SORIN CRM CRTD9750 Implantable cardioverter defibrillator User Manual TABLE OF CONTENTS

SORIN CRM Implantable cardioverter defibrillator TABLE OF CONTENTS

Users Manual

Download: SORIN CRM CRTD9750 Implantable cardioverter defibrillator User Manual TABLE OF CONTENTS
Mirror Download [FCC.gov]SORIN CRM CRTD9750 Implantable cardioverter defibrillator User Manual TABLE OF CONTENTS
Document ID1737998
Application ID7rDyzOG8sfYjb0oWizTcmA==
Document DescriptionUsers Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize49.83kB (622879 bits)
Date Submitted2012-07-06 00:00:00
Date Available2012-12-29 00:00:00
Creation Date2012-05-24 14:24:27
Producing SoftwareMicrosoft® Word 2010
Document Lastmod2012-05-24 14:24:27
Document TitleTABLE OF CONTENTS
Document CreatorMicrosoft® Word 2010
Document Author: Anne Couturier

TABLE OF CONTENTS
1. General description .......................................................................... 6
2. Indications ......................................................................................... 6
3. Contraindications ............................................................................. 7
4. Warnings and precautions .............................................................. 7
4.1. Risks related to medical environment .................................. 9
4.2. Sterilization, storage and handling ..................................... 11
4.3. Implantation and device programming .............................. 11
4.4. Lead evaluation and lead connection ................................. 13
4.5. Generator explant and disposal .......................................... 14
5. Adverse events ............................................................................... 15
5.1. MSP study ............................................................................. 15
5.2. Potential adverse events ..................................................... 18
6. Clinical studies ............................................................................... 20
6.1. MSP clinical study ................................................................ 20
7. Patient selection and treatment .................................................... 26
7.1. Individualization of treatment .............................................. 26
7.2. Specific patient populations ................................................ 27
8. Patient counselling information .................................................... 28
9. Conformance to standards ............................................................ 28
10. Physician guidelines .................................................................... 32
10.1. Physician training ............................................................... 32
10.2. Directions for use ............................................................... 33
10.3. Maintaining device quality ................................................. 33
10.4. V-V Programming Recommendation ................................ 33
11. Patient information ....................................................................... 34
12. How supplied ................................................................................ 34
12.1. Sterility................................................................................. 34
12.2. Warranty and replacement policy ..................................... 34
13. Device description ........................................................................ 35
14. Implant procedure ........................................................................ 38
14.1. Necessary equipment ........................................................ 38
14.2. Packaging ............................................................................ 38
14.3. Optional equipment ............................................................ 39
14.4. Before opening the package ............................................. 39
14.5. Prior to implantation........................................................... 39
14.6. Device placement ............................................................... 40
14.7. Choosing the type of lead .................................................. 40
14.8. Measurement of thresholds at implant................................. 41
14.9. Lead connection ................................................................. 42
14.10. Device implantation .......................................................... 43
14.11. Tests and programming ................................................... 44
15. Special modes............................................................................... 44
15.1. Safety mode (nominal values) ........................................... 44
15.2. Magnet mode....................................................................... 45
15.3. Response in the presence of interference ................................ 46
15.4. Detection characteristics in the presence of
electromagnetic fields ...................................................... 46
15.5. Protection against short-circuits ...................................... 47
16. Main functions .............................................................................. 47
16.1. Automatic lead measurements ......................................... 47
16.2. Atrial tachyarrhythmia management ................................ 47
16.3. Ventricular tachyarrhythmia management....................... 48
16.4. Pacing .................................................................................. 48
16.5. Sensing ................................................................................ 49
16.6. Follow-up function.............................................................. 50
16.7. Remote monitoring function ............................................. 50
17. Patient follow-up ........................................................................... 54
17.1. Follow-up recommendations............................................. 54
17.2. Elective Replacement Indicator (ERI) ............................... 56
17.3. Explantation ........................................................................ 57
17.4. Defibrillator identification .................................................. 58
18. Supplemental information ........................................................... 59
18.1. Adverse events in the SafeR study................................... 59
18.2. SafeR clinical study ............................................................ 61
19. Physical characteristics ............................................................... 64
19.1. Materials used ..................................................................... 64
20. Electrical characteristics ............................................................. 65
20.1. Table of delivered shock energy and voltage .................. 66
20.2. Battery ................................................................................. 66
20.3. Longevity ............................................................................. 67
21. Programmable parameters .......................................................... 69
21.1. Antibradycardia pacing ...................................................... 69
21.2. Ventricular tachyarrhythmia detection ............................. 74
21.3. Ventricular tachyarrhythmia therapies ............................. 76
21.4. Remote alerts and warnings.............................................. 80
22. Non programmable parameters .................................................. 82
23. Limited Warranty .......................................................................... 83
23.1. Article 1 : Terms of Limited warranty ............................... 83
23.2. Article 2 : Terms of replacement ....................................... 85
24. Patents ........................................................................................... 86
25. Explanation of symbols ............................................................... 87
1. GENERAL DESCRIPTION
PARADYM RF CRT-D 9750 is an implantable cardioverter defibrillator
for the recognition and treatment of ventricular tachycardia and
fibrillation, with ventricular resynchronization, in patients with
spontaneous or inducible tachyarrhythmias. PARADYM RF CRT-D
9750 is equipped with an accelerometer to allow adaptation of pacing
to suit the patient’s activity.
PARADYM RF CRT-D 9750 provides high energy shocks (42 J) for
enhanced safety, as well as automatic lead measurements to monitor
system integrity.
PARADYM RF CRT-D 9750 is protected against high-frequency signals
emitted by cellular telephones.
2. INDICATIONS
PARADYM RF CRT-D 9750 is indicated for ventricular antitachycardia
pacing and ventricular defibrillation for automated treatment of life
threatening arrhythmias.
The device is also indicated for the reduction of heart failure symptoms
in medically optimized NYHA Functional Class III and IV patients with
left ventricular ejection fraction of 35% or less, and a QRS duration of
150 ms or longer.
6 – US-ENGLISH
3. CONTRAINDICATIONS
Implantation of PARADYM RF CRT-D 9750 is contraindicated in patients:
─ whose ventricular tachyarrhythmias may have transient or
reversible causes such as: acute myocardial infarction, digitalis
intoxication, drowning, electrocution, electrolyte imbalance,
hypoxia, sepsis, or unstable ischemic episodes,
─ who present incessant tachyarrhythmia,
─ who have an internal pacemaker,
─ whose primary disorder is bradyarrhythmias, or atrial
tachyarrhythmias.
Dual-chamber and single chamber atrial pacing is contraindicated in
patients with chronic refractory atrial tachyarrhythmias.
4. WARNINGS AND PRECAUTIONS
The patient should be warned of the potential risks of defibrillator
malfunction if he is exposed to external magnetic, electrical, or
electromagnetic signals.
These potential interference sources may cause conversion to inhibited
mode (because of noise detection), erratic delivery of VT or VF
therapies, nominal programming, or much more rarely, irreversible
damage to the device’s circuits.
The main sources of high magnitude electrical interference are:
powerful radiofrequency equipment (radar), industrial motors and
transformers, induction furnaces, resistance and arc-welding
equipment, and high power loudspeakers.
US-ENGLISH – 7
Be aware that the changes in the patient’s condition, drug regimen, and
other factors may change the defibrillation threshold (DFT) which may
result in non-conversion of the arrhythmia post-operatively. Successful
conversion of ventricular fibrillation or ventricular tachycardia during
arrhythmia conversion testing is no assurance that conversion will
occur post-operatively.
Resuscitation Availability: Do not perform device testing unless an
external defibrillator and medical personnel skilled in cardiopulmonary
resuscitation (CPR) are readily available.
Electrical Isolation: Do not permit the patient to contact grounded
equipment that could produce hazardous leakage current. Ensuing
arrhythmia induction could result in the patient’s death.
Disable the ICD During Handling: Program Shock Therapy to OFF
during surgical implant and explant or post mortem procedures. The
device can deliver a serious high energy shock should accidental
contact be made with the defibrillation electrodes.
Antitheft gates: Since antitheft devices at the entrance to stores are
not subject to any safety standards, it is advisable to spend as little
time as possible in their vicinity.
Airport detection systems: Since airport detection systems are not
subject to any safety standards, it is advisable to spend as little time as
possible in their vicinity.
High voltage power transmission lines: High voltage power
transmission lines may generate enough EMI to interfere with
defibrillator operation if approached too closely.
Communication equipment: Communication equipment such as
microwave transmitters, linear power amplifiers, or high-power amateur
transmitters may generate enough EMI to interfere with defibrillator
operation if approached too closely.
8 – US-ENGLISH
Home appliances: Home appliances that are in good working order
and properly grounded do not usually produce enough EMI to interfere
with defibrillator operation. There are reports of device disturbances
caused by electric hand tools or electric razors used directly over the
device implant site.
CAUTION: Do not tap sharply on the ICD can after implant, because
the ICD's sensing circuits can detect this as P-waves or R-waves, and
such oversensing could result in inappropriate pacing, inhibition, or
therapy. Normal activities after implant do not result in such
oversensing.
4.1. RISKS RELATED TO MEDICAL ENVIRONMENT
It is advisable to carefully monitor defibrillator operation prior to and
after any medical treatment during which an electrical current from an
external source passes through the patient's body.
Magnetic Resonance Imaging: MRI is strictly contraindicated in
cardiac defibrillator patients.
Radiofrequency ablation: A radio frequency ablation procedure in a
patient with a generator may cause device malfunction or damage.
RF ablation risks may be minimized by: 1. Programming Shock Therapy
and ATP to OFF. 2. Avoiding direct contact between the ablation catheter
and the implanted lead or generator. 3. Positioning the ground, placing it
so that the current pathway does not pass through or near the device, i.e.
place the ground plate under the patient’s buttocks or legs. 4. Having
external defibrillation equipment available.
Electrocautery or diathermy device: Diathermy and electrocautery
equipment should not be used. If such devices must be used: 1. Keep
the current path and ground plate as far away from the device and the
leads as possible (a minimum of 15 cm [six inches]). 2. Before
procedure, deactivate ATP and shock therapies. 3. During the
procedure, keep the electrocautery device as far as possible from the
cardiac defibrillator. Set it at minimum intensity. Use it briefly. 4. After
US-ENGLISH – 9
the procedure, check for proper implant function. The device should
never be exposed directly to the diathermy source.
External defibrillation: PARADYM RF CRT-D 9750 is protected from
external defibrillation shocks. Before external defibrillation, deactivate
ATP and shock therapies. During external defibrillation, it is advisable
to avoid placing the defibrillating paddles directly over the casing or
over the leads. The defibrillating paddles should preferably be placed in
an anteroposterior position. Avoid any direct contact between the
defibrillation paddles and the conductive parts of the implanted leads or
casing of the implanted device. After external defibrillation, check for
proper device function.
Radiation therapy: Avoid exposure to ionizing radiation. Betatrons are
contraindicated. If high doses of radiation therapy cannot be avoided,
the defibrillator should be protected from direct exposure with a screen.
ATP and shock therapies should be disabled during exposure and
proper device function should be checked regularly afterwards.
Resulting damage may not be immediately detectable.
If irradiation of tissues close to the implantation site is necessary, it is
recommended that the cardiac defibrillator be moved. As a safety
measure, an external defibrillator should be immediately available.
Lithotripsy: Lithotripsy may permanently damage the device if it is at
the focal point of the lithotripsy beam. If lithotripsy must be used, keep
the defibrillator at least 2.5 to 5 cm (1-2 inches) away from the focal
point of the lithotripsy beam.
Diagnostic ultrasound (echography): The defibrillator is not affected
by ultrasound imaging devices.
Scales with body fat monitors and electronic muscle stimulators:
A patient with an implanted PARADYM RF CRT-D 9750 should not use
these devices.
10 – US-ENGLISH
4.2. STERILIZATION, STORAGE AND HANDLING
Resterilization: Do not resterilize and re-implant explanted ICDs.
"Use Before" Date: A "Use Before" date is printed on the outer storage
package and on the sterile package. Do not implant the device after this
date because the battery may have reduced longevity and sterility may
be affected. It should be returned to Sorin CRM.
