SORIN CRM DR9550 Implantable cardioverter defibrillator User Manual

SORIN CRM Implantable cardioverter defibrillator Users Manual

Users Manual

TABLE OF CONTENTS 1. General description .................................................................... 6 2. Indications ................................................................................... 6 3. Contraindications ....................................................................... 6 4. Warnings and precautions ........................................................ 7 4.1. Risks related to medical environment ........................... 8 4.2. Sterilization, storage and handling .............................. 10 4.3. Implantation and device programming ........................ 11 4.4. Lead evaluation and lead connection .......................... 12 4.5. Generator explant and disposal ................................... 13 5. Adverse events ......................................................................... 14 5.1. Defender study ............................................................... 14 5.2. SafeR study .................................................................... 17 6. Clinical studies ......................................................................... 19 6.1. Defender study ............................................................... 19 6.2. SafeR study .................................................................... 23 7. Patient selection and treatment .............................................. 28 7.1. Individualization of treatment ....................................... 28 7.2. Specific patient populations ......................................... 29 8. Patient counselling information .............................................. 30 9. Conformance to standards ...................................................... 30 10. Physician guidelines ................................................................ 34 10.1. Physician training .......................................................... 34 10.2. Directions for use........................................................... 34 10.3. Maintaining device quality ............................................ 34 11. Patient information ................................................................... 35 12. How supplied ............................................................................ 35 12.1. Sterility ............................................................................ 35 12.2. Warranty and replacement policy ................................ 35
13. Device description.................................................................... 35 14. Implant procedure .................................................................... 38 14.1. Necessary equipment .................................................... 38 14.2. Packaging ....................................................................... 39 14.3. Optional equipment ....................................................... 39 14.4. Before opening the package ......................................... 39 14.5. Prior to implantation ...................................................... 40 14.6. Device placement ........................................................... 40 14.7. Choosing the type of lead ............................................. 41 14.8. Measurement of thresholds at implant ........................ 42 14.9. Lead connection............................................................. 43 14.10. Device implantation ....................................................... 45 14.11. Tests and programming ................................................ 45 15. Special modes .......................................................................... 46 15.1. Safety mode (nominal values) ...................................... 46 15.2. Magnet mode .................................................................. 46 15.3. Response in the presence of interference .................. 47 15.4. Detection characteristics in the presence of electromagnetic fields ................................................... 48 15.5. Protection against short-circuits .................................. 49 16. Main functions .......................................................................... 49 16.1. Automatic lead measurements ..................................... 49 16.2. Atrial tachyarrhythmia management ............................ 49 16.3. Ventricular tachyarrhythmia management .................. 50 16.4. Pacing ............................................................................. 51 16.5. Sensing ........................................................................... 51 16.6. Follow-up functions ....................................................... 52 16.7. Remote Monitoring function ......................................... 53 17. Patient follow-up ....................................................................... 56 17.1. Follow-up recommendations ........................................ 56 17.2. Holter Function ............................................................... 57 17.3. Elective Replacement Indicator (ERI) .......................... 58 17.4. Explantation .................................................................... 59 17.5. Defibrillator identification ............................................. 60
18. Physical characteristics .......................................................... 61 18.1. Materials used ................................................................ 61 19. Electrical characteristics ......................................................... 62 19.1. Table of delivered shock energy and voltage ............. 63 19.2. Battery ............................................................................. 64 19.3. Longevity ........................................................................ 64 20. Programmable parameters ...................................................... 66 20.1. Antibradycardia pacing ................................................. 66 20.2. Ventricular tachyarrhythmia detection ........................ 71 20.3. Ventricular tachyarrhythmia therapies ........................ 73 20.4. Remote alerts and warnings ......................................... 77 21. Non programmable parameters .............................................. 79 22. Limited warranty ....................................................................... 80 22.1. Article 1 : Terms of limited warranty ............................ 80 22.2. Article 2 : Terms of replacement .................................. 82 23. Patents ....................................................................................... 83 24. Explanation of symbols ........................................................... 84
6 – US-ENGLISH 1. GENERAL DESCRIPTION PARADYM RF DR 9550 is an implantable dual-chamber cardioverter defibrillator. It is equipped with an accelerometer to allow adaptation of pacing to suit the patient’s activity. 2. INDICATIONS This  device  is  indicated for  use  in  patients  who  are  at  high  risk  of sudden  cardiac death  due  to  ventricular  tachyarrhythmias  and  who have experienced one of the following situations:   Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia,  Recurrent, poorly tolerated sustained ventricular tachycardia (VT). NOTE:  The  clinical  outcome  for  hemodynamically  stable  VT patients  is  not fully known.  Safety and effectiveness studies  have not been conducted. 3. CONTRAINDICATIONS Implantation  of  PARADYM  RF  DR  9550  is  contraindicated  in patients:  whose  ventricular  tachyarrhythmias  may  have  transient  or reversible  causes  such  as:  acute  myocardial  infarction,  digitalis intoxication,  drowning,  electrocution,  electrolyte  imbalance, hypoxia, sepsis, or unstable ischemic episodes,  who present incessant tachyarrhythmia,  who have an internal pacemaker,  whose  primary  disorder  is  bradyarrhythmias,  or  atrial tachyarrhythmias. Dual-chamber and single chamber atrial pacing is contraindicated in patients with chronic refractory atrial tachyarrhythmias.
US-ENGLISH – 7 4. WARNINGS AND PRECAUTIONS The  patient  should  be  warned  of  the  potential  risks  of  defibrillator malfunction  if  he  is  exposed  to  external  magnetic,  electrical,  or electromagnetic signals. These  potential  interference  sources  may  cause  conversion  to inhibited mode (because of noise detection), erratic delivery of VT or VF  therapies,  nominal  programming,  or  much  more  rarely, irreversible damage to the device’s circuits. The  main  sources  of  high  magnitude  electromagnetic  interference are:  powerful  radiofrequency  equipment  (radar),  industrial  motors and transformers, arc-welding equipment, high power loudspeakers. Resuscitation Availability: Do not perform device testing unless an external  defibrillator  and  medical  personnel  skilled  in cardiopulmonary resuscitation (CPR) are readily available. Electrical Isolation:  Do not  permit the  patient to  contact  grounded equipment  that  could  produce  hazardous  leakage  current.  Ensuing arrhythmia induction could result in the patient’s death. Disable the ICD During Handling: Program Shock Therapy to OFF during surgical implant and explant or post mortem procedures. The device  can  deliver  a  serious  high  energy  shock  should  accidental contact be made with the defibrillation electrodes. Antitheft gates: Since antitheft devices at the entrance to stores are not subject to any safety standards, it is advisable to spend as little time as possible in their vicinity. Airport detection systems: Since airport detection systems are not subject to any safety standards, it is advisable to spend as little time as possible in their vicinity. High  voltage  power  transmission  lines:  High  voltage  power transmission  lines  may  generate  enough  EMI  to  interfere  with defibrillator operation if approached too closely.
8 – US-ENGLISH Communication  equipment:  Communication  equipment  such  as microwave  transmitters,  linear  power  amplifiers,  or  high-power amateur  transmitters  may  generate  enough  EMI  to  interfere  with defibrillator operation if approached too closely. Home appliances: Home appliances that are in good working order and  properly  grounded  do  not  usually  produce  enough  EMI  to interfere  with  defibrillator  operation.  There  are  reports  of  device disturbances  caused  by  electric  hand  tools  or  electric  razors  used directly over the device implant site. CAUTION: Do not tap sharply on the ICD can after implant, because the  ICD's  sensing  circuits  can  detect  this  as  P-waves  or  R-waves, and such oversensing could result in inappropriate pacing, inhibition, or  therapy.  Normal  activities  after  implant  do  not  result  in  such oversensing. 4.1. RISKS RELATED TO MEDICAL ENVIRONMENT It is advisable to carefully monitor defibrillator operation prior to and after any medical treatment during which an electrical current from an external source passes through the patient's body. Magnetic  Resonance  Imaging:  MRI  is  strictly  contraindicated  in cardiac defibrillator patients. A radio frequency ablation: A radio frequency ablation procedure in a patient with a generator may cause device malfunction or damage. RF  ablation  risks  may  be  minimized  by:  1. Programming  Shock Therapy  and  ATP  to  OFF.  2. Avoiding  direct  contact  between  the ablation catheter and the implanted lead or generator. 3. Positioning the  ground,  placing  it  so  that  the  current  pathway  does  not  pass through  or  near  the  device,  i.e.  place  the  ground  plate  under  the patient’s buttocks or legs. 4. Having external defibrillation equipment available.
US-ENGLISH – 9 Electrocautery or diathermy device: Diathermy and electrocautery equipment  should  not  be  used.  If  such  devices  must  be  used: 1. Keep  the  current  path  and  ground  plate  as  far  away  from  the device and the leads as possible (a minimum of 15 cm [six inches]). 2. Before procedure, deactivate ATP and shock therapies. 3. During the procedure, keep the electrocautery device as far as possible from the  cardiac  defibrillator.  Set  it  at  minimum  intensity.  Use  it  briefly. 4. After the procedure, check for proper implant function. The device should never be exposed directly to the diathermy source. External  defibrillation:  PARADYM  RF  DR  9550  is  protected  from external defibrillation shocks. Before external defibrillation, deactivate ATP and shock therapies. During external defibrillation, it is advisable to avoid placing the defibrillating paddles directly over the casing or over the leads. The defibrillating paddles should preferably be placed in an anteroposterior position. Avoid any direct contact between the defibrillation paddles and the conductive parts of the implanted leads or casing of the implanted device. After external defibrillation, check for proper device function. Radiation therapy:  Avoid  exposure  to  ionizing  radiation. Betatrons are  contraindicated.  If  high  doses  of  radiation  therapy  cannot  be avoided,  the  defibrillator  should  be  protected  from  direct  exposure with a screen. ATP  and  shock  therapies should  be  disabled  during exposure  and  proper  device  function  should  be  checked  regularly afterwards. Resulting damage may not be immediately detectable. If irradiation of tissues close to the implantation site is necessary, it is recommended  that  the  cardiac  defibrillator  be  moved.  As  a  safety measure, an external defibrillator should be immediately available. Lithotripsy: Lithotripsy may  permanently  damage the  device if  this one is at the focal point of the lithotripsy beam. If lithotripsy must be used,  keep  the  defibrillator  at  least  2.5  to  5  cm  (1-2  inches)  away from the focal point of the lithotripsy beam.
10 – US-ENGLISH Diagnostic  ultrasound  (echography):  The  defibrillator  is  not affected by ultrasound imaging devices. Scales  with  body  fat  monitors  and  electronic  muscle stimulators:  A  patient  with  an  implanted  PARADYM RF DR 9550 should not use these devices. 4.2. STERILIZATION, STORAGE AND HANDLING Resterilization: Do not resterilize and re-implant explanted ICDs. "Use  Before"  Date:  A  "Use  Before"  date  is  printed  on  the  outer storage package and on the sterile package. Do not implant the device after  this  date  because  the battery  may  have  reduced  longevity  and sterility may be affected. It should be returned to Sorin CRM. If Package Is Damaged: Do not use the device or accessories if the packaging  is  wet,  punctured,  opened  or  damaged  because  the integrity  of  the  sterile  packaging  may  be  compromised.  Return  the device to the manufacturer. Device  Storage:  Store  the  device  in  a  clean  area,  away  from magnets,  kits  containing  magnets,  and  sources  of  electromagnetic interference to  avoid device damage. Store  the  device  between  0  - 50 °C (32 - 122 °F). Temperatures outside the specified range may damage the device. Equilibration:  Allow  the  device  to  reach  room  temperature  before programming  or  implanting  the  device  because  rapid  temperature changes may affect initial device function.
