Operations Manual

SDS Data Logger
MODEL SDS 8020/8040

Universal Data Logger
Please read, understand, and follow these instructions before operating this equipment.
Super Systems, Inc. is not responsible for damages incurred due to a failure to comply with these
instructions. If at any time there are questions regarding the proper use of this data logger, please
contact us at (800) 666-4330 for assistance.

7205 Edington Drive
Cincinnati, OH 45249
513-772-0060 / 800-666-4330
Fax: 513-772-9466
www.supersystems.com

Super Systems Inc.

Page 1

SDS Data Logger Manual #4565 Rev D

Table of Contents

Product Overview _____________________________________________________________________________ 5
Product Description ________________________________________________________________________ 5
Product Specifications ______________________________________________________________________ 5
Software Installation _______________________________________________________________________ 6
Calibration _______________________________________________________________________________ 6
Getting Started ____________________________________________________________________________ 7
Analog Inputs _____________________________________________________________________________ 7
Thermocouple connections _________________________________________________________________ 7
Voltage connections ______________________________________________________________________ 7
4 – 20 mA. Current Loop connections ________________________________________________________ 7
Survey Templates__________________________________________________________________________ 8
Installation _________________________________________________________________________________ 10
Installing the SDS Software ________________________________________________________________ 10
Connecting The SDS Data Logger To A Network ______________________________________________ 11
SDS 8020/8040 Data Logger ___________________________________________________________________ 13
Menu List _______________________________________________________________________________ 13
Keypad Assignments ______________________________________________________________________ 16
1. Overview Display _____________________________________________________________________ 17
2. Display Survey Information _____________________________________________________________ 17
3. Display Input Settings __________________________________________________________________ 18
4. Display Input Offsets __________________________________________________________________ 19
5. Display Logged Data___________________________________________________________________ 19
6. Help ________________________________________________________________________________ 20
7. Select Input Channels __________________________________________________________________ 20
8. Modify Input Settings __________________________________________________________________ 21
9. Modify Input Offsets ___________________________________________________________________ 22
10. Select Survey Template________________________________________________________________ 22
11. Battery Status _______________________________________________________________________ 23
12. About/Sign-On ______________________________________________________________________ 24
13. Revision Display _____________________________________________________________________ 24
14. Set Display Values ___________________________________________________________________ 25
15. Language/Lengua ____________________________________________________________________ 25
16. Quick Chart _________________________________________________________________________ 26
17. Overview Display Selected _____________________________________________________________ 26
18. Quick Chart Setup ____________________________________________________________________ 27
19. Modify Survey Information ____________________________________________________________ 27
20. Set the Date and Time _________________________________________________________________ 28
21. Port Setup __________________________________________________________________________ 29
22. Input Calibration _____________________________________________________________________ 29
TC Type mV Range Chart ______________________________________________________________ 30
23. Cold Junction Trim ___________________________________________________________________ 31
Firmware revision 1.09 or below _________________________________________________________ 32
Firmware Revision 1.10 and above _______________________________________________________ 33
Firmware Revision 1.11 and above _______________________________________________________ 34
Super Systems Inc.

Page 2

SDS Data Logger Manual #4565 Rev D

To Perform a Cold Junction Calibration____________________________________________________ 34
24. System Setup ________________________________________________________________________ 35
25. Set Pass Codes ______________________________________________________________________ 36
26. Set IP Address _______________________________________________________________________ 36
32. Survey Deviation Alarm Log ___________________________________________________________ 37
34. Clear Logged Data ___________________________________________________________________ 37
SDS Software _______________________________________________________________________________ 39
SDS Reporter ____________________________________________________________________________ 39
File Menu Options ________________________________________________________________________ 39
File Æ New / New Button_________________________________________________________________ 39
File Æ Open / Open Button _______________________________________________________________ 40
Manually Creating a Survey Record _________________________________________________________ 42
File Æ Save / File Æ Save As / Save Button __________________________________________________ 52
File Æ Export __________________________________________________________________________ 52
File Æ Print____________________________________________________________________________ 52
Print Button ____________________________________________________________________________ 54
File Æ Print Preview _____________________________________________________________________ 54
File Æ Exit ____________________________________________________________________________ 54
SDS Menu Options________________________________________________________________________ 54
SDS Æ Manage Survey Templates __________________________________________________________ 54
SDS Æ Download Surveys and Data ________________________________________________________ 55
SDS Æ Launch Direct FTP________________________________________________________________ 58
FTP Login For Access to the A Drive________________________________________________________ 60
SDS Æ View Real-time Data ______________________________________________________________ 61
SDS Æ Manage Survey Records ___________________________________________________________ 63
SDS Æ Perform Comms Test ______________________________________________________________ 65
Options Menu Options ____________________________________________________________________ 67
Options Æ Multiple SDS Mode ____________________________________________________________ 67
Options Æ Show Labels On Chart __________________________________________________________ 67
Help Menu Options _______________________________________________________________________ 67
Help Æ Check for Updates ________________________________________________________________ 67
Help Æ About __________________________________________________________________________ 68
SDS Data Tags ___________________________________________________________________________ 69
Creating a Personalized Template ___________________________________________________________ 74
SDS Template Manager ___________________________________________________________________ 77
SDS Reporter Versions 1.101.0.71 and below __________________________________________________ 77
Description/Function of Template Manager Screen buttons/menu options _________________________ 77
New Button / File Î New ______________________________________________________________ 77
Open Button / File Î Open _____________________________________________________________ 78
Save Buttons / File Î Save _____________________________________________________________ 79
File Î Save As _______________________________________________________________________ 79
Exit Button / File Î Exit _______________________________________________________________ 80
Add New Template Button______________________________________________________________ 80
Save Changes to Template Button ________________________________________________________ 81
Delete Template Button ________________________________________________________________ 82
Description of Template Manager Screen Tabs ________________________________________________ 83
Template Information Tab ______________________________________________________________ 83
Survey Information Tab ________________________________________________________________ 84
Active TCs Tab ______________________________________________________________________ 85
Controller Information Tab _____________________________________________________________ 86

Super Systems Inc.

Page 3

SDS Data Logger Manual #4565 Rev D

SDS Reporter Versions 1.101.0.76 and above __________________________________________________ 87
Description/Function of Template Manager Screen buttons/menu options _________________________ 87
New Button / File Î New ______________________________________________________________ 87
Open Button / File Î Open _____________________________________________________________ 87
Save Buttons / File Î Save _____________________________________________________________ 88
File Î Save As _______________________________________________________________________ 88
Exit Button / File Î Exit _______________________________________________________________ 89
Add Template Button __________________________________________________________________ 89
Delete Button ________________________________________________________________________ 89
Copy Button _________________________________________________________________________ 89
Arrow Buttons _______________________________________________________________________ 89
Description of Template Manager Screen Tabs ________________________________________________ 90
Template Information Tab ______________________________________________________________ 90
Survey Information Tab ________________________________________________________________ 91
Active TCs Tab ______________________________________________________________________ 92
Controller Information Tab _____________________________________________________________ 93
Description of Template Manager Screen Fields _____________________________________________ 94
Notes ___________________________________________________________________________________ 95
Appendix A – Standard Calibration Range _______________________________________________________ 97
Revision History _____________________________________________________________________________ 98

Super Systems Inc.

Page 4

SDS Data Logger Manual #4565 Rev D

Product Overview
Product Description
You’ll find all the necessary information related to how to use the product in this manual. The product is
broken up into 2 parts, the software and hardware components. This user guide is also written in this
manor.
SDS8020 / SDS8040 provides the hardware and software needed to meet industry standards for
thermocouple uniformity by logging and generating temperature and data log reports.
It is easy to use and addresses the many industrial applications where datalogging, trending, and
reporting is required. The product is available in either the 20 or 40-channel version with user-defined
inputs for each channel.
The easy to use Survey Template and Report Manager provides users with all the necessary tools to
generate uniformity reports. The Report Manager includes a report writer that allows the user to setup
reports to meet their specific requirements. The Template Manager allows survey specific data to be
created, updated and used during the reporting process. All reports can be printed, saved electronically
and emailed using the software utilities. The SDS software tools along with the data logger, combine to
create a powerful solution for data logging and reporting temperature uniformity surveys.
Product Specifications
•

User defined “logging” intervals as frequently as 10 seconds

•

Isolated inputs

•

AC ( 90 – 240 V) and DC Power

•

Rechargeable battery – Approximately 12 hours of “runtime”

•

Dimensions: 16”D X 20”L X 8”H

•

Password protected menu options

•

Easy calibration

•

Web-page enabled real-time view

•

ETHERNET and RS232 serial communications

•

16-Bit A/D converter

•

128 Mbytes of on-board data

Super Systems Inc.

Page 5

SDS Data Logger Manual #4565 Rev D

Degrees F.

A/D

Accuracy

Accuracy

Accuracy

TC Type

Min

Max

Full scale

Full Scale

At 1600 Degrees F.

At 1600 Degrees F with
thermister error

B

32

3308

0.0124%

0.0179%

0.29

0.65

C

32

4208

0.0062%

0.0067%

0.11

0.47

E

-328

1832

0.0031%

0.0032%

0.05

0.41

J

-346

2192

0.0031%

0.0036%

0.06

0.42

K

-328

2502

0.0031%

0.0045%

0.07

0.43

N

-328

2372

0.0031%

0.0052%

0.08

0.44

NNM

0

1409

0.0031%

0.0033%

0.05

0.41

R

-58

3214

0.0062%

0.0118%

0.19

0.66

S

-58

3214

0.0124%

0.0133%

0.21

0.57

T

-328

752

0.0062%

0.0119%

0.19

0.55

*** Note: Accuracy at 1600 Degrees F with thermister error requires SDS firmware level 1.10 or later ***
Software Installation
You will find the SDS Recorder software CD that ships with the unit includes all the reporting and
administrative utilities. Insert the CD into the CD ROM device on your computer and click the setup.exe
file. You will be given a step-by-step instruction for completing the installation process. See Installing the
SDS Software section for more details.
Calibration
The SDS data logger is calibrated prior to shipment. The
calibration process is available through menu option 22 on the
data logger. Instructions for calibration are included in this
manual. Super Systems calibrates the SDS data logger using
NIST traceable instruments that are listed on the calibration
certificate included with the device. The default calibration is
performed using a zero and span routine for the millivolt range of
0 – 80mV. Each range that is being used should be calibrated.
For a list of suggested mV ranges by thermocouple types, please
refer to the section on Input Calibration in the manual. Other
calibration ranges are available upon request.
An input device will be used to input a specific type K
thermocouple value and the data logger value will be recorded.
This will be performed at different temperatures and printed with the certificate.
Super Systems Inc.

Page 6

SDS Data Logger Manual #4565 Rev D

Getting Started
When logging data, the operator will use the on-board display to start and stop the data logging / survey
process. The first steps related to data logging are connecting the inputs to the terminal strips on the
data logger. Each terminal strip represents a single analog board with 5 inputs. You can remove these
terminal strips by pulling up on both ends of the terminal strip.
Depending on the model number, you have either 20 or 40 input channels for data logging. For each
channel there will be a positive and negative connection which must be attached to the appropriate input.
The operator defines an input type for each channel. This is performed through menu option 8 (Modify

Input Settings) on the data logger.

Once the inputs have been connected, the operator
selects which inputs will be included in the survey.
This is done in 2 ways. The template provides
operators with input selection but can be updated on
the data logger using the Select Input Channel option.
When a template has been selected, the input
channels setup for the template, as defined in the
Template Manager, will be the selected channels for
the survey. As mentioned, this can be updated using menu option 7 Select Input Channel. Always
select a template prior to running a survey. If you want to create new templates, please see
section SDS Template Manager.
Analog Inputs
The Super Systems, Inc. 31541 Analog Board contains a group of five channels isolated from the main
DC power source. The board can be connected to thermocouples, voltage sources from 20mV full scale to
1.28 Volts full scale, or 4 – 20 mA current loops.

Thermocouple connections
Thermocouple wires can be connected directly to the terminal blocks. The thermocouple junctions should
not be grounded. If they do touch a ground reference, all thermocouples on a board must have a
common ground reference. If multiple thermocouples are connected to different ground reference points,
the accuracy of all thermocouples on the board cannot be guaranteed to be accurate.

Voltage connections
Voltages from 0 mV to 1.28 Volts can be directly connected to the terminal blocks. When measuring
ground-referenced voltages, all references must share a common ground reference. If the voltage
sources are connected to different ground reference points, the accuracy of all the voltage sources
connected to the board need to be checked for accuracy.

4 – 20 mA. Current Loop connections
Before connecting the current loop, insert the shorting jumper on the board for each channel used to
measure current loops. This jumper inserts the 62-ohm shunt resistor across the input of the A/D. If
multiple current loops are connected to one board, all must share the same power supply and ground
reference points or the accuracy of all the current loops need to be checked for accuracy.
Super Systems Inc.

