Texas Instruments Series 525 Rainfall Sensors Users Manual CWS

TR-525USW to the manual 490df9a4-26a3-4bbd-a52b-90a42306cc78

2015-02-05

: Texas-Instruments Texas-Instruments-Series-525-Rainfall-Sensors-Users-Manual-400442 texas-instruments-series-525-rainfall-sensors-users-manual-400442 texas-instruments pdf

Open the PDF directly: View PDF PDF.
Page Count: 21

DownloadTexas-Instruments Texas-Instruments-Series-525-Rainfall-Sensors-Users-Manual CWS
Open PDF In BrowserView PDF
COMPLETE WEATHER STATION
USER’S MANUAL

TEXAS ELECTRONICS, INC.
5529 Redfield St. Dallas, TX 75235
Phone (214) 631-2490

P.O. Box 7225 Dallas, TX 75209
Fax (214) 631-4218

Toll Free (800) 424-5651

SERIES 525 RAINFALL SENSORS
DESCRIPTION
The sensor consists of a gold anodized aluminum collector funnel with a knife-edge that diverts the water to a
tipping bucket mechanism. The models TR-525I and TR-525USW are calibrated in inches (.01” per tip) and
model TR-525M is calibrated in millimeters (.1mm per tip). A magnet is attached to the tipping bucket, which,
as the bucket tips, actuates a magnetic switch. Thus, a momentary switch closure takes place with each tip of
the bucket. Connecting the sensor to an event counter on an electronic datalogger or display module will allow
record keeping of accumulated rainfall. If an analog signal representing rainfall accumulation is required,
Texas Electronics, Inc. manufactures a suitable conditioning circuit.
The spent water drains out of the bottom of the housing; hence, the sensor requires no attention or servicing of
any sort. It is completely automatic. The aluminum sensor housing is finished with a white baked enamel paint
to withstand years of exposure to the environment.
SPECIFICATIONS
Resolution:
Accuracy:
English
Metric
Average Switch Closure Time:
Maximum Bounce Settling Time:
Maximum Switch Rating:
Temperature Limits:
Humidity Limits:
Height:
Weight:
Receiving Orifice Diameter:
Cable:
Installation:

0.01” or 0.1 mm
1.0% at 1”/hr or less
1.0% at 10 mm/hr or less
135 ms
0.75 ms
30 VDC @ 2 A, 115 VAC @ 1 A
+32°F to +125°F
0 to 100%
10.125”
2.5 pounds
6.060” (English)
9.664” (Metric)
8.000” USW (English)
25 feet, 2-conductor
Consists of attaching the three sensor support legs to a firm
platform or securing the side bracket to a stable vertical structure
such as the lower end of weather station mast. Sensor cable is then
connected to monitoring equipment.

Maintenance:

Occasional cleaning of debris from filter screen may be required.

Warranty:

Three Years

ORDERING INFORMATION
Model #

Description

TR-525I

Rain Gauge, 6.06” collector, English
(Please specify for calibration of 0.2 mm/tip)
TR-525USW Rain Gauge, 8.00” collector, English
TR-525M
Rain Gauge, 25 mm collector, Metric
Optional Parts / Accessories
HOBO
MB-525
FC-525
BB-525
HT-525
Cable

Event Datalogger and Software
Pole Mounting Base
Field Calibration Kit
Bird Repellant
Heater, 120 VAC
Additional Cable

MODEL 525
TIPPING BUCKET RAIN GAUGE TRANSMITTER
INSTALLATION INSTRUCTIONS
A clear and unobstructed mounting location is necessary to obtain accurate rainfall readings.
This transmitter has provisions for mounting two ways, surface mounting and mast mounting. Surface
mounting is recommended where possible. The transmitter housing MUST be mounted in a LEVEL position
and in a location free from vibration. If mast mounted, make sure that the mast is properly guyed so that
vibration in high winds is kept to a minimum.
THE FOLLOWING IS VERY IMPORTANT:
After the final transmitter installation has been made, remove the top gold funnel portion of the transmitter and
observe the black tipping bucket. It should NOT (repeat NOT) be held in a dead center position by the
magnetic attraction of the bucket magnet and the hermetically sealed magnetic switch. Press either end of the
bucket down against the stop to be sure that it is not centered.
The transmitter to indicator connecting cable may be either shortened or lengthened as required.
The funnel and tipping bucket mechanism should be cleaned periodically. An accumulation of dirt, bugs, etc.
on the tipping bucket will adversely affect the calibration.

