ShenZhen Jumper Medical Equipment JPDFR400 Infrared Thermometer User Manual EAR THEMOMETER LEAFLET A W

ShenZhen Jumper Medical Equipment Co.,Ltd Infrared Thermometer EAR THEMOMETER LEAFLET A W

Users Manual

Download: ShenZhen Jumper Medical Equipment JPDFR400 Infrared Thermometer User Manual EAR THEMOMETER LEAFLET A W
Mirror Download [FCC.gov]ShenZhen Jumper Medical Equipment JPDFR400 Infrared Thermometer User Manual EAR THEMOMETER LEAFLET A W
Document ID3532451
Application IDjrxdsIHd5uHdjfEBH9FVcg==
Document DescriptionUsers Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize38.64kB (483010 bits)
Date Submitted2017-08-29 00:00:00
Date Available2017-09-05 00:00:00
Creation Date2017-08-24 02:42:56
Document Lastmod2017-08-24 02:42:56
Document TitleEAR THEMOMETER LEAFLET A-W
Document Author: 01

JPD-FR400
0482
Infrared Thermometer
Instruction Manual
Shenzhen Jumper Medical Equipment Co., Ltd
Manual Ver.: 1.0
Issuing Date: 2016/09/12
Introduction
Thank you for choosing the Dual-mode Digital Infrared Thermometer of
Jumper Medical, Infrared thermometer for short.
The infrared thermometer JPD-FR400 can be used to read the body
temperature by measure the ear and forehead temperature, which is
suitable for adult and children(ear test mode only for the child above 3
months).
Please read the instructions carefully before using the product, and put it
in a safe and secure place for reference.
Contents
1. The advantages of infrared thermometer JPD-FR400 ............ 1
2. Necessary safety instructions ................................... 2
3.Instructions for product designs .................................. 3
4.How to measure the forehead temperature ........................ 4
5.How to measure the ear temperature ............................. 4
6.Instructions for display and operation ............................. 6
7.Cleaning and Disinfection ....................................... 9
8.Maintenance .................................................. 9
9. Replace the battery. .......................................... 10
10.Troubleshooting ............................................. 10
11. Technical specifications ...................................... 11
12. After-sale service ........................................... 12
13. Security type ............................................... 13
14. Authorized European Representative: ......................... 13
15. Symbols ................................................... 13
16.Declaration ................................................. 13
17. Appendix A: EMC Information-Guidance and Manufacture‟s
Declaration .................................................... 14
1. The advantages of
infrared thermometer JPD-FR400
Multiple use (measurement of forehead temperature and ear
temperature)
Dual-mode design, you are free to take body temperature by detecting
the infrared heat given off by the forehead or eardrum,The Measuring
range of the thermometer in forehead /ear mode is between 32.0°F 212.0°F (0 °C- 100 °C) .
The product consists of ABS plastics, temperature sensor, Infrared
temperature measuring element, microcomputer controlled circuit, LCD ,
backlight and buzzer.
Convenient for use
●Special ergonomic design to facilitate operation.
●No inconvenience would be caused to your daily life. It‟s available for
measurement when your children‟s sleeping.
●It provides comfortable user experience compared with anus
thermometer for children, rapid reading and simple operation compared
with mouth thermometer.
Memory recall
A maximum of 20 previous readings is available to help you track the
changes of your body temperature.
Safe and hygienic
●Compared to mercury thermometer, there is no danger of breaking the
glass or swallowing mercury.
●It is totally safe for children to use.
Fever warning
When the reading exceeds 37.5℃/99.5℉, this product shall warn the
user that he/she may have a fever by 7 rapid and short rings with LCD
flicker.(For normal forehead temperature, the signal is long ring with LCD
no flicker)
Extensive clinical data of hospital
The cooperation with designated hospitals is subjected to precise clinical
verification, with the support of extensive clinical data and professional
medical experts.
Application scope
The thermometer takes human body temperature for clinical or
household use. It applies to all age groups and those who are suffering
from a fever.
Operating principle
The infrared temperature sensor detects infrared energy emitted by the
eardrum. A built-in lens focuses the collected energy, which is then
converted into a temperature reading by the thermopiles and
measurement circuits.
2. Necessary safety instructions
●The device is not used for newborn baby.
●The device is not continuous monitoring device.
●JPD-FR400 is not waterproof. Please do not douse it into water or other
liquids. Cleaning and disinfection procedure shall be in comply with the
instructions specified in [cleaning and storage]
●Please do not use the product if the temperature sensor or the
Dual-mode infrared thermometer shows any sign of damage. Do not try
to repair the product if it is damaged. Please contact the nearest Jumper
Customer Service.
