True Systems MSAMINI Smart Beauty Analyzer User Manual

True Systems Co., Ltd. Smart Beauty Analyzer Users Manual

Users Manual

Download: True Systems MSAMINI Smart Beauty Analyzer User Manual
Mirror Download [FCC.gov]True Systems MSAMINI Smart Beauty Analyzer User Manual
Document ID3830862
Application IDuPIpCc9FUOowc32p1ZKPnw==
Document DescriptionUsers Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize192.79kB (2409929 bits)
Date Submitted2018-04-26 00:00:00
Date Available2018-04-26 00:00:00
Creation Date2018-03-16 10:59:13
Producing SoftwareMicrosoft: Print To PDF
Document Lastmod2018-04-18 17:56:41
Document TitleMicrosoft Word - MSA mini_Manual (ver.0.95)_ìł’í–“ì€œì¶œ
Document Author: tifun

Portable Counseling Device
MSA mini
User’s Manual
ver.0.95
Contents
1. Safety Information
…………………………………………………………………………
2. Starting MSA mini
………………………………………………………………………………
…………….
…………………………………………………………………………………
2.2 System Components
…………………………………………………………………..
2.3 Parts Name
…………………………………………………………………..
…………………………………………………………………..
……………………………………………………………………….
……………………………………………………………….
4.1 Device Booting
……………………………………………………………………
………....
……………………………………………………………………
………....
…………………………………………………………………
…………..
…………………………………………………………………………...
4.2 Main Screen
……………………………………………………………………………
4.3 Start Measurement
……………………………………………………………………………
2.1 Introduction
3. Main Unit Operation
3.1 Power On/Off
3.2 Charging on Cradle
3.3 Battery Replacement
3.4 Adjusting Focus
4. Program Preparation
5. Measurement Procedure
5.1
5.2
5.3
5.4
5.5
5.6
………………………………………………………………
….
Measurement Item Selection ……………………………………………………………
………….
…………………………………………………………………
Scope Mode
…….
Moisture Measurement ……………………………………………………………………
….
……………………………………………
Counseling Procedure in SKIN CARE
………………………….
Counseling Procedure in SCALP/HAIR ……………………………………………
………………………….
Analysis of Results
……………………………………………………………………….
10
11
23
13
19
22
6. Product Specifications
…………………………………………………………………….
25
7. Warranty and Service
…………………………………………………………………….
26
…………………………………………………………………..
26
7.1 Terms of Warranty
7.2 Warranty Card
………………………………………………………………….. 27
1. Safety Information
Warning
Indicates a danger that may result in death or serious injury if
used improperly.
Caution
Indicates that a danger that may result in injury or properly
damage if used improperly.
Note
Indicates that a possibility that may result in reducing the
performance of the MSA mini, or disabling some functions or MSA
mini itself.
Warning
●Should the product have an abnormality such as producing smoke or an
unpleasant odor, stop operation. If you continue use, this may cause fire or
electric shock. If you find any abnormality, unplug the connector from the power
source or switch power off, and contact the supplier/dealer where purchased or
our company for repair. NEVER service this product by yourself as this is
extremely dangerous and will void the product warranty.
●Should any liquid such as water enter the inside of the MSA mini, unplug the
connector from the power source or switch power off, and contact the
supplier/dealer where purchased or our company for repair. If you continue
using it, this may cause fire or electric shock.
●Do not modify or disassemble the product; otherwise, this may result in a fire or
an electric shock and will void the product warranty.
●Do not use a damaged cable. If the cable is damaged, contact the
supplier/dealer where purchased or our company. If you continue use, this may
cause fire or electric shock
●Do not damage, break, modify, bend forcibly, pull, twist, or bind the cable. Or
do not place a heavy object on the cable or heat it; otherwise, the cable may be
damaged, resulting in fire or electric shock.
●Do not plug or unplug the cable with wet hands; otherwise, you may suffer an
electric shock.
●When the product is transported from a cold location to a warm location, this
may cause condensation to occur. If this happens, do not connect the product to
the power source until moisture disappears; otherwise, this may result in fire or
electric shock.
