Olympus Medical Systems EC-S10 Low Power Transmitter User Manual

Olympus Medical Systems Corp. Low Power Transmitter Users Manual

Users manual

OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                   1            INSTRUCTIONS  OLYMPUS CAPSULE ENDOSCOPE SYSTEM CAPSULE ENDOSCOPE SYSTEM SET MAJ-Y0136 MAJ-Y0136 CAPSULE ENDOSCOPE   EC-Y0005               Refer  to  the  companion  manual,  the  “WORKSTATION  MANUAL”  and  “Cautions  for  Capsule Endoscopy Patient” for operating information of the MAJ-Y0134. USA:CAUTION:Investigational device. Limited by Federal (or United State) Law to investigational use.
   2   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Contents Labels and Symbols.................................................................................................................................1 Symbols...................................................................................................................................................1 Labels......................................................................................................................................................2 Important Information — Please Read Before Use...............................................................................9 Intended use ...........................................................................................................................................9 Instruction manual...................................................................................................................................9 User qualifications.................................................................................................................................10 Instrument compatibility ........................................................................................................................10 Repair and modification ........................................................................................................................10 Signal words..........................................................................................................................................11 Dangers, warnings and cautions ..........................................................................................................12 Danger, warning and cautions for patients ...........................................................................................21 Expiration date of the capsule endoscope............................................................................................23 Color Tone of the Capsule Endoscope Image......................................................................................24 EMC (Electromagnetic Compatibility)...................................................................................................27 Limits of the technology ........................................................................................................................27 Chapter 1 Checking the Package Contents.........................................................................................28 1.1 Capsule endoscope system MAJ-Y0136 (MAJ-Y0136) .................................................................29 1.2 Capsule endoscope (EC-Y0005)....................................................................................................30 Chapter 2 Nomenclature and Functions ..............................................................................................32 2.1 System components .......................................................................................................................32 2.2 Capsule endoscope (EC-Y0005)....................................................................................................34 2.3 Antenna unit (X432262-21).............................................................................................................35 2.4 Recorder unit (X432262-11) ...........................................................................................................36 2.5 Recorder unit cradle (X432262-31) ................................................................................................38 2.6 Recorder unit harness (X432251-40) .............................................................................................39 2.7 Antenna harness (X432252-51)......................................................................................................40 2.8 Capsule activator (MAJ-1478) ........................................................................................................41 2.9 Battery pack (X432262-11-12)........................................................................................................41 2.10 Workstation (MAJ-Y0134).............................................................................................................41 Chapter 3 Installation and Connection.................................................................................................42 3.1 Preparation......................................................................................................................................42 3.2 Exterior............................................................................................................................................43 3.3 Capsule endoscope ........................................................................................................................43 3.4 Preparation of the recorder unit......................................................................................................44 3.5 Preparation of the recorder unit cradle ...........................................................................................46 Chapter 4 Operation ...............................................................................................................................48 4.1 Preparation and Inspection before examination.............................................................................48 4.2 Attaching the equipment to the patient ...........................................................................................54 4.3 Preparing the capsule endoscope ..................................................................................................62 4.4 Starting the examination .................................................................................................................82 4.5 During the examination...................................................................................................................83 4.6 Ending the examination ..................................................................................................................85 Chapter 5 Care, Storage and Disposal.................................................................................................91 5.1 Care.................................................................................................................................................91 5.2 Storage............................................................................................................................................92 i
Contents    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            3 5.3 Disposal ..........................................................................................................................................93 Chapter 6 Troubleshooting....................................................................................................................94 6.1 Troubleshooting ..............................................................................................................................94 6.2 Returning the system for repair ......................................................................................................98 Appendix..................................................................................................................................................99 Transportation, storage, and operation environment ...........................................................................99 Specifications ......................................................................................................................................103  ii
OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                   1      Labels and Symbols  Safety-related labels and symbols are attached at the locations shown below. If labels or symbols are missing or illegible, contact Olympus.  Symbols  The meaning(s) of the symbol(s) shown on the package, the back cover of this instruction manual and/or this equipment are as follows:  Caution, refer to instructions.  Refer to instructions.  When this equipment is in the proximity of portable RF (Radio Frequency) communication devices, electromagnetic interference that affects cardiac pacemakers may occur. Users of cardiac pacemakers must maintain a sufficient distance from the equipment.  Power switch (ON/OFF)   TYPE BF applied part  Single use only  Use by (expiration date)    Sterilized using ethylene oxide  Non-sterile  Manufacturer  Authorized representative in the European Community
   2   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Labels Capsule endoscope   Capsule endoscope storage case (Label 1)  OLYMPUS EC-Y0005OLYMPUS EC-Y0005OLYMPUS EC-Y0005OLYMPUS EC-Y0005IC : 4763B-ECS10IC : 4763B-ECS10IC : 4763B-ECS10IC : 4763B-ECS10       Sterilization lot number Expiration date TYPE BF applied part marking Sterilized with ethylene oxide Product name and model name Protection against fluid ingress Refer to instructions Single use only Manufacture details FCC ID number IC    number
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            3 Antenna unit                                       X432262-21X432262-21X432262-21X432262-21Label Label Unit Unit Model name Manufacturer details IC note Refer to instructions FCC marking CSA marking
   4   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Recorder unit     INPUT 5V-3A(from X432262-31) INPUT 5V-3A(from X432262-31) INPUT 5V-3A(from X432262-31) INPUT 5V-3A(from X432262-31)RECORDER OLYMPUS RE-10RECORDER OLYMPUS RE-10RECORDER OLYMPUS RE-10RECORDER OLYMPUS RE-10MODEL X432262-11MODEL X432262-11MODEL X432262-11MODEL X432262-11INPUT 3.7V-500mA(FromX432262-11-12)INPUT 3.7V-500mA(FromX432262-11-12)INPUT 3.7V-500mA(FromX432262-11-12)INPUT 3.7V-500mA(FromX432262-11-12)        Recorder unit (back panel) Recorder unit (front panel) MENU button BACK button OK button “”, “”, “⊳” , “” : Cursor button Model Manufacturer details Electrical rating / Input power Power switch IC note FCC marking CSA marking Label
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            5 Recorder unit cradle                              MODEL X432262-31MODEL X432262-31MODEL X432262-31MODEL X432262-31This device complies with Part 15 of the FCC Rules.This device complies with Part 15 of the FCC Rules.This device complies with Part 15 of the FCC Rules.This device complies with Part 15 of the FCC Rules.Operation is subject to the following two condtions:Operation is subject to the following two condtions:Operation is subject to the following two condtions:Operation is subject to the following two condtions:(1)This device may not cause harmful interference,and(1)This device may not cause harmful interference,and(1)This device may not cause harmful interference,and(1)This device may not cause harmful interference,and(2)this device must accept any interference received,(2)this device must accept any interference received,(2)this device must accept any interference received,(2)this device must accept any interference received,Including interference that may cause undesired operation.Including interference that may cause undesired operation.Including interference that may cause undesired operation.Including interference that may cause undesired operation.This Class B digital apparatus complies with CanadianThis Class B digital apparatus complies with CanadianThis Class B digital apparatus complies with CanadianThis Class B digital apparatus complies with CanadianICES-003. Cet appareil numerique de la Classe B estICES-003. Cet appareil numerique de la Classe B estICES-003. Cet appareil numerique de la Classe B estICES-003. Cet appareil numerique de la Classe B estconforme a la norme NMB-003 du Canada.conforme a la norme NMB-003 du Canada.conforme a la norme NMB-003 du Canada.conforme a la norme NMB-003 du Canada.Label Recorder unit cradle (bottom panel) Refer to instructions Electrical rating / Input power FCC note Manufacturer details Model name IC note CSA marking FCC marking
   6   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Battery pack                    Battery pack        MODEL X432262-11-12OUTPUT  3.7V   2860mAh   10.5WhWARNING: Use this Battery pack only X432262-11Keep away from heat or frame.Don’t make the battery pack charged, used, or lift under high temperature conditions.Don’t short circuit, disassemble or reconstruct.Don’t drop, avoid physical shock.Don’t make the Battery pack contacted withwater, chemical or any other liquid.Keep away from metal objects.OLYMPUS MEDICAL SYSTEMS CORP.S/N  110K001 MADE  IN  JAPANBattery pack        MODEL X432262-11-12OUTPUT  3.7V   2860mAh   10.5WhWARNING: Use this Battery pack only X432262-11Keep away from heat or frame.Don’t make the battery pack charged, used, or lift under high temperature conditions.Don’t short circuit, disassemble or reconstruct.Don’t drop, avoid physical shock.Don’t make the Battery pack contacted withwater, chemical or any other liquid.Keep away from metal objects.OLYMPUS MEDICAL SYSTEMS CORP.S/N  110K001 MADE  IN  JAPAN                   Label Refer to instructions Manufacturer details WARNING Serial number Electrical rating / Input power Model name
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            7           Recorder unit harness          Label Model No.
   8   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Antenna harness         ManufacturModel name Label
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            9  Important  Information  —  Please  Read Before Use  Intended use The OLYMPUS CAPSULE ENDOSCOPE SYSTEM has been designed to be used for visualization of the small intestine mucosa. The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas. Do not use this system for any purpose other than its intended use.  Instruction manual This instruction manual contains essential information on using the system safely and effectively. Before use, thoroughly review this manual and the manuals of all equipment that will be used during the procedures, and use the equipment as instructed. Note that the complete instruction manual for this system consists of following parts: this manual, the “WORKSTATION MANUAL”, and the “Cautions for Capsule Endoscopy Patient”. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments about any information in this manual, please contact Olympus.  Terms used in this manual DC magnetic field Magnetic field formed by permanent magnets or electromagnets that are used in health devices and audio speakers.  RTC (Real time clock) Refers to the recorder unit’s internal clock.
   10   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          User qualifications The operator of this instrument must be a physician or medical personnel under the supervision of a physician and must have received sufficient training in capsule endoscopy procedures. This manual, therefore, does not explain or discuss capsule endoscopy procedures.  Physicians performing capsule endoscopy should:  − Have completed a formal training program that includes training in the recognition and management of small intestinal diseases − Have competency and privileges to perform gastrointestinal endoscopy − Be familiar with the Endo Capsule hardware and software − Have formal training in capsule endoscopy that includes hands-on training and supervised review of initial cases by a competent capsule endoscopist − Meet pertinent local hospital/clinic requirements and professional society credentialing guidelines.  Instrument compatibility Refer to the “System chart” in the Appendix to confirm that the system is compatible with the ancillary equipment being used. Using incompatible equipment can result in patient injury and/or equipment damage.  Repair and modification This system does not contain any user-serviceable parts. Do not disassemble, modify or attempt to repair it. Doing so may result in patient or user injury and/or equipment damage. Some problems that appear to be malfunctions may be correctable by referring to Chapter 5, “Troubleshooting”. If the problem cannot be resolved using the information in Chapter 5, contact Olympus.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            11   Signal words The following signal words are used throughout this manual.   Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury.   Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.   Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage.   Indicates additional helpful information.
