WTW PH 3210 Operating Manual WTW_p H_3210 P H

User Manual: Pdf WTW_pH_3210

Open the PDF directly: View PDF PDF.
Page Count: 68

DownloadWTW PH 3210 Operating Manual WTW_p H_3210 P H
Open PDF In BrowserView PDF
Operating manual

pH 3210

pH

Men
u

6.09

3

15.1
1
08:3.2008
7

25.0

°C

pH meter
For Sales & Service Contact

2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242

1-800-833-7958
ba75793e05

09/2011

www.geotechenv.com

pH 3210

Accuracy when
going to press

Copyright

The use of advanced technology and the high quality standard of our
instruments are the result of a continuous development. This may result in
differences between this operating manual and your meter. Also, we cannot
guarantee that there are absolutely no errors in this manual. Therefore, we
are sure you will understand that we cannot accept any legal claims resulting
from the data, figures or descriptions.

© Weilheim 2008, WTW GmbH
Reproduction in whole - or even in part - is prohibited without the express
written permission of WTW GmbH, Weilheim.
Printed in Germany.

ba75793e05

09/2011

pH 3210

Contents

pH 3210 - Contents

1

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.1
1.2
1.3

2

Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1
2.2

3

4.3

4.4

09/2011

Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Initial commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2.1 Inserting the batteries. . . . . . . . . . . . . . . . . . . . .
3.2.2 Switching on the meter. . . . . . . . . . . . . . . . . . . .
3.2.3
Setting the date and time . . . . . . . . . . . . . . . . .

13
13
13
14
14

Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1
4.2

ba75793d05

Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 10

Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1
3.2

4

Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Socket field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Switching on the meter . . . . . . . . . . . . . . . . . . . . . . . . . .
General operating principles . . . . . . . . . . . . . . . . . . . . . .
4.2.1 Operating modes . . . . . . . . . . . . . . . . . . . . . . . .
4.2.2 Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.2.3 Example 1 on navigation: Setting the language.
4.2.4 Example 2 on navigation: Setting the date and
time
................................
Sensor-independent settings . . . . . . . . . . . . . . . . . . . . .
4.3.1 System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.3.2 Data storage. . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.3.3 Automatic Stability control . . . . . . . . . . . . . . . . .
pH value / ORP voltage. . . . . . . . . . . . . . . . . . . . . . . . . .
4.4.1 General information . . . . . . . . . . . . . . . . . . . . . .
4.4.2 Measuring the pH value . . . . . . . . . . . . . . . . . . .
4.4.3 Measuring the ORP . . . . . . . . . . . . . . . . . . . . . .
4.4.4 Settings for pH and ORP measurements. . . . . .
4.4.5 pH calibration . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.4.6 Calibration interval . . . . . . . . . . . . . . . . . . . . . . .
4.4.7 Carrying out an automatic calibration (AutoCal)
4.4.8 Carrying out a manual calibration (ConCal) . . . .
4.4.9 Displaying calibration records . . . . . . . . . . . . . .
4.4.10 Continuous measurement control (CMC
function)
............................

15
16
16
17
19
22
24
24
25
26
27
27
28
30
32
34
38
39
42
45
46

3

Contents

pH 3210

4.5

4.6

5

47
48
48
50
51
51
52

Maintenance, cleaning, disposal . . . . . . . . . . . . . . . . . 53
5.1
5.2
5.3
5.4

Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1.1 Replacing the batteries . . . . . . . . . . . . . . . . . . . .
Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Packing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

53
53
54
54
54

6

What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55

7

Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
7.1
7.2

8

4

Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.5.1 Manual storage . . . . . . . . . . . . . . . . . . . . . . . . . .
4.5.2 Editing the measured value storage . . . . . . . . . .
4.5.3 Erasing the measurement data storage . . . . . . .
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4.6.1 Resetting the measurement settings . . . . . . . . .
4.6.2 Resetting the system settings . . . . . . . . . . . . . . .

General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Measuring ranges, resolution, accuracy . . . . . . . . . . . . . 60

Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63

ba75793d05

09/2011

pH 3210

Overview

1

Overview

The compact pH 3210 precision pH meter enables you to perform pH
measurements rapidly and reliably. The pH 3210 provides the
maximum degree of operating comfort, reliability and measuring
certainty for all applications.
The proven calibration procedures and automatic stability control
function (AR) support your work with the pH meter.

3

pH

Men
u

6.09

3

15.1
1
08:3.2008
7

25.0

2

°C

1

ba75793e05

09/2011

1

Keypad

2

Display

3

Socket field

5

Overview

pH 3210

1.1

Keypad

F1

F2

M
CAL

AR
RCL

STO

ENTER

pH 3210

In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. ) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
).

_

F2

Softkey providing situation dependent functions , e.g.:
/[Menu]: Opens the menu for measurement settings
/[Menu]: Opens the menu for system settings

:

Switches the meter on or off

:

Selects the measured parameter

STO

RCL

_

:
:

Calls up the calibration procedure
Displays the calibration data

:


Stores a measured value manually

:


Displays the manually stored measured values

<▲>:

Increments values, scrolls

<▼>:

Decrements values, scrolls

_

CAL

6

:
:
:
:

_
_

F1

ba75793e05

09/2011

pH 3210

Overview

ENTER

AR

:
:

Opens the menu for measurement settings / confirms
entries
Opens the menu for system settings



Freezes the measured value (HOLD function)
Switches the AutoRead measurement on or off

_

1.2

Display

4

3
14

0

pH

6.093

2

25.0 °C

1

HOLD

AR

Menu

Function display
indicators

ba75793e05

09/2011

22.09.2008
08:00

5
6
7
8

1

Status information

2

Measured value (with unit)

3

Measured parameter

4

Continuous measurement control (CMC function)

5

Sensor symbol (calibration evaluation, calibration
interval)

6

Measured temperature (with unit)

7

Status line

8

Softkeys and date + time

AutoCal
e.g. TEC

Calibration with automatic buffer recognition, e.g.
with the buffer set: Technical buffers

ConCal

Calibration with any buffers

Error

An error occurred during calibration

LoBat

Batteries are almost empty

AR

Stability control (AutoRead) is active

HOLD

Measured value is frozen ( key)

7

Overview

pH 3210

1.3

Socket field

1

4

2

3

Connectors:
1

pH electrode

2

Reference electrode

3

Temperature sensor

4

Service interface

Caution
Only connect sensors to the meter that cannot return any voltages
or currents that are not allowed (> SELV and > current circuit with
current limiting).
Almost all customary sensors fulfill these conditions.

8

ba75793e05

09/2011

pH 3210

Safety

2

Safety

This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating
manual before working with the meter.
The operating manual must always be available within the
vicinity of the meter.
Target group

Safety instructions

The meter was developed for work in the field and in the laboratory.
Thus, we assume that, as a result of their professional training and
experience, the operators will know the necessary safety precautions
to take when handling chemicals.