If Package is damaged: Do not use the device or accessories if the
packaging is wet, punctured, opened or damaged because the integrity
of the sterile packaging may be compromised. Return the device to the
manufacturer.
Device Storage: Store the device in a clean area, away from magnets, kits
containing magnets, and sources of electromagnetic interference to avoid
device damage. Store the device between 0 - 50 °C (32 - 122 °F).
Temperatures outside the specified range may damage the device.
Equilibration: Allow the device to reach room temperature before
programming or implanting the device because rapid temperature
changes may affect initial device function.
4.3. IMPLANTATION AND DEVICE PROGRAMMING
Use only a Sorin CRM programmer to communicate with the device.
Do not inadvertently position any magnet over the ICD; this suspends
tachyarrhythmia detection and treatment.
Replace the device when the programmer displays an ERI (defined by
a battery voltage of 2.66 ± 0.01 V or a magnet rate lower than or equal
to 80 bpm).
Program device parameters such as sensitivity threshold and VT and
VF detection intervals as specified in the device manuals.
Lead System: Do not use a lead system other than those with
demonstrated compatibility because undersensing cardiac activity and
failure to deliver necessary therapy may result.
US-ENGLISH – 11
In situations where an ICD and a pacemaker are implanted in the same
patient, interaction testing should be completed. If the interaction
between the ICD and the pacemaker cannot be resolved through
repositioning of the leads or reprogramming of either the pacemaker or
the ICD, the pacemaker should not be implanted (or should be
explanted if previously implanted).
Failure to properly insert the torque screwdriver into the perforation at
an angle perpendicular to the connector receptacle may result in
damage to the sealing system and its self-sealing properties.
It is recommended that a security margin of at least 10 J be
demonstrated between the effective shock energy and maximum
programmable energy. Carefully confirm that true ventricular fibrillation
has been induced because the DFT for ventricular tachycardia or flutter
may be lower.
The defibrillator should be implanted with the engraved side facing outwards
in order to facilitate telemetric communication with the programming head
and to display the radiographic identification correctly.
12 – US-ENGLISH
4.4. LEAD EVALUATION AND LEAD CONNECTION
PARADYM RF CRT-D 9750 has two DF-1 and three IS-1 connector
ports. IS-1 refers to the international standard whereby leads and
generators from different manufacturers are assured a basic fit (ISO
5841-3:2000). DF-1 refers to the international standard for defibrillation
lead connectors (ISO 11318:2002).
Do not tie a ligature directly to the lead body, tie it too tightly, or
otherwise create excessive strain at the insertion site as this may
damage the lead. Use the lead stabilizer to secure the lead lateral to
the venous entry site.
Do not grip the lead with surgical instruments.
Do not use excessive force or surgical instruments to insert a stylet into
a lead.
Use ventricular transvenous leads with caution in patients with either a
mechanical or bioprosthetic tricuspid valvular prosthesis.
Use the correct suture sleeve (when needed) for each lead, to
immobilize the lead and protect it against damage from ligatures.
Never implant the system with a lead system that has a measured
shock impedance of less than 30 ohms. A protection circuit in the
defibrillator prevents shock delivery when impedance is too low. If the
shock impedance is less than 30 ohms, reposition the lead system to
allow a greater distance between the electrodes.
Do not kink leads. Kinking leads may cause additional stress on the
leads, possibly resulting in lead fracture.
Do not insert a lead connector pin into the connector block without first
visually verifying that the setscrews are sufficiently retracted.
Do not tighten the setscrews unless a lead connector pin is inserted
because it could damage the connector block.
Lead electrodes in contact during a cardioversion or defibrillation therapy
will cause current to bypass the heart, possibly damaging the ICD and the
US-ENGLISH – 13
leads. While the ICD is connected to the leads, make sure that the metal
portions of any electrodes do not touch each other.
If a pacing lead is abandoned rather than removed, it must be capped to
ensure that it is not a pathway for currents to or from the heart.
If a thoracotomy is required to place epicardial patches, it should be done
during a separate procedure to reduce the risk of morbidity and mortality.
Do not place the patch lead over nerve tissue as this may cause nerve
damage.
Place the patch lead with the conducting coil side facing the heart to
ensure delivery of energy to the heart.
Place the sutures well outside the coil of the patch lead or in the area
between the coils to avoid possible coil fracture.
If countershock is unsuccessful using external paddles, adjust the
external paddle position (e.g., anterior-lateral to anterior-posterior) and
be sure that the external paddle is not positioned over the patch.
Do not fold, alter, or remove any portion of the patch as it may
compromise electrode function or longevity.
If a header port is unused on the generator, the port must be plugged
to protect the generator.
4.5. GENERATOR EXPLANT AND DISPOSAL
Interrogate the device, and program shock therapy off prior to explanting,
cleaning or shipping the device to prevent unwanted shocks.
Return all explanted generators and leads to the manufacturer.
Never incinerate the device due to the potential for explosion.
The device must be explanted before cremation.
14 – US-ENGLISH
5. ADVERSE EVENTS
Clinical data presented in this section are from the MSP clinical study.
PARADYM RF CRT-D is similar in design and clinical function to the
ALTO 2 MSP and OVATIO CRT-D devices. The data provided are
applicable to PARADYM RF CRT-D.
5.1. MSP STUDY
Sorin CRM conducted an international, multi-center, randomized clinical
trial of its cardiac resynchronization therapy system. Investigators
attempted to implant study devices in 190 patients. A total of 182 patients
received study devices and had an exposure of over 165 device years. Of
those patients, 19 received OVATIO CRT-D, 160 received ALTO 2 MSP,
and 3 received ALTO MSP. The clinical data collected on ALTO MSP,
ALTO 2 MSP and OVATIO CRT-D are applicable to PARADYM RF CRTD. The table below summarizes the adverse events observed for the
CRT-D system. No deaths were related to the system.
US-ENGLISH – 15
Event
# of
Patients
% of
Patients
# of
Events
Events/
100 DeviceYears
Deaths not related to the
system
16
8.4
16
0.8
Cardiac arrest
2.6
0.3
Worsening CHF / CHF
decompensation
1.6
0.2
Multi-organ dysfunction
1.1
0.1
Complications related to the
system
28
14.7
35
2.1
Dislodgment or migration
4.7
11
0.6
Extracardiac stimulation
(e.g., phrenic stim)
4.7
0.5
Complications related to the
implant procedure
18
9.5
21
1.3
Dislodgment or migration
2.1
0.2
Observations related to the
system
23
12.1
27
1.7
Extracardiac stimulation
(e.g., phrenic stim)
12
7.9
15
0.8
Observations related to the
implant procedure
24
12.6
28
1.7
Heart block
3.2
0.3
16 – US-ENGLISH
Extracardiac stimulation
(e.g., phrenic stim)
1.5
0.3
Event
# of
Patients
% of
Patients
# of
Events
Events/
100
DeviceYears
Serious adverse events not
related to the system
85
44.7
176
10.8
Worsening CHF/CHF
decompensation
24
12.6
42
2.1
Atrial fibrillation/flutter
14
7.4
14
0.7
Not Serious events not related
to the system
58
30.5
121
7.4
Pain (in back, arms, chest,
shoulder, groin, head, other)
10
5.3
13
0.7
Worsening CHF/CHF
decompensation
13
6.8
16
0.8
Atrial fibrillation/flutter
3.7
0.4
Ventricular tachycardia
3.7
0.4
US-ENGLISH – 17
5.2. POTENTIAL ADVERSE EVENTS
Adverse events (in alphabetical order), including those reported in the
previous tables, associated with ICD systems include:
─ Acceleration of arrhythmias (caused by device),
─ Air embolism,
─ Bleeding,
─ Chronic nerve damage,
─ Erosion,
─ Excessive fibrotic tissue growth,
─ Extrusion,
─ Fluid accumulation,
─ Formation of hematomas or cysts,
─ Inappropriate shocks,
─ Infection,
─ Keloid formation,
─ Lead abrasion or fracture,
─ Lead migration/dislodgment,
─ Myocardial damage,
─ Pneumothorax,
─ Shunting current or insulating myocardium during defibrillation with
internal or external paddles,
─ Potential mortality due to inability to defibrillate or pace,
─ Thromboemboli,
─ Venous occlusion,
─ Venous or cardiac perforation.
18 – US-ENGLISH
Patients susceptible to frequent shocks despite antiarrhythmic medical
management may develop psychological intolerance to an ICD system
that may include the following:
─ Dependency,
─ Depression,
─ Fear of premature battery depletion,
─ Fear of shocking while conscious,
─ Fear that shocking capability may be lost,
─ Imagined shocking (phantom shock).
US-ENGLISH – 19
6. CLINICAL STUDIES
Clinical data presented in this section are from the MSP clinical study.
PARADYM RF CRT-D is similar in design and function to the ALTO 2
MSP and OVATIO CRT-D devices. The data provided are applicable to
PARADYM RF CRT-D.
6.1. MSP CLINICAL STUDY
OVATIO CRT-D and earlier models were evaluated clinically in an
international, multi-center, randomized clinical trial of Sorin CRM’s
cardiac resynchronization therapy (CRT-D) system. Investigators
attempted to implant study devices in 190 patients.
A total of 182 patients received study devices and had an exposure of
over 165 device years. Of those patients, 19 received OVATIO CRT-D,
160 received ALTO 2 MSP, and 3 received ALTO MSP.
Objectives: The primary objectives of the study were to demonstrate:
─ Greater improvement in a composite endpoint (percent improvement in
peak VO2 percent improvement in quality of life) for CRT-D patients
than for control patients.
─ System complication-free rate ≥ 67 % at six months.
Methods: Patients were New York Heart Association class III or IV and
had one or more indications for an implantable cardioverter defibrillator
(ICD). Patients performed cardiopulmonary exercise testing at baseline
and six-months after randomization. Patients were implanted with a Sorin
CRM ICD with CRT-D, a Situs UW28D left ventricular lead, and
commercially available right atrial and ventricular leads. Routine followups were at pre-discharge, randomization (3-14 days post-implant), one
month, three months, and six months post randomization.
20 – US-ENGLISH
 Results
IMPROVEMENT IN COMPOSITE ENDPOINT
Patients were included in the analysis if complete (peak VO 2 and
quality of life) baseline and six-month data were available.
Number of
patients
contributing to
analysis
Mean percent
improvement
in composite
endpoint for
control group
Mean percent
improvement
in composite
endpoint for
CRT-D group
Percent
greater
improvement
for CRT-D
group
p-value
132
15.5 %
24.9 %
9.4 %
0.046
SIX-MONTH SYSTEM COMPLICATION-FREE RATE
Number of
patients
contributing to
analysis
Kaplan-Meier sixmonth
complication-free
estimate
One-sided lower 95% confidence
bound for six-month complicationfree estimate
190
89.5 %
84.1 %
US-ENGLISH – 21
 Absolute Differences in Peak VO2 and QOL
The tables below show the absolute differences between the control
and test groups’ peak VO2 and QOL over the 6 month follow-up period
in the clinical trial.
Change in Peak VO2
(mL/min/Kg)
Absolute difference between test and control groups’ change in peak
V02 over 6 months
Contr
ol
group
(n=41)
Baseline
Mean ± SD
(range)
6-month
Mean ± SD
(range)
Difference
within group
Difference
between
groups
13.39 ±
4.58
(5.02,
24.10)
13.12 ±
3.99
(3.30,
20.70)
- 0.28
1.85
13.41 ±
4.28
(6.18,
27.67)
1.57
Test
11.84 ±
group 3.90
(n=91) (3.50, 26.3)
Change in QOL
Absolute difference between test and control groups’ change in QOL
score over 6 months
Baseline
Mean ± SD
(range)
6-month
Mean ± SD
(range)
Difference
within group
Difference
between
groups
Control
group
(n=41)
47.5 ±
19.29
(9, 90.3)
31.21 ±
23.96
(0, 95)
16.29
1.28
Test group
(n=91)
52.81 ±
21.84
(9, 92)
35.24 ±
23.73
(0, 93)
17.57
22 – US-ENGLISH
The table below presents the percentage of patients in each group who
improved, worsened, or remained unchanged in each element of the
composite score and the composite score itself.
