SharperImage.com
T.E.N.S. Massager with Handheld Control

.

Operating Manual

Contents
Introduction……………………………………………………..……………………….3
Safety warnings………………………………………………………………….....…4-5
Set up…………………………………………………….………..…………………….6
Operating the product…………………………………………..……………………..7-8
Cleaning and maintenance…………….………………………….……………………..9
Trouble shooting………………………………………………….……………………10
Positions for use………………………………..………………….…………………...11

.

2

Introduction
The Sharper Image T.E.N.S. Massager with Handheld Control delivers electric pulses to
tired and sore muscles. These pulses are generated by the device and delivered through
electrodes on the skin to trigger contractions of the muscles. It is portable, compact and
stylish in design. The device has six auto operational programs represented by six
selection buttons of body areas. The LCD remote displays information about the mode,
intensity and time.

.

Indications for Use
To be used for temporary relief of pain associated with sore and aching muscles in the
shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and
work activities.
Features of the T.E.N.S. Massager with Remote Control:

•
•
•
•
•
•
•
•

Touchscreen with backlight
Lock screen function
Power: DC 6V, 4 AAA batteries
Size: 2.28x7.48x0.79 inch
Weight: 4.87 oz
Two output channels
Adjustable intensity
Accessories included: 4 electrode pads, 2 electrode wires and 4 AAA batteries

3

Safety warnings

Do not use this device if you have an implanted defibrillator or implanted metallic
devices. Such use could cause electrical shock, burns, electrical interference or death.

.

If you have one of the following conditions, please consult with your physician before
purchasing or using this device.
Acute disease, malignant tumor, infective disease, pregnant, heart disease, high fever,
abnormal blood pressure, lack of skin sensation or an abnormal skin condition, any
condition requiring the active supervision of a physician.

Do not use this device while driving.
Do not use this device while sleeping.
Do not use this device in high humidity areas such as a bathroom.
Keep the device away from wet, high temperature and direct-sunlight place.
Keep this device out of reach of children.
Stop using this device at once if you feel pain, discomfort, dizziness or nausea and
consult your physician.
Do not attempt to move the electrode pads while the device is operating.
Do not use the device around the heart, on the head, mouth, pudendum or blemished
skin areas.
Do not apply stimulation of this device in the following conditions:
(1) across the chest because the introduction of electrical current into the chest may
cause rhythm disturbances to the heart, which could be lethal;
(2) over painful areas. Please consult with your physician before using this device if you
have painful areas;
(3) over open wounds or rashes, or over swollen, red, infected, or inflamed areas or skin
eruptions (e.g., phlebitis, thrombophlebitis, varicose veins). Apply stimulation only to
normal, intact, clean, healthy skin;
(4) in the presence of electronic monitoring equipment (e.g., cardiac monitors, ECG
alarms). The electronic Stimulator may not operate properly when the electrical
stimulation device is in use;
(5) while operating machinery, or during any activity in which electrical stimulation can
put you at risk of injury;
(6) on children.
Be aware of the following.
(1) to consult with your physician before using this device. The simulation with the
device may:
i. cause lethal rhythm disturbances to the heart in susceptible individuals, and,
ii. disrupt the healing process after a recent surgical procedure;
4

(2) that the device is not effective for pain of central origin, including headache;
(3) that the device is not a substitute for pain medications and other pain management
therapies;
(4) that the device has no curative value;
(5) that the device is a symptomatic treatment and, as such, suppresses the sensation of
pain that would otherwise serve as a protective mechanism;
(6) that the long-term effects of electrical stimulation are unknown;
(7) that the user may experience skin irritation, burns or hypersensitivity due to the
electrical stimulation or electrical conductive medium (gel);
(8) if the user has suspected or diagnosed epilepsy, the user should follow precautions
recommended by his or her physician;
(9) to use caution if the user has a tendency to bleed internally, such as following an
injury or fracture;
(10) use caution if stimulation is applied over the menstruating uterus;
(11) use caution if stimulation is applied over areas of skin that lack normal sensation;
(12) stop using the device if the device does not provide pain relief; and,
(13) use this device only with the leads, electrodes, and accessories that the
manufacturer recommends.

