Truly Instrument TBG-460 Blood Glucose / Cholesterol 2 in 1 Monitoring System User Manual

Truly Instrument Ltd. Blood Glucose / Cholesterol 2 in 1 Monitoring System

User manual

Blood Glucose / Cholesterol 2 in 1 Monitoring SystemUser's Manual2015/07/14  Rev.01www.trulyinstrument.comModel: TBG-460GC
Chapter 1 TRULY TBG-460GC  Monitoring System1.1 Introduction..............................................................................1.2 Precautions..............................................................................1.3 Intended Use............................................................................1.4 Labelling and Informations.....................................................1.5 System Components...............................................................Chapter 2 Setup the meter2.1 Set the Year/Month/Day/Hour/Minute/ Measurement unit........ Chapter 3 How to perform Test3.1 Before Testing..........................................................................3.2  Connect iOS device via Bluetooth..........................................3.3 Perform the Blood Glucose Test......................................................................................3.5 Special messages..................................................................3.6 Care after the test..................................................................Chapter 4  ............................................................Chapter 5 Meter Memory Function..............................................Chapter 6 Installing/Replacing the Batteries................................Chapter 7 Storage and Handling of Meter and Strips..................Chapter 8 Error Message and Trouble Shooting.......................... Chapter 9 Specification................................................................FCC Statement.............................................................................19 3.4 Perform the Total Cholesterol TestQuality Control223446791011121213 1415161618-1-Table of Contents
-2-Chapter 1 TRULY TBG-460GC  Monitoring System1.1 IntroductionThank you for choosing TRULY Blood Glucose / Cholesterol Monitoring System. Please read the "User's Manual" for your convenient and for safe use.The TRULY Blood Glucose / Cholesterol Monitoring System provides youwith accurate, plasma-calibrated results based on electrochemical sense technology and the superior capillary sampling. The TRULY Blood Glucose / Cholesterol Monitoring System require only a small amount of blood, the result is precisely calculated and displayed.Please read the entire "User's Manual" carefully before using the TRULY Blood Glucose / Cholesterol Monitoring System. The "User's Manual" guides you through the operation procedures with illustrated directions and help you find the required informations.Please contact the local distributor for questions and enquiries. WARNING:Please carefully read the instructions in this "User's Manual" and become familiar with the test procedures before using the system. User should do the quality check and consult with a healthcare professional before doing any important medical decision.1.2 PrecautionsPlease observe the following limitations for obtaining accurate results of TRULY Blood Glucose / Cholesterol Monitoring System.- The measurements are only for reference purposes. For an analysis     of the measurements, consult specialists.- For In-vitro diagnostic use only (External use only).- Single use only, please do not reuse the test strip.- Do not use this product for those who cannot express their opinions     properly including infants or babies.
-3-- Handle the strip with clean and dry hands; lancing site for the test       should be cleaned and dried thoroughly.   - Do not perform the test while a cell phone is used beside; prevent     the electronic signal interference.- Oxygen concentration in air is low may cause the test inaccurate;    there is no effect from altitude up to 10,000 feet above sea level.- Medical therapy for intravenous injection, such as immunoglobulin     preparations and peritoneal dialysis solutions,containing sugar     compounds may cause higher test results.- Hematocrit (Hct) range from 30-55% is acceptable.Hct less than   30% may cause elevated result.Hct above 55% may cause reduced   glucose results versus a whole blood reference. - Do not drop the product.- Keep the lancets and lancing device out of the reach of the children. o- Keep the meter and test strips at room temperature (10-30 C).1.3 Intended UseTruly TBG-460GC  Blood Glucose and Cholesterol Monitoring System is intended to be used for quantitative measurement of blood glucose level and cholesterol level in fresh capillary whole blood drawn from thefingertips. It is intended for use outside of body (In Vitro diagnosticuse). It should be use only as an aid to monitor blood glucose and cholesterol level by individuals at home and by healthcare professionalsat clinical settings.The devices features include Bluetooth function, the measurement result could be send out via Bluetooth protocol.
