ST Microelectronics S R L MHBGW1 Bodyguardian Control Unit User Manual EW 40

ST Microelectronics S.R.L. Bodyguardian Control Unit EW 40

Contents

User Manual

     Bodyguardian Control Unit Base Kit      Operator Manual  SPMHBGW1-MAN Rev. A            EDITION: November 2011
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 2   Contents CHAPTER 1 ........................................................................................................................................ 4 INFORMATION ABOUT SAFETY ................................................................................................... 4 1.1 INFORMATION ABOUT THE MANUAL ......................................................................... 4 1.1.1 CONVENTIONS ........................................................................................................... 5 1.2 DECLARATION OF RESPONSIBILITY BY THE MANUFACTURER .......................... 6 1.3 USAGE RESTRICTIONS AND SAFETY PRECAUTIONS .............................................. 7 1.3.1 ELECTRIC SAFETY .................................................................................................... 7 1.3.2 SAFETY OF THE OPERATING ENVIRONMENT .................................................. 10 1.3.3 OTHER PRECAUTIONS ............................................................................................ 12 1.4 GRAPHIC SYMBOLS IN COMPLIANCE WITH THE IEC 60601-1      STANDARD . 13 1.5 OTHER GRAPHIC SYMBOLS ......................................................................................... 15 1.6 ATTENTION SYMBOL ..................................................................................................... 18 1.7 PRODUCT TRACEABILITY ............................................................................................ 18 1.8 VIGILANCE SYSTEM ...................................................................................................... 18 1.9 INFORMATION ABOUT RECYCLING OF MATERIALS ............................................ 22 1.10 ELECTROMAGNETIC COMPATIBILITY .................................................................. 23 1.10.1 RECOMMENDED DISTANCES FROM RADIOFREQUENCY (RF) COMMUNICATION SYSTEMS .............................................................................................. 27 1.11 BIOCOMPATIBILITY AND INFECTIONS CONTROL ............................................. 29 1.12 CAUTION FOR THE U.S. MARKET ............................................................................ 29 CHAPTER 2 ...................................................................................................................................... 30 DESCRIPTION OF THE DEVICE ................................................................................................... 30 2.1 GENERAL OVERVIEW .................................................................................................... 30 2.2 BODYGUARDIAN CONTROL UNIT DESCRIPTION ................................................... 34 2.2.1 PATIENT CONNECTION .......................................................................................... 35 2.2.2 BLUETOOTH CONNECTION ................................................................................... 36 2.2.3 SIGNALING LEDS ..................................................................................................... 36 2.2.4 MULTIFUNCTION PUSH BUTTON ........................................................................ 38 2.3 BODYGUARDIAN CHARGING CRADLE ..................................................................... 39 2.4 AC/DC MEDICAL POWER SYPPLY ............................................................................... 40 2.5 DISPOSABLE ADHESIVE ELECTRODES PATCH ....................................................... 40 CHAPTER 3 ...................................................................................................................................... 41 POWERING UP THE DEVICE ........................................................................................................ 41 3.1 BATTERY CHARGING .................................................................................................... 41 3.1.1 RECORDING AUTONOMY ...................................................................................... 42 3.2 SWITCHING ON/OFF THE DEVICE ............................................................................... 43 CHAPTER 4 ...................................................................................................................................... 44 WORKING MODE............................................................................................................................ 44 4.1 PREPARING THE PATIENT ............................................................................................ 44 4.1.1 DISPOSABLE ADHESIVE ELECTRODES PATCH APPLICATION SITE ........... 44 4.1.2 PREPARING THE SKIN ............................................................................................ 44 4.1.3 PLACING THE DISPOSABLE ADHESIVE ELECTRODES PATCH ..................... 45 4.2 OPERATIVE MODES ........................................................................................................ 46 4.2.1 MONITORING MODE ............................................................................................... 46 4.2.2 STREAMING MODE.................................................................................................. 47 4.2.3 EVENT MONITORING .............................................................................................. 47 4.3 INSTALLATION AND INSTRUCTIONS FOR THE PATIENT ..................................... 48
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  3  `   CHAPTER 5 ...................................................................................................................................... 49 MAINTENANCE .............................................................................................................................. 49 5.1 GENERAL INFORMATION ABOUT MAINTENANCE ................................................ 49 5.2 SAFETY CHECKS ............................................................................................................. 50 5.2.1 CONNECTORS ........................................................................................................... 50 5.2.2 BATTERY PACK........................................................................................................ 50 5.3 CLEANING THE DEVICE ................................................................................................ 51 5.4 PARTICULAR WARNINGS FOR CRITICAL COMPONENTS ..................................... 52 CHAPTER 6 ...................................................................................................................................... 53 TECHNICAL CHARACTERISTICS................................................................................................ 53 6.1 BODYGUARDIAN CONTROL UNIT .............................................................................. 53 6.2 DISPOSABLE ADHESIVE ELECTRODES PATCH ....................................................... 56 6.3 AC/DC MEDICAL POWER SUPPLY ............................................................................... 57 6.3.1 OPTION 1 .................................................................................................................... 57 6.4 BODYGUARDIAN CHARGING CRADLE ..................................................................... 57 CHAPTER 7 ...................................................................................................................................... 58 REQUEST FOR ASSISTANCE ........................................................................................................ 58 7.1 OBTAINING SERVICE ..................................................................................................... 58 7.2 PREVENTICE MAIN OFFICES ........................................................................................ 59 OPERATING OFFICES ............................................................................................................ 59
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 4   CHAPTER 1 INFORMATION ABOUT SAFETY 1.1  INFORMATION ABOUT THE MANUAL This document contains proprietary information. No part of this publication may be  photocopied  or  reproduced  without  the  prior  written  permission  of  the manufacturer PREVENTICE. Information in this document is subject to change and revision without notice.  Issues:  First edition: SPMHBGW1-MAN - Rev. A - November 2011   This manual is to be considered as a component of the equipment. When installing the equipment for the first time, the user should accurately check the content of the Manual in order to verify its integrity and completeness.  In  the  event  the  Operator  Manual  should  be  ruined,  incomplete  or  inadequate, please  contact  PREVENTICE  in  order  to  immediately  restore  or  replace  the uncompliant Manual.  The official version of the Operator Manual, of which PREVENTICE is directly responsible, is the English versions. For countries in which languages other than English  are  spoken,  the  official  Manual  is  the  one  in  the  English  version. PREVENTICE does not undertake any responsibility for any translations in other languages made by distributors or users.  The observance of the operating procedures and of the warnings described in this Manual is  a basic requirement for the  correct working of the equipment and to guarantee the patient’s and the user’s safety.  The Manual must be read in every part in front of the equipment before using it, in  order  to  become  familiar  with  the  operating  procedures,  the  commands,  the connections to the peripheral instruments, and the precautions for a correct and safe usage.  The Operator Manual should be kept, complete and readable in every part, in a safe place, and, at the same time, it should be rapidly accessible to the user when using the equipment.  This Operator Manual is intended for System Builder and not for the end-user of the device.  The equipment Service Manual is available on request. This Manual contains all information directed to the qualified staff in charge for servicing.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  5  `   1.1.1  CONVENTIONS  In this Operator Manual the following conventions are used:   NOTE  The NOTE messages contain important information, which must be noticeable with respect to the regular text. Usually they have useful information for the operator: detailed data on the correct operating procedures of the instrument.    WARNING  The WARNING messages show in the manual before operations  and procedures, which must be strictly observed in order to avoid possible loss of data or damage to the equipment.     ATTENTION  The ATTENTION messages show in the manual in correlation with the description of procedures and operations, which could cause injury to the operator or to the patient, if not correctly performed.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 6  1.2  DECLARATION OF RESPONSIBILITY BY THE MANUFACTURER  MANUFACTURER:    PREVENTICE 1652 Greenview Drive SW Rochester, MN 55092 Website www.preventice.com Tel: +1 866-830-4043    PREVENTICE  is  responsible  for  safety,  reliability  and  performances  of  the equipment  only  when  the  equipment  is  used  in  compliance  with  the  following conditions:  Calibrations, modifications or servicing must be performed by qualified staff expressly authorized by PREVENTICE.  The equipment must be opened and its internal parts must be accessed to by maintenance qualified staff only expressly authorized by PREVENTICE.  The environment where the equipment is used must be in compliance with the safety prescriptions.  The electric wiring of the building must be designed according to the standards and perfectly working.  Parts of the equipment that can be replaced by the user and accessories must be replaced with items of the same kind and with the same characteristics.  The connection of the equipment with peripherals or other instruments supplied by  the  mains  must  be  performed  according  to  the  IEC  60601-1-1  standards (standards  for  electrical  safety  of  medical  electric  systems)  and  to  the  IEC 60601-1-2 standards (standards for electromagnetic compatibility).  Usage  and  maintenance  of  the  equipment  and  of  its  accessories  must  be performed in compliance with the instructions described in this Manual.  This Manual must be kept complete and readable in every part.  The equipment is used and serviced until its “End of Life”.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  7  `   1.3  USAGE RESTRICTIONS AND SAFETY PRECAUTIONS  In  order  to  guarantee  the  patient’s  and  the  user’s  safety  as  well  as  a  correct working  of  the  equipment,  it  is  necessary  to  operate  within  the  consented restrictions and adopt all the precautions listed below:   ATTENTION  Prior to usage, verify that all the safety requirements are satisfied. The  equipment must  not be supplied  by or  connected  to  other instruments until such safety conditions are restored.  1.3.1  ELECTRIC SAFETY  Leakage current The maximum patient leakage current from the equipment, measured according to the IEC 60601-1 standard (for Type BF) is less than 100  A. However  take  care  when  using  the  equipment  at  the  same  time  with  other instruments.  In  the  event  the  patient  is  connected  to  several instruments  at  the same  time,  it  is  necessary  to  remember  that  the  sum  of  the  leakage  currents determined by each instrument may exceed this value.  Patient Connection All patient connections to the equipment are through the device using the proper adhesive electrode patch provided. Any patient electrodes connected to the device by any other means may constitute an unsafe condition that could result in injury or death to the patient.   ATTENTION  All patient connections on the device are isolated from AC power ground. Do NOT join these connections to earth ground or AC power ground since such an action  constitute  an  unsafe  condition  that  could  result  in  serious  injury  or accidental death to the patient.    ATTENTION  The electrode through which the signal is captured from the body of the patient are not  part  of  the  amplifier  system,  in  any  case  it  is  MANDATORY  to  use  only electrode or sensor approved for commercial use by FDA (USA) or/and CE marked (93/42 EEC directive and following amendment 2007/47/EC).