   12   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Dangers, warnings and cautions  Follow the dangers, warnings and cautions given below when handling the system. This information is to be supplemented by the dangers, warnings and cautions given in each chapter.   • This system is contraindicated to patients with the following conditions. − Patients with cardiac pacemakers or other implanted electronic devices           (e.g., defibrillators) (Cardiac pacemakers or other implanted electronic devices may malfunction due to the RF (radio frequency) interference.) − Patients with known intestinal strictures, adhesions, diverticulum, obstruction, or fistulas that may block the passage of the capsule endoscope (Patients with these physical features risk retention of the capsule endoscope.) − Patients with known dysphagia (Patients with dysphagia may have a risk of aspiration into the tracheal branches and could have difficulty ingesting the capsule endoscope.) − Patients with significant difficulty in swallowing a tablet as large as capsule endoscope (Patients that may have difficulty ingesting a capsule as large as the capsule endoscope risk getting the capsule lodged in their throat.) − Patients who are nonsurgical candidates Open surgery is required if the capsule endoscope is retained in the patient’s body. − Patients who are pregnant Use of the capsule endoscope has not yet been proven to be safe on pregnant patients. − Patients with significant gastrointestinal tract delay Long-term retention of the capsule endoscope may be observed in patients with known radiation enteritis due to intestinal strictures or adhesions. − Patients diagnosed with radiation enteritis Long-term retention of the capsule endoscope may be observed in patients with known radiation enteritis due to intestinal strictures or adhesions.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            13   • Instruct of the patient to stay away from equipment that generates strong electromagnetic fields (such as MRI equipment), between ingestion and excretion of the capsule endoscope. Otherwise, injury within the body cavity may occur. If the excretion of the capsule endoscope has not been confirmed, within 2 weeks after ingestion, attempt to locate the capsule endoscope within the patient’s body by X-ray examination.   • Safety of the Capsule Endoscope has been confirmed by durability and biocompatibility testing for a patient retention period of 30 days. Patient safety has not been verified if the capsule endoscope remains within the patient’s body for a period of 30 days or more. If excretion of the capsule endoscope has not been confirmed within 30 days after ingestion, conduct an X-ray examination in an attempt to locate the capsule endoscope. Once found, remove the capsule endoscope. First of all, perform procedure except for surgical intervention. For methods to remove the capsule endoscope, please follow the respective guidelines in your country. • This system is contraindicated to patients with the following conditions. − Patients who have an obstruction in their digestive tract as a result of past digestive bypass surgery A patient who has undergone bypass surgery in the digestive tract usually has an obstruction that could result in retention of the capsule endoscope. − Patients with instruments that could interfere with the capsule endoscope’s passage in the digestive tract such as a stent in the digestive tract A patient with an instrument in their digestive tract (such as a stent) may have sections that are too narrow for the capsule endoscope to pass through. • Prior to the examination, the physician should explain to the patient the risks of capsule endoscope retention and obtain the patient’s informed consent.
   14   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                           • Prior to using the capsule endoscope, the physician should consider performing a contrasted X-ray series in patients with the following conditions. Patients with suspected intestinal strictures, adhesions, diverticulum, obstructions, fistulas that may block the passage of the capsule endoscope, and patients with suspected gastrointestinal tract delay. − Patients with a history of ileus or small intestinal strictures − Patients diagnosed with, or suspected to have, Crohn’s disease in small intestine − Patients with any history of abdominal surgery, pelvic surgery, and/or radiological treatment • The physician should confirm that the contrast medium reaches the ileocecal valve and should continuously observe/trace the entire small-intestine during the contrasted X-ray series. Do not use the capsule endoscope on the patient if any of the following conditions is observed during the contrast X-ray series. − Patients with known intestinal strictures, adhesions, diverticulum, obstructions, or fistulas that may block the passage of the capsule endoscope. − Patients with significant gastrointestinal tract delay. • The peristaltic motion in the small intestine varies among individuals. Therefore, the recording time in the small intestine may differ among patients. Though this system is designed to obtain the images for 12 hours, image recording in a patient with slower peristaltic motion may finish before the capsule endoscope reaches the patient’s large intestine. • In order to implement the examination smoothly, the physician should hand the “Cautions for capsule endoscopy patient” that comes packed with the capsule endoscope   (EC-Y0005) to the patient at least one day before the examination and verify that the patient understands it thoroughly. • The physician should explain to the patient the risks of retention in the examination and obtain the patient’s informed consent before the examination. The capsule endoscope may cause unexpected intestinal obstruction that may require surgery. • Write your emergency contact number in the “Please provide your emergency contact number” section on the back page of the “Cautions for capsule endoscopy patient” that is packed with the capsule endoscope (EC-Y0005), hand it to the patient and instruct him/her to immediately contact the number if he/she feels abdominal pain, discomfort, or nausea after ingesting the capsule endoscope. If treatment is delayed, patient injury may result.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            15   • A negative or normal result obtained by the imaging capsule does not exclude the presence of pathology and if symptoms persist, further evaluation should be performed. • The capsule endoscope must be ingested under the supervision of the physician or medical personnel authorized by the physician. • Never install and/or operate this system in locations where: − Concentration of oxygen is high. − Oxidizing agents (such as nitrous oxide (N2O)) are present in the atmosphere. − Flammable anesthetics are present in the atmosphere. Otherwise, explosion or fire may result because this system Is not explosion-proof. • The capsule endoscope is a TYPE BF applied part, which means that application to the heart is prohibited. Do not use it for the purpose of observing or treating the heart. • The capsule endoscope is a sterilized single-use product. Reusing a capsule endoscope may cause infection to the patient and/or medical personnel and internal injury to the patient. Also, severe equipment malfunctions may occur. Never reuse a capsule endoscope. • Do not use a capsule endoscope that has been dropped, bitten, or subjected to excessive pressure. Doing so may result in internal injury to the patient. • When using the capsule endoscope, avoid using other medical electrical devices simultaneously. The safety of the use of the capsule endoscope in combination with other medical electrical devices has not been verified. • The capsule endoscope may accidentally enter the trachea. Before the examination, check whether the patient is able to swallow the capsule endoscope safely. If accidental entry occurs, immediately remove the capsule endoscope from the trachea. • Do not disassemble or modify the battery pack. The battery pack is equipped with a safety circuit and a protective circuit. Disabling either or both of these circuits may result in excessive heat, release of smoke, explosion, and/or a fire. • Do not connect the positive (+) and negative (–) terminals of the battery pack using pieces of metal. In addition, do not store or transport the battery pack alongside other metallic objects. The battery pack may be short-circuited, causing an overflow of current, resulting in excessive heat, release of smoke, explosion, and/or a fire from the battery, or excessive heating of the metallic item.
   16   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          • Do not heat the battery pack or bring it in contact with fire. The insulating material may melt, gas could be released, the safety circuit may be damaged, and electrolytic solution may be ignited, causing excessive heat, release of smoke, explosion, and/or a fire. • Do not use or store the battery pack near flames or a source of high temperature (over 80°C (176°F)) such as a stove. If the plastic separator is damaged by heat, the battery pack will be short-circuited internally, causing excessive heat, release of smoke, explosion, and/or a fire. • Keep the battery pack away from water or any other liquids. If the internal protective circuit becomes wet and/or damaged, it may result in excessive heat, release of smoke, explosion, and/or a fire. • Do not recharge the battery pack near an open flame, or under direct sunlight. The protective circuit is activated under high heat, and may prevent proper recharging. If the protection circuit fails, recharging may occur at abnormal current and/or voltage, causing unexpected chemical reactions within the battery pack, which may result in excessive heat, release of smoke, explosion, and/or a fire. • When recharging the battery pack, use only the provided Recorder unit cradle (X432262-31) under the specified recharging conditions. Recharging under conditions other than those specified (temperature outside the specified range) may cause abnormalities or overcharging, causing unexpected chemical reactions within the battery pack, which may result in excessive heat, release of smoke, explosion, and/or a fire. • Do not strike, pierce, drop, or step on the battery pack. A ruptured or damaged battery pack may be short-circuited internally, which could cause excessive heat, release of smoke, explosion, and/or a fire. • Do not apply strong shocks to the battery pack. They could cause the battery pack to leak, generate excessive heat, release smoke, explode, or ignite. If the protection circuit fails, recharging may occur at abnormal current and/or voltage, causing unexpected chemical reactions within the battery pack, which may result in excessive heat, release of smoke, explosion, and/or a fire. • Do not use a battery pack that is visibly damaged or deformed. Excessive heat, release of smoke, explosion, and/or a fire may result. • Use the battery pack only for this equipment. Using the battery pack in other equipment could cause its performance to deteriorate rapidly, or, depending on the equipment, cause damage from overcurrent, excessive heat, release of smoke, explosion, and/or a fire.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            17    • If a battery pack is leaking, take care that the liquid does not come in contact with your skin or eyes. If liquid enters the eye, do not rub it; immediately rinse with clean water, then consult a physician. Failure to take appropriate action may result in injury. • If recharging is not completed after the given recharging time, do not continue recharging. Excessive heat, release of smoke, explosion, and/or a fire may result. • Do not place the battery pack in a microwave or a high-pressure cooker. Rapid heating or breaking of the battery pack’s seal may result in excessive heat, release of smoke, explosion, and/or a fire. In this case, the battery pack may be damaged. Do not use this battery pack and replace it with a new one. • If the operator notices a leak or an abnormal odor from the battery pack, immediately remove it from any heat source in the vicinity. The electrolytic solution may ignite, causing release of smoke, explosion, and/or a fire. Do not use this battery pack, and replace it with a new one. • If the operator notices an abnormal odor, excessive heat, change of color, deformation, or other abnormalities during use, recharging, or storage of the battery pack, immediately remove the battery pack from the recorder unit, and do not use the battery   pack. Continued use of the battery pack may result in excessive heat,   release of smoke, explosion, and/or a fire. Use a new battery pack instead.   • Any ancillary diagnostic tests or examinations (e.g. upper GI endoscopy and colonoscopy) required prior to or in conjunction with this device should be performed at the physician’s discretion. • Do not use the capsule endoscope after its expiration date because the life of the battery may be insufficient to perform the entire exam. The expiration date is placed on the label of the capsule endoscope storage case (see page 2). • To prevent the capsule endoscope battery from deteriorating, store the capsule endoscope within a temperature range of 0 – 25°C (32 – 77°F). Do not store the capsule endoscope in a refrigerator or a freezer because condensation may occur inside the cover dome. •Avoid long term use or storage at temperatures exceeding 40°C (104°F). Battery pack performance may deteriorate rapidly under these conditions. •Avoid long term use at temperatures below 0°C (32°F). In case of long   term use at temperatures below 0°C (32°F), you should not start the examination.
