Unfors RaySafe 8603021 Dose Aware Base station User Manual Dose aware

Unfors RaySafe AB Dose Aware Base station Dose aware

User manual Dose aware

Download: Unfors RaySafe 8603021 Dose Aware Base station User Manual Dose aware
Mirror Download [FCC.gov]Unfors RaySafe 8603021 Dose Aware Base station User Manual Dose aware
Document ID1228983
Application IDSzAObC1+teo8+4+Yepvqqw==
Document DescriptionUser manual Dose aware
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize300.4kB (3754973 bits)
Date Submitted2010-01-18 00:00:00
Date Available2010-01-18 00:00:00
Creation Date2010-01-12 11:01:20
Producing SoftwareMac OS X 10.5.6 Quartz PDFContext
Document Lastmod2010-01-12 11:01:20
Document TitleMicrosoft Word - DoseAware_Base_Station_Package_User_Manual-100111.docx
Document CreatorMicrosoft Word
Document Author: Hilda Harrysson

DoseAware
Base Station Package
Instructions for Use
Table of Contents
1 Introduction ....................................................................................4
1.1 Intended Use of the DoseAware System ................................................... 4
1.2 About the DoseAware System ................................................................... 5
1.3 About this User Manual............................................................................. 6
1.4 Other User Manuals................................................................................... 6
1.5 System Requirements................................................................................ 6
2 Base Station Package Installation ...................................................7
3 Use the Base Station .......................................................................8
3.1 Introduction............................................................................................... 8
3.2 Getting Started .......................................................................................... 8
3.2.1 Wall Mounted Base Station .................................................................................... 8
3.2.2 MCS Mounted Base Station .................................................................................... 8
3.2.3 Troubleshooting...................................................................................................... 8
3.3 Base Station Features ................................................................................ 9
3.3.1 Online View – View Current Dose Rates ............................................................... 10
3.3.2 Walk‐Up View – View Detailed Dose Data ............................................................ 12
3.3.3 Base Station Settings View.................................................................................... 16
3.4 Shut Down the Base Station .................................................................... 20
4 Use DoseView ...............................................................................21
4.1 Introduction to DoseView........................................................................ 21
4.2 Getting Started with DoseView ............................................................... 22
4.2.1 Troubleshooting.................................................................................................... 22
4.3 Home Toolbar Overview.......................................................................... 23
4.4 PDM Panel ............................................................................................... 24
4.4.1 PDM Information .................................................................................................. 24
4.4.2 Dose Information .................................................................................................. 25
4.4.3 Selection Summary ............................................................................................... 25
4.4.4 Change PDM Options ............................................................................................ 25
5 PDM ..............................................................................................30
5.1 Introduction............................................................................................. 30
5.2 Getting Started ........................................................................................ 30
6 Cradle ............................................................................................31
6.1 Introduction............................................................................................. 31
6.2 Getting Started ........................................................................................ 31
6.3 Troubleshooting....................................................................................... 31
9896 002 16621
4.5 View Dose Graph ..................................................................................... 27
4.6 View Dose Table ...................................................................................... 28
4.7 Specify Password and Language .............................................................. 29
7.1 Calibrating the Base Station Screen......................................................... 32
7.2 Cleaning the Base Station Package Components..................................... 32
7.3 Disinfection.............................................................................................. 32
8 Product Disposal ...........................................................................33
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
7 Maintenance .................................................................................32
9 Technical Data...............................................................................34
9.1 PDM Radiology ........................................................................................ 34
9.2 PDM Memory .......................................................................................... 34
9.3 Base Station Memory .............................................................................. 34
9.4 Dose Data Transfer from PDM to Base Station........................................ 35
9.5 PDM and Base Station/DoseView Memories .......................................... 35
9.6 Time Management .................................................................................. 36
9.7 Technical Specifications........................................................................... 36
9.7.1 Radio Communication........................................................................................... 36
9.7.2 Base Station .......................................................................................................... 36
9.7.3 PDM ...................................................................................................................... 36
9.7.4 Cradle .................................................................................................................... 37
10 Safety ..........................................................................................38
10.1 Warnings................................................................................................ 38
10.1.1 Maintenance and Faults...................................................................................... 38
10.1.2 Safety Awareness................................................................................................ 38
10.1.3 Adequate Training............................................................................................... 38
10.1.4 Safety Devices ..................................................................................................... 38
10.1.5 Intended Use and Compatibility ......................................................................... 39
10.2 Electrical Safety ..................................................................................... 39
10.3 Mechanical Safety.................................................................................. 39
10.4 Fire Safety .............................................................................................. 39
10.5 Electrostatic Discharge (ESD)................................................................. 39
10.6 Electromagnetic Compatibility (EMC).................................................... 40
10.6.1 Portable and Mobile Phones............................................................................... 40
Philips Healthcare
9896 002 16621
10.7 Modality Specific Safety......................................................................... 41
DoseAware Base Station Package Document version 1.0
1
Introduction
1 Introduction
1.1 Intended Use of the DoseAware System
The DoseAware System is an electronic X‐ray dose monitoring system. The intended
use is to improve the awareness of people, who work with or are in the presence of
X‐Ray imaging equipment, about their occupational dose (also known as staff dose).
The awareness focuses on:
• a graphical visualization of the real‐time staff dose rate while working with
X‐Ray equipment in examination rooms during medical procedures;
• instant access to historical staff dose for reporting and analysis purposes.
The benefits of the DoseAware System are to:
• make people aware of their received staff dose during clinical work with X‐ray
imaging equipment;
• instantly visualize the result of reducing measures of occupational dose by, for
example, changing a person’s position in the examination room.
The DoseAware System may not be used as a legal staff dose recording solution. The
DoseAware System is not intended for patient use.
NOTE: Do not expose the PDMs in direct X‐ray beam. They are designed to be exposed
to scattered radiation only.