QOL score
VO2 Score
Composite
Score
Contro Test
GROUP
GROU
Control Test
Control
GROU GROU GROU
Test
GROU
% Improved
75.6
74.7
48.8
67.0
62.2
70.9
% Worsened
24.4
25/03/09
51.2
31.9
37.8
28.6
% Unchanged
0.0
0.0
0.0
1.1
0.0
0.0
Histograms for Respiratory exchange rate (RER) at peak VO2 at
baseline and 6 month follow-up are provided below:
RER at peak VO2 at baseline
Test
Control
60
Percent of patients
50
40
30
20
10
≤ 0.79
0.80-0.89
0.90-0.99
1.0-1.09
1.10-1.19
1.20-1.29
≥ 1.30
RER
US-ENGLISH – 23
RER at peak VO2 at six months
Test
Control
60
Percent of patients
50
40
30
20
10
≤ 0.79
0.80-0.89
0.90-0.99
1.0-1.09
1.10-1.19
1.20-1.29
≥ 1.30
RER
 Clinical Results V-V timing
V-V programmable settings were available for the clinical study devices as
follows: ALTO MSP model 617 (not programmable for V-V delay), ALTO 2
MSP model 627 values (0, 31, 39, 47, 55 and 63 ms) and OVATIO CRT-D
6750 values (0 to 63 ms in steps of 8 ms).
The graph below shows the programmed V-V settings at randomization
by percentage of patients programmed to each combination of
Synchronous BiV pacing and V-V delay.
Programmed V-V Delays
Programmed settings for the 149 of 154 patients optimized
35
30
% Patients
25
Synchronous BiV
20
Synchronous BiV (Left then Right)
15
Synchronous BiV (Right then Left)
10
16
24
31
37
39
V-V Delay (msec)
24 – US-ENGLISH
55
60
63
The optimization protocol in the clinical study specified that each
patient randomized should undergo echo guided V-V optimization. Per
the investigational plan for the MSP Clinical Trial, a uniform protocol
was used for V-V programming. This protocol required all patients to
undergo echo-guided V-V delay optimization before randomization (2 to
14 days post-implant). The optimal V-V delay was determined by
finding the programmable V-V delay and ventricular chamber pacing
order (RV then LV, or LV then RV) providing the maximum time velocity
integral (TVI or VTI) across the left ventricular outflow tract (LVOT).
Only those patients randomized to the Test arm were required to be
programmed per the optimization protocol for the V-V delay.
Of the 177 patients that presented at randomization, 3 had Model 617
which does not have V-V programmability hence the inability to
optimize. Of the remaining 174 patients, 154 (89%) were tested per the
V-V optimization protocol. One hundred forty-nine (149) of the 154
patients who were tested per the V-V optimization protocol were
programmed per the recommended or randomized V-V delay (97%).
Thirty-one (31) patients were programmed to BiV synchronous (V-V
delay 0ms), 46 were programmed to Sequential BiV (LV then RV), 22
were programmed to Sequential (RV then LV), and the remaining 50
patients were randomized to RV only.
A sub-analysis of the composite endpoint comparing the subset of
CRT-D patients with optimized V-V delays vs. the subset of patients
that did not undergo V-V delay optimization demonstrated similar
results in both groups. The CRT-D patients who did not undergo V-V
delay optimization showed a smaller improvement in the composite
endpoint, although the sample size did not permit conclusions based
on data from this subset.
US-ENGLISH – 25
7. PATIENT SELECTION AND TREATMENT
7.1. INDIVIDUALIZATION OF TREATMENT
Exercise stress testing. If the patient’s condition permits, use
exercise stress testing to:
─ Determine the maximum rate of the patient’s normal rhythm,
─ Identify any supraventricular tachyarrhythmias,
─ Identify exercise-induced tachyarrhythmias.
The maximum exercise rate or the presence of supraventricular
tachyarrhythmias may influence selection of programmable
parameters. Holter monitoring or other extended ECG monitoring also
may be helpful.
CAUTION: When a parameter is reprogrammed during an exercise
stress test, PARAD/PARAD+ algorithm forces acceleration to
"ventricular". During conducted sinus tachycardia within the
programmed Tachy zone, the device detects a 1:1 fast rhythm.
Assuming that acceleration was set to ventricular by reprogramming,
the device concludes for a VT, and immediately applies the
corresponding therapy. This event could have been avoided with
appropriate device handling during tests.
Electrophysiologic (EP) testing: EP testing may be useful for ICD
candidates. EP testing may identify the classifications and rates of all
the ventricular and atrial arrhythmias, whether spontaneous or during
EP testing.
Drug resistant supraventricular tachyarrhythmias (SVTs): Drug
resistant supraventricular tachyarrhythmias (SVTs) may initiate
frequent unwanted device therapy. A careful choice of programming
options is necessary for such patients.
26 – US-ENGLISH
Antiarrhythmic drug therapy: If the patient is being treated with
antiarrhythmic or cardiac drugs, the patient should be on a
maintenance drug dose rather than a loading dose at the time of ICD
implantation. If changes to drug therapy are made, repeated arrhythmia
inductions are recommended to verify ICD detection and conversion.
The ICD also may need to be reprogrammed.
Changes in a patient’s antiarrhythmic drug or any other medication that
affects the patient’s normal cardiac rate or conduction can affect the
rate of tachyarrhythmias and/or efficacy of therapy.
Direct any questions regarding the individualization of patient therapy
to Sorin CRM’s representative.
7.2. SPECIFIC PATIENT POPULATIONS
Pregnancy: If there is a need to image the device, care should be
taken to minimize radiation exposure to the foetus and the mother.
Nursing Mothers: Although appropriate biocompatibility testing has
been conducted for this implant device, there has been no quantitative
assessment of the presence of leachables in breast milk.
Pediatric Patients: This device has not been studied in patients
younger than 18 years of age.
Geriatric Patients: Most of the patients receiving this device in clinical
studies were over the age of 60 years.
Handicapped and Disabled Patients: Special care is needed in using
this device for patients using an electrical wheel chair or other electrical
(external or implanted) devices.
US-ENGLISH – 27
8. PATIENT COUNSELLING INFORMATION
The physician should consider the following points in counselling the
patient about this device:
─ Persons administering CPR may experience tingling on the patient’s
body surface when the patient’s ICD system delivers a shock.
─ Advise patients to carry Sorin CRM ID cards and/or ID bracelets
documenting their ICD system.
9. CONFORMANCE TO STANDARDS
This device was developed in conformance with all or parts of the
following standards:
─ EN 45502-1: 1998 – Active implantable medical devices.
General requirements for safety, marking and information to be
provided by the manufacturer.
─ EN 45502-2-1: 2003 - Active implantable medical devices.
Part 2-1: Particular requirements for active implantable medical
devices intended to treat bradyarrhythmia (cardiac pacemakers).
─ EN 45502-2-2: 2008 – Active implantable medical devices.
Part 2-2: Particular requirements for active implantable medical devices
intended to treat tachyarrhythmia (includes implantable defibrillators).
─ ISO 5841-3: 2000 Low profile connectors (IS1) for implantable
pacemakers.
─ ISO 11318 (DF-1): Cardiac defibrillator: connector assembly for
implantable defibrillators - Dimensional and test requirements, August
2002.
─ ANSI/AAMI PC69:2007 Active implantable Medical Devices Electromagnetic compatibility - EMC test protocols for implantable
cardiac pacemakers and implantable Cardioverter Defibrillators.
─ IEC 60601-1-2 (2007): Electromagnetic compatibility - Medical
electrical equipment. General requirements for basic safety and
essential performance - Collateral standard
28 – US-ENGLISH
─ EN 50371 (2002) : Generic standard to demonstrate the compliance of
low power electronic and electrical apparatus with the basic restrictions
related to human exposure to electromagnetic fields (10 MHz - 300
GHz)
─ EN 301 489-1 (v1.8.1) & EN 301 489-27 (v1.1.1): Electromagnetic
compatibility and Radio spectrum Matters (ERM); Electromagnetic
Compatibility (EMC) standard for radio equipment and services - Part 1
: Technical Requirements and Part 27: Specific conditions for Ultra Low
Power Active Medical Implants (ULP-AMI) and related peripheral
devices (ULP-AMI-P)
─ EN 301839-1 (v1.3.1) & EN 301839-2 (v1.2.1): Electromagnetic
compatibility and Radio spectrum Matters (ERM); Short Range Devices
(SRD); Ultra Low Power Active Medical Implants (ULP-AMI) and
Peripherals (ULP-AMI-P) operating in the frequency range 402 MHz to
405 MHz; Part 1: Technical characteristics and test methods and Part
2: Harmonized EN covering essential requirements of Article 3.2 of the
R&TTE Directive
─ EN 62311 (2008) : Assessment of electronic and electrical equipment
related to human exposure restrictions for electromagnetic fields (0Hz
to 300 GHz)
─ EN 62209-2 (2010) : Human exposure to radio frequency fields from
hand-held and body-mounted wireless communication devices –
Human models, instrumentation and procedures – Part 2: Procedure to
determine the specific absorption rate (SAR) for wireless
communication devices used in close proximity to the human body
(frequency range of 30MHz to 6 GHz)
This information should not be used as a basis of comparisons among
devices since different parts of the standards mentioned may have
been used.
Sorin CRM declares that this device is in conformity with the essential
requirements of Directive 1999/5/EC on Radio and Telecommunications
Terminal Equipment, with the mutual recognition of their conformity
(R&TTE).
US-ENGLISH – 29
Federal Communication Commission Interference Statement 47 CFR
Section 15.19 and 15.105(b)
- The FCC product ID is YSGCRTD9750.
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference
in a residential installation. This equipment generates uses and can radiate
radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation.
This device complies with Part 15 of the FCC Rules. Operation is subject to
the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
FCC Interference Statement 47 CFR Section 15.21 - No Unauthorized
Modifications
CAUTION: This equipment may not be modified, altered, or changed in any
way without signed written permission from SORIN. Unauthorized
modification may void the equipment authorization from the FCC and will
void the SORIN warranty.
Identification of the equipment according Section 95.1217(a)
This transmitter is authorized by rule under the Medical Device
Radiocommunication Service (in part 95 of the FCC Rules) and must not
cause harmful interference to stations operating in the 400.150-406.000
MHz band in the Meteorological Aids (i.e., transmitters and receivers used
to communicate weather data), the Meteorological Satellite, or the Earth
Exploration Satellite Services and must accept interference that may be
caused by such stations, including interference that may cause undesired
operation. This transmitter shall be used only in accordance with the FCC
Rules governing the Medical Device Radiocommunication Service. Analog
30 – US-ENGLISH
and digital voice communications are prohibited. Although this transmitter
has been approved by the Federal Communications Commission, there is
no guarantee that it will not receive interference or that any particular
transmission from this transmitter will be free from interference.
IC Requirements for canada
- The IC product ID is 10270A-CRTD9750
This class B digital apparatus meets all requirements of the Canadian
Interference- causing equipment regulations.
This device complies with Industry Canada licence-exempt RSS
standard(s). Operation is subject to the following two conditions: (1) this
device may not cause interference, and (2) this device must accept any
interference, including interference that may cause undesired operation of
the device.
Under Industry Canada regulations, this radio transmitter may only operate
using an antenna of a type and maximum (or lesser) gain approved for the
transmitter by Industry Canada. To reduce
potential radio interference to other users, the antenna type and its gain
should be so chosen that the equivalent isotropically radiated power (e.i.r.p.)
is not more than that necessary for successful communication.
This device may not interfere with stations operating in the 400.150–
406.000 MHz band in the Meteorological Aids, Meteorological Satellite, and
Earth Exploration Satellite Services and must accept any interference
received, including interference that may cause undesired operation.
US-ENGLISH – 31
10. PHYSICIAN GUIDELINES
10.1. PHYSICIAN TRAINING
Physicians should be familiar with sterile pulse generator and left
ventricular pacing lead implant procedures. They must apply these
procedures according to professional medical training and experience.
Physicians should be familiar with follow-up evaluation and
management of patients with an implantable defibrillator (or referral to
such a physician).