●JPD-FR400 consists of precision parts with high quality. Please prevent
the product from falling off. Protection shall be provided for no intense
shock and vibration. Do not twist the Dual-mode infrared thermometer
and the temperature sensor.
●Operating Conditions:
Temperature: 10C to 40C
Humidity: 15-95% RH, non-condensing
Atmospheric pressure: 86kPa to 106kPa
●Storage and Shipping Conditions:
Temperature: -20C to 55C
Humidity: 0-95% RH, non-condensing
Atmospheric pressure: 50kPa to 106kPa
●Battery Installation:
Put the two AAA batteries into battery
compartment in correct polarities.
Push the battery cover horizontally along the
arrow.
Notes:
☆ If you are the first time using the thermometer, please pull out the
Plastic sheet .
☆ Battery polarities should be correctly installed. Otherwise, damage
may be caused to the device.
☆ Please put in or remove batteries in right order, or may cause damage
to the device bracket.
☆ Please remove the batteries if the thermometer will not be used for a
long time.
Warnings
●Please keep this infrared thermometer out of children‟s reach
●Medical assistance can‟t be replaced by the use of infrared
thermometer
●The infrared thermometer is not waterproof! Keep it away from water.
3.Instructions for product designs
(1) Probe cover (put the cover on when using forehead mode)
(2) Probe (take off the cover when using ear mode)
(3) Forehead mode
(4) LCD display screen
(5) Ear mode
(6) Battery cover
Display description
1. Ambient temperature
2. Ready for measurement
3. Ear mode
4. Head mode
5. ˚C / ˚F
6. Low power indicator
7. Memory mode
8. Temperature value
4.How to measure the forehead temperature
▀ Spot measurement
1. With the sensor/probe cover attached, position the thermometer at the
center of the forehead, just above the eyebrow. Ensure that the
thermometer is in contact with the forehead.
2. Press and release the HEAD button.The temperature will be displayed
on the screen instantly.
▀ Continuous scanning measurement
Point the temperature sensing probe at the middle part of forehead, with a
distance of 1~5 mm(0.04~0.2 inches).Hold HEAD button, scan the forehead
from one side to the other and then release the button,the system will
read sequentially and record the maximum temperature value.
5.How to measure the ear temperature
1. Gently remove the sensor/probe cover to reveal the ear probe.
2. Insert the ear probe into the ear canal.
3. Press and release the EAR button. The temperature will be displayed
on the screen instantly.
Note:
Children under 1 year: Pull the ear straight back.
Children aged 1 year to adult: Pull the ear up and back.
In order to avoid the inaccuracy:
(1)Please make sure that there is no dirt in on the temperature sensor
(2)Please make sure that the device only be used in room, and there is
no strongly conversation of wind.
(3)Before you measuring the forehead temperature, there shall be no
water or any shade on the forehead.
(4) Before you measuring the ear temperature, please clean the ear
canal first.
(5)Please make sure that there is no intense emotion and movement
before measuring.
(6)If JPD-FR400 is transferred from one condition to another which has
different ambient temperature, it is suggested to deposit for more than 30
minutes, and please follow the rule (2).
(7)If the tester is transferred from one condition to another which has
different ambient temperature, it is suggested to have a rest for more
than 10 minutes, and please follow the rule (2).
(8)Do not hold JPD-FR400 for long time as it is highly sensitive to heat.
JPD-FR400 has undergone clinical test, it is safe and accurate when
using in accordance with operation manual.
Connecting the thermometer to an APP application of a
mobile phone
The thermometer has a built-in Bluetooth module and can be connected
to a smartphone by using the Bluetooth 4.0 wireless technology. View the
measurement data in real time by using mobile phone software
"MyThermo". The software automatically saves the measurement data
and draws a temperature curve graph, which helps view the temperature
change status in real time.
For a method for connecting the thermometer to the mobile phone, see
Operation Instruction.
6.Instructions for display and operation
LCD
display
Operational method and
instruction for displays
1.Measurement of forehead
temperature:
Position the thermometer at the
center of the forehead above the
eyebrow. With the thermometer
touching the forehead, press and
release the HEAD button. The
temperature will be displayed on
the LCD screen.
2.Measurement of ear
temperature:
Insert the thermometer ear probe
into the ear canal. Press and
release the EAR button. The
temperature will be displayed on
the LCD screen.
3. To measure again, simply
press the HEAD/EAR button
accordingly.
