WTW Cond 3210 Operating Manual WTW_Cond_3210

User Manual: Pdf WTW_Cond_3210

Open the PDF directly: View PDF PDF.
Page Count: 58

OPERATING MANUAL
ba75794e05 06/2013
Cond 3210
CONDUCTIVITY METER
25.0
1417
°C
15.11.2008
08:37
ӕ
μS/cm
Menu
[Tr25][nLF][ 0.835 1/cm]
For Sales & Service Contact
2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242
1-800-833-7958
www.geotechenv.com
2ba75794e05 06/2013
Cond 3210
Copyright © Weilheim 2009, WTW GmbH
Reproduction in whole - or even in part - is prohibited without the
express written permission of WTW GmbH, Weilheim.
Printed in Germany.
For the most recent version of the manual, please visit
www.WTW.com.
Cond 3210 Contents
3
ba75794d05 06/2013
Cond 3210 - Contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 10
3 Commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2 Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.2.1 Inserting the batteries. . . . . . . . . . . . . . . . . . . . . 13
3.2.2 Switching on the meter. . . . . . . . . . . . . . . . . . . . 14
3.2.3 Setting the date and time . . . . . . . . . . . . . . . . . 14
4 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Switching on the meter . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.2 General operating principles . . . . . . . . . . . . . . . . . . . . . . 16
4.2.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . . 16
4.2.2 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.2.3 Example 1 on navigation: Setting the language. 19
4.2.4 Example 2 on navigation: Setting the date and
time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.3 Sensor-independent settings . . . . . . . . . . . . . . . . . . . . . 24
4.3.1 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4.3.2 Data storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.3.3 Automatic Stability control . . . . . . . . . . . . . . . . . 26
4.4 Conductivity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4.4.1 General information . . . . . . . . . . . . . . . . . . . . . . 27
4.4.2 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
4.4.3 Temperature compensation . . . . . . . . . . . . . . . . 30
4.4.4 Settings for conductivity measuring cells . . . . . . 31
4.4.5 Determining the cell constant
(calibration in control standard) . . . . . . . . . . . . . 35
4.4.6 Displaying calibration records . . . . . . . . . . . . . . 36
4.5 Data storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.5.1 Manual storage . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.5.2 Displaying and editing the measurement data
storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Contents Cond 3210
4ba75794d05 06/2013
4.5.3 Erasing the measurement data storage . . . . . . . 41
4.6 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
4.6.1 Resetting the measurement settings . . . . . . . . . 42
4.6.2 Resetting the system settings. . . . . . . . . . . . . . . 43
5 Maintenance, cleaning, disposal . . . . . . . . . . . . . . . . . 45
5.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
5.1.1 Replacing the batteries . . . . . . . . . . . . . . . . . . . . 45
5.2 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.3 Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7.2 Measuring ranges, resolution, accuracy . . . . . . . . . . . . . 50
8 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Cond 3210 Overview
5
ba75794e05 06/2013
1Overview
The Cond 3210 compact precision conductivity meter enables you to
perform conductivity measurements quickly and reliably.
The Cond 3210 provides the maximum degree of operating comfort,
reliability and measuring certainty for all applications.
The proven procedures for determining or adjusting the cell constant
support your work with the conductivity meter.
1Keypad
2Sample display
3Socket field
25.0
1417
°C
15.11.2008
08:37
ӕ
μS/cm
1
2
3
Menu
[Tr25][nLF][ 0.835 1/cm]
Overview Cond 3210
6ba75794e05 06/2013
1.1 Keypad
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <ENTER>) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
<ENTER_>).
ENTER
M
F1 F2
AR
CAL
STO RCL
Cond 3210
<F1>:
<F1_>:
<F2>:
<F2_>:
Softkey providing situation dependent functions , e.g.:
<F1>/[Menu]: Opens the menu for measurement settings
<F1_>/[Menu]: Opens the menu for system settings
<On/Off>: Switches the meter on or off
<M>: Selects the measured parameter
<CAL>:
<CAL_>:
Calls up the calibration procedure
Displays the calibration data
<STO>:
<STO_>
Saves a measured value manually
<RCL>:
<RCL_>
Displays the manually stored measured values
<>: Increments values, scrolls
<>: Decrements values, scrolls
F1
F2
CAL
STO
RCL
Cond 3210 Overview
7
ba75794e05 06/2013
1.2 Display
Function display
indicators
<ENTER>:
<ENTER_>:
Opens the menu for measurement settings / confirms
entries
Opens the menu for system settings
<AR> Freezes the measured value (HOLD function)
Switches the AutoRead measurement on or off
AR
1Status information
2Measured value (with unit)
3Measured parameter
4Sensor symbol (calibration evaluation, calibration
interval)
5Measured temperature (with unit)
6Status line
7Softkeys and date + time
HOLD AR
25.0
1417
°C
3
2
4
5
6
7
1
ӕ
μS/cm
[Tr25][nLF][ 0.835 1/cm]
Menu 22.09.2008
08:00
Error An error occurred during calibration
AR Stability control (AutoRead) is active
HOLD Measured value is frozen (<AR> key)
Overview Cond 3210
8ba75794e05 06/2013
1.3 Socket field
Caution
Only connect measuring cells to the meter that cannot return any
voltages or currents that are not allowed (> SELV and > current
circuit with current limiting).
Almost all customary measuring cells fulfill these conditions.
Connectors:
1Conductivity measuring cell
2Service interface
21
Cond 3210 Safety
9
ba75794e05 06/2013
2 Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating
manual before working with the meter.
The operating manual must always be available within the
vicinity of the meter.
