OrthoAccel Technologies 3 AcceleDent Optima User Manual

OrthoAccel Technologies, Inc. AcceleDent Optima

User Manual

OPTIMA™ 405-1307-001 Rev AOrthoAccel®, AcceleDent®, SoftPulse Technology® are registered trademarks and Optima™ is a trademark of OrthoAccel Technologies, Inc.Patents: www.acceledent.com/patents FPODIRECTIONS FOR USE93% RH30% RH1060 hPa700 hPa   50
2BACK TO CONTENTSPage      of 24Contents1.0 Introduction 2.0 About the manufacturer3.0 Intended Use/Indications for Use and Contraindication 4.0 Component summary 5.0 Use 5.1. Set up and charging5.2. Indicators for the AcceleDent Optima Device5.3. Use schedule6.0 Care6.1. Cleaning6.2. Storage6.3. Disposal7.0 Warranty 7.1. Express Limited Warranty7.2. Legal Limitations7.3. Customer Service and Replacement, including troubleshooting and FAQs location.8.0 Safety8.5. Warnings8.6. Precautions9.0 Help 10.0 Technical Description 11.0 Symbols and Definitions11.1 Manufacturer info, form document number 33445577889999111111121314142224
3BACK TO CONTENTSPage      of 241.0 IntroductionThank you for choosing AcceleDent® Optima™. Added to your current treatment, AcceleDent Optima optimizes your orthodontic experience with faster tooth movement and reduced pain. As an FDA-cleared medical device, AcceleDent Optima uses SoftPulse Technology® to safely accelerate tooth movement by as much as 50 percent. It does this by generating small vibrations called micropulses to gently accelerate the movement of your teeth as they are guided by your orthodontic appliances. With AcceleDent Optima, you will experience faster treatment and less pain than traditional treatment methods. 2.0 About the ManufacturerContact the manufacturer, OrthoAccel®, about any questions or concerns that are specifically related to AcceleDent Optima and its components.Customer Service Center:North America: 1-866-866-4919Spain: +34900839096United Kingdom: +448000988430France: + 33805080086Italy: +39800596833Germany: +498007244532Outside North America: +442033181915EMERGO EUROPE: +31703458570Email: customerservice@orthoaccel.comOnline: www.acceledent.com
4BACK TO CONTENTSPage      of 243.0 Intended Use/Indications for      Use and ContraindicationAcceleDent Optima is an orthodontic accessory intended for use during orthodontic treatment. It is used in conjunction with orthodontic appliances such as braces and aligners to help facilitate minor anterior tooth movement.Contra indications for Use• Use of osteoporosis drugs• Poor oral hygiene• Periodontal disease that is not under full control for at least 3-4months prior to the start of treatment.4.0 AcceleDent Optima Component Summary with Images 7562341891 MOUTHPIECE2 ACTIVATOR3 ON/OFF BUTTON (STAND-BY)4 STATUS INDICATOR5 CHARGING CASE6 USB CHARGING PORT7 USB CHARGING CABLE8 POWER ADAPTER9 TRAVEL POUCH
5BACK TO CONTENTSPage      of 245.0 Use5.1. Set-Up and Charging5.1.1. Power the Charging Case by connecting the USB Charging Cable to an electrical source. 5.1.2. Open Charging Case by pushing down on the top of the latch (Unlock symbol). Check to make sure the AcceleDent Optima Activator is secure in the Charging Case for proper charging. To close the lid securely, push down on the front of the latch (Lock symbol). Note: The lid does not need to be closed for proper charging.  5.1.3. The Status Indicator on the AcceleDent Optima Activator will be solid orange while charging and turn o when fully charged. 5.1.4. Prior to first use, charge for at least 8 hours.
6BACK TO CONTENTSPage      of 245.1.5.  To begin your treatment session, press the on/o button. The Activator will vibrate, and the Status Indicator will be solid blue. The device will automatically turn o after 20 minutes of treatment.2 0 min5.1.6. Place the Mouthpiece in your mouth and bite down gently. AcceleDent Optima features a half-time indicator, which will pulse 2 times at the 10-minute mark.  After 20 minutes of treatment, the Activator and Status Indicator light will automatically turn o. 5.1.7. To pause your treatment session for up to 5 minutes, press the on/o button once. During the pause, the Status Indicator will pulse blue slowly. When you are ready to resume, press the on/o button again to resume treatment.  AcceleDent Optima will remember how long you used it as long as the interruption does not exceed 5 minutes.  After 5 minutes of being paused, AcceleDent Optima will operate for a full session of 20 minutes at its next use.5.1.8. Recharge the Activator after each treatment by returning it to the powered Charging Case.  Place the AcceleDent Optima Activator securely in the Charging Case to ensure proper charging.5.1.9. The Status Indicator will blink orange when the battery is low.
