Sonova USA WFT1 Wireless Functional Test Device (WFT) User Manual

Phonak Inc Wireless Functional Test Device (WFT)

User Manual

 Work Instruction / User Guide Wireless Functional Test DeviceFile: 000.000.015.286.pdf Version 2Page: 1 of 15 Number of local copies:Version: 1.00Written: 2015-04-10 11bwestcottDate of print: Uncontrolled copy if printed Approved: 2015-04-20 1bkrajnovicContact: technical.support@phonak.com Released: 2015-04-21 11zbudrovicHealth and SafetyTable of ContentsGeneral Description1Wireless Functional Test Device2Wireless Functional Test Setup3Wireless Functional Test Device Self Test4Wireless Functional Test Device Test Modes5Wireless Functional Test Device Operation6Compliance Information7Information and Description of Symbols8Important Safety InformationRevision HistoryElectrostatic discharge sensitive devicesTake ESD precautionary measures whenever handling electronic devicesWorkplace Ergonomics:Adjust table and chair to elbow height.Elbow height = table The angle between your thigh and lower leg should be at least 90°.
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 2 / 15The Wireless Functional Test (WFT) device is used to perform a wireless functional quality test (Go/NoGo test) of the proprietary HIBAN link (10.6 MHz) by generating an audio stream from the WFT device to a hearing instruments or compatible Digital Wireless Accessories (DWA).WFT device interacts as a simple remote control and can be used to test remote control functionality by decreasing the volume and changing the program of a hearing instrument. Following platforms are supported:•  P3:  Palio 3 Platform with a new chip realizing the HiBAN 2 with 8-DPSK (DQPSK) modulation.   All hearing aids starting by Audeo V and Digital Wireless Accessories of the “II” generation.•  P2:  Palio 2 Platform with a chip realizing the HiBAN 1 with F1D / FSK  modulation. All wireless hearing aids till Audeo V.Please read this user guide carefully to benefit from all the features of the WFT.The WFT device is designed for trained and certified custom product manufacturing and repair centers within Phonak and Unitron Group Companies (GCs) and by qualified distributors.The WFT device is used for performing a functional test of the HIBAN link of a Device Under Test (DUT). The WFT device is not capable to program nor manipulate any setting of hearing instrument or a DWA. The WFT does not give pass/fail output. Rating needs to be done by operator.PurposeThis document is an outline of the user interface menu for the Wireless Functional Test (WFT) of the WFT device. This test is compatible with both Palio 2 (P2) & Palio 3 (P3) hearing instruments (HI) & Palio 3 Digital Wireless Accessories (DWA).  There are different tests available for the two platforms due to the wireless processor used. Palio 2 HI remote control test & the Palio 3 remote control test & wireless functional test for both P3 HI & P3 DWA. The HI will always have a side designation, left or right.  The WFT device is able to test the HI without the operator selecting a specific side.  The WFT device will automatically connect to a DWA under test regardless of the fitting side status, left or right, or binaural.General DescriptionIntended UseWireless Functional Test
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 3 / 15Power on WFT DevicePower off WFT DeviceSelect PlatformPlatform change needed for next orderAll devices testedPalio 3Select Test ModeP3 HI Select Test ModeP3 DWASelect Test ModeHI Passed TestP3 CROSSelect Test ModePalio 2HI Remote Control Test Palio 3 Hearing Instrument TestP3 DWA Remote Control TestStart HI TestP3 DWA Wireless Functional TestEnd HI Test YesNoNoReturn to Select Platform menu for the next testTest the next deviceHI needs repair action.  May be kicked out to troubleshootingP3 HI Remote Control TestP3 HI Wireless Functional TestPalio 3 CROS TestPalio 3 Digital Wireless Accessory TestP3 CROS Remote Control TestP3 CROS Wireless Functional Test
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 4 / 15P2 & P3 HI  Remote Control TestThe WFT device will send a RC command to the HI.  The operator will listen to the HI to confirm the correct function. i.e. Change Program – HI changes program.P3 HI Wireless Functional TestThe WFT device will send a sine wave to the HI. The operator listens to the WFT device output via a Stethoclip & judges the quality of the audio streamed to the WFT device.P3 DWA & P3 CROS II Remote Control TestThe WFT device will receive the remote control command from the DWA or CROS II & display what the command was.  The operator will compare the displayed result to the action they performed.  i.e. VC up is pressed, WFT device displays "Volume UP".P3 DWA & P3 CROS Wireless Functional TestThe DWA or P3 CROS II streams an audio signal to the WFT device. The operator has to judge the quality of the audio based on what he/she hears at the receiver of the WFT device.Hearing InstrumentsThe HI can have either a generic or product firmware loaded.  In either FW the default settings (also called manufacturing settings) should be set.  In the default settings configuration all audio inputs (omni mic, dir mic, T’coil etc.) & all user controls are active.  The HI is set to full on gain output, therefore the volume remote control command is to decrease the HI volume.  The program change command is to toggle to the next program in a loop. Once the last program is reached the HI will change to the first program.Digital Wireless AccessoriesFor the DWA remote control function to work the client settings must be used. The remote control can be setup for a monaural or binaural fitting.For streaming function the DWA can be in a "Generic grouping" or a fitted state.  Regardless of the fitting state of the DWA the WFT device will detect the DWA a stream the audio accordingly.Failure TestIn the event that a HI or DWA fails the test repair actions are to be taken.  These actions are not in the scope of this document.Test PreconditionsFunctional Test
