Rayence RY1417WCA Medical Image Processing Unit User Manual

Rayence Co., Ltd. Medical Image Processing Unit User Manual

User_Manual

       Release Version 1.0.0 Medical Image Processing Unit 1417WCA User Manual
  R-M-014-130312                         © Copyright 2012, Rayence Co. Ltd.   All  pages  of  this  document  contain  proprietary  and  confidential information of Rayence Corporation and are intended for exclusive use by Rayence Corporation personnel or customers. Copying, disclosure to others  or  other  use  is  prohibited  without  the  express  written authorization from Rayence Corporation. Please report any violations of requirement to Rayence Corporation.
  R-M-014-130312     Attention For improvement of product performance, supplementation, or follow-up of information; the contents of this manual are subject to change without separate prior notice.   Please note that our company has neither responsibility for any accidents nor obligation to do free repair service for any damage of the equipment due to user's mistake, which resulted from failure  to follow  the  contents  in  this  manual. Make  sure to be familiar  with  the  safety precautions  and  usage procedures. Also  note that  the product  may  slightly differ  from  the contents of this manual depending on specification.  The following marks are used for the effective use of the product in this manual.            Attention, consult accompanying documents.       This is used to emphasize essential information. Be sure to read this information to avoid incorrect operation. This indicates hazardous situation which, if not heeded, may result in minor or moderate injury to you or others, or may result in machine damage.  This indicates a potentially hazardous situation which, if not heeded, could result in death or serious injury to you or others.   Federal Law restricts this device to sale by or the order of a radiologist  or any other practitioners licensed by the law of the state in which that person practices to use or order the use of the device.
  R-M-014-130312      Contents Introduction ........................................................................................ 1 Intended use ...................................................................................... 1 Product features ................................................................................ 2 Product components ......................................................................... 3 Components Description................................................................... 5 Warning .............................................................................................. 8 Caution ............................................................................................. 10 Marking and labeling symbols ........................................................ 12 Notes for Using the Detector ............................................................ 14 Preparing .......................................................................................... 14 Handling ........................................................................................... 15 Handle Assembly ............................................................................. 15 Before Exposure .............................................................................. 17 During Exposure .............................................................................. 17 Limit of Load .................................................................................... 19 Disinfection and Cleaning ............................................................... 20 Others ............................................................................................... 20 Technical Features ........................................................................... 21 Mechanical Features ........................................................................ 21 Electrical Features ........................................................................... 21 Environmental requirement ............................................................. 24 PC Requirements ............................................................................. 25 Installation........................................................................................ 26 General interface ............................................................................. 26 Connection ....................................................................................... 27 IP set up ........................................................................................... 32 Using Web Manager (IP, SSID Change / Upgrade FW) ................... 33 Calibration ........................................................................................ 37 General Principle ............................................................................. 37 Calibration ........................................................................................ 39 Image Acquisition Test ..................................................................... 44 Program setup ................................................................................. 44 Get Image ......................................................................................... 45 View Images ..................................................................................... 46 Operation .......................................................................................... 51 Recommend ..................................................................................... 51 Switching power on / off .................................................................. 51 Storage ............................................................................................. 51 Information ....................................................................................... 52 Safety standard ................................................................................ 52 Electromagnetic Compatibility Information .................................... 54 Radio Frequency compliance.......................................................... 59 Maintenance ..................................................................................... 61 Maintenance ..................................................................................... 61 Cleaning ........................................................................................... 61 Inspection ........................................................................................ 62 Disposal or Recycling...................................................................... 63 Appendix ........................................................................................... 64
 1  R-M-014-130312      Introduction  Overview The 1417WCA is a wireless digital X-ray flat panel detector that can  generate  images  of  any  part  of  the  body.  The  wireless LAN((IEEE  802.11a/g/n)  communication  feature  improves  the operability,  and  high-speed  processing.  This  X-ray  imaging system  consists  of  a  scintillator  directly  coupled to  an  a-Si  TFT sensor. It makes high-resolution, high-sensitive digital images.  Intended use For U.S.A. 1417WCA  Digital  Flat  Panel  X-Ray  Detector  is  indicated  for digital  imaging  solution  designed  for  human  anatomy  including head,  neck,  spinal  column,  arm,  leg  and peripheral  (foot,  hand, wrist,  fingers,  etc.).  It  is  intended  to  replace  film  based radiographic  diagnostic  systems  and  provide  a  case  diagnosis and  treatment  planning  for  physicians  and  other  health  care professionals. Not to be used for mammography. For European Union This  device  provides  digital  X-ray  imaging  for  diagnosis  of disease,  injury,  or  any  applicable  health  problem.  The  image  is obtained  as  the  result  of  imaging  X-rays  passed  through  the human  body  with  an  X-ray  flat  panel  detector  and  importing  a digital signal output from the detector into the image processor.
