ShenZhen Jumper Medical Equipment JPDHA121 Electronic Blood Pressure Monitor User Manual

ShenZhen Jumper Medical Equipment Co.,Ltd Electronic Blood Pressure Monitor

User Manual

JPD-HA120Electronic Blood Pressure MonitorUser ManualVersion: V1.0Revision Date: May 2018
ForewordDear user:Thank you for purchasing our Electronic Blood PressureMonitor. Please read the User Manual carefully and usecorrectly. Please keep the User Manual properly for checkingand reference at any time. The Electronic Blood PressureMonitor is suitable for adults .Product Profile:The product is called Electronic Blood Pressure Monitor,comprised of the host machine and the cuff, suitable formeasuring the systolic pressure, diastolic pressure and pulse ofhuman body. The blood pressure value measured by theElectronic Blood Pressure Monitor is equivalent to thatmeasured by UOTT.By use of the intelligent pressurization method, theElectronic Blood Pressure Monitor can automatically pressurizeto a suitable pressure value based on your blood pressure, thuseffectively reducing the discomfort caused by incorrectpressurization, shortening the measurement time andprolonging the service life of cuff.The Electronic Blood Pressure Monitor can provide thememory function of 99 groups of measurement data from eachof two users, and can save the data of two users respectively.
ContentsFCC Statement........................................................................... - 1 -1. Open-box Inspection.............................................................. - 2 -2. Packing List............................................................................- 3 -3. Authorized European Representative:.................................... - 3 -4. Safety Precautions.................................................................. - 4 -5. Product Composition..............................................................- 5 -6. Application Scope...................................................................- 5 -7. Product Components.............................................................. - 5 -8. Pre-use Installation.................................................................- 7 -9. Function Setting Mode........................................................... - 9 -10. Correct Method of Use....................................................... - 13 -11. General Knowledge of Blood Pressure.............................. - 20 -12. Contraindications, Precautions, Warnings and PromptInstructions............................................................................... - 21 -13. Question and Answer..........................................................- 28 -14. Abnormal Phenomena and Handling..................................- 32 -15. Cleaning and Disinfection..................................................- 34 -16. Upkeep and Maintenance................................................... - 36 -17. Precautions for Keeping..................................................... - 37 -18. Product Specifications........................................................- 38 -
- 1 -FCC StatementThis device complies with part 15 of the FCC Rules. Operationis subject to the following two conditions:(1) This device may not cause harmful interference, and(2) This device must accept any interference received, includinginterference that may cause undesired operation.Notes:This equipment has been tested and found to comply with thelimits for a Class B digital device, pursuant to part 15 of theFCC Rules. These limits are designed to provide reasonableprotection against harmful interference in a residentialinstallation. This equipment generates, uses and can radiateradio frequency energy and, if not installed and used inaccordance with the instructions,may cause harmfulinterference to radio communications.However, there is noguarantee that interference will not occur in a particularinstallation. If this equipment does cause harmful interferenceto radio or television reception, which can be determined byturning the equipment off and on, the user is encouraged to tryto correct the interference by one or more of the followingmeasures:
- 2 -■ Reorient or relocate the receiving antenna.■ Increase the separation between the equipment and receiver.■ Connect the equipment into an outlet on a circuit differentfrom that to which the receiver is connected.■ Consult the dealer or an experienced radio/TV technician forhelp.This equipment complies with FCC radiation exposure limitsset forth for an uncontrolled environment. This equipmentshould be installed and operated with a minimum distance of20cm between the radiator & your body. This transmitter mustnot be co-located or operating in conjunction with any otherantenna or transmitter.Caution: Any changes or modifications to this device notexplicitly approved by manufacturer could void your authorityto operate this equipment.1. Open-box InspectionBefore use, please unpack carefully, check whether all theaccessories are intact and whether the components are damageddue to transportation against the following packing list, andperform the installation and operation in strict accordance with
- 3 -the User Manual. In case of any damage or operating problem,please contact our vendor or directly contact our company. Atthe time of contact, you shall provide the purchased equipmentmodel, serial number, your purchase date as well as yourcontact information and address.2. Packing ListNo.NameQuantity1Electronic Blood Pressure Monitor12Cuff13Dry battery (AA)44User Manual15Jumper Health User manual16Pouch13. Authorized European Representative:Wellkang LtdSuite B, 29 Harley Street , London W1G9QR,UK
- 4 -4. Safety PrecautionsThe warnings and illustrations shown in the User Manualenable you to use the product safely and correctly, thuspreventing you and others from being injured, specifically asfollows:Legend, mark and meaningWarning messageAnti-electric shock degree is Type BF of theapplication partWhen the product life expires and the end usersdiscard the products, send them to the designatedcollecting and separating place for disposalaccording to the requirements from the localenvironmental protection authority.Non-ionizing radiationConsult the instructions for use.