NOTE: The Dose Manager, DoseView, Cradle and PDM Rack are not intended to be
used inside examination rooms.
NOTE: The DoseAware product is not a replacement for a TLD (ThermoLuminescent
Dosimeter) or similar product.
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
9896 002 16621
The components of the Base Station Package are intended to be used in combination
with a number of PDMs. In this configuration, the Base Station Package is designed as
a monitoring aid for hospital staff.
Introduction
1.2 About the DoseAware System
Philips Healthcare
9896 002 16621
Figure 1 DoseAware System overview with Base Station Package components highlighted
The DoseAware System consists of:
• Personal Dose Meter (PDM)
• Base Stations (display unit)
• DoseView (computer software)
• Dose Manager (computer software)
• Cradle (dock station used to connect PDMs and computer)
• PDM Rack (PDM storage)
The components of the system are delivered in two main configurations and as
separate PDMs:
• Base Station Package, which consists of:
o Base Station, power adaptor, MCS bracket and wall mount kit
o DoseView (CD)
o User manual (CD)
o Cradle, USB cable
DoseAware Base Station Package Document version 1.0
1
Introduction
•
•
•
Dose Manager Package (optional), which consists of:
o Dose Manager licensed software (CD)
o User Manual (CD)
o Cradle, USB cable
PDMs in packages of one, three or ten units
PDM Rack, optional
1.3 About this User Manual
This user manual describes the Base Station Package. In order to get an immediate
hands‐on experience of the Base Station Package, we recommended that you interact
with the Base Station, DoseView and Cradle in parallel to reading this user manual.
1.4 Other User Manuals
•
•
Dose Manager is described in a separate user manual, which can be found on
the CD delivered in the box together with the Dose Manager package.
The PDMs are described in the PDM Quick Guide, a leaflet that is delivered in
the box together with the PDM.
1.5 System Requirements
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
9896 002 16621
DoseView has the following system requirements:
• Operating systems: Windows Vista or Windows XP
• .NET 3.0
• At least one USB port available
• At least 1 GB of system memory available
• At least 40 GB hard drive with at least 15 GB of memory available
• Recommended screen resolution at least 1024 x 768
Base Station Package Installation
2 Base Station Package Installation
Philips Healthcare
9896 002 16621
The Base Station Package is installed and configured by a Philips service engineer and
a local application specialist as part of delivery and hand‐over. During its use or due to
changes in the place of installation, modification to the Base Station Package or its
configuration may be necessary. A trained Philips service engineer will carry this out.
NOTE: Installation of third‐party software in the Base Station is prohibited. Installation
of such software may cause the Base Station to malfunction.
DoseAware Base Station Package Document version 1.0
3
Use the Base Station
3 Use the Base Station
3.1 Introduction
The Base Station is the display unit for the
DoseAware System. It communicates wireless with
PDMs within radio range in order to collect,
present and store PDM dose data.
3.2 Getting Started
3.2.1 Wall Mounted Base Station
1. Connect the output connector of the AC‐
DC adaptor to the back of the Base Station, if needed.
2. Connect the input connector to an AC electrical outlet socket. This will switch
on the Base Station. A Philips start up screen will be displayed for a few
seconds. The Online View will be displayed and within 30 seconds a maximum
of 8 PDMs (at a time) within range of the Base Station will appear.
NOTE: The Base Station screen is a touch screen. Operate the Base Station by
tapping buttons and graphs on the screen.
WARNING: Do not cover the Base Station to prevent excessive temperatures.
WARNING: Be careful when moving other equipment in the vicinity of the Base
Station.
1. Switch on the X‐ray system and the Base Station will automatically switch on.
A Philips start up screen will be displayed for a few seconds. The Online View
will be displayed and within 30 seconds a maximum of 8 PDMs (at a time)
within range of the Base Station will appear.
NOTE: The Base Station screen is a touch screen. Operate the Base Station by
tapping buttons and graphs on the screen.
2. Position the Base Station similar to the other monitors in the MCS as much as
possible.
WARNING: Do not cover the Base Station to prevent excessive temperatures.
WARNING: Be careful when moving the MCS or other equipment because the Base
Station may stick out and can be hit.
9896 002 16621
3.2.2 MCS Mounted Base Station
If a PDM does not appear in the Online View, use DoseView or Dose Manager to check
that the PDM’s:
• battery Status is OK (see chapter 4.4.4).
• power Mode is set to ‘On’ (see chapter 4.4.4).
• Base Station Mode is set to ‘Show’ (see chapter 4.4.4).
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
3.2.3 Troubleshooting
Use the Base Station
3.3 Base Station Features
The Base Station offers:
• An Online View (see chapter 3.3.1) where you can:
o view current Personal Dose Rate information for several PDMs at a
time.
o access the Walk‐Up View by tapping the Displayed Name tag button of
that PDM in the Online View.
• A Walk‐Up View (see chapter 3.3.2) where you can view historical dose data
for a PDM chosen from the OnLine View.
• A Base Station Settings View (see chapter 3.3.3) for administrator settings.
Philips Healthcare
9896 002 16621
See chapter 9 for technical details about how the Base Station displays dose data. In
addition, the Base Station:
• Serves as an interim storage for the PDMs’ dose data to be further analyzed in
the Dose Manager application.
• Is provided with a USB port for dose data download to USB memory (see the
Dose Manager Package user manual).
NOTE: When the Base Station memory is full, the oldest data will be overwritten with
newer data. For storage capacity examples, see chapter 9.3.
DoseAware Base Station Package Document version 1.0
3
Use the Base Station
3.3.1 Online View – View Current Dose Rates
Figure 2 The Online View
The Displayed Name button identifies the PDMs. By tapping the Displayed Name
button of a PDM, you will enter the Walk‐Up View of that PDM.
The dose rate scale is divided in three zones:
• Green zone indicates good working habits. Proper actions have been taken to
avoid exposure for unnecessary radiation.
