Captured Hip Screw Surg Technique
2016-04-01
: Pdf Captured Hip Screw Surg Technique Captured_Hip_Screw_Surg_Technique 3 2016 pdf
Open the PDF directly: View PDF  .
.
Page Count: 24

Surgical Technique
Options:
Supracondylar Plate
Trochanteric Plate
Captured Hip® 
Screw System
TRAUMA

Captured Hip® Screw System
Contents
Introduction .................................................................................................................................................................. 1
Femoral Neck and Intertrochanteric Fractures of the Femur ...................................................................................... 3
Supracondylar Fractures and Fractures of the Lower Third Femur ............................................................................ 9
Catalog Numbers and Descriptions ........................................................................................................................... 18
Ordering Information .................................................................................................................................................. 20
2

Written by Richard F. Kyle, MD
The implants in the Captured Hip Screw with Trochanteric 
and Supracondylar Plate System are made of Ti-6Al/4V. 
Titanium alloy, when compared to 316L stainless steel, 
offers biocompatibility, and strength.1 
The captured hip screw is very simple to insert. It is a keyless 
system. A single instrument is used to insert the assembled 
captured hip screw and plate in one step. This concept has 
been carried forward into the design of the instruments for 
the keyed supracondylar plate. Following placement of the 
central lag screw, the plate is simply guided over the driver 
onto the lag screw and against the bone. 
Introduction
The captured hip screw uses the principle of sliding 
impaction, rather than compression with a compression 
screw. By firmly impacting the fracture after placement of 
the hip screw, a compressive load that is approximately four 
times greater than that of a compression screw is applied to 
close the fracture site.² This method eliminates the potential 
of stripping the lag screw threads in osteoporotic bone and 
avoids the need for a compression screw.
The supracondylar plate utilizes screw-controlled 
compression. Since the load is perpendicular to the sliding 
mechanism, distal femoral fractures do not impact naturally. 
A keyed mechanism is utilized to enhance rotational stability. 
This system provides simple solutions to the surgeon’s 
common hip and knee fracture needs. All implants, 
screws and instruments fit into one tray. The system is 
sensibly designed to address ease of use and inventory 
reduction, two important cost factors when considering 
managed care.
The Captured Hip Screw System System was designed and developed in 
conjunction with Richard F. Kyle, M.D.
This hip fracture surgical technique is utilized by Richard F. Kyle, M.D. Biomet 
as the manufacturer of this device, does not practice medicine and does not 
recommend this device or technique. Each surgeon is responsible for determining 
the appropriate device and technique to utilize on each individual patient.
1

Description
The captured hip screw system is designed for use in 
intertrochanteric fractures. Intracapsular fractures may 
be managed with the addition of solid or cannulated 
cancellous screws to prevent rotation. 
Subtrochanteric fractures may be treated with the device 
provided that a good medial buttress is achieved and 
weight-bearing is not allowed until the fracture shows 
evidence of union.
The trochanteric plate option can be used to control 
comminution of the greater trochanter as well as reverse 
oblique and high subtrochanteric fractures by preventing 
medialization of the femoral shaft.
The supracondylar plate is designed for use in 
supracondylar fractures, and fractures and osteotomies of 
the proximal and distal femur.
Captured Hip® Screw System
2

Surgical Technique
Patient Positioning
Place the patient supine on a standard fracture table. 
Reduce and align the fracture using traction with external 
rotation followed by approximately 20 degrees of internal 
rotation to compress the fracture (Figure 1). Verify reduction 
using dual-plane image intensification. Prepare and drape 
the hip in the usual manner.
Incision
Make a 10 cm incision in the lateral aspect of the hip, with 
dissection beginning at the flare of the greater trochanter 
and extending distally (Figure 1). Carry the dissection 
sharply down through the skin and subcutaneous tissue to 
the fascia lata. Split the fascia lata longitudinally, exposing 
the vastus lateralis. Retract the vastus lateralis anteriorly 
and expose the lateral aspect of the femoral shaft.
Femoral Neck and Intertrochanteric Fractures of the Femur
20°
Traction
Incision
Guide Pin Positioning
Place a 3.2 mm guide pin on the anterior aspect of the 
femoral shaft and neck, 3-4 cm distal to the flare of the 
greater trochanter. Grooves are provided on the outer 
surfaces of the guide pin jig to indicate the pin angle 
options (and the subsequent captured screw assembly 
angle) and thus the proper placement of the hole for the 
internal guide pin. The guide pin should lie at the highest 
angle necessary to position the pin next to the medial 
cortex and in the center of the femoral head in the anterior/
posterior (A/P) plane (Figure 2).
Drill the 3.2 mm guide pin through the corresponding hole 
in the guide pin jig. Advance it into the center of the femoral 
head under image control in both the A/P and lateral planes 
to within 5-7 mm of the subchondral bone (Figure 3). 
Note: Central placement of the guide pin is the single 
most important step in secure fixation of the proximal 
fragment.
Figure 2Figure 1 Figure 3
5–7 mm
3