This training guideline for implantation and follow-up of ICD and CRT-D
devices comes from the Heart Rhythm Society to provide standards for
hospital credentialing bodies to help ensure appropriate patient care
and lead to improved patient outcomes. The following is a summary of
requirements for an alternate training pathway for ICD and CRT-D
implantations(1):
─ Documentation of current experience: 35 pacemaker implantations
per year and 100 implantations over the prior 3 years
─ Proctored ICD implantation experience: 10 Implantations, 5 Revisions
─ Proctored CRT-D implantation experience: 5 implantations
─ Completion of didactic course and/or IBHRE® ExAM
─ Monitoring of patient outcomes and complication rates
─ Established patient follow-up
─ Maintenance of competence: 10 ICD and CRT-D procedures per
year, 20 patients per year in follow-up
Please consult full text of both publications for details. 2004 Heart Rhythm
Society Clinical Competency Statement and the 2005 Addendum on Training
Pathways for Implantation of Cardioverter Defibrillators and Cardiac
Resynchronization Devices. Heart Rhythm (2004) 3, 371-375; Heart Rhythm
(2005) 2(10), 1161-1163.
32 – US-ENGLISH
(1)
10.2. DIRECTIONS FOR USE
ICD operating characteristics should be verified at the time of
implantation and recorded in the patient file. Complete the Patient
Registration Form and return it to Sorin CRM, as it provides necessary
information for warranty purposes and patient tracking.
Additional programming instructions can be found by accessing Online
Help (click the “?” on the screen) on the Sorin CRM dedicated
programmer. Paper copies of Online Help can be obtained by
contacting your Sorin CRM representative.
10.3. MAINTAINING DEVICE QUALITY
This device is FOR SINGLE USE ONLY. Do not resterilize and
reimplant explanted ICDs.
Do not implant the device when:
─ It has been dropped on a hard surface because this could have
damaged pulse generator components.
─ Its sterility indicator within the inner package is not green, because
it might not have been sterilized.
─ Its storage package has been pierced or altered, because this
could have rendered it non-sterile.
─ It has been stored or transported outside the environmental
temperature limits: 32 °F (0 °C) to 122 °F (50 °C) as an electrical
reset condition may occur.
─ "Use Before" date has expired, because this can adversely affect
pulse generator longevity or sterility.
10.4. V-V PROGRAMMING RECOMMENDATION
It is recommended that V-V optimization testing be performed and used
to set the V-V delay for this device to optimize the potential for RF
CRT-D benefit to the patient.
US-ENGLISH – 33
11. PATIENT INFORMATION
Information for the patient is available in the patient booklet, contained
in the outer storage package. Additional copies can be obtained by
contacting your Sorin CRM representative or on the Sorin CRM's web
site: http://www.sorin.com. This information should be given to each
patient with their first ICD and offered to the patient on each return visit
or as deemed appropriate.
12. HOW SUPPLIED
12.1. STERILITY
The PARADYM RF defibrillators are supplied one per package in a sterile
package.
12.2. WARRANTY AND REPLACEMENT POLICY
Sorin CRM warrants its defibrillators. Refer to the section “Warranty” for
additional information. Please see the following labelling sections for
information concerning the performance of this device: Indications,
Contraindications, Warnings and Precautions, and Adverse Events.
34 – US-ENGLISH
13. DEVICE DESCRIPTION
The PARADYM RF CRT-D system includes the model 9750 ICD device
and programming system. The programming system includes the Sorin
CRM Orchestra Plus programmer with the SMARTVIEW programming
software connected to a CPR3 programming head. The programming
system is configured and furnished by Sorin CRM.
The PARADYM RF CRT-D 9750 can serves as a defibrillation
electrode (active housing) with a total surface area of 76 cm².
The PARADYM RF CRT-D 9750 is designed to recognize and treat slow
or fast VT and VF by continuously monitoring atrial and ventricular activity
to identify persistent ventricular arrhythmias and to deliver appropriate
therapies. PARADYM RF CRT-D 9750 features the PARAD/PARAD+
algorithm, which is specifically designed to differentiate ventricular
tachycardias from fast rhythms of supraventricular origin.
PARAD/PARAD+ continuously monitors R-R interval stability, searches
for long cycles, assesses the degree of P-R association, evaluates
sudden onset and determines the chamber of arrhythmia acceleration.
In addition to the advanced detection scheme, PARADYM RF CRT-D
9750 offers programmable single, dual or triple-chamber pacing
therapy (DDD, DDI, VVI or SafeR modes) with or without rateresponsive capabilities (DDDR, DDIR, VVIR, DDD/DDIR and SafeRR modes) using an acceleration sensor. An automatic AV delay
algorithm as well as a mode switching function are available.
PARADYM RF CRT-D 9750 enables an adjustment of the interventricular
delay, and provides the possibility of adapting pacing to each ventricle.
The ICD is intended to resynchronize uncoordinated contraction of the
heart by simultaneously or sequentially pacing both ventricles.
US-ENGLISH – 35
PARADYM RF CRT-D 9750 offers tiered therapy. Therapies can be
programmed independently in each zone:
─ in the Slow VT and VT zones: two ATP programs, up to two shocks
with programmable energy and up to four shocks with maximum
energy can be programmed;
─ in the VF zone: one ATP program, up to two shocks with
programmable energy and up to four shocks with maximum energy
can be programmed.
The ATP can be applied in RV, LV or RV and LV pacing with a VV
delay equal to 0 ms. ATP pacing configuration is independent of
ventricular pacing configuration.
When the rhythm changes from one zone to another, the device
delivers the therapy programmed in this zone, starting with the same or
more aggressive program for the area. The ATP program in the VF
zone will only be applied if the VT coupling interval is longer than the
programmed fast VT cycle length.
The PARADYM RF CRT-D 9750 offers biphasic shocks with a
maximum stored energy of 42 J; all automatic shocks are delivered in a
non-committed way. The shock configuration (electrodes used to apply
the shock) can be chosen by programming one of the following
combinations: can and one coil, can and 2 coils, 2 coils only.
Other features are as follows:
─ Automatic ventricular sensitivity control
─ Electrophysiological studies (EPS) with real-time markers or
electrograms:
─ Programmer-controlled VT induction sequences,
─ Programmer-controlled VF inductions (30 Hz rapid pacing or
shock on T),
─ Programmable electrogram vectors (EGM A, EGM V, RVcoilCAN, SVC-CAN, RVcoil-SVC, LV bip, LV tip-RV),
─ Real-time annotations displayed with the markers and
indicating the majority rhythm,
36 – US-ENGLISH
─ Manual ATP sequences,
─ Manual shocks.
─ Rescue shock
─ Follow-up tests:
─ Pacing lead impedance,
─ Coil impedance,
─ Capacitor charge time,
─ Sensitivity test,
─ Pacing threshold tests.
─ Data storage:
─ Therapy History Report,
─ Statistics (pace/sense, therapy, shocks, and battery voltage),
─ Up to 14 complete Holter records with event logs, marker
channel notation, and electrogram records.
The connector head has five ports: atrial bipolar pace/sense, right
ventricular bipolar pace/sense, left ventricular bipolar pace and two
ports for RV and SVC defibrillation coils. Both pace/sense ports and the
pace port are compatible with the IS-1 standard and both defibrillation
ports are compatible with the DF-1 standard. Distal lead terminal
connections are secured with set-screws accessed via self-sealing
silicone plugs. All lead connections pass through the header into the
device via feedthroughs.
Programming System: The Sorin CRM programmer is used in
conjunction with specific programmer software to interrogate and
program the implanted device at implant and during patient follow-up
procedures.
Remote Monitoring: PARADYM RF CRT-D 9750 is also equipped with
the RF wireless technology which enables to remotely monitor the
patients who have the Sorin CRM SMARTVIEW Monitor installed at
home.
US-ENGLISH – 37
14. IMPLANT PROCEDURE
14.1. NECESSARY EQUIPMENT
Implantation of PARADYM RF CRT-D 9750 requires the following
equipment:
─ Sorin CRM dedicated programmer, equipped with the
SMARTVIEW software interface and with the programming head,
─ pacing system analyzer, as well as its sterile connecting cables, to
evaluate the pacing and sensing thresholds,
─ a complete set of leads with corresponding introducers,
─ physiological signal monitor capable of displaying simultaneously
the surface ECG and arterial pressure,
─ an external defibrillator with sterile external paddles,
─ sterile cover for the telemetry head.
14.2. PACKAGING
 Contents
The PARADYM RF CRT-D 9750 and its accessories are ethylene oxide
sterilized and hermetically sealed in two-ply clear packaging meeting
international requirements.
The sterile packaging contains a defibrillator, one screwdriver, and an
insulating plug for the DF-1 defibrillation connector.
The non-sterile items contained in the outer storage package are the
implant manual, the ICD Registration Form and its envelope, the
patient booklet, the ICD ID card and 12 identification labels.
Once delivered, PARADYM RF CRT-D 9750 is programmed to asshipped values that are different from nominal values (see Chapter
“Programmable Parameters” for details).
38 – US-ENGLISH
14.3. OPTIONAL EQUIPMENT
The following equipment may be required during implantation
of PARADYM RF CRT-D 9750:
─ an IS-1 insulating plug to close the atrial port
─ sterile water to clean traces of blood. Any parts cleaned with sterile
water must be thoroughly dried.
─ mineral oil to lubricate if necessary
─ a lead cap to isolate a lead which is not used
14.4. BEFORE OPENING THE PACKAGE
Before opening the package, check the "Use Before" date printed on the
labels on the box and on the sterile package. Defibrillators that have not
been implanted before that date should be returned to Sorin CRM.
Devices MUST NOT be interrogated and programmed within the vicinity of
other devices.
Also check the integrity of the sterile package. The sterility of the
contents is no longer guaranteed if the package has been pierced or
altered. If the defibrillator is no longer sterile, it should be returned in its
packaging to Sorin CRM. Any re-sterilization of the unit is at the
discretion of Sorin CRM.
14.5. PRIOR TO IMPLANTATION
Use the programmer to verify the defibrillator can be interrogated
before implantation.
Verify all shock therapies are disabled in order to avoid accidental
discharge during implantation.
It is not advisable to program the Smoothing function before
implantation, since the defibrillator may detect noise and pace at a rate
higher than the programmed basic rate.
US-ENGLISH – 39
CAUTION: Do not shake or tap sharply on the ICD package with the
ICD inside, because the ICD's sensing circuits can interpret this as Pwaves or R-waves and record these as an arrhythmia episode. If
unusual shaking or tapping of the package results in a stored
arrhythmia episode, erase the recording before using the ICD.
14.6. DEVICE PLACEMENT
The pocket should be prepared in the left pectoral position, either
subcutaneously or submuscularly. Subcutaneous device implantation is
recommended for optimal RF communication efficacy.
Implantation in an abdominal position is not advisable.
In its final position, the defibrillator should be no more than 4 cm below
the skin surface.
14.7. CHOOSING THE TYPE OF LEAD
The defibrillator should be connected to:
─ one bipolar atrial sensing/pacing lead
─ one right ventricular lead with bipolar sensing/pacing electrodes,
and one or two defibrillation coils
─ one unipolar or bipolar left ventricular pacing lead.
The choice of leads and their configuration is left to the implanting
physician’s judgment.
Note: In case no atrial lead is implanted, the atrial port should be
plugged with IS-1 insulating plug and a single chamber mode (VVIVVIR) should be programmed. PARAD and PARAD+ should not be
used.
Connectors: The unipolar and bipolar pacing/sensing connectors are
compatible with the IS-1 standard and the defibrillation connectors are
compatible with the DF-1 standard.
40 – US-ENGLISH
SHOCK CONFIGURATION (+ -> -)
The shock configuration is the energy pathway between the
defibrillation electrodes. If an atrial coil is present, the shock
configuration can be programmed for bi-directional shocks.
Programming: When active case and SVC are both programmed to
Yes, the shock configuration can be programmed to:
─ RV to Case (or Case to RV),
─ or RV to SVC (or SVC to RV),
─ or RV to Case+SVC (or Case+SVC to RV).
The polarity of shock is determined by the parameter itself.
