SORIN CRM DR1510 Implantable cardioverter defibrillator with RF wireless communication availability User Manual

SORIN CRM Implantable cardioverter defibrillator with RF wireless communication availability

User manual

Implantable cardioverter defibrillatorDR modelsImplant manual
  TABLE OF CONTENTS 1. General description................................................................................................................ 5  2. Indications............................................................................................................................... 6  3. Contraindications....................................................................................................................7  4. Warnings and precautions.....................................................................................................8  4.1. Risks related to medical environment.......................................................................................9  4.2. Sterilization, storage and handling..........................................................................................10  4.3. Implantation and device programming....................................................................................10  4.4. Lead evaluation and lead connection......................................................................................11  4.5. Generator explant and disposal..............................................................................................12  5. Adverse events...................................................................................................................... 13  5.1. Defender study........................................................................................................................ 13  5.2. SafeR (AAI <> DDD) study.....................................................................................................14  6. Clinical studies...................................................................................................................... 16  6.1. Defender study........................................................................................................................ 16  6.2. SafeR (AAI <> DDD) study.....................................................................................................18  7. Patient selection and treatment...........................................................................................22  7.1. Individualization of treatment..................................................................................................22  7.2. Specific patient populations....................................................................................................23  8. Patient counselling information..........................................................................................24  9. Declaration of conformity.....................................................................................................25  10. Physician guidelines.............................................................................................................27  10.1. Physician training.................................................................................................................... 27  10.2. Directions for use.................................................................................................................... 27  10.3. Maintaining device quality.......................................................................................................27  11. Patient information............................................................................................................... 28  12. How supplied......................................................................................................................... 29  12.1. Sterility.................................................................................................................................... 29  12.2. Warranty and replacement policy............................................................................................29  13. Device description................................................................................................................ 30  14. Implant procedure................................................................................................................. 32  14.1. Necessary equipment.............................................................................................................32  14.2. Packaging............................................................................................................................... 32  14.3. Optional equipment.................................................................................................................32  14.4. Before opening the package...................................................................................................33  14.5. Prior to implantation................................................................................................................33  14.6. Device placement.................................................................................................................... 33  14.7. Choosing the type of lead.......................................................................................................33  14.8. Shock configuration (+ -> -).....................................................................................................34  14.9. Measurement of thresholds at implant....................................................................................35  14.10.Lead connection ..................................................................................................................... 35  14.11. Device implantation.................................................................................................................36  14.12.Tests and programming...........................................................................................................37  15. Special modes....................................................................................................................... 38  15.1. Safety mode (nominal values).................................................................................................38 SORIN – PLATINIUM DR – U460A 3
   15.2. Magnet mode.......................................................................................................................... 38  15.3. Response in the presence of disturbance...............................................................................38  15.4. Detection characteristics in the presence of electromagnetic fields........................................38  15.5. Protection against short-circuits..............................................................................................40  16. Main functions....................................................................................................................... 41  16.1. Automatic lead measurements................................................................................................41  16.2. Atrial tachyarrhythmia management.......................................................................................41  16.3. Ventricular tachyarrhythmia management...............................................................................41  16.4. Pacing..................................................................................................................................... 42  16.5. Sensing................................................................................................................................... 42  16.6. Follow-up function................................................................................................................... 43  16.7. Remote Monitoring function....................................................................................................43  17. Patient follow-up................................................................................................................... 46  17.1. Follow-up recommendations...................................................................................................46  17.2. Holter Function........................................................................................................................ 46  17.3. Recommended Replacement Time (RRT)..............................................................................47  17.4. Explantation............................................................................................................................ 47  17.5. Defibrillator identification.........................................................................................................48  18. Physical characteristics.......................................................................................................49  18.1. Materials used......................................................................................................................... 49  19. Electrical characteristics......................................................................................................50  19.1. Table of delivered shock energy and voltage..........................................................................50  19.2. Battery..................................................................................................................................... 50  19.3. Longevity................................................................................................................................. 51  20. Programmable parameters...................................................................................................53  20.1. Antibradycardia pacing............................................................................................................53  20.2. Ventricular tachyarrhythmia detection.....................................................................................56  20.3. Ventricular tachyarrhythmia therapies.....................................................................................57  20.4. Remote alerts and warnings...................................................................................................60  21. Non programmable parameters...........................................................................................62  22. Limited warranty................................................................................................................... 63  23. Patents................................................................................................................................... 64  24. Explanation of symbols........................................................................................................65 4SORIN – PLATINIUM DR – U460A
 1.  GENERAL DESCRIPTION 1. GENERAL DESCRIPTIONPLATINIUM DR is an implantable dual-chamber cardioverter defibrillator. It is equipped with an accelerometer to allow adaptation of pacing to suit the patient’s activity.PLATINIUM DR is also equipped with the RF wireless technology which enables:─Remote monitoring of patients who have the Sorin SMARTVIEW Monitor installed at home,─wireless interrogation and device programming by Orchestra Plus programmer equipped with ORCHESTRA PLUS LINK accessory.PLATINIUM   DR   provides   high   energy   shocks   (42 J)   for   enhanced   safety,   as   well   as automatic lead measurements to monitor system integrity.PLATINIUM DR is protected against high-frequency signals emitted by cellular telephones.Device and lead connections:PLATINIUM DR 1510 2*IS-1 bipolar, 2*DF-1PLATINIUM DR 1540 1*IS-1 bipolar, 1*DF4SORIN – PLATINIUM DR – U460A 5
 2.  INDICATIONS 2. INDICATIONSPLATINIUM DR is indicated for use in patients who are at high risk of sudden cardiac death  due   to   ventricular   tachyarrhythmias   and   who   have   experienced   one   of   the   following situations:─Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia,─Recurrent, poorly tolerated sustained ventricular tachycardia (VT).6SORIN – PLATINIUM DR – U460ANOTE: The clinical outcome for hemodynamically stable VT patients is not fully known. Safety and effectiveness studies have not been conducted.
 3.  CONTRAINDICATIONS 3. CONTRAINDICATIONSImplantation of PLATINIUM DR is contraindicated in patients:─whose ventricular tachyarrhythmias may have transient or reversible causes such as: acute myocardial infarction, digitalis intoxication, drowning, electrocution, electrolyte imbalance, hypoxia, sepsis, or unstable ischemic episodes,─who present incessant tachyarrhythmia,─who have an internal pacemaker,─whose primary disorder is bradyarrhythmias, or atrial tachyarrhythmias.The   use   of   the   dual-chamber   pacing   mode   is   contraindicated   in   patients   with   chronic refractory atrial tachyarrhythmias.SORIN – PLATINIUM DR – U460A 7
 4.  WARNINGS AND PRECAUTIONS 4. WARNINGS AND PRECAUTIONSThe   patient  should   be   warned   of   the   potential   risks  of   defibrillator  malfunction   if   he   is exposed to external magnetic, electrical, or electromagnetic signals.These potential interference sources may cause conversion to inhibited mode (because of noise detection), erratic delivery of VT or VF therapies, nominal programming, or much more rarely, irreversible damage to the device’s circuits.The main sources of high magnitude disturbance are: powerful radiofrequency equipment (radar),   industrial   motors   and   transformers,   induction   furnaces,   resistance,   arc-welding equipment and high power loudspeakers.Electrical Isolation:Do  not  permit the patient to contact  grounded  equipment  that  could  produce  hazardous leakage current. Ensuing arrhythmia induction could result in the patient’s death.Antitheft gates:Since antitheft devices at the entrance to stores are not subject to any safety standards, it is advisable to spend as little time as possible in their vicinity.Airport detection systems:Since airport detection systems are not subject to any safety standards, it is advisable to spend as little time as possible in their vicinity.Work environment:The patient's work environment may be an important source of disturbance. In that case, specific recommendations may be required.High voltage power transmission lines:High voltage power transmission lines may generate enough disturbance to interfere with defibrillator operation if approached too closely.Communication equipment:Communication equipment such as microwave transmitters, linear power amplifiers, or high-power amateur transmitters may generate enough disturbance to interfere with defibrillator operation if approached too closely.Home appliances:Home  appliances  that  are in  good working  order  and   properly  grounded do  not  usually produce   enough   disturbance   to   interfere   with   defibrillator   operation.   However,   there   are reports of device interferences caused by electric hand tools or electric razors used directly over the device implant site. Patient should also avoid using induction ovens and cookers.8SORIN – PLATINIUM DR – U460ACAUTION: Do not tap sharply on the ICD can after implant, because the ICD's sensing circuits can detect this as P-waves or R-waves, and such oversensing could result in inappropriate pacing, inhibition, or therapy. Normal activities after implant do not result in such oversensing.
