PhoCheck Tiger Select Instrument User Manual V1.1 Tiger_Select_V1_1 V1 1

User Manual: Pdf Tiger_Select_V1_1

Open the PDF directly: View PDF PDF.
Page Count: 19

DownloadPhoCheck Tiger Select Instrument User Manual V1.1 Tiger_Select_V1_1 V1 1
Open PDF In BrowserView PDF
PhoCheck

TIGER SELECT
Instrument User Manual V1.1

For Sales & Service Contact

2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242

1-800-833-7958

www.geotechenv.com

Register your instrument online to
receive your Extended 5 year
Warranty. See page 44 of the
Standard TIGER Manual for details.

rt Number: 861489

PhoCheck TIGER Select

Ion Science Ltd

About this Manual
This manual describes the function and operation of the TIGER Select
instrument.
It is a supplement to the standard Phocheck Tiger manual (Part number:
861265) and should be used together with the standard manual.
Please read and understand both user manuals completely before
operating the TIGER Select instrument.

Page 1 of 19

PhoCheck TIGER Select

Ion Science Ltd

Contents Page
About this Manual
Contents
Statements
Responsibility for use
Caution
Quality Assurance
Disposal
Calibration Facility
Ion Science Contact Details
Introduction to PhoCheck TIGER Select
Getting Started
Understanding the Display Screen
Start up screens
Soft Key areas
Using your TIGER Select
TIGER Select mode
TIGER PC Software
Maintenance
Calibration
Probes
TIGER SELECT Parts
Main Assembly
Front End Filter Assembly
Accessories
Instrument Warranty and Service
Technical Specification
Manual History

1
2
3
3
3
3
3
3
3
4
5
6
6
6
7
8-9
10
11-12
11
12
13-14
13
14
Please refer to Tiger Manual
Please refer to Tiger Manual
15
16

Page 2 of 19

PhoCheck TIGER Select

Ion Science Ltd

Statements
RESPONSIBILITY FOR USE
Inadequate performance of the gas detection equipment described in this manual may not necessarily
be self-evident and consequently equipment must be regularly inspected and maintained. Ion Science
recommends that personnel responsible for equipment use institute a regime of regular checks to
ensure it performs within calibration limits, and that a record be maintained which logs calibration check
data. The equipment should be used in accordance with this manual, and in compliance with local
safety standards.
LEGAL NOTICE
Whilst every attempt is made to ensure the accuracy of the information contained in this manual, Ion
Science accepts no liability for errors or omissions, or any consequences deriving from the use of
information contained herein. It is provided "as is" and without any representation, term, condition or
warranty of any kind, either express or implied. To the extent permitted by law, Ion Science shall not be
liable to any person or entity for any loss or damage which may arise from the use of this manual. We
reserve the right at any time and without any notice to remove, amend or vary any of the content which
appears herein.
!CAUTION!
It is essential that the PhoCheck TIGER is always used with a supplied PTFE 0.5 micron filter fitted to
the front of the instrument. Without a filter, particles of debris and dust can be drawn into the detector
inhibiting the function of the instrument. These filters are consumable and should be changed after
every 100 hours of use. The frequency of replacement should be increased for dusty or moisture
laden environments. Filters are available from your distributor or at www.ionscience.com .
Quality Assurance
PhoCheck TIGER has been manufactured in compliance with ISO9001:2000, which ensures that the
equipment supplied to our customers has been designed and assembled reproducibly, from traceable
components, and leaves Ion Science calibrated to stated standards.
Disposal
Dispose of PhoCheck TIGER, its components and any used batteries in accordance with all local and
national safety and environmental requirements. This includes the European WEEE (Waste Electrical
and Electronic Equipment) directive. Ion Science Ltd offers a take back service. Please contact us for
more information. The PhoCheck TIGER field case material is recyclable polypropylene.
Calibration Facility
Ion Science Ltd offers a calibration service including the issue of certification confirming calibration
with equipment traceable to national standards. A PhoCheck TIGER calibration kit is available from
your distributor or service centre or at www.ionscience.com . Ion Science recommends annual return
of all instruments for yearly service and calibration.
Ion Science Contact Details
UK Head Office
Ion Science Ltd
The Way,
Fowlmere
Cambridge
SG8 7UJ
UK
Tel: +44 (0)1763 207206
Fax: +44 (0) 1763 208814
Email: info@ionscience.com
Web: www.ionscience.com