- 5 -BluetoothThis product complies with the MDD93/42/EECrequirements.5. Product CompositionThe product is comprised of the host machine and the cuff ofElectronic Blood Pressure Monitor.6. Application ScopeThe Electronic Blood Pressure Monitor is used to measure thediastolic pressure, systolic pressure and pulse rate of adults byoscillometric method for diagnostic reference.7. Product Components(1) Device Body
- 6 -
- 7 -(2) Display Screen8. Pre-use Installation(1) Install the batteryOpen the battery cover and load thebattery into the battery compartment.The battery compartment is located atthe bottom of the Electronic Blood Pressure Monitor.
- 8 -a)Open the battery cover as shown in the figure (as shown inthe figure on the top right)b)Place 4 AA dry batteries. Pay attention to the batteryelectrode indication.(2) Battery power indication and replacementAfter the product is turned on, if low power symbol appears onthe screen , the measurement cannot be performed, and thebattery must be replaced.Please use 4 AA alkaline dry batteries.Do not use any expired battery.If the product is not used for a long time(over 3 months),please take out the batteries.(3) USB connection for power supplyUSB can be inserted for power supply of the product withoutbattery.
- 9 -(4) CuffThe applicable arm circumference range of the cuff is 22-36cm.Connection: Insert the air plug on the cuff into the air plug jackon the left side of the Electronic Blood Pressure Monitor. Asshown in the following figure:-The Cuff is the application part.JackRubber air guide pipeCuffAir plug9. Function Setting Mode(1) User modeSelect "OFF" or upon completion of measurement, short-pressthe "User" key, and the screen will display "User 1" or "User 2".
- 10 -Press the "User" key again to switch to another user group.(2) Year/Month/Date settingIn the "OFF" state, long-press the "User" key for 3~5s to enterthe setting mode.After entering the "Year" setting mode, the screen will display20XX and keep flashing. Press the "Memory" key, and thevalue will be incremented. The value can be switched between2018~2099. Press the "User" key for confirmation and enter the"Month" setting mode.After entering the "Month" setting mode, the screen willdisplay 1/01 and keep flashing. Press the "Memory" key, andthe value will be incremented. The value can be switchedbetween 1~12. Press the "User" key for confirmation and enterthe "Date" setting mode.
- 11 -After entering the "Date" setting mode, the screen will display1/01 and keep flashing. Press the "Memory" key, and the valuewill be incremented. The value can be switched between 01~31.Press the "User" key for confirmation and enter the "Time"setting mode.After entering the "Time" setting mode, the screen will displayXX: XX and keep flashing. Press the "Memory" key, and thevalue will be incremented. The value can be switched between1~12 am/pm. Press the "User" key for confirmation and enterthe "Minute" setting mode. Press the "Memory" key, and thevalue will be incremented. The value can be switched between00~59. Press the "User" key for confirmation and enter the"Language Broadcast Switch" setting.
- 12 -(3) Language Broadcast Switch settingAfter entering the "Language Broadcast Switch" setting mode,the high-pressure area will display the "SP", and thelow-pressure area will display the current voice state"ON"/"OFF" and the voice icon (not shown if turned off). Pressthe "Memory" key to switch the language state. Press the"User" key for confirmation and saving, and enter the "UnitDisplay" setting.(4) Unit display settingAfter entering the "Language Broadcast Switch" setting mode,press the "User" key, and the screen will display 0 or 0.0, i.e.enter the setting mode. If display of 0, it indicates that themmHg unit is selected; if display of 0.0, it indicates that the kPaunit is selected. At this time, when you press the "Memory" keyonce, the unit will be switched once. After the unit is selected,
- 13 -press the "User" key again for confirmation and saving.10. Correct Method of UseRequirement:Please read the user manuals carefullybefore use , user must follow user manuals to use.(1) Measure the blood pressure - preparation beforemeasurement- Before measurement, avoid eating, smoking or any form ofheavy physical labor, as it will affect the measurement result.- Before measurement, sit on a chair to have a rest for about10min in a quiet environment.- Before measurement, remove the clothing fettering themeasured arm.- Measure the same arm (generally the left arm).- Perform fixed-time measurement daily, because the bloodpressure on the same day is also changing.