 4.  WARNINGS AND PRECAUTIONS4.1. RISKS RELATED TO MEDICAL ENVIRONMENTIt   is   advisable   to  carefully  monitor   defibrillator   operation   prior   to   and  after   any  medical treatment during  which an electrical  current from  an external source passes through the patient's body.Magnetic Resonance Imaging:MRI is strictly contraindicated in cardiac defibrillator patients.Radiofrequency ablation:A  radiofrequency   ablation   procedure   in   a   patient   with   a   generator   may   cause   device malfunction or damage. RF ablation risks may be minimized by:1. Programming Shock Therapy and ATP to OFF.2. Avoiding direct contact between the ablation catheter and the implanted lead or generator.3. Positioning the ground, placing it so that the current pathway does not pass through or near the device, i.e. place the ground plate under the patient’s buttocks or legs.4. Having external defibrillation equipment available.Electrocautery or diathermy device:Diathermy and electrocautery equipment should not be used. If such devices must be used:1. Keep the current path and ground plate as far away from the device and the leads as possible (a minimum of 15 cm [six inches]). 2. Before procedure, deactivate ATP and shock therapies.3. During the procedure, keep the electrocautery device as far as possible from the cardiac defibrillator. Set it at minimum intensity. Use it briefly.4.  After   the   procedure,   check  for   proper   implant   function. The   device  should   never   be exposed directly to the diathermy source.External defibrillation:PLATINIUM DR is protected from external defibrillation shocks.1. Before external defibrillation, deactivate ATP and shock therapies.2.  During  external  defibrillation,  it  is  advisable  to  avoid  placing   the  defibrillating  paddles directly over the casing or over the leads. The defibrillating paddles should preferably be placed in an anteroposterior position.3. Avoid any direct contact between the defibrillation paddles and the conductive parts of the implanted leads or casing of the implanted device.4. After external defibrillation, check for proper device function.Radiation therapy:Avoid exposure to ionizing radiation. Betatrons are contraindicated. If high doses of radiation therapy cannot be avoided, the defibrillator should be protected from direct exposure with a protection shield. ATP and shock therapies should be disabled during exposure and proper device   function   should   be   checked   regularly   afterwards.   Resulting   damage   may   not   be immediately detectable. If irradiation of tissues close to the implantation site is necessary, it is recommended that the cardiac defibrillator be moved. As a safety measure, an external defibrillator should be immediately available.Lithotripsy:Lithotripsy may permanently damage the device if it is at the focal point of the lithotripsy beam. If lithotripsy must be used, keep the defibrillator at least 2.5 to 5 cm (1-2 inches) away from the focal point of the lithotripsy beam.SORIN – PLATINIUM DR – U460A 9
 4.  WARNINGS AND PRECAUTIONSDiagnostic ultrasound (echography):The defibrillator is not affected by ultrasound imaging devices.Scales with body fat monitors and electronic muscle stimulators:A patient with an implanted PLATINIUM DR should not use these devices.4.2. STERILIZATION, STORAGE AND HANDLINGResterilization:Do not resterilize and re-implant explanted ICDs."Use Before" Date:A "Use Before" date is printed on the outer storage package and on the sterile package. Do not implant the device after this date because the battery may have reduced longevity and sterility may be affected. It should be returned to Sorin.If Package is damaged:Do   not   use   the   device   or   accessories   if   the   packaging   is   wet,   punctured,   opened   or damaged because the integrity of the sterile packaging may be compromised. Return the device to the manufacturer.Device Storage:Store the device in a clean area, away from magnets, kits containing magnets, and sources of electromagnetic disturbance to avoid device damage. Store the device between 0 - 50 °C (32 - 122 °F). Temperatures outside the specified range may damage the device.Equilibration:Allow the device to reach room temperature before programming or implanting the device because rapid temperature changes may affect initial device function.4.3. IMPLANTATION AND DEVICE PROGRAMMINGUse only a Sorin programmer to communicate with the device.Do   not   inadvertently   position   any   magnet   over   the   ICD;   this   suspends  tachyarrhythmia detection and treatment.Replace the device when the RRT (Recommended Replacement Time*) point (defined by a battery voltage of 2.66 ± 0.01 V or a magnet rate lower than or equal to 80 bpm) is reached.Program device parameters such as sensitivity threshold and VT and VF detection intervals as specified in the device manuals.Lead System:Do   not   use   a   lead   system   other   than   those   with   demonstrated   compatibility   because undersensing cardiac activity and failure to deliver necessary therapy may result.In situations where an ICD and a pacemaker are implanted in the same patient, interaction testing should be completed. If the interaction between the ICD and the pacemaker cannot be resolved through repositioning of the leads or reprogramming of either the pacemaker or the  ICD, the  pacemaker   should   not   be   implanted (or  should  be   explanted   if   previously implanted).Failure   to   properly   insert   the   torque   screwdriver   into   the   perforation   at   an   angle perpendicular to the connector receptacle may result in damage to the sealing system and its self-sealing properties.10 SORIN – PLATINIUM DR – U460A
 4.  WARNINGS AND PRECAUTIONSIn   the   event   of   a   warning   on   a   low   shock   impedance,   and   after   lead   replacement   or reconnection:   it   is   recommended   to   check   the   system   integrity   (sensing   and   pacing thresholds and the impedance of the shock electrodes)It is recommended that a security margin of at least 10 J be demonstrated between the effective  shock   energy   and   maximum   programmable   energy.   Carefully   confirm   that   true ventricular fibrillation has been induced because the DFT for ventricular tachycardia or flutter may be lower.The defibrillator should  be implanted with  the engraved side facing  outwards in  order to facilitate   telemetric   communication   with   the   programming   head   and   to   display   the radiographic identification correctly.*: corresponds to ERI (Elective Replacement Indicator) previously used. 4.4. LEAD EVALUATION AND LEAD CONNECTIONPLATINIUM DR 1510 has two DF-1 and two IS-1 connector ports.PLATINIUM DR 1540 has one IS-1, and one DF4 connector ports.IS-1   refers   to   the   international   standard   whereby   leads   and   generators   from   different manufacturers are assured a basic fit (ISO 5841-1:2000).DF-1 refers to the international standard for defibrillation lead connectors (ISO 11318:2002).DF4 refers to the international standard for defibrillation lead connectors (ISO 27186:2010).Use only DF4-LLHH or DF4-LLHO standard lead connector types according to ISO 27186: 2010.Do not tie a ligature directly to the lead body, tie it too tightly, or otherwise create excessive strain at the insertion site as this may damage the lead. Use the lead stabilizer to secure the lead lateral to the venous entry site.Do not immerse the leads in mineral oil, silicone oil, or any other liquid.Do not grip the lead with surgical instruments.Do not use excessive force or surgical instruments to insert a stylet into a lead.Use   ventricular   transvenous   leads   with   caution   in   patients   with   either   a   mechanical   or bioprosthetic tricuspid valvular prosthesis.Use the correct  suture sleeve (when needed)  for each lead,  to immobilize the lead and protect it against damage from ligatures.Never implant the system with a lead system that has a measured shock impedance of less than   30   ohms.   A   protection   circuit   in   the   defibrillator   prevents   shock   delivery   when impedance is too low. If the shock impedance is less than 30 ohms, reposition the lead system to allow a greater distance between the electrodes.Do not kink leads. Kinking leads may cause additional stress on the leads, possibly resulting in lead fracture.Do not insert a lead connector pin into the connector block without first visually verifying that the   setscrews   are   sufficiently   retracted.   Do   not   tighten   the   setscrews   unless   a   lead connector pin is inserted because it could damage the connector block.Lead electrodes in contact during a cardioversion or defibrillation therapy will cause current to bypass the heart, possibly damaging the ICD and the leads. While the ICD is connected to the leads, make sure that the metal portions of any electrodes do not touch each other.If a pacing lead is abandoned rather than removed, it must be capped to ensure that it is not a pathway for currents to or from the heart.If a thoracotomy is required to place epicardial patches, it should be done during a separate procedure to reduce the risk of morbidity and mortality.SORIN – PLATINIUM DR – U460A 11
 4.  WARNINGS AND PRECAUTIONSDo not place the patch lead over nerve tissue as this may cause nerve damage.Place the patch lead with the conducting coil side facing the heart  to ensure delivery of energy to the heart.Place the sutures well outside the coil of the patch lead or in the area between the coils to avoid possible coil fracture.If countershock is unsuccessful using external paddles, adjust the external paddle position (e.g.,   anterior-lateral   to   anterior-posterior)   and   be   sure   that   the   external   paddle   is   not positioned over the patch.Do  not   fold, alter,   or  remove  any  portion   of  the  patch  as   it  may compromise   electrode function or longevity.If   a   header   port   is   unused  on   the   generator,  the   port   must  be   plugged   to   protect   the generator.4.5. GENERATOR EXPLANT AND DISPOSALInterrogate   the   device,   and   program   shock   therapy   off   prior   to   explanting,   cleaning   or shipping the device to prevent unwanted shocks.Return all explanted generators and leads to the manufacturer.Never incinerate the device due to the potential for explosion. The device must be explanted before cremation.12 SORIN – PLATINIUM DR – U460A
 5.  ADVERSE EVENTS 5. ADVERSE EVENTSClinical  data  presented  in this section are from the Defender and SafeR  (AAI  <> DDD) clinical studies.PLATINIUM DR is similar in design and function to the Defender devices.SafeR (AAI <> DDD) operation in PLATINIUMis similar to that in the Symphony pacemaker.The data provided are applicable to PLATINIUM DR.5.1. DEFENDER STUDYClinical study of Defender IV DR 612 included 60 devices implanted in 60 patients, 38 in Europe   (37   patients   followed   for   a   minimum   of   3   months),   and   22   in   the   U.S.   (IDE G970282/S15) with a total device exposure of 228.7 and 30.3 device months, respectively. No deaths, serious adverse experiences or complications were judged to be device-related, as determined by the investigator. The following tables summarize the safety data for this study.There was 1 death in the study that was classified as arrhythmic. The cause of death was recurrent VT/VF which occurred 19 days post implant.In the following tables, complications are defined as adverse device effect, which cannot be treated or resolved by simple adjustments (e.g. reprogramming) and requires intervention.Observations  are   defined  as   symptomatic  or  asymptomatic  clinical  events  with  potential adverse   device   effects   that   do   not   require   intervention   or   can   be   corrected   by   simple adjustments.Two of the 38 Defender IV DR 612 patients in Europe (37 patients followed for a minimum of 3   months)   experienced   a   total   of   three   complications,   including   device   failures   and replacements. Fourteen of the 38 Defender IV DR 612 patients experienced a total of 18 observations. Complications and observations are reported in Tables 1 and 2. It should be noted that a patient can have more than one observation or complication. There were no observations or complications in the U.S.Table 1: Summary of European Clinical Complications(Including Device Failures and Replacements)All complications, 2 of 38 Defender IV DR 612 patients in EuropeEvent # of Patients % of Patients # of Events Events/100 Device-Years*Hematoma 1 2.6 1 5.2SORIN – PLATINIUM DR – U460A 13NOTE: The company classified as complications those adverse device effects that were treated with surgery or with external defibrillation of a ventricular cardiac event.NOTE: The company classified as observations those adverse device effects that were treated with programming changes, medication, or other method that was not classified as a complication.
