Orex Computed Radiography CR-SCANNER Scanner with RFID module User Manual manual revised

Orex Computed Radiography Ltd. Scanner with RFID module manual revised

manual revised

Download: Orex Computed Radiography CR-SCANNER Scanner with RFID module User Manual manual revised
Mirror Download [FCC.gov]Orex Computed Radiography CR-SCANNER Scanner with RFID module User Manual manual revised
Document ID1432189
Application ID7RvmbplCDOPFEp/FzWi1+Q==
Document Descriptionmanual revised
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize44.86kB (560801 bits)
Date Submitted2011-03-16 00:00:00
Date Available2011-03-16 00:00:00
Creation Date2011-03-15 09:38:38
Producing SoftwareAcrobat Distiller 8.1.0 (Windows)
Document Lastmod2011-03-15 17:01:49
Document TitleCS 7600 Safety, Regulatory & Technical Specification Guide (8J4070)
Document CreatorFrameMaker 9.0

DRAFT CURRENT AS OF:
15/3/11
CS 7600
Safety, Regulatory & Technical
Specification User Guide
Notice
The Regulatory Information & Technical Specifications Guide for the CS 7600
includes information on the safety instructions, regulatory information and the
technical specifications of the device. We recommend that you thoroughly familiarize
yourself with this Guide in order to make the most effective use of your system.
The information contained in this Guide may be subject to modification without
notice, justification or notification to the persons concerned.
No part of this Guide may be reproduced without the express permission of
Carestream Health, Inc.
U.S. Federal law restricts this device to sale by or on the order of a dentist or
physician.
This document is originally written in English.
Manual Name: CS 7600 System Safety, Regulatory and Technical Specifications
User Guide
Part Number: 8J4070
Revision Number: 01
Print Date: 2011-04
The Brand names and logos reproduced in this Guide are copyright.
0086
Manufacturer
Care stream Hea lth, Inc.
150 Verona Street
Roche ster NY 14 608
Authorized Representative in the European Community
EC REP
Carestream Health France
1, rue Galilée
93192 NOISY-LE-GRAND CEDEX
FRANCE
Contents
Chapter 1
Safety Information
Indications for Use . . . . . . . . 1
Conventions in this Guide . . . . . 1
Warning and Safety Instructions . . 2
Laser Safety Instructions . . . . . 4
Hygiene and Disinfection . . . . . 5
General Notice . . . . . . . . . . 5
Marking and Labeling Symbols TBD .6
IEC Symbols Used . . . . . . . y7
Safety Labels . . . . . . . . . . y7
Label Locations . . . . . . . . . . 8
Chapter 2
Regulatory
Information
General Regulatory Information . . 9
Compliance with International
Regulations . . . . . . . . . . . 16
Compliance with FCC Rules . . . 16
Declaration of Conformity with the
R-TTE Directives TBD -- WILL
NEED THE DECLARATION OF
CONFORMITY FOR THE CS 7600
TO REPLACE THIS. . . . . . . . . 18
Chapter 3
Technical
Specifications
TBD Factory/Country of Origin . 19
Manufacturer . . . . . . . . . . 19
Model. . . . . . . . . . . . . . 19
CS 7600 Technical Specifications 19
System specifications. . . . . y19
Plate Specifications . . . . . . y20
RF Information. . . . . . . . . y21
CR Environmental Requirements .
y21
Minimum Computer System
Requirements . . . . . . . . . . 22
DRAFT CS 7600 _Safety, Regulatory & Technical Specifications
iii
(8J4070)_Ed01
iv
1
Safety Information
Indications for Use
The CS 7600 is intended for digital dental radiography using a storage
phosphor screen for radiographic diagnostic intraoral images.
Conventions in this Guide
The following special messages emphasize information or indicate potential risk
to personnel or equipment.
WARNING: Warns you to avoid injury to
yourself or others by following the safety
instructions precisely.
Important: Alerts you to a condition that might
cause problems.
Tip: Provides extra information and hints.
Note: Emphasizes important information.
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
(8J4070)_Ed01
Warning and Safety Instructions
WARNINGS
Device:
•
Read and understand this Safety Information before using the CS
7600.