Sound and backlit
When the temperature is
between 32.0°F/ 0°C and
99.5°F/ 37.5°C, there will
be a long beep and a
green backlit will be
displayed for 3 seconds.
When the temperature is
between 99.6°F/37.6°C
and 107.9°F/42.2°C, there
will be 7 short beeps and
the reading will flicker with
a red backlit for 3
seconds. This indicates
that the subject may have
a fever. Please consult
your doctor if you are not
sure.
Forehead measurement mode
Eardrum measurement mode
The measured value exceeds
212 °F/100°C.
The measured value falls below
32.0°F/0°C.
3 short beeps with a red
backlit for 3 seconds.
3 short beeps with a red
backlit for 3 seconds.
Inquiry for memory data, Storing 20 temperature
readings
LCD
Operational method and instruction for
display displays
Sound and
backlit
Press and hold button EAR for 3-6 seconds
and the LCD will display “- - -” with the M
None
signal blinking.
Press button EAR again and the LCD will
display the number „1‟ with the M signal.
After 1 second, the measured temperature
with the mode of measurement icon will be
displayed. Note that this is the latest reading.
To recall the next recorded reading, press
None
button EAR again. The number „2‟ with the M
signal will be displayed. After 1 second, the
measured temperature with the mode of
measurement icon will be displayed.
Repeat to recall earlier recordings (up to 20
total) if necessary.
The LCD will display “- - -” with the M signal
blinking if there is no test data.
None
°C/°F conversion
LCD
Operational steps
Sound and
display
backlit
When the thermometer is off, press and hold
the EAR button for 6-9 seconds. The screen
will display “- - - ” with “M” at the bottom right.
Continue to hold the button until “- - -°C/°F”
appears on the screen. Release the EAR
None
button and the “- - -°C/°F” temperature unit
will start to blink. Press the EAR button
again within 5 seconds to change the
temperature unit to your choice.
Error message & Low voltage
LCD screen will display “Er1” when ambient
temperature exceeds 104.0°F/40.0°C or
drops below 50.0°F/10.0°C.
LCD screen will display “ErC” if there is
EEPROM data reading error or the
calibrating process is not finished. Please
contact your supplier.
Low-voltage signal when the battery voltage
is below2.51V±0.15V. Please replace
battery.
3 short ticks
and red backlit
for 3 seconds.
3 short ticks
and red backlit
for 3 seconds.
None
Power Off Mode
The thermometer will power off automatically if no activity is detected for
10 seconds.
Attention:
●Electromagnetic interference: JPD-FR400 contains sensitive electronic
component and shall not use under the condition with electromagnetic
interference,( such as the place nearby the mobile phones and
microwaves )
●Please dispose the used products and batteries in accordance with
local regulation requirements when the products and batteries are not
available.
●Please take out the battery if the device unused for long time.
7.Cleaning and Disinfection
Cleaning
1) Clean the temperature probe with a
soft cloth. Clean the lens of the
temperature probe with a cotton
swab.
2) Wipe the thermometer body with a
slightly damp soft cloth, and gently
dry the body with a piece of tissue
paper.
Keep water off the lens during the cleaning process. Otherwise, the
lens may be damaged.
The lens may be scratched if it is cleaned with a piece of tissue
paper, resulting in inaccurate readings.
Do not clean the thermometer with corrosive cleansers. During the
cleaning process, do not touch the lens using hard objects, immerse any
part of the thermometer into liquid, or allow liquid to enter the
thermometer.
Disinfection
1) Disinfect the thermometer body and the area around the
temperature probe with a cloth slightly moistened with 75% medical
alcohol.
Do not use hot steam or ultraviolet radiation for disinfection.
Otherwise, the thermometer may be damaged or quickly aged.
8.Maintenance
1)
2)
After each use, clean the temperature probe as described in
"Cleaning and Disinfection".
Store the thermometer in a dry, dust-free, and well-ventilated place.
Ensure that the thermometer is not exposed to sunlight. Ensure that
3)
4)
the storage and transportation environments meet the requirements
Check whether safety risks exist on a regular basis.
Remove the batteries if the thermometer will not be used for more
than two months.
9. Replace the battery.
Slide the battery cover off along the marked direction. Put two AAA
batteries correctly into the compartment.
Remove the batteries if the thermometer will not be used for more than
two months.
10.Troubleshooting
Symptom
Possible Cause
Solution
The battery level is
Use new batteries of the
extremely low.
same model or
specifications.
Polarities of the batteries
Ensure that the batteries
are reversed.
are correctly installed
The thermometer
according to the polarity
fails to power on.
symbols in the battery
compartment.