Page 7

SDS Data Logger Manual #4565 Rev D

Survey Templates
The data logger will default to the last survey template used. To modify the template being used, the
operator can go to the Select Survey Template menu
option. This data will be carried over for the reporting
process to make the reporting quick and easy. On the
Survey Template screen there are parameters that can
be modified for a specific survey such as setpoint.
After setting the required parameters, the start survey
process can be selected. This will timestamp the
selection and the data logger will log all channels that have been selected for logging.
The operator may receive deviation alarms depending on the
current input and setpoint as well as the alarm settings. The
alarm will notify the operator when an input is outside the
plus/minus tolerance for the setpoint. The alarm information
will display on the data logger and show the number of
channels above tolerance and below the tolerance. The
survey continues to run, it is simply providing the operator
with the information on the display.
The operator can view the input channel data on a number of
different screens. The overview display will display all
channels on the device but by using the circular arrow key,
, the operator views the overview display with only the
“selected” input channels. By pressing the circular arrow key
again, the operator will get the Quick Chart View screen
displaying trend data for a specific channel.
At anytime, the operator can view the data in real-time from
the computer using the SDS Recorder software. The real-time
data option when selected displays the data every 10 seconds. The system will update 10-second
intervals every minute. Follow instructions for connecting the datalogger to the PC.
The survey can be stopped from the Display Survey Information
screen. On the Action option, entering a “0” and pressing the
Enter key will stop the survey.
After stopping the survey, other surveys can be taken by
repeating the above steps. At anytime, all survey data can be
downloaded to the PC using the Download Surveys And Data
option on SDS Recorder. Follow instructions for
connecting the datalogger to the PC.
Once the data has been downloaded, it is available
for reporting. By clicking on the file open menu
option, the operator will be given the Report dialog
to generate a survey report. Reports use logged
data and the user specified interval in the report.
The report template selected will form the final
report view with all the template data, survey data
and additional report parameters that are entered.
Super Systems Inc.

Page 8

SDS Data Logger Manual #4565 Rev D

The report templates are user defined and should be configured to meet your needs. All reports can be
saved and displayed at a later date.

Super Systems Inc.

Page 9

SDS Data Logger Manual #4565 Rev D

Installation
Installing the SDS Software

Note – The person installing the SDS Software must have administrative rights on the
computer(s) where the installation is taking place for the installation to be successful.
Double-click on the setup file, SDSSetup.msi to start the auto-installation. If there is no version of the
SDS Reporter running on the computer, the installation will prompt the user for the download location
(figure 1.1).

Figure 1.1 SDS Reporter setup screen
The default location is “C:\SSI\SDS”. Click on the “Browse” button to select an alternate location.
Clicking on the “Disk Cost…” button will display the available computer drives onto which the SDS
Reporter can be downloaded, as well as the total space, available space, and required space for each
drive. Selecting the “Everyone” option will install the SDS Reporter for all users on the computer, while
selecting the “Just me” option will only install the SDSReporter for the current user. Clicking on the
“Cancel” button will prompt the user to confirm exiting the setup without continuing. Clicking on the “<
Back” button will display the previous screen, which is the introduction screen for the installer. Clicking
the “Next >” button will continue with the installation process (figure 1.2).

Super Systems Inc.

Page 10

SDS Data Logger Manual #4565 Rev D

Figure 1.2 Installation ready screen
Clicking on the “Cancel” button will prompt the user to confirm exiting the setup without continuing.
Clicking on the “< Back” button will display the previous page, which is the installation setup page (figure
1.1). Clicking on the “Next >” button will install the software to the specified location. Once the
software has been installed, the installer can be closed. The SDS Reporter software will now be ready to
use.
If there is a version of the SDS Reporter already installed on the computer, the installer will either reinstall the software (Repair SDS Reporter option) or remove the software from the computer (Remove
SDS Reporter option) (figure 1.3). Clicking on the “Finish” button will proceed with the selected action.
Note: the installer does not remove any subfolders in the main SDS folder. These files will have to be
manually removed.

Figure 1.3 Repair/Remove SDS Reporter screen
Connecting The SDS Data Logger To A Network
Every SDS Data Logger shipped includes an Ethernet crossover
cable that will connect the data logger to a network. Once the
data logger is connected to a network, the SDS Reporter software
will be able to find it during any searches.
Connecting the data logger to your network or directly to
a PC is accomplished using the Ethernet port on your data logger.

Ethernet Port

If you are connecting the data logger to your network, you will need an Ethernet
cable (the cable provided is a crossover cable for direct connections to a PC). The
cable is plugged into the data logger Ethernet plug and then other end should be
plugged into a network hub. If the IP Address of the data logger needs to be
changed, this can be done on the data logger through the Set IP Address menu
option.
If you are not putting the data logger on the network, you should use the Ethernet crossover cable that
has been provided with the device. Ethernet crossover cables are most often used when connecting two

Super Systems Inc.

Page 11

SDS Data Logger Manual #4565 Rev D

Ethernet computers without a hub. An Ethernet crossover cable has it's send and receive wires crossed.
When using a hub or switch, this is automatically done for you.
With a crossover cable, you are forming a network between the computer that you are directly plugged
into and the data logger. There will be some network settings on the computer that you will have to
configure for the 2 devices to communicate. The data logger will have the network setting already setup
with the following IP address – 192.168.1.215. This can be modified through the Set IP Address screen
on the data logger.
Network settings can be found through the Control Panel in Microsoft Windows. By selecting Network
Setting, the operator will be
given a list of the current
available connection types.
Using the crossover cable will
require the Local Area
Connection as seen in the below
diagram to be modified. The
Properties can be changed by
highlighting the connection and
using the right mouse button to
click and select the Properties tab
or by highlighting the connection and clicking on Change setting of this
connection.
Once the Local Area Connection Properties screen is displayed, the
Internet Protocol (TCP/IP) tab should be highlighted. Click the
Properties button to display Internet Protocol
(TCP/IP) Properties.
On the Internet Protocol (TCP/IP) Properties tab,
you will need to select the option for Use The
Following IP Address. Enter in the following data
on these fields.
IP Address: 192.168.1.209
Subnet Mask: 255.255.255.0
Default Gateway: 192.168.1.1

To change the network settings on your computer you may need addition information so please refer to
the computer manual.

Super Systems Inc.

Page 12

SDS Data Logger Manual #4565 Rev D

SDS 8020/8040 Data Logger
Menu List
The menu list shows the available pages displayed six at a time. To access the list, press the Esc key.
Depending upon where you are starting from, it may require pressing this button more than once. The
up (Ï) and down (Ð) arrows are used to scroll through the selections, which are repeated below. To go
to a specific page, either type in the page number and press Enter, or use the arrow keys to highlight
the selection and press Enter.

Operator Level – No Pass Code Required
1. Overview Display
2. Display Input Settings
3. Display Input Settings
4. Display Input Offsets
5. Display Logged Data
6. Help
7. Select Input Channels
8. Modify Input Settings
9. Modify Input Offsets
10.

Select Survey Template

11.

Battery Status

12.

About/Sign-On

13.

Revision Display

14.

Set Display Values

15.

Language/Lengua

16.

Quick Chart

17.

Overview Display Selected

18.

Quick Chart Setup

19.

Modify Survey Information

Supervisor Level – Level 1 Pass Code Required

Super Systems Inc.

Page 13

SDS Data Logger Manual #4565 Rev D

20.

Set the Date and Time

21.

Port Setup

22.

Input Calibration

23.

Cold Junction Trim

24.

System Setup

Configuration Level – Level 2 Pass Code Required
25. Set Pass Codes
26. Set IP Address
Menu Pages 27, 28 and 29 are reserved for future use and do not exist at this time. To minimize the
possibility of unintended modifications to the instrument, certain menu pages will require the entry of a
pass code to access them. Pages 1 through 18 are Operator level screens that do not require any
security codes. Pages 20 through 24 are Supervisor screens requiring a Level 1 pass code (default = 1).
Pages 25 and 26 are Configuration screens requiring a Level 2 pass code (default = 2). At the bottom of
the menu screen is a status bar, which tells the current date and time. The status bar also displays the
internal temperature (IT) of the instrument. This internal temperature should never exceed 50.
Note About Menu Numbers
Each menu screen has a unique number that is displayed in the upper left-hand corner of the screen.
This number is shown for reference. If you know the menu number of the screen that you would like to
go to, the screen number can be typed and the enter key pushed to access that particular screen directly
from the Main Menu list or the Overview Display screen.

Super Systems Inc.

Page 14

SDS Data Logger Manual #4565 Rev D

Date and Time Quick Reference Chart
Code

Month
January

1

Day
Sunday

Code
0

Code

Hour

Code

12 am

0

12 pm 12

1

1

1

13

February

2

Monday

March

3

Tuesday

2

2

2

2

14

April

4

Wednesday

3

3

3

3

15

May

5

Thursday

4

4

4

4

16

June

6

Friday

5

5

5

5

17

July

7

Saturday

6

6

6

6

18

7

7

7

19

8

8

8

20

9

9

9

21

August
September
October

1

Hour

8
9
10

November

11

10

10

10

22

December

12

11

11

11

23

Super Systems Inc.

Page 15

SDS Data Logger Manual #4565 Rev D

Keypad Assignments
All of the functions of the SDS Data Logger are controlled from the keypad. The functions of each button
are as follows:
0 – 9 are used to enter numeric data that
can vary with each page
“.” Is used to enter a decimal point where
necessary for data entry
Enter is used when entering numeric data
from the keypad or initiating automatic
functions
Ï and Ð are used to navigate through the
menu options or scroll through data on a
specific page
ESC is the escape key. This clears any
entered text, and if continuously pressed,
toggles between the main page and the
menu list
The circular arrow key, , is used to scroll through data on various pages and to change the sign, +/-,
of values on some pages

Super Systems Inc.

Page 16

SDS Data Logger Manual #4565 Rev D

1. Overview Display

The Overview Display shows the current readings of the five thermocouples on all four or eight boards,
depending on the number of channels. Pressing the circular arrow key will display the Quick Chart screen
(menu option 16). Pressing the up or down arrow keys will display the Modify Input Offsets screen
(menu option 9). Pressing any of the number keys (1-9) will display the corresponding menu option.

2. Display Survey Information

The Display Survey Information shows the survey information on the top half of the screen. Included is
the Template name, Company name, Furnace ID, Action (start or stop the survey), Survey Temperature
Setpoint, A field to edit the survey values, The Date/Time the survey was started and the Date/Time the
survey was stopped. Pressing the up or down arrow keys will scroll through the Action, Setpoint and Edit
survey fields. Pressing Enter on the line “To edit survey values Press ’Enter’” will display the Modify
Survey Information screen (menu option 19), which allows the user to modify some of the information
used in the survey. Currently, the only information that can be modified are the Setpoint, TC units, and
TC spool correction.
To start the survey, highlight the Action field, enter a 1 and then press Enter. Once a survey has been
started, “Testing” will appear in the highlighted area and the start Date/Time will be filled in the “Started
at” field.

Super Systems Inc.

Page 17

SDS Data Logger Manual #4565 Rev D

To stop a survey, highlight the Action field, enter a 0 and then press Enter. Once the survey has
finished, “Finished” will appear in the highlighted area and the stop Date/Time will be filled in the
“Stopped at” field.
To change the Setpoint, highlight the Setpoint field, enter in the new Setpoint and press Enter. Pressing
the circular arrow key while the setpoint field is highlighted will change the sign of the value entered.
The bottom half of the screen contains the full template data displayed five lines at a time: Template
Name, Company Name, Furnace ID, Furnace Make/Model, Furnace Type, Furnace Use, Furnace
Dimensions, Temperature Range, Date, Setpoint, Number of TCs, TC Type, TC Spool SN, TC Cal, TC Cal
Date, TC Spool Correction, SDS Calibrated By, SDS Cal Date, Duration, Operator, Interval, Specification,
TC Units, Simulated Load, Tolerance, Next Survey, Cont MFG, Cont Model, Cycle Time, Dead Band,
Output Limit, Prop Band, Rate, Reset, TC Gauge, and Notes. Pressing the circular arrow key will scroll
down through the template data, one line at a time. When the end of the template is reached, the scroll
will start over at the beginning of the template.

3. Display Input Settings

The Display Input Settings shows the input settings for all five thermocouples on all four/eight boards.
Pressing any of the number keys (1-9) will display the corresponding menu option. Pressing the down
arrow key will display the Modify Input Settings screen (menu option 8).

Super Systems Inc.

Page 18

SDS Data Logger Manual #4565 Rev D

4. Display Input Offsets

The Display Input Offsets screen displays the offsets for all five thermocouples on all four /eight boards,
depending on the SDS type. Pressing the down arrow key will display the Modify Input Offsets screen
(menu option 9). Pressing any of the number keys (1-9) will display the corresponding menu option.

5. Display Logged Data

The Display Logged Data shows the logged data from a specific date, with the default setting being the
current hour of the current date. Pressing the down arrow key will allow the user to change the Year,
Month, Day, and Hour. To change the Year, highlight the area and enter any four-digit year between
1980 and 2179. To change the Month, enter a number corresponding to the appropriate month – i.e. 1
for January, 12 for December. To change the Day, enter a number between 1 and 31. To change the
Hour, enter a number between 0 and 23 corresponding to the appropriate hour – i.e. 0 for 12am and 23
for 11pm. Hours are based on the 24-hour clock. For a detailed reference for all of the codes, see the
Date and Time Quick Reference Chart section (Page 19). Pressing Enter on the line that says “To
Display Press ’Enter’ here” will display the logged data.