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 5 / 15        Screen displayP3 CROS WFTStart testStop test        Screen displayP3 DWA WFTStart testStop testScreen display•  Palio 2•  Palio 3Button FunctionMenu functions Test functionsNavigate list Decrease volumeSelect Program change / Start / Stop testBack to previous menu Back to previous menuPalio 2Screen displayP2 HI RCPalio 3        Screen display•  P3 HI •  P3 DWAP3 CROS        Screen display•  P3 CROS RC•  P3 CROS WFTP3 DWA        Screen display•  P3 DWA RC•  P3 DWA WFTScreen displayP3 CROS RCScreen displayP3 DWA RCSelect PlatformSelect test modeSelect test modeSelect test mode        Screen displayP3 HI WFTStart testStop testP3 HI        Screen display•  P3 HI RC•  P3 HI WFTScreen displayP3 HI RCSelect test mode
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 6 / 15The Wireless Functional Test device should be used at any process step to check the wireless function of the hearing instrument or digital wireless accessory. For example, diagnose, component exchange & quality control as a part of the repair process & kit preparation or drilling & wiring, closing & quality control as apart of the ITE new production process.The WFT device is to replace the Lab myPilot for checking the wireless function of Palio 2 & Palio 3 hearing instruments.  In addition it is also possible to check the function of Palio 3 digital wireless accessories.The WFT device has a display and 3 touch pads to control the device. It is powered by a mini USB cable connected to a PC or to the power socket. There is also an acoustic out to connect a stethoclip to listen to the streamed audio signal from DWAs & CROS II products.  The WFT device has an external HIBAN antenna which must be connected to the Palio 2 or Palio 3 socket depending on which platform product is being tested.Wireless Functional Test Device1Product Compatibility Table WH22Palio 2 HIWH30Palio 3 HI Palio 3 DWAAmbra Bolero Audéo V ComPilot IISolana Virto Bolero V ComPilot Air IICassia Naìda Q PilotOne IINaìda S Sky QNios S Audéo QAudéo SFunctional TestPalio3 / Palio2POWER3.3VLINKACTMarconiPalio3Acoustic OutMini-Din Mini USBiWiProPalio2WIRELESS FUNCTIONAL TESTWireless Functional Test Setup2A template will be provided to produce a control guide for the WFT. The control guide will show the acceptable working range for both Palio2 & Palio3 hearing instruments.  The test distance for the WFT device with hearing instruments is 18 to 22cm.•  Power the WFT device via a mini-USB connected to a free USB port on a computer. Use a powered USB hub if there are no free USB ports on your computer. •  For a hearing instrument test connect the stethoclip to the HI.The HI functions correctly within the control range of 18 – 22 cmThe HI HIBAN antenna & the WFT device antenna must be alignedPalio 2Vol-down    ProgPOWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio2
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 7 / 15The WFT device can be used to test the remote control & the wireless streaming function of the P3 DWA. •  The remote control test distance for PilotOne II & ComPilot II (without neckloop) is still 105cm.•  For the ComPilot II streaming test the WFT device should be placed inside the ComPilot II neckloop.•  For testing the ComPilot Air II for both remote control & streaming function the test distance in 35 cm.•  For a DWA streaming test connect the stethoclip to the WFT deviceComPilot Air II remote control test 35cm in either direction as shown.P3 DWA RCGoTo Next    POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3P3 DWA WFTON    POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3ComPilot II streaming test, place the WFT device in the middle of the neck loopWireless Functional Test Device Self Test3During Boot-up touch all three pads simultaneously to enter the “Self Test” mode.•    - will be displayed in two rows across the LCD display. •  Press any pad to continue.•  Check LED – The “Link” LED is blinking•  Press any pad to continue.•  Audio Check – Clear beep is heard without distortion•  Press any pad to continue.•  Press & Release – touch the pad indicated with the #•  After pressing all 3 pads the self test is complete.•  When “Functional Test – Palio3” is displayed the device is ready for use.Functional TestPalio3    POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 8 / 15DWA RC – Test mode•  Use the remote control buttons on the DWA•  Read the WFT Ddvice display to confirm the button functionP3 DWA RCGoTo Next    POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3WIRELESS FUNCTIONAL TESTP3 DWA RCVolume DOWN   POWER3.3VLINKACTAntennaPalio3WIRELESS FUNCTIONAL TESTP3 DWA RCVolume UP    POWER3.3VLINKACTAntennaPalio3Palio3P3 CROS    POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3P3 CROS RCReady    POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3Palio 3 CROSPalio 3 CROS RCBegin with the P3 CROS RC. For the wireless function test a Stethoclip must be connected to the WFT device to listen to the streamed signal from the CROS II hearing instrument. Use a good battery, position the CROS II HI within test range before turning on the CROS II HI.  Do not start the test until the CROS II HI is powered on.Enter the P3 CROS RC test mode.  The WFT device will search for a CROS II hearing instrument.  “Ready” will be displayed once the connection with the HI is made.If no compatible device is found an error message will be displayed.Wireless Functional Test Device Test Modes4HI RC – Test mode•  Listen to the hearing instrument under test.•  Use the WFT device to decrease the HI volume•  Use the WFT device to change the HI programWIRELESS FUNCTIONAL TESTP3 HI RCVol-down    ProgPOWER3.3VLINKACTAntennaPalio3
