PRODUCT INSTRUCTION SHEET UVT-LED Sensor Installation and Operation Manual UVT-LED-H UVT-LED-PW UVT-LED-SW Patent No,: US 8,927,922 B2, Jan. 6, 2015 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 1 of 30 11751 Markon Drive Garden Grove, CA. 92841 USA Document Number 480437 Ph +1(714) 895-4344 Fax +1(714) 894-4839 sales@sensorex.com www.sensorex.com FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Table of Contents 1. Welcome 4 2. Getting to know your UVT Sensor 4 3. Warnings and Precautions 6 4. Introduction to UV transmittance (UVT) 7 5. UVT-LED Sensor Specifications 5.1 Operating Specifications 8 5.2 Dimensional Specifications 9 6. UVT-LED-PW 6.1 UVT-LED-PW Sensor overview 11 6.2 First Time Setup and Preparation 12 6.2 Cleaning Procedure 13 6.3 Clean in Place (CIP) 13 6.4 Calibration 14 6.5.1 Installation a) Weld on Flange(UVT0009) 14 b) 2" NPT Adapter (UVT0008, UVT0021) 15 c) 3" Sanitary Tee (UVT0015), 2" Sanitary Tee (UVT0024) 16 d) 2" PVC tee complete assembly (UVT0023) 17 7. UVT-LED-SW 7.1 UVT-LED-SW sensor overview 18 7.2 First Time Setup and Preparation 19 7.3 Cleaning Procedure 20 7.4 Calibration 21 7.5 Taking Readings 21 7.6 Installation 22 8. UVT-LED-H 8.1 Calibration 26 8.3 Taking Measurements 26 8.4 Cleaning 26 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 2 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Table of Contents 9. Troubleshooting (-PW & -H) 27 10. Troubleshooting (-SW only) 28 10. Ordering Information 29 11. Product Warranty 30 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 3 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Welcome to Sensorex UVT-LED! Thank you for choosing our UVT-LED UV-transmittance sensor. For over forty years, Sensorex has been the preeminent supplier of leading edge technology and innovative products to measure and monitor the quality of water. Today, we remain focused on the continual advancement of water quality monitoring products and technology. The UVT-LED is the latest Sensorex innovation and our first optical-based measurement system designed specifically for simple, reliable UV-transmittance monitoring. The Sensorex UVT-LED is the world’s first family of UV-T monitors that uses a UV-C LED instead of a mercury-based lamp as a light-source. The use of a UV-C LED in the UVT-LED ensures extremely stable readings in all conditions, over an extended lifetime. Getting to know your UVT-LED Sensors Backlit LCD display Backlit LCD display READ/CALIBRATE button READ/CALIBRATE button 1.5” Sanitary Connection Wiper Assembly Optics UVT-LED-PW 5-Pin Connector Connects to UVT0002 cable assembly (UVT-PW) or charging cable (UVT-LED-H) Optics UVT-LED-H Used for all on-line applications. Unit is wiped using patented technology. Use in water with a higher concentration of dissolved and suspended solids. Used for benchtop and off-line field measurements. Unit comes standard with a rechargable internal battery and charging cord for cordless operation. Product Label Includes: Version, Serial No., patent information Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 4 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Getting to know your UVT-LED Sensors (cont.) 1” NPT. Attach to UVT-SW-004 or -005 mounting hardware Touch magnet to this point to take a reading or calibrate UVT-LED-SW UVT-SW cable assembly Wiper Assembly Optics UVT-LED-SW Used for all open channel applications in which the unit is completely submerged. Unit is wiped using patented technology. Use in water with a higher concentration of dissolved and suspended solids. Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 5 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Warnings To avoid risk of damage to monitor, do not open the device. Refer servicing to qualified personnel only. Do not use the provided power/data cord if it is damaged. Use of a damaged cord may result in damage to the device or fire. Do not place device in a location near heat sources, or in a place subject to direct sunlight, excessive dust or mechanical shock. Do not pinch the power/data cord between heavy objects Do not operate sensor above 50°C because this can damage the UV Light source Device should only be used with provided 24V DC power supply or used with a site supplied 24V DC source. Monitor contains a UV LED emitting device. Do not look at the probe of an operating monitor without wearing proper eye and skin protection. Install unit so that power cord is protected from mechanical damage and can be easily unplugged from device. Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 6 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Introduction UV-T is a measure of the amount of UV energy remaining after passing through a material (e.g., water sample or quartz) over a specific distance. UV Absorbance is a measure of the amount of UV light that is absorbed by a substance (e.g., water, microbial DNA, lamp envelope, quartz sleeve) at a specific wavelength (e.g., 254 nm). UV-Transmittance is related to UV Absorbance by the following equation (for a 1cm path length): % UVT = 100 x 10-A. UV-Transmittance is typically; • measured at 254 nm, corresponding to the peak output of mercury-based UV lamps • measured over a path length of 1-cm • represented as a percentage UV-T has a strong effect on the dose delivery of a UV treatment system. As UV-T decreases, the intensity throughout the reactor decreases, which reduces the dose the treatment system delivers. UV treatment systems are sized to deliver the required UV dose under specified UV-T conditions for the application. UV-T will typically vary over time due to changing concentrations of compounds, seasonal effects and changes in biological activity of microorganisms within the water source. UV-T monitors are commonly used as an input to a dose control strategy of a UV treatment system, helping to reduce over or under dosing. They are also used to monitor water quality to help diagnose operational problems. Water with a measured value of 90% UV-T, demonstrating the rate in which available UV light degrades, as distance from light source increases Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 7 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Specifications Measurement Technique: UV absorption method (single beam technique), reagent-free Measurement Wavelength /Pathlength: 254nm/1cm Measurement Range: 10 -100% UV transmittance (UVT) Accuracy: Resolution: 0.1% T +/- 1.0% T (full scale) Sample Flow Rate/ Maxium Pressure: n/a, 150psig Display: LCD with LED backlight Measuring Interval: 60 seconds (UVT-LED-PW & -SW only) Inspection Interval: 6 months(UVT-LED-PW & -SW only) User Maintenance: 1hr/month typical Power Supply: 20-28V DC, maximum 200Ohm load, customer supplied (UVT-LED-PW & -SW) 24V DC to charge battery (UVT-LED-H) Output: 0-100% UVT (UVT-LED-H-WD, UVT-LED-H-NA only) 4-20mA (UVT-LED-PW & -SW) Temperature Range: 32-122 °F (0-50 °C) Weight: UVT-LED-H - 0.78lbs (353g), UVT-LED-PW - 1.58lbs (717g), UVT-LED-SW - 1.78lbs(807g) Dimensions: 7.60"H x 2.60" cap x 1.25" probe (19.1cm H x 6.4cm Dia) - UVT-LED-H & -PW only, 9.4"H x 3.00" cap x 1.25" probe (23.9cm H x 7.6cm Dia) IP Ingress Rating: IP66 housing, IP68 measurement head Certifications: CE Warranty: 2 Year Limited Warranty Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 8 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Outline and Dimensions UVT-LED-H FRONT LEFT READ/CALIBRATE button 2.60” RIGHT % UVT display (66 mm) REAR 5-pin connector for charging cable 7.60” (193 mm) Product label with Serial Number & Version Number UV-C LED Light source photodiode detector 1.25” (32 mm) UVT-LED-PW FRONT LEFT READ/CALIBRATE button 2.60” (66 mm) REAR RIGHT 5-pin connector for charging cable % UVT display 7.60” (193 mm) Product label with Serial Number & Version Number 3.00” (76 mm) Wiper assembly UV-C LED Light source 2 inch sanitary flange fitting (seal provided with installation fittings) photodiode detector 2.00” (50mm) 1.25” (32mm) Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 9 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Outline and Dimensions(cont.) UVT-LED-SW UVT-SW-001, -002, -003 3-conductor cable UVT-SW-001 = 10mtr UVT-SW-002 = 20mtr UVT-SW-003 = 30mtr Top View of 8-Pin connector O-rings to seal against UVT-SW-001, -002 or -003 Cable cord grip Threads to attach UVT-LED-SW cable using knurled nut 5.9” (150mm) 9.4” (239mm) Touch magnet provided to this surface to activate and calibrate the sensor 3.5” (89mm) 1” NPT thread. Connect to UVT-SW-004 or UVT-SW-005 mounting hardware 5.0” (127mm) Knurled nut to attach cable to UVT-LED-SW sensor. 1.43” (36.3mm) 1.25” (31.8mm) 3.00” (76mm) Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 10 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-PW Overview (-PW) The Sensorex UVT-LED-PW can be used in open or closed water systems, for either continuous on-line measurement of a process, or off-line samples, of any type of water, including wastewater and drinking water. This patented UV-T monitoring device includes a UV-C LED light source, a UV sensor and wipers attached to a rotating measuring head. It is operated by placing the measurement head into a water sample. The water sample may be stationary or flowing. The water sample flows through the measurement head, in between the UV light source and the UV sensor. The UV sensor measures the amount of UV light that the UV sensor is exposed to, measuring the amount of UV light that passes through the fluid sample. After the UV measurement of the water sample is collected, a quartz cylinder having a known UVT value is moved in between the UV light source and the UV sensor and another measurement is taken. By comparing the UV measurement taken through the quartz cylinder and the UV measurement taken through the fluid sample, a UVT value for the fluid is determined. In addition a third measurement is taken with the UV light source powered off providing a zero point calibration measurement for the UV sensor. When the quartz cylinder is rotated into a position between the UV light source and the UV sensor, a wiper, attached to the rotating mechanism, wipes the windows in between the UV light source and the water sample and the UV sensor and the water sample, thus removing any contaminants or buildup of constituents from the water sample on the windows. See FIG. 1. UV-C Light Source Measurement Head Wiper Quartz Cylinder UV sensor FIG. 1 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 11 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-PW First Time Setup and Preparation Step 1 - Unpacking your UVT-LED-PW Make sure you have all parts in the box. Review the Packing List included in the product box for details. Step 2: Cable Assembly Preparation A 10meter (33ft) cable assembly is included with each UVT-LED-PW (see FIG 2.) 