69d143cc Bdc1 492e 911c 8e586f539ec8

2017-10-20

: Pdf 69D143Cc-Bdc1-492E-911C-8E586F539Ec8 69d143cc-bdc1-492e-911c-8e586f539ec8 10 2017 pdf

Open the PDF directly: View PDF PDF.
Page Count: 16

The surgical technique shown is
for illustrative purposes only. The
technique(s) actually employed in each
case will always depend upon the medical
judgment of the surgeon exercised
before and during surgery as to the best
mode of treatment for eachpatient.
Please see the package insert for the
complete list of indications, warnings,
precautions, and other important medical
information.
Consult instructions for use at this
website www.medtronic.com/manuals.
Note: Manuals can be viewed using a
current version of any major internet
browser. For best results, use Adobe
Acrobat Reader with the browser.
Medtronic Sofamor Danek USA, Inc.
1800 Pyramid Place
Memphis, TN 38132
(901) 396-3133
(800) 876-3133
Customer Service: (800) 933-2635
Medtronic
Spinal and Biologics Business
Worldwide Headquarters
2600 Sofamor Danek Drive
Memphis, TN 38132
medtronic.com
SURGICAL
TECHNIQUE
OsteoCool™ RF Ablation System
and Bone Access Kits
© 2017 Medtronic. All rights reserved. UC201605782b EN PMD016254-4.0
OSTEOCOOL RF ABLATION SYSTEM AND BONE ACCESS KITS
SUBHEAD
3 1
CONTENTS
3 System Setup
6 Patient Positioning
6 OsteoMAPSM Technique
Using Bone Access Kits — Spine
8 OsteoMAPSM Technique Using
Bone Access Kits — Sacrum,
Iliac, and Acetabulum
9 Sacrum
10 Iliac
11 Acetabulum
12 Use of the Precision Drill —
Spine
14 Use of the Precision Drill —
Sacrum, Iliac, and Acetabulum
16 Probe Selection and Tube Kit
Preparation
17 Filling the OsteoCool™ Tube Kit
19 Connecting the Probe to
the Tube Kit
20 Connecting to the Connector Hub
21 Performing RF Ablation — Spine
23 Performing RF Ablation — Sacrum,
Iliac, and Acetabulum
25 Optional Track RF Ablation
26 Optional Temperature
Monitoring with the Independent
Thermocouple
26 Performing RF Ablation Using
Custom Settings
27 Bone Access for Subsequent
Procedures
27 General Safety Guidelines
28 Product Ordering Information
28 Important Product Information
2 3
The OsteoCool™ RF ablation probe
uses a coaxial, bipolar technology that
delivers localized tumor ablation and
automatically moderates power to keep
RF heating within the desired treatment
range, reducing risks of potential thermal
damage to adjacent structures.
The active tip of the RF ablation probe
is internally cooled with circulating
water. RF energy heats the tissue
while circulating water moderates the
temperature in close proximity to the
active tip. This combination creates
large volume lesions without excessive
heating at the active tip.
The OsteoCool™ RF ablation probes are
sterile and intended for single use.
The OsteoCool™ Bone Access Kit is a
single-use device intended to contact
body tissues. Do not reuse, reprocess, or
resterilize.
The OsteoCool™ independent
thermocouple is for optional
temperature monitoring at or near the
site of ablation.
A physician using this equipment must
be familiar with bony anatomy, image-
guided procedures, and bone access
techniques.
Refer to the OsteoCool™ Instructions
for Use for complete generator and
pump setup.
The OsteoCool™ RF Ablation System is
indicated for palliative treatment in spinal
procedures by ablation of metastatic
malignant lesions in a vertebral body.
It also is intended for coagulation and
ablation of tissue in bone during surgical
procedures, including palliation of pain
associated with metastatic lesions
involving bone in patients who have
failed or are not candidates for standard
therapy.
The system contains an RF generator,
peristaltic pump, and the connector
hub, which provides two channels
for the use of the OsteoCool™ RF
ablation probes and two channels for
use of the OsteoCool™ independent
thermocouples.