•
To ensure safety, read all user manuals carefully before using the
system and observe all Cautions, Important and Notes located
throughout the manuals.
•
Keep this manual with the equipment.
•
You are responsible for the operation and maintenance of this device.
Only legally qualified persons can operate this device. When
necessary, have a trained authorized service technician carry out
inspection and maintenance operations.
•
The CS 7600 is Class I, continuous operated stationary equipment
without applied parts and has one signal input/output part. Product is
provided with ordinary protection against the harmful ingress of
water.
•
DO NOT use this device in the presence of a flammable anesthetics
mixture with air or with oxygen or with nitrous oxide.
•
DO NOT remove or open system covers or plugs. Internal circuits
use high voltage capable of causing serious injury. Fuses blown
within 36 hours of being replaced by a qualified technician may
indicate malfunctioning electrical circuits within the system. Have the
system checked by qualified service personnel. Do not attempt to
replace any fuse. Fluids that seep into the active circuit components
of the system may cause short circuits that can result in electrical
fires. Therefore, do not place any liquid or food on any part of the
system.
•
DO NOT operate the device if there is the threat of an earthquake.
Following an earthquake, ensure that the device is operating
satisfactorily before using it again. Failure to observe this precaution
may expose patients to hazards.
•
DO NOT place objects within the field of operation of the device.
•
Connect this equipment ONLY to a mains power supply with
protective ground to avoid any risk of electric shock.
Chapter 1 Safety Information
•
The appliance coupler/inlet of the flexible power cord is used as a
disconnecting device from the mains.
•
To achieve grounding reliability in USA installations, the equipment
must be connected to an equivalent receptacle marked “Hospital
Only” or “Hospital Grade”.
•
Disinfect the detachable panel of the CS 7600 scanner after TBD
cycles.
•
To dispose of the device or its components, contact a service
technician.
•
No modification of this equipment is allowed.
•
DO NOT block air circulation around the unit. Always maintain at
least 15 cm clearance around the unit to prevent overheating and
damage to the system.
Plates:
•
To prevent damage to the plates and the possibility of image artiacts,
avoid contact between the plates and the following
materials/solutions/solvents: Isopropyl alcohol, hydrogen peroxide
and other peroxides, citrus-based cleaners, hand lotions and
water-less hand sanitizers, as well as surfactants and lubricants.
•
The plate contains Barium, and therefore may be considered
hazardous or special waste at the end of its useful service life. For
disposal or recycling information, contact your local authorities.
•
Do not soak plate in any cleaning or disinfecting solutions. Do not
autoclave; autoclaved plates must be discarded.
•
Plates should be stored in their original packing or storage box when
they are not in use. Always store plates in a dark and dry place.
•
Do not expose the plates to light for long periods as this can have a
degrading effect.
•
Do not store plates in hot or moist conditions.
•
Do not fold, crease, or bend the plates.
•
Avoid touching the phosphor side of the plates. and be careful not to
drag the phosphor side of the plate across any surface as this will
damage the plate.
•
Do not leave plates where they can become damaged by liquid or
chemical spills.
•
Read and follow the instructions in Material Safety Data Sheets
(MSDS) for the CS Screen Cleaner.
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
(8J4070)_Ed01
•
If a commercially prepared equivalent solution of diluted bleach is
used, it should be used according to its manufacturer’s instructions.
Computer:
•
DO NOT place the computer and the peripheral equipment
connected to it in the immediate vicinity of the patient in the unit.
Leave at least 1.5 m distance between the patient and the unit. The
computer and the peripheral equipment must conform to the
IEC60950 standard.
•
See your computer installation guide for details of the data
processing system and screen. Leave a sufficient amount of clear
space around the CPU to ensure that it is properly ventilated.
•
To obtain maximum image quality and visual comfort, position the
screen to avoid direct light reflections from internal or external
lighting.
•
In order to guarantee medical-grade leakage current, the computer
connected to the system should be a medical-grade computer or
connected to the mains through medical-grade power supply. This is
not relevant if the system is connected to a laptop by USB through a
USB isolator or network cable.