Target group The meter was developed for work in the field and in the laboratory.
Thus, we assume that, as a result of their professional training and
experience, the operators will know the necessary safety precautions
to take when handling chemicals.
Safety instructions Safety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g.
"Caution") indicates the level of danger:
Warning
indicates instructions that must be followed precisely in order to
avoid possibly great dangers to personnel.
Caution
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the meter or
the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating
manuals.
Safety Cond 3210
10 ba75794e05 06/2013
2.1 Authorized use
Authorized use of the meter consists exclusively of the measurement
of conductivity, resistivity, salinity, TDS (total dissolved solids) and
temperature in a laboratory or field environment.
The technical specifications as given in chapter 7 TECHNICAL DATA must
be observed. Only the operation and running of the meter according to
the instructions given in this operating manual is authorized.
Any other use is considered unauthorized.
2.2 General safety instructions
This meter is constructed and tested in compliance with the IEC 1010
safety regulations for electronic measuring instruments.
It left the factory in a safe and secure technical condition.
Function and
operational safety
The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during
operation.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 TECHNICAL DATA.
If the meter was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty
functioning of the meter. In this event, wait until the temperature of the
meter reaches room temperature before putting the meter back into
operation.
Caution
The meter is only allowed to be opened by authorized personnel.
Cond 3210 Safety
11
ba75794e05 06/2013
Safe operation If safe operation is no longer possible, the meter must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
-has been damaged in transport
-has been stored under adverse conditions for a lengthy period of
time
-is visibly damaged
-no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the meter.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
-EEC directives for protective labor legislation
-National protective labor legislation
-Safety regulations
-Safety datasheets of the chemical manufacturers.
Caution
In addition to the safety instructions mentioned here, also follow
the safety instructions of the sensors used.
The operating manuals of the sensors are available on the
supplied CD and on the Internet under www.WTW.com.
Safety Cond 3210
12 ba75794e05 06/2013
Cond 3210 Commissioning
13
ba75794e05 06/2013
3 Commissioning
3.1 Scope of delivery
-Conductivity meter Cond 3210
-4 batteries 1.5 V Mignon type AA
-Short instructions
-CD-ROM with detailed operating manual
3.2 Initial commissioning
Perform the following activities:
-Insert the supplied batteries
-Switch on the meter
-Set the date and time
3.2.1 Inserting the batteries
1Unscrew the two screws (1) on the underside of the meter.
2 Open the battery compartment (2) on the underside of the
meter.
3Place four batteries (type Mignon AA) in the battery
compartment.
2
1
Commissioning Cond 3210
14 ba75794e05 06/2013
Note
Alternatively, you can also use Ni-MH rechargeable batteries (type
Mignon AA). In order to charge the batteries, an external charging
device is required.
Caution
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the ± signs in the
battery compartment.
3.2.2 Switching on the meter
Note
The meter has an energy saving feature to avoid unnecessary battery
depletion.
The energy saving feature switches off the measuring instrument if no
key is pressed during the adjusted interval. (How to set the switch-off
interval, see section 4.3.1).
3.2.3 Setting the date and time
4Close the battery compartment (2) and tighten the screws (1).
1Press the <On/Off> key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test
is being performed.
Subsequently, the meter switches to the measuring mode
(measured value display).
1See section 4.2.4
Cond 3210 Operation
15
ba75794e05 06/2013
4 Operation
4.1 Switching on the meter
Switching on Press the <On/Off> key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test is being
performed.
The measured value display appears.
Switching off Press the <On/Off> key.
Automatic switch-off The instrument has an automatic switch-off function in order to save the
batteries (see section 4.3.1). The automatic switch-off switches off the
measuring instrument if no key is pressed for an adjustable period.
Display illumination The meter automatically switches off the display illumination if no key
is pressed for 15 seconds. The illumination is switched on with the next
keystroke again.
You can also generally switch the display illumination on or off (see
section 4.3.1).
ϰ
1417
µS/cm
25.0
°C
[Tr25][nLF][ 0.835 1/cm]
22.09.2008
08:00
Menu
Operation Cond 3210
16 ba75794e05 06/2013
4.2 General operating principles
This section contains basic information on the operation of the
Cond 3210.
Operating elements,
display
An overview of the operating elements and the display is given in
section 1.1 and section 1.2.
Operating modes,
navigation
An overview of the operating modes and navigation of the Cond 3210
is given in section 4.2.1 and section 4.2.2.
4.2.1 Operating modes
The meter has the following operating modes:
-Measuring
The measurement data of the connected sensor is shown in the
measured value display
-Calibration
The course of a calibration with calibration information, functions
and settings is displayed
-Storage
The meter stores measuring data manually
-Setting
The system menu or a sensor menu with submenus, settings and
functions is displayed
Cond 3210 Operation
17
ba75794e05 06/2013
4.2.2 Navigation
Measured value display In the measured value display, you can
-Open the respective measuring menu with <F1> (short keystroke).
-Open the Storage & config menu with the sensor-independent
settings by pressing <F1_> (long keystroke, approx. 2 s).
-Change the display in the measurement window by pressing <M>
(e.g. conductivity −> resistivity −> −> −> ).
Menus and dialogs The menus for settings and dialogs in procedures contain further
submenus. The selection is made with the <><> keys. The current
selection is displayed with a frame.
-Submenus
The name of the submenu is displayed at the upper edge of the
frame. Submenus are opened by confirming with <ENTER>.
Example:
-Settings
Settings are indicated by a colon. The current setting is displayed on
the right-hand side. The setting mode is opened with <ENTER>.
