Vios Medical CS2050 Vios Monitoring System (VMS) User Manual

Vios Medical Inc Vios Monitoring System (VMS)

User manual

Download: Vios Medical CS2050 Vios Monitoring System (VMS) User Manual
Mirror Download [FCC.gov]Vios Medical CS2050 Vios Monitoring System (VMS) User Manual
Document ID3583932
Application IDNy+csclxag0ArfhoLDWwEA==
Document DescriptionUser manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize111.1kB (1388724 bits)
Date Submitted2017-09-28 00:00:00
Date Available2018-01-03 00:00:00
Creation Date2017-09-22 09:23:31
Producing SoftwareMicrosoft® Word 2010
Document Lastmod2017-09-27 19:56:28
Document TitleUser manual
Document CreatorMicrosoft® Word 2010
Document Author: Ben Meyer

Chest Sensor
Instructions For Use
Model CS2050
FL00002046
Revision A
Vios Medical, Inc.
Vios Monitoring System
1 Introduction
Revision History
The cover page of this document has the revision letter at the bottom of the page. The revision letter identifies the
document’s level of revision. The revision history of this document is summarized below.
Revision
Description
Initial release.
Package Contents


One (1) Chest Sensor
One (1) Chest Sensor Instructions For Use
Vios Monitoring System Description
The Vios Monitoring System includes:


Chest Sensor: A plastic encased, Bluetooth-enabled reusable sensor. The Chest Sensor is used in
conjunction with the Adapter and BSM to monitor heart rate, respiratory rate, pulse rate, SpO 2, and patient
posture.
Adapter: A plastic encased, reusable adapter with four (4) patient ECG cables and a one (1) pulse oximetry
cable. The Adapter is used in conjunction with the Chest Sensor and BSM to monitor heart rate, respiratory
rate, pulse rate, SpO2, and patient posture. The following Adapters may be used with the Chest Sensor:
Model: L2050F Adapter (Finger Pulse Oximetry)
Model: L2050E Adapter (Ear Pulse Oximetry)
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 2 of 28
Pulse Oximetry:
Pulse Oximetry:
 Measurement Site: Finger
 Measurement Site: Ear
 Pulse Oximetry Cable Max Length: 36.75 inches
 Pulse Oximetry Cable Max Length: 22.5 inches
ECG:
ECG:
 7-Lead
 7-Lead
 RA Cables Max Length: 9.05 inches
 RA Cables Max Length: 9.05 inches
 LL Cable Max Length: 11.78 inches
 LL Cable Max Length: 11.78 inches
 V Cable Max Length: 14.78 inches
 V Cable Max Length: 14.78 inches
 Bedside Monitor (BSM) Software: Analyzes and displays vitals data from one or more compatible sensors.
May operate in stand-alone mode or communicate analyzed data across standard networking protocols.
 Central Station Monitor (CSM) Software: Simultaneously displays analyzed data from the Bedside Monitor
software for multiple patients.
 Central Station (CS) Software: A communication hub that transfers data between the BSM software and
CSM software.
Intended Use
The Vios Monitoring System (VMS) is intended for use by medically qualified personnel for physiological vital signs
monitoring of adult (18+) patients in healthcare facilities. It is indicated for use in monitoring of 7-Lead ECG, heart
rate, respiratory rate, pulse rate, functional oxygen saturation of arterial hemoglobin, non-invasive blood pressure,
and patient posture and activity. VMS allows for the input of body temperature, and can display data from peripheral
devices. VMS can generate alerts when rate-based cardiac arrhythmias are detected and when physiological vital
signs fall outside of selected parameters.
Intended Audience
This manual is intended for clinical professionals. Clinical professionals are expected to have a working knowledge of
medical procedures, practices, and terminology, as required for monitoring patients.
Equipment Symbols
Warning / Caution
Do not use if package is damaged
RoHS Compliant
Date of Manufacture
Manufacturer
Type CF applied part symbol for shock protection
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 3 of 28
Non-ionizing radiation
Not for general waste
IP22
Ingress Protection. First 2 – Protection against fingers
or similar objects > 12.5 mm. Second 2 – Protection
against dripping water when tilted up to 15º
Serial Number
Bluetooth Low Energy MAC Address
Consult Instructions for Use
Not for use with MRI
Rx Only
For prescription use only
Refer to Instruction Manual
Electrostatic Sensitive Device
Contraindications