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 8  To ensure the safety of the patient and the operator, please follow all the warnings and caution listed in this manual.   Connect to the equipment the proper specified power supply only. In order to guarantee the electrical safety requirements, the recharge of internal battery pack of equipment, must be performed by means the proper medical AC/DC adapter  only.  The  power  supply  is  supplied  by  PREVENTICE  with  the equipment. The allowed models are:   Trademark FRIWO, Model FW7662M/06  To recharge the  equipment place it  on the Charging Cradle according  to the appropriate orientation (only one is possible due to the mechanical constraints) and only after this operation connect the Charging Cradle to the AC/DC Power supply and this last to the mains.  After  the  recharging  operation  is  complete  disconnect  the  AC/DC  Power supply from the mains and only after this operation remove the equipment from the Charging Cradle.  Take  care  when  using  the  equipment  at  the  same  time  with  other instruments. In the event the patient is connected to several instruments at the same time,  it is  necessary to  remember that  the  sum  of the leakage currents determined by each instrument may endanger his life.  Take care when using the equipment at the same time with other radio-frequency instruments. In the event the equipment is used in a surgery room at the same time with a radio knife (Radio-Frequency instrument = RF), it is necessary to hold the radio knife point as far as possible from the electrodes, in order to reduce as much as possible the risk of RF currents making on  such electrodes and the consequent burns. Therefore it is necessary to use electrodes with  a  larger  surface  contacting  the  patient  body,  in  order  to  limit  the  RF current density to acceptable values. In case it is not possible to use the proper electrodes,  it  is  recommended  to  disconnect  the  patient  from  the  equipment before using radio-frequency instruments.  The equipment is not protected against the defibrillator discharges. Please remember  that  the  equipment  is  not  protected  against  the  defibrillator discharges;  for  this  reason,  in  the  event  it  should  be  necessary  to  use  the defibrillator,  it  is  necessary  to  disconnect  the  patient  from  the  equipment  in order to avoid the patient being burned in the electrode contact areas and the equipment undergoing sever and irreversible damages.   Avoid contact of patient and electrodes with conductive metal items. When the  equipment  is  connected  to  other  instruments  supplied  by  the  mains,  the whole  input  circuit  to  which  the  patient  is  connected  is  electrically  isolated (floating isolation). It is necessary to avoid the patient and any conductive part connected to the patient (electrodes, connectors, and transducers) coming into contact with conductive parts (ground included). Please observe this precaution to avoid compromising the equipment isolation level. This precaution must be observed in order to avoid that accessible metal parts of the device get in touch
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  9  `   with  external  conductive  parts  thus  damaging  the  isolation  level  of  the equipment.   Do  not  connect  additional  Multiple  Portable  Socket-Outlet  or  extension cord. Multiple portable socket-outlet or extension cord shall not be connected to the system.   Observe the  IEC 60601-1-1 and  the IEC 60601-1-2  standards  in  case of connection  with  other  instruments.  The  connection  of  the  equipment  with other devices is allowed only when the safety requirements for the patient, the user and the environment are not compromised. If the Manual does not contain enough information about the possibility of interconnection with other devices, the  user  should  contact  the  manufacturer  or  the  nearest  authorized  servicing center to have information about the effects that coupling devices may have on the patient, the user and the environment.  Replace  damaged  parts  immediately.  Cables,  connectors,  accessories,  or other parts of the equipment must be replaced immediately when damaged or not working correctly. In these cases, contact the nearest authorized servicing center.  Do not connect items (accessories and peripherals) which are not specified as part of the system expressley indicated by PREVENTICE. In order to guarantee all the safety requirements, it is necessary to use only the accessories and peripherals specified in this Manual as part of the system, which have been tested  with  the  equipment.  The  usage  of  accessories  and  consumer  goods supplied  by  other  manufacturers  or  not  specifically  indicated  by PREVENTICE does not  guarantee the safety and  the  correct  working of the equipment. Use only peripherals in compliance with the standards of the class they belong to.  Check the functionality of the system before starting any recording. It is strongly recommended to check the overall functionality of the system before starting  any  recording.  In  case  any  anomalies  or  malfunctioning  should  be noticed,  immediately  disconnect  the  patient  from  the  system  (if  a  patient  is already  connected),  switch  off  the  system  and  ask  for  service  to  qualified personnel.  In  particular  (for  example)  if,  with  a  patient  connected  to  the system, some  anomalous  tracing, like  isoelectric or  greatly artefacted signal, should be noticed on the monitor during recording: in this case if the problem should  not  be  solved  with  the  assembly  standard  technique  (poor  electrode connection, broken lead etc) immediately acts as above, disconnect the patient, do not use the system and ask for servicing.  Periodically check that all the system works regularly during  “long term recording”. During “long term recording” (more than one hour) it is strongly recommended to periodically check that all the system works regularly without any sign of malfunctioning. If any anomalies or flat traces should be noted act as in the previous warning. In particular any electrode site used for long term must be checked for irritation and redness. Check each electrode periodically to
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 10  evaluate  the  skin  condition  under  the  electrode.  Redness,  blistering  and permanent skin scarring can occur if electrodes are not regularly monitored.  Using the equipment on patients with a heart pace-maker is not allowed. It is  not  allowed  using  the  equipment  in  the  case  of  patients  with  implanted electric  devices  (or  bed  partners  with  implantable  devices),  especially  heart pace-makers,  because  the  equipment  may  cause  the  cardiac  stimulator malfunctions.  Patients  with  cardiac  pacemakers  should  not  undergo  any examination  with  this  equipment without  authorization  and  under  the  severe control of a specialized physician. For the same reason it is also necessary to be careful when  using the  equipment  in  proximity of operators or  persons  with implanted electric devices  Rechargeable Battery  Pack. The  rechargeable  battery  pack  installed  in  the equipment  contains  one  cell  that  is  not  accessible  to  the  user  and  its replacement should be  performed only by qualified personnel, authorized by PREVENTICE. Anyway, you should consider the following general warnings (e.g. in case of disposal of replaced parts with technical assistance).  Do not let the ends of the battery pack to come in contact with metal objects.  Keep away the battery pack from heat or flames.  Do  not  immerse  the  battery  pack  and  avoid  exposing  it  to  rain  or moisture.  Avoid direct mechanical trauma to the battery pack.  Do not attempt to disassemble, puncture, incinerate or short-circuit the battery.  These  operations  may  cause  a  fire  or  the  emission  of  toxic chemicals.   Charge  the  battery  pack.  In  order  to  ensure  the  safety  requirements,  the battery pack must be recharged only by using the proper Charging Cradle and its medical AC/DC adapter, specifically provided by PREVENTICE together with the equipment. The model of adapter to be used is:     Trademark FRIWO, Model FW7662M/06 1.3.2  SAFETY OF THE OPERATING ENVIRONMENT  The equipment is not designed to be used in locations with inflammable vapors or gases that may cause explosions. The equipment must not be used in  atmospheres  with  a  high  concentration  of  oxygen  or  in  buildings  where inflammable  substances  or  anesthetic  agents  are  present.  The  atmosphere  is considered  as  oxygen-saturated  when  the  oxygen  or  nitrous  oxide  (NO2) concentration contained in the environment is over 24%.  The  equipment  is  not  designed  to  be  used  in  MRI  area.  The  equipment should be removed before Magnetic Resonance Imaging (MRI).  The equipment and its internal parts are protected against the inflow of liquids according to IPX4 degree of protection. The equipment is protected
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  11  `   against  the  dripping,  spraying  and  splashing  of  water  and  relevant  harmful effects. Avoid  submitting  the  equipment  to  the  risk  of  water  jetting  or temporary  and  continue  immersion  because  its  protection  degree  do  not guarantees protection of internal parts against ingress of liquids. Do not use the equipment  where  such  risks  are  present.  Devices  in  which  liquids  have accidentally  penetrated  must  be  immediately  cleaned  and  checked  by authorized qualified staff.  Use of  the  equipment in  humid Environment  is  allowed  if  conditions  are compliant to the environmental limits defined in the following bullet, which are in  accordance with  the  requirements of  the applicable  general  and  particular IEC 60601 standards.  Use  the  equipment  within  the  environmental  limits  of  specified temperature, humidity and pressure. The equipment is designed to work in environmental conditions that, in compliance with the IEC 60601-1 directions and the 60601-2-47, are defined as standard: - temperature        +10°C / +45°C - relative humidity      10% / 95% RH - atmospheric pressure   700 / 1060 hPa  The equipment could heat up during its normal use or battery recharge. This aspect should be considered as a normal characteristic of the equipment due to the  high  integration  of  the  electronic  circuitry  inside.  Never  the  equipment heating  up  should  be  considered  as  a  potential  fault  or  as  a  defect  of  the equipment itself.  Make sure the electric wiring of the building is efficient before connecting the  power  supply  to  the  mains.  When  the  equipment  (power  supply  of Charging  Cradle)  is  connected  the  environmental  mains,  make  sure  that  the building wiring is correctly functioning and efficient and compliant to the local regulations and standards.  Be careful using the equipment in locations disturbed by strong magnetic fields.  The  equipment  is  compliant  with  the  EMC  requirements (Electromagnetic  Compatibility)  according  to  what  specified  by  the  IEC 60601-1-2  standard  and  93/42/EEC  European  Directive.  In  every  case  it  is recommended  to  keep  the  equipment  away  from  disturbance  sources  and induced electromagnetic fields surpassing the values prescribed by the standard in  order  to  avoid  any  possible  instabilities  and  malfunctioning  of  the equipment. For more detail about device classification and minimum distances, please refer to  paragraph  1.11 “Electromagnetic  compatibility” of the  present manual.  