 5.  ADVERSE EVENTSVentricular lead migration/dislodgment2 5.3 2 10.5* There were 228.7 device months in this study.Table 2: Summary of European Clinical Complications(Including Patient Complaints)All complications, 14 of 38 Defender IV DR 612 patients in EuropeEvent # of Patients* % of Patients # of Events Events/100 Device-Years**Change in ventricular sensing threshold1 2.6 1° 5.2Device reset*** 1 2.6 1° 5.2Inappropriate therapy for EMI 1 2.6 1° 5.2Pneumothorax 1 2.6 1° 5.2Pocket hematoma 2 5.3 2° 10.5Pocket infection/hematoma 1 2.6 1° 5.2Pocket infection from previous pacemaker1 2.6 1° 5.2Prolonged implant procedure 1 2.6 1 5.2Sensor acceleration during telemetry***1 2.6 1 5.2Shock for VT in VF Zone 1 2.6 1° 5.2Slow VT not converted by ATP therapy1 2.6 2° 10.5Unsatisfactory sensing threshold test***2 5.3 2 10.5Ventricular oversensing 3 7.9 3 15.7* A patient can have more than one observation.** There were 228.7 device months in this study.***These observations would not have happened with the currently marketed device and programmer.°Investigator indicated that Defender IV DR did not cause or contribute to the event.5.2. SAFER (AAI <> DDD) STUDYClinical study of the SafeR (AAI <> DDD) included 45 Symphony 2550 devices implanted in 45  patients.   No serious   adverse events   were  device-  or  feature-related.  There  were  no deaths in the study.Table 1: Summary of Symphony safety data during studyPatients Number of eventsNumber of patients% of patients Number of eventsEvents per device year (a)14 SORIN – PLATINIUM DR – U460A
 5.  ADVERSE EVENTSDeaths 0 0 0 0Explants 0 0 0 0Serious pacemaker related events outside the use of SafeR (AAI <> DDD)0 0 0 0Non-serious pacemaker related events outside the use of SafeR (AAI <> DDD)0 0 0 0Serious events due to the use of SafeR (AAI <> DDD)0 0 0 0Non-serious events related due to the use SafeR (AAI <> DDD)13 28.9 15 3.2Serious non-pacemaker related events6 13.3 9 1.9Non-serious non-pacemaker related events8 17.8 8 1.7(a) 4.74 device yearsNon-serious events due to the use of SafeR 2 (AAI <> DDD) included: delay in switching on 2nd degree AV block, inappropriate classification of a PAC, disagreement between markers and recorded EGM, atrial pacing above the maximum rate, recycling on an R-wave in a refractory   period,   and   disagreement   in   the   statistics   for   switches   to   DDD.   No   patient symptoms were associated with these events. SORIN – PLATINIUM DR – U460A 15
 6.  CLINICAL STUDIES 6. CLINICAL STUDIESClinical  data  presented  in this section are from the Defender and SafeR  (AAI  <> DDD) clinical studies.PLATINIUM DR is similar in design and function to the Defender devices.SafeR (AAI <> DDD) operation in PLATINIUM is similar to that in the Symphony pacemaker.The data provided are applicable to PLATINIUM DR.6.1. DEFENDER STUDYObjectives:The primary objectives of this study were to demonstrate a complication free rate (CFR) comparable to that of historical controls, to demonstrate, using a chronotropic assessment exercise protocol (CAEP), a rate response proportional to and appropriate for the level of exercise, and to evaluate and report the incidence of adverse events.Materials:Each patient received one Defender IV DR 612 defibrillator, an atrial pacing and sensing lead,   and   a   Medtronic,   Angeion,   or   Biotronik   defibrillation   lead   in   the   U.S.   or   any commercially available defibrillator lead outside the U.S.Methods:Investigators   selected   patients   who   survived   at   least   one   episode   of   cardiac   arrest (manifested by loss of consciousness) presumably due to a ventricular tachyarrhythmia or exhibited recurrent, poorly tolerated,  sustained ventricular tachycardia (VT).  The protocol required evaluation of performance and adverse events at pre-discharge, one month, three months, six months, and (in the U.S.) every three months thereafter. At the one-month visit, eligible patients performed a chronotropic assessment exercise protocol (CAEP) maximal exercise test.Study Population.The table below summarizes inclusions.Region Date of first implantDate of last implantData cut-off dateNumber of centersNumber of patientsUS 14-Dec-99 08-Mar-00 14-Mar-00 6 22Europe 04-May-99 26-Jul-99 14-Apr-00 11 38All 04-May-99 08-Mar-00 14-Apr-00 (Eur), 14-Mar-00 (US)17 606.1.1. Complication-free rateOnly European patients followed for at least 3 months:Symbol Parameter Defender IV DR 612N Overall number of patients 3716 SORIN – PLATINIUM DR – U460A
 6.  CLINICAL STUDIESPe*N Number of successes 35Pe Observed experimental proportion0.95Ps Null hypothesis success rate 0.76ES Estimated standard error of Pe 0.04z´ Test statistic (1) 4.75p Associated p-value < 0,0001(1) Statistical test: z´ = (Pe-Ps)/SE where SE = sqrt(Pe(1-Pe)/N)6.1.2. Rate responseEuropean patients only:GROUP Number of patients includedMean slope %SRR on %MRSTD of slopes %SRR on %MRSE of mean slope %SRR on %MRLower 95% CIUpper 95% CIEurope 20 0.77 0.17 0.04 0.69 0.84Small Centers9 0.79 0.18 0.06 0.67 0.91Large Centers11 0.75 0.15 0.05 0.66 0.84Males 17 0.77 0.16 0.04 0.70 0.85Females 3 0.73 0.22 0.13 0.47 0.98SRR: Sensor Rate ReserveMR: Metabolic ReserveSTD: Standard DeviationSE: Standard ErrorCI: Confidence Interval6.1.3. Adverse eventsEvent US (N=22) Number of events* Number of patients Percent of patientsIntent to treat but did not0 0 0.0Non-device related death0 0 0.0Explant 0 0 0.0Complication 0 0 0.0Observation 0 0 0.0Serious non-related other than death1 1 4.5SORIN – PLATINIUM DR – U460A 17
 6.  CLINICAL STUDIESEvent Europe (N=38) Number of events* Number of patients Percent of patientsIntent to treat but did not0 0 0.0Non-device related death1 1 2.6Explant 1 1 2.6Complication 3 2 5.3Observation 18 14 36.8Serious non-related other than death12 7 18.4Event All (N=60) Number of events* Number of patients Percent of patientsIntent to treat but did not0 0 0.0Non-device related death1 1 1.7Explant 1 1 1.7Complication 3 2 3.3Observation 18 14 23.3Serious non-related other than death13 8 13.3* A patient can have more than one complication, observation, or serious adverse event, not device-related.Device Failures and Replacements:No device failures or replacements occurred with Defender IV DR 612 during the study.6.2. SAFER (AAI <> DDD) STUDYSafeR (AAI <> DDD) mode in PLATINIUM is similar to that in Symphony.The differences in SafeR (AAI <> DDD) mode between the two devices are:─To prevent long RR intervals during VT/VF, SafeR (AAI <> DDD) has no effect during VT/VF therapy, electrophysiologic studies, and post-shock recovery.─The maximum acceptable AV delay for first degree AV block varies as a function of pacing rate.─PLATINIUM requires a ventricular sensed event to atrial paced event (RA) interval of at least 100 ms. Therefore, the device lengthens the atrial escape interval so that it ends at least 102 ms after the ventricular event.─During atrial fibrillation episode, pause criterion is fixed to 2s to avoid long bradycardia episodes in switching to DDD mode.18 SORIN – PLATINIUM DR – U460A
 6.  CLINICAL STUDIESDespite   these   differences,   the   data   collected   on   Symphony   devices   are   applicable   to PLATINIUM because the principles of SafeR (AAI <> DDD) operation did not change. The  criteria for switching from AAI to DDD (or vice versa) did not change. The device’s method for evaluating the presence of AV conduction did not change.Methods:All patients were implanted  with  a Symphony  Model 2550  dual-chamber  rate-responsive pacemaker with SafeR (AAI  <> DDD)  mode. A variety of  marketed atrial and ventricular pacing   leads   were   used.   The   pacemaker   was   programmed   and   interrogated   via   bi-directional telemetry using a Sorin dedicated programmer and a CPR3 programming head.The   study’s  routine   evaluation   consisted   of  enrollment,   pre-discharge   evaluation,   and   a scheduled follow-up visit at one month. At pre-discharge, a 24-hour Holter recording was performed and pacemaker memory was read. At one month, pacemaker memory was read. Investigators also documented adverse events.Patients studied:A total of 45 patients from 12 centers had Symphony 2550 pacemakers with SafeR (AAI <> DDD). Of these, 14 (31 %) were female and 31 (69 %) were male. Mean patient age (± SD)  was 74 ± 9 years.Primary indications for implant were: 1st degree AV block (11.1 %), 2nd degree AV block (6.7 %), 3rd degree AV block (22.2 %), sinus node dysfunction (62.2 %) or other (6.7 %).Effectiveness results:To determine the effectiveness of SafeR (AAI <> DDD) mode, the percentage of ventricular pacing provided over one month was recorded from pacemaker memory.Thirty-five   patients   contributed   data   to   evaluate   the   percentage   of   ventricular   pacing provided with SafeR (AAI <> DDD). Twenty-nine patients had 1 % or less ventricular pacing and six patients had a range of 28-97 % ventricular pacing. The graph below shows the distribution of ventricular pacing observed in patients with and without AV block as a primary indication for implant.The graph shows that many patients programmed to SafeR (AAI <> DDD) had less than 1% ventricular pacing:─84 % of patients without AV block at implant.─63 % of patients with AV block at implant.SORIN – PLATINIUM DR – U460A 19
 6.  CLINICAL STUDIESIn a representative reference group(1) of patients programmed to DDD, none had less than 1 % ventricular pacing and only 10 % had less than 90 % ventricular pacing regardless of AV block indication at implant.The actual reduction of ventricular pacing that SafeR (AAI <> DDD) provides in an individual will depend on the amount of time that the patient spends in AV block. SafeR (AAI <> DDD) cannot and should not provide any decrease in ventricular pacing while the patient is in AV block.(1) Pioger G, Jauvert G, Nitzsché R, Pozzan J, Laure H, Zigelman M, Leny G, Vandrell M,  Ritter   P,   and   Cazeau   S.   Incidence   and   predictive   factors   of   atrial   fibrillation   in   paced  patients. PACE, 28, Supp 1: S137-141; January 2005. This was a prospective observational  study of 377 patients with a functionally similar device programmed to DDD. The primary  indications  for   implant were:  AV block   (49 %),   sinus  node  disease  (16  %),  brady-tachy  syndrome (5 %), AV block + sinus node disease (19 %), AV block + brady-tachy syndrome (6 %), and brady-tachy syndrome + sinus node disease (5 %).Adverse eventsClinical study of the SafeR (AAI <> DDD) included 45 Symphony 2550 devices implanted in 45  patients.   No serious   adverse events   were  device-  or  feature-related.  There  were  no deaths in the study. Table 1 summarizes the safety data for this study.Table 1: Summary of Symphony safety data during studyPatients Number of eventsNumber of patients% of patients Number of eventsEvents per device year (a)Deaths 0 0 0 0Explants 0 0 0 0Serious pacemaker related events outside the use of SafeR (AAI <> DDD)0000Non-serious pacemaker related events outside the use of SafeR (AAI <> DDD)0000Serious events due to the use of SafeR (AAI <> DDD)0000Non-serious events related due to the use SafeR (AAI <> DDD)13 28.9 15 3.2Serious non-pacemaker related events6 13.3 9 1.9Non-serious non-pacemaker related events8 17.8 8 1.720 SORIN – PLATINIUM DR – U460A
 6.  CLINICAL STUDIES(a) 4.74 device yearsNon-serious events due to the use of SafeR 2 (AAI <> DDD) included: delay in switching on 2nd degree AV block, inappropriate classification of a PAC, disagreement between markers and recorded EGM, atrial pacing above the maximum rate, recycling on an R-wave in a refractory   period,   and   disagreement   in   the   statistics   for   switches   to   DDD.   No   patient symptoms were associated with these events. SORIN – PLATINIUM DR – U460A 21
 7.  PATIENT SELECTION AND TREATMENT 7. PATIENT SELECTION AND TREATMENT7.1. INDIVIDUALIZATION OF TREATMENTExercise stress testing:If the patient’s condition permits, use exercise stress testing to:─Determine the maximum rate of the patient’s normal rhythm,─Identify any supraventricular tachyarrhythmias,─Identify exercise-induced tachyarrhythmias.The   maximum   exercise   rate   or   the   presence   of   supraventricular   tachyarrhythmias   may influence selection of programmable parameters. Holter monitoring or other extended ECG monitoring also may be helpful.Electrophysiologic (EP) testing:EP testing may be useful for ICD candidates.EP   testing   may   identify   the   classifications   and   rates   of   all   the   ventricular   and   atrial arrhythmias, whether spontaneous or during EP testing.Drug resistant supraventricular tachyarrhythmias (SVTs):Drug   resistant   supraventricular  tachyarrhythmias   (SVTs)   may   initiate   frequent   unwanted device therapy.A careful choice of programming options is necessary for such patients.Antiarrhythmic drug therapy:If the patient is being treated with antiarrhythmic or cardiac drugs, the patient should be on a maintenance   drug  dose   rather   than  a   loading   dose  at   the   time   of   ICD   implantation.  If changes to drug therapy are made, repeated arrhythmia inductions are recommended to verify ICD detection and conversion. The ICD also may need to be reprogrammed.Changes in a patient’s antiarrhythmic drug or any other medication that affects the patient’s normal cardiac rate or conduction can affect the rate of tachyarrhythmias and/or efficacy of therapy.Direct   any   questions   regarding   the   individualization   of   patient   therapy   to   Sorin’s representative.22 SORIN – PLATINIUM DR – U460ACAUTION: To avoid inappropriate therapy during an exercise stress test, do not reprogram any parameter during the test. When a parameter is reprogrammed, algorithm forces acceleration to "ventricular". During conducted sinus tachycardia within the programmed Tachy zone, the device detects a 1:1 fast rhythm. Assuming that acceleration was set to ventricular by reprogramming, the device may identify this as a VT, and may immediately apply the corresponding therapy.