This device is contraindicated for use in the presence of Magnetic Resonance Imaging (MRI) devices.
Warnings





The Vios Chest Sensor and Adapter may cause interference with pacemakers/ICDs that utilize minuteventilation respiration detection
The Vios Chest Sensor and Adapter to not detect pacing spikes generated from implanted pacemakers
Remove the Vios Chest Sensor and Adapter from any patient prior to magnetic resonance imaging (MRI)
Do not place the Vios Chest Sensor and Adapter on patients under the age of 18
Federal Law (US) restricts this device to sale by or on the order of physician
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 4 of 28










Do not apply the Vios Chest Sensor, Adapter, or electrodes on breached or compromised skin surfaces or on
mucosal membranes
Vios Medical recommends using Medico Electrodes MSGST-06 to adhere the Adapter and Chest Sensor to
the patient
No modification of this equipment internal or external is allowed
Do not destroy this equipment without authorization of the manufacturer.
Hazards can result from unauthorized modification of the equipment.
If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued
safe use of the equipment.
Pins of connectors identified with the ESD warning symbol should not be should and that connections
should not be made to these connectors unless ESD precautionary procedures are used.
This equipment is intended for use by healthcare professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation
measures, such as re-orientating or relocating the ME equipment or shielding the location.
Use of adaptor, cables other than those specified, may result in increased emissions or decreased immunity.
The CS2050 should not be worn and used adjacent to another wireless device such as mobile phone.
Cautions






The Vios Chest Sensor is a Bluetooth® low energy (BLE) device and should not be used in BLE sensitive areas
The Vios Chest Sensor is a radio frequency (RF) emission device. Remove all components of the Vios Bedside
Monitoring System in radio frequency sensitive areas.
The VMS should not be used near direct X-ray exposure. The Vios Chest Sensor and Adapter must be
removed from a patient prior to an upper torso X-ray
The VMS should not be used near X-ray computed tomography (CT) equipment. The Vios Chest Sensor and
Adapter must be removed from patient prior to a CT scan
Never autoclave, sterilize, or immerse the Chest Sensor in liquid of any kind
The following may degrade pulse oximeter performance:
o Excessive ambient light
o Excessive motion
o Improperly applied sensor
o Residue (e.g. dirt, dried blood) in the optical path
o Blood pressure cuffs
o Fingernail polish
Notes: The following notice safety messages apply to this monitoring system.





If VMS components have been transported or stored outside operating temperature, allow them to stabilize
back to the operating temperature range before use.
Portable and mobile RF communications equipment can affect the Chest Sensor and its communication with
the VMS.
Service and repairs are allowed for authorized Vios personnel only.
The Chest Sensor is Internally Powered ME Equipment per IEC 60601-1
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 5 of 28
FCC Compliance
FCC compliance statements,
This Device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, this including interference that may cause undesired
operation.
FCC Caution!!!
• Any changes or modifications not expressly approved by the party
Responsible for compliance could void the user's authority to operate this
Equipment
Part 15B compliance statements for digital devices:
NOTE: This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy
And, if not installed and used in accordance with the instructions, may
Cause harmful interference to radio communications. However, there is no
Guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television
Reception, which can be determined by turning the equipment off and on,
The user is encouraged to try to correct the interference by one or more of
The following measures:
—Reorient or relocate the receiving antenna.
—Increase the separation between the equipment and receiver.
—Connect the equipment into an outlet on a circuit different from that to
Which the receiver is connected
—Consult the dealer or an experienced radio/TV technician for help
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 6 of 28
2 Patient Application
Caution: Examine the product package upon receipt and do not use components if there appears to be any
damage. Contact Vios for a replacement product. See Vios Customer Service section.
Notes: For instructions on how to use the Chest Sensor with the Vios Bedside Monitor Software (Model BSM2050),
please refer to the Vios Bedside Monitor (Model BSM2050) Instructions For Use.
Overview
Component
LED Indicator
Patient Call Button
Chest Sensor MAC ID (Last 4 Digits)
30-Pin Connector
USB Connector
Back Label
Description
Indicates various states of the Chest Sensor through red, yellow, green, and
blue light with varying rates of blinking
User actuation of this button drives various functionalities of the device
(turning the device on/off, sending a distress signal to the BSM)
Provides the last 4 digits of the Chest Sensor MAC ID for pairing to the BSM
Connector to which the Adapter is attached
Connector for charging the Chest Sensor battery
Provides information pertaining to the MAC ID, serial number,
manufacturing
date,
manufacturing
lot,
model
number,
and
warning/cautionary symbols
Patient Preparation
Required Equipment:
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 7 of 28