Important
This guide does not replace the information in the Instructions for Use provided
with the components of the OsteoCool™ RF Ablation System, OsteoCool™ RF
Ablation Probes, OsteoCool™ Bone Access Kits, and the OsteoCool™ Independent
Thermocouple. The Instructions for Use includes important information such as
warnings, precautions, contraindications, and troubleshooting. It is important to read
the Instructions for Use and these precautions carefully prior to device operation.
Additional help information can be accessed through the “help” button located on the
generator screen.
Please note that the anatomy included in this guide reects some of the most
common locations for metastatic bone tumors — spine, pelvis, and hip — but it is not
all inclusive of areas in which the OsteoCool™ system can be used. The OsteoCool™
RF Ablation System is indicated for the palliative treatment in spinal procedures
by ablation of metastatic malignant lesions in a vertebral body. It is also intended
for coagulation and ablation of tissue in bone during surgical procedures, including
palliation of pain associated with metastatic lesions involving bone in patients who
have failed or are not candidates for standard therapy.
SYSTEM
SETUP
The system may be mounted on a rolling cart or a desk stand (see following instructions).
Desk Stand Assembly and Equipment Mounting
1. Assemble the desk stand by sliding the pole into
the base and tightening the set screw with the
provided Allen wrench (Figure 1).
2. Position the desk stand on a stable surface and in
close proximity to an electrical source.
3. Mount the pump by sliding the quick connect
bracket on the back of the pump down the
opposite mounting bracket located on the
bottom of the desk stand (Figure 2a). Mount the
generator by sliding the quick connect bracket
on the back of the generator down the opposite
mounting bracket located on the top of the desk
stand (Figure 2b).
4. Proceed to cable connection.
Figure 1 Figures 2a (Pump Bracket) and
2b (Generator Bracket)
Front View
Pump
Mount
RF-Generator
Mount
Isometric View
2a
2b
4 5
Cart Mounting
1. Position the cart in proximity to an electrical source.
2. Lock the wheels on the cart before mounting
the devices.
3. Mount the pump by sliding the quick connect bracket
on the back of the pump down the opposite mounting
bracket located on the bottom of the cart (Figure 3a).
Mount the generator by sliding the quick connect
bracket on the back of the generator down the
opposite mounting bracket located on the top of the
cart (Figure 3b).
4. Do not adjust the position of the mounting brackets.
5. Proceed to cable connection.
Connect Cables and Power on the Generator and Pump
1. After assembly, connect pump and the generator with the
connector cable to USB ports on the back of both pump and
generator (Figures 4a and 4b).
Plug the USB type A plug to the generator, and the USB type
B plug into the pump.
Generator
Quick Connect
and Release
Pump
Quick Connect
and Release
3b
3a
Figures 3a and 3b
Figure 5a Figure 5b
Figure 4b
Power Switch
USB Port
(for connecting
generator to
pump)
Figure 4a
Type B
Type A
Figure 6
2. Connect power cables to the back of both the
generator and pump.
a. Connecting Power Cables with the Desk Stand:
Route the cable to the backside of the pole and use
the clips on the desk stand to manage the power and
USB cables (Figure 5a).
b. Connecting Power Cables with the Cart:
The cart is provided with cables that are fed through
the cart pole. Attach the connections located near the
mounting brackets to the back of the generator and
pump. The socket end of the cables, which come out
near the bottom of the cart pole, should be connected
directly to the power cables provided with the generator
and pump.
3. Plug both power cables into a surge protector/power strip.
4. Turn on the generator using the power switch located on the
back of the device (Figure 5b).
Note
The circular power button located on the front (top right
corner) of the device is to initiate radiofrequency ablation,
not to turn on power to the generator.
5. After turning on the generator, turn on the pump using the
power switch located on the back of the device.
Note
Turning on the pump before turning on the generator is not
recommended as it could lead to system error.
6. Following the on-screen setup instructions, connect the
connector hub to the front of the generator (Figure 6).
Connect Cables and Power on the Generator and Pump
6 7
PATIENT POSITIONING
1. Place the patient in the prone position for access to the spine.
Note: For access to other areas of bone, a supine or lateral
patient positioning may be more eective for gaining
appropriate access.