Caution
●NEVER clean the MSA mini with pure benzene, thinner, or alcohol; otherwise
this may cause the discoloration or deformation. If it is heavily smeared, wipe it
off with a wrung-out lint-free cloth moistened with a diluted neutral detergent or a
very mild water & alcohol solution, and then dry it with a soft lint-free cloth.
NEVER wash the MSA mini directly with water.
●Do not allow a strong impact to the MSA mini by dropping or hitting it;
otherwise, this may result in a breakage and/or malfunction.
●A stained lens may cause unclear images. In this case, contact the
supplier/dealer where purchased or our company for repair
FCC Information to User :
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed
to provide reasonable protection against harmful interference in a residential
installation. This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this
equipment does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Caution :
Modifications not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment.
FCC Compliance Information :
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including interference
that may cause undesired operation.
2. Starting MSA mini
2.1 Introduction
MSA mini is a high quality portable device for skin, hair/scalp measurement and
analysis. Equipped with scope and moisture sensor, MSA mini performs image
capturing and analyzing to evaluate skin condition on a scale from 0 to 99.
Moisture sensor at MSA mini enables user to measure the degree of hydration of
skin stratum corneum.
2.2 System Components
- MSA mini Main Device with built-in lens ( 30x/120x dual lens )
- mini USB cable for charging
- User’s Manual
Main Unit
Manual
Mini USB cable
2.3 Parts Name
Moisture sensor ( pin type )
USB charging
Power button
Rotary lever for adjusting lens
magnification and focus
3. Main Unit Operation
3.1 Power On/Off
Power On/Off of MSA mini is done by a simple button operation. By pressing the
power button for about 2~3 seconds, the MSA mini will be turned on with beep
sound. It will take about 40 ~50 seconds to boot up the Android OS. To power off
the MSA mini, press the power button for about 2~3 seconds until the power off
message is shown and select ‘Yes’ button.
3.2 Charging the MSA mini
The MSA mini can be charged by connecting the usb port with external USB
power cable. When user plugs USB power cable on MSA mini, the green LED
indicator will be turned on left top of LCD, and charging symbol on the battery
symbol on screen will be displayed.
Note: The LCD screen of MSA mini will be turned off automatically after 10
minutes without any operation to save the battery consumption. And the LCD will
be On if power button is pressed slightly.
⋇ Battery charge status icon
Battery charge status icon is located in the right top of LCD Screen.
On using
On charging
3.3 Battery Replacement
If the battery is not charging or discharged too quickly, the battery needs to be
replaced. You may contact your supplier for battery replacement or follow these
instructions after getting consulting from the provider of MSA mini.
To replace the battery, unscrew the battery cover on the rear side of MSA mini
and simply change the battery. The screw may not be seen, because it is taped in
white sticker. You should detach the sticker before unscrewing the screw on
battery cover.
Unscrewing
3.4 Adjusting lens focus
MSA mini has dual lens with 30x and 120x magnification lens which can be
adjusted by rotating knob. To change the magnification 30x or 120x, pull the knob
for 30x, push the knob for 120x. To adjust the lens focus, rotate the knob slightly in
the direction to see skin image clearly.
4. Program Preparation
4.1 Device Booting
To boot on the device, press the power button for 2~3 seconds until hearing a
beep sound. The booting image will be displayed for
approximately 40 seconds. Wait until the Main screen will
appear.
Booting Screen
4.2 Main Screen
The main screen has two modes. One is the SKIN CARE mode and the other is
the SCALP/HAIR mode. You can choose SKIN CARE or SCALP/HAIR by
selection of the button.
4.3 Start Measurement
: Customer Information
By choosing [SKIN CARE] or [SCALP/HAIR] in the Main screen, “CUSTOMER
INFORMATION” screen will be displayed as below. User needs to check the
gender and age and press ‘START’ to start the measurement.
5. Measurement Procedure
5.1 Measurement Item Selection
User can choose individual measurement item for counseling.
The [SKIN CARE] measurement items include
Whole Measurement : to analyze whole items
Basic : to measure skin texture, sebum and moisture
Color : to measure skin color and pigmentation
Functional : to measure sebum, pore and elasticity
Eye Wrinkle : to compare the eye wrinkle with standard sample image
Scope : to see the image depending on each lighting mode
( Normal, PL, UV mode )
Sensor : to measure moisture on skin
1) Select any measurement menu and put the MSA mini on cheek or eye
wrinkle to be measured. And wait for a second or press the ‘Analysis’ button
2) All measurements will be done by predefined step automatically, so just put
the scope on skin for some seconds.