 7.  PATIENT SELECTION AND TREATMENT7.2. SPECIFIC PATIENT POPULATIONSPregnancy:If there is a need to image the device, care should be taken to minimize radiation exposure to the foetus and the mother.Nursing Mothers:Although appropriate biocompatibility testing has been  conducted for this implant device, there has been no quantitative assessment of the presence of leachables in breast milk.Pediatric Patients:This device has not been studied in patients younger than 18 years of age.Geriatric Patients:Most of the patients receiving this device in clinical studies were over the age of 60 years.Handicapped and Disabled Patients:Special care is needed in using this device for patients using an electrical wheel chair or other electrical (external or implanted) devices.SORIN – PLATINIUM DR – U460A 23
 8.  PATIENT COUNSELLING INFORMATION 8. PATIENT COUNSELLING INFORMATIONThe  physician should   consider  the  following  points  in counselling   the patient   about  this device:─Persons administering CPR may experience tingling on the patient’s body surface when the patient’s ICD system delivers a shock.─Advise patients to carry Sorin ID cards and/or ID bracelets documenting their ICD system.24 SORIN – PLATINIUM DR – U460A
 9.  DECLARATION OF CONFORMITY 9. DECLARATION OF CONFORMITYSorin declares that this device is in conformity with the essential requirements of Directive 1999/5/EC   on   Radio   and   Telecommunications   Terminal   Equipment,   with   the   mutual recognition of their conformity (R&TTE).Federal  Communication  Commission  Interference   Statement  47 CFR Section  15.19 and 15.105(b)The FCC product ID is :─PLATINIUM DR 1510: YSGDR1510─PLATINIUM DR 1540: YSGDR1540This equipment has been tested and found to comply with the limits for a Class B digital device,   pursuant   to   Part   15   of   the   FCC   Rules.   These   limits   are   designed   to   provide reasonable   protection   against   harmful   interference   in   a   residential   installation.   This equipment generates uses and can radiate radio frequency energy and, if not installed and used   in   accordance   with   the   instructions,   may   cause   harmful   interference   to   radio communications.   However,   there   is   no   guarantee   that   interference   will   not   occur   in   a particular installation.This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept   any   interference   received,   including   interference   that   may   cause   undesired operation.FCC Interference Statement 47 CFR Section 15.21 - No Unauthorized ModificationsIdentification of the equipment according Section 95.1217(a)This transmitter is authorized by rule under the Medical Device Radiocommunication Service (in part 95 of the FCC Rules) and must not cause harmful interference to stations operating in the 400.150-406.00 MHz band in the Meteorological Aids (i.e., transmitters and receivers used to communicate weather data), the Meteorological Satellite, or the Earth Exploration Satellite   Services   and  must   accept   interference   that   may   be   caused   by   such   stations, including interference that may cause undesired operation. This transmitter shall be used only in accordance with the FCC Rules governing the Medical Device Radiocommunication Service. Analog and digital voice communications are prohibited. Although this transmitter has been approved by the Federal Communications Commission, there is no guarantee that it will not receive interference or that any particular transmission from this transmitter will be free from interference.IC Requirements for CanadaThe FCC product ID is :─PLATINIUM DR 1510: 10270A-DR1510─PLATINIUM DR 1540: 10270A-DR1540This class B digital apparatus meets all requirements of the Canadian Interference- causing equipment regulations.SORIN – PLATINIUM DR – U460A 25CAUTION: This equipment may not be modified, altered, or changed in any way without signed written permission from SORIN. Unauthorized modification may void the equipment authorization from the FCC and will void the SORIN warranty.
 9.  DECLARATION OF CONFORMITYThis device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.Under   Industry   Canada   regulations,   this   radio   transmitter   may   only   operate   using   an antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry Canada. To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropically radiated power (e.i.r.p.) is not more than that necessary for successful communication.This device may not interfere with stations operating in the 400.150–406.000 MHz band in the Meteorological Aids, Meteorological Satellite, and Earth Exploration Satellite Services and must accept any interference received, including interference that may cause undesired operation.Cet appareil numérique de la classe B respecte toutes les exigences du règlement sur le matériel brouilleur du Canada.Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils radio exempts de licence. L’exploitation est autorisée aux deux conditions suivantes: (1) il ne doit pas produire de brouillage, et (2) l’utilisateur du dispositif doit être prêt a accepter tout brouillage radioélectrique reçu, même si ce brouillage est susceptible de compromettre le fonctionnement du dispositif.Conformément   à   la   réglementation   d’Industrie   Canada,   le   présent   émetteur   radio   peut fonctionner avec une antenne d’un type et d’un gain maximal (ou inférieur) approuvé pour l’émetteur   par   Industrie   Canada.   Dans   le   but   de   réduire   les   risques   de   brouillage radioélectrique à l’ intention d’autres utilisateurs, il faut choisir le type d’antenne et son gain de sorte que la puissance isotrope rayonnée équivalente (p.i.r.e.) ne dépasse pas l’intensité nécessaire à l’établissement d’une communication satisfaisante.Le présent dispositif ne doit pas causer de brouillage aux stations du service des auxiliaires de la météorologie, des satellites météorologiques, du service d’exploration de la terre par satellite, exploitées dans la bande 400,150-406,000 MHz, et il doit accepter tout brouillage reçu, y compris le brouillage pouvant entraîner un mauvais fonctionnement du dispositif.26 SORIN – PLATINIUM DR – U460A
 10.  PHYSICIAN GUIDELINES 10. PHYSICIAN GUIDELINES10.1. PHYSICIAN TRAININGPhysicians should be familiar with sterile pulse generator implant procedure and familiar with follow-up evaluation and management of patients with an implantable defibrillator (or referral to such a physician).10.2. DIRECTIONS FOR USEICD operating characteristics should be verified at the time of implantation and recorded in the patient file. Complete the Patient Registration Form and return it to Sorin, as it provides necessary information for warranty purposes and patient tracking.Additional programming instructions can be found by accessing Online Help (click the “?” on the   screen)   on   the   Sorin   dedicated   programmer.   Paper   copies  of   Online   Help   can   be obtained by contacting your Sorin representative.10.3. MAINTAINING DEVICE QUALITYThis device is FOR SINGLE USE ONLY. Do not resterilize and reimplant explanted ICDs.Do not implant the device when:─It has been dropped on a hard surface because this could have damaged pulse generator components.─Its sterility indicator within the inner package is not green, because it might not have been sterilized.─Its storage package has been pierced or altered, because this could have rendered it non-sterile.─It has been stored or transported outside the environmental temperature limits: 32 °F (0 °C) to 122 °F (50 °C) as an electrical reset condition may occur.─"Use Before" date has expired, because this can adversely affect pulse generator longevity or sterility.SORIN – PLATINIUM DR – U460A 27
 11.  PATIENT INFORMATION 11. PATIENT INFORMATIONInformation for the patient is available in the patient booklet, contained in the outer storage package. Additional copies can be obtained by contacting your Sorin representative.This information should be given to each patient with their first ICD and offered to the patient on each return visit or as deemed appropriate.28 SORIN – PLATINIUM DR – U460A
 12.  HOW SUPPLIED 12. HOW SUPPLIED12.1. STERILITYThe PLATINIUM defibrillators are supplied one per package in a sterile package.12.2. WARRANTY AND REPLACEMENT POLICYSorin warrants its defibrillators. Refer to the section "Warranty" for additional information. Please see the following labelling sections for information concerning the performance of this device: Indications, Contraindications, Warnings and Precautions, and Adverse Events.SORIN – PLATINIUM DR – U460A 29
 13.  DEVICE DESCRIPTION 13. DEVICE DESCRIPTIONThe   PLATINIUM   DR   ICD   device   and   programming   system.   The   programming   system includes   the   Sorin   dedicated  programmer   with   the   SMARTVIEW   programming   software connected   to  a   CPR3   programming  head.   The   programming   system   is   configured   and furnished by Sorin.The  PLATINIUM  DR can serve as a defibrillation  electrode  (active  housing)  with  a  total surface area of 63 cm².The   PLATINIUM   DR   is   designed   to   recognize   and   treat   slow   or   fast   VT   and   VF   by continuously   monitoring   atrial   and   ventricular   activity   to   identify   persistent   ventricular arrhythmias   and   to   deliver   appropriate   therapies.   PLATINIUM   DR   features   the PARAD/PARAD+   algorithm,   which   is   specifically   designed   to   differentiate   ventricular tachycardias   from   fast   rhythms  of   supraventricular   origin.   PARAD/PARAD+  continuously monitors   R-R   interval   stability,   searches   for   long   cycles,   assesses   the   degree   of   P-R association,   evaluates   sudden   onset   and   determines   the   chamber   of   arrhythmia acceleration.In addition to the advanced detection scheme, PLATINIUM DR offers programmable dual or single-chamber  pacing therapy (DDD,  DDI, VVI or SafeR  (AAI  <> DDD)  modes) with or without rate-responsive capabilities (DDDR, DDIR, VVIR, DDD/DDIR and SafeR-R (AAIR <> DDDR) modes) using an acceleration sensor. An automatic AV delay algorithm as well as a mode switching function are available.PLATINIUM DR offers tiered therapy. Therapies can be programmed independently in each zone:─in the Slow VT and VT zones: two ATP programs, up to two shocks with programmable energy and up to four shocks with maximum energy can be programmed;─in the VF zone: one ATP program, up to two shocks with programmable energy and up to four shocks with maximum energy can be programmed.When   the   rhythm   changes   from   one   zone   to   another,   the   device   delivers   the   therapy programmed in this zone, starting with the same or more aggressive program for the area. The ATP program in the VF zone will only be applied if the VT coupling interval is longer than the programmed fast VT cycle length.The PLATINIUM DR offers biphasic shocks with a maximum stored energy of 42 J. The shock configuration (electrodes used to apply the shock) can be chosen by programming one of the following combinations: can and one coil, can and two coils, two coils only.Other features are as follows:─Automatic ventricular sensitivity control─Non-committed shocks─Electrophysiological studies (EPS) with real-time markers or electrograms: ─Programmer-controlled VT induction sequences,─Programmer-controlled VF inductions (30 Hz rapid pacing or shock on T),─Programmable electrogram vectors (A / RV coil-CAN / SVC coil-CAN /RV coil-SVC coil /A ring-CAN / RV tip-CAN / RV ring-CAN) and RV EGM, ─Real-time annotations displayed with the markers and indicating the majority rhythm,─Manual ATP sequences,─Manual shocks.30 SORIN – PLATINIUM DR – U460A
 13.  DEVICE DESCRIPTION─Rescue shock─Follow-up tests: ─Pacing lead impedance,─Coil impedance,─Capacitor charge time,─Pacing threshold tests.─Data storage: ─Therapy History Report,─Statistics (pace/sense, therapy, shocks, and battery voltage),─Up to 16 complete Holter records with event logs, marker channel notation, and electrogram records.The PLATINIUM DR 1510 connector has four ports:─atrial bipolar pace/sense,─ventricular bipolar pace/sense and─wo ports for RV and SVC defibrillation coils.The PLATINIUM DR 1540 connector head has two ports:─atrial bipolar pace/sense,─one ventricular bipolar pace/sense and RV & SVC defibrillation coils.For PLATINIUM DR 1510, both pace/sense ports are compatible with the IS-1 standard and both defibrillation ports are compatible with the DF-1 standard.For   PLATINIUM   DR   1540,   pace/sense   port   is   compatible   with   the   IS-1   standard   and defibrillation ports is compatible with the DF4 standard.Distal   lead   terminal   connections   are   secured  with   set-screws   accessed   via   self-sealing silicone   plugs.   All   lead   connections   pass   through   the   header   into   the   device   via feedthroughs.Programming System:The   Sorin   programmer   is   used   in   conjunction   with   specific   programmer   software   to interrogate   and   program   the   implanted   device   at   implant   and   during   patient   follow-up procedures.Remote Monitoring:PLATINIUM DR is also equipped with the RF wireless technology which enables to remotely monitor the patients who have the Sorin SMARTVIEW Monitor installed at home.SORIN – PLATINIUM DR – U460A 31
 14.  IMPLANT PROCEDURE 14. IMPLANT PROCEDURE14.1. NECESSARY EQUIPMENTImplantation of PLATINIUM DR requires the following equipment:─Sorin ORCHESTRA programmer, equipped with the SMARTVIEW software interface and inductive telemetry head,─Sorin ORCHESTRA PLUS programmer, equipped with the SMARTVIEW software interface, inductive telemetry head and optionally ORCHESTRA PLUS LINK,─pacing system analyzer, as well as its sterile connecting cables, to evaluate the pacing and sensing thresholds,─a complete set of leads with corresponding introducers,─physiological signal monitor capable of displaying simultaneously the surface ECG and arterial pressure,─an external defibrillator with sterile external paddles,─sterile cover for the telemetry head.14.2. PACKAGING14.2.1. ContentsPLATINIUM DR and its accessories are ethylene oxide sterilized and hermetically sealed in two-ply clear packaging meeting international requirements.The sterile packaging contains:─the defibrillator─a ratcheting screwdriver─a DF-1 defibrillating connector insulating plug for 1510 modelOnce delivered, PLATINIUM DR is programmed to as-shipped values that are different from nominal values (see Chapter “Programmable Parameters” for details).14.3. OPTIONAL EQUIPMENTThe following equipment may be required during implantation of PLATINIUM DR:─an IS-1 insulating plug to close the atrial port─a DF4/DF-1 adaptor in case of replacement and use of DF-1 lead in a DF4 connector─sterile water to clean traces of blood. Any parts cleaned with sterile water must be thoroughly dried.─mineral oil to lubricate if necessary─a lead cap to isolate a lead which is not used32 SORIN – PLATINIUM DR – U460ANOTE: In case you’re implanting a DF4 lead, please verify its compatibility with standard alligators pin; please refer to the lead user’s manual for more details.