 2  R-M-014-130312      Product features          Wi-Fi (802.11a/g/n)   Based on  a-Si TFT active matrix   Compact (15.9mm thickness) and light weight (Typ. 3.6kg)   Limiting Resolution : 3.9 lp/mm   14-bit digital output   Easy integration
 3  R-M-014-130312      Product components   Medical Image Processing Unit Photo Item Part Name Quantity        Detector  SD1417WCA  1 Handle  -  1       Battery pack  RB37WH  2 Battery charger  RC120W  1       Charger adapter PMP120-13-3  1     AGI  RA001A  1
 4  R-M-014-130312        Cables  Item Part Name Length Quantity      Link cable VRH076A  6m  1      P-Interface cable VRH017A  8m  1     USB cable(A to B) VRH078A  1.8m  1     AC Power cord  VRH018A/019A  1.8m  1    Installation CD o  Manual o  Detector Library    Option o  Additional Battery o  Charger for the Car o  Cover Bag o  AP package
 5  R-M-014-130312      Components Description The detector is designed to capture radiographic images. Captured images are transmitted to PC using the wireless/wired connection. The SparkLAN WPEA-121N 802.11 a/b/g/n half mini PCI-e module is implemented. It supports 2T2R (2 transmit 2 receive) MIMO technology, which delivers throughput up to 300Mbps. 1417WCA (FCC ID: QIIRY1417WCA) in the RF module (FCC ID: PPD-AR5BHB116) does not use DFS band. The module adopts the latest 802.11n Dual-Band technology (2.4Ghz and 5Ghz). The transmitter of the module is powered by host equipment(Detector). The antennas are 2 printed-dipole antennas.  (1)  Detector             1.  LED indicator: The LED indicates the state of detector. i.  Green on : Power on. ii.  Blue blinking : Wireless Connection. iii.  Orange on : Low battery. 2.  Handle  :  Hold  this  handle  when  carrying  the  sensor  unit.  It  is removable. 3.  Handle unlock-lever : This is an unlock-lever to remove handle.
 6  R-M-014-130312      4.  Link cable connector : This is a connector for Wire communication and  power  supplying.  Connect  the  detector  to  PC  and  SMPS(not provided) using Link cable. 5.  CFRP(Carbon Fiber Reinforced Plastic) : The part of the patient’s body to which an image is to be taken should be placed against this plate. 6.  Power button : Power on /  Power off switch. 7.  Shock  sensor  :  Detector  has  built-in  4 Shock  sensors.  It  detects and records impact and mishandling of fragile 8.  Battery unlock-lever : This is an unlock-lever to remove battery.  (2)  Battery & Charger          1.  Battery  :  Lithium ion  battery.  The number of times being  acquired image  is  600  images(@  cycle  time  :  15s).  The  batteries  last  2.5 hours and are rechargeable. 2.  Charger : Two port cradle type.  3.  LED indicator :  i.  Orange on : charging ii.  Green on : Charging is completed.
 7  R-M-014-130312      (3)  AGI  1.  Trigger  connector:  This  is  a  connector  for  synchronization between detector and generator. Connect the AGI to the generator using P-interface cable.. 2.  USB  connector:  This  is  a  connector  for  communication  between AGI and PC. Connect the AGI to the PC using USB cable.
 8  R-M-014-130312       Warning Environment of Use and Storage Follow the specified process of operational instructions written in this manual for the safety of the users and patients. Does  not  use  or  store  the  instrument  near  any  flammable chemicals such as thinner, benzene, etc. Also, this instrument is not a category AP or APG equipment. If chemicals are spilled or evaporate,  it may result  in fire or  electric  shock  through contact with  electric  parts  inside  the  instruments.  Also,  some disinfectants  are  flammable.  Be  sure  to  take  care  when  using them. Connection Do not connect the instrument with anything other than specified. Otherwise, it may result in fire or electric shock. Handling Always be sure to keep checking the condition of the system and the  patient  to  ensure  they  are  normal  during  the  use  of  the instrument.  If any problem  is found,  take  appropriate  measures, such as stopping the operation of the instrument, as required. Never disassemble or modify the product as  it may result in fire or  electric  shock.  Also,  since  the  instrument  incorporates  parts that  may  cause  electric  shocks  and  other  hazardous  parts, touching them may cause death or serious injury. Do  not  hit  or  drop  the  instrument.  The  instrument  may  be damaged  if  it  receives  a  strong  jolt,  which  may  result  in  fire  or electric shock if the instrument is used without being repaired.                                    Make sure                  to observe the following right
 9  R-M-014-130312      When Problem Occurs Should  any  of  the  following  occur,  immediately  turn  OFF  the power  of  each  instruments,  unplug  the  power  supply  cord  from the AC outlet, and contact Rayence representative or distributor.    When there is smoke, odd smell or abnormal sound.    When liquid  has been spilled  into the  instrument or  a  metal object has entered through an opening.    When the instrument has been dropped and it is damaged.  Maintenance and Inspection For  safety  reasons,  be  sure  to  turn  OFF  the  power  of  each instrument  when  the  inspections  indicated  in  this  manual  are going to be performed. Otherwise, it may result in electric shock. When the instrument is going to be cleaned, be sure to turn OFF the power of each instrument, and unplug the power supply cord from the AC outlet.  The instrument must be repaired by a qualified engineer only. If it is  not  repaired  properly,  it  may  cause  fire,  electric  shock,  or accident.                                      Make sure                  to observe the following right.
 10  R-M-014-130312      Caution Environment of Use and Storage Do  not  install  the  instrument  in  a  location  with  the  conditions listed  below.  Otherwise,  it  may  result  in  failure  or  malfunction, cause fire or injury.   Close to facilities where water is used.   Where it will be exposed to direct sunlight.   Close to air-conditioner or ventilation equipment.   Close to heat source such as a heater.   Prone to vibration.   Insecure place.   Dusty environment.   Saline or sulfurous environment.   High temperature or humidity.   Freezing or condensation. Do  not  place  the  storage  case  in  a  location  with  the  conditions listed below.   Where  the  cable  of  the  sensor  unit  will  be  strongly  pulled when the sensor unit is put into the case, otherwise, the cable may be damaged, resulting in fire or electric shock.    Where someone might get their foot caught in the cable of the sensor unit. Otherwise they could trip over, resulting in injury
 11  R-M-014-130312      Handling Do  not  spill  liquid or  chemicals onto  the  instrument or,  in  cases where the patient is injured, allow it to become wet with blood or other body fluids, as doing so may result in fire or electric shock. In such situation, protect the instrument with disposable covering as necessary. Wipe  the  CFRP  plate  of  the  sensor  unit  with  ethanol  or glutaraldehyde solution to disinfect it each time a different patient uses the instrument, in order to prevent infection. Turn  off  the  power  of  each  instrument  for  safety  when  they  are not going to be used. Maintenance and Inspection For  safety  reasons,  be  sure  to  inspect  the  instrument  before using it. In addition, carry out a regular inspection at least once a year. Modifications Any changes or modifications in construction of this device which are  not  expressly  approved  by  the  party  responsible  for compliance  could  void  the  user's  authority  to  operate  the equipment.