US-ENGLISH – 11 4.3. IMPLANTATION AND DEVICE PROGRAMMING Use only a Sorin CRM programmer to communicate with the device. Do  not  position  any  magnet  over  the  ICD;  this  suspends tachyarrhythmia detection and treatment. Replace the device when the programmer displays an ERI* (defined by a battery voltage of 2.66 ± 0.01 V or a magnet rate lower than or equal to 80 bpm). Program device parameters such as sensitivity threshold and VT and VF detection intervals as specified in the device manuals. Lead  System:  Do  not  use  a  lead  system  other  than  those  with demonstrated  compatibility  because  undersensing  cardiac  activity and failure to deliver necessary therapy may result. In  situations  where  an  ICD  and  a  pacemaker  are  implanted  in  the same  patient,  interaction  testing  should  be  completed.  If  the interaction between the ICD and the pacemaker cannot be resolved through  repositioning  of  the  leads  or  reprogramming  of  either  the pacemaker or the ICD, the  pacemaker should  not  be  implanted (or should be explanted if previously implanted). Failure to properly insert the torque wrench into the perforation at an angle  perpendicular  to  the  connector  receptacle  may  result  in damage to the sealing system and its self-sealing properties. A safety margin of at least 10 J in the defibrillation threshold (DFT) is recommended. Carefully  confirm  that  true  ventricular  fibrillation  has been induced because  the DFT for  ventricular tachycardia or flutter may be lower. The  defibrillator  should  be  implanted  with  the  engraved  side facing  outwards  in  order  to  facilitate  telemetric  communication with  the  programming  head  and  to  display  the  radiographic identification correctly. *: corresponds to Recommended Replacement Time (RRT) / End of Service (EOS) as referred in the EN45502-2-2 standard.
12 – US-ENGLISH 4.4. LEAD EVALUATION AND LEAD CONNECTION PARADYM RF DR 9550 has two DF-1 and two IS-1 connector ports. IS-1  refers  to  the  international  standard  whereby  leads  and generators  from  different  manufacturers  are  assured  a  basic  fit (ISO 5841-1:2000).  DF-1  refers  to  the  international  standard  for defibrillation lead connectors (ISO 11318:2002). Do  not  tie  a  ligature  directly  to  the  lead  body,  tie  it  too  tightly,  or otherwise  create  excessive  strain  at  the  insertion  site  as  this  may damage the lead. Use the lead stabilizer to secure the lead lateral to the venous entry site. Do  not  immerse  the  leads  in  mineral  oil,  silicone  oil,  or  any  other liquid. Do not grip the lead with surgical instruments. Do not use excessive force or surgical instruments to insert a stylet into a lead. Use ventricular transvenous leads with caution in patients with either a mechanical or bioprosthetic tricuspid valvular prosthesis. Use  the  correct  suture  sleeve  (when  needed)  for  each  lead,  to immobilize the lead and protect it against damage from ligatures. Never implant the system with a  lead  system that has  a measured shock  impedance  of  less  than  30  ohms.  A  protection  circuit  in  the defibrillator prevents shock delivery when impedance is too low. If the shock impedance is less than 30 ohms, reposition the lead system to allow a greater distance between the electrodes. Do not kink leads. Kinking leads may cause additional stress on the leads, possibly resulting in lead fracture. Do not  insert  a lead connector  pin into the connector block without first visually verifying that the setscrews are sufficiently retracted. Do not  tighten  the  setscrews  unless  a  lead  connector  pin  is  inserted because it could damage the connector block.
US-ENGLISH – 13 Lead  electrodes  in  contact  during  a  cardioversion  or  defibrillation therapy will cause current to bypass the heart, possibly damaging the  ICD  and  the  leads.  While  the  ICD  is  connected  to the  leads, make  sure  that  the  metal  portions  of any electrodes  do not  touch each other. If a pacing lead is abandoned rather than removed, it must be capped to ensure that it is not a pathway for currents to or from the heart. If  a  thoracotomy  is  required  to  place  epicardial  patches,  it  should be  done  during  a  separate  procedure  to  reduce  the  risk  of morbidity and mortality. Do  not  place  the  patch  lead  over  nerve  tissue  as  this  may  cause nerve damage. Place the patch lead with the conducting coil side facing the heart to ensure delivery of energy to the heart. Place the sutures well outside the coil of the patch lead or in the area between the coils to avoid possible coil fracture. If  countershock  is  unsuccessful  using  external  paddles,  adjust  the external  paddle  position  (e.g.,  anterior-lateral  to  anterior-posterior) and be sure that the external paddle is not positioned over the patch. Do  not  fold,  alter,  or  remove  any  portion  of  the  patch  as  it  may compromise electrode function or longevity. If a header port is unused on the generator, the port must be plugged to protect the generator. 4.5. GENERATOR EXPLANT AND DISPOSAL Interrogate the device, and program shock therapy off prior to explanting, cleaning or shipping the device to prevent unwanted shocks. Return all explanted generators and leads to the manufacturer. Never  incinerate  the  device  due  to  the  potential  for  explosion.  The device must be explanted before cremation.
14 – US-ENGLISH 5. ADVERSE EVENTS Clinical  data  presented  in  this  section  are  from  the  Defender  and SafeR clinical  studies.  PARADYM RF  DR  9550  is  similar in  design and  clinical  function  to  the  Defender  devices.  SafeR  operation  in PARADYM  RF  is  similar  to  that  in  the  Symphony  pacemaker.  The data provided are applicable to PARADYM RF DR 9550. 5.1. DEFENDER STUDY Clinical study of Defender IV DR 612 included 60 devices implanted in 60 patients, 38 in Europe (37 patients followed for a minimum of 3 months), and 22 in the U.S. (IDE G970282/S15) with a total device exposure of 228.7 and 30.3 device months, respectively. No deaths, serious  adverse  experiences  or  complications  were  judged  to  be device-related,  as  determined  by  the  investigator.  The  following tables summarize the safety data for this study. There was 1 death in the study that was classified as arrhythmic. The cause of death was recurrent VT/VF which occurred 19 days post implant. In the following tables, complications are defined as adverse device effect,  which  cannot  be  treated  or  resolved  by  simple  adjustments (e.g. reprogramming) and requires intervention. NOTE:  The  company  classified  as  complications  those  adverse device  effects  that  were  treated  with  surgery  or  with  external defibrillation of a ventricular cardiac event. Observations  are  defined  as  symptomatic  or  asymptomatic  clinical events  with  potential  adverse  device  effects  that  do  not  require intervention or can be corrected by simple adjustments. NOTE: The company classified as observations those adverse device effects  that  were  treated  with  programming  changes,  medication,  or other method that was not classified as a complication. Two of  the 38  Defender  IV  DR 612 patients in  Europe (37 patients followed  for  a  minimum  of  3  months)  experienced  a  total  of  three
US-ENGLISH – 15 complications, including device failures and replacements. Fourteen of  the  38  Defender  IV  DR  612  patients  experienced  a  total  of  18 observations. Complications and observations are reported in Tables 1 and 2. It  should be noted  that a  patient  can have more  than one observation  or  complication.  There  were  no  observations  or complications in the U.S.  Table 1: Summary of European Clinical Complications (Including Device Failures and Replacements) All complications, 2 of 38 Defender IV DR 612 patients in Europe Event # of Patients % of Patients # of Events Events/100 Device-Years* Hematoma 1 2.6 1 5.2 Ventricular lead migration/dislodgment 2 5.3 2 10.5 * There were 228.7 device months in this study.  Table 2: Summary of European Clinical Complications (Including Patient Complaints) All complications, 14 of 38 Defender IV DR 612 patients in Europe Event # of Patients* % of Patients # of Events Events/100 Device-Years** Change in ventricular sensing threshold 1 2.6 1° 5.2 Device reset*** 1 2.6 1° 5.2 Inappropriate therapy for EMI 1 2.6 1° 5.2 Pneumothorax 1 2.6 1° 5.2
16 – US-ENGLISH Event # of Patients* % of Patients # of Events Events/100 Device-Years** Pocket hematoma 2 5.3 2° 10.5 Pocket infection/hematoma 1 2.6 1° 5.2 Pocket infection from previous pacemaker 1 2.6 1° 5.2 Prolonged implant procedure 1 2.6 1 5.2 Sensor acceleration during telemetry*** 1 2.6 1 5.2 Shock for VT in VF Zone 1 2.6 1° 5.2 Slow VT not converted by ATP therapy 1 2.6 2° 10.5 Unsatisfactory sensing threshold test*** 2 5.3 2 10.5 Ventricular oversensing 3 7.9 3 15.7 * A patient can have more than one observation. ** There were 228.7 device months in this study. ***These  observations  would  not  have  happened  with  the  currently marketed device and programmer. °Investigator  indicated  that  Defender  IV  DR  did  not  cause  or contribute to the event.
US-ENGLISH – 17 5.2. SAFER STUDY Clinical  study  of  the  SafeR  included  45  Symphony  2550  devices implanted in 45 patients. No serious adverse events were device- or feature-related. There were no deaths in the study. Table 1 summarizes the safety data for this study.  Table 1: Summary of Symphony safety data during study  Patients  Number of events   Number of patients % of patients Number of events Events per device year (a) Deaths 0 0 0 0 Explants 0 0 0 0 Serious pacemaker related events outside the use of SafeR 0 0 0 0 Non-serious pacemaker related events outside the use of SafeR 0 0 0 0 Serious events due to the use of SafeR 0 0 0 0
18 – US-ENGLISH  Patients  Number of events   Number of patients % of patients Number of events Events per device year (a) Non-serious events related due to the use SafeR 13 28.9 15 3.2 Serious non-pacemaker related events 6 13.3 9 1.9 Non-serious non-pacemaker related events 8 17.8 8 1.7 (a) 4.74 device years Non-serious  events  due  to  the  use  of  SafeR  included:  delay  in switching  on  2nd  degree  AV  block,  inappropriate  classification  of  a PAC,  disagreement  between  markers  and  recorded  EGM,  atrial pacing  above  the  maximum  rate,  recycling  on  an  r-wave  in  a refractory  period,  and  disagreement  in  the  statistics  for  switches  to DDD. No patient symptoms were associated with these events.
US-ENGLISH – 19 6. CLINICAL STUDIES Clinical  data  presented  in  this  section  are  from  the  Defender  and SafeR clinical  studies.  PARADYM RF  DR  9550  is  similar in  design and  function  to  the  Defender  devices.  SafeR  operation  in PARADYM RF  is  similar  to  that  in  the  Symphony  pacemaker.  The data provided are applicable to PARADYM RF DR 9550. 6.1. DEFENDER STUDY Objectives:  The  primary  objectives  of  this  study  were  to demonstrate  a  complication  free  rate  (CFR)  comparable  to  that  of historical controls, to demonstrate, using a chronotropic assessment exercise  protocol  (CAEP),  a  rate  response  proportional  to  and appropriate for the level of exercise, and to evaluate and report the incidence of adverse events. Materials:  Each  patient  received  one  Defender  IV  DR  612 defibrillator,  an  atrial  pacing  and  sensing  lead,  and  a  Medtronic, Angeion,  or  Biotronik  defibrillation  lead  in  the  U.S.  or  any commercially available defibrillator lead outside the U.S. Methods: Investigators selected patients who survived at  least one episode  of  cardiac  arrest  (manifested  by  loss  of  consciousness) presumably  due  to  a  ventricular  tachyarrhythmia  or  exhibited recurrent,  poorly  tolerated,  sustained  ventricular  tachycardia  (VT). The protocol required evaluation of performance and adverse events at pre-discharge, one  month,  three months,  six  months, and (in the U.S.) every  three months  thereafter.  At  the one-month  visit,  eligible patients  performed  a  chronotropic  assessment  exercise  protocol (CAEP) maximal exercise test.