 14.  IMPLANT PROCEDURE14.4. BEFORE OPENING THE PACKAGEBefore opening the package, check the "Use Before" date printed on the labels on the box and  on the sterile  package.  Defibrillators  that have not  been  implanted  before  that date should be returned to Sorin.Interrogate the device: ─if a warning is displayed, do not implant the device and contact your Sorin representative.─if battery voltage is below 3V, and if the last reforming/charge occurred more than one week ago, do not implant the device. Otherwise wait for one more week before checking the voltage.NOTE : The battery voltage can decrease before the expiration date is reached. However, the battery voltage should be equal to or higher than 3V at the time of implant.Devices MUST NOT be interrogated and programmed within the vicinity of other devices.Also check the  integrity of the  sterile package.  The sterility of the  contents is no longer guaranteed if the package has been pierced or altered. If the defibrillator is no longer sterile, it  should   be   returned in   its packaging  to   Sorin.  Any re-sterilization   of the   unit  is  at  the discretion of Sorin.14.5. PRIOR TO IMPLANTATIONUse the programmer to verify the defibrillator can be interrogated before implantation.Verify   all   shock   therapies   are   disabled   in   order   to   avoid   accidental   discharge   during implantation.It   is   not   advisable   to   program   the   Smoothing   function   before   implantation,   since   the defibrillator may detect noise and pace at a rate higher than the programmed basic rate.14.6. DEVICE PLACEMENTThe   pocket   should   be   prepared   in   the   left   pectoral   position,   either   subcutaneously   or submuscularly.   Subcutaneous   device   implantation   is   recommended   for   optimal   RF communication efficacy.Implantation in an abdominal position is not advisable.In its final position, the defibrillator should be no more than 4 cm below the skin surface.14.7. CHOOSING THE TYPE OF LEADThe defibrillator should be connected to:─one bipolar atrial sensing/pacing lead─one right ventricular lead with bipolar sensing/pacing electrodes and one or two SORIN – PLATINIUM DR – U460A 33CAUTION:Do not shake or tap sharply on the ICD package with the ICD inside, because the ICD's sensing circuits can interpret this as P-waves or R-waves and record these as an arrhythmia episode.High voltage capacitors charge performed on ICD without connected leads using wireless telemetry can generate false P-waves or R-waves detection. It is recommended to reset the memory data and statistics before implanting the ICD.
 14.  IMPLANT PROCEDUREdefibrillation coilsThe choice of leads and their configuration is left to the implanting physician’s judgment.Connectors:PLATINIUM DR 1510:The   pacing/sensing   connectors   are   compatible   with   the   IS-1   standard   and   the   right ventricular defibrillation connectors are compatible with DF-1 standard.PLATINIUM DR 1540:The   atrial   connector   is   compatible   with   the   IS-1   standard   and   the   quadripolar   right ventricular connector is compatible with the DF4 standard.14.8. SHOCK CONFIGURATION (+ -> -)The shock configuration is the energy pathway between the defibrillation electrodes. If an atrial coil is present, the shock configuration can be programmed for bi-directional shocks.Programming:When active case and SVC are both programmed to Yes, the shock configuration can be programmed to: ─RV to Case (or Case to RV),─or RV to SVC (or SVC to RV),─or RV to Case+SVC (or Case+SVC to RV).        RV to Case+SVC RV to Case RV to SVCThe polarity of shock is determined by the parameter itself.34 SORIN – PLATINIUM DR – U460ANOTE1: Please note that DF-1 standard compliant lead is not compatible with DF4 connector. In the reverse, DF4 standard compliant lead is not compatible with DF-1 connector. In case of defibrillator replacement, choose the appropriate device compatible with DF-1 or DF4 leads. For any other lead type that requires an adaptor for this device, please contact your Sorin representative for any information on lead / connector compatibility question.NOTE2: In the event that no atrial lead is implanted, the atrial port should be plugged with IS-1 insulating plug and a single chamber mode (VVI-VVIR) should be programmed. PARAD and PARAD+ should not be used.NOTE3: In the event of a warning on a low shock impedance, and after lead replacement or reconnection: it is recommended to check the system integrity (sensing and pacing thresholds and the impedance of the shock electrodes).
 14.  IMPLANT PROCEDURE14.9. MEASUREMENT OF THRESHOLDS AT IMPLANTPacing and sensing thresholds should be measured at implant.Pacing thresholds:Acute  thresholds  should  be lower than 1 V  (or  2 mA)  for  a 0.35 ms  pulse  width, in the ventricle and in the atrium.Sensing thresholds:For appropriate ventricular sensing, the amplitude of the R-wave should be greater than 5 mV.For appropriate atrial sensing, the amplitude of the P-wave should be greater than 2 mV.Pacing impedance measurements:Ventricular and atrial pacing impedances should range from 200 to 3000 ohms (refer to the lead characteristics, especially if high impedance leads are used).Please   refer   to   the   leads   user   manuals   for   more   details   on   the   expected   electrical performances of the leads.14.10. LEAD CONNECTION Implant the ventricular lead, then the atrial lead.Each lead must be connected to the corresponding connector port. The position of each connector is indicated on the casing.Tighten only the distal inserts. To connect each lead, proceed as follows:1. Clean the lead terminal pins thoroughly, if necessary (device replacement).2. Lubricate the lead terminal pins with sterile water, if necessary.3. Do not insert a lead connector pin into the connector block without first visually verifying that the lead port is not filled with any obstacle.4. Insert the screwdriver into the pre-inserted screw socket of the appropriate port (in order to allow excess air to bleed out and to make the insertion of the lead pin easier).5. Insert the lead pin all the way into the port (check that the pin protrudes beyond the distal  insert).6. Tighten, check the tightness and ensure the lead pin still protrudes beyond the distal insert, and has not move.1. Do not tighten the pre-inserted screws when there is no lead (this could damage the connector). 2. Do not loosen the screws before inserting the connector (subsequent risk of being unable to reinsert the screw).3. When mineral oil or sterile water is used to make lead insertion easier, the screwdriver should remain inserted into the pre-inserted screw socket when checking the tightness. SORIN – PLATINIUM DR – U460A 35CAUTION:CAUTION:
 14.  IMPLANT PROCEDUREAs a matter of fact, when the lead port is filled with a liquid, the physics piston effect can  give the feeling the lead is properly tightened. 4. One single set screw is located on the side of the connection header.5. Use  only  the  screwdriver  provided  with   the  defibrillator.  Keep   the  screwdriver's  shaft perpendicular to the plane of the defibrillator (see figure below).6. Removing the screwdriver: to avoid all risk of loosening screws during removal, hold the screwdriver by its metal part and not by the handle.To ensure full insertion, push the screwdriver's hex tip smoothly into the setscrew until it reaches the bottom of the hex chamber in the screw, which can be felt as a solid metallic contact. Do not implant the defibrillator if there is no feeling of solid metallic contact. Do not implant the defibrillator if the wrench does not click when attempting to tighten the setscrew on the lead pin.In the  case of an external  defibrillation  shock delivered to the  patient, always check the programming and functioning of the device, in particular its capacity to deliver shocks.14.11. DEVICE IMPLANTATIONPLATINIUM   DR  should  be  implanted  with  the   device  identification  engraved   side  facing outwards   for   optimal   communication   with   the   programming   head   and   radiographic identification.Carefully wind excess lead and place in a separate pocket to the side of the defibrillator.It is recommended not to place any excess wire between the can and the heart.Suture the casing connector to the muscle using the hole provided for this purpose, in order to avoid potential migration of the device into the pectoral muscle.36 SORIN – PLATINIUM DR – U460AWARNING: Ensure that the screwdriver's tip is fully inserted in the setscrew; otherwise the screwdriver can damage the setscrew and prevent connection with or disconnection from the lead.NOTE: To optimize cardioversion/defibrillation shocks, electrodes must be positioned so that the electric field between anode(s) and cathode covers the largest myocardial mass. In normal conditions, the anode and cathode are adequately separated. In case of a short-circuit, the shock may be aborted to prevent damaging the defibrillator.