FIELD CALIBRATION
TR-525USW
Absolutely accurate calibration can be obtained only with laboratory equipment, but an approximate field check
can be easily made. The tipping bucket mechanism is a simple and highly reliable device. The transmitter must
be located in a clear area, away from trees, buildings, etc. It must also be mounted level. Accurate readings
will not be obtained unless the transmitter is mounted in a level position. The mechanism must be clean. Any
accumulation of foreign material, dust, etc. will alter the calibration of this unit. The transmitter must be
calibrated with the rate of flow of water through the tipping bucket mechanism under control. At least 36
seconds should be allowed to fill one side of the tipping bucket. This represents a maximum flow rate of on
inch of rain per hour. If the flow rate is increased, then the instrument will read low (if properly calibrated).
Decreasing the rate of flow will not materially affect the calibration. The reason for this is obvious if the
tipping bucket assembly is observed when the weight if this water starts to tip the bucket. Some time is
required for the bucket to tip (a few milliseconds). During the first 50% of this time, water flows into the empty
bucket. The amount of water flowing during the first 50% of the time is error, the faster the flow rate, the
greater the error. Now at flow rates of one inch per hour (100 bucket fillings) or less, the water actually drips
into the bucket rather than flowing. Under this condition, the bucket tips between drips, and no error water is
added to a full moving bucket.
TO CHECK CALIBRATION:
A field calibration kit is available from Texas Electronics, Inc. Model No. FC-525 is a kit that consists of a
holding tank, pre-measured beaker, two orifices, cleaning tools and instructions.

FIELD CALIBRATION
TR-525I & TR-525M
Absolutely accurate calibration can be obtained only with laboratory equipment, but an approximate field check
can be easily made. The tipping bucket mechanism is a simple and highly reliable device. The transmitter must
be located in a clear area, away from trees, buildings, etc. It must also be mounted level. Accurate readings
will not be obtained unless the transmitter is mounted in a level position. The mechanism must be clean. Any
accumulation of foreign material, dust, etc. will alter the calibration of this unit. The transmitter must be
calibrated with the rate of flow of water through the tipping bucket mechanism under control. At least 36
seconds should be allowed to fill one side of the tipping bucket. This represents a maximum flow rate of on
inch of rain per hour. If the flow rate is increased, then the instrument will read low (if properly calibrated).
Decreasing the rate of flow will not materially affect the calibration. The reason for this is obvious if the
tipping bucket assembly is observed in operation…with water falling into one side of the tipping bucket, there
comes a point when the weight of this water starts to tip the bucket. Some time is required for the bucket to tip
(a few milliseconds). During the first 50% of this tipping time, water continues to flow into the filled bucket;
the last 50% of this tipping time, water flows into the empty bucket. The amount of water flowing during the
first 50% of the time is error, the faster the flow rate the greater the error. Now at flow rates of one inch per
hour (100 bucket fillings) or less, the water actually drips into the bucket rather than flowing. Under this
condition, the bucket tips between drips, and no error water is added to a full moving bucket.
TO CHECK CALIBRATION:
A field calibration kit is available from Texas Electronics, Inc. Model No. FC-525 is a kit that consists of a
holding tank, pre-measured beaker, two orifices, cleaning tools and instructions.

MODEL TTH-1315 TEMPERATURE HUMIDITY SENSOR
DESCRIPTION
The Texas Electronics, Inc. Model TTH-1315 Sensor utilized a Rotronics HygroClip S3 Humidity Temperature
Probe with interchangeable sensing elements that do not require calibration.
The unit is encased in a corrosive-resistant spun aluminum radiation shield that allows for wind aspiration and
protection from the sun's UV rays. The shield is finished in white baked-enamel to provide for virtual cosmetic
invisibility while reflecting much of the radiant heat from surrounding objects and the sun.
Overcurrent protection is provided in a NEMA 4X enclosure that is mounted to the angled mounting bracket on
the sensor. Tranzorbs are utilized to protect the sensor and signal conditioning units in an overcurrent situation.
With DC voltage applied to the sensor, the unit will return signal voltages that are linear to the range of the
instrument. Signal conditioning is applied to ensure proper voltages are returned and can be amplified or
modified to a current output to connect to any analog signal processing unit.

SPECIFICATIONS
Humidity Sensor:
Temperature Sensor:
Sensor Protection:
Measuring Range:
Operating Limits:
Accuracy at
73°F / 23°C:
Repeatability:
Analog output signals:

ROTRONIC HygromerÒ C94
(capacitive)
RTD Pt100 Ohm, 1/3 DIN
Wire Mesh filter (standard) /
foam filter (optional)
0-100 % Relative Humidity
-40° to +140°F (-40° to +60°C)
Temperature
Same as measuring range
+/- 1.5% RH and +/- 0.5°F /
0.3°C - standard calibration
+/- 1.0% RH and +/- 04°F / 0.2°C
- special calibration (optional)
Better than 0.5% RH and 0.1°C/
32.18°F
0…100% RH = 0…1V
(minimum load >10k ohm)40…+140°F = 0…1V (40…+60°C)
Analog Signal Resolution 0.02%
RH and 0.1°F

