Thermo Scientific Orion Star A211 Benchtop and
Star A221 and Star A321 Portable pH Meters
Reference Guide
For Sales & Service Contact
2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242
1-800-833-7958
www.geotechenv.com
Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and
its subsidiaries.
AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect,
PerpHecT, pHISA, pHuture, Pure Water, Sage, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS, and
TURBO2 are registered trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.
A+, All in One, Aplus, AUTO-BAR, AUTO-CAL, Auto-ID, AUTO-READ, AUTO-STIR,
Auto-Test, AutoTration, CISA, digital LogR, DuraProbe, EZ Startup, ISEasy,
Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, Optimum Results, Orion Dual
Star, Orion Star, SAOB, SMART AVERAGING, SMART STABILITY, Star LogR, Star Navigator 21,
Stat Face, Triode are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher Scientific, Inc.
and its subsidiaries.
© 2011 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of
Thermo Fisher Scientific, Inc. and its subsidiaries.
The specifications, descriptions, drawings, ordering information and part numbers within this
document are subject to change without notice.
This publication supersedes all previous publications on this subject.
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meters
Chapter 1 Introduction ..........................................................1
Chapter 2 Meter Overview.......................................................3
Meter Connections and Inputs.......................................... 3
Power Source....................................3
Electrodes and Other Connections....................4
Connecting Electrode Arm to the Benchtop Meter........5
Meter Display ........................................................ 6
Meter Keypad......................................................... 7
Meter Maintenance ................................................... 8
Chapter 3 Meter Setup ..........................................................9
Navigating the Setup Menu ............................................ 9
Setup Menu Overview................................................. 9
pH Channel Menu................................10
Settings Menu...................................12
View Log Menu..................................13
Diagnostics Menu................................13
Chapter 4 pH, RmV and ORP Calibration ..........................................15
pH Calibration ....................................................... 15
Calibration Editing................................16
RmV Calibration ..................................................... 16
ORP Calibration ..................................................... 17
Chapter 5 Temperature Calibration ..............................................19
Chapter 6 Measurement........................................................21
Chapter 7 Methods ............................................................23
Saving the Current Meter Settings as a New Method..................... 23
Loading an Existing Method........................................... 23
Editing, Copying or Deleting an Existing Method......................... 24
Chapter 8 Data Storage and Retrieval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Data Storage Settings ................................................ 25
Read Type Selection..............................25
Computer and Printer Export Settings.................26
Data Log Settings................................26
Viewing the Calibration Log and Data Log .............................. 27
Viewing the Calibration Log.........................27
Viewing the Data Log..............................27
Exporting the Data Log to a Computer or Printer ......................... 28
Clearing the Data Log................................................. 29
Interfacing a Printer.................................................. 30
Interfacing a Computer ............................................... 30
Chapter 9 Customer Services ...................................................31
Meter Errors and Recommended Actions ............................... 31
Meter Self Test and Checkout ......................................... 32
Meter Self Test..................................32
Meter Checkout (pH)..............................32
Electrode Stability Test............................................... 33
Meter Factory Reset ................................................. 33
Meter User Reset .................................................... 33
About Meter......................................................... 34
Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Warranty and Registration ............................................ 34
Ordering Information ................................................. 35
Chapter 10 Meter Compliance....................................................39
Notice of Compliance................................................. 39
WEEE Compliance ................................................... 39
Declaration of Conformity............................................. 40
Meter Specifications................................................. 41
Appendix 1 Advanced Meter Features .............................................43
Automatic pH Buffer Recognition Feature............................... 43
Testing a pH Electrode for Automatic Buffer Recognition . 43
Electrode Condition Icon ............................................. 44
Multipoint Calibration Curve Feature .................................. 45
This page intentionally left blank.
1
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 1 Introduction
Thank you for your purchase of the Orion Star A211 benchtop pH meter, Star A221 portable pH meter or Star A321
portable pH meter. These meters are capable of measuring pH, mV, relative mV (RmV), ORP, and temperature in °C or °F.
The Orion Star A211 benchtop pH meters feature
tQ)SBOHFPGUPQ)VOJUTXJUIPSSFTPMVUJPOBOEVQUPDBMJCSBUJPOQPJOUT
tN7SFMBUJWFN7BOE031SBOHFPGUPN7XJUIN7SFTPMVUJPO
t"6503&"%DPOUJOVPVTXJUIIPMEPQUJPOBOEUJNFENFBTVSFNFOUNPEFT
tDBMJCSBUJPOFEJUJOH
tQPJOUEBUBMPH
tNBOVBMSFBEZ"6503&"%BOEUJNFEEBUBMPHHJOHGVODUJPOT
t*1SBUFEEVTUBOETQMBTISFTJTUBOUIPVTJOH
The Orion Star A221 portable pH meters feature
tQ)SBOHFPGUPQ)VOJUTXJUIPSSFTPMVUJPOBOEVQUPDBMJCSBUJPOQPJOUT
tN7SFMBUJWFN7BOE031SBOHFPGUPN7XJUIN7SFTPMVUJPO
t"6503&"%BOEDPOUJOVPVTXJUIIPMEPQUJPONFBTVSFNFOUNPEFT
tQPJOUEBUBMPH
tNBOVBMBOEBVUPNBUJDXJUI"6503&"%EBUBMPHHJOHGVODUJPOT
t*1SBUFEEVTUQSPPGBOEXBUFSQSPPGIPVTJOH
The Orion Star A321 portable pH meters feature
tQ)SBOHFPGUPQ)VOJUTXJUIPSSFTPMVUJPOBOEVQUPDBMJCSBUJPOQPJOUT
tN7SFMBUJWFN7BOE031SBOHFPGUPN7XJUIN7SFTPMVUJPO
t"6503&"%DPOUJOVPVTXJUIIPMEPQUJPOBOEUJNFENFBTVSFNFOUNPEFT
tDBMJCSBUJPOFEJUJOH
tQPJOUEBUBMPH
tNBOVBMSFBEZ"6503&"%BOEUJNFEEBUBMPHHJOHGVODUJPOT
t*1SBUFEEVTUQSPPGBOEXBUFSQSPPGIPVTJOH
1MFBTFSFBEUIJTSFGFSFODFHVJEFUIPSPVHIMZ"OZVTFPVUTJEFPGUIFTFJOTUSVDUJPOTNBZJOWBMJEBUFZPVSXBSSBOUZBOE
cause permanent damage to the meter.
2Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
This page intentionally left blank.
3
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 2 Meter Overview
Meter Connections and Inputs
Power Source
 1PXFSBEBQUFSJODMVEFEXJUICFODIUPQNFUFSTTPMETFQBSBUFMZGPSQPSUBCMFNFUFST
B 4FMFDUUIFBQQSPQSJBUFXBMMTPDLFUQMVHQMBUF
b. Slide off the clear plastic cover from the plug plate.
D 4MJEFUIFQMVHQMBUFJOUPUIFHSPPWFPOUIFCBDLPGUIFQPXFSBEBQUFS
E $POOFDUUIFQPXFSBEBQUFSUPUIFNFUFSBOEQPXFSPVUMFU
 #BUUFSJFTJODMVEFEXJUIQPSUBCMFNFUFSTTPMETFQBSBUFMZGPSCFODIUPQNFUFST
a. Select four AA alkaline batteries.
C $POmSNUIBUUIFNFUFSJTQPXFSFEPGG
c. Remove the battery compartment cover.
J #FODIUPQNFUFSToQVTIEPXOPOUIFCBUUFSZDPNQBSUNFOUUBCBOEMJGUUIFCBUUFSZDPWFSVQ
JJ1PSUBCMFNFUFSToMPPTFOUIFTDSFXTIPMEJOHUIFCBUUFSZDPWFSSFMFBTFUIFUPQQPSUJPOPGUIF
battery cover from the meter (use a coin or your finger) and release the bottom portion of the
battery cover.
E 0SJFOUBUFUIFCBUUFSJFTBTTIPXOJOUIFCBUUFSZDPNQBSUNFOUIPVTJOHBOEJOTFSUCBUUFSJFT
F 3FQMBDFUIFCBUUFSZDPNQBSUNFOUDPWFS'PSQPSUBCMFNFUFSToSFQMBDFUIFTDSFXT
1VTIEPXOPOUBCBOEMJGUUIF
battery cover
Orion Star A Series Benchtop
3FNPWFTDSFXT
Orion Star A Series Portable
4Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Electrodes and Other Connections
1. Prepare the pH electrode and any other applicable electrodes according to the directions in the
electrode user guide.
 $POOFDUUIFBQQSPQSJBUFJUFNTBTMBCFMFEPOUIFNFUFSBOEBTTIPXOJOUIFmHVSFCFMPX
t #/$o*OQVUGPSQ)FMFDUSPEFPS031SFEPYFMFDUSPEFXJUI#/$PSXBUFSQSPPG#/$DPOOFDUPS
t 3&'o*OQVUGPSSFGFSFODFFMFDUSPEFXJUITUBOEBSENNQJOUJQDPOOFDUPS
t "5$$0/o*OQVUGPS"5$QSPCFXJUIQJO.JOJ%*/DPOOFDUPS
t (SPVOEo*OQVUGPSBHSPVOEXJSFXJUIBTUBOEBSEQJOUJQDPOOFDUPSOrion Star A211 benchtop meters only.
t 1PXFSo*OQVUGPSUIFQPXFSBEBQUFS
t 4UJSSFSo*OQVUGPSUIFTUJSSFSQSPCF$BU/POrion Star A211 benchtop meters only.
t 64#o*OQVUGPSUIF64#DBCMFVTFEUPDPOOFDUUIFNFUFSUPBQSJOUFSPSDPNQVUFS
t 34o*OQVUGPSUIF34DBCMFVTFEUPDPOOFDUUIFNFUFSUPBQSJOUFSPSDPNQVUFS
Power Stirrer USB RS232
BNCATC/CONDO/RDO
Ground Ref.
DO/
RDO
ATC/
COND
POWER
USB
RS232
REF BNC
Star A211 Meter Back Panel
Star A221 and Star A321 Meter Back Panel
5
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Connecting Electrode Arm to an Orion Star A Series Benchtop Meter
Orion Star A series benchtop meters include an electrode arm that can be attached to either side of the meter.
 6OQBDLUIFFMFDUSPEFBSNBOECBTF
2. Choose the side of the meter to attach the arm.
 6TFBDMFBOTVSGBDFBOEUVSOUIFNFUFSPWFS
 3FMFBTFUIFFYJTUJOHTDSFXGSPNUIFCBDLPGUIFNFUFS
 "MJHOUIFFMFDUSPEFBSNCBTFXJUIUIFDJSDMFTBUUIFCPUUPNPGUIFNFUFS5IFNFUBMQPTUPOUIFFMFDUSPEFBSNCBTF
should be on the same side as the meter display.
 5BLFUIFTDSFXUIBUXBTSFNPWFEBOEVTFJUUPTFDVSFUIFFMFDUSPEFBSNCBTFUPUIFNFUFS
 5VSOUIFNFUFSPWFS
 1MBDFUIFIPMFBUUIFCPUUPNPGUIFFMFDUSPEFBSNPOUPUIFNFUBMQPTUPOUIFFMFDUSPEFBSNCBTF
6Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Meter Display
Display Icon Description
4IPXOXIFOUIFNFUFSJTSVOOJOHPO"$QPXFS
4IPXOXIFOUIFNFUFSIBTCBUUFSJFTJOTUBMMFE
*OEJDBUFTEBUBJTCFJOHTFOUUPBDPNQVUFSPSQSJOUFS
*OEJDBUFTEBUBJTCFJOHTFOUUPUIFEBUBMPH
4IPXOXIFOBOBMBSNJTTFUBOEUIFBMBSNWBMVFJTSFBDIFE
*OEJDBUFTUIFNFUFSJTTFUUPCFJOUFSGBDFEXJUIBQSJOUFSPSDPNQVUFSWJBUIF34QPSU
*OEJDBUFTUIFNFUFSJTTFUUPCFJOUFSGBDFEXJUIBQSJOUFSPSDPNQVUFSWJBUIF64#QPSU
%JTQMBZTUIFUJNFBOEEBUFFOUFSFEJOUIFTFUVQNFOV
25.0˚C
ATC
%JTQMBZTUIFDVSSFOUUFNQFSBUVSFCBTFEPOUIFUFNQFSBUVSFQSPCFSFBEJOHPSFOUFSFEUFNQFSBUVSFWBMVF4IPXTUIF
PSJHJOPGUIFUFNQFSBUVSFBT."/FOUFSFEUFNQFSBUVSFPS"5$UFNQFSBUVSFQSPCF
HOLD 4IPXOXIFO is pressed and the displayed measurement is frozen.
*OEJDBUFTBDBMJCSBUJPOXBTTVDDFTTGVMMZDPNQMFUFE*DPOXJMMCMJOLXIFOUIFDBMJCSBUJPOBMBSNJTTFUBOEUIF
alarm value is reached.
*OEJDBUFTUIFQ)FMFDUSPEFDPOEJUJPOBTHPPEUXPCBSTGBJSPOFCBSPSCBETMBTIUISPVHIJUCBTFEPOUIFMBTU
saved calibration and measurement stability.