Subsequently, the setting can be changed with <><> and
<ENTER>. Example:
General
Clock
Service information
Reset
System
22.09.2008
08:00
Back
Language: Deutsch
Beep:Off
Illumination:On
Contrast:48 %
Switchoff time:30 min
General
22.09.2008
08:00
Back
Operation Cond 3210
18 ba75794e05 06/2013
-Functions
Functions are designated by the name of the function. They are
immediately carried out by confirming with <ENTER>.
Example: Display the Calibration record function.
Messages Information is marked by the
i
symbol. It cannot be selected.
Example:
Note
The principles of navigation are explained in the two following sections
by reference of examples:
-Setting the language (section 4.2.3)
-Setting the date and time (see section 4.2.4).
Calibration record
Calibration data storage
Calibration interval:150 d
ϰ
22.09.2008
08:00
Back
Measuring cell
Temp. comp. (TC)
TDS factor 1.00
Stability control On
Temperature unit °C
i
ϰ = 1413 µS/cm
ϰ
22.09.2008
08:00
Back
Cond 3210 Operation
19
ba75794e05 06/2013
4.2.3 Example 1 on navigation: Setting the language
1Press the <On/Off> key.
The measured value display appears.
The instrument is in the measuring mode.
2Using <F1_>/[Menu] open the Storage & config menu.
The instrument is in the setting mode.
3Select the System submenu with <><>.
The current selection is displayed with a frame.
4 Open the System submenu with <ENTER>.
ϰ
1417
µS/cm
25.0
°C
[Tr25][nLF][ 0.835 1/cm]
22.09.2008
08:00
Menu
System
Data storage
Storage & config
22.09.2008
08:00
Back
Operation Cond 3210
20 ba75794e05 06/2013
5Select the General submenu with <><>.
The current selection is displayed with a frame.
6 Open the General submenu with <ENTER>.
7Open the setting mode for the Language with <ENTER>.
8Select the required language with <><>.
General
Clock
Service information
Reset
System
22.09.2008
08:00
Back
Language: Deutsch
Beep:Off
Illumination:On
Contrast:48 %
Switchoff time:30 min
General
22.09.2008
08:00
Back
Language: Deutsch
Beep:Off
Illumination:On
Contrast:48 %
Switchoff time:30 min
General
22.09.2008
08:00
Back
Cond 3210 Operation
21
ba75794e05 06/2013
9 Confirm the setting with <ENTER>.
The meter switches to the measuring mode.
The selected language is active.
Operation Cond 3210
22 ba75794e05 06/2013
4.2.4 Example 2 on navigation: Setting the date and time
The measuring instrument has a clock with a date function. The date
and time are indicated in the status line of the measured value display.
When storing measured values and calibrating, the current date and
time are automatically stored as well.
The correct setting of the date and time and date format is important for
the following functions and displays:
-Current date and time
-Calibration date
-Identification of stored measured values.
Therefore, check the time at regular intervals.
Note
After a fall of the supply voltage (empty batteries), the date and time are
reset to 01.01.2008, 00:00 hours.
Setting the date, time
and date format
The data format can be switched from the display of day, month, year
(dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or
mm.dd.yyyy).
1In the measured value display:
Using <F1_>/[Menu], open the Storage & config menu.
The instrument is in the setting mode.
2 Select and confirm the System / Clock menu with <><> and
<ENTER>.
The setting menu for the date and time opens up.
3Select and confirm the Time menu with <><> and
<ENTER>.
The hours are highlighted.
Date format:dd.mm.yyyy
Date:30.10.2008
Time:14:53:40
Clock
22.09.2008
08:00
Back
Cond 3210 Operation
23
ba75794e05 06/2013
4Change and confirm the setting with <><> and <ENTER>.
The minutes are highlighted.
5 Change and confirm the setting with <><> and <ENTER>.
The seconds are highlighted.
6Change and confirm the setting with <><> and <ENTER>.
The time is set.
7 If necessary, set the Date and Date format. The setting is made
similarly to that of the time.
8To make further settings, switch to the next higher menu level
with [Back]<F1>.
or
Switch to the measured value display with <M>.
The instrument is in the measuring mode.
Operation Cond 3210
24 ba75794e05 06/2013
4.3 Sensor-independent settings
The Storage & config menu comprises the following settings:
-System (see section 4.3.1).
-Data storage (see section 4.3.2)
4.3.1 System
Overview The following sensor-independent meter characteristics can be
adjusted in the Storage & config/System menu:
-Menu language
-Illumination
-Display contrast
-Interval of the automatic switch-off
-Clock and date function
-Reset of all sensor-independent system settings to the default
condition
Settings To open the Storage & config menu, press the <F1_>[Menu] key in
the measured value display. After completing the settings, switch to the
measured value display with <M>.
Menu item Setting Description
System / General /
Language
Deutsch
English
(further)
Select the menu language
System / General /
Illumination
Auto
On
Off
Switching the display
illumination on/off
System / General /
Contrast
0 ... 100 % Changing the display
contrast
System / General /
Switchoff time
10 min ... 24 h Adjust the switch-off time
System / Clock Time
Date
Date format
Settings of time and date.
For details, see section
4.2.4
System / Service
information
Hardware version and
software version of the
meter are displayed.
Cond 3210 Operation
25
ba75794e05 06/2013
4.3.2 Data storage
This menu contains all functions to display, edit and erase stored
measured values and calibration records.
Note
Detailed information on the storage functions of the Cond 3210 is given
in section 4.5.
System / Reset - Resets the system settings
to the default values.
For details, see section
4.6.2
Menu item Setting Description
Operation Cond 3210
26 ba75794e05 06/2013
4.3.3 Automatic Stability control
The automatic Stability control (AutoRead) function continuously
checks the stability of the measurement signal.
