Operating Instructions

ORP Calibration

User Reset

Calibration should be done regularly, preferred once in a week. ORP tester
allows one point calibration.

Reset the ORP calibration to the default using the user reset function.

1. Press ON/OFF ( ) to power on if needed.

ORPTestr® 50 Pocket Tester

2. D
 ip electrode in about 2 cm to 3 cm into the calibration standard solution.
3. P
 ress the CAL/ESC key to enter calibration mode. The CAL indicator will be
displayed.

Applications
• Aquaculture
• Aquariums
• Boilers
• Car washes
• Ecology

• Spas
• Studies
•W
 ater and wastewater
treatment
 ater quality testing
•W
in pools and more

• Food sectors
• Hydroponics
• Labs
• Sanitation plants
• Steam generators

Getting Started
The ORPTestr 50 Pocket Tester has been factory calibrated and usually works
well out of the box. However, after extended periods of non-use, it is best to
remove the sensor cap and soak the sensor in warm tap water or ORP buffer
for 10 minutes or so. A brief rinse with deionized (DI) water is OK, but avoid
soaking or storing in deionized water as this will shorten ORP electrode life.
Prior to taking the measurements, periodic calibration with certified standards
is recommended for best accuracy.

4. A
 llow the reading to stabilize. The timer icon blinks during the reading
stabilization. Once the reading stabilized the timer stops blinking.

 ower display shows current measured mV with last calibration (if no
6. L
calibration is performed earlier both display shows the same measurement
value).
7. Press MENU/v to adjust the mV value to match the calibration standard
value.
8. Press HOLD/ to confirm the calibration or press CAL/ESC key to exit the
calibration without confirmation.
CAL

CAL
cal
esc

Place the tester in
the buffer to
start or press
to escape.
cal
esc

189
189

1. Press ON/OFF ( ) to power on if needed.

1. Press ON/OFF ( ) to power on if needed.

4. Scroll down by pressing MENU/v to toggle between °C and °F.
5. Press HOLD/ to select a temperature unit. The display shows the selected
temperature setting with a .
menu

Key Info

hold

Temp Set

Set °C/°F

Accept HOLD

Backlight

Temp Cal

EXIT

Reset

CAL

hold

hold

MENU

Scroll

°C

°C
°F

RmV

844
25.0°C

2. P
 ress MENU/v to enter setup window. Scroll down by pressing
MENU/v to select Backlight. Press HOLD/ to select Settings. The
display shows Backlight.
3. Press HOLD/ to select Backlight. The display shows ON and OFF.

For Sales & Service Contact

4. Scroll down by pressing MENU/v to toggle between ON and OFF.
Backlight ON increases readability in low-light conditions.
5. P
 ress HOLD/ to select the desired backlight option. The display
shows the selected backlight option with a .
menu

Key Info
MENU

Temp Set

Accept HOLD

Backlight

Scroll

EXIT

hold

hold

CAL

Reset

1065O100_MAN_35634-45.indd 1

hold

Backlight
Backlight

ON
OFF

OFF

Scroll

EXIT

hold

hold

MENU

Accept HOLD
CAL

Temp Set

hold

User Reset

Settings
Reset

No

Factory Reset

Yes

Factory Reset
Reset the tester to the factory default using the factory reset function.
1. Press MENU/v to enter setup window. Scroll down by pressing the
MENU/v to select Reset. Press HOLD/ to select Reset. The display shows
User Reset and Factory Reset.

menu

2. D
 ip the electrode about 2 cm to 3 cm into the test solution. Stir and let the
reading stabilize. The timer icon blinks during this time. Once the reading is
will appear to indicate the stability
stabilized, the timer stops blinking and
of the reading.
CAUTION: Testing dry samples is not accurate and can
lead to sensor damage or breakage. Soils must be wet
and free of particulates that may scratch the glass
sensor. Excessive force into dry samples can cause
glass breakage.

Key Info

menu

4. Press HOLD/ to confirm either No or Yes. The display shows the factory
reset option with a .

ORP
RmV

844
25.0°C

4. P
 ress ON/OFF ( ) for 3 seconds to turn off tester. If a key is not pressed for
8.5 minutes, the tester will automatically shut off to conserve batteries.