7BACK TO CONTENTSPage      of 245.1.10.  A Travel Pouch has been provided to store your Activator, Charging Case and accessories when not in use or when traveling. 5.2.  Indicators for the AcceleDent Optima DeviceSolid orange Charging batteryBlinking orange Low batterySolid blue Active treatmentSolid blue and pulsing vibration Mid-point treatmentSlow pulsing blue Paused treatmentFading blue to none End of treatmentAlternating orange and blue blinking End of device life5.3.  Use Schedule5.3.1.  By using AcceleDent Optima for 20 minutes per day, you will experience faster treatment and less pain than traditional treatment methods. 5.3.2.  Select a consistent time for use each day. This way using AcceleDent Optima becomes a routine, and you may be less likely to forget a daily session. 5.3.2.1.  Some patients prefer using AcceleDent Optima just after they wake up and brush their teeth in the morning. Others use the appliance while engaged in a consistent daily activity, such as reading, watching TV, or using a computer/mobile device. Try several approaches and decide on the approach that works best for you. 5.3.2.2.  You may find that occasionally using AcceleDent Optima right after you receive a wire change or switch to a new aligner tray is a good time to help reduce pain. 5.3.3. Try to find a time that you can use AcceleDent Optima without interruption as it is best not to stop use in the middle of a session.5.3.4. You may easily pause and resume use if you are interrupted during a session. To pause, press the power button on the Activator. Refer to 5.1.7 for further information.
8BACK TO CONTENTSPage      of 246.0 Care6.1. Cleaning6.1.1. Simple hand cleaning is sucient to remove accumulation of dirt or other build-up on the Activator and Charging Case. 6.1.2. For the Charging Case:6.1.2.1. DO NOT rinse or submerge the Charging Case in water.6.1.2.2.  As needed, unplug from power source and wipe external and internal surfaces using a soft moist cloth. Dry with a soft cloth before reconnecting to power source.6.1.3. For the Mouthpiece:6.1.3.1.  After each use, rinse the Mouthpiece with warm running water and wipe with a dry cloth prior to storage or charging. 6.1.3.2.  Activator is waterproof, but it is recommended that the Activator is not submerged in water or cleaning solutions for an extended period of time.
9BACK TO CONTENTSPage      of 246.1.3.3.  Occasional cleaning with dish soap, dental cleaning solutions, and/ or brushing with toothpaste is acceptable.6.2. Storage6.2.1. Store AcceleDent Optima and Accessories away from pets, pests, and children to avoid inadvertent damage.  6.2.2. Activator should be stored in the Charging Case when not in use and store the entire system in the Travel Pouch when traveling, to keep the Activator and Charging Case free from dust, lint and moisture.6.2.3. When not charging, unplug the Charging Case from the USB Cable and wall mount power supply.6.2.4. Clean and store all components in a cool, dry place.6.3. Disposal6.3.1. The rechargeable battery inside of your Activator cannot be replaced.6.3.2.  In order to protect and preserve our environment, the Activator Assembly, Charging Case, USB Cable and Power Adapter should be disposed of at a designated collection facility for batteries and other electrical devices like cell phones, computers, video games and electronic tools.6.3.3. These designated collection facilities are free of charge to the consumer because  OrthoAccel and its distributors have acquired the appropriate registrations. The separated collection of electronic equipment waste allows for proper recycling, saves energy and resources and prevents hazardous material from going to a landfill. The internet can be a helpful tool to locate the nearest recycling center for batteries and electrical devices, or contact OrthoAccel Technologies, Inc. at 1-866-866-4919 (North America) or +44 (0) 203-318-1915 (outside of North America) or customerservice@orthoaccel.com for assistance.7.0 Warranty7.1. Express Limited Warranty7.1.1. OrthoAccel Technologies, Inc. (“OrthoAccel”) warrants that the product packaged with this warranty (the “Product”) will be free from significant defects