-4-1.4 Labelling and InformationsDo not re-useKeep dryOperation temperaturelimitationIn-vitro diagnosticLot numberCE certificationRead instructionsCaution, consultaccompanying documentsStorage temperaturelimitationUse byEU RepresentativeCatalogue number(Product number)Manufactured by01231.5 System Components 设定Test Strip Port:Insert test strip here.The meter will turn on automatically.LCD Screen :For display of operating information and test results.Memory Button :Enter memory mode operation.Down ButtonUp Button Meter (Front Side)
Cholesterol Test Strip(Yellow Color)Electronic Contact Bar :Insert Electronic Contact Bar of thetest strip into the Test Port of themeter to activate it.   Sample Inlet :The inlet where the whole blood or control solution is drawn in. Touching a bloodsample or control solution will result in acapillary action automatically. At least 10uL of blood sample is essential.Reaction Zone :Reaction zone ofTotal Cholesterol CHOLGlucose Test Strip (Blue Color)Sample Inlet : The inlet where the whole blood or control solution is drawn in. Touching a blood sample or control solution will result in a capillary action automatically  Electronic Contact Bar :Insert Electronic Contact Bar of thetest strip into the Test Port of themeter to activate it.  GLUCTag :Display information about the meter.Meter (Back Side)Battery Cover :For holding a 3V lithium battery-5-
Code Strip ( Front View ) -6-Chapter 2 Setup the meter2.1 Set the Year/ Month/ Day/ Hour/ Minute/ Measurement unitPress and hold the “memory” button for over three seconds, the meter will sound one beep and enter the setting mode. Please follow the directions below to set the meter.Under the setup mode, the LCD screen of the meter shows all setting items start from a blinking year “20XX” and the rest setting items are listed as follows:   Set the Year1. When the meter enter setting mode, the first     item on the screen such  as “2012” will blink on     screen, which means Year.2. Press the up/ down button to .3. After you set the desired year number, Press the    “memory” button again then the month segment     starts blinking.until the correct     year appears Set the Month1. Press the up/ down button until the correct     month appears.2. Press the “memory” button, and then the day     segment starts blinking.Contact Bar :Insert this end intothe Test Port of themeter.  Code NumberTest Mode :GLUC - GlucoseCHOL - Total Cholesterol  CHOL  A412GLUC  0709
  Set the Day1. Press up/dowm button until the correct day     appears.2. Press the “memory” button, and the hour segment     will start blinking.Set the Hour1. Press up/ down button until the correct hour     appears.2. Press the “memory” button, and the minute     segment will start blinking. Set the Minute1. Press up/ down button until the correct minute     appears.2. Press the “memory” button then measurement unit    segment starts blinking.  1. Press the up/dowm button until     the correct measurement unit     appears.2. Press the “memory" button    will show "OFF" on the screen     and power off.Chapter 3 How to perform Test3.1 Before Testing-7-Coding the MeterTRULY TRULY  TBG-460GC meter should be coded before testing with   test Strips for the first time, or every time a new box of test strips is used. Each vial of test strips has an assigned code number. The code strip is packaged with the test strip in order to calibrate/code the meter.Every time a test strip is inserted, please be sure to check the code number appearing on the screen is the same as the code number marked on the test strip vial.
设定-8-Directions : 1. Insert code strip into the test port of     the meter. Wait for the code number     and test mode to appear on the     screen. 2. Make sure the code number on the     screen is the same as the code     number on the code strip, and the     test mode is also the same as     marked on the test vial. Remove the     code strip. “    ” should appear on the     screen, indicating that the meter has     been successfully coded.     Cautions :If “E-E” appears on the screen during coding procedure, there may be a problem with the code strip or a wrong category code strip may be misused. Perform the coding procedures again, and if “E-E” continues to appear, please contact the authorized distributor for service.    Use of Lancing Device     1. 2. 3.