You can activate or switch off the automatic Stability control function
(see section 4.4.4).
The measured parameter flashes on the display
-as soon as the measured value is outside the stability range
-if you switch over between the measured parameters with <M>.
-if the automatic Stability control is switched off.
Cond 3210 Operation
27
ba75794e05 06/2013
4.4 Conductivity
4.4.1 General information
You can measure the following variables:
-Conductivity
-Specific resistance
-Salinity
-Total dissolved solids (TDS)
The measuring instrument is supplied with the following functions:
-AutoRange (automatic switchover of the measurement range). If a
measuring range is exceeded, AutoRange causes the measuring
instrument to automatically change to the next higher measuring
range and back again. Therefore, the instrument always measures
in the measuring range with the highest possible resolution.
-The AutoRead (stability control) function for checking the stability of
the measurement signal. The display of the measured parameter
flashes until a stable measured value is available.
Temperature
measurement
The TetraCon 325, KLE 325, LR 325/01 and LR 325/001 conductivity
measuring cells have an integrated temperature sensor.
Preparatory activities Perform the following preparatory activities when you want to measure:
Note
The selection of the measuring cell and the setting of the cell constant
is done in the measurement settings menu for conductivity (see section
4.4.4). The cell constant to be set must either be taken from the
operating manual of the measuring cell or is printed on the measuring
cell.
1Connect a conductivity measuring cell to the measuring
instrument.
The conductivity measuring window is displayed.
2 Check whether the Measuring cell and cell constant settings
are suitable for the connected conductivity measuring cell. If
necessary, correct the settings.
Operation Cond 3210
28 ba75794e05 06/2013
4.4.2 Measuring
You can carry out conductivity measurements as follows:
Selecting the
displayed
measured variable
You can switch between the following displays with <M>:
-Conductivity [μS/cm] / [mS/cm]
-Resistivity [Ω·cm] / [kΩ·cm] / [MΩ·cm]
-Salinity SaL [ ]
-Total dissolved solids TDS [mg/l] / [g/l]
The factor to calculate the total dissolved solids is set to 1.00 in the
factory. You can adjust this factor to meet your requirements in the
range of 0.40 to 1.00. The factor can be set in the TDS measuring
menu.
Freezing the measured
value (HOLD function)
With the HOLD function, you can freeze the current measured value.
The displayed measured value stops changing until you switch the
HOLD function off.
Note
If the HOLD function is active, you can, e.g. start a manual
measurement with stability control.
1Perform the preparatory activities according to section 4.4.1.
2 Immerse the conductivity measuring cell in the test sample.
ϰ
1417
µS/cm
25.0
°C
[Tr25][nLF][ 0.835 1/cm]
22.09.2008
08:00
Menu
1Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
2Release the frozen measured value again with <AR>.
The HOLD function is switched off.
The [HOLD] status display disappears.
Cond 3210 Operation
29
ba75794e05 06/2013
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
The measured parameter flashes on the display
-as soon as the measured value is outside the stability range
-when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of
the setting for automatic Stability control (see page 26) in the
Measurement menu.
Note
You can prematurely terminate the Stability control function manually
with <ENTER> at any time.
Criteria for a stable
measured value
The Stability control function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
1Freeze the measured value with <AR>.
The [HOLD] status indicator is displayed.
The HOLD function is active.
2 Using <ENTER>, activate the Stability control function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized.
3Using <ENTER>, start a further measurement with Stability
control.
or
Release the frozen measured value again with <AR>.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Measured
parameter
Time interval Stability in the time interval
Conductivity ϰ10 seconds Δ : Better than
1.0% of measured value
Temperature 15 seconds Δ : Better than 0.5 °C
Operation Cond 3210
30 ba75794e05 06/2013
4.4.3 Temperature compensation
The calculation of the temperature compensation is based on the
preset reference temperature, 20 °C or 25 °C. It appears on the display
as Tr20 or Tr25.
You can select one of the following temperature compensation
methods:
-Nonlinear temperature compensation (nLF)
according to EN 27 888
-Linear temperature compensation (lin)
with adjustable coefficients of 0.000 ... 3.000 %/K
-No temperature compensation (off)
Note
The reference temperature and temperature compensation are set in
the conductivity measuring menu (see section 4.4.4).
Application tips Select the following temperature compensations given in the table
according to the respective test sample:
Test sample Temperature compensation Display
indicator
Natural water
(ground water, surface
water, drinking water)
nLF
according to EN 27 888
nLF
Ultrapure water nLF
according to EN 27 888
nLF
Other aqueous
solutions
lin
Set linear temperature
coefficient
0.001 ... 3.000 %/K
lin
Salinity (seawater) Automatic nLF according to
IOT
Sal, nLF
Cond 3210 Operation
31
ba75794e05 06/2013
4.4.4 Settings for conductivity measuring cells
Overview The following settings are possible for conductivity measuring cells:
-Calibration record (display)
-Calibration interval
-Measuring cell/cell constant
-Reference temperature
-Temperature compensation
-TDS factor
-Unit of the temperature
-Automatic Stability control
Settings The settings are made in the measuring menu of the conductivity
measurement. To open the settings, activate the relevant measuring
window in the measured value display and press the <F1>/[menu] or
<ENTER> key shortly. After completing the settings, switch to the
measured value display with <M>.
Menu item Possible
setting
Description
Calibration /
Calibration record
- Displays the calibration
record of the last calibration.
Calibration /
Calibration interval
1 ... 999 d Calibration interval for the
measuring cell (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.