20 – US-ENGLISH Study Population. The table below summarizes inclusions. Region Date of first implant Date of last implant Data cut-off date Number of centers Number of patients US 14-Dec-99 08-Mar-00 14-Mar-00 6 22 Europe 04-May-99 26-Jul-99 14-Apr-00 11 38 All 04-May-99 08-Mar-00 14-Apr-00 (Eur), 14-Mar-00 (US) 17 60  Complication-free rate Only European patients followed for at least 3 months: Symbol Parameter Defender IV DR 612 N Overall number of patients 37 Pe*N Number of successes 35 Pe Observed experimental proportion 0.95 Ps Null hypothesis success rate 0.76 ES Estimated standard error of Pe 0.04 z´ Test statistic (1) 4.75 p Associated p-value < 0,0001 (1) Statistical test: z´ = (Pe-Ps)/SE where SE = sqrt(Pe(1-Pe)/N)
US-ENGLISH – 21  Rate response European patients only: GROUP Number of patients included Mean slope %SRR on %MR STD of slopes %SRR on %MR SE of mean slope %SRR on %MR Lower 95% CI Upper 95% CI Europe 20 0.77 0.17 0.04 0.69 0.84 Small Centers 9 0.79 0.18 0.06 0.67 0.91 Large Centers 11 0.75 0.15 0.05 0.66 0.84 Males 17 0.77 0.16 0.04 0.70 0.85 Females 3 0.73 0.22 0.13 0.47 0.98 SRR: Sensor Rate Reserve MR: Metabolic Reserve STD: Standard Deviation SE: Standard Error CI: Confidence Interval
22 – US-ENGLISH  Adverse events Event US (N=22) Number of events* Number of patients Percent of patients Intent to treat but did not 0 0 0.0 Non-device related death 0 0 0.0 Explant 0 0 0.0 Complication 0 0 0.0 Observation 0 0 0.0 Serious non-related other than death 1 1 4.5  Event Europe (N=38) Number of events* Number of patients Percent of patients Intent to treat but did not 0 0 0.0 Non-device related death 1 1 2.6 Explant 1 1 2.6 Complication 3 2 5.3 Observation 18 14 36.8 Serious non-related other than death 12 7 18.4
US-ENGLISH – 23 Event All (N=60) Number of events* Number of patients Percent of patients Intent to treat but did not 0 0 0.0 Non-device related death 1 1 1.7 Explant 1 1 1.7 Complication 3 2 3.3 Observation 18 14 23.3 Serious non-related other than death 13 8 13.3 *  A  patient  can  have  more  than  one  complication,  observation,  or serious adverse event, not device-related. Device  Failures  and  Replacements:  No  device  failures  or replacements occurred with Defender IV DR 612 during the study. 6.2. SAFER STUDY SafeR mode in PARADYM RF is similar to that in Symphony. The differences in SafeR mode between the two devices are:  To prevent long RR intervals during VT/VF, SafeR has no effect during VT/VF therapy, electrophysiologic studies, and post-shock recovery.  The  maximum  acceptable  AV  delay  for  first  degree  AV  block varies as a function of pacing rate.  PARADYM RF requires a ventricular sensed event to atrial paced event  (RA)  interval  of  at  least  100  ms.  Therefore,  the  device lengthens the atrial escape interval so that it ends at least 102 ms after the ventricular event.
24 – US-ENGLISH  During atrial fibrillation  episode,  pause criterion  is  fixed  to  2s  to avoid long bradycardia episodes in switching to DDD mode. Despite these differences, the  data  collected on Symphony devices are  applicable  to  PARADYM  RF  because  the  principles  of  SafeR operation did not change. The criteria for switching from AAI to DDD (or vice versa)  did  not  change. The  device’s method for evaluating the presence of AV conduction did not change. Methods: All patients were implanted with a Symphony Model 2550 dual-chamber  rate-responsive  pacemaker  with  SafeR  mode.  A variety  of  marketed  atrial  and  ventricular  pacing  leads  were  used. The pacemaker was programmed and interrogated via bi-directional telemetry  using  a  Sorin CRM  dedicated  programmer  and  a  CPR3 programming head. The study’s routine evaluation consisted of enrollment, pre-discharge evaluation,  and  a  scheduled  follow-up  visit  at  one  month.  At  pre-discharge, a 24-hour Holter recording was performed and pacemaker memory  was  read.  At  one  month,  pacemaker  memory  was  read. Investigators also documented adverse events. Patients  studied:  A  total  of  45  patients  from  12  centers  had Symphony 2550 pacemakers with SafeR. Of these, 14 (31 %) were female and 31 (69 %) were male. Mean patient age (± SD) was 74 ± 9 years. Primary indications for implant were: 1st degree AV block (11.1 %), 2nd degree AV block (6.7 %), 3rd degree AV block (22.2 %), sinus node dysfunction (62.2 %) or other (6.7 %). Effectiveness  results:  To  determine  the  effectiveness  of  SafeR mode, the percentage of ventricular pacing provided over one month was recorded from pacemaker memory.
US-ENGLISH – 25 Thirty-five  patients  contributed  data  to  evaluate  the  percentage  of ventricular pacing  provided  with  SafeR. Twenty-nine  patients  had  1 % or less ventricular pacing and six patients had a range of 28-97 % ventricular  pacing.  The  graph  below  shows  the  distribution  of ventricular pacing observed in patients with and without AV block as a primary indication for implant.  The graph shows that many patients programmed to SafeR had less than 1% ventricular pacing:  84 % of patients without AV block at implant.  63 % of patients with AV block at implant.
26 – US-ENGLISH In  a  representative  reference  group(1)  of  patients  programmed  to DDD, none had less than 1 % ventricular pacing and only 10 % had less than 90 % ventricular pacing regardless of AV block indication at implant. The actual reduction of ventricular pacing that SafeR provides in an individual will depend on the amount of time that the patient spends in AV  block.  SafeR cannot and should not  provide any  decrease in ventricular pacing while the patient is in AV block.  Adverse events Clinical  study  of  the  SafeR  included  45  Symphony  2550  devices implanted in 45 patients. No serious adverse events were device- or feature-related.  There  were  no  deaths  in  the  study.  Table  1 summarizes the safety data for this study. 1.                                                       (1) Pioger G, Jauvert G, Nitzsché R, Pozzan J, Laure H, Zigelman M, Leny G, Vandrell M, Ritter P, and Cazeau S. Incidence and predictive factors of atrial fibrillation in paced patients. PACE, 28, Supp 1: S137-141; January 2005. This was a prospective observational study of 377 patients with a functionally similar device programmed to DDD. The primary indications for implant were: AV block (49 %), sinus node disease (16 %), brady-tachy syndrome (5 %), AV block + sinus node disease (19 %), AV block + brady-tachy syndrome (6 %), and brady-tachy syndrome + sinus node disease (5 %).
US-ENGLISH – 27 Table 1: Summary of Symphony safety data during study  Patients Number of events  Number of patients % of patients Number of events Events per device year (a) Deaths 0 0 0 0 Explants 0 0 0 0 Serious pacemaker related events outside the use of SafeR 0 0 0 0 Non-serious pacemaker related events outside the use of SafeR 0 0 0 0 Serious events due to the use of SafeR 0 0 0 0 Non-serious events related due to the use SafeR 13 28.9 15 3.2 Serious non-pacemaker related events 6 13.3 9 1.9 Non-serious non-pacemaker related events 8 17.8 8 1.7 (a) 4.74 device years Non-serious  events  due  to  the  use  of  SafeR  2  included:  delay  in switching  on  2nd  degree  AV  block,  inappropriate  classification  of  a PAC,  disagreement  between  markers  and  recorded  EGM,  atrial pacing  above  the  maximum  rate,  recycling  on  an  r-wave  in  a refractory  period,  and  disagreement  in  the  statistics  for  switches  to DDD. No patient symptoms were associated with these events.
28 – US-ENGLISH 7. PATIENT SELECTION AND TREATMENT 7.1. INDIVIDUALIZATION OF TREATMENT Exercise  stress  testing: If  the  patient’s  condition  permits,  use exercise stress testing to:  Determine the maximum rate of the patient’s normal rhythm,  Identify any supraventricular tachyarrhythmias,  Identify exercise-induced tachyarrhythmias. The  maximum  exercise  rate  or  the  presence  of  supraventricular tachyarrhythmias  may  influence  selection  of  programmable parameters.  Holter  monitoring  or  other  extended  ECG  monitoring also may be helpful. CAUTION: To avoid inappropriate therapy during an exercise stress test,  do  not  reprogram  any  parameter  during  the  test.  When  a parameter  is  reprogrammed,  PARAD/PARAD+  algorithm  forces acceleration  to  "ventricular".  During  conducted  sinus  tachycardia within  the  programmed  Tachy  zone,  the  device  detects  a  1:1  fast rhythm.  Assuming  that  acceleration  was  set  to  ventricular  by reprogramming,  the  device  may  identify  this  as  a  VT,  and  may immediately apply the corresponding therapy.  Electrophysiologic (EP) testing: EP testing may be useful for ICD candidates. EP testing may identify the classifications and rates of all the ventricular and atrial arrhythmias, whether spontaneous or during EP testing.
US-ENGLISH – 29 Drug  resistant  supraventricular  tachyarrhythmias  (SVTs):  Drug resistant supraventricular tachyarrhythmias (SVTs) may initiate frequent unwanted device therapy. A careful choice of programming options is necessary for such patients. Antiarrhythmic  drug  therapy:  If  the  patient  is  being  treated  with antiarrhythmic  or  cardiac  drugs,  the  patient  should  be  on  a maintenance drug dose rather than a loading dose at the time of ICD implantation.  If  changes  to  drug  therapy  are  made,  repeated arrhythmia inductions are recommended to verify ICD detection and conversion. The ICD also may need to be reprogrammed. Changes in  a patient’s antiarrhythmic  drug or  any  other  medication that affects the patient’s normal cardiac rate or conduction can affect the rate of tachyarrhythmias and/or efficacy of therapy. Direct any questions regarding the individualization of patient therapy to Sorin CRM’s representative. 7.2. SPECIFIC PATIENT POPULATIONS Pregnancy: If  there is  a  need to  image the  device, care should be taken to minimize radiation exposure to the fœtus and the mother. Nursing Mothers: Although appropriate biocompatibility testing has been  conducted  for  this  implant  device,  there  has  been  no quantitative assessment of the presence of leachables in breast milk. Pediatric  Patients:  This  device  has  not  been  studied  in  patients younger than 18 years of age. Geriatric  Patients:  Most  of  the  patients  receiving  this  device  in clinical studies were over the age of 60 years. Handicapped  and  Disabled  Patients:  Special  care  is  needed  in using this device for patients using an electrical wheel chair or other electrical (external or implanted) devices.
30 – US-ENGLISH 8. PATIENT COUNSELLING INFORMATION The physician should consider the following points in counselling the patient about this device:  Persons  administering  CPR  may  experience  tingling  on  the patient’s body  surface when  the patient’s ICD  system delivers a shock.  Advise patients to carry Sorin CRM ID cards and/or ID bracelets documenting their ICD system. 9. CONFORMANCE TO STANDARDS This  device  was  developed  in  conformance  with  all  or  parts  of  the following standards:  EN  45502-1:  1998  –  Active  implantable  medical  devices.  General  requirements  for  safety,  marking  and  information  to  be provided by the manufacturer.  EN  45502-2-1:  2003  -  Active  implantable  medical  devices.  Part  2-1:  Particular  requirements  for  active  implantable  medical devices intended to treat bradyarrhythmia (cardiac pacemakers).   EN  45502-2-2:  2008  –  Active  implantable  medical  devices.  Part  2-2:  Particular  requirements  for  active  implantable  medical devices  intended  to  treat  tachyarrhythmia  (includes  implantable defibrillators).  ISO  5841-3:  2000  Low  profile  connectors  (IS1)  for  implantable pacemakers.  ISO  11318  (DF-1):  Cardiac  defibrillator:  connector  assembly  for implantable  defibrillators  -  Dimensional  and  test  requirements, August 2002.  ANSI/AAMI  PC69:2007  Active  implantable  Medical  Devices  - Electromagnetic compatibility - EMC test protocols for implantable cardiac pacemakers and implantable Cardioverter Defibrillators.
