SORIN CRM CRTDSOR1841 Implantable cardioverter defibrillator with RF wireless communication availability User Manual

SORIN CRM Implantable cardioverter defibrillator with RF wireless communication availability

User manual

Implantable cardioverter defibrillator with cardiac resynchronization therapySonR CRT -D modelsImplant manual
  TABLE OF CONTENTS 1. General description................................................................................................................5  2. Indications...............................................................................................................................6  3. Contraindications....................................................................................................................7  4. Warnings and precautions..................................................................................................... 8  4.1. Risks related to medical environment.......................................................................................9  4.2. Sterilization, storage and handling..........................................................................................10  4.3. Implantation and device programming....................................................................................10  4.4. Lead evaluation and lead connection......................................................................................11  4.5. Generator explant and disposal..............................................................................................12  5. Adverse events...................................................................................................................... 13  5.1. MSP study...............................................................................................................................13  5.2. Potential adverse events.........................................................................................................14  6. Clinical studies......................................................................................................................16  6.1. MSP clinical study...................................................................................................................16  7. Patient selection and treatment...........................................................................................20  7.1. Individualization of treatment..................................................................................................20  7.2. Specific patient populations....................................................................................................21  8. Patient counselling information..........................................................................................22  9. Declaration of conformity.....................................................................................................23  10. Physician guidelines.............................................................................................................25  10.1. Physician training....................................................................................................................25  10.2. Directions for use....................................................................................................................25  10.3. Maintaining device quality.......................................................................................................25  10.4. V-V Programming Recommendation.......................................................................................26  11. Patient information...............................................................................................................27  12. How supplied.........................................................................................................................28  12.1. Sterility....................................................................................................................................28  12.2. Warranty and replacement policy............................................................................................28  13. Device description................................................................................................................29  14. Implant procedure.................................................................................................................31  14.1. Necessary equipment.............................................................................................................31  14.2. Packaging...............................................................................................................................31  14.3. Optional equipment.................................................................................................................31  14.4. Before opening the package...................................................................................................32  14.5. Prior to implantation................................................................................................................32  14.6. Device placement.................................................................................................................... 32  14.7. Choosing the type of lead.......................................................................................................32  14.8. Shock configuration (+ -> -).....................................................................................................33  14.9. Measurement of thresholds at implant....................................................................................34  14.10.Leads connection....................................................................................................................34  14.11. Device implantation.................................................................................................................36  14.12.Tests and programming...........................................................................................................36  15. Special modes.......................................................................................................................37  15.1. Safety mode (nominal values).................................................................................................37 SORIN – PLATINIUM SonR CRT-D – U463A 3
   15.2. Magnet mode..........................................................................................................................37  15.3. Response in the presence of disturbance...............................................................................37  15.4. Detection characteristics in the presence of electromagnetic fields........................................37  15.5. Protection against short-circuits..............................................................................................39  16. Main functions.......................................................................................................................40  16.1. Automatic lead measurements................................................................................................40  16.2. Atrial tachyarrhythmia management.......................................................................................40  16.3. Ventricular tachyarrhythmia management...............................................................................40  16.4. Pacing.....................................................................................................................................41  16.5. Sensing...................................................................................................................................41  16.6. SonR CRT optimization...........................................................................................................42  16.7. Follow-up function...................................................................................................................42  16.8. Remote Monitoring function....................................................................................................42  17. Patient follow-up...................................................................................................................45  17.1. Follow-up recommendations...................................................................................................45  17.2. Holter Function........................................................................................................................45  17.3. Recommended Replacement Time (RRT)..............................................................................46  17.4. Explantation............................................................................................................................46  17.5. Defibrillator identification.........................................................................................................47  18. Supplemental Information....................................................................................................48  18.1. Adverse events in the SafeR (AAI <> DDD) study..................................................................48  18.2. SafeR (AAI <> DDD) clinical study..........................................................................................49  19. Physical characteristics.......................................................................................................51  19.1. Materials used.........................................................................................................................51  20. Electrical characteristics......................................................................................................52  20.1. Table of delivered shock energy and voltage..........................................................................52  20.2. Battery.....................................................................................................................................52  20.3. Longevity................................................................................................................................. 53  21. Programmable parameters...................................................................................................55  21.1. Antibradycardia pacing............................................................................................................55  21.2. Ventricular tachyarrhythmia detection.....................................................................................60  21.3. Ventricular tachyarrhythmia therapies.....................................................................................61  21.4. Remote alerts and warnings...................................................................................................64  22. Non programmable parameters...........................................................................................66  23. Limited warranty...................................................................................................................67  24. Patents...................................................................................................................................68  25. Explanation of symbols........................................................................................................69 4SORIN – PLATINIUM SonR CRT-D – U463A
 1.  GENERAL DESCRIPTION 1. GENERAL DESCRIPTIONPLATINIUM SonR CRT-D is an implantable cardioverter defibrillator for the recognition and treatment   of   ventricular tachycardia  and   fibrillation,   with  ventricular  resynchronization,  in patients with spontaneous or inducible tachyarrhythmias. PLATINIUM SonR CRT-D is equipped with an accelerometer to allow adaptation of pacing to suit the patient’s activity. PLATINIUM SonR CRT-D is also equipped with the RF wireless technology which enables:─Remote monitoring of patients who have the Sorin SMARTVIEW Monitor installed at home,─wireless interrogation and device programming by Orchestra Plus programmer equipped with ORCHESTRA PLUS LINK accessory.PLATINIUM SonR CRT-D provides high energy shocks (42 J) for enhanced safety, as well as automatic lead measurements to monitor system integrity.PLATINIUM  SonR  CRT-D  is  protected against  high-frequency  signals  emitted by  cellular telephones.Device and lead connections:PLATINIUM SonR CRT-D 1811 2*IS-1 bipolar (LV unipolar compatible), 2*DF-1, 1*SonR (IS-1 bipolar compatible)PLATINIUM SonR CRT-D 1841 1*IS-1 bipolar (unipolar compatible), 1*DF4, 1*SonR (IS-1 bipolar compatible)SORIN – PLATINIUM SonR CRT-D – U463A 5
 2.  INDICATIONS 2. INDICATIONSPLATINIUM SonR CRT-D is indicated for ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life threatening arrhythmias.The   device   is   also   indicated   for   the   reduction   of   heart   failure   symptoms   in   medically optimized NYHA Functional Class III and IV patients with left ventricular ejection fraction of 35% or less, and a QRS duration of 150 ms or longer.6SORIN – PLATINIUM SonR CRT-D – U463A
 3.  CONTRAINDICATIONS 3. CONTRAINDICATIONSImplantation of PLATINIUM SonR CRT-D is contraindicated in patients:─whose ventricular tachyarrhythmias may have transient or reversible causes such as: acute myocardial infarction, digitalis intoxication, drowning, electrocution, electrolyte imbalance, hypoxia, sepsis, or unstable ischemic episodes,─who present incessant tachyarrhythmia,─who have an internal pacemaker,─whose primary disorder is bradyarrhythmias, or atrial tachyarrhythmias.The   use   of   the   dual-chamber   pacing   mode   is   contraindicated   in   patients   with   chronic refractory atrial tachyarrhythmias.SORIN – PLATINIUM SonR CRT-D – U463A 7
 4.  WARNINGS AND PRECAUTIONS 4. WARNINGS AND PRECAUTIONSThe   patient  should   be   warned   of   the   potential   risks  of   defibrillator  malfunction   if   he   is exposed to external magnetic, electrical, or electromagnetic signals.These potential interference sources may cause conversion to inhibited mode (because of noise detection), erratic delivery of VT or VF therapies, nominal programming, or much more rarely, irreversible damage to the device’s circuits.The main sources of high magnitude disturbance are: powerful radiofrequency equipment (radar),   industrial   motors   and   transformers,   induction   furnaces,   resistance,   arc-welding equipment and high power loudspeakers.Electrical Isolation:Do  not  permit the  patient to  contact  grounded  equipment  that could produce  hazardous leakage current. Ensuing arrhythmia induction could result in the patient’s death.Antitheft gates:Since antitheft devices at the entrance to stores are not subject to any safety standards, it is advisable to spend as little time as possible in their vicinity.Airport detection systems:Since airport detection systems are not subject to any safety standards, it is advisable to spend as little time as possible in their vicinity.Work environment:The patient's work environment may be an important source of disturbance. In that case, specific recommendations may be required.High voltage power transmission lines:High voltage power transmission lines may generate enough disturbance to interfere with defibrillator operation if approached too closely.Communication equipment:Communication equipment such as microwave transmitters, linear power amplifiers, or high-power amateur transmitters may generate enough disturbance to interfere with defibrillator operation if approached too closely.Home appliances:Home  appliances that  are  in  good  working  order  and  properly  grounded  do   not  usually produce   enough   disturbance   to   interfere   with   defibrillator   operation.   However,   there   are reports of device interferences caused by electric hand tools or electric razors used directly over the device implant site. Patient should also avoid using induction ovens and cookers.8SORIN – PLATINIUM SonR CRT-D – U463ACAUTION: Do not tap sharply on the ICD can after implant, because the ICD's sensing circuits can detect this as P-waves or R-waves, and such oversensing could result in inappropriate pacing, inhibition, or therapy. Normal activities after implant do not result in such oversensing.