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 9 / 15P3 CROS RCGo To Next    Palio 3P3 CROS WFT    P3 CROS WFT     On    POWER3.3VPOWER3.3VPOWER3.3VLINKACTLINKACTLINKACTWIRELESS FUNCTIONAL TESTWIRELESS FUNCTIONAL TESTWIRELESS FUNCTIONAL TESTAntennaPalio3AntennaPalio3AntennaPalio3Palio 3 CROS Wireless Functional TestTo check the remote control function use the user controls on the CROS II HI, PS & VC.The WFT device will display the received command. i.e. Program button = Go To Next.Test all available user controls.Repeatedly pressing the same button will not change the display.  Alternating between user controls is recommended when possible.Enter the P3 CROS WFT test mode. The WFT device will search for a CROS II HI & set the CROS device to the streaming program.  “On” will be displayed once the connection made.“Wait . . . . “ will be displayed during setup.  If no compatible device is found an error message will be displayed.Due to the noise environment & test set up it might not be possible to clearly hear the “Streamed audio” from the CROS II HI. Using you finger lightly rub across the microphone inlets.  Alternatively lightly blow across the microphone inlets to check the streaming function.Functional Test Palio3Wireless Functional Test Device Operation5•  Use down arrow to scroll between P3 HI & P3 DWA.•  Use carriage return to select.•  Use back arrow to return to the previous menu.Palio3P3 HI    POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 10 / 15P3 HI•  Use down arrow to scroll between P3 HI RC test & P3 HI WFT test.•  Use carriage return to select.•  Use back arrow to return to the previous menu.P3 HIP3 HI RC Hearing instrument remote control test•  Use down arrow to decrease the volume.•  Use carriage return to change the program.•  Use back arrow to return to the previous menu.P3 DWAP3 DWA RCPOWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3P3 HIP3 HI WFT Streaming test from the WFT Device to the HI•  Use carriage return to turn on & off the sine stream.•  Use back arrow to return to the previous menu.P3 HI WFT    On  POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3P3 DWAP3 DWA RC Digital Wireless Accessory remote control test•  Use down arrow to scroll between         P3 DWA RC test & P3 DWA WFT test.•  Use carriage return to select.•  Use back arrow to return to the previous menu.P3 HI RCVol-down    Prog POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3P3 HIP3 HI RC    POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 11 / 15P3 DWAP3 DWA RC Digital Wireless Accessory remote control test•  Use carriage return to turn on & off the test mode.•  The WFT device displays which DWA button was pressed.•  Use back arrow to return to the previous menu.P3 DWA RCGoTo NEXT  POWER3.3VLINKACTAntennaPalio3P3 DWAP3 DWA WFT Streaming test from the DWA to the WFT device•  Use carriage return to turn on & off the test mode.•  Use back arrow to return to the previous menu.P3 DWA WFT    On  POWER3.3VLINKACTAntennaPalio3WIRELESS FUNCTIONAL TESTPalio2•  Use down arrow to scroll between Palio3 & Palio2 test modes.•  Use carriage return to select.•  Use back arrow to return to the previous menu.P2P2 HI RC Hearing instrument remote control test•  Use down arrow to decrease the volume.•  Use carriage return to change the program.•  Use back arrow to return to the previous menu.Palio2Vol-down    Prog POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3Functional TestPalio2  POWER3.3VLINKACTWIRELESS FUNCTIONAL TESTAntennaPalio3
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 12 / 15Declaration of conformityHereby Phonak AG declares that this Phonak product is in compliance with the essential requirements of the Radio and Telecommunications Terminal Equipment Directive 1999/5/EC. The full text of the Declaration of Conformity can be obtained from the manufacturer.Notice 1:This device complies with Part 15 of the FCC Rules and with RSS-210 of Industry Canada. Operation is subject to the following two conditions:1)  this device may not cause harmful interference, and2)  this device must accept any interference received, including interference that may cause undesired operation.Ce produit répond aux exigences de la section 15 de la réglementation FCC et à la norme RSS-210 d’Industrie Canada. Son fonctionnement est soumis aux deux conditions suivantes: (1)  cet appareil ne doit pas causer d’interférences nuisibles, et(2)  cet appareil doit accepter toutes les interférences reçues, y compris celles pouvant provoquer un fonctionnement préjudiciable.Notice 2:Changes or modifications made to this device not expressly approved by Phonak may void the FCC authorization to operate this device.Compliance Information6Europe:Notice 3:This device has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules and ICES-003 of Industry Canada. These limits are designed to provide reasonable protection against harmful interference in a residential installation. However, there is no guarantee that interference will not occur in a particular installation. If this device does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: •  Reorient or relocate the receiving antenna.•  Increase the separation between the device and receiver.•  Connect the device into an outlet on a circuit different from that to which the receiver is connected.