Safety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g.
"Caution") indicates the level of danger:

Warning
indicates instructions that must be followed precisely in order to
avoid possibly great dangers to personnel.

Caution
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the meter or
the environment.
Further notes
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating
manuals.

ba75793e05

09/2011

9

Safety

pH 3210

2.1

Authorized use

This meter is authorized exclusively for pH and ORP measurements in
a laboratory or field environment.
The technical specifications as given in chapter 7 TECHNICAL DATA must
be observed. Only the operation and running of the meter according to
the instructions given in this operating manual is authorized.
Any other use is considered unauthorized.

2.2

General safety instructions

This meter is constructed and tested in compliance with the IEC 1010
safety regulations for electronic measuring instruments.
It left the factory in a safe and secure technical condition.
Function and
operational safety

The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during
operation.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 TECHNICAL DATA.
If the meter was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty
functioning of the meter. In this event, wait until the temperature of the
meter reaches room temperature before putting the meter back into
operation.

Caution
The meter is only allowed to be opened by authorized personnel.

10

ba75793e05

09/2011

pH 3210

Safety

Safe operation

If safe operation is no longer possible, the meter must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
 has been damaged in transport
 has been stored under adverse conditions for a lengthy period of
time
 is visibly damaged
 no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the meter.

Obligations of the
purchaser

The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
 EEC directives for protective labor legislation
 National protective labor legislation
 Safety regulations
 Safety datasheets of the chemical manufacturers.

Caution
In addition to the safety instructions mentioned here, also follow
the safety instructions of the sensors used.
The operating manuals of the sensors are available on the
supplied CD and on the Internet under www.WTW.com.

ba75793e05

09/2011

11

Safety

12

pH 3210

ba75793e05

09/2011

pH 3210

Commissioning

3

Commissioning

3.1

Scope of delivery

 pH meter, pH 3210
 4 batteries 1.5 V Mignon type AA
 Short instructions
 CD-ROM with detailed operating manual

3.2

Initial commissioning

Perform the following activities:
 Insert the supplied batteries
 Switch on the meter
 Set the date and time

3.2.1

Inserting the batteries

1

Unscrew the two screws (1) on the underside of the meter.

2

Open the battery compartment (2) on the underside of the
meter.

1

2

3

ba75793e05

09/2011

Place four batteries (type Mignon AA) in the battery
compartment.

13

Commissioning

pH 3210

Note
Alternatively, you can also use Ni-MH rechargeable batteries (type
Mignon AA). In order to charge the batteries, an external charging
device is required.

Caution
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the ± signs in the
battery compartment.

4

3.2.2

1

Close the battery compartment (2) and tighten the screws (1).

Switching on the meter

Press the  key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test
is being performed.
Subsequently, the meter switches to the measuring mode
(measured value display).

Note
The meter has an energy saving feature to avoid unnecessary battery
depletion.
The energy saving feature switches off the meter if no key is pressed
during the adjusted interval. (How to set the switch-off interval, see
section 4.3.1).

3.2.3

1

14

Setting the date and time

See section 4.2.4

ba75793e05

09/2011

pH 3210

Operation

Switching on

4

Operation

4.1

Switching on the meter

Press the  key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test is being
performed.
The measured value display appears.

pH

6.949
25.0
Menu

Switching off

°C

22.11.2010
08:00

Press the  key.

Automatic switch-off

The instrument has an automatic switch-off function in order to save the
batteries (see section 4.3.1). The automatic switch-off switches off the
meter if no key is pressed for an adjustable period.

Display illumination

The meter automatically switches off the display illumination if no key
has been pressed for 15 seconds. The illumination is switched on with
the next keystroke again.
You can also generally switch the display illumination on or off (see
section 4.3.1).

ba75793e05

09/2011

15

Operation

pH 3210

4.2

General operating principles

This section contains basic information on the operation of the
pH 3210.
Operating elements,
display
Operating modes,
navigation

An overview of the operating elements and the display is given in
section 1.1 and section 1.2.
An overview of the operating modes and navigation of the pH 3210 is
given in section 4.2.1 and section 4.2.2.

4.2.1

Operating modes

The meter has the following operating modes:
 Measurement
The measurement data of the connected sensor is shown in the
measured value display
 Calibration
The course of a calibration with calibration information, functions
and settings is displayed
 Storage in memory
The meter stores measuring data manually or automatically
 Setting
The system menu or a sensor menu with submenus, settings and
functions is displayed

16

ba75793e05

09/2011

pH 3210

Operation

4.2.2
Measured value display

Navigation

In the measured value display, you can
 open the menu for calibration and measurement settings with 
(short keystroke).
 open the Storage & config menu with the sensor-independent
settings by pressing  (long keystroke, approx. 2 s on ).

_

 change the display in the selected measuring window
(e. g. pH <−> mV) by pressing .
Menus and dialogs

The menus for settings and dialogs in procedures contain further
submenus. The selection is made with the <▲><▼> keys. The current
selection is displayed with a frame.
 Submenus
The name of the submenu is displayed at the upper edge of the
frame. Submenus are opened by confirming with .
Example:
System
General
Measurement
Interface
Clock
Service information
Reset

Back

22.11.2010
08:00

 Settings
Settings are indicated by a colon. The current setting is displayed on
the right-hand side. The setting mode is opened with .
Subsequently, the setting can be changed with <▲><▼> and
. Example:
General
Language:
Beep:
Illumination:
Contrast:
Switchoff time:

Back

ba75793e05

09/2011

Deutsch
Off
On
50 %
1h

22.11.2010
08:00

17

Operation

pH 3210

 Functions
Functions are designated by the name of the function. They are
immediately carried out by confirming with .
Example: Display the Calibration record function.
pH
Calibration record
Data storage
Buffer:
One point calibration:
Calibration interval:
Unit for slope:

Back

Messages

TEC
Yes
7d
mV/pH

22.11.2010
08:00

Information is marked by the i symbol. It cannot be selected.
Example:
pH
Calibration record
Data storage
Buffer:
One point calibration:
Calibration interval:
Unit for slope:

TEC
Yes
7d
mV/pH

i 2.00 4.01 7.00 10.01
Back

22.11.2010
08:00

Note
The principles of navigation are explained in the two following sections
by reference of examples:
 Setting the language (section 4.2.3)
 Setting the date and time (see section 4.2.4).

18

ba75793e05

09/2011

pH 3210

Operation

4.2.3

1

Example 1 on navigation: Setting the language

Press the  key.
The measured value display appears.
The instrument is in the measuring mode.

pH

6.949
25.0
Menu

2

°C

22.11.2010
08:00

_

Using /[Menu] open the Storage & config menu.
The instrument is in the setting mode.