The thermometer is
If the warranty period
faulty.
does not expire, contact
Jumper or a local
distributor.
Only the battery
The battery level is low.
Use new batteries of the
symbol is displayed
same model or
after the
specifications.
thermometer
powers on.
"Er1" is displayed.
The ambient temperature
Take a measurement
is lower than 10°C
under an ambient
10
Symptom
Possible Cause
Solution
(50.0°F) or higher than
temperature between
40°C (104°F).
10°C (50.0°F) and 40°C
(104°F).
An error occurs when
Contact Jumper.
data is being read from or
"ErC" is displayed.
written into the memory,
or the temperature
correction is not
complete.
The lens of the
Clean the lens using a
temperature probe is
cotton swab.
dirty.
The temperature
The distance between the
Move the thermometer
reading is lower
temperature probe and
closer to the target.
than the typical
the target is too long.
body temperature
The thermometer is not
Wait for more than 30
range.
used within 30 minutes
minutes after the
after being taken from a
thermometer is moved
cold environment.
into the measurement
environment.
The temperature
The temperature probe is
reading is higher
faulty.
Contact Jumper.
than the typical
body temperature
range.
11. Technical specifications
Items
Standards
Models
1.Applicable
regulations and laws
JPD-FR400
2.Temperature units
℃/℉, adjustable
ASTM E 1965
11
3.Measurement range
Forehead temperature mode:0.0℃-100.0℃ /32.0F –
212.0 F
Ear temperature mode:0.0℃-100.0℃ /32.0F – 212.0
F
4.Accuracy
(Laboratory)
±0.2℃ /±0.4F(36℃-39℃/96.8℉-102.2℉)
±0.3/±0.5℉(less than 36 °C (96.8°F) or greater than
39 °C (102.2 °F)
5.Display resolution
0.1℃/0.1℉
6.Measure time
About 1 second
7.Latency Time
About 1 second
8.Measuring distance
Forehead temperature:1~5 mm(0.04~0.2 inches)
9.Sound
volume≧50 db (the perpendicular distance from dB
Volume sensor to thermometer is 10cm )
10.Automatic
10s±1s
shutdown function
11.Low-voltage display The product shall display low-voltage signal if the voltage
function
is below 2.51V±0.15V.
12.Memory function
Memorize 20 groups of measured temperature.
13.LED backlit
specifications
≧1.2cd/m2
14.Type of measuring Applicable for forehead temperature and ear temperature
15.Power supply
Internal power supply
16.Operating voltage
DC 3V
17.Battery
18.Battery life
AAA ×2
More than 3000 times
19.The date of
production
See the label
20.Life
5 years
12. After-sale service
The device is under warranty for one year since the date of acquisition.
Application for repairing should be presented during the warranty period.
The damage caused by improper use is not under warranty scope.
Batteries and packaging are not under warranty scope as well.
12
13. Security type
The signal indicates that the thermometer is type BF applied part.
14. Authorized European Representative:
Wellkang Ltd
Suite B, 29Harley Street, London, W1G9QR,UK.
15. Symbols
Symbol
Description
Type BF applied part.
Attention must be paid.
Information about a manufacturer, such as name and
address.
Please read the instructions carefully.
Waste electrical materials should be sent to a dedicated
collection point for recycling.
Warning
Attention
A personal injury or thermometer damage may occur if
the thermometer is not correctly used.
Inaccurate reading or thermometer damage may occur if
the thermometer is not correctly used.
16.Declaration
EMC of this product complies with IEC60601-1-2 standard.
The materials which the user can come into contact have no toxicity
and no action on tissues comply with ISO10993-1, ISO10993-5 and
ISO10993-10.
13
17. Appendix A: EMC Information-Guidance and
Manufacture’s Declaration
CAUTION:
●
JPD-FR400 Infrared Thermometer needs special precautions
regarding EMC and needs to be installed and put into service according
to the EMC information provided for in the ACCOMPANYING
DOCUMENTS.
●
Portable and mobile RF communications equipment can affect
JPD-FR400 Infrared Thermometer
●
The JPD-FR400 Infrared Thermometer should not be used
adjacent to or stacked with other equipment.
Guidance and manufacturer´s declaration – electromagnetic
emission –for all EQUIPMENT AND SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic emission
The JPD-FR400 Infrared Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of the JPD-FR400 Infrared
Thermometer should assure that it is used in such an environment.
Emissions
Compliance
Electromagnetic environment - guidance
test
The JPD-FR400 Infrared Thermometer uses
RF energy only for its internal function.