• Yellow zone indicates higher doses, which can be acceptable for shorter
periods of time, for example if you need to stand closer to the patient during a
procedure. If you are exposed to radiation in the yellow zone frequently, you
need to take actions to minimize the dose exposure.
10
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
Figure 3 The Online View elements
9896 002 16621
The Online View displays the current personal dose rates and peak dose rate
indicators, for up to 8 PDMs within range of the Base Station.
The icon in the upper right corner gives you access to the Base Station settings view
(see chapter 3.3.3).
Use the Base Station
Red zone dose indications should not occur during normal procedures. If you
are exposed to radiation in the red zone, you need to take actions to minimize
the dose exposure.
When a PDM moves out of range of the Base Station, its button will remain visible in
the Online View for two minutes. The button will then remain grayed out for another
eight minutes before it disappears. This feature allows people to temporarily leave the
room without loosing their position in the list. When a button is grayed out, it is not
possible to access the Walk‐Up View.
If no PDM has been within range of the Base Station for 30 minutes, the Online View
will enter screen saver mode. In screen saver mode the Base Station displays a black
screen. The Base Station screen starts up when a PDM connects to the Base Station or
when the user taps the screen.
Philips Healthcare
9896 002 16621
•
DoseAware Base Station Package Document version 1.0
11
3
Use the Base Station
3.3.2 Walk‐Up View – View Detailed Dose Data
The Walk‐Up View also consists of the Personal Dose Overview (see chapter 3.3.2.1),
which is the default Walk‐Up view. It has two sub views:
o the Annual Personal Dose view (see chapter 3.3.2.2)
o the Personal Dose History view (see chapter 3.3.2.3)
12
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
Enter the Walk‐Up View of a PDM by tapping the Displayed Name button of that PDM
in the Online View. On the left hand side of the Walk‐Up View a Personal Dose Meter
Info panel is displayed (see Figure 4). It displays:
• Full Name
• Displayed Name ‐ the name that is displayed in the Base Station Online View
• Clinical Role ‐ Doctor, Nurse, Technician or Other
• PDM Position ‐ Head, Torso, Hand, Belly, Leg or Other
• PDM ID ‐ a unique ID per PDM
• Battery status
o 2‐4 blocks indicate normal use.
o 1 block indicates that you need to change the battery.
• The PDM’s date and time
9896 002 16621
Figure 4 Personal Dose Meter Information panel
Use the Base Station
3.3.2.1
Personal Dose Overview
Philips Healthcare
9896 002 16621
Figure 5 The Personal Dose Overview view
When you enter the Walk‐Up View you will see the following information (see Figure
5):
• The Accumulated Personal Dose since January 1st of the current year, or since
last reset. This value is also shown as a percentage of the yearly max dose.
• Access to the Annual Personal Dose sub view (1), see chapter 3.3.2.2.
• Dose Session 1 and 2 values. Use the Dose Sessions to measure dose for
specific time spans, for example a specific procedure or a working day.
The value shows the accumulated dose for a session since last Dose Session
reset, measured in Sv. Use the reset buttons (2) to reset a Dose Session to
zero. Information about dose session reset will appear as an event in Dose
Manager.
• Total Personal Dose since the last Dose History reset.
• The current day’s Personal Dose History graph. You can also access the
Personal Dose History sub view (3), see chapter 3.3.2.3.
DoseAware Base Station Package Document version 1.0
13
3
Use the Base Station
3.3.2.2
Annual Personal Dose
Figure 6 The Annual Personal Dose view
14
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
9896 002 16621
In this screen you can find the following information:
• Today’s date (white line).
• Annual dose limit (red line). Set this value in DoseView or Dose Manager.
• The annual dose limit distributed over the full year (dashed line). As long as
the accumulated dose stays below this value, the annual dose limit will not be
exceeded for the full year.
• Accumulated dose in relation to the annual dose limit for the current year
(blue/orange line). When the accumulated dose exceeds the distributed
annual dose limit, the color will change from blue to orange in order to alert
the user to take actions.
Use the Base Station
3.3.2.3
Personal Dose History
Philips Healthcare
9896 002 16621
Figure 7 The Personal Dose History view, with Day selected
In this view, you can view Hour, Day and Week overviews by tapping the respective
buttons on top of the screen.
The view lets you:
• For the Hour time span:
view a Dose Rate graph spanning ½ hours. Each data point in the graph
represents the maximum dose rate during the surrounding 15 seconds.
• For the Day/Week time spans:
view accumulated dose value bars. Each bar represents the accumulated dose
during one/four hours, respectively.
Within the Day and Week time spans, tapping a dose bar in the diagram will
zoom into the larger underlying time scale (that is Week ‐> Day and Day ‐>
Hour).
Within each time span (Hour/Day/Week) you can step forward and backward in time
with the arrow buttons on the bottom of the screen. By pressing and holding the
arrow buttons you will scroll along the time axis within the chosen zoom level.
• in the upper left corner of the chart, view the value of the Accumulated Dose
during the chosen time span
DoseAware Base Station Package Document version 1.0
15
3
Use the Base Station
3.3.3 Base Station Settings View
The Base Station Settings View is where administrators configure the Base Station.
Follow the instructions below to enter the Base Station Settings View:
1. Tap the settings menu symbol in the upper right corner of the Online View.
2. Enter configuration values by tapping the buttons in the respective menus.
In the lower part of Base Station Settings, the Base Station ID/serial number and the
software version is displayed. This information can be used as assistance for support
issues.
3.3.3.1
User Settings Menu
16
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
The User Settings menu consists of one screen.
Adjust the backlight level of the screen by moving the bar horizontally. This change will
take effect immediately.
9896 002 16621
Figure 8 The User Settings menu
Use the Base Station
3.3.3.2
Admin Settings Menu
Figure 9 The Base Station Name and Base Station Location menu.
Philips Healthcare
9896 002 16621
The Admin Settings menu consists of four screens. Access these screens by tapping the
arrow buttons in the lower right corner of each screen.