2. Using appropriate image guidance, conrm the target location
for treatment and proceed with the necessary skin incision(s).
OSTEOMAP
TECHNIQUE USING
BONE ACCESS KITS — SPINE
1. Insert the stylet into the cannula to
form the osteointroducer.
2. Using uoroscopic image guidance,
insert the osteointroducer at the
desired site. Note
Markings on the cannulas should
be used as reference marks
only. They are not intended to
replace the use of uoroscopic
observation.
Note
The OsteoCool™ osteointroducer
oers an internal stylet that extends
beyond the distal end of the cannula.
This design is intended to provide a
visual indication under image guidance
of where the posterior margin of the
ablation zone with the OsteoCool™ RF
ablation probe will stop.
Caution
To maintain structural integrity, do
not advance the cannula without the
stylet fully inserted.
3. Using manual control and appropriate
image guidance, advance the
osteointroducer through the soft
tissues into the selected bone to the
desired depth.
4. A surgical mallet may be used
to augment the insertion of the
osteointroducer.
5. While holding the cannula, rotate
the handle 180 degrees counter-
clockwise to remove the stylet.
Proceed with the OsteoCool™
precision drill.
For bi-pedicular ablation, repeat the
steps above.
8 9
OSTEOMAP
TECHNIQUE USING
BONE ACCESS KITS — SACRUM,
ILIAC, AND ACETABULUM
1. Insert the stylet into the cannula to form the osteointroducer.
2. Using CT or uoroscopic image guidance, insert the osteointroducer at the desired site.
3. Using manual control and appropriate image guidance, advance the osteointroducer through the soft tissues into the
selected bone to the desired depth.
4. A surgical mallet may be used to augment the insertion of the osteointroducer.
5. While holding the cannula, rotate the handle 180 degrees counterclockwise to remove the stylet. Proceed with the
OsteoCool™ precision drill.
For ablation using two probes, repeat these steps.
SACRUM
One-probe approach:
Two-probe approach:
10 11
ILIAC
One-probe approach:
ACETABULUM
One-probe approach:
Two-probe approach:
Two-probe approach:
Note
Markings on the cannula should be used
as reference marks only. They are not
intended to replace the use of CT or
uoroscopic observation.
Note
The OsteoCool™ osteointroducer
oers an internal stylet that extends
beyond the distal end of the cannula.
The design is intended to provide a
visual indication under image guidance
of where the posterior margin of the
ablation zone with the OsteoCool™ RF
ablation probe will stop.
Caution
To maintain structural integrity, do not
advance the cannula without the stylet
fully inserted.
12 13
USE OF THE
PRECISION DRILL — SPINE
1. After gaining access to the vertebral
body using the osteoinducer, advance
the precision drill down the cannula
lumen into the vertebral body.
2. Using manual control and image
guidance, rotate clockwise and
advance the precision drill to the
desired depth. This depth will serve as
the anterior boundary for the ablation
zone.
Note
The color markings on the proximal
shaft of the OsteoCool™ precision
drill correlate with sizes of the
OsteoCool™ RF ablation probes.
3. Read the color marking on the
proximal end of the drill to aid in probe
selection. If the drill positioning is in
between color markings, advance or
retreat to a color marking. Select the
probe size that corresponds with the
color marking.
Note
Markings on the drill should be used
as reference marks only. They are
not intended to replace the use of
uoroscopic observation.
4. Remove the precision drill from
the cannula lumen using clockwise
rotation. The target site is now ready
for ablation probe placement.
For bi-pedicular ablation, repeat
the steps above. Product testing
recommends a distance of 8–10 mm
between the distal tips of the probes to
yield the largest ablation zone.
14 15
USE OF THE
PRECISION DRILL — SACRUM,
ILIAC, AND ACETABULUM
1. After gaining access into bone using the osteointroducer,
advance the precision drill down the cannula lumen into the
tissue.
2. Using manual control and image guidance, rotate clockwise
and advance the precision drill to the desired depth. This
depth will serve as the anterior boundary for the ablation
zone.
Note
The color markings on the proximal shaft of the OsteoCool™
precision drill correlate with sizes of the OsteoCool™ RF
ablation probes.