RV to Case+SVC
RV to Case
RV to SVC
14.8. MEASUREMENT OF THRESHOLDS AT IMPLANT
Pacing and sensing thresholds should be measured at implant.
Pacing thresholds: Acute thresholds should be lower than 1 V (or
2 mA) for a 0.35 ms pulse width, in both ventricles and in the atrium.
Sensing thresholds: For proper right ventricular sensing, the
amplitude of the R-wave should be greater than 5 mV. For proper atrial
sensing, the amplitude of the P-wave should be greater than 2 mV.
US-ENGLISH – 41
Pacing impedance measurements: Right ventricular, left ventricular
and atrial pacing impedances should range from 200 to 3000 ohms
(refer to the lead characteristics, especially if high impedance leads are
used).
14.9. LEAD CONNECTION
Implant the ventricular leads, then the atrial lead.
Each lead must be connected to the corresponding connector port. The
position of each connector is indicated on the casing.
CAUTION: Tighten only the distal inserts.
To connect each lead, proceed as follows:
1. Clean the lead terminal pins thoroughly, if necessary (device
replacement).
2. Lubricate the lead terminal pins with sterile water, if necessary.
3. Do not insert a lead connector pin into the connector block without
first visually verifying that the lead port is not filled with any
obstacle.
4. Insert the screwdriver into the pre-inserted screw socket of the
appropriate port (in order to allow excess air to bleed out and to
make the insertion of the lead pin easier).
5. Insert the lead pin all the way into the port (check that the pin
protrudes beyond the distal insert).
6. Tighten, check the tightness and ensure the lead pin still protrudes
beyond the distal insert, and did not move.
42 – US-ENGLISH
Caution: 1. One single set screw is located on the side of the
connection header. 2. Do not tighten the pre-inserted screws when
there is no lead (this could damage the connector). 3. Do not loosen
the screws before inserting the connector (subsequent risk of being
unable to reinsert the screw). 4. Removing the screwdriver: to avoid all
risk of loosening screws during removal, hold the screwdriver by its
metal part and not by the handle. 5. When mineral oil or sterile water is
used to make lead insertion easier, the screwdriver should remain
inserted into the pre-inserted screw socket when checking the
tightness. As a matter of fact, when the lead port is filled with a liquid,
the physics piston effect can give the feeling the lead is properly
tightened.
NOTE: To optimise cardioversion/defibrillation shocks, electrodes must
be positioned so that the electric field between anode (s) and cathode
covers the largest myocardial mass. In normal conditions, the anode
and cathode are adequately separated. In case of a short-circuit, the
shock may be aborted to prevent damaging the defibrillator.
In the case of an external defibrillation shock delivered to the patient,
always check the programming and functioning of the device, in
particular its capacity to deliver shocks.
14.10. DEVICE IMPLANTATION
PARADYM RF CRT-D 9750 should be implanted with the engraved
side facing outwards for optimal communication with the programming
head and radiographic identification.
Carefully wind excess lead and place in a separate pocket to the side
of the defibrillator. It is recommended to not place any excess wire
between the can and the heart.
Suture the casing connector to the muscle using the hole provided for
this purpose, in order to avoid potential migration of the device into the
pectoral muscle.
US-ENGLISH – 43
14.11. TESTS AND PROGRAMMING
During the implant testing procedure, it is recommended that a security
margin of at least 10 J be demonstrated between the effective shock
energy and maximum programmable energy.
Enable shock therapies, then program the defibrillator.
Verify that the defibrillation lead impedance for each shock delivered
ranges from 30 to 150 ohms. Check the lead connection if the values
are outside these boundaries.
Save the programming data on the programmer’s hard disk and on an
external storage device (if desired).
15. SPECIAL MODES
15.1. SAFETY MODE (NOMINAL VALUES)
Nominal values may be rapidly restored by pressing the following
button on the programming head or programmer keyboard:
or via the “Emergency” button on the SMARTVIEW screen.
In safety mode, the defibrillator operates with the parameters
underlined in the table of programmable parameters.
44 – US-ENGLISH
15.2. MAGNET MODE
When the magnet is applied:
─ antiarrhythmia functions are inhibited (detection of rhythm
disturbances, charging, and therapy),
─ hysteresis, VV delay and AVD paced/sensed offset are set to 0,
─ pacing amplitude is set to 6 V,
─ pulse width is set to maximum,
─ pacing rate is set to the magnet rate,
─ the following functions are disabled: ventricular arrhythmia
prevention, Mode Switch, Anti-PMT, Smoothing, Rate Response.
When the magnet is removed:
─ the sensor rate is forced to the basic rate,
─ arrhythmia detection algorithms and sequential therapies are
reinitialized,
─ therapies start with the least aggressive program for each area.
The other parameters remain at their programmed value, including the
ventricular paced chamber parameter.
NOTE: The magnet is inactive during telemetry.
The magnet rate values are as follow:
Magnet rate (bpm)
96
94
91
89
87
85
Magnet period (ms)
625
641
656
672
688
703
Magnet rate (bpm)
83
82
80
78
77
Magnet period (ms)
719
734
750
766
781
US-ENGLISH – 45
15.3. RESPONSE IN THE PRESENCE OF INTERFERENCE
If the defibrillator senses electrical noise at a frequency above 16 Hz, it
switches to an asynchronous mode at the basic rate. The programmed
mode is restored as soon as the noise is no longer detected.
Ventricular pacing is also inhibited by ventricular noise. It can be
restored by setting the parameter V pacing on noise to Yes.
15.4. DETECTION CHARACTERISTICS IN THE
PRESENCE OF ELECTROMAGNETIC FIELDS
Per Clause 27.4 of Standard EN 45502-2-2, detection in the presence
of electromagnetic fields is characterized as follows:
Differential mode:
Common mode rejection ratio:
16.6 Hz
50 Hz
60 Hz
Atrial channel
≥ 75 dB
67 dB
67 dB
Ventricular channel
≥ 69 dB
≥ 69 dB
≥ 69 dB
46 – US-ENGLISH
Modulated interference: For atrial sensitivity setting of 0.2 mV,
compliance to the Cenelec standard 45502-2-2 is met for a maximum
test signal amplitude of 8 V for the frequency of 60 MHz. 0.4 mV
complies with the standard for the whole frequency range.
15.5. PROTECTION AGAINST SHORT-CIRCUITS
The defibrillator can undergo a short-circuit if the anode and cathode
are not adequately separated.
In this case, the shock is aborted to prevent damaging the defibrillator
and a warning will indicate that a short circuit (shock impedance
< 20 ohms) was detected during the last shock.
16. MAIN FUNCTIONS
16.1. AUTOMATIC LEAD MEASUREMENTS
Automatic pacing lead impedance measurement: A lead impedance
measurement is automatically performed on atrial and ventricular leads
every 6 hours. The daily mean impedance is stored for each chamber.
Automatic coil impedance measurement: A coil impedance
measurement is automatically performed on RV and SVC coils once a
week. The result is stored in the device memory.
16.2. ATRIAL TACHYARRHYTHMIA MANAGEMENT
Mode Switch: This function is designed to limit the acceleration and
variation of ventricular rate in the presence of atrial arrhythmia.
US-ENGLISH – 47
16.3. VENTRICULAR TACHYARRHYTHMIA
MANAGEMENT
Ventricular tachyarrhythmia prevention: Set of algorithms that can be
used to avoid the circumstances of ventricular tachyarrhythmia onset.
Searching for a long cycle (P And R based Arrhythmia Detection+:
PARAD+): Additional arrhythmia classification criterion to improve
identification of atrial fibrillation and avoid inappropriate shocks.
Fast VT treatment: Applies detection criteria on fast ventricular
tachycardia, that are different from those of the VT zone, as well as
different therapies. The fast VT zone is included in the VF zone: its
lower limit is determined by the programmed value for the VF zone and
its upper limit by the programmed value for the fast VT zone.
Polarity alternation on Max shock: Reverses the programmed polarity of
every second shock set at maximum energy. The number, type, and energy
of shocks is independently programmable by detection zone.
16.4. PACING
BTO (Brady Tachy Overlap): Enables cardiac resynchronization
therapy within the slow VT zone to preserve patient exercise capacity,
without affecting detection or treatments of slow VTs.
Post-shock mode: After any automatic shock therapy, the post-shock
mode makes it possible to apply a pacing mode other than the
standard antibradycardia pacing mode and/or with different pacing
parameters.
SafeR (AAI <> DDD) mode: Is intended to minimize deleterious
ventricular pacing. The defibrillator functions in AAI mode, and
temporarily switches to DDD mode upon the occurrence of AVB III,
AVB II, AVB I and ventricular pause.
48 – US-ENGLISH
Anti-PMT protection: Is intended to protect the patient from
Pacemaker-Mediated Tachycardia (PMT) without reducing atrial
sensing capability of the device.
16.5. SENSING
Automatic Refractory Periods: Optimize sensing and make the
implant progamming easier. These periods are composed of a minimal
Refractory Period and a triggerable Refractory Period. The duration of
the refractory periods lengthens automatically as needed.
Committed period: In DDI or DDD modes, the committed period is a
non-programmable 95 ms ventricular relative refractory period that
starts with atrial pacing. If a ventricular event is sensed during the
committed period, but outside the blanking period, the ventricle is
paced at the end of the committed period. The committed period
prevents inappropriate ventricular inhibition if crosstalk occurs.
Protection against noise: Allows the distinction between ventricular
noise and ventricular fibrillation. If the device senses ventricular noise, the
ventricular sensitivity is decreased until noise is no longer detected.
Ventricular pacing can be inhibited to avoid a potential paced T-wave.
Automatic sensitivity control: Optimizes arrhythmia detection and
avoids late detection of T-waves and over-detection of wide QRS
waves. The device automatically adjusts the sensitivities based on the
ventricular sensing amplitude. In case of arrhythmia suspicion or after a
paced event, the programmed ventricular sensitivity will be applied.
The minimum ventricular sensitivity threshold is 0.4 mV (minimum
programmable value).
US-ENGLISH – 49
16.6. FOLLOW-UP FUNCTION
Storage of memory data: AIDA+ (Automatic Interpretation for
Diagnosis Assistance) software provides access up to 6 months of
patient follow-up with day by day data collection, or up to 24 hours with
hourly data collection. Episodes of ventricular tachyarrhythmia are
recorded with the programmable EGM channels: either by selecting up
to two traces, or by selecting "Double V" which enables a one-channel
recording that is twice as long.
Alerts / Warnings: The device routinely performs security self-checks
and technical measurements to ensure system integrity. When system
integrity is found to be at risk outside a follow-up, alerts are stored in
the device memory. When system integrity is found to be at risk during
a follow-up, the information is managed as a warning (pop-up
message) to notify immediately the user. For example, the following
types of event can trigger a warning or an alert: technical problem
during a shock, pacing lead impedance or coil impedance
measurements out-of-range, battery depletion,…
16.7. REMOTE MONITORING FUNCTION
Remote monitoring enables the automatic remote transmission of implant
data to the physician thanks to the wireless Radio Frequency (RF)
communication ability of the implant in order to provide a comprehensive
report to the physician about device functioning and patient cardiac status
without having the patient physically in the clinic.
The data is transmitted from the implant and the SMARTVIEW monitor,
a small transmitter placed in the patient home.
Implant data are first transmitted to the SMARTVIEW monitor via RF.
Data are then rooted through the phone network to an internet website.
This website is responsible for transforming the implant data into a
comprehensive report that can be consulted by the physician.
50 – US-ENGLISH
 SMARTVIEW Monitor
The SMARTVIEW monitor is a small device equipped with an RF
transmission module to communicate with the implant and a modem to
export data through the internet.
The SMARTVIEW monitor is delivered to the patient who has to install
it at home. Preferably the SMARTVIEW monitor will be placed on
the nightstand of the patient, as close as possible to the side of the bed
the patient usually sleeps. The SMARTVIEW monitor shall be
connected to the phone network and the power plug. Regular
transmissions are done during the night when the patient is asleep next
to the SMARTVIEW monitor without any intervention from the patient.
 Transmission trigger
There are 3 different triggers for a remote transmission:
─ the remote follow-up transmission is scheduled by the physician to
occur regularly (according to the programming).