Storage & config
System
Data storage

Back

ba75793e05

09/2011

22.11.2010
08:00

3

Select the System submenu with <▲><▼>.
The current selection is displayed with a frame.

4

Open the System submenu with .

19

Operation

pH 3210

System
General
Measurement
Interface
Clock
Service information
Reset

Back

22.11.2010
08:00

5

Select the General submenu with <▲><▼>.
The current selection is displayed with a frame.

6

Open the General submenu with .

General
Language:
Beep:
Illumination:
Contrast:
Switchoff time:

Back

7

Deutsch
Off
On
50 %
1h

22.11.2010
08:00

Open the setting mode for the Language with .

General
Language:
Beep:
Illumination:
Contrast:
Switchoff time:

Back

8

20

Deutsch
Off
On
50 %
1h

22.11.2010
08:00

Select the required language with <▲><▼>.

ba75793e05

09/2011

pH 3210

Operation

9

ba75793e05

09/2011

Confirm the setting with .
The meter switches to the measuring mode.
The selected language is active.

21

Operation

pH 3210

4.2.4

Example 2 on navigation: Setting the date and time

The meter has a clock with a date function. The date and time are
indicated in the status line of the measured value display.
When storing measured values and calibrating, the current date and
time are automatically stored as well.
The correct setting of the date and time and date format is important for
the following functions and displays:
 Current date and time
 Calibration date
 Identification of stored measured values.
Therefore, check the time at regular intervals.
Note
After a fall of the supply voltage (empty batteries), the date and time are
reset to 01.01.2008 00, 00:00 hours.
Setting the date, time
and date format

The date format can be switched from the display of day, month, year
(dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or
mm.dd.yyyy).
1

In the measured value display:
Using /[Menu], open the Storage & config menu.
The instrument is in the setting operating mode.

2

Select and confirm the System / Clock menu with <▲><▼> and
.
The setting menu for the date and time opens up.

3

Select and confirm the Time menu with <▲><▼> and
.
The hours are highlighted.

_

Clock
Date format:
Date:
Time:

Back

22

dd.mm.yyyy
30.10.2008
14:53:40

22.11.2010
08:00

ba75793e05

09/2011

pH 3210

ba75793e05

Operation

09/2011

4

Change and confirm the setting with <▲><▼> and .
The minutes are highlighted.

5

Change and confirm the setting with <▲><▼> and .
The seconds are highlighted.

6

Change and confirm the setting with <▲><▼> and .
The time is set.

7

If necessary, set the Date and Date format. The setting is made
similarly to that of the time.

8

If necessary, select and set the Date with <▲><▼> and
.

9

To make further settings, switch to the next higher menu level
with /[Back].
or
Switch to the measured value display with .
The instrument is in the measuring mode.

23

Operation

pH 3210

4.3

Sensor-independent settings

The Storage & config menu comprises the following settings:
 System (see section 4.3.1).
 Data storage (see section 4.3.1)

4.3.1
Overview

System

The following sensor-independent meter characteristics can be
adjusted in the Storage & config/System menu:
 Menu language
 Illumination
 Display contrast
 Interval of the automatic switch-off
 Clock and date function
 Reset of all sensor-independent system settings to the default
condition

Settings

_

To open the Storage & config menu, press the /[Menu] or
key in the measured value display. After completing the
settings, switch to the measured value display with .
Menu item

Setting

Description

System / General /
Language

Deutsch
English
(further)

Select the menu language

System / General /
Illumination

Auto
On
Off

Switching the display illumination on/off

System / General /
Contrast

0 ... 100 %

Changing the display contrast

System / General /
Switchoff time

10 min ... 24 h

Adjust the switch-off time

System / Clock

Time
Date
Date format

Settings of time and date.
For details, see section
4.2.4

System / Service
information

24

Hardware version and software version of the meter
are displayed.

ba75793e05

09/2011

pH 3210

Operation

Menu item

Setting

Description

System / Reset

-

Resets the system settings
to the default values.
For details, see section
4.6.2

4.3.2

Data storage

This menu contains all functions to display, edit and erase stored
measured values and calibration records.
Note
Detailed information on the storage functions of the pH 3210 is given in
section 4.5.

ba75793e05

09/2011

25

Operation

pH 3210

4.3.3

Automatic Stability control

The automatic Stability control function continuously checks the
stability of the measurement signal. The stability has a considerable
impact on the reproducibility of measured values.
You can activate or switch off the automatic Stability control function
(see section 4.3.1).
The measured parameter flashes on the display
 as soon as the measured value is outside the allowed stability range
 if you switch over between the measured parameters with .
 if the automatic Stability control function is switched off.

26

ba75793e05

09/2011

pH 3210

Operation

4.4

pH value / ORP voltage

4.4.1

General information

You can measure the following variables:
 pH value [ ]
 ORP [mV]
Temperature
measurement

For reproducible pH measurements, it is essential to measure the
temperature of the test sample.
You have the following possibilities to measure the temperature:
 Automatic measurement of the temperature by the temperature
sensor (NTC30 or Pt1000) integrated in electrode.
 Manual determination and input of the temperature.
The meter recognizes whether a suitable sensor is connected and
automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature
measuring mode:

Preparatory activities

ba75793e05

09/2011

Temperature
sensor
yes

Resolution of the
temp. display
0.1 °C

-

1 °C

Mode
Automatic with temperature
sensor
Manual

Perform the following preparatory activities when you want to measure:
1

Connect a pH or ORP electrode to the meter.
The pH measuring window is displayed.

2

If necessary, select the pH or mV display with .

3

Adjust the temperature of the solutions and measure the current temperature if the measurement is made without a temperature sensor.

4

Calibrate or check the meter with the combination electrode.

27

Operation

pH 3210

4.4.2

Measuring the pH value

1

Perform the preparatory activities according to section 4.4.1.

2

Immerse the pH electrode in the test sample.

pH

6.949
25.0
Menu

3
Stability control
(AutoRead )

°C

22.11.2010
08:00

Select the pH or mV display with .

The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
The measured parameter flashes on the display
 as soon as the measured value is outside the stability range
 when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of
the setting for automatic Stability control (see page 26) in the System
menu.
1

Freeze the measured value with .
The [HOLD] status indicator is displayed.
The HOLD function is active.

2

Using , activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized.

Note
You can prematurely terminate the Stability control function manually
with  at any time.

28

ba75793e05

09/2011

pH 3210

Operation

3

Criteria for a stable
measured value

Using , start a further measurement with stability
control.
or
Release the frozen measured value again with  or .
The [AR] status display disappears. The display switches back
to the previous representation.