Base Station Name and Base Station Location Menu
Enter name and location of the Base Station (see Figure 9). This information is used to
identify the Base Station so that it can be detected in Dose Manager.
DoseAware Base Station Package Document version 1.0
17
3
Use the Base Station
Figure 10 The Reserved Slots menu.
18
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
9896 002 16621
Reserved Slots Menu
This is an optional setting, which you can use to reserve a number of slots in the
Online View for the clinical roles Doctor, Nurse and Technician (see Figure 10).
If you, for example, have reserved three slots for Doctors but the Base Station detects
only one, there will be two empty slots before the first Nurse appears.
If you want to reserve slots for Nurses, you also have to reserve slots for Doctors. If
you want to reserve slots for Technicians, you also have to reserve slots for Doctors
and Nurses.
The default behavior “Auto” is first‐come, first‐served, which means that PDMs will
appear on the screen in the order the Base Station detects them, sorted after their
clinical role.
Use the Base Station
Figure 11 The Network setup menu.
Philips Healthcare
9896 002 16621
Network Setup Menu
Configure the network connection between the Base Station and the Dose Manager
(see Figure 11). If you change this information, the Base Station requires a restart.
NOTE: Network setup is only applicable if you are using Dose Manager.
DoseAware Base Station Package Document version 1.0
19
3
Use the Base Station
Figure 12 The Time and Date menu.
Time and Date Menu
Set time, date, time zone and manage daylight saving time for the Base Station (see
Figure 12).
For daylight saving time, you can select to manage it manually or by selecting a
daylight settings region for your time zone. The possible regions available for daylight
saving time are different depending on your current time zone.
If you choose to manage daylight saving time manually, you can choose to set it to
wintertime (+0 hours) or summer time (+1 hours). If a region is selected, the change
between winter/summer time will be done automatically.
For an MCS mounted Base Station:
2. Switch off the X‐ray system and the Base Station will be switched off.
20
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
For a wall mounted Base Station:
1. Unplug the AC‐DC adaptor.
9896 002 16621
3.4 Shut Down the Base Station
Use DoseView 4
4 Use DoseView
4.1 Introduction to DoseView
The DoseView application lets you, for one PDM at a time:
• view the PDM’s dose history
• change the PDM options, for example full name and displayed name
Figure 13 DoseView overview
Philips Healthcare
9896 002 16621
The following items are available when you start DoseView:
Item
Description
Application menu
Access the DoseView options dialog, see chapter 4.7.
Home toolbar
Access tools to navigate in dose graphs and dose tables, see chapter 4.3.
PDM panel
Show and manage options for a PDM that is placed in a Cradle. You can only
access and change PDM options when the PDM is in a Cradle, see chapter
4.4.
Dose graph
View the dose history as a graph, see chapter 4.5.
Dose table
View the dose history as a table, see chapter 4.6.
DoseAware Base Station Package Document version 1.0
21
4
Use DoseView
4.2 Getting Started with DoseView
NOTE: Do not connect the Cradle to the computer unless DoseView and the Cradle
driver are installed on the computer.
Follow the steps below to get started with DoseView:
1. Start the DoseView application.
2. Connect a Cradle to your computer’s USB port.
3. Put a PDM in the Cradle. Within a few seconds, DoseView will detect the PDM.
This is indicated at the top of the DoseView window by a note “Connected to
PDM 100001158”, where PDM 100001158 is an example of a PDM ID.
DoseView will start loading the PDM dose history. This may take up to a few
minutes. You can follow the progress on the progress bar at the bottom of the
DoseView window.
4.2.1 Troubleshooting
22
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
9896 002 16621
If the PDM does not appear:
• Check that the PDM is correctly fitted into the Cradle.
If the Cradle is not detected:
• Check the USB connection between the Cradle and the computer.
• Install the Cradle driver manually. All the driver files are located on the
installation CD in the folder “CradleDriver”. These files are also copied to the
application installation folder when DoseView is installed. When Windows
detect a connected Cradle and the dialog about driver installation is displayed,
select to use the driver files located on the CD or in the application installation
folder.
If you have forgotten your password:
• Contact your local administrator, who will have to re‐install the software.
Use DoseView 4
4.3 Home Toolbar Overview
Figure 14 Home toolbar
The home toolbar is where you find tools to navigate in dose graphs and dose tables.
The following items are available in the home toolbar:
Function
Description
Navigation group
Access tools for dose history navigation.
Undo zoom button
Move one step back in a sequence of zoom actions, showing the last
selected time range.
Redo zoom button
Move one step forward in a sequence of zoom actions, showing the time
range that was selected before the last undo zoom action.
Show all button
Show all available data for the selected PDM. The time range will start the
first date any PDM begun to measure dose and stop the last date any
PDM was synchronized.
Left button
Shift the time range one step backward. If you have selected
year/month/week/day, the time range will move one
year/month/week/day backward. If you have selected another time
range, the time range will move approximately 10% backward.
Right button
Shift the time range one step forward. If you have selected
year/month/week/day, the time range will move one
year/month/week/day forward. If you have selected another time range,
the time range will move approximately 10% forward.
Day button
View dose history for the current day.
Week button
View dose history for the current week.
Month button
View dose history for the current month.
Year button
View dose history for the current year.
Philips Healthcare
9896 002 16621
Advanced navigation
group
Start time button
The viewed time range start day.
End time button
The viewed time range end day.
Data filter group
Select which information to be visible in the graph.
Accumulated dose
check box
Show/hide the accumulated dose in the graph.
Dose rate check
box
Show/hide the dose rate in the dose graph.
Show details
check box
Checked: The graphs will display dose rate samples for every second,
where such data is available.
Unchecked: the graphs will display mean dose rate values per hour.
DoseAware Base Station Package Document version 1.0
23
4
Use DoseView
4.4 PDM Panel
Figure 15 PDM Panel
4.4.1 PDM Information
Description
Full name
Full name of the person using the PDM.