US-ENGLISH – 31  IEC  60601-1-2  (2007):  Electromagnetic  compatibility  -  Medical electrical  equipment.  General  requirements  for  basic  safety  and essential performance - Collateral standard  EN 50371 (2002) : Generic standard to demonstrate the compliance of  low  power  electronic  and  electrical  apparatus  with  the  basic restrictions related to human exposure to electromagnetic fields (10 MHz - 300 GHz)  EN  301  489-1  (v1.8.1)  &  EN  301  489-27  (v1.1.1):  Electromagnetic compatibility  and  Radio  spectrum  Matters  (ERM);  Electromagnetic Compatibility  (EMC)  standard  for  radio  equipment  and  services    - Part 1 : Technical Requirements and Part 27: Specific conditions for Ultra  Low  Power  Active  Medical  Implants  (ULP-AMI)  and  related peripheral devices (ULP-AMI-P)  EN  301839-1  (v1.3.1)  &  EN  301839-2  (v1.2.1):  Electromagnetic compatibility  and  Radio  spectrum  Matters  (ERM);  Short  Range Devices (SRD); Ultra Low Power Active Medical Implants (ULP-AMI) and Peripherals  (ULP-AMI-P) operating  in  the  frequency  range 402 MHz to 405 MHz; Part 1: Technical characteristics and test methods and  Part  2:  Harmonized  EN  covering  essential  requirements  of Article 3.2 of the R&TTE Directive  EN 62311 (2008) : Assessment of electronic and electrical equipment related  to  human  exposure  restrictions  for  electromagnetic  fields (0Hz to 300 GHz)  EN 62209-2 (2010) : Human exposure to radio frequency fields from hand-held  and  body-mounted  wireless  communication  devices  – Human models, instrumentation and procedures – Part 2: Procedure to  determine  the  specific  absorption  rate  (SAR)  for  wireless communication  devices  used  in  close proximity  to  the  human  body (frequency range of 30MHz to 6 GHz) This  information should  not be  used  as  a  basis  of comparisons  among devices since different parts of the standards mentioned may have been used. Sorin CRM declares that this device is in conformity with the
32 – US-ENGLISH essential  requirements  of  Directive  1999/5/EC  on  Radio  and Telecommunications  Terminal  Equipment,  with  the  mutual recognition of their conformity (R&TTE).  Federal Communication Commission Interference Statement 47 CFR Section 15.19 and 15.105(b)  - The FCC product ID is YSGDR9550. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are  designed  to  provide  reasonable  protection  against  harmful interference in a residential  installation. This  equipment generates  uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.  This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:  (1)  This device may  not cause  harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.  FCC Interference Statement 47 CFR Section 15.21 - No Unauthorized Modifications CAUTION: This equipment may not be modified, altered, or changed in any  way  without  signed  written  permission  from  SORIN.  Unauthorized modification may void the equipment authorization from the FCC and will void the SORIN warranty.  Identification of the equipment according Section 95.1217(a) This transmitter is authorized by rule under the Medical Device  Radiocommunication Service (in part 95 of the FCC Rules) and must not cause harmful interference to stations operating in the 400.150 406.00 MHz band in the Meteorological Aids (i.e., transmitters and receivers used
US-ENGLISH – 33 to communicate weather data), the Meteorological Satellite, or the Earth Exploration Satellite Services and must accept interference that may be caused by such stations, including interference that may cause undesired operation. This transmitter shall be used only in accordance with the FCC Rules governing the Medical Device Radiocommunication Service. Analog and digital voice communications are prohibited. Although this transmitter has been approved by the Federal Communications Commission, there is no guarantee that it will not receive interference or that any particular transmission from this transmitter will be free from interference.  IC Requirements for Canada - The IC product ID is 10270A-DR9550 This  class  B  digital  apparatus  meets  all  requirements  of  the  Canadian Interference- causing equipment regulations. This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device  may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Under Industry Canada  regulations,  this radio  transmitter may only  operate using an antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry Canada. To reduce potential  radio  interference  to  other  users,  the  antenna  type  and  its  gain should be so chosen that the equivalent isotropically radiated power (e.i.r.p.) is not more than that necessary for successful communication. This device may not interfere with stations operating in the 400.150–406.000 MHz  band  in  the  Meteorological  Aids,  Meteorological  Satellite,  and  Earth Exploration  Satellite  Services  and  must  accept  any  interference  received, including interference that may cause undesired operation.
34 – US-ENGLISH 10. PHYSICIAN GUIDELINES 10.1. PHYSICIAN TRAINING Physicians  should  be  familiar  with  sterile  pulse  generator  implant procedure  and  familiar  with  follow-up  evaluation  and  management  of patients with an implantable defibrillator (or referral to such a physician). 10.2. DIRECTIONS FOR USE ICD  operating  characteristics  should  be  verified  at  the  time  of implantation  and  recorded  in  the  patient  file.  Complete  the  Patient Registration  Form  and  return  it  to  Sorin CRM,  as  it  provides necessary information for warranty purposes and patient tracking. Additional  programming  instructions  can  be  found  by  accessing Online Help (click the “?” on the screen) on the Sorin CRM dedicated programmer.  Paper  copies  of  Online  Help  can  be  obtained  by contacting your Sorin CRM representative. 10.3. MAINTAINING DEVICE QUALITY This  device  is  FOR  SINGLE  USE  ONLY.  Do  not  resterilize  and reimplant explanted ICDs. Do not implant the device when:  It has been dropped on a hard surface because this could have damaged pulse generator components.  Its  sterility  indicator  within  the  inner  package  is  not  green, because it might not have been sterilized.  Its  storage  package  has  been  pierced  or  altered,  because  this could have rendered it non-sterile.  It  has  been  stored  or  transported  outside  the  environmental temperature limits: 32 °F (0 °C) to 122 °F (50 °C) as an electrical reset condition may occur.
US-ENGLISH – 35  "Use Before" date has expired, because this can adversely affect pulse generator longevity or sterility. 11. PATIENT INFORMATION Information  for  the  patient  is  available  in  the  patient  booklet, contained  in  the  outer  storage  package.  Additional  copies  can  be obtained  by  contacting  your  Sorin CRM  representative  or  on  the Sorin CRM's web site: http://www.sorin.com. This information should  be given to  each patient  with their  first ICD and  offered  to  the  patient  on  each  return  visit  or  as  deemed appropriate. 12. HOW SUPPLIED 12.1. STERILITY The PARADYM RF defibrillators are supplied one per package in  a sterile package. 12.2. WARRANTY AND REPLACEMENT POLICY Sorin CRM warrants its defibrillators. Refer to the section "Warranty" for additional information. Please see the following labelling sections for  information  concerning  the  performance  of  this  device: Indications,  Contraindications,  Warnings  and  Precautions,  and Adverse Events. 13. DEVICE DESCRIPTION The PARADYM RF DR system includes the model 9550 ICD device and  programming  system.  The  programming  system  includes  the Sorin CRM  Orchestra  Plus  programmer  with  SMARTVIEW programming software connected to a CPR3 programming head. The programming system is configured and furnished by Sorin CRM.
36 – US-ENGLISH The PARADYM RF DR 9550 can serve as a defibrillation electrode (active housing) with a total surface area of 76 cm². The PARADYM RF DR 9550 is designed to recognize and treat slow or fast  VT  and  VF  by  continuously  monitoring  atrial  and  ventricular activity  to  identify  persistent  ventricular  arrhythmias  and  to  deliver appropriate  therapies.  PARADYM  RF  DR  9550  features  the PARAD/PARAD+  algorithm,  which  is  specifically  designed  to differentiate  ventricular  tachycardias  from  fast  rhythms  of supraventricular origin. PARAD/PARAD+ continuously monitors R-R interval stability, searches for long cycles, assesses the degree of P-R association, evaluates sudden onset and determines the chamber of arrhythmia acceleration. In addition to the advanced detection scheme, PARADYM RF DR 9550 offers  programmable  dual  or  single-chamber  pacing  therapy  (DDD, DDI,  VVI  or  SafeR  modes)  with  or  without  rate-responsive capabilities  (DDDR,  DDIR,  VVIR,  DDD/DDIR  and  SafeR-R  modes) using  an  acceleration  sensor.  An  automatic  AV  delay  algorithm  as well as a mode switching function are available. PARADYM  RF  DR  9550  offers  tiered  therapy.  Therapies  can  be programmed independently in each zone:  in  the  Slow  VT  and  VT  zones:  two  ATP  programs,  up  to  two shocks  with  programmable  energy  and  up  to  four  shocks  with maximum energy can be programmed;  in  the  VF  zone:  one  ATP  program,  up  to  two  shocks  with programmable  energy  and  up  to  four  shocks  with  maximum energy can be programmed. When  the  rhythm  changes  from  one  zone  to  another,  the  device delivers the therapy programmed in this zone, starting with the same or more  aggressive  program for the  area. The  ATP  program  in  the VF zone will only be applied if the VT coupling interval is longer than the programmed fast VT cycle length.
US-ENGLISH – 37 The PARADYM RF DR 9550 offers biphasic shocks with a maximum stored energy of 42 J.  The  shock  configuration (electrodes  used to apply the shock) can be chosen by programming one of the following combinations: can and one coil, can and 2 coils, 2 coils only. Other features are as follows:  Automatic ventricular sensitivity control  Non-committed shocks  Electrophysiological  studies  (EPS)  with  real-time  markers  or electrograms:   Programmer-controlled VT induction sequences,  Programmer-controlled VF inductions (30 Hz rapid pacing or shock on T),  Programmable  electrogram  vectors  (EGM  A,  EGM  V, RVcoil-CAN, SVC-CAN, RVcoil-SVC)  Real-time  annotations  displayed  with  the  markers  and indicating the majority rhythm,  Manual ATP sequences,  Manual shocks.  Rescue shock  Follow-up tests:   Pacing lead impedance,  Coil impedance,  Capacitor charge time,  Pacing threshold tests.    Data storage:   Therapy History Report,  Statistics  (pace/sense,  therapy,  shocks,  and  battery voltage),
38 – US-ENGLISH  Up to 14 complete Holter records with event logs, marker channel notation, and electrogram records. The  connector  head  has  four  ports:  atrial  bipolar  pace/sense, ventricular  bipolar  pace/sense  and  two  ports  for  RV  and  SVC defibrillation coils. Both pace/sense ports are compatible with the IS-1 standard and both defibrillation ports are compatible with the DF-1 standard.  Distal  lead  terminal  connections  are  secured  with  set-screws accessed via self-sealing silicone plugs. All lead connections pass through the header into the device via feedthroughs. Programming  System:  The  Sorin CRM  programmer  is  used  in conjunction with specific programmer software to interrogate and program the implanted device at implant and during patient follow-up procedures. Remote Monitoring:  The  PARADYM  RF  DR  9550  is  also  equipped with the  RF  wireless  technology  which  enables to remotely monitor the patients who have the Sorin CRM SMARTVIEW Monitor installed at home. 14. IMPLANT PROCEDURE 14.1. NECESSARY EQUIPMENT Implantation  of PARADYM  RF  DR  9550  requires  the  following equipment:  Sorin  CRM  dedicated  programmer,  equipped  with  the SMARTVIEW software interface and with the programming head,  pacing system analyser, as well as its sterile connecting cables, to evaluate the pacing and sensing thresholds,  a complete set of leads with corresponding introducers,  physiological signal monitor capable of displaying simultaneously the surface ECG and arterial pressure,  an external defibrillator with sterile external paddles,  sterile cover for the telemetry head.