Laser Safety Instructions
WARNINGS
The CS 7600 is a CLASS 1 Laser product.
•
During normal operation, always keep the unit enclosed in its
protective cover to prevent the outside area from being exposed to
laser radiation.
•
During normal operation, do not remove the cover. Only authorized
service personnel may remove the cover.
•
Do not operate the system while the access door is open.
WARNING
Laser radiation when cover is removed. Avoid direct exposure to beam.
Class 3B laser inside.
Chapter 1 Safety Information
Hygiene and Disinfection
WARNINGS
•
Do not use chemical autoclave for disinfecting the
detachable panel.
•
To prevent cross-contamination, use a new envelope for
each new patient.
WARNING
Do not operate the equipment in the presence of
explosive liquids, vapors, or gases. Do not plug in or
turn on the system if hazardous substances are detected in the
environment. If these substances are detected after the system has been
turned on, do not attempt to turn off the unit or unplug it. Evacuate and
ventilate the area before turning off the system.
General Notice
WARNING
If the product does not operate properly or fails to
respond to the controls as described in the product accompanying
documentation:
•
•
•
Follow the safety precautions as specified in this guide.
Stop using the unit and prevent any changes to it.
Immediately contact the service office, report the problem, and await
further instructions
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
(8J4070)_Ed01
Marking and Labeling Symbols TBD
In the European Union, this symbol
indicates: DO NOT discard this product in
a trash receptacle; use an appropriate
recovery and recycling facility.
Contact your local sales representative for
additional information on the collection
and recovery programs available for this
product.
Warning: General warning sign
Using accessories other than those specified
in this document with the exception of those
sold by Carestream Health may result in a
lower level of security for the entire system.
The ON/OFF button
General mandatory action sign
Follow operating instructions sign
Manufactured Date
Manufacturer’s address
Chapter 1 Safety Information
IEC Symbols Used
The system may have labels with one or more of the following symbols.
Caution—consult accompanying documents
Power ON
Caution—Electrical shock hazard
Safety Labels
Label
Description
Caution—Electrical
shock hazard
Laser-emitting product
Class 3B laser product
inside scanner
Location
PM board
Optical head
Optical head
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
(8J4070)_Ed01
Label Locations
The following figures illustrate the label locations of the CS 7600 components.
Figure 1
CS 7600 Labels
Chapter 1 Safety Information
2
Regulatory
Information
General Regulatory Information
Compliance with European and International Standards
EN 60601-1 / IEC
60601-1
Medical Electrical Equipment, Part 1: General
Requirements for Safety
EN 60601-1-2 / IEC
60601-1-2
Medical Electrical Equipment, Part 1-2:
Electromagnetic Compatibility
EN 60601-1-6 / IEC
60601-1-6
Medical Electrical Equipment, Part 1-6: General
requirements for basic safety and essential
performance - Collateral Standard: Usability
EN ISO 14971
Medical devices - Application of risk management to
medical devices
EN 980
Symbols for use in the labeling of medical devices
EN 1041
Information supplied by the manufacturer of medical
devices
EN 62304/IEC 62304 Medical device software - Software life cycle
processes
EN 300 328
Electromagnetic Compatibility and Radio Spectrum
Matters
EN 301 489-17
ElectroMagnetic Compatibility and Radio Spectrum
Matters (ERM); ElectroMagnetic Compatibility (EMC)
Standard for Radio Equipment and Services; Part 17 :
Specific Conditions for 2.4 GHz Wideband
Transmission Systems and 5 GHz High Performance
RLAN Equipment
EN 50371
Generic Standard to demonstrate the compliance of
low power electronic and electrical apparatus with the
basic restrictions related to human exposure to
electromagnetic fields (10 MHz - 300 MHz). General
Public
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
(8J4070)_Ed01
Classification in Accordance with EN/IEC 60601-1
Type of protection against Class 1 equipment
electric shock
Degree of protection
against electric shock
Type B – No Applied Parts
Protection against harmful Ordinary equipment, IPX0
ingress of water
Operation mode
Continuous operation
Electromagnetic Compatibility Precautions
Conformity with EN/IEC 60601-1-2
Electromagnetic Compatibility Precautions
10
•
Medical electrical equipment requires special precautions regarding
electromagnetic compatibility (EMC).