   18   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          • Keep the capsule endoscope away from permanent magnets or electromagnets. Magnets can turn the capsule endoscope ON, resulting in the consumption of battery power. • Keep the capsule activator away from implant devices, such as cardiac pacemakers, defibrillators, nerve stimulators, and other equipment that may be affected by DC magnetic fields. • Before use, ensure that there are no rough or sharp edges and/or projections on the surface of the capsule endoscope. If the surface of the capsule endoscope is not smooth, patient injury may occur. •Keep the recorder unit, antenna unit, and recorder unit cradle from water, alcohol, and any other liquids.Liquids entering these components may damage them and/or result in the system failure. • Connect the cables according to the “System chart” on page 103. Be sure to use only the specified cables for the proper connection of system components. • Before using the recorder unit, check that no error icon is displayed on the recorder unit’s display panel.   If an error icon is displayed, refer to the “Recorder unit error messages” section in Chapter 5, “Troubleshooting”,   to evaluate and attempt to correct the problem. • While initializing the recorder unit, do not remove the recorder unit   from the recorder unit cradle or unplug the USB cable and/or the AC adapter. Doing so may damage the recorder unit’s   internal memory. • The examination cannot proceed if there is insufficient space on the recorder unit’s internal memory and “Download” icon is indicated. Use the workstation to initialize the recorder unit in advance. • While initializing the recorder unit, do not turn OFF the workstation or the recorder unit. Doing so will interrupt initialization. • Prior to the examination, use the recorder unit cradle to fully charge the battery pack in the recorder unit. The examination cannot proceed if the battery pack is not fully charged. • Do not remove the battery pack while the receiver unit’s power is ON. Doing so may damage the recorder unit. • Avoid prolonged use of the system at or below 0°C (32°F). In such temperatures, the capsule endoscope may not be able to operate for 12 hours, and the battery packs may not be fully recharged. • Do not use an antenna sheet of the antenna unit that has been   completely folded or creased. Its reception performance may be   degraded, causing noise to appear in the transmitted images. • The capsule endoscope should only be ingested with water. Use of other liquids may interfere with the examination. • Do not make a diagnosis using images displayed on the recorder unit.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            19 • If more than one capsule endoscope is ingested at the same time, image recording may fail because of radio interference. Before use, the physician has to ensure that the patient has not swallowed any other capsule endoscope. Do not use the capsule endoscope with any other swallowable diagnostic devices because the safety of concomitant use has not been verified. • Do not turn ON a capsule endoscope near another   patient already undergoing an examination with an ingested capsule endoscope. Also do not allow patients undergoing such examination sit next to each other. It may cause capsule endoscope images to become distorted. • While downloading, do not remove either the recorder unit or the USB cable and/or AC adapter from the recorder unit cradle.   Doing so will result in loss of image data. If you do remove the recorder   or the USB cable and AC adapter, connect the recorder to the workstation through the recorder unit cradle and the USB cable and restart the workstation before downloading again. • Do not drop or strike either product. Equipment damage may occur.       • During the capsule endoscope exam instruct the patient to stay away from generators of strong electromagnetic fields (such as MRI equipment, amateur (ham) radio, etc.) otherwise image recording may not occur due to radio interference. • Please advise the patients not to use radio transmitting devices such as keyless entry devices (frequency range of possible interference: 306 - 322 MHz) or stay in an area where other persons may use them frequently, in order to reduce the possibility of dropped video frames caused by the use of these devices. • Since the capsule endoscope is a radio-transmitting device, it could interfere or experience interference with other radio-transmitting devices (e.g., telemetry, keyless entry devices). • On rare occasions, interference may result in the need to repeat the capsule endoscope procedure. In this case, the physician should advise the patient to stay within the premises of the medical facility during the examination to prevent this problem from reoccurring. • The capsule endoscope contains a radio transmitter; therefore, the patient cannot fly in an airplane until the examination is over. • In rare cases, the interference could occur due to the military   use of the frequency band in the USA. • This device has not been authorized as required by the rules of the       Federal Communications Comission. This device is not, and may not be,     offerd for sale or lease, or sold or leased, until authorization is obtained.
   20   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                            • This system is used to examine the small intestine by the patient’s ingestion of a capsule endoscope, that continuously takes pictures as it moves down the gastrointestinal tract by peristalsis. The captured images are radio transmitted to the recorder unit, and are observed using the workstation. After the examination, the capsule endoscope is excreted naturally. • Have a spare capsule endoscope ready for use in case the primary device is determined to be unusable prior to patient ingestion. • After turning ON the capsule endoscope, bring it near the antenna unit on the patient, and confirm that the radio reception indicator lamp blinks in green. If the indicator does not blink, the capsule endoscope and/or the recorder unit may be malfunctioning. • You can check for proper operation of the equipment by periodically inspecting the status of the indicator lamp.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            21  Danger, warning and cautions for patients  Provide the following information to the patients at least one day before the examination, so that the examination may be performed properly. Failure to follow these instructions may result in injury to the patient and/or may interfere with the examination. Please provide the patient with the separate “Caution for Capsule Endoscopy Patient” packed with the capsule endoscope   (EC-Y0005).   •Instruct the patient to stay away from generators of strong electromagnetic fields (such as MRI equipment), between ingestion and excretion of the capsule endoscope. Failure to follow instructions may result in injury to the body cavity.    • Instruct the patient to consult a physician immediately if abdominal pain, discomfort, or nausea is experienced after ingesting the capsule endoscope. If treatment for these symptoms is delayed, it may result in injury to the body cavity. • The safety of the patient has not been verified if the capsule endoscope remains within his/her body for a period of 30 days or more. If the capsule endoscope is not excreted from the body, it must be collected using an endoscope or by surgery, within 30 days from the date of ingestion. To determine the location of the capsule endoscope within the patient’s body, an X-ray examination may be required. Be sure to explain this risk of endoscopic or other surgical operation for collection to the patient.    • Instruct the patient to abstain from eating, beginning 8 hours before the examination until 4 hours after ingesting the capsule endoscope. If necessary, small quantities of water may be consumed. Failure to follow instructions may interfere with the examination. • Instruct the patient to handle the external equipment attached to his/her body with care, and to avoid sudden movements. Failure to follow instructions may interfere with the examination.
   22   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                            • Patients who are currently taking medication must not take any medication during the period beginning two hours before and ending two hours after the beginning of the examination. If the medication must be taken on a set schedule, instruct the patient to consult the prescribing physician about shifting the schedule. Failure to follow instructions may interfere with the examination. • Instruct the patient not to turn the recorder unit’s power OFF, remove the battery pack from the recorder unit, disconnect the antenna unit, nor remove the antenna harness during the examination. Failure to follow instructions may interfere with the examination. • During the examination, the patient puts on the antenna harness   from the clothes of the thin nature material. In consideration of   the examination situation, please appoint the clothes of the patient   on the examination day before. • During the examination of the capsule endoscope, stay away from generators of strong electromagnetic fields (such as MRI equipment, amateur-radio, etc.) and avoid pointing radio-transmitting devices such as keyless entry at the antenna leads. Otherwise, images may not be taken due to radio interference. As the capsule endoscope contains a radio transmitter, you cannot take an airplane until the examination is completed. • Instruct the patient to keep the recorder unit, and the antenna unit,   from water, alcohol, and any other liquids. Liquids entering these components may damage them and/or result in system failure. • Instruct the patient to avoid using the recorder unit at or below 0°C (32°F). Failure to follow instructions may cause the battery pack to deteriorate rapidly, resulting in the examination ending prematurely. • Provide the physician’s emergency contact number for the patient on the “Caution for Capsule Endoscopy Patient” packed with the capsule endoscope system set MAJ-Y0136 (MAJ-Y0136).
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            23  Expiration date of the capsule endoscope  The battery inside the capsule endoscope gradually loses life during long term storage. A storage environment with temperatures over 25°C (77°F) may cause the capsule’s battery to deplete rapidly. Loss of battery life may shorten the operation time (continuously recording and transmitting the images) of the capsule endoscope. If stored in the recommended storage environment (0 – 25°C (32 – 77°F)) the capsule endoscope is designed to record 12 continuous hours by the expiration date printed on the package.   • The storage environment of the capsule endoscope is restricted to between 0 – 25°C (32 – 77°F). Storage of the capsule endoscope in temperatures over 25°C (77°F) may cause the battery to lose life. As a result, the capsule endoscope may not operate for 12 continuous hours. • Never store the capsule endoscope in a refrigerator or freezer. The rapid temperature change that occurs when the capsule endoscope is removed from the refrigerator or freezer may cause condensation on the inside the capsule endoscope. Such condensation may cause the battery to lose life and lead to severe equipment damage. • Do not use the expired capsule endoscope. The expired capsule endoscope may not operate for the assured operation time (12 hours). • The capsule endoscope typically operates for 12 continuous hours. Its passage to the cecum depends on the peristalsis movement of the patient. If peristalsis is slower than normal, the capsule endoscope may not reach the ileocecal valve before the 12 hours are up. Therefore, not every case is guaranteed to view the entire small intestine. • To maximize operation time, the capsule endoscope should be swallowed by the patient immediately after the capsule endoscope is turned on.
   24   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Color Tone of the Capsule Endoscope Image  Images photographed by the CMOS incorporated in the capsule endoscope (EC-Y0005) are wirelessly transmitted to the antenna harness attached to the patient and then recorded on the recorder unit . The images stored in the recorder unit can be played on the workstation for observation. Once the capsule endoscope is dispensed to the patient, it cannot be controlled from outside of the patient. The capsule endoscope’s images may be interfered or unusual with depending on the operating environment or the settings of the devices used in conjunction with the capsule endoscope system. The following are foreseeable image interference cases and unusual image cases and the corresponding preventative methods or solutions.   • If the capsule endoscope gets close to the object rapidly, halation will occur. As a result, the image may whiteout. In this case, refer to the images before and after the whiteout image. Compare Figure 1 (halation) and Figure 2 (normal).    Figure 1                                                                                                                         Figure 2
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            25   • Since the capsule endoscope transmits images by radio, noise may appear in the recorded images depending on the surrounding electromagnetic environment (see Figure 3 (an example of the noise is indicated by the arrow)). If an image with noise is displayed, observe the image carefully.                      Figure 3 • Some debris in the small intestine may look like bleeding in color and shape (see Figure 4 (in the circle)). To determine if the image is debris or bleeding, observe the surrounding condition and the images before and after the image in question.                  Figure 4 • Avoid touching the top cover of the capsule endoscope. Do not wipe with gauze or other clothes. Doing so may damage the top cover and interrupt observation.
   26   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                           • Do not change the color setting of Windows® installed in the workstation. Doing so may change the color tone of the image. • Connect the LCD monitor and the workstation hardware securely. If the connection is loose, the images may not be displayed properly. • Use monitors designated by Olympus for observation of the capsule endoscopic images. Non-designated monitors may display the images in different color tones. • Due to the characteristics of an LCD monitor, color tones may look different depending on the angle of observation. When observing capsule endoscopic images, look straight in the LCD monitor. • Some of the pixels on the LCD monitor display can be lost due to gradual deterioration of the monitor. The portion of the monitor that has lost pixels may appear as white or black. Before starting the observation, make sure that the color tone of the color bar is appropriate, refer to “Color Settings” in the Chapter 8 of the workstation MAJ-Y0134 instruction manual. • Do not observe the image in sunlight whether direct or indirect since the color tone may look different because of the LCD monitor.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            27  EMC (Electromagnetic Compatibility) • In order to provide the intended functionality, the system emits RF (Radio Frequency) energy while in operation. This may affect electrical devices in the vicinity.  • The system is equipped with a function for device identification via radio transmission, and is thus subject to electromagnetic interference from other equipment in the vicinity.    Limits of the technology  • Effect of the capsule endoscope on the body is unknown when it remains in the patient for over 30 days. • Even if a full 12 hours of images are recorded, there is no guarantee that the whole small intestine is photographed. • Since the capsule endoscope is a radio-transmitting device, it could interfere/ or experience interference with other radio-transmitting devices (e.g., telemetry, keyless entry devices). • For some patients, there is a possibility that the capsule endoscope may remain in the stomach for a longer period of time. Therefore, Olympus recommends using the real time viewer to confirm that the capsule endoscope has passed completely through the stomach. • A normal diagnosis does not exclude a clinically important abnormality in the small bowel.   (recorder unit, antenna unit)
   28   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                           Chapter 1 Checking the Package Contents  Match all items in the package with the components shown in this chapter. Inspect each item for damage. If the instrument is damaged, a component is missing or you have any questions, do not use the system; immediately contact Olympus.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            29  1.1 Capsule endoscope system MAJ-Y0136 (MAJ-Y0136)  Instruction manual Antenna Unit (X432262-21) Antenna harness (X432252-51)  Recorder unit harness (X432251-40)  Battery pack (X432262-11-12)   AC cable   AC adapter Recorder unit cradle         (X432262-31)    USB cable Capsule Activator   (MAJ-1478) Recorder unit (X432262-11)
   30   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          1.2 Capsule endoscope (EC-Y0005)
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            31  The product model of each item and its content are as follows:                                        Product name  Model  Comments Capsule  endoscope  system set MAJ-Y0136 MAJ-Y0136  Recorder unit, antenna unit, battery pack, recorder unit harness,   recorder unit cradle, antenna harness, AC adapter, AC cable ,USB cable, capsule activator Capsule endoscope  EC-Y0005  1 pc. Capsule activator  MAJ-1478  1 pc. Recorder unit  X432262-11  1 pc., with 1 battery pack Antenna unit  X432262-21  1 pc. Recorder unit cradle  X432262-31  1  pc.,  with  1  AC  adapter  and  1  AC cable   Antenna harness  X432252-51  1 pc. Recorder unit harness  X432251-40  1 pc.