10BACK TO CONTENTSPage      of 24in materials and workmanship under normal use and service for a period of one year from the date of purchase (the “Warranty Period”). This limited warranty is non-transferable and applies only to the original purchaser and the initial orthodontic patient using the Product. This limited warranty also applies only when the Product is used in accordance with the Product manual and directions for use found on the Product website.7.1.2. This limited warranty covers all defects encountered in normal use and service but does not apply if: (1) the Product is modified or tampered with or disassembled, (2) the Product is damaged by an act of God, misuse, abuse, neglect, accident, or mishandling, (3) the Product is not used or maintained in accordance with the accompanying user documents, or (4) the serial number on the Product is defaced, altered or removed. In addition, this limited warranty does not cover normal wear such as discoloration or fading, and normal wear that does not compromise use of the appliance such as chips, scratches, and abrasions. Your exclusive remedy for breach of this limited warranty during the Warranty Period shall be, at the option of OrthoAccel, the replacement of the Product with a new Product, as determined by the OrthoAccel Warranty. Replacement shall not extend the original Warranty Period.7.1.3. All returns for warranty service will require an RMA (Return Merchandise Authorization). To receive an RMA for obtaining warranty services, you must call 1-866-866-4919 in North America or +44 (0) 203-318-1915 if outside of North America. In addition, if your product has not been registered you must submit proof of the date of original purchase such as a copy of your dated invoice and must insure, pack and ship the Product to an authorized service center in accordance with OrthoAccel’s instructions within 7 days after your receipt of an RMA. The package must include your RMA. RMA’s will not be extended or reissued.7.1.4. No Implied Warranties: Limitations on Damages7.1.5. The Express Limited Warranty provided above is the only expressed warranty made to you and is in lieu of all other warranties, whether express or implied. Without limiting the generality of the foregoing, OrthoAccel disclaims any and all other warranties, express or implied. With respect to the product including its condition, the existence of any latent or patent defects, its non-infringement of third party rights and its merchantability or fitness for any particular use. In no event shall OrthoAccel or any of its aliated or subsidiary companies be liable for any special, incidental, or consequential damages based upon breach of warranty, breach of contract, negligence, tort, or any other legal theory. Such damages include, without limitation, loss of savings or revenue; loss of profit; loss of use; the claims of third parties including, without limitation, dental professionals; and cost of any substitute equipment or services.
11BACK TO CONTENTSPage      of 247.2. Legal Limitations7.2.1. Some jurisdictions do not permit or allow limitations on how long implied warranties last or the exclusion or limitation of incidental or consequential damages. If any terms of this limited warranty, including, without limitation, the exclusion of damages, are limited or prohibited by your jurisdiction, the prohibited provision shall not apply but the remainder of this limited warranty shall remain in full force and eect. You may have rights in addition to this limited warranty under the laws of your jurisdiction, which may vary from jurisdiction to jurisdiction.7.3.  Customer Service and Replacement7.3.1. For additional support during use, please reference the Frequently Asked Questions (FAQ) and Troubleshooting Guide located at acceledent.com/FAQ.7.3.2. If you have any questions or comments after reading the provided materials, please contact the OrthoAccel Customer Service:North America: 1-866-866-4919Outside North America: +44 (0) 203-318-1915Email: customerservice@orthoaccel.comOnline: www.acceledent.comEMERGO EUROPE: +31-70-345-85708.0 Safety8.1.  Patients undergoing orthodontic treatment under the supervision of an orthodontist or dental professional are the intended operators of this product. While not exclusive, the intended user profile primarily comes from the following demographics: male and female teens (13 - 17 years old with adult consent or supervision, as required), and male and female adults (18 - 65 years old).  No age or eduction level restrictions. 8.2.  This product is designed to enhance your orthodontic treatment only. Use this product only for its intended use as described in this manual.8.3. (For healthcare professionals): Please instruct patients to read these Directions For Use for precautions to be taken in the event of changes in the performance of the device and precautions to be taken regarding the exposure of the device to reasonably foreseeable environmental conditions.