1. Press ON/OFF ( ) to power on if needed.

4. Press HOLD/ to confirm either No or Yes. The display shows the user
reset option with a .

3. Scroll down by pressing MENU/v to toggle between No and Yes.

3. Note the ORP value or press HOLD/ to freeze the reading. To release the
reading, press HOLD/ again.

Backlight Settings

3. Scroll down by pressing MENU/v to toggle between No and Yes.

2. S
 croll down by pressing the MENU/v to toggle between the resets.
Press HOLD/ to select Factory Reset. The display automatically shows
No and Yes.

CAL
mV

ORP Measurement

3. Press HOLD/ to select Set °C/°F. Display automatically shows °C and °F.

2. Press HOLD/ to select User Reset. The display automatically shows
No and Yes.

5. U
 pper display shows current measured value based on the factory default.

Temperature Settings
2. P
 ress MENU/v to enter setup window. Press HOLD/ to select Temp Set.
The display shows Set °C/°F and Temp Cal.

1. Press MENU/v to enter setup window. Scroll down by pressing MENU/v
to select Reset. Press HOLD/ to select Reset. The display shows User
Reset and Factory Reset.

Key Info
Scroll

Accept HOLD
EXIT

hold

hold

MENU

Temp Set

User Reset

Settings

Factory Reset

Reset

CAL

hold

No
Yes

Manual Temperature Setting
The MTS range is 0°C to 60.0°C (32.0°F to 140.0°F). User reset will set
temperature to default value 25°C or 77°F.
1. Press MENU/v to select setup window. Scroll down by pressing MENU/v
to select Temp Set. Press HOLD/ to select Temp Set. The display shows
Set °C/°F and Temp Cal.
2. Scroll down by pressing MENU/v to toggle between Set °C/°F and
Temp Cal. Press HOLD/ to select Temp Cal.
3. T
 he lower display shows the current measured temperature reading based
on the last set offset and the upper display shows the current measured
temperature reading based on factory default calibration.
4. Dip the tester into a solution of known temperature and allow time for the
built-in temperature sensor to stabilize.
5. P
 ress MENU/v to adjust the temperature value or press the HOLD/ to
confirm the calibrated value as new temperature value of the solution.
Note: To exit without confirming the calibration, press CAL/ESC.

2650 E. 40th Ave. • Denver, CO 80205
Phone 303-320-4764 • Fax 303-322-7242

1-800-833-7958

menu

Key Info
Scroll

Accept HOLD
EXIT

hold

hold

MENU

CAL

Temp Set
Backlight
Reset

Set °C/°F
Temp Cal

hold

Temp Cal
23.4°C
23.4°C

23.4°C

www.geotechenv.com

9/5/2017 3:04:37 PM

HOLD Function
1. Press HOLD/ button to freeze
the measurement.
mV

mV
hold

RmV
189

mV
189
25.0°C

Self-Diagnostic Messages

Replacing the Batteries

This feature helps to freeze the display for a delayed observation.
2. P
 ress HOLD/ again to
release the measurement.
mV

RmV

mV
hold

189

Press Hold

Press Hold

RmV

to release

to release

189
25.0°C

25.0°C

25.0°C

Sensor Maintenance
1. R
 inse the sensor with sensor storage solution after each measurement. Care
has to be taken not to damage the sensor’s glass electrode.
2. In aggressive chemicals, dirty or viscous solutions, and solutions with
heavy metals or proteins, take readings quickly and rinse electrode
immediately afterward.
3. If possible, keep a small amount of clean water or electrode storage
solution (NOT de-ionized water) in sensor cap to keep sensor hydrated.

Sensor Replacement
Replace the sensor module at the fraction of the cost of a new tester.
When the tester fails to calibrate or gives fluctuating readings in calibration
standards, need to change the electrode.
1. With dry hands, grip the ring with sensor facing you. Twist the ring
counterclockwise. Save the ring for later use.
2. Pull the old sensor module away from the tester.
3. Align the four tabs on the new sensor module so that they match the
four slots on the tester.
4. Gently push the module onto the slots to sit it in position. Push the
smaller O-ring fully onto the new sensor module. Push the other O-ring
over the module and thread sensor ring into place by firmly twisting
clockwise.
Note: It is necessary that you recalibrate the tester prior to measurement
after a sensor replacement.

The pH tester 50 uses four AAA
1.5 V batteries.
1. T
 o remove the battery cover,
See Figure 1. Clear the front
catch and then the back
catch, before sliding the
cover off.