Be careful using the equipment near short-wave or micro-wave devices. If the equipment is used in an area where there are also short-waves or micro-wave devices, it is necessary to remember that these may cause instability and interfere with the correct working of the equipment. Do not use the equipment near X-ray or diathermy devices.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 12  1.3.3  OTHER PRECAUTIONS  Take care when using the equipment on patients who are pregnant. It is necessary to be careful when using the equipment in the case of patients who are  pregnant.  These  patients  should  not  undergo  any  examination  with  this equipment without authorization and under the severe control of a specialized physician.  The  equipment  is  intended for  adult  use  only  with  a  weight  greater  than  10 Kgs.  Take  care  when  using  the  equipment  on  patients  with  potentially  life-threatening  arrhythmias  requiring  hospitalization.  These  patients  should not  undergo  any examination  with  this  equipment  without  authorization  and under the severe control of a specialized physician.  Take care when using the equipment on patients with known skin allergies or sensitivities to acrylic, hydrogel or silicone adhesives.  Take care  when  using  the equipment  on patients with  friable skin.  It  is necessary to be careful when using the equipment in the case of patients with sensitive skin or skin disease, because the adhesive electrode may cause skin irritation. Patients with sensitive skin should not undergo any examination with this  equipment  without  authorization  and  under  the  severe  control  of  a specialized physician.  Do  not  apply  creams  or  lotion  to  the  skin  prior  to  use  equipment.  The application of creams or lotion could cause a bad contacts of the electrodes and a bas adhesion of the patch and consequently a bed signal acquisition  Bodyguardian is not waterproof. The device should be removed by patient before bathing, showering or swimming.  Bodyguardian has a usage limited to one patient at a time.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  13  `   1.4  GRAPHIC  SYMBOLS  IN  COMPLIANCE  WITH  THE  IEC  60601-1      STANDARD  The following table shows description and localization of all graphic symbols in compliance  with  the  IEC  60601-1  safety  standards  present  on  the  equipment panels and/or on any other instruments or external devices to which the equipment may be used in cojunction to or present in the same environment.  IEC 60601-1 SYMBOL DESCRIPTION POSITION  Alternating current Symbol placed on the connection points between the equipment and the mains (alternating current source).   Direct current Symbol placed on the connection points to direct current source.    Equipotential terminal Symbol placed on the outlet connecting the equipment to the equipotential node of the building, if any.   Protective earth (ground) Symbol placed on the connection points between the equipment and the protective grounding.   High voltage Symbol placed on circuits or equipment parts with high voltage.    Attention! Refer to the attached instructions. Symbol placed on items for which it is important to read the Operator Manual for relevant information (see ATTENTION paragraph).   Device with CF-type applied parts   Symbol placed on applied parts  to the patient with a CF-protection level.   Device with BF-type applied parts       Symbol placed on applied parts  to the patient with a BF-protection level.   Device with B-type applied parts     Symbol placed on applied parts to the patient with a B-protection level.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 14   IEC 60601-1 SYMBOL DESCRIPTION POSITION     Off (disconnected from the mains) Symbol placed on the off/on positions of the whole equipment general power switch.     On (connected to the mains) Symbol placed on the off/on positions of the whole equipment general power switch.  Off (for a single part of equipment)   Symbol placed on the off/on switch of a single part of the equipment.  On (for a single part of the equipment)   Symbol placed on the off/on switch of a single part of the equipment.  Device with Class II protection type against electric shock  Symbol placed in the identification label of the equipment.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  15  `   1.5  OTHER GRAPHIC SYMBOLS  The following table shows description and localization of all symbols placed on the equipment panels and/or on any other instruments or external devices to which the equipment may be used in cojunction to or present in the same environment.  SYMBOL  DESCRIPTION POSITION    Input Symbol placed on the signal input or mains voltage input connectors of the equipment.   Output Symbol placed on the signal output or the mains voltage output connectors of the equipment.  Rx Only Prescription Only Symbol placed on the identification label of the medical device indicating that Federal (USA) law restricts this product to sale by or on the order of a physician. IPX4 Degree of protection against ingress of water (spashing) Symbol placed on the identification label of the medical device indicating that the device is protected against the splahing of water.    Functional Mode and Communications status Symbol placed close the central led of device to indicate the functioning and communication status of the device    Radio Frequency emitting device (non-ionizing electromagnetic radiation) Symbol placed on the identification label of the medical device to indicate that the device emittes Radio-Frequency for its normal functioning.    Lot number Symbol placed on the identification label of the medical device together with the device lot number.  Battery charge status Symbol placed close the led of device to indicate the battery charge status and re-charge condition.  REF Reference number Symbol placed on the identification label of the medical device together with the device reference number.  SN Serial number Symbol placed on the identification label of the medical device together with the device serial number.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 16     Date of manufacture Symbol placed on the identification label of the medical device together with the device manufacture date.   Manufacturer Symbol placed on the identification label of the medical device together with the name and address of the device Manufacturer  Authorized Representative in the European Community  Symbol placed on the identification label of the medical device together with the name and address of the device Authorized Representative in the European Community.   Crossed-out wheeled bin   Symbol placed on the identification label of the medical device. This symbol indicates the prohibition of throw the medical device in the household wheeled bin device when at its “end of life”.  Recyclable Symbol placed on Battery Pack. The symbol indicates the components of the object are recyclable at the end of life.   Use by Symbol placed on the identification label of the medical device together with the device expiration date.     Do not reuse Symbol placed on the identification label of the medical device. This symbol indicates that the device is a disposable one and cannot be used more than once.    Sterile Symbol placed on the identification label of the medical device indicating a sterile device.   Sterilization with steam or dry heat Symbol placed on the identification label of the medical device indicating a sterile device and the sterilization method used (steam or dry heat).      Sterilization with ethylene oxide Symbol placed on the identification label of the medical device indicating a sterile device and the sterilization method used (ethylene oxide).      Sterilization by irradiation  Symbol placed on the identification label of the medical device indicating a sterile device and the sterilization method used (irradiation).     Refer to the instructions for Use/functioning. Symbol placed on the identification label of the medical device recommending to refer to the instruction for use/functioning for more information about the usage of the device.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  17  `    Temperature Limit   Symbol placed on the identification label of the medical device together with the indication of temperature limits (high and low limits) for the usage/storage of device.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 18  1.6  ATTENTION SYMBOL  The ATTENTION symbol shown below, placed on the equipment casing, refers the user to the Operator Manual for information, warnings and suggestions which are particularly important for a correct and safe use of the equipment.    In  particular,  when  it  is  placed  on  connecting  points  or  commands  or  led indicators, this symbol refers the user to carefully read the Operator Manual for instructions concerning the nature and safety of such connection and/or detailed description of the commands and meaning of indicated events/situations for the operator.  For location of the ATTENTION symbols placed on the equipment, please refer to  chapter  2  “Description of the Device”  of  this  Operator  Manual.  This  chapter shows  the  pictures  of  the  equipment  panels  with  the  corresponding commands, connections,  symbols,  and  labels.  Each  attention  symbol  comes  with  a  detailed explanation of its meaning. 1.7  PRODUCT TRACEABILITY  In order to guarantee the traceability of the product, according to what stated in the ISO 13485 quality standard, QSR 21 CFR Parte 820 FDA, and the 93/42/EEC European  Directive  on  Medical  Devices  (and  its  revised  version  2007/47/EC directive), PREVENTICE kindly requests the original owner of the equipment to give communication to our main offices of any conveyance of the product to third parts,  by  sending  a  photocopy  of  the  proper  duly  filled-in  Product  traceability form (see enclosure 1.7), or by communicating in writing the data indicated in the form. The data concerning the device can be found on its identification label.  The  form  shall  be  sent  either  directly  or  through  any  subsidiary  or  the  nearest authorized distributor to the any PREVENTICE operating office. The list of the main PREVENTICE head and branch offices is contained in chapter “Request for assistance” of this manual. 1.8  VIGILANCE SYSTEM  The  device  is  subject  to  a  vigilance  system  (post-marketing  vigilance)  that PREVENTICE and their distributors and retailers apply to the products they put on  the  market  to  safeguard  the  patient  and  the  physician  from  serious  or