- 14 -(2) Measure the blood pressure - common factors resultingin measurement error- Comparative measurement of blood pressure must beperformed under the same condition! Generally perform themeasurement in a quiet environment!- The blood pressure might be increased due to forced supportof arm; therefore, please ensure that you are in a comfortableand relaxed state before measurement, and do not move themuscle of the measured arm during measurement. Ifnecessary, you can use a soft cushion to support your arm.- If the position of cuff is lower or higher than the heart position,you will get the incorrect measurement results.(3) Measure the blood pressure - attentions- Please use the original cuff that has passed the clinical trial!Any replacement of original component with the componentnot provided by the manufacturer might result inmeasurement error.- Cuff looseness or air bag exposure might result inmeasurement error.- Repeated measurement might result in blood aggregation inthe arm, which will affect the measurement result. Therefore
- 15 -you should have a rest for 3min or lift the arm for a while tomake the aggregated blood flow away before repeatedmeasurement.(4) Measure the blood pressure - how to use the cuff1. Place the cuff flat on the table, leave the hook & loopdownwards, and pass the end of cuff through the metal ringto form a loop. At this time, the hook & loop of cuff willpoint outwards (if the cuff has formed a loop before, pleaseskip this step).2. Pull the cuff through the upper arm to be measured, and wearthe cuff correctly based on the downward icon  , leavingthe air jet of cuff to point to the lower arm.3. Hook up the cuff on the upper arm according to theillustration, ensure that the lower edge of cuff is 2~3cm awayfrom the elbow joint, and leave the air jet inside the arm.4. Tighten the free edge of cuff to stick the hook & loop.5. The cuff should twine round the upper arm comfortably,where you should be able to put two fingers in between them.Before measurement, you should remove any clothing thatmight confine the arm (e.g. overpull).6. Place the lower arm flat on the desktop, leaving the center of
- 16 -palm naturally upwards, sit straight, feet flat on the groundand leave the center of cuff and the heart at the same height.Note that the rubber catheter on the cuff cannot be folded orbent.Note: If you cannot use the left arm for measurement, pleaseuse the right arm for measurement. But anyway, all themeasurements must be performed on the same arm forcomparison.
- 17 -(5) Start measurementMeasure the blood pressure -- measurement stepsAfter you wear the cuff, you can start the measurement:1. Press the "ON/OFF" key, and the machine will return to zeroautomatically, the air pump will start to inflate the cuff, andthe screen will display the change in the pressure in the cuff.2. When reaching the stable pressure upon inflation, the airpump will stop the inflation, and the pressure in the cuff willbe reduced gradually and displays in the screen. If theinflated pressure is insufficient, the machine will re-inflateautomatically to a higher pressure;3. When pulse is detected, the screen will display the "heart"symbol and start flashing.4. Upon completion of measurement, the measured values ofsystolic pressure, diastolic pressure and pulse will bedisplayed on the screen.5. The screen will continue to display the measurement results,unless you long-press the "ON/OFF" key to turn off themachine. If there is no operation, the machine will be turnedoff automatically in 30s.
- 18 -Notes:At the end of measurement, the measurement results andinformation (high pressure, low pressure, heart rate,measurement completion time, current user group andwhether arrhythmia) will be automatically saved to thememory. The number of groups is 01, and the previousmemory group number will be extrapolated back insequence.The default measurement result display unit is mmHg.In case of error in the measurement result, the high-pressurearea will display the "Err", and the low-pressure area willdisplay the value.During measurement, if during the voice broadcast, you canpress the "Memory" key to skip the voice. If you press the"Memory" key during the inflation/automatic deflation, therewill be no reaction.At the end of measurement, if during the voice broadcast,you can press the "Memory" key to stop the voice.During measurement, if you short-press the "ON/OFF" key,the measurement will be stopped and the Electronic BloodPressure Monitor will return to the home page; if youlong-press the "ON/OFF" key for 3s, the Electronic Blood
- 19 -Pressure Monitor will be turned off.The turn-off time is started from the value display uponcompletion of measurement (no voice) or from the valuedisplay and voice broadcast upon completion ofmeasurement (voice).Upon completion of measurement, short-press the "Memory"key to switch to the memory mode, and short-press the"User" key to perform the user group switching.(6) Use Bluetooth functionSee the User manual of Jumper Health.(7) Use memory functionThe Electronic Blood Pressure Monitor can automatically store99 groups of memory values of the last two persons, and if the99 groups of memory data are full, they will be replacedautomatically. After the Electronic Blood Pressure Monitor isused for several times, in the OFF state, you can press the"Memory" key to display the recent measurement, and press the"Memory" key again to display the another recent measurementagain and so on until display the last measurement(8) Clear all memoryIn the memory mode, long-press the "Memory" key for 3s todelete all the measurement data of two user groups.