M
100 *OEJDBUFTBNFUIPEJTJOVTFBOEUIFOVNCFSPGUIFNFUIPECFJOHVTFE
pH *OEJDBUFTUIFUZQFPGNFBTVSFNFOUBOEEFUFSNJOFTUIFUZQFPGDBMJCSBUJPOUIBUXJMMCFQFSGPSNFE
ready Specifies the stability of the electrode as stabilizing or ready.
4IPXOXIFOUIFNFUFSJTJO"6503&"%NPEF5IF JDPOXJMMCMJOLXIJMFUIFSFBEJOHJTTUBCJMJ[JOHBOETUPQ
CMJOLJOHXIFOUIFSFBEJOHJTTUBCMFBOEUIFNFBTVSFNFOUJTMPDLFEPOUIFEJTQMBZ
7.000 pH %JTQMBZTUIFNFBTVSFNFOUWBMVFCBTFEPOUIFMBTUTBWFEDBMJCSBUJPOBOEDVSSFOUFMFDUSPEFSFBEJOH6OJUTBSF
TIPXOUPUIFSJHIUPGUIFWBMVF
0.0mV 4IPXTUIFSBXNJMMJWPMUSFBEJOHPGUIFFMFDUSPEF
BUFFERS: 1.68,
4IPXTUIFCVGGFSWBMVFTVTFEGPSUIFMBTUTBWFEDBMJCSBUJPO
4IPXTUIFPQFSBUPSBTTJHOFETBNQMF*%OVNCFSOrion Star A211 and Star A321 meters only.
4IPXTUIFPQFSBUPSBTTJHOFEVTFS*%OVNCFSOrion Star A211 and Star A321 meters only.
cal %JTQMBZTUIFBDUJPOUIBUXJMMCFQFSGPSNFEXIFOf1 is pressed.
sample ID %JTQMBZTUIFBDUJPOUIBUXJMMCFQFSGPSNFEXIFOf2 is pressed.
channel %JTQMBZTUIFBDUJPOUIBUXJMMCFQFSGPSNFEXIFOf3 is pressed.
7
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Meter Keypad
Press the f1, f2 and f3GVODUJPOLFZTUPQFSGPSNUIFBDUJPOTIPXOBCPWFFBDILFZPOUIFEJTQMBZ
Press to turn the meter on.
8IFOUIFNFUFSJTPOQSFTTBOERVJDLMZSFMFBTFUPUVSOUIFEJTQMBZCBDLMJHIUPOPSPGGPSIPMEEPXOUP
turn the meter off.
*OUIFNFBTVSFNFOUNPEFQSFTTUPUBLFBNFBTVSFNFOU
*OUIFTFUVQDBMJCSBUJPOBOEPUIFSNFOVTQSFTTUPFTDBQFUIFDVSSFOUNFOVBOESFUVSOUPUIF
measurement mode.
*OUIFNFBTVSFNFOUNPEFQSFTTUPFOUFSUIFTFUVQNFOV
*OUIFTFUVQDBMJCSBUJPOBOEPUIFSNFOVTQSFTTUPTDSPMMVQUISPVHIBMJTUPGPQUJPOT
*OUIFDPOUJOVPVTNFBTVSFNFOUNPEFQSFTTUPIPMEGSFF[FUIFEJTQMBZFENFBTVSFNFOUBOEQSFTT
again to release (unfreeze) the measurement.
*OUIFTFUVQDBMJCSBUJPOBOEPUIFSNFOVTQSFTTUPTDSPMMMFGUUISPVHIBMJTUPGPQUJPOT
*OUIFNFBTVSFNFOUNPEFQSFTTUPDIBOHFUIFEJTQMBZFENFBTVSFNFOUNPEF0QUJPOTBSFQ)N73N7
(relative mV) and ORP.
*OUIFTFUVQDBMJCSBUJPOBOEPUIFSNFOVTQSFTTUPTDSPMMSJHIUUISPVHIBMJTUPGPQUJPOT
*OUIFNFBTVSFNFOUNPEFQSFTTUPMPHPSQSJOUBNFBTVSFNFOU
*OUIFTFUVQDBMJCSBUJPOBOEPUIFSNFOVTQSFTTUPTDSPMMEPXOUISPVHIBMJTUPGPQUJPOT
1SFTTUPWJFXUIFEBUBMPHBOEDBMJCSBUJPOMPHOrion Star A211 benchtop meters only.
Press to start or stop the stirrer probe. Orion Star A211 benchtop meters only.
log/print
mode
setup
hold
measure
(esc)
Orion Star A200
and A300 Series
Portable Keypad
Orion Star A200 Series
Benchtop Keypad
8Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
1. Press UPQPXFSUIFNFUFSPO8IFOUIFNFUFSJTPOQSFTTBOERVJDLMZSFMFBTF to turn the backlight
on or off or press and hold GPSBCPVUUISFFTFDPOETUPQPXFSUIFNFUFSPGG
2. Press UPFYJUBOZNFUFSGVODUJPOBOESFUVSOUPUIFNFBTVSFNFOUNPEF
3. The f1, f2, and f3GVODUJPOLFZTQFSGPSNBWBSJFUZPGNFUFSPQFSBUJPOT5IFNFOVTQFDJmDPQFSBUJPOJTTIPXO
BCPWFFBDILFZPOUIFEJTQMBZ'PSFYBNQMFQSFTTf1 in the measurement mode to start a calibration.
 5IF , , or LFZTBSFVTFEBTOBWJHBUJPOLFZTVQSJHIUEPXOMFGUXIFOTFMFDUJOHGSPNBmYFE
MJTUPSHSJEPGNFUFSPQUJPOT*OUIFNFBTVSFNFOUNPEFUIFTFLFZTBSFVTFEUPBDDFTTUIFTFUVQNFOVDIBOHF
the measurement mode, manually log or print a measurement and hold (freeze) a displayed measurement in the
continuous measurement mode.
 *OUIFDPOUJOVPVTNFBTVSFNFOUNPEFXIFO is pressed and the measurement is being held (frozen), only the
LFZXJMMCFGVODUJPOBMVOUJMUIFNFBTVSFNFOUJTSFMFBTFEVOGSP[FO
 1SFTT UPUVSOPOPSPGGUIFTUJSSFSQSPCF$BU/POrion Star A211 benchtop meters only.
 1SFTT to access the calibration log and data log. Orion Star A211 benchtop meters only.
Meter Maintenance
'PSSPVUJOFNFUFSNBJOUFOBODFEVTUBOEXJQFUIFNFUFSXJUIBEBNQDMPUI*GOFDFTTBSZXBSNXBUFSPSBNJME
XBUFSCBTFEEFUFSHFOUDBOCFVTFE.FUFSNBJOUFOBODFDBOCFQFSGPSNFEPOBEBJMZXFFLMZPSNPOUIMZCBTJTBT
SFRVJSFECZUIFFOWJSPONFOUJOXIJDIUIFNFUFSJTPQFSBUFE*NNFEJBUFMZSFNPWFBOZTQJMMFETVCTUBODFGSPNUIF
meter using the proper cleaning procedure for the type of spill.
9
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 3 Meter Setup
Navigating the Setup Menu
 *OUIFNFBTVSFNFOUNPEFQSFTT to enter the main setup menu.
2. Press , , or to scroll through the main setup menu options. Press f3 (select) to
select a main setup menu option.
3. Press or to scroll through setup submenu options. Press f3 (select) to select a setup submenu option.
 1FSGPSNUIFBQQSPQSJBUFBDUJPOTUPTFUUIFEFTJSFEQBSBNFUFSJOUIFTFUVQTVCNFOVT
a. To select a value from a list of options, press or to highlight the desired value and press
f3 (select) to set the value.
b. To enter a numeric value, use the numeric entry screen.
i. Select the value to be entered by pressing f3 (select) or f3 (edit)5IFOVNFSJDFOUSZTDSFFOXJMMQPQVQPOUIFEJTQMBZ
ii. Press , , or to highlight a number, decimal place or negative sign; press
f3 (enter)UPTFMFDUUIFIJHIMJHIUFEJUFNBOESFQFBUVOUJMUIFEFTJSFEWBMVFJTTIPXOPOUIFUPQPGUIFOVNFSJD
entry screen.
iii. Press f2 (done)UPTBWFUIFWBMVFBOEFYJUUIFOVNFSJDFOUSZTDSFFO
 1SFTTf1 (back) and then to return to the measurement mode at any time.
Setup Menu Overview
pH Channel Settings Log View Diagnostics
Method (Star A211 and Star A321 meters only)
Mode & Settings
t.FBTVSF.PEF
t3FBE5ZQF
t3FTPMVUJPO
t#VGGFS(SPVQ
t4UBCJMJUZ(Star A211 and Star A321 meters only)
t"WFSBHJOH(Star A211 and Star A321 meters only)
t-JNJU"MBSN
t$BM%VF"MBSN
Temperature
t.BOVBM5FNQ7BMVF
t5FNQFSBUVSF6OJU
t5FNQFSBUVSFCalibration
t&YQPSU%BUB
t%BUB-PH
t%BUF5JNF
t-BOHVBHF
t,FZ1SFTT#FFQ
t"MBSN#FFQ
t4UJSSFS(Star A211 meters only)
t$POUSBTU
t"VUP4IVU0GG
t6TFS*%(Star A211 and Star A321 meters only)
t4BNQMF*%(Star A211 and Star A321 meters only)
t%BUB-PH
t$BMJCSBUJPO-PH
t'BDUPSZ3FTFU
t6TFS3FTFU
t4FMG5FTU
t4UBCJMJUZ5FTU
t"CPVU.FUFS
10 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
pH Channel Menu
5IFQ)$IBOOFMTFUVQNFOVJTVTFEUPSFWJFXBOEDIBOHFQBSBNFUFSTSFMBUJOHUPQ)N7SFMBUJWFN7BOE031
measurements and calibrations.
Main Setup Menu 2nd Level 3rd Level Menu Options Menu Description
pH Channel
Method List of Methods
.FOVBMMPXTUIFDVSSFOUNFUFSTFUUJOHTUPCFTBWFEBTBOFXNFUIPE
BOFYJTUJOHNFUIPEUPCFSVOPOUIFNFUFSPSBOFYJTUJOHNFUIPEUPCF
WJFXFEFEJUFEDPQJFEEFMFUFEPSQSJOUFE
Star A211 and Star A321 meters only.
Mode and Settings
Measure Mode
tQ)
tN7
t3N7
t031
.FOVVTFEUPTFMFDUUIFEJTQMBZFENFBTVSFNFOUNPEFXIJDI
EFUFSNJOFTUIFUZQFPGDBMJCSBUJPOUIFNFUFSXJMMQFSGPSN
Read Type
t"VUP
t$POUJOVPVT
t5JNFE
(Star A211 and Star
A321 meters only)
.FOVVTFEUPTFMFDUIPXUIFNFUFSXJMMUBLFNFBTVSFNFOUT
Auto ( )– Press to take a measurement. When the
measurement is stable, it is locked on the display until is
pressed again.
Continuous The measurement is constantly updated on the
display and stabilizing or readyJTTIPXOUPJOEJDBUFUIF
measurement stability.
Timed – Measurements are taken at a set time interval. To enter
the time interval, highlight tTimed press to highlight the
hours (HH), minutes (MM) or seconds (SS), press f3 (edit) and use
the numeric entry screen to change the values.
Resolution
t%FDJNBM1MBDF

t%FDJNBM1MBDFT

t%FDJNBM1MBDFT

Menu used to select the resolution for the displayed measurement values.
Buffer Group t64"
t%*/
Menu used to select the buffer group used for automatic buffer
recognition during pH calibrations.
64"oQ)CVGGFST
%*/oQ)CVGGFST
11
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
pH Channel Menu (cont.)
Main Setup Menu 2nd Level 3rd Level Menu Options Menu Description
pH Channel
(cont.)
Mode and Settings
(cont.)
Stability
t4NBSU4UBCJMJUZ
t'BTU
t.FEJVN
t4MPX
.FOVVTFEUPTFMFDUIPXUIFTUBCJMJUZPGUIFEJTQMBZFENFBTVSFNFOU
values is indicated. Smart Stability automatically compensates for
measurement conditions and optimizes the meter response time.
Star A211 and Star A321 meters only.
Averaging t0GG
t"VUPNBUJD4NBSU
Menu used to turn the averaging function on or off. Automatic Smart
automatically compensates for measurement conditions and optimizes
the meter response time.
Star A211 and Star A321 meters only.
-JNJU"MBSN
t-JNJU"MBSN
On
Off
t"MBSN4FUUJOHT
-PX
High
)JHI-PX
t)JHI-JNJU
t-PX-JNJU
Menu used to turn the limit alarm on or off and to set the limit alarm
parameters. The limit alarm is triggered if the measurement goes above
BIJHIWBMVFBOEPSCFMPXBMPXWBMVFTFUCZUIFPQFSBUPS1SFTT or
to change the setting on each line, press or to select
another line and use the numeric entry screen to enter limit values.
Star A211 and Star A321 meters only.
$BM%VF"MBSN
t$BM%VF"MBSN
On
Off
t$BM%VF-JNJU
Menu used to turn the calibration due alarm on or off and to set the
calibration due alarm time. The calibration due alarm is triggered if a set
BNPVOUPGUJNFIBTQBTTFEXJUIPVUBDBMJCSBUJPOCFJOHQFSGPSNFE1SFTT
or to change the setting on the top line, or to
select another line and use the numeric entry screen to enter the
calibration due value in hours.