Guide Pin Depth  
Gauge Reading
70–100 mm
100–130 mm
130–160 mm
Table 1 Figure 5
Figure 6
Figure 4
Captured Screw Assembly Selection
Place the guide pin depth gauge over the pin up to the 
femoral cortex (Figure 4). The measurement read at the 
end of the guide pin is a direct measurement of the length 
of the pin that extends into the femoral neck and head.
Using this measurement, select the proper length screw. 
Table 1 provides screw selection for corresponding depth 
gauge readings.
Note: In some cases a reading between 60 and 70 mm 
may be taken. In this situation, a special short barrel 
hip screw should be utilized. This special captured 
screw assembly is available in a 135-degree angle. 
Once the measurement is read and the proper captured  
screw assembly is determined, advance the guide pin  
into subchondral bone, thereby anchoring it for reaming  
and tapping.
Reaming and Tapping
The long adjustable reamer and calibrated tap can be set in  
5 mm increments. In patients with healthy bone, set the 
reamer at 10 mm less than the position just below the 
actual guide pin depth gauge reading (i.e., a reading 
between 90 and 95 mm means that the reamer should 
be set at 80 mm). This ensures that the reamer never 
advances closer than 1 cm to the subchondral bone. 
Note: In patients with osteoporotic bone, set the 
reamer at 70 mm (Figure 5).
Place the reamer over the guide pin and advance it into 
the proximal femur under image control, assuring that the 
guide pin does not move (Figure 6).
Note: The conical portion of the reamer only needs 
to remove bone in the distal part of the lateral hole to 
prepare the bone for the captured screw assembly.
Recommended Captured 
Screw Assembly Length
100 mm
130 mm
160 mm
Captured Hip® Screw System
4

Figure 9
Figure 10
5–7 mm
Figure 7
Note: To set the reamer to the desired depth, depress 
the button and slide the reamer head until the arrows 
line up with the matching value.
Set the calibrated tap at the position just below the actual 
guide pin depth gauge reading (i.e., a reading between  
90 and 95 mm means that the tap should be set at  
90 mm). Place the calibrated tap over the guide pin and 
tap the neck and head of the femur to within 5–7 mm of 
subchondral bone under image control (Figure 7).
Hip Screw Insertion
Ensure that the side plate is long enough to allow for the 
placement of four screws (eight cortices) below the most 
proximal extent of the fracture. While this makes the four-
hole plate the most common selection, a two-hole plate 
may be selected for a nondisplaced femoral neck fracture. 
Assemble the side plate and captured screw assembly by 
sliding the barrel flange into the plate slot (Figure 8).
Figure 8
Insert the T-wrench inserter into the hexagonal base of the 
hip screw and place the entire assembly over the central 
guide pin into the prepared lateral cortex.
Advance the hip screw by turning the T-wrench inserter 
in a clockwise manner (Figure 9). While the side plate can 
remain loose during initial advancement, it is important 
to hold the plate parallel to the femoral shaft as it nears 
the bone. Verify the depth of the hip screw using image 
intensification. Remove the guide pin.
To ensure that the barrel portion of the captured screw 
assembly is fully seated, position the assembled captured 
hip screw impactor head and impactor handle onto the 
angled superior portion of the plate. With a mallet, lightly 
tap the impactor (Figure 10). 
5

Figure 11
Figure 14a Figure 14bFigure 12 Figure 13
Caution: Only light tapping of the screw is necessary. 
Forceful use of the impactor after the barrel has been 
inserted into the femoral shaft and prior to fixation of 
the side plate to the lateral aspect of the shaft may 
result in greater trochanteric comminution.
Screw Placement
Ensuring that the side plate or trochanteric plate is fully 
engaged to the captured screw assembly, place the 3.8 
mm drill guide into the proximal screw slot. Drill both 
cortices using a 3.8 mm drill bit (Figure 11). 
Measure the proper screw length using the hook depth  
gauge (Figure 12). 
Assemble the power adaptor or quick couple T-handle to 
the 4.5 mm solid hex driver shank and drive the 4.5 mm 
self-tapping screw through the hole, securing the plate to 
the femoral shaft (Figure 13). 
Note: In osteoporotic bone or in the metaphyseal 
portion of the distal femur, 5.0 mm cancellous bone 
screws are provided for fixation.
Always perform final tightening by hand. Using this 
technique, place the screws in succession in the plate 
slots to fasten the plate to the bone (Figure 14a).
Note: The proximal hole in the side plate may be used to 
place an angled 6.5 mm solid or cannulated cancellous  
bone screw if lagging of medial bone fragment is  
required (Figure 14b).
Captured Hip® Screw System
6