   32   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Chapter 2 Nomenclature and Functions 2.1 System components  Capsule endoscope ( EC-Y0005)Antenna unit (X432262-21)Recorder unit (X432262-11)Antenna harness (X432252-51) Recorder unit harness(X432251-40) Battery pack(X432262-11-12) AC cableAC adapterRecorder unit cradle (X432262-31)  USB cableCapsule activator(MAJ-1478)Workstation (MAJ-Y0134) LCD monitorPrinterUSB HDDCapsule endoscope system set  MAJ-Y0136 (MAJ-Y0136)
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            33  1. Capsule endoscope (EC-Y0005) Takes endoscopic pictures of the gastrointestinal tract and transmits the image data by radio transmission.  2. Antenna unit (X432262-21) Receives transmissions from the capsule endoscope, and transfers the image data to the recorder unit.  3. Recorder unit (X432262-11) Records the image data transmitted by the capsule endoscope via the antenna unit.  4. Battery pack (X432262-11-12) The power supply for the recorder unit.  5. Recorder unit cradle (X432262-31) Holds the recorder unit while connecting to the workstation. Used for initialization of the recorder unit, registration of the patient ID, and downloading of image data.  6. Antenna harness (X432252-51) Holds the Antenna unit during the examination.  7. Recorder unit harness (X432251-40) Holds the recorder unit during the examination.  8. Capsule Activator (MAJ-1478) Holds the recorder unit during the examination.  9. USB cable Connects the recorder unit cradle to the workstation for transmission of patient ID, image data, etc.  10. Workstation (MAJ-Y0134) Registers patient information and patient IDs onto the recorder unit. Also used to download the endoscopic images from the recorder unit for observation, diagnosis, and reporting (refer to the “WORKSTATION MANUAL”).
   34   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         2.2 Capsule endoscope (EC-Y0005) Capsule endoscope (EC-Y0005)  1. Exterior package Exterior package of the capsule endoscope. 2. Top cover A transparent observation port covering the objective lens and illumination light emitting diode (LED). 3. Objective lens Lens for endoscopic observation. 4. Illumination LED Illuminates the area being observed. 5. Battery Internal battery for operating the capsule endoscope.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            35 2.3 Antenna unit (X432262-21)               1. Antenna sheet Receives signals from the capsule endoscope. 2. Antenna cable Transfers the received signals to the connection unit. 3. Connection unit Connects to the recorder unit. 4. Hole Use it for alining the antenna sheet.   5. Antenna unit eject button Slide the button located on the backside of the recorder unit to disconnect the antenna unit.        1. Antenna sheet 2. Antenna cable 3. Connection unit 5. Antenna unit eject button 4. hole
   36   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         2.4 Recorder unit (X432262-11)                                          1. Indicator lamp Indicates the reception/communication status with the capsule endoscope and the workstation.  2. Power lamp 1. Indicator lamp 3. Display panel 2. Power lamp 7. Power switch 4. Battery cover 5. Battery pack case 6. Antenna unit connector 8. Eject button 9.Hole
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            37 Illuminates green when the recorder unit is turned ON. 3. Display panel Displays patient ID, battery pack level, etc. 4. Battery cover Covers the battery pack and holds the battery pack in place. 5. Battery pack case Insert the battery pack here.   6. Antenna unit connector Attach the antenna unit here. Connect to the recorder unit cradle.   7. Power switch Turn the recorder unit ON by holding down the switch for 1 second or more. Turn the recorder unit OFF by holding down the switch for 2 seconds or more. 8. Battery cover eject button Push it to remove the battery cover 9. Hole Insert the nail of the battery pack into this hole when inserting the battery pack into the battery pack case.
   38   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         2.5 Recorder unit cradle (X432262-31)                            1. AC adapter Connects to an AC power supply. 2. USB cable Connects between the USB connector and the workstation. 3. Recorder unit connector Connect to the recorder unit.   4. Power connector Connect to the AC adapter. 5. USB connector Connect to USB cable. 6.Charge completion lamp Show the status that battery is charged completely.. 7. Charge lamp Show the status that battery is charging. 2. USB cable 3. Recorder unit connector 4. Power connector 5. USB connector  1. AC adapter 6.Charge completion lamp 7.Charge lamp
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            39 2.6 Recorder unit harness (X432251-40)                  1. Suspender Adjusts the length according to the patient’s physique. 2. Pouch Stores the recorder unit during the examination. 3. Pouch cover Secures the recorder unit and the antenna cables to the pouch.         1. Suspender 2. Pouch 3. Pouch cover
   40   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          2.7 Antenna harness (X432252-51)     1.Antenna pouch Housing the antenna sheet.. 2. Fastener Zip up to fix the antenna sheet.. 3. Alignment hole Adjust this hole to the umbilical position. 4. Belt ring Through a belt, and fix the hook and loop fastener.. 5. Hook and loop fastener Connect the both ends of the belt. 6. Belt coupler Please use it to connect the long belt and the short belt. 7. Short belt It fixes the antenna pouch to the patient’s body. If necessary, connect to the long belt. 8. Long belt It fixes the antenna pouch to the patient’s body. If necessary, connect to the short belt.   5. Hook and loop fastener 6. Belt coupler 7. Short belt 3.Alignment hole 1. Antenna pouch 4. Belt ring 8. Long belt 2. Fastener
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            41  2.8 Capsule activator (MAJ-1478)  1. Capsule endoscope insertion hole To turn ON the capsule endoscope, insert it into this hole. 2. Boss Projected portion that indicates direction of the activator. It is like a guide when inserting the activator.  2.9 Battery pack (X432262-11-12)     1. Terminal Connects to the connector in the battery case of the recorder unit.      2. Clawe Insert this nail into the hole of the recorder unit when inserting the battery pack into the battery pack case.     2.10 Workstation (MAJ-Y0134) For a diagram of the workstation configuration, please refer to the “WORKSTATION MANUAL”  1.Terminal 2.Clawe
   42   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Chapter 3 Installation and Connection  3.1 Preparation The following items are required for the capsule endoscope examination. Inspect them the day before the examination as described in this chapter.                            Table 3.1              Equipment  Quantity  Capsule endoscope  2 (Including 1 spare)  Antenna unit  1  Recorder unit  1  Battery pack  1  Recorder unit cradle  1  USB cable  1  Recorder unit harness 1  Workstation  1  Capsule activator  1  Antenna harness  1  Drinking water  As required  Gauze  As required  Ethyl or isopropyl alcohol As required
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            43   3.2 Exterior  Do not use if the equipment exterior is abnormal.It may result in failure of the examination.  Inspect the equipment as described in Table 3.2 below.                  Table 3.2  3.3 Capsule endoscope   Perform the following inspections before use. Should any irregularity be observed, do not use this system. Damage to or irregularity of this system may cause malfunction or patient injury.  1. Confirm that the expiration date has not passed. 2. Check the sterile package for ruptures, damage to the seal or water that might have entered the sterile package. 3. Before removing the capsule endoscope from the sterile package, inspect the capsule’s surface for cracks, deterioration, deformation, and other damage.    Equipment  Description Antenna unit    Check that the unit is free of cracks, deterioration, deformation, And other damage. Recorder unit  Check that the unit is free of cracks, deterioration, deformation, And other damage. Recorder unit cradle  Check that the unit is free of cracks, deterioration, deformation,  and other damage.   AC adapter  Check that the unit is free of cracks, deterioration, deformation, And other damage. Check the power supply cable for scratches and bends,. Check the connectors for cracks, deterioration,   deformation, and other damage. USB cable  Check the cable for scratches and bends. Check the connectors  for  cracks,  deterioration,  deformation,  and otherdamage. Battery pack  Check that the unit is free of cracks, deterioration, deformation,  and other damage.
   44   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                           3.4 Preparation of the recorder unit Attaching the battery pack 1.  Open the battery cover by pressing the battery cover eject button.              Figure 3.1                                            Figure 3.2   2.  Align the two claws on the battery pack with the hole in the battery pack case, and insert the battery pack to the battery pack case.               figure 3.3                                                              Figure 3.4                                                           Hole
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            45  3.  Close the battery cover by pressing it down over the battery pack until the battery cover is locked. (Refer Figure 3.5)                                      Figure 3.5    Don’t use the recorder unit if there are cracks, deterioration, deformation, and other damage. It may interfere with the examination.   • The battery pack is not charged when it is shipped. Charge it before use.      • To remove the battery pack, remove the battery cover for reference ‘1’ of “Attaching the Battery pack” of the previous page. And lift the edge of the battery pack which is surrounded with the dotted line of the figure below.                                                                                           Figure 3.6
   46   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          3.5 Preparation of the recorder unit cradle 1. Connect the AC adapter to the power connector.         Figure 3.7  2. Connect the USB cable to the USB connector.                Figure 3.8
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            47  3. Connect the AC cable to the AC Adapter.                    Figure 3.9 4. Connect the AC cable to the AC power supply. 5. Connect the USB cable to the USB connector on the workstation. Refer to “Workstation manual”.
   48   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Chapter 4 Operation  4.1 Preparation and Inspection before examination   Before starting an examination, be sure to read and clearly understand the information in this instruction manual.   Instruct the patient to stay away from equipment that generates strong electromagnetic fields (such as MRI equipment), between ingestion and excretion of the capsule endoscope. Otherwise, injury within the body cavity may occur. If the excretion of the capsule endoscope has not been confirmed 30 days after ingestion, attempt to locate the capsule endoscope within the patient’s body by X-ray examination.   Instruct the patient to consult their physician immediately if symptoms such as abdominal pain, discomfort, or nausea are experienced after ingesting the capsule endoscope. Injury to the patient may result if treatment for such symptoms is delayed.   • Instruct the patient not to turn the recorder unit’s power OFF,     push any buttons on the recorder unit, disconnect the   antenna unit, or remove an antenna harness during the examination.   Failure to follow the instructions may interfere with the examination. • During the capsule endoscope exam instruct the patient to stay away from generators of strong electromagnetic fields (such as MRI equipment, amateur (ham) radio, etc.), otherwise image recording may not occur due to radio interference.  • Please advise the patients not to use radio transmitting devices such as keyless entry devices (frequency range of possible interference: 306 - 322 MHz) or stay in an area where other persons may use them frequently, in order to reduce the possibility of dropped video frames caused by the use of these devices. • Since the capsule endoscope is a radio-transmitting device, it could interfere/ or experience interference with other radio-transmitting devices (e.g., telemetry, keyless entry devices).
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            49     • On rare occasions, interference may result in the need to repeat the capsule endoscope procedure. In this case, the physician should advise the patient to stay within the premises of the medical facility during the examination to prevent this problem from reoccurring. • The capsule endoscope contains a radio transmitter; therefore, the patient cannot fly in an airplane until the examination is over. •The life span of a battery pack is approximately 1 year from the date of purchase. When the battery pack approaches its end-of-life, the recorder unit will prompt for replacement. Order a new battery pack and replace.   Otherwise the recorder unit may not be able to operate for 12 hours.   Exterior Inspect the equipment as described in Table 4.1 below.                            Equipment  Description Antenna unit    Check  that  the  Antenna  unit  is  free  of  cracks, deterioration, deformation, and other damage. Recorder unit  Check  that  the  recorder  unit  is  free  of  cracks, deterioration, deformation, and other damage. Recorder unit cradle  Check  that  the  recorder unit cradle  is  free  of cracks, deterioration, deformation, and other damage.   AC adapter  Check  that  the  AC  adapter  is  free  of  cracks, deterioration, deformation, and other damage. Check the  power  supply  cable  for  scratches  and  bends,. Check  the  connectors  for  cracks,  deterioration, deformation, and other damage. USB cable  Check the cable for scratches and bends. Check the connectors for cracks, deterioration, deformation, and other damage. Antenna harness  Check  that  the  Antenna  harness  is  free  of  rips, deterioration, and other damage. Recorder unit harness  Check  that  the  recorder  unit  harness  is  free  of  rips, deterioration, and other damage.