Operation Cond 3210
32 ba75794e05 06/2013
Measurement /
Measuring cell /
Type
Measuring cell used
Cal Measuring cells the cell
constant of which is
determined by calibration in
the KCL control standard
solution.
Calibration ranges:
0.450 to 0.500 cm-1 and
0.800 to 0.880 cm-1
The currently valid cell
constant is displayed in the
status line.
LR325/01 LR 325/01 measuring cell,
nominal cell constant
0.100 cm-1.
The cell constant can be
adjusted in the range from
0.090 to 0.110 cm-1.
LR325/001 LR 325/001 measuring cell,
nominal cell constant
0.010 cm-1.
The cell constant is
permanently set.
man Any measuring cells with
freely adjustable cell
constants in the range from
0.250 to 25.000 cm-1.
Measurement /
Measuring cell /
Cell const. man
0.250 to
25.000 cm-1 Display and setting option of
the cell constant of any
measuring cells (man).
Measurement /
Measuring cell /
Cell const. LR325/01
0.090 to
0.110 cm-1 Display and setting option of
the cell constant of the
LR 325/01 measuring cell.
Measurement /
Temp. comp. (TC) /
Method
nLF
lin
Off
Procedure for temperature
compensation (see section
4.4.3).
This setting is only available
for the measured
parameters, ϰ und ρ .
Menu item Possible
setting
Description
Cond 3210 Operation
33
ba75794e05 06/2013
Measurement /
Measuring cell /
Type
Measuring cell used
Cal Measuring cells the cell
constant of which is
determined by calibration in
the KCL control standard
solution.
Calibration ranges:
0.450 to 0.500 cm-1 and
0.800 to 0.880 cm-1
The currently valid cell
constant is displayed in the
status line.
LR325/01 LR 325/01 measuring cell,
nominal cell constant
0.100 cm-1.
The cell constant can be
adjusted in the range from
0.090 to 0.110 cm-1.
LR325/001 LR 325/001 measuring cell,
nominal cell constant
0.010 cm-1.
The cell constant is
permanently set.
man Any measuring cells with
freely adjustable cell
constants in the range from
0.250 to 25.000 cm-1.
Measurement /
Measuring cell /
Cell const. man
0.250 to
25.000 cm-1 Display and setting option of
the cell constant of any
measuring cells (man).
Measurement /
Measuring cell /
Cell const. LR325/01
0.090 to
0.110 cm-1 Display and setting option of
the cell constant of the
LR 325/01 measuring cell.
Measurement /
Temp. comp. (TC) /
Method
nLF
lin
Off
Procedure for temperature
compensation (see section
4.4.3).
This setting is only available
for the measured
parameters, ϰ und ρ .
Menu item Possible
setting
Description
Operation Cond 3210
34 ba75794e05 06/2013
Measurement /
Temp. comp. (TC) /
Linear coeff.
0.000 ...
3.000 %/K
Coefficient of the linear
temperature compensation.
This setting is only available
when the linear temperature
compensation is set.
Measurement /
Temp. comp. (TC) /
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available
for the measured
parameters, ϰ und ρ .
Measurement /TDS
factor
0.40 ... 1.00 Factor for TDS value
Measurement /
Stability control
On /Off Switches on or off the
automatic stability control
during measurement (see
section 4.3.3)
Measurement /
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperatures are
displayed with the selected
unit.
Reset - Resets all sensor settings to
the delivery condition (see
section 4.6.1).
Menu item Possible
setting
Description
Cond 3210 Operation
35
ba75794e05 06/2013
4.4.5 Determining the cell constant
(calibration in control standard)
Why determine the cell
constant?
Aging slightly changes the cell constant, e. g. by coatings. As a result,
an inexact measured value is displayed. The original characteristics of
the cell can often be restored by cleaning the cell. Calibration
determines the current value of the cell constant and stores this value
in the meter.
Thus, you should calibrate at regular intervals (we recommend: every
6 months).
Procedure You can determine the actual cell constant of the conductivity
measuring cell by calibrating with the control standard in the following
ranges:
- 0.450 ... 0.500 cm-1
(e.g. TetraCon 325, nominal cell constant 0.475)
- 0.800 ... 0.880 cm-1
(cells with a cell constant of approx. 0.840)
The cell constant is determined in the control standard, 0.01 mol/l KCl.
Cell constants outside the ranges quoted above cannot be calibrated.
In the delivery condition, the calibrated cell constant of the meter is set
to 0.475 cm-1 (conductivity measuring cell, TetraCon 325).
AutoRead
(stability control)
The calibration procedure automatically activates the AutoRead
(stability control) function.
Display calibration data You can have the data of the last calibration displayed (see section
4.4.6).
Calibration evaluation After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the
display and in the calibration record.
Display Calibration
record
Cell constant [cm-1]
+++ within the ranges
0.450 ... 0.500 cm-1
or
0.800 ... 0.880 cm-1
Error Error outside the ranges
0.450 ... 0.500 cm-1
or
0.800 ... 0.880 cm-1
Eliminate the error according to
chapter 6 WHAT TO DO IF...
Operation Cond 3210
36 ba75794e05 06/2013
Determining the cell
constant
For this calibration procedure, the Measuring cell setting must be set to
cal in the measuring menu. Proceed as follows to determine the cell
constant:
4.4.6 Displaying calibration records
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the <CAL_> key.
The calibration records of the last calibrations (up to 10) are available
in the menu, <F1>/[Menu] / Calibration / Calibration data storage and
in the menu, <F1_>/[Menu] / Storage & config/Data storage / Calibra-
tion data storage.
1Connect a conductivity measuring cell to the measuring
instrument.