The Stability control function checks whether the measured values are
stable within the monitored time interval.
Measured
parameter

Time interval

Stability in the time interval

pH value

15 seconds

Δ pH: better than 0.01

Temperature

15 seconds

Δ T (° C): Better than 0.02

The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.

ba75793e05

09/2011

29

Operation

pH 3210

4.4.3

Measuring the ORP

Note
ORP electrodes are not calibrated. However, you can check ORP
electrodes using a test solution.

1

Perform the preparatory activities according to section 4.4.1.

2

Submerse the ORP electrode in the sample.

U

157.0

mV

24.8 °C

Menu

3
Stability control
(AutoRead )

22.11.2010
08:00

Select the mV display with .

The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.
The measured parameter flashes on the display
 as soon as the measured value is outside the stability range
 when the automatic Stability control is switched off.
You can start the Stability control manually at any time, irrespective of
the setting for automatic Stability control (see page 26) in the System
menu.
1

30

Freeze the measured value with .
The [HOLD] status indicator is displayed.
The HOLD function is active.

ba75793e05

09/2011

pH 3210

Operation

2

Using , activate the Stability control function manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The current measurement data
are output to the interface.
Measurement data meeting the stability control criterion are
marked by AR.

Note
You can prematurely terminate the Stability control function manually
with  at any time.

3

Criteria for a stable
measured value

Using , start a further measurement with stability
control.
or
Release the frozen measured value again with  or .
The [AR] status display disappears. The display switches back
to the previous representation.

The Stability control function checks whether the measured values are
stable within the monitored time interval.
Measured
parameter

Time interval

Stability in the time
interval

ORP

15 seconds

Δ mV: better than 0.3

Temperature

15 seconds

Δ T (° C): Better than 0.02

The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.

ba75793e05

09/2011

31

Operation

pH 3210

4.4.4
Overview

Settings for pH and ORP measurements

The following settings are possible for pH and ORP measurements:
 Resolution
 Calibration interval
 Buffers for calibration
 Unit of the temperature
 Automatic stability control
 Unit for slope
 Calibration record (display)

Settings

32

The settings are made in the measuring menu of the pH/ORP
measurement. To open the settings, activate the relevant measuring
window in the measured value display and press the  key
shortly. After completing the settings, switch to the measured value
display with .
Menu item

Possible
setting

Description

Calibration /
Calibration record

-

Displays the calibration
record of the last calibration.

Calibration /Buffer

TEC
NIST/DIN
ConCal
...

Buffer sets to be used for pH
calibration.
More buffers and details,
see section 4.4.5.

Calibration /One
point calibration

Yes
No

Rapid calibration with 1 buffer

Calibration /
Calibration interval

1 ... 999 d

Calibration interval for the
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.

Calibration /Unit for
slope

mV/pH
%

Unit of the slope.
The % display refers to the
Nernst slope of
-59.16 mV/pH (100 x determined slope/Nernst slope).

Man. temperature

-25 ... +130 °C

Entry of the manually determined temperature. For
measurements without temperature sensor only.

ba75793e05

09/2011

pH 3210

ba75793e05

Operation

09/2011

Menu item

Possible
setting

Description

Temperature unit

°C
°F

Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperatures are displayed with the selected
unit.

Resolution pH

0.001
0.01
0.1

Resolution of the pH display:

Resolution mV

0.1
1

Resolution of the mV display:

Stability control

On / Off

Switches on or off the automatic stability control during
measurement (see section
4.3.3)

Reset

-

Resets all sensor settings to
the delivery condition (see
section 4.6.1).

33

Operation

pH 3210

4.4.5
Why calibrate?

When do you have to
calibrate?
Buffer sets for
calibration

34

pH calibration

pH combination electrodes age. This changes the zero point
(asymmetry) and slope of the pH combination electrode. As a result, an
inexact measured value is displayed. Calibration determines and
stores the current values of the zero point and slope of the electrode.
Thus, you should calibrate at regular intervals.
 After connecting another combination electrode
 When the calibration interval has expired
You can use the buffer sets quoted in the table for an automatic
calibration. The pH values are valid for the specified temperature
values. The temperature dependence of the pH values is taken into
consideration during the calibration.
No.

Buffer set*

pH values

at

1

ConCal

Any

Any

2

NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers

1,679
4,006
6,865
9,180
12,454

25 °C

3

TEC
WTW Technical buffers

2,000
4,010
7,000
10,011

25 °C

4

Merck 1*

4,000
7,000
9,000

20 °C

5

Merck 2 *

1,000
6,000
8,000
13,000

20 °C

6

Merck 3 *

4,660
6,880
9,220

20 °C

7

Merck 4 *

2,000
4,000
7,000
10,000

20 °C

8

Merck 5 *

4,010
7,000
10,000

25 °C

ba75793e05

09/2011

pH 3210

ba75793e05

Operation

09/2011

No.

Buffer set*

pH values

at

9

DIN 19267

1,090
4,650
6,790
9,230

25 °C

10

Mettler Toledo USA *

1,679
4,003
7,002
10,013

25 °C

11

Mettler Toledo EU *

1,995
4,005
7,002
9,208

25 °C

12

Fisher *

2,007
4,002
7,004
10,002

25 °C

13

Fluka BS *

4,006
6,984
8,957

25 °C

14

Radiometer *

1,678
4,005
7,000
9,180

25 °C

15

Baker *

4,006
6,991
10,008

25 °C

16

Metrohm *

3,996
7,003
8,999

25 °C

17

Beckman *

4,005
7,005
10,013

25 °C

18

Hamilton Duracal *

4,005
7,002
10,013

25 °C

19

Precisa *

3,996
7,003
8,999

25 °C

35

Operation

pH 3210

No.

Buffer set*

pH values

at

20

Reagecon TEC *

2,000
4,010
7,000
10,000

25 °C

21

Reagecon 20 *

2,000
4,000
7,000
10,000
13,000

20 °C

22

Reagecon 25 *

2,000
4,000
7,000
10,000
13,000

25 °C

23

Chemsolute *

4,000
7,000
10,000

20 °C

24

USABlueBook *

4,000
7,000
10,000

20 °C

* Brand names or trade names are trademarks of their respective owners
protected by law.

Note
The buffers are selected in the menu, pH / /[Menu] / Calibration /
Buffer (see page 32).
Calibration points

Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The meter determines the
following values and calculates the calibration line as follows:

1-point

Determined
values

Displayed calibration data

Asy

 Zero point = Asy
 Slope = Nernst slope
(-59.16 mV/pH at 25 °C)

2-point

36

Asy
Slp.

 Zero point = Asy
 Slope = Slp.

ba75793e05

09/2011

pH 3210

Operation

3-point to
5-point

Determined
values

Displayed calibration data

Asy
Slp.