Displayed name
The name that is displayed in the Base Station Online View.
Clinical role
One of Doctor, Nurse, Technician or Other.
PDM ID
A unique PDM serial number.
PDM position
One of Head, Torso, Hand, Belly, Leg or Other.
Battery status
The PDM's battery status:
•
•
•
•
24
Document version 1.0
Green: normal use.
Yellow: normal use.
Red: the PDM need to be replaced in 4‐6 months.
Crossed battery: there is no battery left. The PDM does not
measure radiation and will not communicate with Base
Stations.
DoseAware Base Station Package
Philips Healthcare
Item
9896 002 16621
The following information is visible when a PDM is placed in the Cradle:
Use DoseView 4
4.4.2 Dose Information
Accumulated personal
dose
The PDM’s total dose measured this year or since last reset, measured in
Sv.
Percentage of annual
dose
The PDM's accumulated annual dose measured this year or since last
manual reset, measured in Sv.
Dose Session 1 and 2
Trip meter for dose values. The accumulated dose for a session since last
Dose Session reset, measured in Sv. You can also reset these values.
NOTE: Dose session reset done in DoseView will not appear as an event in
Dose Manager.
Total personal dose
The total dose exposure for a PDM since last dose history reset.
4.4.3 Selection Summary
View a summary of the dose data selection you might have done either in the dose
table or in the dose graph (see chapters 4.5 and 4.6). You can also clear the contents
of the selection summary field by tapping clear selection.
4.4.4 Change PDM Options
Philips Healthcare
9896 002 16621
Figure 16 PDM Options dialog box
Follow the instructions below to change PDM options:
1. Make sure that a Cradle is connected to your computer’s USB slot.
2. Insert a PDM in the Cradle. The computer will detect the PDM automatically
and the PDM information will appear in DoseView.
3. Access the PDM options dialog by clicking the PDM options (see Figure 16).
The PDM options dialog consists of two tabs, see below.
NOTE: If a password has been set (see chapter 4.7), access to the PDM options
dialog box will be password protected.
In the Info tab you can:
• Edit PDM information (full name, displayed name, PDM position, clinical role,
and annual dose limit value, see chapter 3.3.2) and select a displayed symbol,
DoseAware Base Station Package Document version 1.0
25
4
Use DoseView
which is displayed in the Base Station interface.
The displayed name is used to identify a PDM in the Base Station. The name is
limited to 16 characters. However, in the Base Station’s online view, the
displayed name may be truncated.
NOTE: The dose measurement for a PDM depends on factors such as where
on the body the PDM is positioned and if x‐ray protection devices (for example
a lead apron) that shield the PDM measurements are used. These factors need
to be considered for the annual dose limit.
In the Settings tab you can:
• View and synchronize the PDM’s clock with the computer’s clock.
WARNING: Make sure that the computer’s clock is correct, otherwise the dose
history data will be shifted in time and therefore not accurate anymore.
NOTE: If you need to change hour‐portion of time, the dose history will have
to be reset. This is done automatically; you just need to confirm the action.
When you synchronize PDM’s time the internal clock will be synchronized to
the same time as the host PC, including the Windows time zone settings. If you
are using multiple PDMs, it is important to synchronize time for the PDMs with
the same PC, because they will have the same time reference.
•
•
•
26
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
9896 002 16621
•
NOTE: If you need to change the time backwards to where there is dose
history stored, the dose history will have to be reset. This is done
automatically; you just need to confirm the action.
NOTE: The synchronization takes immediate effect and you do not need to
press the Save button.
Set Base Station Mode to Show/Hide, which makes the PDM to be shown/not
shown on a Base Station.
Turn Power Mode On/Off, which puts the PDM in operating or power saving
mode. In operating mode, communication with Base Stations will take place
and registration of dose data will occur. In power saving mode no
communication with Base Stations will take place and no registration of dose
data will occur.
Reset the PDM’s Dose History.
NOTE: This will permanently delete the PDM’s entire dose data
View Battery Status and Firmware Version.
Use DoseView 4
4.5 View Dose Graph
Figure 17 Dose Graph with selection (vertical red dotted lines)
Philips Healthcare
9896 002 16621
Use the data filter panel in the home menu toolbar to select which information to be
visible in the graph:
• Accumulated Dose graph – blue graph
• Dose Rate graph – orange graph, displays dose rate samples for every second,
where such data is available
• Show details
Checked: The graphs will display dose rate samples for every second, where
such data is available.
Unchecked: the graphs will display mean dose rate values per hour.
The graphs are covering a time span that you choose, either:
• from the Advanced Navigation panel,
• from the Navigation panel,
• or by left‐clicking and dragging in the graph (zooming).
You can also select a time span to be summarized in the Selection Summary field of
the Personal Dose Meter panel. You do the selection by shift‐left‐clicking and dragging
in the graph. Two red, dotted vertical lines in the graph will indicate the selected time
span.
DoseAware Base Station Package Document version 1.0
27
4
Use DoseView
4.6 View Dose Table
Figure 18 Dose Table with selection
In the Dose Table tab (see Figure 18) you can view a table of
• Accumulated Dose values
• Dose Rate values
The tables are covering a time span that you choose either from the:
• Navigation panel,
• or the Advanced Navigation panel.
28
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
9896 002 16621
You can also select one or several rows to be summarized in the Selection Summary
field of the Personal Dose Meter panel.
With the Show details check box checked, the table will list second data, where such
data is available.
Use DoseView 4
4.7 Specify Password and Language
Figure 19 Program Options dialog box
Philips Healthcare
9896 002 16621
In the Application Menu ‐> Program Options dialog box (see Figure 19) you can:
• specify a password to protect the access to the PDM options dialog. The
password is only used when you are making PDM options changes. Others can
still view the PDM data.
NOTE: Contact your local administrator for password guidelines.