12BACK TO CONTENTSPage      of 248.4.  Contact your physician, orthodontist or dentist if you have any medicalconcerns about AcceleDent Optima. Discontinue use of this product andimmediately contact your orthodontist if any discomfort or pain isexperienced.8.5. AcceleDent Optima has been tested and is in compliance with safety standards for electromechanical devices. AcceleDent Optima meets the required electrical safety and emissions standards; however, the appliance has not been tested for eects on any other specific medical devices (such as pacemakers, cochlear implants, and/or nerve stimulators). Please contact the other device manufacturers with any questions or concerns about eects on their devices.8.6. This appliance is not intended for use by persons (including children)with reduced physical, sensory, or mental capabilities unless they have beengiven supervision or instruction concerning use by a person responsible fortheir safety, or unless otherwise indicated by an orthodontist or dentist.8.5. Warnings8.5.1. WARNING: DO NOT plug the Power Adapter into an outlet with a voltage other than that specified on the Power Adapter as it may cause electric shock and permanent damage to your device. This product is designed to operate within a range of 100 – 240 volts. Voltage converters DO NOT guarantee voltage compatibility. 8.5.2. WARNING: NEVER force the Power Adapter into an outlet; if it does not easily fit into the outlet, discontinue attempts as it may cause electric shock and permanent damage to your device. 8.5.3. WARNING: NEVER charge your device using a Power Adapter that does not meet specifications as listed in the Technical Description of this manual as it may cause electric shock and permanent damage to your device. 8.5.4. WARNING: AcceleDent Optima should not be stacked or located on or around other equipment that may create electromagnetic interferences.8.5.5. WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of AcceleDent Optima, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.8.5.6. WARNING: NEVER use a damaged Activator, Charging Case, USB Charging Cable, or Power Adapter as it may cause electric shock and/or further damage to your device. Refer to the section in this manual titled “Customer Service and Replacement” if your AcceleDent Optima or any of its components no longer work properly.
13BACK TO CONTENTSPage      of 248.5.7. WARNING: DO NOT place or store the device or accessories near any heated surfaces, as it could cause fire or permanent damage to your device. 8.5.8. WARNING: DO NOT place or store the Power Adapter or Charging Case where it will sit in a pool of water or where it can fall or be pulled into a bathtub, sink, or toilet. DO NOT reach for a Charging Case that has fallen into water. Unplug immediately.8.5.9. WARNING: DO NOT use accessories and cables other than those specified or provided by the manufacturer of this equipment. This could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.8.5.10. WARNING: DO NOT interconnect the Charging Case USB connector to equipment not specified in this manual.8.5.11. WARNING: The Federal Communications Commission does not allow any modifications or changes to the unit EXCEPT those specified in this manual. Failure to comply with this government regulation could void your right to operate this equipment.8.5.11.1.  No modification of the equipment is allowed. 8.5.11.2.  DO NOT attempt to replace the battery.8.6. Precautions8.6.1.  DO NOT use AcceleDent Optima without it being prescribed by an orthodontist or dentist. This is a prescription device which must be prescribed by an orthodontist or dentist. As with many medical devices, use by an individual without the proper issuance from an orthodontist or dentist can result in unintended consequences.8.6.2.  CONSULT your orthodontist or dentist prior to use of this product if you have had any recent craniofacial surgery (above the neck). 8.6.3. DO NOT share your AcceleDent Optima device. AcceleDent Optima is a single patient device (one patient to one device). Use of the device by more than one person may result in the transmission of viral and bacterial infective agents, even if an eort has been made to sanitize it.8.6.4.  DO NOT use a Mouthpiece that is damaged. Consult your orthodontist or dentist if the Mouthpiece no longer fits comfortably for any reason. Inappropriate cleaning may also cause damage to the Mouthpiece (see “Care” section).8.6.5.  DO NOT use AcceleDent Optima outside of the environmental conditions specified in the Technical Description of this manual.