Sampling Time:
Maximum cable length:

Protection grade:
Supply voltage:
Current consumption:

<0.7 sec
5 m / 15 ft. for direct output
Up to 100 m / 325 ft. with signal
amplifier
Bayonet cap on mounting
connector
IP65 (probe to connector)
3.5 … 50VDC
<4 mA

Minimum excitation time:

<4 sec

EMC compatibility (CE):

EN-50081-2, EN-50082-2

Material:
Dimensions:

Polycarbonate, white
Length: 100 mm (3.94")
Diameter: 15 mm (0.59")
14 g / 0.5 oz.

Connection Type:

Weight:

FEATURES & BENEFITS
•
•
•
•
•
•

Combines Temperature and Humidity onto one sensing unit
Rotronics HygroClip S3 has interchangeable sensing elements requiring no calibration
Stacked plate construction of shelter provides natural ventilation
Quick-Release mounting bracket allows for easy installation and maintenance
Aluminum radiation shield is lightweight and extremely durable
White baked-enamel finish reflects most radiant heat from sun and surrounding objects

INSTALLATION & MAINTENANCE
The radiation shield with sensing element can be pole or mast mounted. Whenever possible, sensors should be
installed at a height of 4 ft. (1.2 meters) or greater over earth or sod at least 100 ft. (30.48 meters) away from
any concrete or other hard-surfaced area and not closer to any other object than four times the height of the
object above the instrument shelter or remote sensors. Avoid roof installations if possible. If it is necessary to
roof-mount shelters and sensors, they should not be closer than 30 ft. (9.14 meters) to any large, vertical
reflecting surface (walls, etc.), exhaust fans, or cooling towers. Electronic remote sensors when roof-mounted
should be at least 9 ft. (2.74 meters) or greater above the roof surface. To minimize radiation effects from the
roof, they can also be mounted on a horizontal boom so that they will extend from the side of the building roof
or tower assembly.
ORDERING INFORMATION
Model #

Description

TTH-1315
TTH-1315A

Temperature & Humidity Sensor
Temperature & Humidity Sensor, 4-20 Ma

Optional Parts / Accessories
H50
Cable

Interior Humidity Sensing element only
Additional Cable

MODEL TD-4 WIND DIRECTION SENSOR
DESCRIPTION
The Texas Electronics, Inc. TD-4 Wind Direction Sensor is a mechanical style wind meter that measures the
horizontal wind azimuth. This unit combines small physical size with superior bearings to meet the EPA's
Prevention of Significant Deterioration (PSD) starting threshold requirements.
The TD-4 wind direction sensor is a freestanding device for measuring the direction of wind. The sensor
consists of a vane and counterweight assembly, which is mechanically coupled to a potentiometer (variable
resistor). As the vane rotates in the wind, the potentiometer changes resistance proportionally to the direction of
wind.
SPECIFICATIONS
Operating Range:
Signal Presentation:

5000 ohm output:
10000 ohm output:
Analog 4-20 mA output:

0-360° mechanical
5000 ohm potentiometer,
10000 ohm potentiometer, or
Analog 4-20 mA
0-355° electrical range
3 VDC excitation minimum
0-357° electrical range
3 VDC excitation minimum
0-355° electrical range
10-30 VDC

Performance:
Accuracy:
Starting Threshold:
Resolution:
Potentiometer Linearity:

+/- 3.0°
0.6 mph (0.27 m/s)
1°
+/- 1.0%

Environmental:
Operational Envelope:
Temperature:
Relative Humidity:

0-135 mph (0 to 60 m/s)
-40 to 160° F (-40 to 70° C)
0-100%

Physical:
Vane Overall Length:
Overall Height:
Turning Radius:
Weight:
Bearings:
Mounting Base:
Cable:
Warranty:

8.5" (21.6 cm)
6.75" (17.2 cm)
13" (33 cm)
0.5 lbs (0.23 kg) less cable
APEC 3 or better
Screw attachment, 10-32 machine
screw
60' (18.3 m) 3 conductor 22 gauge
3 year

FEATURES & BENEFITS
•
•
•
•
•
•
•
•

Superior low starting threshold
Long life hybrid dual wiper potentiometer
No plastic parts for extremely long life
Precision stainless steel bearings for stability and repeatability
Crossarm included with purchase of matching wind speed sensor
Easy installation and maintenance
Over 25 years in production
Lightweight and rugged anodized aluminum exterior