NOTE: if you have lost your password, contact your local administrator, who
will have to re‐install the software.
• change application language.
• view the DoseView program version.
DoseAware Base Station Package Document version 1.0
29
5
PDM
5 PDM
5.1 Introduction
The PDM is an active dose meter designed for maintenance‐free
usage throughout its lifetime.
You can personalize the PDM’s appearance by attaching one of the 8
Figure 20 The PDM
inlays of different color, which are delivered together with the PDM
(see the PDM Quick Guide).
The intended PDM position is outside the lead apron. If it is used under the lead apron
the recorded dose will have to be adjusted.
5.2 Getting Started
30
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
9896 002 16621
Follow the instructions below to start using your PDM:
1. Make sure that the PDM’s power mode is set to “on”, see chapter 4.4.4.
2. Attach the PDM on your clothes, using the metallic clip, which is located on
the back of the PDM. The PDM now records dose values and transmits them
to Base Stations within range. You can also read out the recorded dose values
by using DoseView via the Cradle (see chapter 4.2).
See chapter 9 to learn more about PDM memory and data transfer between PDM and
Base Stations.
Cradle 6
6 Cradle
6.1 Introduction
The Cradle is a dock station that lets you connect a PDM to a
computer for data read out as well as PDM Options writing into the
PDM. Use the Cradle in combination with the DoseView and/or Dose
Manager applications.
6.2 Getting Started
Figure 21 The Cradle
NOTE: Do not connect the Cradle to the computer unless DoseView
and the Cradle driver are installed on the computer.
1. Start the DoseView application.
2. Connect a Cradle to your computer’s USB port.
3. Put a PDM in the Cradle. Within a few seconds, DoseView will detect the PDM.
This is indicated at the top of the DoseView window by a note “Connected to
PDM 100001158”, where PDM 100001158 is an example of a PDM ID.
DoseView will start loading the PDM dose history. This may take up to a few
minutes. You can follow the progress on the progress bar at the bottom of the
DoseView window.
6.3 Troubleshooting
Philips Healthcare
9896 002 16621
If the PDM does not appear:
• Check that the PDM is correctly fitted into the Cradle.
If the Cradle is not detected:
• Check the USB connection between the Cradle and the computer.
• Install the Cradle driver manually. All the driver files are located on the
installation CD in the folder “CradleDriver”. These files are also copied to the
application installation folder when DoseView is installed. When Windows
detect a connected Cradle and the dialog about driver installation is displayed,
select to use the driver files located on the CD or in the application installation
folder.
DoseAware Base Station Package Document version 1.0
31
7
Maintenance
7 Maintenance
7.1 Calibrating the Base Station Screen
If the touch screen does not respond correctly to user interaction, a touch screen re‐
calibration may be needed. Follow the instructions below to perform a re‐calibration:
1. Tap and hold on the Philips start up screen that appears during Base Station
startup until the Setup view appears.
2. Follow the instruction on the bottom of the screen and the subsequent
instructions in the next views.
7.2 Cleaning the Base Station Package Components
Clean the components of the Base Station Package when necessary.
WARNING: Always isolate the components from the mains electrical supply before
cleaning to prevent electric shocks.
WARNING: Never allow water or other liquids to enter the components, since these
may cause electrical short‐circuits or metal corrosion.
Clean the Base Station, Cradle, PDMs, PDM Rack and connecting cables in the
following way:
1. Wipe enameled parts and aluminum surfaces clean with a damp cloth and
mild detergent.
2. Rub down with a dry woolen cloth.
Disinfect the components of the Base Station Package when necessary. Disinfect the
Base Stations, Cradle, PDMs, PDM Rack and connection cables in the following way:
1. Switch off the equipment and allow it to cool. This prevents convection
currents drawing disinfectant mist into the equipment (this concerns only the
Base Station).
2. Wipe the component with a cloth dampened with disinfectant.
Philips Healthcare
9896 002 16621
7.3 Disinfection
DoseAware Base Station Package Document version 1.0
32
Product Disposal
8 Product Disposal
Final disposal is when the user disposes of the DoseAware products in such a way that
it can no longer be used for its intended purpose(s).
In the European Union (the WEEE directive), this label
indicates that the product should not be disposed of together with household waste. It
should be disposed of at an appropriate facility to enable recovery and recycling.
WARNING Do not dispose of any parts of this product with industrial or domestic
waste. Incorrect disposal of DoseAware products may lead to serious environmental
pollution.
NOTE: Make sure that the power mode is set to “off” when returning a PDM.
NOTE: PDMs and Base Stations contain personal data. Remove or disable this data for
third parties to read before removal of the part when it is no longer used (end‐of‐life
or broken).
Philips Healthcare
9896 002 16621
Philips supports users in:
• Recovering reusable parts
• Recycling of useful materials by competent disposal companies
• Safe and effective disposal of product
For advice and information, contact your Philips Service Organization.
DoseAware Base Station Package Document version 1.0
33
9
Technical Data
9 Technical Data
9.1 PDM Radiology
Characteristic
Operational Quantity*
Reproducibility
Dose Rate Range
Energy dependence X‐/g‐rays
Angular dependence
Measure
Hp(10)
5% or 1 µSv
40 µSv/h – 500 mSv/h
+/‐ 30% within N40 – N120
+/‐ 5% within +/‐ 5°
+/‐ 30% within +/‐ 50°
+200%/‐100% within +/‐ 180°
+/‐ 5% within 20‐26°C
+/‐ 25% within 15‐35°C
+/‐ 1% from fully charged until low
battery shutdown
Less than 1s above 100µSv/h, less than 4s
otherwise
On torso outside lead apron
Temperature dependence
Battery voltage dependence
Response time
Position on body**
*) Hp(10): Personal dose equivalent at a depth of 10 mm according to ISO 4037.
**) Position on body: The Hp(10) measurement is only valid for a position on the torso outside any lead
apron or other protection. (To estimate effective dose to a user, one must use other means to estimate
things like the environmental radiation situation and the effectiveness of protection.)