RF emissions
Therefore, its RF emissions are very low and
Group 1
CISPR 11
are not likely to cause any interference in
nearby electronic
equipment.
The JPD-FR400 Infrared Thermometer is
suitabl e for use in all establishments,
including domestic establish ments and those
RF emissions
Class B
directly connected to the public low-voltage
CISPR 11
powe r supply network that supplies buildings
used for domestic
purposes.
Guidance and manufacturer's declaration – electromagnetic
immunity –for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The JPD-FR400 Infrared Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of the JPD-FR400 Infrared
Thermometer should assure that it is used in such an environment.
Immunity test
IEC
Compliance
Electromagnetic environment60601
level
guidance
test level
Electrostatic
±6 kV
±6 kV contact
Floors should be wood,
14
discharge
(ESD)
IEC 61000-4-2
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-4-8
contact
±8 kV air
±8 kV air
3 A/m
3 A/m
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Guidance and manufacturer´s declaration – electromagnetic
immunity –for EQUIPMENT and SYSTEM that are not
LIFE-SUPPORTING
Guidance and manufacturer´s declaration – electromagnetic immunity
The JPD-FR400 Infrared Thermometer is intended for use in the electromagnetic
environment specified below. The customer or the user of the JPD-FR400 Infrared
Thermometer should assure that it is used in such an environment.
Immunity test
IEC
Compliance
Electromagnetic environment 60601
level
guidance
test level
Portable and mobile RF
communications quipment
should be used no closer to
any part of the JPDFR400
Infrared Thermometer,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
3 V/m
frequency of the transmitter.
Radiated RF
80 MHz
3 V/m
Recommended separation
IEC 61000-4-3
to 2.5
distance
GHz
where p is the maximum
15
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).b
Field strengths from fixed RF
transmitters, as determined by
an
Electromagnetic site survey,a
should be less than the
compliance level in each
frequency range.b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic is
affected by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in
the location in which the JPD-FR400 Infrared Thermometer is used exceeds the
applicable RF compliance level above, the JPD-FR400 Infrared Thermometer
should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating
the JPD-FR400 Infrared Thermometer.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than
3 V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM
-for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF
communications equipment and the JPD-FR400 Infrared Thermometer.
16
The JPD-FR400 Infrared Thermometer is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the
user of the JPD-FR400 Infrared Thermometer can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the JPD-FR400 Infrared
Thermometer as recommended below, according to the maximum output power of
the communications equipment.
Rated
Separation distance according to frequency of transmitter m
maximum
150 kHz to 80 MHz
80 MHz to 800
800 MHz to 2,5 GHz
output
MHz
power
of
transmitter
0.01
0.12
0.23
0.1
0.38
0.73
1.2
2.3
10
3.8
7.3
100
12
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
17
FCC IC Caution:
This device complies with part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including
interference that may cause undesired operation.
Note: This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more of
the following measures:
-Reorient or relocate the receiving antenna.
-Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected.
-Consult the dealer or an experienced radio/TV technician for help.
Any Changes or modifications not expressly approved by the party responsible
for compliance could void the user's authority to operate the equipment.
The device has been evaluated to meet general RF exposure requirement.
The device can be used in portable exposure condition without restriction.
- English:
This device complies with Industry Canada licence-exempt RSS standard(s).
Operation is subject to the following two conditions: (1) This device may not ca
use interference, and (2) This device must accept any interference, including i
nterference that may cause undesired operation of the device.
- French:
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux
appareils radio exempts de licence. L'exploitation est autorisée aux deux
conditions suivantes:
(1) l'appareil ne doit pas produire de brouillage, et
(2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi,
même si le brouillage est susceptible d'en compromettre le fonctionnement.
Shenzhen Jumper Medical Equipment Co., Ltd.
Address: D Building, No. 71, Xintian Road, Fuyong Street,
Baoan,Shenzhen,Guangdong,China
Tel: +86-755-26692192, 26696279
Fax: +86-755-26852025
Website:www.jumper-medical.com
Postal Code: 518102

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 22
Language                        : zh-CN
Tagged PDF                      : Yes
Title                           : EAR THEMOMETER LEAFLET A-W
Author                          : 01
Creator                         : Microsoft® Office Word 2007
Create Date                     : 2017:08:24 02:42:56
Modify Date                     : 2017:08:24 02:42:56
Producer                        : Microsoft® Office Word 2007
EXIF Metadata provided by EXIF.tools
FCC ID Filing: 2ADYL-JPDFR400

Navigation menu