- 20 -11. General Knowledge of Blood Pressure(1) General knowledge of blood pressure - definition ofblood pressureBlood pressure is the pressure that the blood is transmitted tothe artery through the heart beat (systole and diastole).When systole, the pressure at which the blood is transmitted tothe artery is called systolic pressure, i.e. the high pressure; theblood circulating the whole body will backflow to the heart,and when diastole, the pressure is called diastolic pressure, i.e.the low pressure.
- 21 -(2) Classification standard for blood pressure conditionThe standard blood pressure of WHO is as shown in the figurebelow.- There is no definition of hypotension yet. Generally, if thehigh pressure is less than 100mmHg (13.3kPa), it ishypotension.12. Contraindications, Precautions, Warnings andPrompt Instructions(1) Warning:Do not let a child below 12 years old and the people who
- 22 -can’t express one’s intention .When it used by the people of12~18 years old ,it should accompanied by the adult .Pregnantwomen use under the guidance of a doctor.Make sure there is no connection tubing kinking beforestart measuring to avoid can not be measured or other injury.For any patient, do not measure more than 3 timescontinuously, it should be at least above 5 minutes of intervalrest between any two measurements,otherwise will causeextravasated blood.Do not apply the cuff over a wound as this can causefurther injury.Do not use cuff in the arm which has invasive treatmentdevice or arteriovenous shunt,otherwise will cause hurt.When the arm use the cuff.,it will make the monitoringequipment failure.Pay attention to the widget, do not swallow it.Please use the cuff follow the use manual,pay attention to
- 23 -the use of the hose,avoid necking due to excessive lengthDo not use the cuff for a long time,avoid allergies.If an incorrect reading appears, check the instructions andcheck for equipment failureDo not make self-diagnosis according to the measurementresults. Please consult your professional doctor with themeasurement result record. The treatment based on theself-judgement of measurement results will be verydangerous;Please give the results of the measurement to the doctorwho knows your health and receive the diagnosis;If you take some medicine, please follow the doctor'sadvice;For those with peripheral circulation disorder caused bycertain diseases, there might be significant difference inthe blood pressure value measured on the wrist and upperarm;Do not make measurements in 1h after meals orimmediately after strenuous exercise, shower, smoking,alcohol drinking, coffee drinking or black tea drinking;Do not use the Electronic Blood Pressure Monitor for thepurpose other than blood pressure measurement;
- 24 -otherwise, it might cause accident or failure;Please use the Electronic Blood Pressure Monitor underthe guidance of doctor;This model of Electronic Blood Pressure Monitor issuitable for the arm circumference range of 22-36cm, andif beyond this range, you might not obtain the correctmeasured value of blood pressure;This model of Electronic Blood Pressure Monitor is notsuitable for newborns or young children;The blood pressure is constantly changing. You cannotrely on only one blood pressure measurement result tojudge the blood pressure condition. The repeatedmeasurement data over a period of time will be morereliable;Do not bind too tightly. It will be the most appropriate ifyou can relax and can put one finger in. If you do not bindcorrectly or put it in place, you might not obtain thecorrect measured value of blood pressure;Do not measure the non-invasive blood pressure (NIBP)for the patients with sickle cell disease or the patientswith skin injury of limb for blood pressure measurement.