3. Read the color marking on the proximal end of the drill to aid
in probe selection. If the drill positioning is in between color
markings, advance or retreat to a color marking. Select the
probe size that corresponds with the color marking.
Note
Markings on the drill should be used as reference marks only.
They are not intended to replace the use of CT or uoroscopic
observation.
4. Remove the precision drill from the cannula lumen using
clockwise rotation. The target site is now ready for ablation
probe placement.
For ablation using two probes, repeat the steps above.
A. Sacrum A1 A2
B1 B2
C1 C2
B. Iliac
C. Acetabulum
16 17
FILLING THE OSTEOCOOL TUBE KIT
INSERTING IT INTO THE PUMP UNIT
Once a probe kit has been chosen, remove the tube kit and
probe from the sterile package and place in the sterile eld.
Fill the burette with sterile water
1. Fill the burette to the 70 mL mark with room temperature sterile water, using one of the following options:
Option 1 – Inject water through the port in the lid.
Option 2 – Remove the lid and pour the water directly into the burette.
Option 1 — Injecting sterile water using the lid port
a. Place sterile syringe in the port.
b. Inject 70 mL of sterile water at room temperature
into the burette.
Option 2 — Removing the lid and pouring sterile water
directly into the burette
a. Open the lid by pressing in and up with your thumbs
around one of the three petals.
b. Observe proper sterile handling technique while lling the
burette; do not place the lid of the burette on a non-sterile
surface.
c. The ll lines on the burette represent 70 mL and 80 mL
respectively.
d. After lling to between the lines, snap the lid back into
place on the burette.
PROBE SELECTION AND
TUBE KIT PREPARATION
Select the OsteoCool™ RF ablation probe size for the procedure.
Note
If the OsteoCool™ precision drill was used, select the probe size that
coordinates with the color marking on the drill shaft.
7 mm Active Tip
11×10 mm Ablation Zone
10 mm Active Tip
17×13 mm Ablation Zone
15 mm Active Tip
23×18 mm Ablation Zone
20 mm Active Tip
29×21 mm Ablation Zone
18 19
CONNECTING
THE PROBE
TO THE TUBE KIT
1. Pass the tubing and electrical connections on the probe out
of the sterile eld.
2. Remove the caps on the male and female luer locks for both
the tube kit and the chosen probe. Connect the appropriate
luer lock on the tube kit to the corresponding luer lock on the
probe.
Warning
DO NOT over tighten the connection. Maintain sterility of the
tubing’s inner pathway so if water is spilled in the sterile eld,
sterility will not be compromised.
2. Put the burette into the pump units burette holder.
3. Open the pump head lid and thread the thicker tubing
from the bottom of the burette into the pump head
tube holder.
4. Ensure that the tubing is properly placed between the
notches and along the center channel beneath the
pump head.
Warning
Improper positioning of the tubing can pinch the tube
and restrict water ow.
5. Close the lid to hold the tubing in place. Leave the luer
lock caps on the tubing until you are ready to connect
the probes, so the inner pathway of the tube kit
remains sterile.
Open lid
Closed lid holds tubing in place
20 21
PERFORMING RF ABLATION —
SPINE
1. The ablation probes come with a pre-
assembled spacer. Discard or keep
the spacer, based on the bone access
tool used. Reference chart below:
2. Place the ablation probe through the
bone access cannula, ensuring it is
fully seated in the cannula.
Warning
Never force the probe if you feel
signicant resistance.
3. Under image guidance, conrm the
proximal radiopaque marker has
fully surpassed the distal end of
the cannula.
Warning
Incorrect use of the spacer may result
in inadequate probe clearance from
the cannula.
Description/Size Part Number Directions
OsteoCool™ Bone Access Kit 10G 090 OCN002 Use with Spacer
OsteoCool™ Bone Access Kit 8G 090 OCN003 Use with Spacer
OsteoCool™ Bone Access Kit 10G 095 OCN004 Discard Spacer
OsteoCool™ Bone Access Kit 13G 100 OCN005 Discard Spacer
Remove spacer for use with
OCN004 and OCN005.
Use spacer with
OCN002 and OCN003.