Figure 17
Figure 18Figure 16
Final Seating and Impaction of the Hip Screw
Following fixation of the side plate to the femoral shaft, 
attain final seating of the hip screw into the femoral head 
by using the T-wrench inserter. Locate the tip of the hip 
screw within 5-7 mm of subchondral bone as verified  
by image intensification. Release traction from the  
effected hip.
Impact the fracture by placing the assembled impactor 
over the proximal aspect of the plate and applying three to 
four firm blows with a mallet (Figure 15).
Note: 6.5 mm cannulated cancellous screws are 
provided in the system if additional fixation is desired 
anterior or posterior to the captured screw assembly. 
Place a 3.2 mm guide wire under image intensification 
(Figure 16).
Insert the self-drilling, self-tapping cannulated screw over 
the guide wire using the assembled 6.5 mm hex driver 
Figure 15
shank and T-handle (Figure 17). Cannulated drills and taps 
are available if hard, dense bone is encountered.
Optional Compression
Alternatively, compression may be applied by placing the  
T-wrench inserter into the screw and locking it there with  
the extractor lock. Grasp the T-wrench inserter and pull on  
it forcefully to compress the fracture (Figure 18).
7

Caution: In patients with intertrochanteric fractures 
with subtrochanteric components, weight-bearing 
is prohibited until callus formation is apparent on 
the X-ray.
Extraction
After performing a routine opening exposure, use the  
4.5 mm solid and 6.5 mm hex driver shanks to remove 
the screws fixing the side plate to the shaft of the femur. 
Attach the extractor lock through the T-wrench inserter 
to the captured screw assembly. Remove the hip screw 
by turning the T-wrench inserter in a counter-clockwise 
manner. Close in a routine fashion.
Captured Hip® Screw System
8

Screw Placement with  
Trochanteric Side Plate Option
Control comminution of the greater trochanter by using the 
trochanteric side plate (Figure 19). This plate also benefits 
reverse oblique and high subtrochanteric fractures by 
preventing medialization of the femur shaft.
Reduce the bony fragments using forceps or clamps and 
proceed in the same fashion as the side plate fixation. The 
quantity and sequence of screw placement in the proximal 
flare is physician dependent.
The most proximal screw hole provides fixation for the 
greater trochanter and the two distal screw holes also 
provide greater trochanter fixation. The two distal screw 
holes may also be used for anti-rotation of the femoral 
head, reverse oblique and subtrochanteric fractures to 
prevent medialization of the femoral shaft.
Figure 19 Figure 21
Figure 22Figure 20
Place the 3.8 mm drill guide into one of the flared screw 
holes and drill to the desired depth using a 3.8 mm drill bit 
(Figure 20).
Measure the proper screw length using the hook depth  
gauge (Figure 21).
Caution: When using the hook depth gauge in a blind 
hole (i.e., one that does not go through both corticies), 
select a screw 2 mm less (or the next smaller size) 
than indicated.
Assemble the power adaptor or quick couple T-handle 
to the solid hex driver shank and drive the 6.5 mm self-
tapping screw through the hole securing the plate to the 
bone (Figure 22). 
9

Figure 23
Note: For placement of the 6.5mm screws, use the 
respective drivers below.
Always perform final tightening by hand. An example 
of screw placement for the fracture indicated is shown 
(Figure 23).
Note: For final seating and impaction of the hip screw, 
please see next section.
Caution: In patients with intertrochanteric fractures 
with subtrochanteric components, weight-bearing is 
prohibited until callus formation is apparent on the 
X-ray.
Extraction
After performing a routine opening exposure, use the 
4.5 mm solid and 6.5 mm hex driver shanks to remove 
the screws fixing the side plate to the shaft of the femur. 
Attach the extractor lock through the T-wrench inserter 
to the captured screw assembly. Remove the hip screw 
by turning the T-wrench inserter in a counter-clockwise 
manner. Close in a routine fashion.
Discontinued Cat. Nos. Driver
14520-xx 14616
14521-xx
14522-xx
Active Cat. Nos. Driver
8157-61-xxx 14541
8157-62-xxx
8157-64-xxx
Captured Hip® Screw System
10