 14.  IMPLANT PROCEDURE14.12. TESTS AND PROGRAMMINGDuring the implant testing procedure:It is recommended that a security margin of at least 10 J be demonstrated between the effective shock energy and maximum programmable energy.Enable shock therapies, then program the defibrillator.Verify that the defibrillation lead impedance for each shock delivered is within the range of 30 to 150 ohms. Check the lead connection if the values are outside these boundaries.Save the programming data  on the programmer’s hard  disk  and on  an external storage device (if desired).Resuscitation Availability:Do   not   perform   device   testing   unless   an   external   defibrillator   is   available   and  medical personnel skilled in cardiopulmonary resuscitation (CPR) are present.Disable the ICD During Handling:Program   Shock   Therapy   to   OFF   during   surgical   implant   and   explant   or   post   mortem procedures. The device can deliver a serious high energy shock should accidental contact be made with the defibrillation electrodes, the device can deliver a very high energy shock.SORIN – PLATINIUM DR – U460A 37
 15.  SPECIAL MODES 15. SPECIAL MODES15.1. SAFETY MODE (NOMINAL VALUES)Nominal   values   may   be   rapidly   restored   by   pressing   the   following   button   on   the programming head or programmer keyboard:or via the Emergency button on the SMARTVIEW screen.In  safety  mode,  the  defibrillator  operates  with  the  parameters  underlined in the table of programmable parameters.15.2. MAGNET MODEWhen the magnet is applied:─antiarrhythmia functions are inhibited (detection of rhythm disturbances, charging, and therapy),─hysteresis and AVD paced/sensed offset are set to 0,─pacing amplitude is set to 6 V,─pulse width is set to maximum,─the following functions are disabled: ventricular arrhythmia prevention, Mode Switch, Anti-PMT, Smoothing, Rate Response.When the magnet is removed:─arrhythmia detection algorithms and sequential therapies are reinitialized,─therapies start with the least aggressive program for each area.The   antiarrhythmia   functions   inhibition   is   extended   after   magnet   removal   if   a   charge occurred  just  before   the application   of the  magnet  in  order   to ease   the  communication between the device and the programmer.The other parameters remain at their programmed value.15.3. RESPONSE IN THE PRESENCE OF DISTURBANCEIf the defibrillator  senses electrical  noise  at a frequency above  16  Hz,  it switches to  an asynchronous mode at the basic rate. The programmed mode is restored as soon as the noise is no longer detected.Ventricular pacing may also be inhibited by ventricular noise. It can be restored by setting the parameter V pacing on noise to Yes.15.4. DETECTION CHARACTERISTICS IN THE PRESENCE OF ELECTROMAGNETIC FIELDSPer Clause 27.4 of Standard EN 45502-2-2, detection in the presence of electromagnetic fields is characterized as follows:─Differential mode:For DF-1 models:38 SORIN – PLATINIUM DR – U460A
 15.  SPECIAL MODESFor DF4 models:─Common mode rejection ratio:For DF-1 Models:16.6 Hz 50 Hz 60 HzAtrial channel ≥74 dB ≥74 dB ≥74 dBVentricular channel ≥68 dB ≥68 dB ≥68 dBFor DF4 Models:16.6 Hz 50 Hz 60 HzAtrial channel ≥74 dB ≥74 dB ≥74dBVentricular channel ≥68 dB ≥68 dB ≥68dBFor atrial sensitivity settings below 0.4mV, the ICD may detect noise lower than the level specified in clause 27.5.1 of standard EN 45502-2-2 for frequencies below 200 Hz.For ventricular sensitivity settings below 0.6mV, the ICD may detect noise lower than the level specified in clause 27.5.1 of standard EN 45502-2-2 for frequencies below 200 Hz.SORIN – PLATINIUM DR – U460A 39
 15.  SPECIAL MODES15.5. PROTECTION AGAINST SHORT-CIRCUITSThe defibrillator can undergo a short-circuit if the anode and cathode are not adequately separated.In this case, the shock is aborted to prevent damaging the defibrillator and a warning will indicate  that  a short circuit  (shock impedance  <  20 ohms) was detected during  the last shock.40 SORIN – PLATINIUM DR – U460A
 16.  MAIN FUNCTIONS 16. MAIN FUNCTIONS16.1. AUTOMATIC LEAD MEASUREMENTSAutomatic pacing lead impedance measurement:A lead impedance measurement is automatically performed on atrial and ventricular leads every 6 hours. The daily mean impedance is stored for each chamber.Automatic coil impedance measurement:A continuity measurement is automatically performed on defibrillation coil(s) once per day. The continuity is stored for each coil.Automatic sensing measurement:The  amplitude  of  P and  R  waves are  automatically  measured at  each cycle.  Every  8.5 minutes, the amplitude of the last 8 P and R detections are averaged and stored.16.2. ATRIAL TACHYARRHYTHMIA MANAGEMENTMode Switch:This function is  designed to  limit the acceleration and  variation  of ventricular rate in the presence of atrial arrhythmia.16.3. VENTRICULAR TACHYARRHYTHMIA MANAGEMENTVentricular tachyarrhythmia prevention:Set of algorithms that can be used to avoid the circumstances of ventricular tachyarrhythmia onset.Arrhythmia discrimination algorithm PARAD and PARAD+ (P And R based Arrhythmia Detection):PARAD is the algorithm used to discriminate sinus tachycardias (ST) and supraventricular tachycardias (SVT) from ventricular tachycardias (VT).PARAD+   is   based   on   the   PARAD   algorithm   but   additionally   takes   into   account   the “AFdetect” discrimination criteria: the occurrence of a "long ventricular cycle" characteristic for   AF   patients   which   is   an   additional   arrhythmia   classification   criterion   to   improve identification of atrial fibrillation and avoid inappropriate shocks.Fast VT treatment:Applies detection criteria on fast ventricular tachycardia that are different from those of the VT zone, as well as different therapies. The fast VT zone is included in the VF zone: its lower limit is determined by the programmed value for the VF zone and its upper limit by the  programmed value for the fast VT zone.Polarity alternation on Max shock:Reverses  the  programmed  polarity  of every second  shock  set  at maximum energy. The number, type, and energy of shocks is independently programmable by detection zone.SORIN – PLATINIUM DR – U460A 41
 16.  MAIN FUNCTIONSDefibrillation threshold (DFT):Be aware that the changes in the patient’s condition, drug regimen, and other factors may change   the   defibrillation   threshold   (DFT)   which   may   result   in   non-conversion   of   the arrhythmia  post-operatively.   Successful  conversion   of  ventricular fibrillation   or ventricular tachycardia during arrhythmia conversion testing is no assurance that conversion will occur post-operatively.16.4. PACINGBTO (Brady Tachy Overlap):Corrects chronotropic atrial incompetence by allowing pacing in the slow VT zone, without affecting arrhythmia detection specificity.Post-shock mode:After any automatic shock therapy, the post-shock mode makes it possible to apply a pacing mode  other  than  the standard  antibradycardia  pacing  mode  and/or with  different pacing parameters.SafeR (AAI <> DDD) mode:Is intended to minimize deleterious effects of ventricular pacing. The defibrillator functions in AAI mode, and temporarily switches to DDD mode upon the occurrence of AVB III, AVB II, AVB I and ventricular pause.Anti-PMT protection:Is  intended   to protect   the   patient   from   Pacemaker-Mediated  Tachycardia   (PMT)  without reducing atrial sensing capability of the device.16.5. SENSINGAutomatic Refractory Periods:Optimize sensing and make the implant programming easier. These periods are composed of  a minimal Refractory Period and a triggerable  Refractory Period. The  duration  of  the refractory periods lengthens automatically as needed.Committed period:In  DDI  or  DDD  modes,  the committed  period   is  a  non-programmable  95 ms  ventricular relative refractory period that starts with atrial pacing. If a ventricular event is sensed during the committed period, but outside the blanking period, the ventricle is paced at the end of the committed period. The committed period prevents inappropriate ventricular inhibition if crosstalk occurs. Protection against noise:Allows   the   distinction   between   ventricular   noise   and   ventricular   fibrillation.   If   the   device senses  ventricular  noise,  the  ventricular  sensitivity  is  decreased  until  noise  is  no longer detected. Ventricular pacing can be inhibited to avoid a potential paced T-wave.Automatic sensitivity control:Optimizes arrhythmia detection and avoids late detection of T-waves and over-detection of wide QRS waves. The device automatically adjusts the sensitivities based on the ventricular sensing amplitude. In case of arrhythmia suspicion or after a paced event, the programmed ventricular sensitivity will be applied. The minimum ventricular sensitivity threshold is 0.4 mV (minimum programmable value).42 SORIN – PLATINIUM DR – U460A
 16.  MAIN FUNCTIONS16.6. FOLLOW-UP FUNCTIONStorage of memory data:AIDA+ (Automatic Interpretation for Diagnosis Assistance) software provides access up to 6 months of patient follow-up with day by day data collection, or up to 24 hours with hourly data   collection.   Episodes   of   ventricular   tachyarrhythmia   are   recorded   with   one programmable EGM channel which can be selected, in addition to RV EGM.Diagnosis of AV conduction:Automatic diagnosis of AV conduction with graphic displays.Alerts / Warnings:The device routinely performs security self-checks and technical measurements to ensure system integrity. When system integrity is found to be at risk outside a follow-up, alerts are stored in the device memory. When system integrity is found to be at risk during a follow-up, the information is managed by a warning (pop-up message) to immediately notify the user. For   example,   the   following  types   of   event   can   trigger   a   warning   or   an   alert:  technical problem during a shock, lead impedance or shock continuity measurements out-of-range, battery depletion, …16.7. REMOTE MONITORING FUNCTIONRemote   monitoring   enables   the   automatic   remote   transmission   of   implant   data   to   the physician thanks to the wireless Radio Frequency (RF) communication ability of the implant in order to provide a comprehensive report to the physician about device functioning and patient cardiac status without having the patient physically in the clinic.The data is transmitted from the implant to the SMARTVIEW monitor, a small transmitter placed in the patient's home.Implant data are first transmitted to the SMARTVIEW monitor via RF. Data are then rooted through the phone line or via GPRS to an internet website. This website is responsible for transforming the implant data into  a comprehensive  report that  can be  consulted by the physician.16.7.1. SMARTVIEW MonitorThe SMARTVIEW monitor is a small device equipped with an RF transmission module to communicate with the implant and a modem to export data through the internet.The SMARTVIEW monitor is delivered to the patient who has to install it at home. Preferably the   SMARTVIEW   monitor  will   be   placed   on   the   nightstand   of  the   patient,   as   close   as possible to the side of the bed where the patient usually sleeps. The SMARTVIEW monitor connects to the phone line of the patient and the power plug. Regular transmissions are done during the night when the patient is asleep next to the SMARTVIEW monitor without  any intervention from the patient. 16.7.2. Transmission triggerThere are 3 different triggers for a remote transmission:─the remote follow-up transmission is scheduled by the physician to occur regularly (according to the programming).─the alert transmission will take place when the implant has recorded an abnormal event. The list of abnormal event is available in a following paragraph. Alert conditions SORIN – PLATINIUM DR – U460A 43
 16.  MAIN FUNCTIONSare checked daily.─the on-demand follow-up transmission is triggered by the patient himself through the use of a specific button on the SMARTVIEW monitor.16.7.3. Data transmittedThe data transmitted are identical to the data available during a standard interrogation with the dedicated programmer. All counters, histograms, IEGMs and diagnosis available in the device are transmitted containing (not exhaustive list):─programmed parameters─information on patient and system implanted─battery status─lead status (brady leads and defibrillation coils) ─pacing counters and mean heart rate (brady)─atrial and ventricular arrhythmia counters and episodes─ventricular therapy countersData   are   presented  in  the   form   of   2  reports   to   the   physician:   the   first  one   contains   a summary of major counters, histograms, warnings and diagnosis. The second one presents the most important IEGM episodes automatically selected based on the degree of severity for the patient.16.7.4. User websiteOn the website, the physician is able to:─consult and schedule the remote follow-ups of their patient─configure additional ways of being notified of alerts (for instance by SMS, fax or e-mail)─consult, print and export patient reports 16.7.5. Alert systemThe   following   set   of   alert   trigger   can   be   independently   programmed   ON/OFF   by   the physician using the dedicated programmer and can trigger an alert transmission:─Low or high impedance (A, V)─Abnormal coil impedance (shock lead)─Low or High shock impedance─Inefficient high energy shock─All shocks programmed OFF─Shock treated VT/VF─ATP treated VT/VF─Suspicion of noise on the V lead─AF occurrence─Fast V rate during AFThe following set of alert trigger (system alerts) cannot be deactivated when the Alerts are programmed “On” and can trigger an alert transmission:─Battery depletion – RRT─Device reset─Excessive charge time (>25s)44 SORIN – PLATINIUM DR – U460A
 16.  MAIN FUNCTIONS─System integritySORIN – PLATINIUM DR – U460A 45WARNING: The use of remote monitoring does not replace regular follow-up. Therefore, when using remote monitoring, the time period between follow-ups visits may not be extended.
 17.  PATIENT FOLLOW-UP 17. PATIENT FOLLOW-UP17.1. FOLLOW-UP RECOMMENDATIONSBefore the patient is discharged and at each subsequent follow-up visit, it is advisable to:─check the occurrence of system warnings─check the battery status,─check the integrity of the pacing and defibrillation leads,─check for proper sensing (sensitivity, crosstalk) and pacing ; set the pacing amplitude to twice the pacing threshold,─interrogate the implant memories (AIDA+),─check the efficacy of the therapies delivered,─keep a printout of programmed parameters, test results, and memory data,─reset the memory data and statistics.These operations should be performed by medical personnel in an appropriate care unit, with resuscitation equipment present.It is recommended that a routine follow-up examination be done one month after discharge, and then every three months until the device nears the replacement date.Refer to the online help for a description of displayed warning, and the necessity to contact Sorin for an evaluation.Implant software upgrade:In   case   a   new   implant   software   is   downloaded   in   the   device   memory   through   the programmer, a warning message could be displayed by the programmer to inform the user and give the correct instructions to follow.17.2. HOLTER FUNCTIONThe Holter records markers and EGM on RV and on 1 programmable channel: A / RV coil – Can / SVC coil – Can / RV coil - SVC coil / RA ring – Can / RV tip – Can / RV ring – Can─Up to 10 episodes and 5 min EGM on significant events: AV block switch, lead impedance out of range. ─Up to 16 tachyarrhythmia episodes as well as the therapy history.46 SORIN – PLATINIUM DR – U460ANOTES: If the last reforming, charge or shock occurred during the week preceding the interrogation, the last battery value may be still impacted by the event. One week post event, the battery will recover its steady state value.Automatic capacitor charging may affect communication between the device and the programmer.