Batteries are weak and need replacement soon.
stable error

Fig. 1: Removing battery cover

Fig. 2: Push to unlock

2. T
 o remove the battery plate,
push the center tab towards
the front of the tester as
shown in Figure 2. Once
unlocked, remove the plate
to access the batteries.
3. Invert the tester upside down
to remove the batteries. Each
side uses two AAA batteries.
Orient each battery with
positive terminal facing
downward.
4. To lock the battery plate, align
the small tabs (Figure 3) into
the guide ribs on the housing and then press down.
See Figure 4.

over range
under range

Appears when calibration is attempted but the reading is not yet stable.
Wait for the reading to stabilize or manually confirm the calibration by
pressing enter.
The reading is above the measuring range of tester.
The reading is below the measuring range of tester.

Specifications

Fig. 3: Align tabs

Fig. 5: Battery plate unlocked

Fig. 4: Push down to lock

Fig. 6: Battery plate locked

Warranty
This instrument is supplied with a warranty against manufacturing defects
for a period of one year from the date of purchase.

Return of Items
Authorization must be obtained from distributor before returning items for
any reason. When applying for authorization, please include information
regarding the reason the item(s) are to be returned.
We reserve the right to make improvements in design, construction and
appearance of products without notice. Prices are subject to change
without notice.

Specifications
ORP/RMV
ORP range
Resolution
Relative accuracy
Calibration points
Calibration window
Calibration type
Temperature
Temperature range
Temperature resolution
Temperature accuracy
Temperature compensation
Temperature calibration window
General Specifications
Display
Backlight
Auto off
Reset
Power requirement
Battery life
Waterproofing
Regulatory certifications
Ambient operating temperature
Relative humidity
Storage temperature
Storage humidity

ORPTestr 50
–1000 to 1000 mV
±1 mV
±2 mV ±1 LSD
One
±200 mV
Offset
0 to 60°C / 32.0 to 140.0°F
0.1°C / 0.1°F
NA
Manual
0 to 60.0°C / 32.0 to 140.0°F
Graphics, dot matrix 80 x100 pixel
Yes, selectable (30 sec from last key press)
8.5 minutes (from last key press)
User / factory
Four AAA 1.5 V batteries
>250 hours
IP67
CE, FCC
5 to 45°C / 41 to 113°F
5% to 85% noncondensing
–20 to 60°C
5% to 85%, noncondensing

Accessories
Ordering Code
35634-45
35634-47
35634-09
09376-00
17101-45
05478-60

Product Description
ORPTestr 50 pocket tester with case, lanyard, and batteries
Replacement sensor module
Replacement sensor cap
Replacement alkaline batteries; AAA, 1.5 V. Pack of 12
NIST-traceable calibration with data for pH testers
YSI® Zobell solution

www.4oakton.com
1065O100_MAN_35634-45

1065O100_MAN_35634-45.indd 2

August, 2017		

9/5/2017 3:04:41 PM



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Author                          : Oakton
Create Date                     : 2017:09:05 15:04:36-05:00
Keywords                        : 1065O100_MAN_35634-45
Modify Date                     : 2018:02:28 12:05:11-07:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 5.6-c015 84.159810, 2016/09/10-02:41:30
Metadata Date                   : 2018:02:28 12:05:11-07:00
Creator Tool                    : Adobe InDesign CS6 (Windows)
Instance ID                     : uuid:df975bab-9f51-084f-a935-a7f3ee22a658
Original Document ID            : xmp.did:5526ACFA893AE311905385DF55ED17D0
Document ID                     : xmp.id:F652F3707592E711A2448160D06E30C7
Rendition Class                 : proof:pdf
Derived From Instance ID        : xmp.iid:9D84663F6892E711A2448160D06E30C7
Derived From Document ID        : xmp.did:3B112A05BE81E7119195C87AC1E5FFD6
Derived From Original Document ID: xmp.did:5526ACFA893AE311905385DF55ED17D0
Derived From Rendition Class    : default
History Action                  : converted
History Parameters              : from application/x-indesign to application/pdf
History Software Agent          : Adobe InDesign CS6 (Windows)
History Changed                 : /
History When                    : 2017:09:05 15:04:36-05:00
Format                          : application/pdf
Title                           : Oakton ORPTestr® 50 Pocket Tester Operating Instructions
Creator                         : Oakton
Description                     : Oakton ORPTestr 50 Pocket Tester Manual
Subject                         : 1065O100_MAN_35634-45
Producer                        : Adobe PDF Library 10.0.1
Trapped                         : False
Page Count                      : 2
EXIF Metadata provided by EXIF.tools

Navigation menu