Field Intelligent Intelligent Device Device Series Series Field Electromagnetic Flowmeter Converter Sanitary Electromagnetic Flowmeter LF232 Introduction The electromagnetic flowmeter uses Faraday’s Law of electromagnetic induction to measure the process flow. The device consists of two units: a detector, through which the fluid to be measured flows and in which low-level signals proportional to flow rates are obtained; and a converter, which supplies excitation current to the detector, and amplifies the signals from the detector and then processes and converts the signals into the 4–20 mA dc current signal. The LF232 offers various outputs such as totalization output and multi-range selection outputs. With it patented Noise-Sentry original noise-suppression circuit and advanced algorithms. The LF232 is highly resistant to noise and provides a stable output even when measuring fluids contain slurries. The LF232 can be used with a separately mounted detector such as the LF150 or LF470. The LF232 can also be used with conventional models of detectors. The AF900 hand-held terminal (HART*1 communicator) can be used to communicate with the flowmeter from a remote place. See the Communications signal specification for details about HART protocol. *1: HART protocol (Highway Addressable Remote Transducer) is a communication protocol for industrial sensors recommended by the HCF (HART Communication Foundation). Specifications Overall Specifications Figure 1. LF232 Converter Accuracy: the accuracy is dependent on the type of detector combined with the LF232. See the following tables for accuracy when combined with specified detectors. ・Detector combined LF470 (Meter size 1/10", 1/6”, 1/4" (2.5 to 6 mm)) Flow rate as a percent of range 0~50% 50~100% Accuracy 0.3–1.0m/s 1.0–10m/s ±0.8%FS ±0.4%FS ±0.8%FS ±0.8% of rate ・Detector combined LF150 (Meter size 28” to 120” (700 to 3000 mm)) Measurement range in terms of flow velocity: Accuracy: ±0.5 % of Rate* Conbination LF150, LF470 0 –1.0 ft/s to 0 – 32.8 ft/s (0 – 0.3 m/s to 0 –10 m/s) * This pulse output error result is established under standard operating conditions at Toshiba’s flow calibration facility, Fuchu Japan. * Individual meter’s measurement error may vary up to ± 0.8 % of Rate at 3.28 ft/s (1.0 m/s) or more and ± 0.4% of Rate ±0.157 inch/s (4mm/s) at 3.28 ft/s (1.0 m/s) or less. * Current output: plus ± 8µA (0.05% of span). * Refer to individual calibration data for each individual meter’s measurement error. Fluid conductivity: 5µS/cm or more Ambient temperature: -4 to 140 °F (-20 to 60 °C) (Storage temperature: -13 to 149 °F (-25 to 65 °C)) Structure: IP67 and NEMA 4X Watertight TIC-LF232A TIC-LF232A Power supply: One of the following can be selected ・100 to 240 Vac (Allowable voltage 80 to 264 Vac, 50/60Hz)*1 ・24 Vdc (Allowable voltage 21 to 27 Vdc)*2 *1 When the 7th digit of specification code is “B”, power supply is 100 to 120Vac (Allowable voltage range is 80 to 132VAC, 50/60Hz) *2 When the 7th digit of specification code is “B”, 24Vdc cannot be selected. Power consumption: 17W (27VA) or less (When 7th digit of specification code is “A”) 35W (120VA) or less (When 7th digit of specification code is “B”) Converter Specifications Input signals Analog signal — the voltage signal from detector, proportional to process flow rate. Digital input DI (opt.) (added when 9th digit of specification code is ”3”) Voltage level: High level 20 to 30 V dc Low level 2 V dc or less Input resistance: 2.7 kΩ Number of inputs: 2 points (DI1 and DI2) DI functions: For digital input function, either of the following functions can be selected. • Range switching One Digital Input is used: switching between large and small ranges of forward/reverse, 2-range measurement. Two Digital Inputs are used: switching between ranges of single direction, 4-range measurement. • Totalizer control input Totalizer Start/Stop control or Reset/Start •Output hold input Fixed with set value for current output and pulse output (loop check) • Zero adjustment Starts zero adjustment (on-stream at zero flow rate) when DI voltage level goes low after remaining high for 10 to 20 seconds. 