Supracondylar Fractures and 
Fractures of the Lower Third Femur    
Patient Positioning
Pay careful attention to the positioning of the patient on 
the standard operating table. Place a sandbag under 
the buttock of the operated extremity with the patient in 
the supine position. This effectively rotates the extremity 
internally and facilitates the surgical approach. Use a roll 
under the leg to provide 20–30 degrees flexion (Figure 
24). The same positioning can also be achieved using a 
fracture table.
Incision
Perform a standard lateral exposure starting at the lateral 
femoral epicondyle and extend it proximally.
Note: The initial incision over the femoral epicondyle may 
be limited to allow insertion of the lag screw and later 
extended for plate application. This reduces overall wound 
exposure and blood loss.
A straight lateral incision allows extension distally and 
proximally for an extensile exposure. Incise the suprapatellar 
pouch at the edge of the epicondyle and reflect medially 
in order to expose the intercondylar notch. The lateral 
superior genicula artery requires cauterization or ligation. 
Retract the vastus lateralis muscle anteriorly to expose the 
fracture. Place a retractor posterior to the femur to protect 
the posterior neurovascular structures.
Convert a three-part fracture into two parts by reducing the 
intercondylar component. Hold the reduction temporarily 
with 3.2 mm Kirschner wires and bone-holding clamps or 
reduction forceps. Take care in placing the K-wires so as 
not to interfere with the placement of the supercondylar lag 
screw. Cannulated screws may be used to permanently 
secure the fragments.
Figure 24
Incision
20 - 30 degrees flexion
Supracondylar Fractures 
and Fractures of the Lower Third Femur
11

Guide Pin
10 degrees
Guide Pin Positioning
Assemble the central pin guide handle to the central pin 
guide and place against the lateral wall of the femoral 
condyles. Drive a 3.2 mm threaded guide pin through 
the guide and extend to, but not through, the medial  
cortex (Figure 25).
Note: This pin should pass 2 cm from the articular 
cartilage.
It is important that the 3.2 mm guide pin is positioned at  
95 degrees to the shaft and parallel to the joint axis 
angled about 10 degrees (parallel to the inclination of 
the patellofemoral joint) anteriolateral to posteriomedial 
(Figures 26a, 26b). Verify the pin position with image 
intensification. Remove the central pin guide.
Lag Screw Size Selection
Measure the central guide pin depth using the guide pin 
depth gauge to determine the appropriate lag screw length 
(Figure 27). Lag screws are provided in 10 mm increments.
When using the depth gauge, the reading will fall within a 
10 mm range (i.e., between 70-80 mm). If the depth gauge 
reading falls in the lower half of the 10 mm range, select 
a screw size 10 mm shorter than the lowest mark in the 
range. For example, if the reading is between 70 and 75 
mm, use a size 60 mm screw.
Using the shorter screw ensures that the screw will not 
pass through the medial condyle and that there will be 
adequate length to compress the fracture. If the depth 
gauge reading falls in the upper half of the 10 mm range, 
select a screw the same size as the lowest reading in 
the range. For example, if the reading is between 75 and 
80 mm use a 70 mm screw. This method allows for a 
minimum of 5 mm of compression for most every screw 
size.
Note: If the measurement falls between 50 and 55 mm, 
then the 50 mm screw must be used.
Figure 25
Figure 26a
Guide Pin
2 cm
Figure 26b
Figure 27
Captured Hip® Screw System
12

Reaming and Tapping
The short adjustable reamer and calibrated tap can be set in 
5 mm increments. If the guide pin depth gauge reads in the 
lower half of the 10 mm range, set the reamer and tap at the 
lower reading (i.e., set at 70 mm if the reading is between 
70 and 75 mm). If the depth gauge reads in the upper half 
of the 10 mm range, set the reamer or tap at a mid-range 
setting (i.e., 75 mm for a reading between 75 and 80 mm). 
For clarity, Table 2 provides instrument settings and lag 
screw selection for corresponding depth gauge readings.
Note: Perform reaming and tapping under image 
intensification to ensure that the far cortex is not 
compromised.
Note: In osteoporotic bone, set the reamer at its 
shortest depth because additional reaming is not 
necessary.
Advance the step reamer over the guide pin and drill the 
hole (Figure 28).
Assemble the tap to the quick couple T-handle and use it  
to prepare the femoral condyle for placement of the lag  
screw (Figure 29).
In the event that the guide pin comes out in either the 
reaming or tapping step, a guide pin centralizer (same as 
the central pin guide handle) is provided for replacement 
of the guide pin (Figure 30).
Figure 28 Figure 29 Figure 30
Guide Pin Depth 
Gauge Reading
50-55 mm
55-60 mm
60-65 mm
65-70 mm
70-75 mm
75-80 mm
80-85 mm
85-90 mm
90-95 mm
95-100 mm
Reamer or 
Tap Setting
50 mm
55 mm
60 mm
65 mm
70 mm
75 mm
80 mm
85 mm
90 mm
95 mm
Recommended 
Lag Screw Length
50 mm
50 mm
50 mm
60 mm
60 mm
70 mm
70 mm
80 mm
80 mm
90 mm
Maximum 
Compression
0 mm
5 mm
10 mm
5 mm
10 mm
5 mm
10 mm
5 mm
10 mm
5 mm
Table 2
13