1 inch of the cable jacket is stripped off for your convenience (see FIG 3). If you want longer leads, remove as much of the gray cable covering to reveal the length of wire you require. Next, strip the red, black ,and white wire tips to desired length and tin with solder to prevent wirings from fraying. Trim off braided shield and string filler even to cable jacket. FIG. 2 FIG. 3 Step 3 - Powering your UVT-LED-PW Connect the 5-pin connector from cable assembly UVT0002 to 5-pin connector on UVT-LED-PW and rotate the knurled part of the conenctor until connection is complete (see FIG 4.). Wire the stripped end of the cable RED, WHITE, BLACK wires to your 4-20mA input device as shown in FIG 5. The LCD display will illuminate and will display "0" and then will take a reading within 10 seconds after the wiper actuates (you will hear a "click" sound). - + Red Black + _ White 24V DC Current Signal(4-20mA) UVT-LED-PW 5-pin connector of UVT0002 cable GND PLC Analog current + FIG. 5 FIG. 4 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 12 of 30 11751 Markon Drive Garden Grove, CA. 92841 USA Document Number 480437 Ph +1(714) 895-4344 Fax +1(714) 894-4839 sales@sensorex.com www.sensorex.com FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-PW Cleaning Procedure Cleaning with Deionized water is recommended every 30days followed by recalibration off-line. When fouling of the quartz windows are suspected, the Sensorex UVT-LED-PW should be cleaned according to the following procedure: 1) Remove Sensorex UVT-LED-PW from the process. 2) Fill up a small container with a 15-20% solution of phosphoric acid (e.g. Calcium Lime and Rust remover CLR) or food grade citric acid. Fill container high enough to completely submerge the measuring portion of the tip in the 15-20% solution of phosphoric acid or food grade citric acid (see FIG. 6). 3) While the unit is still powered, allow unit to soak for 30 minutes. 4) After 30 minute soak, rinse measuring tip with de-ionized water to rinse off excess cleaning solution. 5) Fill container with isopropanol, high enough to fully submerge measuring tip. 6) Submerge Measuring tip in isopropanol for 10 minutes while unit is powered on. 7) Using the UVT-LED cleaning swab (part #UVT0011), gently clean the quartz measuring windows in front of the sensor and the UV-C light source. 8) Place unit in de-ionized water high enough to fully submerge measuring tip. 9) Submerge measuring tip in de-ionized water for 10 minutes. 10) Re-calibrate monitor according to recalibration procedure. Fill all solutions above level shown with red arrow Clean In Place (CIP) FIG. 6 When performing a CIP with the Sensorex UVT-LED-PW, the temperature of the water should not exceed 50˚ C. If the temperature exceeds 50˚ C, then the Sensorex UVT-LED-PW should have power removed. Operating the Sensorex UVT-LED-PW above temperatures of 50˚ C can damage the UV-C LED. Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 13 of 30 11751 Markon Drive Garden Grove, CA. 92841 USA Document Number 480437 Ph +1(714) 895-4344 Fax +1(714) 894-4839 sales@sensorex.com www.sensorex.com FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-PW Calibration Calibration should be done every 30 days after cleaning with deionized water. It is also recommended to verify sensor accuracy weekly using a hand-held UVT-LED-H sensor. Step 1 - Place your UVT-LED-PW into de-ionized (DI) water. Make sure it is inserted as shown in FIG 7 so that it is above the opening for the wiper and LED. Shake to remove air bubbles. Step 2 - Press and hold the "READ/CAL" button for at least 3 seconds as shown in FIG 8. ! USE DE-IONIZED WATER ONLY FIG. 7 FIG. 8 Installation The UVT-LED-PW can mount into a few different installation types: A) Weld on SS316 sanitary flange adapter (UVT0009) to pipe. Weld #4 to pipe --> Install #2 into groove on top of #4 --> Insert #1 (UVT-LED-PW) into #4 --> Clamp in place with #3 by installing clamp around #4 and flange of #1, then tighten wing nut of #3 until tight and no leaking occurs. (see FIG. 9) FIG. 9 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 14 of 30 11751 Markon Drive Garden Grove, CA. 92841 USA Document Number 480437 Ph +1(714) 895-4344 Fax +1(714) 894-4839 sales@sensorex.com www.sensorex.com FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-PW Installation Cont. B) 2" NPT threaded adapter (UVT0008 - SS316, UVT0021 - PVC). Apply threaded sealing tape or paste to threads of #4 --> Thread #4 into #5 (customer supplied) --> Install #3 gasket into groove of #4. --> Insert #1(UVT-LED-PW) into #4 --> Clamp in place with #2 by installing clamp around #4 and flange of #1 then tighten wing nut of #2 until tight and no leaking occurs (see FIG. 10). FIG 11 shows part detail. FIG. 10 1.99” Groove for sanitary gasket 1.5” Sanitary flange Wrench flats x4 2” male NPT thread FIG. 11 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 15 of 30 11751 Markon Drive Garden Grove, CA. 92841 USA Document Number 480437 Ph +1(714) 895-4344 Fax +1(714) 894-4839 sales@sensorex.com www.sensorex.com FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-PW Installation Cont. C) 3" Sanitary tee kit (UVT0015) - refer to FIG. 12. 