- 25 -Please do not use mobile phone, computer,electric kettleand the other can cause interference device around thedevice.Please do not use the device around the magneticfield.Please keep to maintain the device ,during use it.(2) Precautions:Do not repair, disassemble or modify the Electronic BloodPressure Monitor without permission;Do not collide or fall down the host to avoid collision orstrong impact of the machine;Do not use the Electronic Blood Pressure Monitorimmediately after it is taken out from the place below 0℃which should be at least placed for more than 1h in theenvironment at the temperature of 5 ℃-40 ℃and thehumidity of 85%;When the ambient temperature is 20°C, the time from thehighest storage temperature 55°C ,after use to the timeabout 1 hour or more that the equipment is ready for itsintended use.There is no need to check whether the operation of the
- 26 -sphygmomanometer will have a lasting effect on thepatient's blood circulation,if used according to theinstructionsDo not use the Electronic Blood Pressure Monitorimmediately after it is taken out from the place below 0℃,which should be at least placed for more than 1h in theenvironment at the temperature of 5 ℃-40 ℃and thehumidity of 85%;Except the cloth stained with warm water or soapy waterfor wiping the host shell, do not use other articles forwiping;Do not pull the cuff forcefully or use the cuff as a rag;Do not be contaminated by the dust or corrosive liquid;Do not touch the host when measurement;Remove the battery when the product is left unused for along time;Do not store the Electronic Blood Pressure Monitor in aplace with high temperature, moisture or direct sunlight;Calibrate the Electronic Blood Pressure Monitorperiodically.Do not combine old and new batteries,different brands
- 27 -batteries for using.(3) Indicative descriptionThe patient is an intended operator.Before measurement, calibrate the Electronic BloodPressure Monitor, check the function of each component,and maintain the good performance of Electronic BloodPressure Monitor.During blood pressure measurement, the measured limbshall be at the same level at the heart.Indoor temperature, environment, noise, subject's bodyposition, speech or exercise might have effect on theblood pressure measurement. The cuff wearing, impropersensor position, body movement or presence of electricfield might have effect on the measurement.We do not recommend that patients with weak pulse,severe upper gastrointestinal bleeding or severearteriosclerosis or elderly with poor peripheral circulation,particularly patients with peripheral vascular lesioncaused by complicated severe diabetes use the product.In case of rescue of seriously ill patients, diagnosis ofhypertension and direction of hypertensive medication,
- 28 -use the Electronic Blood Pressure Monitor with caution.Keep the power supply enough.After product packaging, it is allowed to use commonvehicle transportation, but rain, moisture, squeezing ormechanical collision should be avoided.13. Question and AnswerQ1. Why the blood pressure measured at home is lowerthan that measured in hospital?1. The blood pressure measured at home might be20mmHg-30mmHg (2.7kPa-4.0kPa) lower than thatmeasured in hospital, because you might be nervous whenmeasurement in hospital, and you tend to be in stable moodwhen measurement at home. It is important to understandthe normal value of blood pressure at home whencalmness.2. If the cuff wearing position is higher than the heart, themeasured blood pressure value will be lower, so pleasemake measurement in the correct posture.Q2. Why the blood pressure measured at home is higherthan that measured in hospital?1. The anti-hypertensive drugs lose the efficacy.
- 29 - Please accept the doctor's instructions.2. The cuff is not twined correctly.If the cuff position is wrong, no arterial signal will be caught,and the measured blood pressure value will be higher.Please confirm whether the cuff position is correct.3. The cuff is rolled too loose.If the cuff is rolled too loose, the compression force mightfail to be transmitted to the artery, resulting that the measuredblood pressure value is higher than the actual value. Adjust the gap between the arm and the cuff, and tighten thecuff.4. The posture is incorrect when measurement.If bending down, sitting cross-legged or bending over thesofa or low table when measurement, the blood pressuremight be increased due to the abdominal pressure or due tothe arm position lower than the heart. Please make measurement in correct posture.Q3.Why I feel painful or numb due to tightening of the cuffwhen measurement?When blood pressure measurement, the cuff shall be tighten
- 30 -until the arterial blood flow is stopped temporarily. So youmight feel painful or numb for a while, but it is not harmful foryour body. Please be assured.Q4. When will the measurement be better?Please make measurement in the morning after you get up tourinate or when your body and mood are in stable state. It isbetter to make measurement at the same time every day.Q5.Why the blood pressure value measured each time isdifferent?1. When systole each time, the blood pressure will change tosome extent. For example, a person with the pulse of 70 beatsper minute will have 100,800 blood pressure changes every day.Because the blood pressure is constantly changing, it is difficultto obtain the correct blood pressure value by measurement onlyonce. Please make measurement for 2~3 times. The firstmeasurement will generally be higher due to nervousness orinadequate preparation, and then when the second measurement,the nervous emotion will be slightly alleviated, so generally, thesecond measurement will be 5mmHg-10mmHg (0.7kPa-1.3kPa)lower than the first time. This will be more obvious for those
- 31 -with higher blood pressure.- When continuous measurement, please note that:There might be extravasated blood because the arm iscompressed, resulting that the finger tip blood does not flowsmoothly, If you continue the measurement in case ofextravasated blood, you cannot obtain the correct measuredvalue. Loosen the arm band, raise your hand over the head,and grasp and stretch your left and right palms for 15 timesrepeatedly. Then the extravasated blood can be dissolved,and you can continue the blood pressure measurement.2. Cuff position and twining method. The measured valuevaries with the cuff size. Particularly, if the cuff is twined roundthe elbow, you cannot obtain the correct measured value.- Please use the correct cuff twining method for measurement.The arm circumference range of the enclosed cuff is22~36cm (center of the upper arm).If the model isinconsistent, please purchase separately.