9.3 Base Station Memory
The Base Station stores dose information and PDM Info for PDMs that have been
connected to it.
As the Base Station does not have a limitation of 3600 s for the dose rate memory, the
information transmitted from the PDM will be more detailed. When a PDM is
connected to a Base Station, it continuously transmits its measured dose exposures.
NOTE: When the Base Station memory is full, the oldest data will be overwritten.
34
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
The PDM has two dose data memories:
• The accumulated dose memory, where the PDM stores accumulated dose
values every hour for the entire lifetime of the PDM.
• The dose rate memory, where the PDM stores dose rate samples. When the
radiation exceeds 40 µSv/h, the PDM stores one sample per second. This
memory is limited to 3600 second‐samples.
NOTE: When the dose rate memory is full, the oldest data will be overwritten
with newer data.
9896 002 16621
9.2 PDM Memory
Technical Data 9
The capacity of the Base Station depends on the number of PDMs that have been On‐
Line and the number of dose rate samples. Storage capacity example:
• 290 hours of dose exposure for 50 PDMs each
9.4 Dose Data Transfer from PDM to Base Station
When a PDM gets within range of a Base Station, it will transfer data to the Base
Station (the accumulated dose values the PDM has collected since last time it was
within range. If a PDM is within range of a Base Station when it is exposed to radiation,
the PDM will also start sending dose rate samples to the Base Station each second.
9.5 PDM and Base Station/DoseView Memories
Dose rate samples that have been overwritten in the PDM’s dose rate memory may
still be available in the Base Station.
If there are no dose rate samples neither in the Base Station, nor in the PDM dose rate
memory, the Base Station and DoseView will instead display mean dose rate values
based on accumulated dose values (see Figure 22).
Lack of dose rate samples in the Base Station and/or DoseView memories occurs when
the:
• PDM is not within range of the Base Station when it is exposed to radiation.
• dose rate samples in the PDM dose rate memory are overwritten.
Philips Healthcare
9896 002 16621
Dose Rate (µSv/h
Accumulated dose (µSv)
06
07
08
09
10
11 (h)
Figure 22 DoseView chart. In lack of dose rate samples, the yellow curve displays mean dose rate values
per hour calculated from the accumulated dose (blue curve).
NOTE: The same effect as illustrated in Figure 22 is also obtained by un‐checking the
Show details check box in the Data filter panel (see chapter 4.5).
DoseAware Base Station Package Document version 1.0
35
9
Technical Data
9.6
Time Management
The PDM logs dose history in local time with no daylight saving time adjustments.
Daylight saving time adjustment is done in the Base Station, DoseView or Dose
Manager when the dose history is presented.
The following happens when the daylight saving time is changed:
• When going to summertime, one extra hour with no dose data will be added
to the dose log.
• When going to wintertime, two hours of dose data will be merged into one
hour. This hour contains no dose rate details. When changing from
summertime to wintertime, dose date details in the two merged hours will not
be displayed.
9.7 Technical Specifications
9.7.1 Radio Communication
• The communication range between a PDM facing a Base Station and the Base
Station is at least 10 meters in open air inside an operating room.
Communication radio, Europe, complies with 1995/5/EC Radio and
Telecommunications Terminal Equipment (R & TTE)
Communication radio, US, complies with FCC Declaration of Conformity
•
9.7.2
Base Station
Characteristic
Measure
Weight
1.45 kg (3,2 lb)
Dimensions
297x243x51 mm (WxHxD)
Display
10.4” touch screen, 640x480 pixels, 65 000 colors
Power supply
12 V, 2 A
Memory
512 Mb
Storage
Approximately 290 hours of dose history for each of 50
PDMs
Backlight life time
Approximately 50000 hours
Maximum heat dissipation
25 W
Network
Ethernet 10/100
USB
1.1 host for USB Mass Storage Device
9896 002 16621
•
Characteristic
Measure
Weight
30 g (1 oz)
Dimensions
44x45x10 mm (WxHxD)
Fastening
Metallic clip and lanyard holder
36
DoseAware Base Station Package
Document version 1.0
Philips Healthcare
9.7.3 PDM
Technical Data 9
Log memory 1
5 years Accumulated Dose with 1 hour resolution,
cyclically overridden
Log memory 2
3600 Dose Rate samples with one‐second resolution,
cyclically overridden. Dose Rate is only measured above
a 40µSv/h threshold limit
Time resolution
1 second
Time accuracy
Maximum error 2 seconds/24 hrs
Power supply
via the Cradle when connected to a computer via USB
Expected battery life
At least 4 years based on “normal use”, which is defined
as the operating conditions under normal use are 8
hours per day, 5 days a week and 52 weeks per year
at an ambient temperature of 20°C.
9.7.4 Cradle
Measure
Weight
50 g
Dimensions
64x61x59 mm (WxHxD)
Cable length
1,5 M
Power
via USB
Communication with
computer
USB 2.0
Philips Healthcare
9896 002 16621
Characteristic
DoseAware Base Station Package Document version 1.0
37
10
Safety
10 Safety
10.1 Warnings
If the DoseAware system is not functioning correct or damage is visible, inform a
Philips service engineer, which will take appropriate actions in order not to harm
personnel or patients.
Handle the hardware and software with care. Make sure that the hardware and
software is used and stored in a secured environment to prevent unauthorized access.
10.1.1 Maintenance and Faults
Do not use the product for any application until you are sure that the user routine‐
checks have been satisfactorily completed, and that the periodic maintenance of the
product is up to date. If any part of the product is known (or suspected) to be
defective or wrongly adjusted, DO NOT USE the product until a repair has been made.
Operation of the product with defective or wrongly adjusted components could
expose the user or the patient to safety hazards. This could lead to fatal or other
serious personal injury.