2 Output signals Current output: 4–20mAdc (load resistance 0 to 1KΩ) Digital output (DO) — One point (std.) and three more point is optionally available as follows. Digital output DO1 (std.): Output type: Transistor open collector Number of outputs: One point Output capacity : 30Vdc, 200mA maximum Digital output DO2~DO4 (opt.) (Added when 9th digit of specification code is “3”) Out put type: Solid-state relay output (non polarity) Number of outputs: 3 points Output capacity: 150Vdc, 150mA maximum or 150Vac (peak to peak), 100mA maximum DO functions—Four digital output function, either of the following functions can be selected. • Totalizer pulse output DO1 or DO2 can be selected (Forward direction pulse and reverse direction pulse can be assigned independently) In the case of DO1 Pulse rate: 3.6 to 3,600,000 pulses/h Pulse width:Settable within the range of 0.3 to 500ms In the case of DO2 Pulse rate: 3.6 to 360,000 pulses/h Pulse width: Settable within the range of 4 to 500ms • Rage switching output One Digital Output (DO) is used ・Single direction, 2-range switching signal ・Forward/reverse direction switching signal Two Digital Outputs (DO) are used ・Single direction, 4-range switching signals ・Forward/reverse 2-range switching signals • High/low limit alarm output An alarm is output when flow rate goes above or below the set-point value Setting range: -10 to 110% of the settable maximum range High limit 2 points, low limit 2 points can be set At the time of alarm output, Normally Open or Normally Closed contact can be selected • Preset point output Contact ON when totalizer count exceeds the set value Setting range: 1 to 99999999 count • Converter error alarm An alarm is output when an operation error is detected by self-diagnosis. At the time of alarm output, Normally Open or Normally Closed contact can be selected TIC-LF232A Communications signal: A digital communications (HART protocol) signal is superimposed on 4 to 20 mA dc analog output signal. Load resistance: 240 Ω to 1 kΩ Load capacitance: 0.25 µF or less Load inductance: 4mH or less (For maximum cable length, about 2km is a guideline length when CVV-S 1.25mm2 is used under standard installation condition.) Note: HART (Highway Addressable Remote Trans-ducer) is a communications protocol for industrial sensors recommended by the HCF (HART Commu-nication Foundation). Output display: 16-character×2-line dot-matrix LCD (With back light). 2 units can be selected from the flowing units: flow velocity, instantaneous flow rate, total flow (forward/reverse/difference flow), total count, %, custom unit. Parameter settings—Parameters can be set as follows •IR Switches: Various parameters can be set without opening the converter housing using 4 infrared switches (password can be set) •Zero adjustment: Zero point adjustment can be started by pressing the switch in the converter. •Digital communication: The AF 900 hand-held terminal is needed to set parameters. •Damping: 0.5 to 60 s (selectable in 1s increments) “Field re-verification” Mag-Prover – Toshiba’s Zero span calibration tool allows unit to be re-calibrated and verified using internal software program. (For more information contact Toshiba International Corp.) Conditions when power fails: Parameter setting values are stored in non-volatile memory and the values will be restored when the power returns to normal condition. The output and display will stay as follows when power fails. Current output: 0 mA Digital output: OFF (contact open) LCD display : No display Surge protection: Surge protectors are installed in the power supply, excitation circuit, current signal output and digital I/O circuit. Terminal block structure: 21-pole, screw connection type (M4 screw) Housing: Aluminum alloy Coating: Acrylic resin-baked coating, pearl-gray colored Cable connection ports : A cable gland is provided for each port. OD of cable φ11 to 13 mm Material Nylon 66 G (PF) 1/2 male screws. Vibration resistance: No resonance to the following levels of vibration: • 10 to 55Hz with amplitude of 0.07mm • Vibration of 30Hz with 29.4 m/s2 in 4h in each direction will not cause any defect to unit. Note: If the flowmeter is intended to be used in a location where vibration is applied constantly contact Toshiba. Dimension and Weights: See Figure 2 3 TIC-LF232A Installation Dimensions 9.33 (237) 8.74 (222) 4-φ12 hole Installation method 据付方法 ・Installed by panel or pipe 据 付 け は 、パ ネ ル ( 壁 面 ) 取 付 け か ・ Appointed bout and nut for パ イプ取 付 けで行 います。 installation 指 定 され た取is付attached. 用 ボ ル ト・ナットが 付 6.10 (155) 2.91 (74) 属 します。 In the caseイofプ2B 2 Bパ 取 pipe 付 けinstallation の場合 15.75 400 (400) 14.57(370) 15.75 (400) M 10 UUボ ルト M10 bolts 2Bm )pipe 2B( 50m パ イプ Output cable ground Power supply cable ground In the case of panel installation パネル取付穴加工寸法 1.57 2.17(55) (40) 2.9174 (74) Grounding terminal 2.99 (76) 2.99 (76) Excitation cable ground I/O cable ground 2.99 (76) Weight : Approximately18 lb (8 kg) (including a mounting bracket) Figure 2. Separate type converter LF232 4 11 to 12 φhole 11~ 12φ 穴 (4ヵ所 ) 4 places 370(370) 14.57 Signal cable ground TIC-LF232A External Connections Terminal box Power supply (CVV) Ⅳ wire 5.5mm2 or more Power supply Grounded with 100Ω or less ground resistance Current output (4 to 20mA) Digital input 2 (opt.) Thick walled steel conduit Digital input 1 (opt.) (20 to 30 Vdc) Digital output 4 (opt.) Digital output 3 (opt.) Digital output 2 (opt.) Detector Signal cable (2-wire shielded