 4.  WARNINGS AND PRECAUTIONS4.1. RISKS RELATED TO MEDICAL ENVIRONMENTIt   is   advisable   to  carefully  monitor   defibrillator   operation   prior   to   and  after   any  medical treatment during which  an electrical current  from an external source  passes through the patient's body.Magnetic Resonance Imaging:MRI is strictly contraindicated in cardiac defibrillator patients.Radiofrequency ablation:A  radiofrequency   ablation   procedure   in   a   patient   with   a   generator   may   cause   device malfunction or damage. RF ablation risks may be minimized by:1. Programming Shock Therapy and ATP to OFF.2. Avoiding direct contact between the ablation catheter and the implanted lead or generator.3. Positioning the ground, placing it so that the current pathway does not pass through or near the device, i.e. place the ground plate under the patient’s buttocks or legs.4. Having external defibrillation equipment available.Electrocautery or diathermy device:Diathermy and electrocautery equipment should not be used. If such devices must be used:1. Keep the current path and ground plate as far away from the device and the leads as possible (a minimum of 15 cm [six inches]). 2. Before procedure, deactivate ATP and shock therapies.3. During the procedure, keep the electrocautery device as far as possible from the cardiac defibrillator. Set it at minimum intensity. Use it briefly.4.  After   the   procedure,   check  for  proper   implant   function.   The   device   should  never  be exposed directly to the diathermy source.External defibrillation:PLATINIUM SonR CRT-D is protected from external defibrillation shocks.1. Before external defibrillation, deactivate ATP and shock therapies.2.  During  external  defibrillation,  it  is  advisable  to  avoid  placing  the  defibrillating  paddles directly over the casing or over the leads. The defibrillating paddles should preferably be placed in an anteroposterior position.3. Avoid any direct contact between the defibrillation paddles and the conductive parts of the implanted leads or casing of the implanted device.4. After external defibrillation, check for proper device function.Radiation therapy:Avoid exposure to ionizing radiation. Betatrons are contraindicated. If high doses of radiation therapy cannot be avoided, the defibrillator should be protected from direct exposure with a protection shield. ATP and shock therapies should be disabled during exposure and proper device   function   should   be   checked   regularly   afterwards.   Resulting   damage   may   not   be immediately detectable. If irradiation of tissues close to the implantation site is necessary, it is recommended that the cardiac defibrillator be moved. As a safety measure, an external defibrillator should be immediately available.Lithotripsy:Lithotripsy may permanently damage the device if it is at the focal point of the lithotripsy beam. If lithotripsy must be used, keep the defibrillator at least 2.5 to 5 cm (1-2 inches) away from the focal point of the lithotripsy beam.SORIN – PLATINIUM SonR CRT-D – U463A 9
 4.  WARNINGS AND PRECAUTIONSDiagnostic ultrasound (echography):The defibrillator is not affected by ultrasound imaging devices.Scales with body fat monitors and electronic muscle stimulators:A patient with an implanted PLATINIUM SonR CRT-D should not use these devices.4.2. STERILIZATION, STORAGE AND HANDLINGResterilization:Do not resterilize and re-implant explanted ICDs."Use Before" Date:A "Use Before" date is printed on the outer storage package and on the sterile package. Do not implant the device after this date because the battery may have reduced longevity and sterility may be affected. It should be returned to Sorin.If Package is damaged:Do   not   use   the   device   or   accessories   if   the   packaging   is   wet,   punctured,   opened   or damaged because the integrity of the sterile packaging may be compromised. Return the device to the manufacturer.Device Storage:Store the device in a clean area, away from magnets, kits containing magnets, and sources of electromagnetic disturbance to avoid device damage. Store the device between 0 - 50 °C (32 - 122 °F). Temperatures outside the specified range may damage the device.Equilibration:Allow the device to reach room temperature before programming or implanting the device because rapid temperature changes may affect initial device function.4.3. IMPLANTATION AND DEVICE PROGRAMMINGUse only a Sorin programmer to communicate with the device.Do   not   inadvertently   position   any  magnet  over   the   ICD;  this   suspends  tachyarrhythmia detection and treatment.Replace the device when the RRT (Recommended Replacement Time*) point (defined by a battery voltage of 2.66 ± 0.01 V or a magnet rate lower than or equal to 80 bpm) is reached.Program device parameters such as sensitivity threshold and VT and VF detection intervals as specified in the device manuals.Lead System:Do   not   use   a   lead   system   other   than   those   with   demonstrated   compatibility   because undersensing cardiac activity and failure to deliver necessary therapy may result.In situations where an ICD and a pacemaker are implanted in the same patient, interaction testing should be completed. If the interaction between the ICD and the pacemaker cannot be resolved through repositioning of the leads or reprogramming of either the pacemaker or the  ICD,  the  pacemaker   should   not   be  implanted  (or   should   be   explanted   if previously implanted).Failure   to   properly   insert   the   torque   screwdriver   into   the   perforation   at   an   angle perpendicular to the connector receptacle may result in damage to the sealing system and its self-sealing properties.10 SORIN – PLATINIUM SonR CRT-D – U463A
 4.  WARNINGS AND PRECAUTIONSIn   the   event   of   a   warning   on   a   low  shock   impedance,   and   after   lead   replacement   or reconnection:   it   is   recommended   to   check   the   system   integrity   (sensing   and   pacing thresholds and the impedance of the shock electrodes)It is recommended that a security margin of at least 10 J be demonstrated between the effective  shock   energy   and   maximum   programmable   energy.   Carefully   confirm   that   true ventricular fibrillation has been induced because the DFT for ventricular tachycardia or flutter may be lower.The defibrillator  should be implanted with the  engraved side  facing outwards in order to facilitate   telemetric   communication   with   the   programming   head   and   to   display   the radiographic identification correctly.*: corresponds to ERI (Elective Replacement Indicator) previously used. 4.4. LEAD EVALUATION AND LEAD CONNECTIONPLATINIUM SonR CRT-D 1811 has two DF-1, two IS-1, and one SonR connector ports. The SonR connector port has been specifically designed by Sorin to accept three connections (tripolar). The two distal connections respect the same dimensions as the IS-1 standard, and an additional proximal connection allows to collect the SonR signal.PLATINIUM SonR CRT-D 1841 has one DF4, one IS-1, and one SonR connector ports. The SonR connector port has been specifically designed by Sorin to accept three connections (tripolar). The two distal connections respect the same dimensions as the IS-1 standard, and an additional proximal connection allows to collect the SonR signal.The SonR port accepts either a conventional atrial lead (without SonR capability) or a SonR atrial lead (bipolar pacing/sensing and SonR capability).IS-1   refers   to   the   international   standard   whereby   leads   and   generators   from   different manufacturers are assured a basic fit (ISO 5841-3:2013).DF-1 refers to the international standard for defibrillation lead connectors (ISO 11318:2002).DF4 refers to the international standard for defibrillation lead connectors (ISO 27186:2010).Use only DF4-LLHH or DF4-LLHO standard lead connector types according to ISO 27186: 2010.Do not tie a ligature directly to the lead body, tie it too tightly, or otherwise create excessive strain at the insertion site as this may damage the lead. Use the lead stabilizer to secure the lead lateral to the venous entry site.Do not grip the lead with surgical instruments.Do not use excessive force or surgical instruments to insert a stylet into a lead.Use   ventricular   transvenous   leads   with   caution   in   patients   with   either   a   mechanical   or bioprosthetic tricuspid valvular prosthesis.Use the  correct suture  sleeve  (when needed)  for each  lead,  to immobilize  the lead and protect it against damage from ligatures.Never implant the system with a lead system that has a measured shock impedance of less than   30   ohms.   A   protection   circuit   in   the   defibrillator   prevents   shock   delivery   when impedance is too low. If the shock impedance is less than 30 ohms, reposition the lead system to allow a greater distance between the electrodes.Do not kink leads. Kinking leads may cause additional stress on the leads, possibly resulting in lead fracture.Do not insert a lead connector pin into the connector block without first visually verifying that the   setscrews   are   sufficiently   retracted.   Do   not   tighten   the   setscrews   unless   a   lead connector pin is inserted because it could damage the connector block.SORIN – PLATINIUM SonR CRT-D – U463A 11
 4.  WARNINGS AND PRECAUTIONSLead electrodes in contact during a cardioversion or defibrillation therapy will cause current to bypass the heart, possibly damaging the ICD and the leads. While the ICD is connected to the leads, make sure that the metal portions of any electrodes do not touch each other.If a pacing lead is abandoned rather than removed, it must be capped to ensure that it is not a pathway for currents to or from the heart.If a thoracotomy is required to place epicardial patches, it should be done during a separate procedure to reduce the risk of morbidity and mortality.Do not place the patch lead over nerve tissue as this may cause nerve damage.Place the patch lead with the conducting  coil side facing the heart to ensure delivery of energy to the heart.Place the sutures well outside the coil of the patch lead or in the area between the coils to avoid possible coil fracture.If countershock is unsuccessful using external paddles, adjust the external paddle position (e.g.,   anterior-lateral   to   anterior-posterior)   and   be   sure   that   the   external   paddle   is   not positioned over the patch.Do  not   fold,  alter,   or  remove  any  portion  of  the  patch  as   it  may compromise   electrode function or longevity.If   a   header   port   is   unused  on   the   generator,  the   port   must  be   plugged   to  protect   the generator.4.5. GENERATOR EXPLANT AND DISPOSALInterrogate   the   device,   and   program   shock   therapy   off   prior   to   explanting,   cleaning   or shipping the device to prevent unwanted shocks.Return all explanted generators and leads to the manufacturer.Never incinerate the device due to the potential for explosion. The device must be explanted before cremation.12 SORIN – PLATINIUM SonR CRT-D – U463A
 5.  ADVERSE EVENTS 5. ADVERSE EVENTSClinical data presented in this section are from the MSP clinical study. PLATINIUM SonR CRT-D is similar in design and function to the ALTO 2 MSP and OVATIO CRT-D devices. The data provided are applicable to PLATINIUM SonR CRT-D.5.1. MSP STUDYSorin   conducted   an   international,   multi-center,   randomized   clinical   trial   of   its   cardiac resynchronization therapy system. Investigators attempted to implant study devices in 190 patients. A total of 182 patients received study devices and had an exposure of over 165 device years. Of those patients, 19 received OVATIO CRT-D, 160 received ALTO 2 MSP, and 3 received ALTO MSP. The clinical data collected on ALTO MSP, ALTO 2 MSP and OVATIO CRT-D are applicable to PLATINIUM SonR CRT-D. The table below summarizes the adverse events observed for the CRT-D system. No deaths were related to the system.Event # of Patients % of Patients # of Events Events/100 Device-YearsDeaths not related to the system16 8.4 16 0.8Cardiac arrest 5 2.6 5 0.3Worsening CHF / CHF decompensation3 1.6 3 0.2Multi-organ dysfunction 2 1.1 2 0.1Complications related to the system28 14.7 35 2.1Dislodgment or migration 9 4.7 11 0.6Extracardiac stimulation (e.g., phrenic stim)9 4.7 9 0.5Complications related to the implant procedure18 9.5 21 1.3Dislodgment or migration 4 2.1 4 0.2Observations related to the system23 12.1 27 1.7Extracardiac stimulation (e.g., phrenic stim)12 7.9 15 0.8Observations related to the implant procedure24 12.6 28 1.7Heart block 6 3.2 6 0.3Extracardiac stimulation (e.g., phrenic stim)3 1.5 5 0.3SORIN – PLATINIUM SonR CRT-D – U463A 13
 5.  ADVERSE EVENTSSerious adverse events not related to the system85 44.7 176 10.8Worsening CHF/CHF decompensation24 12.6 42 2.1Atrial fibrillation/flutter 14 7.4 14 0.7Not Serious events not related to the system58 30.5 121 7.4Pain (in back, arms, chest, shoulder, groin, head, other)10 5.3 13 0.7Worsening CHF/CHF decompensation13 6.8 16 0.8Atrial fibrillation/flutter 7 3.7 8 0.4Ventricular tachycardia 7 3.7 7 0.45.2. POTENTIAL ADVERSE EVENTSAdverse   events   (in   alphabetical   order),  including  those   reported  in  the   previous   tables, associated with ICD systems include:─Acceleration of arrhythmias (caused by device),─Air embolism,─Bleeding,─Chronic nerve damage,─Erosion,─Excessive fibrotic tissue growth,─Extrusion,─Fluid accumulation,─Formation of hematomas or cysts,─Inappropriate shocks,─Infection,─Keloid formation,─Lead abrasion and fracture,─Lead migration/dislodgment,─Myocardial damage,─Pneumothorax,─Shunting current or insulating myocardium during defibrillation with internal or external paddles,─Potential mortality due to inability to defibrillate or pace,─Thromboemboli,─Venous occlusion,─Venous or cardiac perforation.Patients susceptible to frequent shocks despite antiarrhythmic medical management may develop psychological intolerance to an ICD system that may include the following:─Dependency,─Depression,14 SORIN – PLATINIUM SonR CRT-D – U463A
 5.  ADVERSE EVENTS─Fear of premature battery depletion,─Fear of shocking while conscious,─Fear that shocking capability may be lost,─Imagined shocking (phantom shock).SORIN – PLATINIUM SonR CRT-D – U463A 15
