c6, uuid:3cf5b001-6173-468f-9d4f-9c6790d455c6, uuid:d6bf8575-2ad0-4f2b-a275-7479404d3d57, uuid:c04f2835-e2b3-4269-8140-f2cdb679fd31, uuid:d6bf8575-2ad0-4f2b-a275-7479404d3d57, uuid:c04f2835-e2b3-4269-8140-f2cdb679fd31, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:3e9c76a7-4e9c-4566-a0d0-9b4be438fb2f, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:22564e96-4cf8-4d08-a59b-66261ba438f6, uuid:2d68e825-c700-41b3-83b5-b836cfec8630, uuid:22564e96-4cf8-4d08-a59b-66261ba438f6, uuid:2d68e825-c700-41b3-83b5-b836cfec8630, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:b21cc7fe-85f8-4426-9d82-beb8cdef06c2, uuid:cb3de6f6-e137-4e58-8871-d3e199bb7e6f, uuid:ba135975-97c9-44c6-b537-cfbc3c57ad06, uuid:a2684e19-4782-482c-a9b1-30161b1c0c75, uuid:2a6cac31-1754-4234-b1d2-b87cc2842404, uuid:df77a215-a528-4878-baab-461075a0e403, uuid:577de756-59ec-4598-8b00-158e57afd233, uuid:431d5394-683c-4b63-bd7e-ac2b4c6867cb, uuid:431d5394-683c-4b63-bd7e-ac2b4c6867cb, uuid:a838e6dd-5326-49b6-9bf2-0a69d3dedff9, uuid:e7501cd8-0e97-40b4-a3b3-547bfa4cf839, uuid:4c0e0c75-0f8c-4290-a27c-a35a859850e3, uuid:fe235ed7-f721-4205-b61c-bcbbebd8d3ea, uuid:fe235ed7-f721-4205-b61c-bcbbebd8d3ea, uuid:03f230fe-8168-41ab-bd7e-ac351125f400, uuid:5c7af7e9-5cde-4ea5-ac53-5557ac6e74d9, uuid:5c7af7e9-5cde-4ea5-ac53-5557ac6e74d9, uuid:953448b8-d1ef-4dc2-8263-0b40196ad0ce, uuid:953448b8-d1ef-4dc2-8263-0b40196ad0ce, uuid:ca3d102b-3ed0-470f-af82-6e6c8b825569, uuid:ca3d102b-3ed0-470f-af82-6e6c8b825569, uuid:29a4e645-8e9e-484d-9f6a-6982d9befee4, uuid:29a4e645-8e9e-484d-9f6a-6982d9befee4, uuid:a68293f2-8373-4f42-85c8-bd020624be5b, uuid:a68293f2-8373-4f42-85c8-bd020624be5b, uuid:e73f1b44-be4f-487c-8fd4-204bb513d9ce, uuid:e73f1b44-be4f-487c-8fd4-204bb513d9ce, uuid:8b73fee7-cfa5-4b5b-93d6-81d752fb20f6, uuid:f6209906-feb3-4750-9e29-27ffa2022610, uuid:18916a9d-78b0-4688-a104-b721d9cc478a, uuid:91e7a197-0e56-48ff-ac3c-8c655d7baa50, uuid:8e0a0299-675e-4575-85cf-e4c9949beed9, uuid:62f8457a-5d35-48c1-b7a0-528cfbe4eb3a, uuid:ef8f6327-c960-4517-b6f2-0a8489f864a9, uuid:681df685-b559-484d-82e6-84094cdad727, uuid:1802e47d-a47a-4564-a41f-1fdf026a94c3, uuid:2423cc54-7783-4c7c-ab36-b1595cb32a1b, uuid:a9724340-c63c-4fd1-898e-6427701a3d47, uuid:44746c3e-047f-4670-bcb6-cf9878f53144, uuid:ca6c6458-1388-46ac-bb45-034ab0959390, uuid:93f6a974-4bbb-4420-aa8f-87fe0ae5e032, uuid:93f6a974-4bbb-4420-aa8f-87fe0ae5e032, uuid:abcd517a-cb0a-4154-bd00-cedc28e5b5a3, uuid:c9107b8f-2011-434d-96a9-9b0c476a5b31, uuid:c9107b8f-2011-434d-96a9-9b0c476a5b31, uuid:b8ca4f44-f463-415f-af4d-e7ec09e99663, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:7250e782-3db2-46b3-bf0e-98374995ea75, uuid:abcd517a-cb0a-4154-bd00-cedc28e5b5a3, uuid:7250e782-3db2-46b3-bf0e-98374995ea75, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:74f6a198-abf5-4224-a31b-5bca8a0527dc, uuid:d3444aac-d8af-4748-a5fc-876001152459, uuid:d3444aac-d8af-4748-a5fc-876001152459, uuid:fadbd6e6-4e7c-4c81-a5a0-7d33dbd3f496, uuid:fadbd6e6-4e7c-4c81-a5a0-7d33dbd3f496 Manifest Reference Document ID : uuid:AA92D7405D46DD1197ACAE89217B0B30, uuid:A892D7405D46DD1197ACAE89217B0B30, uuid:AC92D7405D46DD1197ACAE89217B0B30, uuid:614AC8FF5C46DD1197ACAE89217B0B30, uuid:C71647C85C46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:C51647C85C46DD1197ACAE89217B0B30, uuid:634AC8FF5C46DD1197ACAE89217B0B30, uuid:B92B4F4F6046DD11A9F5E6DF9C74764A, uuid:9CCD98925B46DD1197ACAE89217B0B30, uuid:2E958A74604DDD11A395F86AAE11BE8B, uuid:9ECD98925B46DD1197ACAE89217B0B30, uuid:120B56E75C46DD1197ACAE89217B0B30, uuid:140B56E75C46DD1197ACAE89217B0B30, uuid:5F4AC8FF5C46DD1197ACAE89217B0B30, uuid:30958A74604DDD11A395F86AAE11BE8B, uuid:30958A74604DDD11A395F86AAE11BE8B, uuid:DA6B399D5C46DD1197ACAE89217B0B30, uuid:D86B399D5C46DD1197ACAE89217B0B30, uuid:DA6B399D5C46DD1197ACAE89217B0B30, uuid:D86B399D5C46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:DC6B399D5C46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:92AEBB025C46DD1197ACAE89217B0B30, uuid:94AEBB025C46DD1197ACAE89217B0B30, uuid:92AEBB025C46DD1197ACAE89217B0B30, uuid:94AEBB025C46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:7997116A5C46DD1197ACAE89217B0B30, uuid:BC123FA75D46DD11A9F5E6DF9C74764A, uuid:D46B399D5C46DD1197ACAE89217B0B30, uuid:D66B399D5C46DD1197ACAE89217B0B30, uuid:BA123FA75D46DD11A9F5E6DF9C74764A, uuid:98AEBB025C46DD1197ACAE89217B0B30, uuid:866C17B7DA47DD11A748873697FBB31A, uuid:32958A74604DDD11A395F86AAE11BE8B, uuid:32958A74604DDD11A395F86AAE11BE8B, uuid:7397116A5C46DD1197ACAE89217B0B30, uuid:6F97116A5C46DD1197ACAE89217B0B30, uuid:7197116A5C46DD1197ACAE89217B0B30, uuid:C11647C85C46DD1197ACAE89217B0B30, uuid:C11647C85C46DD1197ACAE89217B0B30, uuid:AE92D7405D46DD1197ACAE89217B0B30, uuid:BE123FA75D46DD11A9F5E6DF9C74764A, uuid:BE123FA75D46DD11A9F5E6DF9C74764A, uuid:BF1647C85C46DD1197ACAE89217B0B30, uuid:BF1647C85C46DD1197ACAE89217B0B30, uuid:3D3823CD5B46DD1197ACAE89217B0B30, uuid:3D3823CD5B46DD1197ACAE89217B0B30, uuid:3B3823CD5B46DD1197ACAE89217B0B30, uuid:3B3823CD5B46DD1197ACAE89217B0B30, uuid:393823CD5B46DD1197ACAE89217B0B30, uuid:393823CD5B46DD1197ACAE89217B0B30, uuid:373823CD5B46DD1197ACAE89217B0B30, uuid:373823CD5B46DD1197ACAE89217B0B30, uuid:DC21992E5D46DD1197ACAE89217B0B30, uuid:100B56E75C46DD1197ACAE89217B0B30, uuid:886B77A80FEFDC119595EDD5F250AAE8, uuid:C8A80E215D46DD1197ACAE89217B0B30, uuid:C6A80E215D46DD1197ACAE89217B0B30, uuid:654AC8FF5C46DD1197ACAE89217B0B30, uuid:BDBD36345D46DD1197ACAE89217B0B30, uuid:BBBD36345D46DD1197ACAE89217B0B30, uuid:0E0B56E75C46DD1197ACAE89217B0B30, uuid:C0123FA75D46DD11A9F5E6DF9C74764A, uuid:0C0B56E75C46DD1197ACAE89217B0B30, uuid:C91647C85C46DD1197ACAE89217B0B30, uuid:7597116A5C46DD1197ACAE89217B0B30, uuid:E1226131174EDD11A6138F14ECE982BC, uuid:E1226131174EDD11A6138F14ECE982BC, uuid:5F4AC8FF5C46DD1197ACAE89217B0B30, uuid:333823CD5B46DD1197ACAE89217B0B30, uuid:333823CD5B46DD1197ACAE89217B0B30, uuid:C31647C85C46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:634AC8FF5C46DD1197ACAE89217B0B30, uuid:5F4AC8FF5C46DD1197ACAE89217B0B30, uuid:634AC8FF5C46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:C1BD36345D46DD1197ACAE89217B0B30, uuid:A3E35F7DBE14DD11A65ABAC307FA1A18, uuid:A3E35F7DBE14DD11A65ABAC307FA1A18, uuid:A4CD98925B46DD1197ACAE89217B0B30, uuid:A4CD98925B46DD1197ACAE89217B0B30 Format : application/pdf Page Count : 99EXIF Metadata provided by EXIF.tools