•
CS 7600 must be installed and put into service according to the
EMC information provided in this document.
•
Communication Equipment: Portable and mobile Radio Frequency
(RF) communications equipment can affect the Electromagnetic
Compatibility of CS 7600s.
•
CS 7600 may be interfered with other equipment even if that other
equipment complies with CISPR emission requirements.
Chapter 2 Regulatory Information
Guidance and Manufacturer’s Declaration - Electromagnetic
Emissions (IEC 60601-1-2)
The CS 7600 is intended for use in the electromagnetic environment specified
below. The customer or the user of the CS 7600 should assure that it is used in
such an environment.
Emissions Test Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class A
Harmonic
emissions
Class A
IEC 61000-3-2
Voltage
fluctuations/
Complies
flicker emissions
IEC 61000-3-3
Electromagnetic Environment - Guidance
The CS 7600 uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The CS 7600 is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity
The CS 7600 is intended for use in the electromagnetic environment specified
below. The customer or the user of the CS 7600 should assure that it is used in
such an environment.
Immunity
Test
Electrostatic
discharge
(ESD)
IEC
61000-4-2
Electrical fast
transient/burst
IEC
61000-4-4
IEC 60601
Test Level
±6 kV
Compliance
±6 kV
Electromagnetic Environment Guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
contact
contact
±8 kV air
±8 kV air
±2 kV for
power supply
lines
±1 kV for
input/output
lines
±2 kV for
power
supply lines Mains power quality should be
that of a typical commercial or
±1 kV for
hospital environment.
input/output
lines
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
(8J4070)_Ed01
11
Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity (Continued)
Surge
IEC
61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
±1 kV line(s)
Mains power quality should be
to line(s)
that of a typical commercial or
Not
hospital environment.
applicable
< 5 % UT
( > 95 % dip
in UT)
for 0.5 cycle
40 % UT
(60 % dip in
UT)
for 5 cycles
70 % UT
(30 % dip in
UT)
for 25 cycles
< 5 % UT
( > 95 % dip
in UT)
for 5 sec
< 5 % UT
( > 95 %
dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in
UT)
for 5 cycles
70 % UT
(30 % dip in
UT)
for 25 cycles
< 5 % UT
( > 95 %
dip in UT)
for 5 sec
Power
frequency
(50/60 Hz)
magnetic field 3 A/m
IEC
61000-4-8
12
3 A/m
Chapter 2 Regulatory Information
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the CS 7600 requires continued
operation during power mains
interruptions, it is recommended
that the CS 7600 be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity (IEC 60601-1-2)
The CS 7600 is intended for use in the electromagnetic environment specified
below. The customer or the user of the CS 7600 should assure that it is used in
such an environment.
Immunity
Test
IEC
Compliance
60601
Level
Test
Level
Electromagnetic Environment Guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the CS 7600, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
Conducted
RF
IEC
61000-4-6
3 Vrms
150
kHz to
80 MHz
3.5
d = ( ------- ) P
V1
[V1]= 3 V
3.5
d = ( ------- ) P
E1
80 MHz to 800 MHz
3 V/m
Radiated RF
[E1]= 3 V/m
d = ( ------ ) P 800 MHz to 2.5 GHz
80 MHz
E1
IEC
to 2.5
61000-4-3 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey should be less than the
compliance level in each frequency
range.b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
(8J4070)_Ed01
13
Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity (IEC 60601-1-2) (Continued)
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the
CS 7600 is used exceeds the applicable RF compliance level above, the CS 7600
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the CS 7600.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3
V/m.
14
Chapter 2 Regulatory Information
Recommended Separation Distances between Portable and
Mobile RF Communications Equipment and the CS 7600
(IEC 60601-1-2)
The CS 7600 is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the CS
7600 can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the CS 7600 as recommended below, according to the
maximum output power of the communications equipment.