Plate and Lag Screw Insertion
Ensure that the supracondylar plate is long enough to 
allow for the placement of four screws (eight cortices)  
above the most proximal extent of the fracture. Slide 
the chosen supracondylar plate onto the shaft of the lag 
screwdriver. Begin driving the lag screw using the lag 
screwdriver (Figure 31).
To assist in determining the insertion of the lag screw,  
0, +5 and +10 mm lines are indicated on the lag screwdriver.
CAUTION: Never advance the lag screwdriver past the 
+10 mm line to ensure proper engagement of the barrel
over the screw (Figure 32).
Note: In good quality bone, the lag screw can simply 
engage the cancellous bone of the femoral condyle. 
In osteoporotic bone, it is beneficial to engage one 
thread of the lag screw in the cortical bone of the 
medial condyle.
Drive the lag screw to the proper depth, making sure that  
the T-handle of the lag screwdriver is parallel to the  
femoral shaft and plate. This positioning ensures that 
the supracondylar plate will align with the femoral shaft  
(Figure 33).
Note: When the handle is parallel to the plate, this 
ensures proper alignment of the keys on the screw to 
the keyways in the barrel.
Verify the position of the lag screw using the image 
intensifier (Figure 34).
Using the driver shank as a guide, slide the supracondylar 
plate along the driver and onto the lag screw (Figure 35).
Ensure that the plate lines up with the femoral shaft. A 
standard plate impactor head has been provided for 
impaction through the barrel hole. Thread the standard 
Figure 35
Figure 33
Figure 34
Figure 31 Figure 32
Captured Hip® Screw System
14

impactor head onto the impactor handle. Prior to 
impaction, ensure that the lag screw does not extend 
outside of the lateral surface of the bone. Gently impact 
the supracondylar plate against the femoral condyle using 
the assembled impactor (Figure 36).
Screw Placement
Hold the supracondylar plate in place with one or twobone 
clamps after re-establishing the proper bone length. 
Perform minimal to no periosteal stripping. If there is shaft 
comminution and indirect reduction has been used, use 
four screws in the proximal four slots to fix the plate to the 
shaft. Screws may also be incorporated for adjunct fixation 
in the region of comminution where good purchase can 
be obtained without periosteal stripping. If there is little or 
no comminution and compression of the shaft is desired, 
place the distal screws first.
Note: The distal hole in the neck of the plate may be used  
to place a 6.5 mm solid or cannulated cancellous bone  
screw if lagging of metaphysical bone to the femur is 
required (Figure 37).
Note: In hard or dense bone, tap with the solid or 
cannulated 6.5 mm cancellous tap.
Place the 3.8 mm drill guide into a screw slot (Figure 38).
Drill both cortices using a 3.8 mm drill bit. Measure the 
proper screw length using the hook depth gauge (Figure 39).
Assemble the power adaptor or quick couple T-handle to the 
6.5 mm hex driver shank and drive the 6.5 mm self-tapping 
screw through the hole, securing the plate to the femoral 
shaft (Figure 40). Always perform final tightening by hand.
Figure 39 Figure 40
Figure 37 Figure 38
Figure 36
15

In a similar fashion, use the 4.5 mm solid hex driver shank 
to place 4.5 mm cortical bone screws in succession in the 
plate slots to fasten the plate (Figure 41).
Note: In osteoporotic bone or in the metaphyseal 
portion of the distal femur, 5.0 mm cancellous screws 
are provided for fixation.
Compressing a Condylar Fracture
Insert the short compression screw into the lag screw and 
turn it clockwise using the 6.5 mm hex driver shank until  
the compression screw engages the supracondylar plate 
(Figure 42). Continue turning clockwise until the condylar 
fracture has closed the desired amount (insets a and b) as 
viewed under image intensification.
Closure and Postoperative Instructions
Confirm final positioning of the implant and proper fracture 
reduction using X-ray or image intensification. If significant 
medial comminution exists, consider bone grafting 
medially. Close in a routine fashion.
Early motion of the knee is preferable if fracture repair 
is stable. Delay weight-bearing until callus formation is 
identified by X-ray.
Extraction
After performing a routine opening exposure, use the  
4.5 mm solid and 6.5 mm hex driver shanks to remove 
screws fixing the plate to the shaft of the femur. Use the  
6.5 mm hex driver to remove the short compression screw. 
Remove the supracondylar plate. Attach the extraction rod 
through the lag screwdriver to the lag screw (Figure 43). 
Remove the lag screw. Close in a routine fashion.
Figure 41
Figure 42
AB
Figure 43
Captured Hip® Screw System
16