Star A211 and Star A321 meters only.
Star A221 meters have the cal due alarm with a fixed value of 12 hours.
Temperature Manual Temp Value %FGBVMUWBMVFPSMBTU
entered value.
.FOVVTFEUPTFUUIFUFNQFSBUVSFWBMVFUIBUXJMMCFVTFEGPSBMM
NFBTVSFNFOUTBOEDBMJCSBUJPOT6TFUIFOVNFSJDFOUSZTDSFFOUPFOUFS
the temperature.
5FNQFSBUVSF6OJU t$FMTJVT
t'BISFOIFJU Menu used to select the temperature units.
Temperature Calibration t"5$ .FOVBMMPXTUIFUFNQFSBUVSFNFBTVSFECZBO"5$QSPCFTFQBSBUFPS
built in to an electrode) to be adjusted by the operator.
5FNQFSBUVSF*OQVU t"5$
t.BOVBM Menu used to select the source of the temperature measurement.
12 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Settings Menu
5IF4FUUJOHTTFUVQNFOVJTVTFEUPSFWJFXBOEDIBOHFQBSBNFUFSTSFMBUJOHUPUIFHFOFSBMPQFSBUJPOPGUIFNFUFS
Main Setup Menu 2nd Level Menu Options Menu Description
Settings
&YQPSU%BUB
t1SJOUJOH
t0O
t0GG
t$PNN4FUVQ
t34
t64#
t%BUB'PSNBU
t1SJOUFS
t1$$47
t$PNN$POGJH
t
t
t
t
t
t
.FOVVTFEUPUVSOEBUBFYQPSUJOHPOPSPGGTFUUIFNFUFSUPCFJOUFSGBDFEVTJOHUIF
34PS64#QPSUTFUUIFFYQPSUFEEBUBGPSNBUGPSBQSJOUFSTUSJOHPGUFYUPS1$
DPNNBEFMJNJUFEUFYUBOETFUUIFCBVESBUF
%BUB-PH t0GG
t0O Menu used to turn on or off the data log feature.
%BUF5JNF t%BUF
t%%..::
t..%%::
t5JNF
t)PVS$MPDL
t)PVS$MPDL
.FOVVTFEUPTFUUIFEBUFGPSNBUBTEBZNPOUIZFBSPSNPOUIEBZZFBSFOUFS
UIFEBUFWBMVFTTFUUIFUJNFGPSNBUBTIPVSPSIPVSDMPDLBOEFOUFSUIFUJNF
WBMVFT6TFUIFOVNFSJDFOUSZTDSFFOUPFOUFSUIFWBMVFT
-BOHVBHF t&OHMJTI
t4QBOJTI
t(FSNBO
t'SFODI
t*UBMJBO
t$IJOFTF
Menu used to set the language for all displayed meter prompts.
,FZ1SFTT#FFQ t0GG
t0O Menu used to turn on or off the beep sound made every time a meter key is pressed.
Alarm Beep t0GG
t0O .FOVVTFEUPUVSOPOPSPGGUIFCFFQTPVOENBEFXIFOBOBMBSNWBMVFJTSFBDIFE
13
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Settings Menu (cont.)
Main Setup Menu 2nd Level Menu Options Menu Description
Settings (cont.)
Stirrer t4QFFE
t4QFFE
t4QFFE
t4QFFE
t4QFFE
Menu used to turn the stirrer probe on or off and to set the stirrer probe speed.
Star A211 meters only.
Contrast t-FWFM
t-FWFM
t-FWFM
t-FWFM
t-FWFM
t-FWFM
Menu used to set the display contrast.
Auto Shut Off t0GG
t0O
.FOVVTFEUPUVSOPOPSPGGUIFNFUFSBVUPNBUJDTIVUPGGGFBUVSFXIJDIUVSOTUIF
NFUFSPGGXIFOOPLFZTBSFQSFTTFEGPSNJOVUFT
6TFS*% %FGBVMUWBMVFPSMBTU
entered value.
.FOVVTFEUPTFUUIFVTFS*%6TFUIFBMQIBOVNFSJDFOUSZTDSFFOUPFOUFSUIFVTFS*%
Star A211 and Star A321 meters only.
4BNQMF*% t0GG
t.BOVBM
t"VUP*ODSFNFOUBM
.FOVVTFEUPUVSOUIFTBNQMF*%PGGPSPO8IFOTBNQMF*%JTTFUUPNBOVBMPSBVUP
JODSFNFOUBMFOUFSUIFTBNQMF*%JOUIFNFBTVSFNFOUNPEFCZQSFTTJOHf2 (sample
ID) and using the alphanumeric entry screen.
Star A211 and Star A321 meters only.
View Log Menu
5IF7JFX-PHTFUVQNFOVJTVTFEUPBDDFTTJOGPSNBUJPOTUPSFEJOUIFEBUBMPHBOEDBMJCSBUJPOMPH3FGFSUPUIF%BUB
Storage and Retrieval chapter for detailed information on using this setup menu.
Diagnostics Menu
5IF%JBHOPTUJDTTFUVQNFOVJTVTFEUPQFSGPSNUIFNFUFSTFMGUFTUSFTFUUIFNFUFSUPUIFGBDUPSZEFGBVMUTFUUJOHTBOE
EJTQMBZNFUFSJOGPSNBUJPOJODMVEJOHTFSJBMOVNCFSBOETPGUXBSFSFWJTJPO3FGFSUPUIF$VTUPNFS4FSWJDFTDIBQUFSGPS
detailed information on using this setup menu.
14 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
This page intentionally left blank.
15
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 4 pH, RmV and ORP Calibration
5IF0SJPO4UBS"BOE4UBS"TFSJFTQ)NFUFSTDBOQFSGPSNBPOFUPmWFQPJOUQ)DBMJCSBUJPOPOFQPJOUSFMBUJWF
N7DBMJCSBUJPOBOEPOFQPJOU031DBMJCSBUJPO4UFQCZTUFQJOTUSVDUJPOTBOEQSPNQUTBSFEJTQMBZFEPOUIFNFUFSUPBTTJTU
PQFSBUPSTXJUIUIFDBMJCSBUJPONFOVT5IFNFUFSDBOTBWFVQUPDBMJCSBUJPOTJOUIFMPHBOEDBMJCSBUJPOEBUBDBOCF
FYQPSUFEGSPNUIFNFUFSUPBQSJOUFSPSDPNQVUFS
pH Calibration
0OFUPmWFQ)CVGGFSTDBOCFVTFEGPSDBMJCSBUJPO"MXBZTVTFGSFTIQ)CVGGFSTBOETFMFDUCVGGFSTUIBUCSBDLFUUIF
sample pH and are one to four pH units apart. Prepare the pH electrode according to the instructions in the electrode use
guide. Connect the pH electrode and any other electrodes to be used (stirrer probe, ATC probe, reference electrode) to the
NFUFS1PXFSPOUIFNFUFSBOETFUUIFNFBTVSFNFOUNPEFUPQ)
Note: In most calibration screens, press f1 (esc) to return to the measurement mode without saving the calibration.
 *OUIFNFBTVSFNFOUNPEFQSFTTf1 (cal).
 3JOTFUIFQ)FMFDUSPEFBOEBOZPUIFSFMFDUSPEFTJOVTFXJUIEJTUJMMFEXBUFSCMPUESZXJUIBMJOUGSFFUJTTVFBOEQMBDF
into the pH buffer.
3. When the electrode and buffer are ready, press f3 (start)*GVTJOHB4UBS"CFODIUPQNFUFSBOETUJSSFSQSPCFUIF
TUJSSFSQSPCFXJMMBVUPNBUJDBMMZTUBSUTUJSSJOHBUUIFTQFFETFMFDUFEJOUIFTFUVQNFOVPODFf3 (start) is pressed and stop
stirring once the reading stabilizes.
 8BJUGPSUIFQ)WBMVFPOUIFNFUFSUPTUBCJMJ[FBOETUPQnBTIJOHBOEQFSGPSNPOFPGUIFGPMMPXJOHBDUJPOT
a. Press f2 (accept) to accept the displayed value.
b. Press f3 (edit) to access the numeric entry screen and edit the value.
i. Press , , or to highlight a number, decimal point or negative sign; press
f3 (enter) to select the highlighted item and repeat until the value at the measured temperature is
TIPXOBCPWFUIFOVNFSJDFOUSZTDSFFO
ii. Press f2 (done)UPFYJUUIFOVNFSJDFOUSZTDSFFO
iii. P r es s f2 (accept) to accept the entered value.
 1SFTTf2 (next) UPQSPDFFEUPUIFOFYUCVGGFSBOESFQFBUTUFQTUISPVHIPSpress f3 (cal done) to save and end
UIFDBMJCSBUJPO*GmWFCVGGFSTBSFVTFEUIFDBMJCSBUJPOXJMMTBWFBOEFOEPODFUIFmGUIWBMVFJTBDDFQUFE
B *GBPOFQPJOUDBMJCSBUJPOJTQFSGPSNFEQSFTTf2 (accept) to accept the displayed slope value or press
f3 (edit) to access the numeric entry screen, enter the slope value and press f2 (accept).
 5IFNFUFSXJMMEJTQMBZUIFDBMJCSBUJPOTVNNBSZJODMVEJOHUIFBWFSBHFTMPQFBOEFYQPSUUIFEBUBUPUIFDBMJCSBUJPOMPH1SFTT
f1 (meas) to proceed to the measurement mode or press f2 (print)UPQSPDFFEUPUIFNFBTVSFNFOUNPEFBOEFYQPSUUIFEBUB
to a printer or computer.
16 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Calibration Editing
Star A211 and Star 321 Meters Only
 *OUIFDBMJCSBUJPOTVNNBSZEJTQMBZQ)DBMJCSBUJPOTUFQpress f3 (cal edit) to edit the calibration points.
2. Press or to highlight the calibration point to be edited and press f2 (select).
3. Press or to highlight Remeasure, Edit or Delete and press f2 (select).
B *G RemeasureJTTFMFDUFESFQFBUQ)DBMJCSBUJPOTUFQTUISPVHI
C *GEdit is selected, press f3 (edit)UPBDDFTTUIFOVNFSJDFOUSZTDSFFOFOUFSUIFOFXWBMVFQSFTTf2
(done) and then press f2 (accept). Select another calibration point to edit or press f1 (back).
D *GDeleteJTTFMFDUFEUIFDBMJCSBUJPOQPJOUXJMMCFEFMFUFE*GEFMFUJOHUIFDBMJCSBUJPOQPJOUDIBOHFTUIF
calibration to a one point calibration, enter the slope using the numeric entry screen.
 5IFNFUFSXJMMEJTQMBZBOVQEBUFEDBMJCSBUJPOTVNNBSZBOEFYQPSUUIFEBUBUPUIFDBMJCSBUJPOMPH1SFTTf1 (meas)
to proceed to the measurement mode, press f2 (print)UPQSPDFFEUPUIFNFBTVSFNFOUNPEFBOEFYQPSUUIFEBUBUPB
printer or computer or press f3 (cal edit) to edit another calibration point.
RmV Calibration
0OFTUBOEBSEDBOCFVTFEGPSBSFMBUJWFN7DBMJCSBUJPO"MXBZTVTFGSFTITUBOEBSE1SFQBSFUIFFMFDUSPEFBDDPSEJOHUPUIF
instructions in the electrode use guide. Connect the electrode and any other electrodes to be used (stirrer probe, ATC probe,
SFGFSFODFFMFDUSPEFUPUIFNFUFS1PXFSPOUIFNFUFSBOETFUUIFNFBTVSFNFOUNPEFUP3N7
Note: In most calibration screens, press f1 (esc) to return to the measurement mode without saving the calibration.
 *OUIFNFBTVSFNFOUNPEFQSFTTf1 (cal).
 3JOTFUIFFMFDUSPEFBOEBOZPUIFSFMFDUSPEFTJOVTFXJUIEJTUJMMFEXBUFSCMPUESZXJUIBMJOUGSFFUJTTVFBOEQMBDFJOUP
the standard.
3. When the electrode and standard are ready, press f3 (start)*GVTJOHB4UBS"CFODIUPQNFUFSBOETUJSSFSQSPCFUIF
TUJSSFSQSPCFXJMMBVUPNBUJDBMMZTUBSUTUJSSJOHBUUIFTQFFETFMFDUFEJOUIFTFUVQNFOVPODF
f3 (start) is pressed and stop stirring once the reading stabilizes.
 8BJUGPSUIFN7WBMVFPOUIFNFUFSUPTUBCJMJ[FJOEJDBUFECZUIFnBTIJOHstabilizing or solid ready icon, and perform
POFPGUIFGPMMPXJOHBDUJPOT
a. Press f2 (accept)UPBDDFQUUIFN7WBMVFTIPXOPOUIFRel Value line.
b. Press f3 (edit)UPBDDFTTUIFOVNFSJDFOUSZTDSFFOBOEFEJUUIFN7WBMVFTIPXOPOUIFRel Value line.
i. Press , , or to highlight a number, decimal point or negative sign; press f3 (enter) to select the
IJHIMJHIUFEJUFNBOESFQFBUVOUJMUIFWBMVFBUUIFNFBTVSFEUFNQFSBUVSFJTTIPXO
ii. Press f2 (done)UPFYJUUIFOVNFSJDFOUSZTDSFFO
iii. P r es s f2 (accept) to accept the entered mV value.