CONNECTING TO
THE CONNECTOR HUB
The connector hub provides two channels for the use of the
ablation probes and two channels for use of the independent
thermocouples.
The white ports (top) accept the RF ablation probes. The black
ports (bottom) accept the independent thermocouples.
1. Connect the male connector on the ablation probe to the
female connector on the connector hub.
2. Conrm via on-screen information that the probe has been
detected and is ready for use.
Note
The generator will quickly detect the probe and indicate a
“High Impedance” error while the probe is outside of tissue.
When the probe is placed in tissue, the “High Impedance”
error will disappear.
22 23
4. When the screen shows “Ready,”
press the RF power button. Ablation
time is pre-set based on the active tip
size of the probe(s) being used.
a. Pre-set ablation times are shown
at right.
Note
Modications to standard procedure
settings can be adjusted on the RF
generator user interface. Refer to
the user manual for instruction on
changing these settings.
5. Allow the generator to complete the
ablation, monitoring for error signs on
the generator screen.
a. The ablation will terminate
automatically as soon as the set
ablation time is reached.
6. After the ablation is complete, remove
the probe from the cannula and
detach from the tube kit. Discard as
biohazardous waste. Detach the tube
kit from the generator and discard as
biohazardous waste.
7. The OsteoCool™ bone access
cannula is still in place and ready to
accept other instrumentation.
Label Color Product Code Active Tip Lesion Size Ablation Time
Yellow OCP107 and OCP207 7 mm 11 × 10 mm 6:30 min
Blue OCP110 and OCP210 10 mm 17 × 13 mm 7:30 min
Orange OCP115 and OCP215 15 mm 23 x 18 mm 11:30 min
Green OCP120 and OCP220 20 mm 29 × 21 mm 15:00 min
PERFORMING RF ABLATION
SACRUM, ILIAC, AND ACETABULUM
Note
The ablation process will remain the same
for all bony anatomy. Follow steps 1–7 under
“PERFORMING RF ABLATION — SPINE” for
detailed instructions regarding this process.
One-probe approaches:
A3
B3
C3
24 25
OPTIONAL TRACK
RF ABLATION
1. To ablate the track, select “Retract” on the generator screen.
2. Press the RF power button.
3. The screen will signal when the track ablation starts.
4. Slowly pull back on the RF ablation probe and cannula
simultaneously (approximately 1–2 mm/second) or pull back and
track ablate while stationary. When the desired end point for track
ablation is reached, press the RF power button to stop RF delivery.
Warning
Track ablation is the use of uncooled RF ablation during which time
the peristaltic pump is not circulating water. Ablating the track all
the way to the skin can result in skin burns. It is recommended to
end the track ablation prior to this point.
Two-probe approaches:
A4
B4
C4
26 27
OPTIONAL
TEMPERATURE
MONITORING
WITH THE INDEPENDENT
THERMOCOUPLE
The independent thermocouple provides real-time temperature
information to gauge ablation zone margins and help prevent
thermal damage to adjacent critical structures. The temperature
detected by the independent thermocouple is displayed on the
generator screen.
1. Insert the thermocouple into the provided introducer.
2. Connect the male connector on the thermocouple to the female
connector on the connector hub.
3. Conrm via on-screen information that the thermocouple
(sensor) has been detected.
4. Access the site. Place the thermocouple in the tissue to be
monitored.
PERFORMING RF
ABLATION USING
CUSTOM SETTINGS
The OsteoCool™ system generator screen includes the ability to
change procedure settings. These settings should only be adjusted
by physician experts in RF ablation. To access the custom settings
on the generator, complete the following steps:
1. Press the “Settings” button.
2. Select “Procedure Settings.”
3. Edit settings based on procedural need. Each probe can be
individually adjusted by time, temperature, power limit, ramp
rate, and impedance cut o. Additionally, each independent
thermocouple sensor can be individually adjusted by
temperature.
4. Once desired settings have been achieved, press the “X” button
to return to the generator’s home screen user interface.
5. Proceed with the RF ablation procedure.
Note
All ablation zone sizing is based on default settings. Using custom
settings by manipulating the procedure settings will change the
ablation zone sizing. To reset the generator settings to default
settings, select the “Refresh” button.