Super Systems Inc.

Page 19

SDS Data Logger Manual #4565 Rev D

The first screen will display the first five inputs and the first ten minutes of data. Pressing the circular
arrow button will allow the user to scroll through the inputs, five inputs at a time. Pressing the down
arrow button will show the next ten minutes of logged data. Pressing the up arrow will show the
previous ten minutes of logged data. Pressing the ESC key will display the Display Logged Data Screen.

6. Help
The Help page can be used as a reference to briefly describe the functions of the various buttons on the
keypad. This manual is a better resource, however if it is not available this screen will be able to offer
information that will help in navigating through the screens on the SDS. To scroll through the screen
pages, press the down arrow key to scroll down and the up arrow key to scroll up.

7. Select Input Channels

The Select Input Channels page allows the user to manually turn a thermocouple input “on” (“Yes”) or
“off” (“No”). The thermocouples are displayed from one to twenty / forty. Use the up or down arrow
keys to select a specific thermocouple. Enter a 1 and then press Enter to turn the thermocouple on, or
enter a 0 and press Enter to turn the thermocouple off. Pressing Enter without entering a 1 or a 0 will

Super Systems Inc.

Page 20

SDS Data Logger Manual #4565 Rev D

turn a thermocouple off as well. Pressing the circular arrow key will display the Overview Display
Selected screen (menu option 17). Pressing the ESC key will also display the Overview Display Selected
screen.

8. Modify Input Settings

The Modify Input Settings allows the user to select a channel (input) and manually set the range/type, or
set the range/type for all of the channels. To select a channel, highlight the area and then enter a
number from 1 to 40 corresponding to the specific channel. Pressing the circular arrow key will scroll
through the input channels. To set the code for the range of one specific channel, scroll down to the
“Enter code for range” line and enter one of the codes from the list at the top of the screen. To enter a
code for all of the channels, scroll down to the “To set ALL, enter code here” line and enter the
appropriate code. The codes for the range are:
Num-Code

Num-Code

Num-Code

Num-Code

Num-Code

Num-Code

0–B

3–J

6 – NNM

9–T

18 – 160mv

21 – 20mv

1–C

4–K

7–R

16 – 2.56v

19 – 80mv

22 – 4-20ma

2–E

5–N

8–S

17 – 1.28v

20 – 40mv

Pressing the up or down arrow keys will scroll up or down through the three input fields: “Select input (1
to 40)” line, “Enter code for range” line, and “To set ALL, enter code here” line.

Super Systems Inc.

Page 21

SDS Data Logger Manual #4565 Rev D

9. Modify Input Offsets

The Modify Input Offsets screen allows the user to modify the offsets of each channel, or to clear all of
the offsets. To select a channel, highlight the “Select Input” line and enter a number between 1 and 40.
To enter the offset value, highlight the “Enter offset value” line and enter a number between –1000 and
1000. The user can toggle the sign (+/-) of the number being entered by pressing the circular arrow
button. To clear out all of the offsets, highlight the “Enter 1 here to clear all offsets” line and enter a 1.

10. Select Survey Template

The Select Survey Template screen allows the user to manually select a template to run a survey from.
The SDS is capable of saving thirty-two templates per template file. Each template will be listed by
template name. To create a template, go to the Template Manager screen section, Manage Survey
Templates, using the SDS software and follow the procedures. To select a template, use the up or down
arrow keys to highlight the template’s name and press Enter. The Display Survey Information screen
will be displayed (menu option 2).

Super Systems Inc.

Page 22

SDS Data Logger Manual #4565 Rev D

11. Battery Status

This page displays information about the instrument’s power supply.
Battery Voltage: This displays the DC voltage from the battery. It will read higher when the instrument is
plugged in, so a true measurement should be taken when it is not connected to AC power.
Battery Condition: This will display the battery status as Very Low, Low, or OK depending on the battery
voltage. When the instrument is plugged in, this will display Charging.
DC Supply: This will display the amount of power that is being generated by the internal power supply.
This will display 0 when the instrument is not plugged in.
Internal Supply: this will display the amount of voltage that the instrument is seeing. This is the greater
of the battery voltage and the DC supply.
If the instrument has been left unattended in the ON position, the battery may be completely dead. It
may be possible to recharge the battery, but it will take a few days to accomplish this. Leave the
instrument plugged in and periodically check the battery voltage to see if it is increasing. If it is increasing
the battery is taking a charge. Remember: it could take three to four days if the battery is completely
dead.
The instrument should not be plugged in after every use. The battery will last longer if it is discharged
for four – five hours before it is charged If the battery gets low, a warning message will appear on the
screen, followed by an estimate of the amount of time until it runs out of power. The operator will
always have at least an hour of run time after the message is displayed. When it is plugged in to
recharge, the instrument can still be used. The instrument will recharge and operate at the same time.
The battery has approximately twelve hours of battery life when fully charged.

Super Systems Inc.

Page 23

SDS Data Logger Manual #4565 Rev D

12. About/Sign-On

The above screen is the sign-on screen that shows the SSi logo, address, and phone. This page will also
display the instrument model and serial number, the date of the last factory calibration, and the number
of hours and minutes the instrument has been in use since the last factory calibration. Pressing the down
arrow key will display the Revision Display screen (menu option 13).

13. Revision Display

This screen displays the firmware revision levels for the instrument. Although the user cannot modify this
information, it may be helpful to report when contacting Super Systems for support. The “Main
Processor” value is the main firmware revision number for the SDS data logger.

Super Systems Inc.

Page 24

SDS Data Logger Manual #4565 Rev D

14. Set Display Values

This screen is used to adjust the display values. The backlight brightness and contrast are factory set to
40% and 50%, respectively, which is appropriate for most viewing conditions. Pressing the circular
arrow key will return these values to their default settings. The values for brightness and contrast can
range from 0 (dark) to 100 (bright). The backlight ON time is a power-saving feature that extends battery
life. The backlight ON feature will turn the backlight off if no buttons are pressed within a predetermined number of minutes. When the backlight goes off but the datalogger is not turned off, the
instrument is still functioning. The backlight can be restored by pressing any key. To disable this
function and keep the backlight always on, enter 0 and press Enter.

15. Language/Lengua

This screen allows the user to select a language to be used on the instrument. The SDS Data Logger
arrives with the necessary languages already downloaded. The selected language will always be listed
above the selections. To select a language, press the up or down arrow keys to highlight the desired
language and then press Enter.

Super Systems Inc.

Page 25

SDS Data Logger Manual #4565 Rev D

16. Quick Chart

This screen displays a trend chart for one active thermocouple over a period of ninety minutes. The
information on the right displays the active thermocouple number (TC), the temperature of that
thermocouple, the Setpoint (SP), the maximum value (Max) and the minimum value (Min). The
information on the left of the screen displays the full-scale value and the middle value as reference
points. Pressing the up or down arrow keys will display the Quick Chart Setup screen (menu option 18).
Pressing the circular arrow key will display the Overview Display Selected screen (menu option 17).

17. Overview Display Selected

This screen displays the active thermocouples on the instrument. Pressing the up or down arrow keys
will display the Select Input Channels screen (menu option 7). Pressing any of the number keys (1-9)
will display the corresponding menu item. Pressing the circular arrow key will cycle through the Survey
Deviation Alarm Log menu item (menu option 32) and the Overview Display screen (menu option 1). The
Deviation Alarm Log menu item will show all deviation alarms that were logged.

Super Systems Inc.

Page 26

SDS Data Logger Manual #4565 Rev D

18. Quick Chart Setup

This screen displays the setup information for the Quick Chart. Valid values for the setup information
are:
Survey Setpoint:

0 through 9999

TC input to chart:

1 through 40

Chart full scale:

0 through 9999

Chart zero scale:

0 through 9999

Reset min/max 1 = yes:

0 or 1

To change a value, highlight the desired value to change, enter in the new value and then press Enter.
Pressing the circular arrow key or the ESC key will display the Quick Chart screen (menu option 16).

19. Modify Survey Information

This screen allows the user to modify some of the information used in the survey. Currently, only the
Setpoint, Temperature type (Fahrenheit or Celsius), and the TC Spool correction factor can be modified.

Super Systems Inc.

Page 27

SDS Data Logger Manual #4565 Rev D

To change any of these values, use the up or down arrow keys to highlight the field that is to be
changed, enter the new value, and press Enter. Valid values are:
Setpoint: -400 through 4000
TC Units: 0 for Fahrenheit and 1 for Celsius
TC Spool Correction: -100 through 100
For the Setpoint and the Spool Correction, the user can change the sign of the value by pressing the
circular arrow key. Pressing ESC will display the Display Survey Information screen (menu option 2).

20. Set the Date and Time

This screen displays the date and time information for the instrument’s internal clock and calendar. Press
the up or down arrows to navigate to the part of the Date/Time that is going to be updated. Once the
desired part is highlighted, enter the new number and press Enter. Months are numbers 1 (January)
through 12 (December). Days are numbers 0 (Sunday) through 6 (Saturday). Hours are based on the
24-hour clock with numbers 0 (12 am) through 23 (11 pm). No changes will take place until the seconds
have been entered. For a detailed reference for all of the codes, see the Date and Time Quick Reference
Chart.

Super Systems Inc.

Page 28

SDS Data Logger Manual #4565 Rev D

21. Port Setup

This screen displays the parameters used for the communications ports. The factory default settings
should not need to be changed by the operator. To change the values that are not “Modbus fixed”, press
the circular arrow key to cycle through the available values. Pressing Enter will set the value as the
lowest value.

22. Input Calibration

This screen displays information to setup the calibration of the inputs. Calibration can be performed 1
channel at a time or from 1 – 5 inputs per board. The user can select board one through eight. The user
can also individually select each input (one through five) to calibrate, where 1 = yes (calibrate) and 0 =
no (do not calibrate). To continue the calibration setup, scroll down to the “Press Enter for calibration
mode” line and press Enter. This will display the setup for the calibration mode.

Super Systems Inc.

Page 29

SDS Data Logger Manual #4565 Rev D

The top of the screen displays the codes for the ranges. The codes are:
0 – 20MV 1 – 40MV 2 – 80MV 3 – 160MV 4 – 1.28V 5 – 2.56V
Below is a listing of the suggested ranges for the various TC types.

TC Type mV Range Chart
TC Type

Range in mV

B

20

C

40

E

80

J

80

K

80

N

80

NNM

80

R

40

S

20

T

20

To select a board to calibrate, highlight the first line (“Board number to cal (1 to 8)”), enter the desired
board number that will be calibrated, then press Enter. The next line is the TCs (CH = Channels) to
calibrate. If there is a Y next to the TC, then it will be calibrated. If there is an N, then the TC will not

Super Systems Inc.

Page 30

SDS Data Logger Manual #4565 Rev D

be calibrated. To select the TCs to calibrate, enter a number between 0 and 31. To determine the
number, follow these steps:
TC1 has a “value” of 1, TC2 has a “value” of 2, TC3 has a “value” of 4, TC4 has a “value” of 8, and TC5
has a “value” of 16.
A TC that is on has a value of 1. A TC that is off has a value of 0.
Multiply the TC’s value by its “value” and add all 5 TC values up.
For example, we wish to calibrate TC1, TC4, and TC5. TC1 = 1 (1 * 1), TC2 = 0 (0 * 2), TC3 = 0 (0 *
4), TC4 = 8 (1 * 8), and TC5 = 16 (1 * 16). The number to enter would be 1 + 0 + 0 + 8 + 16 = 25.
The line will now look like this: CH’s to cal 1Y 2N 3N 4Y 5Y.
To select the range to calibrate, select one of the codes above (0 through 5) and press Enter. The mV
ranges displayed above are ideal because they cover the mV range for the selected thermocouple type.
On the next line, enter a 0 for ZERO or a 1 for SPAN. Enter the target value (in MV) as a value ranging
from 0 to 320. The SPAN value should be 90% of full scale. That means for 80mV, the span value
should be 72mV. Scroll down to the next line and enter 1 to start the calibration. DO NOT START THE
CALIBRATION until you have entered the proper value in the calibration device. Also be sure that the
polarity of the wires is hooked up properly.
When the calibration is finished, the “Enter 1 here to start” line will read “Ready 0”. Pressing ESC will
display the Input Calibration first screen (menu option 22).
The calibration process can be run multiple times and if initial calibration is off significantly, running
multiple times is recommended.

23. Cold Junction Trim
The instructions for performing the Cold Junction Trim on the inputs will depend on the firmware level of
the instrument that you have. You can determine the firmware version be checking menu option 13
Revision Display. The number to the right of the first line, “Main Processor”, will be the firmware level.
Please select the instructions below based on the firmware level of the instrument.

Super Systems Inc.