And [SCALP/HAIR] measurement items include
Whole Measurement : to analyze Sebum, Keratin, Sensitivity and Density
on scalp
Scope : to see the image depending on each lighting mode
( Normal, PL, UV mode, 120x mode )
1) Select ‘Whole Measurement ‘to start analysis or ‘Scope’ menu to see the
live image
2) Put the MSA mini on scalp and wait for a second or press the ‘Measure’
button
3) Measurements will be done by predefined step automatically, so just put the
scope on scalp for some seconds.
5.2 Scope mode
In the Scope mode of SKIN CARE, lighting mode can be changed, Normal, PL or
UV lighting. And in the Scope mode of SCALP/HAIR, lighting mode can be
changed to Normal, PL, UV lighting and 120x mode.
[Measurement items and the tools for each measurement]
Parameter
Lighting Mode
Measurement Target
Skin Texture
Normal
Skin
Pigmentation
PL
Skin
Porphyrin on Pore
UV
Skin, Scalp
Hair condition
120x Normal
Hair
Live image can be frozen by [FREEZE] button and released by [LIVE]. Lighting
mode can be changed by sliding up menu.
[ Lighting mode ]
NORMAL : texture mode lighting
PL : pigmentation mode lighting
UV : UV mode lighting
FREEZE : to stop the live image
LIVE : to release the frozen image
10
Step 1: Position
Place the scope on target. For [Skin texture], [Pigmentation], [Skin Color], [Pore],
[Eye Wrinkle], skin should be a direct target for measurement. Meanwhile, for
[Sebum], [Keratin], skin indicators (sebum and keratin tapes) should be the target.
To get the best image, press lens slightly to skin and change the pressure until
get clean image. If the captured image is blurry, adjust focus of lens by rotating
the knob.
⋇ When captured image is blurry;
- Apply only slight pressure to MSA mini during the lens contacts on skin.
Strong pressure will reduce image quality
- Adjust focus of lens by rotating knob
- No make-up is recommended in the clients’ skin. Any make-up may cause
inaccurate measurements
Step 2: Freeze, Analysis
By pressing “Freeze” button, live images will be changed to frozen image.
5.3 Moisture Measurement
When user selects [SENSOR] in the menu, MSA mini prepares moisture
measurement. LCD Screen will look like the picture below.
Step 1: Preparation
Put the MSA mini on skin for the sensor pins to touch the skin equally and press
the sensor surface with same pressure and press [MEASURE] button.
11
Step 2: Moisture Sensor Ready
Press [MEASURE] button to start the moisture measurement. A beep sound
should be heard.
When moisture sensor is ready to measure on skin,
press the [MEASURE] button and will show the
result with beep sound.
12
5.4 Counseling Procedure in SKIN CARE
In [SKIN CARE] measurement screen, different measurement course can be
selected.
1. WHOLE MEASUREMENT
Select the [WHOLE MEASUREMENT] button and MSA mini will be ready to
measure the skin.
Put the MSA mini on face skin as in the picture and wait for camera mode
for some seconds or press [MEASURE] to start the measurement and wait
until beep sounds.
Whole
measurement
will take
images
and
analyze
skin
texture,
pigmentation, skin tone, pore, elasticity, keratin, sebum and moisture at one
time measurement. Elasticity is calculated based on roughness of skin by
image. Keratin and sebum are also calculated by image, which are related
to color and shining etc. ( These may not accurate comparing to use tapes )
After measurement, the result will be shown as following result screen.
13
Summary of skin condition with skin
level between 1 ~ 5 levels.
Result bar graph. Higher number
means better condition.
To go to SKIN CARE menu
To see the detail image, press [+] button in
each result screen
Skin Texture [+]
Pigmentation [+]
Porphyrin [+]
14
Moisture [+]
2. BASIC
Select [BASIC] button and MSA mini will be ready to measure skin texture,
sebum and moisture. Place the MSA mini on cheek and press the
[MEASURE] button to start the imaging and analysis.