10.1.2 Safety Awareness
Do not use the product for any application until you have read, understood and know
all the safety information, safety procedures and emergency procedures contained in
this safety section. Operation of the product without a proper awareness of how to
use it safely could lead to fatal or other serious personal injury.
WARNING: During communication between the Base Station and PDMs, personal data
is transmitted in open air.
10.1.4 Safety Devices
Never attempt to remove, modify, or over‐ride or frustrate any safety device on the
product. Interfering with safety devices could lead to fatal or other serious personal
injury.
Changes or modifications on the equipment, not expressly approved by the party
responsible for compliance, could void the user’s authority to operate the equipment.
38
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
Do not use the product for any application until you have received adequate and
proper training in its safe and effective operation. If you are unsure of your ability to
operate this product safely and effectively DO NOT USE IT. Operation of this product
without proper and adequate training could lead to fatal or other serious personal
injury.
Do not operate the product with patients unless you have an adequate understanding
of its capabilities and functions. Using this product without such an understanding may
compromise its effectiveness and/or reduce the safety of the patient, you and others.
9896 002 16621
10.1.3 Adequate Training
Safety 10
10.1.5 Intended Use and Compatibility
Do not use the product for any purpose other than those for which it is intended. Do
not use the product with any product other than that which Philips Healthcare
recognizes as compatible. Operation of the product for unintended purposes, or with
incompatible product, could lead to fatal or other serious injury.
Do not move a PDM to an unknown environment (for example another hospital). If
you are visiting unknown environments, there is a risk that personal data is registered
there. For correct registration of staff dose data, only use the PDM within designated
environment.
10.2 Electrical Safety
WARNING: Do not remove covers or cables from this product. Dangerous electrical
voltages are present within this product. Removing covers or cables could lead to
serious or fatal personal injury.
Covers or cables should only be removed by qualified and authorized service
personnel. Use this product in rooms or areas that comply with all applicable law (or
regulations having the force of law) concerning electrical safety for this type of
product.
Electrically isolate this product from the mains electrical supply before cleaning or
disinfecting it.
10.3 Mechanical Safety
WARNING: Do not remove covers from this product. Removing covers could lead to
serious or fatal personal injury.
Covers should only be removed by qualified and authorized service personnel. In this
context, qualified means those legally permitted to work on this type of medical
electrical product in the jurisdiction(s) in which the product is being used, and
authorized means those authorized by the user of the product.
Philips Healthcare
9896 002 16621
10.4 Fire Safety
Use of electrical product in an environment for which it was not designed can lead to
fire or explosion.
Fire regulations for the type of medical area being used should be fully applied,
observed and enforced. Fire extinguishers should be available for both electrical and
non‐electrical fires.
WARNING: Only use extinguishers on electrical or chemical fires, which are specifically
labeled for those purposes. Using water or other liquids on an electrical fire can lead
to fatal or other serious personal injury.
If it is safe to do so, attempt to isolate the product from electrical and other supplies
before attempting to fight a fire. This will reduce the risk of electric shocks.
10.5 Electrostatic Discharge (ESD)
CAUTION: Always wait at least ten seconds after the product is switched OFF before
switching the product back to ON.
DoseAware Base Station Package Document version 1.0
39
10
Safety
CAUTION: Always use proper static procedures, protection, and product prior to
opening and during handling of this product. This product contains components that
are electrostatic sensitive. Failure to use ESD procedures may cause damage to these
components. Such damage to components is not covered by Philips warranties.
Connections to sensitive parts are identified by the ESD warning symbol as
shown.
Electrostatic discharge (ESD) can amount to a significant voltage, which may cause
damage to PCBs or other system components.
ESD damage is cumulative and may not be apparent at first, as indicated by a hard
failure, but can cause degraded performance. Therefore, always use proper ESD
handling procedures. ESD can result from low humidity conditions, use of electrical
equipment on carpeting, linens, and clothing.
10.6.1 Portable and Mobile Phones
CAUTION: Portable and mobile RF communications can affect electrical equipment.
Use caution when using such communication devices within the specified range of
electrical devices.
WARNING: Non‐radiation sources, for example a mobile phone, may influence the
measurement of data.
40
Document version 1.0
DoseAware Base Station Package
Philips Healthcare
This Philips product complies with relevant international and national law and
standards on EMC (electro‐magnetic compatibility) for this type of product when used
as intended. Such laws and standards define both the permissible electromagnetic
emission levels from product and its required immunity to electromagnetic
interference from external sources.
Other electronic products exceeding the limits defined in such EMC standards could,
under unusual circumstances, affect the operation of the product.
Medical electrical products needs special precautions regarding EMC, and needs to be
installed and put into service according to EMC information provided in the
accompanying documents.
The use of accessories and cables other than those specified, may result in increased
emission or decreased immunity levels.
The product should not be used adjacent to or stacked with other products and that if
adjacent or stacked use is necessary, it should be observed to verify normal operation.
The Base Station and PDM comply with part 15 of the FCC Rules. Operation is subject
to the following conditions:
(1) This device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may
cause undesired operation.
9896 002 16621
10.6 Electromagnetic Compatibility (EMC)
Safety 10
10.7 Modality Specific Safety
Philips Healthcare
9896 002 16621
Be careful when using a PDM while being near a patient and make sure that the PDM
does not fall or comes in contact with other equipment (such as a catheter) to
endanger the procedure.
DoseAware Base Station Package Document version 1.0
41

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
Linearized                      : No
Page Count                      : 41
PDF Version                     : 1.4
Title                           : Microsoft Word - DoseAware_Base_Station_Package_User_Manual-100111.docx
Author                          : Hilda Harrysson
Creator                         : Microsoft Word
Producer                        : Mac OS X 10.5.6 Quartz PDFContext
Create Date                     : 2010:01:12 11:01:20Z
Modify Date                     : 2010:01:12 11:01:20Z
EXIF Metadata provided by EXIF.tools
FCC ID Filing: XWK8603021

Navigation menu