INSTALLATION & MAINTENANCE
Installation consists of threading the 10-32 mounting base into our crossarm or any other suitable beam. If a
crossarm is used, the entire unit can be bolted to a mast or attached via U-bolts.
The sensor is dynamically calibrated at the factory and due to the nature of its operation should not require field
calibration. Calibration is a matter of proper orientation during installation. A magnetic compass is
recommended for proper orientation. Field maintenance should include occasional cleaning of the vane
assembly and inspection of the internal mechanism to make sure it is free from insects and debris. In some
applications users may need to occasionally verify and document sensor accuracy with a calibration dial/vane.
Possible bearing and potentiometer replacement every three to five years to maintain low starting threshold.
ORDERING INFORMATION
Model #

Description

TD-4
Wind Direction Sensor, Light Industrial
TD-410
Wind Direction Sensor, Light, 10K
TD-4A
Wind Direction Sensor, Light, 4-20 mA
* Sensor is designed to work with TV-4 series wind speed sensors.
Optional Parts / Accessories
Cable

Additional Cable

MODEL TV-4 WIND SPEED SENSOR
DESCRIPTION
The Texas Electronics, Inc. TV-4 Wind Speed Sensor is a mechanical style anemometer that measures the
horizontal velocity of wind. This unit combines small physical size with superior bearings to meet the EPA's
Prevention of Significant Deterioration (PSD) starting threshold requirements.
The TV-4 wind speed sensor is a freestanding device for measuring air velocity. The sensor consists of a
lightweight 3-cup anemometer, which electromechanically converts wind speed into a measurable electronic
signal.
The output signal can be presented in 3 optional forms: a pulsed DC signal, an AC frequency, or a conditioned
analog signal. Each output has a specific application. The pulsed DC signal is used where high-accuracy is
needed and continuous power is not a problem. The AC frequency output is used in situations where power
consumption is critical. And finally, the conditioned analog signal is used to easily and quickly communicate
with virtually all digital control systems such as PLC's or SCADA systems.
SPECIFICATIONS
Operating Range:
Signal Presentation:
Pulsed DC output:
AC Frequency output:
Analog 4-20 mA out:

Performance:
Accuracy:
Distance Constant:
Starting Threshold:
Environmental:
Operational Envelope:
Temperature:
Relative Humidity:

0-100 mph
Pulsed DC output - light chopper
AC Frequency, or Analog, 4-20 mA
(Please specify)
20-slot disc
Input Power:
1 MPH = 520 pulses/min.
100 MPH = 52000 pulses/min.
26 mV/MPH (typical)
Input Power:
0.133 Hz/MPH
4 mA = 0 MPH
Input Power:
20 mA = 100 MPH

+/- 2.0 mph (0.89 m/s)
> 21.7' (6.6 m)
0.6 mph (0.27 m/s)

0-135 mph (0 to 60 m/s)
-40 to 160° F (-40 to 70° C)
0-100%

+5.0 VDC @ 5mA (typical)
(Other voltages available upon request)
None (self-generating)
10-36 VDC

Physical:
Cup Wheel Diameter:
Overall Height:
Turning Radius:
Cup Diameter:
Bearings:
Mounting Base:
Weight:
Cable:

6.0" (15.3 cm)
4.75" (12.1 cm)
3.0" (7.6 cm)
2.0" (5.1 cm)
APEC 3 or better
Screw attachment, 10-32 machine screw
0.5 lbs (0.23 kg) less cable
60' (18.3 m) 3 conductor 22 gauge

Warranty:

3 year

FEATURES & BENEFITS
•
•
•
•
•
•
•

Superior low starting threshold due to small physical size
No plastic parts for extremely long life
Precision stainless steel bearings for stability and repeatability
Crossarm included with purchase of matching wind direction sensor
Easy installation and maintenance
Over 5 years in production
Lightweight and rugged anodized aluminum exterior

INSTALLATION & MAINTENANCE
Installation consists of threading the 10-32 mounting base into our crossarm or any other suitable beam. If a
crossarm is used, the entire unit can be bolted to a mast or attached via U-bolts.
The sensor is dynamically calibrated at the factory and due to the nature of its operation should not require field
calibration. Field maintenance should include occasional cleaning of the cup assembly and inspection of the
internal mechanism to make sure it is free from insects and debris. In some applications users may need to
occasionally verify and document sensor accuracy with a synchronous test motor. Other possible routine
maintenance is to replace the bearing housing assembly every three to five years to maintain low starting
threshold.
ORDERING INFORMATION
Model #

Description

TV-4

Wind Speed Sensor, Light Industrial
(Specify supply voltages other than 5 VDC)
Wind Speed Sensor, AC Generator
Wind Speed Sensor, Analog 4-20 mA

TV-4AC
TV-4A

*Sensor is designed to work with TD-4 wind direction sensor.