 12  R-M-014-130312      Marking and labeling symbols Symbols  Meaning  Caution : “Attention, see instructions for use”  Manufacturer  Date of manufacture  Serial number   WEEE  :  Waste  Electrical  and  Electronic Equipment  Authorized  representative  in  the  European Community  CE  symbol  grants  the  product  compliance  to the  European  Directive  for  Medical  Devices 93/42/EEC as a class Ⅱa device. Authorized by Notified Body SGS (code no.:0120) of British  non-ionizing radiation
 13  R-M-014-130312      Labels Detector label   Battery label
 14  R-M-014-130312      Notes for Using the Detector  Preparing Fully charge the battery pack.  Charge the  battery on the day of examination or on the previous day.   Battery  slowly  discharges  even  of  not  in  use.  The  battery pack  may  have  expired  if  it  discharges  immediately  after being  fully  charged.  You  can  purchase  an  optional  battery pack to replace an exhausted one.   Be sure to fully charge the battery before use. ※ The  battery  charger,  RC120W  is  designed  for  the dedicated battery pack. ※ When the detector will not be used for some time, remove the battery pack. Attach  the battery  pack. Align  the  claw on  the  battery pack and the groove on the battery bay. Insert the battery pack fully. Push down the battery pack. Slide the lock lever toward (lock) side and lock it.
 15  R-M-014-130312      Handling          Handle  the  instrument  carefully,  as  it  may  be  damaged  if something is hit against it, dropped, or receives a strong jolt.  Handle Assembly          ※ Insert the handle always in the same direction.   Handle Assembly
 16  R-M-014-130312                        Insert the handle in the same direction                                  If the  handle is  not locked, the detector can be dropped. Please confirm lever position. When the handle is locked, the lever should be same position with the right picture.
 17  R-M-014-130312      Before Exposure Be  sure  to check  the  equipment  daily  and  confirm  that  it  works properly. Sudden  heating  of  the  room  in  cold  areas  will  cause condensation  to  form  on  the  instrument.  In  this  case,  wait  until condensation  disappears  before  performing  exposure.  If  the instrument is used with condensation formed on it, problems may occur in the quality of the instrument. When an air-conditioner is going  to  be  used,  be  sure  to  raise/lower  the  temperature gradually  so that a difference in temperature in the  room and  in the  instrument  does  not  occur,  to  prevent  forming  of condensation.  During Exposure Do  not  use  the  detector  near  devices  generating  a  strong magnetic field. Doing so may produce image noise or artifacts. Do not apply excessive weight to the sensor unit. Otherwise, the sensor may be damaged.
 18  R-M-014-130312        Sleep Mode/Wake up                     Control “e” parameter : sleep time - 0 : disabled (default) - 1 : 1 sec - 2 : 2 sec … - 600 : 600 sec  If  you  set  sleep  time,  the  detector  goes  to  the  sleep  mode  after  “N”  sec  after acquiring image  If you want to wake up the detector, Press the power button(less than 1 second)                                  Do not  change  parameters except “Control e”
 19  R-M-014-130312      Limit of Load          Uniform load: 150 kg over the whole area of sensor window.         Local load: 100 kg on an area 40 mm in diameter.
 20  R-M-014-130312        Be sure to use the sensor unit on a flat place so it will not bend. Otherwise, the sensor may be damaged.  Disinfection and Cleaning Do not spray the detector directly with disinfectants or detergents.  Do not use anything other than neutral detergent for cleaning the cover of the instrument. Otherwise, the coating will be corroded.  Others Be  sure  to  reconnect  the  cables  to  the  proper  connectors. Otherwise, the instrument may malfunction or may be damaged.