 6.  CLINICAL STUDIES 6. CLINICAL STUDIESClinical data presented in this section are from the MSP clinical study. PLATINIUM SonR  CRT-D is similar in design and function to the ALTO 2 MSP and OVATIO CRT-D devices. The data provided are applicable to PLATINIUM SonR CRT-D.6.1. MSP CLINICAL STUDYOVATIO CRT-D and earlier models were evaluated clinically in an international, multi-center, randomized   clinical   trial   of   Sorin’s   cardiac   resynchronization   therapy   (CRT-D)   system. Investigators  attempted  to implant study  devices  in  190  patients. A total  of 182  patients received study devices and had an exposure of over 165 device years. Of those patients, 19 received OVATIO CRT-D, 160 received ALTO 2 MSP, and 3 received ALTO MSP.6.1.1. ObjectivesThe primary objectives of the study were to demonstrate:─Greater improvement in a composite endpoint (percent improvement in peak VO2 percent improvement in quality of life) for CRT-D patients than for control patients.─System complication-free rate ≥ 67 % at six months.6.1.2. MethodsPatients were New York Heart Association class III or IV and had one or more indications for an implantable cardioverter defibrillator (ICD). Patients performed cardiopulmonary exercise testing at baseline and six-months after randomization. Patients were implanted with a Sorin ICD with CRT-D, a Situs UW28D left ventricular lead, and commercially available right atrial and ventricular leads. Routine follow-ups were at pre-discharge, randomization (3-14 days post-implant), one month, three months, and six months post randomization.6.1.3. ResultsImprovement in composite endpointPatients were included in the analysis if complete (peak VO2 and quality of life) baseline and six-month data were available.Number of patients contributing to analysisMean percent improvement in composite endpoint for control groupMean percent improvement in composite endpoint for CRT-D groupPercent greater improvement for CRT-D groupp-value132 15.5 % 24.9 % 9.4 % 0.04616 SORIN – PLATINIUM SonR CRT-D – U463A
 6.  CLINICAL STUDIESSix-month system complication-free rateNumber of patients contributing to analysisKaplan-Meier six-month complication-free estimateOne-sided lower 95% confidence bound for six-month complication-free estimate190 89.5 % 84.1 %6.1.4. Absolute Differences in Peak VO2 and QOLThe tables below show the absolute differences between the control and test groups’ peak VO2 and QOL over the 6 month follow-up period in the clinical trial.Absolute difference between test and control groups’ change in peak V02 over 6 monthsBaseline Mean ± SD (range)6-month Mean ± SD (range)Difference within groupDifference between groupsChange in Peak VO2 (mL/min/Kg)Control group (n=41)13.39 ± 4.58 (5.02, 24.10)13.12 ± 3.99 (3.30, 20.70)- 0.28 1.85Test group (n=91)11.84 ± 3.90 (3.50, 26.3)13.41 ± 4.28 (6.18, 27.67)1.57Absolute difference between test and control groups’ change in QOL score over 6 monthsBaseline Mean ± SD (range)6-month Mean ± SD (range)Difference within groupDifference between groupsChange in QOL Control group (n=41)47.5 ± 19.29 (9, 90.3)31.21 ± 23.96 (0, 95)16.29 1.28Test group (n=91)52.81 ± 21.84 (9, 92)35.24 ± 23.73 (0, 93)17.57The   table   below   presents   the   percentage   of   patients   in   each   group   who   improved, worsened,   or   remained   unchanged   in   each   element   of   the   composite   score   and   the composite score itself.QOL score VO2 Score Composite ScoreControl GROUPTest GROUP Control GROUPTest GROUP Control GROUPTest GROUP% Improved 75.6 74.7 48.8 67.0 62.2 70.9% Worsened 24.4 25.3 51.2 31.9 37.8 28.6% Unchanged0.0 0.0 0.0 1.1 0.0 0.0Histograms for Respiratory Exchange Rate  (RER) at peak VO2  at baseline and 6 month follow-up are provided below:SORIN – PLATINIUM SonR CRT-D – U463A 17
 6.  CLINICAL STUDIES6.1.5. Clinical Results V-V timingV-V programmable settings were available for the clinical study devices as follows: ALTO MSP model 617 (not programmable for V-V delay), ALTO 2 MSP model 627 values (0, 31, 39, 47, 55 and 63 ms) and OVATIO CRT-D 6750 values (0 to 63 ms in steps of 8 ms).The graph below shows the programmed V-V settings at randomization by percentage of patients programmed to each combination of Synchronous BiV pacing and V-V delay.The optimization protocol in the clinical study specified that each patient randomized should undergo  echo guided  V-V  optimization.  Per the  investigational  plan  for the  MSP Clinical 18 SORIN – PLATINIUM SonR CRT-D – U463A
 6.  CLINICAL STUDIESTrial, a uniform protocol was used for V-V programming. This protocol required all patients to undergo   echo-guided   V-V   delay   optimization   before   randomization   (2   to   14   days   post-implant). The optimal V-V delay was determined by finding the programmable V-V delay and ventricular chamber pacing order (RV then LV, or LV then RV) providing the maximum time velocity integral (TVI or VTI) across the left ventricular outflow tract (LVOT).Only those patients randomized to the Test arm were required to be programmed per the optimization protocol for the V-V delay.Of the 177 patients that presented at randomization, 3 had Model 617 which does not have V-V  programmability  hence  the  inability  to  optimize.  Of  the remaining  174   patients,  154 (89%) were tested per the V-V optimization protocol. One hundred forty-nine (149) of the 154 patients who were tested per the V-V optimization protocol were programmed per the recommended or randomized V-V delay (97%). Thirty-one (31) patients were programmed to BiV synchronous (V-V delay 0ms), 46 were programmed to Sequential BiV (LV then RV), 22  were  programmed   to  Sequential  (RV  then  LV),   and the   remaining   50 patients   were randomized to RV only.A sub-analysis  of  the  composite  endpoint  comparing  the   subset of  CRT-D  patients   with optimized V-V delays vs. the subset of patients that did not undergo V-V delay optimization demonstrated similar results in both groups. The CRT-D patients who did not undergo V-V delay optimization showed a smaller improvement in the composite endpoint, although the sample size did not permit conclusions based on data from this subset.SORIN – PLATINIUM SonR CRT-D – U463A 19
 7.  PATIENT SELECTION AND TREATMENT 7. PATIENT SELECTION AND TREATMENT7.1. INDIVIDUALIZATION OF TREATMENTExercise stress testing:If the patient’s condition permits, use exercise stress testing to:─Determine the maximum rate of the patient’s normal rhythm,─Identify any supraventricular tachyarrhythmias,─Identify exercise-induced tachyarrhythmias.The   maximum   exercise   rate  or   the   presence   of   supraventricular   tachyarrhythmias   may influence selection of programmable parameters. Holter monitoring or other extended ECG monitoring also may be helpful.Electrophysiologic (EP) testing:EP testing may be useful for ICD candidates.EP   testing   may   identify   the   classifications   and   rates   of   all   the   ventricular   and   atrial arrhythmias, whether spontaneous or during EP testing.Drug resistant supraventricular tachyarrhythmias (SVTs):Drug   resistant   supraventricular  tachyarrhythmias   (SVTs)   may   initiate   frequent   unwanted device therapy.A careful choice of programming options is necessary for such patients.Antiarrhythmic drug therapy:If the patient is being treated with antiarrhythmic or cardiac drugs, the patient should be on a maintenance   drug  dose   rather   than   a  loading   dose   at  the   time   of   ICD  implantation.   If changes to drug therapy are made, repeated arrhythmia inductions are recommended to verify ICD detection and conversion. The ICD also may need to be reprogrammed.Changes in a patient’s antiarrhythmic drug or any other medication that affects the patient’s normal cardiac rate or conduction can affect the rate of tachyarrhythmias and/or efficacy of therapy.Direct   any   questions   regarding   the   individualization   of   patient   therapy   to   Sorin’s representative.20 SORIN – PLATINIUM SonR CRT-D – U463ACAUTION: To avoid inappropriate therapy during an exercise stress test, do not reprogram any parameter during the test. When a parameter is reprogrammed, algorithm forces acceleration to "ventricular". During conducted sinus tachycardia within the programmed Tachy zone, the device detects a 1:1 fast rhythm. Assuming that acceleration was set to ventricular by reprogramming, the device may identify this as a VT, and may immediately apply the corresponding therapy.
 7.  PATIENT SELECTION AND TREATMENT7.2. SPECIFIC PATIENT POPULATIONSPregnancy:If there is a need to image the device, care should be taken to minimize radiation exposure to the foetus and the mother.Nursing Mothers:Although appropriate biocompatibility testing has been  conducted for this implant device, there has been no quantitative assessment of the presence of leachables in breast milk.Pediatric Patients:This device has not been studied in patients younger than 18 years of age.Geriatric Patients:Most of the patients receiving this device in clinical studies were over the age of 60 years.Handicapped and Disabled Patients:Special care is needed in using this device for patients using an electrical wheel chair or other electrical (external or implanted) devices.SORIN – PLATINIUM SonR CRT-D – U463A 21
 8.  PATIENT COUNSELLING INFORMATION 8. PATIENT COUNSELLING INFORMATIONThe  physician  should  consider  the  following   points   in  counselling   the  patient   about  this device:─Persons administering CPR may experience tingling on the patient’s body surface when the patient’s ICD system delivers a shock.─Advise patients to carry Sorin ID cards and/or ID bracelets documenting their ICD system.22 SORIN – PLATINIUM SonR CRT-D – U463A
 9.  DECLARATION OF CONFORMITY 9. DECLARATION OF CONFORMITYSorin declares that this device is in conformity with the essential requirements of Directive 1999/5/EC   on   Radio   and   Telecommunications   Terminal   Equipment,   with   the   mutual recognition of their conformity (R&TTE).Federal  Communication  Commission  Interference  Statement   47  CFR  Section  15.19 and 15.105(b)The FCC product ID is :─PLATINIUM SonR CRT-D 1811: YSGCRTDSOR1811─PLATINIUM SonR CRT-D 1841: YSGCRTDSOR1841This equipment has been tested and found to comply with the limits for a Class B digital device,   pursuant   to   Part   15   of   the   FCC   Rules.   These   limits   are   designed   to   provide reasonable   protection   against   harmful   interference   in   a   residential   installation.   This equipment generates uses and can radiate radio frequency energy and, if not installed and used   in   accordance   with   the   instructions,   may   cause   harmful   interference   to   radio communications.   However,   there   is   no   guarantee   that   interference   will   not   occur   in   a particular installation.This device complies with Part 15 of the FCC Rules. Operation is subject to the following  two conditions: (1) This device may not cause harmful interference, and (2) this device must accept   any   interference   received,   including   interference   that   may   cause   undesired operation.FCC Interference Statement 47 CFR Section 15.21 - No Unauthorized ModificationsIdentification of the equipment according Section 95.1217(a)This transmitter is authorized by rule under the Medical Device Radiocommunication Service (in part 95 of the FCC Rules) and must not cause harmful interference to stations operating in the 400.150-406.00 MHz band in the Meteorological Aids (i.e., transmitters and receivers used to communicate weather data), the Meteorological Satellite, or the Earth Exploration Satellite   Services   and  must   accept   interference   that  may   be   caused   by  such   stations, including interference that may cause undesired operation. This transmitter shall be used only in accordance with the FCC Rules governing the Medical Device Radiocommunication Service. Analog and digital voice communications are prohibited. Although this transmitter has been approved by the Federal Communications Commission, there is no guarantee that it will not receive interference or that any particular transmission from this transmitter will be free from interference.IC Requirements for CanadaThe FCC product ID is :─PLATINIUM SonR CRT-D 1811: 10270A-CRTDSOR1811─PLATINIUM SonR CRT-D 1841: 10270A-CRTDSOR1841This class B digital apparatus meets all requirements of the Canadian Interference- causing equipment regulations.SORIN – PLATINIUM SonR CRT-D – U463A 23CAUTION: This equipment may not be modified, altered, or changed in any way without signed written permission from SORIN. Unauthorized modification may void the equipment authorization from the FCC and will void the SORIN warranty.
 9.  DECLARATION OF CONFORMITYThis device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.Under   Industry   Canada   regulations,   this   radio   transmitter   may   only   operate   using   an antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry Canada. To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropically radiated power (e.i.r.p.) is not more than that necessary for successful communication.This device may not interfere with stations operating in the 400.150–406.000 MHz band in the Meteorological Aids, Meteorological Satellite, and Earth Exploration Satellite Services and must accept any interference received, including interference that may cause undesired operation.Cet appareil numérique de la classe B respecte toutes les exigences du règlement sur le matériel brouilleur du Canada.Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils radio exempts de licence. L’exploitation est autorisée aux deux conditions suivantes: (1) il ne doit pas produire de brouillage, et (2) l’utilisateur du dispositif doit être prêt a accepter tout brouillage radioélectrique reçu, même si ce brouillage est susceptible de compromettre le fonctionnement du dispositif.Conformément   à   la   réglementation   d’Industrie   Canada,   le   présent   émetteur   radio   peut fonctionner avec une antenne d’un type et d’un gain maximal (ou inférieur) approuvé pour l’émetteur   par   Industrie   Canada.   Dans   le   but   de   réduire   les   risques   de   brouillage radioélectrique à l’ intention d’autres utilisateurs, il faut choisir le type d’antenne et son gain de sorte que la puissance isotrope rayonnée équivalente (p.i.r.e.) ne dépasse pas l’intensité nécessaire à l’établissement d’une communication satisfaisante.Le présent dispositif ne doit pas causer de brouillage aux stations du service des auxiliaires de la météorologie, des satellites météorologiques, du service d’exploration de la terre par satellite, exploitées dans la bande 400,150-406,000 MHz, et il doit accepter tout brouillage reçu, y compris le brouillage pouvant entraîner un mauvais fonctionnement du dispositif.24 SORIN – PLATINIUM SonR CRT-D – U463A