─ the alert transmission will take place when the implant has recorded an
abnormal event. The list of abnormal event is available in a following
paragraph. Alert conditions are checked daily.
─ the on-demand follow-up transmission is triggered by the patient
himself through the use of a specific button on the remote-monitor.
US-ENGLISH – 51
 Data transmitted
The data transmitted are identical to the data available during a
standard interrogation with the Orchestra Plus programmer. All
counters, histograms, IEGMs and diagnosis available in the device are
transmitted containing (not exhaustive list):
─ programmed parameters
─ Information on patient and system implanted
─ battery status
─ lead status (brady leads and defibrillation coils)
─ pacing counters and mean heart rate (brady)
─ atrial and ventricular arrhythmia counters and episodes
─ ventricular therapy counters
─ heart failure monitoring
Data are presented in the form of 2 reports to the physician: the first
one contains a summary of major counters, histograms, warnings and
diagnosis. The second one presents the 3 most important IEGM
episodes automatically selected based on the degree of severity for
the patient.
 User website
On the website, the physician is able to:
─ consult and schedule the remote follow-ups of their patient
─ configure additional ways of being notified of alerts (for instance by
SMS, fax or e-mail
─ consult, print and export patient reports
52 – US-ENGLISH
 Alert system
The following set of alert trigger can be independently programmed
ON/OFF by the physician using the Orchestra Plus programmer and
can trigger an alert transmission:
─ Reset of the device
─ ERI reached
─ Low or high impedance (A, RV, LV)
─ Abnormal coil impedance (shock lead)
─ Low or High shock impedance
─ Long charge time
─ Inefficient high energy shock
─ All shocks programmed OFF
─ Shock treated VT/VF
─ Lack of V pacing in CRT device
─ Suspicion of noise on the V lead
─ AT/AF occurrence
─ Fast V rate during AT/AF
WARNINGS
The use of remote monitoring does not replace regular follow-up.
Therefore, when using remote monitoring, the time period between
follow-ups visits may not be extended.
When ERI mode is reached, this information is transmitted via the
remote monitoring facility and then the remote-monitoring is switched
off to preserve battery life.
US-ENGLISH – 53
17. PATIENT FOLLOW-UP
17.1. FOLLOW-UP RECOMMENDATIONS
Before the patient is discharged and at each subsequent follow-up visit,
it is advisable to:
─ check the occurrence of system warnings,
─ check the battery status,
─ check the integrity of the pacing and defibrillation leads,
─ check for proper sensing (sensitivity, crosstalk) and pacing ;
set the pacing amplitude to twice the pacing threshold,
─ interrogate the implant memories (AIDA+),
─ check the efficacy of the therapies delivered,
─ keep a printout of programmed parameters, test results, and memory
data,
─ reset the memory data and statistics.
These operations should be performed by medical personnel in an
appropriate care unit, with resuscitation equipment present.
It is recommended that a routine follow-up examination be done one
month after discharge, and then every three months until the device
nears the replacement date.
After a device reset, the magnet rate is equal to 87 ppm; it will be
updated within the next 24 hours.
Refer to the online help for a description of displayed warning, and the
necessity to contact Sorin CRM for an evaluation.
Implant software upgrade: in case a new implant software is
downloaded in the device memory through the programmer, a warning
message could be displayed by the programmer to inform the user and
give the proper instructions to follow.
54 – US-ENGLISH
17.2. HOLTER FUNCTION
The Holter records up to 14 tachyarrhythmia episodes as well as the
therapy history.
STORED EPISODES
PARADYM RF CRT-D 9750 stores up to 14 episodes (VF, VT, Slow VT,
SVT/ST, nonsustained).
For each episode four levels of details are presented:
─ Tachogram (to visualize PP and PR intervals)
─ Event log for the entire episode:
─ PARAD/PARAD+ analysis for each majority,
─ Delivered therapies,
─ Markers: Atrial, ventricular and biventricular markers, sensed, paced
and in relative refractory periods,
─ EGM: onset and detection of the arrhythmia, on two therapies, and the
return to slow rhythm by recording electrogram.
Therapy history: For each arrhythmia detection, each therapy
delivered (either automatically or during an electrophysiological study)
and at the end of each arrhythmia, PARADYM RF CRT-D 9750 records
the type of majority rhythm, the number of ATP sequences delivered,
the energy and the number of shocks delivered.
US-ENGLISH – 55
17.3. ELECTIVE REPLACEMENT INDICATOR (ERI)
Elective Replacement Indicators (ERI)(1) are:
─ magnet rate equal to 80 ± 1 min-1 or
─ battery voltage equal to 2.66 V ± 0.01 V
Caution: The defibrillator should be replaced as soon as the Elective
Replacement Indicator (ERI) point is reached.
Between the ERI and the EOL (End of Life)(2), PARADYM RF CRT-D
9750 can still function for:
─ 7.4 months (100% atrial and biventricular pacing in DDD mode,
500 ohms, with as-shipped settings), and deliver 7 shocks at 34 J
or
─ 6.4 months (0% pacing, sensor OFF, one 42 J shock every
2 weeks).
Once the Elective Replacement Indicator (ERI) point has been
reached, the device operates normally, except that the charge time
increases. Under normal conditions (and without programmer use) the
charge times are as as follows:
Shock energy
Charge time (sec)
BOL
42 J
10 (± 2)
ERI
42 J
13 (± 3)
(1) Elective Replacement Indicators (ERI) corresponds to Recommended
Replacement Time (RRT) as referred in the EN45502-2-2 standard.
(2) End of Life (EOL) corresponds to End of Service (EOS) as referred in the
EN45502-2-2 standard.
56 – US-ENGLISH
17.4. EXPLANTATION
The defibrillator should be explanted in the following cases:
─ The Elective Replacement Indicator (ERI) point is reached
─ Confirmed malfunction
─ Burial of the patient (for environmental reasons, the local regulation
may require the explantation of the devices containing a battery supply)
─ Cremation of the patient (the defibrillator may explode if placed in an
incinerator).
The explanted defibrillator should not be reused in another patient.
All explanted defibrillators should be returned to Sorin CRM, carefully
cleaned of all traces of contamination. This may be done by immersing
them in an aqueous sodium hypochlorite containing at least 1%
chlorine, followed by rinsing copiously with water.
The defibrillator should be protected against mechanical impact and the
temperature variations that may occur during shipping.
Before explantation, it is advisable to:
─ print out all programmed parameters, statistics and Holter function
report,
─ save Patient data on floppy disk or hard disk,
─ disable shock therapies (VT and VF) to avoid any risk of untimely
shock.
US-ENGLISH – 57
17.5. DEFIBRILLATOR IDENTIFICATION
The defibrillator can be interrogated and programmed via telemetry,
using the programming head interfaced with the Sorin CRM dedicated
programmer.
Position the programming head over the telemetry antenna located in
the upper part of the device, in order to communicate effectively via
telemetry (see diagram below).
The device can be non-invasively identified as follows:
1.
Take an X-ray to identify the name of the manufacturer and
model, printed on the device (x-ray ID is SDD : S = SORIN ; D =
Defibrillator ; D = PARADYM RF CRT-D 9750).
2.
Interrogate the device using the Sorin CRM dedicated
programmer. The model and serial number of the device are
automatically displayed. The first figure in the serial number
corresponds to the last figure in the year of manufacture.
58 – US-ENGLISH
18. SUPPLEMENTAL INFORMATION
Clinical data presented in this section are from the SafeR clinical study.
SafeR operation in PARADYM RF is similar to that in the Symphony
pacemaker. The data provided are applicable to PARADYM RF CRTD.
18.1. ADVERSE EVENTS IN THE SAFER STUDY
Clinical study of the SafeR included 45 Symphony 2550 devices
implanted in 45 patients. No serious adverse events were device- or
feature-related. There were no deaths in the study. Table 1
summarizes the safety data for this study.
US-ENGLISH – 59
Table 1: Summary of Symphony safety data during study
Patients
Number of events
Number of
patients
% of
patients
Number of
events
Events per
device
year (a)
Deaths
Explants
Serious pacemaker
related events outside
the use of SafeR
Non-serious
pacemaker related
events outside the use
of SafeR
Serious events due to
the use of SafeR
Non-serious events
related due to the use
SafeR
13
28.9
15
3.2
Serious nonpacemaker related
events
13.3
1.9
Non-serious nonpacemaker related
events
17.8
1.7
(a) 4.74 device years
60 – US-ENGLISH
Non-serious events due to the use of SafeR included: delay in
switching on 2nd degree AV block, inappropriate classification of a
PAC, disagreement between markers and recorded EGM, atrial pacing
above the maximum rate, recycling on an r-wave in a refractory period,
and disagreement in the statistics for switches to DDD. No patient
symptoms were associated with these events.
18.2. SAFER CLINICAL STUDY
SafeR mode in PARADYM RF is similar to that in Symphony.
The differences in SafeR mode between the two devices are:
─ To prevent long RR intervals during VT/VF, SafeR has no effect during
VT/VF therapy, electrophysiologic studies, and post-shock recovery.
─ The maximum acceptable AV delay for first degree AV block varies
as a function of pacing rate.
─ PARADYM RF requires a ventricular sensed event to atrial paced
event (RA) interval of at least 100 ms. Therefore, the device
lengthens the atrial escape interval so that it ends at least 102 ms
after the ventricular event.
─ During atrial fibrillation episode, pause criterion is fixed to 2s to
avoid long bradycardia episodes in switching to DDD mode.
Despite these differences, the data collected on Symphony devices are
applicable to PARADYM RF because the principles of SafeR operation
did not change. The criteria for switching from AAI to DDD (or vice
versa) did not change. The device’s method for evaluating the
presence of AV conduction did not change.
Methods: All patients were implanted with a Symphony Model 2550 dualchamber rate-responsive pacemaker with SafeR mode. A variety of
marketed atrial and ventricular pacing leads were used. The pacemaker
was programmed and interrogated via bi-directional telemetry using a Sorin
CRM dedicated programmer and a CPR3 programming head.
The study’s routine evaluation consisted of enrollment, pre-discharge
evaluation, and a scheduled follow-up visit at one month.
US-ENGLISH – 61
At pre-discharge, a 24-hour Holter recording was performed and
pacemaker memory was read. At one month, pacemaker memory was
read. Investigators also documented adverse events.
Patients studied: A total of 45 patients from 12 centers had Symphony
2550 pacemakers with SafeR. Of these, 14 (31 %) were female and 31
(69 %) were male. Mean patient age (± SD) was 74 ± 9 years.
Primary indications for implant were: 1st degree AV block (11.1 %), 2nd
degree AV block (6.7 %), 3rd degree AV block (22.2 %), sinus node
dysfunction (62.2 %) or other (6.7 %).
Effectiveness results: To determine the effectiveness of SafeR mode,
the percentage of ventricular pacing provided over one month was
recorded from pacemaker memory.
Thirty-five patients contributed data to evaluate the percentage of ventricular
pacing provided with SafeR. Twenty-nine patients had 1 % or less
ventricular pacing and six patients had a range of 28-97 % ventricular
pacing. The graph below shows the distribution of ventricular pacing
observed in patients with and without AV block as a primary indication for
implant.
62 – US-ENGLISH
The graph shows that many patients programmed to SafeR had less than
1% ventricular pacing:
 84 % of patients without AV block at implant.
 63 % of patients with AV block at implant.
In a representative reference group1(1) of patients programmed to DDD,
none had less than 1 % ventricular pacing and only 10 % had less than
90 % ventricular pacing regardless of AV block indication at implant.
The actual reduction of ventricular pacing that SafeR provides in an
individual will depend on the amount of time that the patient spends in AV
block. SafeR cannot and should not provide any decrease in ventricular
pacing while the patient is in AV block.
(1) Pioger G, Jauvert G, Nitzsché R, Pozzan J, Laure H, Zigelman M, Leny G,
Vandrell M, Ritter P, and Cazeau S. Incidence and predictive factors of atrial
fibrillation in paced patients. PACE, 28, Supp 1: S137-141; January 2005. This
was a prospective observational study of 377 patients with a functionally similar
device programmed to DDD. The primary indications for implant were: AV block
(49 %), sinus node disease (16 %), brady-tachy syndrome (5 %), AV block +
sinus node disease (19 %), AV block + brady-tachy syndrome (6 %), and bradytachy syndrome + sinus node disease (5 %).