 Zero point = Asy
 Slope = Slp.
The calibration line is calculated
by linear regression.

Note
You can display the slope in the units, mV/pH or % (see page 24).
Stability control

The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be
terminated at any time (accepting the current value).

Calibration record

The new calibration values are displayed when the calibration is
finished.

Display calibration

You can have the data of the last calibration displayed (see page 50).

Calibration evaluation

After calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account. The evaluation appears on the display and
in the calibration record.
Display

Calibration
record

Zero point
[mV]

Slope
[mV/pH]

+++

-15 ... +15

-60.5 ... -58

++

-20 ... +20

-58 ... -57

+

-25 ... +25

-61 ... -60.5
or
-57 ... -56

-

-30 ... +30

-62 ... -61
or
-56 ... -50

< -30 or
> 30

... -62 or
... -50

Clean the combination electrode
according to the electrode operating
manual
Error

ba75793e05

09/2011

Error

37

Operation

pH 3210

Display

Calibration
record

Zero point
[mV]

Slope
[mV/pH]

Eliminate the error according to chapter
6 WHAT TO DO IF... (page 55)
Preparatory activities

Perform the following preparatory activities when you want to calibrate:
1

Connect the pH combination electrode to the meter.
The pH measuring window is displayed.

2

Keep the buffer solutions ready. Adjust the temperature of the
buffer solutions, or measure the current temperature, if you
measure without a temperature sensor.

4.4.6

Calibration interval

The calibration evaluation is indicated on the display as a sensor
symbol.
The sensor symbol flashes after the adjusted calibration interval has
expired. It is still possible to measure.
Note
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.

Setting the calibration
interval

38

The calibration interval is set to 7 days (d7) in the factory.
You can change the interval (1 ... 999 days):
1

Open the menu for measurement settings with /[Menu].

2

In the Calibration / Calibration interval menu, set the calibration
interval with <▲><▼>.

3

Confirm the setting with .

4

Exit the menu with .

ba75793e05

09/2011

pH 3210

Operation

4.4.7

Carrying out an automatic calibration (AutoCal)

Make sure that in the sensor menu, Buffer menu, the buffer set is correctly selected (see page 32).
Use any one to five buffer solutions of the selected buffer set in ascending or descending order.
Below, calibration with Technical buffers (TEC) is described. When
other buffer sets are used, other nominal buffer values are displayed.
Apart from that, the procedure is identical.
Note
If single-point calibration was set in the menu, the calibration procedure
is automatically finished with the measurement of buffer solution 1 and
the calibration record is displayed.

1

In the measured value display, select the measured parameter,
pH or mV with .

2

Start the calibration with .
The calibration display for the first buffer appears (voltage display).
Buffer

pH

-180.0
24.8

1
mV
°C

AutoCal TEC

22.11.2010
08:00

ba75793e05

09/2011

3

Thoroughly rinse the electrode with deionized water.

4

Immerse the electrode in buffer solution 1.

5

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <▲><▼>.

6

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

39

Operation

pH 3210

Buffer

pH

-180.0
24.8

10.011
mV
°C

AutoCal TEC

AR
22.11.2010
08:00

7

Wait for the end of the measurement with stability control or
accept the calibration value with .
The calibration display for the next buffer appears.

8

If necessary, terminate the calibration as a single-point calibration with .
The calibration record is displayed.

Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the zero point of the
electrode.
Continuing with twopoint calibration

9

Thoroughly rinse the electrode with deionized water.

10

Immerse the electrode in buffer solution 2.

11

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <▲><▼>.

12

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

pH

Buffer

0.0
24.8

7.000
mV
°C

AutoCal TEC

AR
22.11.2010
08:00

40

ba75793e05

09/2011

pH 3210

Operation

Continuing with threeto five-point calibration

13

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with .
The calibration display for the next buffer appears.

14

If necessary, terminate the calibration as a two-point calibration
with .
The calibration record is displayed.

15

Thoroughly rinse the combination electrode with deionized
water.

16

Immerse the electrode in buffer solution 3.

17

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <▲><▼>.

18

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.
Buffer

pH

180.0
24.8

4.010
mV
°C

AutoCal TEC

AR
22.11.2010
08:00

19

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with .
The calibration display for the next buffer appears (voltage display).

20

If necessary, use  to finish the calibration.
The calibration record is displayed.
or
Switch to calibration with the next buffer with .

Note
Calibration is automatically completed after the last buffer of a buffer
set has been measured. Then the calibration record is displayed.
The calibration line is determined by linear regression.
ba75793e05

09/2011

41

Operation

pH 3210

4.4.8

Carrying out a manual calibration (ConCal)

Make sure that in the sensor menu, Buffer menu, the ConCal buffer set
is correctly selected (see page 32).
Use any one to five buffer solutions in ascending or descending order.
Note
If single-point calibration was set in the menu, the calibration procedure
is automatically finished with the measurement of buffer solution 1 and
the calibration record is displayed.

1

In the measured value display, select the measured parameter
pH or mV with .

2

Start the calibration with .
The calibration display appears.
Buffer

pH

0.0
24.8

1

mV
°C

ConCal

22.11.2010
08:00

42

3

Thoroughly rinse the combination electrode with deionized
water.

4

Immerse the electrode in buffer solution 1.

5

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <▲><▼>.

6

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

ba75793e05

09/2011

pH 3210

Operation

7.000

Buffer

pH

0.0
24.8

mV
°C

ConCal

AR
22.11.2010
08:00

7

Wait for the measurement with stability control to be completed.
Set the nominal buffer value for the measured temperature with
<▲><▼>.
Buffer

pH

7.000
24.8

1
▲
▼

°C

ConCal

22.11.2010
08:00

8

Set the nominal buffer value for the measured temperature with
<▲><▼>.

9

Accept the calibration value with .
The calibration display for the next buffer appears (voltage display).

10

If necessary, terminate the calibration as a single-point calibration with .
The calibration record is displayed.

Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the zero point of the
electrode.

ba75793e05

09/2011

43

Operation

Continuing with twopoint calibration

pH 3210

11

Thoroughly rinse the combination electrode with deionized
water.

12

Immerse the electrode in buffer solution 2.

13

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <▲><▼>.

14

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

15

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with .
The pH value of the buffer solution is displayed.
Buffer

pH

4.035
24.8

2
▲
▼

°C

ConCal

22.11.2010
08:00

Continuing with threeto five-point calibration

44

16

Set the nominal buffer value for the measured temperature with
<▲><▼>.

17

Accept the calibration value with .
The calibration display for the next buffer appears (voltage display).

18

If necessary, finish the calibration procedure as a two-point
calibration with .
The calibration record is displayed.

19

Thoroughly rinse the electrode with deionized water.

20

Immerse the electrode in the next buffer solution.

