إرشادات إدارة الغذاء والدواء (FDA) حول التطبيقات الطبية للأجهزة المحمولة UCM263366

User Manual: Pdf إرشادات إدارة الغذاء والدواء (FDA) حول التطبيقات الطبية للأجهزة المحمولة توفر التطبيقات الطبية للأجهزة ال٠ح٠ولة إ٠كانية الوصول الى اى ٠كان

Open the PDF directly: View PDF PDF.
Page Count: 45

Downloadإرشادات إدارة الغذاء والدواء (FDA) حول التطبيقات الطبية للأجهزة المحمولة UCM263366
Open PDF In BrowserView PDF
The 21st Century Cures Act (12/13/2016) amended the definition of “device” in
the Food, Drug and Cosmetic Act to exclude certain software functions,
including some described in this guidance document. FDA is assessing how to
revise this guidance to represent our current thinking on this topic. For
additional information, contact digitalhealth@fda.hhs.gov or refer to
https://www.fda.gov/MedicalDevices/DigitalHealth/default.htm.

Contains Nonbinding Recommendations

Mobile Medical Applications
__________________________
Guidance for Industry and Food
and Drug Administration Staff
Document issued on February 9, 2015.
This document supersedes “Mobile Medical Applications: Guidance for
Food and Drug Administration Staff” issued on September 25, 2013.
This document was updated to be consistent with the guidance document
“Medical Devices Data Systems, Medical Image Storage Devices, and Medical
Image Communications Devices” issued on February 9, 2015.
For questions about this document regarding CDRH-regulated devices, contact Bakul Patel at
301-796-5528 or by electronic mail at Bakul.Patel@fda.hhs.gov or contact the Office of the
Center Director at 301-796-5900.
For questions about this document regarding CBER-regulated devices, contact the Office of
Communication, Outreach and Development (OCOD), by calling 1-800-835-4709 or 240-4027800.

U.S. Department of Health and Human Services
Food and Drug Administration
Center for Devices and Radiological Health
Center for Biologics Evaluation and Research

-1-

Contains Nonbinding Recommendations

Preface
Public Comment
You may submit written comments and suggestions at any time for Agency consideration to
http://www.regulations.gov. Submit written comments to the Division of Dockets Management,
Food and Drug Administration, 5630 Fishers Lane, Room 1061, (HFA-305), Rockville, MD,
20852. Identify all comments with the docket number FDA-2011-D-0530. Comments may not
be acted upon by the Agency until the document is next revised or updated.

Additional Copies
CDRH
Additional copies are available from the Internet. You may also send an e-mail request to
CDRH-Guidance@fda.hhs.gov to receive a copy of the guidance. Please use the document
number (1741) to identify the guidance you are requesting.

CBER
Additional copies are available from the Center for Biologics Evaluation and Research (CBER)
by written request, Office of Communication, Outreach and Development (OCOD), 10903 New
Hampshire Ave., Bldg. 71, Room 3128, Silver Spring, MD 20903, or by calling 1-800-835-4709
or 240-402-7800, by e-mail, ocod@fda.hhs.gov, or from the Internet at
http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/
Guidances/default.htm.

-2-

Contains Nonbinding Recommendations

Table of Contents
I.

INTRODUCTION ............................................................................................................................................ 4

II.

BACKGROUND ............................................................................................................................................... 6

III.

DEFINITIONS .................................................................................................................................................. 7

A.
B.
C.
D.
E.

MOBILE PLATFORM .......................................................................................................................................... 7
MOBILE APPLICATION (MOBILE APP) .............................................................................................................. 7
MOBILE MEDICAL APPLICATION (MOBILE MEDICAL APP) .............................................................................. 7
REGULATED MEDICAL DEVICE ........................................................................................................................ 9
MOBILE MEDICAL APP MANUFACTURER ......................................................................................................... 9

IV.

SCOPE ............................................................................................................................................................. 12

V.

REGULATORY APPROACH FOR MOBILE MEDICAL APPS............................................................. 13
A.
B.

VI.

MOBILE MEDICAL APPS: SUBSET OF MOBILE APPS THAT ARE THE FOCUS OF FDA’S REGULATORY
OVERSIGHT ..................................................................................................................................................... 13
MOBILE APPS FOR WHICH FDA INTENDS TO EXERCISE ENFORCEMENT DISCRETION (MEANING THAT
FDA DOES NOT INTEND TO ENFORCE REQUIREMENTS UNDER THE FD&C ACT) ............................................. 15
REGULATORY REQUIREMENTS ............................................................................................................ 19

APPENDIX A

EXAMPLES OF MOBILE APPS THAT ARE NOT MEDICAL DEVICES........................ 20

APPENDIX B
EXAMPLES OF MOBILE APPS FOR WHICH FDA INTENDS TO EXERCISE
ENFORCEMENT DISCRETION ................................................................................................................. 23
APPENDIX C
EXAMPLES OF MOBILE APPS THAT ARE THE FOCUS OF FDA’S
REGULATORY OVERSIGHT (MOBILE MEDICAL APPS) .................................................................. 27
APPENDIX D

EXAMPLES OF CURRENT REGULATIONS ....................................................................... 30

APPENDIX E

BRIEF DESCRIPTION OF CERTAIN DEVICE REGULATORY REQUIREMENTS .... 33

APPENDIX F

FREQUENTLY ASKED QUESTIONS (FAQS) ...................................................................... 38

APPENDIX G

ADDITIONAL RESOURCES ................................................................................................... 43

-3-

Contains Nonbinding Recommendations

Mobile Medical Applications
Guidance for Industry and
Food and Drug Administration Staff
This guidance represents the Food and Drug Administration's (FDA's) current thinking on
this topic. It does not create or confer any rights for or on any person and does not operate to
bind FDA or the public. You can use an alternative approach if the approach satisfies the
requirements of the applicable statutes and regulations. If you want to discuss an alternative
approach, contact the FDA staff responsible for implementing this guidance. If you cannot
identify the appropriate FDA staff, call the appropriate number listed on the title page of this
guidance.

I.

Introduction

The Food and Drug Administration (FDA) recognizes the extensive variety of actual and
potential functions of mobile apps, the rapid pace of innovation in mobile apps, and the potential
benefits and risks to public health represented by these apps. The FDA is issuing this guidance
document to inform manufacturers, distributors, and other entities about how the FDA intends to
apply its regulatory authorities to select software applications intended for use on mobile
platforms (mobile applications or “mobile apps”). Given the rapid expansion and broad
applicability of mobile apps, the FDA is issuing this guidance document to clarify the subset of
mobile apps to which the FDA intends to apply its authority.
Many mobile apps are not medical devices (meaning such mobile apps do not meet the definition
of a device under section 201(h) of the Federal Food, Drug, and Cosmetic Act (FD&C Act)), and
FDA does not regulate them. Some mobile apps may meet the definition of a medical device but
because they pose a lower risk to the public, FDA intends to exercise enforcement discretion
over these devices (meaning it will not enforce requirements under the FD&C Act). The
majority of mobile apps on the market at this time fit into these two categories.
Consistent with the FDA’s existing oversight approach that considers functionality rather than
platform, the FDA intends to apply its regulatory oversight to only those mobile apps that are
medical devices and whose functionality could pose a risk to a patient’s safety if the mobile app
were to not function as intended. This subset of mobile apps the FDA refers to as mobile
medical apps.
FDA is issuing this guidance to provide clarity and predictability for manufacturers of mobile
medical apps. This document has been updated to be consistent with the guidance document
entitled “Medical Device Data Systems, Medical Image Storage Devices, and Medical Image
Communications Devices” issued on February 9, 2015.
-4-

Contains Nonbinding Recommendations
http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocu
ments/UCM401996.pdf. Examples on the Mobile Medical Apps Web site
http://www.fda.gov/medicaldevices/productsandmedicalprocedures/connectedhealth/mobilemedi
calapplications/default.htm which were added after September 25, 2013, were incorporated into
the appropriate appendices of this document for consistency.
FDA's guidance documents, including this guidance, do not establish legally enforceable
responsibilities. Instead, guidances describe the Agency's current thinking on a topic and should
be viewed only as recommendations, unless specific regulatory or statutory requirements are
cited. The use of the word should in Agency guidances means that something is suggested or
recommended, but not required.

-5-

Contains Nonbinding Recommendations

II. Background
As mobile platforms become more user friendly, computationally powerful, and readily
available, innovators have begun to develop mobile apps of increasing complexity to leverage
the portability mobile platforms can offer. Some of these new mobile apps are specifically
targeted to assisting individuals in their own health and wellness management. Other mobile
apps are targeted to healthcare providers as tools to improve and facilitate the delivery of patient
care.
In 1989, FDA prepared a general policy statement on how it planned to determine whether a
computer-based product and/or software-based product is a device, and, if so, how the FDA
intended to regulate it. The document, “FDA Policy for the Regulation of Computer Products,”
became known as the “Draft Software Policy.” After 1989, however, the use of computer and
software products as medical devices grew exponentially and the types of products diversified
and grew more complex (and that trend has continued). As a result, the FDA determined that the
draft policy did not adequately address all of the issues related to the regulation of all medical
devices containing software. Therefore, in 2005, the Draft Software Policy was withdrawn.1
Although the FDA has not issued an overarching software policy, the Agency has formally
classified certain types of software applications that meet the definition of a device and, through
classification, identified specific regulatory requirements that apply to these devices and their
manufacturers. These software devices include products that feature one or more software
components, parts, or accessories (such as electrocardiographic (ECG) systems used to monitor
cardiac rhythms), as well as devices that are composed solely of software (such as laboratory
information management systems). On February 15, 2011, the FDA issued a regulation downclassifying certain computer- or software-based devices intended to be used for the electronic
transfer, storage, display, and/or format conversion of medical device data – called Medical
Device Data Systems (MDDSs) – from Class III (high-risk) to Class I (low-risk).2
The FDA has previously clarified that when stand-alone software is used to analyze medical
device data, it has traditionally been regulated as an accessory to a medical device3 or as medical
device software.
As is the case with traditional medical devices, certain mobile medical apps can pose potential
risks to public health. Moreover, certain mobile medical apps may pose risks that are unique to
the characteristics of the platform on which the mobile medical app is run. For example, the
1

Annual Comprehensive List of Guidance Documents at the Food and Drug Administration (70 FR 824 at 890)
(January 5, 2005).
2
Medical Devices; Medical Device Data Systems Final Rule (76 FR 8637) (Feb. 15, 2011).
3
See, for example, Content of a 510(k) -http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourDevice/PremarketSubmissio
ns/PremarketNotification510k/ucm142651.htm (“Accessories to classified devices take on the same classification as
the "parent" device. An accessory such as software that accepts input from multiple devices usually takes on the
classification of the "parent" device with the highest risk, i.e., class.”); See also final Rule, Medical Devices,
Medical Device Data Systems, 76 FR 8637 at 8643-8644 – comment 16 and FDA’s response (Feb. 15, 2011).

-6-

Contains Nonbinding Recommendations
interpretation of radiological images on a mobile device could be adversely affected by the
smaller screen size, lower contrast ratio, and uncontrolled ambient light of the mobile platform.
FDA intends to take these risks into account in assessing the appropriate regulatory oversight for
these products.
This guidance clarifies and outlines the FDA’s current thinking. The Agency will continue to
evaluate the potential impact these technologies might have on improving health care, reducing
potential medical mistakes, and protecting patients.