 10.  PHYSICIAN GUIDELINES 10. PHYSICIAN GUIDELINES10.1. PHYSICIAN TRAININGPhysicians should be familiar with sterile pulse generator and left ventricular pacing lead implant procedures. They must apply these procedures according to professional medical training and experience.Physicians should be familiar with follow-up evaluation and management of patients with an implantable defibrillator (or referral to such a physician).This training guideline for implantation and follow-up of ICD and CRT-D devices comes from the  Heart  Rhythm  Society  to  provide  standards  for  hospital  credentialing  bodies  to  help ensure appropriate patient care and lead to improved patient outcomes. The following is a summary   of   requirements   for   an   alternate   training   pathway   for   ICD   and   CRT-D implantations(1):─Documentation of current experience: 35 pacemaker implantations per year and 100 implantations over the prior 3 years─Proctored ICD implantation experience: 10 Implantations, 5 Revisions─Proctored CRT-D implantation experience: 5 implantations─Completion of didactic course and/or IBHRE® ExAM─Monitoring of patient outcomes and complication rates─Established patient follow-up─Maintenance of competence: 10 ICD and CRT-D procedures per year, 20 patients per year in follow-up(1) Please consult full text of both publications for details. 2004 Heart Rhythm Society Clinical  Competency Statement and the 2005 Addendum on Training Pathways for Implantation of  Cardioverter Defibrillators and Cardiac Resynchronization Devices. Heart Rhythm (2004) 3,  371-375; Heart Rhythm.10.2. DIRECTIONS FOR USEICD operating characteristics should be verified at the time of implantation and recorded in the patient file. Complete the Patient Registration Form and return it to Sorin, as it provides necessary information for warranty purposes and patient tracking.Additional programming instructions can be found by accessing Online Help (click the “?” on the   screen)   on   the   Sorin   dedicated   programmer.   Paper   copies   of   Online   Help   can   be obtained by contacting your Sorin representative.10.3. MAINTAINING DEVICE QUALITYThis device is FOR SINGLE USE ONLY. Do not resterilize and reimplant explanted ICDs.Do not implant the device when:─It has been dropped on a hard surface because this could have damaged pulse generator components.─Its sterility indicator within the inner package is not green, because it might not have been sterilized.SORIN – PLATINIUM SonR CRT-D – U463A 25
 10.  PHYSICIAN GUIDELINES─Its storage package has been pierced or altered, because this could have rendered it non-sterile.─It has been stored or transported outside the environmental temperature limits: 32 °F (0 °C) to 122 °F (50 °C) as an electrical reset condition may occur.─"Use Before" date has expired, because this can adversely affect pulse generator longevity or sterility.10.4. V-V PROGRAMMING RECOMMENDATIONIt is recommended that V-V optimization testing be performed and used to set the V-V delay for this device to optimize the potential for RF SonR CRT-D benefit to the patient.26 SORIN – PLATINIUM SonR CRT-D – U463A
 11.  PATIENT INFORMATION 11. PATIENT INFORMATIONInformation for the patient is available in the patient booklet, contained in the outer storage package. Additional copies can be obtained by contacting your Sorin representative.This information should be given to each patient with their first ICD and offered to the patient on each return visit or as deemed appropriate.SORIN – PLATINIUM SonR CRT-D – U463A 27
 12.  HOW SUPPLIED 12. HOW SUPPLIED12.1. STERILITYThe PLATINIUM defibrillators are supplied one per package in a sterile package.12.2. WARRANTY AND REPLACEMENT POLICYSorin warrants its defibrillators. Refer to the section "Warranty" for additional information. Please see the following labelling sections for information concerning the performance of this device: Indications, Contraindications, Warnings and Precautions, and Adverse Events.28 SORIN – PLATINIUM SonR CRT-D – U463A
 13.  DEVICE DESCRIPTION 13. DEVICE DESCRIPTIONThe  PLATINIUM   SonR  CRT-D   ICD device   and   programming   system.  The  programming system   includes   the   Sorin   dedicated   programmer   with   the   SMARTVIEW   programming software connected to a CPR3 programming head. The programming system is configured and furnished by Sorin.The PLATINIUM SonR CRT-D can serve as a defibrillation electrode (active housing) with a total surface area of 63 cm².The PLATINIUM SonR CRT-D is designed to recognize and treat slow or fast VT and VF by continuously   monitoring   atrial   and   ventricular   activity   to   identify   persistent   ventricular arrhythmias  and  to   deliver  appropriate  therapies.  PLATINIUM  SonR  CRT-D  features  the PARAD/PARAD+   algorithm,   which   is   specifically   designed   to   differentiate   ventricular tachycardias   from   fast   rhythms  of  supraventricular   origin.   PARAD/PARAD+   continuously monitors   R-R   interval   stability,   searches   for   long   cycles,   assesses   the   degree   of   P-R association,   evaluates   sudden   onset   and   determines   the   chamber   of   arrhythmia acceleration.In   addition   to   the   advanced   detection   scheme,   PLATINIUM   SonR   CRT-D   offers programmable single, dual or triple-chamber pacing therapy (DDD, DDI, VVI or SafeR (AAI <>   DDD)   modes)   with   or   without   rate-responsive   capabilities   (DDDR,   DDIR,   VVIR, DDD/DDIR   and   SafeR-R   (AAIR   <>   DDDR)   modes)   using   an   acceleration   sensor.  An automatic AV delay algorithm as well as a mode switching function are available.PLATINIUM SonR CRT-D enables an adjustment of the interventricular delay, and provides the possibility of adapting pacing to each ventricle. The ICD is intended to resynchronize uncoordinated   contraction   of   the   heart   by   simultaneously   or   sequentially   pacing   both ventricles.PLATINIUM   SonR   CRT-D   offers   tiered   therapy.   Therapies   can   be   programmed independently in each zone:─in the Slow VT and VT zones: two ATP programs, up to two shocks with programmable energy and up to four shocks with maximum energy can be programmed;─in the VF zone: one ATP program, up to two shocks with programmable energy and up to four shocks with maximum energy can be programmed.The ATP can be applied in RV, LV or RV and LV pacing with a VV delay equal to 0 ms. ATP pacing configuration is independent of ventricular pacing configuration.When   the   rhythm   changes   from   one   zone   to   another,   the   device   delivers  the   therapy programmed in this zone, starting with the same or more aggressive program for the area. The ATP program in the VF zone will only be applied if the VT coupling interval is longer than the programmed fast VT cycle length.The PLATINIUM SonR CRT-D offers biphasic shocks with a maximum stored energy of 42 J. The   shock   configuration   (electrodes   used   to   apply   the   shock)   can   be   chosen   by programming one of the following combinations: can and one coil, can and two coils, two coils only.Other features are as follows:─Automatic ventricular sensitivity control─Non-committed shocks─Electrophysiological studies (EPS) with real-time markers or electrograms: SORIN – PLATINIUM SonR CRT-D – U463A 29
 13.  DEVICE DESCRIPTION─Programmer-controlled VT induction sequences,─Programmer-controlled VF inductions (30 Hz rapid pacing or shock on T),─Programmable electrogram vectors (A / RV coil-CAN / SVC coil-CAN /RV coil-SVC coil / LV bip / LV tip-RV ring / A ring-CAN / RV tip-CAN / RV ring-CAN / LV tip-CAN / LV ring-CAN) and RV EGM, ─Real-time annotations displayed with the markers and indicating the majority rhythm,─Manual ATP sequences,─Manual shocks.─Rescue shock─Follow-up tests: ─Pacing lead impedance,─Coil impedance,─Capacitor charge time,─Sensitivity test,─Pacing threshold tests.─Data storage: ─Therapy History Report,─Statistics (pace/sense, therapy, shocks, and battery voltage),─Up to 16 complete Holter records with event logs, marker channel notation, and electrogram records.The PLATINIUM SonR CRT-D 1811 connector has five ports:─Atrial “SonR” port: performs atrial bipolar pace/sense if a conventional IS-1 lead is connected as well as SonR capability if the SonR atrial lead is connected,─RV “IS-1” port: performs right ventricular bipolar pace/sense,─LV “IS-1” port: performs left ventricular bipolar pace,─RV “DF-1” port for RV defibrillation coil,─SVC “DF-1” port for SVC defibrillation coil.The PLATINIUM SonR CRT-D 1841 connector head has three ports:─Atrial “SonR” port: performs atrial bipolar pace/sense if a conventional IS-1 lead is connected as well as SonR capability if the SonR atrial lead is connected,─LV “IS-1” port: performs left ventricular bipolar pace,─RV “DF4” port: performs right ventricular bipolar pace/sense, port for RV/SVC defibrillation coils.Distal   lead   terminal   connections   are   secured  with   set-screws   accessed   via   self-sealing silicone   plugs.   All   lead   connections   pass   through   the   header   into   the   device   via feedthroughs.Programming System:The   Sorin   programmer   is   used   in   conjunction   with   specific   programmer   software   to interrogate   and   program   the   implanted   device   at   implant   and   during   patient   follow-up procedures.Remote Monitoring:PLATINIUM SonR CRT-D is also equipped with the RF wireless technology which enables to remotely monitor the patients who have the Sorin SMARTVIEW Monitor installed at home.30 SORIN – PLATINIUM SonR CRT-D – U463A
 14.  IMPLANT PROCEDURE 14. IMPLANT PROCEDURE14.1. NECESSARY EQUIPMENTImplantation of PLATINIUM SonR CRT-D requires the following equipment:─Sorin ORCHESTRA programmer, equipped with the SMARTVIEW software interface and inductive telemetry head,─Sorin ORCHESTRA PLUS programmer, equipped with the SMARTVIEW software interface, inductive telemetry head and optionally ORCHESTRA PLUS LINK,─pacing system analyzer, as well as its sterile connecting cables, to evaluate the pacing and sensing thresholds,─a complete set of leads with corresponding introducers,─physiological signal monitor capable of displaying simultaneously the surface ECG and arterial pressure,─an external defibrillator with sterile external paddles,─sterile cover for the telemetry head.14.2. PACKAGING14.2.1. ContentsPLATINIUM SonR CRT-D and its accessories are ethylene oxide sterilized and hermetically sealed in two-ply clear packaging meeting international requirements.The sterile packaging contains:─the defibrillator─a ratcheting screwdriver─a DF-1 defibrillating connector insulating plug for 1811 modelOnce delivered,  PLATINIUM  SonR CRT-D  is programmed  to  as-shipped  values that  are different from nominal values (see Chapter “Programmable Parameters” for details).14.3. OPTIONAL EQUIPMENTThe following equipment may be required during implantation of PLATINIUM SonR CRT-D:─an IS-1 insulating plug to close the atrial port─a DF4/DF-1 adaptor in case of replacement and use of DF-1 lead in a DF4 connector─sterile water to clean traces of blood. Any parts cleaned with sterile water must be thoroughly dried.─mineral oil to lubricate if necessary─a lead cap to isolate a lead which is not usedSORIN – PLATINIUM SonR CRT-D – U463A 31NOTE: In case you’re implanting a DF4 lead, please verify its compatibility with standard alligators pin; please refer to the lead user’s manual for more details.
 14.  IMPLANT PROCEDURE14.4. BEFORE OPENING THE PACKAGEBefore opening the package, check the "Use Before" date printed on the labels on the box  and  on the  sterile  package. Defibrillators that  have not  been implanted before that  date should be returned to Sorin.Interrogate the device: ─if a warning is displayed, do not implant the device and contact your Sorin representative.─if battery voltage is below 3V, and if the last reforming/charge occurred more than one week ago, do not implant the device. Otherwise wait for one more week before checking the voltage.NOTE : The battery voltage can decrease before the expiration date is reached. However, the battery voltage should be equal to or higher than 3V at the time of implant.Devices MUST NOT be interrogated and programmed within the vicinity of other devices.Also check  the integrity  of the  sterile package.  The  sterility of  the  contents is  no longer guaranteed if the package has been pierced or altered. If the defibrillator is no longer sterile, it  should   be  returned  in  its  packaging   to  Sorin.  Any re-sterilization  of   the  unit   is  at   the discretion of Sorin.14.5. PRIOR TO IMPLANTATIONUse the programmer to verify the defibrillator can be interrogated before implantation.Verify   all   shock   therapies   are   disabled   in   order   to   avoid   accidental   discharge   during implantation.It   is   not   advisable   to   program   the   Smoothing   function   before   implantation,   since   the defibrillator may detect noise and pace at a rate higher than the programmed basic rate.14.6. DEVICE PLACEMENTThe   pocket   should   be   prepared   in   the   left   pectoral   position,   either   subcutaneously   or submuscularly.   Subcutaneous   device   implantation   is   recommended   for   optimal   RF communication efficacy.Implantation in an abdominal position is not advisable.In its final position, the defibrillator should be no more than 4 cm below the skin surface.14.7. CHOOSING THE TYPE OF LEADThe defibrillator should be connected to:─one bipolar atrial sensing/pacing lead with or without dedicated SonR sensor─one right ventricular lead with bipolar sensing/pacing electrodes and one or two 32 SORIN – PLATINIUM SonR CRT-D – U463ACAUTION:Do not shake or tap sharply on the ICD package with the ICD inside, because the ICD's sensing circuits can interpret this as P-waves or R-waves and record these as an arrhythmia episode.High voltage capacitors charge performed on ICD without connected leads using wireless telemetry can generate false P-waves or R-waves detection. It is recommended to reset the memory data and statistics before implanting the ICD.
 14.  IMPLANT PROCEDUREdefibrillation coils─one unipolar or bipolar left ventricular pacing leadThe choice of leads and their configuration is left to the implanting physician’s judgment.Replacement of an existing atrial lead:If the initial lead which was connected to the atrial port of the PLATINIUM SonR CRT-D was  not a SonR lead (i.e. a conventional lead or plug), then do not implant a SonR lead as replacement; possible blood infiltration at the entrance of the connector port may prevent proper measurement of the SonR signal.Connectors:PLATINIUM SonR CRT-D 1811:The   pacing/sensing   connectors   are   compatible   with   the   IS-1   standard   and   the   right ventricular defibrillation connectors are compatible with DF-1 standard.PLATINIUM SonR CRT-D 1841:The  atrial  and left  ventricular connectors  are  compatible  with the IS-1  standard  and  the quadripolar right ventricular connector is compatible with the DF4 standard.14.8. SHOCK CONFIGURATION (+ -> -)The shock configuration is the energy pathway between the defibrillation electrodes. If an atrial coil is present, the shock configuration can be programmed for bi-directional shocks.Programming:When active case and SVC are both programmed to Yes, the shock configuration can be programmed to: ─RV to Case (or Case to RV),─or RV to SVC (or SVC to RV),─or RV to Case+SVC (or Case+SVC to RV).SORIN – PLATINIUM SonR CRT-D – U463A 33NOTE1: Please note that DF-1 standard compliant lead is not compatible with DF4 connector. In the reverse, DF4 standard compliant lead is not compatible with DF-1 connector. In case of defibrillator replacement, choose the appropriate device compatible with DF-1 or DF4 leads. For any other lead type that requires an adaptor for this device, please contact your Sorin representative for any information on lead / connector compatibility question.NOTE2: In the event that no atrial lead is implanted, the atrial port should be plugged with IS-1 insulating plug and a single chamber mode (VVI-VVIR) should be programmed. PARAD and PARAD+ should not be used.NOTE3: In the event of a warning on a low shock impedance, and after lead replacement or reconnection: it is recommended to check the system integrity (sensing and pacing thresholds and the impedance of the shock electrodes).
