PRIMETECH UCD300 Micro Infusion Pump Data Communication Unit Transmitter User Manual

PRIMETECH CORPORATION Micro Infusion Pump Data Communication Unit Transmitter Users Manual

Users Manual

 SMP/UCD-300
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e    Read  me  Thank you for choosing iPRECIO®, programmable infusion pump for Lab animal for your research/experiments.    This  quick  guide  show  you  how  to  use  iPRECIO®  pump  and  operate  iPRECIO® management software.      The iPRECIO® is for use in  Laboratory Animal Research ONLY. Not for huma use.    Please  obey  the  Guide  for  the  Care  and  use  of  laboratory  animal by your organization and or institute when use iPRECIO®.  NOTICE TO THE USER One or more of the following statements may be applicable for this equipment.  Notice (For Model UCD-300 only) The provided LAN Cable with a ferrite core shall be used when connect UCD-300 to a PC and do not replace the LAN cable with other LAN cable.  FCC WARNING This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.  Changes  or  modifications  not  expressly  approved  by  the  manufacturer  could  void  the user’s authority to operate the equipment.          For more information : PRIMETCH CORPORATION TEL:+81-3-3816-0851  E-mail : iprecio@primetech.co.jp URL :www.iprecio.com 2 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j    INFORMATION TO THE DIGITAL DEVICE USER REQUIREMENT by FCC (For Model UCD-300 only)  This equipment has been tested and found to comply with the limits for a Class B digital device,  pursuant  to  part  15  of  the  FCC  Rules.  These  limits  are  designed  to  provide reasonable  protection  against  harmful  interference  in  a  residential  installation.  This equipment generates, uses and can radiate radio frequency energy and, if not installed and used  in  accordance  with  the  instructions,  may  cause  harmful  interference  to  radio communications.  However,  there  is  no  guarantee  that  interference  will  not  occur  in  a particular installation. If this equipment does cause harmful interference  to  radio  or television reception, which can be determined by turning the equipment off and on, the user  is  encouraged  to  try  to  correct  the  interference  by  one  or  more  of  the  following measures: —Reorient or relocate the receiving antenna. —Increase the separation between the equipment and receiver. —Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. —Consult the dealer or an experienced radio/TV technician for help.3  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e    Contents  READ ME ..................................................................................................................................... 2 CONTENTS .................................................................................................................................. 4 ABOUT IPRECIO ...................................................................................................................... 5 ®THE WORLD FIRST WIRELESS PROGRAMMABLE INFUSION PUMP ......................................................... 5 ACCESSORIES WHICH INCLUDED IN IPRECIO  STD SYSTEM ........................................................... 5 ®IPRECIO  SYSTEM CONSTRUCTION ................................................................................................ 6 ®WORK FLOW .............................................................................................................................. 7 PRODUCT WARRANTY ............................................................................................................. 8 BATTERY LIFE............................................................................................................................ 9 INSTALL MANAGEMENT APPLICATION ............................................................................. 10 PREPARATION BEFORE START......................................................................................................... 10 INSTALL IPRECIO  MANAGEMENT SOFTWARE................................................................................ 10 ®CABLE CONNECTION OF DATA COMMUNICATION UNIT UCD-300 .................................................... 12 TCP/IP CONNECTION BETWEEN DATA COM UNIT AND CONTROL PC ................................................ 13 PREPARATION BEFORE SOFTWARE USE .......................................................................... 14 FOR THE FIRST TIME EXECUTE MANAGEMENT SOFTWARE ................................................................ 14 USER REGISTRATION .................................................................................................................... 14 ACCESS CONTROL OF SOFTWARE................................................................................................... 15 CHANGE LANGUAGE ...................................................................................................................... 15 NETWORK SETTING ...................................................................................................................... 