21

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <▲><▼>.

ba75793e05

09/2011

pH 3210

Operation

22

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

23

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with .
The pH value of the buffer solution is displayed.
Buffer

pH

3
▲
▼

9.958
24.8

°C

ConCal

22.11.2010
08:00

24

Set the nominal buffer value for the measured temperature with
<▲><▼>.

25

Accept the calibration value with .
The calibration display for the next buffer appears (voltage display).

26

If necessary, use  to finish the calibration.
The calibration record is displayed.
or
Continue calibrating using the next buffer with .

Note
After the fifth buffer has been measured the calibration is automatically
finished. Then the calibration record is displayed.
The calibration line is determined by linear regression.

4.4.9
Displaying the
calibration record

Displaying calibration records

The calibration record of the last calibration is to be found under the
menu item, Calibration / Calibration record. To open it in the measured
value display, press the  key.

_

The calibration records of the last calibrations (up to 10) are available
in the menu, /[Menu] / Calibration / Calibration data storage and
in the menu, /[Menu] / Storage & config/Data storage / Calibration data storage.

_

ba75793e05

09/2011

45

Operation

pH 3210

4.4.10 Continuous measurement control (CMC function)
The continuous measurement control (CMC function) enables to
quickly and safely evaluate the current measured value at a glance.
After each successful calibration, the scale of the pH measuring range
is displayed in the measured value display. Here you can very easily
see whether or not the current measured value is in the calibrated part
of the measuring range.
The following information is displayed:

1
2

pH

0

6.093

14

25.0 °C

Menu

3
4

22.11.2010
08:00

1

Measuring range for which a valid calibration is available (white).
Measured values in this range are suitable for documentation.

2

Measuring range for which no valid calibration is available (shaded). Measured values in this range are not suitable for documentation. If necessary, calibrate the meter with buffers
covering this measuring range.
If the current measured value is outside the calibrated range,
this area is shaded stronger.
If a measured value is outside the measuring range pH 0 - 14,
overflow arrows are displayed at the left or right edge of the
measuring range.

3

Current measured pH value (pointer)

4

Line marks for all nominal buffer values used for the last valid
calibration

The limits of the calibrated range are determined by the buffers used
for calibration:
Lower limit:
Upper limit:

46

Buffer with the lowest pH value minus 2 pH units
Buffer with the highest pH value plus 2 pH units

ba75793e05

09/2011

pH 3210

Operation

4.5

Storage

You can transmit measured values (datasets) to the data storage:
 Manual storage (see section 4.5.1)
Measurement dataset

A complete dataset consists of:
 Date/time
 Measured value of the connected sensor
 Measured temperature value of the connected sensor
 AutoRead info: AR appears with the measured value if the
AutoRead criterion was met while storing (stable measured value).
Otherwise, the AR display is missing.
 Calibration evaluation: +++, ++, +, -, or no evaluation

Storage locations

ba75793e05

09/2011

The pH 3210 meter has a measurement data memory.
Storage

Maximum number of datasets

Manual data storage

200

47

Operation

pH 3210

4.5.1

Manual storage

You can transmit a measurement dataset to the data storage as
follows:
1

Press the  key shortly.
The menu for manual storage appears.
Manual data storage 4 From 200

30.10.2008 11:24:16
pH 7.000 24.8 °C AR +++

ID number:
Continue

Back

2

If the storage is full

1

22.11.2010
08:00

If necessary, change and confirm the ID number (1 ... 10000)
with <▲><▼> and .
The dataset is stored. The instrument switches to the measured value display.

The following window appears if all 200 storage locations are occupied:
Warning
Data storage full. Erase?

Yes
No
Back

22.11.2010
08:00

You have the following options:
 To erase the entire storage, confirm Yes.
 To cancel the storage process and switch to the measured value
display, confirm No.
4.5.2

Editing the measured value storage

The contents of the manual or automatic measurement data storage
can be shown on the display.
The measurement data storage has a function to erase the entire
contents.
48

ba75793e05

09/2011

pH 3210

Operation

Editing the data storage

Settings

The storage is edited in the menu, Storage & config/ Data storage. To
open the Storage & config menu, press the /[Menu] key in the
measured value display.

_

Menu item

Setting/
function

Description

Data storage /
Manual data storage /
Display

-

Displays all measurement
datasets page by page.
Further options:
 Scroll through the
datasets with <▲><▼>.
 Quit the display with
/[Back].

Data storage /
Manual data storage /
Erase

-

Erases the entire manual
measurement data storage.
Note:
All calibration data remains
stored when this action is
performed.

Display presentation
of a dataset

Manual data storage

3 of 64

30.10.2008 11:24:16 ID number: 1

Quitting the display

pH 7.000

24.8 °C AR +++

Back

22.11.2010
08:00

To quit the display of stored measuring datasets, you have the
following options:
 Switch directly to the measured value display with .
 Quit the display and move to the next higher menu level with /
[Back].

ba75793e05

09/2011

49

Operation

pH 3210

4.5.3

Erasing the measurement data storage

How to erase the measurement data storage is described in section
4.5.2 EDITING THE MEASURED VALUE STORAGE.

50

ba75793e05

09/2011

pH 3210

Operation

4.6

Reset

You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.

4.6.1

Resetting the measurement settings

Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH

The following settings for pH measurements are reset to the default
settings with the Reset function:
Setting

Default settings

Buffer

TEC

Cal. interval

7d

Unit for slope

mV/pH

Measured parameter

pH

Resolution pH

0.001

Resolution mV

0.1

Asymmetry

0 mV

Slope

-59.16 mV

Man. temperature

25 °C

One point calibration

Off

The sensor settings are reset under the Reset menu item in the
measuring menu. To open the settings, activate the relevant measuring
window in the measured value display and press the /[Menu] key
shortly.

ba75793e05

09/2011

51

Operation

pH 3210

4.6.2

Resetting the system settings

The following system settings can be reset to the delivery status:
Setting

Default settings

Language

English

Temperature unit

°C

Beep

On

Contrast

50 %

Illumination

On

Switchoff time

1h

The system settings are reset in the menu, Storage & config / System
/ Reset. To open the Storage & config menu, press the /[Menu]
key in the measured value display.

_

52

ba75793e05

09/2011

pH 3210

Maintenance, cleaning, disposal

5

Maintenance, cleaning, disposal

5.1

Maintenance

The only maintenance activity required is replacing the batteries.
Note
See the relevant operating manuals of the combination electrodes for
instructions on maintenance.

5.1.1

Replacing the batteries

1

Unscrew the two screws (1) on the underside of the meter.

2

Open the battery compartment (2) on the underside of the
meter.

1

2

3

Remove the four batteries from the battery compartment.

4

Place four new batteries (type Mignon AA) in the battery
compartment.