Vios Chest Sensor, Model CS2050
Vios L2050F Adapter or L2050E Adapter
Recommended electrodes
Disinfecting wipes such as isopropyl alcohol or cleaning solution (not included)
Warning: Perform visual inspection of the Chest Sensor, Adapter, and electrodes for any signs of damage
prior to use. The components should not have any damage. If any damage is observed, then use a new set of
components and contact Vios to replace damaged parts.
Observe the chest and limb lead placement sites. Trim or clip hair if necessary. See Figure 1.
Figure 1
Use disinfecting wipes to clean the Chest Sensor placement sites and limb lead placement sites.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 8 of 28
Figure 2
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 9 of 28
Check Chest Sensor Battery
Always check the Chest Sensor battery status prior to monitoring a patient. You may check the Chest Sensor battery
status by pressing the Patient Call Button when the device is off. Do not press and hold.
Briefly after pressing the Patient Call Button, the Chest Sensor LED will briefly display a color for 4 seconds. The
displayed color will indicate the battery status – see the table below for the color representations:
Use the battery status indication to guide your decision about whether that Chest Sensor has enough charge to be
used for a certain period of time. If the Chest Sensor indicates low battery (red or blinking red), please consider
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 10 of 28
charging the Chest Sensor or using a different device. See the ‘Charging the Chest Sensor’ section for instructions on
charging.
Device Cleaning
Use disinfecting wipes to clean the Chest Sensor and Adapter prior to patient application. The following may be used
for cleaning the Chest Sensor and Adapter:




Isopropyl Alcohol (70% Alcohol/water mixture)
Hydrogen Peroxide
Ethyl Alcohol (50% Alcohol/water mixture)
Water
Figure 3
Attach an Adapter
Depending on where you want to monitor pulse oximetry, you may choose either a L2050F Adapter or L2050E
Adapter. Slide the Chest Sensor into the Adapter so that the 30-Pin connectors mate. Ensure that the Chest Sensor is
properly aligned on the Adapter rails.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 11 of 28
Figure 4
Caution: Ensure when sliding the Chest Sensor into the Adapter that the 30-Pin connectors are fully mated.
Snap two (2) electrodes into the back of the Adapter. Ensure there is no overlap between the two electrodes.
Caution: Ensure that when snapping the two (2) electrodes into the Adapter that there is no overlap between
the electrodes.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 12 of 28
Snap four (4) electrodes into the RA, LL, and V lead snaps.
Device Placement
Peel off the adhesive liner from all six (6) electrodes.
Attach the Chest Sensor with Adapter to the upper left pectoral area of the patient’s chest. Ensure that the device is
positioned correctly per the image below.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 13 of 28
Affix the two Right Arm (RA) electrodes to the upper right pectoral area of the patient’s chest. Ensure the leads are
positioned correctly per the image below.
Affix the Left Leg (LL) electrode to the patient’s left abdominal area, below the Chest Sensor. Ensure the lead is
positioned correctly per the image below.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 14 of 28
Lastly, affix the precordial (V) electrode to the patient. Placement of the electrode will determine which precordial lead
is being monitored. The VMS may monitor one of the six (V1, V2, V3, V4, V5, V6) leads at a time – please see the
image below for appropriate placement.
Pulse Oximeter Placement
Next, place the pulse oximeter at the appropriate site. The site will depend on which adapter was connected to the
Chest Sensor – please refer to the corresponding section below:
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 15 of 28
L2050F Adapter
Place the pulse oximeter finger clip over a finger on the patient’s left hand (index finger preferred). Ensure that the
patient’s finger reaches the back of the pulse oximeter enclosure.
It is recommended to tape down the cable slack to prevent accidentally pulling or yanking of the cable. See the
picture below for recommended tape sites.
L2050E Adapter
Place the pulse oximeter ear clip over the patient’s earlobe. Ensure that the ear clip fully covers the earlobe and is
secure.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 16 of 28
It is recommended to tape down the cable slack to prevent accidentally pulling or yanking of the cable. See the
picture below for recommended tape sites.
3 Charging the Chest Sensor
Overview
The Chest Sensor uses a micro-USB connector for connecting to a charger. The charger must be connected to mains
power to charge the Chest Sensor internal battery.
Vios recommends using the following power supplies for charging the Chest Sensor:


TRUMPower Medical Grade Power Supply, Model TRM15-S05-E-UB-10F
ASUS Switching Power Supply, Model PSM06A-050Q
Note: Please do not charge the Chest Sensor at temperatures less than or equal to 32°F.
LED Indications
When properly connected to a powered charger, the Chest Sensor LED will indicate a color based on the remaining
battery life. See the images below for descriptions of each color representation.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 17 of 28
Use the battery status indication to guide your decision about whether that particular Chest Sensor should be used or
not. If the Chest Sensor indicates low battery (red), please consider waiting for the Chest Sensor to fully charge or use
a different, fully charged device. See the ‘Charging the Chest Sensor’ section for instructions on charging.
Pairing to a Bedside Monitor
Overview
Before pairing the Chest Sensor with a Bedside Monitor (BSM), you must initiate Bluetooth Low Energy (BLE)
advertising on the Chest Sensor.
Initiating Bluetooth Low Energy (BLE) Advertising
To initiate Bluetooth Low Energy (BLE) Advertising, press and hold the Patient Call Button for at least seconds (4)
seconds. For the first four (4) seconds, you will see the battery status indicated as described in the ‘Battery Check’
section. If you continue to hold the Patient Call Button for more than four (4) seconds, the Chest Sensor LED will begin
blinking blue at a rate of once per second.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 18 of 28
Pairing to a Bedside Monitor (BSM)
Once you have initiated Bluetooth Low Energy advertising, pair the Chest Sensor to the Bedside Monitor (BSM) using
the Bedside Monitor software. For instructions on how to do this, please refer to the Bedside Monitor (BSM2050)
Instructions for Use.
Confirmation of Pairing
After pairing the Chest Sensor to the Bedside Monitor (BSM), the Chest Sensor LED will slowly blink green (once per
10 seconds) to indicate that it is paired.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 19 of 28
4 Troubleshooting
Overview
The troubleshooting section consists of several parts which should help you resolve the most common problems with
the Chest Sensor. The ‘Common Issues’ section lists out the most likely issues you may encounter and the steps that
should be taken to resolve them.
Common Issues
The Chest Sensor won’t fit into the Adapter.
Ensure that the 30-Pin male connector on the Chest Sensor is properly aligned with the 30-Pin female connector on
the Adapter. Ensure that the slots on the Chest Sensor are properly aligned with the rails on the Adapter.
The Chest Sensor / Adapter does not adhere to the patient.
Ensure that the steps in the ‘Patient Preparation’ section have been followed. Ensure that the Chest Sensor / Adapter
have been placed per the ‘Device Placement’ section.
The Chest Sensor will not turn on.
Ensure that the steps in the ‘Check Chest Sensor Battery’ section have been followed. If the device is still not turning
on, the battery may be depleted – try plugging it into a Vios recommended charger.
Service & Maintenance
If the device is damaged, please contact Vios for a replacement. Do not attempt to repair or service the device.
See Customer Service information below.
If you are suspicious of damage to the device, please inspect the following:



Ensure there are no breaks or tears in any of the cables
Ensure the snaps (both on the leads and on the back of the adapter) are not loose
Ensure the 30-Pin connector is not loose
To clean the Chest Sensor:



Disconnect the Adapter from the Chest Sensor
Dampen a clean cloth or gauze pad with appropriate cleaning solution and wipe all exposed surfaces
including patient ECG cables and pulse oximetry cable/enclosure
Dry all exposed surfaces with a clean, dry cloth or gauze pad
For proper disposal, please return the device to Vios at the end of its service life. See Customer Service information
below.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 20 of 28
5 Technical Specifications
Materials and Dimensions
Materials
Dimensions
High impact PC, ABS, or a ABS/PC blend
74.6mm X 62.0mm X 14.5mm
Performance Specifications
Minimum amplitude of QRS
complex in the patient
physiological signal
Duration of use
Use life
Service Life
Degree of protection
Enclosure degree of Ingress
protection
0.15 mV
At least 28 hours on a new, fully-charged battery
 The Chest Sensor shall have up to 365
attach/detach events with the Adapter.
 The Chest Sensor shall have up to 365
attach/detach events with the Vios
recommended charger.
Expected to remain suitable for its intended use for up
to 12 months
Type CF applied part symbol for shock protection
IP22
Caution: To prevent detection of P-waves or baseline noises as QRS complexes, the minimum detection level
for QRS complexes as mentioned in the table above will be required for proper operation.
Operating Environment
Operating Temperature
Storage and Transport
Temperature
Operating Humidity
Storage and Transport
Humidity
Atmospheric Pressure
Wireless Frequency and
maximum radiated power
Battery Specification
Type
Chemistry
Capacity
Voltage
10°C to 40°C
-40°C to +70°C
Relative humidity range from 10% to 95% (Noncondensing)
Relative humidity range from 10% to 100%
80 kPa to 100 kPa
f = 2402 + k*2 MHz, k = 0, … , 39
power: 0.01 mW (-20 dBm) … 10 mW (+10 dBm)
Rechargeable
Lithium-ion Polymer
1250 mAh (Typical), 1150 mAh (Minimum)
3.7 V (Nominal)
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 21 of 28
Dimension
Safety Certification
Battery Life
52.0mm X 34.5mm x 7.3mm
IEC 62133:2012
28 hours of continuous use
6 Customer Service & Contact
Information
Vios Medical value’s our Customer’s feedback regarding our products, please call Vios Medical Inc. for any Customer
Feedback, Complaints, or Return/Replacement Inquiries. Customer Service is available 24 hours a day, 7 days a week.
Vios Medical USA: +1-651-764-8467
Vios Medical India: +91-(0)80-4115-0546
Vios Medical Marketing: +91-8427439000
support@viosmedical.com
7 Safety & Compliance
Safety Notes
The Vios Chest Sensor:

Uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.

Is suitable for use in all establishments other than domestic and those directly connected to the public lowvoltage power supply network that supplies buildings used for domestic purposes.

May need special precautions regarding EMC and needs to be installed and put into service according to the
EMC information.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 22 of 28

ECG signals must have a minimum QRS amplitude of 5uV. ECG signal amplitude lower than this will be
inhibited by the system

May be affected by portable and mobile RF communications equipment.