Figure 44a Figure 44b
Reamer Shaft
Keyway
Key
Reamer Head
Reamer Disassembly
Depress the button and slide the reamer head along the 
shaft away from the drill flutes.
Once the reamer head is off, remove the button for cleaning 
purposes (Figure 44a).
Do not disassemble the button and spring from one another.
Reamer Assembly
Depress the button and slide the reamer shaft through the 
hole of the reamer head.
Take care to align the keyway slot of the reamer shaft with 
the key inside the reamer head (Figure 44b).
Button
Spring
Reamer Head
17

Captured Hip Screw Implants
Side Plate (Cat. Nos. 140xx)
•  Available in two – 14-hole configurations (two-hole 
increments)
•  Flexibility allows plate conformity to the femur
•  Low profile design
•  Oval compression plate screw holes allow optimum 
positioning of the bone screws (important in the securing of 
medial fragments)
•  Proximal hole allows for use of 6.5 mm screw
Trochanteric Side Plate (Cat. Nos. 14252-4, 6)
•  Available in four and six-hole configurations
•  Same features as standard side plate
•  Added benefit of proximal buttress flange that conforms to 
the greater trochanter
•  Proximal screw holes to stabilize severe comminution of 
greater trochanter and for reverse oblique fracture patterns
Captured Screw Assembly (Cat. Nos. 14033-x, Cat. Nos. 14007-x)
•  Angled barrels available in 135, 140, 145 and 150 degrees
•  Captured screws for each angle accommodate lengths of 90 
to 160 mm in three assemblies, thus reducing inventory
•  Short barrel accommodating 60 to 90 mm length available in 
the most commonly used 135-degree angle
Supracondylar Plate Implants
95-degree Standard Supracondylar Plate (Cat. Nos. 14552-xx)
•  Available in six – 14-hole configurations (two-hole increments)
•  Low profile plate
•  Distal hole allows for use of 6.5 mm screw
Standard Lag Screw (Cat. Nos. 14553-xx)
•  Available in 50 to 90 mm lengths (10 mm increments)
•  Unitized insertion with the supracondylar plate
Short Compression Screw (Cat. Nos. 8113-05-003)
•  Facilitates up to 10 mm of compression
•  Smooth radiused compression screw/plate profile
Captured Hip Screw with Supracondylar and Trochanteric Plate System
Catalog Numbers and Descriptions
Captured Hip® Screw System
18

Self-Tapping Solid Screws
6.5 mm Solid 22 mm Threaded Cancellous Lag Screw  
(Cat. Nos. 8157-62-xxx)
•  Available in 5 mm increments from 40–120 mm lengths
6.5 mm Solid 40 mm Threaded Cancellous Lag Screw  
(Cat. Nos. 8157-64-xxx)
•  Available in 5 mm increments from 60–120 mm lengths
6.5 mm Solid Fully-Threaded Cancellous Bone Screw 
(Cat. Nos. 8157-61-xxx)
•  Available in 5 mm increments from 25–120 mm lengths
5.0 mm Solid Cancellous Fully-Threaded Bone Screw  
(Cat. Nos. 14224-xx)
•  Available in 2 mm increments from 20–60 mm lengths 
and 5 mm increments from 65–80 mm lengths
4.5 mm Solid Cortical Fully-Threaded Bone Screw  
(Cat. Nos. 14022-xx)
•  Available in 2 mm increments from 26–58 mm lengths 
•  Double lead thread design for bone
Self-Tapping, Self-Drilling Cannulated Screws
6.5 mm Cannulated 22 mm Thread Cancellous Lag Screw  
(Cat. Nos. 14196-xx)
•  Available in 5 mm increments from 25–120 mm lengths
•  Self-drilling and self-tapping
•  Reverse radial tapping flutes for ease of removal
•  3.2 mm guide wire
6.5 mm Cannulated 40 mm Thread Cancellous Lag Screw  
(Cat. Nos. 14197-xx)
•  Available in 5 mm increments from 40–120 mm lengths
•  Self-drilling and self-tapping
•  Reverse radial tapping flutes for ease of removal
•  3.2 mm guide wire
19