 5IFNFUFSXJMMEJTQMBZUIFDBMJCSBUJPOTVNNBSZJODMVEJOHUIFN7PGGTFUBOEFYQPSUUIFEBUBUPUIFDBMJCSBUJPOMPH1SFTT
f1 (meas) to proceed to the measurement mode or press f2 (print)UPQSPDFFEUPUIFNFBTVSFNFOUNPEFBOEFYQPSU
the data to a printer or computer.
17
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
ORP Calibration
The mV reading of an ORP standard is dependent on the ORP electrode reference system, reference material and filling
TPMVUJPOBTXFMMBTUIF031TUBOEBSEUFNQFSBUVSF"O031FMFDUSPEFDBOCFBVUPNBUJDBMMZDBMJCSBUFEUPSFBEN7
SFMBUJWFUPUIF4UBOEBSE)ZESPHFO&MFDUSPEF&HBU¡$XIFOUIF0SJPO031TUBOEBSE$BU/PPS0SJPO
031FMFDUSPEF$BU/P#/81#/81#/.%PS#/.%BOE0SJPO.,$MmMMJOHTPMVUJPO$BU/P
PS.,$MHFMJO$BU/P#/.%BSFVTFE5IFSBXN7SFBEJOHPGBOFMFDUSPEFXJUIUIFTBNFDPOEJUJPOTJT
N7BU¡$
5IF5IFSNP4DJFOUJmD0SJPO031TUBOEBSEJTOPOIB[BSEPVTTUBCMFBOEXJMMOPUDIBOHFPWFSUJNFTPJUJTBOJEFBM
TUBOEBSEGPSBOZ031NFBTVSFNFOUTZTUFN5IF0SJPO4UBS"BOE4UBS"TFSJFTQ)NFUFSTPGGFSBTJNQMF
BVUPNBUJDDBMJCSBUJPOUPUIF&HWBMVFJOUIF031NPEFXIFOVTJOH0SJPO031FMFDUSPEFTBOETUBOEBSE7JTJU
XXXUIFSNPTDJFOUJmDDPNXBUFSGPS031BQQMJDBUJPOOPUFTUIBUJODMVEFBUBCMFPG031TUBOEBSEWBMVFTBUEJGGFSFOU
temperatures.
0OFTUBOEBSEDBOCFVTFEGPS031DBMJCSBUJPO"MXBZTVTFGSFTI031TUBOEBSE1SFQBSFUIF031FMFDUSPEFBDDPSEJOHUP
the instructions in the electrode use guide. Connect the electrode and any other electrodes to be used (stirrer probe, ATC
QSPCFSFGFSFODFFMFDUSPEFUPUIFNFUFS1PXFSPOUIFNFUFSBOETFUUIFNFBTVSFNFOUNPEFUP031
Note: In most calibration screens, press f1 (esc) to return to the measurement mode without saving the calibration.
 *OUIFNFBTVSFNFOUNPEFQSFTTf1 (cal).
 3JOTFUIF031FMFDUSPEFBOEBOZPUIFSFMFDUSPEFTJOVTFXJUIEJTUJMMFEXBUFSCMPUESZXJUIBMJOUGSFFUJTTVFBOEQMBDF
into the standard.
3. When the electrode and standard are ready, press f3 (start)*GVTJOHB4UBS"CFODIUPQNFUFSBOETUJSSFSQSPCF
UIFTUJSSFSQSPCFXJMMBVUPNBUJDBMMZTUBSUTUJSSJOHBUUIFTQFFETFMFDUFEJOUIFTFUVQNFOVPODF
f3 (start) is pressed and stop stirring once the reading stabilizes.
 8BJUGPSUIFN7WBMVFPOUIFNFUFSUPTUBCJMJ[FJOEJDBUFECZUIFnBTIJOHstabilizing or solid ready icon, and press f2
(accept) once the value is stable.
 5IFNFUFSXJMMEJTQMBZUIFDBMJCSBUJPOTVNNBSZJODMVEJOHUIFN7PGGTFUBOEFYQPSUUIFEBUBUPUIFDBMJCSBUJPOMPH
Press f1 (meas) to proceed to the measurement mode or press f2 (print) to proceed to the measurement mode and
FYQPSUUIFEBUBUPBQSJOUFSPSDPNQVUFS
18 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
This page intentionally left blank.
19
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 5 Temperature Calibration
5IFNFUFS"5$UFNQFSBUVSFEJTQMBZIBTBSFMBUJWFBDDVSBDZPG¡$"5$QSPCFTCPUITFQBSBUFBOEUIPTFCVJMUJOUP
FMFDUSPEFTIBWFWBSZJOHUFNQFSBUVSFBDDVSBDJFTVTVBMMZ¡$UP¡$6TFUIJTGVODUJPOPOMZJGJUJTOFDFTTBSZ
to calibrate the temperature readings taken by the ATC probe. Since the temperature offset calculated during the
calibration is applied to all future temperature measurements, recalibrate if a different ATC probe is used.
Temperature calibration is for the calibration of an ATC probe (separate or built in to a pH electrode or ORP electrode.)
*GNBOVBMUFNQFSBUVSFDPNQFOTBUJPOJTCFJOHVTFEBOEUIFEJTQMBZTIPXT."/BTUIFUFNQFSBUVSFTPVSDFJDPO
set the manual temperature value using the Manual Temp Value menu. The temperature calibration menu is not
accessible unless an ATC probe is connected to the meter.
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight pH Channel and press f3 (select).
3. Press or to highlight Temperature and press f3 (select).
 1SFTT or to highlight Temperature Calibration and press f3 (select).
 1SFTT or to highlight ATC UPJEFOUJGZUIF"5$QSPCFUIBUXJMMCFDBMJCSBUFEBOEQSFTTf3 (select).
 1MBDFUIF"5$QSPCFBOEUIFSNPNFUFSTJOUPBTPMVUJPOXJUIBLOPXOTUBCMFUFNQFSBUVSF*UJTSFDPNNFOEFEUIBUUXP/*45
traceable thermometers be used to measure and verify the temperature of the solution.
 8BJUGPSUIFUFNQFSBUVSFSFBEJOHTPGUIF"5$QSPCFBOEUIFSNPNFUFSTUPTUBCJMJ[FVTVBMMZUXPUPmWFNJOVUFT
 1SFTTf3 (edit) to access the numeric entry screen and edit the temperature value.
a. Press , , or to highlight a number, decimal point or negative sign; press f3 (enter)
to select the highlighted item and repeat until the temperature value read by the thermometers is
TIPXOPOUIFEJTQMBZ
b. Press f2 (done) UPFYJUUIFOVNFSJDFOUSZTDSFFO
 5IFNFUFSXJMMEJTQMBZUIFFOUFSFEUFNQFSBUVSFWBMVFPOUIFReference Temp line, the original temperature read
by the ATC probe on the ATC MJOFBOEUIFUFNQFSBUVSFPGGTFUWBMVFUIBUXJMMCFBQQMJFEUPBMMGVUVSFUFNQFSBUVSF
readings on the Offset line. Press f2 (accept)UPQSPDFFEXJUIUIFEJTQMBZFEUFNQFSBUVSFEBUBTBWFBOEFOEUIF
DBMJCSBUJPOBOEFYQPSUUIFEBUBUPUIFDBMJCSBUJPOMPH
1SFTT f1 (meas) to proceed to the measurement mode or press f2 (print) to proceed to the measurement mode
BOEFYQPSUUIFEBUBUPBQSJOUFSPSDPNQVUFS
20 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
This page intentionally left blank.
21
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 6 Measurement
5IF0SJPO4UBS"4UBS"BOE4UBS"NFUFSTBSFDBQBCMFPGTIPXJOHQ)N73N7SFMBUJWFN7PS031
measurements on the display.
Press XIJMFUBLJOHBNFBTVSFNFOUJOUIFDPOUJOVPVTNFBTVSFNFOUNPEFUPGSFF[FUIFEJTQMBZBOEQSFTT
a second time to unfreeze the display and continue the measurement. Press XIJMFUBLJOHBNFBTVSFNFOUUP
NBOVBMMZFYQPSUUIFNFBTVSFNFOUUPUIFEBUBMPHJGUIFEBUBMPHJTFOBCMFEJOUIFTFUVQNFOV
 3JOTFUIFQ)FMFDUSPEFBOEBOZPUIFSFMFDUSPEFTJOVTFXJUIEJTUJMMFEXBUFSCMPUESZXJUIBMJOUGSFFUJTTVFBOEQMBDF
into the sample.
2. Star A211 meters onlyo*GUIFTUJSSFSQSPCFJTJOVTFBOEUIFNFUFSJTJODPOUJOVPVTPSUJNFENPEFQSFTT
UPUVSOPOUIFTUJSSFSQSPCF*GUIFTUJSSFSQSPCFJTJOVTFBOEUIFNFUFSJTJO"6503&"%NPEFUIFTUJSSFSQSPCF
XJMMTUBSUTUJSSJOHXIFO JTQSFTTFEBOETUPQTUJSSJOHXIFOUIFSFBEJOHTUBCJMJ[FT
 4UBSUUIFNFBTVSFNFOUBOEXBJUGPSJUUPTUBCJMJ[F
B *GUIFNFUFSJTJO"6503&"%NPEFEFGBVMUTFUUJOHQSFTT to start the measurement. When
the JDPOTUPQTnBTIJOHSFDPSEUIFBQQMJDBCMFNFBTVSFNFOUQBSBNFUFSTBOEUFNQFSBUVSFPGUIF
sample. Press BHBJOUPTUBSUBOFXNFBTVSFNFOU
C *GUIFNFUFSJTJODPOUJOVPVTNPEFUIFNFUFSXJMMJNNFEJBUFMZTUBSUUBLJOHBNFBTVSFNFOUBOE
VQEBUFUIFEJTQMBZXIFOFWFSUIFNFBTVSFNFOUDIBOHFT8BJUGPSUIFEJTQMBZUPTIPXready and
record the applicable measurement parameters and temperature of the sample.
c. Star A211 and Star A321 meters only*GUIFNFUFSJTJOUJNFENPEFUIFNFUFSXJMMMPHNFBTVSFNFOUT
BUUIFQSFTFMFDUFEUJNFJOUFSWBMSFHBSEMFTTPGUIFNFBTVSFNFOUTUBCJMJUZ5IFNFUFSXJMMVQEBUF
UIFEJTQMBZXIFOFWFSUIFNFBTVSFNFOUDIBOHFTTPUIFBQQMJDBCMFNFBTVSFNFOUQBSBNFUFSTBOE
UFNQFSBUVSFPGUIFTBNQMFDBOCFSFDPSEFEXIFOUIFEJTQMBZTIPXTready.
 Star A211 meters onlyo*GUIFTUJSSFSQSPCFJTJOVTFBOEUIFNFUFSJTJODPOUJOVPVTPSUJNFENPEFQSFTT
to turn off the stirrer probe.
 3FNPWFUIFFMFDUSPEFTGSPNUIFTBNQMFSJOTFXJUIEJTUJMMFEXBUFSCMPUESZBOEQMBDFJOUPUIFOFYUTBNQMF
 3FQFBUTUFQTUISPVHIGPSBMMTBNQMFT
 8IFOBMMTBNQMFTIBWFCFFONFBTVSFETUPSFUIFFMFDUSPEFTBDDPSEJOHUPUIFJSVTFSHVJEFT
22 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
This page intentionally left blank.
23
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 7 Methods
The Orion Star A211 and Star A321 meters can save up to ten methods per channel and the corresponding calibration
data for each method, for the fast and easy recollection of different meter setup parameters. The default meter
TFUUJOHJTUIBUOPNFUIPEJTVTFEBOEUIFNFUIPEOVNCFSXJMMCFTIPXOBT.JGOPNFUIPEJTJOVTF5IFDVSSFOU
NFUFSTFUUJOHTDBOCFTBWFEJOUIFTFUVQNFOV&YJTUJOHNFUIPETDBOCFTFMFDUFEGSPNUIFNFUIPETMJTUBOESVOPO
UIFNFUFS.FUIPETDBOCFWJFXFEFEJUFEDPQJFEUPBOFXNFUIPEPSEFMFUFE
Saving the Current Meter Settings as a New Method
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight pH Channel and press f3 (select).
3. Press or to highlight Method and press f3 (select).
 5IFNFUFSXJMMEJTQMBZCurrent SettingsBOEBMJTUPGNFUIPET5IFMJTUXJMMTIPXUIFTFRVFOUJBMOVNCFSPGUIFNFUIPET
(M101) and the date, time and mode of each method (07/01/11 12:45 pH) if methods have been previously created.
 1SFTT or to highlight Current Settings and press f3 (save).
 1SFTT or UPIJHIMJHIUBNFUIPEOVNCFSUIBUUIFDVSSFOUTFUUJOHTXJMMCFTBWFEBTBOEQSFTTf2 (accept).