Page 31

SDS Data Logger Manual #4565 Rev D

Firmware revision 1.09 or below

This screen displays the setup information to trim the cold junction value of each board. Highlight the
first line (“Select board number”), enter the desired board number to trim, then press the Enter key. The
acceptable values for the board number are: 1 – 8. Any number entered over 8 will default to 8 and any
number entered less than 1 will default to 1. Press the down arrow key to highlight the next line (“Clear
trims, 1 = yes”). This line will allow an operator to clear out the previous cold junction trims for each
input. Enter a 1 for “yes” (clear input trims) or a 0 for “no” (do not clear input trims), and press the
Enter key. Note – Clearing the trims will only clear out the cold junction trims for the selected board. If
the operator selected 1 (“yes”), the trims will be cleared once the Enter key has been pressed. Press the
down arrow key to highlight the “Modify brd CJ by (+/-)” line. This will allow an operator to modify a
board’s overall cold junction value. See the section To Perform a Cold Junction Calibration below for
details on performing the cold junction calibration. Enter the desired cold junction trim value, and, if
necessary, press the circular arrow key to toggle the sign of the cold junction value. Note – Pressing the
circular arrow key will also toggle the “Clear trims” value between “yes” and “no”. Press the Enter key
to begin the cold junction trim. When the trim has finished, the line will display 0.0, instead of the value
that was entered. This is because the offset is actually a recalibration of the board and after the
calibration process there will be no offset. If there are any trim values for the inputs, they must be
cleared before modifying the cold junction value for the board. If an operator attempts to change the
cold junction value without clearing the trim values, the “Modify brd CJ by (+/-)” line will display “Clr
Trims”. To set each individual input’s trim value, press the down arrow key to highlight the specific input.
The inputs are listed from top to bottom starting with input five and ending with input one. Enter the
desired trim value, and, if necessary, press the circular arrow key to toggle the sign of the trim value.
Press the Enter key to set the new trim value. The input trim values range from –18.00 to 18.00. Note –

The overall Cold Junction value is modified by adding or subtracting a value to the current value. The
individual input trim values are set to the entered value.

Super Systems Inc.

Page 32

SDS Data Logger Manual #4565 Rev D

Firmware Revision 1.10 and above
Depending on the revision level of the internal input board, the screen will resemble one of
the two following screens. One version uses one Cold Junction Trim value that is applied to
the entire board. The other version uses a unique Cold Junction Trim value for each of the
five inputs on that board. The instrument will automatically determine which screen should
be displayed based on the internal boards that are installed inside it.
Input Board Version 1.07 and below

This screen displays the setup information to trim the cold junction value of each board. Note – Since

the Cold Junction screen’s functionality changes slightly based upon the input boards’ version, it is
important to know which version of boards is in the SDS Data Logger. The input board version number

will be displayed in the top right corner. Highlight the first line (“Select board number”), enter the
desired board number to trim, then press the Enter key. The acceptable values for the board number
are: 1 – 8. Any number entered over 8 will default to 8 and any number entered less than 1 will default
to 1. The next section of the menu screen will look like the following:
Input 5 trim value
Input 4 trim value
Cold Junction Trim
Input 2 trim value
Input 1 trim value
The line for “Input 3 trim value” is shown as “Cold Junction Trim” because this line controls the actual
cold junction trim. Input boards with version 1.07 and below only contain one thermister per board; so
only one cold junction trim will be needed.
See the section To Perform a Cold Junction Calibration below for details on performing the cold junction
calibration. The “Cold Junction Trim” will adjust all of the inputs simultaneously. Any trim value entered
for inputs 5, 4, 2, or 1 will not be reset during the cold junction trim; therefore, those inputs would be
even more offset after the cold junction trim. For best results, perform the cold junction trim first, then
perform any necessary input trims. Press the down arrow key to highlight the “Cold Junction Trim” line.
This will allow an operator to modify a board’s overall cold junction value. Enter the desired cold junction
trim value, and, if necessary, press the circular arrow key to toggle the sign of the cold junction value.
Press the Enter key to begin the cold junction trim. While the system is working, the word “Busy” will
appear below the “Input 1 trim value” line. When the trim has finished, “Busy” will disappear. To set
each individual input’s trim value, press the down arrow key to highlight the specific input. The inputs

Super Systems Inc.

Page 33

SDS Data Logger Manual #4565 Rev D

are listed from top to bottom starting with input five and ending with input one (except for input three).
Enter the desired trim value, and, if necessary, press the circular arrow key to toggle the sign of the trim
value. Press the Enter key to set the new trim value. Note – The overall Cold Junction value, as well as

any trim value, is modified by adding or subtracting a value to the current value.
Input Board Version 1.08 and higher

The functionality of the Cold Junction Trim menu screen is similar to the Cold Junction trim for the input
board versions 1.07 and below. The difference is that the new boards have five thermisters, instead of
only one, so each board must have five cold junction trims performed. The middle of the screen will look
like the following:
Input
Input
Input
Input
Input

5
4
3
2
1

trim
trim
trim
trim
trim

value
value
value
value
value

See the section To Perform a Cold Junction Calibration below for details on performing the cold junction
calibration. Press the down arrow to highlight the input to be trimmed. Enter the desired cold junction
trim value, and, if necessary, press the circular arrow key to toggle the sign of the cold junction value.
Press the Enter key to begin the cold junction trim. While the system is working, the word “Busy” will
appear below the “Input 1 trim value” line. When the trim has finished, “Busy” will disappear.

Firmware Revision 1.11 and above
With Firmware revision 1.11, the title of the menu option 23 has become “CJ Calibration” instead of “Cold
Junction Trim”.

To Perform a Cold Junction Calibration
The user will need a thermocouple calibrator capable of outputting a thermocouple signal to calibrate the
cold junction value of the data logger. The user will need to connect the calibrator to one of the inputs
on the data logger for the channel that will be calibrated. It is recommended to let everything (calibrator
and datalogger) sit for approximately thirty minutes to allow the temperature to achieve equilibrium. Set
up the calibrator for the specific thermocouple type of the thermocouples in the datalogger, i.e. type K,
type J, etc. Then, source a specific temperature, like 1000 °F to the connected input. It is recommended
that the actual temperature used be similar to an appropriate process temperature. For example, if your

Super Systems Inc.

Page 34

SDS Data Logger Manual #4565 Rev D

equipment normally operates at 1700 °F, then perform the cold junction calibration using a 1700 °F
signal.
Firmware Version 1.09 and below
The user can source temperature to any one of the inputs, since the cold junction value is separate from
the input trim values, even though there is only one thermister per board. If the temperature being
sourced does not match the temperature displayed on the screen, the user will need to adjust the cold
junction value (“Modify brd CJ by (+/-) line) to match. After the cold junction value has been modified,
the user will be able to modify any of the input trims, if necessary.
Firmware Version 1.10 and above
Input Board Version 1.07 and below
The user will want to source the temperature to input three, since input boards with a version of 1.07 or
earlier only have one thermister, so the cold junction calibration will apply to the whole board. If the
temperature being sourced does not match the temperature displayed on the screen, then the user will
need to adjust the cold junction trim value to match. After the cold junction value has been modified,
the user will be able to adjust the trim values for inputs one, two, four, and five, if necessary.
Input Board Version 1.08 and above
The user will have to source the temperature to all five inputs, since input boards with a version of 1.08
or later have five thermisters per board, one for each input, so the cold junction calibration will have to
be made to each input separately. If the temperature being sourced does not match the temperature
displayed on the screen, then the user will need to adjust the input trim value to match.

24. System Setup

This screen displays any system setup information. Currently, the only information available is the
Survey Deviation Alarm Smart and the Temperature Resolution. To turn the Survey deviation Alarm
Smart on, enter a 1 and press Enter. To turn the Survey Deviation Alarm Smart off, enter a 0 and press
Enter. The Temperature resolution field accepts a 0, 1, or 2 as a value. The temperature resolution
affects the resolution of the temperatures displayed on the data logger.

Super Systems Inc.

Page 35

SDS Data Logger Manual #4565 Rev D

25. Set Pass Codes

This screen allows the user to set the pass codes to access the higher-level functions on the SDS. Menu
options 20 through 24 are considered Supervisor Level and they require a Level 1 pass code. Menu
options 25 and 26 are considered Configuration Level and they require a Level 2 pass code. The default
pass code for Level 1 is 1 and the default pass code for Level 2 is 2. There are 3 pass codes currently on
the SDS (there is no pass code for operator-level functions). To change the Level 1 Supervisor Pass
Code, use the up or down arrow keys to scroll to this line, enter the new pass code and press Enter.
The same is true to change the Level 2 Configuration Pass Code. Either of these pass codes may be
changed to a number between 0 and 512. Note: The Level 2 pass code will also work on all Level 1
menu options. The Level 3 menus are not visible, and Super Systems uses them when the instrument is
being set up prior to shipment.

26. Set IP Address

This screen displays the Ethernet communications address information. To change the IP Address, Net
Mask, or IP Gateway, use the up or down arrow keys to scroll to the appropriate section, enter the new
number and press Enter. To set the changes, scroll down to Set and press Enter. Note – if the
changes are not set, then no changes will be saved.

Super Systems Inc.

Page 36

SDS Data Logger Manual #4565 Rev D

32. Survey Deviation Alarm Log

The survey deviation alarm log will log all instances where the temperature of one of the survey channels
goes above or below the deviation band. The alarm log is on a one-minute update, so any alarm would
have to be present for more than a minute to be logged. The start and stop time of the alarm is listed at
the top of the screen on the line “AL 1 Start at”. The screen will display the each channel and which
channel was high or low. This log will be cleared out when the next survey is run.

34. Clear Logged Data

This screen will allow the user to clear out the logged data. This menu option is only accessible with the
SSi Special pass code that can be obtained from Super Systems Inc support. To begin clearing out the
datalog data, enter a 1 and press Enter. The middle-left of the screen will display “*** Busy ***” while
the SDS instrument is clearing out the data. When the instrument is finished clearing out the datalog
data, the message will change to “Clearing complete.” If there was an error, the SDS instrument will
display “File error -13”. For SDS firmware versions 1.09 or lower (SDS menu 13 – Revision Display), it
is suggested that the datalog files be periodically deleted from the SDS data logger. All survey data that
is captured on the data logger is transferred to the PC using the SDS Reporter software and can be
reviewed at any time from that PC. Any data logger that retains datalog files for more than one hundred
ten (110) days, or roughly three and a half (3 ½) months, will begin to experience issues with the data
logger. Datalog files can also be deleted by using the SDS → Download Surveys and Data menu option
on the SDS Reporter software to bring up the survey download screen. From the Download screen, the
Super Systems Inc.

Page 37

SDS Data Logger Manual #4565 Rev D

Options → After Downloading → Delete datalog files from SDS menu option will delete the data log files
from the SDS after they have been downloaded to the local computer. Note: Once the datalog files have
been downloaded to the local computer, they will not be automatically deleted by the SDS Reporter
software. The only way to delete those datalog files will be to delete them manually. Note – With

firmware version 1.10, when the file space begins to get low, every screen will begin to flash the
message “Flash Card Space Low!” at the bottom of the screen. If the file space eventually does get filled
up without any data log files begin deleted, the SDS Data Logger will begin to delete the oldest files.
When this happens, every screen will begin to flash the message “File Deletion is occurring!”

Super Systems Inc.

Page 38

SDS Data Logger Manual #4565 Rev D

SDS Software
SDS Reporter

Figure 2.1 SDS Reporter initial screen
When the SDS Reporter software is started up, the user sees the above screen.(figure 2.1). The four
menu options to choose from are: File, SDS, Options and Help. There are four buttons to choose from:
New

, Open

, Save

, and Print

.

File Menu Options

File Æ New / New Button
The New option allows the user to create a new report that will include the report properties, the trend
chart, and the TC data (figure 2.2). Note: the selected tab will be in Red when it is active only for the
three main tabs – Report Properties, Chart, and Data.

Super Systems Inc.

Page 39

SDS Data Logger Manual #4565 Rev D

Figure 2.2 Report Properties Tab

File Æ Open / Open Button
The Open menu option will allow the user to choose an existing report from a common Windows Open
dialog box. The software will open the dialog box in the “SDSReports” folder (typically C:\SSi\SDS\SDSReports) for existing reports (.SDSReport) to open.
The Report Properties tab contains several tabs, which in turn contain the information that makes up the
report. The first tab is the Report tab (figure 2.3), which contains the data to use for the report, the
report template to use, the report heading or title, the survey results, and any notes about the report.

Super Systems Inc.

Page 40

SDS Data Logger Manual #4565 Rev D

Figure 2.3 Report Properties Report tab

, next to the “Data” field will display a screen from which the user can
Clicking on the open box,
select the interval time (ten seconds, twenty seconds, thirty seconds, one minute, two minutes, or five
minutes) and also the specific survey to use (figure 2.4). This screen will only display survey data that
has already been downloaded (SDS → Download Surveys and Data).

Figure 2.4 Load Survey Data screen
The default interval is one minute. Select the survey to use and click on the “Use” button. Information
included with the data sources is: Template Name, [Company Name - Furnace Id], (Survey box Serial
number), and the date/time range of the survey data. Any manual entries will have [Manual Entry] at
the beginning. See the section Manually Creating a Survey Record for information on how to add a
survey record manually. Clicking on the open box next to the “Report Template” field will open a dialog
box from which the user can select the specific report template to use for the report. The software will
open the dialog box in the “SDSReportTemplates” folder (typically C:\SSi\SDS\ReportTemplates) for
existing report templates to use. The “Report Heading / Title” field is for the heading or title of the
Super Systems Inc.