US-ENGLISH – 39 14.2. PACKAGING  Contents The PARADYM RF DR 9550 and its accessories are ethylene oxide sterilized and hermetically sealed in two-ply clear packaging meeting international requirements. The  sterile  packaging  contains  a  defibrillator,  one  screwdriver,  and an insulating plug for the DF-1 defibrillation connector. The non-sterile items contained in the outer storage package are the implant  manual,  the  ICD  Registration  Form  and  its  envelope,  the patient booklet, the ICD ID card and 12 identification labels. Once  delivered,  PARADYM  RF  DR  9550  is  programmed  to  as-shipped values  that are  different from  nominal values (see Chapter “Programmable Parameters” for details). 14.3. OPTIONAL EQUIPMENT The  following  equipment  may  be  required  during  implantation of PARADYM RF DR 9550:  an IS-1 insulating plug to close the atrial port  sterile  water  to  clean  traces  of  blood.  Any  parts  cleaned  with sterile water must be thoroughly dried.  mineral oil to lubricate if necessary  a lead cap to isolate a lead which is not used 14.4. BEFORE OPENING THE PACKAGE Before opening the package, check the "Use Before" date printed on the labels  on the box  and  on the sterile package.  Defibrillators that have  not  been  implanted  before  that  date  should  be  returned  to Sorin CRM.
40 – US-ENGLISH Devices  MUST  NOT  be  interrogated  and  programmed  within  the vicinity of other devices. Also  check  the  integrity  of  the  sterile  package.  The  sterility  of  the contents is no longer guaranteed if the package has been pierced or altered. If the defibrillator is no longer sterile, it should be returned in its packaging  to  Sorin CRM.  Any re-sterilization of the unit is at  the discretion of Sorin CRM. 14.5. PRIOR TO IMPLANTATION Use  the  programmer  to  verify  the  defibrillator  can  be  interrogated before implantation. Verify  all  shock  therapies  are  disabled  in  order  to  avoid  accidental discharge during implantation. It  is  not  advisable  to  program  the  Smoothing  function  before implantation, since the defibrillator  may  detect  noise  and  pace  at a rate higher than the programmed basic rate. CAUTION: Do not shake or tap sharply on the ICD package with the ICD inside, because the ICD's sensing circuits can interpret this as P-waves  or  R-waves  and  record  these  as  an  arrhythmia  episode.  If unusual  shaking  or  tapping  of  the  package  results  in  a  stored arrhythmia episode, erase the recording before using the ICD. 14.6. DEVICE PLACEMENT The  pocket  should  be  prepared  in  the  left  pectoral  position,  either subcutaneously or submuscularly. Subcutaneous device implantation is recommended for optimal RF communication efficacy. Implantation in an abdominal position is not advisable. In  its  final  position,  the  defibrillator  should  be  no  more  than  4  cm below the skin surface.
US-ENGLISH – 41 14.7. CHOOSING THE TYPE OF LEAD The defibrillator should be connected to:  one bipolar atrial sensing/pacing lead   one  ventricular  defibrillation  lead  with  sensing/pacing  bipolar electrodes, and one or two defibrillation electrodes. The  choice  of  leads  and  their  configuration is  left  to  the  implanting physician’s judgment. Note:  In  case  no  atrial  lead  is  implanted,  the  atrial  port  should  be plugged with IS-1 insulating plug and a single chamber mode (VVI-VVIR) should be programmed. PARAD and PARAD+ should not be used. Connectors:  The  bipolar  pacing/sensing  connectors  are  compliant with the IS-1 standard and the defibrillation connectors are compliant with the DF-1 standard.  Shock configuration (+ -> -) The  shock  configuration  is  the  energy  pathway  between  the defibrillation  electrodes.  If  an  atrial  coil  is  present,  the  shock configuration can be programmed for bi-directional shocks.
42 – US-ENGLISH Programming: When active case and SVC are both programmed to Yes, the shock configuration can be programmed to: 1. RV to Case (or Case to RV), 2. RV to SVC (or SVC to RV), 3. RV to Case+SVC (or Case+SVC to RV). The polarity of shock is determined by the parameter itself.    RV to Case+SVC RV to Case RV to SVC 14.8. MEASUREMENT OF THRESHOLDS AT IMPLANT Pacing and sensing thresholds should be measured at implant. Pacing thresholds: Acute thresholds should  be  lower than 1 V  (or 2 mA)  for  a  0.35  ms  pulse  width,  both  in  the  ventricle  and  in  the atrium. Sensing  thresholds:  For  proper  ventricular sensing,  the  amplitude of  the  R-wave  should  be  greater  than  5  mV.  For  proper  atrial sensing, the amplitude of the P-wave should be greater than 2 mV. Pacing  impedance  measurements:  Ventricular  and  atrial  pacing impedances should range from 200 to 3000 ohms (refer to the lead characteristics, especially if high impedance leads are used).
US-ENGLISH – 43 14.9. LEAD CONNECTION Implant the ventricular lead, then the atrial lead. Each lead  must  be  connected to the corresponding connector  port. The position of each connector is indicated on the casing. CAUTION: Tighten only the distal inserts. To connect each lead, proceed as follows: 1. Clean  the  lead  terminal  pins  thoroughly,  if  necessary  (device replacement). 2. Lubricate the lead terminal pins with sterile water, if necessary. 3. Do  not  insert  a  lead  connector  pin  into  the  connector  block without first  visually  verifying  that  the lead  port is  not  filled  with any obstacle. 4. Insert  the  screwdriver  into  the  pre-inserted  screw  socket  of  the appropriate port (in order to allow excess air to bleed out and to make the insertion of the lead pin easier). 5. Insert  the  lead  pin  all  the  way  into  the  port  (check  that  the  pin protrudes beyond the distal insert). 6. Tighten,  check  the  tightness  and  ensure  the  lead  pin  still protrudes beyond the distal insert, and did not move.
44 – US-ENGLISH CAUTION: 1. One  single  set  screw  is  located  on  the  side  of  the connection  header.  2. Do  not  tighten  the  pre-inserted  screws  when there is no lead (this could damage the connector). 3. Do not loosen the screws before inserting the connector (subsequent risk of being unable to reinsert the screw). 4. Removing the screwdriver: to avoid all risk of loosening screws during removal, hold the screwdriver by its  metal  part  and  not  by  the  handle.  5. When  mineral  oil  or  sterile water is used to  make lead insertion easier, the screwdriver should remain inserted into the pre-inserted screw socket when checking the tightness. As a matter of fact, when the lead port is filled with a liquid, the  physics  piston  effect  can  give  the  feeling  the  lead  is  properly tightened. NOTE: To optimise cardioversion/defibrillation shocks, electrodes must be positioned so that the electric field between anode (s) and cathode covers  the largest myocardial mass.  In normal  conditions,  the anode and cathode  are  adequately separated. In case  of a short-circuit, the shock may be aborted to prevent damaging the defibrillator. In the case of an external defibrillation shock delivered to the patient, always  check  the  programming  and  functioning  of  the  device,  in particular its capacity to deliver shocks.
US-ENGLISH – 45 14.10.   DEVICE IMPLANTATION PARADYM RF DR 9550 should be implanted with the engraved side facing  outwards  for  optimal  communication  with  the  programming head and radiographic identification. Place the device in the pocket. Once in place, the defibrillator should be no more than 4 cm below the skin surface. Carefully wind excess lead and place in a separate pocket to the side of the defibrillator. It is recommended to not place any excess wire between the can and the heart. Suture the  casing  connector to  the  muscle  using  the  hole  provided for  this  purpose, in  order  to  avoid  potential  migration  of  the  device into the pectoral muscle. 14.11.  TESTS AND PROGRAMMING During  the  implant  testing  procedure,  it  is  recommended  that  a security  margin  of  at  least  10  J  be  demonstrated  between  the effective shock energy and maximum programmable energy. Enable shock therapies, then program the defibrillator. Verify that the defibrillation lead impedance for each shock delivered ranges from 30 to 150 ohms. Check the lead connection if the values are outside these boundaries. Save the programming data on  the  programmer’s hard disk and on an external storage device (if desired).
46 – US-ENGLISH 15. SPECIAL MODES 15.1. SAFETY MODE (NOMINAL VALUES) Nominal  values  may  be  rapidly  restored  by  pressing  the  following button on the programming head or programmer keyboard:  or via the "Emergency" button on the SMARTVIEW screen. In  safety  mode,  the  defibrillator  operates  with  the  parameters underlined in the table of programmable parameters. 15.2. MAGNET MODE When the magnet is applied:  antiarrhythmia  functions  are  inhibited  (detection  of  rhythm disturbances, charging, and therapy),  hysteresis and AVD paced/sensed offset are set to 0,  pacing amplitude is set to 6 V,  pulse width is set to maximum,  pacing rate is set to the magnet rate,  the  following  functions  are  disabled:  ventricular  arryhtmia prevention, Mode Switch, Anti-PMT, Smoothing, Rate Response. When the magnet is removed:  the sensor rate is forced to the basic rate,  arrhythmia  detection  algorithms  and  sequential  therapies  are reinitialized,  therapies start with the least aggressive program for each area. The other parameters remain at their programmed value.
US-ENGLISH – 47 The magnet rate values are as follow: Magnet rate (bpm) 96 94 91 89 87 85 Magnet period (ms) 625 641 656 672 688 703 Magnet rate (bpm) 83 82 80 78 77 Magnet period (ms) 719 734 750 766 781 15.3. RESPONSE IN THE PRESENCE OF INTERFERENCE If the defibrillator senses electrical noise at a frequency above 16 Hz,  it switches  to  an asynchronous mode  at the basic rate.  The programmed mode is restored as soon as the noise is no longer detected. Ventricular  pacing  is  also  inhibited  by  ventricular  noise.  It  can  be restored by setting the parameter V pacing on noise to Yes.
48 – US-ENGLISH 15.4. DETECTION CHARACTERISTICS IN THE PRESENCE OF ELECTROMAGNETIC FIELDS Per  Clause  27.4  of  Standard  EN  45502-2-2,  detection  in  the presence of electromagnetic fields is characterized as follows:  Differential mode:   Common mode rejection ratio:  16.6 Hz 50 Hz 60 Hz Atrial channel ≥ 75 dB 67 dB 67 dB Ventricular channel ≥ 69 dB ≥ 69 dB ≥ 69 dB Modulated  interference:  For  atrial  sensitivity  setting  of  0.2 mV, compliance to the Cenelec standard 45502-2-2 is met for a maximum test  signal  amplitude  of  8 V  for  the  frequency  of  60 MHz.  0.4 mV complies with the standard for the whole frequency range.
US-ENGLISH – 49 15.5. PROTECTION AGAINST SHORT-CIRCUITS The defibrillator can undergo a short-circuit if the anode and cathode are not adequately separated. In  this  case,  the  shock  is  aborted  to  prevent  damaging  the defibrillator  and  a  warning  will  indicate  that  a  short  circuit  (shock impedance < 20 ohms) was detected during the last shock. 16. MAIN FUNCTIONS 16.1. AUTOMATIC LEAD MEASUREMENTS Automatic pacing lead impedance measurement: A lead impedance measurement is automatically performed on atrial  and ventricular  leads every 6 hours. The daily mean impedance is stored for each chamber. Automatic  coil  impedance  measurement:  A  coil  impedance measurement is automatically performed on RV and SVC coils once a week. The result is stored in the device memory. 16.2. ATRIAL TACHYARRHYTHMIA MANAGEMENT Mode Switch: This function is designed to limit the acceleration and variation of ventricular rate in the presence of atrial arrhythmia.