2 In the measured value display, select the conductivity
parameter with <M>.
3Start the calibration with <CAL>.
The cell constant that was calibrated last is displayed.
4Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
5 Start the measurement with <ENTER>.
The [AR] status indicator is displayed. The measured
parameter flashes.
6Wait for the end of the AutoRead measurement or accept the
calibration value with <ENTER>.
The calibration record is displayed.
7Using <F1>/[Continue] or <ENTER>, switch to the measured
value display.
ϰ
0.839
1/cm
25.1
°C
[Tr25][nLF][ 0.835 1/cm]
22.09.2008
08:00
Menu
Cond 3210 Operation
37
ba75794e05 06/2013
Operation Cond 3210
38 ba75794e05 06/2013
4.5 Data storage
You can transmit measured values (datasets) to the data storage:
-Manual storage (see section 4.5.1)
Measurement dataset A complete dataset consists of:
-ID number
-Date/time
-Measured value of the connected sensor
-Measured temperature value of the connected sensor
-AutoRead info: AR appears with the measured value if the
AutoRead criterion was met while storing (stable measured value).
Otherwise, the AR display is missing.
-Calibration evaluation: +++, ++, +, -, or no evaluation
Storage locations The Cond 3210 meter has a measurement data storage.
Storage Maximum number of datasets
Manual data storage 200
Cond 3210 Operation
39
ba75794e05 06/2013
4.5.1 Manual storage
You can transmit a measurement dataset to the data storage as
follows.
If the storage is full The following window appears if all 200 storage locations are occupied:
You have the following options:
-To erase the entire storage, confirm Yes.
-To cancel the storage process and switch to the measured value
display, confirm No.
4.5.2 Displaying and editing the measurement data storage
The contents of the manual measurement data storage can be shown
on the display.
The measurement data storage has a function to erase the entire
1Press the <STO> key shortly.
The menu for manual storage appears.
2If necessary, change and confirm the ID number (1 ... 10000)
with <><> and <ENTER>.
The dataset is stored. The meter switches to the measured
value display.
30.10.2008 11:24:16
pH 7.000 24.8 °C AR +++
ID number: 1
Continue
Manual data storage 4 From 200
22.11.2010
08:00
Back
Data storage full. Erase?
Yes
No
Warning
22.09.2008
08:00
Back
Operation Cond 3210
40 ba75794e05 06/2013
contents.
Editing the data storage The storage is edited in the menu, Storage & config/ Data storage. To
open the Storage & config menu, press the <F1_>[Menu] key in the
measured value display. Open the manual storage directly with the
<RCL> key.
Settings
Display presentation
of a dataset
Quitting the display To quit the display of stored measurement datasets, you have the
following options:
-Switch directly to the measured value display with <M>.
-Quit the display and move to the next higher menu level with <F1>/
[Back].
Menu item Setting/
function
Description
Data storage /
Manual data storage /
Display
- Displays all measurement
datasets page by page.
Further options:
-Scroll through the
datasets with <><>.
-Quit the display with
<F1>/[Back].
Data storage /
Manual data storage /
Erase
- Erases the entire manual
measurement data
storage.
Note:
All calibration data remains
stored when this action is
performed.
Manual data storage 3 of 64
30.10.2008 11:24:16 ID number: 1
ϰ 1415 µS/cm 25.1 °C AR +++
C = 0.835 1/cm, Tref 25, nLF
22.09.2008
08:00
Back
Cond 3210 Operation
41
ba75794e05 06/2013
4.5.3 Erasing the measurement data storage
How to erase the measurement data storage is described in section
4.5.2 DISPLAYING AND EDITING THE MEASUREMENT DATA STORAGE.
Operation Cond 3210
42 ba75794e05 06/2013
4.6 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.6.1 Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
The following settings for conductivity measurements are reset to the
default settings with the Reset function:
The sensor settings are reset under the Reset menu item in the
measuring menu. To open the settings, activate the relevant measuring
window in the measured value display and press the <F1>[Menu] key
shortly.
Setting Default settings
Cal. interval 150 d
Measured parameter ϰ
Cell constant (c) 0.475 cm-1 (calibrated)
0.475 cm-1 (set up)
Temperature compensation nLF
Reference temperature 25 °C
Temperature coefficient (TC) of
the linear temperature
compensation
2.000 %/K
TDS factor 1.00
Stability control On
Temperature unit °C
Cond 3210 Operation
43
ba75794e05 06/2013
4.6.2 Resetting the system settings
The following system settings can be reset to the delivery status:
The system settings are reset in the menu, Storage & config / System
/ Reset. To open the Storage & config menu, press the <F1_>[Menu]
key in the measured value display.
Setting Default settings
Language English
Beep On
Contrast 50 %
Illumination On
Switchoff time 1 h
Operation Cond 3210
44 ba75794e05 06/2013
Cond 3210 Maintenance, cleaning, disposal
45
ba75794e05 06/2013
5 Maintenance, cleaning, disposal
5.1 Maintenance
The only maintenance activity required is replacing the batteries.
Note
See the relevant operating manuals of the measuring cells for
instructions on maintenance.
5.1.1 Replacing the batteries
Note
Alternatively, you can also use Ni-MH rechargeable batteries (type
Mignon AA). In order to charge the batteries, an external charging
device is required.
Caution
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the ± signs in the
battery compartment.
1Unscrew the two screws (1) on the underside of the meter.
2 Open the battery compartment (2) on the underside of the
meter.
3Remove the four batteries from the battery compartment.
4 Place four new batteries (type Mignon AA) in the battery
compartment.