 5IFNFUFSXJMMQSPNQUGPSBNFUIPETQFDJmDQBTTXPSEUPCFDSFBUFE5IFQBTTXPSENVTUCFBUMFBTUDIBSBDUFSTMPOH
a. Press f3 (edit) to access the numeric entry screen.
b. Press , , or to highlight a number, press f3 (enter) to select the
 IJHIMJHIUFEOVNCFSBOESFQFBUVOUJMUIFOFXQBTTXPSEJTTIPXO
c. Press f2 (done)UPFYJUUIFOVNFSJDFOUSZTDSFFO
d. Press f2 (accept)UPBDDFQUUIFFOUFSFEQBTTXPSE
 5IFNFUFSXJMMBVUPNBUJDBMMZQSPDFFEUPUIFNFBTVSFNFOUNPEF
Loading an Existing Method
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight pH Channel and press f3 (select).
3. Press or to highlight Method and press f3 (select).
 5IFNFUFSXJMMEJTQMBZCurrent SettingsBOEBMJTUPGNFUIPET5IFMJTUXJMMTIPXUIFTFRVFOUJBMOVNCFSPGUIF
methods (M101) and the date, time and mode of each method (07/01/11 12:45 pH) if methods have been previously
created.
 1SFTT or to highlight the method to be loaded and press f2 (load).
 5IFNFUFSXJMMBVUPNBUJDBMMZQSPDFFEUPUIFNFBTVSFNFOUNPEF
24 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Editing, Copying or Deleting an Existing Method
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight pH Channel and press f3 (select).
3. Press or to highlight Method and press f3 (select).
 5IFNFUFSXJMMEJTQMBZCurrent SettingsBOEBMJTUPGNFUIPET5IFMJTUXJMMTIPXUIFTFRVFOUJBMOVNCFSPGUIF
methods (M101) and the date, time and mode of each method (07/01/11 12:45 pH) if methods have been
previously created.
 1SFTT or to highlight the desired method and press f3 (options).
 &OUFSUIFNFUIPETQFDJmDQBTTXPSE
a. Press f3 (edit) to access the numeric entry screen.
b. Press , , or to highlight a number, press f3 (enter) to select the highlighted
 OVNCFSBOESFQFBUVOUJMUIFQBTTXPSEJTTIPXO
c. Press f2 (done)UPFYJUUIFOVNFSJDFOUSZTDSFFO
d. Press f2 (accept)UPBDDFQUUIFFOUFSFEQBTTXPSE
 1SFTT or to highlight Edit, Copy or Delete and press f2 (accept).
B *GEdit is selected – press or to highlight any setup menu parameters that need to be
changed and press f3 (select)3FWJFXBOEDIBOHFBMMBQQMJDBCMFTFUVQNFOVQBSBNFUFST1SFTTf1
(back) to return to the list of methods and then press to return to the measurement mode.
C *GCopy is selected – press or to highlight a method number that the selected
NFUIPEXJMMCFDPQJFEUPBOEQSFTT f2 (accept)5IFNFUFSXJMMBVUPNBUJDBMMZQSPDFFEUPUIF
measurement mode.
D *GDelete is selected – press f2 (yes) to delete the method or press f3 (no) to keep the method.
5IFNFUFSXJMMBVUPNBUJDBMMZQSPDFFEUPUIFNFBTVSFNFOUNPEF
25
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 8 Data Storage and Retrieval
Data Storage Settings
5IFEBUBBOEDBMJCSBUJPOTUPSBHFTFUUJOHTJODMVEFUIFNFBTVSFNFOUSFBEUZQFDPNQVUFSBOEQSJOUFSFYQPSUTFUUJOHT
BOEEBUBMPHTFUUJOHT5IFNFBTVSFNFOUSFBEUZQFEFUFSNJOFTXIFOUIFNFUFSTFOETNFBTVSFNFOUTUPUIFEBUBMPH
if the data log is enabled in the setup menu, and to a printer or computer, if a printer or computer is connected to the
NFUFSBOEFOBCMFEJOUIFTFUVQNFOV3FWJFXUIFEBUFBOEUJNFTFUUJOHTBOEVQEBUFBTOFFEFE
Read Type Selection
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight pH Channel and press f3 (select).
3. Press or to highlight Mode and Settings and press f3 (select).
 1SFTT or to highlight Read Type and press f3 (select).
 1SFTT or to highlight Auto, Continuous or Timed and press f3 (select).
a. Star A211 and Star A321 meters only –
*GTimed is selected and the time interval needs to be changed – highlight Timed; press to highlight
hours (HH), minutes (MM) or seconds (SS); press f3 (edit) to access the numeric entry screen; use the
numeric entry screen to change the values and press f1 (back) XIFOUIFUJNFJOUFSWBMJTDPSSFDU
 1SFTT to return to the measurement mode.
26 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Computer and Printer Export Settings
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight Settings and press f3 (select).
3. Press or to highlight Export Data and press f3 (select).
 1SFTT or to highlight Printing and press f3 (select). Press or to highlight On or Off and
press f3 (select). Select OnUPBMMPXEBUBUPCFTFOUUPBDPNQVUFSPSQSJOUFS
 1SFTT or to highlight Comm Setup and press f3 (select). Press or to highlight RS232 or
USB and press f3 (select). Select RS232JGUIFNFUFSXJMMCFJOUFSGBDFEVTJOHUIF34QPSUBOETFMFDUUSB
JGUIFNFUFSXJMMCFJOUFSGBDFEVTJOHUIF64#QPSU
 1SFTT or to highlight Data Format and press f3 (select). Press or to highlight Printer or
PC (CSV) and press f3 (select). Select PrinterUPFYQPSUUIFEBUBBTBTUSJOHPGUFYUBOETFMFDUPC (CSV) to
FYQPSUUIFEBUBJODPNNBEFMJNJUFEGPSNBU
 1SFTT or to highlight Comm Config and press f3 (select). Press or to highlight 1200, 2400,
4800, 9600, 19200 or 38400 and press f3 (select). Select the baud rate best for the data transfer requirements.
5IFmYFENFUFSPVUQVUTFUUJOHTBSFCJUTTUPQCJUBOEQBSJUZOPOF
 1SFTTf1 (back) and then press to return to the measurement mode.
Data Log Settings
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight Settings and press f3 (select).
3. Press or to highlight Data Log and press f3 (select).
 1SFTT or to highlight Off or On and press f3 (select).
 1SFTTf1 (back) and then press to return to the measurement mode.
27
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Viewing the Calibration Log and Data Log
5IF0SJPO4UBS"BOE4UBS"TFSJFTNFUFSTTBWFVQUPUFOPGUIFNPTUSFDFOUDBMJCSBUJPOTQFSDIBOOFM5IF4UBS
"NFUFSTBWFTVQUPTFUTPGEBUBQPJOUTUIF4UBS"NFUFSTBWFTVQUPTFUTPGEBUBQPJOUTBOEUIF
4UBS"NFUFSTBWFTVQUPTFUTPGEBUBQPJOUT"MPHOVNCFSJTBTTJHOFEUPFBDIEBUBQPJOU5IFSFBEUZQFPG
UIFEJTQMBZFEDIBOOFMPSDIBOOFMTEFUFSNJOFTXIFONFBTVSFNFOUTBSFFYQPSUFEUPUIFEBUBMPH8IFOBEBUBQPJOUJT
FYQPSUFEUPBDPNQVUFSPSQSJOUFSUIFUJNFTUBNQXJMMJODMVEFUIFIPVSNJOVUFTBOETFDPOET
Viewing the Calibration Log
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight View Log and press f3 (select).
3. Press or to highlight Calibration Log and press f2 (accept).
 1SFTT or to highlight pH – Channel and press f2 (select).
 1SFTT or to highlight pH, RmV or ORP and press f2 (select).
 5IFNFUFSXJMMEJTQMBZBMJTUPGDBMJCSBUJPOTGPSUIFTFMFDUFEDIBOOFMBOEDBMJCSBUJPOUZQF5IFMJTUTIPXTUIF
TFRVFOUJBMOVNCFSPGUIFDBMJCSBUJPOBOEUIFEBUFBOEUJNFJUXBTTBWFE(07/01/2011 12:45).
 5PWJFXUIFDBMJCSBUJPOEBUBQSFTT or to highlight a specific calibration and press f2 (select). Press
f2 (print) to print the calibration, press f3 (info)UPWJFXUIFFMFDUSPEFTMPQFCFUXFFOQ)CVGGFSQPJOUTPSQSFTT
f1 (back) to return to the list of calibrations.
 1SFTT to return to the measurement mode.
Viewing the Data Log
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight View Log and press f3 (select).
3. Press or to highlight Data Log and press f2 (accept).
 1SFTT or to highlight pH – Channel and press f2 (select).
 5IFNFUFSXJMMEJTQMBZBMJTUPGUIFEBUBQPJOUT5IFMJTUTIPXTUIFTFRVFOUJBMOVNCFSPGUIFEBUBQPJOUBOEUIF
EBUFBOEUJNFUIFEBUBQPJOUXBTTBWFE(07/01/2011 12:45).
 5PWJFXUIFNFBTVSFNFOUJOGPSNBUJPOGPSBOJOEJWJEVBMEBUBQPJOUQSFTT or to highlight the data point
and press f2 (select). Press f2 (print) to print the data point or press f1 (back) to return to the list of data points.
 1SFTT to return to the measurement mode.
28 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Exporting the Data Log to a Computer or Printer
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight View Log and press f3 (select).
3. Press or to highlight Data Log and press f2 (accept).
 1SFTT or to highlight pH – Channel and press f3 (select).
 5IFNFUFSXJMMEJTQMBZBMJTUPGUIFEBUBQPJOUT5IFMJTUTIPXTUIFTFRVFOUJBMOVNCFSPGUIFEBUBQPJOUBOEUIF
EBUFBOEUJNFUIFEBUBQPJOUXBTTBWFE(07/01/2011 12:45).
 1SFTTf3 (options).
 1SFTT or to highlight Log Export and press f2 (accept).
 1SFTT or to highlight Today, Last 10, Range or AllBTUIFEBUBMPHQPJOUTUPCFFYQPSUFE
and press f2 (accept).
B *GToday, Last 10 or All is selected – press f2 (yes)UPFYQPSUUIFEBUB
C *GRangeJTTFMFDUFEoFOUFSUIFmSTUBOEMBTUEBUFQPJOUT
i. Press f3 (edit) to display the numeric entry screen.
ii. Press , , or to highlight a number, press f3 (enter) to select the highlighted
OVNCFSBOESFQFBUVOUJMUIFOVNCFSPGUIFmSTUEBUBQPJOUUPCFFYQPSUFEJTTIPXOPOUIF
display.
iii. Press f2 (done)UPFYJUUIFOVNFSJDFOUSZTDSFFO
iv. Press to enter the last data point.
v. Press f3 (edit) to display the numeric entry screen.
vi. Press , , or to highlight a number, press f3 (enter) to select the highlighted
OVNCFSBOESFQFBUVOUJMUIFOVNCFSPGUIFMBTUEBUBQPJOUUPCFFYQPSUFEJTTIPXOPOUIFEJTQMBZ
vii. Press f2 (done)UPFYJUUIFOVNFSJDFOUSZTDSFFO
viii. Press f2 (accept) and then press f2 (yes)UPFYQPSUUIFEBUB
 "GUFSFYQPSUJOHUIFEBUBUIFNFUFSXJMMBVUPNBUJDBMMZSFUVSOUPUIFNFBTVSFNFOUNPEF
29
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Clearing the Data Log
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight View Log and press f3 (select).
3. Press or to highlight Data Log and press f2 (accept).
 1SFTT or to highlight pH – Channel and press f3 (select).
 5IFNFUFSXJMMEJTQMBZBMJTUPGUIFEBUBQPJOUT5IFMJTUTIPXTUIFTFRVFOUJBMOVNCFSPGUIFEBUBQPJOUBOEUIF
EBUFBOEUJNFUIFEBUBQPJOUXBTTBWFE(07/01/2011 12:45).
 1SFTTf3 (options).
 1SFTT or to highlight Log Clear and press f2 (accept).
 1SFTTf2 (yes)UXJDFUPDMFBSUIFFOUJSFEBUBMPH
 &OUFSUIFQBTTXPSE5IFEFGBVMUJT111111 and pressing f2 (accept)XJMMBEWBODFUIFNFUFSUPUIFOFYUTUFQ
1SFTTf1 (back) and then press to return to the measurement mode.
30 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Interfacing a Printer
5IF0SJPO4UBSQSJOUFS$BU/PJTSFDPNNFOEFEGPSVTFXJUIUIF0SJPO4UBS"BOE4UBS"TFSJFT
NFUFST5IF0SJPO4UBSQSJOUFSIBTBCBVESBUFBOEJUJTQBDLBHFEXJUIUIF34QSJOUFSDBCMF$BU/P
XIJDIJTSFRVJSFEUPJOUFSGBDFBO0SJPO4UBS"NFUFSUPBQSJOUFS5IFCBVESBUFPGUIF0SJPO4UBS"
BOE4UBS"TFSJFTNFUFSTDBOCFTFUUPPSGPSDPNNVOJDBUJPOUPEJGGFSFOU
QSJOUFST5IFmYFENFUFSTFUUJOHTBSFOVNCFSPGCJUTTUPQCJUTQBSJUZOPOF5IFnPXDPOUSPMGPSUIFQSJOUFS
TIPVMECFTFUUP90/90''5IFNFUFSXJMMTFOENFBTVSFNFOUTBOEDBMJCSBUJPOEBUBUPUIFQSJOUFSJGUIFQSJOUFS
option is turned on in the setup menu. Alternatively, measurements and calibration data can be sent to the data log
and calibration log. From the data log and calibration log, the operator can chose to print a single data log point, a
range of data log points or the entire data log.