And the result will be shown as
15
Summary of skin condition
Result score between 1 ~ 5.
Higher number means better
condition.
3. COLOR
In COLOR measurement, MSA mini will measure skin color and
pigmentation.
Place the MSA mini on cheek and press the [MEASURE] button to start the
imaging and analysis.
16
4. FUNCTIONAL
[FUNCTIONAL] will measure the sebum, pore and elasticity.
5. EYE WRINKLE
[EYE WRINKLE] will take image of eye wrinkle and compare the image with
standard sample image
Select [EYE WRINKLE] and place the MSA mini on eye rim and wait for
17
camera be ready. Freeze the eye wrinkle image to be compared and open
the standard sample images as in the picture. In case you want to image
the eye wrinkle again, press the [LIVE] or [MY SKIN] button.
6. SCOPE
In [SKIN CARE], three different lighting mode can be selected as Normal,
PL or UV mode to see the skin texture, pigmentation or porphyrin on pore.
Normal mode
PL mode
UV mode
7. SENSOR
To measure the moisture on skin, select the [SENSOR] and put the MSA
mini on skin to be measured and press the [MEASURE] button to start the
sensor measurement.
To measure the moisture again, MSA mini should be detached from skin
once to get accurate measurement.
18
5.5 Counseling Procedure in SCALP/HAIR
In [SCALP/HAIR] measurement screen, there are two measurement courses.
1. WHOLE MEASUREMENT
Select the [WHOLE MEASUREMENT] button and MSA mini will be ready to
measure the scalp/hair, such as sebum, keratin, sensitivity on scalp and
density of hair.
19
Place the MSA mini on scalp to be measured and press [MEASURE]
button to start the measurement. MSA mini will automatically take pictures
in different lighting conditions.
After a few seconds, your images for
sebum and keratin will be show to be compared
and
selected
with
reference images.
Others such as porphyrin and sensitivity on scalp and hair density will be
calculated automatically and shown in result screen. To see the images t
aken and compared, press [+] button in each graph.
20
Sensitivity[+]
Keratin[+]
Porpyrin[+]
Density[+]
Sebum[+]
2. SCOPE
Same as in Skin, the [SCOPE] function views scalp or hair according to
different lighting condition, Normal, PL, UV or 120x magnification.
21
With 120x lens mode, enlarged hair can be seen.
22
5.6 Analysis of Results
1. Grading and interpretation
1) Analysis steps
1. Capture of
skin image
2. Image
Processing
3. Analysis of
b/w image
4. Normalize to
score
5. Categorize by
Look-up Table
•Normal lighting
mode
•Polarized lighting
mode
•Conversion to
black and white
image
•calculate white in
% of black area
• all calculated
values will be
normalized to
values between 0
... 100
• A, B or C
2) Look-up Table for grades [Example]
[Image]
Grade
Age Group
<= 19
0 – 20
21 – 40
41 – 60
61 – 80
81 - 99
20 - 29
0 – 19
20 – 39
40 – 59
60 -79
80 - 99
30 - 39
0 - 18
19 - 38
39 - 58
59 - 78
79 - 99
40 - 49
0 - 17
18 - 37
38 - 57
58 - 77
78 - 99
50 - 59
0 - 16
17 - 36
37 - 56
57 - 76
77 - 99
>=60
0 - 15
16 – 35
36 - 55
56 – 75
76 – 99
23
2. Determination of skin types
In addition to moisture, MSA mini evaluates the amount of sebum on the
particular area of skin. This allows users to check the balance of both parameters
and determine the resulting skin type. MSA mini categorize the six skin types as
following diagram.