Page 41

SDS Data Logger Manual #4565 Rev D

report and the “Survey Results (Pass/Fail)” field is for the results of the survey. The “Notes” field is for
any general notes for the report.

Manually Creating a Survey Record
Clicking on the “Advanced: Manually Create Survey Record” link from the Load Survey Data screen will
allow the user to create a survey record to use. Clicking on this link will display the “Manually Create
Survey Record” form (figure 2.5). Five pieces of data are required to manually create a record: the SDS
model, serial number, the start time and date of the survey, the end time and date of the survey, and the
TC Spool Offset.

Figure 2.5 Manually Create Survey Record form
The drop-down list next to “Survey Box Model” contains a list of the available models, the SDS8020 and
the SDS8040. Select they model type. Enter the SDS serial number on the next line, “Survey Box Serial
Number”. The SDS model and serial number can be found on the About / Sign-on menu, menu option 12
on the SDS instrument. The process for selecting the start date/time and the end date/time is the same.
To select a date, click on the down arrow next to the time listed. This will bring up a calendar from
which the user can select a new date (figure 2.6). Select the new date by clicking on it.

Figure 2.6 Select new date
The date of the survey start /end is listed to the right of the drop-down list. To select a time, click on the
hours, minutes, seconds or “AM/PM”. The selected field will be highlighted. Each field can be changed
by using the up or down arrows on the keyboard. If the survey start date is later than the survey end
date, a message will be displayed to the user and the record will not be added (figure 2.7).

Super Systems Inc.

Page 42

SDS Data Logger Manual #4565 Rev D

Figure 2.7 Survey start/end date error message
Enter the TC spool offset and click on the “Add Record” button to add the record. Clicking on the
“Cancel” button will cancel the add process and bring the user back to the “Load Survey Data” dialog
box. If the add was successful, a message box will be displayed (figure 2.8).

Figure 2.8 Record Add successful message
The added record will be displayed chronologically in the top of the list. Any record manually added will
begin with “[Manual Entry]” (figure 2.9). If the manually created survey is less than one second, it will
not be displayed in the list.

Figure 2.9 Manual entry on Load Survey Data form
Note – Most of the “Report Properties” tabs (Report, Furnace, Survey Box) will need to be filled out
manually as well.
The next tab is the Thermocouples tab (figure 2.10). This tab contains the information about the
thermocouple settings and it also lists the active thermocouples. The information in the thermocouple
settings includes: the temperature type (F or C), the number of thermocouples, the number of TCs used
in the report, the thermocouple type, the thermocouple gauge, the thermocouple spool number, whom
the thermocouple was calibrated by, the date the thermocouple was calibrated, and the thermocouple
spool correction factor. Clicking on the “Change” button will allow the user to change the spool
correction (figure 2.11).

Super Systems Inc.

Page 43

SDS Data Logger Manual #4565 Rev D

Figure 2.10 Report Properties Thermocouples Tab

Figure 2.11 Spool Correction screen

All values displayed to the users are corrected. That means if there is a defined offset for the TC
Spool and a channel offset, the user would be shown the net result of those offsets into the value
displayed.

There are checkboxes for a possible forty thermocouples, but only the active thermocouples will have
checks in the checkboxes. Clicking on the “Define Control TC” link will allow the user to select the TC or
TCs that will be the control TC(s) (figure 2.12). The operator can select any of the thermocouples to be
the control TC(s) and provide an optional description for that TC.

Super Systems Inc.

Page 44

SDS Data Logger Manual #4565 Rev D

Figure 2.12 Define Control TC screen
Clicking on the Accept button will set the selected control TC(s). Once a control TC(s) has been selected,
the TC’s number will be in Red on the “Active Thermocouples” section of the Thermocouples tab. The

thermocouple(s) that is identified as the control TC will not be used for the TC summary when identifying
hottest and coldest channels.

Clicking on the “View Offsets used in Survey” link will display a list of the offsets that were used on the
survey (figure 2.13). All values that are displayed to the users are corrected values. That means
the Offset defined for these TCs have already been incorporated into the value the operator would see.
IF there is a value for the TC Spool Correction Factor, that value will also be incorporated providing a
corrected view for the user.

Figure 2.13 TCOffsets screen
Super Systems Inc.

Page 45

SDS Data Logger Manual #4565 Rev D

Note: Checking or unchecking TCs on the Thermocouples Tab will affect the TCs plotted on the Chart
Tab.
The next tab is the Furnace tab (figure 2.14a and 2.14b), which contains the furnace ID, the make and
model of the furnace, the furnace type, the furnace use, the furnace’s operating range, the furnace
dimensions, the furnace class (newer versions only), and an optional image of the furnace.

Figure 2.14a Report Properties Furnace Tab (Version 1.101.0.76 & below)

Figure 2.14b Report Properties Furnace Tab (Version 1.101.0.80 & above)

Super Systems Inc.

Page 46

SDS Data Logger Manual #4565 Rev D

Clicking on the open box next the image frame will open a dialog box where the user can search for an
image of the furnace to use.
The next tab is the Survey Box tab, which contains information about the survey box itself, such as the
make and model, the serial number, whom the survey box was calibrated by, and the calibration date
(figure 2.15).

Figure 2.15 Report Properties Survey Box Tab
The next tab is the Survey Details Tab (figure 2.16). The Survey Details tab contains the company name,
the survey start date, the survey duration, the actual duration of the survey, the survey specifcation(s),
whom the survey was performed by, whom the survey was approved by, the date range for the survey,
and the due date of the next survey.

Figure 2.16 Report Properties Survey Details Tab

Super Systems Inc.

Page 47

SDS Data Logger Manual #4565 Rev D

The next tab is the Survey Parameters Tab (figure 2.17). The Survey Parameters tab contains: the
survey tolerance, the survey setpoint, the overtemp setpoint, the simulated load in pounds, and the
microns.

Figure 2.17 Report Properties Survey Parameters Tab
The next tab is the Controller Tab (figure 2.18a and 2.18b). This tab contains the Controller
manufacturer, the controller model, and the instrument type (newer versions only) in the “Temperature
Controller Settings” area and this tab also contains the PID Settings: cycle time, dead band, output limit,
PB (Gain), rate, and reset.

Figure 2.18a Report Properties Controller Tab (Version 1.101.0.76 & below)

Super Systems Inc.

Page 48

SDS Data Logger Manual #4565 Rev D

Figure 2.18b Report Properties Controller Tab (Version 1.101.0.80 & above)
The last tab in the Report Properties tab is the User Defined Tab (figure 2.19a and 2.19b). This tab
contains ten (version 1.101.0.76 & below) or twenty (version 1.101.0.80 & above) fields that can be
defined by the user and included in the report.

Figure 2.19a Report Properties User Defined Tab (Version 1.101.0.76 & below)

Super Systems Inc.

Page 49

SDS Data Logger Manual #4565 Rev D

Figure 2.19b Report Properties User Defined Tab (Version 1.101.0.80 & above)
To create a user defined field, click on the “Modify User Defined Fields”. This will display a screen that
will allow the user to edit the fields (figure 2.20). The user can type in the name or description of the
field and click the “Save” button.

Figure 2.20 Edit User Defined Fields screen

Super Systems Inc.

Page 50

SDS Data Logger Manual #4565 Rev D

The Chart Tab (figure 2.21) contains the charted information from the time specified on the survey. This
tab is similar in design and function to the Real-Time chart (figure 2.40). When the chart tab is active,
the chart toolbar buttons also appear and these buttons can be used to zoom in on a section of the chart,
pan the chart up, down, left or right, and refresh the chart back to its original values. The screen will
display 2 horizontal lines that represent the high and low temperature tolerance levels based on the
setpoint. This display will change based on the tolerance level and setpoint defined in the Report
Parameters. For more information about the function of this screen, see the section “SDS Æ View
Real-time Data.”

Figure 2.21 Chart Tab
Note: checking or unchecking TCs on the Chart Tab will affect the Active TCs on the Report Properties –
Thermocouples tab.
The Data Tab (figure 2.23) contains the data from each thermocouple for the each interval selected from
the Load Survey Data screen (figure 2.4).

Figure 2.23 Data Tab
Super Systems Inc.

Page 51

SDS Data Logger Manual #4565 Rev D

Clicking on the “Round this data off to the nearest degree” link will pop up a message box confirming the
action (figure 2.24). The rounding only affects imported data. The data in the logged files will still
contain data rounded to one decimal place. Clicking on the “Yes” button will round the data.

Figure 2.24 Round data confirmation

File Æ Save / File Æ Save As / Save Button
This will save any changes made to a new or existing report. If an existing file was opened, then the
software will automatically save the file without prompting. If a new file was created, then the software
will display a common Windows Save dialog box that will allow the user to save the report to the
“SDSReports” folder. Once the report has been saved to the “SDSReports” folder, the software will
automatically save the file the next time the save command is chosen. Clicking the Save As menu option
will automatically bring up the Save dialog box.

File Æ Export
There are three sub-menu options available for the export function:
Export Report to Word
Export Report to RTF
Export Survey Data to CSV
The Export Report to Word menu option will export the report to a Word document format, which will
provide more detail than the standard .rtf file format, if that option is desired. When this menu option is
clicked, the software will automatically begin to export the report selected to a word document. Once the
report has been exported, the new document will be displayed. Note – no save dialog box will be
displayed to the user. Initially, the report is saved as “temp.doc” to the “C:\Temp” directory. The user
can rename and save this report to any other desired location. Note – any time this menu option is used,

the resulting report will be saved as “temp.doc”, so any previous report that has not been re-saved will
be lost. The Export Report to RTF menu option will export the report to a rich-text format (RTF). When

this menu option is clicked, the software will display a Windows Save dialog box that will save the .rtf file
to the “SDSReports” folder. The Export Survey Data to CSV menu option will export all of the survey
data (as seen on the Data tab) to a comma-separated value format file. When this menu option is
clicked, the software will display a Windows Save dialog box that will save the .csv file to the
“SDSReports” folder.

File Æ Print
This menu option is slightly different from the print button. There are four sub-menu options available:
Print
Print
Print
Print

Report
Survey TC Data
Approach TC Data
All

Super Systems Inc.

Page 52

SDS Data Logger Manual #4565 Rev D

The Print Report menu option will allow the user to print out a copy of the .SDSReport file. The user will
have to select the printer to print the report (figure 2.26).

Figure 2.26 Print prompt screen
The Print Survey TC Data menu option will print out a copy of the TC data (as seen on the Data tab). The
user will have the option to configure the page settings and select a printer to print to (figure 2.26a).

Figure 2.26a Page Setup screen
The Print Approach TC Data menu option will allow the user to just print the approach data in a tabular
format. The Print All menu option will print all of the options – Report, Survey TC Data, Approach TC
Data – at once. The user will be able to select the printer (figure 2.26). All three options will be printed
as separate reports.

Super Systems Inc.

Page 53

SDS Data Logger Manual #4565 Rev D

Print Button
When the user clicks on the Print button, the software will display the print preview screen, which is
similar in design and function to the print preview screen on the View Real-time chart (figure 2.41 or
figure 2.27 below).

File Æ Print Preview
The print preview screen is similar in design and function to the print preview screen on the View Realtime chart (figure 2.41) and is displayed here (figure 2.27). See the section “Print Button” under the
SDS Æ View Real-time Data menu option for an explanation of the screen. There are three sub-menu
options available:
Print Preview Report
Print Preview Survey TC Data
Print Preview Approach TC Data
Each option will display a print preview of the desired report.

Figure 2.27 Print Preview screen

File Æ Exit
This menu option will exit the application.
SDS Menu Options

SDS Æ Manage Survey Templates
The Manage Survey Templates menu option will display the SDS Template Manger screen, which will
allow a user to add new templates and modify existing templates. See the section SDS Template
Manager Screen for a more detailed description of the functions of this screen.

Super Systems Inc.

Page 54

SDS Data Logger Manual #4565 Rev D

SDS Æ Download Surveys and Data
The Download Surveys and Data menu option will allow the user to download surveys from a specific SDS
instrument. When the menu option is clicked, the SDS software will begin to search for any and all
dataloggers located on the network (figure 2.29).

Figure 2.29 SDS80XX Data log Extractor screen
Any datalogger found on the network will be added to the drop-down list at the top of the screen. If no
devices are found, a message box will pop up letting the user know that no devices were found on the
network (figure 2.30). Check the network connections to verify that any datalogger and computer is
properly connected to the network.

Figure 2.30 No SDS devices found
Select the instrument from which to download the data files from the drop-down list box labeled “SDS:”.

Note: Currently, the drop-down list will also populate with any Video Recorder data loggers as well. Use
caution when selecting the instrument to download data from, since downloading/deleting data from a
Video Recorder using the SDS Reporter software could cause undesirable consequences. Under the
Options menu, the Download menu option (figure 2.31) will allow the user to choose to download only
data that has been used in surveys, or all of the logged data. The SDS data logger will log data
continuously when it is on regardless if a survey is running or not. There will be a check mark next to
the menu option selected. The default menu option is Only data used in surveys.

Figure 2.31 Download menu option
Super Systems Inc.

