Philips Medical Systems North America WLANBV1 WLAN Module User Manual G 0 addendum

Philips Medical Systems North America Co. WLAN Module G 0 addendum

User Manual addendum

Addendum to the Instructions for Use31English  Addendum to the Instructions for UseThis addendum provides additional or corrected information for the following IntelliVue Instructions for Use (software revision G.01.xx):• MP20-MP90 Patient Monitor (M8000-9001K)• MP5 Patient Monitor (M8105-9001C)• MP2 Patient Monitor (M8102-9001B)• X2 Multi-Measurement Module (M3002-9001B)If your Instructions for Use has a different part number, you may discard this addendum, otherwise please store it with your monitor documentation.CO2 Alarm SpecificationsFor Microstream CO2 in the MP5 and for the M3015A, the CO2 alarm specifications are as follows:IntelliVue Telemetry Transceiver User InformationFor complete user information, including accessory listings and specifications, for the M4841A/M4851A TRx/TRx+ IntelliVue Telemetry Transceiver, please refer to the Instructions for Use provided with the device. MP2/X2 Including the Wireless IntelliVue AdapterRadiofrequency Radiation Exposure InformationFor body worn operation, this equipment has been tested and meets the FCC RF exposure guidelines when used with the accessories supplied or those approved for use with this product. Use of other accessories may not ensure compliance with FCC RF exposure guidelines. CO2 Alarm Specifications Range Adjustment DelayetCO2 High 20 to 95 mmHg (2 to 13 kPa) 1 mmHg (0.1 kPa) less than 21 seconds.etCO2 Low 10 to 90 mmHg (1 to 12 kPa)imCO2 High 2 to 20 mmHg (0.3 to 3.0 kPa)steps of 1 mmHg (0.1 kPa) less than 21 seconds.awRR High Adult/pedi: 10 to 100 rpmNeo: 30 to 150 rpmunder 20 rpm: 1 rpm stepsover 20 rpm:5 rpm steps less than 21 seconds.awRR Low Adult/pedi: 0 to 95 rpmNeo: 0 to 145 rpmsettings <20 rpm: less than 8 seconds>20 rpm: less than 21 secondsApnea delay 10 to 40 seconds 5 second steps set apnea delay time + 8 seconds (with 2 m sample lines) or 11 seconds (with 4 m sample lines).
Addendum to the Instructions for Use4MP5/MP5T - Intended UseThe monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment. The MP5 and MP5T monitors are also intended for use during patient transport inside the hospital environment; only the MP5 monitor for use during patient transport outside of the hospital environment. The MP5 and MP5T when used with TRx 4841A/4851A IntelliVue Telemetry System Transceiver are intended for use in a hospital environment and during patient transport inside the hospital environment.Measurement ModulesYou can plug in and unplug modules during monitoring. Insert the module until the lever on the module clicks into place. Remove a module by pressing the lever upwards and pulling the module out. A measurement automatically switches on when you plug the module in, and switches off when you unplug it. Reconnecting a module to the same monitor restores its label and measurement settings, such as alarms limits. If you connect it to a different monitor, the module remembers only its label.

Navigation menu