Captured Screw Assemblies
Cat. No. Description Length mm
14033-0 135° 90
14033-1 135° 100
14033-2 135° 130
14033-3 135° 160
14007-1 140° 100
14007-2 140° 130
14007-3 140° 160
14033-4 145° 100
14033-5 145° 130
14033-6 145° 160
14007-4 150° 100
14007-5 150° 130
14007-6 150° 160
Side Plates
14027 2 Hole Plate 67
14000 4 Hole Plate 97
14029 6 Hole Plate 128
14030 8 Hole Plate 158
14031 10 Hole Plate 189
14034 12 Hole Plate 220
14032 14 Hole Plate 250
142524 Troch Plate 4 hole 112
142526 Troch Plate 6 hole 114
95° Standard Supracondylar Plates
14552-6 6 slot 131
14552-8 8 slot 161
14552-10 10 slot 203
14552-12 12 slot 239
14552-14 14 slot 275
Lag Screws
Cat. No. Length mm
14553-50 50
14553-60 60
14553-70 70
14553-80 80
14553-90 90
Compression Screw
Cat. No. Description
8113-05-003 Short Compression Screw
Screws
Cat. No. Description
4.5 mm Solid Cortical Fully-Threaded Bone Screws*
14022-xx 26 – 58 mm (2 mm increments)
5.0 mm Solid Cancellous Fully-Threaded Bone Screws*
14224-xx 20 – 60 mm (2 mm increments)
14224-xx0 65 – 80 mm (5 mm increments)
6.5 mm Solid 22 mm Thread Cancellous Lag Screws*
8157-62-xxx 40 – 110 mm (5 mm increments)
6.5 mm Solid 40 mm Thread Cancellous Lag Screws*
8157-64-xxx 60 – 110 mm (5 mm increments)
6.5 mm Solid Cancellous Fully-Threaded Bone Screws*
8157-61-xxx 25 – 110 mm (5 mm increments)
6.5 mm Cannulated 22 mm Thread Cancellous Lag Screws*
14196-xx 25 – 120 mm (5 mm increments)
6.5 mm Cannulated 40 mm Thread Cancellous Lag Screws*
14197-xx 40 – 120 mm (5 mm increments)
Instruments
Cat. No. Description
14008 T-wrench Inserter
14011 Extractor Lock
14052 Guide Pin Jig
14142 Guide Pin Depth Gauge
14315 Power Adaptor
14316 T-handle
8242-00-120 Hook Depth Gauge
14540 3.8 mm Drill Guide
14541 4.5 mm Solid Hex Driver Shank
14549 Central Pin Guide - Standard
14566 Extraction Rod
14569 Calibrated Lag Screw Tap
14570 Lag Screw Driver
14577 Central Pin Guide Handle/Repositioner
14578 Impactor Handle
14580 Captured Hip Screw Impactor Head
14581 Standard Plate Impactor Head
14585-1 Adjustable Reamer - Short
14585-3 Adjustable Reamer - Long
14616 6.5 mm Hex Driver Shank
Disposables
14012-9 3.2 mm x 9 in. Guide Pin
14545 3.8 mm Solid Twist Drill, 7 in.
14584 6.5 mm Solid Cancellous Tap
14627 6.5 mm Cancellous Tap
14629 5.0 mm Cannulated Drill
Modules and Cases
14530 Solid and Cannulated Screw Module
14532 Captured Hip Screw Module
14533 Standard Supracondylar Plate Module
14534 General Instruments Module
14535 Screw Instruments Module
14602 Sterilization Case (10 in. x 10 in. x 5 in.)
14603 Sterilization Case (10.5 in. x 20 in. x 5 in.)
Ancillary Part
14599 Reamer Button Spring Assembly
*
Warning: This device is not approved for screw attachment or fixation 
to the posterior (pedicles) of the cervical, thoracic or lumbar spine.
Ordering Information
Captured Hip® Screw System
20