- 32 -14. Abnormal Phenomena and Handling* When abnormal measurement, the following symbols willappear. Please use the correct method for measurement.LCD Display InformationReason/meaningSBP Display shows Err,DBP Display shows 1Sensor exception.SBP Display shows Err,DBP Display shows 2No pulse detectedSBP Display shows Err,DBP Display shows 4Air leakage is detected in theearly stage of decompressionSBP Display shows Err,DBP Display shows 5Trachea is obstructed asdetected in the early stage ofpressurizationSBP Display shows Err,DBP Display shows 6There is too muchinterference as detectedwhen deflationSBP Display shows Err,DBP Display shows 7Pressure value exceeds295mm (40.0 kPa)The high-pressure display areadisplays Err and malfunction iconArm movement
- 33 -* TroubleshootingAnomalyInspection ItemsCountermeasuresThere is nodisplay after thepower switch ispressedWhether the batterypower is insufficientReplace the battery orinsert the USB forchargingWhether positiveand negative polesof the battery areinstalled reverselyInstall in the positiveand negative poledirection of batteryNo pressurizingWhether the air tubeplug is insertedtightlyInsert the air tube plugfirmly into the jackWhether the air tubeis broken or leaked?Purchase a new cuffUnable tomeasure due todisplay errorWhether the arm ismoved whenpressurizationKeep your arm andbody still staticWhether you talkduring measurementKeep quiet whilemeasuring the bloodpressureAir leakage ofcuffWhether the cuff istwined too loosePlease tighten the cuffWhether the air bagof cuff is rupturedReplace with new cuffIf the blood pressure still cannot be measured normally bythe above methods, please contact the vendor. Please do notdisassemble it by yourself!
- 34 -15. Cleaning and Disinfection(1) Cleaning:Recommended cleaning agent:Medical cleaning agentHousehold neutral cleaning agentCleaning steps:1)Remove the batteries before cleaning.2)The body of Electronic Blood Pressure Monitor can becleaned with a clean soft cloth.3)The body can be wiped with a clean soft cloth stained with asmall amount of neutral detergent or water.Do not use any corrosive cleaning agent. Whencleaning, be careful not to immerse any part of theElectronic Blood Pressure Monitor to avoid liquid flowinto the instrument.It is suggested to clean once every week. Complete thecleaning in 3min each time. The number of repeatedcleaning each time shall not exceed 3 times.(2) Disinfection:Recommended disinfecting agent:Isopropanol solution with 70% concentrationMedical alcohol with 75% concentration
- 35 -Sodium hypochlorite solution with 3% concentrationDisinfection steps:1)Wipe the machine body with a clean soft cloth stained with asmall amount of the above disinfectant, and dry itimmediately.2)The body of the Electronic Blood Pressure Monitor can bewiped with a clean cloth stained with a small amount of75% medical alcohol for disinfection.Do not disinfect by such method as high temperaturesteam or ultraviolet irradiation, which might damagethe instrument or accelerate the aging!It is suggested to disinfect the Electronic BloodPressure Monitor before and after use each time. Eachtime of disinfection shall be completed within 1min.The number of repeated disinfection each time shallnot exceed 2 times.Cleaning and disinfection shall be carried out in thefollowing environment: temperature: +5℃~+40℃(50-104 ), relative humidity: 15%~85%RH,non-condensing, atmospheric pressure: 70kPa~106kPa.
- 36 -16. Upkeep and Maintenance- Always keep the surface of Electronic Blood Pressure Monitorclean and tidy, helpful to prolong the service life of ElectronicBlood Pressure Monitor.- If the host is dirty, please wipe with a dry soft cloth. If the dirtcannot be eliminated easily, wipe with a soft cloth stained withwater or neutral detergent, and then dry with a dry cloth.-No maintenance or maintenance required when using thedevice.Do not allow water or other liquids to flow into the host.