hard-rubber sheathed cable) Excitation cable (3-wire hard-rubber sheathed cable) Digital output Grounded with 100Ω or less ground resistance Output cable (CVV-S) Digital I/O cable (CVV-S) Figure 3. External Wiring Diagram 5 TIC-LF232A Wiring Precautions (1)Be sure to use thick walled steel conduit (22mm) for signal and excitation cable wiring between the detector and converter. Use flexible conduit at the cable outlets of the detector and converter. (2) Connect the grounding wire (IV wire 5.5mm² or more) to a good earth ground (100Ω or less ground resistance). Make the wire as short as possible. Do not use a common ground shared with other equipment where earth current may flow. An independent earth ground is recommended. (3) The allowable cable lengths between the detector and converter for the separate type flowmeter depend on the electrical conductivity of the object fluid. See Figure 4 below. (4) Digital output 1, digital output 2 to 4 , digital output 1 to 2 are insulated mutually. 300 200 Allowable cable length (m) ケーブル許容長さ 100 50 30 ① 20 ② 10 〔m〕 5 3 3 5 10 20 30 50 100 200 Conductivity (µS/cm) 導電率〔μS/cm〕 ① LF130形、LF131形、LF150形検出器と接続の場合 (1)When combined with LF150. ② LF470形検出器と接続の場合 (2)When combined with LF470. Figure 4. Electrical Conductivity and Cable length 6 About establishment environment Do not store or install the flowmeter: • Where there is direct sunlight. • Where excessive vibration or mechanical shock occurs. • Where high temperature or high humidity conditions exist. • Places where corrosive atmospheres exist. • Places submerged under water. • Place where there is sloped floor. To put the flowmeter temporarily on the floor, place it carefully with something, such as stopper, to support it so that the flowmeter will not topple over. In areas like the following, there may be the case that infrared switches do not function correctly. (If these are unavoidable, use an appropriate cover.) (1) Where unit (operation panel) is exposed to direct sunlight, reflection of light onto window pane and diffused light reflection. (2) Where smoke and steam may occur. (3) Where exposed to direct snow, ice or mud. Ordering Information 1. When ordering the LF232 converter, refer to Tables 1 (Type Specification Code). An entry must be made for each of the columns in each of these tables. 2. Measuring range 3. I/O function specifications. 4. Ordering scope: Actual flow calibration data: (required or not) 5. Specification of combined detector 6. Other scope: Specification other than standard items. Consult a Toshiba representative before ordering. TIC-LF232A Table 1. Specification Code for converters (LF232) Model Specification Code 1 2 3 4 5 6 7 8 9 10 11 12 13 14 L F 2 3 2 Separate type converter A A B C E 1 3 1 1 2 A Contents Purpose Standard Compatible detectors (Note1,2,3) Medium to small-size detectors: Meter size 1/10” to 18” (2.5 to 450 mm) Large-size detectors: Meter size 28” to 120” (700 to 3000 mm) Mounting nuts and bolts Panel, wall mounting (BNP material : SUS304) Pipe mounting (BNP material : SUS304) Digital input/output Current output + Digital output points (1 point) Current output + Digital output points (4 points) + Digital input points (2 points) Communication function HART communication Power supply 100Vac-240Vac, 50/60Hz (Note 2) 24Vdc (Note 3) Standard Note 1: For applicable detector code, select one from the following table. To combine with an existing detector, select one depending on the type of replaceable converters. For combination with detectors not listed below, contact Toshiba. Applicable detector code Applicable detector A LF470 type B LF150 type Applicable detector Replaceable converters (discontinued model) 334, 372, LF200, LF220, LF420, 335 type (Converters of LF230* A type types shown right) 337, 373, 378, LF230* B type 335 type (Converters of types shown right) Note 2: When the 7th digit is “B” and power supply specification code is “1”, power supply specification is 100 to 120VAC. Note 3: 24VDC can be specified only when the 7th digit is “A”. 7 TIC-LF232 ISO9001 and ISO14001 are certified. Misuse of this product can result in damages to property or human injury. Read related manuals carefully before using this product. TIC-LF232A Specifications are subject to change without notice. Printed in Japan 2008-5 (TDOC) © TOSHIBA Corporation 2008 All Rights Reserved. http://www.toshiba.com/ind/
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.4 Linearized : Yes Page Count : 8 XMP Toolkit : XMP toolkit 2.9.1-13, framework 1.6 About : uuid:40258f03-6c91-4864-babc-f345b1adb45e Producer : Acrobat Distiller 6.0 (Windows) Create Date : 2008:05:12 17:10:35+09:00 Modify Date : 2008:05:12 17:10:35+09:00 Document ID : uuid:b8e91c9a-807e-4e0e-859b-740ef370471f Format : application/pdfEXIF Metadata provided by EXIF.tools