Separation Distance (m) According to Frequency of
Transmitter
Rated Maximum
Output Power of
Transmitter (W)
150 kHz to
80 MHz
3.5
d = ( ------- ) P
V1
80 MHz to
800 MHz
3.5
d = ( ------- ) P
E1
800 MHz to
2.5 GHz
d = ( ------ ) P
E1
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Note: The communication without breaking has
been determined to be essential performance with
regard to electromagnetic compatibility.
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
(8J4070)_Ed01
15
Compliance with International Regulations
The product conforms to the following safety standards: EN 60601-1:90 +
A1(93) + A2(95) + A13(96)/IEC 60601-1:88 + A1(91) + A2(95)/EN/IEC
60601-1-1, EN/IEC 60825-1:94 + a1(97) + a2(01) + Cor1(02), EN
60601-1:06/IEC 60601-1:05, UL 60601-1:03, rev.06, IEC
60825-1:93+A1:97+A2:2001 (Foe LED module), IEC 60825-1:2007 (For
Laser module CAN\CSA-C22.2 No. 601.1-M90, JIS T 0601-1:1999.
This device complies with 21CFR 1040.10.
•
Medical Device directives 93/42/ European Economic Community
(EEC), Class IIa follow the rule 16 as amended by 2007/47/EEC.
•
The CS 7600 is an active device specifically intended for scanning of
X-ray diagnostic images.
•
ElectroMagnetic Compatibility (EMC) directive 89/336/EEC, Group
1, Class A
•
Radio and Telecommunications Terminal Equipment directives
1999/5/EEC
•
FCC rules part 15/EN 300 330-2
V1.3.1/47CFRp15sb.C/RSS/210:07 Issue 7
Compliance with FCC Rules
This equipment has been tested and found compliant with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate
radio frequency energy and if not installed and used in accordance
with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that the interference will not
occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
•
•
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that
to which the receiver is connected.
•
Consult the dealer or an experienced radio/TV technician for help.
This device complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions:
16
Chapter 2 Regulatory Information
•
•
This device may not cause harmful interference.
This device must accept any interference received, including
interference that may cause undesired operation.
Warning: Any changes or modifications not
expressly approved by the party responsible for
compliance could void the user’s authority to
operate this equipment.
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
(8J4070)_Ed01
17
Declaration of Conformity with the R-TTE Directives
TBD -- WILL NEED THE DECLARATION OF CONFORMITY FOR
THE CS 7600 TO REPLACE THIS.
18
Chapter 2 Regulatory Information
3
Technical
Specifications
TBD Factory/Country of Origin
SANMINA TBC
Manufacturer
Carestream health, Inc.
150 Verona Street
Rochester,
New York - USA 14608
Model
CS 7600
CS 7600 Technical Specifications
System specifications
Components
Gray scale
CS 7600
4096 bits
Dimensions
(without bracket) 266.5 mm (H), 236.6 mm (W), 259.4 mm (D)
Weight
Approximately 6 kg
Power supply
100–240 V AC, 50/60 Hz, 1.5 A
Laser Wavelength 635 nm– 650 nm (Class 3B)
Output Power
Up to 12 mW
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
19
(8J4070)_Ed01
Plate Specifications
Plate Dimensions
SHR
HR
HS
Plate
Size
Height
(mm)
22
24
31
27
54
Width
(mm)
31
40
41
54
76
Height
(pixels)
1552
2003
2053
2704
3805
Width
(pixels)
1101
1202
1552
1352
2704
Height
(pixels)
884
1141
1169
1540
2167
Width
(pixels)
627
684
884
770
1540
Height
(pixels)
499
643
659
868
1222
Width
(pixels)
354
386
499
434
868
Plate Scanning Resolution
Resolution Level
Laser Spot
Size
Practical
Resolution
Pixel Size
Super high resolution
(SHR)
20.0 micron
20 LP/mm*
0.019 mm
High resolution (HR)
35.0 micron
14 LP/mm*
0.035 mm
High speed (HS)
62.5 micron
8 LP/mm*
0.062 mm
*LP/mm: Line pairs per millimeter (when scanning Resolution Phantom)
20
Chapter 3 Technical Specifications
RF Information.