- 37 -17. Precautions for Keeping- Do not wipe the host and cuff withvolatile oil, diluent or gasoline.- Do not clean or moistenthe cuff.-Do not store the product at a placewhich has direct sunlight, hightemperature, humidity, dust orcorrosive gas.- If the Electronic BloodPressure Monitor is not inuse for a long time (over 3months), take out of thebatteries from the host.The Company will not assume any quality responsibility for themachine failure due to your non-compliance with the aboveconsiderations for safekeeping or other correct use methods.
- 38 -18. Product SpecificationsProduct NameElectronic Blood Pressure MonitorModelJPD-HA120Display ModeDigital display modeMeasuring ModeOscillographic methodMeasuring BodyPartUpper armMeasuring RangeBloodpressurevalue0-295mmHg(0kPa-39.2kPa)Pulse Count40-199 pulse beats/minStaticmeasurementaccuracyPressurevalue±3mmHg(±0.4kPa)Pulse Count± 5% of read valueLCD displayPressureUnit : mmHg/ kPaPulsePulse beats per minute,displaying three digitsStorage capacityIt can store 99 groups of measurement datafor each of two personsPower supply4 AA dry batteries/DC6VPower Off ModeManual closing/ 30s delay automaticclosing
- 39 -Device weight(without batteries)About 285gMonitor Size138mm (length) *120mm (width) * 59mm(height)Screen Size60mm (length) *77mm (width)CuffUpper-arm-type cuff(measured arm circumference 22-36cm)AnnexureCuff, User Manual, Jumper Health Usermanual, Dry battery,PouchBattery LifeHigh-performance dry battery can be usedfor about 300 times at normal temperatureService Life5 yearsDate of ProductionSee labelOperatingenvironmentTemperatureCondition5℃-40℃If stored orused beyondthedesignatedtemperatureandhumidityrange, it willnot be usedHumidityCondition15%-85%RHAtmosphericCondition70kPa-106kPa
- 40 -properlyTransportation andStorageEnvironmentAvoid strong impact, direct impact,exposure or rain during transportation. Thepackaged Electronic Blood PressureMonitor shall be stored indoors at thetemperature of -20℃~55℃and the relativehumidity of 15%~93% without corrosivegas and with good ventilation.Appendix 1 Guidance and Manufacturer Declaration TablesGuidance and manufacturer´s declaration – ElectromagneticemissionThe Electronic Blood Pressure Monitor is intended for use in theelectromagnetic environment specified below.The customer orthe user of the Electronic Blood Pressure Monitor should assurethat it is used in such an environment.EmissionsComplianceElectromagnetic environment -guidanceRFemissionsCISPR 11Group 1The Electronic Blood PressureMonitor uses RF energy only forits internal function. Therefore,
- 41 -its RF emissions are very low andare not likely to cause anyinterference in nearby electronicequipment.RFemissionsCISPR 11Class BThe Electronic Blood PressureMonitor is suitable for use in allestablishments,includingdomestic establishments andthose directly connected to thepublic low-voltage power supplynetwork that supplies buildingsused for domestic purposes.HarmonicemissionsIEC61000-3-2N.A.Voltagefluctuations/flickeremissionsIEC61000-3-3N.A.
- 42 -Guidance and manufacturer´s declaration – ElectromagneticimmunityThe Electronic Blood Pressure Monitor is intended for use in theelectromagnetic environment specified below.The customer orthe user of The Electronic Blood Pressure Monitor should assurethat it is used in such an environment.ImmunitytestIEC60601testlevelCompliance levelElectromagneticenvironment-guidanceElectrostaticdischarge(ESD)IEC61000-4-2±8kVcontact±2 kV,±4kV,±8 kV,±15KVair±8 kVcontact±2 kV,±4 kV,±8 kV,±15 KVairFloors should be wood,concrete or ceramic tile.If floors are coveredwith synthetic material,the relative humidityshould be at least 30 %.Powerfrequency(50/60 Hz)magnetic30A/m,50/60Hz30A/m,50/60HzPower frequencymagneticfields should be atlevels
- 43 -fieldIEC61000-4-8characteristic of atypicallocation in a typicalcommercial or hospitalenvironment.NOTE: UT is the a.c. mains voltage prior to application of thetest levelGuidance and manufacturer´s declaration – ElectromagneticimmunityThe Electronic Blood Pressure Monitor is intended for use in theelectromagnetic environment specified below. The customer or theuser ofThe Electronic Blood Pressure Monitor should assure that itis used in such an environment.ImmunitytestIEC60601testlevelCompliancelevelElectromagneticenvironment-guidanceRadiated RFIEC61000-4-33Vrms150 kHzto 80N/APortable and mobileRFcommunicationsequipment should be