USA Office
Ion Science Americas LLC
33 Commercial Drive
Waterbury
VT 05676
USA
Tel: +1 802 244 5153
Fax: +1 802 244 8942
Email: info@ionscienceamericas.com
Web: www.ionscience.com
Page 3 of 19

German Office
Ion Science Messtechnik
GMBH
Laubach 30
Metmann-Neandertal
40822
GERMANY
Tel: +49 2104 14480
Fax: +49 2104 144825
Email: info@ism-d.de
Web: www.ism-d.de

PhoCheck TIGER Select

Ion Science Ltd

Introduction to PhoCheck TIGER Select
Benzene vapour is carcinogenic. Benzene is a constituent of certain types of fuel, and used as a
solvent, and in the production of a wide range of chemicals and plastics.
PID technology can easily detect Benzene vapour as well as many of other volatile organic
compounds.
The Tiger Select uses a combination of a ultraviolet and chemical filtering to deliver a response to
Benzene free from interferents.
When the Select upgrade is activated, data logging, H&S mode and ppb automatically become
available.
At start up a specific Select screen appears however the instrument initially runs like a standard Tiger.
When in Select mode the instrument takes control over many of the standard features.
The Tiger Select uses a standard 10.6 eV lamp and a selective 10 eV filter incorporated in the
pellet/stack.
The lower eV output prevents sensing many of the gases co-present in Benzene containing
atmospheres.
10.0eV

10.6eV

The Tiger Select starts up in normal running mode.
Ion Science recommend operating the instrument in the working environment for 10 to 15 minutes
before use. This allows the tubes and instrument to acclimatize and ensures the best results.
Ion Science suggest that the TIGER Select is calibrated before each use and a new PID-Pre –Filter
Tube is used.
With no tube fitted, the customer can check the general VOC level, if there are no VOC's present then
no tube test is required.

Page 4 of 19

PhoCheck TIGER Select

Ion Science Ltd

Getting started
The TIGER Select has all the features and functions of the standard TIGER instrument with the
additional feature of Benzene detection using the PID-Pre-Filter Tube in TIGER Select mode..
This manual describes the operation of the TIGER in the TIGER Select mode indicated by the following
symbol.
When first turned on, the TIGER Select will run through a start up routine as described below in
‘Understanding the display screen’. The TIGER Select will then run in standard mode with all the
standard TIGER functions. All standard functions are described in the Standard TIGER manual (part
number 861265).
When in TIGER Select mode the instrument will default to the Benzene calibration.
PID-Pre-Filter Tube for Benzene.
The PID-Pre-Filter Tube is used in a specifically designed way with the TIGER Select. The instructions
supplied with the tubes do not affect the operation of the TIGER Select.
Please note:
1. If your TIGER Select has just been calibrated please ensure that a new PID-Pre-Filter Tube is
used when in TIGER Select mode.
2. A 10eV filter Pellet is fitted to the TIGER Select is used for Benzene and Isobutylene detection.
To revert to the full standard TIGER function with the whole range of gases the 10eV filter
should be replaced with a standard 10.6eV pellet. Ensure that the 10.6eV lamp is selected from
the Tiger PC Software. A custom calibration should then be carried out.

Page 5 of 19

PhoCheck TIGER Select

Ion Science Ltd

Understanding the display screen
Start up screens
Start up the instrument by pressing the ON/Off key
A sequence of screens will appear displaying the following:Screen 1:
Test screen where all fixed icons are shown
Screen 2:
The Ion Science logo is presented
Screen 3
The TIGER Select logo.
Screen 4:
Instrument identification; PhoCheck Tiger (default), IRN number, Firmware version
Screen 5:
Lamp strike routine. Animated dots indicate progress. When the lamp strikes, 'OK' will
appear

Soft key areas
The following icons will appear in the soft
key areas as soft key options. They are
selected using the UP and DOWN keys.