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  19  `   potentially serious hazards during the normal use of the equipment, in order to be able to remove the source of such hazards with the best efficiency and timing.  To the purpose of helping PREVENTICE take any timely and effective corrective measure, it is extremely important that the user performs a careful inspection of the  equipment  performances  in  order  to  identify  or  foresee  any  dangerous situation for the patient’s and the user’s health.  For this reason, the user shall give immediate communication of any malfunction or deterioration of the characteristics or the performances of the equipment or any mistake found in these instructions that caused or could cause serious damages to the patient’s and the user’s health. In  this  case,  the  user  may  send  a  photocopy  of  the  proper  duly  filled-in  Post-Marketing  Vigilance  Form (see  enclosure 1.8),  or  communicate  in  writing  the data indicated in the form. The instrument’s data can be collected from it’s identification label.  The  form  shall  be  sent  either  directly  or  through  any  subsidiary  or  the  nearest authorized distributor to the any PREVENTICE operating office. The list of the main PREVENTICE head and branch offices is contained in chapter “Request for assistance” of this manual.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 20   Enclosure 1.7    PRODUCT TRACEABILITY FORM    To:  PREVENTICE    1652 Greenview Drive SW   Rochester, MN 55092     c.a. Quality Assurance Department     System/device name……….................................................................................................  Device code / reference number (REF)  ............................................................................  Device serial (SN) / lot number (LOT) .................................................................................  Name and address of the former owner  .................................................................…….... ............................................................................................................................................. ............................................................................................................................................. .............................................................................................................................................  Name and address of the present owner .................................................................…....... ....................................................................................................................................…..... ................................................................................................................…......................... ...............................................................................................................…..........…............   Date:...........................   Signature    ...............................................  (please name in full)
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  21  `     Enclosure 1.8    POST-MARKETING VIGILANCE FORM    To:  PREVENTICE  1652 Greenview Drive SW Rochester, MN 55092   c.a. Quality Assurance Department    System/device name.......................................................................................………..........  Device code/reference number (REF) ................................................................................  Device serial (SN)/lot number(LOT) .....................................…....................................…….  Description of the real or potential hazard…………………………………………………….. ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. .............................................................................................................................................  User’s comments/suggestions ............................................................................................ ............................................................................................................................................. ............................................................................................................................................. ............................................................................................................................................. .............................................................................................................................................  User’s address.................................................................................................................... Phone......................................................... Fax .............................................…................  Department where the device is installed............................................................................. Person in charge of the department.....................................................................................   Data:...........................     Signature  ...............................................  (please name in full)
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 22   1.9  INFORMATION ABOUT RECYCLING OF MATERIALS  In  accordance  with  the  specific  worldwide  regulations,  PREVENTICE  aims  to continuously improve the design and the fitting of electromedical devices in order to reduce as much as possible any negative impact on the environment caused by the  management  of  component  parts,  consumer  materials,  packaging  and discharge of devices when at their “end of life”.  Packaging materials were conceived and produced so as to allow the re-usage and the  salvage,  including  recycling,  of  most  part  of  materials  and  to  reduce  the quantity  of  garbage  or  residual  products  for  discharge  as  much  as  possible.  In particular, packaging materials have been produced so as to limit the presence of harmful  metals  and  of  other  dangerous  substances  to  minimum  quantities  in emissions, ashes or lixiviation residual products. The total concentration levels of heavy metals such as Lead, Cadmium, Mercury and hexavalent Chrome contained in  the  packaging  materials are  in  accordance  with  the  limits  established  by the directives in force related to this subject.  In order to cause minimum consequences to the environment, the design of the device includes the highest possible miniaturization of the circuits, with the least possible  differentiation  of  materials  and  components,  with  a  selection  of substances  that  guarantee  the  highest  possibility  to  recycle  and  re-use  the components and to discharge them without risks for the environment.  The device is designed to guarantee the easy separation or  disassembling of the materials containing polluting substances from the others, in particular during the operations  of  servicing  and  replacing  parts.  In  particular,  the  largest  plastic components  are  marked  according  to  their  plastic  contents  in  order  to  make  it easier to recycle the product.     ATTENTION  Please  refer  to  local  codes  and  regulations  for  proper  disposal/recycle requirements of packaging and consumer materials and of the device when at its “end of life”.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  23  `   1.10  ELECTROMAGNETIC COMPATIBILITY This  medical  device  is  designed  for  use  in  the  electromagnetic  environments declared in the tables below, in compliance with the IEC 60601-1-2:2001 (second edition)  standard.  The  operator  must  assure  that  the  device  is  used  in  an environment compliant to this standard.  This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. The device is labeled FCC-ID S9NMHBGW1.   Table 1 - Electromagnetic Emissions  Emission Test  Compliance Electromagnetic Environment Radiated and conducted RF emissions  CISPR 11 Group 1 This medical device use RF energy only for its internal function. Therefore, the RF emission is very low and not likely to cause any interference in nearby electronic equipment.  Radiated and conducted RF emissions  CISPR 11 Class B This medical device is suitable for use in all establishments, including domestic directly connected to the public low voltage power supply network that supplies buildings used for domestic purpose.  Harmonic emissions  IEC 61000-3-2 Complies  Class A Voltage fluctuations/ flicker emissions  IEC 61000-3-3 Complies
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 24  Table 2 - Electromagnetic Immunity  Immunity Test IEC 60601-1-2 Test Level Compliance Electromagnetic environment Electrostatic Discharge (ESD)  IEC 61000-4-2 6 kV in contact  8 kV on air IEC 60601-1-2 Test Levels  Residential/Hospital (Note 1)  Electrical fast transient/burst  IEC 61000-4-4 2 kV for power supply line  1 kV for input/output lines >3m  IEC 60601-1-2 Test Levels Residential/Hospital (Note 2) (Note 3)   Surge  IEC 61000-4-5 1/0.5 kv differential mode  2/1/0.5 kV common mode  IEC 60601-1-2 Test Levels  Residential/Hospital (Note 2) (Note 3)   Voltage dips, short interruptions and voltage variations on power supply input lines  IEC 61000-4-11 0 % of rated voltage (voltage dip 100 %) for 0.5 cycles  40 % of rated voltage (voltage dip 60 %) for 5 cycles  70 % of rated voltage (voltage dip 30 %) for 25 cycles  0 % of rated voltage (voltage dip 100 %) for 5 seconds
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  25  `   Power frequency (50/60 Hz) magnetic field  IEC 61000-4-8  3 A/m
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 26  Table 3 - Electromagnetic Immunity for non-life supporting equipment  Immunity Test IEC 60601-1-2 Test Level Compliance Electromagnetic environment    Portable and mobile RF communications equipment should be used no closer to any part of this medical device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter (table 4 - Recommended separation distance).  Radiated RF fields  IEC 61000-4-3  3 V/m from 80 MHz to 2.5 GHz IEC 60601-1-2 Test Levels  3 V/m  d = 1.2 P   80 MHz a 800 MHz  d = 2.3 P   800 MHz a 2.5 GHz Radiated RF fields  IEC 61000-4-6  3 V from 150 kHz to 80 MHz IEC 60601-1-2 Test Levels  3 V  d = 1.2 P    Where P is the maximum output rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).  Field strengths for fixed RF transmitter, as determined by an electromagnetic site survey, should be less then the compliance level in each frequency range.  Interference may occur in the vicinity of equipment marked with the following symbol:
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  27  `   Measures to be taken Note 1: The floor should be in antistatic material (wood, ceramic, ect.). If covered by synthetic material, relative humidity should be maintained at least at 30%.  Note 2: The quality of the electrical power supply and the mains frequency magnetic fields should be typical of domestic, commercial and hospital environments.  Note 3: If the operator has to work without a break while power supply is interrupted, it is necessary to have power supplied through a UPS (Uninterruptible Power Supply) unit.   1.10.1  RECOMMENDED DISTANCES FROM RADIOFREQUENCY (RF) COMMUNICATION SYSTEMS  As stated in the chapter 1 “Information about safety” of this operator manual, it is recommended  to  not  use  Radiofrequency  (RF)  transmission  system  near  the medical device. RF systems  can cause  interference which may cause instability and interferences with the correct working of the equipment and it may alters the signal acquired tracing.  The  operator  can  prevent  interference  caused  by  electromagnetic  field  by maintaining  a  minimum  distance  between  the  medical  device  and  the  RF communication system being used (cell phones, mobile phones, etc,). The  following  table  shows  the  minimum  distances  in  meters,  according  to  the maximum power at RF system output.  Table 4 – Recommended separation distance for non-life supporting equipment  Rated maximum output power of the transmitter (W) Separation distance according to frequency of transmitter (m) 150KHz to 80MHz d=1.2× P 80Mhz to 800MHz d=1.2× P 800MHz a 2.5GHz d=1.2× P 0.01 0.12 0.12 0.23 0.1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23 For transmitters rated at the maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated the equation applicable to the frequency of