 14.  IMPLANT PROCEDURE        RV to Case+SVC RV to Case RV to SVCThe polarity of shock is determined by the parameter itself.14.9. MEASUREMENT OF THRESHOLDS AT IMPLANTPacing and sensing thresholds should be measured at implant.Pacing thresholds:Acute thresholds should be lower than 1 V (or 2 mA) for a 0.35 ms pulse width, in both ventricles and in the atrium.Sensing thresholds:For appropriate right ventricular  sensing, the amplitude of the R-wave should be  greater than 5 mV.For appropriate atrial sensing, the amplitude of the P-wave should be greater than 2 mV.Pacing impedance measurements:Right ventricular,  left ventricular  and atrial  pacing impedances  should range from  200 to 3000 ohms (refer to the lead characteristics, especially if high impedance leads are used).Please   refer   to   the   leads   user   manuals   for   more   details   on   the   expected   electrical performances of the leads.14.10. LEADS CONNECTIONImplant the ventricular leads, then the atrial lead.Each lead must be connected to the corresponding connector port. The position of each connector is indicated on the casing.Tighten only the distal inserts. To connect each lead, proceed as follows:1. Clean the lead terminal pins thoroughly, if necessary (device replacement).2. Lubricate the lead terminal pins with sterile water, if necessary.3. Do not insert a lead connector pin into the connector block without first visually verifying that the lead port is not filled with any obstacle.4. Insert the screwdriver into the pre-inserted screw socket of the appropriate port (in order to allow excess air to bleed out and to make the insertion of the lead pin easier).5. Insert the lead pin all the way into the port (check that the pin protrudes beyond the distal  insert).34 SORIN – PLATINIUM SonR CRT-D – U463ACAUTION:
 14.  IMPLANT PROCEDURE6. Tighten, check the tightness and ensure the lead pin still protrudes beyond the distal insert, and has not move.1. Do not tighten the pre-inserted screws when there is no lead (this could damage the connector). 2. Do not loosen the screws before inserting the connector (subsequent risk of being unable to reinsert the screw).3. When mineral oil or sterile water is used to make lead insertion easier, the screwdriver should remain inserted into the pre-inserted screw socket when checking the tightness. As a matter of fact, when the lead port is filled with a liquid, the physics piston effect can  give the feeling the lead is properly tightened. 4. One single set screw is located on the side of the connection header.5. Use  only  the  screwdriver  provided   with  the defibrillator.   Keep  the  screwdriver's shaft perpendicular to the plane of the defibrillator (see figure below).6. Removing the screwdriver: to avoid all risk of loosening screws during removal, hold the screwdriver by its metal part and not by the handle.To ensure full insertion, push the screwdriver's hex tip smoothly into the setscrew until it reaches the bottom of the hex chamber in the screw, which can be felt as a solid metallic contact. Do not implant the defibrillator if there is no feeling of solid metallic contact. Do not implant the defibrillator if the wrench does not click when attempting to tighten the setscrew on the lead pin.In the  case of an  external defibrillation  shock delivered to the patient, always check the programming and functioning of the device, in particular its capacity to deliver shocks.SORIN – PLATINIUM SonR CRT-D – U463A 35CAUTION:WARNING: Ensure that the screwdriver's tip is fully inserted in the setscrew; otherwise the screwdriver can damage the setscrew and prevent connection with or disconnection from the lead.NOTE: To optimize cardioversion/defibrillation shocks, electrodes must be positioned so that the electric field between anode(s) and cathode covers the largest myocardial mass. In normal conditions, the anode and cathode are adequately separated. In case of a short-circuit, the shock may be aborted to prevent damaging the defibrillator.
 14.  IMPLANT PROCEDURE14.11. DEVICE IMPLANTATIONPLATINIUM SonR CRT-D should be implanted with the device identification engraved side facing outwards for optimal communication with the programming head and radiographic identification.Carefully wind excess lead and place in a separate pocket to the side of the defibrillator.It is recommended not to place any excess wire between the can and the heart.Suture the casing connector to the muscle using the hole provided for this purpose, in order to avoid potential migration of the device into the pectoral muscle.14.12. TESTS AND PROGRAMMINGDuring the implant testing procedure:It is recommended that a security margin of at least 10 J be demonstrated between the effective shock energy and maximum programmable energy.Enable shock therapies, then program the defibrillator.Verify that the defibrillation lead impedance for each shock delivered is within the range of 30 to 150 ohms. Check the lead connection if the values are outside these boundaries.Save the  programming  data on  the programmer’s hard disk and on an  external storage device (if desired).Resuscitation Availability:Do   not   perform   device   testing   unless   an   external   defibrillator   is   available   and  medical personnel skilled in cardiopulmonary resuscitation (CPR) are present.Disable the ICD During Handling:Program   Shock   Therapy   to   OFF   during   surgical   implant   and   explant   or   post   mortem procedures. The device can deliver a serious high energy shock should accidental contact be made with the defibrillation electrodes, the device can deliver a very high energy shock.36 SORIN – PLATINIUM SonR CRT-D – U463A
 15.  SPECIAL MODES 15. SPECIAL MODES15.1. SAFETY MODE (NOMINAL VALUES)Nominal   values   may   be   rapidly   restored   by   pressing   the   following   button   on   the programming head or programmer keyboard:or via the Emergency button on the SMARTVIEW screen.In  safety  mode,  the  defibrillator  operates  with  the  parameters  underlined in  the  table  of programmable parameters.15.2. MAGNET MODEWhen the magnet is applied:─antiarrhythmia functions are inhibited (detection of rhythm disturbances, charging, and therapy),─hysteresis, VV delay and AVD paced/sensed offset are set to 0,─pacing amplitude is set to 6 V,─pulse width is set to maximum,─the following functions are disabled: ventricular arrhythmia prevention, Mode Switch, Anti-PMT, Smoothing, Rate Response.When the magnet is removed:─arrhythmia detection algorithms and sequential therapies are reinitialized,─therapies start with the least aggressive program for each area.The   antiarrhythmia   functions   inhibition   is   extended   after   magnet   removal   if   a   charge occurred  just  before  the  application   of  the   magnet  in   order  to  ease  the  communication between the device and the programmer.The other parameters remain at their programmed value, including the ventricular paced chamber parameter.15.3. RESPONSE IN THE PRESENCE OF DISTURBANCEIf the  defibrillator  senses  electrical noise  at a frequency above  16  Hz, it  switches  to an asynchronous mode at the basic rate. The programmed mode is restored as soon as the noise is no longer detected.Ventricular pacing may also be inhibited by ventricular noise. It can be restored by setting the parameter V pacing on noise to Yes.15.4. DETECTION CHARACTERISTICS IN THE PRESENCE OF ELECTROMAGNETIC FIELDSPer Clause 27.4 of Standard EN 45502-2-2, detection in the presence of electromagnetic fields is characterized as follows:─Differential mode:For DF-1 models:SORIN – PLATINIUM SonR CRT-D – U463A 37
 15.  SPECIAL MODESFor DF4 models:─Common mode rejection ratio:For DF-1 Models:16.6 Hz 50 Hz 60 HzAtrial channel ≥74 dB ≥74 dB ≥74 dBVentricular channel ≥68 dB ≥68 dB ≥68 dBFor DF4 Models:16.6 Hz 50 Hz 60 HzAtrial channel ≥74 dB ≥74 dB ≥74dBVentricular channel ≥68 dB ≥68 dB ≥68dBFor atrial sensitivity settings below  0.4mV, the ICD may detect noise lower than the level specified in clause 27.5.1 of standard EN 45502-2-2 for frequencies below 200 Hz.For ventricular sensitivity settings below  0.6mV, the ICD may detect noise lower than the level specified in clause 27.5.1 of standard EN 45502-2-2 for frequencies below 200 Hz.38 SORIN – PLATINIUM SonR CRT-D – U463A
 15.  SPECIAL MODES15.5. PROTECTION AGAINST SHORT-CIRCUITSThe defibrillator can undergo a short-circuit if the anode and cathode are not adequately separated.In this case, the shock is aborted to prevent damaging the defibrillator and a warning will indicate  that  a short  circuit  (shock impedance  <  20 ohms) was  detected  during the last shock.SORIN – PLATINIUM SonR CRT-D – U463A 39
 16.  MAIN FUNCTIONS 16. MAIN FUNCTIONS16.1. AUTOMATIC LEAD MEASUREMENTSAutomatic pacing lead impedance measurement:A lead impedance measurement is automatically performed on atrial and ventricular leads every 6 hours. The daily mean impedance is stored for each chamber.Automatic coil impedance measurement:A continuity measurement is automatically performed on defibrillation coil(s) once per day. The continuity is stored for each coil.Automatic sensing measurement:The  amplitude of  P  and  R waves  are  automatically  measured at  each cycle. Every 8.5 minutes, the amplitude of the last 8 P and R detections are averaged and stored.16.2. ATRIAL TACHYARRHYTHMIA MANAGEMENTMode Switch:This function  is designed to  limit the acceleration and variation of  ventricular rate in  the presence of atrial arrhythmia.16.3. VENTRICULAR TACHYARRHYTHMIA MANAGEMENTVentricular tachyarrhythmia prevention:Set of algorithms that can be used to avoid the circumstances of ventricular tachyarrhythmia onset.Arrhythmia discrimination algorithm PARAD and PARAD+ (P And R based Arrhythmia Detection):PARAD is the algorithm used to discriminate sinus tachycardias (ST) and supraventricular tachycardias (SVT) from ventricular tachycardias (VT).PARAD+   is   based   on   the   PARAD   algorithm   but   additionally   takes   into   account   the “AFdetect” discrimination criteria: the occurrence of a "long ventricular cycle" characteristic for   AF   patients   which   is   an   additional   arrhythmia   classification   criterion   to   improve identification of atrial fibrillation and avoid inappropriate shocks.Fast VT treatment:Applies detection criteria on fast ventricular tachycardia that are different from those of the VT zone, as well as different therapies. The fast VT zone is included in the VF zone: its lower limit is determined by the programmed value for the VF zone and its upper limit by the  programmed value for the fast VT zone.Polarity alternation on Max shock:Reverses  the  programmed polarity of  every  second  shock set at maximum  energy.  The number, type, and energy of shocks is independently programmable by detection zone.40 SORIN – PLATINIUM SonR CRT-D – U463A
 16.  MAIN FUNCTIONSDefibrillation threshold (DFT):Be aware that the changes in the patient’s condition, drug regimen, and other factors may change   the   defibrillation   threshold   (DFT)   which   may   result   in   non-conversion   of   the arrhythmia  post-operatively.   Successful  conversion   of ventricular   fibrillation  or   ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post-operatively.16.4. PACINGBTO (Brady Tachy Overlap):Enables   cardiac   resynchronization   therapy   within   the   slow   VT  zone   to   preserve   patient exercise capacity, without affecting detection or treatment of slow VTs.Post-shock mode:After any automatic shock therapy, the post-shock mode makes it possible to apply a pacing mode  other than  the  standard  antibradycardia  pacing  mode  and/or  with  different pacing parameters.SafeR (AAI <> DDD) mode:Is intended to minimize deleterious effects of ventricular pacing. The defibrillator functions in AAI mode, and temporarily switches to DDD mode upon the occurrence of AVB III, AVB II, AVB I and ventricular pause.Anti-PMT protection:Is  intended   to  protect   the   patient   from   Pacemaker-Mediated  Tachycardia   (PMT)  without reducing atrial sensing capability of the device.16.5. SENSINGAutomatic Refractory Periods:Optimize sensing and make the implant programming easier. These periods are composed of  a minimal Refractory  Period and a  triggerable  Refractory  Period. The  duration of  the refractory periods lengthens automatically as needed.Committed period:In  DDI  or DDD  modes,  the  committed   period  is  a  non-programmable  95  ms ventricular relative refractory period that starts with atrial pacing. If a ventricular event is sensed during the committed period, but outside the blanking period, the ventricle is paced at the end of the committed period. The committed period prevents inappropriate ventricular inhibition if crosstalk occurs. Protection against noise:Allows   the   distinction   between   ventricular   noise   and   ventricular   fibrillation.   If   the   device senses  ventricular  noise, the  ventricular  sensitivity  is decreased  until  noise  is  no  longer detected. Ventricular pacing can be inhibited to avoid a potential paced T-wave.Automatic sensitivity control:Optimizes arrhythmia detection and avoids late detection of T-waves and over-detection of wide QRS waves. The device automatically adjusts the sensitivities based on the ventricular sensing amplitude. In case of arrhythmia suspicion or after a paced event, the programmed ventricular sensitivity will be applied. The minimum ventricular sensitivity threshold is 0.4 mV (minimum programmable value).SORIN – PLATINIUM SonR CRT-D – U463A 41