   50   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Recharging  Fully recharge the battery packs to prevent low battery charge, which could cause the examination to end prematurely. In preparation for the examination, recharge the battery packs on the previous day.  The battery pack can not be recharged at temperatures below 0℃.   1.  Check that the recorder unit is off and the display panel is blank.  2.  Suppress the recorder unit cradle by hand, and insert the recorder unit to the recorder unit cradle as shown in Figure 4.1. The recorder unit is turned on, and displays starting screen as shown in Figure 4.2 on the display panel.                            Figure 4.1                         Figure 4.2  When the antenna unit is connected to the recorder unit, the recorder unit cannot be connected to the recorder unit cradle. Remove the antenna unit from the recorder unit to insert the recorder unit to the recorder unit cradle. Starting screen
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            51  3. Wait a few seconds, and check that the charge lamp illuminates and Battery charge screen as shown in Figure 4.4 is displayed on the display panel. The recharging time is approximately 2 hours.                   Figure 4.3                                              Figure 4.4  4. When recharging is complete, the charge lamp goes off, the charge completion lamp illuminates and Battery full screen as shown in Figure 4.6 is displayed on the display panel.                                                                                                                                        Figure 4.5                                                                                            Figure 4.6                                                       Battery charging   screen Charge lamp Charge completion lamp Battery full   screen
   52   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Initial setup of the recorder unit Before initializing, to complete the connection and setup of the Workstation, refer to the “WORKSTATION MANUAL”.  The recorder unit is connected to the workstation via the recorder unit cradle.  While setting up the recorder unit, do not remove the recorder unit from the recorder unit cradle or remove the USB cable and/or AC adapter. Doing so may result in failure of the examination. If the recorder unit is taken off the recorder unit cradle, or the USB cable and/or AC adapter is disconnected, reconnect the recorder unit to the workstation by placing it on the recorder unit cradle or by reattaching the USB cable and AC adapter. Restart the workstation before setting up the recorder unit again.      1. The workstation is used to initialize the recorder unit and to enter patient information. For more information on the initialization procedure and entering patient information, refer to the “WORKSTATION MANUAL”.   During the initial setup, recorder unit is displayed Initializing screen as shown in Figure 4.7 on the display panel, and the recorder unit’s indicator lamp illuminates orange.                      Figure 4.7
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            53  2. When initialization is completed, patient ID and the patient’s name are displayed on the Battery full screen as shown in Figure 4.8 on the display panel and the recorder unit’s indicator lamp change green.                  Figure 4.8   3. While holding the recorder unit cradle with one hand, lift the recorder unit up with your other hand to remove it from the recorder unit cradle. The recorder unit is turned OFF automatically.   None
   54   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          4.2 Attaching the equipment to the patient  Patient confirmation Before attaching the antenna unit to the patient, confirm the patient’s identity. Also confirm that the patient has not eaten or drunk for at least 8 hours before the examination.    Inserting the antenna sheet  •When inserting the antenna sheet to the antenna harness, take care not to bend the antenna sheet. Otherwise, receiving sensitivity of the antenna sheet may decrease so that noise may appear in the images or the images may not be transmitted properly.     • Do not reverse the antenna sheet upside down, left to right, and inside out when inserting it to the antenna harness. If antenna sheet is inserted to antenna harness improperly, the recorder unit may not be able to receive accurate data from the capsule endoscope. • In case of accidental drop of the recorder unit, push the power button to restart the recorder unit, and confirm if the error message is displayed or not.    The examination can be continued unless the error message is displayed.  1. Open the fastener of the antenna harness and insert the antenna sheet in the antenna       harness as shown in the figure 4.9. 2. Zip up the fastener of the antenna harness.                                                                                          Figure 4.9   Please insert the antenna unit in the antenna harness so that the brown aspect of the antenna unit and the label of the antenna harness are the same side.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            55 Preparation of antenna unit harness Select the appropriate belt to fit the body form of the patient. If it is necessary, connect two belts in the following procedures. (Refer to figure 4.10, figure 4.11) 1.  Turn down a belt in a belt coupler of the short belt. And close a hook and loop fastener.                                           Figure 4.10                                                      Figure 4.11  2.     Put the end of the long belt through the belt coupler of the short belt like figure 4.12.                                            Figure 4.12  3.  Close the hook and loop fastener of the long belt. (Refer to figure 4.13)  Figure 4.13
   56   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Attaching the antenna unit   • Refer to the following information to attach the antenna unit to the patent’s body.   Otherwise, noise may appear in the images or the images may not be transmitted properly. • Allow only a thin clothing of natural materials to be placed between the body and the antenna harness. • Do not tuck metals, such as buckle of belt, between the antenna unit and surfaces of the body.    Otherwise, noise may appear in the images or the images may not be transmitted properly. • Attach the antenna unit and the recorder unit so that antenna cable is not caught on things such as a knob of a door. Otherwise it may damage the antenna unit and/or the recorder unit and may result in failure of the examination. • Attach the antenna unit and the recorder unit so that antenna cable is not laid on or near the antenna sheet. It may prevent the reception of images so that noise may appear in the images or the images may not be transmitted properly.  1.  Instruct the patient to wear only a thin piece of clothing around abdominal area.     2.  Put the belt through the belt ring. And fix the belt to the belt ring of the porch with a hook and loop fastener. Set a porch to the abdomen so that a mesh side becomes the patient side and so that an antenna cable comes to the left side of the patient’s body.        Figure 4.14
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            57    3. Let the other end of the belt into a belt ring. Pull out the both ends of the belt in right and left to attach antenna harness to the patient. (Refer to figure 4.16)                               Figure 4.15                                                            Figure 4.16  4. Adjust the position of the antenna harness so that the alignment hole of the antenna harness to the umbilical position of the patient.(Refer to figure 4.17)                                                                                                                      Figure 4.17  5.    Reattach the antenna harness in the case such as follows when the patient walks and/or sits down. - A patient feels unpleasantness. - The antenna harness move or is separate from the patient’s body  ・  Tighten the belt so that the antenna harness is not separate from the surface of the patient’s body, and so that the antenna harness is believed not to move from the attached position. Otherwise, noise may appear in the images, or the images may not be transmitted properly.
   58   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                            ・Instruct the patient to be careful so that the antenna harness does not move from the attached position and so that the antenna harness is not separated from the surface of the patient’s body. Otherwise, noise may appear in the images, or the images may not be transmitted properly. ・  Instruct the patient to be careful so that the antenna cable is not caught on things such as knob of doors.   Instruct the patient to confirm that the position of the antenna harness is same as it’s attached position after going to a restroom.       Attaching the recorder unit harness  Attach the recorder unit harness as shown in the figure 4.18.     Figure 4.18
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            59 Connecting the recorder unit and the antenna unit    1. Connect the antenna unit to the recorder unit by inserting the connection unit into the recorder unit until they click (Refer to Figure 4.19).                                                                       Figure 4.19    2. Hold down the recorder unit’s power switch for at least 1 second to turn ON the recorder unit (Refer to 4.20). The starting screen is displayed as shown in Figure 4.21 on the display panel.  3. After the starting screen is displayed, The ID view screen is displayed as shown in Figure on the display panel. Check that battery icon is green. (Refer to Figure4.21, Figure 4.22)                                                                             Figure 4.20
   60   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                                                                  Figure 4.21      4. Confirm the patient’s identity with the patient ID and the patient’s name displayed on the               recorder unit (Refer to Fig 4.22).                                                                          Figure 4.22          Recording time Patient Name Patient ID ID View screen Battery icon Starting screen
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            61    • If the Battery icon is orange, the battery may run out before 12 hour. Recharge the battery pack, or replace with a fully-charged battery pack.      • Start the examination immediately after turning the recorder unit ON. Failure to do so may result in premature termination of the examination due to an empty or low battery.  • If any of error messages are appeared on the display panel, refer to “Troubleshooting”.  • The recording time is displayed at the lower left of the display panel.    Recording time is displayed when data are remained without being downloaded. Download such data and initialize the recorder unit if necessary.
   62   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          IInserting the Recorder Unit in the Recorder Unit Harness  • Attach the antenna unit and the recorder unit so that antenna cable is not caught on things such as a knob of a door. Otherwise it may damage the antenna unit and/or the recorder unit and may result in failure of the examination. • Attach the antenna unit and the recorder unit so that antenna cable is not laid on or near the antenna sheet. It may prevent the proper reception of images.  1.  Insert the recorder unit into the pouch of recorder unit harness and close the pouch cover. ( Refer to Figure 4.23)    Please attach them so that an antenna cable is between the body of the patient and the suspenders of the recorder unit.                                                                           Figure 4.23   4.3 Preparing the capsule endoscope   • Do not use capsule endoscopes that have been dropped, bitten, or been subjected to excessive pressure. Doing so may result in an infection control risk as well as cause injury to the patient due to damage to the capsule endoscope.   • Inspect the sterile package before you open it. If it is already opened or damaged, the sterility of the capsule endoscope may have been compromised. Use a new capsule endoscope instead. • To prevent the capsule endoscope battery from deteriorating, store the capsule endoscope under a temperature of 0 – 25°C (32 – 77°F). Do not store in a refrigerator or a freezer, as condensation inside of the top cover may result.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            63  Turning the power ON  1. Check the expiration date. 2. Slowly remove the sealing paper and take care that the capsule endoscope does not fall out of the container.   Figure 4.24 3. Hold the sterile container, taking care not to squeeze it. Adjust the direction of the activator’s boss to the closest groove to the handle of the inner lid of the sterile container. Insert the capsule activator straight along with the groove. The capsule endoscope will turn ON (see Figure 4.25). When the capsule endoscope is turned ON, the LED illuminates.                                                    Figure 4.25
   64   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                            • The capsule activator is not sterilized. Do not allow the capsule endoscope to touch the activator. • Keep the capsule endoscope away from DC magnetic fields such as magnets. Magnets can turn the capsule endoscope ON, resulting in the consumption of battery power. • Do not look directly at the capsule endoscope’s LED for a prolonged duration. You may feel dizzy due to the afterimage.      Inserting the capsule activator again will turn the capsule endoscope OFF.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            65   Confirming Operation  Confirm the operation of the capsule endoscope according to the following steps.       1. Bring the capsule endoscope in the sterile container close to the antenna sheet (the patient’s abdomen).      2. Confirm that the receiving lamp of the recorder unit blinks green.