Should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is
necessary, the Vios Chest Sensor should be observed to verify normal operation in the configuration in which
it will be used.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 23 of 28
Electromagnetic Emissions
Table A: Guidance and manufacturer’s declaration – electromagnetic emissions – for all medical electrical equipment
and medical electrical systems
Guidance and manufacturer’s declaration – electromagnetic emissions
The CS2050 is intended for use in the electromagnetic environment specified below. The customer or the user of the
CS2050 should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
EN 55011
RF emissions
Class B
EN 55011
Harmonic emissions
EN 61000-3-2
Voltage fluctuations /
flicker emissions
Class A
Electromagnetic environment - guidance
The CS2050 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
The CS2050 is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low voltage power supply network that supplies buildings
used for domestic purposes.
Complies
EN 61000-3-3
Table B: Guidance and manufacturer’s declaration – electromagnetic immunity – for all medical electrical equipment
and medical electrical systems
Electromagnetic Immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The CS2050 is intended for use in the electromagnetic environment specified below. The customer or the user of
the CS2050 should assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
Error! Reference source
not found. 60601 test
level
Compliance level
± 6kV contact
± 6kV contact
± 8kV air
± 8kV air
Electrical fast
transient/burst
± 2 kV for power
supply lines
± 2 kV for power
supply lines
EN 61000-4-4
± 1 kV for input/output
lines
EN 61000-4-2
Electromagnetic
environment – guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 24 of 28
Surge
± 1 kV line(s) to line(s)
± 1 kV line(s) to line(s)
Mains power quality should be that
of a typical commercial or hospital
environment.
< 5 % UT
(> 95 % dip in UT)
for 0,5 cycle
< 5 % UT
(> 95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of CS2050
requires
continued
operation
during power mains interruptions, it
is recommended that the CS2050
be powered from an uninterruptible
power supply or a battery.
< 5 % UT
(> 95 % dip in UT)
for 5 s
< 5 % UT
(> 95 % dip in UT)
for 5 s
3 A/m
3 A/m
EN 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
EN 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
EN 61000-4-8
NOTE
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial
or
hospital
environment.
UT is the a.c. mains voltage prior to application of the test level.
Table C: Guidance and manufacturer’s declaration – electromagnetic immunity – for all medical electrical equipment
and medical electrical systems that are not life-supporting
Guidance and manufacturer’s declaration – electromagnetic immunity
The CS2050 is intended for use in the electromagnetic environment specified below. The customer or the user of the
CS2050 should assure that it is used in such an environment.
Immunity test
Error! Reference
source not found.
60601 test level
Compliance level
Electromagnetic
environment – guidance
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 25 of 28
Portable and mobile RF communications
equipment should be used no closer to any
part of the CS2050, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
Conducted RF
3 Vrms
EN 61000-4-6
150 kHz to 80 MHz
3 Vrms
d  [3,5 / V 1] P
d  [3,5 / E1] P
d  [7 / E1] P
Radiated RF
3 V/m
EN 61000-4-3
80 MHz to 2,5 GHz
3 V/m
80 MHz to 800 MHz
800 MHz to 2,5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the CS2050 is used exceeds the
applicable RF compliance level above, the CS2050 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the CS2050.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table D: Recommended separation distances between portable and mobile RF communications equipment and the
medical electrical equipment and medical electrical systems – for medical electrical equipment and medical electrical
systems that are not life-supporting
Recommended separation distances between
portable and mobile RF communications equipment and the CS2050
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 26 of 28
The CS2050 is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the CS2050 can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the CS2050
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d  [3,5 / V 1] P
d  [3,5 / E1] P
d  [7 / E1] P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
The Vios Monitoring System (VMS) has been designed in compliance with applicable Safety Standards given below.
Safety Standards










60601-1: Safety Requirements for Medical Electrical Systems
60601-1-2: General Requirements for Safety – Electromagnetic Compatibility
60601-2-27: Particular requirements for the basic safety and essential performance of
electrocardiographic monitoring equipment
60601-2-49: Particular requirements for the safety of Multifunction Patient
Monitoring Equipment
ISO 80601-2-61: Particular requirements for basic safety and essential performance of
pulse oximeter equipment
ISO 10993-1: Biological Evaluation of medical devices
EN 300 328 v1.8.1, Electromagnetic compatibility and Radio spectrum Matters (ERM);
Wideband transmission systems; Data transmission equipment operating in the 2,4
GHz ISM band and using wide band modulation techniques
ASTM D4169-14: Standard Practice for Performance Testing of Shipping Containers
and Systems
IEC 60529 standard Ingress Protection Marking – IP22
EC53 ECG Trunk Cables and Patient Leadwires
Vios™ is a trademark of Vios Medical Incorporated.
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 27 of 28
®
Bluetooth is a registered trademark of Bluetooth SIG, Incorporated.
Medico™ is a trademark of Medico Electrodes International Ltd.
Manufactured by Vios Medical, Inc.
7300 Hudson BLVD N 140
St Paul MN 55128
+1 651-764-8467
_____________________________________________________________________________________________________________________________
Chest Sensor IFU
Page 28 of 28

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
Author                          : Ben Meyer
Create Date                     : 2017:09:22 09:23:31+05:30
Modify Date                     : 2017:09:27 19:56:28+08:00
Language                        : en-US
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Format                          : application/pdf
Creator                         : Ben Meyer
Creator Tool                    : Microsoft® Word 2010
Metadata Date                   : 2017:09:27 19:56:28+08:00
Producer                        : Microsoft® Word 2010
Document ID                     : uuid:52fa61b7-8b80-4a3a-a075-331a7d188ba7
Instance ID                     : uuid:7dc66334-3bdb-4d50-bc83-fdc5364d757a
Page Count                      : 28
EXIF Metadata provided by EXIF.tools
FCC ID Filing: 2ANL9CS2050

Navigation menu