Note
Alternatively, you can also use Ni-MH rechargeable batteries (type
Mignon AA). In order to charge the batteries, an external charging
device is required.

Caution
Make sure that the poles of the batteries are positioned correctly.
The ± signs on the batteries must correspond to the ± signs in the
battery compartment.
ba75793e05

09/2011

53

Maintenance, cleaning, disposal

pH 3210

5

5.2

Close the battery compartment (2) and tighten the screws (1).

Cleaning

Occasionally wipe the outside of the meter with a damp, lint-free cloth.
Disinfect the housing with isopropanol as required.

Caution
The housing is made of synthetic material (ABS). Thus, avoid
contact with acetone or similar detergents that contain solvents.
Remove any splashes immediately.

5.3

Packing

This meter is sent out in a protective transport packing.
We recommend: Keep the packing material. The original packing
protects the meter against damage during transport.

5.4

Disposal

Note
This meter contains batteries. Batteries that have been removed must
only be disposed of at a recycling facility set up for this purpose or via
the retail outlet.
It is illegal to dispose of it in household refuse.

54

ba75793e05

09/2011

pH 3210

What to do if...

6
Error message
OFL, UFL

Error message
Error

What to do if...

Cause

Remedy

pH combination electrode:
– Measured value outside the
measuring range

– Use suitable combination
electrode

– Air bubble in front of the
junction

– Remove air bubble

– Air in the junction

– Extract air or moisten
junction

– Cable broken

– Replace combination
electrode

– Gel electrolyte dried out

– Replace combination
electrode

Cause

Remedy

pH combination electrode:
– The values determined for zero
point and slope of the
combination electrode are
outside the allowed limits.

– Recalibrate

– Junction contaminated

– Clean junction

– Combination electrode broken

– Replace combination
electrode

Buffer solutions:

ba75793e05

09/2011

– Incorrect buffer solutions

– Change calibration
procedure

– Buffer solutions too old

– Use only once.
Note the shelf life

– Buffer solutions depleted

– Change solutions

55

What to do if...

pH 3210

No stable measured
value

Cause

Remedy

pH combination electrode:
– Junction contaminated

– Clean junction

– Membrane contaminated

– Clean membrane

Test sample:
– pH value not stable

– Measure with air excluded if
necessary

– Temperature not stable

– Adjust temperature if
necessary

Combination electrode + test
sample:

Sensor symbol flashes

– Conductivity too low

– Use suitable combination
electrode

– Temperature too high

– Use suitable combination
electrode

– Organic liquids

– Use suitable combination
electrode

Cause
– Calibration interval expired

Display

Cause
– batteries almost empty

56

Remedy
– Recalibrate the measuring
system

Remedy
– Replace the batteries (see
section 5.1 MAINTENANCE)

ba75793e05

09/2011

pH 3210

What to do if...

Obviously incorrect
measured values

Cause
pH combination electrode:
– pH combination electrode
unsuitable

– Use suitable combination
electrode

– Temperature difference
between buffer and test sample
too high

– Adjust temperature of buffer
or sample solutions

– Measurement procedure not
suitable

– Follow special procedure

Meter does not react to
keystroke

Cause

You want to know which
software
version is in the meter

Cause

ba75793e05

09/2011

Remedy

– Operating condition undefined
or EMC load unallowed

– E. g., a question by the service
department

Remedy
– Processor reset:
Press the  and
 key
simultaneously

Remedy
– Switch on the meter.
Open the menu,
[Menu] / Storage &
config / System / Service
information. The instrument
data is displayed.

_

57

What to do if...

58

pH 3210

ba75793e05

09/2011

pH 3210

Technical data

Dimensions
Weight
Mechanical structure

7

Technical data

7.1

General data

approx. 180 x 80 x 55 mm
approx. 0.4 kg
Type of protection

IP 67

Electrical safety

Protective class

III

Test certificates

CE, cETLus

Ambient
conditions

Power
supply

Sensor input

Guidelines
and norms used

ba75793e05

09/2011

Storage

- 25 °C ... + 65 °C

Operation

-10 °C ... + 55 °C

Allowable relative
humidity

Annual mean: < 75 %
30 days/year: 95 %
Other days: 85 %

Batteries

4 x 1.5 V alkali-manganese batteries, type
AA

Rechargeable
batteries

4 x 1,2 V NiMH rechargeable batteries,
type AA (no charging function)

Operational life

up to 1000 h without / 150 h with illumination

Input resistance

> 5 * 1012 ohm

Input current

< 1 * 10-12 A

EMC

EC directive 2004/108/EC
EN 61326-1
EN 61000-3-2
EN 61000-3-3
FCC Class A

Meter safety

EC directive 2006/95/EC
EN 61010-1

IP protection class

EN 60529

59

Technical data

pH 3210

7.2
Measuring ranges,
resolution

Measuring ranges, resolution, accuracy

Variable

Measuring range

Resolution

pH

- 2.0 ... + 20.0

0.1

- 2.00 ... + 20.00

0.01

- 2.000 ... + 19.999

0.001

- 1200.0 ... + 1200.0

0.1

- 2500 ... + 2500

1

T [°C]

- 5.0 ... + 105.0

0.1

T [°F]

23.0 ... + 221.0

0.1

Variable

Range

Increment

Tmanual [°C]

- 25 ... + 130

1

Tmanual [°F]

-13 ... + 266

1

Variable

Accuracy

Temperature of the test
sample

- 2.0 ... + 20.0

± 0.1

+ 15 °C ... + 35 °C

- 2.00 ... + 20.00

± 0.01

+ 15 °C ... + 35 °C

- 2.000 ... + 19.999

± 0.005

+ 15 °C ... + 35 °C

- 2500 ... + 2500

±1

+ 15 °C ... + 35 °C

-1200.0 ... +1200.0

± 0.3

+ 15 °C ... + 35 °C

U [mV]

Manual
temperature input

Accuracy (± 1 digit)

pH / range *

U [mV] / range

T [°C] / temperature sensor
NTC 30

± 0.1

PT 1000

± 0.1

* when measuring in a range of ± 2 pH around a calibration point

Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the combination electrodes and buffer solutions has to be
taken into account additionally.

60

ba75793e05

09/2011

pH 3210

Technical data

FCC Class A Equipment Statement
Note: This equipment has been tested and found to comply with
the limits for a Class A digital device, pursuant to Part 15 of the
FCC Rules. These limits are designed to provide reasonable
protection against harmful interference when the equipment is
operated in a commercial environment. This equipment
generates, uses, and can radiate radio frequency energy and, if
not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause
harmful interference in which case the user will be required to
correct the interference at his own expense.
Changes or modifications not expressly approved by the
manufacturer could void the user‘s authority to operate the
equipment.

ba75793e05

09/2011

61

Technical data

62

pH 3210

ba75793e05

09/2011

pH 3210

Lists

8

Lists

This chapter provides additional information and orientation aids.
Specialist terms

Index

The glossary briefly explains the meaning of the specialist terms.
However, terms that should already be familiar to the target group are
not described here.