   66   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                            Confirming image You can confirm if the image transmitted from capsule endoscope by Image View screen. 1. Push the “MENU” button to display the Menu screen. 2.Choose  the  “Image  View”  by pushing  “Cursor”  button,  and  push  “OK”  button.(Refer  to Figure 4.26)   The capsule endoscope image is displayed on the Image View screen. (Refer to Figure 4.27)   Fig 4.26   Fig 4.27 Patient ID Patient NAME Battery icon Present time  The capsule endoscope image Average color bar: Display the average color of the image data. “Image View” “MENU” button  “OK” button Cursor button     “”, “”, “⊳” ,   “” MENU screen Image View screen
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            67     When the ID View screen with Image View icon (Refer to Figure 4.28) is displayed, Image View screen can be displayed with the pushing “OK” button.                                            Figure 4.28      •  When  approximately  two  minutes  pass,  the  display  panel  of  the recorder unit turn off.  • When “MENU” button is pushed, the MENU screen is displayed.  • When “Back” button is pushed, the previous screen is displayed.    • You can capture a displaying image with the pushing the “OK” button. Refer to the “Capture of the image”.    •  Do  not  touch  the  display  panel  with  Image  View  screen.  You  may  not operate the recorder unit normally.  ID View screen ID View icon
   68   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Confirming the recorded image  You can display recorded images on Playback mode of the Image View screen. 1. Push the “MENU” button to display the Menu screen. Choose the “Image View” by pushing “Cursor” button, and push “OK” button. The capsule endoscope image is displayed by Image View screen. 2. Push the ”⊳” on Image View screen shown below. (See Figure 4.29)   As shown figure 4.30, a stored image is displayed by playback screen.      Fig 4.29  3. You can confirm the image after one piece by pushing ”⊳” one piece ago by pushing ””.    Fig4.30  Cursor button     “”, “”, “⊳” ,    “” Image View screen The position of the image displayed nowin the average color bar. The  time  when  the image is recorded   Present time Image View screen   (Playback mode) Expansion area
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            69    • Expansion and the reduction of the average color bar are possible by pushing “”or ””.   The average color bar can display three following conditions.   ―    All mode displays the average color bar of all images which are stored. ―    1hour  mode  displays  30  minutes  before  and  after  an  image  chosen. (Displays average color bar of 1 hour.) ―    10 minutes mode displays 5 minutes before and after an image chosen. (Displays average color bar of 10 minutes.)  • You can choose 30 minutes before and after image by pushing the “⊳” or“” in all mode.  • You can choose 5 minutes before and after image by pushing in the“⊳” or“” in 1 hour mode.  • You can choose 5 minutes before and after image by pushing in the“⊳” or“” in 10 minutes mode.  •You  can  rewind  or  fast  forward  when  keep  pushing  in  the“⊳”  or“”  in  10 minutes mode.  • “[]” shows the spreading range in average color bar.  • When approximately two minutes pass, the display panel of the recorder unit turn off.  • When “MENU” button is pushed, the MENU screen is displayed.  • When “Back” button is pushed, the previous screen is displayed.  • You  can  capture  a  displaying  image  with  the  pushing  the  “OK”  button. Refer to the “Capture of the image”.   • Do not touch the display panel with Image View screen. You may not operate the recorder unit normally.
   70   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Capture of the image During Image View screen is displayed, you can capture the image shown on the screen. 1. Push “OK” button when the image to be captured is displayed. Such screen as in the Figure 4.31 is displayed. Set icon and Cancel icon are displayed in the screen.      Figure 4.31  2.  Push  “OK”  button  to  capture  images.  When  you  do  not  capture  it,  push  “BACK” button. After operation, the recorder unit displays Image View screen.     •  The  display  panel  of  the  recorder  unit  turns  off  approximately  two minutes after the last operation.  ・When “MENU” button is pushed, the MENU screen is displayed.  ・When “Back” button is pushed, the previous screen is displayed.  ・Max 15 images are captured. Selection image time   number of images already captured / max number of images to be captured   A position of the captured image on average color bar   “OK” button Set icon Cancel icon
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            71    ・When number of the capture image exceeds 15,    Figure 4.32 is displayed. Choose  “Capture  cancel”  to  stop  capturing  image  or  “Erase others”  to  capture more image by ”⊳”,”” and “OK” button.  ・When  you  choose  “Erase  others”,  Capture  list  is  displayed.  Please  refer  to cancellation method “capture for the later operation.”  Figure 4.32   • Do not touch the display panel with Image View screen. You may not operate the recorder unit normally.
   72   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Confirming capture image You can confirm the image which is captured by following operation from Capture list. 1. Push “MENU” button to display Menu screen.  2. Choose “Capture View” and push “OK” button on MENU screen.  Figure 4.33   3. The images which are captured are displayed. The chosen image is surrounded in a blue line.(Refer to Figure 4.34) Choose an image using Cursor button.   Figure 4.34  “Capture View” “MENU” button  “OK” button Cursor button     “”, “”, “⊳” ,   “” Capture List Selection image
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            73  4.  The  image  which  you  choose  is  displayed  when  push  “OK”  button.  (Refer  to Figure4.35) You can display previous and following images by pushing a “⊳”, ””.    Figure 4.35   ・The  display  panel  of  the  recorder  unit  turn  off  approximately  two minutes after the last operation.  ・When “MENU” button is pushed, the MENU screen is displayed.  ・When “Back” button is pushed, the previous screen is displayed.  ・  Image capture is cancelled by pushing with screen ”” on the screen of the Figure 4.35. Please refer to “Cancellation of the capture”.  ・  You  can  view  the  image  which is  captured  by  the  recorder  unit  on Workstation.     number of the capture image/ capture possibility number of sheets
   74   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Cancellation of the captured image By the following operation, cancellation of the capture is possible. 1. Choose an image to erase from Capture list and push "OK" button. 2. Push ”” when the captured image which you choose is displayed on the screen.  Fig 4.36  3. Choose "Yes" and push "OK" button on the following screen. (Refer to Figure 4.37)  Figure 4.37       “MENU” button  “OK” button Cursor button     “”, “”, “⊳” ,    “”
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            75    ・The  display  panel  of  the  recorder  unit  turn  off  approximately  two minutes after the last operation.  ・When “MENU” button is pushed, the MENU screen is displayed.  ・When “Back” button is pushed, the previous screen is displayed.
   76   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         Confirming the guidance  The  recorder  unit  gives  guidance  to  a  patient  during  the  examination  by  registering patient and examination data in Workstation. By the following operation, guidance to a patient is displayed.   1. Push “MENU” button to display MENU screen.  2. Choose the “Guidance List” by the Cursor button, and push “OK” button. (Refer to Figure 4.38)                         Figure 4.38       “Guidance List” “MENU” button  “OK” button Cursor button     “”, “”, “⊳” ,   “”
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            77  3. Guidance List is displayed. (Refer to Figure 4.39)  Please call a nurse.You can take a light meal.Please come back to the procedure room.09:0010:0012:0016:00 Please Come back to the hPlease come back to the pPlease call a nurse.You can take a light meal.Please come back to the procedure room.09:0010:0012:0016:00 Please Come back to the hPlease come back to the pPlease call a nurse.You can take a light meal.Please come back to the procedure room.09:0009:0010:0010:0012:0016:00 Please Come back to the hPlease come back to the p Figure 4.39   ・If the guidance has already been given, the icon shown beside the time green from white. ・The latest guidance is displayed in the lower part of the screen.  (Refer to Figure 4.39)   ・When “MENU” button is pushed, the MENU screen is displayed.  ・When “Back” button is pushed, the previous screen is displayed.  ・The  display  panel  of  the  recorder  unit  turn  off  approximately  two minutes after the last operation.  The latest guidance
   78   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Handling the recorder unit when guidance is given    ・Instuct  the  patient  not  to  operate  the  recorder  unit  other  than  guidance occurs and to stop beep and vibration. ・Instruct  a  flow  of  the  examination to the  patient  to  patient  cope  even  if  a patient does not notice guidance.   1. Instruct  a patient  how  to  stop  beep  and  vibration  of guidance by pushing “OK”  button when guidance is given.  2.  ID  View  screen  is  displayed  when  guidance  is  given.  Instruct  patient  to  confirm  the guidance after stopping beep and vibration.     ・The beep and vibration of the guidance stop automatically 30 seconds later.  ・The latest guidance information is displayed on the ID View screen. (Refer to figure 4.40)   01[h] 00[min]Please come back to the procedure room.01[h] 00[min]01[h] 00[min]Please come back to the procedure room. Figure 4.40
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            79  Confirmation of the memo You can confirm the memo which is inputted in the recorder unit. 1.   Push the “MENU” button. 2. Choose “Memo” by the Cursor button and push “OK” button. (Refer to Figure 4.41)  Figure 4.41  3.  Such screen as figure 4.42 is displayed so that you can read the memo.  Figure 4.42  “MEMO” “MENU” button  “OK” button Cursor button     “”, “”, “⊳” ,   “”
   80   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                              ・The  display  panel  of  the  recorder  unit  turn  off  approximately  two minutes after the last operation.  ・When “MENU” button is pushed, the MENU screen is displayed. ・When “Back” button is pushed, the previous screen is displayed.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            81  Editing of the memo  You can edit the memo into a recorder unit.  1. Choose the inputting position on the Memo screen by Cursor button. 2. Input a letter by touching it with a keyboard displayed on the screen.   Figure 4.43 Button  Function   Press to move cursor up in input.   Press to move cursor down in input. ⊳  Press to move cursor left in input.   Press to move cursor right in input. OK  Save and Exit to Menu screen. BACK  Exit to Menu screen, Memo not saved. MENU  Exit to Menu screen, Memo not saved.     ・When “MENU” button is pushed, the MENU screen is displayed.  ・You can confirm the memo which is inputted with the recorder unit only. You can not confirm it in Workstation.
   82   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          4.4 Starting the examination  Even when the guidance function is used, explain the patient how the examination is performed, so that the examination can proceed if the patient fails to be aware of the guidance.       Ingesting capsule endoscope      ・The capsule endoscope should only be ingested with water. Use of other liquids may interfere with the examination. ・Avoid touching the top cover (LED side) of the capsule endoscope. Do not wipe with gauze or other cloths.   1. Prepare a glass of water  2. Remove the inner case, and hand it to the patient.  3. Confirm that the capsule endoscope’s LED is blinking.  4. Instruct the patient to take the capsule endoscope out of the inner case without touching the top cover.  5. Instruct the patient to carefully ingest the capsule endoscope with a sip of water.   Confirming Passage    1. If necessary, you can confirm on the Image view screen that the capsule endoscope has passed through the esophagus and entered the stomach. To display the Image View screen, follow the instruction of “Confirming image.”  2. If necessary, you can confirm on the Image view screen that the capsule endoscope has passed through the stomach and entered the duodenum by using the recorder unit If it takes too long for the capsule endoscope to pass to the duodenum, the battery may run out before the capsule goes through the entire small intestine.      The passage of the capsule endoscope can be confirmed on the Image View screen. To display the Image View screen, follow the instruction of “Confirming image.”
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            83   4.5 During the examination  Familiarize the patient with the following warnings and cautions.   Instruct the patient to stay away from equipment that generates strong electromagnetic fields (such as MRI equipment), between ingestion and excretion of the capsule endoscope. Otherwise, injury within the body cavity may occur. If the excretion of the capsule endoscope has not been confirmed within 2 weeks from the time of ingestion, attempt to locate the capsule endoscope within the patient’s body by X-ray examination.   Instruct the patient to consult their physician immediately if symptoms such as abdominal pain, discomfort, or nausea are experienced after ingesting the capsule endoscope. If treatment of such symptoms is delayed, patient injury may result.    • Patients with allergies may experience reddening or irritation of the skin caused by the adhesive on the antenna unit covers. • Instruct the patient not to turn the recorder unit’s power OFF,     to push any buttons on the recorder unit, disconnect the   antenna unit, or remove an antenna harness during the examination.   Failure to follow the instructions may interfere with the examination. • During the capsule endoscope exam instruct the patient to stay away from generators of strong electromagnetic fields (such as MRI equipment, amateur (ham) radio, etc.) otherwise image recording may not occur due to radio interference. • Please advise the patients not to use radio transmitting devices such as keyless entry devices (frequency range of possible interference: 306 - 322 MHz) or stay in an area where other persons may use them frequently, in order to reduce the possibility of dropped video frames caused by the use of these devices.