 21  R-M-014-130312      Technical Features Mechanical Features Size  450 x 417 x 15.9 mm Weight  3.6 kg (not incl. Handle) Encapsulation Material Mg Window Material  Carbon fiber plate with 1.2 mm thickness  Electrical Features Detector Sensor Type  Amorphous Silicon with TFT (Single Panel) X-ray Converter  CsI:Tl Total Pixel Number  2816 × 3328 pixels Active Pixel Area  35.8 cm × 42.3 cm Active Pixel Number  2270 × 3280 pixels Pixel Pitch  127 μm Limiting Resolution  Max. 3.9 lp/mm Energy Range  40 - 150 kV A/D Conversion  14 bits Frame Rate  240 fph  Preview Time  < 3.0 sec Charge Integration Time  upto 4 sec Max. Linear Dose  Typ. 60 μGy @ SID 1500 mm  Saturation Dose  Typ. 70 μGy @ SID 1500 mm  Data Interface - Wireless - Wired  Wi-Fi (802.11a/g/n) Gigabit Ethernet   Under RQA5 condition (70kVp, 21mmAl)   Preview time may vary by complex factors
 22  R-M-014-130312      Battery Size  232.5 x 132.8 x 7 mm Weight  Typ. 0.3 kg Input  12.5 VDC Output  11.1 VDC Charging time  Typ. 2.5 hours Capacity  Typ. 3400 mAh The number of times being acquired image  600 images  The Battery level can be displayed on the LED status of battery.  If  the  battery  level  goes  down  under  25%,  please  charge  the battery Display  Battery level  75~100 %  50~75 %  25~50 %  0~25 %  Charger Size  267.5 x 100 x 77 mm Weight  Typ. 0.86 kg Input  20 VDC Output  11.1 VDC  Adapter Size  160 x 76 x 43.7mm (cable length: 1.28m) Weight  Typ. 0.92 kg  Input  100~240 VAC, 47~63 Hz, 1.4~0.6A Output  20 VDC, Max 6.0A
 23  R-M-014-130312      Wireless Spec Standard  802.11a/g/n compliance Peak Rate  300Mbps Frequency  2.4 GHz / 5 GHz Bandwidth  20MHz/40MHz MIMO  2x2   ※ Recommended Maximum operable distance : 7m (From the Access Point)  ※ 5.15~5.25 GHz band is restricted to indoor operations only. (for FCC) 5.15~5.350 GHz band is restricted to indoor operations only. (for CE) 5.15~5.350 GHz is indoor use only. (for Japan)
 24  R-M-014-130312      Environmental requirement Storage condition Item  Min.  Typ.  Max.  Unit  Note Temperature  -20    55  ℃  Humidity  10    80  % H.R.   Pressure  70    106  kPa   Shocks (Wrapping condition)  10G (25G)     16ms, 1000times, in 6directions, non-Driving Vibrations (Wrapping condition)  2G (10G)     10-150Hz, 10Sweeps, 1min/Octave, XYZ axis  Operation condition Item  Min.  Typ.  Max.  Unit  Note Temperature  5    35  ℃  Humidity  30    75  % H.R.   Pressure  70    106  kPa    ※ Regularly changed parts : Battery (warranty 6 months)
 25  R-M-014-130312      PC Requirements   Processor  :  At  least  Intel  Pentium  IV  HT  with  2.8GHz,  Intel Core Duo / Core 2 or comparable AMD Dual Core processor   At least 2 GB RAM   At  least  40  GB hard  disk for  the  software,  in  addition  to  the required archive capacity for image storage   Upper that Intel® Gigabit Network card   Graphics card/monitor  : Resolution of at  least 1,280 x  1,024 pixels in True Color mode   OS : Windows XP, vista, 7(32/64 bit)
 26  R-M-014-130312      Installation  The  Detector  is  composed  of  sensitive  electronic  parts  and components.  It  is  recommended  to  use  the  product  in  a  clean place and to exercise caution to ensure that it is not affected by dust or liquids. It is recommended to Use a dry and soft cloth to clean the detector housing.  General interface                         Portable                  Imaging                 processing unit must be installed in a way that enables the user to achieve optimal use
 27  R-M-014-130312      Connection  Power Connection  A. Connect the battery pack or power cable to the equipment. B. Connect the USB cable from your PC to AGI. ※ Be sure to sure only the dedicated battery pack, RB37WH for 1417WCA. Wireless Communication  A. AP Router(Line sharer) setting -  SSID : Griffon -  Internal network   - IP address : 2.2.2.1   - Subnet mask : 255.255.255.0   - Dynamic IP allocation range : 2.2.2.2~2.2.2.254 - Pre-Shared Key(Password) : project302   - Authentication methods : WPAPSK  or WPA2PSK   - Password methods : TKIP/AES - AP IP : 2.2.2.1 -  Channel (Frequency) - Avoid crowded channel (Using wireless detector under crowded channel result in low performance) -  If  available,  Use  ‘Auto-Channel  Selection’  function of router to find optimal channel
 28  R-M-014-130312       B. Reception Indicator Link LED flickering Blink Speed : Slow – Low link quality     Fast – High link quality  C. Checking Link Quality - After wireless connection is established, perform ‘Get Bright’ in ‘Calibration’ tap. - Check the value named ‘Wireless Signal’ in black log screen. Wireless Signal = Link Quality (Max. 100)     The value ‘Wireless Signal’ will be displayed here
 29  R-M-014-130312      Trigger Connection A. Connect the P-interface cable to the generator     X-ray Generator Connection Connect  the  P-interface  cable  between  the  AGI  box  and  X-ray generator.  A. Wired Mode 1 : P-interface cable mode        Operating description  Make assurance doubly sure   SIGNAL RATING before connection.
 30  R-M-014-130312        The window time can be changed. Refer to the following pictures
 31  R-M-014-130312                          Exposure                 Time(*clk)                 1333333 is designate to 0.5sec
 32  R-M-014-130312      IP set up  [My  Network  Places]  →  [Properties]  →  [Local  Area  Connection] → [Properties] → [Internet Protocol (TCP/IP)]  → [Use the following IP address] IP address : Obtain an IP address automatically                      IP address : Obtain an IP address automatically
 33  R-M-014-130312      Using Web Manager (IP, SSID Change / Upgrade FW) Change IP Address of Detector A. Turn on Detector and connect to PC (wired connection is recommended) B. After detector boot up, Launch web-browser (Optimized for Chrome/Internet Explorer) C. Connect to “http://[Detector’s IP]“  D. Select IP address to change from drop-down menu
 34  R-M-014-130312       E. Click “SUBMIT” F. Restart detector(Turn Off then On) Change SSID and PSK(Pre-Shared Key) A. Turn on Detector and connect to PC (wired connection is recommended) B. After detector boot up, Launch web-browser (Optimized for Chrome/Internet Explorer) C. Connect to “http://[Detector’s IP]“ D. Select SSID to set (This value should match to Router’s setting)
 35  R-M-014-130312       E. Type Pre-Shared Key to set (Password) (This value should match to Router’s setting Default:project302 )  F. Click “SUBMIT” G. Restart detector(Turn Off then On) Upgrade Firmware A. Turn on Detector and connect to PC (wired connection is recommended)
 36  R-M-014-130312      B. After detector boot up, Launch web-browser (Optimized for Chrome/Internet Explorer) C. Connect to “http://[Detector’s IP]“ D. Select firmware file by click “Choose File” button. (Released file is named ‘fw_ppc’)  E. Click ‘Upload File’ button   F. Click ‘Flash Write’ button   G. Restart detector(Turn Off then On)
 37  R-M-014-130312      Calibration  General Principle  Notation Calibration  can  be  done  by  image  acquisition  S/W.  The  gain-offset correction (under calibration) will be done with one dark, at least one bright and object frame.            Parameter  Description Offset  Dark image, acquired image without X-ray exposure Bright  Acquired image with X-ray exposure Object  Bright image with object, will be calibrated Gain  Gain of imaging system, offset subtracted image Offset correction  Offset subtract Gain correction  Compensate gain variance of pixel                 X-ray                  detector                  should be used at stable state within driving temperature range. Acquire the X-ray images after power on and 5 minutes warming up to obtain high quality images.