16 CREATE INFUSION PROFILE ........................................................................................................... 16 CUSTOMIZE INFORMATION............................................................................................................. 20 START STUDY........................................................................................................................... 21 START NEW STUDY....................................................................................................................... 21 CATHETER CONNECTION, PLOMB ADMIN LIQUID AND ACTIVATION................................................... 23 PUMP DETECTION.......................................................................................................................... 25 KVO SETTING, START AND COMPLETE ............................................................................................ 25 INFUSION PROTOCOL SETTING ...................................................................................................... 27 START INFUSION PROTOCOL.......................................................................................................... 27 MONITOR THE RUNNING PUMP STATUS........................................................................................... 28 REPLACE AND/OR REFILL THE MEDICAL SOLUTION ......................................................................... 29 END OF STUDY ............................................................................................................................. 30 SAVE STUDY FILE.......................................................................................................................... 30 EXPORT STUDY FILE...................................................................................................................... 30 PRODUCT SPECIFICATION ................................................................................................... 31    4 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j  5  www.iprecio.com About iPRECIO®  The world first wireless programmable infusion pump  iPRECIO® SMP-300 model is the world first programmable pump which is complete implantable into a laboratory mouse.     Utilize high precision  mechanical  pump  technology that  was unique developed  and controlled by built in microprocessor. An administration protocol can be programmed with  dedicated  management  application software and pump will be work by the programmed  protocol  with  accurate  low  flow  rate  (L/H order) medical solution infusion. In addition the medical solution in the reservoir can be complimented and or replace through septum by transcutaneous even after implanted in an animal.     Accessories which included in iPRECIO® STD System  iPRECIO® system configured with the following accessories.  Small Lab animal micro infusion pump iPRECIO® SMP-300  iPRECIO® management system IMS-300  Data communication    unit UCD-300 (AC Adapter, LAN cable included)  Management Software  iPRECIO Quick guide (this guide)      (L) Pump in pre-sterilized pack and 5pcs box.    (R)Packaged data UCD-300
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e   iPRECIO® System construction     Number Top surface Bottom surface name 󰧈  Refill septum 󰧉  Outlet tube 󰧊  Auxiliary Wire antenna for radio communication 󰧋  Reservoir 󰧌  Micro cam 󰧍  Pump ID label 󰧎  Power switch 󰧏  LAN cable connect port   󰧐  AC Adaptor connect port 󰧑  Antenna (The antenna is permanently fixed)  6 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j   Work  flow  A study performed with iPRECIO®を, normally follow work flow below. 7  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e   Product Warranty  PRIMETECH CORPORATION (hereinafter PRIMETECH) guarantee the product and it’s material are free from defect within the below period which was distributed through the official sales channel. PRIMETECH do not responsible with any problem that caused by using  with  other company product. Also the product performance is not guaranteed when a customer uses it beyond the original specification.    The defect part will be replaced and or repair by free of charge except, in-transit accident, customer’s fault, breakage caused by abnormal/improper use and damnification cause by PRIMETECH recommendation of use.    Any defect parts back to PRIMETCH for replace, always need the RMA format and ship back through original sales channel.   Duration of guarantee  iPRECIO® Micro Infusion Pump SMP-300: 24 months after manufacturer  iPRECIO® Management System IMS-300: 12 months after shipping date   This  warrantee  represent  any  other  warrantee  and  include  an  implied  consent  of merchantability and dedicated fitness for purpose, but not limited to.    Liability Limitation Except the warrantee describe in above, PRIMTECH do not responsible  any  direct, indirect and special loss caused by violence of use condition within the maximum LAW protection. This includes any return loss and or opportunity loss (includes profit loss by contract) with direct, indirect, and damage.   