2" Sanitary tee kit (UVT0024) refer to FIG 12 Install #5 (x2) to grooves of #4 --> Install #6 (x2) to 4 with #5 (gasket) in-between --> Use clamp #7 (x2) to hold #6 (x2) to #4. Install #3 into groove on top of #4. Insert #1 into #4 . Clamp #2 around #1 and #4 and tighten using wing nut of #2 until #1 is installed tightly (no leaks) into #4. Thread #8 (x2) into #6 (x2) using sealing tape or paste to avoid leaks. It is recommended to install this into a bypass loop with valves on each side of the assembly and a flow meter on the inlet side. See FIG 13. 1. 2. 5. 6. 6. 1. UVT-LED-PW sensor 2. 1.5” Sanitary clamp 3. 1.5”Sanitary gasket 4. Sanitary tee (3“(UVT0015), 2”(UVT0024)) 5. Sanitary gasket x2 (3“(UVT0015), 2”(UVT0024)) 6. Flange cap x2 (3“(UVT0015), 2”(UVT0024)) 7. Sanitary clamp x2 (3“(UVT0015), 2”(UVT0024)) 8. 1/4” NPT Barbed fitting x2 3. 4. 8. 7. FIG. 12 Suggested By-pass Installation for UVT-LED-PW Sensor Flow Main line Back to main line (UVT0015 only) Flow meter Sensor + Flow Cell (UVT0015 or UVT0023) Valve To drain (UVT0023 only) UVT-LED-PW Sensor 1.5” Sanitary Clamp (included with UVT0023) 1.5” Sanitary Clamp (included with UVT0015 & UVT0024) UVT0015 (3”) UVT0024 (2’) UVT0023 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 16 of 30 UVT-LED-PW Sensor FIG. 13 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-PW Installation Cont. D) 2" NPT PVC tee kit (UVT0023) - refer to FIG. 12 Note: Use sealing tape or paste on all threaded connections to avoid leaks. Thread #6 (x2) into #5 --> Thread #7 (x2) into #6 (x2) --> Thread #8 (x2) into #7 (x2). Thread #4 into #5. Install #3 into groove on top of #4. Insert #1 into #4 . Clamp #2 around #1 and #4 and tighten using wing nut of #2 until #1 is installed tightly (no leaks) into #4. It is recommended to install this into a bypass loop with valves on each side of the assembly and a flow meter on the inlet side. See FIG 13. See FIG 14 for all parts inluded in UVT0023. Suggested mounting orientation is shown in FiG 15. 1. 1. UVT-LED-PW sensor 2. 1.5” Sanitary clamp 3. 1.5”Sanitary gasket 4. 2“ PVC adapter 5. 2” NPT PVC SCH 80 TEE 6. 2“ x 1/2” PVC reducer 7. 1/2“ x 1/4” reducer 8. 1/4” NPT to barb quick disconnect fitting 2. 20 DEG UVT-LED-PW sensor Position Wiper slot upward and perpedicular to flow. Wall or panel 3. 4. 5. 6. 7. 8. UVT0023 flow cell Mounting Strap x 2 FIG. 14 FIG. 15 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 17 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-SW Overview (-SW) The Sensorex UVT-LED-SW is designed for use in open channel/submersion UVT monitoring of any type of water, including wastewater and drinking water. This patented UV-T monitoring device includes a UV-C LED light source, a UV sensor and wiper attached to a rotating measuring head. It is operated by submerging the sensor mounted to installation hardware into a flowing water channel. The water sample flows through the measurement head, in between the UV light source and the UV sensor. The UV sensor measures the amount of UV light that the UV sensor is exposed to, measuring the amount of UV light that passes through the fluid sample. After the UV measurement of the water sample is collected, a quartz cylinder having a known UVT value is moved in between the UV light source and the UV sensor and another measurement is taken. By comparing the UV measurement taken through the quartz cylinder and the UV measurement taken through the fluid sample, a UVT value for the fluid is determined. In addition a third measurement is taken with the UV light source powered off providing a zero point calibration measurement for the UV sensor. The sensor is calibrated by actuating with a magnet and placing in a sample of DI water. Measurements are taken by actuating the sensor again with a magnet and then measurements are automatic every 1 minute and are output as 4-20mA scaled to 0-100% UVT. When the quartz cylinder is rotated into a position between the UV light source and the UV sensor, a wiper, attached to the rotating mechanism, wipes the windows in between the UV light source and the water sample and the UV sensor and the water sample, thus removing any contaminants or buildup of constituents from the water sample on the windows. See FIG. 1. Note that UVT-LED-SW does not have a display but all other components match FIG 1. Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 18 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-SW First Time Setup and Preparation Step 1 - Unpacking your UVT-LED-SW Make sure you have all parts in the box. Review the Packing List included in the product box for details. Your UVT-LED-SW sensor is supplied with an alcohol cleaning swab and a 10mtr cable (UVT-SW-001). Step 2: Cable Assembly Preparation: A 10meter (33ft) cable assembly is included with each UVT-LEDSW. Assemble the cable to the 1" adapter of the UVT-SW mounting assembly. After the assembly is completely mounted then attach the sensor to the cable by aligning the connector parts and tightening the knurled nut clockwise until it stops (FIG 16). 