-9-3.2  Connect iOS device via Bluetooth Prior to first use, Download and install "TRULY" application to your iOS device. (Bluetooth 4.0 capabilities, e.g. iPhone 4S, iPhone 5, NewiPad and iPadMini)1. Installation1. Remove the end cap by gently pushing it down and away from the     lancing device.2. Insert the lancet firmly into the lancet holder until it comes to a stop     and remove the protective lancet cover.3. Replace the end cap by snapping it back onto the lancing device     body.4. Select the skin penetration depth by turning the depth selector until     the desired number can be seen in the selection window. 5 levels of     skin penetration: The higher the number, the deeper it pricks.5. Pull the part located at the back of the device until it comes to a stop    and then release.6. Place the lancing device firmly against the finger and press the     release button to obtain a blood sample. After finishing your blood    test, dispose the used lancet safely in an appropriate container.84. 5. 6.2.     on Measure screen.Open "TRULY" application, and then press the Blood Pressure icon
3.3  Perform the Blood Glucose TestGLUC1. Insert a Blood Glucose test strip (blue     color) into the test strip port of the     meter and immediately cover the vial     to keep the strips in dry condition.     The meter will turn on automatically.Directions :2. Be careful to make sure themarked re-code test     mode and code number on the     screen are the same as the one      on the test strip vial. Icon    “GLUC” and “CHOL” indicate     measuring mode of glucose and     total cholesterol respectively. If not,     remove the test strip and   the     meter.设定3. Press the “Glucose & Cholesterol”     button.3. Obtain a drop of blood with a       device.    Do not squeeze your fingertip hard     to obtain a drop of blood. It may     influence the result.lancing-10-
1. Insert a Total Cholesterol test strip (yellow color) into the test strip     port of the meter and follow the procedure as performing the blood     glucose test.2. Make sure to get enough blood sample volume.3. The screen will display a count down from 26 seconds for cholesterol     testing.3.4 Perform the Total Cholesterol Test-11-4. Touch the sample Inlet of the strip to     the drop of blood sample until the     meter sounds a “beep”, which     means the sample has drawn into     strip successfully.  5. The screen will display a count down    from 10 seconds for glucose testing.6. When the measurement is complete,    the glucose value will display on     Blood Glucose monitor & save on    iOS application screen.
-12-Note :The TBG-460GC Monitoring System includes glucose and     total cholesterol test that are all plasma calibrated. The user does     not have to convert it into plasma result.2. Use the strips before the   date, the strips can only be used     within 3 months remaining shelf life after the first opening of the    vial.3. Apply the blood sample onto the correct position of the strip     precisely, follow the directions to 4. Prohibited applying the blood sample to the test strip twice, it will     cause inaccurate test results.        1. expiryavoid incorrect operations.3.6 Care after the TestDirections:1. Improper usage of the meter, such as dropping, bumping or other     violent impact may cause the meter damaged.2. Please keep meter away by interferences of magnetic,     electromagnetic, and radioactive.  Warning :Do not disassemble the meter for any reason.2. Please follow local regulations to discard used test strips     and lancets.3. Used test strips, lancets and any other materials that have    come in contact with blood should be treated as potential     biohazards. If a user has an infectious disease, used test     strips and other materials could be a sources of infection.  1. 3.5 Special messagesMode                                          Lo                                             HiGlucose                 < 20 mg/dL  (1.11 mmol/L)     > 600 mg/dL (33.3 mmol/L)Total Cholesterol   < 100 mg/dL  (2.59 mmol/L)   > 400 mg/dL  (10.35 mmol/L)
Intended to use:The control solution is used to check the performance of the TRULY system, including Blood Glucose and Cholesterol Monitor System, Glucose and Cholesterol strip, user's test skill. The system is performing adequately if the control solution test result falls within the indicated control range listed on the test strips vial label.A control test should be proceeded:1. When you suspect that the meter or test strips are not working     properly.2. When your blood glucose or cholesterol test results are not     consistent with how you feel, or when you think your results are not     accurate.3. When you drop or bump your meter.4. Upon the advice of the healthcare professional.How to perform a control test.1. Insert a test strip into the test strip port of the meter and immediately     cover the vial to keep the strips in dry condition.2. Be careful to make sure the test mode and code number on the     screen