III. Definitions
A. Mobile Platform
For purposes of this guidance, “mobile platforms” are defined as commercial off-the-shelf
(COTS) computing platforms, with or without wireless connectivity, that are handheld in nature.
Examples of these mobile platforms include mobile computers such as smart phones, tablet
computers, or other portable computers.

B. Mobile Application (Mobile App)
For purposes of this guidance, a mobile application or “mobile app” is defined as a software
application that can be executed (run) on a mobile platform (i.e., a handheld commercial off-theshelf computing platform, with or without wireless connectivity), or a web-based software
application that is tailored to a mobile platform but is executed on a server.

C. Mobile Medical Application (Mobile Medical App)
For purposes of this guidance, a “mobile medical app” is a mobile app that meets the definition
of device in section 201(h) of the Federal Food, Drug, and Cosmetic Act (FD&C Act) 4; and
either is intended:
·
·

to be used as an accessory to a regulated medical device; or
to transform a mobile platform into a regulated medical device.

4

Products that are built with or consist of computer and/or software components or applications are subject to
regulation as devices when they meet the definition of a device in section 201(h) of the FD&C Act. That provision
defines a device as “…an instrument, apparatus, implement, machine, contrivance, implant, in vitro reagent, or other
similar or related article, including any component, part, or accessory”, that is “… intended for use in the diagnosis
of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, in man …” or “…
intended to affect the structure or any function of the body of man or other animals ….” Thus, software
applications that run on a desktop computer, laptop computer, remotely on a website or “cloud,” or on a handheld
computer may be subject to device regulation if they are intended for use in the diagnosis or the cure, mitigation,
treatment, or prevention of disease, or to affect the structure or any function of the body of man. The level of
regulatory control necessary to assure safety and effectiveness varies based upon the risk the device presents to
public health. (See Appendix D for examples).

-7-

Contains Nonbinding Recommendations
The intended use of a mobile app determines whether it meets the definition of a “device.” As
stated in 21 CFR 801.4,5 intended use may be shown by labeling6 claims, advertising materials,
or oral or written statements by manufacturers or their representatives. When the intended use of
a mobile app is for the diagnosis of disease or other conditions, or the cure, mitigation, treatment,
or prevention of disease, or is intended to affect the structure or any function of the body of man,
the mobile app is a device.
One example is a mobile app that makes a light emitting diode (LED) operate. If the
manufacturer intends the system to illuminate objects generally (i.e., without a specific medical
device intended use), the mobile app would not be considered a medical device. If, however,
through marketing, labeling, and the circumstances surrounding the distribution, the mobile app
is promoted by the manufacturer for use as a light source for doctors to examine patients, then
the intended use of the light source would be similar to a conventional device such as an
ophthalmoscope.
In general, if a mobile app is intended for use in performing a medical device function (i.e. for
diagnosis of disease or other conditions, or the cure, mitigation, treatment, or prevention of
disease) it is a medical device, regardless of the platform on which it is run. For example,
mobile apps intended to run on smart phones to analyze and interpret EKG waveforms to detect
heart function irregularities would be considered similar to software running on a desktop
computer that serves the same function, which is regulated under 21 CFR 870.2340
(“Electrocardiograph”). FDA’s oversight approach to mobile apps is focused on their
functionality, just as we focus on the functionality of conventional devices. Our oversight is not
determined by the platform. Under this guidance, FDA would not regulate the sale or
general/conventional consumer use of smartphones or tablets. FDA’s oversight applies to mobile
apps performing medical device functions, such as when a mobile medical app transforms a
mobile platform into a medical device. However, as previously noted, we intend to apply this
oversight authority only to those mobile apps whose functionality could pose a risk to a patient’s
safety if the mobile app were to not function as intended.

5

“The words ‘intended uses’ or words of similar import … refer to the objective intent of the persons legally
responsible for the labeling of devices. The intent is determined by such persons' expressions or may be shown by
the circumstances surrounding the distribution of the article. This objective intent may, for example, be shown by
labeling claims, advertising matter, or oral or written statements by such persons or their representatives. It may be
shown by the circumstances that the article is, with the knowledge of such persons or their representatives, offered
and used for a purpose for which it is neither labeled nor advertised. The intended uses of an article may change
after it has been introduced into interstate commerce by its manufacturer. If, for example, a packer, distributor, or
seller intends an article for different uses than those intended by the person from whom he received the devices,
such packer, distributor, or seller is required to supply adequate labeling in accordance with the new intended uses.
But if a manufacturer knows, or has knowledge of facts that would give him notice that a device introduced into
interstate commerce by him is to be used for conditions, purposes, or uses other than the ones for which he offers it,
he is required to provide adequate labeling for such a device which accords with such other uses to which the article
is to be put.” 21 CFR 801.4.
6

“The term ‘labeling’ means all labels and other written, printed, or graphic matter (1) upon any article or any of its
containers or wrappers, or (2) accompanying such article.” Section 201(m) of the FD&C Act, 21 U.S.C. 321(m).

-8-

Contains Nonbinding Recommendations

D. Regulated Medical Device
For purposes of this guidance, a “regulated medical device” is defined as a product that meets the
definition of device in section 201(h) of the FD&C Act and that has been cleared or approved by
the FDA review of a premarket submission or otherwise classified by the FDA.
This definition can include novel devices, whether or not on a mobile platform, that the FDA will
clear or approve by the review of a premarket submission or otherwise classify. Examples of
regulated medical devices are identified in Appendix D.

E. Mobile Medical App Manufacturer
For purposes of this guidance, a “mobile medical app manufacturer” is any person or entity that
manufactures mobile medical apps in accordance with the definitions of manufacturer in 21 CFR
Parts 803, 806, 807, and 820.7 A mobile medical app manufacturer may include anyone who
initiates specifications, designs, labels, or creates a software system or application for a regulated
medical device in whole or from multiple software components. This term does not include
persons who exclusively distribute mobile medical apps without engaging in manufacturing
functions; examples of such distributors may include owners and operators of “Google play,”
“iTunes App store,” and “BlackBerry App World.” Examples of mobile medical app
manufacturers include any person or entity that:
·

Creates, designs, develops, labels, re-labels, remanufactures, modifies, or creates a
mobile medical app software system from multiple components. This could include a
person or entity that creates a mobile medical app by using commercial off the shelf
(COTS) software components and markets the product to perform as a mobile
medical app;

·

Initiates specifications or requirements for mobile medical apps or procures product
development/manufacturing services from other individuals or entities (second party)
for subsequent commercial distribution. For example, when a “developer” (i.e., an
entity that provides engineering, design, and development services) creates a mobile
medical app from the specifications that were initiated by the “author,” the “author”
who initiated and developed specifications for the mobile medical app is considered a
“manufacturer” of the mobile medical app under 21 CFR 803.3. For purposes of this
guidance, manufacturers of a mobile medical app would include persons or entities
who are the creators of the original idea (initial specifications) for a mobile medical

7

Regulatory definitions of the term “manufacturer” or “manufacture” appear in 21 CFR Parts 803, 806, 807, and
820. For example -- under FDA’s 21 CFR 807.3(d)-- establishment registration and device listing for manufacturers
and initial importers of devices-- ,“Manufacture, preparation, propagation, compounding, assembly, or processing
of a device means the making by chemical, physical, biological, or other procedures of any article that meets the
definition of device in section 201(h) of the act.” These terms include the following activities: (1) Repackaging or
otherwise changing the container, wrapper, or labeling of any device package in furtherance of the distribution of
the device from the original place of manufacture to the person who makes final delivery or sale to the ultimate
consumer; (2) Initial importation of devices manufactured in foreign establishments; or (3) Initiation of
specifications for devices that are manufactured by a second party for subsequent commercial distribution by the
person initiating specifications.”

-9-

Contains Nonbinding Recommendations
app, unless another entity assumes all responsibility for manufacturing and
distributing the mobile medical app, in which case that other entity would be the
“manufacturer.”8 Software “developers” of a mobile medical app that are only
responsible for performing design and development activities to transform the
author’s specifications into a mobile medical app would not constitute manufacturers,
and instead the author would be considered the manufacturer;
·

Creates a mobile medical app and hardware attachments for a mobile platform that
are intended to be used as a medical device by any combination of the mobile medical
app, hardware attachments, and the mobile platform;

·

Creates a mobile medical app or a software system that provides users access to the
medical device function through a website subscription, software as a service,9 or
other similar means.

In contrast, the following are examples of persons or entities that are NOT considered to be
mobile medical app manufacturers (i.e., persons not within the definition of manufacturer in 21
CFR Parts 803, 806, 807, and 820). Because they are not manufacturers, none of the persons or
entities in these examples would have to register their establishments, list their products with the
FDA10 or submit a premarket application:
·

Manufacturers or distributors of mobile platforms who solely distribute or market their
platform and do not intend (by marketing claims -- e.g., labeling claims or advertising
material) the platform to be used for medical device functions. When mobile medical
apps are run on a mobile platform, the mobile platform is treated as a component of the
mobile medical app’s intended use. 11 Therefore the mobile platform manufacturer is
exempt from the Quality System regulation and registration and listing requirements. 12
For example, if it is possible to run mobile medical apps on BrandNamePhone but
BrandNamePhone is not marketed by BrandNameCompany as intended for use as a
medical device, then BrandNameCompany would not be considered a mobile medical
app manufacturer or a medical device manufacturer. Also, in this example, the
BrandName Phone sold to consumers would not be regulated by FDA as a medical
device. FDA does not consider entities that exclusively distribute mobile medical apps,
such as the owners and operators of the “iTunes App store” or the “Android market,” to
be medical device manufacturers. FDA also does not consider mobile platform
manufacturers to be medical device manufacturer just because their mobile platform
could be used to run a mobile medical app regulated by FDA.