1 inch of the cable jacket is stripped off for your convenience (see FIG 3). If you want longer leads, remove as much of the gray cable covering to reveal the length of wire you require. Next, strip the brown, white, and green wire tips to desired length and tin with solder to prevent wirings from fraying. Trim off braided shield and string filler even to cable jacket. Step 3 - Powering your UVT-LED-SW After the cable assembly has been connected to the UVT-LED-SW in step 2, wire the stripped and of the cable BROWN, WHITE, GREEN wires to your 4-20mA input device as shown in FIG 17. UVT-SW cable Knurled nut to attach cable to UVT-LED-SW sensor STEP 1: Align connector from UVT-SW cable with connector of UVT-LED-SW 24V DC + _ Current Signal(4-20mA) Green Threads to attach UVT-SW cable using knurled nut - + White STEP 2: Rotate knurled nut clockwise to tighten UVT-SW cable to UVT-LED-SW sensor Brown O-rings to seal against UVT-SW-001, -002 or -003 UVT-LED-SW UVT-LED-SW sensor GND PLC Analog current + FIG. 16 FIG. 17 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 19 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-SW Cleaning Procedure Cleaning with Deionized water is recommended every 30days followed by recalibration off-line. When fouling of the quartz windows are suspected, The Sensorex UVT-LED-SW should be cleaned according to the following procedure: 1) Remove Sensorex UVT-LED-SW from the process. 2) Fill up a small container with a 15-20% solution of phosphoric acid (e.g. Calcium Lime and Rust remover CLR) or food grade citric acid. Fill container high enough to completely submerge the measuring portion of the tip in the 15-20% solution of phosphoric acid or food grade citric acid (see FIG. 18). 3) While the unit is still powered, allow unit to soak for 30 minutes. 4) After 30 minute soak, rinse measuring tip with de-ionized water to rinse off excess cleaning solution. 5) Fill container with isopropanol, high enough to fully submerge measuring tip. 6) Submerge measuring tip in isopropanol for 10 minutes while unit is powered on. 7) Using the UVT-LED cleaning swab (part #UVT0011), gently clean the quartz measuring windows in front of the sensor and the UV-C light source. 8) Place unit in de-ionized water high enough to fully submerge measuring tip. 9) Submerge measuring tip in de-ionized water for 10 minutes. 10) Re-calibrate monitor according to recalibration procedure. Fill all solutions above level shown with red arrow FIG. 18 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 20 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-SW Calibration Calibration should be done every 30 days after cleaning with deionized water. It is also recommended to verify sensor accuracy weekly using a hand-held UVT-LED-H sensor. Step 1 - Place your UVT-LED-SW into de-ionized (DI) water. Make sure it is inserted as shown in FIG 19 so that it is above the opening for the wiper and LED. Shake to remove air bubbles. Step 2 - Hold Magnet over white dot sensor actuation location for at least 3 seconds as shown in FIG 19. Wait for wiper to click. Remove magnet. At completion of calibration, UVT-LED-SW will beep 7x. FIG. 19 Taking Readings Your UVT-LED-SW sensor will take readings every 1 minute after actuation of the wiper once the sensor is calibrated. You can take a reference reading at any time. To do this: Step 1 - Place your UVT-LED-SW into a sample of your water. Make sure it is inserted as shown in FIG 18 so that it is above the opening for the wiper and LED. Shake to remove air bubbles. Step 2 - Hold Magnet over white dot sensor actuation location as shown in FIG 19. Wait for 1 second. Remove magnet. UVT-LED-SW will beep 1x after reading is taken. Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 21 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-SW Installation The UVT-LED-SW is designed for submersion mounting. Mounting hardware options include: 1. UVT-SW-005: Complete installation kit that attaches to wall of channel. See below and follwing pages for instructions and FIGURE references. 2. UVT-SW-004: Retrofit installation kit that attaches to existing installation brackets from other suppliers that require 1 1/4" pipe. See FIG 32 for description of hardware. UVT-SW cable Concrete Channel Wall Wall Mounting Bracket x 2 (installed on >1ft(0.3mtr) wall clamp with anchor bolt and washer) 1.5” threaded plastic top cap with 3/8” NPT cord grip Wall of open channel pipe clamp x 2 Water Level FLOW 6ft L SS 1.5” threaded pipe FLOW UVT-LED-SW sensor SS 90 deg Elbow 1.5 “ threaded SS 1.5” to 1” reducer UVT-LED-SW sensor Note: Suggested sensor orientation is parallel to flow with wiper oriented upward Note: Suggested sensor orientation is parallel to flow with wiper oriented upward. FIG. 20 UVT-SW-001 cable FIG. 21 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 22 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-SW Installation (cont.) 1) UVT-SW-005 step by step installation procedure: A) B) C) C) D) E) F) G) H) I) J) K) L) M) Attach provided steel 90 degree elbow to pipe with sealing tape or paste (see FIG 22) Atttach 1.5" x 1" reducing fitting to elbow with sealing tape or paste (see FIG 23) Attach UVT-SW cable assembly to reducer (1” threads on both) and seal with sealing tap or paste. See FIG 24 Feed cable up through pipe and insert into