2
1
Maintenance, cleaning, disposal Cond 3210
46 ba75794e05 06/2013
5.2 Cleaning
Occasionally wipe the outside of the measuring instrument with a
damp, lint-free cloth. Disinfect the housing with isopropanol as
required.
Caution
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain solvents.
Remove any splashes immediately.
5.3 Packing
This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing
protects the meter against damage during transport.
5.4 Disposal
Note
This meter contains batteries. Batteries that have been removed must
only be disposed of at a recycling facility set up for this purpose or via
the retail outlet.
It is illegal to dispose of it in household refuse.
5Close the battery compartment (2) and tighten the screws (1).
Cond 3210 What to do if...
47
ba75794e05 06/2013
6 What to do if...
Error message,
OFL, UFL
Error message,
Error
Sensor symbol flashes
Display
Meter does not react to
keystroke
You want to know which
software
version is in the meter
Cause Remedy
Measured value outside the
measuring range
Use suitable measuring cell
Cause Remedy
Measuring cell contaminated Clean cell and replace it if
necessary
Calibration solution not suitable Check calibration solutions
Measuring cell not connected Connect measuring cell
Cause Remedy
Cleaning interval expired Recalibrate the measuring
system
Cause Remedy
batteries almost empty Replace the batteries (see
section 5.1 MAINTENANCE)
Cause Remedy
Operating condition undefined
or EMC load unallowed
Processor reset:
Press the <ENTER> and
<On/Off> key
simultaneously
Cause Remedy
E. g., a question by the service
department
Switch on the meter.
Open the menu,
<F1_>[Menu] / Storage &
config / System / Service
information. The instrument
data is displayed.
What to do if... Cond 3210
48 ba75794e05 06/2013
Cond 3210 Technical data
49
ba75794e05 06/2013
7 Technical data
7.1 General data
Dimensions approx. 180 x 80 x 55 mm
Weight approx. 0.4 kg
Mechanical structure Type of protection IP 67
Electrical safety Protective class III
Test certificates CE, cETLus
Ambient
conditions
Storage - 25 °C ... + 65 °C
Operation -10 °C ... + 55 °C
Allowable relative
humidity
Annual mean: < 75 %
30 days/year: 95 %
Other days: 85 %
Power
supply
Batteries 4 x 1.5 V alkali-manganese batteries, type
AA
Rechargeable
batteries
4 x 1,2 V NiMH rechargeable batteries,
type AA
(no charging function)
Operational life up to 800 h without / 100 h with illumination
Guidelines
and norms used
EMC EC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A
Meter safety EC directive 2006/95/EC
EN 61010-1
IP protection class EN 60529
Technical data Cond 3210
50 ba75794e05 06/2013
7.2 Measuring ranges, resolution, accuracy
Measuring ranges,
resolution
Parameter Measuring range Resolution
ϰ [µS/cm] 0.000 ... 1.999*
0.00 ... 19.99**
0.0 ... 199.9
200 ... 1999
0.001
0.01
0.1
1
ϰ [mS/cm] 2.00 ... 19.99
20.0 ... 199.9
200 ... 1000
0.01
0.1
1
ρ (Resistivity)
[Ohm*cm]
1,000 ... 1,999
2,00 ... 19,99
20,0 ... 199,9
200 ... 1999
0,001
0,01
0,1
1
ρ (Resistivity)
[kOhm*cm]
2,00 ... 19,99
20,0 ... 199,9
200 ... 1999
0,01
0,1
1
ρ (Resistivity)
[MOhm*cm]
2,00 ... 19,99 **
20,0 ... 199,9 *
0,01
0,1
SAL 0.0 ... 70.0
according to the IOT
table
0.1
TDS 0 ... 1999 mg/l
2.00 ... 19.99 g/l
20.0 ... 199.9 g/l
1
0.01
0.1
T [°C] 5.0 ... + 105.0 0.1
T [°F] + 23.0 ... + 221.0 0.1
* only possible with cells of the cell constant, 0.010 cm-1
** only possible with cells of the cell constant, 0.010 cm-1 or 0.090 ... 0.110 cm-1
Cell constants Cell constant C Values
Can be calibrated in
the ranges
0.450 ... 0.500 cm-1
0.800 ... 0.880 cm-1
Adjustable 0.010 cm-1 (fixed)
0.090 ... 0.110 cm-1
0.250 ... 25.000 cm-1
Reference temperature Reference
temperature
Values
Adjustable 20 °C (Tref20)
25 °C (Tref25)
Cond 3210 Technical data
51
ba75794e05 06/2013
Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the measuring cell has also to be taken into account.
Accuracy (± 1 digit) Parameter Accuracy Temperature of the test
sample
ϰ and
ρ
/ temperature compensation
None (Off) ± 0.5 %
Nonlinear (nLF) ± 0.5 % 0 °C ... + 35 °C
according to EN 27 888
± 0.5 % + 35 °C ... + 50 °C
enhanced nLF function
Linear (lin) ± 0.5 % + 10 °C ... + 75 °C
SAL / range
0.0 ... 42.0 ± 0.1 + 5 °C ... + 25 °C
± 0.2 + 25 °C ... + 30 °C
TDS [mg/l]
± 0.5 %
T [°C] / temperature sensor
NTC 30 ± 0.2
PT 1000 ± 0.3
Technical data Cond 3210
52 ba75794e05 06/2013
FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is
operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to
correct the interference at his own expense.
Changes or modifications not expressly approved by the
manufacturer could void the user‘s authority to operate the
equipment.
Cond 3210 Lists
53
ba75794e05 06/2013
8Lists
This chapter provides additional information and orientation aids.
Specialist terms The glossary briefly explains the meaning of the specialist terms.