 38  R-M-014-130312      Bright Calibration Point To  gain  correction,  bright  frame  and  dark  frame  should  be acquired.  The dark  frame is  needed  only  one frame.  The  bright frame  is  recommended  to  be  acquired  more  than  8  different levels  of  median  values  of  bright  frames.  The  X-ray  source condition  will  be  recommended  the  tube  energy  level  of  70kVp with variable tube current and exposure time. To acquire at least 3  frames  at  same  condition  will  be  recommended.  The  median values of bright frames are below.  (Additional 21mm Al filter is recommended for calibration)   The Purpose of Bright Calibration The center of the non calibrated image is brighter than the edge due to hill effect of X-ray exposure. Generally, the intensity of X-ray  flux  at  center  region  of  exposed  area  is  higher  than surroundings  due  to  the  X-ray  expose  like  cone  shape.  A calibration  process  is  used  to  compensate  for  this  effect. Generally, called it ‘Flat Field Correction’(Bright calibration).     Point  1  2  3  4  5  6  7  8 Median Value [LSB] 850 ~ 1150 1850~ 2150 2850 ~ 3150 3850 ~ 4150 5850~ 6150 7850 ~ 8150 9850 ~ 10150 11850~ 12150                 The                  calibration                  range of bright is can be select by which exposure level is maximum level that user want to use. If the maximum level of user want to use is ‘20000’ in this case the level is contained in Bright point of ‘4’(refer ‘Table 5 : Median value’). The meaning is you don’t have to make bright point for ‘5’ and ‘6’(In this case, it will be does not working if you get image on higher level than maximum bright point.)
 39  R-M-014-130312      Calibration Describe the calibration step by step. 1st Step Move  to  “Calibration”  tap,  and  push  “Get  Dark”  button.  Acquire dark frame, the “dark.raw” will be generated at “\cal\” folder.
 40  R-M-014-130312      2nd Step Push “Get Bright” button at different six of X-ray condition. The X-ray condition should be set or tested before, same as the level of ‘1.2’. Push “Get Bright” button at least 3 times at same condition, and then the offset subtracted bright (gain) is generated which of filename is “xNNNNNA.raw(Refer to NOTE)                      Click                  button [Get                  Bright].  It will produce frame with name %CAL% xNNNNNA.raw, where NNNNN is median pixel’s value within current image borders after offset calibration (cut frame edges are never used during calibration). Suffix ‘A’ (it also could be ‘B’,’C’ etc) avoids casual coincidence of file names.
 41  R-M-014-130312      3rd step After 2nd  step,  the  “Generate”  button  will  be  activated.  Click  the button  “Generate”,  and  then  calibration  point  will  be  generated which of file name is “A ‘# of point’_ ‘median value of generated point’” like file of bright frame. The acquired bright frames within tolerance  value  which  is  variance  of  median  level  of  acquired bright  frames  will  be  averaged  and  generated  to  a  calibration point. The tolerance value can be edited.     After making calibration point the “Generate” button will be disabling.
 42  R-M-014-130312      4th step   After 3rd step, Change Bad Pixels Removal Tab, Click the button “Generate  Auto  BPM”,  and  then  Defect  Map  will  be  generated which of file name is “BPM.raw “ at the “\cal\” folder.
 43  R-M-014-130312      5th step For additional Defect correction, if  “BPMM.raw” is existed at the install CD, copy to the “\cal\” folder. 6th step On  Acquisition  Tab.  Check  the  box  “Offset  Calibration”,  “Gain Calibration”,  and  “Bad  Pix  Map”  for  activate  to  each  calibration and  Bad  Pixels  Removal.  Otherwise,  it  will  does  not  working when going to pre-processing .
 44  R-M-014-130312      Image Acquisition Test  Program setup  To acquire images, run _vadav.lnk program. Please set the following figures   Detector’s IP : 2.2.2.100   Detector‘s number of ADC : 26   Detector’s size of image : 2816 X 3328
 45  R-M-014-130312      Get Image  On  Acquisition  tab,  click  the  “Get  Image”  button  to  get  image. After  click  the  button,  you  can  see  pop-up  window,  which  is display window time and process of acquiring image.
 46  R-M-014-130312      View Images Frame-  and  image-files  have  extension  “raw”  and  contain  pixel data  in signed 14-bits little-endian format. One could view those files in Photoshop or another image editor.
 47  R-M-014-130312      Common controls and displayed statistics    Pixel_Min – minimum pixel value in frame- or image- data   Pixel_Max – maximum pixel value   Pixel_Black  – if  a pixel  ≤ Pixel_Black then  it  is  displayed as black (RGB 0, 0, 0)   Pixel_White –  if a  pixel  ≥ Pixel_Black then  it is displayed as black (RGB 255, 255, 255)
 48  R-M-014-130312      Histogram’s presentation Relative  Histogram  Scale  [H]=1000  means  that  the  distance depicted  as  “H”  on  the  drawing  matches  1%  of  total  number  of pixels.  Respectively  [H]=100  means  that  “H”  matches  0.1%  of pixels and [H]=500 means that “H” matches 0.5% of pixels.