Optional Parts / Accessories
Cable

Additional Cable

MODEL SP-LITE SOLAR RADIATION SENSOR
DESCRIPTION
The Texas Electronics, Inc. SP-Lite Solar Radiation Sensor utilizes a Kipp & Zonen Silicon pyranometer
mounted in a white baked-enamel finished aluminum bracket that provides a stable upward-facing installation.
It measures the solar energy that is received from the entire hemisphere (180 degrees field of view). The output
is expressed in Watts per square meter.
The pyranometer is designed for continuous outdoor use. Its calibration is valid for natural sunlight only, but
not for artificial light. In its most frequent application, the pyranometer is used for measuring the solar radiation
emitting on the horizontal surface.
The sensor consists of a photodiode; housing, mounting bracket with cable junction box attached, and cable. A
resistance shunts the photodiode. This is done to generate a voltage output. The photodiode and the material on
top of it determine most electrical specifications. It is encapsulated in the housing in such a way that it has a
field of view of 180 degrees and that its angular characteristics fulfill the "Cosine Response".
The nominal output resistance of the pyranometer is 50 Watts. This implies that the input impedance of the
readout equipment should be at least 5000 Ohms in order to make an error of less than 0.1%.
Cable can be extended without problems to a length of 328 ft. or 100 meters, provided that cable resistance is
less than 0.1% of the input impedance of the readout equipment.
The electrical sensitivity of the photodiode changes with the temperature. A nominal value for this is 0.2%
change per degree Celsius. Calibration is done at 20°C (68°F).

SPECIFICATIONS
Electrical:

Impedance (nominal): 50 Ohms
Response time: < 1 sec.
Sensitivity (nominal): 100 uV/W/m2
Expected signal range under atmospheric conditions: 0 to 0.2V
Stability: <+/- 2% per year
Non-linearity: < 1% up to 1000 W/m2
Temperature dependence of sensitivity: +/- 0.15%/°C

Spectral:

Spectral range: 0.4 to 1.1 nm
Detector type: SILICON photo diode

Directional:

Cosine corrected between 80° angle of incidence, error: within +/- 10%
Cosine errors averaged over opposite azimuth error (at 60° angle of
incidence): within +/- 10%
Tilt response: no error

Mechanical:

Environmental:

Material of housing: Anodized aluminum contained in white bakedenamel finished aluminum mounting bracket
Cable length: 60 ft.
Dimensions: Height from surface to top of level Pyranometer - - 6.25"
(15.87 cm)
Width - 2.75 " (6.98 cm)
Length - 8.25 " (20.95 cm)
Weight: 3 lbs. (1.36 kg.) with 60 ft. cable
Working temperature range - -30° to +70°C (-22° to +158°F)

Cable:

60 ft. (18.28 meters)

Warranty:

3 year

FEATURES & BENEFITS
•
•
•
•
•

SP-Lite is an all-weather instrument
Designed for continuous outdoor use
Complies with "Cosine Response"
Full 180-degree field of view for complete hemispheric measurement
Contained in lightweight and rugged white baked-enamel finished aluminum mounting bracket

INSTALLATION & MAINTENANCE
Installation:
The site for an upward-facing pyranometer should be free from any significant obstructions above the plane of
the sensing element and should be readily accessible. If practicable, instrument should be located so that (1.) a
shadow will not be cast on it at any time (e.g. by radio masts, etc.); (2.) it is not close to light-colored walls or
other objects likely to reflect sunlight onto it; and (3.) it is not exposed to artificial radiation sources. A flat roof
provides the best location, or a rigid stand with a horizontal upper surface some distance from building
structures or other obstructions. The stand should be sufficiently rigid that the horizontal position of the
receiving surface does not change, especially during high winds. Precautions should be taken to avoid
subjecting the instrument to severe shocks or vibration.
Calibration / Cleaning Frequency:
Recalibration is suggested every two years, preferably by letting a higher standard run parallel during two sunny
days and comparing daily totals.
The sensor should be kept clean, using water or alcohol.
ORDERING INFORMATION
Model #

Description

SP-Lite
SP-Lite-A

Solar Radiation Sensor
Solar Radiation Sensor, 4-20 mA

Optional Parts / Accessories
Cable

Additional Cable available upon request

MICRO
SWITCH
USA
U1

Texas Electronics, Inc.

P1

P2

5529 Redfield St.
Dallas, Tx 75235

C3
D3
TB1

Remove cover of TB-2012M and adjust

D2

for local pressure.

R21
D4

Refer to Pressure vs Voltage/Current

D1

Table under the Barometer section of
+12V

COM.

SIG.

COM.

the service manual.