14BACK TO CONTENTSPage      of 248.6.6. DO NOT use AcceleDent Optima outside of the electromagnetic environment specified in the Technical Description of this manual. 8.6.7. DO NOT clean any components or accessories in the dishwasher or microwave.8.6.8.  DO NOT boil, steam or dry heat sterilize any components or accessories. 8.6.9.  DO NOT rinse or submerge the Charging Case or accessories in water.8.6.10.  DO NOT attempt to service or perform maintenance on the device. This product contains no serviceable parts.8.6.11.  DO NOT attempt to charge the Activator with a Qi charger or any other wireless charger than the one supplied by OATI.8.6.12.  DO NOT expose Activator or Charging Case to UV light for an extended period of time as it may discolor.9.0 Help 9.1.  If you have any questions, concerns, or to report an unexpected operation or events regarding your AcceleDent Optima product, please contact the Customer Care Center for your country (visit www.acceledent.com for contact information). If you need immediate assistance, please contact the OrthoAccel Customer Service. 10.0 Technical Description10.1.   AcceleDent Optima complies with IEC 60601-1 and 3rd Edition Amendment 1; IEC 60601-1-11; IEC 60601-1-2 and 4th Edition; IEC 60601-1-6/62366; IEC 60601-1 2nd Edition Amendment 1 and 2. This information is intended to provide environmental operating conditions, transport/storage conditions, as well as IEC 60601 equipment ratings, electrical safety classification for the included equipment. Conditions should fall between the ranges detailed below.10.2.  The expected service life of AcceleDent Optima is two years.10.3. Environmental Operating ConditionsAmbient Temperature Range: 5-35°CMaximum temperature of any external component at maximum ambient
15BACK TO CONTENTSPage      of 24temperature: 50°CRelative Humidity Range: 15-90%Atmospheric Pressure Range: 700-1060 hPa10.4. Transport and Storage ConditionsAmbient Temperature Range: -20 to 50°CRelative Humidity Range: 30 - 93% RH 10.4.1. If AcceleDent Optima has been transported or stored at its maximum or minimum transport or storage conditions, the product shall be allowed to cool or warm for at least  two hours within the environmental operating conditions before use.10.5.  The Activator is a Type BF Applied Part, meaning this medical device has direct contact with the patient. Such devices shall be marked with   symbol.10.6.  A Power Adapter intended for connection to the USB port of the Charging Case shall comply with the relevant product standard e.g. IEC 60950-1 or IEC 62368-1 for IT-equipment and the IEC 60601-series for Medical Electrical Equipment. In addition, all such combinations – Medical Electrical Systems – shall comply with the safety requirements stated in the general standard IEC 60601-1, edition 3, clause 16. Any equipment not complying with the leakage current requirements in IEC 60601-1 shall be kept outside the patient environment i.e. at least 1.5 m from the patient support or shall be supplied via a separation transformer to reduce the leakage currents.10.7.  Use the provided Power Adapter or one that is certified as Class II adapter with double or reinforced insulation. Such Power Adapter shall be marked with   symbol or labeled Class II. The specifications of an appropriate Power Adapter are detailed below:10.7.1.  Certification: Class II, IEC 60950  Output Voltage: 5 V  Output Current Limit: 1 A  AC Input Voltage Range: 100-240 V  AC Input Current: 0.2 A  AC Input Frequency: 50/60 Hz10.8. Any person who connects external equipment to the USB connector has formed a Medical Electrical System and is therefore responsible for the system to comply with the requirements. If in doubt, contact a qualified medical technician or your local representative.
16BACK TO CONTENTSPage      of 2410.9.  The USB port located on the back of the Charging Case is for battery charging purposes only. The Charging Case has no data communication ability. The Charging Case shall have two means of protection, as Class II equipment, in order to prevent the parts from exceeding voltage, current, or energy limits defined by IEC 60601-1. As a precaution, the USB connector should not be touched or manipulated when connected to the Power Adapter and when the Power Adapter is connected to the supply mains. The Power Adapter can be isolated from the supply mains by unplugging the device. Do not position the charging case to make it dicult to operate the disconnection device (USB cable).10.10. Ingress Protection Class10.10.1.  The Charging Case is rated IP32; it is protected from intrusion of solid foreign objects greater than 2.5 millimeters, except dust or lint, and against vertically falling water drops when enclosure is tilted up to 15°.10.10.2.  The Activator is rated IP67; it is protected from total dust ingress and from immersion between 15 centimeters and 1 meter in depth.10.11. FCC Compliance Statement10.11.1.  This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.10.11.1.1. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment o and on, the user is encouraged to try to correct the interference by one or more of the following measures:- Reorient or relocate the receiving antenna.- Increase the separation between the equipment and receiver.- Connect the equipment into an outlet on a circuit dierent from that              to which the receiver is connected.- Consult the dealer or an experienced radio/TV technician for help.