are the same as the one marked on the test strip vial. If not,     remove the test strip and re-code the meter to correct test mode and     code number.3. Shake the control Solution vial vigorously then open the cap to     discard the first three drops of control solution, after wiping the     dispense tip and cleaning with tissue then drop some control solution     onto a clean object such as transparency. If it is a cholesterol test,     Please drop out  enough control solution (at least 10uL).4. Touch the Sample Inlet of the Glucose or Cholesterol Test Strip to the     drop of Control Solution until the Meter sounds a “beep”,     The screen will display a count down from 10 seconds for Glucose     Test Strip or 26 seconds for Cholesterol Test Strip.    5. After counting down, the Screen will display the control Test results,     Compare the results to the control range listed on the Test Strip Vial.Warning:-13-Chapter 4 Quality Control
-14-Truly TBG-460GC meter automatically stores up 360 blood glucose and 50 cholesterol test results. It also provides you with 7-, 14-, 21- and 28-day averages of the blood glucose test results. Please follow these directions to recall stored test results.Directions :1. Make sure that no test strip is inserted in the meter.2. Press the “memory” button once to turn the meter on. After a short     “beep” sound a full display will appear on screen then the meter     enters standby mode.3. Press the “memory” button once again and the meter stays at     memory mode. The 7-day average (7 DAY.A) will display on the     screen.4. Followed by 14, 21 and 28-day average when press the “memory”    button repeatedly.Note :The day average function can be applied to glucose test mode only, total cholesterol test mode do not have the day average function. Chapter 5 Meter Memory Function. Please read this user's manual for control solution test carefully     before use.. Do not reuse the Test Strips and close the Test Strip Vial lid tightly     after use.. The first three drops of Control Solution must be discarded and the     vial should be shaked well before use when the vial is newly opened.. Newly opened vial of Control Solution and Test Strips must be     marked with the opening date.. Glucose Control Solution is used for glucose test mode And     Cholesterol Control Solution is used for cholesterol test mode. . If the test results fall outside the control range, please reconfirm and     repeat the test, if it is still fall outside the range, please contact     authorized distributor for help and service.. Control solutions are not included in the system, please contact your     distributor if you want to do a quality control test.
-15-Chapter 6 Installing/Replacing the BatteriesBatteriesA 3V lithium battery (CR2032) is included in the meter package. Use only 3V lithium batteries. Do not use or replace the battery with different types of batteries. Doing so will damage the electronics of the meter. Instructions :1. Hold the meter with one hand, slightly lift the back cover lid to open     the back cover. 2. Insert a new battery in the meter with the "+" side facing up, you     will hear a "beep" sound which indicates that meter is  .3. Place back the cover onto the meter.  ready5. After the 28day-average, the latest stored result M01 appears on the     screen. Then you can read the stored result from M01 to     M360(Glucose mode) or from M01 to M50(Cholesterol mode) by     pressing Down or UP Button.6. Press the “memory” button for over 3 seconds and the meter will turn     off after a long “beep” sound. Meter will also automatically shut down     after 3 minutes without any operation.       Note :If 360 test results are already stored and a new one is added, theoldest test result is automatically overwrited. The newest result will be shown in M01.
-16-Chapter 8 Error Message and Trouble Shooting Message   Cause                        SolutionThere will be a problem with the test strip. Repeat the test with a new teststrip. If the problem persists, please contact the dealer for service.  E-0E-b The power of the battery is too low to run a test. Replace the battery at once.E-A The reference voltage orother circuit is error. Call local authorized distributorfor help and service. The test strip may be removed after applying blood the absorbent hole.  Revlew the instructiona and tryagain with a new test strip.If the problem persists, pleasecontact the dealer for service.  E-9Chapter 7 Storage and Handling of Meter and StripsDue to sensitive to heat and humidity, the performance of test strips may be affected by improper storage and handling. Please store thetest strips and meter carefully according to the specifications listed inchapter 8.Note:● Store the meter, code strip, test strips at room temperature. Do      not expose them beyond the range described in Chapter 8.● Do not freeze or refrigerate the meter and strips.● Keep the meter clean by wiping the exterior of the meter with      a tissue or lint-free cloth.