·

Third parties who solely provide market access to mobile medical apps (i.e. solely
distribute mobile apps), but do not engage in any manufacturing functions as defined in

8

See 21 CFR 803.3 (definition of manufacturer) and 21 CFR 807.20(a)(2).
By this we mean to include any “server software application” that provides a service to a client software
application on a mobile platform.
10
21 CFR 807.65 and 21 CFR 807.85
11
See 21 CFR 820.3(c) which defines a component as “any raw material, substance, piece, part, software, firmware,
labeling, or assembly which is intended to be included as part of the finished, packaged, and labeled device.”.
12
21 CFR 807.65(a) and 21 CFR 820.1(a).
9

- 10 -

Contains Nonbinding Recommendations
21 CFR Parts 803, 806, 807, and 820. Examples of such third parties may include
owners and operators that are only engaged in providing an online market place that
allow mobile medical app manufacturers to commercially distribute their mobile medical
apps. Specific examples of such online market places include “Google play,” “iTunes
store,” and “BlackBerry App World”;
·

Providers of tools, services or infrastructure used in the development, distribution, or use
of a mobile medical app. Examples include providers of internet connectivity (i.e.,
internet service), providers of general purpose computer or information technology,
providers that host the web service for content or software application. Other examples
of providers of tools, services, or infrastructure include customer support services, data
center hosting services, cloud hosting services, application hosting services, wireless
carriers, or providers of software development kits. However, a creator of a mobile
medical app or a software system that provides users access to the medical device
function through a website subscription, software as a service,13 or other similar means is
considered a mobile medical app manufacturer;

·

Licensed practitioners, including physicians, dentists, and optometrists, who manufacture
a mobile medical app or alter a mobile medical app solely for use in their professional
practice and do not label or promote their mobile medical apps to be generally used by
other licensed practitioners or other individuals.14,15 For example, if Dr. XYZ, a licensed
practitioner, creates a mobile medical app called the “XYZ-recorder” which enables
attaching an ECG electrode to a smartphone, and provides the “XYZ-recorder” to his/her
patient to use it to record the patient’s electrocardiographic readings for 24 hours, Dr.
XYZ is not considered a mobile medical app manufacturer. If Dr. XYZ is in a group
practice (including a telehealth network) and permits other physicians in the practice to
provide the XYZ-recorder to their patients, Dr. XYZ is not considered a mobile medical
apps manufacturer. However, if Dr. XYZ, the licensed practitioner, distributes the
“XYZ-recorder” and, through labeling or promotion intends to make it generally
available to or to be generally used by other physicians (or other specially qualified
persons), Dr. XYZ would be considered a mobile medical app manufacturer;

·

Persons who manufacture mobile medical apps solely for use in research, teaching, or
analysis and do not introduce such devices into commercial distribution. We note that
while persons conducting research using mobile medical apps involving human subjects
are exempt from registration and listing, they may instead be subject to investigational
device exemption regulations.16,17

13

See footnote 9.
Section 510(g)(2) of the FD&C Act: - Registration of producers of drugs or devices – Exclusions from
application of section: “practitioners licensed by law to prescribe or administer drugs or devices and who
manufacture, prepare, propagate, compound, or process drugs or devices solely for use in the course of their
professional practice.”
15
See 21 CFR 807.65(d).
16
See 21 CFR 807.65(f).
17
See 21 CFR 812.1.
14

- 11 -

Contains Nonbinding Recommendations

IV. Scope
This guidance explains the FDA’s intentions to focus its oversight on a subset of mobile apps.
Mobile medical apps as defined in section III include only those mobile apps that meet the
statutory definition of a device and either are intended:
·
·

to be used as an accessory to a regulated medical device; or
to transform a mobile platform into a regulated medical device.

Appendix A provides examples of mobile apps that FDA does NOT consider to meet the
definition of medical device and, therefore, are NOT mobile medical apps for the purposes of
this guidance.
Section V-B and Appendix B provide examples of mobile apps that MAY meet the definition of
a medical device but for which the FDA intends to exercise enforcement discretion because they
pose a low risk to patients.18
This guidance does not address the approach for software that performs patient-specific analysis
to aid or support clinical decision-making.
FDA’s policies regarding accessories to medical devices are not unique to mobile medical apps
and go beyond the scope of this guidance. Specifically this guidance does not address FDA’s
general approach for accessories to medical devices.
If you are developing a mobile medical app with an entirely new intended use, we encourage you
to contact FDA to discuss what regulatory requirements may apply.

18

This indicates that for certain mobile medical app devices, such as those in Appendix B, the FDA intends not to
pursue enforcement action for violations of the FD&C Act and applicable regulations by a manufacturer of a mobile
app that meets the definition of a device in section 201(h) of the FD&C Act as specified in this guidance. This does
not constitute a change in the requirements of the FD&C Act or any applicable regulation.

- 12 -

Contains Nonbinding Recommendations

V.

Regulatory approach for mobile medical apps

As described in this guidance, FDA intends to apply its regulatory oversight to only those mobile
apps that are medical devices and whose functionality could pose a risk to a patient’s safety if the
mobile app were to not function as intended. This approach to overseeing mobile medical apps is
consistent with our existing approach to overseeing medical device functionality of a product and
the risks it poses to patients regardless of the shape, size or the platform. The FDA believes that
this subset of mobile medical apps poses the same or similar potential risks to the public health
as currently regulated devices if they fail to function as intended.
The FDA strongly recommends that manufacturers of all mobile apps that may meet the
definition of a device follow the Quality System19 regulation (which includes good
manufacturing practices) in the design and development20 of their mobile medical apps and
initiate prompt corrections to their mobile medical apps, when appropriate, to prevent patient and
user harm.
For mobile medical apps, manufacturers must meet the requirements associated with the
applicable device classification. If the mobile medical app, on its own, falls within a medical
device classification, its manufacturer is subject to the requirements associated with that
classification. A mobile medical app, like other devices, may be classified as class I (general
controls), class II (special controls in addition to general controls), or class III (premarket
approval).21

A. Mobile medical apps: Subset of mobile apps that are the focus of
FDA’s regulatory oversight
Mobile apps may take a number of forms, but it is important to note that the FDA intends to
apply its regulatory oversight to only the subset of mobile apps identified below and in Appendix
C. These mobile apps can transform a mobile platform into a regulated medical device by using
attachments, display screens, sensors, or other such methods. Regardless of the mechanism
behind the transformation, FDA considers such mobile apps to be mobile medical apps.
The following are mobile apps that FDA considers to be mobile medical apps subject to
regulatory oversight:

19

See 21 CFR part 820.
The FDA has found that the majority of software-related device failures are due to design errors. In one study, the
most common problem was failure to validate software prior to routine production. See Medical Devices; Current
Good Manufacturing Practice (CGMP) Final Rule); Quality System Regulation (61 FR 52602) (October 7, 1996).
21
See footnotes 3 and 4.
20

- 13 -

Contains Nonbinding Recommendations
1. Mobile apps that are an extension of one or more medical devices by connecting22
to such device(s) for purposes of controlling23 the device(s) or for use in active
patient monitoring or analyzing medical device data.
o Examples of displays of patient-specific medical device data include: display of
medical images directly from a Picture Archiving and Communication System
(PACS) server and remote display of data from bedside monitors (note that
mobile medical apps that display medical device data to perform active patient
monitoring such as bedside monitors are subject to regulations associated with
such devices) ..
o Examples of mobile apps that control medical devices include: apps that provide
the ability to control inflation and deflation of a blood pressure cuff through a
mobile platform and mobile apps that control the delivery of insulin on an
insulin pump by transmitting control signals to the pumps from the mobile
platform.
Mobile medical apps of this type are considered an accessory to the connected
device and are required to comply with the controls applicable to that connected
device. The FDA considers such mobile medical apps to extend the intended use
and functionality of the connected medical device. As a result, the mobile
medical app would be required to comply with the regulations applicable to the
connected medical device in order to address any associated risks.
2. Mobile apps that transform the mobile platform into a regulated medical device
by using attachments, display screens, or sensors or by including functionalities
similar to those of currently regulated medical devices. Mobile apps that use
attachments, display screens, sensors or other such similar components to
transform a mobile platform into a regulated medical device are required to
comply with the device classification associated with the transformed platform.
o Examples of these types of mobile apps include: a mobile app that uses a mobile
platform for medical device functions, such as attachment of a blood glucose
strip reader to a mobile platform to function as a glucose meter; or attachment
of electrocardiograph (ECG) electrodes to a mobile platform to measure, store,
and display ECG signals; a mobile app that uses the built-in accelerometer on a
mobile platform to collect motion information for monitoring sleep apnea; a
mobile app that uses sensors (internal or external) on a mobile platform for
creating electronic stethoscope function is considered to transform the mobile
platform into an electronic stethoscope; manufacturers of such a mobile app are
required to follow the requirements of 21 CFR 870.1875(b) (Electronic
Stethoscope); and similarly a mobile app that displays radiological images for
diagnosis transforms the mobile platform into a class II Picture Archiving and
Communications System (PACS) under 21 CFR 892.2050.
22

To meet this criterion, the mobile medical apps need not be physically connected to the regulated medical device
(i.e. the connection can be wired or wireless).
23
Controlling the intended use, function, modes, or energy source of the connected medical device.

- 14 -

Contains Nonbinding Recommendations
The FDA has cleared several mobile medical apps with attachments to a mobile
platform. Specifically, patient monitoring mobile apps that monitors a patient
for heart rate variability from a signal produced by an electrocardiograph,
vectorcardiograph, or blood pressure monitor are classified as cardiac
monitoring software under 21 CFR 870.2300 (Cardiac monitor). Other mobile
medical apps that use a hardware attachment or interface to a monitoring system
that have been cleared include an automatic electronic blood pressure monitor
under 21 CFR 870.1130 and a perinatal monitoring system under 21 CFR
884.2740.
3. Mobile apps that become a regulated medical device (software) by performing
patient-specific analysis and providing patient-specific diagnosis, or treatment
recommendations. These types of mobile medical apps are similar to or perform
the same function as those types of software devices that have been previously
cleared or approved.
o Examples of mobile apps that perform sophisticated analysis or interpret data
(electronically collected or manually entered) from another medical device
include: apps that use patient-specific parameters and calculate dosage or create
a dosage plan for radiation therapy; Computer Aided Detection software (CAD)
image processing software24; and radiation therapy treatment planning software
25
. We believe that these types of software present the same level of risk to
patients regardless of the platform on which they run.
The FDA encourages manufacturers of such mobile medical apps that perform patient-specific
analysis to contact FDA to discuss what, if any, regulatory requirements may apply to their
mobile app. For additional examples see 0.

B. Mobile Apps for which FDA intends to exercise enforcement
discretion (meaning that FDA does not intend to enforce
requirements under the FD&C Act)
FDA intends to exercise enforcement discretion for mobile apps that:
·
·
·
·
·

24
25

Help patients (i.e., users) self-manage their disease or conditions without providing
specific treatment or treatment suggestions;
Provide patients with simple tools to organize and track their health information;
Provide easy access to information related to patients’ health conditions or treatments;
Help patients document, show, or communicate potential medical conditions to health
care providers;
Automate simple tasks for health care providers;

21 CFR 892.2050.
21 CFR 892.5050.

- 15 -

Contains Nonbinding Recommendations
·
·

Enable patients or providers to interact with Personal Health Record (PHR) or Electronic
Health Record (EHR) systems; or
Intended to transfer, store, convert format, and display medical device data in its original
format from a medical device (as defined by MDDS regulation 880.6310 OUG).

Some mobile apps in the above categories and listed below may be considered mobile medical
apps, and others might not. For those mobile apps listed below that are devices, FDA intends to
exercise enforcement discretion because they pose a low risk to patients.
The following examples represent mobile apps for which the FDA intends to exercise
enforcement discretion:
1. Mobile apps that provide or facilitate supplemental clinical care, by coaching or
prompting, to help patients manage their health in their daily environment. – These are
apps that supplement26 professional clinical care by facilitating behavioral change or
coaching patients with specific diseases or identifiable health conditions in their daily
environment. Examples include:
o Apps that coach patients with conditions such as cardiovascular disease,
hypertension, diabetes or obesity, and promote strategies for maintaining a
healthy weight, getting optimal nutrition, exercising and staying fit, managing salt
intake, or adhering to pre-determined medication dosing schedules27 by simple
prompting.
2. Mobile apps that provide patients with simple tools to organize and track their health
information – These are apps that provide patients with tools28 to organize and track health
information without providing recommendations to alter or change a previously prescribed
treatment or therapy. Examples include:
o Apps that provide simple tools for patients with specific conditions or chronic
disease (e.g., obesity, anorexia, arthritis, diabetes, heart disease) to log, track, or
trend their events or measurements (e.g., blood pressure measurements, drug
intake times, diet, daily routine or emotional state) and share this information with
their health care provider as part of a disease-management plan.