Page 55

SDS Data Logger Manual #4565 Rev D

The After Downloading option will allow the user to delete the data files from the SDS instrument, or
keep them on the instrument when the files have finished downloading. Choosing the After Downloading
→ Keep datalog files on SDS menu option will keep the datalog files on the SDS data logger. Choosing
the After Downloading → Delete datalog files from SDS will delete the datalog files from the SDS data
logger. Note: Once the datalog files have been downloaded to the local computer, they will not be
automatically deleted by the SDS Reporter software. The only way to delete those datalog files will be to
delete them manually. For SDS firmware versions 1.09 or lower (SDS menu 13 – Revision Display), it is
suggested that the datalog files be periodically deleted from the SDS data logger. All survey data that is
captured on the data logger is transferred to the PC using the SDS Reporter software and can be
reviewed at any time from that PC. Any data logger that retains datalog files for more than one hundred
ten (110) days, or roughly three and a half (3 ½) months, will begin to experience issues with the data
logger. Datalog files can also be deleted by using menu option 34 on the SDS data logger (Clear Logged
Data). Note – With firmware version 1.10, when the file space begins to get low, every screen will begin

to flash the message “Flash Card Space Low!” at the bottom of the screen. If the file space eventually
does get filled up without any data log files begin deleted, the SDS Data Logger will begin to delete the
oldest files. When this happens, every screen will begin to flash the message “File Deletion is occurring!”
The SDS software will begin to download the survey data once the user has clicked on the “Download”
button (figure 2.34). The button will read “Abort” while the data files are downloading. If a connection
to the SDS device cannot be established, the software will display a message box informing the user
(figure 2.32).

Figure 2.32 Cannot establish a connection message box
If the user chose to only download data files used in surveys, and no data files are found, the software
will display an error message (figure 2.33).

Figure 2.33
The files will be downloaded to: SDS\(SDS Model number)\log(year), i.e., SDS\SDS505000\log2006.

Super Systems Inc.

Page 56

SDS Data Logger Manual #4565 Rev D

Figure 2.34 SDS Data Log Extractor screen
The software will display a continuous progress of the download status (figure 2.35). Note: since there
may a large number of files to download, this process may take several minutes to complete.

Figure 2.35 SDS Data Log Extractor download progress
When all of the files have been downloaded, the words “Operation Complete” will be at the bottom of the
list, and the button at the bottom will read “Download”. A message box will also be displayed reading
“Operation Complete”. Clicking on the OK button on the message box will close down the download data
screen.
Clicking on the “Abort” button will stop the download and close the SDS Data Log Extractor screen. Any
files that have already been downloaded will not be erased. The next time the user wishes to download
survey data from the same device, the files that have already been downloaded will be skipped and not
be downloaded again.
The Drive menu option will allow the user to select the drive on the datalogger to download data from.
The choices are E or A (Troubleshooting Only). The E drive is the default location for logged data, but
sometimes when there is an issue with the E drive, the datalogger will begin to log data to the A drive.
The most common issue with the E drive is that the flash card is not firmly secured in its socket.

Super Systems Inc.

Page 57

SDS Data Logger Manual #4565 Rev D

SDS Æ Launch Direct FTP
The Launch Direct FTP tool is used for administrative purposes. If you wanted to clean up old data
log files, this tool will allow you to perform this action. Only data stored in the E folder should be
manipulated or removed. The Launch Direct FTP menu option will start a new FTP session. This will
open a new window that displays all of the files and directories on a specific SDS instrument (figure
2.36).

Figure 2.36 FTP session window with Windows Explorer
Please note that if Microsoft Internet Explorer 7 is installed on the computer, then the ftp session window
will be displayed as a link-style, instead of the folder-style. If only Internet Explorer 6 or earlier is
installed, then the ftp session window will be in a folder-style window. If Internet Explorer 7 is installed,
it is possible to open the ftp session in a folder-style window. Click on the Launch Direct FTP menu
option, then with Internet Explorer 7 open, click on the Page button in the top right of the screen and
select the option Open FTP Site in Windows Explorer (figure 2.36b). This will open a folder-style window.

Super Systems Inc.

Page 58

SDS Data Logger Manual #4565 Rev D

Figure 2.36a FTP session window with Internet Explorer 7

Figure 2.36b – Open IE7 FTP site with Windows Explorer
The version of Internet Explorer can be found in the Help → About Internet Explorer menu option of
Internet Explorer. Note: if the SDS Æ Multiple SDS Mode menu option is checked, then the software will
display a selection box for the user to select which SDS instrument to FTP into (figure 2.37). See the
section “SDS Æ Multiple SDS Mode” for a more detailed explanation of the Multiple SDS Mode menu
option.

Figure 2.37 SDS instrument selection box

Super Systems Inc.

Page 59

SDS Data Logger Manual #4565 Rev D

Clicking the “OK” button will open the FTP window. Clicking the “Cancel” button will cancel the FTP
session creation. If the Multiple SDS Mode menu option is not checked, the software will start an FTP
session with the first SDS device that is found on the network.

FTP Login For Access to the A Drive
To gain access to the A drive on the SDS instrument, enter the following text in the address bar on
Internet Explorer 6 or earlier, or open up the My Computer icon and type the following into the address
bar if Internet Explorer 7 is installed:
ftp://LANG:SSI772@
where  is the IP address of the SDS instrument. For example, if the IP address is
192.168.1.202, then the text would look like: ftp://LANG:SSI772@192.168.1.202 (Figure 2.39). Note –
The username of LANG and the password of SSI772 must all be capitalized in order for the login to be
successful. If the username or the password is not capitalized, the user will receive a message box
requesting login information, which will not log the user in (figure 2.38).

Figure 2.38 Re-login message box

Figure 2.39 FTP login information in Internet Explorer’s address bar
Once the user has been successfully logged in, the user will have access to the A drive. Any language file
that will be placed on the SDS instrument needs to be located on the A drive.

Super Systems Inc.

Page 60

SDS Data Logger Manual #4565 Rev D

SDS Æ View Real-time Data

Figure 2.40 Real-time Data screen
The View Real-time Data menu option will display the most current real-time data available. This screen
will update itself with the most current real-time data every 60 seconds. The amount of time left until
the next update will be displayed in the bottom left-hand corner of the screen. When it does update, the
software will connect to the SDS, download the data, and re-draw the screen. Any screen settings, such
as the size of the graph and how zoomed in the user is, will not be changed. Note: if the Multiple SDS
Mode menu option is checked, then the SDS software will display a device selection box that will allow
the user to choose which device to gather the real-time data from (figure 2.37). If the Multiple SDS
Mode menu option is not checked, then the SDS software will gather real-time data from the first SDS
device that it finds on the network. Note: it is possible to view real-time data from more than one SDS
instrument at a time. To view multiple real-time data, click on the View Real-time data menu option
again and select the next SDS instrument.
Cursor Button
This is the default cursor used on the screen. The user can select the various TCs to plot and move the
scrollbar on the left side of the screen up or down in order to see more TCs.
Zoom Button
This will allow the user to zoom in the graph of the real-time data. Once the cursor is over the graph it
will change to a crosshair, ¨. The user can then hold down the left mouse button and drag the cursor to

Super Systems Inc.

Page 61

SDS Data Logger Manual #4565 Rev D

highlight the section of the graph that is to be enlarged. Letting go of the mouse button will enlarge the
graph.
Pan Button

This will allow the user to pan to a different part of the graph. Once the cursor is over the graph, it will
. By holding down the left mouse button and moving the mouse
change to a crosshair with arrows,
left, right, up, or down, the user can move the scale of the graph by minutes or temperature, or both.
Refresh Button
This will refresh the graph and reset the graph’s display values to their default values. For the values
along the y-axis, the refresh sets the range from 0 to 2400. For the minutes along the x-axis, the refresh
encompasses approximately two hours. Note: when the Real-time screen first loads up, the graph
encompasses roughly one hour of data. The graph is marked with fifteen-minute intervals along the xaxis.
Print Button
This will create a print preview of the graph for printing purposes. When the user clicks on the Print
button, the Print Preview screen will be displayed (figure 2.41).

Figure 2.41 Print Preview screen
From the Print Preview screen, the user can choose to zoom in on the preview by pre-defined
percentages (10% - 500%) or leave the zoom on the default setting (auto). The user can also determine

Super Systems Inc.

Page 62

SDS Data Logger Manual #4565 Rev D

how many pages to view at once (one, two, three, four or six). The default is one page. The upper
right-hand corner of the Print Preview screen contains a page counter, which will allow the user to select
which page to preview. Clicking on the “Close” button will close the Print Preview screen without printing
the chart. Clicking on the Print button ( ) will print the document. Note: the Zoom function is for
preview purposes only and will not affect the actual print size. Once the document has been printed, the
user can close the Print Preview screen.
Real-time button
This button will allow the user to keep the cursor in a real-time mode, which will automatically update the
TC values on the left of the screen after every update. Clicking on this button will start the real-time
mode. Clicking on this button a second time will take the screen out of real-time mode. Note: The graph
will still continue to update once a minute, but the values on the left will no longer be up-to-the-minute
accurate.
The checkboxes under the “Plot” section will allow the user to determine which TCs to plot on the graph.
Checking a specific TC will display that TC’s data on the graph. The data under the “Value” section is the
current temperature of the TC. Note: even if the Pan button or the Zoom button has been selected, the
user can still select or de-select the TCs. This will not affect the zoom status or the pan status.

SDS Æ Manage Survey Records

Figure 2.42 Manage Survey Records – Local Survey Records Tab
The Manage Survey Records screen will allow the operator to delete survey records from the survey
database file either locally on the computer or remotely on the SDS data logger. The “Local Survey
Records” tab (Figure 2.42) will work with survey records saved locally to the computer. To select
multiple sequential records, hold down the Shift key and click on the desired survey records. To select
non-sequential records, hold down the Ctrl key and click on the desired survey records. The operator will
have to confirm the deletion by clicking on the “Yes” button on the message box (Figure 2.43).

Super Systems Inc.

Page 63

SDS Data Logger Manual #4565 Rev D

Figure 2.43 Confirm delete of local records

Figure 2.44 Remote SDS80xx Survey Records Tab
The “Remote SDS80XX Survey Records” tab (Figure 2.44) will allow the operator to delete survey records
directly from the SDS data logger. The drop-down list at the top of the screen will list all of the SDS data
loggers found on the network. Note – the drop-down list may take a few moments to search the network
for SDS data loggers. Note – Currently, the drop-down list will also be populated with Video Recorder

dataloggers. Please use caution when selecting a datalogger from the list; deleting data from a Video
Recorder could produce undesirable consequences. When the operator has selected the correct data

logger, he/she can click on the “Load” button to load the survey data. To select multiple sequential
records, hold down the Shift key and click on the desired records. To select multiple non-sequential
records, hold down the Ctrl key and click on the desired records. Clicking on the “Delete” button will
remove the survey records from the list. Note – The survey records will be removed from the list, but not

actually deleted from the SDS data logger until the “Save Changes” button has been clicked. If one or
more survey record is accidentally deleted, click on the “X” in the top right of the screen to close the
screen down, and start over. Clicking on the “Save Changes” button will save the changes back to the
SDS data logger. No confirmation is needed for this action. Once the changes have been saved, a
message box will be displayed (Figure 2.45).

Super Systems Inc.

Page 64

SDS Data Logger Manual #4565 Rev D

Figure 2.45 Changes Saved message box

SDS Æ Perform Comms Test
SDS Æ Perform Comms Test
The Perform Comms Test menu option will send out a broadcast over the network to determine if any
SDS instruments are located on the network. If an instrument is located on the network, the software
will attempt to ping the device. This function will let the user know if communication with an SDS
instrument through the SDS Reporter software is possible. When the test has finished, one of three
options will happen:
1. No SDS found
No SDS instruments were found on the network. Check the network connections. A message
will be displayed to the user (figure 2.46). Click the “OK” button to close the form

Figure 2.46 No SDS found on network message
2. One or more SDS were found.
One or more SDS instruments were found on the network, and all responded to the ping. A
message will be displayed saying (figure 2.47).

Figure 2.47 Communications OK message

Super Systems Inc.

Page 65

SDS Data Logger Manual #4565 Rev D

This means that all communications appear to be ok (figure 2.48). Click the “Close” button to
close the form.

Figure 2.48 Communications ok message on Test Comms dialog box
3. One or more SDS were found, but one or more failed to respond to a ping.
SDS instruments were found on the network, but one or more did not respond to a ping. For
each instrument that does not successfully respond to the ping, the user will be informed that
there was a problem testing the communications and the software will give the IP address of the
SDS instrument(s) that failed to respond as well as the IP address of the SDS instrument(s) that
did respond successfully to the ping (figure 2.49). The most common cause is different subnets
between the SDS and the computer (figure 2.50). The user may need to change the IP address
of the SDS through menu option 26, Set IP Address. Click the “Close” button to close the form.

Figure 2.49 One or more SDS did not respond message on Test Comms dialog box

Figure 2.50 One or more SDS did not respond message
If the user is attempting to communicate with the SDS directly using a laptop that also has a
wireless network on it, the wireless network could interfere with some of the SDS Reporter
functions. The UDP error could happen if the user tried to ping a datalogger while a wireless
network was enabled (figure 2.50a). Any wireless network should be disabled before running
the SDS Reporter software.

Super Systems Inc.