C99

C1

Model TB-2012M Barometer

R3
R4
R2
R1
R18
R19
R20

C98
R13
R14
R16

R22

REF

C5

C7

R17
C6

ELECTRIC BAROMETER
MODEL NO. TB-2012M
S/N 102795

C4

U2
R11
R12
R8
R9

SPAN

C2
U3

R15
NULL

MODEL TB-2012M BAROMETRIC PRESSURE SENSOR
(Shown with cover removed)

DESCRIPTION
The Texas Electronics, Inc. TB-2012M Barometric Pressure Sensor uses an active solid-state device to sense
barometric pressure. Self-contained electronics provide a regulated voltage to the solid-state sensor and
amplification for the signal output. The unit's range of 26" to 32" or 878 mb to 1080 mb of mercury allows it to
be used at elevations up to 1800 feet or 548.64 meters above sea level. The unit is temperature compensated
from -18° to +50°C. if elevations above 1800 feet or 548.64 meters are required, contact the factory for higher
elevation calibration.
SPECIFICATIONS
Calibration Range:
Supply Voltage:
Current Draw:
Accuracy:
Operating Temperature Range:
Calibrated Temperature Range:
Output:
Optional Output:
Weight:
Dimensions:

26" to 32" or 878 mb to 1080 mb
12 to 15 VDC
<15 mA
+/- 1.3mb
-40° to +50°C
-18° to +50°C
0-1 VDC
4-20 mA
2 lb. or 8.9 N or .907 kg. w/ 60 ft. or 18.3 meters cable
6" or 15.24 cm H x 5" or 12.7 cm W x 3" or 7.62 cm D
(Single enclosure: double enclosure also available)

Warranty:

3 year

FEATURES & BENEFITS
•
•
•
•
•
•
•

Interfaces to virtually all data acquisition systems
Can be used up to 1800 ft. elevation without factory modification
Over 1800 ft. elevation applications available
Easy installation and maintenance
Over 5 years in production
Weatherproof Nema Enclosure for superior outdoor protection
Wide range of 26" to 32" Hg. or 878 to 1080 mb

INSTALLATION & MAINTENANCE
Select a site where the instrument will not be subject to rapid fluctuations of temperature or to jarring and
continuous vibration. Avoid exposing the instrument to direct sunlight or radiant heaters and to direct drafts
such as open windows and doors. A mounting bracket with hardware is attached to the Nema enclosure of the
sensor.
ORDERING INFORMATION
Model #

Description

TB-2012M
TB-2012MA

Barometric Pressure Sensor
Barometric Pressure Sensor, 4-20 mA

Optional Parts / Accessories
Cable
High Elevation:

Additional Cable
Applications of higher than 1800 ft. or 548.64 meters above sea-level require factory
modification.

PROPER EXPOSURE OF METEOROLOGICAL INSTRUMENTS
Generally recognized guidelines follow which depict "ideal" sensor mounting locations. These guidelines or
"rules of thumb" are only suggestive in nature in an attempt to aid the user to selecting optimum representative
sampling locations for a particular sensor.
Reference was made to US Weather Bureau Installation criteria in preparing this data (See Reference 1).

WIND EQUIPMENT:
So far as available sites permit, wind sensors should be placed above the ground on a freely-exposed tower (20
feet or higher) and over terrain that is relatively level and free from obstructions to wind flow. When a
compromise must be made, sensing units should be exposed at least 12 feet above any obstruction within 100
feet and at least as high as any obstruction within 100 to 200 feet of the wind equipment. Support towers or
masts should not be of such bulk or shape as to create an appreciable obstruction to wind flow. Avoid sites
where local obstructions may create up-or-down drafts, eddy currents or jet-flow effects. When sensors are
roof-mounted, they should be installed at least 10 feet (or greater) from the roof surface depending upon the
particular installation site. Turbulence and other local effects can be reduced somewhat by mounting sensors on
the upwind and of the building (that end of the building exposed to the most common local prevailing winds).
Horizontal-mount booms which extend from existing towers should be fabricated so that sensors will extend a
distance of 5 to 10 feet from the tower assembly (dependent on tower thickness).
Wind direction sensors are oriented upon installation in reference to either true north or magnetic north. True
north is obtained by applying a local magnetic variation correction factor to a magnetic north compass
indication (magnetic variation for a particular locality is obtainable from the nearest Weather Bureau Branch
Office). Indicator readings for a true north sensor orientation will then be in terms of true geographic compass
points. All U.S. Weather Bureau surface wind data used for observational network reporting purposes and
general public use is given in reference to this true north format. Indicator readings for a magnetic north sensor
orientation will be in terms of actual readings as would be obtained from directly viewing a magnetic compass
instrument. Wind direction data at Federal Aviation Agency and other aircraft reporting facilities (for direct
control tower-to-pilot utilization) is always made in reference to this magnetic north format.
REMOTE TEMPERATURE/HUMIDITY SENSORS AND INSTRUMENT SHELTERS:
Whenever possible, instrument shelters* as well as remote temperature and/or humidity sensors should be
installed at a height of 4 feet (or greater) over earth or sod at least 100 feet from any concrete or other hardsurfaced area and not closer to any other object than four times the height of the object above the instrument
shelter or remote sensors. Avoid roof installations if possible. If it is necessary to roof-mount shelters and
sensors, they should not be closer than 30 feet to any large, vertical reflecting surface (walls, etc.), exhaust fans,
or cooling towers. Electronic remote sensors when roof-mounted should be at least 9 feet (or greater) above the
roof surface. To minimize radiation effects from the roof, they can also be mounted on a horizontal boom so
that they will extend from the side of a building roof or existing tower. Horizontal booms should extend
approximately 5 to 10 feet from the side of the building roof or tower assembly.