 49  R-M-014-130312      Marker type “S” Displays local surround of selected location  Marker type “R” Display profile chart of a row.
 50  R-M-014-130312      Marker type “C” Display profile chart of a column.
 51  R-M-014-130312      Operation  Recommend X-ray  detector  should  be  used  at  stable  state  within  driving temperature range. Acquire the X-ray images after power on and 5 minutes warming up to obtain high quality images.  Switching power on / off   All connection should be done, before turn on the power.   Press  the  power  button  by  more  than  3  sec,  when  power on/off.   The green light of the LED indicator on the detector is on, the detector power is on.   The blue light of the LED indicator on the detector is blinking, the detector is getting prepared to work and initialize.   After power off, separate the battery.  Storage Store the sensor unit in clean and dry place. Ensure that storage place should be not affected by dust or liquids.
 52  R-M-014-130312      Information  Safety standard This  equipment  has  been  tested  and  found  to  comply  with  the limits  for  medical  devices  in  IEC  60601-1-2:1994.    These  limits are  designed  to  provide  reasonable  protection  against  harmful interference in a typical medical installation. This equipment generate, uses and  can radiate radio frequency energy  and,  if  not  installed  and  used  in  accordance  with  the instructions,  may cause  harmful  interference  to other devices  in the vicinity.  However, there is no guarantee that interference will not occur in a particular installation.  If this equipment does cause harmful  interference  to  other  devices,  which  can  be  determined by turning the equipment off and on, the user is encouraged to try to  correct  the  interference  by  one  or  more  of  the  following measures:   Reorient or relocate the receiving device.   Increase the separation between the equipment.   Connect  the  equipment  into  an  outlet  on  a  circuit  different from that to which the other devices are connected.   Consult the manufacturer or field service technician for help  (e.g., IEC 60950-1 for IT equipment and IEC 60601-1 series for medical electrical equipment.) In  addition,  all  such  combination  system  shall  comply  with  the standard  IEC  60601-1  and/or  IEC  60601-1  harmonized  national standard  or  the  combination.  If,  in  doubt,  contact  qualified technician or your local representative.                    Do not                   touch                  signal input, signal output or other connectors, and the patient simultaneously. External equipment intended for connection to signal input, signal output or other connectors, shall comply with relevant IEC Standard.
 53  R-M-014-130312        Type of protection against electric shock: Class I equipment   Classification  according  to  the  degree  of  protection  against ingress of water as detailed in the current edition of IEC 529: IPX0, ordinary equipment   This  equipment  is  not  suitable  for  use  in  the  presence  of flammable anesthetic s or oxygen.   Mode of operation: continuous operation
 54  R-M-014-130312      Electromagnetic Compatibility Information   Guidance and manufacturer’s declaration - electromagnetic emissions    The EUT is intended for use in the electromagnetic environment specified below.  The customer or the user of the EUT should assure that it is used in such an environment.  Immunity Test  Compliance   Electromagnetic Environment - Guidance  RF Emissions  CISPR 11  Group 1   The EUT uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment  RF Emissions  CISPR 11  Class A  The EUT is suitable for use in ail establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes  Harmonic emissions IEC 61000-3-2 Complies  The EUT is suitable for use in ail establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes  Voltage fluctuations/  Flicker emissions   IEC 61000-3-3 Complies  The EUT is suitable for use in ail establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes
 55  R-M-014-130312      Guidance and manufacturer’s declaration - electromagnetic immunity  The EUT is intended for use in the electromagnetic environment specified below.  The customer or the user of the EUT should assure that it is used in such an environment.   The EUT is intended for use in the electromagnetic environment specified below.  The customer or the user of the EUT should assure that it is used in such an environment.  Immunity test   IEC 60601-1-2  Test level  Compliance level Electromagnetic environment -guidance  Electrostatic discharge (ESD) IEC 61000-4-2 ±6kV Contact ±8kV air ±6kV Contact ±8kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.  Electrical fast transient/burst IEC 61000-4-4 ±2kV for power supply lines ± 1kV for input/output lines ±2kV for power supply lines ± 1kV for input/output lines Mains power quality should be that of a typical commercial or hospital environment.  Surge IEC 61000-4-5 ±1kV differential mode ±2kV common mode ±1kV differential mode ±2kV common mode Mains power quality should be that of a typical commercial or hospital environment.  Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 <5% Uт  (>95% dip in Uт)  for 0.5 cycle  40% Uт  (60% dip in Uт )  for 5 cycle  70% Uт  (30% dip in Uт)  for 25 cycle  <5% Uт  (<95% dip in Uт ) for 5 s  <5% Uт  (>95% dip in Uт)  for 0.5 cycle  40% Uт  (60% dip in Uт )  for 5 cycle  70% Uт  (30% dip in Uт)  for 25 cycle  <5% Uт  (<95% dip in Uт ) for 5 s  Mains power quality should be that of a typical commercial or hospital environment. If the user of the EUT image intensifier requires continued operation during power mains interruptions, it is recommended that the EUT image intensifier be powered from an uninterruptible power supply or a battery.  Power frequency (50/60Hz) magnetic field IEC 61000-4-8 3 A/m  3 A/m  Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.  NOTE Uт is the a.c. mains voltage prior to application of the test level.