17BACK TO CONTENTSPage      of 2410.11.2. Recommended separation distances:10.11.2.1.  AcceleDent Optima is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of AcceleDent Optima can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the AcceleDent Optima device as recommended below, according to the maximum output power of the communications equipment.Rated maximum output power of transmitterWSeparation distance according to frequency of transmittedm1150 kHz to 80 MHzd=1.2√P80 MHz to 800 MHzd=1.2√P800 MHz to 2.5 GHzd=2.3√P0.01 0.12 0.12 .23For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.10.12. CAN ICES -3 (B)/NMB-3(B) 10.12.1. This device complies with Industry Canada’s licence -exempt RSSs. Operation is subject to the following two conditions:(1) This device may not cause interference; and(2) This device must accept any interference, including interference that may cause undesired operation of the device.10.13. Environmental Specifications10.13.1.  Electromagnetic Requirements:10.13.1.1.  AcceleDent Optima complies with all applicable and required standards for electromagnetic compatibility. AcceleDent Optima is suitable for the home healthcare environment meaning all environments except for near active HF SURGICAL EQUIPMENT and areas of high intensity of EM DISTURBANCE. 10.13.1.2.  AcceleDent Optima transmits and receives electromagnetic energy at 288 kHz +/- 10 kHz for the purposes of wireless charging.
18BACK TO CONTENTSPage      of 24AcceleDent Optima complies with the Bluetooth Low Energy (BLE) standard; it transmits and receives electromagnetic energy through at 2.4 to 2.4835 GHz. The band is split into 40 2 MHz channels. The maximum transmit power is less than 1 mW.10.13.1.3.  See IEC 60601-1-11 clause 4.2.3.1 for applicable conditions. This device has been tested and found to comply with requirements for CISPR 11 Class B digital devices, which is designed to provide reasonable protection against interference in a residential setting. However, there is no guarantee that interference will not occur in a particular installation.10.14. Electromagnetic Emissions10.14.1. The Activator is intended for use in the electromagnetic environment specified below. The customer or the user of the unit should assure that it is used in such an environment. 10.14.2.  In accordance with Clause 4.3 of IEC 60601-1, AcceleDent Optima does not have Essential Performance.Emissions Test Compliance The  AcceleDent Optima uses RF energy only for its internal function; therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.RF Emissions CISPR 11Group 1RF Emissions CISPR 11Class B The system is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.Harmonic  emissions IEC61000-3-2N/AVoltage fluctuations / Flicker emissions IEC61000-3-3N/A10.14.3.  Bluetooth Low Energy:10.14.3.1.  Bluetooth, Bluetooth Low Energy, Bluetooth Smart and the Bluetooth logo are registered trademarks of Bluetooth SIG.10.15. Electromagnetic Immunity10.15.1.  Electromagnetic Immunity Part 110.15.1.1. The Activator is intended for use in the electromagnetic environment specific below. The customer or the user of the Activator should assure that it is used in such an environment.
19BACK TO CONTENTSPage      of 24IMMUNITY Test IEC60601 TestLevelCompliance Level Electromagnetic Environment - GuidanceElectrostatic discharge (ESD) IEC 61000-4-2±8 kV contact±2, 4, 8, 15 kV air±8 kV contact± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV airFloors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.Electrical fast transient/burstIEC 61000-4-4± 2 kV for power supply lines ± 1 kVfor input/output lines± 2 kV for power supply lines ± 1 kV for input/output linesMains power quality should be that of a typical domestic environment.Surge IEC 61000-4-5 ± 1 kV line(s) to line(s) ± 2 kV line(s) to earth ± 1 kV dierential mode± 2 kV common mode Mains power quality should be that of a typical domestic environment.Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11100% drop, 0.5 periods, 0º, 45º, 90º, 135º, 180º, 225º, 270º, 315º 100% dip, 1 period 30% dip, 25/30 periods Voltage Interruptions (all input current)100% drop, 0.5 periods, 0º, 45º, 90º, 135º, 180º, 225º, 270º, 315º 100% dip, 1 period 30% dip, 25/30 periods 100% drop, 5 secondsMains power quality should be that of a typical domestic environmentPower frequency (50/60 Hz) magnetic field IEC 61000-4-830 A/m 30 A/m Power frequency magnetic fields should be at levels characteristic of a typical domestic environment.Tested in X, Y, and Z axes.Can be waived due to a justification that there are no magnetic components. Assumes a minimum distance of 15 cm to magnetic source in actual usageNOTE: UT is the A.C. mains voltage prior to application of the test level.10.15.2.  Electromagnetic Immunity Part 210.15.2.1. The Activator is intended for use in the electromagnetic environment specific below. The customer or the user of the Activator should assure that it is used in such an environment.