 16.  MAIN FUNCTIONS16.6. SONR CRT OPTIMIZATIONSonR   is   a   specific   sensor,   located   at   the   tip   of   the   atrial   lead,   that   picks-up   micro-accelerations of the heart walls to derive information pertaining to cardiac contractility. The signal picked-up  by the SonR  sensor can  be processed by PLATINIUM  SonR CRT-D  in order to automatically adjust VV delays and AV delays during rest and exercise for optimal resynchronization therapy. In addition, the SonR signal is recorded during tachyarrhythmia episodes to depict acute variations  of cardiac contractility.  PLATINIUM SonR CRT-D can also transmit real-time SonR signals via telemetry.16.7. FOLLOW-UP FUNCTIONStorage of memory data:AIDA+ (Automatic Interpretation for Diagnosis Assistance) software provides access up to 6 months of patient follow-up with day by day data collection, or up to 24 hours with hourly data   collection.   Episodes   of   ventricular   tachyarrhythmia   are   recorded   with   one programmable EGM channel which can be selected, in addition to RV EGM.Alerts / Warnings:The device routinely performs security self-checks and technical measurements to ensure system integrity. When system integrity is found to be at risk outside a follow-up, alerts are stored in the device memory. When system integrity is found to be at risk during a follow-up, the information is managed by a warning (pop-up message) to immediately notify the user. For   example,   the   following  types  of   event   can  trigger   a  warning   or   an  alert:   technical problem during a shock, lead impedance or shock continuity measurements out-of-range, battery depletion, …16.8. REMOTE MONITORING FUNCTIONRemote   monitoring   enables   the   automatic   remote   transmission   of   implant   data   to   the physician thanks to the wireless Radio Frequency (RF) communication ability of the implant in order to provide a comprehensive report to the physician about device functioning and patient cardiac status without having the patient physically in the clinic.The data is transmitted from the implant to the SMARTVIEW  monitor, a small transmitter  placed in the patient's home.Implant data are first transmitted to the SMARTVIEW monitor via RF. Data are then rooted through the phone line or via GPRS to an internet website. This website is responsible for transforming the  implant data  into a comprehensive  report that  can be  consulted by the physician.16.8.1. SMARTVIEW MonitorThe SMARTVIEW monitor is a small device equipped with an RF transmission module to communicate with the implant and a modem to export data through the internet.The SMARTVIEW monitor is delivered to the patient who has to install it at home. Preferably the   SMARTVIEW   monitor   will   be  placed   on   the  nightstand  of   the  patient,   as   close   as possible to the side of the bed where the patient usually sleeps. The SMARTVIEW monitor connects to the phone line of the patient and the power plug. Regular transmissions are done during the night when the patient is asleep next to the SMARTVIEW monitor without any intervention from the patient. 42 SORIN – PLATINIUM SonR CRT-D – U463A
 16.  MAIN FUNCTIONS16.8.2. Transmission triggerThere are 3 different triggers for a remote transmission:─the remote follow-up transmission is scheduled by the physician to occur regularly (according to the programming).─the alert transmission will take place when the implant has recorded an abnormal event. The list of abnormal event is available in a following paragraph. Alert conditions are checked daily.─the on-demand follow-up transmission is triggered by the patient himself through the use of a specific button on the SMARTVIEW monitor.16.8.3. Data transmittedThe data transmitted are identical to the data available during a standard interrogation with the dedicated programmer. All counters, histograms, IEGMs and diagnosis available in the device are transmitted containing (not exhaustive list):─programmed parameters─information on patient and system implanted─battery status─lead status (brady leads and defibrillation coils) ─pacing counters and mean heart rate (brady)─atrial and ventricular arrhythmia counters and episodes─ventricular therapy countersData   are   presented  in  the  form   of   2   reports  to  the  physician:   the   first   one  contains  a summary of major counters, histograms, warnings and diagnosis. The second one presents the most important IEGM episodes automatically selected based on the degree of severity for the patient.16.8.4. User websiteOn the website, the physician is able to:─consult and schedule the remote follow-ups of their patient─configure additional ways of being notified of alerts (for instance by SMS, fax or e-mail)─consult, print and export patient reports 16.8.5. Alert systemThe   following   set   of   alert   trigger   can   be   independently   programmed   ON/OFF   by   the physician using the dedicated programmer and can trigger an alert transmission:─Low or high impedance (A, RV, LV)─Abnormal coil impedance (shock lead)─Low or High shock impedance─Inefficient high energy shock─All shocks programmed OFF─Shock treated VT/VF─ATP treated VT/VF─Lack of V pacing in CRT device─Suspicion of noise on the V leadSORIN – PLATINIUM SonR CRT-D – U463A 43
 16.  MAIN FUNCTIONS─AT/AF occurrence─Fast V rate during AT/AFThe following set of alert trigger (system alerts) cannot be deactivated when the Alerts are programmed “On” and can trigger an alert transmission:─Battery depletion – RRT─Device reset─Excessive charge time (>25s)─System integrity44 SORIN – PLATINIUM SonR CRT-D – U463AWARNING: The use of remote monitoring does not replace regular follow-up. Therefore, when using remote monitoring, the time period between follow-ups visits may not be extended.
 17.  PATIENT FOLLOW-UP 17. PATIENT FOLLOW-UP17.1. FOLLOW-UP RECOMMENDATIONSBefore the patient is discharged and at each subsequent follow-up visit, it is advisable to:─check the occurrence of system warnings─check the battery status,─check the integrity of the pacing and defibrillation leads,─check for proper sensing (sensitivity, crosstalk) and pacing ; set the pacing amplitude to twice the pacing threshold,─interrogate the implant memories (AIDA+),─check the efficacy of the therapies delivered,─keep a printout of programmed parameters, test results, and memory data,─reset the memory data and statistics.These operations should be performed by medical personnel in an appropriate care unit, with resuscitation equipment present.It is recommended that a routine follow-up examination be done one month after discharge, and then every three months until the device nears the replacement date.Refer to the online help for a description of displayed warning, and the necessity to contact Sorin for an evaluation.Implant software upgrade:In   case   a   new   implant   software   is   downloaded   in   the   device   memory   through   the programmer, a warning message could be displayed by the programmer to inform the user and give the correct instructions to follow.17.2. HOLTER FUNCTIONThe Holter records markers and EGM on RV and on 1 programmable channel: A / RV coil – Can / SVC coil – Can / RV coil - SVC coil / LV bip / LV tip – RV ring / RA ring – Can / RV tip – Can / RV ring – Can / LV tip – Can / LV ring – Can: ─Up to 10 episodes and 5 min EGM on significant events: AV block switch, lead impedance out of range. ─Up to 16 tachyarrhythmia episodes as well as the therapy history.SORIN – PLATINIUM SonR CRT-D – U463A 45NOTES: If the last reforming, charge or shock occurred during the week preceding the interrogation, the last battery value may be still impacted by the event. One week post event, the battery will recover its steady state value.Automatic capacitor charging may affect communication between the device and the programmer.
 17.  PATIENT FOLLOW-UPStored Tachyarrythmia Episodes:PLATINIUM   SonR   CRT-D   stores   up   to   16   episodes   (VF,   VT,   Slow   VT,   SVT/ST,   non-sustained) with a total of 25.6 min of high resolution EGM and 5 min of SonR signal.For each episode four levels of details are presented:─Tachogram (to visualize PP and PR intervals)─Event log for the entire episode: ─PARAD/PARAD+ analysis for each majority,─Delivered therapies,─Markers: Atrial, ventricular and biventricular markers, sensed, paced and in relative refractory periods,─EGM: onset and detection of the arrhythmia, on two therapies, and the return to slow rhythm by recording electrogram.Therapy historyFor each  arrhythmia detection,  each therapy delivered  (either automatically or  during an electrophysiological study)  and at the end of each arrhythmia, PLATINIUM SonR CRT-D records the type of majority rhythm, the number of ATP sequences delivered, the energy and the number of shocks delivered.17.3. RECOMMENDED REPLACEMENT TIME (RRT)Recommended Replacement Time (RRT)(1) is controlled by: battery voltage equal to 2.62 V ± 0.01 VBetween   the   RRT  and   the   EOS   (End   of   Service)(2),  PLATINIUM   SonR   CRT-D   can   still function for:─10.8 months (100% atrial and biventricular pacing in DDD mode, 500 ohms, with as-shipped settings), and deliver 11 shocks at 34 J or─10,4 months (0% pacing, sensors OFF, one shock every 2 weeks) and deliver 21 shocks at 34 J. Once  the  Recommended  Replacement  Time  (RRT)  point  has  been reached,  the  device operates  normally,  except  that the  charge  time increases.  Under normal conditions (and without programmer use) the charge times are as follows:Shock energy Charge time (sec)BOS(3) 42 J 10 (± 2)RRT 42 J 13 (± 3)(1)   Recommended   Replacement   Time   (RRT)   corresponds   to   Elective   Replacement Indicators (ERI) previously used.(2) End of Service (EOS) corresponds to End of Life (EOL) previously used.(3) Beginning of Service (BOS) corresponds to Beginning of Life (BOL) previously used.17.4. EXPLANTATIONThe defibrillator should be explanted in the following cases:46 SORIN – PLATINIUM SonR CRT-D – U463ACAUTION: The defibrillator should be replaced as soon as the Recommended Replacement Time (RRT) point is reached.
 17.  PATIENT FOLLOW-UP─The Recommended Replacement Time (RRT) point is reached─Confirmed malfunction─Burial of the patient (for environmental reasons, the local regulation may require the explantation of the devices containing a battery supply)─Cremation of the patient (the defibrillator may explode if placed in an incinerator)The explanted defibrillator should not be reused in another patient.All explanted  defibrillators  should  be returned to  Sorin, carefully  cleaned  of all  traces  of contamination.   Cleaning   may   be   done   by   immersing   them   in   an   aqueous   sodium hypochlorite containing at least 1% chlorine, followed by rinsing copiously with water.The   defibrillator   should   be   protected   against   mechanical   impact   and   the   temperature variations that may occur during shipping.Before explantation, it is advisable to:─Print out all programmed parameters, statistics and Holter function report,─Save Patient data on floppy disk or hard disk,─Disable shock therapies (VT and VF) to avoid any risk of untimely shock.17.5. DEFIBRILLATOR IDENTIFICATIONThe defibrillator can be interrogated and programmed via telemetry, using the programming head interfaced with the Sorin dedicated programmer.Position the programming head over the telemetry antenna located in the upper part of the device, in order to communicate effectively via telemetry (see diagram below).The device can be non-invasively identified as follows:1. Take an x-ray to identify the name of the manufacturer and model, printed on the device (X-ray ID is SEA for PLATINIUM range). 2. Interrogate  the   device   using   the   Sorin  dedicated  programmer.  The   model  and  serial number of the device are automatically displayed. The first figure in the serial number corresponds to the last figure in the year of manufacture.SORIN – PLATINIUM SonR CRT-D – U463A 47