 56  R-M-014-130312      Guidance and manufacturer’s declaration - electromagnetic immunity  The EUT is intended for use in the electromagnetic environment specified below.  The customer or the user of the EUT should assure that it is used in such an environment.   Immunity test  IEC 60601-1-2  test level  Compliance  level  Electromagnetic environment - guidance        Conducted RF  IEC 61000-4-6    Radiated RF  IEC 61000-4-3         3 Vrms  150 kHz to 80MHz   3 V/m  80  MHz  to 2.5GHz          V1=3Vrms    E1=3V/m Portable and mobile RF communications equipment should be used no closer to any part of the EUT, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.    Recommended separation distance : PVd15.3  PEd15.3  80MHz to 800MHz PEd17 800MHz to 2.5GHz  where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).  Field strengths from fixed RF transmitters, as deter-mined by an electromagnetic site survey, a  should be less than the compliance level in each frequency range. b  Interference may occur in the vicinity of equipment marked with the following symbol :   NOTE 1) At 80MHz and 800MHz, the higher frequency range applies.  NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and  land  mobile  radios,  amateur  radio,  AM  and  FM  radio  broadcast  and  TV  broadcast  cannot  be predicted  theoretically  with  accuracy.  To  assess  the  electromagnetic  environment  due  to  fixed  RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the EUT is used exceeds the applicable RF compliance level above, the EUT should be  observed  to  verify  normal  operation.  If  abnormal  performance  is  observed,  additional  measures may be necessary, such as re-orienting or relocating the EUT. b Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.
 57  R-M-014-130312       Recommended  separation  distances  between  portable  and  mobile  RF communications equipment and the EUT   There is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The  customer  or  the  user  of  the  EUT  can  help  Prevent  electromagnetic  interference  by maintaining  a  minimum  distance  between  portable  and  mobile  RF  communications  equipment (transmitters) and the EUT as recommended below, according to the maximum output power of the communications equipment.   Separation distance according to frequency of transmitter [m] IEC 60601-1-2 Frequency of Transmitter  Equation   Rated maximum output power of transmitter [W] 150kHz to 80MHz   PVd15.3    V1=3Vrms  Separation Distance (meters) 80MHz to 800MHz   PEd15.3    E1=3V/m  Separation Distance (meters) 800MHz to 2.5GHz  PEd17   E1=3V/m  Separation Distance (meters) 0.01   0.116   0.1166   0.2333  0.1   0.368   0.3687   0.7378 1   1.166   1.1660   2.3333 10   3.687   3.6872   7.3785 100   11.660   11.6600   23.333 For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended  separation distance  d  in  meters  (m)  can  be  estimated  using  the  equation  applicable  to  the  frequency  of  the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.   NOTE 1) At 80MHz and 800MHz, the separation distance for the higher frequency range applies.  NOTE 2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and  land  mobile  radios,  amateur  radio,  AM  and  FM  radio  broadcast  and  TV  broadcast  cannot  be predicted  theoretically  with  accuracy.  To  assess  the  electromagnetic  environment  due  to  fixed  RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the EUT is used exceeds the applicable RF compliance level above, the EUT should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the EUT. b Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.  Name  Shield Type  Length  Notes Link cable  Shielded  6 m  Provided with wiring unit. P-interface cable  Non-shielded 8 m  Provided with X-ray interface unit. USB cable (A to B)  Non-shielded 1.8 m  Provided with AGI unit. AC power cord (220V)  Shielded  1.8 m  Provided with charger unit.
 58  R-M-014-130312         Immunity and Compliance Level  Immunity test   IEC 60601-1-2  Test Level  Actual Immunity Level  Compliance Level Conducted RF  IEC 61000-4-6  3Vrms  150kHz to 80MHz 3Vrms  3Vrms  Radiated RF IEC 61000-4-3  3Vrms  80MHz to 2.5GHz 3V/m  3V/m   Guidance and manufacturer’s declaration - electromagnetic immunity   The EUT is intended for use in the electromagnetic environment specified below.  The  customer  or  the  user  of  the  EUT  should  assure  that  it  is  used  in  such  an  electromagnetic environment.  Immunity test  IEC 60601-1-2 Test level Compliance level Electromagnetic environment - guidance Conducted RF IEC 61000-4-6     Radiated RF IEC 61000-4-3 3 Vrms 150kHz to 80MHz    3V/m 80MHz to 2.5GHz 3 Vrms 150 kHz to 80MHz    3V/m 80MHz to 2.5GHz The  EUT must  be  used  only in  a  shielded  location with a minimum RH shielding effectiveness nad, for each  cable  that  enters  the  shielded  location  with  a minimum  RF  shielding  effectiveness  and,  for  each cable that enters the shielded location.  Field  strengths  outside  the  shielded  location  from fixed  RF  transmitters,  as  determined  by  an electromagnetic  site  survey,  should  be  less  than 3V/m.a  Interference may  occur in  the vicinity of  equipment marked with the following symbol:   NOTE 1)  These guidelines  may not  apply in  all  situations.  Electromagnetic propagation is affected  by absorption and reflection from structures, objects and people.  NOTE  2)  It  is  essential  that  the  actual  shielding  effectiveness  and  filter  attenuation  of  the  shielded location be verified to assure that they meet the minimum specification. a Field  strengths from fixed transmitters,  such as  base stations for  radio  (cellular/cordless) telephones and  land  mobile  radios,  amateur  radio,  AM  and  FM  radio  broadcast  and  TV  broadcast  cannot  be predicted  theoretically  with  accuracy.  To  assess  the  electromagnetic  environment  due  to  fixed  RF transmitters, an electromagnetic site survey should be considered. If the measured field strength outside the  shielded  location  in  which  the EUT is  used  exceeds  3V/m,  the  EUT  should  be  observed  to  verify normal operation.  If  abnormal  performance  is  observed,  additional  measures  may  be  necessary,  such  as  relocating  the EUT or using a shielded location with a higher RF shielding effectiveness and filter attenuation.