Page 66

SDS Data Logger Manual #4565 Rev D

Figure 2.50a UDP Error
Options Menu Options

Options Æ Multiple SDS Mode
This is a toggle switch for the application. If there are multiple SDS instruments located on the network,
click this menu option. Once clicked, there will be a check mark next to the menu item. If this menu
option is not checked and there are multiple SDS instruments on the network, the software will use the
first SDS device that is found. This menu option is checked by default.

Options Æ Show Labels On Chart
This is a toggle switch for the application. When checked, this will display the “Begin Survey” label for
the survey on the real-time graph (figure 2.51).

Figure 2.51 Label from real-time graph
Help Menu Options

Help Æ Check for Updates
The Check for Updates menu option will check for updates over the Internet and automatically update
the SDS firmware and software. If an update is found, the software will display a message box asking
the user for update confirmation (figure 2.52).

Figure 2.52 Update found message box

Super Systems Inc.

Page 67

SDS Data Logger Manual #4565 Rev D

The software will automatically update the files and restart the application (figure 2.53)

Figure 2.53 Automatic Update screen
If no updates are available, then the software will display a message box informing the user (figure
2.54).

Figure 2.54 No updates message box

Help Æ About
The About menu option displays the SDS Reporter version that is running, and all of the previous versions
with any version notes (figure 2.55).

Super Systems Inc.

Page 68

SDS Data Logger Manual #4565 Rev D

Figure 2.55 SDSReporter About box
SDS Data Tags
SDS Data tags are created so that all of the template information and survey data can automatically be
used for the report process. Data that is captured during the survey is summarized to provide the overall
results for the report. This can be in the form of a graph, tabular data, text, etc. Data tags have been
specifically created to address the uniformity requirements for AMS and other standards. We have tags
that address, overshoot, minimum TC with value, maximum TC with value, trend data, tabular data, etc.
An example of an output from the data tags can be seen by opening the Example report from the SDS
Recorder provided with the installation. The only data not generated from data tags is the tabular print
out of the actual temperatures. This data is generated from the Survey TC Data option and will print all
data points that are displayed on the graph between the “start” and “stop” selected by the user in the
survey.
In the figure below (Output From Data Tags), you can see and example of the data tags for the graph,
survey setpoint, minimum and maximum TCs with values and deviation. The data tags used to generate
the information below are;

Super Systems Inc.

Page 69

SDS Data Logger Manual #4565 Rev D

TC Survey Summary
Temperature Setpoint: °#
Minimum TC number: #mtcn#

Maximum TC number: #xtcn#

Minimum TC Value: #mtcv#

Maximum TC Value: #xtcv#

Min deviation from setpoint: #md#

Max deviation from setpoint: #xd#

Figure Output From Data Tags
Tag
SDS:FILE
SDS:PD
SDS:PT
SDS:FID
SDS:FM
SDS:FT
SDS:FU
SDS:FD
SDS:FC
SDS:FOR
SDS:SD
SDS:ST
SDS:SDR
SDS:DUR
SDS:OP
SDS:APP
SDS:TITLE
SDS:NOTE
SDS:TOL
SDS:OTS
Super Systems Inc.

Description
The filename the report is saved as
The current date when the report is printed
The current time when the report is printed
The furnace ID
Furnace make/model
Furnace type
Furnace use
Furnace dimensions
Furnace Class
Furnace operating range
Survey date
Survey time
Survey date range
Duration of the survey
Operator. Survey performed by
Survey approved by
The heading, or title, of the report
Notes for the survey
Survey tolerance / uniformity required
Overtemp setpoint
Page 70

Usage




















SDS Data Logger Manual #4565 Rev D

SDS:SP
SDS:TCN
SDS:TCT
SDS:TCS
SDS:TCC
SDS:TCCD
SDS:TCCF
SDS:SDN
SDS:SDSM
SDS:SDSS
SDS:SDSC
Tag
SDS:SDSCD
SDS:SDSCF
SDS:SPEC
SDS:INT
SDS:T
SDS:SIM
SDS:OVS
SDS:MIC
SDS:RES
SDS:GRAPH600x400
SDS:COM
SDS:UDUR
SDS:UTCN
SDS:PCT
SDS:PDB
SDS:POL
SDS:PPB
SDS:PR
SDS:PRE
SDS:CMF
SDS:CMA
SDS:IT
SDS:TCG
SDS:OV
SDS:OD
SDS:OVT
SDS:CTD
SDS:CTN
°#
SDS:ED
SDS:ET
SDS:TCM
SDS:UD1
SDS:UD2
SDS:UD3
SDS:UD4
SDS:UD5
SDS:UD6

Super Systems Inc.

Survey setpoint

Number of thermocouples

Thermocouple type

Thermocouple spool number

Thermocouple calibrated by

Thermocouple calibration date

Thermocouple spool correction factor

Next survey due date

Survey box make/model

Survey box serial number

Survey box calibrated by

Description
Usage
Survey box calibration date

Survey box correction factor

Specifications the survey meets

Sample interval

Temperature character – F or C

Simulate load, in pounds

Overshoot (Deprecated)

Microns

Survey result

Inserts a 600X400 image of the graph

Company name

User defined survey duration

User defined number of thermocouples

PID cycle time

PID dead band

PID output limit

PID PB (gain)

PID rate

PID reset

Temperature controller manufacturer

Temperature controller make/model

Controller instrument type

Thermocouple gauge

Overshoot - yes or no

Outputs TC and temp if overshoot occurred

Time the overshoot occurred

Inserts a note regarding the control

Thermocouple. Inserts nothing if no control
Thermocouple was chosen
Control thermocouple. Inserts “None” if 
None is defined
°degree type
°#
Survey end date

Survey end time

Thermocouple map (image)

User defined field
1

User defined field
2

User defined field
3

User defined field
4

User defined field
5

User defined field
6


Page 71

SDS Data Logger Manual #4565 Rev D

SDS:UD7
SDS:UD8
SDS:UD9
SDS:UD10
SDS:UD11
SDS:UD12
SDS:UD13
SDS:UD14
SDS:UD15
SDS:UD16
SDS:UD17
SDS:UD18
SDS:UD19
SDS:UD20
Sample Output
Tag
SDS:FILE
SDS:PD
SDS:PT
Tag
SDS:FID
SDS:FM
SDS:FT
SDS:FU
SDS:FD
SDS:FC
SDS:FOR
SDS:SD
SDS:ST
SDS:SDR
SDS:DUR
SDS:OP
SDS:APP
SDS:TITLE
SDS:NOTE
SDS:TOL
SDS:OTS
SDS:SP
SDS:TCN
SDS:TCT
SDS:TCS
SDS:TCC
SDS:TCCD
SDS:TCCF
SDS:SDN
SDS:SDSM
SDS:SDSS
SDS:SDSC
SDS:SDSCD
SDS:SDSCF
SDS:SPEC
SDS:INT

Super Systems Inc.

User
User
User
User
User
User
User
User
User
User
User
User
User
User

defined
defined
defined
defined
defined
defined
defined
defined
defined
defined
defined
defined
defined
defined

field
field
field
field
field
field
field
field
field
field
field
field
field
field

7
8
9
10
11
12
13
14
15
16
17
18
19
20
















Sample Output
SDSReport1.sdsreport
2/16/06
11:30:00 AM
Sample Output
Batch 1
Furnace Mfg.
Batch Gas Draw
Temper
24 x 15 x 12
A
650 - 1000°C
2/16/06
11:30:00 AM
2/13 THRU 2/14
00:30:00
Shaun Scott
Scott Brown
Test Survey 1
This is a test survey
[+/-] 10
+10
750
20
K
Z458
Soandso
1/15/06
-1
2/28/06
SSI SDS8020
SDS60293201
Super Systems Inc
1/15/06
-1.0
AMS 2750-C and D
10s

Page 72

SDS Data Logger Manual #4565 Rev D

SDS:T
F
SDS:SIM
500
SDS:OVS
None
SDS:MIC
N/A
SDS:RES
Passed
SDS:GRAPH600x400 [An image of the graph]
SDS:COM
Company Name
SDS:UDUR
30 mins
SDS:UTCN
20
SDS:PCT
20
SDS:PDB
2
SDS:POL
2
Sample Output
Tag
SDS:PPB
2
SDS:PR
2
SDS:PRE
2
SDS:CMF
Super Systems
SDS:CMA
7EK
SDS:IT
B
SDS:TCG
16
SDS:OV
Yes
SDS:OD
TC6 - 1700°
SDS:OVT
5:30
SDS:CTD
TC5 is the control TC, and is not used in uniformity survey
results
SDS:CTN
TC5
°#
°F
SDS:ED
2/16/06
SDS:ET
11:30:00 AM
SDS:TCM
[Thermocouple image]
TC Offsets
Tag
#o1#

Description
Offset for TC1

Tag
#o1#

Sample Output
-1

Usage
#o1#

Min/Max/Mean/Spread/Deviation
Description
Tag
#n1#
Minimum temp for TC1 within the selected
Survey region. Substitute numbers for other
TCs
#x1#
Max temp for TC1
#m1#
Mean temp for TC1
#s1#
Spread for TC1
#vn1#
Deviation @min temp from setpoint for TC1
Description
Tag
#vx1#
Deviation @max temp from setpoint for TC1
#vm1#
Deviation @mean temp from setpoint for
TC1
#mtcn#
Minimum TC number
#mtan#
Minimum TC number during approach

Super Systems Inc.

Page 73

Usage
#n1#
#x1#
#m1#
#s1#
#vn1#
Usage
#vx1#
#vm1#
#mtcn#
#mtan#

SDS Data Logger Manual #4565 Rev D

#mtav#
#mtcv#
#mtcx#
#md#
#rect#
#xtcn#
#xtan#
#xtav#
#xtcm#
#xtcv#
#xd#
Tag
#n1#
#x1#
#m1#
#s1#
#vn1#
#vx1#
#vm1#
#mtcn#
#mtan#
#mtav#
#mtcv#
#mtcx#
#md#
#rect#
#xtcn#
#xtan#
#xtav#
#xtcm#
#xtcv#
#xd#
#ts#

segment
Minimum TC value during approach segment #mtav#
Minimum TC value
#mtcv#
The maximum temperature of the minimum #mtcx#
TC
Minimum calculated survey TC absolute
#md#
deviation from setpoint
Recovery Time – the time between the time #rect#
when the first TC comes into tolerance and
the time when the last TC comes into tolerance
Maximum TC number
#xtcn#
Maximum TC number during approach
#xtan#
segment
Maximum TC value during approach segment #xtav#
Minimum temperature of the maximum TC
#xtcm#
Maximum TC value
#xtcv#
Maximum calculated survey TC absolute
#xd#
deviation from setpoint
Sample Output
183
183
183
12.5
383
383
383
2
2
150
1
500
201.0
3 mins 30 secs
1
6
1500
500
183
383.0
13.8

Creating a Personalized Template
To create a personalized template, the user needs to open a Word document and save it to the
“ReportTemplates” folder in the main installed folder as a Rich Text Format (.rtf) file. Any of the SDS
data tags listed in the above section can be used to create a company-specific template. For example,
suppose that someone wanted to create a simple template called “NewTemplate” which displays the
current date and time the report is printed, the company name, the title of the report, the number of
thermocouples used, and the results of the survey. First, the user would open a Word document and
save the file as “NewTemplate.rtf” in the “ReportTemplates” folder in the main installed folder. Next, the

Super Systems Inc.

Page 74

SDS Data Logger Manual #4565 Rev D

user would set up any initial formatting settings, such as margin sizes, page layout (portrait or
landscape), etc. Now, the user can set up the template to his or her company’s specifications. Suppose
that the current date and time will be in the top right corner of the report and should be 10-point Arial
font. On the first line, the user can select the right-justified option, the 10-point font size option, and the
Arial font option. This will leave the cursor in the top right corner. Next, the user will enter the following
exactly (including the brackets): . This will be where the current date is displayed. On the
next line, the user can enter . This will be where the current time is displayed. Now suppose
the title of the report should be centered, bold, 14-point Arial font. The user will select the centerjustified option, and the 14-point font size option. This will place the cursor in the middle of the line.
The user will then enter . The rest of the report - company name, number of
thermocouples used, and the results of the survey - can be left-justified, non-bold, 12-point font. On the
next line, the user will select the left-justified option, the 12-point font size option, and deselect the bold
option. The user can then enter: Company name: . On the next line the user can enter:
Number of thermocouples used: . On the next line, the user will enter: Survey results:
. Next, make sure the document is saved and close the document out. Note: it is always
good practice to save a document in progress often in case of power failure. The resulting template
should resemble figure 2.56 below.

Figure 2.56 Completed NewTemplate report template

Next, the user will select the report template to use from the SDSReporter application’s Open option.
, next to the Report Template field will allow the user to choose which
Clicking on the Open button,
report template to use (figure 2.3). The user can choose the NewTemplate.rtf file. Next, the user will
choose the “Export to RTF” print menu option on the File menu and choose the location to save the .rtf
file. The actual report should resemble figure 2.57 below.

Super Systems Inc.

Page 75

SDS Data Logger Manual #4565 Rev D

Figure 2.57 Actual NewTemplate report

Super Systems Inc.