20BACK TO CONTENTSPage      of 24IMMUNITY Test IEC60601 TestLevelCompliance LevelElectromagnetic Environment - GuidanceConducted RF IEC 61000-4-6Radiated RF IEC 61000-4-36V Amateur 0,15 MHz – 80 MHzEnclosure 10V/m 80% AM at 1 kHz or risk frequency 80 MHz – 2700 MHz6 VEnclosure  10V/mPortable and mobile RF communications equipment should be used no closer to any part of AcceleDent Optima, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. d=1.2√P80 MHz to 800 MHzd=2.3√P800 MHz to 2,5 GHzwhere P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a) should be less than the compliance level in each frequency range.b) Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.a) Field strengths from fixed transmitters, such as base stations for radio (cellular/ cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the AcceleDent Optima is used exceeds the applicable RF compliance level above, the AcceleDent Optima should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating AcceleDent Optima.b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.10.16 Material:10.16.1. AcceleDent Optima is made of medical grade, biocompatible materials. The AcceleDent Optima Activator has been tested and verified to have no Latex proteins. The AcceleDent Optima Activator has been tested and verified to have less than 0.1% by mass of DEHP.
21BACK TO CONTENTSPage      of 2410.17. Recommended separation distances10.17.1.1.  AcceleDent Optima is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of AcceleDent Optima can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the AcceleDent Optima device as recommended below, according to the maximum output power of the communications equipment.Rated maximum output power of transmitterWSeparation distance according to frequency of transmittedm1150 kHz to 80 MHzd=1.2√P80 MHz to 800 MHzd=1.2√P800 MHz to 2.5 GHzd=2.3√P0.01 0.12 0.12 .23For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.
22BACK TO CONTENTSPage      of 2411.0 Symbols and DefinitionsSymbol Description Lot number or batch code Reference or Catalog number S/N Serial Number P/N Part Number, Model Number or Reference Date of manufacture Manufacturer Use by date Follow Instructions for use Refer to instruction manual Rx Only Caution: Federal law restricts this device to sale by or on the order of a physician. Certification that the electromagnetic interference from the device is under limits approved by the Federal Communications Commission. Class II Equipment Type BF Applied Part: Devices that have direct contact with the patient, or parts that have long term contact with the patient. Interference may occur in the vicinity of equipment marked with the following symbol. Bluetooth enabled Authorized representative in the European Community European Conformity (CE) Mark, with notified body number
23BACK TO CONTENTSPage      of 24 Temperature limitation Separate collection for electric and electrical equipment Keep Dry Fragile Stand-by 2 0 min20 minutes for treatment Open Charging Case by pushing down on this part of the latchClose Charging Case by pushing down on this part of the latchHumidity limitationAtmospheric pressure limitation Not waterproof IP32 Is protected from intrusion of solid foreign objects greater than 2.5 millimeters and against vertically falling water drops when enclosure tilted up to 15°. IP67 Is protected from total dust ingress and from immersion between 15 centimeters and 1 meter in depth.
24BACK TO CONTENTSPage      of 24 OrthoAccel Technologies, Inc.6575 West Loop South Suite 200Bellaire, TX 77401, USANorth America: 1 866-866-4919 Spain: +34900839096United Kingdom: +448000988430France: + 33805080086Italy: +39800596833Germany: +498007244532Outside North America: +44 (0) 203-318-1915www.acceledent.comEmergo EuropeMolenstraat 15, 2513 BH, The Hague, The NetherlandsPhone: +31-70-345-8570 / Fax: +31-70-346-7299Australia RepresentativeEmergo Australia201 Sussex Street, Darling Park, Tower II, Level 20Sydney, NSW 2000, Australia Phone: + 61 (0) 2-9006-1662 / Fax: +61 (0) 2-9006-1010