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 28  the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.  Note: (1)  at 80MHz and 800MHz, the separation distance for the higher frequency range applies. (2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people    The  operator  must  remember  that  the  intensity  of  the  electromagnetic  fields generated by fixed transmitters (radio-base stations for cellular or cordless phone, TV  and  radio  transmissions,  amateur  radio  transmission,  etc.)  cannot  be predicated on theoretical basis. Consequently,  a  direct  measure  may  be  necessary  in  the  environment  where  is used the medical device.  If  the  intensity  of  the  electromagnetic  fields  exceeds  that  specified  in  the immunity levels  shown  in  the  previous  tables,  and  the  medical  device  behaves incorrectly  working,  additional  measures  may  be  necessary.  I.e.  orienting  or locating the medical device in a different way.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  29  `   1.11  BIOCOMPATIBILITY AND INFECTIONS CONTROL  No  system  component  is  intended  to  be  in  contact  with  the  patient  during  the usage. Electrodes and sensors are not intended to be parts of the Bodyguardian Control Unit product. All the body contacting material are not part of the MD, in any case remember that electrodes and sensors MUST meet the requirements of 93/42/EEC Medical Devices  Directive  and  its  revised  version  (CE  marked)  for  the  European Community and MUST have FDA clearance/approval for the U.S. market.  1.12  CAUTION FOR THE U.S. MARKET  Federal law  (USA)  restricts  the  device to  sale by or on the order  of  a licensed practitioner or therapist.  Medical professionals prescribe the Bodyguardian system to obtain physiological data  from  their  patients.  It  does  not  replace  direct  communication  between physicians and patients and does not summon physicians or emergency personnel.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 30  CHAPTER 2 DESCRIPTION OF THE DEVICE 2.1  GENERAL OVERVIEW The Bodyguardian Control Unit is a wearable battery power device intended for use as a part of a Multi-Parameter Analysis System and is designed to be used by a System Builder in a Multi-Parameter Analysis Application. The Bodyguardian Control Unit is worn on the chest for the acquisition, recording and  transmission  via  a  Bluetooth  radio  link  of  physiological  parameters  to external  host  devices  which  can  analyze  by  the  suitable  proper  Application Software and/or forward the data to additional storage elements or system.  The  Bodyguardian  Control  Unit  is  also  capable  to  record  symptomatic  and asymptomatic  events.  The  Bodyguardian  Control  Unit  continuously  records, stores and periodically transmits the following physiological data:   Single lead ECG  Heart rate  Respiratory rate  Posture  Activity  Event marker  The Bodyguardian Control Unit is used for ambulatory monitoring of non-lethal cardiac arrhythmias. Bodyguardian Control Unit is a wearable electronic device that  is  worn  on  the  chest.  The  device  includes  a  disposable  fabric  adhesive component  that  attaches  to  the  subject  and  connects  to  the  enclosed  electronic components.  The device will be worn intermittently or continuously for up to a 30-day period. The recordings will be stored on the Bodyguardian Control Unit and transmitted by  the  device  via  Bluetooth  communication  to  an  external  Associated  Device (Android smartphone or other). The device is prescription only.  The Bodyguardian Control Unit recorder is intended to be used in a:   “Home Care” environment without clinicians surveillance.  “Clinic Care” environment with clinicians surveillance  The device is intended to be used by two different categories of end-users: 1. The patient which will use the device for continuous monitoring applications and who will have to:  apply the device to the body by connecting the Bodyguardian Control Unit to the disposable patch and attaching it to the selected body area;
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  31  `    turn it on and off by using the appropriate button;  manage  the  recharging  of  the  Bodyguardian  Control  Unit  using  the dedicated charging station and power supply cable;  monitor the status of the device by looking at the led lights and colors  manage the devices by using the mobile phone application 2. The  clinicians  who  will  eventually  subscribe  the  use  of  the  device  to  the patient and who will be in charge of:  Setting  the  initial  thresholds  and  baselines  for  the  parameters  based  on patient’s conditions to personalize it (on Multi-Parameter Analysis System side)  Reviewing the data acquired and transmitted from the device to the remote server  location  (Multi-Parameter  Analysis  System)  for  diagnosis,  trends monitoring and in general clinical evaluations.  Depending  on  the  Environment,  optional  algorithms  and/or  thresholds  (settable via host software) can be used to generate measurements on an advisory basis for patients. These are presented for:   Review and interpretation by the clinician, based upon knowledge of the patient.  Results of the physical examination. The data provided by the Bodyguardian Control Unit are exclusively intended to be used by trained medical personnel to assist patients that requires monitoring of physiological parameters with reference of the followings context:   Screening  of  patients  with  symptoms  suggesting  arrhythmia  over  a minimum 24-hour period.  Screening and off line evaluation of HR and Breath variability if the nature of this variability cannot determine a Life-threatening for the patient.  Not intended to allow direct diagnosis or monitoring of vital physiological processes parameters (for instance cardiachearth rate, respiration), where the nature of variations is such that it could result in immediate danger to the patient.  Not intended for therapeutic purpose.  The  “Body  Guardian”  Recorder  is  intended  for  adult  use  only  with  a  weight greater  than  10  Kgs,  without  race  exclusion.  Patient  with  implanted  device  are excluded to wear this device. The Bodyguardian Control Unit is intended for System builder user in order to create a complete remote ambulatory monitoring system including Bodyguardian Control Unit kit, communication hub (Android smartphone or other) and System Server.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 32  All  aspects  related  to  safety  and  effectiveness  of  the  whole  ambulatory  multi parametric  monitoring  system  is  under  the  complete  responsibility  of  System Builder.  The  Bodyguardian  Control  Unit  is  provided  with  a  Base  Kit  containing  the following base elements:  The Bodyguardian Control Units (2 units)   The Bodyguardian Charging Cradle  The AC/DC medical power supply  The Disposable adhesive electrodes patch  The User Manual  The Android Smartphone The  Bodyguardian  Disposable  adhesive  electrodes  patch,  the  AC/DC  medical power supply, the User Manual and eventually the Smartphone  are provided by the System Builder. Here below  the  pictures  of  the  different  elements composing the Bodyguardian Control Unit base kit:
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  33  `    Figure 2-1 Components of Bodyguardian Control Unit base kit
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 34  2.2  BODYGUARDIAN CONTROL UNIT DESCRIPTION The following figures (2-2 and 2-3) show the Bodyguardian Control Unit:  Figure 2.2 – Bodyguardian Control Unit: front view               Figure 2.3 - Bodyguardian Control Unit: back view  In the previous figures the following components can be found: 1. Signaling Leds 2. Push button  3. Charger connectors (n° 2 contacts) 4. Patient  connectors  (n°  4  snap  contacts  to  connect  to  Disposable  adhesive electrodes patch) with the following configuration reading in clockwise sense from the Top-Left angle: INJ2, INJ1, SENSE1, SENSE2. The connection of the  device  to  the  disposable  adhesive  patch  is  possible,  due  to  mechanical constraints, only in one correct way. 1  3  4  2  INJ2 INJ1 SENSE1 SENSE2
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  35  `   2.2.1  PATIENT CONNECTION Biologic  input  signals  are  acquired  by  Bodyguardian  Control  Unit  through  the patient connectors (fig. 2.4).    Figure 2.4 - Patient connectors  which uses the 4 ECG snaps to interconnect the Bodyguardian Control Unit with the Disposable electrodes patch. The connection is guaranteed through the four snaps on the SnapStrip.   Figure 2.5 - Patient connection    ATTENTION  All  the  patient  applied  parts  and  corresponding  input  sockets  of  Bodyguardian Control  Unit  (patient  inputs)  are  electrically  isolated  according  to  IEC  60601-1 standard  requirements  for  internally  powered,  Type  BF  equipments.  This characteristic  is  indicated  to  the  operator  with  the  proper  symbol  placed  on  the external cover of the device.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 36  2.2.2  BLUETOOTH CONNECTION The Bodyguardian Control Unit is equipped with an integrated Bluetooth network interface  (for  Protocol  Specification  please  refer  to  the  document BGP110006_Commend_Interface_Specifciations)  that  allows  to  establish  a pairing connection with an external Associated Device (e.g. Android smartphone or other). The Bluetooth interface can be used to configure the Bodyguardian Control Unit and to transmit the data registered and stored on the Bodyguardian Control Unit to the Associated Device. The  communication  between  Bodyguardian  Control  Unit  and  the  external Associated Device is handled via software. 2.2.3  SIGNALING LEDS The following figure shows the signalling Leds placed on the front cover of the Bodyguardian Control Unit.   Figure 2-9 – Signaling Leds  The 3 Monocolor Leds indicate device status and conditions. The Leds indicates the following conditions: LED 1 – Yellow: indicates the Device Charging Status; LED 2 – Green: indicates the Device Operative Mode Status; LED 3 – Yellow: indicates special Events occurrences. The  following  table  summarizes  the  possible  Leds  configuration  status  and relative significance: Signalling LEDS
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  37  `            Led states OFF Led Off  SLOW Led Turned on every  10 Sec & OnTime 1000 ms  MEDIUM Led Turned on every    1 Sec & OnTime 500 ms  FAST Led Turned on every  0.5 Sec & OnTime 100 ms  ON Led On   d.n.c. Do Not Care          LED Priority High H     Medium M     Low L                   HW/SW driven SW driven SW driven   LED1 LED2   LED3     Charging Led Operative Led   Event Led     Yellow Green Yellow   STATE STATE Priority STATE Priority Device BGD OFF OFF OFF   OFF   ON charge ON (by HW) d.n.c.   d.n.c.   FULL charge OFF d.n.c.   d.n.c.   Identification d.n.c. Medium-Q H Medium-QN H        Modes IDLE d.n.c. ON L OFF   ENGAGED d.n.c. L d.n.c.   SERVICE ON (by SW) Medium-Q H Medium-QN H ---> SERVICE (FOTA) ON (by HW) d.n.c.   d.n.c.   ---> SERVICE (ACCEPT-TEST) Test driven MONITORING d.n.c. SLOW L d.n.c.   STREAMING d.n.c. FAST M d.n.c.          Events Symptomatic Event Not considered Electrodes Detached d.n.c. d.n.c.   FAST M Low Battery Level SLOW (by SW) d.n.c.   d.n.c.          Communic. Active Connection (In/Out) d.n.c. FAST M d.n.c.   Data Transfer Failure d.n.c. d.n.c.   ON L