26

By this we mean that the app can be safely used by a patient without active oversight by a medical professional
and, when used for serious conditions necessitating professional medical care, use of the app is not intended to
replace or discourage seeking treatment from a health care provider.
27
We consider these mobile medical apps as “medication reminders – Product code NXQ” currently defined as “A
medication reminder is a device intended for medical purposes to provide alerts to patients or healthcare providers
for pre-determined medication dosing schedules. The device may incorporate wireless communication.” The FDA
intends to exercise enforcement discretion for this specific product code (NXQ) identified under 21 CFR 890.5050 –
Daily activity assist device.
28
We consider these mobile apps to be simple tools which are not intended to provide specific treatment
recommendations. For such simple tools, even when exceeding the limitations of 510(k) exemptions referred to in
21 CFR 8XX.9, FDA does not intend to enforce compliance with the regulatory controls. For example, to the extent
that these limitations apply, FDA does not intend to enforce compliance with regulatory controls for these simple
tools, in vitro devices that are intended to be used in diabetes management (21 CFR 8xx.9(c)(5)) or for assessing
cardiovascular disease (21 CFR 8xx.9(c)(4)).

- 16 -

Contains Nonbinding Recommendations

3. Mobile apps that provide easy access to information related to patients’ health
conditions or treatments (beyond providing an electronic “copy” of a medical
reference) – These are apps that provide contextually-relevant information to users by
matching patient-specific information (e.g., diagnosis, treatments, allergies, signs or
symptoms) to reference information routinely used in clinical practice29 (e.g., practice
guidelines) to facilitate a user’s assessment of a specific patient. Examples include:
o Apps that use a patient’s diagnosis to provide a clinician with best practice
treatment guidelines for common illnesses or conditions such as influenza;
o Apps that are drug-drug interaction or drug-allergy look-up tools.

4. Mobile apps that are specifically marketed to help patients document, show, or
communicate to providers potential medical conditions – These are apps that in their
labeling or promotional materials are not promoted for medical uses but which, by virtue of
other circumstances surrounding their distribution, may meet the definition of a medical
device. These products either pose little or no risk, or are the sole responsibility of the health
care providers who have used them in medical applications. Examples include:
o Apps that serve as videoconferencing portals specifically intended for medical use
and to enhance communications between patients, healthcare providers, and
caregivers;
o Apps specifically intended for medical uses that utilize the mobile device’s builtin camera or a connected camera for purposes of documenting or transmitting
pictures (e.g., photos of a patient’s skin lesions or wounds) to supplement or
augment what would otherwise be a verbal description in a consultation between
healthcare providers or between healthcare providers and patients/caregivers.
5. Mobile apps that perform simple calculations routinely used in clinical practice – These
are apps that are intended to provide a convenient way for clinicians to perform various
simple medical calculations taught in medical schools30 and are routinely used in clinical
practice. These apps are generally tailored for clinical use, but retain functionality that is
similar to simple general purpose tools such as paper charts, spread sheets, timers or generic
mathematical calculators. Examples of such general purpose tools include medical
calculators for:
o Body Mass Index (BMI)
o Total Body Water / Urea Volume of Distribution
o Mean arterial pressure
o Glascow Coma Scale score
o APGAR score
o NIH Stroke Scale
o Delivery date estimator

29

The type of information provided in these apps is from authoritative medical sources, as recognized by the field or
discipline that is the subject of the app.
30
The types of information in these calculators are available in medical sources which includes medical textbooks
used in the curriculum of accredited medical schools.

- 17 -

Contains Nonbinding Recommendations

6. Mobile apps that enable individuals to interact with PHR systems or EHR systems -These are apps that provide patients and providers with mobile access to health record
systems or enables them to gain electronic access to health information stored within a PHR
system or EHR system. Applications that only allow individuals to view or download EHR
data are also included in this category. These mobile apps are generally meant to facilitate
general patient health information management and health record-keeping activities.
7. Mobile apps that meet the definition of Medical Device Data Systems – These are apps
that are intended to transfer, store, convert format, and display medical device data, without
controlling or altering the functions or parameters of any connected medical device, as
defined in the MDDS classification regulation (21 CFR 880.6310). These mobile apps
include those that are used as a secondary display to a regulated medical device when these
apps are not intended to provide primary diagnosis, treatment decisions, or to be used in
connection with active patient monitoring (i.e., mobile apps that meet the MDDS definition).

See Appendix B for additional examples for the above categories.

- 18 -

Contains Nonbinding Recommendations

VI. Regulatory requirements
This guidance, including 0 and existing medical device regulatory classifications in Appendix D,
is intended to assist manufacturers in determining if a product is a mobile medical app and
FDA’s expectations for that product. Additional information can be found in “Device Advice:
Classify Your Medical Device” at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/Overview/ClassifyYourDev
ice/default.htm. This section describes in greater detail the regulatory requirements applicable to
mobile medical apps under this guidance (as described in Section V).
Manufacturers of mobile medical apps are subject to the requirements described in the applicable
device classification regulation below. Depending on the classification and the associated
regulation for the mobile medical apps, manufacturers of mobile medical apps are required to
follow associated controls established by the regulation.
In general, the associated controls for each class of device are outlined below.
Class I devices: General Controls, including:
·
·
·
·
·
·
·

Establishment registration, and Medical Device listing (21 CFR Part 807);
Quality System (QS) regulation (21 CFR Part 820);
Labeling requirements (21 CFR Part 801);
Medical Device Reporting (21 CFR Part 803);
Premarket notification (21 CFR Part 807);
Reporting Corrections and Removals (21 CFR Part 806); and
Investigational Device Exemption (IDE) requirements for clinical studies of
investigational devices (21 CFR Part 812).

Class II devices: General Controls (as described for Class I), Special Controls, and (for most
Class II devices) Premarket Notification.
Class III devices: General Controls (as described for Class I), and Premarket Approval (21 CFR
Part 814).
Appendix E provides a brief summary of the above requirements. Additional information is
available at http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/default.htm,
under “Overview of Medical Device Regulation” and “How to Market Your Device.”
If you need further assistance, you may contact the Division of Industry and Consumer
Education: Email: DICE@fda.hhs.gov; phone: 301-796-7100 or 800-638-2041.

- 19 -

Contains Nonbinding Recommendations

Appendix A Examples of mobile apps that are NOT
medical devices
This Appendix provides a representative list of mobile app functionalities to illustrate the types
of mobile apps that could be used in a healthcare environment, in clinical care or patient
management, but are not considered medical devices. Because these mobile apps are not
considered medical devices, FDA does not regulate them. The FDA understands that there may
be other unique and innovative mobile apps that may not be covered in this list that may also
constitute healthcare related mobile apps. This list is not exhaustive; it is only intended to
provide clarity and assistance in identifying when a mobile app is not considered to be a medical
device.
Specific examples of mobile apps that FDA does not consider to be devices and with no
regulatory requirements under the current laws administered by FDA include:
1. Mobile apps that are intended to provide access to electronic “copies” (e.g., e-books, audio
books) of medical textbooks or other reference materials with generic text search capabilities.
These are not devices because these apps are intended to be used as reference materials and
are not intended for use in the diagnosis of disease or other conditions, or in the cure,
mitigation, treatment, or prevention of disease by facilitating a health professional’s
assessment of a specific patient, replacing the judgment of clinical personnel, or performing
any clinical assessment. Examples include mobile apps that are:
o Medical dictionaries;
o Electronic copies of medical textbooks or literature articles such as the Physician’s
Desk Reference or Diagnostic and Statistical Manual of Mental Disorders (DSM);
o Library of clinical descriptions for diseases and conditions;
o Encyclopedia of first-aid or emergency care information;
o Medical abbreviations and definitions;
o Translations of medical terms across multiple languages.
2. Mobile apps that are intended for health care providers to use as educational tools for
medical training or to reinforce training previously received. These may have more
functionality than providing an electronic copy of text (e.g., videos, interactive diagrams),
but are not devices because they are intended generally for user education and are not
intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation,
treatment, or prevention of disease by facilitating a health professional’s assessment of a
specific patient, replacing the judgment of clinical personnel, or performing any clinical
assessment. Examples include mobile apps that are:
o Medical flash cards with medical images, pictures, graphs, etc.;
o Question/Answer quiz apps;
o Interactive anatomy diagrams or videos;
o Surgical training videos;
o Medical board certification or recertification preparation apps;
o Games that simulate various cardiac arrest scenarios to train health professionals in
advanced CPR skills.

- 20 -

Contains Nonbinding Recommendations
3. Mobile apps that are intended for general patient education and facilitate patient access to
commonly used reference information. These apps can be patient-specific (i.e., filters
information to patient-specific characteristics), but are intended for increased patient
awareness, education, and empowerment, and ultimately support patient-centered health care.
These are not devices because they are intended generally for patient education, and are not
intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation,
treatment, or prevention of disease by aiding clinical decision-making (i.e., to facilitate a
health professional’s assessment of a specific patient, replace the judgment of a health
professional, or perform any clinical assessment). Examples include mobile apps that:
o Provide a portal for healthcare providers to distribute educational information (e.g.,
interactive diagrams, useful links and resources) to their patients regarding their
disease, condition, treatment or up-coming procedure;
o Help guide patients to ask appropriate questions to their physician relevant to their
particular disease, condition, or concern;
o Provide information about gluten-free food products or restaurants;
o Help match patients with potentially appropriate clinical trials and facilitate
communication between the patient and clinical trial investigators;
o Provide tutorials or training videos on how to administer first-aid or CPR;
o Allow users to input pill shape, color or imprint and displays pictures and names of
pills that match this description;
o Find the closest medical facilities and doctors to the user’s location;
o Provide lists of emergency hotlines and physician/nurse advice lines;
o Provide and compare costs of drugs and medical products at pharmacies in the user’s
location.

4. Mobile apps that automate general office operations in a health care setting and are not
intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation,
treatment, or prevention of disease. Examples include mobile apps that:
o Determine billing codes like ICD-9 (international statistical classification of
diseases);
o Enable insurance claims data collection and processing and other apps that are
similarly administrative in nature;
o Analyze insurance claims for fraud or abuse;
o Perform medical business accounting functions or track and trend billable hours and
procedures;
o Generate reminders for scheduled medical appointments or blood donation
appointments;
o Help patients track, review and pay medical claims and bills online;
o Manage shifts for doctors;
o Manage or schedule hospital rooms or bed spaces;
o Provide wait times and electronic check-in for hospital emergency rooms and urgent
care facilities;
o Allow healthcare providers or staff in healthcare settings to process payments (for
example, using a HIPAA compliant app to process payments);
o Track or perform patient satisfaction survey after an encounter or a clinical visit.

- 21 -

Contains Nonbinding Recommendations

5. Mobile apps that are generic aids or general purpose products. These apps are not considered
devices because they are not intended for use in the diagnosis of disease or other conditions,
or in the cure, mitigation, treatment, or prevention of disease. Examples include mobile apps
that:
o Use the mobile platform as a magnifying glass (but are not specifically intended for
medical purposes31);
o Use the mobile platform for recording audio, note-taking, replaying audio with
amplification, or other similar functionalities;
o Allow patients or healthcare providers to interact through email, web-based
platforms, video or other communication mechanisms (but are not specifically
intended for medical purposes);
o Provide maps and turn-by-turn directions to medical facilities;
o Allow health care providers to communicate in a secure and protected method (for
example using a HIPAA compliant app to send messages between health care
providers in a hospital).