50 – US-ENGLISH 16.3. VENTRICULAR TACHYARRHYTHMIA MANAGEMENT Ventricular  tachyarrhythmia  prevention:  Set  of  algorithms  that can  be  used  to  avoid  the  circumstances  of  ventricular tachyarrhythmia onset. Searching  for  a  long  cycle  (P  And  R  based  Arrhythmia Detection+:  PARAD+):  Additional  arrhythmia  classification criterion  to  improve  identification  of  atrial  fibrillation  and  avoid inappropriate shocks. Fast  VT  treatment:  Applies  detection  criteria  on  fast  ventricular tachycardiathat are different  from  those  of the VT  zone,  as  well  as different therapies. The fast VT zone is included in the VF zone: its lower limit is determined by the programmed value for the VF zone and its upper limit by the programmed value for the fast VT zone. Polarity  alternation  on  Max  shock:  Reverses  the  programmed polarity of every second shock set at maximum energy. The number, type,  and  energy  of  shocks  is  independently  programmable  by detection zone.
US-ENGLISH – 51 16.4. PACING BTO  (Brady  Tachy  Overlap):  Corrects  chronotropic  atrial incompetence  by  allowing  pacing  in  the  slow  VT  zone,  without affecting detection specificity. Post-shock  mode:  After  any  automatic  shock  therapy,  the  post-shock mode makes it possible to apply a pacing mode other than the standard  antibradycardia  pacing  mode  and/or  with  different  pacing parameters. SafeR  (AAI  <>  DDD)  mode:  Is  intended  to  minimize  deleterious ventricular  pacing.  The  defibrillator  functions  in  AAI  mode,  and temporarily switches to DDD  mode upon the occurrence of AVB III, AVB II, AVB I and ventricular pause. Anti-PMT  protection:  Is  intended  to  protect  the  patient  from Pacemaker-Mediated  Tachycardia  (PMT)  without  reducing  atrial sensing capability of the device. 16.5. SENSING Automatic  Refractory  Periods:  Optimize  sensing  and  make  the implant  progamming  easier.  These  periods  are  composed  of  a minimal  Refractory Period  and a triggerable Refractory Period.  The duration of the refractory periods lengthens automatically as needed. Committed period: 1. In DDI or DDD modes, the committed period is  a  non-programmable  95  ms  ventricular  relative  refractory  period that starts  with  atrial  pacing. If a ventricular event  is  sensed  during the committed period, but outside the blanking period, the ventricle is paced  at  the  end  of  the  committed  period.  The  committed  period prevents inappropriate ventricular inhibition if crosstalk occurs.  Protection against noise: Allows the distinction between ventricular noise  and  ventricular  fibrillation.  If  the  device  senses  ventricular noise, the ventricular sensitivity is decreased until noise is no longer
52 – US-ENGLISH detected.  Ventricular  pacing  can  be  inhibited  to  avoid  a  potential paced T-wave. Automatic sensitivity control: Optimizes arrhythmia detection and avoids  late  detection  of  T-waves  and  over-detection  of  wide  QRS waves.  The  device  automatically  adjusts  the  sensitivities  based  on the ventricular sensing amplitude. In case of arrhythmia suspicion or after  a  paced  event,  the  programmed  ventricular  sensitivity  will  be applied.  The  minimum  ventricular  sensitivity  threshold  is  0.4  mV (minimum programmable value). 16.6. FOLLOW-UP FUNCTIONS Storage  of  memory  data:  AIDA+  (Automatic  Interpretation  for Diagnosis  Assistance)  software provides  access  up  to  6  months  of patient follow-up  with day  by  day data  collection,  or up  to 24  hours with  hourly  data  collection.  Episodes  of  ventricular  tachyarrhythmia are  recorded  with  the  programmable  EGM  channels:  either  by selecting up to two traces, or by selecting "V-Double" which enables a one-channel recording that is twice as long. Diagnosis  of  AV  conduction:  Automatic  diagnosis  of  AV conduction with graphic displays. Alerts  /  Warnings:  The  device  routinely  performs  security  self-checks  and  technical  measurements  to  ensure  system  integrity. When system integrity is found to be at risk outside a follow-up, alerts are stored in the device memory. When system integrity is found to be  at  risk  during  a  follow-up,  the  information  is  managed  as  a warning  (pop-up  message)  to  notify  immediately  the  user.  For example,  the  following  types  of  event  can  trigger  a  warning  or  an alert:  technical  problem  during  a  shock,  pacing  lead  impedance  or coil impedance measurements out-of-range, battery depletion, …
US-ENGLISH – 53 16.7. REMOTE MONITORING FUNCTION Remote  monitoring  enables  the  automatic  remote  transmission  of implant data to the physician thanks to the wireless Radio Frequency (RF)  communication  ability  of  the  implant  in  order  to  provide  a comprehensive report to the physician about device functioning and patient  cardiac  status  without  having  the  patient  physically  in  the clinic. The  data  is  transmitted  from  the  implant  and  the  SMARTVIEW monitor, a small transmitter placed in the patient home. Implant data are first transmitted to the SMARTVIEW monitor via RF. Data  are  then  rooted  through  the  phone  network  to  an  internet website. This website is responsible for transforming the implant data into a comprehensive report that can be consulted by the physician.  SMARTVIEW Monitor The  SMARTVIEW  monitor  is  a  small  device  equipped  with  an  RF transmission module to communicate with the implant and a modem to export data through the internet. The  SMARTVIEW  monitor  is  delivered  to  the  patient  who  has  to install it at home. Preferably the SMARTVIEW monitor will be placed on the nightstand of the patient, as  close as possible to the side of the  bed  the patient  usually  sleeps.  The  SMARTVIEW  monitor  shall be  connected  to  the  phone  network  and  the  power  plug.  Regular transmissions are done  during  the  night  when  the  patient is  asleep next  to  the  SMARTVIEW  monitor  without  any  intervention from  the patient.
54 – US-ENGLISH  Transmission trigger There are 3 different triggers for a remote transmission:  the remote follow-up transmission is scheduled by the physician to occur regularly (according to the programming).  the  alert  transmission  will  take  place  when  the  implant  has recorded  an  abnormal  event.  The  list  of  abnormal  event  is available  in  a  following paragraph. Alert  conditions  are  checked daily.  the on-demand follow-up transmission is triggered by the patient himself  through  the  use  of  a  specific  button  on  the remote-monitor.  Data transmitted The  data  transmitted  are  identical  to  the  data  available  during  a standard  interrogation  with  the  Orchestra  Plus  programmer. All counters,  histograms,  IEGMs  and  diagnosis  available  in  the device are transmitted containing (not exhaustive list):  programmed parameters  Information on patient and system implanted  battery status  lead status (brady leads and defibrillation coils)   pacing counters and mean heart rate (brady)  atrial and ventricular arrhythmia counters and episodes  ventricular therapy counters  heart failure monitoring Data are presented in the form of 2 reports to the physician: the first one  contains  a  summary  of  major  counters,  histograms,  warnings and diagnosis. The second one presents the 3 most important IEGM episodes automatically selected based on the degree of severity for the patient.
US-ENGLISH – 55  User website On the website, the physician is able to:  consult and schedule the remote follow-ups of their patient  configure additional ways of being notified of alerts (for instance by SMS, fax or e-mail)  consult, print and export patient reports   Alert system The following set of  alert trigger can be independently programmed ON/OFF by the physician using the Orchestra Plus programmer and can trigger an alert transmission:  Reset of the device  ERI reached  Low or high impedance (A, RV, LV)  Abnormal coil impedance (shock lead)  Low or High shock impedance  Long charge time  Inefficient high energy shock  All shocks programmed OFF  Shock treated VT/VF  Lack of V pacing in CRT device  Suspicion of noise on the V lead  Fast V rate during AF WARNINGS The  use  of  remote  monitoring  does  not  replace  regular  follow-up. Therefore,  when  using  remote monitoring,  the time  period  between follow-ups visits may not be extended. When  ERI  mode  is  reached,  this  information  is  transmitted  via  the remote monitoring facility and then the remote-monitoring is switched off to preserve battery life.
56 – US-ENGLISH 17. PATIENT FOLLOW-UP 17.1. FOLLOW-UP RECOMMENDATIONS Before the  patient  is  discharged  and  at  each  subsequent  follow-up visit, it is advisable to:  check the occurrence of system warnings  check the battery status,  check the integrity of the pacing and defibrillation leads,  check  for  proper  sensing  (sensitivity,  crosstalk)  and pacing ;  set  the  pacing  amplitude  to  twice  the  pacing threshold,  interrogate the implant memories (AIDA+),  check the efficacy of the therapies delivered,  keep  a  printout  of  programmed  parameters,  test  results, and memory data,  reset the memory data and statistics. These  operations  should  be  performed  by  medical  personnel  in  an appropriate care unit, with resuscitation equipment present. It is recommended that a routine follow-up examination be done one month after discharge, and then every three months until the device nears the replacement date. After a  device reset, the  magnet  rate  is  equal to  87  ppm;  it  will  be updated within the next 24 hours. Refer to the online help for a description of displayed warning, and the necessity to contact Sorin CRM for an evaluation. Implant  software  upgrade:  In  case  a  new  implant  software  is downloaded  in  the  device  memory  through  the  programmer,  a warning  message  could  be  displayed  by  the  programmer  to  inform the user and give the proper instructions to follow.
US-ENGLISH – 57 17.2. HOLTER FUNCTION The Holter records up to 14 tachyarrhythmia episodes as well as the therapy history. STORED EPISODES PARADYM RF DR 9550 stores up to 14 episodes (VF, VT, Slow VT, SVT/ST, nonsustained). For each episode four levels of details are presented:  Tachogram (to visualize PP and PR intervals)  Event log for the entire episode:   PARAD/PARAD+ analysis for each majority,  Delivered therapies,  Markers:  Atrial  and  ventricular  markers,  sensed,  paced  and  in relative refractory periods,  EGM:  onset  and  detection  of  the  arrhythmia,  on  two  therapies, and the return to slow rhythm by recording electrogram. Therapy  history:  For  each  arrhythmia  detection,  each  therapy delivered  (either  automatically  or  during  an  electrophysiological study) and  at the end of each  arrhythmia,  PARADYM  RF DR 9550 records the  type of  majority  rhythm,  the number  of ATP sequences delivered, the energy and the number of shocks delivered.
58 – US-ENGLISH 17.3. ELECTIVE REPLACEMENT INDICATOR (ERI) Elective Replacement Indicators (ERI)(1) are:  magnet rate equal to 80 ± 1 min-1 or  battery voltage equal to 2.66 V ± 0.01 V CAUTION:  The  defibrillator  should  be  replaced  as  soon  as  the Elective Replacement Indicator (ERI) point is reached.  Between  the  ERI  and  the  EOL  (End  of  Life)(2),  PARADYM  RF  DR 9550 can still function for:  8.4  months  (100%  atrial  and  ventricular  pacing  in  DDD  mode, 500 ohms, with as-shipped settings), and deliver 7 shocks at 34 J or  6.4  months  (0%  pacing,  sensor  OFF,  one  42 J  shock  every 2 weeks). Once  the  Elective  Replacement  Indicator  (ERI)  point  has  been reached, the device operates normally, except  that the  charge time increases.  Under  normal  conditions  (and  without  programmer  use) the charge times are as follows:  Shock energy Charge time (sec)  BOL 42 J 10 (± 2)  ERI 42 J 13 (± 3)  (1) Elective  Replacement  Indicators  (ERI)  corresponds  to Recommended  Replacement  Time  (RRT)  as  referred  in  the EN45502-2-2 standard. (2) End  of  Life  (EOL)  corresponds  to  End  of  Service  (EOS)  as referred in the EN45502-2-2 standard.