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 38  2.2.4   MULTIFUNCTION PUSH BUTTON The  following  figure  shows  the  multifunction  push  button  placed  on  the  front cover of the Bodyguardian Control Unit.  Figure 2-10 – Multifunction push button  The  push  Button  turns  ON  and  OFF  the device  and  insert  symptomatic  events notification from the patient. The button has the following functional modes: - Power on/off  When the device is OFF, press the button with a short pressure to Turn ON the device (T>0sec).  When the device is ON, press the button with a long pressure to Turn OFF the device (T>10sec). - Additional functions  When  the  device  is  ON  and  operative,  click  the  button  to  record  a symptomatic  event.  The  pressure  time  must  be  <  10sec  to  prevent  the device to Turn OFF. Multifunction  push button
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  39  `   2.3  BODYGUARDIAN CHARGING CRADLE The Bodyguardian Control Unit is provided with a proprietary Charging Cradle which must be used to re-charge the internal battery (Figure 2.11).           Figure 2-11 – Bodyguardian Charging Cradle – Upper view                              Figure 2-12 – Bodyguardian Charging Cradle – Side view   In the previous figures the following components can be found:  5. Recharging contacts (to charge the Bodyguard Control Unit) 6. Snaps (to properly place and hold the Bodyguardian Control Unit) 7. Input voltage connector (connection of AC/DC medical adapter)  7  5  6
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 40  2.4  AC/DC MEDICAL POWER SYPPLY The  following  figure  shows  the  specified  medical  AC/DC  adapter  for Bodyguardian Charging Cradle powering up and their parts: 8. AC/DC Medical power supply 9. Mains plug 10. Isolated voltage output connector with cable  Figure 2-13 – AC/DC Medical Power Supply 2.5  DISPOSABLE ADHESIVE ELECTRODES PATCH The  following  figure  shows  the  disposable  adhesive  electrodes  patch,  which integrates four electrodes used to acquire ECG and Bio-impedance signals.  Figure 2-14 – Disposable adhesive electrodes patch It is attached to the patient’s skin and is connected to the Bodyguardian Control Unit using four snaps. Each SnapStrip has a water resistant layer, ECG electrodes and snaps that create an electrical connection to the Bodyguardian Control Unit. 8 10 9
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  41  `   CHAPTER 3 POWERING UP THE DEVICE  The  Bodyguardian  Control  Unit  is  powered  through  an  internal  battery.  The battery is rechargeable by means of the Bodyguardian Charging Cradle connected to the mains trough the AC/DC power supply. 3.1  BATTERY CHARGING While operating, if the Bodyguardian Control Unit detects that the battery level is below  a  certain  level    (Low  Battery  Level  as  defined  in  the  Preventice Requirement Document) of the full capacity, LED 1 ( ) starts blinking with a slow frequency, 1 second every 10 seconds, indicating that the device needs to be recharged.  In this case, switch off Bodyguardian Control Unit and remove it from the strip placed on the patient. Additionally,  if  the  Bodyguardian  Control  Unit  detects  that  the  battery level  is below  a  safety  level  (Critical  Battery  Level  as  defined  in  the  Preventice Requirement Document) of the full capacity, the Bodyguardian Control Unit starts an automatic power off procedure to safely close the application and guarantee the integration of the data sampled and stored up to that moment.    ATTENTION  If the device is not switched off manually by the user, when approaching a very low battery level, the  device will  initiate  an  automatic power down procedures which will allow uploading all the data acquired and safely switch off the device.   To  recharge  the  Bodyguardian  Control  Unit  place  it  on  the  Charging  Cradle according  to  the  appropriate  orientation  (only  one  is  possible  due  to  the mechanical constraints) and only after this operation connect the Charging Cradle to the AC/DC Power supply and this last to the mains. If the Bodyguardian Control Unit is properly placed and  connected, the LED 1 ( ) turns ON indicating that the device is charging. When the device battery is fully charged the Led 1 turns OFF. After the recharging operation is complete disconnect the AC/DC Power supply from  the  mains  and  only  after  this  operation  remove  the  equipment  from  the Charging Cradle. .
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 42      Figure 3-1 – Connection to Bodyguardian Charging Cradle 3.1.1  RECORDING AUTONOMY The autonomy of  operating  and  recording of  the  Bodyguardian Control  Unit is determined by many factors linked each other. Main factors determining the duration are the following:  Battery charge level  Batteries performance  Data Sampling frequencies  Operative Mode. The Rechargeable battery integrated in the device has a capacity of 380mA. When fully  charged, the  battery  guarantees  an  autonomy  which  span  from  3  hours,in continuous Streaming mode of all the data at the maximum sampling frequency, to 24 hours ,in Monitoring mode with data sampled at the lowest frequency and stored in the internal memory for uploading at the end of the usage.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  43  `   3.2  SWITCHING ON/OFF THE DEVICE In order to switch ON the Bodyguardian Control Unit, push with a short pressure the button area in middle of the upper face (Push button). LED 2 and LED 3 will flash when the Bodyguardian Control Unit is turned on. After the Bodyguardian Control Unit completes its initialization, the middle led ( ) will flash green. The Bodyguardian Control Unit can be switched OFF by pressing the Push button with a long pressure (more than 10 seconds). All the LEDS will turn OFF.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 44  CHAPTER 4 WORKING MODE 4.1  PREPARING THE PATIENT The  stage  of  preparing  the  patient  is  very  important  in  order  to  obtain  a  good acquisition of the biological signals and can be divided in the following phases:  Identification of the Disposable adhesive electrodes patch application site.  Preparation of the skin.  Application of the Disposable adhesive electrodes patch. 4.1.1  DISPOSABLE ADHESIVE ELECTRODES PATCH APPLICATION SITE   Attach the BodyGuardian Device at one of three positions on your chest.   Your healthcare provider  will  show  you correct placement on  your chest. Follow  the instructions  carefully.  Correct  placement  is  very  important  to  get  accurate readings from your heart.          T 4.1.2  PREPARING THE SKIN  The phase of preparing the body area on which placing the Bodyguardian Control Unit, has an important role in establishing a good electric contact, thus allowing the best kind of recording.  Here below the recommended procedure: Figure 4-1  Preferred Bodyguardian Placement
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  45  `    Prepare the skin area by cleaning with soap and water.  If hair is present, shave the area.  Make sure application area is dry with no lotion or cream applied. 4.1.3  PLACING THE DISPOSABLE ADHESIVE ELECTRODES PATCH   Follow  the  following  instructions  in  order  to  correctly  place  the  Bodyguardian Control Unit. 1. Before removing the Bodyguardian Control Unit from its charging station, ensure that the control unit is fully charged; the light LED 1 ( ) should be OFF indicating a full charge. 2. Attach the Bodyguardian Control Unit to the four snaps on the Electrodes Patch.  Make  sure  all  four  snaps  are  connected;  see  Error!  Reference source not found.2.    Figure 4-2  Connecting Bodyguardian Control Unit to Electrodes Patch  3. Peel plastic off the back of the Electrodes Patch. 4. Ensure that the arrow on the Electrodes Patch points towards your left arm if the patch is placed vertically (Error! Reference source not found.) or up towards  your  head  if  placed  horizontally  (Error!  Reference  source  not found.). 5. Apply adhesive side of the Disposable adhesive electrodes patch to the skin on the established placement (§ 4.1.1), and gently press.   NOTE  The  Disposable adhesive electrodes  patch  should be  comfortable when properly positioned on the chest.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 46  4.2  OPERATIVE MODES The Bodyguardian Control Unit  can be configured  to run in different operative modes depending from the conditions (i.e. presence of Bluetooth connection with an Associated Device, configuration settings received, etc.) as described below:  IDLE Mode: is when the Bodyguardian Control Unit is powered ON and it is waiting to establish a Bluetooth pairing connection with the Associated Device.  ENGAGED Mode: is when the Bodyguardian Control Unit is powered ON, waiting  for  a  command  from  a  paired  Associated  Device  or  a  request  for pairing from a new Associated Device.  MONITORING Mode:  is  when the  Bodyguardian Control Unit  is  powered ON,  gathering  and  storing  in  the  internal  memory  data  at  the  frequency specified in the Configuration Settings, and sending the data as requested by the Associated Device.  STREAMING Mode: is when the Bodyguardian Control Unit is powered ON and sampling the data as specified in the Configuration Settings, sending them directly  to  the  Associated  Device  with  periodicity  specified  by  the  last configuration commands received from the Associated Device. The two main operative modes which characterize the functionalities and intended use of the device are Monitoring Mode and Streaming Mode. 4.2.1  MONITORING MODE In monitoring mode the Bodyguardian Control Unit operates as a recorder of the physiological data. The acquired signals are stored in the internal memory of the device and works as an Holter. The data are sent from the Bodyguardian Control Unit to the Associated Device according to the configuration parameters received from  the  Associated  Device  during  the  configuration  settings.  The  Associated Device is responsible for forwarding the data to a remote storage server for later analysis and evaluation. In  Monitoring  Mode  the  Bodyguardian  Control  Unit  can  acquire  and  store  the following data, depending on the configuration received:  Raw Data o ECG (128 or 256 Hz)  Derived values o Activity level o Body position values o Breathing rate values o Heart rate values o Heart rate reliability o RR interval variability o Battery level