 59  R-M-014-130312      Radio Frequency compliance     FCC/IC Notice (U.S.A and CANADA)  FCC Part 15 Subpart C §15.247 and IC RSS-210 Issue 7 FCC Part 15 Subpart E §15.407 and IC RSS-210 Issue 7   FCC ID: QIIRY1417WCA  5.15- 5.25 GHz band is restricted to indoor operations only. Host device of the approved module shall be marked with the following item: Contains Transmitter Module FCC ID: PPD-AR5BHB116 Compliance with FCC requirement 15.407(c) Data transmission  is always  initiated by software,  which is the  passed down through the MAC, through the digital and analog baseband, and finally to the RF chip. Several special packets are initiated by the MAC. These are the only ways the digital baseband portion will turn on the RF transmitter, which it  then  turns  off  at  the  end  of the packet. Therefore, the transmitter will be on only while one of the aforementioned packets is  being  transmitted.  In  other  words,  this  device  automatically discontinues  transmission  in  case  of  either  absence  of  information  to transmit or operational failure. This  device  complies  with  part  15  of  the  FCC  Rules.  Operation  is subject  to the following two conditions:  (1)  This device may  not cause harmful interference,  and  (2)  this  device must  accept  any  interference received, including interference that may cause undesired operation.   FCC CAUTION Changes  or  modifications  not  expressly  approved  by  the  party responsible for compliance could void the user’s authority to operate the equipment.  This  transmitter  must  not  be  co-located  or  operated  in conjunction with any other antenna or transmitter.   When installing it in a mobile equipment This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment and meets the FCC radio frequency (RF) Exposure  Guidelines in  Supplement  C  to  OET65.  This  equipment  has very low levels of RF energy that it deemed to comply without maximum
 60  R-M-014-130312      permissive exposure evaluation (MPE). But it is desirable that it should be  installed  and  operated  keeping  the  radiator  at  least  20cm  or  more away from person’s body (excluding extremities: hands, wrists, feet and ankles).    When  you  use  the  detector  with  wire  mode,  the  wireless  function  is automatically off.
 61  R-M-014-130312      Maintenance Maintenance   Maintenance of the detector should be done by an authorized service provider    If  the  Detector  Panel  is  defective,  the  detector  will  be returned as is to the manufacturer for repair    Clean  the  equipment  with  a  dry  soft  cloth,  or  a  soft  cloth lightly moistened with mild detergent solution. Do not use any type of solvent, such as benzene    This equipment and accessories are to be disposed of safely after  the  life  span  of  them  and  national  regulation  must  be observed.   For safety reasons, be  sure to  inspect the  instrument before using  it.  In  addition,  carry  out  a  regular  inspection  at  least once a year.   Arrange  the detector and power supply  link cable  to  prevent the damage of  the cable’s  rubber  tube.  For example, do not press the cable under the legs of the table or the people.   This  equipment  must  only  be  connected  to  a  supply  mains with protective earth. Cleaning Cover Clean the cover by the left warning box, if it is dirty. CFRP of Detector Wipe  the  CFRP  plate  of  the  sensor  unit  with  ethanol  or glutaraldehyde solution to disinfect it each time a different patient uses the instrument, in order to prevent infection. If you are using disinfectant other  than those specified above, or you are  mixing another disinfectant with ethanol, please also consult a specialist, because they may harm the CFRP plate.                When the                 instrument                               is going to be cleaned, be sure to turn OFF the power of each instrument, and unplug the power cable from the AC outlet. Otherwise, fire or electric shock may result.
 62  R-M-014-130312       Inspection In order to ensure that the instrument is used safely and normally, please  be  sure  to  inspect  the  instrument  before  use.  If  any problem  is  found  during  the  inspection,  please  take  measures indicated  in  this  chapter.  If  problem  still  cannot  be  corrected, please  contact  Rayence  representative  or  distributor.  It  is recommended that a record of the inspection be kept by making copies  of  the  check  lists  in  this  section,  or  making  a  separate check list. Inspection chart Inspection  User Vendor Cycle Check  that  cables  are  not damaged  or  cover  of  cables  is not torn O    Daily Check  that  the  plugs  and  locks of connectors are not loose  O    Daily Check  that  the  cover  or  parts are not damaged and not loose  O    Daily Check the LED indicator  O    Daily Check the bad pixel map    O  Half year Check  the  performance  of  the instrument  by  performing exposures using a phantom or a resolution chart   O  Yearly                      For                 safety                  reasons, be sure to turn OFF the power of the detector when the  following inspections are going to be performed. Otherwise, it may result in electric shock.
 63  R-M-014-130312      Disposal or Recycling  Follow local governing ordinances and recycling plans regarding the disposal or recycling of device components.  Disposal of old Electrical & Electronic Equipment    (Application  in  the  European  Union  and  other European  countries  with  separate  collection system.)  This  symbol  indicates  that  this  product  shall  not  be treated  as household  waste.  Instead,  it  shall be handed  over  to the  applicable  collection  point  for  the  recycling  of  electrical  and electronic  equipment.  By  ensuring  this  product  is  disposed  of correctly,  you  will help prevent potential negative consequences for the environment and human health, which could otherwise be caused by inappropriate waste handling of this product. For more detailed information  about recycling  this product, please  refer to local governing ordinances and recycling plans.
 64  R-M-014-130312      Appendix Dimension                 [unit : mm]
 65  R-M-014-130312         Rayence Co., Ltd. 14, Samsung 1ro 1-gil,  Hwaseong-si, Gyeonggi-do, Korea www.rayence.com Medical Image Processing Unit 1417WCA

Navigation menu