.

Medical Electrical Equipment needs special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the
EMC information provided. Portable and mobile radio frequency (RF) communication
equipment can affect Medical Electrical Equipment.
Environmental condition for transport and storage

Easily fragile product

Keep the product in the dry place away from water and rain.
Product package should be recycled.
Stacked up 4 cartons at most

Temperature range: 0oc to 48oc
Humidity: 10% to 85%
Atmospheric pressure: 400hPa to 1060hPa

5

Setup
Unpack the box of the product, take the product and batteries out, place the batteries
into the device at its back compartment, and connect the electrode pads into the device
through the wires.

.
6

Operating the product
To begin
1. Install 4 AAA batteries into the device. Match the positive electrode and
negative electrode in the right direction;
2. Connect 1 pair of electrode wire into the upper left hole of the device and
another pair of electrode wire in the upper right hole.
3. Attach the electrode pads (connecting to the electrode wire) to the treatment area,
such as shoulder.
4. Turn the power/intensity switch knob to 1. The screen should be bright with blue
light. This means the device is ready to begin treatment.
5. Press one of the function keys to start treatment, such as Shoulder or Mode1.
6. Rotate the Switch Knob to adjust the intensity.
7. Press TIMER to adjust the time.

.

Note: Start from the lowest speed and intensity, and then gradually adjust to a
comfortable level on a scale level from 1 to 10.

7

Right output hole

Left output hole

LCD screen

Auto program buttons

Function buttons

.

Timer

Auto program buttons

Electrode wire

Electrode
E
lectrode pads

Recommended practices:
1. Duration is 10-15 minutes for each treatment area.
2. Frequency is 2 times per day per area.
3. Intensity and speed selections are based upon the level of your comfort.

8

Cleaning and maintenance

.

Please use water or neutral detergent to clean the device first then using the dry cloth to
wipe it again. The electrode pads coming with the device are disposable, and should be
replaced when their adhesiveness becomes worse. Contact the seller for replacements.
Do not let the sticky side of the pad touch anything, including the greasy finger tips.

.
9

Trouble Shooting
If your device is not operating properly, please check it for the following problems or
conditions.
MALFUNCTION
Stimulation is weak or not
existent

CAUSE
Are the electrode heads in
close contact with the skin?

SOLUTION
Remove, clean and place
back on the skin.

Do the electrode heads
touch each other?

If so, then separate.

Are the wires connected to
the electrode heads and the
main device properly?

Disconnect, clean and
reconnect all the parts.

Skin turns red

Are the batteries dead?
Is treatment time too long?

Skin feels uncomfortable

Are electrode pads dirty?

Replace the batteries.
Stop and consult your
physician if symptom
lasted.
Clean with damp cloth or
replace.

Device does not turn on

Pads are not sticky

Customer Service

.

Do the electrode pads touch
the skin closely?
Batteries installed
correctly?

Remove and place again on
the skin
Remove and install
batteries.

Are batteries dead?
Pads get deteriorative

Replace the batteries.
Contact vendor for
replacements.

Sharper Image branded items purchased from SharperImage.com include a 1-year
limited warranty. For assistance, please call 1 (877) 210-3449.
.

10

Positions for use

.
11



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Create Date                     : 2015:07:22 13:19:24Z
Creator                         : Word
Modify Date                     : 2015:07:22 09:21:14-04:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Creator Tool                    : Word
Metadata Date                   : 2015:07:22 09:21:14-04:00
Keywords                        : 
Producer                        : Mac OS X 10.10.1 Quartz PDFContext
Format                          : application/pdf
Title                           : Microsoft Word - 204058 TENS Unit-Manual.doc
Document ID                     : uuid:61983900-3792-d640-b7d5-0521fe6337f5
Instance ID                     : uuid:50907494-8333-b54b-b284-0797bd5362f4
Page Count                      : 11
EXIF Metadata provided by EXIF.tools

Navigation menu