 E-E 1. There may be a problem    with self-check of the     meter.2. Use of the wrong or     damaged code strip or     the code strip was     inserted improperly.      Power off then retry, if the errorappears again, contact the authorized distributor for service.Check that the code number on the meter display matches the code number on the test strip vial. Code the meter or retest as necessary. If the error message appears again, please contact the dealer for service.   E-t &TemperatureIcon Sign The temperature is belowor above the systemoperating range.Repeat the test after the meterand the test strip have reached at temperature within the operating range.   E-U It could be caused by  a used test strip or atemporary or permanentelectronic problem.  Repeat the test with a new test strip. If the error message appears again, please contact the dealer for service. Message   Cause                        SolutionHI 1. Blood glucose level is    higher than 600 mg/dL.2. Test strip has problem. 1. Review the instructions and     try again with a new test     strip.2. Check the test strip vial to     make sure that it is not     marked beyond the expiration     date. Lo 1. Blood glucose level is   lower than 20 mg/dL.2. Test strip has problem.1. Review the instructions and     try again with a new test     strip.2. Check the test strip vial to   make sure that it is not     marked beyond the expiration    date.-17-
-18-WeightDimensions                   Battery Life                    Battery Type                 Memory                      Relative Humidity          Operation Temperature Strip Storage ConditionHematocrit Range        Sample Volume              Measuring Range         Measuring Time              About 10 seconds (Glucose),26 seconds (Total Cholesterol)Test Sample                      Fresh Capillary whole bloodCalibration                    Principle                           System Accuracy           ± 20% at Glucose level ≥75 mg/dL (4.17mmol/L) ± 20% at Total Cholesterol ≥150mg/dL(3.88mmol/L)Electrochemical Biosensor Plasma-equivalentGlucose 20 - 600 mg/dL (1.1 - 33.3 mmol/L)Total Cholesterol 100 - 400 mg/dL (2.59 - 10.35 mmol/L)About 1 - 1.5 µL (Glucose)About 10 µL (Total Cholesterol)30 - 55% (for Glucose)35 - 50% (for Total Cholesterol) 10 - 30ºC (50 - 86ºF)10 - 40ºC  (50 - 104ºF)Less than 95%360 for Glucose, 50 for Total CholesterolOne 3V (CR2032) lithium batteryApproximately 1,000 tests82*54*21.5mm (Length, Width, Height)About 45 g (with battery)Chapter 9 Specification
FCC Statement -19-This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:1) This device may not cause harmful interference, and2) This device must accept any interference received, including     interference that my cause undesired operation.This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. Theselimits are designed to provide reasonable protection against harmfulinterference in a residential installation. This equipment generates, uses and can radiate radio frequency energy.However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:• Reorient or relocate the receiving antenna.• Increase the separation between the equipment and receiver.• Connect the equipment into an outlet on a circuit different from that towhich the receiver is connected.• Consult the dealer or an experienced technician for help.MODIFICATION:Any changes or modifications not expressly approved by the grantee of this device could void the user’s authority to operate the device.FCC ID: 2ABG7-TBG-460
Truly Instrument Ltd2/F., Chung Shun Knitting Centre,1-3 Wing Yip St, Kwai Chung, N.T., HONG KONGTel: (852) 2487 9803Fax: (852) 2146 5756Http: www.trulyinstrument.com: EC - Representative Wellkang Ltd                                   t/a Wellkang Tech Consulting                                   Suite B, 29 Harley Street,                                    LONDON W1G 9QR,                                    United Kingdom