 17.  PATIENT FOLLOW-UPStored Tachyarrythmia Episodes:PLATINIUM DR stores up to 16 episodes (VF, VT, Slow VT, SVT/ST, non-sustained) with a total of 25.6 min of high resolution EGM.For each episode four levels of details are presented:─Tachogram (to visualize PP and PR intervals)─Event log for the entire episode: ─PARAD/PARAD+ analysis for each majority,─Delivered therapies,─Markers: Atrial and ventricular markers, sensed, paced and in relative refractory periods,─EGM: onset and detection of the arrhythmia, on two therapies, and the return to slow rhythm by recording electrogram.Therapy historyFor each  arrhythmia detection,  each  therapy delivered (either automatically or  during an electrophysiological study) and at the end of each arrhythmia, PLATINIUM DR records the type   of   majority   rhythm,   the   number   of  ATP   sequences   delivered,   the   energy   and   the number of shocks delivered.17.3. RECOMMENDED REPLACEMENT TIME (RRT)Recommended Replacement Time (RRT)(1) is controlled by: battery voltage equal to 2.62 V ± 0.01 VBetween the RRT and the EOS (End of Service)(2), PLATINIUM DR can still function for:─12 months (100% atrial and ventricular pacing in DDD mode, 500 ohms, with as shipped settings), and deliver 13 shocks at 34 J or─10,4 months (0% pacing, sensors OFF, one shock every 2 weeks) and deliver 21 shocks at 34 J. Once  the  Recommended  Replacement  Time  (RRT)  point  has  been  reached,  the  device operates  normally, except that  the charge  time increases. Under  normal conditions  (and without programmer use) the charge times are as follows:Shock energy Charge time (sec)BOS(3) 42 J 10 (± 2)RRT 42 J 13 (± 3)(1)   Recommended   Replacement   Time   (RRT)   corresponds   to   Elective   Replacement Indicators (ERI) previously used.(2) End of Service (EOS) corresponds to End of Life (EOL) previously used.(3) Beginning of Service (BOS) corresponds to Beginning of Life (BOL) previously used.17.4. EXPLANTATIONThe defibrillator should be explanted in the following cases:─The Recommended Replacement Time (RRT) point is reachedSORIN – PLATINIUM DR – U460A 47CAUTION: The defibrillator should be replaced as soon as the Recommended Replacement Time (RRT) point is reached.
 17.  PATIENT FOLLOW-UP─Confirmed malfunction─Burial of the patient (for environmental reasons, the local regulation may require the explantation of the devices containing a battery supply)─Cremation of the patient (the defibrillator may explode if placed in an incinerator)The explanted defibrillator should not be reused in another patient.All explanted defibrillators should  be  returned  to Sorin, carefully cleaned  of  all  traces of contamination.   Cleaning   may   be   done   by   immersing   them   in   an   aqueous   sodium hypochlorite containing at least 1% chlorine, followed by rinsing copiously with water.The   defibrillator   should   be   protected   against   mechanical   impact   and   the   temperature variations that may occur during shipping.Before explantation, it is advisable to:─Print out all programmed parameters, statistics and Holter function report,─Save Patient data on floppy disk or hard disk,─Disable shock therapies (VT and VF) to avoid any risk of untimely shock.17.5. DEFIBRILLATOR IDENTIFICATIONThe defibrillator can be interrogated and programmed via telemetry, using the programming head interfaced with the Sorin dedicated programmer.Position the programming head over the telemetry antenna located in the upper part of the device, in order to communicate effectively via telemetry (see diagram below).The device can be non-invasively identified as follows:1. Take an x-ray to identify the name of the manufacturer and model, printed on the device  (X-ray ID is SEA for PLATINIUM range). 2. Interrogate  the   device   using   the   Sorin dedicated  programmer.  The   model  and  serial number of the device are automatically displayed. The first figure in the serial number corresponds to the last figure in the year of manufacture.48 SORIN – PLATINIUM DR – U460A
 18.  PHYSICAL CHARACTERISTICS 18. PHYSICAL CHARACTERISTICS1510 MODEL:Dimensions 73 x 54.3 x 11.1 mmWeight 86 gVolume 33 cm3Active surface area of casing 62.7cm2Connector Atrium: IS-1 bipolar. Ventricle: IS-1 bipolar, 2*DF-1.1540 MODEL:Dimensions 72.3 x 54.3 x 11.1 mmWeight 86 gVolume 33 cm3Active surface area of casing 62.8cm2Connector Atrium: IS-1 bipolar. Ventricle: DF4.18.1. MATERIALS USEDActive surface area of casing 99% pure titaniumConnectors Polyurethane* and silicone elastomer**Medical-grade materials that have undergone "in vitro" and "in vivo" qualifications.SORIN – PLATINIUM DR – U460A 49
 19.  ELECTRICAL CHARACTERISTICS 19. ELECTRICAL CHARACTERISTICSAtrial input impedance 80 kilohms ± 30 %Ventricular input impedance 80 kilohms ± 30 %D.C. capacitance 149 µF ± 8 %Capacitor reformation No reformation requiredRate limit 192 ppm ± 10 ppmPacing waveformDefibrillation waveform19.1. TABLE OF DELIVERED SHOCK ENERGY AND VOLTAGEThe relationship between stored energies, maximum voltages and delivered energies (at 37 °C, 50 ohm load) for the minimum, low, mean and maximum programmed energy values is as follows:Stored energy (J) 0.5 10 20 34 42V1 (Volt) 71 332 471 617 686V2 (Volt) 35 167 235 309 342Delivered E: Phase 1 (J) 0.32 6.94 14.0 23.8 29.6Delivered E: Phase 2 (J) 0.08 1.75 3.4 6.0 7.4Delivered E: Total (J) 0.4 8.7 17.4 30 37Tolerances are 12% for voltage (25% at 0.5 J) and 30% for energy.19.2. BATTERYManufacturer GreatbatchType Quasar High Rate (QHR)Model GB 3070Number of batteries 150 SORIN – PLATINIUM DR – U460A
 19.  ELECTRICAL CHARACTERISTICSTotal capacity 2192 mAhUsable capacity Between BOS and RRT: 1530 mAh. Between BOS and EOS: 1910 mAh.Voltage BOS: 3.24 V. RRT: 2.62 V. EOS: 2.5 V.19.3. LONGEVITYThe longevities are calculated by taking into account 6 months storage with the following conditions:─Basic rate: 60 ppm─Pulse width (A, RV, LV): 0.35 ms─EGM:ON─2 battery reformings per year (at 34J), replaced by shocks if any─Remote monitoring: ON, daily check, 4 follow-ups and 5 full alert reports per year─RF telemetry: ON, 45min at implantation + 15min at discharge + 15min for in-clinic quarterly follow-upsLongevity projection at 500 Ω pacing impedance:Mode DDD DDD DDD DDD DDD SafeR -A pacing (%) 100 100 15 15 15 30 0V pacing (%) 100 100 1 15 1 6 0Pacing amplitude (V) 3.5 3.5 3.5 4.5 2.5 2.5 -Pulse width (ms) 0.35 0.35 0.35 0.50 0.35 0.35 -Sensor OFF ON OFF OFF OFF OFF OFFMax shocks (42J) per year 4 4 4 4 0 4 4Longevity (years) 8.6 8.4 13.1 11.6 17 13 13.7Longevity projection at 600 Ω pacing impedance:Mode DDD DDD DDD DDD DDD SafeR -A pacing (%) 100 100 15 15 15 30 0V pacing (%) 100 100 1 15 1 6 0Pacing amplitude (V) 3.5 3.5 3.5 4.5 2.5 2.5 -Pulse width (ms) 0.35 0.35 0.35 0.50 0.35 0.35 -Sensor OFF ON OFF OFF OFF OFF OFFMax shocks (42J) per year 4 4 4 4 0 4 4Longevity (years) 9.1 8.9 13.2 11.8 17 13.1 13.7SORIN – PLATINIUM DR – U460A 51
 19.  ELECTRICAL CHARACTERISTICSLongevity projection at 700 Ω pacing impedance:Mode DDD DDD DDD DDD DDD SafeR -A pacing (%) 100 100 15 15 15 30 0V pacing (%) 100 100 1 15 1 6 0Pacing amplitude (V) 3.5 3.5 3.5 4.5 2.5 2.5 -Pulse width (ms) 0.35 0.35 0.35 0.50 0.35 0.35 -Sensor OFF ON OFF OFF OFF OFF OFFMax shocks (42J) per year 4 4 4 4 0 4 4Longevity (years) 9.5 9.3 13.3 12.0 17.1 13.2 13.71h of additional RF programming session reduces the device longevity from 1 to 2 weeks depending on the device functioning mode (no pacing, 100% pacing). The mean longevity as a function of shocks delivered at maximum energy, with and without pacing, is as follows:The mean longevity as a function of yearly remote follow-ups1, with and without pacing, is as follows:1. An excessive number of remote follow-ups can have a non-negligible impact on device longevity.52 SORIN – PLATINIUM DR – U460A
 20.  PROGRAMMABLE PARAMETERS 20. PROGRAMMABLE PARAMETERSMeasured at 37 °C under a 500 ohm loadLegend:Value in bold: "as shipped" valueUnderlined value: nominal value20.1. ANTIBRADYCARDIA PACINGBasic parameters ValuesMode VVI-VVIR-DDD-DDDR-DDD/DDIR-DDI-DDIR-SafeR (AAI <=> DDD)-SafeR-R (AAIR <=> DDDR)Basic rate (ppm) (1) From 30 to 90 by steps of 5 ; 60 (± 4 %)Maximum rate (ppm) From 100 to 145 by steps of 5 ; 120 (± 6 %)Rate hysteresis (%) 0-5-10-20-35 (± 18 ms)Rest AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms)Exercise AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms)AVD Paced/Sensed Offset (ms) 0-10-15-25-30-40-45-55-65-70-80-85-95-100-110-115-125 (± 1 ms)(1) The corresponding periods are (in ms): 2000-1714-1500-1333-1200-1091-1000-923-857-800-750-706-667 ms.Special features ValuesRate smoothing OFF-Very slow-Slow-Medium-FastMode Switch ON-OFFPhysical activity Very low-Low-Medium-High-Very highSORIN – PLATINIUM DR – U460A 53
 20.  PROGRAMMABLE PARAMETERSPacing/Sensing ValuesAtrial sensitivity (mV) (1) From 0.2 to 4 by steps of 0.2 ; 0.4 (± 50 %)Atrial amplitude (V) (2) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)Atrial pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)Ventricular sensitivity (mV) (1) From 0.4 to 4 by steps of 0.2 ; 0.4 (± 50 %)Ventricular amplitude (V) (2) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)Ventricular pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)(1) Values are measured using a positive and negative triangular signal of 2/13 ms.(2) The  correlation  between  the   programmed  amplitudes,   the  stored  amplitudes  and  the  mid-pulse delivered amplitudes under a 500 ohm load are given in the following table:Programmed amplitude (V) Stored amplitude (V) Mid-pulse delivered amplitude (V)1 1,11 0,941,5 1,63 1,382 2,1 1,782,5 2,6 2,23 3,15 2,673.5 3,65 3,094 4,2 3,554,5 4,69 3,975 5,25 4,446 6,3 5,33Ventricular arrhythmia algorithms ValuesAtrial pacing on PVC Yes-NoPost extrasystolic pause suppression Yes-NoAcceleration on PVC Yes-NoMax acceleration rate (ppm) From 60 to 145 by steps of 5 ; 10054 SORIN – PLATINIUM DR – U460A