BONE ACCESS
FOR SUBSEQUENT
PROCEDURES
The bone access can be used for subsequent physician-
directed procedures such as cementoplasty, sacroplasty,
vertebroplasty, or kyphoplasty. Conrm such procedures are
compatible with the OsteoCool™ Bone Access Introducer
length and gauge.
GENERAL
SAFETY GUIDELINES
For desired tissue heating and anatomical access:
Avoid advancing the probe into bone tissue without rst using
the introducer and drill to create a clear channel for the probe.
Forcing the probe through bone tissue may cause damage to
the probe.
Indications for Use
Kyphon Xpede Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis,
cancer, or benign lesions using a cementoplasty (i.e., kyphoplasty or vertebroplasty) procedure. It is also indicated for the xation
of pathological fractures of the sacral vertebral body or ala using sacral vertebroplasty or sacroplasty. Cancer includes multiple
myeloma and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include
hemangioma and giant cell tumor. Pathologic fracture may include a symptomatic vertebral body microfracture (as documented by
appropriate imaging and/or presence of a lytic lesion) without obvious loss of vertebral body height.
Kyphon HV-R Bone Cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis,
cancer, or benign lesions using a cementoplasty (i.e., kyphoplasty or vertebroplasty) procedure. Cancer includes multiple myeloma
and metastatic lesions, including those arising from breast or lung cancer, or lymphoma. Benign lesions include hemangioma and
giant cell tumor. Pathologic fracture may include a symptomatic vertebral microfracture (as documented by appropriate imaging
and/or presence of a lytic lesion) without obvious loss of vertebral body height.
28 29
OsteoCool™ RF Ablation Capital Equipment
OsteoCool™ RF Generator OC01
OsteoCool™ RF Pump OC02
OsteoCool™ RF Pump Cable OC03
OsteoCool™ Connector Hub OC04
OsteoCool™ RF Ablation Probes
Probe Kit OsteoCool™ RF 17G 7 mm OCP107
Probe Kit OsteoCool™ RF 17G 10 mm OCP110
Probe Kit OsteoCool™ RF 17G 15 mm OCP115
Probe Kit OsteoCool™ RF 17G 20 mm OCP120
Dual Probe Kit OsteoCool™ RF 17G 7 mm OCP207
Dual Probe Kit OsteoCool™ RF 17G 10 mm OCP210
Dual Probe Kit OsteoCool™ RF 17G 15 mm OCP215
Dual Probe Kit OsteoCool™ RF 17G 20 mm OCP220
OsteoCool™ Bone Access Kits
OsteoCool™ Bone Access Kit 10G 090 OCN002
OsteoCool™ Bone Access Kit 8G 090 OCN003
OsteoCool™ Bone Access Kit 10G 095 OCN004
OsteoCool™ Bone Access Kit 13G 100 OCN005
OsteoCool™ Carts and Stands
OsteoCool™ Cart Stand OCA01
OsteoCool™ Desk Stand OCA02
Temperature Monitoring
Thermocouple OsteoCool™ ITC 180 mm OCN001
PRODUCT ORDERING
INFORMATION IMPORTANT PRODUCT
INFORMATION
§ The OsteoCool™ RF Ablation System is intended
for the palliative treatment in spinal procedures by
ablation of metastatic malignant lesions in a vertebral
body. It is also intended for coagulation and ablation
of tissue in bone during surgical procedures, including
palliation of pain associated with metastatic lesions
involving bone in patients who have failed or are not
candidates for standard therapy.
Risks of the system include damage to surrounding
tissue through iatrogenic injury as a consequence of
electrosurgery; pulmonary embolism; nerve injury
including thermal injury, puncture of the spinal cord
or nerve roots potentially resulting in radiculopathy,
paresis, and paralysis.
§ OsteoCool™ Bone Access Kits: Indicated for
percutaneous access to bone.
§ OsteoCool™ Independent Thermocouple: Intended
for measuring tissue temperature throughout an RF
ablation procedure.
OSTEOCOOL RF ABLATION SYSTEM
AND BONE ACCESS KITS
NOTES:

Navigation menu