This warrantee will be no longer valid when customer (include stuff, member and any other  bystander)  use  iPRECIO® for any other purpose against PRIMETECH sales conditions.        8 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j  9  www.iprecio.com  Battery life  iPRECIO® SMP-300 can continuous operate time below table during product warranty period (After 24 months from manufacturing date) Flow rate (μl/hr)  Battery life (hr) 0.1  408 0.5  384 1.0  360 5.0  216 10.0  144  Please note that below condition makes shorter battery life. :  In case of kept certain period of time without use after power on. (After power on, internal circuit becomes active and consume battery power by standby power requirement)    In  case  of  not  storage  under  proper  circumstances.  (Under  frequent temperature swing or exposed in over storage temp range in spec may cause increasing self-discharge of battery)
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e   Install management application  Preparation before start   iPRECIO® Management software install CD  Windows 7(32bit OS) installed and LAN port ready PC  Data communication unit UCD-300 (data com unit)  UCD-300 main body    AC adapter  LAN cable (Cross over cable with Ferrite core)   Install iPRECIO® management software   PC setting should be:  Off the screen saver  Get into the control panel Power option of “Change when PC sleep” and set “Put the computer to sleep” = “Never”   1. Run the “setup.exe” in a install CD. 2. When the startup window appeared, click “Next>”. Then setup wizard window comes up, click “Next>” again.     3. Chose the  program files store  folder  and click “Next>”. Subsequently confirmation screen open, then click “Next>”. For this step the installation will be started. 10 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j    4. When install complete windows appears, click “Close”.   5. Check start menu that the following software installed.  iPRECIO® Management Software IMS-300  Utility Base Station Manager  Utility Pump Database Manager  6. Three of those software has been installed, then the management software install is successful. 11  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e   Cable connection of data communication unit UCD-300  1. Plug in the AC Adapter to data communication unit and connect with the PC which the management software installed by cross-over type LAN cable.  Connect to LAN port of Control PC Plug in the AC Adapter to outlet(100-240V, 50/60Hz)         Ferrite core attached end shall be connected to UCD-300. 12 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j   TCP/IP connection between data com unit and control PC   1. Connect data com unit and control PC with LAN cable 2. Open Network and Sharing Center from control panel. 3. Click Change adapter setting, then right click on the data com Local Area Control and open Property.   4. Open Internet Protocol version 4(TCP/IPv4) property and input IP address below IP Address: 192.168.1.11 Subnet mask: 255.255.255.0 Then press OK to finalize.    The  Data  communication unit  factory default  IP address  configuration  is 192.168.1.10.  If  IP address  doesn’t  work  by  usage  environment  with  default  configuration,  please  change  IP address  in  accordance  with  “Appendix  A:  How  to  change  IP  address  of  the  Data communication unit”    5. With that the Data communication unit and control PC is connected by Internet Protocol. Subsequently network connection should be made in the management software. 13  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e    Preparation  before  software  use  For the first time execute management software   For the first time  run the  management software, the user  registration window will open automatically. To register user information, user management and access control function will be available. The  language  default  setting  is  in  English  and  after  user  registration the  language selectable function will be active to change.     User Registration   1. Input required information in the user registration window. The asterisk (*) marked item below are mandatory.  Institution & Department  Administrator 2. In addition, at least one operator registration required at this time. Input operator’s User Name, Login ID and Password (Confirmation required), then click on Add tab. The operator will be listed in right table.     3. To use Access control, check the “Enable Access Control” check box. Detail of Access Control function, please refer the nest page “Access control of software”. 4. After all fill in required column, click OK to finalize.   14 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j    Access Control of software  To utilize Access Control function, administrator can confine and or control of user who can reach the management software. To permit to access the software, check the “Enable Access Control” in the user registration window. Once  use  control  function  is  active, every  time  Login  ID  and  Password  is  required whenever the software executed.    Please note Login ID and Password. When forgot password, then software access will be denied.       