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 13 / 15Information and Description of Symbols7With the CE symbol, Phonak AG confirms that this Phonak product meets the requirements of the R&TTE Directive 1999/5/EC on radio and telecommunications equipmentEMC and radio communications compliance label Australia.This symbol indicates that it is important for the user to read and take into account the relevant information in this user guide.OperatingconditionsThe device is designed such that it functions without problems or restrictions if used as intended, unless otherwise noted in this user guide.This symbol indicates that it is important for the user to pay attention to the relevant warning notices in these user guides.Temperature transport and storage:–20° to +60° Celsius(–4° to +140° Fahrenheit).Temperature operation:15° to 40° Celsius(+60° to +104° Fahrenheit)Important information for handling and product safety. Keep dryHumidity transportation and storage:<90% (non-condensing). Humidity operation: <95% (non condensing)The symbol with the crossed-out garbage bin is to make you aware that this device may not be thrown away as normal household waste. Please dispose of old or unused device, at waste disposal sites intended for electronic waste, or return it to your group company for disposal. Proper disposal protects the environment and health.
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 14 / 15Hazard warningsImportant Safety Information8The following pages contain important safety information.For persons with active implantable medical devices(i.e. pacemakers, defibrillators, etc.) the use of this device is not permitted.Only use accessories approved by Phonak AG, to avoid any electrical shock.Keep this device out of reach of children and mentally challenged people or pets.Using the device, its connectors or cables in any way contradictory to their intended purpose as described in this user guide can cause injury.Changes or modifications to the device that were not explicitly approved by Phonak are not permitted. Phonak is not responsible for situations resulting from a modified device.The mini-USB port is to be used for the described purpose only.Opening the device might damage it. If problems occur which cannot be resolved by following the remedy guidelines consult your Phonak representative.Do not use the device in explosive areas (mines or industrial areas with danger of explosions, oxygen rich environments or areas where flammable anesthetics are handled) or where electronic equipment is prohibited.Dispose of electrical components in accordance with your local regulations.Information on product safetyProtect the device from excessive moisture or heat  Keep the device of at least 10 cm away from any kind of magnets.Do not drop the device. Dropping it onto a hard surface can damage the device.Do not connect an USB cable exceeding 3 meters (9 feet) in length to the device.Protect the device connectors and plugs from dirt and debris. Clean the device using a damp cloth. Never use household cleaning products (washing powder, soap, etc.) or alcohol to clean the device. Never use a microwave or other heating devices to dry the device.The mini-USB port is to be used for the described purpose only.The digitally-coded, inductive transmission technology used in the device is highly reliable and experiences virtually no interference from other devices. It should be noted, however, that when operating the hearing system near computer equipment, larger electronic installation or other strong electromagnetic fields, it may be necessary to be at least 60 cm (24”) away from the interfering device to ensure proper operation.When the device is not in use, disconnect from power cable.
File: 000.000.015.286.pdf Version: 2Title: Work Instruction Wireless Functional Test Device Page: 15 / 15Other Important InformationIf you use several devices in the same building,be aware that their inductive field can disturb one ano-ther if within a range of 2 meters (6 feet).This may even occur when used in different rooms.High-powered electronic equipment, larger electronic installations and metallic structures may impair and significantly reduce the operating range.Version: Details: Written: Approved: Released:1 First Release  see page 1 see page 1 see page 12 Added French text page 12 (Notice 1) 2015-05-26  11mgraeterRevision History