These icons can be reorganised using the
TIGER Select PC software (refer to
Standard TIGER manual).

Button A

Button B

Information

Options

Tiger Select mode

Average

Multiple data logging (optional)

Data log (optional)

Gas selection

Zone(s)

Zeroing

Sleep

Health & safety (optional)

Peak hold

Refer to the Standard TIGER manual for a description of all the soft key functions.
This manual now describes, in detail, the operation of the TIGER Select Function only.

Page 6 of 19

PhoCheck TIGER Select

Ion Science Ltd

Using your TIGER Select
INFORMATION
Pressing the Information soft key allows access to a number of other
information screens. Use the DOWN key to scroll through the screens. Press ESCAPE to return to the
main screen:
First Screen

Gas selected
Response Factor
Upper alarm
Lower alarm

Second screen

Lamp selected
Date of factory calibration
Date of custom calibration
Date of benzene calibration

Third Screen

SPAN 1

(gas concentration set in TIGER PC)

SPAN 2

(gas concentration set in TIGER PC)

PID detector
A/D reading
Temperature
Fourth Screen

STEL
TWA
Internal Reference Number (IRN)
Firmware version
Bootloader version
Battery type and status

Fifth Screen

Memory Status
Gas table date.
Date and time

Sixth Screen

Features. An icon will appear for each feature available on this instrument.

Page 7 of 19

PhoCheck TIGER Select

Ion Science Ltd

Using your TIGER Select
TIGER Select mode
Please note: If your TIGER Select has just been calibrated
please ensure that a new PID-Pre-Filter Tube is used. Ion Science suggest that the TIGER Select is
calibrated before each use.
Press the TIGER Select soft key to display the icon that
indicates that the PID-Pre-Filter should be inserted
(part number 5/FD-01). Do not insert the PID-Pre-Filter until
Both ends of the tube have been removed as below.
Remove both ends of the PID-Pre-Filter tube
with the Glass Tube Cutter supplied with your
TIGER Select (part number 5/FD-02).

Insert the PID-Pre-Filter tube into the glass tube cutter and turn the tube to cut the glass.

Unscrew the probe and insert the PID-Pre-Filter Tube. Now replace the probe.

Press the Enter button.
Page 8 of 19

PhoCheck TIGER Select

Ion Science Ltd

Using your TIGER Select (continued)
The TIGER Select will now take a sample over 150 seconds.
During this time the other soft keys will be disabled.

At the end of the sample the display will show the level of
Benzene detected in the in the location that the instrument
has been sampling.

To return to the main screen press Esc and confirm with the
Enter button.

An idication of the gas selected is for normal Tiger operation
is then shown.

If the Health & Safety option is enabled (via Tiger PC) the instrument
will run for a 15 minutes STEL after the initial Benzene test. The icon in
the bottom left corner will flash as indication to the user to press the
enter button to observe
the STEL routine.

At the end of the routine the final result will be displayed.

To escape the routine at any time press the Esc button and confirm
by pressing the Enter button.
IMPORTANT: Tiger Select STEL.
While the Tiger Select calculates the 15 minute STEL the instrument displays the ongoing STEL as it is
calculated. This ongoing calculation is for indication purposes only to allow the user to gauge how the
calculation is progressing. Only the final reading at the end of the calculation is logged by the
instrument and should be referenced by the user.

Page 9 of 19

PhoCheck TIGER Select

TIGER PC Software

Ion Science Ltd

________________________________________

Please Refer to the standard TIGER manual for the TIGER PC Software user information.

The TIGER Select Configuration screen will show the window below. The calibration default is 10 ppm
benzene.
To run the Health & Safety mode tick the
STEL calculation after the initial Benzne test.

box. This will then allow the option to run a 15 minute

Page 10 of 19

PhoCheck TIGER Select

Ion Science Ltd

Maintenance
CALIBRATION

Please Refer to the standard TIGER manual for Factory and
Customer calibration.