Change Language 15  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e    1. Click Language menu 2. Select language from pull down menu, then click Change. 3. Now the language has been changed. (Below example shows Japanese)     Network Setting   1. Click network setting menu in tab. 2. Input default IP address(192.168.1.10) in the  Host IP  address column  for the data com unit.  3. Click “connect” 4.  After confirm “Status: connected”, then close the window. 5. If the status column shows “disconnected”, check the IP address setting. Please refer the “TCP/IP connection between data com unit and control PC”.      Create Infusion Profile  Administration protocol that perform in vivo pump of a subject animal of experiments, it is required create in the management software in advance and store as a “administration protocol”. The administration profile should be determined by each group of dosing preparation and concentrations and assign administration schedule.    1. Open Administration profile menu by clicking. 2. Administration window comes up. 16 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j    3. Add new administration profile, click “Add” button. 4. Group profile creating window will be opened.  17  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e   5. Input and or  select basic information. When chose  the injected  amount unit  as       μg/kg/hr(Dose) or mg/kg/hr(Dose), amount set by dose. It is therefore the animal weight range and concentration input are mandatory. 6. Input dose amount and duration from Step1 to create the infusion schedule. Input the value to the each boxes and press Enter to finalize dose. The starting time and finish time will be shown in elapsed time that accumulate the injection period. The battery life shown in the bottom of window indicate the expected battery life which calculated as a new battery. It may help to create an administration schedule.   Weight range should be input by maximum to minimum.  Please do not forget to finalize the data by pressing “Enter” key.  Do not refer the battery life when the pump was re-use and or re-programming. The indicated battery life was calculated from the battery capacity as a new one.   7. When  ditto the same schedule, the use the “Repeat Setting”.  Check  the “Repeat  Start” box for start of repeated schedule step then check the “Repeat  End” box for termination ditto schedule. For number of cycle, input “Times” of “Repeat Start” column and finalize to press Enter key.   8. After all information input, press OK then new administration profile will be registered in the Infusion Profile list. 18 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j    9. When to change the existing Infusion Profile, select intended Infusion Profile and click “Edit”button. If want  to  eliminate the existing Infusion Profile  on the list, select  the intended Infusion profile to delete and click the “Delete” button.  10. Close the Infusion profile window by clicking “Close” button.  19  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e   Customize information Pull down menu registration and elimination are possible for Study Setting. And KVO setting of flow rate and duration also set in advance.   1. Open customize window from “Customize(Z)” menu in the tool bar. 2. Input  a  desire  information  in  the  Animal  Species,  Animal  Strain  and   Administration Route, then click Add button in the below of each boxes. 3. If would like to delete existing information, select information and click Del button in the below of each boxes. 4. Input the numerical value of KVO Flow Rate and Duration. 5. To complete the input value, click OK button in the left bottom of the Window.  20 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j   Start Study  Start New Study iPRECIO® Management software has a function to abort the study during the KVO and Administration with pumps which implanted in under study animals. For example, to be able to except individual animal from study that is slow-healing than others during KVO. Also even during Administration to be able to deselect individual animal that is difficult to continue Administration by some trouble.      KVOChange medical solutionAdministration The Data access Radio system of iPRECIO®is not Broadband type in use. For that reason, the management system can control only one pump in the same time, then it may require long time to complete the following data communication.   Pump detection  Start KVO  Abort KVO  Complete KVO  Start Administration protocol  Abort Administration protocol  Finish Administration protocol  Extract executed study log 21  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e    1. Click New from file menu bar. 2. Appear Study Header window. (*) marked items are mandatory. 3. Click Next after fill the information input boxes.   4. Pumps and Animals setting window comes up. Input intend animal number in the Number of Animals and click Set button. Default Number of Animal is 20.   The information input column will be active in answer to input animal number.   5. Input and select by pull down for except Pump ID and Calibration Factor, such as individual information. (Animal ID, weight[g] etc.)   6. Can copy and paste of the column which already input individual information to another column.   