- 44 -MHz6 Vrms150kHz to80MHZoutsideISMbandsaused no closer to anypart of the JPD-HA120Electronic BloodPressureMonitor ,includingcables, than therecommendedseparation distancecalculated from theequation applicable tothe frequency of thetransmitter.Recommendedseparation distancePVd15.3Radiated RFIEC61000-4-310 V/m80 MHzto 2.7GHz10 V/mpEd15.380MHz to800MHzPEd17800MHz to2.7GHz
- 45 -where P is themaximum outputpower rating of thetransmitter in watts(W) according to thetransmittermanufacturerand d is therecommendedseparation distance inmetres(m).Field strengths fromfixed RFtransmitters,asdetermined by anelectromagnetic sitesurvey, ashould be lessthan thecompliance level ineach frequency range bInterference may occurin the vicinity
- 46 -of equipment markedwith the followingsymbol:NOTE 1 At 80 MHz and 800 MHz, the higher frequency rangeapplies.NOTE 2 These guidelines may not apply in all situations.Electromagnetic is affected by absorption and reflection fromstructures, objects and people.a The ISM (industrial, scientific and medical) bands between 0,15MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to13,567 MHz; 26,957 MHz to 27,283MHz; and 40,66 MHz to40,70 MHz. The amateur radio bands between 0,15 MHz and 80MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to5,4 MHz, 7MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to14,2 MHz, 18,07 MHz to 18,17MHZ, 21,0 MHz to 21,4 MHz,24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHzto 54,0 MHz.b The compliance levels in the ISM frequency bands between 150kHz and 80MHz and in the frequency range 80 MHz to 2,7 GHz are intended
- 47 -to decreasethe likelihood that mobile/portable communications equipmentcould causeinterference if it is inadvertently brought into patient areas. Forthis reason, anadditional factor of 10/3 has been incorporated into the formulaeused incalculating the recommended separation distance for transmittersin thesefrequency ranges.c Field strengths from fixed transmitters, such as base stations forradio (cellular/cordless) telephones and land mobile radios,amateur radio, AM and FM radio broadcast and TV broadcastcannot be predicted theoretically with accuracy.To assess theelectromagnetic environment due to fixed RF transmitters, anelectromagnetic site survey should be considered. If the measuredfield strength in the location in which the Electronic BloodPressure Monitor is used exceeds the applicable RF compliancelevel above, the Series Electronic Blood Pressure Monitor shouldbe observed to verify normal operation. If abnormal performanceis observed, additional measures may be necessary, such asre-orienting or relocating the Electronic Blood Pressure Monitor.
- 48 -d Over the frequency range 150 kHz to 80 MHz, field strengthsshould be less than 3 V/m.Recommended separation distances between portable andmobile RF communications equipment and ElectronicBlood Pressure MonitorThe Series Electronic Blood Pressure Monitor is intended foruse in an electromagnetic environment in which radiated RFdisturbances are controlled. The customer or the user of theSeries Electronic Blood Pressure Monitor can help preventelectromagnetic interference by maintaining a minimumdistance between portable and mobile RF communicationsequipment (transmitters) and the Series Electronic BloodPressure Monitor as recommended below, according to themaximum output power of the communications equipment.RatedmaximumoutputpoweroftransmitterWSeparation distance according to frequencyof transmitterm150 kHz to80 MHzPVd15.380 MHz to800MHz800 MHz to2.7 GHzPEd17
- 49 -pEd15.30.010.120.120.230.10.380.380.7311.21.22.3103.83.87.3100121223For transmitters rated at a maximum output power not listedabove, the recommended separation distance d in metres (m)can be estimated using the equation applicable to the frequencyof the transmitter, where P is the maximum output power ratingof the transmitter in watts (W) according to the transmittermanufacturer.NOTE 1 At 80 MHz and 800 MHz, the separation distance forthe higher frequency range applies.NOTE 2 These guidelines may not apply in all situations.Electromagnetic propagation is affected by absorption andreflection from structures, objects and people.
Shenzhen Jumper Medical Equipment Co., Ltd.Address: D Building, No. 71, Xintian Road,Fuyong Street, Baoan, Shenzhen ,Guangdong,ChinaTel: +86-755-26692192, 26696279Fax: +86-755-26852025Website:www.jumper-medical.comPostal Code: 518103