   84   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          • Since the capsule endoscope is a radio-transmitting device, it could interfere/ or experience interference with other radio-transmitting devices (e.g., telemetry, keyless entry devices). • On rare occasions, interference may result in the need to repeat the capsule endoscope procedure. In this case, the physician should advise the patient to stay within the premises of the medical facility during the examination to prevent this problem from reoccurring. • The capsule endoscope contains a radio transmitter; therefore, the patient cannot fly in an airplane until the examination is over. • Do not make a diagnosis using images displayed on the Image View screen. •  Instruct  the  patient  not  to  operate  the  recorder  unit  other  than  guidance         occurs and to stop beep and vibration. • Instruct the patient to be careful so that the antenna cable is not caught on a knob and the lever of the door. •  Instruct  a  flow  of  the  examination  to  the  patient  to  patient  cope  even  if  a patient does not notice guidance. • Instruct the patient to contact the physician or medical personnel, when the patient dropped the recorder unit during the examination. After having received the recorder unit, physician or medical personnel push the power button to restart the recorder unit, and confirm the error message is not displayed. If error message is not displayed, please continue the examination. • Please attach the antenna unit to the body firmly, and set the alignment hole of the antenna harness to the umbilical position. Otherwise, noise may appear in the images, or the images may not be transmitted properly.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            85  4.6 Ending the examination Removing equipment  1. About 8 hours after the start of the examination, check that the recorder unit’s indicator lamp has stopped blinking, and end the examination. If the indicator lamp is blinking, confirm the image view screen and determine whether to continue or stop the examination.  2. Take the recorder unit out of the pouch of the recorder unit harness.    In order to prevent damage to the antenna unit and/or the recorder unit, be sure to hold the connection unit when disconnecting the antenna unit. Do not disconnect the antenna unit while holding only the antenna cable.
   86   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                           3. Hold down the recorder unit’s power switch for 2 seconds or more to turn OFF the recorder unit (see Figure4.44), then disconnect the antenna unit. Slide the antenna unit eject button on the backside of the recorder unit, and disconnect the connection unit by sliding it in the direction of the arrow (Refer to Figure4.45).                 Figure 4.44                   Figure 4.45                                                                           4. Remove the recorder unit harness from the patient. 5. Remove the antenna unit from the patient.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            87  Downloading image data Download the image data from the recorder unit to the workstation according to the following procedure:   • While downloading, do not remove the recorder unit from the recorder unit cradle or turn OFF the workstation. Doing so may result in loss of image data. • While downloading, do not unplug the power cable. Doing so may result in loss of image data. • While downloading, do not turn OFF the recorder unit. Doing so will interrupt the download, and require you to start the download again from the beginning. • While downloading, do not remove the recorder unit from the recorder cradle or remove the USB cable and/or AC adapter. Doing so may result in loss of image data. If the recorder unit or cradle are disconnected from the workstation, ensure that the USB cable and AC adapter is connected to the workstation and reconnect the recorder unit to the workstation through the recorder unit cradle. Restart the workstation before downloading again.   1. Check that the recorder unit cradle is connected to the workstation using the USB cable.  2.Turn the workstation ON and log in, as described in “WORKSTATION MANUAL”.  3. Insert the recorder unit into the recorder unit cradle. The recorder unit will be turned ON   automatically.
   88   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                           4. The patient’s ID and name are displayed on the recorder unit’s display panel.                                     Figure 4.46   If an error is detected on the recorder unit, the indicator lamp will blink yellow. To troubleshoot, refer to Chapter 6, “Troubleshooting”.  5. The workstation is used to download image data. For specific information on downloading image data, refer to the “WORKSTATION MANUAL”.   • While downloading the image data, the indicator lamp on the recorder unit turns from green to yellow and blinks once every second. The recorder unit will be turned OFF when the download is complete. • While downloading the image data, the icon in Figure 4.47 is displayed on the recorder unit’s display panel.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            89  Figure 4.47
   90   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Confirming excretion of the capsule endoscope Familiarize the patient with the following danger.   Instruct the patient to stay away from equipment that generates strong electromagnetic fields (such as MRI equipment), between ingestion and excretion of the capsule endoscope. Otherwise, injury within the body cavity may occur. If the excretion of the capsule endoscope has not been confirmed within 2 weeks after ingestion, attempt to locate the capsule endoscope within the patient’s body by X-ray examination.  1. The capsule endoscope will be excreted with the patient’s feces. 2. Contact the patient to ensure that the patient observes the capsule endoscope in his/her feces.  Cleaning and storing the equipment For instructions on cleaning and storing the equipment, refer to Chapter 4, “Care, Storage and Disposal”.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            91    Chapter 5 Care, Storage and Disposal 5.1 Care After using the system, immediately perform the following cleaning procedure.   • After wiping with a piece of moistened gauze, dry this system thoroughly before using it again. If it is used while still wet, there is the risk of an electric shock. • During cleaning, wear appropriate personal protective equipment, such as eye wear, face mask, moisture-resistant clothing and chemical-resistant gloves that fit properly and are long enough so that the operator’s skin is not exposed. Blood, mucous and other potentially infectious material adhering to this system could pose an infection control risk.   • Do not soak the system components in water, or sterilize them using autoclave or gas. Doing so will damage the system. • Do not wipe the external surfaces with hard or abrasive wiping material. Doing so will scratch the surface. • Do not wipe the connector of the recorder unit cradle with wet gauze. A terminal becomes short and defective, and the charging the   recorder unit and downloading the data may not be possible.  1. Remove dust, dirt, and other stains on the surface of the antenna unit   using a piece of gauze moistened with a neutral detergent, then wipe the   surfaces using a piece of gauze moistened with70% ethyl or isopropyl alcohol.  2. Remove dust, dirt, and other stains on the surface of the following equipment using a piece of gauze moistened with a neutral detergent, then wipe the surfaces using a piece of gauze moistened with 70% ethyl or isopropyl alcohol. After doing so, dry them thoroughly.  • Recorder unit     • Recorder unit cradle • USB cable      • Antenna harness • Recorder unit harness    • Capsule activator   Do not soak the equipment in any liquid. It may damage the equipment.
   92   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                             ・Do not iron or dry clean and do not use a bleach for the recorder unit harness and antenna harness. Doing so will damage the surface.   Note that machine-washing may cause the recorder unit harness to shrink slightly. 3. If the recorder unit harness becomes wet, be sure to dry it completely.  5.2 Storage After cleaning the equipment according to the procedure given in Section 4.1, “Care”, be sure to dry it off thoroughly. When the equipment is dry, place it into its designated positions in the carrying cases for storage. Store the carrying cases in a clean and dry location. For details on storage environment, refer to “Transportation, storage, and operation environment” on page 99 in the Appendix.   • To prevent the capsule endoscope battery from deteriorating, store the capsule endoscope within the temperature range of 0 – 25°C (32 – 77°F). Do not store the capsule endoscope in a refrigerator or a freezer. ・Avoid long term use or storage at temperatures exceeding 40°C (104°F). Battery pack performance may deteriorate rapidly under these conditions. • Keep water and any other liquid away from this system and components. It may cause damage to the system and components. • Do not store the system in a location exposed to direct sunlight, UV rays, X-rays, radio activity, or strong electromagnetic radiation (e.g. near microwave therapy equipment, shortwave therapy equipment, MRI equipment, radios, or mobile phones). Doing so may damage the system. • Do not store the system in a location where it may be subjected to impacts or excessive vibrations. Doing so may damage the system. • Do not store the capsule endoscope in a location where it may be accessed and/or accidentally swallowed by children.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            93 5.3 Disposal When disposing of any item, follow all applicable national and local laws and guidelines. The capsule endoscope is excreted naturally through the patient’s feces.
   94   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Chapter 6 Troubleshooting  If, during the inspections described in Chapter3, “Installation and connection” and Chapter4 “Operation, the system or any of its components appears to have problems,refer to Section 6.1, “Troubleshooting” to correct the problem.If the problem persists, stop using the system, and refer to Section 6.2,“Returning the system for repair”    to correct the situation. If the problem cannot be resolved by these countermeasures, contact Olympus.   Never use the system if any abnormality is observed. Damage to the system or irregularities with the system not only cause equipment malfunction, but may also cause injury to the patient’s body cavity.  6.1 Troubleshooting  The following table lists the possible causes of and countermeasures for problems that may occur due to equipment setting errors or to the deterioration of replaceable parts. For more information, refer to “Recorder unit error messages”. Problems or failures other than those listed in the following table require repair. Since repairs performed by persons who are not qualified by Olympus could cause further damage to the equipment and result in operator and/or user injury, be sure to contact Olympus for repair (refer to Section 6.2).  Irregularity  Possible cause  Solution The power is not turned ON. Use the capsule activator to turn the power ON. Capsule endoscope’s LED fails to illuminate.  The capsule endoscope is broken. Use a new capsule endoscope. The battery pack is missing. Insert a charged battery pack. Recorder unit’s power fails to come ON. The battery pack is low.  Recharge the battery pack. Recorder unit’s power lamp fails to illuminate. The recorder unit is broken. Try turning on the recorder unit power again. If the problem persists, contact Olympus.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            95  Irregularity  Possible cause  Solution The recorder unit is broken. Contact Olympus. The capsule endoscope is broken. Use a new capsule endoscope. The antenna cable is broken. Contact Olympus. The antenna unit is not connected to the recorder unit. Connect the antenna unit to the recorder unit properly. Images from the capsule endoscope are not displayed by the Image View screen.  The antenna unit is too far from the capsule endoscope. Reduce the distance between the antenna and the capsule endoscope. The recorder unit fails to turn OFF. The power switch is not held down long enough. Hold down the power switch for 2 seconds or more. Display layout is out of place. The PC’s DPI setting is not set to 96DPI or 120DPI. Right-click on the desktop and select [Properties]. Select [Settings]. Click on the [Advanced] button. Set [DPI setting] to either [96DPI] or [120DPI].   You can not operate the recorder unit by pushing any button. Troubles occur in software of the recorder unit. Remove the battery pack from the recorder unit, and attach it again. Then turn ON the recorder unit by pushing the power button. USB cable or AC adapter is not connected normally. Connect USB cable and AC adapter once again. WS does not recognize the recorder unit The USB cable or AC adapter   is broken. Contact Olympus
   96   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Recorder unit error messages  Error code on recorder Unit’s display panel Possible cause  Solution   System error  1. Restart the system. 2. If the error persists, contact Olympus.   Internal memory error  Contact Olympus.   RTC error    1. Connect the recorder unit to the workstation, and perform initialization. 2. If error persists after initialization, contact Olympus.   Communication error with  the battery pack 1.  Replace the battery pack again. 2.  Replace with a new battery pack.   Internal memory full  Connect the recorder unit to the workstation, and download the data from the recorder unit.   Uncharged battery pack Recharge the battery pack, or replace it with a charged battery pack.   Patient Information not registered Connect the recorder unit to the workstation, and perform initialization.   Low battery pack  Recharge the battery pack.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            97    Error code on recorder Unit’s display panel Possible cause  Solution   Time to replace battery pack∗1 Order a new battery pack from Olympus, and replace it after the new battery pack arrives. The recorder unit may not be able to operate for 12 hours      Antenna unit disconnected Connect the antenna unit to the recorder unit.  ∗1 The expiration of a battery pack is approximately 1 year from the date of purchase When the battery pack approaches its end-of-life, the recorder unit will prompt for replacement. Order a new battery pack and replace. The battery pack is configured to provide more than 12 continuous hours of power. Recorder unit does not function with an expired battery pack.