31

Medical purpose magnifiers are regulated either under 21 CFR 886.5840 - Magnifying spectacles (“devices that
consist of spectacle frames with convex lenses intended to be worn by a patient who has impaired vision to enlarge
images”), or under 21 CFR 886.5540 - Low-vision magnifiers (“a device that consists of a magnifying lens intended
for use by a patient who has impaired vision. The device may be held in the hand or attached to spectacles”).

- 22 -

Contains Nonbinding Recommendations

Appendix B Examples of mobile apps for which FDA
intends to exercise enforcement discretion
This Appendix provides examples of mobile apps that MAY meet the definition of medical
device but for which FDA intends to exercise enforcement discretion. These mobile apps may be
intended for use in the diagnosis of disease or other conditions, or in the cure, mitigation,
treatment, or prevention of disease. Even though these mobile apps MAY meet the definition of
medical device, FDA intends to exercise enforcement discretion for these mobile apps because
they pose lower risk to the public.
The FDA understands that there may be other unique and innovative mobile apps that may not be
covered in this list that may also constitute healthcare related mobile apps. This list is not
exhaustive; it is only intended to provide clarity and assistance in identifying the mobile apps
that will not be subject to regulatory requirements at this time.
·

Mobile apps that help patients with diagnosed psychiatric conditions (e.g., post-traumatic
stress disorder (PTSD), depression, anxiety, obsessive compulsive disorder) maintain
their behavioral coping skills by providing a “Skill of the Day” behavioral technique or
audio messages that the user can access when experiencing increased anxiety;

·

Mobile apps that provide periodic educational information, reminders, or motivational
guidance to smokers trying to quit, patients recovering from addiction, or pregnant
women;

·

Mobile apps that use GPS location information to alert asthmatics of environmental
conditions that may cause asthma symptoms or alert an addiction patient (substance
abusers) when near a pre-identified, high-risk location;

·

Mobile apps that use video and video games to motivate patients to do their physical
therapy exercises at home;

·

Mobile apps that prompt a user to enter which herb and drug they would like to take
concurrently and provide information about whether interactions have been seen in the
literature and a summary of what type of interaction was reported;

·

Mobile apps that help asthmatics track inhaler usage, asthma episodes experienced,
location of user at the time of an attack, or environmental triggers of asthma attacks;

·

Mobile apps that prompt the user to manually enter symptomatic, behavioral or
environmental information, the specifics of which are pre-defined by a health care
provider, and store the information for later review;

·

Mobile apps that use patient characteristics such as age, sex, and behavioral risk factors
to provide patient-specific screening, counseling and preventive recommendations from
well-known and established authorities;

- 23 -

Contains Nonbinding Recommendations
·

Mobile apps that use a checklist of common signs and symptoms to provide a list of
possible medical conditions and advice on when to consult a health care provider;

·

Mobile apps that guide a user through a questionnaire of signs and symptoms to provide a
recommendation for the type of health care facility most appropriate to their needs;

·

Mobile apps that record the clinical conversation a clinician has with a patient and sends
it (or a link) to the patient to access after the visit;

·

Mobile apps that are intended to allow a user to initiate a pre-specified nurse call or
emergency call using broadband or cellular phone technology;

·

Mobile apps that enable a patient or caregiver to create and send an alert or general
emergency notification to first responders;

·

Mobile apps that keep track of medications and provide user-configured reminders for
improved medication adherence;

·

Mobile apps that provide patients a portal into their own health information, such as
access to information captured during a previous clinical visit or historical trending and
comparison of vital signs (e.g., body temperature, heart rate, blood pressure, or
respiratory rate);

·

Mobile apps that aggregate and display trends in personal health incidents (e.g.,
hospitalization rates or alert notification rates);

·

Mobile apps that allow a user to collect (electronically or manually entered) blood
pressure data and share this data through e-mail, track and trend it, or upload it to a
personal or electronic health record;

·

Mobile apps that provide oral health reminders or tracking tools for users with gum
disease;

·

Mobile apps that provide prediabetes patients with guidance or tools to help them
develop better eating habits or increase physical activity;

·

Mobile apps that display, at opportune times, images or other messages for a substance
abuser who wants to stop addictive behavior;

- 24 -

Contains Nonbinding Recommendations
·

Mobile apps32 that are intended for individuals to log, record, track, evaluate, or make
decisions or behavioral suggestions related to developing or maintaining general fitness,
health or wellness, such as those that:
o Provide tools to promote or encourage healthy eating, exercise, weight loss or other
activities generally related to a healthy lifestyle or wellness;
o Provide dietary logs, calorie counters or make dietary suggestions;
o Provide meal planners and recipes;
o Track general daily activities or make exercise or posture suggestions;
o Track a normal baby’s sleeping and feeding habits;
o Actively monitor and trend exercise activity;
o Help healthy people track the quantity or quality of their normal sleep patterns;
o Provide and track scores from mind-challenging games or generic “brain age” tests;
o Provide daily motivational tips (e.g., via text or other types of messaging) to reduce
stress and promote a positive mental outlook;
o Use social gaming to encourage healthy lifestyle habits;
o Calculate calories burned in a workout.

·

Mobile apps that transfer, store, convert formats, and display medical device data without
modifying the data and do not control or alter the functions or parameters of any
connected medical device (i.e., mobile apps that meet the definition of MDDS under 21
CFR 880.6310).

·

Mobile apps that meet the definition of MDDS and connect to a nursing central station
and display medical device data to a physician’s mobile platform for review. Product
code: OUG (21 CFR 880.6310).

·

Mobile apps that are not intended for diagnostic image review such as image display for
multidisciplinary patient management meetings (e.g., rounds) or patient consultation (and
include a persistent on-screen notice, such as “for informational purposes only and not
intended for diagnostic use”). Such devices would be considered medical image
communications devices under 21 CFR 892.2020, product code LMD,

·

Mobile apps for providers that help track or manage patient immunizations by assessing
the need for immunization, consent form, and immunization lot number;

·

Mobile apps that provide drug-drug interactions and relevant safety information (side
effects, drug interactions, active ingredient) as a report based on demographic data (age,
gender), clinical information (current diagnosis), and current medications;

32

When these items are not marketed, promoted or intended for use in the diagnosis of disease or other conditions,
or in the cure, mitigation, treatment, or prevention of disease, or do not otherwise meet the definition of medical
device, FDA does not regulate them. When they are marketed, promoted or intended for use in the diagnosis of
disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, or otherwise meet the
definition of medical device, FDA intends to exercise enforcement discretion.

- 25 -

Contains Nonbinding Recommendations
·

Mobile apps that enable, during an encounter, a health care provider to access their
patient’s personal health record (health information) that is hosted on a web-based or
other platform;

·

Mobile apps that allow a user to, collect, log, track and trend data, such as blood glucose,
blood pressure, heart rate, weight or other data from a device to eventually share with a
heath care provider, or upload it to an online (cloud) database, personal or electronic
health record.

- 26 -

Contains Nonbinding Recommendations

Appendix C Examples of mobile apps that are the focus of
FDA’s regulatory oversight (mobile medical
apps)
This Appendix provides examples of mobile apps that are considered medical devices (i.e.,
mobile medical apps), on which FDA will focus its regulatory oversight. These mobile apps meet
the definition of medical device in the FD&C Act and their functionality poses a risk to a
patient’s safety if the mobile app were to not function as intended. Each example below
provides a list of possible relevant product code(s) and/or regulation number.
FDA also encourages mobile medical app manufacturers to search FDA’s public databases, such
as the “Product Classification” database and the “510(k) Premarket Notification” database, to
determine the level of regulation for a given device and for the most up-to-date information
about the relevant regulatory requirements. These databases can be accessed through the
following links:
(http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPCD/classification.cfm) and
(http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfPMN/pmn.cfm.)

Mobile apps that transform a mobile platform into a regulated medical device and
therefore are mobile medical apps: These mobile apps use a mobile platform’s built-in features
such as light, vibrations, camera, or other similar sources to perform medical device functions
(e.g., mobile medical apps that are used by a licensed practitioner to diagnose or treat a disease).
Possible product codes: Varies depending on the intended use and function of the mobile medical
app; see additional examples below.
·

Mobile apps that use a sensor or lead that is connected to a mobile platform to measure
and display the electrical signal produced by the heart (electrocardiograph or ECG).
Possible product code(s): DPS, MLC, OEY (21 CFR 870.2340), MLO, MWJ (21 CFR
870.2800).

·

Mobile apps that use a sensor or electrode attached to the mobile platform or tools within
the mobile platform itself (e.g., microphone and speaker) to electronically amplify and
“project sounds associated with the heart, arteries and veins and other internal organs”
(i.e., an electronic stethoscope). Possible product code: DQD (21 CFR 870.1875(b)).

·

Mobile apps that use a sensor or electrode attached to the mobile platform or tools within
the mobile platform itself (e.g., accelerometer) to measure physiological parameters
during cardiopulmonary resuscitation (CPR) and give feedback about the quality of CPR
being delivered. Possible product code: LIX (21 CFR 870.5200).

·

Mobile apps that use a sensor attached to the mobile platform or tools within the mobile
platform itself to record, view, or analyze eye movements for use in the diagnosis of
balance disorders (i.e., nystagmograph). Possible product code: GWN (21 CFR
882.1460).

- 27 -

Contains Nonbinding Recommendations
·

Mobile apps that use tools within the mobile platform (e.g., speaker) to produce
controlled levels of test tones and signals intended for use in conducting diagnostic
hearing evaluations and assisting in the diagnosis of possible otologic disorders (i.e., an
audiometer). Possible product code: EWO (21 CFR 874.1050).

·

Mobile apps that use a sensor attached to the mobile platform or tools within the mobile
platform itself (e.g., accelerometer) to measure the degree of tremor caused by certain
diseases (i.e., a tremor transducer). Possible product code: GYD (21 CFR 882.1950).

·

Mobile apps that use a sensor attached to the mobile platform or tools within the mobile
platform itself (e.g., accelerometer, microphone) to measure physiological parameters
(e.g., limb movement, electrical activity of the brain (EEG)) during sleep and are
intended for use in diagnosis of specific diseases or conditions such as sleep apnea.
Possible product code(s): OLV (21 CFR 882.1400), LEL, MNR (21 CFR 868.2375),
FLS, NPF (21 CFR 868.2377).

·

Mobile apps that use an attachment to the mobile platform to measure blood oxygen
saturation for diagnosis of specific disease or condition. Possible product code(s): DQA,
NLF, MUD, NMD (21 CFR 870.2700) or DPZ (21 CFR 870.2710).

·

Mobile apps that present donor history questions to a potential blood donor and record
and/or transmit the responses to those questions for a blood collection facility to use in
determining blood donor eligibility prior to collection of blood or blood components.
Possible product code: MMH

·

Mobile apps that use an attachment to the mobile platform to measure blood glucose
levels. Possible product code: NBW (21 CFR 862.1345).

·

Mobile apps that use that use an attachment to the mobile platform (e.g., light source,
laser) to treat acne, reduce wrinkles, or remove hair. Possible product code: OLP, OHT,
OHS (21 CFR 878.4810), OZC (21 CFR 890.5740).