However, terms that should already be familiar to the target group are
not described here.
Index The index will help you to find the topics that you are looking for.
Glossary
Adjusting To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains
within the tolerance.
AutoRange Name of the automatic selection of the measuring range.
Calibration Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct.
Often, this expression is also used when the measuring system is
adjusted at the same time (see adjusting).
Cell constant, k Characteristic quantity of a conductivity measuring cell, depending on
the geometry.
Conductivity Short form of the expression, specific electrical conductivity.
It corresponds to the reciprocal value of the resistivity.
It is a measured value of the ability of a substance to conduct an
electric current. In water analysis, the electrical conductivity is a
dimension for the ionized substances in a solution.
Measured parameter The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D.O. concentration.
Measured value The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).
Molality Molality is the quantity (in Mol) of a dissolved substance in 1000 g
solvent.
Lists Cond 3210
54 ba75794e05 06/2013
Reference
temperature
Fixed temperature value to compare temperature-dependent
measured values. For conductivity measurements, the measured
value is converted to a conductivity value at a reference temperature
of 20 °C or 25 °C.
Reset Restoring the original condition of all settings of a measuring system.
Resistance Short name for the electrolytic resistivity. It corresponds to the
reciprocal value of the electrical conductivity.
Resolution Smallest difference between two measured values that can be
displayed by a meter.
Salinity The absolute salinity SA of seawater corresponds to the relationship
of the mass of dissolved salts to the mass of the solution (in g/Kg). In
practice, this dimension cannot be measured directly. Therefore, the
practical salinity according to IOT is used for oceanographic
monitoring. It is determined by measuring the electrical conductivity.
Salt content General designation for the quantity of salt dissolved in water.
Stability control Function to control the measured value stability.
Standard solution The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Temperature
coefficient
Value of the slope of a linear temperature function.
Temperature
compensation
Name of a function that considers the temperature influence on the
measurement and converts it accordingly. Depending on the
measured parameter to be determined, the temperature
compensation functions in different ways. For conductimetric
measurements, the measured value is converted to a defined
reference temperature. For potentiometric measurements, the slope
value is adjusted to the temperature of the test sample but the
measured value is not converted.
Temperature function Name of a mathematical function expressing the temperature
behavior of a test sample, a probe or part of a probe.
Test sample Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
α
TRef
=
Meas*1
1+ a(T - )
*TRef
Cond 3210 Lists
Index
55
ba75794d05 06/2013
A
Authorized use . . . . . . . . . . . . . . . . . . . . . . 10
Automatic switch-off . . . . . . . . . . . . . . . . . . 15
AutoRead . . . . . . . . . . . . . . . . . . . . . . . . . . 29
B
Battery compartment . . . . . . . . . . . . . . 13, 45
C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . 35
Calibration evaluation . . . . . . . . . . . . . . . . 35
Cell constant . . . . . . . . . . . . . . . . . . . . . . . 35
Control standard . . . . . . . . . . . . . . . . . . . . 35
Copyright . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
D
Dataset . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Date and time . . . . . . . . . . . . . . . . . . . . . . . 22
Default settings
Measured parameter . . . . . . . . . . . . . . 42
System settings . . . . . . . . . . . . . . . . . . 43
E
Energy saving feature . . . . . . . . . . . . . . . . 14
I
Initial commissioning . . . . . . . . . . . . . . 13, 14
Initialize . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
K
Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
M
Measured value display . . . . . . . . . . . . . . . 17
Measurement data memory
Edit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Erase . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Storage locations . . . . . . . . . . . . . . . . . 38
Measurement dataset . . . . . . . . . . . . . . . . 38
Measuring . . . . . . . . . . . . . . . . . . . . . . . . . 28
Measuring menu . . . . . . . . . . . . . . . . . . . . 31
Menus (navigation) . . . . . . . . . . . . . . . . . . 17
Messages . . . . . . . . . . . . . . . . . . . . . . . . . . 18
O
Operational safety . . . . . . . . . . . . . . . . . . . 10
P
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . 9
R
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
S
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Sample display . . . . . . . . . . . . . . . . . . . . . . 7
Saving . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Scope of delivery . . . . . . . . . . . . . . . . . . . . 13
Setting the date . . . . . . . . . . . . . . . . . . . . . 14
Setting the time . . . . . . . . . . . . . . . . . . . . . 14
Socket field . . . . . . . . . . . . . . . . . . . . . . . . . 8
Stability control
automatic . . . . . . . . . . . . . . . . . . . . . . . 26
T
Temperature compensation . . . . . . . . . . . . 30
Temperature measurement . . . . . . . . . . . . 27
Lists Cond 3210
56 ba75794d05 06/2013
Wissenschaftlich-Technische
Werkstätten GmbH
Dr.-Karl-Slevogt-Straße 1
D-82362 Weilheim
Germany
+49 881 183-0
+49 881 183-420
info@wtw.com
www.wtw.com
Tel:
Fax:
E-Mail:
Internet:
1) The tissue in plants that brings water upward from the roots;
2) a leading global water technology company.
We're 12,500 people unified in a common purpose: creating innovative solutions
to meet our world's water needs. Developing new technologies that will improve
the way water is used, conserved, and re-used in the future is central to our work.
We move, treat, analyze, and return water to the environment, and we help people
use water efficiently, in their homes, buildings, factories and farms. In more than
150 countries, we have strong, long-standing relationships with customers who
know us for our powerful combination of leading product brands and applications
expertise, backed by a legacy of innovation.
For more information on how Xylem can help you, go to www.xyleminc.com
Xylem | m|ˈəzīl

Navigation menu