 18.  SUPPLEMENTAL INFORMATION 18. SUPPLEMENTAL INFORMATIONClinical data presented in this section are from the SafeR (AAI <> DDD) clinical study. SafeR (AAI <> DDD) operation in PLATINIUM is similar to that in the Symphony pacemaker. The data provided are applicable to PLATINIUM SonR CRT-D.18.1. ADVERSE EVENTS IN THE SAFER (AAI <> DDD) STUDYClinical study of the SafeR (AAI <> DDD) included 45 Symphony 2550 devices implanted in 45  patients.   No  serious  adverse   events were   device- or   feature-related. There  were  no deaths in the study.Table 1: Summary of Symphony safety data during studyPatients Number of eventsNumber of patients% of patients Number of eventsEvents per device year (a)Deaths 0 0 0 0Explants 0 0 0 0Serious pacemaker related events outside the use of SafeR (AAI <> DDD)0000Non-serious pacemaker related events outside the use of SafeR (AAI <> DDD)0000Serious events due to the use of SafeR (AAI <> DDD)0000Non-serious events related due to the use SafeR (AAI <> DDD)13 28.9 15 3.2Serious non-pacemaker related events6 13.3 9 1.9Non-serious non-pacemaker related events8 17.8 8 1.7(a) 4.74 device yearsNon-serious events due to the use of SafeR 2 (AAI <> DDD) included: delay in switching on 2nd degree AV block, inappropriate classification of a PAC, disagreement between markers and recorded EGM, atrial pacing above the maximum rate, recycling on an R-wave in a 48 SORIN – PLATINIUM SonR CRT-D – U463A
 18.  SUPPLEMENTAL INFORMATIONrefractory   period,   and   disagreement   in   the   statistics   for   swuitches   to   DDD.   No   patient symptoms were associated with these events.18.2. SAFER (AAI <> DDD) CLINICAL STUDYSafeR (AAI <> DDD) mode in PLATINIUM is similar to that in Symphony.The differences in SafeR (AAI <> DDD) mode between the two devices are:─To prevent long RR intervals during VT/VF, SafeR (AAI <> DDD) has no effect during VT/VF therapy, electrophysiologic studies, and post-shock recovery.─The maximum acceptable AV delay for first degree AV block varies as a function of pacing rate.─PLATINIUM requires a ventricular sensed event to atrial paced event (RA) interval of at least 100 ms. Therefore, the device lengthens the atrial escape interval so that it ends at least 102 ms after the ventricular event.─During atrial fibrillation episode, pause criterion is fixed to 2s to avoid long bradycardia episodes in switching to DDD mode.Despite   these   differences,   the   data   collected   on   Symphony   devices   are   applicable   to PLATINIUM because the principles of SafeR (AAI <> DDD) operation did not change. The  criteria for switching from AAI to DDD (or vice versa) did not change. The device’s method for evaluating the presence of AV conduction did not change.Methods:All patients  were  implanted  with a Symphony  Model 2550  dual-chamber rate-responsive pacemaker with  SafeR (AAI <> DDD) mode. A variety of  marketed atrial  and ventricular pacing   leads   were   used.   The   pacemaker   was   programmed   and   interrogated   via   bi-directional telemetry using a Sorin dedicated programmer and a CPR3 programming head.The   study’s  routine   evaluation   consisted   of   enrollment,   pre-discharge   evaluation,  and   a scheduled follow-up visit at one month. At pre-discharge, a 24-hour Holter recording was performed and pacemaker memory was read. At one month, pacemaker memory was read. Investigators also documented adverse events.Patients studied:A total of 45 patients from 12 centers had Symphony 2550 pacemakers with SafeR (AAI <> DDD). Of these, 14 (31 %) were female and 31 (69 %) were male. Mean patient age (± SD) was 74 ± 9 years.Primary indications for implant were: 1st degree AV block (11.1 %), 2nd degree AV block (6.7 %), 3rd degree AV block (22.2 %), sinus node dysfunction (62.2 %) or other (6.7 %).Effectiveness results:To determine the effectiveness of SafeR (AAI <> DDD) mode, the percentage of ventricular pacing provided over one month was recorded from pacemaker memory.Thirty-five   patients   contributed   data   to   evaluate   the   percentage   of   ventricular   pacing provided with SafeR (AAI <> DDD). Twenty-nine patients had 1 % or less ventricular pacing and six patients had a range of 28-97 % ventricular pacing. The graph below shows the distribution of ventricular pacing observed in patients with and without AV block as a primary indication for implant.SORIN – PLATINIUM SonR CRT-D – U463A 49
 18.  SUPPLEMENTAL INFORMATIONThe graph shows that many patients programmed to SafeR (AAI <> DDD) had less than 1% ventricular pacing:─84 % of patients without AV block at implant.─63 % of patients with AV block at implant.In a representative reference group(1) of patients programmed to DDD, none had less than 1 % ventricular pacing and only 10 % had less than 90 % ventricular pacing regardless of AV block indication at implant.The actual reduction of ventricular pacing that SafeR (AAI <> DDD) provides in an individual will depend on the amount of time that the patient spends in AV block. SafeR (AAI <> DDD) cannot and should not provide any decrease in ventricular pacing while the patient is in AV block.(1) Pioger G, Jauvert G, Nitzsché R, Pozzan J, Laure H, Zigelman M, Leny G, Vandrell M,  Ritter   P,   and   Cazeau   S.   Incidence   and   predictive   factors   of   atrial   fibrillation   in   paced  patients. PACE, 28, Supp 1: S137-141; January 2005. This was a prospective observational  study of 377 patients with a functionally similar device programmed to DDD. The primary  indications  for   implant  were:  AV  block  (49  %),  sinus  node  disease   (16  %),  brady-tachy  syndrome (5 %), AV block + sinus node disease (19 %), AV block + brady-tachy syndrome  (6 %), and brady-tachy syndrome + sinus node disease (5 %).50 SORIN – PLATINIUM SonR CRT-D – U463A
 19.  PHYSICAL CHARACTERISTICS 19. PHYSICAL CHARACTERISTICS1811 MODEL:Dimensions 73 x 54.3 x 11.1 mmWeight 86 gVolume 33 cm3Active surface area of casing 62.7cm2Connector Atrium: SonR (IS-1 bipolar compatible). Right ventricle: IS-1 bipolar, 2*DF-1. Left ventricle: IS-1 bipolar (unipolar compatible).1841 MODEL:Dimensions 72.3 x 54.3 x 11.1 mmWeight 87 gVolume 33 cm3Active surface area of casing 62.8cm2Connector Atrium: SonR (IS-1 bipolar compatible). Right ventricle: DF4. Left ventricle: IS-1 bipolar (unipolar compatible).19.1. MATERIALS USEDActive surface area of casing 99% pure titaniumConnectors Polyurethane* and silicone elastomer**Medical-grade materials that have undergone "in vitro" and "in vivo" qualifications.SORIN – PLATINIUM SonR CRT-D – U463A 51
 20.  ELECTRICAL CHARACTERISTICS 20. ELECTRICAL CHARACTERISTICSAtrial input impedance 80 kilohms ± 30 %Ventricular input impedance 80 kilohms ± 30 %D.C. capacitance 149 µF ± 8 %Capacitor reformation No reformation requiredRate limit 192 ppm ± 10 ppmPacing waveformDefibrillation waveform20.1. TABLE OF DELIVERED SHOCK ENERGY AND VOLTAGEThe relationship between stored energies, maximum voltages and delivered energies (at 37 °C, 50 ohm load) for the minimum, low, mean and maximum programmed energy values is as follows:Stored energy (J) 0.5 10 20 34 42V1 (Volt) 71 332 471 617 686V2 (Volt) 35 167 235 309 342Delivered E: Phase 1 (J) 0.32 6.94 14.0 23.8 29.6Delivered E: Phase 2 (J) 0.08 1.75 3.4 6.0 7.4Delivered E: Total (J) 0.4 8.7 17.4 30 37Tolerances are 12% for voltage (25% at 0.5 J) and 30% for energy.20.2. BATTERYManufacturer GreatbatchType Quasar High Rate (QHR)Model GB 3070Number of batteries 152 SORIN – PLATINIUM SonR CRT-D – U463A
 20.  ELECTRICAL CHARACTERISTICSTotal capacity 2192 mAhUsable capacity Between BOS and RRT: 1530 mAh. Between BOS and EOS: 1910 mAh.Voltage BOS: 3.24 V. RRT: 2.62 V. EOS: 2.5 V.20.3. LONGEVITYThe longevities are calculated by taking into account 6 months storage with the following conditions:─Mode: DDD─Basic rate: 60 ppm─Pulse width (A, RV, LV): 0.35 ms─EGM:ON─SonR: ON─2 battery reformings per year (at 34J), replaced by shocks if any─Remote monitoring: ON, daily check, 4 follow-ups and 5 full alert reports per year─RF telemetry: ON, 120min at implantation + 15min at discharge + 15min for in-clinic quarterly follow-upsLongevity projection at 500 Ω pacing impedance:A pacing (%) 100 100 1 15 1 30 0BiV pacing (%) 100 100 100 100 100 100 0A, RV Pacing amplitude (V)3.5 3.5 3.5 4.5 2.5 2.5 -LV Pacing amplitude (V) 3.5 3.5 3.5 4.5 2.5 3.0 -Sensor OFF ON ON ON OFF OFF OFFMax shocks (42J) per year 4 4 4 4 0 3 4Longevity (years) 6.8 6.7 7.8 6.3 11.1 9.6 12.2Longevity projection at 600 Ω pacing impedance:A pacing (%) 100 100 1 15 1 30 0BiV pacing (%) 100 100 100 100 100 100 0A, RV Pacing amplitude (V)3.5 3.5 3.5 4.5 2.5 2.5 -LV Pacing amplitude (V) 3.5 3.5 3.5 4.5 2.5 3.0 -Sensor OFF ON ON ON OFF OFF OFFMax shocks (42J) per year 4 4 4 4 0 3 4Longevity (years) 7.3 7.1 8.2 6.7 11.6 10.0 12.2SORIN – PLATINIUM SonR CRT-D – U463A 53
 20.  ELECTRICAL CHARACTERISTICSLongevity projection at 700 Ω pacing impedance:A pacing (%) 100 100 1 15 1 30 0BiV pacing (%) 100 100 100 100 100 100 0A, RV Pacing amplitude (V)3.5 3.5 3.5 4.5 2.5 2.5 -LV Pacing amplitude (V) 3.5 3.5 3.5 4.5 2.5 3.0 -Sensor OFF ON ON ON OFF OFF OFFMax shocks (42J) per year 4 4 4 4 0 3 4Longevity (years) 7.6 7.5 8.5 7.1 11.9 10.3 12.21h of additional RF programming session reduces the device longevity from 1 to 2 weeks depending on the device functioning mode (no pacing, 100% pacing). The mean longevity as a function of shocks delivered at maximum energy, with and without pacing, is as follows:The mean longevity as a function of yearly remote follow-ups1, with and without pacing, is as follows:1. An excessive number of remote follow-ups can have a non-negligible impact on device longevity.54 SORIN – PLATINIUM SonR CRT-D – U463A
 21.  PROGRAMMABLE PARAMETERS 21. PROGRAMMABLE PARAMETERSMeasured at 37 °C under a 500 ohm loadLegend:Value in bold: "as shipped" valueUnderlined value: nominal value21.1. ANTIBRADYCARDIA PACINGBasic parameters ValuesMode VVI-VVIR-DDD-DDDR-DDD/DDIR-DDI-DDIR-SafeR (AAI <=> DDD)-SafeR-R (AAIR <=> DDDR)Basic rate (ppm) (1) From 30 to 90 by steps of 5 ; 60 (± 4 %)Maximum rate (ppm) From 100 to 145 by steps of 5 ; 120 (± 6 %)Rate hysteresis (%) 0-5-10-20-35 (± 18 ms)Rest AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms)Exercise AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms)AVD Paced/Sensed Offset (ms) 0-10-15-25-30-40-45-55-65-70-80-85-95-100-110-115-125 (± 1 ms)(1) The corresponding periods are (in ms): 2000-1714-1500-1333-1200-1091-1000-923-857-800-750-706-667 ms.Special features ValuesRate smoothing OFF-Very slow-Slow-Medium-FastMode Switch ON-OFFPhysical activity Very low-Low-Medium-High-Very highExercise AV opt. rate (ppm) From 70 to 120 by steps of 5 ; 90CRT optimization OFF-Monitor-AV-AV+VVSORIN – PLATINIUM SonR CRT-D – U463A 55
