DIGITAL MULTIMETERS & SYSTEMS
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY (U.S. only)
The 7½-digit Model 2010 Low Noise Multi meter combines high
resolution with the high speed and accuracy needed for produc-
tion applications such as testing precision sensors, transducers,
A/D and D/A con verters, regulators, references, connectors,
switches, and relays. It is based on the same high speed, low
noise A/D converter technology as the Models 2000, 2001,
and 2002.
High Measurement Flexibility
The 2010 has 15 built-in measurement functions, including DCV,
ACV, DCI, ACI, 2W
W
, 4W
W
, dry circuit resistance, temperature
(with either thermocouples or RTDs), frequency, period, ratio,
continuity meas ure ment, and diode testing. This multi-functional
design minimizes added equip ment costs.
Creating a self-contained multipoint measurement solution is as
simple as plugging a 2000-SCAN or 2001-TCSCAN scanner card into the option slot in the 2010’s back
panel. This “plug-in” approach eliminates the need for a separate scanner and significantly reduces
programming and setup time in applications involving a limited number of test points. For larger
applications, the 2010 is compatible with Keithley’s Series 7000 switch matrices and cards.
Unique Resistance Measurement Functions
Characterizing the resistance, linearity, or isolation of contacts, connectors, switches, or relays
complete ly and efficiently demands an uncommon combination of ohms measurement capa bilities.
The 2010 offers:
Low-power ohms measurement mode. Low-level resistance measurements can be made with
source current as low as 100µA, an order of magnitude lower than is possible with other DMMs,
so device self-heating is minimized. Among other benefits, this low-power measurement capability
makes the 2010 suitable for end-of-life contact testing per ASTM B539-90.
Dry circuit test function. When measuring contact and connector resistances, it is important to
control the test voltage carefully in order to avoid puncturing any oxides or films that may have
formed. A built-in clamp limits the open circuit test voltage to 20mV to ensure dry circuit conditions.
Offset compensated ohms function. This function eliminates thermal effects that can create
errors in low-level resistance measurements in system environments.
Extended ohms measurement capability. The 2010 provides a 10W range for more precise
measurements of low resistances.
Optional Multiplexer Cards
Creating a self-contained multipoint measure-
ment solution is as simple as plugging a scanner
card into the option slot on the 2010’s back
panel. This ap proach eliminates the complexities
of triggering, timing, and processing issues and
helps reduce test time significantly. For appli-
cations involving more than 10 measurement
points, the 2010 is com patible with Keithley’s
Series 7000 switch matrices and cards.
Model 2000-SCAN Scanner Card
Ten analog input channels (2-pole)
Configurable as 4-pole, 5-channel
7½-digit resolution
100nV rms noise floor
7ppm DCV repeatability
Built-in 10-channel scanner
mainframe
Dry circuit and low power
measurement mode
15 measurement functions
including support for RTD and
thermocouple temperature
measurements
Built-in ratio measurement
function
GPIB and RS-232 interfaces
Ordering Information
2010 Autoranging DMM
Accessories Supplied
Model 1751 Safety Test Leads,
User Manual, Service Manual
ACCESSORIES AVAILABLE
TEST LEADS
5804/5 4-Wire/Kelvin Test Lead Sets
SWITCH/SCANNER CARDS
2000-SCAN 10-channel Scanner
2001-TCSCAN 9-channel Thermocouple Scanner
CABLES/ADAPTERS
7007-1 Shielded IEEE-488 Cable, 1m (3.3 ft)
7007-2 Shielded IEEE-488 Cable, 2m (6.6 ft)
7009-5 RS-232 Cable
RACK MOUNT KITS
4288-1 Single Fixed Rack Mount Kit
4288-2 Dual Fixed Rack Mount Kit
GPIB INTERFACES
KPCI-488LPA IEEE-488 Interface/Controller for the PCI Bus
KUSB-488B IEEE-488 USB-to-GPIB Interface Adapter
2010
Low Noise 7½-Digit
Autoranging Multimeter
SERVICES AVAILABLE
2000-SCAN-3Y-EW 1-year factory warranty extended to 3 years
from da
te of shipment
2001-TCSCAN-3Y-EW 1-year factory warranty extended to 3 years
from date of shipment
2010-3Y-EW 1-year factory warranty extended to 3 years
from date of shipment
C/2000-3Y-ISO 3 (ISO-17025 accredited) calibrations within
3 years of purchase for Model 2000-SCAN*
C/2001-3Y-ISO 3 (ISO-17025 accredited) calibrations
within 3 years of purchase for Model
2001-TCSCAN*
C/2010-3Y-ISO 3 (ISO-17025 accredited) calibrations within
3 years of purchase for Mo
del 2010*
*Not available in all countries
DMM optimized for resistance measurement applications
DMM optimized for resistance measurement applications
DIGITAL MULTIMETERS & SYSTEMS
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Greater Measure of Confidence
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
Voltage Range Resolution
Frequency
Range
Accuracy (1 Year)
23°C ±5°C
±(% of reading + % of range)
100 mV to 750 V 0.1 µV to 1 mV
3 Hz–10 Hz 0.35 + 0.03
10 Hz–20 kHz 0.06 + 0.03
20 kHz–50 kHz 0.12 + 0.05
50 kHz–100 kHz 0.60 + 0.08
100 kHz–300 kHz 4 + 0.5
AC OPERATING CHARACTERISTICS
Function Digits Readings/s Rate Bandwidth
ACV (all ranges), and
ACI (all ranges)
0.5 (0.4) SLOW 3 Hz–300 kHz
1.4 (1.5) MED 30 Hz–300 kHz
4.0 (4.3) MED 30 Hz–300 kHz
2.2 (2.3) FAST 300 Hz300 kHz
35 (30) FAST 300 Hz300 kHz
FREQUENCY AND PERIOD CHARACTERISTICS
ACV
Range
Frequency
Range
Period
Range
Gate
Time
Resolution
±(ppm of
reading)
Accuracy
90 Day/1 Year
±(% of reading)
100 mV to
750 V 3 Hz to
500 kHz 333 ms to
2 µs 1 s 0.3 0.01
TEMPERATURE CHARACTERISTICS
Thermocouple Accuracy 1
90 Day/1 Year (2C ± 5°C)
Type Range Resolution
Relative to
Reference Junction
USING
2001-TCSCAN2
J–200 to + 760°C 0.001°C ±0.5°C ±0.65°C
K–200 to + 1372°C 0.001°C ±0.5°C ±0.70°C
N–200 to + 1300°C 0.001°C ±0.5°C ±0.70°C
T–200 to + 400°C 0.001°C ±0.5°C ±0.68°C
4-WIRE RTD Accuracy 3
90 Day/1 Year
(23°C ± 5°C)
Accuracy 3
2 Years
(23°C ± 5°C)