The index will help you to find the topics that you are looking for.

Glossary
Adjusting

To manipulate a measuring system so that the relevant value (e. g. the
displayed value) differs as little as possible from the correct value or
a value that is regarded as correct, or that the difference remains
within the tolerance.

Asymmetry

see zero point

AutoRange

Name of the automatic selection of the measuring range.

Calibration

Comparing the value from a measuring system (e. g. the displayed
value) to the correct value or a value that is regarded as correct.
Often, this expression is also used when the measuring system is
adjusted at the same time (see adjusting).

Electromotive force of
a combination
electrode

The electromotive force U of the combination electrode is the
measurable electromotive force of a combination electrode in a
solution. It equals the sum of all the galvanic voltages of the
combination electrode. Its dependency on the pH results in the
electrode function, which is characterized by the parameters, slope
and zero point.

Junction

Measured parameter
Measured value

ba75793e05

09/2011

The junction is a porous body in the housing wall of reference
electrodes or electrolyte bridges. It arranges the electrical contact
between two solutions and makes the electrolyte exchange more
difficult. The expression, junction, is also used for ground or junctionless transitions.
The measured parameter is the physical dimension determined by
measuring, e. g. pH, conductivity or D.O. concentration.
The measured value is the special value of a measured parameter to
be determined. It is given as a combination of the numerical value and
unit (e. g. 3 m; 0.5 s; 5.2 A; 373.15 K).

63

Lists

pH 3210

Molality
ORP voltage

The ORP is caused by oxidizing or reducing substances dissolved in
water, if these substances become effective at an electrode surface
(e. g. a gold or platinum surface).

pH value

The pH is a measure of the acidic or basic effect of an aqueous
solution. It corresponds to the negative decadic logarithm of the molal
hydrogen ions activity divided by the unit of the molality. The practical
pH value is the value of a pH measurement.

Potentiometry

Name of a measuring technique. The signal (depending on the
measured parameter) of the electrode is the electrical potential. The
electrical current remains constant.

Reset
Resolution
Slope
Stability control
(AutoRead )
Standard solution

64

Molality is the quantity (in Mol) of a dissolved substance in 1000 g
solvent.

Restoring the original condition of all settings of a measuring system.
Smallest difference between two measured values that can be
displayed by a meter.
The slope of a linear calibration function.
Function to control the measured value stability.
The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.

Test sample

Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.

Zero point

The zero point of a pH combination electrode is the pH value at which
the electromotive force of the pH combination electrode at a specified
temperature is zero. Normally, this is at 25 °C.

ba75793e05

09/2011

pH 3210

Lists

Index

A

Battery compartment . . . . . . . . . . . . . . 13, 53

Storage locations . . . . . . . . . . . . . . . . . 47
Measurement dataset . . . . . . . . . . . . . . . . 47
Measuring
ORP . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Measuring menu
pH/ORP . . . . . . . . . . . . . . . . . . . . . . . . 32
Menus (navigation) . . . . . . . . . . . . . . . . . . 17
Messages . . . . . . . . . . . . . . . . . . . . . . . . . 18

C

O

Authorized use . . . . . . . . . . . . . . . . . . . . . . 10
Automatic switch-off . . . . . . . . . . . . . . . . . . 15
AutoRead
pH . . . . . . . . . . . . . . . . . . . . . . . . . 28, 30

B

Calibration
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Calibration evaluation
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Calibration interval . . . . . . . . . . . . . . . . . . . 38
Calibration points
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

D

P
pH buffer sets . . . . . . . . . . . . . . . . . . . . . . 34
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . 9

R
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51

Dataset . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Date and time . . . . . . . . . . . . . . . . . . . . . . . 22
Default settings
Measured parameter . . . . . . . . . . . . . . 51
System settings . . . . . . . . . . . . . . . . . . 52
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

E
Energy saving feature . . . . . . . . . . . . . . . . 14

I
Initial commissioning . . . . . . . . . . . . . . 13, 14
Initialize . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Interval
calibration . . . . . . . . . . . . . . . . . . . . . . . 38

K
Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

M
Measured value display . . . . . . . . . . . . . . . 17
Measurement accuracy . . . . . . . . . . . . . . . 38
Measurement data memory
Edit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Erase . . . . . . . . . . . . . . . . . . . . . . . . . . 49

ba75793d05

Operational safety . . . . . . . . . . . . . . . . . . . 10

09/2011

S
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Saving . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Scope of delivery . . . . . . . . . . . . . . . . . . . . 13
Setting the date . . . . . . . . . . . . . . . . . . . . . 14
Setting the time . . . . . . . . . . . . . . . . . . . . . 14
Single-point calibration
pH . . . . . . . . . . . . . . . . . . . . . . . . . 40, 43
Slope
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Socket field . . . . . . . . . . . . . . . . . . . . . . . . . 8
Stability control
Automatic . . . . . . . . . . . . . . . . . . . . . . . 26

T
Temperature measurement
pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Three-point calibration
pH . . . . . . . . . . . . . . . . . . . . . . . . . 41, 44
Two-point calibration
pH . . . . . . . . . . . . . . . . . . . . . . . . . 40, 44

Z
Zero point of pH combination electrode . . . 34

65

Lists

66

pH 3210

ba75793d05

09/2011

Wissenschaftlich-Technische Werkstätten GmbH
Dr.-Karl-Slevogt-Straße 1
D-82362 Weilheim
Germany
Tel:
Fax:
E-Mail:
Internet:

+49 (0) 881 183-0
+49 (0) 881 183-100
+49 (0) 881 183-420
Info@WTW.com
http://www.WTW.com



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Author                          : direim
Create Date                     : 2011:09:14 15:52:19Z
Modify Date                     : 2015:09:17 14:09:06-06:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 5.6-c015 81.157285, 2014/12/12-00:43:15
Creator Tool                    : FrameMaker 10.0.1
Metadata Date                   : 2015:09:17 14:09:06-06:00
Format                          : application/pdf
Title                           : WTW pH 3210 Operating Manual
Creator                         : direim
Producer                        : Acrobat Elements 9.0.0 (Windows)
Document ID                     : uuid:607207ef-fa62-4fde-ab9d-1b9bb7322f60
Instance ID                     : uuid:5119dc3d-c82b-3e48-9ab6-ef5d1997bee4
Page Mode                       : UseOutlines
Page Count                      : 68
EXIF Metadata provided by EXIF.tools

Navigation menu