   98   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          6.2 Returning the system for repair  Olympus is not liable for any injury or damage occurring as a result of repairs attempted by non-Olympus personnel.  Before sending the system for repair, contact Olympus. With the system, include a description of malfunction or damage and the name and telephone number of the individual at your location who is most familiar with the problem.   Refer to Labels Section in Chapter 1 to easily locate the model name and serial number when sending equipment to Olympus for repair. To purchase accessories and consumables, contact Olympus.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            99  Appendix Transportation, storage, and operation environment  Capsule endoscope  Ambient temperature 0 – 25°C (32 – 77°F)∗1 Relative humidity  30 – 85%∗2 Transportation and storage environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) Ambient temperature 10 – 40°C (50 – 104°F) Relative humidity  N/A Operation environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) ∗1 Storage outside the storage environment range may cause condensation and/or deterioration of the battery. ∗2 No condensation (inside the capsule endoscope)  Antenna unit  Ambient temperature –20 to +70°C (–4 to +158°F)  Relative humidity  10 – 95%∗1 Transportation and storage environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) Ambient temperature 0 – 50°C (32 – 122°F) Relative humidity  30 – 90%∗1 Operation environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) ∗1 No condensation
   100   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Recorder unit  Ambient temperature –20 to +70°C (–4 to +158°F)  Relative humidity  10 – 95%∗1 Transportation and storage environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) Ambient temperature 0 – 50°C (32 – 122°F) Relative humidity  30 – 90%∗1 Operation environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) ∗1 No condensation  Recorder unit cradle  Ambient temperature –20 to +70°C (–4 to +158°F) Relative humidity  10 – 95%∗1 Transportation and storage environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) Ambient temperature 10 – 40°C (50 – 104°F) Relative humidity  30 – 85%∗1 Operation environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) ∗1 No condensation
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            101  Battery pack  Ambient temperature 0 – 50°C (32 – 122°F) Avoid prolonged storage at temperatures exceeding 40°C (104°F).  Relative humidity  10 – 95%∗1 Transportation and storage environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) Ambient temperature 0 – 50°C (32 – 122°F) Recharging: 0 – 40°C (32 – 104°F) Discharging: 0 to +50°C (32 to +122°F) Relative humidity  30 – 90%∗1 Operation environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) ∗1 No condensation
   102   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Recorder unit harness  Ambient temperature –20 to +70°C (–4 to +158°F) Relative humidity  10 – 95% Transportation and storage environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) Ambient temperature 0 – 50°C (32 – 122°F) Relative humidity  30 – 90% Operation environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) ∗1 No condensation  Antenna harness   Ambient temperature –20 to +70°C (–4 to +158°F) Relative humidity  10 – 95% Transportation and storage environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia) Ambient temperature 0 – 50°C (32 – 122°F) Relative humidity  30 – 90% Operation environment Barometric pressure 700 – 1060 hPa (10.2 – 15.4 psia)
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            103     Specifications System chart  Capsule endoscope ( EC-Y0005)Antenna unit (X432262-21)Recorder unit (X432262-11)Antenna harness (X432252-51) Recorder unit harness(X432251-40) Battery pack(X432262-11-12) AC cableAC adapterRecorder unit cradle (X432262-31)  USB cableCapsule activator(MAJ-1478)Workstation (MAJ-Y0134) LCD monitorPrinterUSB HDDCapsule endoscope system set  MAJ-Y0136 (MAJ-Y0136)
   104   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         The system (components within the bold dotted line) is IEC 60601-1 certified in a configuration that includes the capsule endoscope, antenna unit, and recorder unit. The recorder unit cradle, and USB cable, are IEC 60950-1 certified.   • In order to provide the intended functionality, the system emits RF energy while in operation. This may affect electrical devices in the vicinity. The patient should keep a distance from such devices. • Before the examination, the capsule endoscope is turned on before it is ingested. This could interfere with other nearby electrical equipment. To prevent problems, keep the capsule endoscope away from equipment that is susceptible to RF interference prior to ingestion. • This system conducts weak radio communication. Images may be lost or distorted when operating in an environment affected by electromagnetic transmissions (such as that of portable RF (Radio Frequency) communication devices). Avoid use in such environments.  FCC (for recorder unit , antenna unit, and   Recorder unit cradle)  This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:  -Reorient or relocate the receiving antenna. -Increase the separation between the equipment and receiver. -Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -Consult the dealer or an experienced radio/TV technician for help.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            105  Capsule endoscope  Product name Capsule endoscope (EC-Y0005) Field of view (maximum)  160°  Optics Depth of field  0 – 20 mm Sampling rate   2 fps Power supply Power source  Internal battery Battery life    12 hours Size Dimensions  ø 11 (diameter) × 26 (length) mm Type of protection against electric shock Internal power supply Degree of protection against electric shock TYPE BF Classification as medical electrical equipment Degree of protection against explosion Use under combustible atmosphere prohibited. Protection against fluid ingress   IEC 60529 IPX-8 EMC Applied standard; IEC 60601-1-2:2001  This instrument complies with the standards listed in the left column. CISPR 11 of emission: Group 1, Class A  Year of manufacturer 1712345  The year of manufacture is indicated by the second digit of the serial number. The year of manufacture indicates the serial number of EC-Y0005. FCC ID  :S8QEC-S10  This device comlies with part 15 of   the FCC Rules. Operation is subject to the following two conditions: (1)This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. IC  :4763B-ECS10  This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference , including interference that may cause undesired operation of the device.
   106   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            107  Recorder unit Product name Recorder unit (X432262-11)  Power source  Battery pack (X432262-11-12) Voltage  3.7 V Power supply   Current  500 mA Battery life    12 hours or longer Weight  282 g Size   Dimensions  83 (W) × 151 (H) × 32 (D) mm Type of protection against electric shock Internal power supply Classification as medical electrical equipment Degree of protection against explosion Use under combustible atmosphere prohibited. Year of manufacture 1712345  The year of manufacture is indicated by the second digit of the serial number. EMC Applied standard; IEC 60601-1-2:2001 This instrument complies with the standards listed in the left column. CISPR 11 of emission: Group 1, Class A
   108   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Recorder unit cradle  Product name Recorder unit cradle (X432262-31) Voltage  DC 5V Power supply  Current  3 A Weight  Main body: 345 g    (AC Adapter: 175 g) Size   Dimensions  133 (W) × 79 (H) × 84 (D) mm Type of protection against electric shock Class  Ⅰ Classification as medical electrical equipment Degree of protection against explosion Use under combustible atmosphere prohibited. Year of manufacture 1712345  The year of manufacture is indicated by the second digit of the serial number. EMC  Applied standard; IEC 60950-1:2005 This instrument complies with the standards listed in the left column. CISPR 22 of emission: Group 1, Class A
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            109  Battery pack  Product name   Battery pack (X432262-11-12) Type    Lithium-Ion storage cell Capacity    2860 mAh Voltage    3.7 V Recharging time   Approx. 2 hours Weight  70 g Size Dimensions  70.4 (W) × 10.2 (H) × 55.2 (D) mm
   110   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          Antenna unit  Product name Antenna unit (X432262-21) Weight  140 g Dimensions  83 (W) × 50 (H) × 13 (D) mm Size   The number of antennas 8 Degree of protection against electric shock TYPE BF Classification as medical electrical equipment Degree of protection against explosion Use under combustible atmosphere prohibited. EMC Applied standards; IEC 60601-1-2: 2001  This instrument complies with the standards listed in the left column. CISPR 11 of emission: Group 1, Class A
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            111  Recorder unit harness  Product name Recorder unit harness (X432251-40)  Weight  70 g Size   Dimensions  100 (W) × 160 (H)  ×  37 (D) mm    Antenna harness  Product name Antenna harness (X432252-51)  Weight  180 g Size   Dimensions  pouch 350(W)×160(H)mm Long belt    90(W)×50(H)mm Short belt    60(W)×50(H)mm
   112   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         EMC information 〇 Magnetic emission compliance information and recommended electromagnetic environments The system (components within the bold dotted line of “System chart” on page 103) is designed for use under the following electromagnetic environment.  Emission standard Compliance Guidance RF emissions CISPR 11 Group 1  This instrument uses RF (Radio Frequency) energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. Radiated emissions CISPR 11 Main terminal conducted emissions CISPR 11 Class A  This product is suitable for use in all buildings besides the residence. Harmonic emissions IEC 61000-3-2 Voltage fluctuations/flicker emissions IEC 61000-3-3 Not applicable  The system is not powered by a commercial power source.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            113  〇 Electromagnetic immunity compliance information and recommended electromagnetic environments  The system (components within the bold dotted solid line of “System chart” on page 103) is designed for use under the following electromagnetic environment.  Immunity test  IEC 60601-1-2 test level Compliance level Guidance Electrostatic discharge (ESD) IEC 61000-4-2  Contact: ±2, ±4, ±6 kV Air: ±2, ±4, ±8 kV Same as left  Floors should be made of wood, concrete, or ceramic tile that hardly produces static. If floors are covered with synthetic material that tends to produce static, the relative humidity should be at least 30%. Electrical fast transient/burst IEC 61000-4-4 ±2 kV for power supply lines ±1 kV for input/output lines Not applicable  The system does not have a cable for this signal. Surge IEC 61000-4-5 Differential mode: ±0.5, ±1 kV Common mode: ±0.5, ±1, ±2 kV Not applicable  The system is not powered by a commercial power source. < 5% UT (> 95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycle 70% UT (30% dip in UT) for 25 cycle Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 < 5% UT (> 95% dip in UT) for 5 seconds Not applicable  The system is not powered by a commercial power source. Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 3 A/m  Same as left  It is recommended to use this instrument by maintaining enough distance from any equipment that operates with high current.   UT is the AC mains power supply prior to application of the test level.
   114   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                          〇 Cautions and recommended electromagnetic environment regarding portable and mobile RF communications equipment such as cellular phones    • Where “P” is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and “d” is the recommended separation distance in meters (m). • This instrument complies with the requirements of IEC 60601-1-2: 2001. However, under electromagnetic environment that exceeds its noise level, electromagnetic interference may occur on this instrument. • Electromagnetic interference may occur on this instrument near a high-frequency electrosurgical equipment and/or other equipment marked with the following symbol:
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            115  ○ Recommended separation distance between portable and mobile RF communications equipment and this instrument  Separation distance according to frequency of transmitter (m) (calculated as V1=3 and E1=3) Rated maximum output power of transmitter P (W) 150 kHz – 80 MHz    80 MHz – 800 MHz 800 MHz – 2.5 GHz  0.01  0.12  0.12  0.23 0.1  0.38  0.38  0.73 1  1.2  1.2  2.3 10  3.8  3.8  7.3 100  12  12  23   The guidance may not apply in some situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Portable and mobile RF communications equipment such as cellular phones should be used no closer to any part of this instrument, including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.   FCC WARNING(for the USA) Change or modifications not expressly approved by the party responsible for Compliance could void the user’s authority to operate the equipment. The shielded interface cable recommended in this manual must be used with this equipment in order to comply with the limits for a digital device pursuant to Subpart B of Part 15 of FCC Rules.         1.2√ P d = 1.2√ P d = P 2.3√ d = Declation of ConformityDeclation of ConformityDeclation of ConformityDeclation of Conformity    Trade Name: OLYMPUS MEDICAL SYSTEMS  Model: X432262-11, X432262-21, X432262-31  Responsible Party: OLYMPUS AMERICA INC.  Address: 3500 Corporate Parkway, P.O. Box 610 Center Valley.                 PA 18034-0610, U.S.A Telephone Number: (484)896-5000
   116   OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                                                         FCC notice           This  device  complies  with  part  15  of  the  FCC  Rules.  Operation  is  subject  to  the following two conditions: (1) This device may not cause harmful interference, and (2) this  device  must accept any interference  received,  including  interference that  may cause undesired operation. IC    notice This  device  complies  with  Industry  Canada  licence-exempt  RSS  standard(s). Operation  is  subject  to  the  following  two  conditions:  (1)  this  device  may  not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.  Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de  To reduce potential radio interference to other users, the antenna type and its gain  should  be  so  chosen  that  the  equivalent  isotropically  radiated  power (e.i.r.p.) is not more than that permitted for successful communication.  Afin de réduire le risque d'interférence aux autres utilisateurs, il faut choisir le type  d'antenne et  son  gain  de  façon  à  ce  que  la  puissance  isotrope  rayonnée équivalente (p.i.r.e.) ne soit pas supérieure au niveau  requis pour  l'obtention d'une communication satisfaisante.
    OLYMPUS CAPSULE ENDOSCOPE SYSTEM SYSTEM MANUAL                            117                   X432262#01