Page 76

SDS Data Logger Manual #4565 Rev D

SDS Template Manager
SDS Reporter Versions 1.101.0.71 and below
The purpose of a template is to save the users time by filling in some of the more generic survey data.
For example, the user could create a template for one customer, including company name, TC type, etc;
then when the survey is run using this template, this information is already present and it won’t need to
be entered for each survey run for that customer. The SDS Template Manager (figure 3.1) allows the
user to see all of the templates currently stored on the SDS instrument, as well as any templates saved to
a local computer. The Template Manager also allows the user to create new templates and save them to
a local file on a computer or network, or to the SDS instrument.

Figure 3.1 SDS Template Manager

Description/Function of Template Manager Screen buttons/menu options
This section will describe the function of the toolbar buttons and menu options for the Template Manager
Screen. Clicking on either the form button or the menu option will produce the same result, so each pair
of button/menu option will be explained in this section.

New Button / File Î New
The New button,
, or New menu option will allow the user to create a new set of templates. The
maximum number of templates that can be added per file is 32. Note: initially, there are zero (0)
templates. The SDS software does not create 32 blank templates. When the New Button is clicked or
Super Systems Inc.

Page 77

SDS Data Logger Manual #4565 Rev D

the New menu option is clicked, the tabs that contain all of the fields for the template will become visible,
as well as the Templates list. Creating a new template file will also enable the Save menu option and the
Save As menu option. Using the keyboard shortcut Ctrl + N will also create a new set of templates.

Open Button / File Î Open
The Open button,
, or the Open menu option will allow the user to open an existing template file
either from the local computer, or from an SDS instrument on the network. Note: Before clicking
“Open”, the user must first select the device to open from the Device menu option (“Local File” or “SDS
Instrument”) (figure 3.2).

Figure 3.2 Device menu option
Once selected, the chosen device will have a check mark next to it on the Device menu option. The
“Device connected:” label at the bottom of the Template Manager screen will also show what device the
user is connected to (either “Local File” or “SDS Instrument”). When the user is not connected, this label
will show “Not Connected”. If the user tries to open a file without selecting a device first, the SDS
software will display a “No Device Type Chosen” message (figure 3.3).

Figure 3.3 No Device Type Chosen message
If the user selects a local file, the SDS software will display a common Windows dialog box for the user to
select which file to open. If the user selects an SDS instrument, the SDS Software will display a selection
box with all of the available SDS instruments on the network (figure 3.4). Note: Since the SDS software
is searching the network for all SDS devices, it may take a couple of minutes for the selection box to be
displayed.
Super Systems Inc.

Page 78

SDS Data Logger Manual #4565 Rev D

Figure 3.4 SDS Instrument selection box
While the SDS software is searching the network for SDS devices, it will display a progress box, which will
detail the progress of the search (figure 3.5).

Figure 3.5 SDS software progress box
When the software is finished searching, the progress box will close automatically. When the Open
Button or the Open menu option is clicked, the tabs that contain all of the fields for the template will
become visible. Opening an existing file will also enable the Save menu option and the Save As menu
option. Using the keyboard shortcut Ctrl + O will also open an existing file. Once the file has been
loaded, the filename will be listed above the Templates list on the right of the form (figure 3.1).

Save Buttons / File Î Save
The Save button,
, Save File button, or Save menu option will allow the user to save the current file
back to the local machine or the SDS instrument, depending on which device was selected. Note: this
option will only save the entire file back; it will not save changes to individual templates. Any changes
to the individual templates that have not been saved will not be saved. See the section “Save
Changes to Template Button” to find out how to save changes to individual templates. If the user selects
an SDS instrument, the SDS software will display a progress box (figure 3.5). When the save has
completed, the SDS software will close the progress box. Once the file has been saved, the filename will
be listed above the Templates list on the right of the form (figure 3.1).

File Î Save As
This option acts the same way as the Save option, only it will display a dialog box or SDS selection box
(depending on the device selected) that will allow the user to select the file to save. If the user selects
“Local File” as the device, the SDS software will display a common Windows dialog box. Clicking “Cancel”
will cancel the save. If the user selects a file that already exists and clicks “Save”, the SDS software will
prompt the user for overwrite confirmation (figure 3.6). Clicking “No” will close the confirmation box
without saving and return the user to the Save As dialog box. Clicking “Yes” will save the file.

Super Systems Inc.

Page 79

SDS Data Logger Manual #4565 Rev D

Figure 3.6 Overwrite confirmation

If the user types in a filename that does not exist and clicks “Save”, the SDS software will prompt the
user for creation confirmation (figure 3.7). Clicking “No” will close the confirmation box without saving
and return the user to the Save As dialog box. Clicking “Yes” will save the file.

Figure 3.7 Creation of a file confirmation
If the user selects an SDS instrument, the SDS software will display a progress box (figure 3.5). When
the save has completed, the SDS software will close the progress box. Once the file has been saved, the
filename will be listed above the Templates list on the right of the form (figure 3.1).

Exit Button / File Î Exit
This option will close the Template Manager screen.
The following buttons are located on the “Template Information” tab, “Survey Information” tab, “Active
TCs” tab, and the “Controller Information” tab, and deal specifically with individual templates, not the
template file as a whole. Clicking on these buttons will not affect the template file.

Add New Template Button
This option will add a template to the collection of templates stored in memory, up to 32 total templates.
The only required field is the Template Name. If the user attempts to create a template without
providing a template name, the SDS software will display a No Template Name error message (figure
3.8).

Figure 3.8 No Template Name error message
The Setpoint field may be left blank or filled in with a numeric number up to 4 digits long. If the user
does enter non-numeric text, the SDS software will display a Setpoint Not Numeric error message (figure
3.9).

Super Systems Inc.

Page 80

SDS Data Logger Manual #4565 Rev D

Figure 3.9 Setpoint Not Numeric error message

Once a template has been added, the first template will be loaded into the fields.

Save Changes to Template Button
This option will save the changes made to a specific template back to the template collection stored in
memory. A valid template must be selected from the “Templates” list to the right of the “Template
Information” tab (figure 3.10). Note: this will only save changes to 1 template at a time and it
will not save changes to the template file. Use the Save buttons or menu options to save the file.
Once the template has been saved, the first template will be loaded into the fields.

Figure 3.10 Templates drop-down list
If a valid template is not chosen, the SDS software will display a No Template Selected error message
(figure 3.11).

Figure 3.11 No Template Selected error message

Super Systems Inc.

Page 81

SDS Data Logger Manual #4565 Rev D

Delete Template Button
This option will delete a template from the collection of 32 templates stored in memory. Note: the first
template cannot be deleted. If the user attempts to delete the first template, the SDS software will
display an error message (figure 3.12).

Figure 3.12 Unable to Delete First Template message
Once the user clicks on the “Delete Template” Button, the SDS software will display a message
confirming the delete (figure 3.13).

Figure 3.13 Confirm delete message
Clicking on “No” will cancel the delete. Clicking on “Yes” will delete the template. After the deletion has
occurred, it will not be necessary to click the “Save Changes to Template” button.

Super Systems Inc.

Page 82

SDS Data Logger Manual #4565 Rev D

Description of Template Manager Screen Tabs

Template Information Tab

Figure 3.14 Template Information tab
This tab contains general information about the template such as the Template Name, the Company
Name, the Number of TCs, the TC Type, the TC Spool Number, the TC Spool Correction, the Setpoint, the
Simulated Load in pounds, and any Notes about the template. This tab also contains the 3 template
buttons: Add New Template, Save Changes to Template, Delete Template. See “Description of Template
Manager Screen Fields” for the maximum number of characters allowed per field.

Super Systems Inc.

Page 83

SDS Data Logger Manual #4565 Rev D

Survey Information Tab

Figure 3.15 Survey Information tab
This tab contains information about the survey such as the Survey Duration in minutes, the Survey
Tolerance, the Specification the survey meets, the Sample Interval in seconds, and the Operator
performing the survey. It also contains information about the specific furnace involved such as the
Furnace ID, the furnace’s Make and Model, the furnace’s Dimensions, the Temperature Type used in the
furnace (Fahrenheit or Celsius), the Temperature Range of the furnace, the Type of furnace, and what
the furnace is used for. This tab also contains the 3 template buttons: Add New Template, Save Changes
to Template, Delete Template.

Super Systems Inc.

Page 84

SDS Data Logger Manual #4565 Rev D

Active TCs Tab

Figure 3.16 Active TCs Tab
This tab contains information about the active TCs such as the type of SDS instrument involved (8020 or
8040), the Gauge of the TC, and a check box for each of the TCs. There is also a check box to allow the
user to use the active settings for the TCs or to individually select the TCs to use on the SDS instrument.
Checking “Use these active TC settings” will set the checked TCs and will enable either the first 20 check
boxes, or all of the check boxes, depending on what type of SDS instrument is selected. Leaving this
check box unchecked will not set any TC and will disable all 40 check boxes. This tab also contains the 3
template buttons: Add New Template, Save Changes to Template, Delete Template.

Super Systems Inc.

Page 85

SDS Data Logger Manual #4565 Rev D

Controller Information Tab

Figure 3.17 Controller Information Tab
This tab contains information about the controllers such as the Controller Manufacturer, and the
Controller Make and Model. It also contains information about the PID settings such as the Cycle Time,
the Dead Band, the Output Limit, the PB (gain), the Rate, and the Reset. This tab also contains the 3
template buttons: Add New Template, Save Changes to Template, Delete Template.

Super Systems Inc.

Page 86

SDS Data Logger Manual #4565 Rev D

SDS Reporter Versions 1.101.0.76 and above
The purpose of a template is to save the users time by filling in some of the more generic survey data.
For example, the user could create a template for one customer, including company name, TC type, etc;
then when the survey is run using this template, this information is already present and it won’t need to
be entered for each survey run for that customer. The new SDS Template Manager (figure 3.18) allows
the user to see all of the templates currently stored on the SDS instrument, as well as any templates
saved to a local computer. The Template Manager also allows the user to create new templates and save
them to a local file on a computer or network, or to the SDS instrument.

Figure 3.18 New SDS Template Manager

Description/Function of Template Manager Screen buttons/menu options
This section will describe the function of the toolbar buttons and menu options for the Template Manager
Screen. Clicking on either the form button or the menu option will produce the same result, so each pair
of button/menu option will be explained in this section.

New Button / File Î New
The New button,
, or New menu option will allow the user to create a new set of templates. The
maximum number of templates that can be added per file is 32. Note: initially, there are no templates in
the file.

Open Button / File Î Open
The Open button,
, or the Open menu option will allow the user to open an existing template file
either from the local computer, or from an SDS instrument on the network (figure 3.19).

Super Systems Inc.

Page 87

SDS Data Logger Manual #4565 Rev D

Figure 3.19 Open template menu option
The use will be able to open a local template file or a template file that is on an SDS. If the user selects
a local file, the user will be able to click on the Browse button to select the file to open. The SDS
software will display a common Windows dialog box for the user to select which file to open. If the user
selects an SDS instrument, the SDS Software will display a selection box with all of the available SDS
instruments on the network. Note: Since the SDS software is searching the network for all SDS devices,
it may take a couple of minutes for the selection box to be displayed. This list will be populated with all
of the available SDS devices found when the “Open Template File” screen is first opened. Clicking on the
Scan button will allow the user to re-scan the network for available SDS instruments. Once a local or
remote file has been selected, click on the Open File button to open the file, or click on the Cancel
button to close down the screen without opening a file. Note: When the Open Button or the Open menu
option is clicked, the tabs that contain all of the fields for the template will not become visible – an
individual template must be selected to view the tabs. Once the file has been loaded, the filename will
be listed above the Templates list on the left of the form (figure 3.18).

Save Buttons / File Î Save
The Save button,
, or Save menu option will allow the user to save the current file back to the local
machine or the SDS instrument. Note: Any changes made to any template will be saved at once when
this option is selected. Note: If the user is working with a new template file, the Save function will act
like the Save As function (see below section).

File Î Save As
This option acts the same way as the Save option, only it will display a dialog box that will allow the user
to select the file to save (figure 3.20). This functionality of this screen is similar to the Open Template
screen (figure 3.19). The template file can be saved locally on a computer or remotely on an SDS
instrument. If the user selects to save the file locally, the user will be able to click on the Browse
button, which will display a common Windows save dialog box. The user will be able to select the
location and give the template file a name. If the user selects to save the file to a remote SDS, the
software will display a selection box with all of the available SDS instruments on the network. Note:
Since the SDS software is searching the network for all SDS devices, it may take a couple of minutes for
the selection box to be displayed. This list will be populated with all of the available SDS devices found

Super Systems Inc.

Page 88

SDS Data Logger Manual #4565 Rev D

when the “Save Template File As” screen is first opened. Clicking on the Scan button will allow the user
to re-scan the network for available SDS instruments. Once a local file or remote file has been selected,
click on the Save File button to save the file, or click on the Cancel button to close down the screen
without saving the file.

Figure 3.20 Save Template File As screen
The software will always display a message box with the success or failure of the save.

Exit Button / File Î Exit
This option will close the Template Manager screen.

Add Template Button
This option will add a template to the collection of templates stored in memory, up to 32 total templates.
The only required field is the Template Name. The default template name is