22 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j   Catheter connection, plomb admin liquid and activation. Activate the pump by switch on to communication enable. If required  connect additional catheter as necessary.   To fill administration liquid till the end of catheter before power on, because once a pump switched on, the internal cam will start moving to the  initial position to block the inner tube.   1. Open pump pre-sterilization pack in clean environment. 2. Cut, replace tube and/or catheter connection of pump as necessary.    Each pump dead volume can be calculated in KVO setting, so keep final tube length and inner diameter or measured dead volume need to be recorded.  3. Fill the admin into the end of tube with 27G syringe needle through upper port.  23  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e   4. Press down the back side power button tenaciously with pointed top.   5. When activation successfully, inner cam rotates about 100 seconds and most rear pin block the tube. (When a bland new pump in use)  6. After activate with switch on, a pump is ready to implant to an animal under test. Factory-configured Block tube mode   Once activate an infusion pump, it is impossible to turn off power of the pump.     24 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j   Pump detection   1. Place pumps by recommended position with data com unit UCD-300. 2. Click  Detect  button.  During  Communication  progress  bar  is  shown,  the communication enable pump will be detected and assign individual animal ID.    If there is no response from pump, please check the followings:  It may take long term to detect pump due to radio wave back ground environment.  Verify  PC  and  UCD-300  connect  properly.  For detail, please refer “TCP/IP connection between data com unit and control PC” and “Network Setting”.  Verify pump activation. For detail, please refer “Catheter connection, plomb admin liquid and activation”.  3. After all pump detected, then click Next button.   KVO setting, start and complete  KVO is abbreviate of Keep Vein Open, semantics to prevent the catheter edge clog by continuous  liquid  dosing  by  low  flow  rate.  Be  normally  administrate  biocompatible liquid such as saline as KVO by low flow rate after implant surgery and prevent clog of catheter end during animal healing period. After animal comeback form the surgery, the KVO liquid in reservoir can be replaced by medicine liquid from the septum. And acknowledge user the timing that when the medicine liquid will reach the catheter end by pump and appurtenant catheter dead volume calculation.   1. KVO setting window pops up. If KVO is not required, then check skip KVO and click Next>> button. 2. Click individual column dead volume setting. 3. Input the required information for dead volume calculation and click OK button. 25  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e     Even if there is no need to change dead volume setting, it need to open every animal  dead  volume  setting  and  determine  the  dead  volume  choice.  If  this determine  was  lost,  the  dead  volume  and  flash  rate  will  not  apply  to  the administration start schedule.  4. Check KVO flow rate and KVO duration and if it’s ok then click to KVO start. The pump status will be KVO after start KVO. Pump status color will change in accordance with communication stats, red YellowGreen. When the color will be Green, then injection is started.   During KVO is running, please DO NOT STOP the management software.  5. When the pre-set KVO period will end, pumps will stop and Pump status changes into Finished.    Click Next>, the injection schedule setting. All of animal looks fully recovered before pre-setting KVO period, you can abort all KVO by clicking abort as a pack and start administration schedule. When some animal can not heal enough during KVO period and need to terminate the experiment of that animal, then click the remove button from individual animal column.     After  KVO stopped, please check the Pump status as “Finish”. Demepnd on the communication condition, “Finish” status appears very slow after prompt to end KVO from management software due to communication condition.   26 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j   Infusion Protocol setting   1. Group information window appears. 2. Chose  infusion  group  name  (Profile)  from Group ID pull down menu and set Flushing Rate and Administration Start Time, then click Set button.   3. When click Stagger, the administration will be staggered from set time in that value. 4. If the proper infusion group name does not found, then create new one. Please refer “Create Infusion Profile”. 5. After all animal are assigned to profile group, click Next>> button.   Start Infusion Protocol   1. Infusion Protocol window comes up. 2. Click Start all or click start for individual animal to start Infusion Protocol.    During  Infusion  Protocol  is  running,  please  DO  NOT  STOP  the  management software.  