General Instruments Module  
Cat. No.14534
Cat. No.                  Description
1 14052 Guide Pin Jig
2 14012-9 3.2 mm X 9 In. Guide Pin*
3 14142 Guide Pin Depth Gauge
4 14585-3 Adjustable Reamer - Long
5 14585-1 Adjustable Reamer - Short
6 14569 Calibrated Lag Screw Tap
7 14578 Impactor Handle
8 14580 Captured Hip Screw Impactor Head
9 14008 T-Wrench Inserter
10 14011 Extractor Lock
12
3
4
10
9
7
5
6
8
11
16 17
18
15
14
19
20
Instruments Contained In  
Standard Supracondylar Plate Module
Cat. No. 14533
 Cat. No. Description
21 14581 Standard Plate Impactor Head
22 14549 Central Pin Guide - Standard
23 14577 Central Pin Guide Handle/Repositioner
24 14566 Extraction Rod
25 14570 Lag Screwdriver
Ancillary Parts
 Cat. No. Description
           14599    Reamer Button Spring Assembly 
                         (Not Shown)
Screw Instruments Module
Cat. No. 14535
Cat. No. Description
11 14540 3.8 mm Drill Guide
12 14545 3.8 mm Solid Twist Drill, 7 in.*
13 14584 6.5 mm Solid Cancellous Tap*
14 14541
4.5 mm Solid Hex Driver Shank
15 14616 6.5 mm Hex Driver Shank
16 14316 T-handle
17 14315 Power Adaptor
18 14629 5.0 mm Cannulated Drill*
19 14627 6.5 mm Cancellous Tap*
20 8242-00-120 Hook Depth Gauge
* Disposables
21

All trademarks herein are the property of Biomet, Inc. or its subsidiaries unless 
otherwise indicated.
This material is intended for the sole use and benefit of the Biomet sales force and 
physicians. It is not to be redistributed, duplicated or disclosed without the express 
written consent of Biomet.
For product information, including indications, contraindications, warnings, 
precautions and potential adverse effects, see the package insert.
Responsible Manufacturer
Biomet Trauma 
P.O. Box 587
56 E. Bell Drive
Warsaw, Indiana 46581-0587 
USA
www.biomet.com
©2014 Biomet Trauma • Form No. BMET0134.0 • REV0814
References
  1.   DVA-107504-DVER.
  2.  Massie, William K.   “Fractures of the Hip.” Journal of Bone and Joint 
Surgery, April 1964: 669. 
Important:  This Essential Product Information does not include all of the informa-
tion necessary for selection and use of a device. Please see full labeling for all 
necessary information. 
The use of metallic surgical appliances provides the orthopaedic surgeon a means 
of bone fixation and helps generally in the management of fractures and reconstruc-
tive surgeries.  These implants are intended as a guide to normal healing, and are 
NOT intended to replace normal body structure or bear the weight of the body in the 
presence of incomplete bone healing.  Delayed unions or nonunions in the presence 
of load bearing or weight bearing might eventually cause the implant to break due 
to metal fatigue.  All metal surgical implants are subjected to repeated stress in use, 
which can result in metal fatigue.
Indications: The Captured Hip Screw is for the internal fixation of hip fractures. The 
Supracondylar plate systems is intended only for fractures and osteotomies of the 
proximal and distal femur. The Trochanteric Side Plate is used in conjunction with 
the Captured Hip Screw and is designed for use in osteotomies, arthrodesis and hip 
fractures; including intertrochanteric, intracapsular and subtrochanteric
Contraindications:   Screws, plates, nails, compression hip screws, pins and 
wires are contraindicated in:  active infection; conditions which tend to retard 
healing such as blood supply limitations, previous infections; insufficient quantity 
or quality of bone to permit stabilization of the fracture complex; conditions that 
restrict the patient’s ability or willingness to follow postoperative instructions during 
the healing process; foreign body sensitivity; and cases where the implant(s) would 
cross open epiphyseal plates in skeletally immature patients. 
Additional Contraindication for Orthopaedic Screws and Plates only:  Cases with 
malignant primary or metastatic tumors which preclude adequate bone support or 
screw fixations, unless supplemental fixation or stabilization methods are utilized.
Additional Contraindications for Compression Hip Screws only: Inadequate implant 
support due to the lack of medial buttress.
Additional Contraindications for Retrograde Femoral Nailing only: A history of septic 
arthritis of the knee; knee extension contracture with inability to attain at least 45° 
of flexion.
Warnings and Precautions: In using partial weight bearing or nonweight bearing ap-
pliances (orthopaedic devices other than prostheses), a surgeon should be aware 
that no partial weight bearing or nonweight bearing device can be expected to 
withstand the unsupported stresses of full weight bearing.
Adverse Events: The following are possible adverse events after fixation with ortho-
paedic screws, plates, nails, compression hip screws, pins and wires:  loosening, 
bending, cracking or fracture of the components or loss of fixation in bone attribut-
able to nonunion, osteoporosis, markedly unstable comminuted fractures; loss of 
anatomic position with nonunion or malunion with rotation or angulation; infection, 
both deep and superficial; and allergies and other reactions to the device material.  
Additional Adverse Events for Compression Hip Screw only:  Screw cutout of the 
femoral head (usually associated with osteoporotic bone).