 20.  PROGRAMMABLE PARAMETERSPost-shock mode ValuesMode OFF-VVI-DDI-DDDDuration 10s-20s-30s-1min-2min-3min-4min-5minBasic rate (ppm) From 50 to 90 by steps of 5 ; 60 (± 4 %)Rest AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms)Exercise AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms)AVD Paced/Sensed Offset (ms) 0-10-15-25-30-40-45-55-65-70-80-85-95-100-110-115-125 (± 1 ms)A amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)A pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)V amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %)V pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)Refractory periods ValuesAtrial refractory period post ventricular sensing (ms)45-65-80-95-110-125-140-155 (± 16 ms)Atrial refractory period post ventricular pacing (ms)80-95-110-125-140-155 (± 4 ms)Sensitivity margins ValuesAtrial post pacing/sensing margin (mV)From 0 to 1 by steps of 0.2 ; 0.4Ventricular post pacing margin (mV) From 0 to 2 by steps of 0.2 ; 0.8Response to noise ValuesAutomatic sensitivity on noise ON-OFFV pacing on noise ON-OFFSafeR parameters ValuesAVB I switch Rest+Exercise-ExerciseLong PR: max (ms) From 200 to 500 by steps of 50 ; 450Long PR: min (ms) From 200 to 500 by steps of 50 ; 250Max. pause (s) 2-3-4SORIN – PLATINIUM DR – U460A 55
 20.  PROGRAMMABLE PARAMETERS20.2. VENTRICULAR TACHYARRHYTHMIA DETECTIONTherapy zones ValuesSlow VT detection zone Slow VT ON-Slow VT OFFVT detection zone VT ON-VT OFFFast VT / VF detection zone Fast VT+VF ON-VF ONSlow VT rate (lower limit) (ppm) From 100 to 200 by steps of 5 ; 190VT rate (lower limit) (ppm) 130-135-140-145-150-155-160-165-170-175-180-185-190-195-200-210-220-230VF rate (lower limit) (ppm) 150-155-160-165-170-175-180-185-190-195-200-210-220-230-240Fast VT rate (upper limit) (ppm) 155-160-165-170-175-180-185-190-195-200-210-220-230-240-255Slow VT persistence (cycles) 4-6-8-12-16-20-30-50-100-200VT persistence (cycles) 4-6-8-12-16-20-30-50-100-200VF persistence (cycles) From 4 to 20 by steps of 1 ; 6Detection criteria ValuesSlow VT and VT detection criteria Rate Only-Stability-Stability+-Stability/Acc-Stability+/Acc-PARAD-PARAD+Fast VT detection criteria Rate+Stability-Rate OnlyMajority: (X/Y), Y (cycles) 8-12-16Majority: (X/Y), X (%) 65-70-75-80-90-95-100Window of RR stability for Slow VT and VT (ms)30-45-65-80-95-110-125Window of RR stability for fast VT (ms)30-45-65Acceleration (%) 6-13-19-25-31-38-44-50Long cycle persistence extension (cycles)From 0 to 16 by steps of 1 ; 10Long cycle gap (ms) 15-30-45-65-80-95-110-125-140-155-170-190-205Atrial monitoring Yes-No56 SORIN – PLATINIUM DR – U460A
 20.  PROGRAMMABLE PARAMETERS20.3. VENTRICULAR TACHYARRHYTHMIA THERAPIESCommon parameters ValuesEnable ATP therapy Yes-NoEnable shock therapy Yes-NoPolarity alternation (42J) Yes-NoAtrial coil (SVC) present Yes-NoActive case Yes-NoShock configuration (+ --> -) Case to RV-SVC to RV-Case + SVC to RV-RV to Case-RV to SVC-RV to Case + SVCSVC exclusion (shock < 15J) Yes-No20.3.1. Therapy parameters in slow VT zoneATP 1 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310SORIN – PLATINIUM DR – U460A 57
 20.  PROGRAMMABLE PARAMETERSATP 2 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310Shock program ValuesShock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Number of Max. Shock (42 J) OFF-1-2-3-420.3.2. Therapy parameters in VT zoneATP 1 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-31058 SORIN – PLATINIUM DR – U460A
 20.  PROGRAMMABLE PARAMETERSATP 2 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310Shock program ValuesShock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Number of Max. Shock (42 J) OFF-1-2-3-420.3.3. Therapy parameters in fast VT / VF zoneATP 1 program ValuesATP program OFF-Burst-Burst+Scan-Ramp-Ramp+ScanNumber of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15Coupling interval (%) 50-55-60-65-70-75-80-85-90-95Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60Time limit 10s-20s-30s-1min-1.5min-2minMinimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310SORIN – PLATINIUM DR – U460A 59
 20.  PROGRAMMABLE PARAMETERSShock program ValuesShock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9-10-12-14-16-18-20-22-24-26-28-30-32-34-42Number of Max. Shock (42 J) 1-2-3-420.4. REMOTE ALERTS AND WARNINGSThe device routinely performs security self-checks and technical measurements to ensure system integrity. When system integrity is found to be at risk outside a follow-up, alerts are stored in the device memory. When system integrity is found to be at risk during a follow-up, the information is managed by a warning (pop-up message) to immediately notify the user. For   example,   the   following  types   of   event   can   trigger   a   warning   or   an   alert:  technical problem during a shock, pacing lead impedance or coil impedance measurements out-of-range,   battery   depletion,   etc.   The   Remote   tab   presents   an   overview   of   all   the   alerts  managed by the device.General parameters ValuesRF communication (1) ON-OFFAlerts (1) ON-OFF(1) RF and Remote alerts are turned on automatically if Shocks are programmed ON.When Alerts are programmed "On", the following System Alerts are automatically activated:─Battery depletion – RRT─Device reset─Excessive charge time (>25s)─System integrityLead Alerts ValuesAbnormal A lead impedance ON-OFFAbnormal A lead low limit (Ohm) 200-250-300-350-400-450-500Abnormal A lead high limit (Ohm) 1500-1750-2000-2500-3000Abnormal V lead impedance ON-OFFAbnormal V lead low limit (Ohm) 200-250-300-350-400-450-500Abnormal V lead high limit (Ohm) 1500-1750-2000-2500-3000Abnormal RV coil impedance ON-OFFAbnormal SVC coil impedance ON-OFFAbnormal Shock impedance (1) ON-OFF(1) Normal impedance range [20 Ohm-200 Ohm]60 SORIN – PLATINIUM DR – U460A
 20.  PROGRAMMABLE PARAMETERSClinical status ValuesV oversensing ON-OFFHigh AT/AF burden ON-OFFAT/AF limit (on 24h) (h) 0.5-1-3-6-12-24Fast V Rate during AT/AF ON-OFFFast V Rate limit (ppm) 80-90-100-110-120Fast V Duration limit (h) 0.5-1-3-6-12-24Therapy information ValuesShock disabled ON-OFFShocks delivered OFF-All shocks-Inefficient shock-Inefficient max shockATP delivered ON-OFFSORIN – PLATINIUM DR – U460A 61
 21.  NON PROGRAMMABLE PARAMETERS 21. NON PROGRAMMABLE PARAMETERSInterval ValuesCommitted period 95 ms (± 5 ms)Atrial refractory periods ValuesPost atrial sensing 47 ms (± 16 ms)Post atrial pacing 109 ms (± 4 ms)Ventricular refractory periods ValuesPost ventricular sensing 95 ms (± 16 ms)Post ventricular pacing 220 ms (± 4 ms)Post atrial pacing (blanking) 16 ms (± 3 ms)Tachycardia criteria ValuesWindow of PR association 63 ms (± 1 ms)Therapies ValuesWaveform (1) Constant tilt (50% - 50%)Stored energy for the Max. shock 42 JPacing amplitude during ATP therapies7 V (Actual value at 300 ms: 5.3 V )Anti-PMT protection Termin(1) The device has 50% tilt in each phase thus delivers 94% of stored energy. Each phase is limited to  10 ms duration.62 SORIN – PLATINIUM DR – U460A
 22.  LIMITED WARRANTY 22. LIMITED WARRANTYThe   PLATINIUM   implantable   cardioverter   defibrillator   is   the   result   of   highly   advanced research and all components have been selected after exhaustive testing.The terms of the limited warranty are available upon request from your Sorin representative or on the CD-ROM of the package content.SORIN – PLATINIUM DR – U460A 63
 23.  PATENTS 23. PATENTSThe PLATINIUM model described in this manual is covered by the following US patents:5 713 928, 5 741 315, 5 776 164, 5 776 165, 5 818 703, 5 836 980, 5 868 793, 5 891 170, 5 891 184, 5 899 931, 5 931 856, 5 935 153, 5 954 660, 5 978 708, 6 181 968, 6 230 058, 6 236 111, 6 251 703, 6 256 206, 6 307 261, 6 337 996, 6 397 105, 6 408 209, 6 487 451, 6 487 452, 6 505 068, 6 532 238, 6 556 866, 6 604 002, 6 622 039, 6 625 491, 6 711 441, 6  738 665, 6 830 548, 6 889 080, 6 898 845, 6 912 421, 6 937 898, 6 975 905, 7 065 402, 7 072 716, 7 076 297, 7 113 826, 7 142 924, 7 164 946, 7 251 526, 7 366 566, 7 400 921, 7  400   922,   7   953   483,   US8064992,   US8043225,   US7792582,   US8798748,   US7890168, US8195293, US7966068, US8768464, US8874209, US8554313, US8214036, US8233981, US8554319, US7966065, US8253279, US8874210, US8219193, US8391976, US8855764, US8359096, US8494629, US8359091, US8489188, US8712526, US8862230, US8641436, US8718765, US8798771, US8868170, US8678843 , US8694098, US8938286, US8874212.64 SORIN – PLATINIUM DR – U460A
 24.  EXPLANATION OF SYMBOLS 24. EXPLANATION OF SYMBOLSThe symbols on product labelling have the following meaning:General symbols Explanation of symbols Defibrillator symbols Explanation of symbolsUse by   ICD (dual chamber, RA, RV)Date of manufacture   DR DF-1 connectorsManufacturer   DR DF4 connectorsCatalogue number ShocksSerial number   ATP Anti-tachycardia pacing, RVImplantable device uncoated SafeR As shipped modePackage content Basic rateSterile package content High voltageOpen here Ready for wireless interrogation and programming by Orchestra Plus programmer equipped with ORCHESTRA PLUS LINK accessoryDo not use if the package is damagedDF-1 defibrillating connector insulating plugDo not reuseDo not resterilizeSterilised using ethylene oxideNon sterileSORIN – PLATINIUM DR – U460A 65
 24.  EXPLANATION OF SYMBOLSTemperature limitationTorque wrenchConsult instructions for useInstructions for use in the CD-ROMThis icon is used to call your attention to a particularly important point.This icon alerts you to a hazard that may result in equipment damage or personal injury. Carefully read the instructions provided with this icon.Eucomed / Advamed Code of Ethical Business PracticeLast revision date of this implant manual: 2015-0366 SORIN – PLATINIUM DR – U460A
BACKCOVER                                                                CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN2015-03 U460A© 2015 Sorin CRM SAS2015-03U460A

Navigation menu