US-ENGLISH – 59 17.4. EXPLANTATION The defibrillator should be explanted in the following cases:  The Elective Replacement Indicator (ERI) point is reached  Confirmed malfunction  Burial of the patient (for environmental reasons, the local regulation  may  require  the  explantation  of  the  devices containing a battery supply)  Cremation  of  the  patient  (the  defibrillator  may  explode  if placed in an incinerator) The explanted defibrillator should not be reused in another patient. All explanted defibrillators should be returned to Sorin CRM, carefully cleaned  of  all  traces  of  contamination.  This  may  be  done  by immersing  them  in  an  aqueous  sodium  hypochlorite  containing  at least 1% chlorine, followed by rinsing copiously with water. The defibrillator should be protected against mechanical impact and the temperature variations that may occur during shipping. Before explantation, it is advisable to:  print out all programmed parameters, statistics and Holter function report,  save Patient data on floppy disk or hard disk,  disable shock therapies (VT and VF) to  avoid any  risk of untimely shock.
60 – US-ENGLISH 17.5. DEFIBRILLATOR IDENTIFICATION The defibrillator can be interrogated and programmed via telemetry, using  the  programming  head  interfaced  with  the  Sorin  CRM dedicated programmer. Position the programming head over the telemetry antenna located in the upper part of the device, in order to communicate effectively via telemetry (see diagram below).  The device can be non-invasively identified as follows: 1. Take  an  X-ray  to  identify  the  name  of  the  manufacturer  and model, printed on the device (x-ray ID is SDC : S = SORIN ; D = Defibrillator ; C = PARADYM RF DR 9550).   2. Interrogate  the  device  using  the  Sorin  CRM  dedicated programmer.  The  model  and  serial  number  of  the  device  are automatically  displayed.  The  first  figure  in  the  serial  number corresponds to the last figure in the year of manufacture.
US-ENGLISH – 61 18. PHYSICAL CHARACTERISTICS Dimensions 69.5 x 73.4 x 11 mm Weight 95 g Volume 38.6 cm3 Active surface area of casing 76 cm2 Connector Atrium: IS-1. Ventricle: IS-1, DF-1. 18.1. MATERIALS USED Active surface area of casing 99% pure titanium Connectors Polyurethane* and silicone elastomer* DF-1 insulating plug silicone elastomer* * Medical-grade materials that have undergone “in vitro” and “in vivo” qualifications
62 – US-ENGLISH 19. ELECTRICAL CHARACTERISTICS Atrial input impedance 80 kilohms ± 30 % Ventricular input impedance 80 kilohms ± 30 % D.C. capacitance 148 µF ± 8 % Capacitor formation No formation required Rate limit 192 min-1 ± 10 min-1 Pacing waveform  Defibrillation waveform
US-ENGLISH – 63 19.1. TABLE OF DELIVERED SHOCK ENERGY AND VOLTAGE The  relationship  between  stored  energies,  maximum  voltages  and delivered  energies  (at  37  °C,  50  ohm  load)  for  the  minimum,  low, mean and maximum programmed energy values is as follows: Stored energy (J) 0.5 10 20 34 42 V1 (Volt) 75 341 483 631 702 V2 (Volt) 37 173 245 318 353 Delivered E: Phase 1 (J) 0.31 7.0 14.0 23.9 29.6 Delivered E: Phase 2 (J) 0.08 1.8 3.6 6.1 7.5 Delivered E: Total (J) 0.4 8.8 17.6 30.0 37.1 Tolerances are 12% for voltage (25% at 0.5 J) and 30% for energy.
64 – US-ENGLISH 19.2. BATTERY Manufacturer Greatbatch Type Quasar High Rate (QHR) Model GB 2593 Number of batteries 1 Total capacity 1964 mAh Usable capacity Between BOL and ERI: 1278 mAh. Between BOL and EOL: 1675 mAh. Voltage BOL: 3.25 V. ERI: 2.66 V. EOL: 2.5 V. 19.3. LONGEVITY The  longevities  mentioned  below  are  calculated  by  taking  into account 6 months storage. 6.0 years Pacing in DDD mode, 100%, 500 ohm, 3.5 V, 0.35 ms, 60 min-1, one 42 J shock per quarter, sensor OFF 5.8 years Pacing in DDD mode, 100%, 500 ohm, 3.5 V, 0.35 ms, 60 min-1, one 42 J shock per quarter, G sensor ON 8.7 years Pacing in DDD mode, 15% in atrium, 1% in ventricle, 500 ohm, 3.5 V, 0.35 ms, 60 min-1, one 42 J shock per quarter, sensor OFF 7.8 years Pacing in DDD mode, 15% in atrium, 15% in ventricle, 500 ohm, 4.5 V, 0.50 ms, 60 min-1, one 42 J shock per quarter, sensor OFF 9.1 years 0% pacing, one 42 J shock per quarter, sensor OFF
US-ENGLISH – 65 The  mean longevity  as  a function  of  shocks  delivered  at  maximum energy, with and without pacing, is as follows:  The mean longevity as a function of yearly remote follow-ups(1), with and without pacing, is as follows:  (1)  An  excessive  number  of  remote  follow-up  can  have  a  non-negligible impact on device longevity.
66 – US-ENGLISH 20. PROGRAMMABLE PARAMETERS measured at 37 °C under a 500 ohm load Legend: Value in bold: “as shipped” value Underlined value: nominal value 20.1. ANTIBRADYCARDIA PACING Basic parameters Values Mode VVI-VVIR-DDD-DDDR-DDD/DDIR-DDI-DDIR-SafeR (AAI <=> DDD)-SafeR-R (AAIR <=> DDDR) Basic rate (min-1) (1) From 30 to 90 by steps of 5 ; 60 (± 4 %) Maximum rate (min-1) From 100 to 145 by steps of 5 ; 120 (± 6 %) Rate hysteresis (%) 0-5-10-20-35 (± 18 ms) Rest AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms) Exercise AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms) AVD Paced/Sensed Offset (ms) 0-10-15-25-30-40-45-55-65-70-80-85-95-100-110-115-125 (± 1 ms) (1) The corresponding periods are (in ms): 2000-1714-1500-1333-1200-1091-1000-923-857-800-750-706-667 ms.
US-ENGLISH – 67 Special features Values Smoothing OFF-Very slow-Slow-Medium-Fast Mode Switch ON-OFF Mode Switch Rate (min-1) From 30 to 90 by steps of 5 ; 60 Anti-PMT protection Termin-Reprog Physical activity Very low-Low-Medium-High-Very high
68 – US-ENGLISH Pacing/Sensing Values Atrial sensitivity (mV) (1) From 0.2 to 4 by steps of 0.2 ; 0.4 (± 50 %) Atrial amplitude (V) (2) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %) Atrial pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %) Ventricular sensitivity (mV) (1) From 0.4 to 4 by steps of 0.2 ; 0.4 (± 50 %) Ventricular amplitude (V) (2) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %) Ventricular pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %) (1)  Values  are  measured  using  a  positive  and  negative  triangular  signal  of 2/13 ms. (2)  The  correlation  between  the  programmed  amplitudes,  the  stored amplitudes  and  the  mid-pulse  delivered  amplitudes  under  a  500  ohm load are given in the following table: Programmed ampl. (V) 1 1.5 2 2.5 3 3.5 Mid-pulse delivered ampl. (V) 0.97 1.39 1.79 2.35 2.84 3.25 Stored amplitude (V) 1.14 1.63 2.1 2.76 3.33 3.82 Programmed ampl. (V) 4 4.5 5 6   Mid-pulse delivered ampl. (V) 3.58 4.23 4.47 5.37   Stored amplitude (V) 4.2 4.96 5.25 6.3
US-ENGLISH – 69 Ventricular arrhythmia prevention Values Atrial pacing on PVC Yes-No Post extrasystolic pause suppression Yes-No Acceleration on PVC Yes-No Max accelerated rate (min-1) From 60 to 145 by steps of 5 ; 100 Post-shock mode Values Mode OFF-VVI-DDI-DDD Duration 10s-20s-30s-1min-2min-3min-4min-5min Basic rate (min-1) From 50 to 90 by steps of 5 ; 60 (± 4 %) Rest AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms) Exercise AV delay (ms) 30-40-45-55-65-70-80-85-95-100-110-115-125-135-140-150-155-165-170-180-190-195-205-210-220-225-235-250 (± 19 ms) AVD Paced/Sensed Offset (ms) 0-10-15-25-30-40-45-55-65-70-80-85-95-100-110-115-125 (± 1 ms) A amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %) A pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %) V amplitude (V) 1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %) V pulse width (ms) 0.12-0.25-0.35-0.5-0.6-0.75-0.85-1 (± 10 %)
70 – US-ENGLISH Refractory periods Values Atrial refractory period post ventricular sensing (ms) 45-65-80-95-110-125-140-155 (± 16 ms) Atrial refractory period post ventricular pacing (ms) 80-95-110-125-140-155 (± 4 ms) Sensitivity margins Values Atrial post pacing/sensing margin (mV) From 0 to 1 by steps of 0.2 ; 0.4 Ventricular post pacing margin (mV) From 0 to 2 by steps of 0.2 ; 0.8 Response to noise Values Automatic sensitivity on noise ON-OFF V pacing on noise ON-OFF SafeR (AAI <=> DDD) parameters Values AVB I switch Rest+Exercise-Exercise Long PR: max (ms) From 200 to 500 by steps of 50 ; 450 Long PR: min (ms) From 200 to 500 by steps of 50 ; 250 Max. pause (s) 2-3-4
US-ENGLISH – 71 20.2. VENTRICULAR TACHYARRHYTHMIA DETECTION Therapy zones Values Slow VT detection zone (1) Slow VT ON-Slow VT OFF VT detection zone VT ON-VT OFF Fast VT / VF detection zone Fast VT+VF ON-VF ON Slow VT rate (lower limit) (min-1) From 100 to 200 by steps of 5 ; 190 VT rate (lower limit) (min-1) 130-135-140-145-150-155-160-165-170-175-180-185-190-195-200-210-220-230 VF rate (lower limit) (min-1) 150-155-160-165-170-175-180-185-190-195-200-210-220-230-240 Fast VT rate (upper limit) (min-1) 155-160-165-170-175-180-185-190-195-200-210-220-230-240-255 Slow VT persistence (cycles) 4-6-8-12-16-20-30-50-100-200 VT persistence (cycles) 4-6-8-12-16-20-30-50-100-200 VF persistence (cycles) From 4 to 20 by steps of 1 ; 6 (1)  The  Slow  VT  zone  should  be  programmed  ON  only  if  the  VT  zone  is programmed ON.