*

Standard U.S. Weather Bureau cotton-region shelter (Spec. No. 450.0615, Rev. 8/67)

PRECIPITATION GAUGES:
Rain gauges should be installed on a level plot of ground, at a distance from any object of at least two and
preferably four times the height of the object above the top of the gauge. All types of gauges must be exposed
with the rim of the receiver in a horizontal plane and at a level well above the average level of snow surfaces.
Roof-mounting of rain gauges should be avoided when possible. Air currents at heights other than at ground
level have been observed to cause an apparent decrease in rainfall catch commensurate with the increase in
mounting height above ground level.
Objects which individually or in small groups constitute a "windbreak" reduce prevailing wind speed in the
vicinity of the gauge. This reduction of wind speed will, as a consequence, also reduce possible eddy currents
and turbulence around the gauge. The presence of such objects is usually beneficial in providing a more
accurate rainfall catch. Ideally, the "windbreak" objects (fences, bushes, etc.) should be generally uniform in
height and distance from the gauge. Height above the gauge should not exceed about twice their distance from
the gauge.
ANEROID BAROMETERS - SELF-CONTAINED MECHANICAL INSTRUMENTS AND
ELECTRONIC REMOTE BAROMETRIC PRESSURE SENSORS:
Select a site where the instrument will not be subject to rapid fluctuations of temperature or to jarring and
continuous vibration. Avoid exposing the instrument to direct sunlight or radiant heaters and to direct drafts
such as open windows and doors.
Reference 1:
U.S. Department of Commerce - National Weather Service Bulletin
LS 5927 Revised, 0-4.12, January, 1963.
SOLAR RADIATION SENSORS:
The Solar Radiation Sensor is normally mounted on a level surface totally remote from trees, poles, or power
lines that might cast a shadow on the sensor at any time of the day. However, there may be occasions, because
of extreme latitudes, when it is desired to mount the sensor at some angle other than level.
The sensors may also be mounted on a sun tracking mechanism or behind a shadow band if diffuse sky
radiation is to be measured.
WIND DIRECTION & SPEED SENSORS
INSTALLATION
These instructions apply to roof-top installation. We advise that you first read over these instructions before
beginning assembly as several referenced items are not supplied with your weather equipment (this is because
most every installation is unique thus these parts are best obtained by the installer). Reference to the "U-Tube
Cross-Bar Installation" figure and ".Sensor Installation" figure may be helpful.
Step 1: Attach the three anemometer cups to the speed sensor head.
Loosen the three set screws on the top of the anemometer (lower unit).
Insert the anemometer cup arms into the holes. Be sure to press the arms all the way in and make
sure that the flat areas on the arms face toward the set screws.
Tighten the set screws.