Interfacing a Computer
5IF0SJPO4UBS"BOE4UBS"TFSJFTNFUFSTDBOTFOENFBTVSFNFOUTBOEDBMJCSBUJPOEBUBUPBDPNQVUFSJOB
DPNNBEFMJNJUFEGPSNBUUIBUJTFBTZUPQBSTFJODPNQVUFSQSPHSBNTMJLF&YDFM$POOFDUUIF34DPNQVUFSDBCMF
$BU/PUPUIF34JOQVUPOUIFNFUFSBOEUIF34JOQVUPOBDPNQVUFS5IFCBVESBUFPGUIF0SJPO
4UBS"TFSJFTNFUFSTDBOCFTFUUPPS5IFmYFENFUFSTFUUJOHTBSFOVNCFS
PGCJUTTUPQCJUTQBSJUZOPOF5IFnPXDPOUSPMGPSUIFDPNQVUFSTIPVMECFTFUUP90/90''5IFNFUFSXJMM
send measurements and calibration data to the computer if the computer option is turned on in the setup menu.
Alternatively, measurements and calibration data can be sent to the data log and calibration log. From the data log
and calibration log, the operator can chose to send a single data log point, a range of data log points or the entire
data log to the computer.
A complimentary copy of general data logging software is available for use with the Orion Star A200 and Star A300 series
meters. This can be found on the CD included with the meter or at www.thermoscientific.com/water.
31
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 9 Customer Services
Meter Errors and Recommended Actions
Error Recommended Actions
%JTQMBZ&SSPS *GUIJTFSSPSPDDVSTDPOUBDU5FDIOJDBM4VQQPSU
)BSEXBSF&SSPS *GUIJTFSSPSPDDVSTDPOUBDU5FDIOJDBM4VQQPSU
$IBOOFM*OQVU&SSPS %JTDPOOFDUBMMPGUIFFMFDUSPEFTGSPNUIFNFUFSDPOOFDUUIF#/$TIPSUJOHDBQUPUIF#/$JOQVUTBOE
repeat the self test.
3FGFSFODF*OQVU&SSPS Make sure that all of the electrodes are disconnected from the meter (including the reference input),
DPOOFDUUIF#/$TIPSUJOHDBQUPUIF#/$JOQVUBOESFQFBUUIFTFMGUFTU
(SPVOE*OQVU&SSPS Make sure that all of the electrodes are disconnected from the meter (including the ground input),
DPOOFDUUIF#/$TIPSUJOHDBQUPUIF#/$JOQVUBOESFQFBUUIFTFMGUFTU
,FZQBE'BJMVSF Repeat the self test. When the meter prompts, press each of the keys one at a time, including the
QPXFSLFZXJUIJOGJWFTFDPOETPGPOFPUIFS*GUIFFSSPSQFSTJTUTDPOUBDU5FDIOJDBM4VQQPSU
Over Range Measurement or
6OEFS3BOHF.FBTVSFNFOU
*GUIFNFBTVSFNFOUJTnBTIJOHBOEPWFSSBOHFPSVOEFSSBOHFUIFNFBTVSFNFOUWBMVFJT
PVUTJEFPGUIFBMMPXBCMFNFBTVSFNFOUSBOHF.BLFTVSFUIBUUIFFMFDUSPEFJTDPOOFDUFEUPUIF
appropriate meter input and the electrode parameters are entered correctly in the setup menu.
$MFBOUIFFMFDUSPEFBDDPSEJOHUPUIFFMFDUSPEFVTFSHVJEFBOESFDBMJCSBUFUIFFMFDUSPEFXJUIOFX
CVGGFSTPSTUBOEBSET*GUIFFSSPSQFSTJTUTQFSGPSNUIFNFUFSTFMGUFTU
Measurement Out of Range
Make sure that the electrode is connected to the appropriate meter input and the electrode
parameters are entered correctly in the setup menu. Clean the electrode according to the electrode
VTFSHVJEFBOESFDBMJCSBUFUIFFMFDUSPEFXJUIOFXCVGGFSTPSTUBOEBSET1FSGPSNUIFNFUFSTFMGUFTU
32 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Meter Self Test and Checkout
Meter Self Test
 *OUIFNFBTVSFNFOUNPEFQSFTTTFUVQ .
2. Press , , or to highlight Diagnostics and press f3 (select).
3. Press or to highlight Self Test and press f2 (accept).
 %JTDPOOFDUBMMPGUIFFMFDUSPEFTBOEQSPCFTGSPNUIFNFUFSBUUBDIUIF#/$TIPSUJOHDBQUPUIF#/$JOQVU
and press f2 (yes).
 1SFTTf2 (yes) to start the self test.
 5IFNFUFSXJMMQFSGPSNBOBDDVSBDZUFTU8IFOUIFNFUFSEJTQMBZTAccuracy Test Passed, press f2 (next).
 5IFNFUFSXJMMQFSGPSNBO&130.UFTU8IFOUIFNFUFSEJTQMBZTE2Prom Test Passed, press f2 (next).
 5IFNFUFSXJMMQFSGPSNBLFZQBEUFTU1SFTTFBDIPGUIFLFZTPOUIFLFZQBEJODMVEJOH ) one at a time in
any order. When the meter displays Keypad Test Passed, press f2 (next).
 5IFNFUFSXJMMEJTQMBZSelf Test Passed. Press to return to the measurement mode.
Meter Checkout (pH)
 "GUFSUIFTFMGUFTUJTDPNQMFUFUIFNFUFSXJMMSFUVSOUPUIFNFBTVSFNFOUNPEF,FFQBMMPGUIFFMFDUSPEFT
EJTDPOOFDUFEGSPNUIFNFUFSBOELFFQUIF#/$TIPSUJOHDBQBUUBDIFEUPUIF#/$JOQVUPOUIFNFUFS
 4FUUIFNFBTVSFNFOUNPEFUPQ)5IFNFUFSTIPVMESFBEBTUFBEZQ)Q)*GUIFNFUFSEPFTOPU
SFBEBTUFBEZQFSGPSNBPOFQPJOUQ)DBMJCSBUJPOXJUIUIF#/$TIPSUJOHDBQBUUBDIFEUPUIF#/$
JOQVU4FUUIFQ)WBMVFUPBOEUIFTMPQFWBMVFUP3FGFSUPUIFQ)$BMJCSBUJPOTFDUJPOGPSEFUBJMFE
instructions.
 *OUIFNFBTVSFNFOUNPEFUIFNFUFSTIPVMESFBEBTUFBEZN7N7*GUIFNFUFSEPFTOPUSFBEB
TUFBEZN7N7XJUIUIF#/$TIPSUJOHDBQBUUBDIFEUPUIF#/$JOQVUDPOUBDU5FDIOJDBM4VQQPSU
33
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Electrode Stability Test
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight Diagnostics and press f3 (select).
3. Press or to highlight Stability Test and press f2 (accept).
 1SFTT or to highlight pH – Channel and press f2 (next).
 .BLFTVSFUIFFMFDUSPEFPSQSPCFUPCFUFTUFEJTDPOOFDUFEUPUIFDIBOOFMTFMFDUFEJOUIFQSFWJPVTTUFQBOE
press f2 (yes).
 1MBDFUIFFMFDUSPEFPSQSPCFJOBOBQQSPQSJBUFTPMVUJPOXJUIBTUBCMFUFNQFSBUVSFBOEQSFTTf2 (yes).
 B0SJPOQ)CVGGFS$BU/PPSJTSFDPNNFOEFEGPSUFTUJOHBQ)FMFDUSPEF
 C0SJPO031TUBOEBSE$BU/PPSJTSFDPNNFOEFEGPSUFTUJOHBO031FMFDUSPEF
 5IFNFUFSXJMMQFSGPSNUIFTUBCJMJUZUFTUBOETIPXUIFN7QFSNJOVUFESJGUBOEOPJTFPOUIFEJTQMBZ
 5IFNFUFSXJMMEJTQMBZStability Test Passed or Stability Test Failed. Press f1 (esc) or to return to the
measurement mode.
Meter Factory Reset
"GUFSBGBDUPSZSFTFUJTQFSGPSNFEBMMPGUIFNFUFSTFUUJOHTXJMMCFSFTFUUPUIFGBDUPSZEFGBVMUTBOEUIFEBUBMPH
DBMJCSBUJPOMPHBOENFUIPETXJMMCFEFMFUFE
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight Diagnostics and press f3 (select).
3. Press or to highlight Factory Reset and press f2 (accept).
 1SFTTf2 (yes) to reset the meter.
 5IFNFUFSXJMMQBVTFQPXFSPGGBOESFTUBSUXJUIUIFGBDUPSZEFGBVMUTFUUJOHT
Meter User Reset
"GUFSBVTFSSFTFUJTQFSGPSNFEBMMPGUIFNFUFSTFUVQNFOVTFUUJOHTXJMMCFSFTFUUPUIFEFGBVMUTIPXFWFSUIFEBUB
MPHDBMJCSBUJPOMPHBOENFUIPETXJMMCFSFUBJOFE
*OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight Diagnostics and press f3 (select).
3. Press or to highlight User Reset and press f2 (accept).
1SFTTf2 (yes) to reset the meter.
5IFNFUFSXJMMQBVTFBOEUIFOSFUVSOUPUIFNFBTVSFNFOUNPEFXJUIUIFEFGBVMUTFUVQNFOVTFUUJOHT
34 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
About Meter
 *OUIFNFBTVSFNFOUNPEFQSFTT .
2. Press , , or to highlight Diagnostics and press f3 (select).
3. Press or to highlight About Meter and press f2 (accept).
 5IFNFUFSNPEFMNFBTVSFNFOUQBSBNFUFSTTFSJBMOVNCFSBOETPGUXBSFSFWJTJPOXJMMCFEJTQMBZFE
 1SFTTf1 (back) to return to the diagnostics menu or press to return to the measurement mode.
Assistance
"GUFSUSPVCMFTIPPUJOHBMMDPNQPOFOUTPGZPVSNFBTVSFNFOUTZTUFNDPOUBDU5FDIOJDBM4VQQPSU8JUIJOUIF6OJUFE
4UBUFTDBMMBOEPVUTJEFUIF6OJUFE4UBUFTDBMMPSGBY*O&VSPQFUIF
.JEEMF&BTUBOE"GSJDBDPOUBDUZPVSMPDBMBVUIPSJ[FEEFBMFS'PSUIFNPTUDVSSFOUDPOUBDUJOGPSNBUJPOWJTJUXXX
UIFSNPTDJFOUJmDDPNXBUFS
Warranty and Registration
5PSFHJTUFSZPVSNFUFSBOEGPSUIFNPTUDVSSFOUXBSSBOUZJOGPSNBUJPOWJTJUXXXUIFSNPTDJFOUJmDDPNXBUFS
35
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Ordering Information
Orion Star A200 Series Benchtop Meters
Cat. No. Description
STARA2110 4UBS"Q)#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE
STARA2115 4UBS"Q)#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE#/6.%3044Q)"5$5SJPEF
3044Q)#VGGFS,JU
STARA2120 4UBS"$POEVDUJWJUZ#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE
STARA2125 4UBS"$POEVDUJWJUZ#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE.%$POEVDUJWJUZ$FMM
$POEVDUJWJUZç4DN4UBOEBSE
STARA2126 4UBS"$POEVDUJWJUZ#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE.%$POEVDUJWJUZ$FMMXJUI
%FUBDIBCMF'MPX5ISPVHI$FMM$POEVDUJWJUZç4DN4UBOEBSE
STARA2130 4UBS"3%0%0#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE
STARA2135 4UBS"3%0%0#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE.%1PMBSPHSBQIJD%01SPCFXJUI
$BMJCSBUJPO4MFFWF1SPCF.BJOUFOBODF,JU#0%'VOOFM4UJSSFSBOE"EBQUFS
STARA2136 4UBS"3%0%0#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE.%"65045*3%0#0%1SPCF
1PMBSPHSBQIJD&MFDUSPMZUF4PMVUJPO1SPCF.BJOUFOBODF,JU
STARA2140 4UBS"Q)*4&#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE
STARA2145 4UBS"Q)*4&#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE#/68130446MUSBQ)&MFDUSPEF
.%"5$1SPCF4UJSSFS1SPCF3044Q)#VGGFS,JU
STARA2146
4UBS"Q)*4&#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE#/68130446MUSBQ)&MFDUSPEF
.%"5$1SPCF)1#/81)JHI1FSGPSNBODF"NNPOJB&MFDUSPEF4UJSSFS1SPCF
"NNPOJBQQN4UBOEBSE-PXMFWFM"NNPOJB*4""NNPOJB&MFDUSPEF
Storage Solution
STARA2147
4UBS"Q)*4&#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE#/68130446MUSBQ)&MFDUSPEF
.%"5$1SPCF#/81'MVPSJEF&MFDUSPEF4UJSSFS1SPCF'MVPSJEFQQN
XJUI5*4"#**4UBOEBSE'MVPSJEFQQNXJUI5*4"#**4UBOEBSE'MVPSJEFQQNXJUI
5*4"#**4UBOEBSE5*4"#**
STARA2148
4UBS"Q)*4&#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE#/68130446MUSBQ)&MFDUSPEF
.%"5$1SPCF#/8130444PEJVN&MFDUSPEFXJUI4PEJVN4UBOEBSET3FBHFOUTBOE
4PMVUJPOT4UJSSFS1SPCF
STARA2150 4UBS"Q)$POEVDUJWJUZ#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE
STARA2155
4UBS"Q)$POEVDUJWJUZ#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE#/6.%3044Q)"5$
5SJPEF.%$POEVDUJWJUZ$FMM3044Q)#VGGFS,JU$POEVDUJWJUZç4DN
Standard
STARA2160 4UBS"Q)3%0%0#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE
STARA2165
4UBS"Q)3%0%0#FODIUPQ.FUFSXJUI&MFDUSPEF4UBOE#/6.%3044Q)"5$5SJPEF
.%1PMBSPHSBQIJD%01SPCFXJUI$BMJCSBUJPO4MFFWF3044Q)#VGGFS,JU
1SPCF.BJOUFOBODF,JU#0%'VOOFM4UJSSFSBOE"EBQUFS
36 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Ordering Information (cont.)