72 – US-ENGLISH Detection criteria Values Slow VT and VT detection criteria Rate Only-Stability-Stability+-Stability/Acc-Stability+/Acc-PARAD-PARAD+ Fast VT detection criteria Rate+Stability-Rate Only Majority: (X/Y), Y (cycles) 8-12-16 Majority: (X/Y), X (%) 65-70-75-80-90-95-100 Window of RR stability for Slow VT and VT (ms) 30-45-65-80-95-110-125-125 Window of RR stability for fast VT (ms) 30-45-65 Prematurity acceleration (%) 6-13-19-25-31-38-44-50 Long cycle persistence extension (cycles) From 0 to 16 by steps of 1 ; 10 Long cycle gap (ms) 15-30-45-65-80-95-110-125-140-155-170-190-205 Atrial monitoring Yes-No
US-ENGLISH – 73 20.3. VENTRICULAR TACHYARRHYTHMIA THERAPIES Common parameters Values Enable ATP therapy Yes-No Enable shock therapy Yes-No Polarity alternation (42J) Yes-No Atrial coil (SVC) present Yes-No Active case Yes-No Shock configuration (+ --> -) Case to RV-SVC to RV-Case + SVC to RV-RV to Case-RV to SVC-RV to Case + SVC SVC exclusion (shock < 15J) Yes-No ♦ Therapy parameters in slow VT zone ATP 1 program Values ATP program OFF-Burst-Burst+Scan-Ramp-Ramp+Scan Number of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Coupling interval (%) 50-55-60-65-70-75-80-85-90-95 Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60 Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60 Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4 Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310
74 – US-ENGLISH ATP 2 program Values ATP program OFF-Burst-Burst+Scan-Ramp-Ramp+Scan Number of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Coupling interval (%) 50-55-60-65-70-75-80-85-90-95 Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60 Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60 Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4 Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310 Shock program Values Shock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9- (± 30 %) 10-12-14-16-18-20-22-24-26-28-30-32-34-42 (± 15 %) Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9- (± 30 %) 10-12-14-16-18-20-22-24-26-28-30-32-34-42 (± 15 %) Number of Max. Shock (42 J) OFF-1-2-3-4
US-ENGLISH – 75 ♦ Therapy parameters in VT zone ATP 1 program Values ATP program OFF-Burst-Burst+Scan-Ramp-Ramp+Scan Number of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Coupling interval (%) 50-55-60-65-70-75-80-85-90-95 Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60 Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60 Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4 Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310 ATP 2 program Values ATP program OFF-Burst-Burst+Scan-Ramp-Ramp+Scan Number of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Coupling interval (%) 50-55-60-65-70-75-80-85-90-95 Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60 Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60 Time limit (min) 0.5-1-1.5-2-2.5-3-3.5-4 Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310
76 – US-ENGLISH Shock program Values Shock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9- (± 30 %) 10-12-14-16-18-20-22-24-26-28-30-32-34-42 (± 15 %) Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9- (± 30 %) 10-12-14-16-18-20-22-24-26-28-30-32-34-42 (± 15 %) Number of Max. Shock (42 J) OFF-1-2-3-4 ♦ Therapy parameters in fast VT / VF zone ATP 1 program Values ATP program OFF-Burst-Burst+Scan-Ramp-Ramp+Scan Number of sequences 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles in first sequence 1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Cycles added per sequence 0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 Coupling interval (%) 50-55-60-65-70-75-80-85-90-95 Ramp decrement (per cycle) (ms) 0-4-8-12-16-20-30-40-50-60 Scan decrement (per sequence) (ms) 0-4-8-12-16-20-30-40-50-60 Time limit 10s-20s-30s-1min-1.5min-2min Minimum cycle length (ms) 95-110-125-140-155-170-190-205-220-235-250-265-280-295-310
US-ENGLISH – 77 Shock program Values Shock 1 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9- (± 30 %) 10-12-14-16-18-20-22-24-26-28-30-32-34-42 (± 15 %) Shock 2 (J) OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-7-8-9- (± 30 %) 10-12-14-16-18-20-22-24-26-28-30-32-34-42 (± 15 %) Number of Max. Shock (42 J) 1-2-3-4 20.4. REMOTE ALERTS AND WARNINGS General parameters Values RF communication (1) ON-OFF Remote alerts (1) ON-OFF (1)  RF  and  Remote  alerts  are  turned  on  automatically  if  Shocks  are programmed ON. System Alerts Values Battery depletion – ERI ON-OFF Device reset ON-OFF Excessive charge time (>25s) ON-OFF System integrity  ON-OFF
78 – US-ENGLISH Lead Alerts Values Abnormal A lead impedance ON-OFF Abnormal A lead low limit (Ohm) 200-250-300-350-400-450-500 Abnormal A lead high limit (Ohm) 1500-1750-2000-2500-3000 Abnormal V lead impedance ON-OFF Abnormal V lead low limit (Ohm) 200-250-300-350-400-450-500 Abnormal V lead high limit (Ohm) 1500-1750-2000-2500-3000 Abnormal RV coil impedance ON-OFF Abnormal SVC coil impedance ON-OFF Abnormal Shock impedance (1) ON-OFF (1) Normal impedance range [20 Ohm-200 Ohm] Clinical status Values V oversensing ON-OFF High AT/AF burden ON-OFF AT/AF limit (on 24h) (h) 0.5-1-3-6-12-24 Fast V Rate during AT/AF ON-OFF Fast V Rate limit (min-1) 80-90-100-110-120 Fast V Duration limit (h) 0.5-1-3-6-12-24 Therapy information Values Shock disabled ON-OFF Shocks delivered OFF-All shocks-Inefficient shock-Inefficient max shock
US-ENGLISH – 79 21. NON PROGRAMMABLE PARAMETERS Interval Values Committed period 95 ms (± 5 ms) Atrial refractory periods Values Post atrial sensing 47 ms (± 16 ms) Post atrial pacing 109 ms (± 4 ms) Ventricular refractory periods Values Post ventricular sensing 95 ms (± 16 ms) Post ventricular pacing 220 ms (± 4 ms) Post atrial pacing (blanking) 16 ms (± 3 ms) Tachycardia criteria Values Window of PR association 63 ms (± 1 ms) Therapies Values Waveform (1) Constant tilt (50% - 50%) Stored energy for the Max. shock 42 J (± 15 %) Pacing amplitude during ATP therapies 7 V (Actual value at 300 ms: 5.3 V) Anti-PMT protection Termin.  (1)  The  device  has  50%  tilt  in  each  phase  thus  delivers  94%  of  stored energy. Each phase is limited to 10 ms duration.
80 – US-ENGLISH 22. LIMITED WARRANTY The PARADYM RF implantable cardioverter defibrillator is the result of highly advanced research and all components have been selected after exhaustive testing. Sorin CRM S.r.l. (identified as “Sorin CRM” hereafter) guarantees the product PARADYM  RF  against  any  damage  caused  by  component failure or  production  defects  during  a  period  of  four years  after  the implantation  date,  and  Sorin  CRM  commits  itself  to  replace  all PARADYM RF devices according to the terms described in article 1 and described in article 2 of this section. Sorin  CRM  makes  no  claim  that  the  human  body  will  not  react unsuitably  to  the  implantation  of  the  PARADYM  RF  device,  or  that failure will never occur. Sorin  CRM  does  not  guarantee  the  suitability  of  PARADYM  RF  in defined  types  of  patients;  selection  of  the  device  is  a  medical decision. Sorin  CRM  shall  not  be  held  liable  for  any  damage  indirectly associated  with  the  PARADYM  RF,  whether  as  part  of  normal  or abnormal  operation,  nor  damage  from  its  explantation  or replacement. Sorin  CRM  does  not  authorise  anyone  to  modify  these  limited warranty conditions. 22.1. ARTICLE 1 : TERMS OF LIMITED WARRANTY 1. The  PARADYM  RF  implantable cardioverter defibrillator is  only guaranteed for one implantation. 2. The EURID/IAPM implant form must be sent to Sorin CRM within 30 days after implantation. 3. The  PARADYM  RF  cardioverter  defibrillator  must  be  implanted prior to the use-before date indicated on the packaging.
US-ENGLISH – 81 4. The limited guarantee  only applies to suspect devices  returned to  the  manufacturer,  carefully  packed  and  accompanied  by  an explantation report duly completed by the hospital or the doctor and considered defective after analysis by Sorin CRM.  The  device  must  be  returned  within  the  30  days  following explantation to Sorin CRM. Any device returned and replaced under the terms of this limited warranty will become the exclusive property of Sorin CRM. Any  rights  under  the  terms  of  this  limited  warranty  will  be forfeited  if  the  PARADYM  RF  device  has  been  opened  by anyone other than Sorin CRM. These  rights  will  also  be  forfeited  if  the  device  has  been damaged by carelessness or accident. This  is  the  case  especially  if  the  device  has  been  exposed  to temperatures above 50°C, to  electrical  abuse  or to  mechanical shock, particularly as  a  result  of  being dropped. Consequently, any expert opinion offered by a third party after the device has been removed also nullifies the guarantee. 5. The limited warranty will be forfeited if it is proven that the device has  been  misused  or  inadequately  implanted,  against  the physicians’manual recommendations of PARADYM RF. 6. The  limited  warranty  does  not  include  leads  and  other accessories used for the implantation. 7. The  replacement  terms  or  conditions  described  in  article  2 include  all  devices  that  shall  be  replaced  within  the  limited warranty period because of battery depletion, without any link to a component failure or a production hazard. The device battery longevity varies with the type and number of delivered therapies. 8. Legal  requirements  of  jurisdictions  where  the  PARADYM  RF device  is  distributed  will  supersede  any  warranty  conditions indicated in this manual that conflict with such laws.
82 – US-ENGLISH 22.2. ARTICLE 2 : TERMS OF REPLACEMENT 1. In case of PARADYM RF failure because of a component failure, a production defect, or a conception error, occurring within two-year  period  starting  from  the  implantation  date,  Sorin  CRM  is committed to:   replacing free of charge the explanted device  by a  Sorin CRM device with equivalent features,  or  issuing  a  replacement  credit  equal  to  the  purchase price  for  the  purchase  of  any  other  Sorin  CRM replacement device. 2. After a two-year period and up to 4 years after the implantation, Sorin  CRM,  because  of  limited  warranty  terms,  will  issue  a replacement credit to the buyer of an amount equivalent to half of  the  initial  purchase  price  minus  prorata  temporis  during  this two-years period. 3. In  any  case  the  credit  issued  by  the  limited  warranty  terms cannot exceed the purchase price of a Sorin CRM replacement device.
US-ENGLISH – 83 23. PATENTS The PARADYM RF model described in this manual is covered by the following US patents:  5 167 224, 5 226 415, 5 271 394, 5 312 451, 5 325 856, 5 339 820, 5 350 406, 5 411 533, 5 462 060, 5 513 645, 5 545 181, 5 558 097, 5 564 430, 5 591 218, 5 626 619, 5 645 574, 5 674 265, 5 697 960, 5 702 424, 5 702 426, 5 713 928, 5 741 315, 5 776 164, 5 776 165, 5 818 703, 5 836 980, 5 868 793, 5 891 170, 5 891 184, 5 899 931, 5 931 856, 5 935 153, 5 954 660, 5 978 708, 6 181 968, 6 230 058, 6 236 111, 6 251 703, 6 256 206, 6 307 261, 6 337 996, 6 397 105, 6 408 209, 6 487 451, 6 487 452, 6 505 068, 6 532 238, 6 556 866, 6 604 002, 6 622 039, 6 625 491, 6 711 441, 6 738 665, 6 830 548, 6 889 080, 6 898 845, 6 912 421, 6 937 898, 6 975 905, 7 065 402, 7 072 716, 7 076 297, 7 113 826, 7 142 924, 7 164 946, 7 251 526, 7 366 566, 7 400 921, 7 400 922, 7 953 483.
84 – US-ENGLISH 24. EXPLANATION OF SYMBOLS The symbols on product labelling have the following meaning:  Use by  Date of manufacture  Serial number  Batch number  For single use only  Sterilised using ethylene oxide  Temperature limitation  High voltage  Consult instruction for use.
US-ENGLISH – 85   FCC ID YSGDR9550 IC : 10270A-DR9550  Last revision date of this manual: 2012-05

Navigation menu