US-ENGLISH – 63
19. PHYSICAL CHARACTERISTICS
Dimensions
69.5 x 73.4 x 11 mm
Weight
95 g
Volume
38.6 cm3
Active surface area of casing
76 cm2
Connector
Atrium: IS-1. Right ventricle: IS-1, DF-1.
Left ventricle: IS-1.
19.1. MATERIALS USED
Active surface area of casing
99% pure titanium
Connectors
Polyurethane* and silicone elastomer*
DF-1 insulating plug
silicone elastomer*
*Medical-grade materials that have undergone “in vitro” and “in vivo”
qualifications.
64 – US-ENGLISH
20. ELECTRICAL CHARACTERISTICS
Atrial input impedance
80 kilohms ± 30 %
Ventricular input impedance
80 kilohms ± 30 %
D.C. capacitance
148 µF ± 8 %
Capacitor formation
No formation required
Rate limit
192 min-1 ± 10 min-1
Pacing waveform
Defibrillation waveform
US-ENGLISH – 65
20.1. TABLE OF DELIVERED SHOCK ENERGY
AND VOLTAGE
The relationship between stored energies, maximum voltages and
delivered energies (at 37 °C, 50 ohm load) for the minimum, low, mean
and maximum programmed energy values is as follows:
Stored energy (J)
0.5
10
20
34
42
V1 (Volt)
75
341
483
631
702
V2 (Volt)
37
173
245
318
353
Delivered E: Phase 1 (J)
0.31
7.0
14.0
23.9
29.6
Delivered E: Phase 2 (J)
0.08
1.8
3.6
6.1
7.5
Delivered E: Total (J)
0.4
8.8
17.6
30.0
37.1
Tolerances are 12% for voltage (25% at 0.5 J) and 30% for energy.
20.2. BATTERY
Manufacturer
Greatbatch
Type
Quasar High Rate (QHR)
Model
GB 2593
Number of batteries
Total capacity
1964 mAh
Usable capacity
Between BOL and ERI: 1278 mAh.
Between BOL and EOL: 1675 mAh.
Voltage
BOL: 3.25 V. ERI: 2.66 V. EOL: 2.5 V.
66 – US-ENGLISH
20.3. LONGEVITY
The longevities mentioned below are calculated by taking into account
6 months storage.
5.1 years
Biventricular pacing in DDD mode, 100%, 500 ohm, 3.5 V, 0.35 ms,
60 min-1, one 42 J shock per quarter, sensor OFF
5.0 years
Biventricular pacing in DDD mode, 100%, 500 ohm, 3.5 V, 0.35 ms,
60 min-1, one 42 J shock per quarter, sensor ON
6.0 years
Biventricular pacing in DDD mode, 1% in atrium, 100% in both
ventricles, 500 ohm, 3.5 V, 0.35 ms, 60 min-1, one 42 J shock per
quarter, sensor OFF
4.2 years
Biventricular pacing in DDD mode, 15% in atrium, 100% in both
ventricles, 500 ohm, 4.5 V, 0.50 ms, 60 min-1, one 42 J shock per
quarter, sensor OFF
9.0 years
0% pacing, one 42 J shock per quarter, sensor OFF
The mean longevity as a function of shocks delivered at maximum
energy, with and without pacing, is as follows:
US-ENGLISH – 67
The mean longevity as a function of yearly remote follow-ups(1), with
and without pacing, is as follows:
(1) An excessive number of remote follow-up can have a non-negligible
impact on device longevity.
68 – US-ENGLISH
21. PROGRAMMABLE PARAMETERS
measured at 37 °C under a 500 ohm load
Legend:
Value in bold: “as shipped” value
Underlined value: nominal value
21.1. ANTIBRADYCARDIA PACING
Basic parameters
Values
Mode
VVI-VVIR-DDD-DDDR-DDD/DDIR-DDIDDIR-SafeR (AAI <=> DDD)-SafeR-R
(AAIR <=> DDDR)
Basic rate (min-1) (1)
From 30 to 90 by steps of 5 ; 60 (± 4 %)
Maximum rate (min-1)
From 100 to 145 by steps of 5 ; 120
(± 6 %)
Rate hysteresis (%)
0-5-10-20-35 (± 18 ms)
Rest AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-115125-135-140-150-155-165-170-180-190195-205-210-220-225-235-250 (± 19 ms)
Exercise AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-115125-135-140-150-155-165-170-180-190195-205-210-220-225-235-250 (± 19 ms)
AVD Paced/Sensed Offset (ms)
0-10-15-25-30-40-45-55-65-70-80-85-95100-110-115-125 (± 1 ms)
(1) The corresponding periods are (in ms): 2000-1714-1500-1333-1200-10911000-923-857-800-750-706-667 ms.
US-ENGLISH – 69
Special features
Values
Smoothing
OFF-Very slow-Slow-Medium-Fast
Mode Switch
ON-OFF
Mode Switch Rate (min-1)
From 30 to 90 by steps of 5 ; 60
Physical activity
Very low-Low-Medium-High-Very high
Exercise AV opt. rate (min-1)
From 70 to 120 by steps of 5; 90
Pacing/Sensing
Values
Atrial sensitivity (mV) (1)
From 0.2 to 4 by steps of 0.2 ; 0.4
(± 50 %)
Atrial amplitude (V) (2)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
Atrial pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
(± 10 %)
Ventricular sensitivity (mV) (1)
From 0.4 to 4 by steps of 0.2 ; 0.4
(± 50 %)
RV amplitude (V) (2)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
RV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
(± 10 %)
LV amplitude (V) (2)
0.25 (± 50 %)
0.5-0.75- (± 30 %)
1-1.25-1.5-1.75-2-2.25-2.5-2.75-3-3.253.5-3.75-4-4.25-4.5-4.75-5-6-7 (± 20 %)
LV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
(± 10 %)
LV pacing polarity
LV bipolar-LV tip to RV-LV ring to RV coil
V chambers
Right-R+L-L+R
VV delay (ms)
0-8-16-24-32-40-48-56-64 (± 3 ms)
70 – US-ENGLISH
(1) Values are measured using a positive and negative triangular signal of 2/13
ms.
(2) The correlation between the programmed amplitudes, the stored amplitudes
and the mid-pulse delivered amplitudes under a 500 ohm load are given in
the following table:
Programmed ampl. (V)
0.25* 0.5* 0.75* 1
1.25* 1.5
Mid-pulse delivered ampl. (V)
0.28
0.49
0.76
0.97
1.18
1.39
Stored amplitude (V)
0.33
0.57
0.89
1.14
1.38
1.63
Programmed ampl. (V)
1.75* 2
2.25* 2.5
2.75* 3
Mid-pulse delivered ampl. (V)
1.66
1.79
2.08
2.35
2.56
2.84
Stored amplitude (V)
1.95
2.10
2.44
2.76
3.01
3.33
Programmed ampl. (V)
3.25* 3.5
3.75* 4
4.25* 4.5
Mid-pulse delivered ampl. (V)
3.05
3.25
3.39
3.58
3.88
4.23
Stored amplitude (V)
3.58
3.82
3.98
4.20
4.55
4.96
Programmed ampl. (V)
4.75* 5
7*
Mid-pulse delivered ampl. (V)
4.36
4.47
5.37
6.26
Stored amplitude (V)
5.12
5.25
6.30
7.35
* For left ventricular amplitude only.
Ventricular arrhythmia
prevention
Values
Atrial pacing on PVC
Yes-No
Post extrasystolic pause suppression
Yes-No
Acceleration on PVC
Yes-No
Max accelerated rate (min-1)
From 60 to 145 by steps of 5; 100
US-ENGLISH – 71
Post-shock mode
Values
Mode
OFF-VVI-DDI-DDD
Duration
10s-20s-30s-1min-2min-3min-4min-5min
Basic rate (min-1)
From 50 to 90 by steps of 5 ; 60 (± 4 %)
Rest AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-115125-135-140-150-155-165-170-180-190195-205-210-220-225-235-250 (± 19 ms)
Exercise AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-115125-135-140-150-155-165-170-180-190195-205-210-220-225-235-250 (± 19 ms)
AVD Paced/Sensed Offset (ms)
0-10-15-25-30-40-45-55-65-70-80-85-95100-110-115-125 (± 1 ms)
A amplitude (V)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
A pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
RV amplitude (V)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)
RV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
LV amplitude (V)
0.25- (± 50 %)
0.5-0.75- (± 30 %)
1-1.25-1.5-1.75-2-2.25-2.5-2.75-3-3.25-3.53.75-4-4.25-4.5-4.75-5-6-7 (± 20 %)
LV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
72 – US-ENGLISH
Refractory periods
Values
Atrial refractory period post ventricular
sensing (ms)
45-65-80-95-110-125-140-155 (± 16 ms)
Atrial refractory period post ventricular
pacing (ms)
80-95-110-125-140-155 (± 4 ms)
Sensitivity margins
Values
Atrial post pacing/sensing margin (mV)
From 0 to 1 by steps of 0.2 ; 0.4
Ventricular post pacing margin (mV)
From 0 to 2 by steps of 0.2 ; 0.8
Response to noise
Values
Automatic sensitivity on noise
ON-OFF
V pacing on noise
ON-OFF
SafeR (AAI <=> DDD)
parameters
Values
AVB I switch
Rest+Exercise-Exercise
Long PR: max (ms)
80-100-125-150-200-250-300-350-400450-500
Long PR: min (ms)
80-100-125-150-200-250-300-350-400450-500
Max. pause (s)
2-3-4
US-ENGLISH – 73
21.2. VENTRICULAR TACHYARRHYTHMIA
DETECTION
Therapy zones
Values
Slow VT detection zone (1)
Slow VT ON-Slow VT OFF
VT detection zone
VT ON-VT OFF
Fast VT / VF detection zone
Fast VT+VF ON-VF ON
Slow VT rate (lower limit) (min-1)
From 100 to 200 by steps of 5 ; 190
VT rate (lower limit) (min-1)
130-135-140-145-150-155-160-165-170175-180-185-190-195-200-210-220-230
VF rate (lower limit) (min-1)
150-155-160-165-170-175-180-185-190195-200-210-220-230-240
Fast VT rate (upper limit) (min-1)
155-160-165-170-175-180-185-190-195200-210-220-230-240-255
Slow VT persistence (cycles)
4-6-8-12-16-20-30-50-100-200
VT persistence (cycles)
4-6-8-12-16-20-30-50-100-200
VF persistence (cycles)
From 4 to 20 by steps of 1 ; 6
(1) The Slow VT zone should be programmed ON only if the VT zone is
programmed ON.