·

Mobile apps that use a microphone or speaker within a mobile platform to serve as a
audiometer to allow healthcare providers to determine hearing loss at different
frequencies. Possible product code: EWO (21 CFR 874.1050)

·

Mobile apps that analyze an image of a skin lesion using mathematical algorithms, such
as fractal analysis, and provide the user with an assessment of the risk of the lesion.

Mobile apps that connect to an existing device type for purposes of controlling its
operation, function, or energy source and therefore are mobile medical apps: These mobile
apps are those that control the operation or function (e.g., changes settings) of an implantable or
body worn medical device. Possible product codes: Varies depending on the intended use and
function of the parent medical device; see additional examples below.

- 28 -

Contains Nonbinding Recommendations
·

Mobile apps that alter the function or settings of an infusion pump. Possible product
codes: MEB, FRN, LZH, LZG, OPP, MEA (21 CFR 880.5725), FIH (21 CFR 876.5820),
LKK.

·

Mobile apps that act as wireless remote controls or synchronization devices for computed
tomography (CT) or X-Ray machines. Possible product code: JAK (21 CFR 892.1750),
IZL (21 CFR 892.1720), KPR (21 CFR 892.1680).

·

Mobile apps that control or change settings of an implantable neuromuscular stimulator.
Possible product code(s): GZC (21 CFR 882.5860).

·

Mobile apps that calibrate, control, or change settings of a cochlear implant. Possible
product code(s): MCM.

·

Mobile apps that control the inflation or deflation of a blood-pressure cuff. Possible
product code: DSJ (21 CFR 870.1100), DSK (21 CFR 870.1110), DXN (21 CFR
870.1130).

·

Mobile apps that are used to calibrate hearing aids and assess the electroacoustic
frequency and sound intensity characteristics emanating from a hearing aid, master
hearing aid, group hearing aid or group auditory trainer. Possible product code ETW (21
CFR 874.3310).

Mobile apps that are used in active patient monitoring or analyzing patient-specific
medical device data from a connected device and therefore are mobile medical apps:
·

Mobile apps that connect to bedside (or cardiac) monitors and transfer the data to a
central viewing station for display and active patient monitoring. Possible product
code(s): DSI, MHX, MLD (21 CFR 870.1025), DRT, MWI, MSX (21 CFR 870.2300).

·

Mobile apps that connect to a perinatal monitoring system and transfer uterine
contraction and fetal heart rate data to another display to allow for remote monitoring of
labor progress. Possible product code(s): HGM (21 CFR 884.2740).

·

Mobile apps that are intended to display images for diagnostic review may be regulated
as a picture archiving and communications system. Possible product code LLZ, (21 CFR
892.2050).

- 29 -

Contains Nonbinding Recommendations

Appendix D Examples of current regulations
This Appendix provides additional examples of classifications for regulated medical devices, the
Class according to which they are regulated, and their regulation numbers as listed in Title 21 of
the Code of Federal Regulations (CFR). This list is intended as a starting point for mobile
medical app manufacturers to assist them in identifying regulated medical devices.
In the table below -- Regulation number 8xx.yyyy refers to regulation 21 CFR 8xx.yyyy; Device
class 1, 2, 3 -- indicates the classification that applies to the device; Submission type “510(k)
exempt,” -- means that the manufacturer is not required to submit a premarket notification (i.e.,
510(k)) prior to marketing the device. However, the 510(k) exemption may be subject to certain
limitations. Submission type “510(k),” -- means that the manufacturer is typically required to
submit a premarket notification.
Regulation
number
862.1345

Regulation
Description
Glucose test system

862.2100

Calculator/data
processing module for
clinical use

868.1850

Monitoring spirometer

868.1920

Esophageal
stethoscope with
electrical conductors
Breathing Frequency
Monitor
Apnea Monitor
Arrhythmia detector
and alarm (including
ST-segment
measurement and
alarm)
Blood-Pressure
Computer
Noninvasive blood
pressure measurement
system
Stethoscope

868.2375
868.2377
870.1025

870.1110
870.1130

870.1875(b)
870.2300

Cardiac Monitor
(including
cardiotachometer and
rate alarm)

870.2340

Electrocardiograph

870.2700

Oximeter

Example Device(s) within the Regulation
(and current product code)
System, Test, Blood Glucose, Over The
Counter (NBW)
Digital Image, Storage And Communications,
Non-Diagnostic, Laboratory Information
System (NVV)

Device
Class
2

Submission
Type
510(k)

1

510(k) exempt

Spirometer, Monitoring (W/Wo Alarm)
(BZK)
Stethoscope, Esophageal, With Electrical
Conductors (BZT)

2

510(k)

2

510(k)

Ventilatory Effort Recorder (MNR)

2

510(k)

Monitor, Apnea, Home Use (NPF)
Detector and Alarm, Arrhythmia (DSI)

2
2

510(k)
510(k)

Computer, Blood-Pressure (DSK)

2

510(k)

System, Measurement, Blood-Pressure, NonInvasive (DXN)

2

510(k)

Lung Sound Monitor (OCR)
Stethoscope, Electronic (DQD)
Monitor, Cardiac (Incl. Cardiotachometer &
Rate Alarm) (DRT)
Monitor, Physiological, Patient(Without
Arrhythmia Detection Or Alarms) (MWI)
System, Network And Communication,
Physiological Monitors (MSX)
Monitor, St Segment (MLC)
Single Lead Over-the-Counter
Electrocardiograph (OEY)
Oximeter (DQA)

2
2
2

510(k)
510(k)
510(k)

2

510(k)

2

510(k)

2

510(k)

2
2

510(k)
510(k)

- 30 -

Contains Nonbinding Recommendations
Regulation
number
870.2770
870.2800

Regulation
Description
Impedance
plethysmograph
Medical magnetic tape
recorder

874.1050

Audiometer

874.3400
874.4770
876.1500

Tinnitus masker
Otoscope
Endoscope and
accessories

876.1725

Gastrointestinal
motility monitoring
system
Surgical camera and
accessories

878.4160

878.4810

880.2400
880.2700
880.2910
880.5580
880.6350

882.1400

882.1550

882.1620
882.1890
882.1900
882.1950
884.2730
884.2740
884.2800

Laser surgical
instrument for use in
general and plastic
surgery and in
dermatology
Bed-patient monitor
Stand-on patient scale
Clinical electronic
thermometer
Acupuncture needle
Battery-powered
medical examination
light
Electroencephalograph

Nerve conduction
velocity measurement
device
Intracranial pressure
monitoring device
Evoked response
photic stimulator
Evoked response
auditory stimulator
Tremor Transducer
Home uterine activity
monitor
Perinatal monitoring
system and accessories
Computerized labor

Example Device(s) within the Regulation
(and current product code)
Analyzer, Body Composition (MNW)

Device
Class
2

Electrocardiograph, Ambulatory, With
Analysis Algorithm (MLO)
Recorder, Event, Implantable Cardiac,
(Without Arrhythmia Detection) (MXC)
Audiometer (EWO)

2

510(k)

2

510(k)

2

Masker, Tinnitus (KLW)
Otoscope (ERA)
Endoscopic Video Imaging
System/Component, GastroenterologyUrology (FET)
Recorder, External, Pressure, Amplifier &
Transducer (FES)

2
1
2

510(k) or
510(k) exempt
510(k)
510(k) exempt
510(k)

2

510(k)

Camera, Cine, Microsurgical, With Audio
(FWK)
Camera, Still, Microsurgical (FTH)
Camera, Television, Endoscopic, With Audio
(FWG)
Light Based Over The Counter Wrinkle
Reduction (OHS)
Over-The-Counter Powered Light Based
Laser For Acne (OLP)

1

510(k) exempt

1
1

510(k) exempt
510(k) exempt

2

510(k)

2

510(k)

Monitor, Bed Patient (KMI)
Scale, Stand-On, Patient (FRI)
Thermometer, Electronic, Clinical (FLL)

1
1
2

510(k) exempt
510(k) exempt
510(k)

Locator, Acupuncture Point (BWJ)
Light, Examination, Medical, Battery
Powered (KYT)

2
1

510(k)
510(k) exempt

Full-montage electroencephalograph (GWQ)
Standard polysomnograph with
electroencephalograph (OLV)
Device, Nerve conduction velocity
measurement (JXE)

2
2

510(k)
510(k)

2

510(k)

Device, Monitoring, Intracranial pressure
(GWM)
Stimulator, Photic, Evoked response (GWE)

2

510(k)

2

510(k)

Stimulator, Auditory, Evoked response (GWJ)

2

510(k)

Transducer, Tremor (GYD)
Monitor, Heart Rate, Fetal, Non-Stress Test
(Home Use) (MOH)
System, Monitoring, Perinatal (HGM)

2
2

510(k)
510(k)

2

510(k)

System, Monitoring, For Progress Of Labor

2

510(k)

- 31 -

Submission
Type
510(k)

Contains Nonbinding Recommendations
Regulation
number
884.2900
884.6120
884.6190

886.1510
886.1570
886.1930
886.5540
892.1560
892.2030

892.2050

Regulation
Description
monitoring system
Fetal stethoscope
Assisted reproductive
accessories
Assisted reproductive
microscopes and
microscope
accessories
Eye movement
monitor
Ophthalmoscope
Tonometer and
Accessories
Low-vision magnifier
Ultrasonic pulsed echo
imaging system
Medical image
digitizer

Picture archiving and
communications
system

Example Device(s) within the Regulation
(and current product code)
(NPB)
Stethoscope, Fetal (HGN)
Accessories, Assisted Reproduction (MQG)

Device
Class

Submission
Type

1
2

510(k) exempt
510(k)

Microscope And Microscope Accessories,
Reproduction, Assisted (MTX)

1

510(k) exempt

Monitor, Eye Movement, Diagnostic (HMC)

2

510(k)

Ophthalmoscope, Battery-powered (HLJ)
Tonometer, Ac-Powered (HPK)

2
2

510(k)
510(k)

Magnifier, Hand-Held, Low-Vision (HJF)
Spectacle Microscope, Low-Vision (HKC)
System, Imaging, Optical Coherence
Tomography (Oct) (NQQ)
Digitizer, Image, Radiological (LMA)

1
1
2

510(k) exempt
510(k) exempt
510(k)

2

Digitizer, Images, Ophthalmic (NFH)

2

Enforcement
Discretion for
510(k)
submission33

System, Image Processing, Radiological
(LLZ)
System, Image Management, Opthalmic (NFJ)

33

2

Enforcement
Discretion for
510(k)
submission 34
510(k)

2

510(k)

See “Guidance for Industry and Food and Drug Administration Staff - Enforcement Policy for Premarket
Notification Requirements for Certain In Vitro Diagnostic and Radiology Devices” (December 20, 2011) available
at http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm283904.htm
34
See footnote 31.

- 32 -

Contains Nonbinding Recommendations

Appendix E Brief description of certain device regulatory
requirements
This Appendix provides a high level description of certain regulatory requirements for medical
devices, including mobile medical apps. The FDA has additional resources and publications
online that describe these and other requirements in detail.