Range Resolution
–100° to +100°C 0.001°C ±0.08°C ±0.12°C
–200° to +630°C 0.001°C ±0.14°C ±0.18°C
TEMPERATURE NOTES
1. For temperatures <–100°C, add ±0.1°C and >900°C add ±0.3°C.
2. Specifications apply to channels 2–6. Add 0.06°C/channel from channel 6.
3. Excluding probe errors.
2010 Low Noise 7½-Digit
Autoranging Multimeter
GENERAL
POWER SUPPLy: 100V / 120V / 220V / 240V.
LINE FREQUENCy: 50Hz to 60Hz and 440Hz, automatically sensed at power-up.
POWER CONSUMPTION: 22VA.
VOLT HERTz PRODUCT: 8 × 107V·Hz.
OPERATING ENVIRONMENT: Specified for 0° to 50°C. Specified to 80% R.H. at 35°C.
STORAGE ENVIRONMENT:40° to 70°C.
ALTITUDE: Up to 2000 meters.
SAFET y: Conforms to European Union Directive 73/23/EEC EN 61010-1, Cat II.
EMC: Complies with European Union Directive 89/336/EEC, EN 61326-1.
VIBRATION: MIL-PRF-28800F Class 3 Random.
WARMUP: 2 hours to rated accuracy.
DIMENSIONS:
Rack Mounting: 89mm high × 213mm wide × 370mm deep (3½ in × 838 in × 14916 in).
Bench Configuration (with handle and feet): 104mm high × 238mm wide × 370mm
deep (418 in × 938 in × 14916 in).
SHIPPING WEIGHT: 5kg (11 lbs).
DC VOLTAGE
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range) Input
ResistanceRange Resolution 90 Day 1 Year
100.00000 mV 10 nV 25 + 9 37 + 9 > 10 GW
1.0000000 V 100 nV 18 + 2 25 + 2 > 10 GW
10.000000 V 1 µV 18 + 4 24 + 4 > 10 GW
100.00000 V 10 µV 25 + 5 35 + 5 10 MW ±1%
1000.0000 V 100 µV 31 + 6 41 + 6 10 MW ±1%
RESISTANCE
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range)
Test CurrentRange Resolution 90 Day 1 Year
10.000000 W1 µW40 + 9 60 + 9 10 mA
100.00000 W10 µW36 + 9 52 + 9 1 mA
1.0000000 kW100 µW33 + 2 50 + 2 1 mA
10.000000 kW1 mW32 + 2 50 + 2 100 µA
100.00000 kW10 mW40 + 4 70 + 4 10 µA
1.0000000 MW100 mW50 + 4 70 + 4 10 µA
10.000000 MW 1 W200 + 4 400 + 4 640 nA//10 MW
100.00000 MW 10 W1500 + 4 1500 + 4 640 nA//10 MW
DC CURRENT
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range) Burden
VoltageRange Resolution 90 Day 1 Year
10.000000 mA 1 nA 300 + 80 500 + 80 < 0.15 V
100.00000 mA 10 nA 300 + 800 500 + 800 < 0.18 V
1.0000000 A 100 nA 500 + 80 800 + 80 < 0.35 V
3.000000 A 1 µA 1200 + 40 1200 + 40 < 1 V
CONTINUITY 2W
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range)
Test CurrentRange Resolution 90 Day 1 Year
1 kW100 mW100 + 100 120 + 100 1 mA
DIODE TEST
Accuracy 23°C ± 5°C
±(ppm of rdg. + ppm of range)
Test CurrentRange Resolution 90 Day 1 Year
10.000000 V 1 µV 30 + 7 40 + 7 1 mA
4.400000 V 1 µV 30 + 7 40 + 7 100 µA
10.000000 V 1 µV 30 + 7 40 + 7 10 µA
DC OPERATING CHARACTERISTICS
Function Digits Readings/s PLCs
DCV (all ranges),
DCI (all ranges), and
Ohms (<10M range)
4(3) 5
30 (27) 1
50 (44) 1
260 (220) 0.1
490 (440) 0.1
1000 (1000) 0.04
2000 (1800) 0.01
DC NOISE PERFORMANCE
Rate Digits
RMS Noise
100mV Range
(2 min.)
RMS Noise
10V Range
(2 min.) NMRR CMRR
5 PLC 110 nV 1.2 µV 60 dB 140 dB
1 PLC 125 nV 1.4 µV 60 dB 140 dB
0.1 PLC 1.9 µV 11.5 µV 80 dB
0.01 PLC 2.9 µV 139 µV 80 dB
Model 2010 specifications
Model 2010 specifications