24
6. Product Specifications
Item
Details
Remarks
CPU
XBurst dual-core CPU, 1.2 GHz
MIPS Ingenic M200
Memory
DDR2
OS
Android 4.4.4
LCD
4.0" touch TFT LCD
Resolution
800 x 480 pixels
Camera
1/3" 2MP CMOS
MT9D131
Lens
Dual Lens( 30x, 120x )
Select by knob
Sensor
Skin Moisture
( 0 ~ 99 )
Pin type, Impedance
measurement
WiFi
802.11 b/g/n
Lighting
Data
Communication
White LEDs, UV LEDs
512MB, NAND 4GB
Normal/PL/UV mode
WiFi
Battery
Li-Polymer, 2,000mAh
3 hours continuous operation
Rechargeable
External I/O
USB
Only for Charging
Charging
USB cable
Dimension
Weight
Operating
Condition
180(L) x 68(W) x 67(T) mm
About 200 gram
0 ~ 40 °C
5 ~ 85 %RH
25
7. Warranty and Service
7.1 Terms of Warranty
1) Warranty information is included with this product. Be sure to record the
supplier name, purchase date, and warranty period when receiving the MSA mini
and keep this information in a safe place.
2) The warranty period of MSA mini is one year from the purchase date.
If any malfunction occurs during proper and normal use within the warranty
period, then free-of-charge servicing will be carried out by True Systems or
representative according to the terms of the warranty.
3) Service after the warranty period has expired will incur charges.
When functions of this device are judged to be restorable through repairs,
service can be completed with charges at the customer’s request.
4) Repairs and replacement of disposable / replacement articles should be
managed by the user and may be completed with charges.
5) Please contact the supplier where purchased or True Systems if there are any
concerns about details of repair items or service.
When requesting service, please provide following information.
 Product Name : MSA mini
 Serial Number :
 Purchase Date :
 Malfunction Description :
 User Name and Contact Information: (E-mail / Phone number)
26
7.2 Warranty Card
Model
MSA mini
Serial No.: MM-12345
Warranty period : 1 year from
Customer
(day)
(month)
(year)
Address :
Tel :
Name :
E-Mail
The product described on this certificate is hereby certified as having passed
stringent quality control and inspection prior to shipment and it is hereby subject
to warranty in accordance with the following conditions.
1. The manufacturer shall repair the MSA mini at the manufacturer’s expense if a
malfunction occurs during the warranty period under normal usage.
* This warranty card will not be reissued, so keep this carefully.
2. Even if the warranty is still valid, repair will be completed at the customer’s
expense under the following conditions:
a) If breakdown or damage occurs as a result of improper usage,
inadvertent handling errors, or modification of the product.
b) Breakdown or damage occurs as a result of fire, earthquake or other
natural calamity.
c) This warranty is not valid if the warranty card items are not completed,
the sales store details are not completed, or there has been any amendment
to the details.
Support Contact Details:
True Systems Co., Ltd.
D-603 Bundang Techno Park, 700 Pangyo-ro,
Bundang-gu, Seongnam-city,Gyeonggi-do, Korea (13516)
Tel
+82 – 31 – 712 – 8031
Fax
+82 – 31 – 712 – 8041
27
FCC Information to User
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation. This equipment generates,
uses and can radiate radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does
cause harmful interference to radio or television reception, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one of the
following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is
con-nected.
• Consult the dealer or an experienced radio/TV technician for help.
Caution
Modifications not expressly approved by the party responsible for compliance could void the
user’s authority to operate the equipment.
FCC Compliance Information : This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions: (1) This device may not cause harmful
interference, and (2) this device must accept any interference received, including interference
that may cause undesired operation.
2017 True Systems Co., Ltd. All rights reserved.
All designs and other registered symbols are the property of True Systems.
True Systems reserves the right to change the specification without prior notice.
Application program can be changed or updated without prior notice.
True Systems Co., Ltd.
www.truesystem.co.kr
info@truesystem.co.kr
28

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.7
Linearized                      : No
Author                          : tifun
Create Date                     : 2018:03:16 10:59:13+09:00
Modify Date                     : 2018:04:18 17:56:41+09:00
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Format                          : application/pdf
Creator                         : tifun
Title                           : Microsoft Word - MSA mini_Manual (ver.0.95)_ìł’í–“ì€œì¶œ
Metadata Date                   : 2018:04:18 17:56:41+09:00
Producer                        : Microsoft: Print To PDF
Document ID                     : uuid:4f3153a2-4a75-470e-961e-6be76213bc59
Instance ID                     : uuid:82a963d6-9cdb-4818-adbb-fe7679603c54
Page Count                      : 31
EXIF Metadata provided by EXIF.tools
FCC ID Filing: 2AO9H-MSAMINI

Navigation menu