1. Establishment Registration and Medical Device Listing
Under 21 CFR Part 807, manufacturers of medical devices are required to annually register their
establishments35 with FDA and provide a list of the devices they market. The registration and
listing requirement is a means of keeping FDA advised of who is manufacturing devices, and of
the types of devices an establishment is manufacturing. Mobile medical app manufacturers are
required to register their establishments with FDA and to list36 by identifying to FDA the mobile
medical apps they are marketing.
Additional information can be found in “Device Advice: Device Registration and Listing” at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourDevice/
RegistrationandListing/default.htm. If you need further assistance, you may contact the Division
of Risk Management Operations, Regulatory Policy and Systems Branch: Email:
reglist@fda.hhs.gov, phone: 301-796-7400. Assistance is also available from, Division of
Industry and Consumer Education: Email: DICE@fda.hhs.gov
phone: 301-796-7100 or 800-638-2041.

2. Investigational Device Exemption (IDE) requirements
An IDE allows an investigational device to be used in a clinical study in order to collect safety
and effectiveness data required to support a Premarket Approval (PMA) application or a
Premarket Notification 510(k) submission to FDA. Clinical studies with devices of significant
risk must be approved by FDA and by an Institutional Review Board (IRB) before the study can
begin. Studies with devices of non-significant risk must be approved by the IRB only before the
study can begin.
Mobile medical app manufacturers who are creating mobile apps with novel technologies are
encouraged to engage in early collaboration meetings with the FDA to receive recommendations
for testing and development of those devices requiring clinical investigations to support
marketing.
Additional information about these meetings is described in guidance issued on February 28,
2001: “Early Collaboration Meetings Under the FDA Modernization Act (FDAMA); Final
Guidance for Industry and for CDRH Staff.” This document is available at
35

Under 21 CFR 807.3(c), “Establishment” is defined as “a place of business under one management at one general
physical location at which a device is manufactured, assembled, or otherwise processed.”
36
See 21 CFR part 807.

- 33 -

Contains Nonbinding Recommendations
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm0
73604.htm.
Further information regarding the investigational device exemption can be found in “Device
Advice: Investigational Device Exemption” at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourDevice/I
nvestigationalDeviceExemptionIDE/default.htm.

3. Labeling requirements
Medical device manufacturers are required to comply with applicable labeling regulations found
in 21 CFR Part 801 for medical devices and Part 809 for in vitro diagnostic products.

4. Premarket submission for approval or clearance
Mobile medical app manufacturers should identify the current classification covering their
mobile medical app. Manufacturers are required to prepare and submit to the FDA an appropriate
premarket submission, as required for their device classification.
Additional information can be found in “Device Advice: Device Registration and Listing” at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/HowtoMarketYourDevice/
RegistrationandListing/default.htm.

5. Quality System Regulation (QS Regulation)
Mobile medical app manufacturers are required to comply with the QS regulation. 37 The QS
regulation does not prescribe in detail how a manufacturer must produce a specific device, but
provides a framework for all manufacturers to develop and follow to help ensure that their
products consistently meet applicable requirements and specifications. As part of this
framework, mobile medical app manufacturers are required to develop requirements for their
products that will result in devices that are safe and effective, and to establish methods and
procedures to design, produce, and distribute their devices.
Furthermore, mobile medical app manufacturers are required, as part of the QS regulation (21
CFR 820.30), to appropriately verify and validate their mobile medical apps along with the
mobile platform to ensure safe and effective operation of the mobile medical app.
Mobile medical app manufacturers are required to ensure that adequate controls and processes
are in place through purchasing controls to ensure safe distribution, installation, and operation of
the mobile medical app.
Additional information regarding the QS regulation and can be found at “Quality System (QS)
Regulation/Medical Device Good Manufacturing Practices”

37

See 21 CFR part 820.

- 34 -

Contains Nonbinding Recommendations
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/PostmarketRequirements/Q
ualitySystemsRegulations/default.htm.

6. Medical Device Reporting (MDR) (Adverse event reporting)
The Medical Device Reporting (MDR) regulation requires manufacturers and importers of
medical devices to submit reports to the FDA whenever they receive or otherwise become aware
of information, from any source, that reasonably suggests that a device they market may have
caused or contributed to a death or serious injury, or has malfunctioned and the device or a
similar device that they market would be likely to cause or contribute to a reportable death or
serious injury if the malfunction were to recur.38 MDR requires medical device manufacturers
to:
· Submit MDR reportable events involving their medical devices as described in 21 CFR
803.10(c) and 803.50;
· Submit 5-day reports as described in 21 CFR 803.53;
· Submit supplemental reports as described in 21 CFR 803.56;
· Develop, maintain, and implement written procedures for the identification and evaluation
of all medical device events to determine whether the event is MDR reportable as
described in 21 CFR 803.17;
· Conduct an investigation of each event and evaluate the cause of the event as described in
21 CFR 803.50(b)(3); and
· Establish and maintain complete files for all complaints concerning adverse medical
device events as described in 21 CFR 803.18.
The MDR report (FDA Form 3500A) must contain all the information described in 21 CFR
803.52 that is reasonably known to the manufacturer. Information reasonably known includes
any information that:
· Can be obtained by contacting a user facility, importer, or other initial reporter;
· Is in the possession of the manufacturer; or
· Can be obtained by analysis, testing, or other evaluation of the device.
For additional instructions on how to complete the 3500A form, refer to the document titled
“Instructions for Completing Form FDA 3500A” at
http://www.fda.gov/Safety/MedWatch/HowToReport/DownloadForms/ucm149238.htm.
For additional guidance on the MDR regulation and the reporting requirements, refer to the
document titled “Medical Device Reporting for Manufacturers” at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm0
94529.htm.
For Questions about Medical Device Reporting, including interpretation of MDR policy:
38

See 21 CFR part 803.

- 35 -

Contains Nonbinding Recommendations
·
·
·

Call: (301) 796-6670 (voice)
Email: RSMB@fda.hhs.gov
Mail: Food and Drug Administration, Center for Devices and Radiological Health,
Reporting Systems Monitoring Branch, 10903 New Hampshire Avenue,
WO Bldg. 66, Room 3217, Silver Spring, MD 20993-0002

7. Correcting Problems
A mobile medical app manufacturer may voluntarily take action at any time or may be requested
to take action by the FDA to correct problems. Voluntary action is usually taken by device
manufacturers. Examples of the types of actions that a mobile medical app manufacturer may be
requested to take include, but are not limited to:
·
·
·
·

Inspecting the device for problems;
Repairing the device;
Adjusting settings on the device; and
Upgrading software to reduce risk from a “bug” or unintended response.

Under certain circumstances, FDA may initiate a request that a manufacturer address a problem
with a device through other means, including by removal of the product from the market. When
recommending corrective action, the FDA intends to take into account the essential role that
certain mobile medical apps take as an integral part of a larger patient care system.
Reporting Corrections to FDA:
In accordance with 21 CFR 806.10, mobile medical app manufacturers are required to promptly
report, within 10 working days from the time the correction is initiated, to the FDA certain
actions concerning device corrections and removals for the mobile medical app. Specifically,
mobile medical app manufacturers are required to report to FDA any corrections made to a
mobile medical app to reduce a risk to health posed by the mobile medical app or to remedy a
violation of the FD&C Act caused by the mobile medical app which may present a risk to health.
The reporting requirement does not extend to all modifications to mobile medical apps. For
example, certain actions that would improve the quality of a mobile medical app but that would
not reduce a risk to health posed by the mobile medical app or remedy a violation of the FD&C
Act are not required to be reported under 21 CFR 806.1(b)39. If there is not a "risk to health"
involved, a report to FDA is not required, but the mobile medical app manufacturer must keep a
record of the correction. An example of such action taken by the manufacturer could be changes

39

Under 21 CFR 806.1(b), the following actions are exempt from the reporting requirements of part 806:
(1) Actions taken by device manufacturers or importers to improve the performance or quality of a device but that
do not reduce a risk to health posed by the device or remedy a violation of the act caused by the device.
(2) Market withdrawals as defined in 21 CFR 806.2(h).
(3) Routine servicing as defined in 21 CFR 806.2(k).
(4) Stock recoveries as defined in 21 CFR 806.2(l).

- 36 -

Contains Nonbinding Recommendations
made to correct a defect that creates a nuisance for the user but does not present a risk to the
health of the user or patient.
More information about reporting requirements under 21 CFR Part 806 is available in “Device
Advice: Recalls, Corrections, and Removals” at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/PostmarketRequirements/R
ecallsCorrectionsAndRemovals/default.htm.

- 37 -

Contains Nonbinding Recommendations

Appendix F Frequently Asked Questions (FAQs)
1)

I have a mobile app not identified in this guidance. What is the best way to get
additional information from the FDA about my product?

Answer: FDA recognizes that this guidance does not describe all types of mobile apps used in
healthcare. Some manufacturers may be unsure whether their mobile app is considered a
medical device which is subject to regulatory oversight, or whether their medical device could be
under FDA’s intent to exercise enforcement discretion. If the device is subject to regulatory
oversight, manufacturers may have questions about which regulatory requirements are applicable
to their specific mobile app.
After reviewing this guidance, FDA encourages mobile app manufacturers to contact the Agency
to obtain more information using one of the following ways:
- Phone or e-mail - For general regulatory information, contact the Division of Industry and
Consumer Education (DICE). Email: DICE@fda.hhs.gov; phone: 301-796-7100 or 800-6382041.
If your question relates to apps used in blood establishments or another area of CBER
regulation, contact the Office of Communication, Outreach and Development, Center for
Biologics Evaluation and Research, 10903 New Hampshire Ave., Bldg. 71, Room 3128,
Silver Spring; e-mail: ocod@fda.hhs.gov; phone: 1-800-835-4709 or 240-402-7800.
- Online – The FDA has several resources and publications online that describe various
regulatory requirements in detail. FDA’s “Device Advice” website
(http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/default.htm) and online
courses at “CDRH Learn” (http://www.fda.gov/Training/CDRHLearn/default.htm) are a
good place to start. Other sections in this guidance provide links to more detailed
information related to more specific topics.
- Letter - For written feedback about the classification and the regulatory requirements that
may be applicable to a mobile medical app that is a device, manufacturers should use the
513(g) process. Specifically, a manufacturer should submit the following for a 513(g)
submission:
· User fee,
· Cover letter,
· Description of the mobile app,
· Description of what the mobile app is to be used for, and
· Any proposed labeling or promotional material for the mobile app and, as applicable,
any labeling or promotional material of a similar, legally marketed device, if
available.
FDA will generally issue a response to the 513(g), in the form of a confidential letter to the
manufacturer, within 60 days of receipt of the request for information. For more specific
information about what to include in a 513(g) and where to send it, refer to FDA’s guidance
document titled “FDA and Industry Procedures for Section 513(g) Requests for Information
- 38 -

Contains Nonbinding Recommendations
Under the Federal Food, Drug, and Cosmetic Act” at:
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/uc
m209841.htm. For more information about 513(g) user fees, refer to FDA’s guidance
document titled “User Fees for 513(g) Requests for Information” at:
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/uc
m209852.htm.
2)

Why does FDA recommend that manufacturers follow the Quality System (QS)
regulation for those mobile apps that MAY be devices and could be mobile
medical apps but for which FDA intends to exercise enforcement discretion?

Answer: FDA believes all manufacturers of medical device software should have in place an
adequate quality management system that helps ensure that their products consistently meet
applicable requirements and specifications and can support the software throughout its total life
cycle. Having and maintaining an adequate quality management system is also important since
the FDA has found that the majority of software-related failures in medical devices are due to
design errors. In one study, the most common problem was failure to validate software prior to
routine maintenance.40
Adequate quality management systems incorporate appropriate risk management strategies, good
design practices, adequate verification and validation, and appropriate methods to correct and
prevent risks to patients and adverse events that may arise from the use of the product. All of
these elements are part of FDA’s QS regulation.
3)

Is FDA’s QS regulation similar to software development practices I already use?