TIGER Select mode will operate against a benzene calibration.
Readings taken against unsatisfactory calibrations may be
unreliable.
For Benzene Calibration, first set up the parameters in TIGER
PC - see page 10.
TIGER Select allows you to custom calibrate using 10ppm
Benzene gas. Each cylinder should be regulated for a flow rate
above 250ml/min.
Note: Have on hand the cylinder(s) of gas, regulator(s) and PIDPre-Filter Benzene Tube before starting the procedure. Please
ensure you are familiar with the entire calibration procedure
before attempting to calibrate your TIGER Select.
Press the Options soft key
access the adjustable features.

on your TIGER Select to

Then use the UP or DOWN key to select calibration.
Press ENTER to confirm selection.
Select TIGER Select Calibration

and press Enter to confirm.

On confirming the selection, the user is presented with a prompt to
insert the PID-Pre-Filter Tube (part number 5/FD-01). Remove both ends
of the PID-Pre-Filter tube as described on page 9.

Now connect the tube from the gas cylinder to the end of the probe.

Now Press the Enter button. A screen will be shown with a countdown
In seconds.
Turn on the gas and press the enter button. The countdown will begin.
When the countdown has finished a tick will indicate that the calibration has finished. Press enter
to confirm the calibration.

Page 11 of 19

PhoCheck TIGER Select

Ion Science Ltd

Maintenance
Probes
Should the probe become cantaminated or damaged, replacement probes may be obtained from your
distributor or from Ion Science Ltd. Please note that a small ‘O ring’ at the base of the probe ensures
the probe is sealed, this can be seen in the transparent filter housing when the probe is removed.
The probe can be removed for cleaning or replacement
by unscrewing it from the transparent filter housing.
Refit the probe using fingers only, avoid using tools as
this may damage the filter housing. To ensure the
assembly is sealed, place a finger over the probe to block
the flow while the instrument is running. A flow alarm
should occur if sealed correctly.

Page 12 of 19

PhoCheck TIGER Select

Ion Science Ltd

TIGER Select Parts
TIGER Select Main Assembly

Page 13 of 19

PhoCheck TIGER Select

Ion Science Ltd

TIGER Select Parts
Front End Filter Assembly

Page 14 of 19

PhoCheck TIGER Select

Ion Science Ltd

Technical Specifications
Please Refer to the standard TIGER manual for the full specification of the Standard
TIGER instrument.
Benzene Detectable Range

0.01 – 40 ppm Benzene.

Page 15 of 19

PhoCheck TIGER Select

Ion Science Ltd

Manual History
Manual version
1.0

1.1

Amendment

Issue Date

First issue
Front cover issue updated to V1.1
Page 9, Health and Safety mode
added.
Page 10 Health and Safety mode
added.

Page 16 of 19

27/02/2011

8/04/2011

Instrument
Firmware
V 0.3.85

V0.3.93

PC
Software
V 1.0.0.42

V1.0.0.45

PhoCheck TIGER Select

Ion Science Ltd
This page is deliberately left blank

Page 17 of 19

PhoCheck TIGER Select

Ion Science Ltd
This page is deliberately left blank

Page 18 of 19



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Author                          : andrew
Create Date                     : 2011:04:15 10:44:04Z
Modify Date                     : 2011:07:22 13:41:41-06:00
Has XFA                         : No
Language                        : en-GB
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 4.2.1-c043 52.372728, 2009/01/18-15:56:37
Format                          : application/pdf
Creator                         : andrew
Title                           : PhoCheck Tiger Select Instrument User Manual V1.1
Creator Tool                    : Microsoft® Office Word 2007
Metadata Date                   : 2011:07:22 13:41:41-06:00
Producer                        : Microsoft® Office Word 2007
Document ID                     : uuid:1d40699a-3242-d14b-b2b8-fcb9a7486710
Instance ID                     : uuid:918985db-efd7-a74e-82f6-d2f9669ac3a4
Page Count                      : 19
EXIF Metadata provided by EXIF.tools

Navigation menu