top cap/cord grip assembly. Make sure cord grip is loose when feeding cable through. See FIG 25. Seal top cap onto pipe (make sure cord grip is loose). See FIG 26. Tighten cord grip after applying some slack to cable. See FIG 27. Attach brackets x 2 to pipe. Make sure bolts are not too tight so the brackets can be adjusted up or down the pipe to the correct height and spacing. Spacing will depend on wall bracket height and distance apart. Attach wall mounting brackets to concrete wall in channel using customer supplied concrete anchor bolts. Note that assembly weighs 33lbs(15kg). Holes in wall mounting bracket are 0.65” (16mm) OD. See FIG 28. Remove pins from brackets. See FIG. 29 Slide pipe assembly onto the wall mounting brackets. See FIG. 30. Adjust pipe depth into water so that UVT-LED-SW measurement pathway is fully submerged. Tighten bolts on pipe clamps to hold in position. Reinstall locking pins into wall mount brackets. See FIG. 31 To remove sensor for maintenance, remove the 2 locking pins and slide the pipe/assembly out of wall bracket channels. Seal with Tape or paste Seal with Tape or paste SS 1.5” to 1” reducer 90 deg SS elbow 1.5” Seal with Tape or paste SS 1.5” to 1” reducer 90 deg SS elbow 1.5” FIG. 23 FIG. 24 FIG. 22 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 23 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-SW Concrete Channel Wall Wall anchor bolts(customer supplied) must be rated to support >30 lbs(14 kg), 2 each per bracket. Mount brackets above expected water level. >1ft(0.3mtr) Loosen nut on cord grip to feed cable through Water Level Tighten cord grip 1.5” threaded plastic top cap with 3/8” NPT cord grip FLOW Seal with Tape or paste FIG. 25 FIG. 26 FIG. 27 Concrete Channel Wall FIG. 28 Concrete Channel Wall Concrete Channel Wall WRAP LOOP AROUND BOTTOM OF PIN Water Level Water Level Water Level FLOW FLOW FLOW FIG. 29 FIG. 30 FIG. 31 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 24 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-SW Installation (cont.) 2) UVT-SW-004: Retrofit installation kit that attaches to existing installation brackets from other suppliers that require 1 1/4" pipe. See FIG 32 for description of hardware. Refer to FIG 22 - 27 for assembly procedure. After assembling, attach to existing clamps in place of 1.25" pipe from other supplier. UVT-SW cable Threaded plastic top cap with 3/8” NPT cord grip 6ft L SS 1.25” threaded pipe UVT-LED-SW sensor (attached to UVT-SW cable) SS 90 deg Elbow 1.25“ Suggested orientation is Note: Suggested sensor with wiper opening orientation is parallel facing upward and to flowparallel with wiper oriented to flow or up to upward 45 degrees to flow SS 1.25” to 1” reducer FIG. 32 Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 25 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET UVT-LED-H Calibration Step 1 - Place your UVT-LED-H into de-ionized (DI) water. Make sure it is inserted as shown in FIG 33 so that it is above the opening for UV light source. Shake the sensor to remove bubbles. Remove from water and check for bubbles outside of the DI water. Re-insert into water if no bubbles are seen. Step 2 - Press and hold the "READ/CAL" button for at least 3 seconds as shown in FIG 34 (Make sure LOW BATTERY indicator is not on before taking measurement). See FIG 35. The display will flash blue 7x after calibrating. When the display reads >99.8%, calibration is complete. This should take approxiamtely 10 seconds. ! USE DE-IONIZED WATER ONLY FIG. 33 FIG. 34 FIG. 35 Taking Meausurements Step 1 - Place your UVT-LED-H into the water sample you want to measure the for UVT. Make sure it is inserted as shown in FIG 33 so that it is above the opening for UV light source. Shake the sensor to remove bubbles. Remove from water and check for bubbles outside of the DI water. Re-insert into water if no bubbles are seen. Step 2 - Reference measurment. Press the "READ/CAL" button and release when the screen flashes 1x. This is done to ensure no bubbles are on the measuring surface. Step 3 - When the screen flashes 2x then the sensor is ready to take a new reading (new readings can only be taken 30 seconds after the previous reading) Cleaning Using the Sensorex UVT-LED cleaning swab (part #UVT0011) gently clean the quartz measuring windows in front of the sensor and the UV-C light source. Place unit in de-ionized water high enough to fully submerge measuring tip. Submerge measuring tip in de-ionized water tor for 10 minutes, according to recalibration procedure. Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 26 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Troubleshooting (UVT-LED-H & -PW) Fault Possible Solution UVT-LED will not switch on - Check connections in power cabinet to ensure 24V DC is being supplied. - Check 5-pin connector on back of sensor to make sure it is securely attached. UVT reading is lower than expected - During measurement an air bubble may have been present in the measurement area. Ensure no air bubbles are present and repeat the measurement. - Follow cleaning procedure on page 13 (-PW), 22 (-H) to remove materials that may foul the sensor. - On battery-operated models, battery is low and unit should be plugged in to recharge battery. You can take measurements while UVT-LED-H is charging. - Unit needs to be recalibrated accoring to outlined calibration procedure. - Make sure flow cell/pipe is full enough to completely submerge the measurement pathway. If not, adjust orientation. UVT reading is higher than expectecd - During calibration, an air bubble may have been present in the measurement area. Recalibrate unit. - Follow cleaning procedure. - Make sure flow cell/pipe is full enough to completely submerge the measurement pathway. If not, adjust orientation. UVT reading is not being transmitted to plant controls (UVT-LED-PW ONLY) - Check to make sure the 4-20mA signal loop has an external power source. - Check 5-pin connector on back of sensor to make sure it is securely attached. Analog scaling is not correct (-PW) ` - Scaling should be set up for 4mA = 0% UVT and 20mA = 100% UVT. UVT reading is ZERO - During measurement an air bubble may have been present in the measurement area. Ensure no air bubbles are present and repeat the measurement. - Ensure signal wires are connected per wiring diagram. - Follow cleaning procedure. - Recalibrate unit. - LED or sensor has stopped working. Return for repair. - Water has high absorbance and is not allowing UV-C light to pass through. Wiper Mechanism does not return - Return sensor to supplier for repair/replacement to home position during cycle (-PW). Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 27 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Troubleshooting (UVT-LED-SW) 3 beeps from sensor indicate a failure for the UVT-LED-SW. See chart below for faults and possible solutions. Fault Possible Solution UVT-LED-SW will not switch on - Check connections in power cabinet to ensure 24V DC is being supplied. - Place magnet over white dot sensor actuation location and wait for wiper to actuate. - During measurement an air bubble may have been present in the measurement UVT reading is lower than expected area. Ensure no air bubbles are present and repeat the measurement. - Follow cleaning procedure to remove materials that may foul the sensor. - Unit needs to be recalibrated accoring to outlined calibration procedure. UVT reading is higher than expectecd - During calibration, an air bubble may have been present in the measurement area, Recalibrate unit. - Follow cleaning procedure on page 20. UVT reading is not being transmitted to plant controls - Check to make sure the 4-20mA signal loop has an external power source. - Ensure signal wires are connected per wiring diagram. Analog scaling is not correct (-PW) - Scaling should be set up for 4mA = 0% UVT and 20mA = 100% UVT. UVT reading is ZERO - During measurement an air bubble may have been present in the measurement area. Ensure no air bubbles are present and repeat the measurement. - Follow cleaning procedure. - Recalibrate unit. - LED or sensor has stopped working. Return for repair. - Water has high absorbance and is not allowing UV-C light to pass through. Wiper Mechanism does not return to home position during cycle (-PW) - Return sensor to supplier for repair/replacement. Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 28 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Ordering Information Part Number Description UVT-LED-PW In-line Mount Process UV Transmittance Monitor (with wiper/auto cal.), 10m (33 ft), cleaning swab UVT-LED-SW Submersion Mount Process UV Transmittance Monitor (with wiper/auto cal.), includes 10m(33ft) detachable cable, cleaning swab UVT-LED-H-WD Handheld, Off-Line Transmittance Monitor with power supply (outside of North America), case, cleaning swab x 1 UVT-LED-H-NA Handheld, Off-Line Transmittance Monitor with power supply (North America only), case, cleaning swab x 1 UVT0001 Carrying Case for UVT-LED-H (included with UVT-LED-H-WD and UVT-LED-H-NA kits) UVT0002 Power/Communication Cable for UVT-LED-PW, 10m (33 ft)- included with UVT-LED-PW UVT0003 Power/Communication Cable for UVT-LED-PW, 20m (66 ft) UVT0004 Power/Communication Cable for UVT-LED-PW, 30m (99 ft) UVT-SW-001 Detachable Power/Communication Cable for UVT-LED-SW, 10m (33 ft)- included with UVT-LED-SW UVT-SW-002 Detachable Power/Communication Cable for UVT-LED-SW, 20m (66 ft) UVT-SW-003 Detachable Power/Communication Cable for UVT-LED-SW, 30m (99 ft) UVT0011 Cleaning Swabs (pack of 50) UVT0008 UVT-LED-PW Installation Fitting, 2" NPT 316SS UVT0009 UVT-LED-PW Installation Fitting, Stainless Steel Pipe (requires welding) UVT0021 UVT-LED-PW Installation Fitting for PVC Pipe (2" NPT) UVT0015 UVT-LED-PW Installation Kit, Stainless Steel Sanitary Flow Cell (3”) UVT0023 UVT-LED-PW Installation Kit, PVC Flow Cell (2") with connections for barbed fitting with quick-disconnect UVT0024 UVT-LED-PW Installation Kit, Stainless Steel Flow cell (2") with 1/4" barbed fittings UVT-SW-004 UVT-LED-SW Installation Kit, retrofit to other supplier installation kits that accept 1.25" diameter pipe. UVT-SW-005 UVT-LED-SW Installation Kit, Complete Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 29 of 30 Document Number 480437 FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation PRODUCT INSTRUCTION SHEET Product Warranty This UVT-LED product, as supplied and by Sensorex, or an authorized Sensorex distributor, and delivered new, in the original carton is warranted by Sensorex against