Answer: Most likely. Though not all of the principles in the QS regulation are applicable to the
development and manufacture of quality mobile medical apps41, the majority of them are
applicable and are consistent with commonly used and accepted good software development
practices, such as those from the Institute of Electrical and Electronics Engineers’ (IEEE),
Software Engineering Body of Knowledge (SWEBOK), and Carnegie Mellon Software
Engineering Institute’s Capability Maturity Model Integration (CMMI) methods.
The FDA’s approach to QS regulation is also harmonized with certain international standards
such as ISO 9001 and ISO 13485.42 Similar to these international standards, the QS regulation
does not prescribe in detail how a manufacturer must produce a specific device but provides a
framework for all manufacturers to develop and follow to help ensure that their products
consistently meet applicable requirements and specifications. The QS regulation can apply to
and be scaled for any size manufacturer and any type of product. It also allows for a
40

See footnote 19.
Certain portions of the QS regulation that apply to medical device hardware (such as the production and process
controls outlined in 21 CFR 820.70) may not clearly apply to mobile medical apps.
42
ISO 9001:2008 “Quality management systems--Requirements,'' and ISO 13485:2003 ``Medical devices-Quality
management systems-Requirements for regulatory purposes.” See also ANSI/AAMI/ISO 13485:2003.
41

- 39 -

Contains Nonbinding Recommendations
manufacturer to choose those requirements most appropriate for its given device and
manufacturing process.43
4)

What are some examples of parts of the QS regulation that are of particular
importance to mobile medical apps and where can I find more information about
them?

Answer: Though not a complete list, some examples of principles within the QS regulation that
are relevant to all mobile medical app manufacturers include risk assessment and management,
design controls, and corrective and preventive actions. Risk assessment and management is a
critical part of good quality management systems. Good design practices are important to the
development and manufacture of safe mobile medical apps. It is also important for
manufacturers to have procedures in place to identify, analyze, correct, and prevent app-related
causes of patient or user harm. References related to these examples are provided in Appendix E
of this guidance. Additional references about these principles which mobile medical app
manufacturers may find useful include the following:
FDA’s “Design Control Guidance for Medical Device Manufacturers” at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/uc
m070627.htm
FDA’s “General Principles of Software Validation” guidance at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/uc
m085281.htm

5)

Do all the mobile medical apps have to submit a premarket submission and
receive FDA clearance or approval before marketing?

Answer: No, not all mobile medical app manufacturers have to submit a premarket submission
(i.e., a 510(k) or PMA) prior to marketing their app. This determination depends on the
classification of the device. Manufacturers of devices that are exempt from 510(k) or PMA
requirements do not have to file a submission with FDA prior to marketing their device. For
example, the majority of class I devices are exempt from the premarket submission requirements
and are subject to the least regulatory control.
Regardless of whether medical devices are subject to the premarket submission requirements,
most medical devices (including Class I devices) have to comply with other basic regulatory
requirements that are called “General Controls.” More information about what “General
Controls” are and what a medical device manufacturer should do to comply with these
requirements, can be found in “Device Advice: General Controls for Medical Devices” at
43

See 21 CFR 820.1 (stating “if a manufacturer engages in only some operations subject to the requirements in this
part, and not in others, that manufacturer need only comply with those requirements applicable to the operations in
which it is engaged.”)

- 40 -

Contains Nonbinding Recommendations
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/Overview/GeneralandSpeci
alControls/ucm055910.htm and at
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/Overview/GeneralandSpeci
alControls/default.htm.
6)

Some FDA classifications state they are “510(k) exempt.” What does 510(k)
exempt mean and how do I know if it applies to my product?

Answer: If a classification states the device type is “510(k) exempt,” this means that the
manufacturer is not required to submit a premarket notification (i.e., a 510(k)) prior to marketing
the device. However, the 510(k) exemption may be subject to certain limitations. Manufacturers
are encouraged to confirm the device’s exempt status and any limitations to that status that may
apply in accordance with 21 CFR Parts 862-892.. Additional information about 510(k) exempt
devices can be found at: http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpcd/315.cfm.

7)

If a 510(k) is required for my mobile medical app, what type of software
documentation does FDA recommend I include in the submission?

Answer: FDA’s recommendations for the software-related documentation that you provide in
your premarket submission are addressed in detail in the FDA’s “Guidance for the Content of
Premarket Submissions for Software Contained in Medical Devices.” This guidance can be
accessed on FDA’s website here:
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm0
89543.htm.
If the mobile medical app uses off-the-shelf software, manufacturers should also refer to FDA’s
“Guidance for Industry, FDA Reviewers, and Compliance on Off-the-Shelf Software Use in
Medical Devices” available at:
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm0
73778.htm.
8)

I am a medical device manufacturer and making my product labeling available
electronically using a mobile app. Is my app considered a mobile medical app?

Answer: Mobile apps that provide electronic access and are intended for use as a digital version
of medical device labeling or instructions for use are not considered medical device on their own
and therefore are not considered mobile medical apps. These are apps from a device
manufacturer that provide information to support the company’s own device. Examples include
apps that provide an electronic copy of cleared or approved medical device labeling or apps that
provide video instruction for how to use a medical device. These types of apps are not
considered devices within themselves, but instead are considered part of the medical device
labeling and are subject to the regulatory labeling requirements relevant to that particular
product.

- 41 -

Contains Nonbinding Recommendations
9)

Does an electronic method of collecting clinical investigations for example
through a mobile app considered a mobile medical app, what requirements apply?

Answer: Mobile apps used for data collection in clinical studies (such as electronic Patient
Reported Outcomes (ePRO) apps) are not considered on its own a mobile medical app.
However, manufacturers and users of this type of mobile app should see FDA’s draft guidance
related to use of computers in clinical trials, “Electronic Source Data in Clinical Investigations,”
44
issued on November 20, 2012. When final, this guidance will represent the FDA’s current
thinking on this topic. For the most recent version of this guidance, check the CDER guidance
webpage at
http://www.fda.gov/Drugs/GuidanceComplianceRegulatoryInformation/Guidances/default.htm.

10)

I am a medical device manufacturer. Is an electronic method of collecting and
storing quality systems information in my manufacturing process considered a
medical device or a mobile medical app?

Answer: Mobile apps used in the production process for medical devices, or for collecting,
storing and maintaining quality system data collection for medical devices (including complaint
submissions) are not considered medical device on their own and therefore are not considered
mobile medical apps. These types of apps do not meet the definition of medical device but are
part of the quality system. However these mobile apps are required to comply with the
appropriate good manufacturing practices (GMP) regulations (see 21 CFR Part 820).

- 42 -

Contains Nonbinding Recommendations

Appendix G Additional Resources
AAMI
ANSI
IEC
IEEE
ISO

= Association for the Advancement of Medical Instrumentation
= American National Standards Institute
= International Electrotechnical Commission
= Institute of Electrical and Electronics Engineers
= International Organization for Standardization

1. Guidance for Industry and FDA Staff - Implementation of Medical Device Establishment
Registration and Device Listing Requirements Established by the Food and Drug
Administration Amendments Act of 2007
(http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/u
cm185871.htm).
2. ISO/IEC 90003:2004, Software engineering -- Guidelines for the application of ISO
9001:2000 to computer software, 2004.
3. ISO 9001:2008, Quality management systems – Requirements, 2008.
4. ISO 13485:2003, Medical devices - Quality management systems - Requirements for
regulatory purposes, 2003. Note: This has also been adopted in the U.S. as ANSI/AAMI/ISO
13485:2003, Medical devices - Quality management systems - Requirements for regulatory
purposes, 2003 (identical adoption).
5. ISO 9000:2005 Quality management systems – Fundamentals and vocabulary, 2005.
6. ISO 14971:2007, Medical Devices - Risk Management - Part 1: Application of Risk
Analysis, 2007. Note: This has also been adopted in the U.S. as ANSI/AAMI/ISO
14971:2007, Medical devices - Application of risk management to medical devices, 2007
(identical adoption).
7. Guidance for Industry - Cybersecurity for Networked Medical Devices Containing Off-theShelf (OTS) Software
(http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/u
cm077812.htm).
8. Information for Healthcare Organizations about FDA's "Guidance for Industry:
Cybersecurity for Networked Medical Devices Containing Off-The-Shelf (OTS) Software" -http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/uc
m070634.htm
9. IEC 62304:2006, Medical device software – Software life cycle processes, 2006. Note: This
is also adopted in the U.S as ANSI/AAMI/IEC 62304:2006, Medical device software Software life cycle processes, 2006 (identical adoption).
10. IEEE Std 1012-2004, IEEE Standard for Software Verification and Validation, 2004.
- 43 -

Contains Nonbinding Recommendations
11. IEEE 1012-2012, IEEE Standard for System and Software Verification and Validation, 2012.
12. IEEE Standards Collection, Software Engineering, 1994. ISBN 1-55937-442-X.
13. ISO/IEC 25051:2006, Software engineering -- Software product Quality Requirements and
Evaluation (SQuaRE) -- Requirements for quality of Commercial Off-The-Shelf (COTS)
software product and instructions for testing, 2006.
14. ISO/IEC 12207:2008, Systems and software engineering – Software life cycle processes,
2008 and IEEE Std 12207-2008, Systems and software engineering – Software life cycle
processes, 2008.
15. ISO/IEC 14598:1999, Information technology - Software product evaluation, 1999.
16. AAMI TIR32:2004, Medical device software risk management, 2004.
17. AAMI TIR36:2007, Validation of software for regulated processes, 2007.
18. ANSI/AAMI/IEC TIR80002-1:2009, Medical device software - Part 1: Guidance on the
application of ISO 14971 to medical device software, 2009. (Identical adoption of IEC/TR
80002-1:2009)
19. Guidance for the Submission of Premarket Notifications for Medical Image Management
Devices
(http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/u
cm073720.htm)
20. Clause 14 of IEC 60601-1:2005, Medical electrical equipment, Part 1: General requirements
for basic safety and essential performance, 2005 OR Clause 14 of ANSI/AAMI ES606011:2005, Medical electrical equipment, Part 1: General requirements for basic safety and
essential performance, 2005 (adoption with national deviations of IEC 60601-1:2005).
21. IEC 61508-2:2010, Functional safety of electrical/electronic/programmable electronic
safety-related systems, 2010.

- 44 -



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.7
Linearized                      : No
Create Date                     : 2015:01:12 09:15:27-05:00
Modify Date                     : 2017:06:09 10:06:43-04:00
NCCL App                        : Office
NCCL Standard                   : Section 508
NCCL Status                     : Passed
Language                        : en-US
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Metadata Date                   : 2017:06:09 10:06:43-04:00
Producer                        : iTextSharp 4.0.3 (based on iText 2.0.2)
Format                          : application/pdf
Document ID                     : uuid:a725aff7-eb86-427d-824d-a633a22ad6e6
Instance ID                     : uuid:d7da74b1-c3d7-4afd-adb7-d919119e89a6
Page Count                      : 45
EXIF Metadata provided by EXIF.tools

Navigation menu