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  47  `   The  upload  frequency  depends  on  the  Configuration  Specifications.  The Bodyguardian Control Unit is equipped with an internal memory capable to store up  to  24  hours  of  data  recording,  enabling  the  device  to  work  as  an  Holter equipment. 4.2.2  STREAMING MODE In Streaming mode the Bodyguardian Control Unit operates as an acquisition and transmitter of the physiological data. The  acquired  signals are not stored in the internal  memory  of  the  device  but  are  sent  directly  from  the  Bodyguardian Control Unit to the Associated Device according to the configuration parameters received  from  the  Associated  Device  during  the  configuration  settings.  The Associated  Device  is  responsible  for  forwarding  the  data  to  a  remote  storage server for later analysis and evaluation. In Streaming Mode the Bodyguardian Control Unit can acquire and transmit the following data, depending on the configuration received:  Raw signals: o ECG (128 or 256 Hz) o 3-axis accelerometer (50 Hz) o dZ bioimpedance (32 Hz) o Z0 bioimpedance (32Hz)  Derived values: o Heart rate values, o Heart rate reliability, o Breathing rate values, o Activity level, o RR interval variability, o Body position values. 4.2.3  EVENT MONITORING The Bodyguardian Control Unit has the capabilities to register symptomatic and asymptomatic events during the functionalities. Symptomatic Events recording: while wearing the Bodyguardian Control Unit the user has the possibility to initiate an event recording, if is feeling symptoms that suggest the need for  further  evaluation,  by  pressing  the  push  button  in  the middle  of  the  device.  The  functionality  is  enabled  either  in  Monitoring  or Streaming  Modes.  In  case  of  symptomatic  event  button  push,  Bodyguardian Control Unit will send a notification of the event to the Associated Device. If the Associated  Device  is  not  available,  the  Bodyguardian  Control  Unit  will  retry according to implementation and configuration requirement and eventually store in the internal memory the undelivered notifications.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 48  Asymptomatic Events recording: while wearing the Bodyguardian Control Unit some asymptomatic events can  be recorded. Asymptomatic events occurs  when the  values  of  physiological  parameters  detected  violates  the  medical  protocol specified by the configured thresholds.  In case of asymptomatic event detection the  Bodyguardian  Control  Unit  will  send  a  notification  of  the  event  to  the Associated Device. If the Associated Device is not available, the  Bodyguardian Control  Unit  shall  retry  according  to  the  implementation  and  configuration requirement  and  eventually  store  in  the  internal  memory  the  undelivered notifications.  4.3  INSTALLATION AND INSTRUCTIONS FOR THE PATIENT   The Bodyguardian Control Unit does not require specific installation procedure, except  for  the  configuration  section  which  must  be  executed  by  the  external Associated Device.  WARNING  In  order  to  configure  the  Bodyguardian  Control  Unit  and  to  know  the  software details please refer to the indications contained in the User Manual provided by the System Builder.   In order to turn ON the Bodyguardian Control Unit push the button area in the middle of the device (Figure 4.4). The LED 2 and LED 3  will flash when it is turned on.  After the  Bodyguardian Control Unit completes  its  initialization, the middle led will flash green. According to the configuration established, the Bodyguardian Control Unit  will start the data acquisition and transmission.  During the monitoring, if you want to mark a symptomatic event, press the push button in the middle of the Bodyguardian Control Unit.   Figure 4-3  Bodyguardian Control Unit Push Button
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  49  `   CHAPTER 5 MAINTENANCE 5.1  GENERAL INFORMATION ABOUT MAINTENANCE In order to keep the Bodyguardian Control Unit working for a long time and to ensure  the  patient’s  and  the  operator’s  safety,  it  is  necessary  that  the  general checks  indicated  below  are  periodically  performed  by  medical  or  paramedical qualified staff or by technical staff authorized by Preventice.  Perform a sight inspection of all the components, the accessories, in order to identify any traces of failure, damage, or disconnection.  Verify that all labels and any warning or instructions printed on the device are readable.  Check that the performances of the device are correct.  Clean  the  external  surface  of  the  device  carefully  with  the  recommended products only.  Replace parts or accessories only with those having the same characteristics or expressly indicated by Preventice.  Discard replaced parts, accessories, and the device at its “end of life” according to the local standards and directives currently in force. For all  ordinary maintenance operations pertaining to  the  components  to which the device is connected or auxiliary components not produced by Preventice such as external Associated Device (Android smartphone or other), please refer to the corresponding user’s manual provided with them.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 50  5.2  SAFETY CHECKS It is essential to periodically check the equipment and the devices or systems it is connected  to  and  all  the  connections  in  order  to  ensure  that  the  equipment continues  working  efficiently  and  safely.  It  is  also  necessary  to  check  the equipment  to  remove  any  dust  deposits.  Preventive  or  corrective  maintenance operations must be performed by qualified technical staff expressly authorized by Preventice. A  sight  inspection  of  the  interconnections,  with  particular  care  for  the  snaps between  equipment  and  the  Disposable  adhesive  electrodes  patch,  can  be performed also by the user in order to remark any breaking or disconnection. In case  of  need,  immediately  contact  a  qualified  technician  to  solve  the  problem detected before continuing to use the equipment or connecting it to other devices. For the technical assistance request procedures, please  refer  to chapter “Request for assistance” of this manual.  ATTENTION  Safety checks must be accurately performed periodically.    5.2.1  CONNECTORS Check the connector on the  Bodyguardian Control  Unit  for broken or  damaged contacts.  If  you  find  a  damaged  connector  refer  to  qualified  technical  support service. 5.2.2  BATTERY PACK The rechargeable battery pack installed on the device is not accessible to the user and its  replacement  should be  performed  only by qualified  personnel  expressly authorized  by  Preventice.  Anyway,  you  should  consider  the  following  general warnings (e.g. in case of disposal of replaced parts with technical assistance). • Do not let the ends of the battery pack come into contact with metal objects. • Keep the battery pack away from heat or flames. • Do not immerse the battery pack and avoid exposing it to rain or moisture. • Avoid direct mechanical trauma to the battery pack. • Do not attempt to disassemble, puncture, incinerate or short-circuit the battery. These operations may cause a fire or emit toxic chemicals.   ATTENTION When a battery is replaced, it cannot be reused and must be discarded according to  the  standards  and  directives  currently  in  force  in  the  country  where  the equipment is used.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  51  `    WARNING  The state of battery charge is not guaranteed for inactivity periods longer than 1 month of storage at temperatures up to +45 ° C. It is generally recommended to perform a cycle of discharge/charging the battery after periods of inactivity.  5.3  CLEANING THE DEVICE It is necessary to keep the equipment clean in order to avoid dust deposits, which could interfere with the efficiency of all the system components. It  is  possible  to  clean  the  equipment  external  surface  with  a  cloth  lightly moistened with warm water and soap. Wipe the washed parts with a dry cloth.  WARNING  Do not immerse the equipment nor its parts in liquids, do not oil any part of it and avoid  cleaning  the  external  surface  with  alcoholic  disinfectants that  could cause damages and discoloration of the printed surfaces.   ATTENTION  Before  cleaning  any  part  of  the  equipment  disconnect  the  equipment  from  the power  supply  and  disconnect  the  device  from  any  other  equipment  or  external devices.    ATTENTION  Make  sure  no  liquid  seeps  into  the  instrument  and  check  it’s  complete  dryness before to reconnect to the patient or before connecting it with other devices, thus switching it on.    WARNING  Use  Isopropyl  Alcohol  or  equivalent  products  in  order  to  disinfect  the BodyGuardian Control Unit, without immersing them into the liquids. Use products according to directives and regulations of the country in which the device is used.  WARNING  All disposable (band-aid) must be destroyed after using them and cannot be used again  in  any  way.  Please  refer  to  the  directives  and  regulations  concerning disposable requirements of the country in which the device is used.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 52  5.4  PARTICULAR WARNINGS FOR CRITICAL COMPONENTS   ATTENTION  The internal battery installed in the equipment cannot be accessed by the user and its replacement must be performed only by qualified staff expressly authorized by Preventice.  In every case please follow the general warnings listed below (for exemple in case of discard of components replaced by servicing technicians).  Avoid terminal parts of the battery pack coming in contact with metal objects.  Keep the battery pack away from heat sources or flames.  Do not immerse the battery and avoid to exposure to rain or humidity.  Avoid directly hitting the battery.  Do not attempt to disassemble, burn or cause short-circuits to the battery. Such operations may cause a fire or emission of toxic chemical substances. In the event that electrolytes escape, make sure you do not touch it. Wash with water for at least 15 minutes any body part that may have been in contact with it. Should you experience any symptom after this period, ask for immediate medical help.  ATTENTION  When a battery is replaced, it cannot be reused and must be discarded according to  the  standards  and  directives  currently  in  force  in  the  country  where  the equipment is used.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  53  `   CHAPTER 6 TECHNICAL CHARACTERISTICS 6.1  BODYGUARDIAN CONTROL UNIT  Product name  Bodyguardian Control Unit  Description    Active, non-invasive medical device  Intended use  The  Bodyguardian  Control  Unit  is  used  for  ambulatory monitoring  of  non-lethal  cardiac  arrhythmias.  Bodyguardian  Control  Unit  is  a wearable  electronic  device  that  is  worn  on  the  chest.  The  device  includes  a disposable fabric adhesive component that attaches to the subject and connects to the enclosed electronic components.  The device will be worn intermittently or continuously for up to a 30-day period. The recordings will be stored on the Bodyguardian Control Unit and transmitted by  the  device  via  Bluetooth  communication  to  an  external  Associated  Device (Android smartphone or other). The device is prescription only.  The Bodyguardian Control Unit recorder is intended to be used in a:   “Home Care” environment without clinicians’ surveillance.  “Clinic Care” environment with clinicians surveillance Depending  on  the  Environment,  optional  algorithms  and/or  thresholds  (settable via host software) can be used to generate measurements on an advisory basis for patients. These are presented for:   Review and interpretation by the clinician, based upon knowledge of the patient.  Results of the physical examination. The data provided by the Bodyguardian Control Unit are exclusively intended to be used by trained medical personnel to assist patients that require monitoring of physiological parameters with reference of the followings context:   Screening  of  patients  with  symptoms  suggesting  arrhythmia  over  a minimum 24-hour period.  Screening and off line evaluation of HR and Breath variability if the nature of this variability cannot determine a Life-threatening for the patient.  Not intended to allow direct diagnosis or monitoring of vital physiological processes parameters (for instance cardiachearth rate, respiration), where the nature of variations is such that it could result in immediate danger to the patient.  Not intended for therapeutic purpose.