27  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e   Monitor the running pump status   1. Open Monitor window by clicking Monitor (M) from menu bar.   2. Upper part of window shows Study Name, Group and Animal. Select Group  ID from pull down menu in Group column and Animal ID from Animal column. 3. There is Refill Log table in the Right side. At first input the medical solution refill time and volume, then click Enter button. For detail, please refer “Replace and/or refill drag solution”. (Nest chapter) 4. The  estimated  remain  solution  volume  is  shown  by  calculated  from  infusion schedule and duration in the bottom of window. Also software alarm will be appear when the calculated volume reaches the Alarm Setting volume. 5. The Infusion Profile Graph locates in the left side of window. Clicking “Time Unit” button upper right, the Graph axis unit can be switched. Also clicking “Schedule” button bottom right, the Infusion schedule table will be shown. TO return, just click “Graph” button.   28 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j   Replace and/or Refill the medical solution   1. Open Monitor (M) window. 2. Verify the estimated remain volume in the bottom area. 3. Make pump implanted animal immobilization by weak anesthesia or retention. 4. When replace the solution, aspirate remain fluid in reservoir by 27G syringe needle through the skin to port. Subsequently fill the replacing solution into syringe and refill it to reservoir. To restock the drag solution, just fill the solution in a loaded syringe through the skin to port. 5. Input Refill Time and Volume, click Enter button. When executed by dose mode, input new weight and click Calculate button, recast New Concentration result will be displayed to keep dose condition.   6. After fix the Refill Log, the table record automatically the Refill time, volume and comments (weight and density). To remove the refill log, select target refill log and click “Delete” button.  29  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e   End of Study  After all infusion protocol finished, click “Study End” makes automatically record study date and finish study. If study should be abort during infusion schedule, click “Finish All” then click “Study End”    Save Study file     1. Click  Save  Study  from  the  File  (F)  in  menu  bar.  The  data  file  stored  in C:\Users\XXX\AppData\Local\iPRECIO\study 2. The save data can be loaded from “Open” in the File (F) in the menu bar.  Export Study file   1. Chose “Export” from the File (F) in menu bar. 2. Select folder to export, save with file name. 3. The  exported  data  format  is  “csv”.  To view the data, simply import it by regular spreadsheet program.   30 www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.j   Product Specification  Micro infusion pump for small laboratory animal iPRECIO®Model:SMP-300 Administration method  Rotary finger type (Patented: Peristaltic mechanism) Reservoir  130μL Built-in reservoir: (Material) Medical grade SEBS The  Maximum  capacity  is  130 μL,  but  out  of  accuracy guarantee  range.  To  keep  accuracy, only within 100μL is guaranteed.     Inner Tube: (Material) Medical grade SEBS Tube Outer Tube: (Material) Medical grade SEBS (3Fr) (Size) ID=0.5mm, L=140mm Outer tube is detachable. Replace/Refill port  2mm size refill port stick out in the pump. (Recommend below 27Ga size needle) Data communication  ISM band radio (Frequency: 916.1MHz) Com unit:Within 3m from UCD-300 Antenna  Wire antenna with pump (L: ca. 50mm) Coating by Polyurethane(1.2Fr) Titanium wire(0.1mm) Programmable Flow Rate (FR 0.1~10.0μL/hr. (setting resolution:0.1μL/hr.  Discharge flow by 360 degree roll of cam:3.4μL Infusion accuracy  ±5 (Under pressure 0~80cmH2O Battery Life  Continuous motion FR 0.1μL/hr.: 19days,   FR 10.0μL/hr.: 1 week Power on Activation  By pressing down of the power on nub switch on the bottom of pump to pump starter. (Coinstantaneously activated) Once activated, it’s not reversible to switch off.   Size  24.8(L)  15.0(W)  7.2 (H) mm (Except accompanying catheter and wire antenna) Weight  3.3g (Include accompanying catheter and wire antenna) Solid content  2.15cc Type of usage  Disposable Package  EOG Pre-Sterilization pack (5 package/box) Sterilization TTL  2yrs (Built in batter causes self-discharge during storage)  Data com unit (Model UCD-300) Interface with control PC  LAN ( Cross type cable with ferrite core) Control through Ethernet by setting fixed IP address to the Data com unit. Communication way with pump ISM band radio (Frequency: 916.1MHz) Not broadband type. One –on one communication between Data com unit and individual pump. One  unit  controllable  20~25 pumps (Theoretical value without time expense ) 31  www.iprecio.com
iPRECIO® SMP/UCD-300 Quick guide ver.1.0.0.e  32 www.iprecio.com pump number  Power source  AC adapter cable included. [Weight] 73g [Cable length] 1.85m [I/P] 100-240V 50/60Hz 25VA  [O/P] DC 5V 2A Size and Weight  120(L)  77 (W)  31(H) mm  167g STD accessory  Anchorage to  cage(Velcro)   LAN cable with ferrite core (3m)  1 cable  Management software Supported OS  Windows XP  Windows 7 (32bit)/Japanese  English