Orion Star A200 and Star A300 Series Portable Meters
Cat. No. Description
STARA2210 Star A221 pH Portable Meter
STARA2215
4UBS"Q)1PSUBCMF.FUFS68..%3044Q)"5$5SJPEFXJUI.FUFS$BCMFQ)#VGGFS
1PVDIFTQ)#VGGFS1PVDIFTQ)#VGGFS1PVDIFT3044&MFDUSPEF4UPSBHF
4PMVUJPO3JOTF4PMVUJPO1PVDIFT45"3"$44UBS"$BSSZJOH$BTF45"3""34UBS".FUFS"SNPS
STARA2220 Star A222 Conductivity Portable Meter
STARA2225
4UBS"$POEVDUJWJUZ1PSUBCMF.FUFS.%$POEVDUJWJUZ$FMMXJUI.FUFS$BCMF
$POEVDUJWJUZç4DN4UBOEBSE1PVDIFT3JOTF4PMVUJPO1PVDIFT45"3"$44UBS"$BSSZJOH$BTF
45"3""34UBS".FUFS"SNPS
STARA2230 4UBS"3%0%01PSUBCMF.FUFS
STARA2235 4UBS"3%0%01PSUBCMF.FUFS.%3%00QUJDBM%01SPCFXJUI.FUFS$BCMF45"3"$44UBS"
$BSSZJOH$BTF45"3""34UBS".FUFS"SNPS
STARA3210 Star A321 pH Portable Meter
STARA3215
4UBS"Q)1PSUBCMF.FUFS68..%3044Q)"5$5SJPEFXJUI.FUFS$BCMFQ)#VGGFS
1PVDIFTQ)#VGGFS1PVDIFTQ)#VGGFS1PVDIFT3044&MFDUSPEF4UPSBHF
4PMVUJPO3JOTF4PMVUJPO1PVDIFT45"3"$44UBS"$BSSZJOH$BTF45"3""34UBS".FUFS"SNPS
STARA3220 Star A322 Conductivity Portable Meter
STARA3225
4UBS"$POEVDUJWJUZ1PSUBCMF.FUFS.%$POEVDUJWJUZ$FMMXJUI.FUFS$BCMF
$POEVDUJWJUZç4DN4UBOEBSE1PVDIFT3JOTF4PMVUJPO1PVDIFT45"3"$44UBS"$BSSZJOH$BTF
45"3""34UBS".FUFS"SNPS
STARA3230 4UBS"3%0%01PSUBCMF.FUFS
STARA3235 4UBS"3%0%01PSUBCMF.FUFS.%3%00QUJDBM%01SPCFXJUI.FUFS$BCMF45"3"$44UBS"
$BSSZJOH$BTF45"3""34UBS".FUFS"SNPS
STARA3240 4UBS"Q)*4&1PSUBCMF.FUFS
STARA3245
4UBS"Q)*4&1PSUBCMF.FUFS68..%3044Q)"5$5SJPEFXJUI.FUFS$BCMFQ)#VGGFS
1PVDIFTQ)#VGGFS1PVDIFTQ)#VGGFS1PVDIFT3044&MFDUSPEF4UPSBHF
4PMVUJPO3JOTF4PMVUJPO1PVDIFT45"3"$44UBS"$BSSZJOH$BTF45"3""34UBS".FUFS"SNPS
STARA3250 4UBS"Q)$POEVDUJWJUZ1PSUBCMF.FUFS
STARA3255
4UBS"Q)$POEVDUJWJUZ1PSUBCMF.FUFS68..%3044Q)"5$5SJPEFXJUI.FUFS$BCMF
.%$POEVDUJWJUZ$FMMXJUI.FUFS$BCMFQ)#VGGFS1PVDIFTQ)#VGGFS
1PVDIFTQ)#VGGFS1PVDIFT3044&MFDUSPEF4UPSBHF4PMVUJPO$POEVDUJWJUZ
ç4DN4UBOEBSE1PVDIFT3JOTF4PMVUJPO1PVDIFT45"3"$44UBS"$BSSZJOH$BTF45"3""3
Star A Meter Armor
STARA3260 4UBS"Q)3%0%01PSUBCMF.FUFS
STARA3265
4UBS"Q)3%0%01PSUBCMF.FUFS68..%3044Q)"5$5SJPEFXJUI.FUFS$BCMF
.%3%00QUJDBM%01SPCFXJUI.FUFS$BCMFQ)#VGGFS1PVDIFTQ)
#VGGFS1PVDIFTQ)#VGGFS1PVDIFT3044&MFDUSPEF4UPSBHF4PMVUJPO45"3"$44UBS
"$BSSZJOH$BTF45"3""34UBS".FUFS"SNPS
STARA3290 4UBS"Q)*4&$POEVDUJWJUZ3%0%01PSUBCMF.FUFS
STARA3295
4UBS"Q)*4&$POEVDUJWJUZ3%0%01PSUBCMF.FUFS68..%3044Q)"5$5SJPEFXJUI.FUFS
$BCMF.%$POEVDUJWJUZ$FMMXJUI.FUFS$BCMF.%3%00QUJDBM%01SPCFXJUI.FUFS
$BCMFQ)#VGGFS1PVDIFTQ)#VGGFS1PVDIFTQ)#VGGFS1PVDIFT
3044&MFDUSPEF4UPSBHF4PMVUJPO$POEVDUJWJUZç4DN4UBOEBSE1PVDIFT
3JOTF4PMVUJPO1PVDIFT45"3"$44UBS"$BSSZJOH$BTF45"3""34UBS".FUFS"SNPS
37
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Ordering Information (cont.)
Orion Star A Series Accessories and Electrodes
Cat. No. Description
STARA-BEA 0SJPO4UBS"4FSJFT&MFDUSPEF"SNBOE)PMEFS"UUBDIFTUPCBTFPGNFUFSPS45"3")#CBTF
STARA-HB 0SJPO4UBS"4FSJFT&MFDUSPEF4UBOE#BTF0OMZ6TFXIFOFMFDUSPEFBSNJTOPUBUUBDIFEUPNFUFS
STARA-CS Orion Star A Series Portable Hard Carrying Case
STARA-AR 0SJPO4UBS"4FSJFT1PSUBCMF.FUFS"SNPSXJUI4UBOEJODMVEFTPOFQ)&MFDUSPEF)PMEFSXJUI
4UPSBHF4MFFWFBOEPOF$POEVDUJWJUZ$FMMBOE%01SPCF)PMEFS
STARA-ESPH 0SJPO4UBS"4FSJFTQ)&MFDUSPEF)PMEFSGPS.FUFS"SNPSSFRVJSFTQ)TUPSBHFTMFFWFQVSDIBTF
separately)
STARA-ESCD 0SJPO4UBS"4FSJFT$POEVDUJWJUZ$FMMBOE%01SPCF)PMEFSGPS.FUFS"SNPS
1010003 0SJPO4UBS"4FSJFT1PXFS"EBQUFSGPS#FODIUPQBOE1PSUBCMF.FUFST
8102BNUWP 30446MUSB(MBTT3FGJMMBCMFQ)&MFDUSPEF
8302BNUMD 3044(MBTT3FGJMMBCMFQ)"5$5SJPEF
8157BNUMD 30443FGJMMBCMF&QPYZQ)"5$5SJPEF
8107UWMMD 3044(FMGJMMFE&QPYZQ)"5$5SJPEFXJUI.FUFS$BCMF
927005MD &QPYZ4UFFM"5$1SPCF
927007MD Stainless Steel ATC Probe
9179BNMD (FMGJMMFE&QPYZ031"5$5SJPEF
9180BNMD 3FGJMMBCMF&QPYZ031"5$5SJPEF
9512HPBNWP )JHI1FSGPSNBODF"NNPOJB&MFDUSPEF
9609BNWP 'MVPSJEF&MFDUSPEF
8611BNWP 30444PEJVN&MFDUSPEFXJUI4PEJVN4UBOEBSET3FBHFOUTBOE4PMVUJPOT
013005MD %VSB1SPCF&QPYZ$FMM,$POEVDUJWJUZ$FMM
013010MD %VSB1SPCF&QPYZ$FMM,$POEVDUJWJUZ$FMMXJUI.FUFS$BCMF
013020MD %VSB1SPCF&QPYZ$FMM,$POEVDUJWJUZ$FMMXJUI.FUFS$BCMF
013025MD %VSB1SPCF&QPYZ$FMM,$POEVDUJWJUZ$FMMXJUI.FUFS$BCMF
013016MD 4UFFM$FMM,$POEVDUJWJUZ$FMMXJUI%FUBDIBCMF'MPX5ISPVHI$FMM
083005MD 1PMBSPHSBQIJD%01SPCFXJUI$BMJCSBUJPO4MFFWF
083010MD 1PMBSPHSBQIJD%01SPCFXJUI$BMJCSBUJPO4MFFWFBOE.FUFS$BCMF
083025MD 1PMBSPHSBQIJD%01SPCFXJUI$BMJCSBUJPO4MFFWFBOE.FUFS$BCMF
086030MD "65045*3%0#0%1SPCF
087010MD 3%00QUJDBM%01SPCFXJUI.FUFS$BCMF
087020MD 3%00QUJDBM%01SPCFXJUI.FUFS$BCMF
087030MD 3%00QUJDBM%01SPCFXJUI.FUFS$BCMF
096019 4UJSSFS1SPCFGPS4UBS"4FSJFT#FODIUPQ.FUFST
38 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
This page intentionally left blank.
39
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Chapter 10 Meter Compliance
Notice of Compliance
This equipment generates, uses, and can radiate radio frequency energy and if not installed and used in
BDDPSEBODFXJUIUIFJOTUSVDUJPONBOVBMNBZDBVTFJOUFSGFSFODFUPSBEJPDPNNVOJDBUJPOT*UIBTCFFOUFTUFEBOE
GPVOEUPDPNQMZXJUIUIFMJNJUTGPSB$MBTT"DPNQVUJOHEFWJDFQVSTVBOUUP4VCQBSU+PG1BSUPG'$$3VMFT
XIJDIBSFEFTJHOFEUPQSPWJEFSFBTPOBCMFQSPUFDUJPOBHBJOTUTVDIJOUFSGFSFODFXIFOPQFSBUFEJOBDPNNFSDJBM
FOWJSPONFOU0QFSBUJPOPGUIJTFRVJQNFOUJOBSFTJEFOUJBMBSFBJTMJLFMZUPDBVTFJOUFSGFSFODFJOXIJDIDBTFUIF
VTFSBUIJTPXOFYQFOTFXJMMCFSFRVJSFEUPUBLFXIBUFWFSNFBTVSFTNBZCFSFRVJSFEUPDPSSFDUUIFJOUFSGFSFODF
 i5IJTEJHJUBMBQQBSBUVTEPFTOPUFYDFFEUIF$MBTT"MJNJUTGPSSBEJPOPJTFFNJTTJPOTGSPNEJHJUBMBQQBSBUVTTFU
PVUJOUIF3BEJP*OUFSGFSFODF3FHVMBUJPOTPGUIF$BOBEJBO%FQBSUNFOUPG$PNNVOJDBUJPOTw
 i-FQSFTFOUBQQBSFJMOVNFSJRVFOFNFUQBTEFCSVJUTSBEJPFMFDUSJRVFTEFQBTTBOUMFTMJNJUFTBQQMJDBCMFTBVY
appareils numeriques (de la class A) prescrites dans le Reglement sur le brouillage radioelectrique edicte par le
NJOJTUFSFEFT$PNNVOJDBUJPOTEV$BOBEBw
WEEE Compliance
5IJTQSPEVDUJTSFRVJSFEUPDPNQMZXJUIUIF&VSPQFBO6OJPOT8BTUF&MFDUSJDBM&MFDUSPOJD&RVJQNFOU8&&&
%JSFDUJWF&$*UJTNBSLFEXJUIUIFTZNCPMPOUIFMFGU
 5IFSNP'JTIFS4DJFOUJmDIBTDPOUSBDUFEXJUIPOFPSNPSFSFDZDMJOHEJTQPTBMDPNQBOJFTJOFBDI&6.FNCFS4UBUF
BOEUIJTQSPEVDUTIPVMECFEJTQPTFEPGPSSFDZDMFEUISPVHIUIFN'VSUIFSJOGPSNBUJPOPODPNQMJBODFXJUIUIFTF
directives, the recyclers in your country, and information on Thermo Scientific Orion products that may assist the
EFUFDUJPOPGTVCTUBODFTTVCKFDUUPUIF3P)4%JSFDUJWFBSFBWBJMBCMFBUXXXUIFSNPTDJFOUJmDDPN
40 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Declaration of Conformity
Manufacturer: Thermo Fisher Scientific Inc.