RF Information
RFID
ISO15693
Frequency
13.56 MHz
Output Power
20 mW
CR Environmental Requirements
Ambient Operating Conditions
Ambient Operating Conditions
Temperatures
5 °C ~35 °C
Relative humidity
30 % ~ 85 % Non-condensing
Atmospheric pressure
700 hPa ~ 1060 hPa
Storage Conditions
Storage Conditions (in packaging)
Temperatures
-10 °C ~ 60 °C
Relative humidity
10 % ~ 95 % Non-condensing
Atmospheric pressure
700 hPa ~ 1060 hPa
Transport Conditions
Transport Conditions
Temperatures
-20 °C ~ 60 °C
Relative humidity
45 %~ 85 %
Atmospheric pressure
700 hPa ~ 1060 hPa
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
21
(8J4070)_Ed01
Minimum Computer System Requirements
The computer and the peripheral equipment must conform to the
IEC60950 standard.
Item
Viewing and Acquisition
CPU
2 GHz Intel Duo Core
RAM
1 GB (2 GB recommended)
Hard disk drive
Graphic board
Monitor
Operating
system
•
•
•
•
•
1 monitor
17" or larger
1024 x 768 minimum screen
resolution, 32-bit color mode
•
•
Windows XP Pro edition SP3
Windows Vista Ultimate 32-bit
SP2
Windows 7 Professional 32-bit
Windows 7 Ultimate 32-bit
Ethernet
interface
10/100 LAN card
USB 2.0
3 ports (4 ports recommended)
Wireless adapter
22
RAM has a major
impact on system
performance.
1.2 GB for software installation
80 GB free space to use the
software
Nvidia/ATI based board supporting
Open Glide 1.2 with 256 MB of video
RAM on AGP x8 video bus (example:
Nvidia GeForce 6800 GT)
•
•
Comments
Wireless 802.11g Adapter (USB,
PCMCIA or PCI)
Chapter 3 Technical Specifications
The video RAM has
major impact on
system
performance.
Your monitor is a
vital component in
displaying quality
images. Low-quality
screens will prevent
you from proper
diagnoses and
treatment.
Second Ethernet
interface is
recommended for
Internet\PACS\
Server connection
option
Item
CD/DVD drive
Viewing and Acquisition
DVD-ROM drive is required to install
the product software
To enable hearing
audible alarms
initiated by the
system
Audio speakers
Blue Tooth
interface on
board
(Optional)
Comments
Recommended for RFID option.
External USB
dongle for BT can
be used also.
DRAFT CS 7600_Safety, Regulatory & Technical Specifications
23
(8J4070)_Ed01
24
Chapter 3 Technical Specifications

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Page Mode                       : UseOutlines
Signing Date                    : 2011:03:15 17:01:49+02:00
Signing Authority               : ARE Acrobat Product v8.0 P23 0002337
Annotation Usage Rights         : Create, Delete, Modify, Copy, Import, Export
XMP Toolkit                     : Adobe XMP Core 4.0-c316 44.253921, Sun Oct 01 2006 17:14:39
Producer                        : Acrobat Distiller 8.1.0 (Windows)
Creator Tool                    : FrameMaker 9.0
Modify Date                     : 2011:03:15 17:01:49+02:00
Create Date                     : 2011:03:15 09:38:38Z
Metadata Date                   : 2011:03:15 17:01:49+02:00
Format                          : application/pdf
Title                           : CS 7600 Safety, Regulatory & Technical Specification Guide (8J4070)
Description                     : Safety, Regulatory & Technical Specification User Guide
Document ID                     : uuid:677fdb23-8f27-4e27-bfaa-b40fd0ee4688
Instance ID                     : uuid:e73dc066-b48a-42cf-a1b1-1fa7c7cbf087
Has XFA                         : No
Page Count                      : 28
Subject                         : Safety, Regulatory & Technical Specification User Guide
Creator                         : FrameMaker 9.0
Keywords                        : regulatory, stella, CS 7600, 8J4070
EXIF Metadata provided by EXIF.tools
FCC ID Filing: ZBE-CR-SCANNER

Navigation menu