Step 2: Attach wind vane and counter-weight to the direction sensor head.
Loosen the two set screws on the top of the wind vane (upper unit).
Insert the vane and counter-weight into the holes. Be sure to press both parts all the way in and
make sure the flat areas on each arm face the set-screws.
Tighten the set screws.
Note: For optimum performance and maximum bearing longevity you may wish to fine-tune the
balance of both wind sensors. Place the U-tube flat on a table such that the sensors hang over the
edge. Rotate the vane and the cup in 10 degree increments. After positioning the vane and cups
verify that there is no movement after releasing your hold (this must be done in a wind-free
environment). Balance adjustments are made by loosening the set screw to the lighter cup,
counter-weight or fin and shifting it slightly away from the sensor head.
Step 3: Attach cross-bar to U-tube.
Spread end clamps and slide over the U-tube.
Insert cross-bar into the ends of both clamps.
Fasten cross-bar in a level position with screws, nuts and washers.
Step 4: Attach U-bolts to cross-bar and U-tube.
Remove the two nuts and reinforcing plate from both U-bolts (do not remove the toothed bracket).
Insert one U-bolt through the two holes in the cross-bar and the other through the two holes in the
bottom of the U-arm (be careful not to damage the wires inside the U-arm).
Replace the tube reinforcing plate on the U-bolt and replace the U-bolt nuts.
Step 5: Slip the U-bolts over the mast and tighten.
Make certain that the anemometer cups do not hit the mast.
Step 6: Attach guy wire clamp just below the U-tube assembly.
Step 7: Attach base mount to the roof or side wall.
Note that the base mount U-bolt will rotate to fit any angle.
Step 8: Install guy wire anchors (not included) or locate secure points for guy wire attachment.
Step 9: Erect mast and install guy wires (not included) and turn-buckles (not-included).
Step 10: Ground the mast to help protect the sensors and structure from lightning hits.
Supplies needed: mast wire clamp, grounding wire, wire supports and grounding rod.
Step 11: Run the sensor wire inside to the console.
Lead in wire is permanently attached to the sensor unit.
Attach to console according to wire color code.
If necessary the cable may be cut down in length or wire may be added with negligible effect on the
calibration. If changing cable lengths more than a few hundred feet you may wish contact the
factory to determine the severity of the effect on calibration.
Additional cable lengths are available from Texas Electronics if needed.
Step 12: Calibrate the Wind Vane.
Be sure console is operating properly first.
This is normally a two man job with one individual watching the direction indicated by the weather
station and the other adjusting the sensor while watching a compass.

Two methods of aligning the vane are available. The first method involves loosening the large set
screw at the bottom of the wind direction sensor so that it will rotate on the U-arm. Turn the bottom
half of the sensor until the compass readings and the indication match then retighten the set screw.
The second method involves rotating the entire mast assembly until proper orientation is achieved;
this technique is usually easier because of the heights involved but will usually necessitate
repositioning (rotating) of the guy-wire clamp.
If winds are creating rapid fluctuations in the vane making calibration difficult, the vane can be
temporarily secured in a fixed position by carefully wedging a thickly folded piece of paper or
cardboard into the gap between the upper and lower halves of the direction sensor. An alternative
technique is to lap a string over the vane and carefully hold it in position (be careful not to bend the
vane when using this approach).

Warranty
Texas Electronics, Inc. (hereafter TEI) warrants the equipment manufactured by it
to be free from defects in material and workmanship. Upon return, transportation
charges prepaid to TEI, within three (3) years of original shipment of sensors and
one (1) year of original shipment of electronics, recorders and indicators, TEI will
repair or replace, at its option, any equipment which it determines to contain
defective material or workmanship, and will return said equipment to purchaser,
F.O.B., TEI. Texas Electronics shall not be obligated however to repair or replace
equipment which has been repaired by others, abused, improperly installed,
altered or otherwise misused or damaged in any way. TEI will not be responsible
for any dismantling, re-assembly, or reinstallation charges.
This warranty is in lieu of all other warranties, expressed or implied. TEI shall
not be liable for any special, indirect, incidental or consequential damages
claimed in connection with any rescission of this agreement by purchaser.

For a list of specific items covered by the extended warranty, see the Three-Year
Warranty Equipment List.

Three-Year Warranty Equipment List
Effective February 1, 1992 all of Texas Electronics, Inc. sensors will carry a Three-Year warranty instead of the previous
One-Year. The remainder of terms and conditions of the warranty remains unchanged. A specific list of items follows.

Sensors Covered by Three-Year Warranty
Parameter
Wind Direction

Wind Speed

Barometric Pressure
Relative Humidity
Rainfall
Temperature

Solar Radiation

Model No.
TD-105 (Synchro)
TD-104D (Potentiometer)
TD-110-L2 (Photo-Chopper)
TD-106 (Potentiometer)
TV-110-L2 (Photo-Chopper)
TV-110-L3 (Photo-Chopper)
TV-114 (A.C. Generator)
TB-2012
TH-2013
TH-2013V
TR-525
TR-6118
TT-101 (Outdoor)
TT-103R (Surface Mount)
TT-103R-W (Water Probe)
TT-309I (Indoor)
TS-100

Systems Covered by Three-Year Warranty
Model No.
WSC-5-S
WSC-5-ST
WSC-5-D
WSC-5-DT
WDC-2

Description
Wind Speed Controller
Single Set Point
Wind Speed Controller
Single Set Point with Time Delay
Wind Speed Controller
Dual Set Point
Wind Speed Controller
Dual Set Point with Time Delay
Wind Direction Controller



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : No
Page Count                      : 21
Producer                        : GPL Ghostscript 8.15
Create Date                     : 2007:08:13 14:53:12
Modify Date                     : 2007:08:13 14:53:12
Title                           : Microsoft Word - CWS.doc
Creator                         : PScript5.dll Version 5.2
Author                          : Jason
EXIF Metadata provided by EXIF.tools

Navigation menu