Address: Ayer Rajah Crescent
#ML
4JOHBQPSF
Singapore
Hereby declares that the following products:
#FODIUPQNFUFSTBSFSBUFEUP7"$)["
)BOEIFMENFUFSTVTFGPVSOPOSFDIBSHFBCMF""CBUUFSJFT
Benchtop Meters Portable Meters
Orion Star A211 pH Orion Star A221 pH
Orion Star A212 Conductivity Orion Star A222 Conductivity
0SJPO4UBS"3%0%0 0SJPO4UBS"3%0%0
0SJPO4UBS"Q)*4& Orion Star A321 pH
0SJPO4UBS"Q)$POEVDUJWJUZ Orion Star A322 Conductivity
0SJPO4UBS"Q)3%0%0 0SJPO4UBS"3%0%0
0SJPO4UBS"Q)*4&
0SJPO4UBS"Q)$POEVDUJWJUZ
0SJPO4UBS"Q)3%0%0
0SJPO4UBS"Q)*4&$POEVDUJWJUZ3%0%0
Equipment Class:
Measurement, control and laboratory
0SJPO4UBS"TFSJFTNFUFSTBSF&.$$MBTT"
Conforms with the following directives and standards:
EN61326-1:2006 Electromagnetic Compatibility (EMC Directive)
 &MFDUSJDBMFRVJQNFOUGPSNFBTVSFNFOU
 DPOUSPMBOEMBCPSBUPSZVTF&.$SFRVJSFNFOUT
EN61010-1:2001 Safety Standards
UL61010-1:2004 Safety requirements for electrical equipment for measurement,
CAN/CSA C22.2 No. 61010-1-04 DPOUSPMBOEMBCPSBUPSZVTFHFOFSBMSFRVJSFNFOUT
$IFPX,XBOH$IBO
2"3FHVMBUPSZ.BOBHFS
1MBDFBOE%BUFPG*TTVF
+VOF
Singapore
41
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Meter Specifications
Meter Operating Conditions
Operating Ambient Temperature UP¡$
Operating Relative Humidity UPOPODPOEFOTJOH
Storage Temperature UP¡$
Storage Relative Humidity UPOPODPOEFOTJOH
Pollution %FHSFF
Overvoltage $BUFHPSZ**
Weight 1PSUBCMFH
#FODIUPQH
Size 1PSUBCMFDN)YDN8YDN%
#FODIUPQDN)YDN8YDN%
Regulatory and Safety $&567'$$$MBTT"
1PXFS3BUJOH %$*OQVU7%$"
#BUUFSZY""
Shock and Vibration 7JCSBUJPOTIJQQJOHIBOEMJOHQFS*45""
4IPDLESPQUFTUJOQBDLBHJOHQFS*45""
&ODMPTVSFEFTJHOFEUPNFFU 1PSUBCMF*1
#FODIUPQ*1
Warranty 3 year meter replacement
Universal Power Adapter Operating Conditions
Operating Ambient Temperature UP¡$
Operating Relative Humidity UPOPODPOEFOTJOH
Storage Temperature UP¡$
Storage Relative Humidity UPOPODPOEFOTJOH
Pollution %FHSFF
Overvoltage $BUFHPSZ**
42 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Meter Parameter Specifications
pH
Range UP
Resolution 
Relative Accuracy 
Calibration Points 6QUP
$BMJCSBUJPO&EJUJOH 4UBS":FT
4UBS"/P
4UBS":FT
mV/RmV
Range N7
Resolution 
Relative Accuracy 
&H ORP Mode :FT
Temperature Measurement
Range UP¡$
Resolution 
Relative Accuracy 
Offset Calibration 1 point
Data Logging
/VNCFSPG1PJOUT 4UBS"XJUIUJNFBOEEBUFTUBNQ
4UBS"XJUIUJNFBOEEBUFTUBNQ
4UBS"XJUIUJNFBOEEBUFTUBNQ
-PH'VODUJPO 4UBS".BOVBM3FBEZ"6503&"%5JNFE
4UBS".BOVBM"VUPNBUJDXJUI"6503&"%
4UBS".BOVBM3FBEZ"6503&"%5JNFE
-PH&EJU 4UBS"%FMFUFJOEJWJEVBMQPJOUTHSPVQPGQPJOUTPSBMM
4UBS"%FMFUFMBTUSFBEJOHPSBMM
4UBS"%FMFUFJOEJWJEVBMQPJOUTHSPVQPGQPJOUTPSBMM
Inputs
Q)&MFDUSPEF #/$SFGFSFODFQJO
ATC Probe QJO.JOJ%*/
Output 3464#
Power
AC Adapter 6OJWFSTBM7"$JODMVEFEXJUI4UBS"
#BUUFSZ1PXFS 'PVS"MLBMJOF""JODMVEFEXJUI4UBS"BOE4UBS"
#BUUFSZ-JGF IPVST
43
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Appendix 1 Advanced Meter Features
Automatic pH Buffer Recognition Feature
5IF0SJPO4UBS"BOE4UBS"TFSJFTQ)NFUFSTBSFDBQBCMFPGBVUPNBUJDBMMZSFDPHOJ[JOHQ)
BOECVGGFSTPSQ)BOECVGGFSTEVSJOHBQ)DBMJCSBUJPOEFQFOEJOHPOUIFQ)CVGGFS
TFUUIBUJTTFMFDUFEJOUIFTFUVQNFOV%VSJOHBDBMJCSBUJPOUIFNFUFSVTFTUIFTFMFDUFECVGGFSTFUBOEUIFSBXN7
reading of the pH electrode in the buffer to recognize and display the buffer value at the measured temperature. The
SBXN7SFBEJOHPGUIFQ)FMFDUSPEFJOUIFCVGGFSNVTUCFBCPVUN7GSPNUIFUIFPSFUJDBMN7SFBEJOHPGUIFCVGGFS
in order for the meter to automatically recognize the buffer.
USA pH Buffer Set DIN pH Buffer Set
Buffer mV Range Buffer mV Range
 UPN7  UPN7
 UPN7  UPN7
 UPN7  UPN7
 UPN7  UPN7
 UPN7
Testing a pH Electrode for Automatic Buffer Recognition
5PWFSJGZUIBUUIFSBXN7SFBEJOHPGUIFQ)FMFDUSPEFJOVTFJTN7GSPNUIFUIFPSFUJDBMN7SFBEJOHPGUIFCVGGFS
and therefore verify that the pH electrode in use is capable of performing automatic buffer recognition, perform the
GPMMPXJOHQSPDFEVSF
1. Set the measurement mode to mV. Prepare the pH electrode according to the instructions in the pH electrode
user guide.
 3JOTFUIFQ)FMFDUSPEFXJUIEJTUJMMFEXBUFSCMPUJUESZXJUIBMJOUGSFFUJTTVFBOEQMBDFUIFQ)FMFDUSPEFJOUPBQ)
CVGGFSBUBQQSPYJNBUFMZ¡$
 8BJUGPSUIFNFBTVSFNFOUUPTUBCJMJ[FBOESFDPSEUIFN7WBMVFPGUIFQ)CVGGFSXIFOUIFNFUFSJOEJDBUFT
UIBUUIFNFBTVSFNFOUJTTUBCMF5IFSFBEUZQFTFMFDUFEJOUIFTFUVQNFOVXJMMEFUFSNJOFIPXUIFN7
measurements are displayed by the meter.
 3FNPWFUIFQ)FMFDUSPEFGSPNUIFQ)CVGGFS
 3JOTFUIFQ)FMFDUSPEFXJUIEJTUJMMFEXBUFSCMPUJUESZXJUIBMJOUGSFFUJTTVFBOEQMBDFUIFQ)FMFDUSPEFJOUPBQ)
CVGGFSBUBQQSPYJNBUFMZ¡$
 8BJUGPSUIFNFBTVSFNFOUUPTUBCJMJ[FBOESFDPSEUIFN7WBMVFPGUIFQ)CVGGFSXIFOUIFNFUFSJOEJDBUFT
that the measurement is stable.
 5IFN7SFBEJOHPGUIFQ)FMFDUSPEFJOQ)CVGGFSTIPVMECFJOUIFSBOHFPGUPN7BOEUIFN7
SFBEJOHPGUIFQ)FMFDUSPEFJOQ)CVGGFSTIPVMECFJOUIFSBOHFPGUPN7*GUIFN7SFBEJOHTPGUIF
pH electrode are in the correct ranges, the pH electrode is capable of performing automatic buffer recognition.
*GUIFN7SFBEJOHTPGUIFQ)FMFDUSPEFBSFOPUJOUIFDPSSFDUSBOHFTUIFWBMVFTPGUIFQ)CVGGFSTXJMMOFFEUPCF
manually entered during a pH calibration.
44 Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Electrode Condition Icon
The electrode condition icon indicates the performance of the pH electrode, based on the last saved calibration
and electrode measurement stability.
Icon pH Electrode Definition of Icon
&MFDUSPEFDPOEJUJPOJTHPPEBOEUIFFMFDUSPEFTMPQFJTUP
&MFDUSPEFDPOEJUJPOJTGBJSBOEUIFFMFDUSPEFTMPQFJTUP
&MFDUSPEFDPOEJUJPOJTCBEBOEUIFFMFDUSPEFTMPQFJTMFTTUIBOPSHSFBUFS
UIBO
$POTVMUUIFFMFDUSPEFVTFSHVJEFGPSJOTUSVDUJPOTPOIPXUPDMFBODPOEJUJPOBOE
troubleshoot the electrode.
45
Thermo Scientific Orion Star A211 Benchtop and Star A221 and Star A321 Portable pH Meter Reference Guide
Multipoint Calibration Curve Feature
5IF0SJPO4UBS"BOE4UBS"TFSJFTQ)NFUFSTVUJMJ[FBQPJOUUPQPJOUTDIFNFGPSNVMUJQPJOUDBMJCSBUJPOTJOUIFQ)
NPEF*OUIJTUFDIOJRVFTUSBJHIUMJOFTFHNFOUTBSFESBXOCFUXFFOFBDITVDDFTTJWFQBJSPGDBMJCSBUJPOQPJOUT4FQBSBUF
TMPQFBOE&PWBMVFTBSFDBMDVMBUFEVTJOHUIFBEKBDFOUQBJSPGEBUBQPJOUT'PSFYBNQMFJOBmWFQPJOUDBMJCSBUJPOGPVS
TMPQFTBOEmWF&PWBMVFTBSFDBMDVMBUFEUPDPSSFTQPOEUPUIFGPVSMJOFTFHNFOUTUIBUDPOOFDUUIFEBUBQ)WBMVFTGPS
TBNQMFTBSFDBMDVMBUFEBDDPSEJOHUPUIFTFHNFOUUIBUUIFJSQPUFOUJBMWBMVFTGBMMJOUP5IPTFBCPWFPSCFMPXUIFSBOHFPG
DBMJCSBUJPOBSFDBMDVMBUFECZFYUSBQPMBUJPOGSPNUIFUPQPSCPUUPNNPTUTFHNFOU
5IFEJTQMBZFETMPQFJTBOBWFSBHFTMPQFPGUIFTFQBSBUFTFHNFOUT5IFTMPQFNBZCFPVUTJEFUIFSBOHFOPSNBMMZFYQFDUFE
GPSBOFMFDUSPEF'PSFYBNQMFJGUIFTMPQFPGPOFTFHNFOUJTMPXUIFOUIFBWFSBHFXJMMCFMPXFSUIBOUIFPSFUJDBMFWFO
UIPVHIUIFSFTQPOTFNBZCF/FSOTUJBOPWFSNPTUPGUIFSBOHF
5IFmHVSFCFMPXJTBOFYBNQMFPGBUZQJDBMDBMJCSBUJPODVSWFVTJOHUIJTNFUIPE
This page intentionally left blank.
9 0 0 1 : 2 0 0 8
68X576701 RevA 10-11
Water Analysis Instruments
North America
166 Cummings Center
Beverly, MA 01915 USA
Toll Free: 1-800-225-1480
Tel: 1-978-232-6000
info.water@thermo.com
Netherlands
Tel: (31) 033-2463887
info.water.uk@thermo.com
India
Tel: (91) 22-4157-8800
wai.asia@thermofisher.com
Japan
Tel: (81) 045-453-9175
wai.asia@thermofisher.com
China
Tel: (86) 21-68654588
wai.asi@thermofisher.com
Singapore
Tel: (65) 6778-6876
wai.asia@thermofisher.com
Australia
Tel: (613) 9757-4300
in Australia (1300) 735-296
InfoWaterAU@thermofisher.com
www.thermoscientific.com/water
© 2011 Thermo Fisher Scientific Inc. All rights reserved.

Navigation menu