74 – US-ENGLISH
Detection criteria
Values
Slow VT and VT detection criteria
Rate Only-Stability-Stability+Stability/Acc-Stability+/Acc-PARADPARAD+
Fast VT detection criteria
Rate+Stability-Rate Only
Majority: (X/Y), Y (cycles)
8-12-16
Majority: (X/Y), X (%)
65-70-75-80-90-95-100
Window of RR stability for Slow VT and 30-45-65-80-95-110-125-125
VT (ms)
Window of RR stability for fast VT (ms)
30-45-65
Prematurity acceleration (%)
6-13-19-25-31-38-44-50
Long cycle persistence extension
(cycles)
From 0 to 16 by steps of 1 ; 10
Long cycle gap (ms)
15-30-45-65-80-95-110-125-140-155170-190-205
Atrial monitoring
Yes-No
US-ENGLISH – 75
21.3. VENTRICULAR TACHYARRHYTHMIA THERAPIES
Common parameters
Values
Enable ATP therapy
Yes-No
Enable shock therapy
Yes-No
ATP pacing chamber
Right-Left-R+L
Polarity alternation (42J)
Yes-No
Atrial coil (SVC) present
Yes-No
Active case
Yes-No
Shock configuration (+ --> -)
Case to RV-SVC to RV-Case + SVC to
RV-RV to Case-RV to SVC-RV to Case +
SVC
SVC exclusion (shock < 15J)
Yes-No
 Therapy parameters in slow VT zone
ATP 1 program
Values
ATP program
OFF-Burst-Burst+Scan-RampRamp+Scan
Number of sequences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles in first sequence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles added per sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95
Ramp decrement (per cycle) (ms)
0-4-8-12-16-20-30-40-50-60
Scan decrement (per sequence) (ms)
0-4-8-12-16-20-30-40-50-60
Time limit (min)
0.5-1-1.5-2-2.5-3-3.5-4
Minimum cycle length (ms)
95-110-125-140-155-170-190-205-220235-250-265-280-295-310
76 – US-ENGLISH
ATP 2 program
Values
ATP program
OFF-Burst-Burst+Scan-RampRamp+Scan
Number of sequences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles in first sequence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles added per sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95
Ramp decrement (per cycle) (ms)
0-4-8-12-16-20-30-40-50-60
Scan decrement (per sequence) (ms)
0-4-8-12-16-20-30-40-50-60
Time limit (min)
0.5-1-1.5-2-2.5-3-3.5-4
Minimum cycle length (ms)
95-110-125-140-155-170-190-205-220235-250-265-280-295-310
Shock program
Values
Shock 1 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-67-8-9- (± 30 %)
10-12-14-16-18-20-22-24-26-28-30-3234-42 (± 15 %)
Shock 2 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-67-8-9- (± 30 %)
10-12-14-16-18-20-22-24-26-28-30-3234-42 (± 15 %)
Number of Max. Shock (42 J)
OFF-1-2-3-4
US-ENGLISH – 77
 Therapy parameters in VT zone
ATP 1 program
Values
ATP program
OFF-Burst-Burst+Scan-RampRamp+Scan
Number of sequences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles in first sequence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles added per sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95
Ramp decrement (per cycle) (ms)
0-4-8-12-16-20-30-40-50-60
Scan decrement (per sequence) (ms)
0-4-8-12-16-20-30-40-50-60
Time limit (min)
0.5-1-1.5-2-2.5-3-3.5-4
Minimum cycle length (ms)
95-110-125-140-155-170-190-205-220235-250-265-280-295-310
ATP 2 program
Values
ATP program
OFF-Burst-Burst+Scan-RampRamp+Scan
Number of sequences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles in first sequence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles added per sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95
Ramp decrement (per cycle) (ms)
0-4-8-12-16-20-30-40-50-60
Scan decrement (per sequence) (ms)
0-4-8-12-16-20-30-40-50-60
Time limit (min)
0.5-1-1.5-2-2.5-3-3.5-4
Minimum cycle length (ms)
95-110-125-140-155-170-190-205-220235-250-265-280-295-310
78 – US-ENGLISH
Shock program
Values
Shock 1 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-67-8-9- (± 30 %)
10-12-14-16-18-20-22-24-26-28-30-3234-42 (± 15 %)
Shock 2 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-67-8-9- (± 30 %)
10-12-14-16-18-20-22-24-26-28-30-3234-42 (± 15 %)
Number of Max. Shock (42 J)
OFF-1-2-3-4
 Therapy parameters in fast VT / VF zone
ATP 1 program
Values
ATP program
OFF-Burst-Burst+Scan-RampRamp+Scan
Number of sequences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles in first sequence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Cycles added per sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95
Ramp decrement (per cycle) (ms)
0-4-8-12-16-20-30-40-50-60
Scan decrement (per sequence) (ms)
0-4-8-12-16-20-30-40-50-60
Time limit
10s-20s-30s-1min-1.5min-2min
Minimum cycle length (ms)
95-110-125-140-155-170-190-205-220235-250-265-280-295-310
US-ENGLISH – 79
Shock program
Values
Shock 1 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-67-8-9- (± 30 %)
10-12-14-16-18-20-22-24-26-28-30-3234-42 (± 15 %)
Shock 2 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-67-8-9- (± 30 %)
10-12-14-16-18-20-22-24-26-28-30-3234-42 (± 15 %)
Number of Max. Shock (42 J)
1-2-3-4
21.4. REMOTE ALERTS AND WARNINGS
General parameters
Values
RF communication (1)
ON-OFF
Remote alerts (1)
ON-OFF
(1) RF and Remote alerts are turned on automatically if Shocks are programmed
ON.
System Alerts
Values
Battery depletion – ERI
ON-OFF
Device reset
ON-OFF
Excessive charge time (>25s)
ON-OFF
System integrity
ON-OFF
80 – US-ENGLISH
Lead Alerts
Values
Abnormal A lead impedance
ON-OFF
Abnormal A lead low limit (Ohm)
200-250-300-350-400-450-500
Abnormal A lead high limit (Ohm)
1500-1750-2000-2500-3000
Abnormal RV lead impedance
ON-OFF
Abnormal RV lead low limit (Ohm)
200-250-300-350-400-450-500
Abnormal RV lead high limit (Ohm)
1500-1750-2000-2500-3000
Abnormal LV lead impedance
ON-OFF
Abnormal LV lead low limit (Ohm)
200-250-300-350-400-450-500
Abnormal LV lead high limit (Ohm)
1500-1750-2000-2500-3000
Abnormal RV coil impedance
ON-OFF
Abnormal SVC coil impedance
ON-OFF
Abnormal Shock impedance (1)
ON-OFF
(1) Normal impedance range [20 Ohm-200 Ohm]
Clinical status
Values
V oversensing
ON-OFF
High AT/AF burden
ON-OFF
AT/AF limit (on 24h) (h)
0.5-1-3-6-12-24
Fast V Rate during AT/AF
ON-OFF
Fast V Rate limit (min-1)
80-90-100-110-120
Fast V Duration limit (h)
0.5-1-3-6-12-24
Limited % of V pacing in CRT
ON-OFF
Limited % of V pacing (%)
50-70-80-85-90-95
US-ENGLISH – 81
Therapy information
Values
Shock disabled
ON-OFF
Shocks delivered
OFF-All shocks-Inefficient shockInefficient max shock
22. NON PROGRAMMABLE PARAMETERS
Interval
Values
Committed period
95 ms (± 5 ms)
Atrial refractory periods
Values
Post atrial sensing
47 ms (± 16 ms)
Post atrial pacing
109 ms (± 4 ms)
Ventricular refractory periods
Values
Post ventricular sensing
95 ms (± 16 ms)
Post ventricular pacing
220 ms (± 4 ms)
Post atrial pacing (blanking)
16 ms (± 3 ms)
Tachycardia criteria
Values
Window of PR association
63 ms (± 1 ms)
Therapies
Values
Waveform
Constant tilt (50% - 50%)
Stored energy for the Max. shock
42 J (± 15 %)
Pacing amplitude during ATP therapies
7 V (Actual value at 300 ms: 5.3 V)
Anti-PMT protection
Termin
82 – US-ENGLISH
23. LIMITED WARRANTY
The PARADYM RF implantable cardioverter defibrillator is the result of
highly advanced research and all components have been selected after
exhaustive testing.
Sorin CRM S.r.l. (identified as “Sorin CRM” hereafter) guarantees the
product PARADYM RF against any damage caused by component
failure or production defects during a period of four years after the
implantation date, and Sorin CRM commits itself to replace all
PARADYM RF devices according to the terms described in article 1
and described in article 2 of this section.
Sorin CRM makes no claim that the human body will not react
unsuitably to the implantation of the PARADYM RF device, or that
failure will never occur.
Sorin CRM does not guarantee the suitability of PARADYM RF in
defined types of patients: selection of the device is a medical decision.
Sorin CRM shall not be held liable for any damage indirectly associated
with the PARADYM RF, whether as part of normal or abnormal
operation, nor damage from its explantation or replacement.
Sorin CRM does not authorise anyone to modify these limited warranty
conditions.
23.1. ARTICLE 1 : TERMS OF LIMITED WARRANTY
1. The PARADYM RF implantable cardioverter defibrillator is only
guaranteed for the first implantation.
2. The EURID/IAPM implant form must be sent to Sorin CRM within
30 days after implantation.
3. The PARADYM RF cardioverter defibrillator must be implanted
prior to the use-before date indicated on the packaging.
US-ENGLISH – 83
4. The limited guarantee only applies to suspect devices returned to
the manufacturer, carefully packed and accompanied by an
explantation report duly completed by the hospital or the doctor and
considered defective after analysis by Sorin CRM.
The device must be returned within the 30 days following
explantation to Sorin CRM.
Any device returned and replaced under the terms of this limited
warranty will become the exclusive property of Sorin CRM.
Any rights under the terms of this limited warranty will be forfeited if
the PARADYM RF device has been opened by anyone other than
Sorin CRM.
These rights will also be forfeited if the device has been damaged
by carelessness or accident.
This is the case especially if the device has been exposed to
temperatures above 50°C, to electrical abuse or to mechanical
shock, particularly as a result of being dropped. Consequently, any
expert opinion offered by a third party after the device has been
removed also nullifies the guarantee.
5. The limited warranty will be forfeited if it is proven that the device
has been misused or inadequately implanted, against the
physicians’manual recommendations of PARADYM RF.
6. The limited warranty does not include leads and other accessories
used for the implantation.
7. The replacement terms or conditions described in article 2 include
all devices that shall be replaced within the limited warranty period
because of battery depletion, without any link to a component
failure or a production hazard. The device battery longevity varies
with the type and number of delivered therapies.
8. Legal requirements of jurisdictions where the PARADYM RF device
is distributed will supersede any warranty conditions indicated in
this manual that conflict with such laws.
84 – US-ENGLISH
23.2. ARTICLE 2 : TERMS OF REPLACEMENT
1. In case of PARADYM RF failure because of a component failure, a
production defect, or a conception error, occurring within two-year
period starting from the implantation date, Sorin CRM is committed
to:
 replacing free of charge the explanted device by a Sorin CRM
device with equivalent features,
 or issuing a replacement credit equal to the purchase price for
the purchase of any other Sorin CRM replacement device.
After a two-year period and up to 4 years after the implantation, Sorin
CRM, because of limited warranty terms, will issue a replacement
credit to the buyer of an amount equivalent to half of the initial
purchase price minus prorata temporis during this two-years period.
In any case the credit issued by the limited warranty terms cannot
exceed the purchase price of a Sorin CRM replacement device.
US-ENGLISH – 85
24. PATENTS
The PARADYM RF model described in this manual is covered by the
following US patents:
5 167 224, 5 226 415, 5 271 394, 5 312 451, 5 325 856, 5 339 820,
5 350 406, 5 411 533, 5 462 060, 5 513 645, 5 545 181, 5 558 097,
5 564 430, 5 591 218, 5 626 619, 5 645 574, 5 674 265, 5 697 960,
5 702 424, 5 702 426, 5 713 928, 5 741 315, 5 776 164, 5 776 165,
5 818 703, 5 836 980, 5 868 793, 5 891 170, 5 891 184, 5 899 931,
5 931 856, 5 935 153, 5 954 660, 5 978 708, 6 181 968, 6 230 058,
6 236 111, 6 251 703, 6 256 206, 6 307 261, 6 337 996, 6 397 105,
6 408 209, 6 487 451, 6 487 452, 6 505 068, 6 532 238, 6 556 866,
6 604 002, 6 622 039, 6 625 491, 6 711 441, 6 738 665, 6 830 548,
6 889 080, 6 898 845, 6 912 421, 6 937 898, 6 975 905, 7 065 402,
7 072 716, 7 076 297, 7 113 826, 7 142 924, 7 164 946, 7 251 526,
7 366 566, 7 400 921, 7 400 922, 7 953 483.
86 – US-ENGLISH
25. EXPLANATION OF SYMBOLS
The symbols on product labelling have the following meaning:
Fo
Te
US-ENGLISH – 87
FCC ID YSGCRTD9750
IC : 10270A-CRTD9750
Last revision date of this implant manual: 2012-05
88 – US-ENGLISH

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 88
Language                        : fr-FR
Title                           : TABLE OF CONTENTS
Author                          : Anne Couturier
Creator                         : Microsoft® Word 2010
Create Date                     : 2012:05:24 14:24:27+02:00
Modify Date                     : 2012:05:24 14:24:27+02:00
Producer                        : Microsoft® Word 2010
EXIF Metadata provided by EXIF.tools
FCC ID Filing: YSGCRTD9750

Navigation menu