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 54  The “Body Guardian” Control Unit is intended from adult without race exclusion. Patient  with  implanted  device  are  excluded  to  wear  this device.  Applied Standards  Standard  Edition / year  Description IEC 60601-1 Ed. 2 + A1 + A2 + Deviation UL/CSA Medical Electrical Equipment – General requirement for safety IEC 60601-1-1 Ed. 2 Collateral standard: Safety requirements for medical electrical systems IEC 60601-1-2 Ed. 2 + A1 Collateral standard: Electromagnetic compatibility Requirements and tests IEC 60601-1-4 Ed. 1 Collateral standard: Programmable electrical medical systems IEC 60601-1-6 Ed. 1 Collateral standard: Usability IEC 60601-2-47 Ed. 1 Particular requirements for the safety, including essential performance, of ambulatory electrocardiographic systems IEC 60601-2-49 Ed 1 + Ec1 Particular requirements for the safety of multifunction patient monitoring equipment IEC 60529 Ed. 2 + A1 Degrees of protection provided by enclosures (IP Code) EC38 2007 Particular requirements for the safety, including essential performance, of ambulatory electrocardiographic systems EC57 Ed. 2 Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms UL 60601-1 Ed. 1 Safety of Medical Electrical Equipment Part 1 – General requirements for Safety CAN/CSA C22.2 Ed. 2 + A1 + A2 Medical Electrical Equipment Part 1 – General requirements for Safety  Type of protection against electric shocks   Internal powered equipment.  Protection level against electrical direct and indirect contacts   BF type  (patient inputs)  Protection level against inflow of solids and liquids   IPX4    Device protected against  Operational mode   Continuous, within the specified limits  Environmental conditions for usage   - Temperature:    from +10°C to +45°C   - Relative humidity:    from 10% to 95% RH   - Atmospheric pressure:  from 700hPA to 1060hPA  Environmental conditions for storage (max. 15 weeks )
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  55  `     - Temperature:    from -30°C to +60°C   - Relative humidity:    from 5% to 95% RH   (excluding condensation)   - Atmospheric pressure:  from 500hPA to 1060hPA  External dimensions and weight   Height:  ~ 17 mm   Width:   ~ 60 mm   Depth:   ~ 50 mm   Max Weight:  ~ 35 g  Case Material:    PC Polycarbonate    Battery Capacity  350 mAh min, 380 mAh typical, 1.5Wh   Battery Charger Input Power  Requirement        100-240 VAC, 50-60 Hz,  Battery Charger Output Power  Requirement        6 VDC, 200mA,       Battery Life        500 Cycles > 70% of initial          capacity  Battery Voltage   3.7V  ECG  Sampling Rate  128/256 Hz  Digital Resolution  12 bit  Input Dynamic Range  ± 10 mV  Input Offset Dynamic Range  ± 300 mV  Bio Impedance  Bio-impedance injection  100µA @ 50KHz  Bio-impedance sampling  32/second  Impedance  0 to 120 Ohms  Maximum allowed Load  7KOhms  Accelerometer  Accelerometer activity sampling 50/second
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 56   Accelerometer  3 axis 12-bit  Sampling Rate  ECG  128/256 Hz  Impedance  32 Hz  Accelerometer  50 Hz  Measurement Ranges  Heart Rate  25 to 240 BPM  Respiration  0 to 30 BrPM  Posture  ± 2g range in x,y,z direction  Data Storage  Capacity  24 hours continuous recording  Type  Internal NAND Flash  Communication Means  Bluetooth communication between    Bodyguardian  Control  Unit  and   Smatphone  Communication Frequency  2.4 GHz  Electrode connections      - 4 ECG snaps connectors   Other interface      - 1 yellow Led for Device Charging Status           - 1 green Led for Device Operative Mode Status          - 1 yellow Led for special Events occurrences           - 1 multifunction push button  Power supply  Battery  powered  (Rechargeable  built-in Lithium Battery)  Absorption during Recharging  <600mW 6.2  DISPOSABLE ADHESIVE ELECTRODES PATCH    The characteristics specification of this component is responsibility of the System Builder
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  57  `   6.3  AC/DC MEDICAL POWER SUPPLY 6.3.1  OPTION 1  Manufacturer      FRIWO Model        AC/DC ADAPTER FW7662M/06 Input        100-240 VAC – 50-60 Hz Output        5.9VDC @ 1A Safety standards      Fulfils Class II SELV for            IEC 60601-1, UL 2601, VDE, CE label, SIQ,           Fulfill medical application class B / BF / CF  Electrical protection level  Class II  Case        Material:    Plastic           Dimensions:  52 x 52 x 35.5 mm 6.4  BODYGUARDIAN CHARGING CRADLE  Input/Output  Operative  5.9VDC  nominal.  Connection  with specified  AC/DC  adapter  (model  FRIWO FW7662M/06) Input connector THT  POWER  SUPPLAY  JACK  4,4mm opening diameter p/n 1613_05 Output connector  Spring-Loaded Contacts Pins Case  Material:  PC Policarbonate Dimensions: 63 x 54 x 21 mm Operational mode    Continuous, within the specified limits  Environmental conditions for usage         - Temperature:    from +10°C to +45°C   - Relative humidity:    from 10% to 95% RH   - Atmospheric pressure:  from 700hPA to 1060hPA  Environmental conditions for storage (max. 6 months)   - Temperature:    from -30°C to +60°C   - Relative humidity:    from 5% to 95% RH   (excluding condensation)   - Atmospheric pressure:  from 500hPA to 1060hPA
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual 58  CHAPTER 7 REQUEST FOR ASSISTANCE 7.1  OBTAINING SERVICE In case of problems such as failure of the device or anyway in case of partial or incorrect  working  that  cannot  be  solved  through  usual  maintenance  operations, please contact  one  of  the main  offices or branches  of  Preventice or the  nearest retailer or authorized servicing center.  ATTENTION  In case of failure of the device or if it starts working in a way not complying with what is written in the manual, especially as far as safety is concerned, STOP USING IT IMMEDIATELY and contact the technical service. Do not use the device until the safety conditions have been checked and restored.   NOTE  In  order  to  speed  up  the  procedures  to  start  the  intervention  of  the  technical service  and  to  make  it  easier  for  the  specialized  technical  staff  to  identify  the problem on the first phone call by the customer, please fill in the form below in this page. The equipment data may be found on the equipment identification label.     REQUEST FOR ASSISTANCE  Preventice Device/system name.....................................................................................  Preventice Device code/reference number (REF) .......................................................  Serial number (SN) or lot number (LOT) ....................................................................  Current software version (Rel) ....................................................................................
BODYGUARDIAN CONTROL UNIT – BASE KIT   Operator Manual  59  `   7.2  PREVENTICE MAIN OFFICES OPERATING OFFICES  PREVENTICE 1652 Greenview Drive SW Rochester, MN 55902 Phone  800-509-0503 Fax  507-281-3630 Website www.preventice.com  Any other authorized assistance center or Technical assistance numbers.