 21.  PROGRAMMABLE PARAMETERSPacing/Sensing ValuesAtrial sensitivity (mV) (1) From 0.2 to 4 by steps of 0.2 ; 0.4 (± 50 %)Atrial amplitude (V) (2) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)Atrial pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)Ventricular sensitivity (mV) (1) From 0.4 to 4 by steps of 0.2 ; 0.4 (± 50 %)RV amplitude (V) (2) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)RV pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)LV amplitude (V) (2) 0.10-0.25-0.35 (± 50 %)0.50-0.60-0.75-0.85 (± 30 %)1.00-1.10-1.25-1.35-1.50-1.60-1.75-1.85-2.00-2.10-2.25-2.35-2.50-2.60-2.75-2.85-3.00-3.10-3.25-3.35-3.50-3.60-3.75-3.85-4.00-4.10-4.25-4.35-4.50-4.60-4.75-4.85-5.00-5.10-5.25-5.35-5.50-5.60-5.75-5.85-6.00-6.10-6.25-6.35-6.50-6.60-6.75-6.85-7.00- (± 20 %)LV pulse width (ms) 0.12-0.25-0.35-0.50-0.60-0.75-0.85-1.00-1.10-1.25-1.35-1.50-1.60-1.75-1.85-2.00 (± 10 %)LV pacing polarity LV bipolar-LV tip to RV ring-LV ring to RV coil-LV tip to RV coil-LV ring to CAN-LV tip to CANV chambers Right-R+L-L+RVV delay (ms) 0-16-24-32-40-48-56-64- (± 3 ms)(1) Values are measured using a positive and negative triangular signal of 2/13 ms.(2) The  correlation   between  the  programmed  amplitudes,  the  stored   amplitudes  and  the  mid-pulse delivered amplitudes under a 500 ohm load are given in the following table:Programmed amplitude (V) Stored amplitude (V) Mid-pulse delivered amplitude (V)0,10* 0,20 0,190,25* 0,26 0,220,35* 0,39 0,330,50* 0,52 0,440,60* 0,72 0,610,75* 0,85 0,720,85* 0,98 0,831,00 1,11 0,941,10* 1,24 1,051,25* 1,37 1,161,35* 1,50 1,271,50 1,63 1,381,60* 1,76 1,4956 SORIN – PLATINIUM SonR CRT-D – U463A
 21.  PROGRAMMABLE PARAMETERS1,75* 1,89 1,601,85* 2,02 1,712,00 2,10 1,782,10* 2,21 1,872,25* 2,34 1,982,35* 2,47 2,092,50 2,60 2,202,60* 2,73 2,312,75* 2,87 2,422,85* 3,00 2,533,00 3,15 2,673,10* 3,26 2,763,25* 3,39 2,873,35* 3,52 2,983.50 3,65 3,093,60* 3,78 3,203,75* 3,91 3,313,85* 4,04 3,424,00 4,20 3,554,10* 4,30 3,644,25* 4,43 3,754,35* 4,56 3,864,50 4,69 3,974,60* 4,82 4,084,75* 4,95 4,194,85* 5,08 4,305,00 5,25 4,445,10* 5,34 4,525,25* 5,47 4,635,35* 5,60 4,745,50* 5,73 4,85SORIN – PLATINIUM SonR CRT-D – U463A 57
 21.  PROGRAMMABLE PARAMETERS5,60* 5,86 4,965,75* 5,99 5,075,85* 6,12 5,186,00 6,30 5,336,10* 6,38 5,406,25* 6,51 5,516,35* 6,64 5,626,50* 6,77 5,736,60* 6,90 5,846,75* 7,03 5,956,85* 7,16 6,067,00 7,35 6,22* For left ventricular amplitude only.Ventricular arrhythmia algorithms ValuesAtrial pacing on PVC Yes-NoPost extrasystolic pause suppression Yes-NoAcceleration on PVC Yes-NoMax acceleration rate (ppm) From 60 to 145 by steps of 5 ; 10058 SORIN – PLATINIUM SonR CRT-D – U463A
 21.  PROGRAMMABLE PARAMETERSPost-shock mode ValuesMode OFF-VVI-DDI-DDDDuration 10s-20s-30s-1min-2min-3min-4min-5minBasic rate (ppm) From 50 to 90 by steps of 5 ; 60 (± 4 %)Rest AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms)Exercise AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms)AVD Paced/Sensed Offset (ms) 0-10-15-25-30-40-45-55-65-70-80-85-95-100-110-115-125 (± 1 ms)A amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)A pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)RV amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)RV pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)LV amplitude (V) 0.1-0.25-0.35 (± 50 %)0.5-0.6-0.75-0.85 (± 30 %)1.00-1.10-1.25-1.35-1.50-1.60-1.75-1.85-2.00-2.10-2.25-2.35-2.50-2.60-2.75-2.85-3.00-3.10-3.25-3.35-3.50-3.60-3.75-3.85-4.00-4.10-4.25-4.35-4.50-4.60-4.75-4.85-5.00-5.10-5.25-5.35-5.50-5.60-5.75-5.85-6.00-6.10-6.25-6.35-6.50-6.60-6.75-6.85-7.00- (± 20 %)LV pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)1.10-1.25-1.35-1.50-1.60-1.75-1.85-2.00 (± 1 %)Refractory periods ValuesAtrial refractory period post ventricular sensing (ms)45-65-80-95-110-125-140-155 (± 16 ms)Atrial refractory period post ventricular pacing (ms)80-95-110-125-140-155 (± 4 ms)Sensitivity margins ValuesAtrial post pacing/sensing margin (mV)From 0 to 1 by steps of 0.2 ; 0.4Ventricular post pacing margin (mV) From 0 to 2 by steps of 0.2 ; 0.8Response to noise ValuesAutomatic sensitivity on noise ON-OFFV pacing on noise ON-OFFSORIN – PLATINIUM SonR CRT-D – U463A 59
 21.  PROGRAMMABLE PARAMETERSSafeR parameters ValuesAVB I switch Rest+Exercise-ExerciseLong PR: max (ms) From 200 to 500 by steps of 50 ; 450Long PR: min (ms) From 200 to 500 by steps of 50 ; 250Max. pause (s) 2-3-421.2. VENTRICULAR TACHYARRHYTHMIA DETECTIONTherapy zones ValuesSlow VT detection zone Slow VT ON-Slow VT OFFVT detection zone VT ON-VT OFFFast VT / VF detection zone Fast VT+VF ON-VF ONSlow VT rate (lower limit) (ppm) From 100 to 200 by steps of 5 ; 190VT rate (lower limit) (ppm) 130-135-140-145-150-155-160-165-170-175-180-185-190-195-200-210-220-230VF rate (lower limit) (ppm) 150-155-160-165-170-175-180-185-190-195-200-210-220-230-240Fast VT rate (upper limit) (ppm) 155-160-165-170-175-180-185-190-195-200-210-220-230-240-255Slow VT persistence (cycles) 4-6-8-12-16-20-30-50-100-200VT persistence (cycles) 4-6-8-12-16-20-30-50-100-200VF persistence (cycles) From 4 to 20 by steps of 1 ; 660 SORIN – PLATINIUM SonR CRT-D – U463A
 21.  PROGRAMMABLE PARAMETERSDetection criteria ValuesSlow VT and VT detection criteria Rate Only-Stability-Stability+-Stability/Acc-Stability+/Acc-PARAD-PARAD+Fast VT detection criteria Rate+Stability-Rate OnlyMajority: (X/Y), Y (cycles) 8-12-16Majority: (X/Y), X (%) 65-70-75-80-90-95-100Window of RR stability for Slow VT and VT (ms)30-45-65-80-95-110-125Window of RR stability for fast VT (ms)30-45-65Acceleration (%) 6-13-19-25-31-38-44-50Long cycle persistence extension (cycles)From 0 to 16 by steps of 1 ; 10Long cycle gap (ms) 15-30-45-65-80-95-110-125-140-155-170-190-205Atrial monitoring Yes-No21.3. VENTRICULAR TACHYARRHYTHMIA THERAPIESCommon parameters ValuesEnable ATP therapy Yes-NoEnable shock therapy Yes-NoATP pacing chamber Right-Left-R+LPolarity alternation (42J) Yes-NoAtrial coil (SVC) present Yes-NoActive case Yes-NoShock configuration (+ --> -) Case to RV-SVC to RV-Case + SVC to RV-RV to Case-RV to SVC-RV to Case + SVCSVC exclusion (shock < 15J) Yes-NoSORIN – PLATINIUM SonR CRT-D – U463A 61
 21.  PROGRAMMABLE PARAMETERS21.3.1. Therapy parameters in slow VT zoneATP 1 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310ATP 2 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310Shock program ValuesShock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Number of Max. Shock (42 J) OFF-1-2-3-462 SORIN – PLATINIUM SonR CRT-D – U463A
 21.  PROGRAMMABLE PARAMETERS21.3.2. Therapy parameters in VT zoneATP 1 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310ATP 2 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310Shock program ValuesShock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Number of Max. Shock (42 J) OFF-1-2-3-4SORIN – PLATINIUM SonR CRT-D – U463A 63
 21.  PROGRAMMABLE PARAMETERS21.3.3. Therapy parameters in fast VT / VF zoneATP 1 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit 10s-20s-30s-1min-1.5min-2minMinimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310Shock program ValuesShock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Number of Max. Shock (42 J) 1-2-3-421.4. REMOTE ALERTS AND WARNINGSThe device routinely performs security self-checks and technical measurements to ensure system integrity. When system integrity is found to be at risk outside a follow-up, alerts are stored in the device memory. When system integrity is found to be at risk during a follow-up, the information is managed by a warning (pop-up message) to immediately notify the user. For   example,   the   following  types  of   event   can  trigger   a  warning   or   an  alert:   technical problem during a shock, pacing lead impedance or coil impedance measurements out-of-range,   battery   depletion,   etc.   The   Remote   tab   presents   an   overview   of   all   the   alerts  managed by the device.General parameters ValuesRF communication (1) ON-OFFAlerts (1) ON-OFF(1) RF and Remote alerts are turned on automatically if Shocks are programmed ON.When Alerts are programmed "On", the following System Alerts are automatically activated:─Battery depletion – RRT─Device reset─Excessive charge time (>25s)64 SORIN – PLATINIUM SonR CRT-D – U463A
 21.  PROGRAMMABLE PARAMETERS─System integrityLead Alerts ValuesAbnormal A lead impedance ON-OFFAbnormal A lead low limit (Ohm) 200-250-300-350-400-450-500Abnormal A lead high limit (Ohm) 1500-1750-2000-2500-3000Abnormal RV lead impedance ON-OFFAbnormal RV lead low limit (Ohm) 200-250-300-350-400-450-500Abnormal RV lead high limit (Ohm) 1500-1750-2000-2500-3000Abnormal LV lead impedance ON-OFFAbnormal LV lead low limit (Ohm) 200-250-300-350-400-450-500Abnormal LV lead high limit (Ohm) 1500-1750-2000-2500-3000Abnormal RV coil impedance ON-OFFAbnormal SVC coil impedance ON-OFFAbnormal Shock impedance (1) ON-OFF(1) Normal impedance range [20 Ohm-200 Ohm]Clinical status ValuesV oversensing ON-OFFHigh AT/AF burden ON-OFFAT/AF limit (on 24h) (h) 0.5-1-3-6-12-24Fast V Rate during AT/AF ON-OFFFast V Rate limit (ppm) 80-90-100-110-120Fast V Duration limit (h) 0.5-1-3-6-12-24Limited % of V pacing in CRT ON-OFFLimited % of V pacing (%) 50-70-80-85-90-95Therapy information ValuesShock disabled ON-OFFShocks delivered OFF-All shocks-Inefficient shock-Inefficient max shockATP delivered ON-OFFSORIN – PLATINIUM SonR CRT-D – U463A 65
 22.  NON PROGRAMMABLE PARAMETERS 22. NON PROGRAMMABLE PARAMETERSInterval ValuesCommitted period 95 ms (± 5 ms)Atrial refractory periods ValuesPost atrial sensing 47 ms (± 16 ms)Post atrial pacing 109 ms (± 4 ms)Ventricular refractory periods ValuesPost ventricular sensing 95 ms (± 16 ms)Post ventricular pacing 220 ms (± 4 ms)Post atrial pacing (blanking) 16 ms (± 3 ms)Tachycardia criteria ValuesWindow of PR association 63 ms (± 1 ms)Therapies ValuesWaveform (1) Constant tilt (50% - 50%)Stored energy for the Max. shock 42 JPacing amplitude during ATP therapies7 V (Actual value at 300 ms: 5.3 V )Anti-PMT protection Termin(1) The device has 50% tilt in each phase thus delivers 94% of stored energy. Each phase is limited to  10 ms duration.66 SORIN – PLATINIUM SonR CRT-D – U463A
 23.  LIMITED WARRANTY 23. LIMITED WARRANTYThe   PLATINIUM   implantable   cardioverter   defibrillator   is   the   result   of   highly   advanced research and all components have been selected after exhaustive testing.The terms of the limited warranty are available upon request from your Sorin representative or on the CD-ROM of the package content.SORIN – PLATINIUM SonR CRT-D – U463A 67
 24.  PATENTS 24. PATENTSThe PLATINIUM model described in this manual is covered by the following US patents:5 713 928, 5 741 315, 5 776 164, 5 776 165, 5 818 703, 5 836 980, 5 868 793, 5 891 170, 5 891 184, 5 899 931, 5 931 856, 5 935 153, 5 954 660, 5 978 708, 6 181 968, 6 230 058, 6 236 111, 6 251 703, 6 256 206, 6 307 261, 6 337 996, 6 397 105, 6 408 209, 6 487 451, 6 487 452, 6 505 068, 6 532 238, 6 556 866, 6 604 002, 6 622 039, 6 625 491, 6 711 441, 6 738 665, 6 830 548, 6 889 080, 6 898 845, 6 912 421, 6 937 898, 6 975 905, 7 065 402, 7 072 716, 7 076 297, 7 113 826, 7 142 924, 7 164 946, 7 251 526, 7 366 566, 7 400 921, 7  400   922,   7   953   483,   US8064992,   US8043225,   US7792582,   US8798748,   US7890168, US8195293, US7966068, US8768464, US8874209, US8554313, US8214036, US8233981, US8554319, US7966065, US8253279, US8874210, US8219193, US8391976, US8855764, US8359096, US8494629, US8359091, US8489188, US8712526, US8862230, US8641436, US8718765, US8798771, US8868170, US8678843 , US8694098, US8938286, US8874212.68 SORIN – PLATINIUM SonR CRT-D – U463A
 25.  EXPLANATION OF SYMBOLS 25. EXPLANATION OF SYMBOLSThe symbols on product labelling have the following meaning:General symbols Explanation of symbols Defibrillator symbols Explanation of symbolsUse by   CRT-D, RA, RV, LVDate of manufacture   SonR, CRT-D DF-1 connectorsManufacturer   SonR, CRT-D DF4 connectorsCatalogue number ShocksSerial number   ATP Anti-tachycardia pacing, RV, LVImplantable device uncoated DDD As shipped modePackage content Basic rateSterile package content High voltageOpen here Ready for wireless interrogation and programming by Orchestra Plus programmer equipped with ORCHESTRA PLUS LINK accessoryDo not use if the package is damagedDF-1 defibrillating connector insulating plugDo not reuseDo not resterilizeSterilised using ethylene oxideNon sterileSORIN – PLATINIUM SonR CRT-D – U463A 69
 25.  EXPLANATION OF SYMBOLSTemperature limitationTorque wrenchConsult instructions for useInstructions for use in the CD-ROMThis icon is used to call your attention to a particularly important point.This icon alerts you to a hazard that may result in equipment damage or personal injury. Carefully read the instructions provided with this icon.Eucomed / Advamed Code of Ethical Business PracticeLast revision date of this implant manual: 2015-0370 SORIN – PLATINIUM SonR CRT-D – U463A
2015-03U463A© 2015 Sorin CRM SAS

Navigation menu