manufacturing defects in materials and workmanship for a limited warranty period of: Two (2) years on all components The limited warranty commences on the date of shipment. To make a warranty claim, the purchaser must contact Sensorex or distributor for problem determination and receipt of an RMA number. The serial number must be presented for warranty processing to begin. Sensorex will repair, or replace, any part found to be defective, in our opinion and at no charge as stipulated herein, with new or reconditioned parts during the limited warranty period specified above. All replaced parts and products become the property of Sensorex and must be returned to Sensorex. Replacement parts and products assume the remaining original warranty, or ninety (90) days, whichever is longer. This limited warranty covers manufacturing defects in materials and workmanship encountered in normal use of this product, and shall not apply to the following, including but not limited to: wear and tear to quartz window wiper blades, damage which occurs in shipment; delivery and installation; applications and uses for which this product was not intended; altered product or serial numbers; cosmetic damage or exterior finish; accidents, abuse, neglect, fire, water, lightning or other acts of nature; use in applications with water temperature and ambient air temperature over 100 degrees F, with water pressure greater than 150 psi (PW only); use of products, equipment, systems, utilities, services, parts, supplies, accessories, applications, installations, repairs, external wiring not supplied or authorized by Sensorex, or which damage this product or result in service problems; incorrect supply voltage, fluctuations and surges in electrical supply; customer adjustments and failure to follow operating instructions, cleaning and maintenance instructions that are covered and prescribed in the Operations and Maintenance manual. There are no express warranties other than those listed and described above, and no warranties whether expressed or implied, including, but not limited to, any implied warranties of merchantability or fitness for a particular purpose, shall apply after the express warranty periods stated above, and no other express warranty or guaranty given by any person, firm or corporation with respect to this product shall be binding on Sensorex. Sensorex shall not be liable for loss of revenue or profits, failure to realize savings or other benefits, or any other special, incidental or consequential damages caused by the use, misuse or inability to use this product, regardless of the legal theory on which the claim is based, and even if Sensorex has been advised of the possibility of such damages. Nor shall recovery of any kind against Sensorex be greater in amount than the purchase price of the product sold by Sensorex and causing the alleged damage. Without limiting the foregoing, purchaser assumes all risk and liability for loss, damage or injury to purchaser and purchaser’s property and to others and their property arising out of the use, misuse or inability to use this product sold by Sensorex not caused directly by the negligence of Sensorex. This limited warranty shall not extend to anyone other than the original purchaser of this product either from Sensorex directly or through an authorized Sensorex distributor, it is non-transferable and states the exclusive remedy. Parts covered by this product instruction sheet include: UVT-LED-PW, UVT-LED-H, UVT-LED-SW, and all accessories and mounting hardware Page 30 of 30 11751 Markon Drive Garden Grove, CA. 92841 USA Document Number 480437 Ph +1(714) 895-4344 Fax +1(714) 894-4839 sales@sensorex.com www.sensorex.com FORM: InstrUVT-LED Instruction Manual-09 [Rev: 2016-10-19] ©2016 Sensorex Corporation
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.4 Linearized : Yes Language : en-US Tagged PDF : Yes XMP Toolkit : Adobe XMP Core 5.6-c123 79.158978, 2016/02/13-01:11:19 Create Date : 2016:10:19 09:37:37-07:00 Metadata Date : 2016:10:19 09:38:05-07:00 Modify Date : 2016:10:19 09:38:05-07:00 Creator Tool : Adobe InDesign CC 2015 (Windows) Instance ID : uuid:2e63ac93-0cad-4239-b9cd-c726f53cc45f Original Document ID : adobe:docid:indd:1d0e12d6-0f51-11df-9e83-f86aa3353eef Document ID : xmp.id:46e53006-ab26-9c4b-b31f-22a05e0051d8 Rendition Class : proof:pdf Derived From Instance ID : xmp.iid:001ef1a7-5d5c-0c4d-8faf-049b6e996025 Derived From Document ID : xmp.did:67fbe760-846f-c94d-ae16-874f6f57b862 Derived From Original Document ID: adobe:docid:indd:1d0e12d6-0f51-11df-9e83-f86aa3353eef Derived From Rendition Class : default History Action : converted History Parameters : from application/x-indesign to application/pdf History Software Agent : Adobe InDesign CC 2015 (Windows) History Changed : / History When : 2016:10:19 09:37:37-07:00 Format : application/pdf Title : UVT-LED Instruction Manual Creator : Sensorex Description : Use and maintenance instructions for UVT-LED sensors Rights : © Copyright 2011 Sensorex. All rights reserved Subject : UVT, UVT-LED, UVC, UVT-LED-H, UVT-LED-PW, UV Web Statement : http://www.sensorex.com Marked : True Startup Profile : Print Producer : Adobe PDF Library 15.0 Trapped : False Page Count : 30 Author : SensorexEXIF Metadata provided by EXIF.tools